Hanna Instruments HI 931402 User Manual

Instruction Manual
HI 931402 HI 931410
Microprocessor
Bench-top
pH Meters
ins
rum
n
s
pH calibration procedure
C
AL
S
tart calibra
tion
/
p
H
°
C
H
I 9
M
IC
R
p
H
M
E
T
MEM
°C
M
R
°C
C
A
L
COMM
C
F
M
R
A
N
G
E
S
elect buffer1 value
C
FM
Confirm
w
hen ready
/
S
elect buffer2 value
C
FM
C
onfirm
w
hen ready
3
1
4
0
2
O
P
R
O
C
E
S
E
S
R
O
R
N
OFF
Function
M
E
M
For storing display result
M
R
R
ecall previous stored
C
O
M
M
data
F
or R
S
232 configuration
R
AN
G
E
Select pH
o
r O
R
P m
easurem
e
nt
This Instrument is in Compliance with the CE Directives
www.hannainst.com
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. These instruments are in compliance with CSA, UL and (EN 50081-1 and EN 50082-1) directives.
TABLE OF CONTENTS
Preliminary Examination ........................... 1
General Description .................................. 1
Functional Description LCD ...................... 3
Functional Description HI 931402 ............. 4
Specifications HI 931402 .......................... 5
Functional Description HI 931410 ............. 6
Specifications HI 931410 .......................... 7
Operational Guide ..................................... 8
pH Calibration ......................................... 15
pH Values at Various Temperature .......... 26
Temperature and mv Calibration............. 27
Interface with PC (HI 931402 only) ......... 28
Additional Calibration Procedures &
Information for HI 931402 only ................ 29
Troubleshooting for HI 931402 ................ 33
Troubleshooting for HI 931410 ................ 34
Electrode Conditioning and Maintenance .. 35 Temperature-Resistance Correlation
for Hanna pH Sensitive Glass ................. 39
Accessories ............................................. 41
Electrode Application Reference Guide .. 50
Warranty.................................................. 51
CE Declaration of Conformity ................. 53
ISO 9000 Certified Company since 1992
PRELIMINARY EXAMINATION
Remove the instrument from the packing ma­terial and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer.
Note: Save all packing material until you are
sure that the instrument functions cor­rectly. All defective items must be re­turned in the original packing with the supplied accessories.
GENERAL DESCRIPTION
Hanna Instruments HI 931402 and HI 931410 are bench pH meters for pH, ORP (Oxidation Reduction Potential), ISE (Ion Selective Elec­trode) and temperature measurements.
Their built-in microprocessor ensures accu­rate and user-friendly operation.
pH measurements are manually or automati­cally compensated for the temperature effect (MTC/ATC).
They feature: large liquid crystal display (LCD), user-friendly keyboard, error codes to guide the user in calibration and troubleshooting and storage of calibration data even after the unit is shut off.
HI 931402 will not allow calibration until the electrode has stabilized to ensure accurate pH measurements.
These models meet the most stringent de­mands of the laboratory: they both automati­cally recognize five standard calibration buff­ers (pH 4.01, 6.86, 7.01, 9.18, 10.01).
In HI 931410 a recorder output of 0 to 2 volts is available and the wide mV range and high resolution allow for ORP or ion-selective mea­surement.
1
In HI 931402, through the RS 232C serial port you can transfer sample measurements to a PC.
HI 931402 comes supplied with HI 1131B glass­body combination pH electrode, HI 7669/2W temperature probe (with 1 meter / 3.3' cable), HI 76405 electrode holder, pH 4.01 and pH 7.01 buffer solutions (30 ml each), HI 7071 reference electrolyte (30 ml), 12VDC adapter (HI 710005 or HI 710006) and a dust cover.
HI 931410 comes supplied with HI 1131B glass­body combination pH electrode, HI 7669/2W temperature probe (with 1 meter / 3.3' cable), 12VDC adapter (HI 710005 or HI 710006) and a dust cover.
2
FUNCTIONAL DESCRIPTION LCD
Primary Display
HI 931402
MEM
CAL
NOT READY
CON WRONG LOW BAT
BUF
2
pH
mV
°C1
pH
Secondary Display
HI 931410
°C
pH
Temperature Display pH/mV Display
3
FUNCTIONAL DESCRIPTION HI 931402
ins rum n
MEM
1
CAL
NOT READY CFM WRONG
BUF
2
pH
mV
°C1
pH
CAL Start calibration
/ Select buffer1 value
CFM Confirm when ready
/ Select buffer2 value
CFM Confirm when ready
HI 931402 MICROPROCESSOR pH METER
MR
CAL
COMM
RANGE
CFM
pH calibration procedure
s
°CMEM
°C
Function
MEM For storing display result
MR Recall previous stored data
COMM For RS232 configuration
RANGE Select pH or ORP measurement
OFF
ON
10
2345 678
1. Liquid Crystal Display
2. MEM key to store a value in memory
3. MR key to recall the stored value
4. CAL key to enter and exit the calibration mode
5. CFM key to confirm calibration values
6. RANGE key to select the measurement range
7. COMM key to set the baud rate
↑↑
8.
°C and
↑↑
↓↓
°C keys to manually set tempera-
↓↓
ture and to select pH buffer values
9. ON to turn the meter on
10. OFF key to turn the meter off
9
DTR
TXD
GND
COMBINATION
pH
REFERENCE
ELECTRODE
ELECTRODE
TEMPERATURE
PROBE
RESET
1
23 4 5 6 7
RXD
1
5
ooooo oooo
96
RS-232
ooooo
oooo
DSR
1. BNC Connector
2. Reference Electrode Connector (Ø 0.08")
3. Reference Electrode Connector (Ø 4 mm)
4. Temperature Probe Connector
5. Reset Button
6. RS 232C Connector
7. DC Power Socket (for HI 710005 or HI 710006)
4
SPECIFICATIONS HI 931402
Range pH 0.00 to 14.00
ISE ±399.9 mV
ORP ±1999 mV
°C 0.0 to 100.0
Resolution pH 0.01
ISE 0.1 mV
ORP 1 mV
°C 0.1
Accuracy pH ±0.01 (@20°C/68°F) ISE ±0.2 mV
ORP ±1mV
°C ±0.5
Typical EMC pH ±0.02 Deviation ISE ±0.5 mV
ORP ±1 mV
°C ±2
pH Calibration Automatic 1 or 2 points with
5 memorized buffers
(pH 4.01, 6.86, 7.01, 9.18, 10.01)
Temperature Automatic or manual Compensation from 0 to 100°C (32 to 212°F)
Electrode HI1131B glass body combination
refillable pH electrode (included)
Temperature HI 7669/2W (included) probe
Input impedance 10
12
Computer RS 232C (optoisolated) Interface
Power supply 12 VDC (CSA, UL & CE approved)
Environment 0 to 50°C (32 to 122°F); 95% RH
Dimensions 230x170x70 mm (9.1x6.7x2.7")
Shipping Weight 1.3 kg (2.9 lb.);
kit with holder: 3.2 kg (7 lb.)
