Dear Customer,
Thank you for chosing a Hanna Instruments
Product.
Please read this instruction manual carefully
before using the instrument.
This manual will provide you with all the necessary information for the correct use of the
instrument, as well as a precise idea of its
versatility in a wide range of applications.
These instruments are in compliance with
directives EN 50081-1 and 50082-1.
CE Declaration of Conformity .................. 35
ISO 9000 Certified
Company since 1992
2
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure
that no damage has occurred during shipping.
If there is any noticeable damage, notify your
Dealer.
Note: Save all packing materials until you
are sure that the instrument functions
correctly. All defective items must be
returned in the original packing materials
together with the supplied accessories.
GENERAL DESCRIPTION
HI 931700 and HI 932700 respectively are pH
and ORP panel-mounted indicators and controllers designed for simplicity of use in a wide
range of industrial applications.
The models are panel mount designed with
membrane keypads on the front panel and a
12x30 mm LCD display. Connections to the
electrodes, power supply, contacts and recorders are made via the plug-in terminal
blocks on the rear panel.
The instruments are equipped with a BNC
socket and accept input direct from pH or ORP
electrode.
Other features include: recorder output 0 to
20mA or 4 to 20 mA as indicated; LED indicators which identify whether the controller is in
operation mode or selection mode.
3
FUNCTIONAL DESCRIPTION HI 931700
FRONT PANEL
HI 931700
pH CONTROLLER
pH
SET
ON
OFFSET
SLOPE
REAR PANEL HI 931700
SETMEA
SET
Keypad
SETTo display and to set the working
point of pH dosage
MEASURE To set HI 931700 to the measure-
ment mode
Trimmers
OFFSETFor Offset Calibration
SLOPEFor Slope Calibration
SETTo adjust the Set Point
3. Connections for dosing pump to dose either
acid or base products. These contacts act
only as a switch for the power to the drive;
4. Recorder output(M1 ! + mA output,
M2 ! - mA output);
5. BNC for pH electrode;
6. Connection for Potential Matching Pin;
7. Connection for Electrode Reference.
Note: Be sure your main line is protected by
a fuse.
5
FUNCTIONAL DESCRIPTION HI 932700
FRONT PANEL
HI 932700
ORP CONTROLLER
mV
SET
ON
SLOPE
REAR PANEL HI 932700
SETMEA
SET
Keypad
SETTo display and to set the working
point of ORP dosage
MEASURE To set HI 932700 to the measure-
ment mode
Trimmers
SLOPEFor Slope Calibration
SETTo adjust the Set Point
Leds
SET ONTo show that oxidant or reduc-
tant dosage is in progress
6
1. Power Supply
15-24 VDC:115 or 220 VAC:
P1 ! Not ConnectedEarth
P2 !NegativeNeutral
P3 !PositiveLive
2. Oxidant/Reductant dosage selection (see
page 10);
3. Connections for dosing pump to dose either oxidant or reductant products. These
contacts act only as a switch for the power
to the drive;
4. Recorder output(M1 ! + mA output,
M2 ! - mA output);
5. BNC for ORP electrode;
6. Connection for Potential Matching Pin;
7. Connection for Electrode Reference.
Note: Be sure your main line is protected by
a fuse.
7
MECHANICAL DIMENSIONS
Front view of the panel-mounted unit.
Dimensions show the cutout size for the installation and also the outside dimensions of the
panel.
42mm
1.65"
49mm
1.93"
73mm
2.87"
79mm
3.11"
Side view of the panel-mounted unit.
79mm
3.11"
0.25/4mm
0.01/0.160"
76mm
3"
95mm MIN
3.74"
ADJUSTABLE
LOCATION
BRACKET
Adjustable location brackets (supplied with
the meter) allow the controller to slide into the
cutout and will hold the unit securely in place.
95 mm (3.74") is the minimum amount of
space required to install the controller with the
cables connected.
