Hanna Instruments HI93501, HI935007, HI935004, HI935001, HI935008 Instruction Manual

Page 1
Instruction Manual
HI93501 HI935001 HI935007 HI935004
HI935008
Portable Microprocessor
Thermometers
www.hannainst.com
1
Page 2
Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using these instruments.
This manual will provide you with the necessary information for correct use of these instruments, as well as a precise idea of their versatility.
If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com or view our worldwide contact list at www.hannainst.com.
WARRANTY
HI93501, HI935001, HI935004, HI935007 and HI935008 are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Electrodes and probes are guaranteed for six months. This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packed for complete protection.
TABLE OF CONTENTS
WARRANTY ..................................................................................2
GENERAL DESCRIPTION ................................................................3
SPECIFICATIONS OF HI93501 ...................................................... 4
SPECIFICATIONS OF HI935001 & HI935007 ................................ 5
SPECIFICATIONS OF HI935004 & HI935008 ................................ 6
OPERATIONAL GUIDE ................................................................... 7
METER SETUP ........................................................................... 11
BATTERY REPLACEMENT ............................................................. 12
TEMPERATURE PROBES .............................................................. 13
2
Page 3
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your dealer or the nearest Hanna Customer Service Center.
Each meter is supplied complete with:
• Temperature probe
• Alkaline batteries: 3 x 1.5V AAA
• Instruction manual. Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the original packing with the supplied accessories.
GENERAL DESCRIPTION
Temperature plays a very important role in the food industry, and it also has to be checked in the warehouses, loading docks and in transportation vehicles.
You can choose the Hanna food-thermometer which best meets your needs, among several models:
HI93501, HI935001, HI935004, HI935007 and HI935008 are pro­fessional, rugged and waterproof thermometers, designed to be used on a daily basis in industrial kitchens and catering. They provide excellent accuracy and can operate with a wide range of interchangeable probes without requiring recalibration.
Standard features include Cal-Check feature for checking the meter calibration status any time, low battery detection, auto-off capability and long battery life.
HI93501 is a NTC10K-thermistor thermometer. It offers a wide temperature range with a resolution of 0.1 °C (0.1 °F).
HI935001 and HI935007 is a K-thermocouple thermometer. It offers a wide temperature range, automatic change of resolution from 0.1 °C (0.1 °F) to 1 °C. (1 °F).
HI935004 and HI935008 is a T-thermocouple thermometer. It offers a wide temperature range, automatic change of resolution from 0.1 °C (0.1 °F) to 1 °C. (1 °F).
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
3
Page 4
SPECIFICATIONS OF HI93501
Main features:
FC762PW NTC1OK-thermistor probes series
• Range -50.0 to 150.0 °C (-58.0 to 302.0 °F)
• Calibration Check feature
• Remaining battery life indication / low battery detection
• Auto-off capability
• Waterproof casing
Range -50.0 to 150.0 °C
-58.0 to 302.0 °F
Resolution 0.1 °C / 0.1 °F Accuracy ±0.4 °C / ±0.7 °F excluding probe error Battery 3 x 1.5V AAA, approx. 5000 hours of continuous use Auto-off user selectable: after 8 min, 60 min, disabled Environment -20 to 50 °C / -4 to 122 °F ; RH 100% Dimensions 140 x 57 x 28 mm (5.5 x 2.2 x 1.1") Weight 175 g (6.17 oz.)
4
Page 5
SPECIFICATIONS OF HI935001 AND
HI935007
Main features:
FC766PW K-thermocouple probe for HI935001
• Fixed K-thermocouple probe for HI935007
• Range -50.0 to 1350.0 °C (-58.0 to 2462.0 °F)
• Remaining battery life indication / low battery detection
• Auto-off capability
• Waterproof casing
Range -50.0 to 199.9 °C / -58.0 to 399.9°F
200 to 1350 °C / 400 to 2462 °F
Resolution 0.1 °C (up to 199.9 °C) / 1°C (outside)
0.1 °F (up to 399.9 °F) / 1 °F (outside)
Accuracy ±0.2% full scale
excluding probe error
Battery 3 x 1.5V AAA, approx. 4000 hours of continuous use Auto-off user selectable: 8 min, 60 min, disabled Environment -20 to 50 °C / -4 to 122 °F; RH 100% Dimensions 140 x 57 x 28 mm (5.5 x 2.2 x 1.1") Weight 178 g (6.27 oz.)
5
Page 6
SPECIFICATIONS OF HI935004 AND
HI935008
Main features:
FC767PW T-thermocouple probe for HI935004
• Fixed T-thermocouple probe for HI935008
• Range -50.0 to 400.0 °C (-58.0 to 752.0 °F)
• Remaining battery life indication / low battery detection
• Auto-off capability
• Waterproof casing
Range -50.0 to 400.0 °C
-58.0 to 752.0 °F
Resolution 0.1 °C / 0.1 °F Accuracy ±0.2% full scale
excluding probe error
Battery 3 x 1.5V AAA, approx. 4000 hours of continuous use Auto-off user selectable: 8 min, 60 min, disabled Environment -20 to 50 °C /-4 to 122 °F; RH 100% Dimensions 152 x 58 x 30 mm (5.5 x 2.2 x 1.1") Weight 178 g (6.27 oz.)
