This Instrument is in Compliance with the CE Directives
www.hannainst.com
Dear Customer,
Thank you for choosing a Hanna Instruments
Product.
Please read this instruction manual carefully
before using the instrument.
This manual will provide you with all the
necessary information for the correct use of
the instrument, as well as a precise idea of its
versatility in a wide range of applications.
These instruments are in compliance with CSA,
UL and (EN 50081-1 and EN 50082-1)
directives.
Remove the instrument from the packing material and examine it to make sure that no
damage has occurred during shipping. If there
is any damage, notify your Dealer.
Note: Save all packing material until you are
sure that the instrument functions correctly. All defective items must be returned in the original packing with the
supplied accessories.
GENERAL DESCRIPTION
Hanna Instruments HI 931402 and HI 931410
are bench pH meters for pH, ORP (Oxidation
Reduction Potential), ISE (Ion Selective Electrode) and temperature measurements.
Their built-in microprocessor ensures accurate and user-friendly operation.
pH measurements are manually or automatically compensated for the temperature effect
(MTC/ATC).
They feature: large liquid crystal display (LCD),
user-friendly keyboard, error codes to guide
the user in calibration and troubleshooting
and storage of calibration data even after the
unit is shut off.
HI 931402 will not allow calibration until the
electrode has stabilized to ensure accurate
pH measurements.
These models meet the most stringent demands of the laboratory: they both automatically recognize five standard calibration buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01).
In HI 931410 a recorder output of 0 to 2 volts
is available and the wide mV range and high
resolution allow for ORP or ion-selective measurement.
1
In HI 931402, through the RS 232C serial port
you can transfer sample measurements to a
PC.
HI 931402 comes supplied with HI 1131B glassbody combination pH electrode, HI 7669/2W
temperature probe (with 1 meter / 3.3' cable),
HI 76405 electrode holder, pH 4.01 and
pH 7.01 buffer solutions (30 ml each), HI 7071
reference electrolyte (30 ml), 12VDC adapter
(HI 710005 or HI 710006) and a dust cover.
HI 931410 comes supplied with HI 1131B glassbody combination pH electrode, HI 7669/2W
temperature probe (with 1 meter / 3.3' cable),
12VDC adapter (HI 710005 or HI 710006) and
a dust cover.
2
FUNCTIONAL DESCRIPTION LCD
Primary Display
HI 931402
MEM
CAL
NOT READY
CON WRONG
LOW BAT
BUF
2
pH
mV
°C1
pH
Secondary Display
HI 931410
°C
pH
Temperature DisplaypH/mV Display
3
FUNCTIONAL DESCRIPTION HI 931402
ins rum n
MEM
1
CAL
NOT READY
CFM WRONG
BUF
2
pH
mV
°C1
pH
CAL Start calibration
/ Select buffer1 value
CFM Confirm when ready
/ Select buffer2 value
CFM Confirm when ready
HI 931402
MICROPROCESSOR
pH METER
MR
CAL
COMM
RANGE
CFM
pH calibration procedure
s
°CMEM
°C
Function
MEM For storing display result
MR Recall previous stored data
COMM For RS232 configuration
RANGE Select pH or ORP measurement
OFF
ON
10
2345678
1. Liquid Crystal Display
2. MEM key to store a value in memory
3. MR key to recall the stored value
4. CAL key to enter and exit the calibration mode
5. CFM key to confirm calibration values
6. RANGE key to select the measurement range
7. COMM key to set the baud rate
↑↑
8.
