This Instrument is in Compliance with the CE Directives
www.hannainst.com
Dear Customer,
Thank you for choosing a Hanna Instruments
Product.
Please read this instruction manual carefully
before using the instrument.
This manual will provide you with all the
necessary information for the correct use of
the instrument, as well as a precise idea of its
versatility in a wide range of applications.
These instruments are in compliance with CSA,
UL and (EN 50081-1 and EN 50082-1)
directives.
Remove the instrument from the packing material and examine it to make sure that no
damage has occurred during shipping. If there
is any damage, notify your Dealer.
Note: Save all packing material until you are
sure that the instrument functions correctly. All defective items must be returned in the original packing with the
supplied accessories.
GENERAL DESCRIPTION
Hanna Instruments HI 931402 and HI 931410
are bench pH meters for pH, ORP (Oxidation
Reduction Potential), ISE (Ion Selective Electrode) and temperature measurements.
Their built-in microprocessor ensures accurate and user-friendly operation.
pH measurements are manually or automatically compensated for the temperature effect
(MTC/ATC).
They feature: large liquid crystal display (LCD),
user-friendly keyboard, error codes to guide
the user in calibration and troubleshooting
and storage of calibration data even after the
unit is shut off.
HI 931402 will not allow calibration until the
electrode has stabilized to ensure accurate
pH measurements.
These models meet the most stringent demands of the laboratory: they both automatically recognize five standard calibration buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01).
In HI 931410 a recorder output of 0 to 2 volts
is available and the wide mV range and high
resolution allow for ORP or ion-selective measurement.
1
In HI 931402, through the RS 232C serial port
you can transfer sample measurements to a
PC.
HI 931402 comes supplied with HI 1131B glassbody combination pH electrode, HI 7669/2W
temperature probe (with 1 meter / 3.3' cable),
HI 76405 electrode holder, pH 4.01 and
pH 7.01 buffer solutions (30 ml each), HI 7071
reference electrolyte (30 ml), 12VDC adapter
(HI 710005 or HI 710006) and a dust cover.
HI 931410 comes supplied with HI 1131B glassbody combination pH electrode, HI 7669/2W
temperature probe (with 1 meter / 3.3' cable),
12VDC adapter (HI 710005 or HI 710006) and
a dust cover.
2
FUNCTIONAL DESCRIPTION LCD
Primary Display
HI 931402
MEM
CAL
NOT READY
CON WRONG
LOW BAT
BUF
2
pH
mV
°C1
pH
Secondary Display
HI 931410
°C
pH
Temperature DisplaypH/mV Display
3
FUNCTIONAL DESCRIPTION HI 931402
ins rum n
MEM
1
CAL
NOT READY
CFM WRONG
BUF
2
pH
mV
°C1
pH
CAL Start calibration
/ Select buffer1 value
CFM Confirm when ready
/ Select buffer2 value
CFM Confirm when ready
HI 931402
MICROPROCESSOR
pH METER
MR
CAL
COMM
RANGE
CFM
pH calibration procedure
s
°CMEM
°C
Function
MEM For storing display result
MR Recall previous stored data
COMM For RS232 configuration
RANGE Select pH or ORP measurement
OFF
ON
10
2345678
1. Liquid Crystal Display
2. MEM key to store a value in memory
3. MR key to recall the stored value
4. CAL key to enter and exit the calibration mode
5. CFM key to confirm calibration values
6. RANGE key to select the measurement range
7. COMM key to set the baud rate
↑↑
8.
↑°C and
↑↑
↓↓
↓°C keys to manually set tempera-
↓↓
ture and to select pH buffer values
9. ON to turn the meter on
10. OFF key to turn the meter off
9
DTR
TXD
GND
COMBINATION
pH
REFERENCE
ELECTRODE
ELECTRODE
TEMPERATURE
PROBE
RESET
1
23 4 567
RXD
1
5
ooooo
oooo
96
RS-232
ooooo
oooo
DSR
1. BNC Connector
2. Reference Electrode Connector (Ø 0.08")
3. Reference Electrode Connector (Ø 4 mm)
4. Temperature Probe Connector
5. Reset Button
6. RS 232C Connector
7. DC Power Socket (for HI 710005 or
HI 710006)
4
SPECIFICATIONS HI 931402
RangepH0.00 to 14.00
ISE±399.9 mV
ORP±1999 mV
°C0.0 to 100.0
Resolution pH0.01
ISE0.1 mV
ORP1 mV
°C0.1
AccuracypH±0.01
(@20°C/68°F) ISE±0.2 mV
ORP±1mV
°C±0.5
Typical EMC pH±0.02
Deviation ISE±0.5 mV
ORP±1 mV
°C±2
pH CalibrationAutomatic 1 or 2 points with
5 memorized buffers
(pH 4.01, 6.86, 7.01, 9.18, 10.01)
TemperatureAutomatic or manual
Compensationfrom 0 to 100°C (32 to 212°F)
1. ON and OFF keys to turn the instrument on
and off. (Warning: Push OFF before unplugging instrument. If unplugged before turned
off, the meter's memories will be erased)
2. CAL key to enter the calibration mode
3. CON key to confirm the calibration data
4.°C and °C keys to set temperature during
manual compensation adjustment
5. pH/mV key to select the mV (mV, ORP or ISE
measurements) or the pH range
6. Liquid Crystal Displays
TEMPERATURE
DC POWER
12V to 20V
COMBINED
pH or ORP
ELECTRODES