5
FUNCTIONAL DESCRIPTION HI 931410
°C
HI 931410
T Buffer Values (pH)
°C 4.01 7.01 10.01
ON
0 4.01 7.13 10.32 10 4.00 7.07 10.18 20 4.00 7.03 10.06
OFF
30 4.02 7.00 9.96
1
40 4.04 6.98 9.88
23
pH
°C
CAL
CON
pH mV
°C
4
5
1. ON and OFF keys to turn the instrument on and off. (Warning: Push OFF before unplug­ging instrument. If unplugged before turned off, the meter's memories will be erased)
2. CAL key to enter the calibration mode
3. CON key to confirm the calibration data
4.°C and °C keys to set temperature during manual compensation adjustment
5. pH/mV key to select the mV (mV, ORP or ISE measurements) or the pH range
6. Liquid Crystal Displays
TEMPERATURE
DC POWER 12V to 20V
COMBINED pH or ORP
ELECTRODES
REFERENCE ELECTRODE
RESET
RECORDER
0-2 VOLT
PROBE
+
6
1
23
4
5
6
1. BNC Connector
2. Reference Electrode Connector
3. RESET button
4. Recorder Output Connector
5. Temperature Probe Connector
6. DC Power Socket (for HI 710005 or HI 710006)
6
SPECIFICATIONS HI 931410
Range pH
ISE
ORP
°C
Resolution pH
ISE
ORP
°C
Accuracy pH
(@ 20°C / 68°F) ISE
ORP
°C
Typical EMC pH DeviationISE/ORP
°C
Recorder pH Output m V
Max. Current
Offset
Slope
Calibration
Temperature Compensation
Electrode
Temp. Probe
Input Impedance
Power
Environment
Dimensions
Shipping Weight
0.00 to 14.00
0.0 to ±399.9mV ±1999mV
-10.0° to 125.0
0.01
0.1mV
0.1/1mV
0.1
±0.01
±0.2 mV
±1 mV
±0.5
±0.01
±0.2mV
±0.7
0 to 2 V 0 to 2 V
20mA
±1 pH
from 85% to 105%
Automatic 1 or 2 points with
3 memorized buffers
(pH 4.01, 7.01, 10.01)
Manual from 0 to 100°C (32 to 122°F)
Automatic from -10 to 125°C (14 to 257°F)
HI 1131B glass body combination
refillable pH electrode (included)
HI 7669/2W (included)
12
10
12VDC (CSA, UL & CE approved)
0 to 50°C (32 to 122°F); 95% RH
230x170x70 mm (9.1x6.7x2.7")
1.5 Kg (3.3 lb.)
7
OPERATIONAL GUIDE
ELECTRODE AND PROBE CONNECTIONS
For combination pH or ORP electrodes (with internal reference) plug the electrode's BNC to the socket provided (see pages 4 and 6, #1).
For electrode with separate reference, con­nect the measuring electrode's BNC to the BNC socket (see pages 4 and 6, #1) and the reference electrode's plug to the socket pro­vided (see page 4, #2 or #3 for HI 931402, or page 6, #2 for HI 931410).
For temperature measurements and auto­matic temperature compensation connect the temperature probe to the °C socket (see page 4, #4 for HI 931402, or page 6, #5 for HI 931410).
POWER CONNECTION
Plug the 12VDC adapter (HI 710005 or HI 710006) into the DC socket (see page 4,
#7 for HI 931402, or page 6, #6 for HI 931410). Plug the adapter to the mains.
Note: Make sure the main line is protected
by a fuse.
HI931402 uses an EEPROM and HI 931410 uses a rechargeable battery to retain the pH calibration, mV calibration, temperature cali­bration as well as the serial communication setting. The instrument will store the respec­tive data after a calibration or serial communi­cation setting, even when it is not plugged-in. In HI 931410 the battery lasts approximately one month when fully charged, and it is auto­matically charged any time the meter is plugged-in (either turned on or off).
8
pH MEASUREMENTS
To take a pH measurement simply submerge the tip (4 cm/1½") of the electrode
(1½ ")
and the temperature probe into the sample to be tested.
Turn the instrument on by pressing the ON key.
ON
ON
The meter automatically defaults to the pH measurement mode.
Allow 1 or 2 minutes for the electrode to stabilize.
In HI 931402 pH is displayed on the primary display and temperature on the secondary one, whereas in HI 931410 pH is displayed on the left LCD.
pH
°C
°C
pH
4 cm
In order to take accurate pH measurements, make sure that the instrument has been cali­brated for pH before use (see page 15).
In HI 931402, if the calibra­tion data is lost or the meter is not calibrated, the "CAL" indicator is displayed to
CAL
prompt the user to perform a calibration. If the user chooses not to do it and presses the CAL key to quit from the cali­bration mode, the offset and
CAL
slope of the instrument are set to 0 mV and 100% of the theoretical value respectively.
9
pH
°C
If measurements are taken in different samples successively, it is recommended that the electrode and the temperature probe are rinsed thoroughly for better conditioning, and then blot dry between each measurements. For this rinsing process, it is recommended to use a liberal amount of the sample to be measured.
Using refillable electrodes
Unscrew the small refill cap to open the refill hole during measurements. After measurement replace the refill hole cap.
The pH reading is directly affected by temperature. In order for the meter to measure the pH accurately, temperature must be taken in consideration.
A perfect equilibrium between the pH electrode and the sample is reached in approximately 15 minutes.
To use the meter's Automatic
Temperature Compensation
feature, submerge the HI7669/2W temperature probe into the sample as close to the electrode as possible and wait for 1-2 minutes.
MANUAL TEMPERATURE COMPENSATION
If you know the temperature of the sample to be tested you can manually compensate for it.
If manual temperature compensation is desired the temperature probe must be disconnected from the instrument.
10
The display will then show the default temperature (25°C in HI 931402 or 20°C in HI 931410) or the last recorded temperature reading (HI 931402 only). "°C" is blinking.
Note the temperature of the sample using a ChecktempC or a glass thermometer.
°C
ºC
°C
°C
ºC
Use the UP or DOWN arrow keys to manually adjust the display reading to the value of the reference thermometer.
°C
If the temperature probe is connected to the meter, the secondary or the temperature LCD figure display the measured temperature read­ing and the "°C" indicator does not blink.
Pressing the °C or °C keys will not alter
the temperature reading.
If measurements are taken in different samples successively, it is recommended that the elec­trode and the temperature probe are rinsed thoroughly with deionized water or distilled water (if not available, tap water may also be used) and then blot dry between each mea­surements.
ORP MEASUREMENTS
HI 931402 and HI 931410 have the capability to take ORP measurements, using an optional ORP electrode. Hanna Instruments offers a variety of ORP electrodes for this purpose (see page 47). Con­tact your Dealer for more information.
Just as pH measurements allow the acidity or the alkalinity of a substance to be expressed in numbers, oxidation-reduction potential (redox) measurements provide the quantification of the
11
°C
oxidizing or reducing power of the sample tested.
To correctly perform an ORP measurement, the surface of the ORP electrode must be clean and smooth.
When not in use, the tip of the electrode should be kept moist and safe from any mechanical stress which might cause damage to the glass/ platinum junction.
To get into the "mV" mode (ORP or ISE measurements) turn the instrument ON and
pHpH mVmV
RANGE
press the RANGE (HI 931402) or the pH/mV (HI 931410) key.
To measure the mV of a solution simply submerge the ORP or ISE electrode tip (4 cm/1½") into the sample to be tested. Allow a few
(1½ ")
minutes for the readings to stabilize.
The display will indicate the absolute mV value (positive or negative).
mV
4 cm
mV
ºC
When values are less than ±400 mV, tenths of mV are displayed; for values greater than ±400 mV, the decimal digit disappears. Note that the change in scale (from 01. mV to 1 mV as resolution) is automatic.
"---" for HI 931402 or "Err2" for HI 931410 appear if the value exceeds ±1999 mV (out of range).