↑°C and
↑↑
↓↓
↓°C keys to manually set tempera-
↓↓
ture and to select pH buffer values
9. ON to turn the meter on
10. OFF key to turn the meter off
9
DTR
TXD
GND
COMBINATION
pH
REFERENCE
ELECTRODE
ELECTRODE
TEMPERATURE
PROBE
RESET
1
23 4 567
RXD
1
5
ooooo
oooo
96
RS-232
ooooo
oooo
DSR
1. BNC Connector
2. Reference Electrode Connector (Ø 0.08")
3. Reference Electrode Connector (Ø 4 mm)
4. Temperature Probe Connector
5. Reset Button
6. RS 232C Connector
7. DC Power Socket (for HI 710005 or
HI 710006)
4
SPECIFICATIONS HI 931402
RangepH0.00 to 14.00
ISE±399.9 mV
ORP±1999 mV
°C0.0 to 100.0
Resolution pH0.01
ISE0.1 mV
ORP1 mV
°C0.1
AccuracypH±0.01
(@20°C/68°F) ISE±0.2 mV
ORP±1mV
°C±0.5
Typical EMC pH±0.02
Deviation ISE±0.5 mV
ORP±1 mV
°C±2
pH CalibrationAutomatic 1 or 2 points with
5 memorized buffers
(pH 4.01, 6.86, 7.01, 9.18, 10.01)
TemperatureAutomatic or manual
Compensationfrom 0 to 100°C (32 to 212°F)
1. ON and OFF keys to turn the instrument on
and off. (Warning: Push OFF before unplugging instrument. If unplugged before turned
off, the meter's memories will be erased)
2. CAL key to enter the calibration mode
3. CON key to confirm the calibration data
4.°C and °C keys to set temperature during
manual compensation adjustment
5. pH/mV key to select the mV (mV, ORP or ISE
measurements) or the pH range
6. Liquid Crystal Displays
TEMPERATURE
DC POWER
12V to 20V
COMBINED
pH or ORP
ELECTRODES
REFERENCE
ELECTRODE
RESET
RECORDER
0-2 VOLT
PROBE
+
6
1
23
4
5
6
1. BNC Connector
2. Reference Electrode Connector
3. RESET button
4. Recorder Output Connector
5. Temperature Probe Connector
6. DC Power Socket (for HI 710005 or HI 710006)
6
SPECIFICATIONS HI 931410
RangepH
ISE
ORP
°C
Resolution pH
ISE
ORP
°C
AccuracypH
(@ 20°C / 68°F) ISE
ORP
°C
Typical EMC pH
DeviationISE/ORP
°C
RecorderpH
Outputm V
Max. Current
Offset
Slope
Calibration
Temperature
Compensation
Electrode
Temp. Probe
Input Impedance
Power
Environment
Dimensions
Shipping Weight
0.00 to 14.00
0.0 to ±399.9mV
±1999mV
-10.0° to 125.0
0.01
0.1mV
0.1/1mV
0.1
±0.01
±0.2 mV
±1 mV
±0.5
±0.01
±0.2mV
±0.7
0 to 2 V
0 to 2 V
20mA
±1 pH
from 85% to 105%
Automatic 1 or 2 points with
3 memorized buffers
(pH 4.01, 7.01, 10.01)
Manual from 0 to 100°C (32 to 122°F)
Automatic from -10 to 125°C (14 to 257°F)
HI 1131B glass body combination
refillable pH electrode (included)
HI 7669/2W (included)
12
10
Ω
12VDC (CSA, UL & CE approved)
0 to 50°C (32 to 122°F); 95% RH
230x170x70 mm (9.1x6.7x2.7")
1.5 Kg (3.3 lb.)
7
OPERATIONAL GUIDE
ELECTRODE AND PROBE CONNECTIONS
For combination pH or ORP electrodes (with
internal reference) plug the electrode's BNC
to the socket provided (see pages 4 and 6, #1).
For electrode with separate reference, connect the measuring electrode's BNC to the
BNC socket (see pages 4 and 6, #1) and the
reference electrode's plug to the socket provided (see page 4, #2 or #3 for HI 931402, or
page 6, #2 for HI 931410).
For temperature measurements and automatic temperature compensation connect the
temperature probe to the °C socket (see
page 4, #4 for HI 931402, or page 6, #5 for
HI 931410).