REFERENCE
ELECTRODE
RESET
RECORDER
0-2 VOLT
PROBE
+
6
1
23
4
5
6
1. BNC Connector
2. Reference Electrode Connector
3. RESET button
4. Recorder Output Connector
5. Temperature Probe Connector
6. DC Power Socket (for HI 710005 or HI 710006)
6
SPECIFICATIONS HI 931410
RangepH
ISE
ORP
°C
Resolution pH
ISE
ORP
°C
AccuracypH
(@ 20°C / 68°F) ISE
ORP
°C
Typical EMC pH
DeviationISE/ORP
°C
RecorderpH
Outputm V
Max. Current
Offset
Slope
Calibration
Temperature
Compensation
Electrode
Temp. Probe
Input Impedance
Power
Environment
Dimensions
Shipping Weight
0.00 to 14.00
0.0 to ±399.9mV
±1999mV
-10.0° to 125.0
0.01
0.1mV
0.1/1mV
0.1
±0.01
±0.2 mV
±1 mV
±0.5
±0.01
±0.2mV
±0.7
0 to 2 V
0 to 2 V
20mA
±1 pH
from 85% to 105%
Automatic 1 or 2 points with
3 memorized buffers
(pH 4.01, 7.01, 10.01)
Manual from 0 to 100°C (32 to 122°F)
Automatic from -10 to 125°C (14 to 257°F)
HI 1131B glass body combination
refillable pH electrode (included)
HI 7669/2W (included)
12
10
Ω
12VDC (CSA, UL & CE approved)
0 to 50°C (32 to 122°F); 95% RH
230x170x70 mm (9.1x6.7x2.7")
1.5 Kg (3.3 lb.)
7
OPERATIONAL GUIDE
ELECTRODE AND PROBE CONNECTIONS
For combination pH or ORP electrodes (with
internal reference) plug the electrode's BNC
to the socket provided (see pages 4 and 6, #1).
For electrode with separate reference, connect the measuring electrode's BNC to the
BNC socket (see pages 4 and 6, #1) and the
reference electrode's plug to the socket provided (see page 4, #2 or #3 for HI 931402, or
page 6, #2 for HI 931410).
For temperature measurements and automatic temperature compensation connect the
temperature probe to the °C socket (see
page 4, #4 for HI 931402, or page 6, #5 for
HI 931410).
POWER CONNECTION
Plug the 12VDC adapter (HI 710005 or
HI 710006) into the DC socket (see page 4,
#7 for HI 931402, or page 6, #6 for HI 931410).
Plug the adapter to the mains.
Note: Make sure the main line is protected
by a fuse.
HI931402 uses an EEPROM and HI 931410
uses a rechargeable battery to retain the pH
calibration, mV calibration, temperature calibration as well as the serial communication
setting. The instrument will store the respective data after a calibration or serial communication setting, even when it is not plugged-in.
In HI 931410 the battery lasts approximately
one month when fully charged, and it is automatically charged any time the meter is
plugged-in (either turned on or off).
8
pH MEASUREMENTS
To take a pH measurement
simply submerge the tip
(4 cm/1½") of the electrode
(1½
")
and the temperature probe into
the sample to be tested.
Turn the instrument on by pressing the ON
key.
ON
ON
The meter automatically defaults to the pH
measurement mode.
Allow 1 or 2 minutes for the electrode to stabilize.
In HI 931402 pH is displayed on the primary
display and temperature on the secondary
one, whereas in HI 931410 pH is displayed
on the left LCD.
pH
°C
°C
pH
4 cm
In order to take accurate pH measurements,
make sure that the instrument has been calibrated for pH before use (see page 15).
In HI 931402, if the calibration data is lost or the meter
is not calibrated, the "CAL"
indicator is displayed to
CAL
prompt the user to perform a
calibration. If the user chooses
not to do it and presses the
CAL key to quit from the calibration mode, the offset and
CAL
slope of the instrument are
set to 0 mV and 100% of the
theoretical value respectively.
9
pH
°C
If measurements are taken in different samples
successively, it is recommended that the
electrode and the temperature probe are rinsed
thoroughly for better conditioning, and then
blot dry between each measurements. For
this rinsing process, it is recommended to
use a liberal amount of the sample to be
measured.
Using refillable electrodes
Unscrew the small refill cap
to open the refill hole during
measurements. After
measurement replace the refill
hole cap.
The pH reading is directly affected by
temperature. In order for the meter to measure
the pH accurately, temperature must be taken
in consideration.
A perfect equilibrium between the pH
electrode and the sample is reached in
approximately 15 minutes.
To use the meter's Automatic
Temperature Compensation
feature, submerge the
HI7669/2W temperature
probe into the sample as
close to the electrode as
possible and wait for 1-2
minutes.
MANUAL TEMPERATURE COMPENSATION
If you know the temperature of the sample to
be tested you can manually compensate for it.
If manual temperature compensation is
desired the temperature probe must be
disconnected from the instrument.
10
The display will then show
the default temperature (25°C
in HI 931402 or 20°C in
HI 931410) or the last recorded
temperature reading (HI 931402
only). "°C" is blinking.
Note the temperature of the
sample using a ChecktempC
or a glass thermometer.