6
Page 7
OPERATIONAL GUIDE
Each meter is supplied with batteries. Remove the battery cap, unwrap the supplied batteries and install them while paying attention to their polarity (see also the "Battery replacement" section for more details).
TAKING MEASUREMENTS
Connect a suitable temperature probe to the meter (see pages 12 for a complete list of Hanna probes).
Note: HI935007 and HI935008 has a fixed probe. To switch the meter ON, press the ON/
OFF button on the front of the meter. If it does not turn on, make sure that the batteries are properly installed in their place.
At start-up the thermometer displays all LCD segments for a few seconds, followed by the percentage indication of the remaining battery life; then it enters Cal Check mode and then enters measure­ment mode.
HI93501
HI935001 and HI935007
HI935004 and HI935008
7
Page 8
Note: The full display can be frozen by keeping the ON/OFF
button pressed.
If a temperature probe is plugged in, the meter displays the measured temperature.
If no temperature probe is plugged in or the reading is over-range, the display shows blinking dashes and NO PROBE” message is
displayed (see WARNING MESSAGES section). If the reading is
slightly beyond the measuring range, the meter flashes the closest full-scale value.
To switch the meter OFF, press the ON/OFF key.
AUTO-OFF FEATURE
To save battery life, each model is provided with an auto-off feature, which switches the instrument off after a certain period of non use. The auto-off delay time can be set in 3 modes from SETUP MENU:
8 minutes
60 minutes
Disabled
8
Page 9
CALIBRATION CHECK
All the models are provided with automatic Cal-Check feature at start­up, which allows the user to check the instrument calibration status.
If the meter is calibrated, the display will stabilize and show “0.0 °C” (±0.3 °C) in a few seconds.
WARNING MESSAGES
NO PROBE message
If the probe is not connected or is interrupted and “NO PROBE” message is displayed.
HI93501
HI935001 and HI935007
HI935004 and HI935008
9
Page 10
CALIBRATION CHECK ERROR MESSAGE
If during CALIBRATION CHECK the meter display more than
±0.3 °C, then a “Err” message is dispayed.
Contact your dealer or your nearest Hanna Service Center for recalibration.
DAMAGED PROBE message
When the readings are very far from the reading range, the probe can be damaged, “NO PROBE” message is displayed (see WARNING MESSAGES section).
10
Page 11
METER SETUP
Enter Setup to change the tempera­ture unit (°C or °F) or to adjust the
auto shut off time. To enter the Setup MODE, press the SET button located inside the battery compartment.
• Use the ON/OFF button to switch between “°C” or “°F”. Long press ON/ OFF button once to save and continue to auto off settings.
• Use the ON/OFF button to cycle through the following Auto Shut Off options: 8 minutes (“d08”, default value), 60 minutes (“d60”), OFF (“d---”). Long press the ON/OFF button once to save and continue to normal operation.
11
Page 12
BATTERY REPLACEMENT
When the remaining battery life is less than 10% the battery tag blinks on the display to warn the user.
If the battery level is 0% enough to cause erroneous readings, the Battery Error Prevention System (BEPS) turns the meter off.
Immediately replace the batteries with new ones. Battery replacement must only take place in a non-hazardous
area. The batteries are accessed by opening the battery compartment
cap on the back of the instrument. Replace the three 1.5V AAA alkaline batteries located in the
battery compartment, observing the polarity on the rear of the instrument.
Reattach the battery compartment cap making sure that the gasket is in place.
12
Page 13
TEMPERATURE PROBES
Hanna offers a wide range of probes to meet your needs with air, liquid, penetration and surface measurements. All Hanna temperature probes are pre-calibrated at the factory and ready-to-use. These interchangeable probes make it possible to switch from one probe to another without requiring recalibration. Hanna probes are also available in different handle colors to avoid cross contamination during testing.
Note: To clean the meters, do not use aggressive detergents. It is
recommended to use water.
FOR HI93501
FC762PW (*) General purpose penetration probe w/1 m (3.3')
white cable & white handle
FOR HI935001
FC766CPW (*) Penetration, K-type thermocouple probe w/1 m
(3.3') white cable & white handle
FOR HI935004
FC767CPW (*) Penetration, T-type thermocouple probe w/1 m
(3.3') white cable & white handle
(*) To be replaced only by authorized service technicians.
13
Page 14
RECOMMENDATIONS FOR USERS
Before using these products, make sure they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential areas could cause unacceptable interferences to radio and TV equipment, requiring the operator to follow all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instruments’ EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24 VAC or 60 VDC.
To avoid damage or burns, do not perform any measurement in microwave ovens.
Hanna Instruments reserves the right to modify the design,
construction or appearance of its products without advance notice.
14
Page 15
15
Page 16
Hanna Instruments Inc.
Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA
Technical Support for Customers
Tel. (800) 426 6287 Fax (401) 765 7575 E-mail tech@hannainst.com www.hannainst.com
Local Sales and Customer Service Office
Printed in ROMANIA
MAN93501Y 12/16
16