POWER CONNECTION
Plug the 12VDC adapter (HI 710005 or
HI 710006) into the DC socket (see page 4,
#7 for HI 931402, or page 6, #6 for HI 931410).
Plug the adapter to the mains.
Note: Make sure the main line is protected
by a fuse.
HI931402 uses an EEPROM and HI 931410
uses a rechargeable battery to retain the pH
calibration, mV calibration, temperature calibration as well as the serial communication
setting. The instrument will store the respective data after a calibration or serial communication setting, even when it is not plugged-in.
In HI 931410 the battery lasts approximately
one month when fully charged, and it is automatically charged any time the meter is
plugged-in (either turned on or off).
8
pH MEASUREMENTS
To take a pH measurement
simply submerge the tip
(4 cm/1½") of the electrode
(1½
")
and the temperature probe into
the sample to be tested.
Turn the instrument on by pressing the ON
key.
ON
ON
The meter automatically defaults to the pH
measurement mode.
Allow 1 or 2 minutes for the electrode to stabilize.
In HI 931402 pH is displayed on the primary
display and temperature on the secondary
one, whereas in HI 931410 pH is displayed
on the left LCD.
pH
°C
°C
pH
4 cm
In order to take accurate pH measurements,
make sure that the instrument has been calibrated for pH before use (see page 15).
In HI 931402, if the calibration data is lost or the meter
is not calibrated, the "CAL"
indicator is displayed to
CAL
prompt the user to perform a
calibration. If the user chooses
not to do it and presses the
CAL key to quit from the calibration mode, the offset and
CAL
slope of the instrument are
set to 0 mV and 100% of the
theoretical value respectively.
9
pH
°C
If measurements are taken in different samples
successively, it is recommended that the
electrode and the temperature probe are rinsed
thoroughly for better conditioning, and then
blot dry between each measurements. For
this rinsing process, it is recommended to
use a liberal amount of the sample to be
measured.
Using refillable electrodes
Unscrew the small refill cap
to open the refill hole during
measurements. After
measurement replace the refill
hole cap.
The pH reading is directly affected by
temperature. In order for the meter to measure
the pH accurately, temperature must be taken
in consideration.
A perfect equilibrium between the pH
electrode and the sample is reached in
approximately 15 minutes.
To use the meter's Automatic
Temperature Compensation
feature, submerge the
HI7669/2W temperature
probe into the sample as
close to the electrode as
possible and wait for 1-2
minutes.
MANUAL TEMPERATURE COMPENSATION
If you know the temperature of the sample to
be tested you can manually compensate for it.
If manual temperature compensation is
desired the temperature probe must be
disconnected from the instrument.
10
The display will then show
the default temperature (25°C
in HI 931402 or 20°C in
HI 931410) or the last recorded
temperature reading (HI 931402
only). "°C" is blinking.
Note the temperature of the
sample using a ChecktempC
or a glass thermometer.
°C
ºC
°C
°C
ºC
Use the UP or DOWN arrow
keys to manually adjust the
display reading to the value of
the reference thermometer.
°C
If the temperature probe is connected to the
meter, the secondary or the temperature LCD
figure display the measured temperature reading and the "°C" indicator does not blink.
Pressing the ↑°C or ↓°C keys will not alter
the temperature reading.
If measurements are taken in different samples
successively, it is recommended that the electrode and the temperature probe are rinsed
thoroughly with deionized water or distilled
water (if not available, tap water may also be
used) and then blot dry between each measurements.
ORP MEASUREMENTS
HI 931402 and HI 931410 have the capability to
take ORP measurements, using an optional
ORP electrode.
Hanna Instruments offers a variety of ORP
electrodes for this purpose (see page 47). Contact your Dealer for more information.
Just as pH measurements allow the acidity or
the alkalinity of a substance to be expressed in
numbers, oxidation-reduction potential (redox)
measurements provide the quantification of the
11
°C
oxidizing or reducing power of the sample tested.
To correctly perform an ORP measurement, the
surface of the ORP electrode must be clean and
smooth.