°C
ºC
°C
°C
ºC
Use the UP or DOWN arrow
keys to manually adjust the
display reading to the value of
the reference thermometer.
°C
If the temperature probe is connected to the
meter, the secondary or the temperature LCD
figure display the measured temperature reading and the "°C" indicator does not blink.
Pressing the ↑°C or ↓°C keys will not alter
the temperature reading.
If measurements are taken in different samples
successively, it is recommended that the electrode and the temperature probe are rinsed
thoroughly with deionized water or distilled
water (if not available, tap water may also be
used) and then blot dry between each measurements.
ORP MEASUREMENTS
HI 931402 and HI 931410 have the capability to
take ORP measurements, using an optional
ORP electrode.
Hanna Instruments offers a variety of ORP
electrodes for this purpose (see page 47). Contact your Dealer for more information.
Just as pH measurements allow the acidity or
the alkalinity of a substance to be expressed in
numbers, oxidation-reduction potential (redox)
measurements provide the quantification of the
11
°C
oxidizing or reducing power of the sample tested.
To correctly perform an ORP measurement, the
surface of the ORP electrode must be clean and
smooth.
When not in use, the tip of the electrode should
be kept moist and safe from any mechanical
stress which might cause damage to the glass/
platinum junction.
To get into the "mV" mode
(ORP or ISE measurements)
turn the instrument ON and
pHpH
mVmV
RANGE
press the RANGE (HI 931402)
or the pH/mV (HI 931410)
key.
To measure the mV of a
solution simply submerge the
ORP or ISE electrode tip
(4 cm/1½") into the sample
to be tested. Allow a few
(1½
")
minutes for the readings to
stabilize.
The display will indicate the
absolute mV value (positive
or negative).
mV
4 cm
mV
ºC
When values are less than ±400 mV, tenths of
mV are displayed; for values greater than ±400
mV, the decimal digit disappears. Note that the
change in scale (from 01. mV to 1 mV as
resolution) is automatic.
"---" for HI 931402 or "Err2" for
HI 931410 appear if the value
exceeds ±1999 mV (out of
range).
Note: ORP measurements are taken without
temperature compensation.
12
mV
°C
TEMPERATURE MEASUREMENTS
Taking a temperature measurement is very
easy.
Connect the temperature probe to the
instrument and turn the instrument ON.
Dip the liquid/general
purpose temperature probe
HI 7669/2W into the sample
and allow the reading on the
secondary display to stabilize
(1 or 2 minutes).
The temperature value will be displayed on
the secondary LCD for HI 931402 and on the
left LCD for HI 931410.
pH
°C
°C
pH
HOLD FUNCTION (for HI 931402 only)
To hold and memorize the
measured value, press the
MEM
MEM key.
The last reading will be frozen on the display until the
key is released.
At the same time, this reading is stored in
the memory.
During this operation, the MEM indicator is
lit.
To recall the stored memory value, press the
MR key button and the memorized reading
will be displayed with the MEM indicator lit.
MR
13
MEM
pH
°C
RESET BUTTON
The RESET button (see page 4, #5 for
HI 931402, or page 6, #3 for HI 931410) is
used when the instrument displays erroneous
messages due to strong electrical interference or when the instrument's power supply
was disconnected before the meter was
switched off.
It is necessary to press the reset button and
restart the entire operation.
Calibration points should remain memorized.
It is recommended to verify calibration before
proceeding.
RECORDER OUTPUT (for HI 931410 only)
HI 931410 is provided with a recorder output
from 0 to 2 volts. Connect the recorder output
(see page 6, #4) and select the pH or mV range
by pressing the pH/mV key. The output signal
is related to the values as follows:
pH output:
0.01 pH = 1.42857 mV
e.g.0.00 pH =0 mV
7.00 pH =1000 mV
14.00 pH =2000 mV
mV output:
0.1 mV =0.05 mV
e.g. -1999 mV =0 mV
0.0 mV =1000 mV
+1999 mV =2000 mV
The output impedance is 100 ohms. The maximum output short-circuit current is 20 mA.
AFTER MEASUREMENTS
Press the OFF key to switch the instrument off.
OFF
OFF
14
pH CALIBRATION
It is recommended to calibrate the instrument
frequently, especially if high accuracy is required.
The instrument should be re-calibrated for
pH:
• Whenever the pH electrode or temperature
probe is replaced.
• At least once a month.
• After testing aggressive chemicals.
• After cleaning procedure and changing the
reference electrolyte.
• If greatest accuracy is required.
PREPARATION
Pour small quantities of the buffer solutions
into clean beakers. If possible use plastic
beakers to minimize any EMC interferences
HI 7004
HI 7004
HI 7007
To get accurate readings, use pH 7.01 and
pH 4.01 if you are going to measure acid
samples or pH 7.01 and pH 10.01 for alkaline
measurements.
For accurate calibration use two beakers for
each buffer solution, the first one for rinsing
the electrode, the second one for calibration.
In this way contamination of the buffers is
minimized.
RINSE
HI 7007
CALIBRATION
HI 7007
15
PROCEDURE FOR HI 931402
Calibration of HI 931402 has a choice of 5
memorized standard buffers: 4.01, 6.86, 7.01,
9.18 and 10.01 pH.
• Make sure that the meter is in pH mode.
• Remove the protective cap from the electrode and rinse it with some of the buffer
calibration solution you are going to use first.