Note: ORP measurements are taken without
temperature compensation.
12
mV
°C
TEMPERATURE MEASUREMENTS
Taking a temperature measurement is very easy.
Connect the temperature probe to the instrument and turn the instrument ON.
Dip the liquid/general purpose temperature probe HI 7669/2W into the sample and allow the reading on the secondary display to stabilize (1 or 2 minutes).
The temperature value will be displayed on the secondary LCD for HI 931402 and on the left LCD for HI 931410.
pH
°C
°C
pH
HOLD FUNCTION (for HI 931402 only)
To hold and memorize the measured value, press the
MEM
MEM key.
The last reading will be fro­zen on the display until the key is released.
At the same time, this reading is stored in the memory.
During this operation, the MEM indicator is lit.
To recall the stored memory value, press the MR key button and the memorized reading will be displayed with the MEM indicator lit.
MR
13
MEM
pH
°C
RESET BUTTON
The RESET button (see page 4, #5 for HI 931402, or page 6, #3 for HI 931410) is used when the instrument displays erroneous messages due to strong electrical interfer­ence or when the instrument's power supply was disconnected before the meter was switched off.
It is necessary to press the reset button and restart the entire operation.
Calibration points should remain memorized. It is recommended to verify calibration before proceeding.
RECORDER OUTPUT (for HI 931410 only)
HI 931410 is provided with a recorder output from 0 to 2 volts. Connect the recorder output (see page 6, #4) and select the pH or mV range by pressing the pH/mV key. The output signal is related to the values as follows:
pH output:
0.01 pH = 1.42857 mV
e.g. 0.00 pH = 0 mV
7.00 pH = 1000 mV
14.00 pH = 2000 mV
mV output:
0.1 mV = 0.05 mV
e.g. -1999 mV = 0 mV
0.0 mV = 1000 mV
+1999 mV = 2000 mV
The output impedance is 100 ohms. The maxi­mum output short-circuit current is 20 mA.
AFTER MEASUREMENTS
Press the OFF key to switch the instrument off.
OFF
OFF
14
pH CALIBRATION
It is recommended to calibrate the instrument frequently, especially if high accuracy is re­quired.
The instrument should be re-calibrated for pH:
• Whenever the pH electrode or temperature probe is replaced.
• At least once a month.
• After testing aggressive chemicals.
• After cleaning procedure and changing the reference electrolyte.
• If greatest accuracy is required.
PREPARATION
Pour small quantities of the buffer solutions into clean beakers. If possible use plastic beakers to minimize any EMC interferences
HI 7004
HI 7004
HI 7007
To get accurate readings, use pH 7.01 and pH 4.01 if you are going to measure acid samples or pH 7.01 and pH 10.01 for alkaline measurements.
For accurate calibration use two beakers for each buffer solution, the first one for rinsing the electrode, the second one for calibration. In this way contamination of the buffers is minimized.
RINSE
HI 7007
CALIBRATION
HI 7007
15
PROCEDURE FOR HI 931402
Calibration of HI 931402 has a choice of 5 memorized standard buffers: 4.01, 6.86, 7.01,
9.18 and 10.01 pH.
• Make sure that the meter is in pH mode.
• Remove the protective cap from the elec­trode and rinse it with some of the buffer calibration solution you are going to use first.
ONE POINT CALIBRATION
One point calibration at pH 6.86 (NBS):
Immerse the pH electrode
and the temperature probe into pH 6.86 buffer solution and shake briefly.
HI 7006
Note: the electrode should be submerged approxi­mately 4 cm (1½") into the solution. The temperature
(1½ ")
probe should be located as close as possible to the pH electrode.
• Press the CAL key. The "CAL" and "
BUF
indicators will be displayed. The second­ary LCD will display "4.01".
4 cm
1
"
CAL
CAL
1
BUF
pH
• Press the °C key. The secondary display will change to “6.86”.
°C
CAL
1
BUF
pH
• When the electrode is submerged in the buffer solution, the meter will notify the user if the reading is not stable. If the readings fluctuate the LCD will blink "NOT READY".
16
NOT READY
Only when the reading is stable, "READY" and "CON"
READY
CON
will blink.
• Press the CFM key to con­firm the calibration. If the reading is not close to the
selected buffer, "WRONG
BUF
1
"
CFM
and "WRONG " will blink alternatively. If the reading is close to the selected
WRONG
WRONG
1
buffer, the meter stores the
BUF
reading (and adjusts the off­set point). The buffer value is then displayed on the primary LCD and the secondary LCD
CAL
1
BUF
will display "4.01".
• Press the CAL key and the
CAL
calibration process is ended with only the offset of the meter calibrated.
For best accuracy however, it is recommended that a two point calibration is performed (see page 19).
One point calibration at pH 7.01 (STAN­DARD):
Immerse the pH electrode
and the temperature probe into pH 7.01 buffer solution
HI 7007
and shake briefly.
Note: the electrode should be submerged approximately 4 cm (1½") into the solution. The temperature probe should be located as close as pos-
(1½ ")
sible to the pH electrode.
• Press the CAL key. The “CAL” and “
BUF
indicators will be displayed. The second­ary LCD will display “4.01”.
CAL
17
CAL
1
BUF
pH
pH
4 cm
1
• Press the °C key twice. The secondary display will change to “7.01”.
°C
CAL
1
BUF
pH
• When the electrode is submerged in the buffer solution the meter will notify the user the reading is not stable. If the readings fluctuate the LCD will blink “NOT
NOT READY
READY”. Only when the reading is stable, “READY” and “CON”
READY
CON
will blink.
• Press the CFM key to con­firm the calibration. If the reading is not close to the
selected buffer, “WRONG
BUF
1
CFM
and “WRONG ” will blink al­ternatively. If the reading is close to the selected buffer
WRONG
WRONG
1
BUF
the meter stores the read­ing (and adjust the offset point). The buffer value is than displayed on the pri­mary LCD and the second-
CAL
1
BUF
ary LCD will display “4.01”.
• Press the CAL key and the calibration process is ended
CAL
with only the offset of the meter calibrated.
pH
pH
For best accuracy however, it is recommended that a two point calibration is performed (see page 19).
18
TWO POINT CALIBRATION
F
A two point calibration is recommended for best accuracy. As a second point use pH 4.01 if you are going to measure acid samples (pH 7 or less), use pH 9.18 or pH 10.01 if you are going to measure alkaline samples.
Two point calibration at pH 6.86 and pH 4.01 or pH 9.18 (NBS):
Proceed as described in “One point calibra­tion at pH 6.86” but do not quit calibration by pressing the CAL key at the end.
After the first calibration point is confirmed, immerse the pH electrode and the temperature probe into the second buffer
HI 7004
solution (pH 4.01 or pH 9.18) and shake briefly.
Note: the electrode should be submerged approxi­mately 4 cm (1½") into the solution. The temperature
(1½ ")
probe should be located as close as possible to the pH electrode.
Select the second buffer value on the secondary display by pressing the
°C
°C
ºC
or °C key.
When the “READY” and “CON” symbols blink on the display the reading is stable
READY
CON
and the calibration can be confirmed.
Press the CFM key. If the
CFM
reading is not close to the selected buffer solution,
WRONG
BU
” and “WRONG
2
will blink alternatively. If the reading is close to the selected buffer, the slope and the offset are calibrated.
WRONG
WRONG
BUF
2
4 cm
19
The values will be stored in memory and the meter will return to the operating mode.