8
SPECIFICATIONS HI 931700
HI 931700
Range0.00 to 14.00 pH
Resolution0.01 pH
Accuracy±0.02 pH
Typical EMC
Deviation
±0.02 pH
Installation Cat.II
Input1012 Ohm
CalibrationOffset: ±2 pH through Offset trimmer
Slope: 80 to 110% through slope trimmer
Readout4-digit LCD plus graphic symbols
Recorder Output 0 to 20mA or 4 to 20mA (not isolated)
Set Point RelayIsolated, 2A, max 240 V,
resistive load, 1.000.000 strokes
Power Supply115±10% or 230±10% VAC, 50/60 Hz
or 15-24VDC (according to the model)
Environment-10 to 50°C (14 to 122°F)
Weight130 g (4.58 oz)
SPECIFICATIONS HI 932700
HI 932700
Range-1000 to +1000 mV
Resolution1 mV
Accuracy±5 mV
Typical EMC
Deviation
Installation Cat.II
Input1012 Ohm
CalibrationSlope: 90 to 110% through slope trimmer
Readout4-digit LCD plus graphic symbols
Recorder Output0 to 20 mA or 4 to 20 mA (not isolated)Set Point RelayIsolated, 2 A, max 240 V,
resistive load, 1.000.000 strokes
Power Supply115±10% or 230±10% VAC, 50/60 Hz
or 15-24VDC (according to the model)
Environment-10 to 50°C (14 to 122°F) Weight130 g (4.58 oz)
±2 mV
9
INITIAL PREPARATION
•Connect a 3-wire (115 or
230VAC versions) or 2-wire
(15-24VDC version) power
Neutral
P1
Earth
P2
P3
Live
cable to the terminal strip
paying attention to the correct live, earth and neutral
(115 or 230VAC version)
or positive and negative
(15-24VDC version) terminal connections.
•Connect the pH or ORP
electrode to the BNC connector marked "INPUT
ELECTRODE".
•To benefit from the differential input, connect the proper electrode wire (if available) or a cable with a potential matching pin (grounding bar) to the relevant
terminal (# 6 at pages 5 and 7).
Note: When it is not possible to immerse
the Potential Matching Pin together with
the electrode in the solution, disable the
differential input by connecting the connection for Potential Matching Pin with
the connection for Electrode Reference
with a jumper wire.
•Recorder Terminals: these
contacts are the output terminals for connection to a
recorder. The output is from
M1
M2
0 to 20 mA or 4 to 20 mA
as indicated and is proportional to the measured pH
or ORP value.
Note: Calibrate the external recorder to be in
accordance with the display reading
(consult recorder manufacturer instruction manual).
•HI 931700 and HI 932700 are single set
point dosing controllers for dosing respectively acid or alkaline / reducing or oxidizing products.
HI 931700:
dose acid (e.g. in
hexavalent chromium re-
If you plan to
S1
S2
duction), make an open circuit between terminals
marked "S1" and "S2" on
the rear panel (description
on page 3).
If you plan to dose base
S1
S2
(e.g. in cyanide oxidation),
short terminals "S1" and
"S2".
HI 932700:
dose oxidants (e.g. in cya-
If you plan to
S1
S2
nide oxidation), make a
short circuit between terminals marked "S1" and "S2"
on the rear panel (description at page 5).
If you plan to dose reductants (e.g. in hexavalent
S1
S2
chromium reduction), make
an open circuit across terminals "S1" and "S2".
•Set Contacts: these contacts (maximum 2A, 220 V)
are used for connection to
dosing pumps. Use HI
931700 for dosage of acidic
C1
C2
or alkaline reagents, and
HI 932700 for dosage of
oxidizing or reducing reagents. Both units act as a
switch for power to the control element (e.g. pump).
Note: all external cables connected to the
rear panel should end with cable lugs.
10
11
OPERATIONAL GUIDE
CALIBRATION
The setting of the various parameters are
made via the front panel keys and trimmers.
When each key is pressed the LED is lighted
indicating to the user that the function is in
operation.
Make sure that the meters and electrodes are
calibrated before operating the instruments
(see page 13).
To set the working point of pH or ORP dosage,
press the "SET" key. The display will show the
set value for the dosage.
SET MEA
SLOPE
SET
pH
Use a small screwdriver to adjust the "SET"
trimmer until the desired set value is displayed.