When not in use, the tip of the electrode should
be kept moist and safe from any mechanical
stress which might cause damage to the glass/
platinum junction.
To get into the "mV" mode
(ORP or ISE measurements)
turn the instrument ON and
pHpH
mVmV
RANGE
press the RANGE (HI 931402)
or the pH/mV (HI 931410)
key.
To measure the mV of a
solution simply submerge the
ORP or ISE electrode tip
(4 cm/1½") into the sample
to be tested. Allow a few
(1½
")
minutes for the readings to
stabilize.
The display will indicate the
absolute mV value (positive
or negative).
mV
4 cm
mV
ºC
When values are less than ±400 mV, tenths of
mV are displayed; for values greater than ±400
mV, the decimal digit disappears. Note that the
change in scale (from 01. mV to 1 mV as
resolution) is automatic.
"---" for HI 931402 or "Err2" for
HI 931410 appear if the value
exceeds ±1999 mV (out of
range).
Note: ORP measurements are taken without
temperature compensation.
12
mV
°C
TEMPERATURE MEASUREMENTS
Taking a temperature measurement is very
easy.
Connect the temperature probe to the
instrument and turn the instrument ON.
Dip the liquid/general
purpose temperature probe
HI 7669/2W into the sample
and allow the reading on the
secondary display to stabilize
(1 or 2 minutes).
The temperature value will be displayed on
the secondary LCD for HI 931402 and on the
left LCD for HI 931410.
pH
°C
°C
pH
HOLD FUNCTION (for HI 931402 only)
To hold and memorize the
measured value, press the
MEM
MEM key.
The last reading will be frozen on the display until the
key is released.
At the same time, this reading is stored in
the memory.
During this operation, the MEM indicator is
lit.
To recall the stored memory value, press the
MR key button and the memorized reading
will be displayed with the MEM indicator lit.
MR
13
MEM
pH
°C
RESET BUTTON
The RESET button (see page 4, #5 for
HI 931402, or page 6, #3 for HI 931410) is
used when the instrument displays erroneous
messages due to strong electrical interference or when the instrument's power supply
was disconnected before the meter was
switched off.
It is necessary to press the reset button and
restart the entire operation.
Calibration points should remain memorized.
It is recommended to verify calibration before
proceeding.
RECORDER OUTPUT (for HI 931410 only)
HI 931410 is provided with a recorder output
from 0 to 2 volts. Connect the recorder output
(see page 6, #4) and select the pH or mV range
by pressing the pH/mV key. The output signal
is related to the values as follows:
pH output:
0.01 pH = 1.42857 mV
e.g.0.00 pH =0 mV
7.00 pH =1000 mV
14.00 pH =2000 mV
mV output:
0.1 mV =0.05 mV
e.g. -1999 mV =0 mV
0.0 mV =1000 mV
+1999 mV =2000 mV
The output impedance is 100 ohms. The maximum output short-circuit current is 20 mA.
AFTER MEASUREMENTS
Press the OFF key to switch the instrument off.
OFF
OFF
14
pH CALIBRATION
It is recommended to calibrate the instrument
frequently, especially if high accuracy is required.
The instrument should be re-calibrated for
pH:
• Whenever the pH electrode or temperature
probe is replaced.
• At least once a month.
• After testing aggressive chemicals.
• After cleaning procedure and changing the
reference electrolyte.
• If greatest accuracy is required.
PREPARATION
Pour small quantities of the buffer solutions
into clean beakers. If possible use plastic
beakers to minimize any EMC interferences
HI 7004
HI 7004
HI 7007
To get accurate readings, use pH 7.01 and
pH 4.01 if you are going to measure acid
samples or pH 7.01 and pH 10.01 for alkaline
measurements.
For accurate calibration use two beakers for
each buffer solution, the first one for rinsing
the electrode, the second one for calibration.
In this way contamination of the buffers is
minimized.
RINSE
HI 7007
CALIBRATION
HI 7007
15
Loading...
+ 38 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.