Two point calibration at pH 7.01 and pH 4.01 or pH 10.01 (STANDARD):
Proceed as described in “One point calibra­tion at pH 7.01” but do not quit calibration by pressing the CAL key at the end.
After the first calibration point is confirmed, immerse the pH electrode and the temperature probe into the second buffer solution (pH 4.01 or pH 10.01)
HI 7004
and shake briefly.
Note: the electrode should be submerged approximately 4 cm (1½") into the solution. The temperature probe should be located as close as pos-
(1½ ")
sible to the pH electrode.
Select the second buffer value on the secondary display by
°C
ºC
pressing the °C or °C key.
When the “READY” and “CON” symbols blink on the display the reading is stable and the
READY
CON
calibration can be confirmed.
Press the CFM key. If the reading is not close to the
CFM
selected buffer solution,
WRONG
BUF
” and “WRONG
2
will blink alternatively. If the reading is close to the se­lected buffer, the slope and the offset are calibrated.
WRONG
WRONG
BUF
2
The values will be stored in memory and the meter will return to the operating mode.
Note: the meter will automatically skip the
buffer that was used for the first cali­bration to avoid erroneous calibration.
4 cm
20
Note: during calibration, the
secondary LCD displays the selected buffer value. By pressing the RANGE key the tem­perature value can be also displayed.
CAL
BUF
pH
1
pH
RANGE
CAL
BUF
pH
°C
1
This will allow you to check the buffer temperature during calibration.
21
PROCEDURE FOR HI 931410
HI 931410 provides three technical pre-pro­grammed calibration buffers: t1, t2, t3 i.e. pH 7.01, 4.01, 10.01 respectively.
When the ATC probe is not used, make sure the temperature is set manually to reflect the tem­perature of the solution (see page 26).
Switch the instrument on after
connecting the pH electrode
ON
and the temperature probe.
ONE POINT CALIBRATION
Immerse the pH electrode and
the temperature probe into a pH 7.01 buffer solution. Shake briefly and wait 1-2 min-
HI 7007
utes for thermal equilibrium.
Note: the electrode should be submerged approximately 4 cm (1½") into the solution. The temperature probe should be located as close as pos-
(1½ ")
sible to the pH electrode.
Press the CAL key. The symbol "t1" will
appear on the display.
CAL
4 cm
Press the CON key. The buffer value will be
displayed on the right LCD.
CON
pH
Note: the "pH" symbol will be flashing through-
out the calibration procedure.
22
Allow 1-2 minutes for the pH reading to stabi-
lize.
The pH reading will be temperature compen­sated.
E.g. if the buffer temperature is at 25°C, the right display will show "pH 7.01". If the buffer temperature is at 20°C, the display will show "pH 7.03" (see page 26).
°C
pH
If "Err4" appears the cali­bration solution is wrong. Repeat the process using fresh buffers
Note: If you are not using the ATC probe,
refer to page 26 to manually correct for temperature.
• Press the CON key again to ac-
cept the buffer value. The first
CON
point calibration is now complete.
• If the pH/mV key is pressed at
this point, the calibration process is ended and only the offset of the
pHpH
mVmV
meter is calibrated. This is known as one point calibration.
For greatest accuracy, however, it is recom­mended that a two-point calibration is per­formed.
TWO POINT CALIBRATION
• Proceed as described in “One point calibration”
but do not quit calibration by pressing the pH/mV key at the end.
• After the first calibration point
is confirmed, immerse the pH electrode and the temperature probe into the second buffer solution (pH 4.01 or pH 10.01)
HI 7004
and shake briefly.
23
Note: the electrode should be submerged approximately 4 cm (1½") into the solution. The temperature probe should be located as close as pos-
(1½ ")
sible to the pH electrode.
• Rinse the electrode and the temperature probe
before immersing them into pH 4.01/ pH10.01 calibration buffer, shake briefly and wait for thermal equilibrium.
The meter automatically rec-
ognizes the buffer value and prompts for the second at "t2" (pH 4.01, acid) or "t3" (pH 10.01, alkaline).
Press the CON key to confirm the pH value
and the display will indicate the temperature compensated buffer value e.g. if the tempera­ture is at 25°C, the display will show "pH 4.01" or "10.01", depending on the solution used.
4 cm
CON
°C
pH
If the temperature is at 20°C, the display will show respectively "pH 4.00" or "pH 10.06" (see page 26).
Press CON key again to confirm calibration.
The pH symbol will stop flashing and the unit is now calibrated and ready for measure­ments.
CON
°C
pH
24
• If the reading is drifting, turn the meter off and
repeat the calibration procedure waiting longer time for the reading to stabilize. However, usually 1 or 2 minutes are enough to prevent drifting of the reading.
Note: If you need to unplug the instrument,
press the OFF key thus switching the instrument off before disconnecting the power cord. In this way the instrument will retain the calibration data. Other­wise, re-calibration might be necessary. In the case of a power failure while the instrument is operational, re-calibration might be required.
25
pH VALUES AT VARIOUS TEMPERATURE
Temperature has an effect on pH. The calibra­tion buffer solutions are effected by temperature changes to a lesser degree than normal solu­tions.
During calibration the meter will automatically calibrate to the pH value corresponding to the measured or set temperature.
For manual temperature compensation please refer to the following chart.
pH Buffers
°C °F 4.01 6.86 7.01 9.18 10.01
0
5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95
32 41 50 59 68 77 86
95 104 113 122 131 140 149 158 167 176 185 194 203
4.01
4.00
4.00
4.00
4.00
4.01
4.02
4.03
4.04
4.05
4.06
4.07
4.09
4.11
4.12
4.14
4.16
4.17
4.19
4.20
6.98
6.95
6.92
6.90
6.88
6.86
6.85
6.84
6.84
6.83
6.83
6.84
6.84
6.85
6.85
6.86
6.86
6.87
6.88
6.89
7.13
7.10
7.07
7.04
7.03
7.01
7.00
6.99
6.98
6.98
6.98
6.98
6.98
6.99
6.99
6.99
7.01
7.02
7.03
7.04
9.46
9.39
9.33
9.27
9.22
9.18
9.14
9.10
9.07
9.04
9.01
8.99
8.97
8.95
8.93
8.91
8.89
8.87
8.85
8.83
10.32
10.24
10.18
10.12
10.06
10.01
9.96
9.92
9.88
9.85
9.82
9.79
9.77
9.76
9.75
9.74
9.73
9.74
9.75
9.76
For instance, if the buffer temperature is at 25°C, the display will show pH 4.01 or 7.01 or
10.01.
If the buffer temperature is at 20°C, the display will show pH 4.00/7.03/10.06.
If the buffer temperature is at 50°C, the display will show pH 4.06/6.98/9.82.
26
TEMPERATURE AND mV CALIBRATION
All Hanna bench-top pH meters have been factory calibrated for the temperature and mV and is ready for measurements.
The temperature probes and ORP electrodes are interchangeable and no temperature or mV calibration is needed when the tempera­ture probe or the ORP electrode is replaced.
If, for any reason, the temperature or the mV measurements are out of accuracy, the tem­perature or the mV re-calibration should be carried out.
In any case, we recommend that you have your bench-top meters re-calibrated in tem­perature and mV once a year.
In HI 931402 it is possible to calibrate the meter for temperature and mV (see page 29), however it is recommendable that these cali­brations are performed by technical personnel only.
For an accurate recalibration, contact your nearest Hanna Service Center.
27
INTERFACE WITH PC (HI931402 ONLY)
Data transmission from the instrument to the PC is now much easier with the new HI 92000 Windows® compatible application software of­fered by Hanna Instruments.