SET MEA
SLOPE
SET
pH
After setting the pH or ORP
value immerse the electrode in
the test solution and press the
SET MEA
SLOPE
SET
"MEASURE" key. The actual
pH or ORP value of the test
solution is displayed.
When the chemicals are dosed,
the "SET ON" LED will be lit.
12
SET MEA
SET
ON
OFFSET
SLOPE
SET
Make sure the meter is in the measurement
mode (MEASURE LED light is on) before
starting the calibration procedure.
HI 931700:
Note the temperature of the buffer
solution using a thermometer.
Immerse the electrode and the Potential Matching Pin in pH 7.01 buffer
solution (HI 7007).
Note: When it is not possible to immerse
the Potential Matching Pin together with the
electrode in the solution, disable the differential input by connecting the connection for
Potential Matching Pin with the connection
for Electrode Reference with a jumper wire.
SET MEA
SET
ON
OFFSET
SLOPE
SET
pH
Shake briefly and
wait one minute before adjusting the
Offset trimmer to
display pH 7.01 on the LCD if the temperature
of the buffer solution is at 25°C.
If the buffer solution temperature is not 25°C,
refer to Table 1 for the appropriate buffer
value to adjust at the noted temperature.
Rinse the electrode and the Potential Matching Pin thoroughly
in water and immerse them in
4.01pH (HI7004) or 10.01pH
(HI7010) buffer solution.
Shake briefly and
wait one minute before adjusting the
slope trimmer to
SET MEA
SET
ON
OFFSET
SLOPE
SET
pH
display pH 4.01 (or
10.01) on the LCD if the temperature of the
buffer solution is at 25°C, if not refer to Table 1
for appropriate buffer value at the corresponding temperature.
using a calibration solution at the same
temperature of the liquid to be measured.
If calibration is performed at a different
temperature, the error is approx.
0.0035pH for each 1°C difference multiplied by the difference between the pH
of the measured liquid and pH7;
e.g. meter calibrated at 25°C/measured
liquid: pH 4 / liquid temperature: 35°C;
the error is approx.:
0.0035x(35-25)x(7-4) = 0.10 pH
The error is added or subtracted to the
reading according to table 2.
HI 932700:
Immerse the electrode and the
Potential Matching Pin in HI 7020
ORP solution.
Note: When it is not possible to
immerse the Potential Matching Pin together
with the electrode in the solution, disable
the differential input by connecting the connection for Potential Matching Pin with the
connection for Electrode Reference with a
jumper wire.
Adjust the slope trimmer until an ORP value
from 200 mV and 275 mV is displayed.
SET
ON
SET MEA
OFFSET
SLOPE
SET
mV
The calibration is now complete and the instrument is ready for use.
.
Temp.above 25°CTemp. below 25°C
pH above 7SubtractAdd
pH below 7AddSubtract
Table 2
14
15
TAKING REDOX MEASUREMENTS
To make correct redox measurements the
following conditions must prevail:
Redox measurements allow the quantification of the oxidizing or reducing power of a
solution, and are commonly expressed in
mV.
Oxidation may be defined as the process
during which a molecule (or an ion) loses
electrons and reduction as the process by
which electrons are gained.
Oxidation is always coupled together with
reduction so that as one element gets oxidized, the other automatically is reduced,
therefore the term oxidation-reduction is frequently used.
Redox potentials are measured by an electrode capable of absorbing or releasing electrons without causing a chemical reaction
with the elements with which it comes into
contact.
The electrodes most usually available for
this purpose have gold or platinum surfaces;
gold possesses a higher resistance than platinum in conditions of strong oxidation, while
platinum is preferred for the measurements
of oxidizing solutions containing halides and
for more general uses.
When a platinum electrode is immersed in
an oxidizing solution a monomolecular layer
of oxygen is developed on its surface. This
layer does not prevent the electrode from
functioning, but it increases the response
time. The opposite effect is obtained when
the platinum surface absorbs hydrogen in
the presence of reducing mediums. This phenomenon is rough on the electrode.
16
-The surface of the electrode must be
cleaned and smooth.
-The surface of the electrode must undergo a preventive treatment depending whether the solution to be measured has oxidizing or reductive characteristics.
Because the Pt/PtO system depends on the
pH, the pretreatment of the electrode may
be determined by the pH and the redox
potential of the solution to be measured.