ONE POINT CALIBRATION
One point calibration at pH 6.86 (NBS):
• Immerse the pH electrode
and the temperature probe
into pH 6.86 buffer solution
and shake briefly.
HI 7006
Note: the electrode should
be submerged approximately 4 cm (1½") into the
solution. The temperature
(1½
")
probe should be located as
close as possible to the pH electrode.
• Press the CAL key. The "CAL" and "
BUF
indicators will be displayed. The secondary LCD will display "4.01".
4 cm
1
"
CAL
CAL
1
BUF
pH
• Press the °C key. The secondary display
will change to “6.86”.
°C
CAL
1
BUF
pH
• When the electrode is submerged in the
buffer solution, the meter will notify the
user if the reading is not stable.
If the readings fluctuate the
LCD will blink "NOT READY".
16
NOT READY
Only when the reading is
stable, "READY" and "CON"
READY
CON
will blink.
• Press the CFM key to confirm the calibration. If the
reading is not close to the
selected buffer, "WRONG
BUF
1
"
CFM
and "WRONG " will blink
alternatively. If the reading
is close to the selected
WRONG
WRONG
1
buffer, the meter stores the
BUF
reading (and adjusts the offset point). The buffer value is
then displayed on the primary
LCD and the secondary LCD
CAL
1
BUF
will display "4.01".
• Press the CAL key and the
CAL
calibration process is ended
with only the offset of the
meter calibrated.
For best accuracy however, it is recommended
that a two point calibration is performed (see
page 19).
One point calibration at pH 7.01 (STANDARD):
• Immerse the pH electrode
and the temperature probe
into pH 7.01 buffer solution
HI 7007
and shake briefly.
Note: the electrode should be
submerged approximately
4 cm (1½") into the solution.
The temperature probe should
be located as close as pos-
(1½
")
sible to the pH electrode.
• Press the CAL key. The “CAL” and “
BUF
indicators will be displayed. The secondary LCD will display “4.01”.
CAL
17
CAL
1
BUF
pH
pH
4 cm
1
”
• Press the °C key twice. The secondary
display will change to “7.01”.
°C
CAL
1
BUF
pH
• When the electrode is submerged in the
buffer solution the meter will notify the
user the reading is not stable.
If the readings fluctuate the
LCD will blink “NOT
NOT READY
READY”.
Only when the reading is
stable, “READY” and “CON”
READY
CON
will blink.
• Press the CFM key to confirm the calibration. If the
reading is not close to the
selected buffer, “WRONG
BUF
1
”
CFM
and “WRONG ” will blink alternatively. If the reading is
close to the selected buffer
WRONG
WRONG
1
BUF
the meter stores the reading (and adjust the offset
point). The buffer value is
than displayed on the primary LCD and the second-
CAL
1
BUF
ary LCD will display “4.01”.
• Press the CAL key and the
calibration process is ended
CAL
with only the offset of the
meter calibrated.
pH
pH
For best accuracy however, it is recommended
that a two point calibration is performed (see
page 19).
18
TWO POINT CALIBRATION
F
A two point calibration is recommended for
best accuracy. As a second point use pH 4.01
if you are going to measure acid samples (pH 7
or less), use pH 9.18 or pH 10.01 if you are going
to measure alkaline samples.
Two point calibration at pH 6.86 and pH 4.01
or pH 9.18 (NBS):
•Proceed as described in “One point calibration at pH 6.86” but do not quit calibration by
pressing the CAL key at the end.
•After the first calibration point
is confirmed, immerse the pH
electrode and the temperature
probe into the second buffer
HI 7004
solution (pH 4.01 or pH 9.18)
and shake briefly.
Note: the electrode should
be submerged approximately 4 cm (1½") into the
solution. The temperature
(1½
")
probe should be located as
close as possible to the pH electrode.
•Select the second buffer
value on the secondary
display by pressing the
°C
°C
ºC
or °C key.
•When the “READY” and
“CON” symbols blink on the
display the reading is stable
READY
CON
and the calibration can be
confirmed.
•Press the CFM key. If the
CFM
reading is not close to the
selected buffer solution,
“WRONG
BU
” and “WRONG ”
2
will blink alternatively. If
the reading is close to the
selected buffer, the slope and
the offset are calibrated.
WRONG
WRONG
BUF
2
4 cm
19
The values will be stored in memory and the
meter will return to the operating mode.
Two point calibration at pH 7.01 and pH 4.01
or pH 10.01 (STANDARD):
•Proceed as described in “One point calibration at pH 7.01” but do not quit calibration by
pressing the CAL key at the end.
•After the first calibration point
is confirmed, immerse the pH
electrode and the temperature
probe into the second buffer
solution (pH 4.01 or pH 10.01)
HI 7004
and shake briefly.
Note: the electrode should
be submerged approximately
4 cm (1½") into the solution.
The temperature probe should
be located as close as pos-
(1½
")
sible to the pH electrode.
•Select the second buffer value
on the secondary display by
°C
ºC
pressing the °C or °C key.
•When the “READY” and “CON”
symbols blink on the display
the reading is stable and the
READY
CON
calibration can be confirmed.
•Press the CFM key. If the
reading is not close to the
CFM
selected buffer solution,
“WRONG
BUF
” and “WRONG ”
2
will blink alternatively. If the
reading is close to the selected buffer, the slope and
the offset are calibrated.