User friendly, HI 92000 offers a variety of features and has an on line help feature to support you throughout all situations.
HI 92000 allows you to use the powerful means of the most diffused spread sheet programs (e.g. Excel©, Lotus 1-2-3©). Simply run your favorite spread sheet and open the file down­loaded by HI 92000. It is then possible to make any elaboration available with your soft­ware (e.g. graphics, statistical analysis).
To install HI 92000 you need a 3.5" drive and few minutes to follow the instructions conve­niently printed on the disk's label.
In your copy of HI92000 you will also find a set of configuration files for your Windows Terminal® program to send commands di­rectly from the PC keyboard (see appendix for commands description and usage).
Contact your Hanna Dealer to request a copy.
To connect your HI 931402 to the PC use HI 920010, available through your Hanna Dealer. Make sure that your meter is switched off and plug the connectors, one into the meter RS232C connector, the other into the serial port of your PC.
Note: Cables different from the HI 920010 may
use a different configuration. In such case any communication between the meter and the PC is not possible.
Excel© Copyright of "Microsoft Co." Lotus 1-2-3© Copyright of "Lotus Co." Windows® and Windows Terminal® are registered Trademark of "Microsoft Co."
28
ADDITIONAL CALIBRATION PROCEDURES
& INFORMATION FOR HI931402 ONLY
(for technical personnel only)
For HI931402 only, you can perform the tem­perature and mV calibration as follows.
TEMPERATURE CALIBRATION OF HI931402
Initial Preparation
• Prepare a vessel containing ice and water and another one containing hot water (at a temperature of at least 50°C). Place insula­tion material around the container to mini­mize temperature changes.
• Use a ChecktempC or
C
°
C
°
a calibrated thermom­eter with a resolution of 0.1 °C as a refer-
0 °C
(32 °F)
50 ºC
(122 ºF)
ence thermometer.
Procedure
• Switch the instrument on while pressing the CAL key. The "CAL" indicator will be lit. The secondary LCD section will show 0.0°C.
ON
CAL
+
CAL
°C
°
• Immerse the temperature probe
C
in the vessel with the ice and water as near to the ChecktempC as possible.
0 °C
(32 °F)
• Wait for temperature reading to stabilize.
• Use the °C and °C keys to set the reading on the sec-
°C
ºC
ondary LCD section to the noted temperature of the ice with the water.
• Press CFM key. The secondary LCD sec­tion will show 50.0°C.
CFM
CAL
°C
29
°
• Immerse the temperature probe
C
in the vessel with hot water.
• When the temperature reading has stabilized, use the °C and
°C
keys to set the reading on
the secondary LCD to the
°C
50 °C
(122 °F)
ºC
noted temperature of the hot water.
• Press CFM key. The primary LCD
CFM
CAL
section will show "390.0 mV".
• Complete your temperature calibra-
CAL
tion by pressing the CAL key if you do not want to perform the mV calibration mode.
mV CALIBRATION OF HI 931402
Initial Preparation
Prepare HI 931001 or a simulator that can
provide ±390.0 mV with accuracy of ±0.1mV and ±1000 mV with accuracy of ±1 mV.
Procedure
Continue from the tempera-
ture calibration, the primary
CAL
LCD section will show "390.0 mV".
Select 390.0 mV of the simulator.
Wait for 3 sec., press the CFM
CFM
key. The primary LCD will display the next reading to be selected on the simulator.
Repeat previous two steps for -390.0 mV,
1000 mV and -1000 mV.
At the end of the mV calibration, the meter
will return to the pH measurement mode.
mV
mV
INTERFACE WITH PC
If you are not using Hanna Instruments HI 92000 application software, please find here below some additional information to help your connection to the PC.
30
Setting the Baud Rate and the Command Prefix
The transmission speed (baud rate) of your HI 931402 and of the external device must be the same. To set the baud rate of the meter press the COMM key, the primary
COMM
LCD shows the cur­rent baud rate.
The following baud rate can be selected through the
°C
°C
ºC
and °C keys: 150, 300, 600, 1200 (factory setting) and 2400.
Press the CFM key to confirm the setting and the primary LCD shows the current com- mand prefix; 16 is the factory setting.
CFM
Note: the Command Prefix has not to be changed using HI92000 Hanna Software.
Select a different command pre­fix (between 0 to 47 decimal)
°C
ºC
by pressing the °C and °C keys .
Press the CFM key to confirm the
CFM
setting.
Sending Commands from PC
With terminal programs such as, for example, Telix®, Windows Terminal®, it is possible to remotely control your HI 931402. Use HI 920010 cable to connect the meter to the PC, start the terminal program and set the communication options as follows: 8, N, 1, no flow control.
Command Types
To send a command to the pH meter the scheme is: <DLE> <command> <CR>
Windows Terminal® are registered Trademark of "Microsoft Co." TELIX® is registered Trademark of "Deltacomm"
31
This line makes the computer send a Data Link Escape character, the command ex­pressed as a number or a 3-character se­quence) and a CR character.
Note: Windows Terminal® and all the other terminal programs that support the ANSI es­cape sequence, represent the DLE character by the string '^P' and the CR character by the string '^M'. E.g. the line '^PPHR^M' sets the range to pH.
Commands not requiring an answer from the pH meter:
PHR sets the range to pH MVR sets the range to mV CAL is equivalent to pressing the CAL key CFM is equivalent to pressing the CFM key UPC is equivalent to pressing the °C key DWC is equivalent to pressing the °C key MEM is equivalent to pressing the MEM key MRR is equivalent to pressing the MR key COM is equivalent to pressing the COMM key OFF is equivalent to pressing the OFF key
Commands requiring an answer:
pH? Causes the meter to send the pH value. If
the reading is out of range "Err 1" is sent. If the range is set to mV, "Err 6" is sent.
MV? Causes the meter to send the mV value. If
the reading is out of range "Err 2" is sent. If the range is set to pH, "Err 6" is sent.
TM? Causes the meter to send the temperature
value. If the reading is out of range "Err 3" is sent.
These commands may be sent with either capital or small letters. Invalid commands will be ignored. The characters sent by HI 931402 are always capital letters. When the meter receives an unknown or corrupted commands, it will send a character CAN (ASCII Code 24).
Windows Terminal® are registered Trademark of "Microsoft Co."
32
TROUBLESHOOTING FOR HI 931402
F
F
Symptoms Problem Solution
The meter does not work with the temperature probe The meter fails to calibrate or gives faulty readings
The meter is slow in responding or gives faulty readouts
The meter does not accept the 2 buffer solution for calibration
The reading drifts
Display shows:
Display shows:
Defective ATC probe
Defective pH electrode
The electrode is not working or the reference junction is clogged
Defective pH
nd
electrode
Defective pH electrode
Out of range-pH scale
pH
Out of range-°C scale
pH
ºC
Replace the probe
Replace the electrode
Leave the electrode in a storage solution after cleaning the junction. If problem persists, replace the electrode
Follow the cleaning procedure. If this doesn't work replace the electrode
Replace the electrode
a) Check the calibration b) Assure the pH sample
is in the range 0 to 14
c) Check the level of
electrolyte and the state of the electrode itself
Make sure the °C is within 0-100° and temperature probe is plugged
Display shows:
Display shows:
1
"WRONG
BU
and/or
"WRONG "
mV
"
Out of range-mV scale
Erroneous buffer solution used for offset cal.