As a general rule, if the ORP mV reading
corresponding to the pH solution value is
higher than the value in the Table below, an
oxidizing pre-treatment is necessary; otherwise a reducing pre-treatment is necessary:
Reducing pre-treatment: immerse the electrode for a few minutes in HI 7091.
Oxidizing pre-treatment: immerse the electrode for a few minutes in HI 7092.
If pre-treatment is not performed, the electrode will take a significantly longer time to
respond.
When working with electrodes of the refilling type, the electrolyte used for filling must
be constantly kept at an adequate level (no
less than 2½ centimeters from the filling
hole) and topped up if necessary with HI 7071
refilling solution.
17
In the event that measurements are performed with solutions containing sulfides or
proteins, the cleaning of the diaphragm of
the reference electrode must be performed
(see page 21, "Cleaning Procedure").
ELECTRODE CONDITIONING
AND MAINTENANCE
In order to have a correct functioning of the
ORP electrode, immerse it into HI 7020 and
measure the response; the obtained value
should be within 200 and 275 mV.
After this functional test, it is suggested to
wash the electrode thoroughly with water
and proceed to the oxidizing or reducing
pre-treatment before taking measurements.
When not in use, the electrode tip should be
kept moist and far from any type of mechanical stress which might cause damage.
For this reason , the use of the protective
cap supplied with the electrode is advised.
Reference
Filling Hole
Fill Hole
Screw
Sensitive
Wire
Reference
Wire
Reference
Junction
Glass
Bulb
Reference
Wire
Reference
Junction
Platinum or
Gold tip
Plastic Body
pH Electrode
Reference
Plastic Body
ORP Electrode
Reference
Filling Hole
Reference
Wire
Sensitive
Wire
Reference
Junction
Glass
Bulb
Glass Body
pH Electrode
Wire
Reference
Junction
Platinum or
Gold tip
Glass Body
ORP Electrode
18
PREPARATION
Remove the protective cap.
DO NOT BE ALARMED IF ANY SALT DE-
POSITS ARE PRESENT.
This is normal with electrodes and they will
disappear when rinsed with water.
During transport tiny bubbles of air may have
formed inside the glass bulb. The electrode
19
cannot function properly under these conditions. These bubbles can be removed by
"shaking down" the electrode as you would
do with a glass thermometer.
If the bulb and/or junction are dry, soak the
electrode in HI70300 Storage Solution for at
least one hour.
For refillable electrodes:
If the fill solution (electrolyte) is more than
1 cm (½") below the fill hole, add HI70823,5M KCl Electrolyte Solution for double
junction or HI7071 3,5M KCl+AgCl Electro-lyte Solution for single junction electrodes.
For a faster response unscrew the fill hole
screw during measurements.
Follow the Preparation Procedure above before taking measurements.
Note: NEVER STORE THE ELECTRODE
IN DISTILLED WATER OR DRY.
PERIODIC MAINTENANCE
Inspect the electrode and the cable. The
cable used for the connection to the meter
must be intact and there must be no points
of broken insulation on the cable or cracks
on the electrode stem or bulb.
Connectors must be perfectly clean and dry.
If any scratches or cracks are present, replace the electrode. Rinse off any salt deposits with water.
For AmpHel® electrodes:
If the electrode does not respond to pH
changes, the battery is run down and the
electrode should be replaced.
MEASUREMENT
Rinse the electrode tip with distilled water.
Immerse the tip (4 cm / 1½") in the sample
and stir gently for approx. 30 seconds.
For a faster response and to avoid cross
contamination of the samples, rinse the electrode tip with a few drops of the solution to
be tested, before taking measurements.
STORAGE
To minimize clogging and assure a quick
response time, the glass bulb and the junction should be kept moist and not allowed to
dry out.
Replace the solution in the protective cap
with a few drops of HI 70300 Storage Solu-tion or, in its absence, Filling Solution
(HI7071 for single junction or HI7082 for
double junction electrodes).
AmpHel® is a registered Trademark of "Hanna Instruments"
20
For refillable electrodes:
Refill the electrode with fresh electrolyte
(HI7071 for single junction or HI7082 for
double junction electrodes). Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.