WRONG
WRONG
BUF
2
The values will be stored in memory and the
meter will return to the operating mode.
Note: the meter will automatically skip the
buffer that was used for the first calibration to avoid erroneous calibration.
4 cm
20
Note: during calibration, the
secondary LCD displays
the selected buffer value.
By pressing the RANGE key the temperature value can be also displayed.
CAL
BUF
pH
1
pH
RANGE
CAL
BUF
pH
°C
1
This will allow you to check the buffer
temperature during calibration.
21
PROCEDURE FOR HI 931410
HI 931410 provides three technical pre-programmed calibration buffers: t1, t2, t3
i.e. pH 7.01, 4.01, 10.01 respectively.
When the ATC probe is not used, make sure the
temperature is set manually to reflect the temperature of the solution (see page 26).
• Switch the instrument on after
connecting the pH electrode
ON
and the temperature probe.
ONE POINT CALIBRATION
• Immerse the pH electrode and
the temperature probe into a
pH 7.01 buffer solution.
Shake briefly and wait 1-2 min-
HI 7007
utes for thermal equilibrium.
Note: the electrode should be
submerged approximately
4 cm (1½") into the solution.
The temperature probe should
be located as close as pos-
(1½
")
sible to the pH electrode.
• Press the CAL key. The symbol "t1" will
appear on the display.
CAL
4 cm
• Press the CON key. The buffer value will be
displayed on the right LCD.
CON
pH
Note: the "pH" symbol will be flashing through-
out the calibration procedure.
22
• Allow 1-2 minutes for the pH reading to stabi-
lize.
The pH reading will be temperature compensated.
E.g. if the buffer temperature is at 25°C, the
right display will show "pH 7.01". If the buffer
temperature is at 20°C, the display will show
"pH 7.03" (see page 26).
°C
pH
If "Err4" appears the calibration solution is wrong.
Repeat the process using
fresh buffers
Note: If you are not using the ATC probe,
refer to page 26 to manually correct for
temperature.
• Press the CON key again to ac-
cept the buffer value. The first
CON
point calibration is now complete.
• If the pH/mV key is pressed at
this point, the calibration process
is ended and only the offset of the
pHpH
mVmV
meter is calibrated. This is known
as one point calibration.
For greatest accuracy, however, it is recommended that a two-point calibration is performed.
TWO POINT CALIBRATION
• Proceed as described in “One point calibration”
but do not quit calibration by pressing the
pH/mV key at the end.
• After the first calibration point
is confirmed, immerse the pH
electrode and the temperature
probe into the second buffer
solution (pH 4.01 or pH 10.01)
HI 7004
and shake briefly.
23
Note: the electrode should be
submerged approximately
4 cm (1½") into the solution.
The temperature probe should
be located as close as pos-
(1½
")
sible to the pH electrode.
• Rinse the electrode and the temperature probe
before immersing them into pH 4.01/ pH10.01
calibration buffer, shake briefly and wait for
thermal equilibrium.
• The meter automatically rec-
ognizes the buffer value and
prompts for the second at "t2"
(pH 4.01, acid) or "t3"
(pH 10.01, alkaline).
• Press the CON key to confirm the pH value
and the display will indicate the temperature
compensated buffer value e.g. if the temperature is at 25°C, the display will show "pH 4.01"
or "10.01", depending on the solution used.
4 cm
CON
°C
pH
If the temperature is at 20°C, the display will
show respectively "pH 4.00" or "pH 10.06"
(see page 26).
• Press CON key again to confirm calibration.
The pH symbol will stop flashing and the unit
is now calibrated and ready for measurements.
CON
°C
pH
24
• If the reading is drifting, turn the meter off and
repeat the calibration procedure waiting longer
time for the reading to stabilize. However,
usually 1 or 2 minutes are enough to prevent
drifting of the reading.
Note: If you need to unplug the instrument,
press the OFF key thus switching the
instrument off before disconnecting the
power cord. In this way the instrument
will retain the calibration data. Otherwise, re-calibration might be necessary.
In the case of a power failure while the
instrument is operational, re-calibration
might be required.
25
pH VALUES AT VARIOUS TEMPERATURE
Temperature has an effect on pH. The calibration buffer solutions are effected by temperature
changes to a lesser degree than normal solutions.
During calibration the meter will automatically
calibrate to the pH value corresponding to the
measured or set temperature.
For manual temperature compensation please
refer to the following chart.
For instance, if the buffer temperature is at
25°C, the display will show pH 4.01 or 7.01 or
10.01.
If the buffer temperature is at 20°C, the display
will show pH 4.00/7.03/10.06.
If the buffer temperature is at 50°C, the display
will show pH 4.06/6.98/9.82.
26
TEMPERATURE AND mV CALIBRATION
All Hanna bench-top pH meters have been
factory calibrated for the temperature and mV
and is ready for measurements.
The temperature probes and ORP electrodes
are interchangeable and no temperature or
mV calibration is needed when the temperature probe or the ORP electrode is replaced.
If, for any reason, the temperature or the mV
measurements are out of accuracy, the temperature or the mV re-calibration should be
carried out.
In any case, we recommend that you have
your bench-top meters re-calibrated in temperature and mV once a year.
In HI 931402 it is possible to calibrate the
meter for temperature and mV (see page 29),
however it is recommendable that these calibrations are performed by technical personnel
only.