Defective
Electrode not connected
Make sure the buffer solution used is pH7 and replace if necessary
Replace the electrode
electrode
Display shows: "WRONG and/or
BU
"
2
"WRONG "
Erroneous buffer solution used for slope cal. Defective electrode
Use fresh buffer solution
Replace the electrode
33
TROUBLESHOOTING FOR HI 931410
Symptoms are:
The instrument does not work with the HI7669/2W probe
The meter fails to calibrate or gives erroneous results
The meter is slow in responding or gives erroneous readouts
The meter does not accept the 2nd buffer solution
The reading drifts
"Err1" on the display
"Err3" on the display
"Err2" on the display
"Err4" on the display
"Err5" on the display
The Problem is:
A defective ATC Probe
A defective pH electrode
The electrode is not working or the diaphragm is clogged
A defective pH electrode or bad buffer solution
Temperature not stable
Out of range on the pH scale
Out of range in °C
Out of range in mV
Erroneous buffer solution used for offset calibration
Defective electrode
Erroneous buffer solution used for slope calibration
Defective electrode
Possible Solutions:
Replace the probe.
Replace the electrode and repeat the pH cali­bration.
Leave the electrode in a storage solution after cleaning the diaphragm. If the problem is not solved, replace the elec­trode and re-calibrate the pH scale.
Try the cleaning proce­dure. If this does not work, replace the elec­trode and re-calibrate the pH. Use fresh buffer solution.
Thermal equilibrium not reached. Wait longer.
a) Verify the pH elec­trode is connected. b) Make sure the pH sample is included be­tween 0 and 14 pH. c) Check the level of the electrode's filling electrolyte and the state of the electrode itself.
Verify the temperature probe is connected and make sure the °C is within -10/125°C.
Verify the temperature probe is connected and make sure the mV is within ±1999 range.
Make sure that the buffer solution used is of pH 7 and replace if necessary.
Replace the electrode and re-calibrate the pH scale.
Be sure the correct pH buffer solution is being used for slope calibra­tion (4.01, 9.18 or 10.01 pH).
Check the electrode and replace if neces­sary.
34
ELECTRODE CONDITIONING
AND MAINTENANCE
Reference Filling Hole
Reference Filling Hole
Sensitive Wire
Reference Wire
Reference Junction
Glass Bulb
Reference Wire
Reference Junction
Platinum or Gold tip
Plastic Body
pH Electrode
Reference
Plastic Body
ORP Electrode
Glass Body
pH Electrode
Wire
Glass Body
ORP Electrode
Reference
Wire
Sensitive
Wire
Reference
Junction
Glass
Bulb
Reference
Junction
Platinum or
Gold tip
PREPARATION
Remove the protective cap. DO NOT BE ALARMED IF ANY SALT DE-
POSITS ARE PRESENT.
This is normal with electrodes and they will disappear when rinsed with water.
During transport tiny bubbles of air may have formed inside the glass bulb. The electrode cannot function properly under these condi­tions. These bubbles can be removed by
35
"shaking down" the electrode as you would do with a glass thermometer.
If the bulb and/or junction are dry, soak the electrode in HI70300 or HI 80300 Storage So­lution for at least one hour.
For refillable electrodes:
If the filling solution (electrolyte) is more than 1 cm (½") below the fill hole, add HI 7082 or HI 8082 3,5M KCl Electrolyte Solution for double junction or HI 7071 or HI8071 3,5M KCl+AgCl Electrolyte Solution for single junction electrodes.
For a faster response, unscrew the fill hole screw during measurements.
For AmpHel electrodes:
If the electrode does not respond to pH changes, the battery is run down and the electrode should be replaced.
MEASUREMENT
Rinse the electrode tip with distilled water. Immerse the tip (4 cm /1½") in the sample and
stir gently for at least 30 seconds. For a faster response
and to avoid cross con­tamination of the samples, rinse the elec­trode tip with a few drops of the solution to be tested, before taking measurements.
STORAGE
To minimize clogging and assuring a quick response time, the glass bulb and the junction should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of HI 70300 or HI 80300
Storage Solution or, in its absence, Filling Solution (HI 7071 or HI 8071 for single junc-
tion or HI 7082 or HI8082 for double junction electrodes). Follow the Preparation Proce­dure above before taking measurements.
Note: NEVER STORE THE ELECTRODE IN
DISTILLED WATER OR DRY.
36
PERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable used for connection to the meter must be intact and there must be no points of broken insulation on the cable or cracks on the electrode stem or bulb. Connectors must be perfectly clean and dry.
If any scratches or cracks are present, re­place the electrode.
Rinse off any salt deposits with water.
For refillable electrodes:
Refill the reference chamber with fresh elec­trolyte (HI7071 or HI 8071 for single junction or HI 7082 or HI 8082 for double junction elec­trodes). Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.
CLEANING PROCEDURE
General Soak in Hanna HI 7061 or HI8061
General Cleaning Solution for
approximately 1 hour.
Removal of films, dirt or deposits on the membrane/junction:
- Protein Soak in Hanna HI7073 or HI8073
Protein Cleaning Solution for 15 minutes.
- Inorganic Soak in Hanna HI 7074 or HI 8074
Inorganic Cleaning Solution
for 15 minutes.
- Oil/grease Rinse with Hanna HI 7077 or
HI 8077 Oil and Fat Cleaning Solution.
IMPORTANT: After performing any of the
cleaning procedures rinse the electrode thor­oughly with distilled water, refill the reference chamber with fresh electrolyte, (not neces­sary for GEL filled electrodes) and soak the electrode in HI 70300 or HI 80300 Storage Solution for at least 1 hour before taking measurements.
37
TROUBLESHOOTING
Evaluate your electrode performance based on the following.
Noise (Readings fluctuate up and down)
could be due to:
- Clogged/Dirty Junction: Refer to the Cleaning Procedure above.
- Loss of shielding due to low electrolyte level (in refillable electrodes only): refill with fresh HI7071 or HI 8071 for single junction or HI7082 or HI 8082 for double junction electrodes.
Dry Membrane/Junction: Soak in Stor­age Solution HI 70300 or HI 80300 for at
least 1 hour.
Drifting: Soak the electrode tip in warm Hanna Solution HI 7082 or HI 8082 for one hour and rinse tip with distilled water. Refill with fresh HI 7071 or HI8071 for single junction electrodes and HI 7082 or HI8082 for double junction electrodes.
Low Slope: Refer to the cleaning proce­dure above.
No Slope: Check the electrode for cracks in glass stem or bulb and replace the electrode.
Slow Response/Excessive Drift: Soak the tip in Hanna Solution HI 7061 or HI8061 for 30 minutes, rinse thoroughly in dis­tilled water and then follow the Cleaning Procedure above.
38
Temperature-Resistance Correlation
20
0 +10+20 +30+40 +50+60+70+80+90
C
for HANNA pH sensitive glass
The resistance of glass electrodes partially depends on the temperature. The lower the temperature, the higher the resistance. It takes longer time for the reading to stabilize if the resistance is higher. In addition, the response time will suffer to a greater degree at tempera­tures below 10°C.
9
2x10
-10
9
1x10
-10
8
-10
2x10
8
1x10
-10
7
2x10
-10
-10
7
1x10
-10
-20-
°
Since the resistance of the pH electrode is in the range of 200 Mohm, the current across the membrane is in the pico Ampere range. Large currents can disturb the calibration of the elec­trode for many hours.
For these reasons high humidity environ-
ments, short circuits and static discharges
are detrimental for a stable pH reading.