CLEANING PROCEDURE
General
Soak in Hanna HI7061 Gen-eral Cleaning Solution for approximately 1 hour.
Removal of films, dirt or deposits on the
membrane/junction:
Protein
Soak in Hanna HI7073 ProteinCleaning Solution for 15 minutes.
Inorganic
Soak in Hanna HI7074 Inor-ganic Cleaning Solution for 15
minutes.
Oil/grease
Rinse with Hanna HI7077 Oiland Fat Cleaning Solution.
21
IMPORTANT: After performing any of the
cleaning procedures rinse the electrode thoroughly with distilled water, drain and refill
the reference chamber with fresh electrolyte, (not necessary for GEL filled electrodes)
and soak the electrode in HI 70300 StorageSolution for at least 1 hour before taking
measurements.
TROUBLESHOOTING
Evaluate your electrode performance based
on the following.
•Noise (Readings fluctuate up and down)
could be due to:
- Clogged/Dirty Junction: Refer to the
Cleaning Procedure above.
- Loss of shielding due to low electro-
lyte level (in refillable electrodes only):
refill with HI7071 for single junction or
HI7082 for double junction electrodes.
•Dry Membrane/Junction: Soak in Stor-
age Solution HI70300 for at least 1 hour.
ter and then follow the Cleaning Procedure above.
•For ORP Electrodes: polish the metal
tip with a light abrasive paper (paying
attention not to scratch the surface) and
wash throughly with water.
•Drifting: Soak the electrode tip in warm
Hanna Solution HI7082 for one hour and
rinse tip with distilled water. Refill with
fresh HI7071 for single junction electrodes
and HI7082 for double junction electrodes.
•Low Slope: Refer to the cleaning proce-
dure above.
•No Slope: Check the electrode for cracks
in glass stem or bulb (replace the electrode if cracks are found).
•Slow Response/Excessive Drift: Soak
the tip in Hanna Solution HI7061 for 30
minutes, rinse thoroughly in distilled wa-
22
23
TEMPERATURE-RESISTANCE
20
0 +10+20+30+40+50+60+70 +80+90
C
CORRELATION FOR HANNA
pH SENSITIVE GLASS
The resistance of glass electrodes partially
depends on the temperature. The lower the
temperature, the higher the resistance. It
takes longer time for the reading to stabilize
if the resistance is higher. In addition, the
response time will suffer to a greater degree
at temperatures below 10°C.
Ω
9
2x10
-10
9
1x10
-10
8
-10
2x10
8
1x10
-10
7
2x10
-10
-10
7
1x10
-10
-20-
Since the resistance of the pH electrode is
in the range of 200 Mohm, the current across
the membrane is in the pico Ampere range.
Large currents can disturb the calibration of
the electrode for many hours.
Typical Electrode Life
Ambient Temperature 1- 3 years
90 °C Less than 4 months
120°C Less than 1 month
High concentrations of sodium ions interfere
with readings in alkaline solutions; the pH at
which the interference starts to be significant depends upon the composition of the
glass. This interference is the alkaline error
and causes the pH readings to be underestimated. Hanna's glass formulations have the
indicated characteristics.
Alkaline Error
Sodium Ion Correction for the Glass
at 20-25°C
ConcentrationpH Error
°
0.1 Mol L-1 Na
+
13.00
13.50
14.00
1.0 Mol L-1 Na
+
12.50
13.00
13.50
14.00
0.10
0.14
0.20
0.10
0.18
0.29
0.40
For these reasons high humidity environ-
ments, short circuits and static discharges are detrimental for a stable pH
reading.
The pH electrode's life also depends on the
temperature. If constantly used at high temperatures, the electrode life is drastically
reduced.
24
25
SUGGESTED INSTALLATIONS
SHORT DISTANCE, INDOOR INSTALLATION
Due to the high impedance of the pH and
ORP electrode's glass membrane (usually
more than 100 MΩ), a very high grade of
insulation is required.
A dry environment is needed in order to
obtain a level of insulation not lower than
12
10
Ω.