For an accurate recalibration, contact your
nearest Hanna Service Center.
27
INTERFACE WITH PC (HI931402 ONLY)
Data transmission from the instrument to the
PC is now much easier with the new HI 92000
Windows® compatible application software offered by Hanna Instruments.
User friendly, HI 92000 offers a variety of
features and has an on line help feature to
support you throughout all situations.
To install HI 92000 you need a 3.5" drive and
few minutes to follow the instructions conveniently printed on the disk's label.
In your copy of HI92000 you will also find a
set of configuration files for your Windows
Terminal® program to send commands directly from the PC keyboard (see appendix
for commands description and usage).
Contact your Hanna Dealer to request a copy.
To connect your HI 931402 to the PC use
HI 920010, available through your Hanna
Dealer. Make sure that your meter is switched
off and plug the connectors, one into the
meter RS232C connector, the other into the
serial port of your PC.
Note: Cables different from the HI 920010 may
use a different configuration. In such
case any communication between the
meter and the PC is not possible.
For HI931402 only, you can perform the temperature and mV calibration as follows.
TEMPERATURE CALIBRATION OF HI931402
Initial Preparation
• Prepare a vessel containing ice and water
and another one containing hot water (at a
temperature of at least 50°C). Place insulation material around the container to minimize temperature changes.
• Use a ChecktempC or
C
°
C
°
a calibrated thermometer with a resolution
of 0.1 °C as a refer-
0 °C
(32 °F)
50 ºC
(122 ºF)
ence thermometer.
Procedure
• Switch the instrument on while pressing
the CAL key. The "CAL" indicator will be lit.
The secondary LCD section will show 0.0°C.
ON
CAL
+
CAL
°C
°
• Immerse the temperature probe
C
in the vessel with the ice and
water as near to the ChecktempC
as possible.
0 °C
(32 °F)
• Wait for temperature reading to stabilize.
• Use the °C and °C keys to
set the reading on the sec-
°C
ºC
ondary LCD section to the
noted temperature of the ice
with the water.
• Press CFM key. The secondary LCD section will show 50.0°C.
CFM
CAL
°C
29
°
• Immerse the temperature probe
C
in the vessel with hot water.
• When the temperature reading
has stabilized, use the °C and
°C
keys to set the reading on
the secondary LCD to the
°C
50 °C
(122 °F)
ºC
noted temperature of the hot
water.
• Press CFM key.
The primary LCD
CFM
CAL
section will show
"390.0 mV".
• Complete your temperature calibra-
CAL
tion by pressing the CAL key if
you do not want to perform the mV
calibration mode.
mV CALIBRATION OF HI 931402
Initial Preparation
•Prepare HI 931001 or a simulator that can
provide ±390.0 mV with accuracy of ±0.1mV
and ±1000 mV with accuracy of ±1 mV.
Procedure
•Continue from the tempera-
ture calibration, the primary
CAL
LCD section will show
"390.0 mV".
•Select 390.0 mV of the simulator.
•Wait for 3 sec., press the CFM
CFM
key. The primary LCD will display
the next reading to be selected on
the simulator.
•Repeat previous two steps for -390.0 mV,
1000 mV and -1000 mV.
•At the end of the mV calibration, the meter
will return to the pH measurement mode.
mV
mV
INTERFACE WITH PC
If you are not using Hanna Instruments
HI 92000 application software, please find here
below some additional information to help
your connection to the PC.
30
Setting the Baud Rate and the Command
Prefix
The transmission speed (baud rate) of your
HI 931402 and of the external device must be
the same. To set the baud rate of the meter
press the COMM
key, the primary
COMM
LCD shows the current baud rate.
The following baud rate can
be selected through the
°C
°C
ºC
and °C keys: 150, 300, 600,
1200 (factory setting) and 2400.
Press the CFM key to confirm the setting
and the primary LCD shows the current com-mand prefix; 16 is the factory setting.
CFM
Note: the Command Prefix has not to be
changed using HI92000 Hanna Software.
Select a different command prefix (between 0 to 47 decimal)
°C
ºC
by pressing the °C and °C keys
.
Press the CFM key to confirm the
CFM
setting.
Sending Commands from PC
With terminal programs such as, for example,
Telix®, Windows Terminal®, it is possible to
remotely control your HI 931402. Use
HI 920010 cable to connect the meter to the
PC, start the terminal program and set the
communication options as follows: 8, N, 1,
no flow control.
Command Types
To send a command to the pH meter the
scheme is: <DLE> <command> <CR>
Windows Terminal® are registered Trademark of "Microsoft Co."
TELIX® is registered Trademark of "Deltacomm"
31
This line makes the computer send a Data
Link Escape character, the command expressed as a number or a 3-character sequence) and a CR character.
Note: Windows Terminal® and all the other
terminal programs that support the ANSI escape sequence, represent the DLE character
by the string '^P' and the CR character by the
string '^M'. E.g. the line '^PPHR^M' sets the
range to pH.
Commands not requiring an answer from
the pH meter:
PHR sets the range to pH
MVR sets the range to mV
CAL is equivalent to pressing the CAL key
CFM is equivalent to pressing the CFM key
UPC is equivalent to pressing the °C key
DWC is equivalent to pressing the °C key
MEM is equivalent to pressing the MEM key
MRR is equivalent to pressing the MR key
COM is equivalent to pressing the COMM key
OFF is equivalent to pressing the OFF key
Commands requiring an answer:
pH? Causes the meter to send the pH value. If
the reading is out of range "Err 1" is sent. If
the range is set to mV, "Err 6" is sent.