The pH electrode's life also depends on the temperature. If constantly used at high tem­peratures, the electrode life is drastically re­duced.
39
Typical Electrode Life Ambient Temperature 1- 3 years 90 °C Less than 4 months 120°C Less than 1 month
High concentrations of sodium ions interfere with readings in alkaline solutions; the pH at which the interference starts to be significant depends upon the composition of the glass. This interference is the alkaline error and causes the pH to be underestimated. Hanna's glass formulations have the indicated characteristics.
Alkaline Error
Sodium Ion Correction for the Glass
at 20-25°C
Concentration pH Error
0.1 Mol L-1 Na
1.0 Mol L-1 Na
+
13.00
13.50
14.00
+
12.50
13.00
13.50
14.00
0.10
0.14
0.20
0.10
0.18
0.29
0.40
40
ACCESSORIES
pH CALIBRATION SOLUTIONS:
HI77400P pH 4.01 & 7.01 Buffer Solution,
20 mL sachet, 5 pcs each
HI 7004M pH 4.01 Buffer Solution, 230 mL HI 7004L pH 4.01 Buffer Solution, 460 mL HI 7006M pH 6.86 Buffer Solution, 230 mL HI 7006L pH 6.86 Buffer Solution, 460 mL HI77700P pH 7.01 Buffer Sol. 10x20 mL HI 7007M pH 7.01 Buffer Solution, 230 mL HI 7007L pH 7.01 Buffer Solution, 460 mL HI 7009M pH 9.18 Buffer Solution, 230 mL HI 7009L pH 9.18 Buffer Solution, 460 mL HI770710P pH 7.01 & 10.01 Buffer Solution,
20 mL sachet, 5 pcs each
HI 7010M pH 10.01 Buffer Solution, 230 mL HI 7010L pH 10.01 Buffer Sol., 460 mL
pH CALIBRATION SOLUTIONS IN FDA APPROVED BOTTLE:
HI 8004L pH 4.01 Buffer Solution, 460 mL HI 8006L pH 6.86 Buffer Solution, 460 mL HI 8007L pH 7.01 Buffer Solution, 460 mL HI 8009L pH 9.18 Buffer Solution, 460 mL HI 8010L pH 10.01 Buffer Solution, 460 mL
ELECTRODE STORAGE SOLUTIONS:
HI 70300M Storage Solution, 230 mL HI 70300L Storage Solution, 460 mL
ELECTRODE STORAGE SOLUTIONS IN FDA APPROVED BOTTLE:
HI 80300M Storage Solution, 230 mL HI 80300L Storage Solution, 460 mL
ELECTRODE CLEANING SOLUTIONS:
HI 7061M General Cleaning Sol., 230 mL HI 7061L General Cleaning Sol., 460 mL HI 7073M Protein Cleaning Sol., 230 mL
41
HI 7073L Protein Cleaning Sol., 460 mL HI 7074M Inorganic Cleaning Sol., 230 mL HI 7074L Inorganic Cleaning Sol., 460 mL HI 7077M Oil & Fat Cleaning Sol., 230 mL HI 7077L Oil & Fat Cleaning Sol., 460 mL
ELECTRODE CLEANING SOLUTIONS IN FDA APPROVED BOTTLE:
HI 8061M General Cleaning Solution, 230 mL HI 8061L General Cleaning Solution, 460 mL HI 8073M Protein Cleaning Solution, 230 mL HI 8073L Protein Cleaning Solution, 230 mL HI 8077M Oil & Fat Cleaning Solution, 230 mL HI 8077L Oil & Fat Cleaning Solution, 460 mL
REFILLING ELECTROLYTE SOLUTIONS:
HI 7071 3.5M KCl + AgCl Electrolyte,
4x50 mL, for single junction elec­trodes
HI 7072 1M KNO3 Electrolyte, 4x50 mL HI 7082 3.5M KCl Electrolyte, 4x50 mL,
for double junction electrodes
REFILLING ELECTROLYTE SOLUTIONS IN FDA APPROVED BOTTLE:
HI 8071 3.5M KCl + AgCl Electrolyte,
4x50 mL, for single junction elec­trodes
HI 8072 1M KNO3 Electrolyte, 4x50 mL HI 8082 3.5M KCl Electrolyte, 4x50 mL,
for double junction electrodes
ORP PRETREATMENT SOLUTIONS:
HI 7091M Reducing Pretreatment Solution,
230 mL
HI 7091L Reducing Pretreatment Solution,
460 mL
HI 7092M Oxidizing Pretreatment Solution,
230 mL
HI 7092L Oxidizing Pretreatment Solution,
460 mL
42
pH & ORP ELECTRODES:
B = BNC CONNECTION PLUG + 1 m (3.3’) CABLE
S = SCREW-TYPE CONNECTOR
HI 1043B / HI1040S
Glass-body, double junction, refillable, combi­nation pH electrode. Use: strong acid/alkali.
12 mm
0.5"
HI 1043
"S" VERSION
9.5mm DIA
120 mm
4.7"
0.37"
HI 1053B / HI1050S
Glass-body, triple ceramic, conic shape, refill­able, combination pH electrode. Use: emulsions.
12 mm
0.5"
HI 1053
"S" VERSION
120 mm
4.7"
HI 1083B
Glass-body, micro, Viscolene, nonrefillable, combination pH electrode. Use: biotechnology, micro titration.
HI 1083
12 mm
0.5"
5 mm
0.2"
120 mm
3 mm
0.12"
"
3.0 mm DIA
0.12"
HI 1131B / HI1111S
Glass-body, single junction, refillable, combi­nation pH electrode. Use: general purpose.
12 mm
0.5"
HI 1131
"S" VERSION
43
9.5mm DIA
120 mm
4.7"
0.37"
HI1330B / HI 1310S
Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: laboratory.
HI 1330
"S" VERSION
120 mm
4.7"
5mm DIA
0.2"
5mm
0.2"
HI1331B / HI 1311S
Glass-body, semimicro, single junction, refill­able, combination pH electrode. Use: flasks.
HI 1331
"S" VERSION
8 mm
0.3"
7.5mm DIA
210 mm
8.25"
0.29"
HI1230B / HI 1210S
Plastic-body (Ultem®), double junction, gel-filled, combination pH electrode. Use: general purpose.
12 mm
0.5"
HI 1230
120 mm
"
"
4.7"
HI2031B / HI 2020S
Glass-body, semimicro, conic, refillable, combina­tion pH electrode. Use: semisolid products.
6 mm
0.25"
HI 2031
75 mm
"
"
2.95"
HI1332B / HI 1312S
Plastic-body (Ultem®), double junction, refillable, combination pH electrode. Use: general purpose.
12 mm
0.5"
HI 1332
44
120 mm
4.7"
"S" VERSION
Ultem® is registered Trademark of "General Electrics Co."
FC100B
Plastic-body (Kynar®), double junction, refill­able, combination pH electrode. Use: general purpose for food industry.
12 mm
0.5"
FC 100
120 mm
4.7"
FC200B / FC200S Plastic-body (Kynar®), single junction, conic, Viscolene, refillable, combination pH electrode. Use: meat & cheese.
6 mm
0.25"
FC 200
75 mm
2.95"
FC210B
Glass-body, double junction, conic, Viscolene, combination pH electrode. Use: milk, yogurt.
12 mm
0.5"
FC 210
120 mm
4.7"
FC220B
Glass-body, single junction, refillable, combi­nation pH electrode. Use: food & wine process- ing.