MAXIMUM 10 METERS
DRY ENVIRONMENT
LESS THAN 80%
RELATIVE HUMIDITY
pH METER
WITH CONVENTIONAL ELECTRODES
MEASUREMENTS CAN BE
TAKEN FROM DISTANCES UP TO
10 METERS (33 FEET)
This type of connection is very delicate and
requires constant attention to maintain proper
operating conditions.
The conventional electrodes should be used
in indoor applications only, and should not
use a cable longer than 10m (33').
MAXIMUM 50 METERS
WET ENVIRONMENT
RELATIVE HUMIDITY UP TO 100%
pH METER
WITH AN AmpHel ELECTRODE
TAKEN FROM DISTANCES UP TO
®
MEASUREMENTS CAN BE
50 METERS (165 FEET)
AmpHel® electrodes have a micro-amplifier
in the electrode cap to boost the signal,
drastically reducing susceptibility to noise
and drift.
With all of the components sealed in the
electrode body, moisture up to 100% RH
will not effect the signal.
MEDIUM DISTANCE,
INDOOR/OUTDOOR INSTALLATION
Since the introduction of AmpHel® these distances are no longer a problem. You are
now able to connect your meter directly to
an AmpHel® electrode, saving the cost of a
transmitter or costly coaxial cable.
The standard cable length of the AmpHel
electrode is 5 m (16.5'). Additional lengths
of regular cable up to 50 m (165'), can be
installed without special connectors.
AmpHel® is a registered Trademark of "Hanna Instruments"
26
®
AmpHel® is a registered Trademark of "Hanna Instruments"
27
M13 x 1.5
DIA
16
mm
7
mm
25
mm
M13 x 1.5
DIA
16
mm
7
mm
25
mm
ACCESSORIES
pH CALIBRATION SOLUTIONS
HI7004M pH 4.01 Buffer Solution, 230 mL
HI7004L pH 4.01 Buffer Solution, 460 mL
HI7006M pH 6.86 Buffer Solution, 230 mL
HI7006L pH 6.86 Buffer Solution, 460 mL
HI7007M pH 7.01 Buffer Solution, 230 mL
HI7007L pH 7.01 Buffer Solution, 460 mL
HI7009M pH 9.18 Buffer Solution, 230 mL
HI7009L pH 9.18 Buffer Solution, 460 mL
HI7010M pH 10.01 Buffer Solution, 230 mL
HI7010L pH 10.01 Buffer Solution, 460 mL
ORP SOLUTIONS
HI 7020M 200-275mV Buffer Solution, 230 mL
HI 7020L 200-275mV Buffer Solution, 460 mL
HI 7091M Pre-Treatment Reducing Solution,
230 mL
HI 7091L Pre-Treatment Reducing Solution,
460 mL
HI 7092M Pre-treatment Oxidizing Solution,
230 mL
HI 7092L Pre-Treatment Oxidizing Solution,
460 mL
ELECTRODE STORAGE SOLUTIONS
HI70300MStorage Solution, 230 mL
HI70300L Storage Solution, 460 mL
ELECTRODE CLEANING SOLUTIONS
HI7061M General Cleaning Sol., 230 mL
HI7061L General Cleaning Sol., 460 mL
HI7073M Protein Cleaning Sol., 230 mL
HI7073L Protein Cleaning Sol., 460 mL
HI7074M Inorganic Cleaning Sol., 230 mL
HI7074L Inorganic Cleaning Sol., 460 mL
HI7077M Oil & Fat Cleaning Sol., 230 mL
HI7077L Oil & Fat Cleaning Sol.,460 mL
REFILLING ELECTROLYTE SOLUTIONS
HI70713.5M KCl+ AgCl Electrolyte,
4x50 mL, for single junction electrodes
HI70721M KNO3 Electrolyte, 4x50 mL
HI70823.5M KCl Electrolyte, 4x50 mL,
for double junction electrodes
28
pH ELECTRODES
HI 1090TScrew cap PG13.5 connector,
double junction, glass-body
PG13.5 THREAD
φ 12mmφ 9.5mm
110mm30mm
HI 1110SScrew connector, single junction,
glass-body
HI 1130B/3 BNC connector, 3 m (9.9') cable,
single junction, glass-body
3/4 x 16 UNF
DIA 20.5mm
38.5mm
HI 1110SHI 1130B/3
DIA 9.5mm
DIA 12mm
110mm
HI 1110TScrew cap PG13.5 connector,