MV? Causes the meter to send the mV value. If
the reading is out of range "Err 2" is sent. If
the range is set to pH, "Err 6" is sent.
TM? Causes the meter to send the temperature
value. If the reading is out of range "Err 3" is
sent.
These commands may be sent with either
capital or small letters. Invalid commands will
be ignored. The characters sent by HI 931402
are always capital letters. When the meter
receives an unknown or corrupted commands,
it will send a character CAN (ASCII Code 24).
Windows Terminal® are registered Trademark of "Microsoft Co."
32
TROUBLESHOOTING FOR HI 931402
F
F
SymptomsProblemSolution
The meter does
not work with the
temperature probe
The meter fails to
calibrate or gives
faulty readings
The meter is slow
in responding or
gives faulty
readouts
The meter does
not accept the 2
buffer solution for
calibration
The reading drifts
Display shows:
Display shows:
Defective ATC
probe
Defective pH
electrode
The electrode is
not working or
the reference
junction is
clogged
Defective pH
nd
electrode
Defective pH
electrode
Out of range-pH
scale
pH
Out of range-°C
scale
pH
ºC
Replace the probe
Replace the electrode
Leave the electrode in
a storage solution after
cleaning the junction. If
problem persists,
replace the electrode
Follow the cleaning
procedure. If this
doesn't work replace
the electrode
Replace the electrode
a) Check the calibration
b) Assure the pH sample
is in the range 0 to 14
c) Check the level of
electrolyte and the
state of the electrode
itself
Make sure the °C is within
0-100° and temperature
probe is plugged
Display shows:
Display shows:
1
"WRONG
BU
and/or
"WRONG "
mV
"
Out of range-mV
scale
Erroneous
buffer solution
used for offset
cal.
Defective
Electrode not connected
Make sure the buffer
solution used is pH7 and
replace if necessary
Replace the electrode
electrode
Display shows:
"WRONG
and/or
BU
"
2
"WRONG "
Erroneous
buffer solution
used for slope
cal.
Defective
electrode
Use fresh buffer solution
Replace the electrode
33
TROUBLESHOOTING FOR HI 931410
Symptoms are:
The instrument
does not work with
the HI7669/2W
probe
The meter fails to
calibrate or gives
erroneous results
The meter is slow
in responding or
gives erroneous
readouts
The meter does
not accept the 2nd
buffer solution
The reading drifts
"Err1" on the
display
"Err3" on the
display
"Err2" on the
display
"Err4" on the
display
"Err5" on the
display
The Problem is:
A defective ATC
Probe
A defective pH
electrode
The electrode is
not working or the
diaphragm is
clogged
A defective pH
electrode or bad
buffer solution
Temperature not
stable
Out of range on
the pH scale
Out of range in °C
Out of range in mV
Erroneous buffer
solution used for
offset calibration
Defective
electrode
Erroneous buffer
solution used for
slope calibration
Defective
electrode
Possible Solutions:
Replace the probe.
Replace the electrode
and repeat the pH calibration.
Leave the electrode in
a storage solution after
cleaning the diaphragm.
If the problem is not
solved, replace the electrode and re-calibrate
the pH scale.
Try the cleaning procedure. If this does not
work, replace the electrode and re-calibrate
the pH. Use fresh
buffer solution.
Thermal equilibrium not
reached. Wait longer.
a) Verify the pH electrode is connected.
b) Make sure the pH
sample is included between 0 and 14 pH.
c) Check the level of
the electrode's filling
electrolyte and the state
of the electrode itself.
Verify the temperature
probe is connected and
make sure the °C is
within -10/125°C.
Verify the temperature
probe is connected and
make sure the mV is
within ±1999 range.
Make sure that the
buffer solution used is
of pH 7 and replace if
necessary.
Replace the electrode
and re-calibrate the pH
scale.
Be sure the correct pH
buffer solution is being
used for slope calibration (4.01, 9.18 or 10.01
pH).
Check the electrode
and replace if necessary.
34
ELECTRODE CONDITIONING
AND MAINTENANCE
Reference
Filling Hole
Reference
Filling Hole
Sensitive
Wire
Reference
Wire
Reference
Junction
Glass
Bulb
Reference
Wire
Reference
Junction
Platinum or
Gold tip
Plastic Body
pH Electrode
Reference
Plastic Body
ORP Electrode
Glass Body
pH Electrode
Wire
Glass Body
ORP Electrode
Reference
Wire
Sensitive
Wire
Reference
Junction
Glass
Bulb
Reference
Junction
Platinum or
Gold tip
PREPARATION
Remove the protective cap.
DO NOT BE ALARMED IF ANY SALT DE-
POSITS ARE PRESENT.
This is normal with electrodes and they will
disappear when rinsed with water.
During transport tiny bubbles of air may have
formed inside the glass bulb. The electrode
cannot function properly under these conditions. These bubbles can be removed by
35
"shaking down" the electrode as you would
do with a glass thermometer.
If the bulb and/or junction are dry, soak the
electrode in HI70300 or HI 80300 Storage Solution for at least one hour.