9.5mm DIA
FC 220
120 mm
4.7"
Kynar® is registered Trademark of "Pennwalt Corp."
45
0.37"
12 mm
0.5"
FC 911B
Plastic-body (Kynar®), double junction, refill­able with built-in amplifier, combination pH electrode. Use: very high humidity.
12 mm
0.5"
FC 911
110 mm
4.3"
HI 1413B / HI1410S
Glass-body, single junction, flat tip, Viscolene, combination pH electrode. Use: surface mea- surement.
12 mm
0.5"
HI 1413
110 mm
4.3"
Half-cell electrodes:
HI 2110B
Glass-body, single half-cell pH electrode. Use: general purpose.
9.5mm DIA
0.37"
12 mm
0.5"
HI 2110
120 mm
"
HI 5311
Glass-body, Ag/AgCl reference half-cell elec­trode, double junction, refillable with 4mm plug with 1m (3.3') cable. Use: general purpose with wide temperature range.
12 mm
0.5"
HI 5311
120 mm
4.7"
Kynar® is registered Trademark of "Pennwalt Corp."
46
HI 5412 Glass-body, single Calomel reference half­cell electrode, refillable with 4mm plug with 1m
(3.3') cable. Use: general purpose with constant temperature range.
12 mm
0.5"
HI 5412
120 mm
4.7"
ORP electrodes:
HI3131B / HI 3111S
Glass-body, refillable, combination platinum ORP electrode. Use: titration.
12 mm
0.5"
HI 3131
150 mm
"S" VERSION
5.9"
HI3230B / HI 3210S
Plastic-body (Ultem®), gel-filled, combination platinum ORP electrode. Use: general purpose.
12 mm
0.5"
HI 3230
120 mm
"S" VERSION
4.7"
HI4430B / HI 4410S
Plastic-body (Ultem®), gel-filled, combination gold ORP electrode. Use: general purpose.
12 mm
0.5"
HI 4430
120 mm
"S" VERSION
Ultem® is registered Trademark of "General Electrics Co."
4.7"
47
EXTENSION CABLES FOR SCREW-TYPE ELECTRODES ONLY (SCREW TO BNC CONNECTOR):
HI 7855 SERIES CABLE CONNECTORS
CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC
CONNECT TO SCREW TYPE
ELECTRODES
CONNECT TO THE
BNC SOCKET
OF THE METER
HI7855/1 Extension cable 1m (3.3') long
HI7855/3 Extension cable 3m (9.9') long
HI7855/5 Extension cable 5m (16.5') long
HI7855/10 Extension cable 10m (33') long
HI7855/15 Extension cable 15m (49.5') long
POWER UNITS:
HI 710005 Voltage adapter from 110 VAC
to 12 VDC
HI 710006 Voltage adapter from 220 VAC
to 12 VDC
48
OTHER ACCESSORIES:
ChecktempC Pocket-size thermometer with
penetration probe and 0.1°C resolution (range -50.0 to
150.0°C)
HI 76405 Electrode holder
HI 7669/2W Temperature probe with 1 m
(3.3') screened cable
HI 8427 pH and ORP electrode simula-
tor with 1 m (3.3') coaxial cable ending in female BNC connec­tors (HI 7858/1)
HI 92000 Windows® compatible applica-
tion software
HI 920010 PC connection cable
HI 931001 pH and ORP electrode simula-
tor with display with 1 m (3.3') coaxial cable ending in female BNC connectors (HI 7858/1)
Windows® is registered Trademark of "Microsoft Co."
49
ELECTRODE APPLICATION
REFERENCE GUIDE
Application Electrodes
1. Aquarium HI 1332B, HI1911B, HI 1312S
2. Bath-water HI 1910B, HI 1130B, HI 1110S
3. Beer HI 1131B, HI 1111S
4. Bread HI2031B, FC 200B, HI 2020S, FC 200S
5. Cheese FC 200B, FC 200S
6. Dairy products FC 911B, FC 100B
7. Dirty water HI 1910B, HI 1912B
8.Emulsions HI1053B, HI 1050S
9. Environment HI1230B, HI 1210S
10.Flasks HI1331B, HI1310S
11. Food industry general use FC 911B, FC 100B
12. Fruit FC 200B, FC220B, FC 200S
13. Fruit juices, organic FC 210B
14. Galvanizing waste solution HI 1130B, HI 1912B, HI 1110S
15. Heavy-duty applications HI 1135B, HI 1115S
16. High purity water HI1053B, HI 1050S
17. Horticulture HI 1053B, FC 200B, HI1050S, FC 200S
18. In-line applications HI1134B, HI 1135B, HI2114B, HI 2910B,
19. Laboratory general use HI 1131B, HI 1230B, HI 1332B, HI1330B
20. Leather HI1413B, HI 1410S
21. Lemon juice FC 100B
22. M eat FC200B, HI 2031B, FC 200S, HI2020S
23. Micro plate sampling of less than HI 1083B, HI 1080S 100 mL
24. Milk and Yogurt FC 210B
25.Paints HI1053B, HI 1050S
26. Paper HI1413B, HI 1410S
27. Photographic chemicals HI 1230B, HI 1210S
28. Quality control HI 1332B, HI 1312S
29.Sausages FC200B, HI 2031B, FC 200S, HI 2020S
30. Semi-solid products HI 2031B, HI 2020S
31.Skin HI1413B, HI 1410S
32. Soil samples HI1230B, HI 1210S
33.Solvents HI1043B, HI 1040S
34. Strong acid HI 1043B, HI 1040S
35. Submersion application HI 1130B, HI 1110S
36. Surface measurements HI 1413B, HI 1410S
37. Swimming pool H 1130B, HI 2114B, HI 2910B
38. Titrations with constant temperature H I 1131B, HI 1111S range
39. Titrations with wide temperature HI 1131B, H I 1111S range
40. Very high humidity FC 911B, HI 1912B, HI 1911B
41. Vials and test tube H I 1330B, HI 1310S
42. Wine processing FC 220B
HI 1114S, HI1115S
HI1111S, H I 1210S, HI 1312S, HI1310S
B = BNC connection plug S = Screw-type connection plug
50
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workmanship
and materials when used for their intended purpose and maintained according to the in­structions.
The electrodes and probes are warranted for a period of six months.
Damages due to accidents, misuse, tampering or lack of prescribed maintenance are not cov­ered. This warranty is limited to repair or re­placement free of charge.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the war­ranty, you will be notified of the charge for repair or replacement. If the instrument is to be re­turned to Hanna Instruments, obtain a Return Goods Authorization No. (RGA) from the Cus­tomer Service Department first and than send it with shipment cost prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
51
CE DECLARATION OF CONFORMITYCE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITYCE DECLARATION OF CONFORMITY
DECLARATION OF CONFORMITY
We
Hanna Instruments Srl V.le delle industrie 12 35010 Ronchi di Villafranca (PD) ITALY
herewith certify that the bench pH meters
HI 931402 HI 931410
have been tested and found to be in compliance with the following regulations:
IEC 801-2 Electrostatic Discharge IEC 801-3 RF Radiated IEC 801-4 Fast Transient EN 55022 Radiated, Class B
Date of Issue: 01-02-1996
D.Volpato - Engineering Manager
On behalf of
Hanna Instruments S.r.l.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences.
The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times.
During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24VAC or 60VDC.
To avoid damages or burns, do not perform any measurement in microwave ovens.
53
http://www.hannainst.com
PRINTED IN
ITALY
MANPHBN3R2
03/96
Loading...