double junction, glass-body
PG13.5 THREAD
φ 12mmφ 9.5mm
110mm30mm
HI 1114SScrew connector, double junc-
tion plastic-body
HI 1134B/3 BNC connector, 3 m (9.9') cable,
double junction plastic-body
3/4 x 16 UNF
DIA 20.5mm
38.5mm
HI 1114SHI 1134B/3
DIA 12mm
110mm
HI 1115SScrew connector, single junction,
glass-body
HI 1135B/3 BNC connector, 3 m (9.9') cable,
single junction, glass-body
DIA 16.5mm
5mm
M13 x 1.5
DIA
16
mm
25
7
mm
mm
HI 1115SHI 1135B/3
DIA 7.6mm
25mm
29
DIA 12mm
150mm
M13 x 1.5
DIA
16
mm
7
mm
25
mm
HI 1210TScrew cap PG13.5 connector,
double junction, plastic-body
PG13.5 THREAD
φ 12mm
HI 2910B/5 BNC connector, 5 m (16.5') cable,
double junction, plastic-body,
built-in amplifier
3/4 x 16 UNF
DIA 20.5mm
DIA 12mm
110mm30mm
HI 1910BBNC connector, 1 m (3.3') cable,
double junction, plastic-body,
built-in amplifier
3/4 x 16 UNF
DIA 20.5mm
38.5mm
DIA 12mm
110mm
HI 1911BBNC connector, 1 m (3.3') cable,
double junction, plastic-body,
built-in amplifier
3/4 x 16 UNF
DIA 20.5mm
38.5mm
DIA 12mm
110mm
HI 1912BBNC connector, 1 m (3.3') cable,
double junction, plastic-body,
built-in amplifier
HI 1912B/5 BNC connector, 5 m (16.5') cable,
double junction, plastic-body,
built-in amplifier
3/4 x 16 UNF
DIA 20.5mm
DIA 9.5mm
DIA 12mm
38.5mm
110mm
ORP ELECTRODES
HI 2930B/5 BNC connector, 5 m (16.5') cable,
Pt, Ultem®-body, built-in amplifier
3/4 x 16 UNF
DIA 20.5mm
38.5mm
DIA 12mm
110mm
HI 3110SScrew-type connector, Pt, glass-
body
HI 3130B/3 BNC connector, 3 m (9.9') cable,
Pt, glass-body
3/4 x 16 UNF
DIA 20.5mm
HI 3110SHI 3130B/3
38.5mm
DIA 12mm
110mm
HI 3110TScrew cap PG13.5 connector,
Pt, glass-body
PG13.5 THREAD
φ 12mmφ 1mm
38.5mm
110mm
HI 2114B/5 BNC connector, 5 m (16.5') cable,
double junction, plastic-body
3/4 x 16 UNF
DIA 20.5mm
38.5mm
30
DIA 12mm
110mm
110mm30mm
HI 3115SScrew-type connector, side-arm,
Pt, glass-body
HI 3135B/3 BNC connector, 3 m (9.9') cable,
side-arm, Pt, glass-body
DIA 16.5mm
5mm
M13 x 1.5
DIA
16
mm
25
7
mm
mm
HI 3115SHI 3135B/3
Ultem® is a registered Trademark of "General Electrics Company"
DIA 7.6mm
25mm
31
DIA 12mm
150mm
HI 3210TScrew cap PG13.5 connector,
M13 x 1.5
DIA
16
mm
7
mm
25
mm
Pt, plastic-body
PG13.5 THREAD
φ 12mm
110mm30mm
HI 3410SScrew connector, Pt, plastic-body
HI 3430B/3 BNC connector, 3 m (9.9') cable,
Pt, plastic-body
3/4 x 16 UNF
DIA 20.5mm
38.5mm
HI 3410SHI 3430B/3
DIA 12mm
110mm
HI 3932B/5 BNC connector, 5 m (16.5') cable,
Pt, Ultem®-body, built-in amplifier
3/4 x 16 UNF
DIA 20.5mm
38.5mm
DIA 12mm
110mm
HI 4110SScrew-type connector, Au, glass-
body
HI 4130B/3 BNC connector, 3 m (9.9') cable,
Au, glass-body
DIA
16
mm
3/4 x 16 UNF
DIA 20.5mm
M13 x 1.5
25
7
mm
mm
HI 4110SHI 4130B/3
38.5mm
DIA 12mm
110mm
HI 4932B/5 BNC connector, 5 m (16.5') cable,
Au, Ultem®-body, built-in amplifier
3/4 x 16 UNF
DIA 20.5mm
38.5mm
DIA 12mm
110mm
EXTENSION CABLES FOR SCREW-TYPE
ELECTRODES ONLY (SCREW TO BNC
CONNECTOR)
HI 7855 SERIES CABLE CONNECTORS
CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC
CONNECT TO
SCREW TYPE
ELECTRODES
CONNECT TO THE
BNC SOCKET
OF THE METER
HI7855/1Extension cable 1m (3.3') long
HI7855/3Extension cable 3m (9.9') long
HI7855/5Extension cable 5m (16.5') long
HI7855/10Extension cable 10m (33') long
HI7855/15Extension cable 15m (49.5') long
OTHER ACCESSORIES
BL PUMPSDosing Pumps with Flow Rate
from 1.5 to 20 LPH
ChecktempC Pocket-size thermometer with
penetration probe and 0.1°C
resolution (range -50.0 to
150.0°C)
HI6050Submersible Electrode Holders
& HI 6051
HI 6054Electrode Holders for In-Line
& HI 6057Applications
HI731326Calibration Screwdriver (20 pcs)
HI 7871Level Controllers
& HI 7873
HI8427pH and ORP Electrode
Simulator with 1 m (3.3') Coaxial
Cable ending in Female BNC
Connectors (HI 7858/1)
HI931001pH and ORP Electrode
Simulator with LCD Display and
1 m (3.3') Coaxial Cable ending
in Female BNC Connectors
(HI 7858/1)
MANMINPRR2Instruction Manual
Ultem® is a registered Trademark of "General Electrics Company"
32
33
WARRANTY
All Hanna Instruments meters are warranted
for two years against defects in
workmanship and materials when used for
their intended purpose and maintained
according to instructions.
The probes and the electrodes are
warranted for a period of six months.
Damages due to accident, misuse, tampering
or lack of prescribed maintenance are not
covered. This warranty is limited to repair or
replacement free of charge.
If service is required, contact the dealer
from whom you purchased the instrument. If
under warranty, report the model number,
date of purchase, serial number and the
nature of the failure. If the repair is not
covered by the warranty, you will be notified
of the charge for repair or replacement. When
shipping any instrument, make sure it is
properly packaged for complete protection.
CE DECLARATION OF CONFORMITY
DECLARATION OF CONFORMITY
We
Hanna Instruments Srl
V.le delle industrie 12
35010 Ronchi di Villafranca (PD)
ITALY
herewith certify that the panel mounted pH / ORP controllers
have been tested and found to be in compliance with the following regulations:
IEC 801-2Electrostatic Discharge
IEC 801-3RF Radiated
IEC 801-4Fast Transient
EN 55022Radiated, Class B
Date of Issue: 16-04-1996
HI 931700 HI 932700
D.Volpato - Engineering Manager
On behalf of
Hanna Instruments S.r.l.
To validate your warranty, fill out and return
the enclosed warranty card within 14 days
from the date of purchase.
All rights are reserved. Reproduction in whole
or in part is prohibited without the written
consent of the copyright owner.
Hanna Instruments reserves the right to
modify the design, construction and
appearance of its products without
advance notice.
34
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment
in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences
to radio and TV equipments, requiring the operator to take all necessary steps to correct
interferences.
The trimmers are sensitive to electrostatic discharges. It is recommended to use anti-static
screwdriver.
Any variation introduced by the user to the supplied equipment may degrade the instruments'
EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement
surface exceed 24VAC or 60VDC.
To avoid damages or burns, do not perform any measurement in microwave ovens.
35
PRINTED IN
PORTUGAL
MANMINPRR2
02/99
http://www.hannainst.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.