For refillable electrodes:
If the filling solution (electrolyte) is more than
1 cm (½") below the fill hole, add HI 7082 orHI 8082 3,5M KCl Electrolyte Solution for
double junction or HI 7071 or HI8071 3,5MKCl+AgCl Electrolyte Solution for single
junction electrodes.
For a faster response, unscrew the fill hole
screw during measurements.
For AmpHel electrodes:
If the electrode does not respond to pH
changes, the battery is run down and the
electrode should be replaced.
MEASUREMENT
Rinse the electrode tip with distilled water.
Immerse the tip (4 cm /1½") in the sample and
stir gently for at least 30 seconds.
For a faster response
and to avoid cross contamination of the samples, rinse the electrode tip with a few drops of the solution to be
tested, before taking measurements.
STORAGE
To minimize clogging and assuring a quick
response time, the glass bulb and the junction
should be kept moist and not allowed to dry out.
Replace the solution in the protective cap
with a few drops of HI 70300 or HI 80300
Storage Solution or, in its absence, Filling
Solution (HI 7071 or HI 8071 for single junc-
tion or HI 7082 or HI8082 for double junction
electrodes). Follow the Preparation Procedure above before taking measurements.
Note: NEVER STORE THE ELECTRODE IN
DISTILLED WATER OR DRY.
36
PERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable
used for connection to the meter must be
intact and there must be no points of broken
insulation on the cable or cracks on the
electrode stem or bulb. Connectors must be
perfectly clean and dry.
If any scratches or cracks are present, replace the electrode.
Rinse off any salt deposits with water.
For refillable electrodes:
Refill the reference chamber with fresh electrolyte (HI7071 or HI 8071 for single junction
or HI 7082 or HI 8082 for double junction electrodes). Allow the electrode to stand upright
for 1 hour.
Follow the Storage Procedure above.
CLEANING PROCEDURE
GeneralSoak in Hanna HI 7061 or HI8061
General Cleaning Solution for
approximately 1 hour.
Removal of films, dirt or deposits on the
membrane/junction:
- ProteinSoak in Hanna HI7073 or HI8073
Protein Cleaning Solution for
15 minutes.
- InorganicSoak in Hanna HI 7074 or HI 8074
Inorganic Cleaning Solution
for 15 minutes.
- Oil/grease Rinse with Hanna HI 7077 or
HI 8077 Oil and Fat Cleaning
Solution.
IMPORTANT: After performing any of the
cleaning procedures rinse the electrode thoroughly with distilled water, refill the reference
chamber with fresh electrolyte, (not necessary for GEL filled electrodes) and soak the
electrode in HI 70300 or HI 80300 StorageSolution for at least 1 hour before taking
measurements.
37
TROUBLESHOOTING
Evaluate your electrode performance based
on the following.
•Noise (Readings fluctuate up and down)
could be due to:
- Clogged/Dirty Junction: Refer to the
Cleaning Procedure above.
- Loss of shielding due to low electrolyte
level (in refillable electrodes only): refill
with fresh HI7071 or HI 8071 for single
junction or HI7082 or HI 8082 for double
junction electrodes.
•Dry Membrane/Junction: Soak in Storage Solution HI 70300 or HI 80300 for at
least 1 hour.
•Drifting: Soak the electrode tip in warm
Hanna Solution HI 7082 or HI 8082 for one
hour and rinse tip with distilled water.
Refill with fresh HI 7071 or HI8071 for single
junction electrodes and HI 7082 or HI8082
for double junction electrodes.
•Low Slope: Refer to the cleaning procedure above.
•No Slope: Check the electrode for cracks
in glass stem or bulb and replace the
electrode.
•Slow Response/Excessive Drift: Soak
the tip in Hanna Solution HI 7061 or HI8061
for 30 minutes, rinse thoroughly in distilled water and then follow the Cleaning
Procedure above.
38
Temperature-Resistance Correlation
20
0 +10+20 +30+40 +50+60+70+80+90
C
for HANNA pH sensitive glass
The resistance of glass electrodes partially
depends on the temperature. The lower the
temperature, the higher the resistance. It takes
longer time for the reading to stabilize if the
resistance is higher. In addition, the response
time will suffer to a greater degree at temperatures below 10°C.
Ω
9
2x10
-10
9
1x10
-10
8
-10
2x10
8
1x10
-10
7
2x10
-10
-10
7
1x10
-10
-20-
°
Since the resistance of the pH electrode is in the
range of 200 Mohm, the current across the
membrane is in the pico Ampere range. Large
currents can disturb the calibration of the electrode for many hours.
For these reasons high humidity environ-
ments, short circuits and static discharges
are detrimental for a stable pH reading.
The pH electrode's life also depends on the
temperature. If constantly used at high temperatures, the electrode life is drastically reduced.
39
Typical Electrode Life
Ambient Temperature 1- 3 years
90 °C Less than 4 months
120°C Less than 1 month
High concentrations of sodium ions interfere
with readings in alkaline solutions; the pH at
which the interference starts to be significant
depends upon the composition of the glass.
This interference is the alkaline error and causes
the pH to be underestimated. Hanna's glass
formulations have the indicated characteristics.
Alkaline Error
Sodium Ion Correction for the Glass
at 20-25°C
ConcentrationpH Error
0.1 Mol L-1 Na
1.0 Mol L-1 Na
+
13.00
13.50
14.00
+
12.50
13.00
13.50
14.00
0.10
0.14
0.20
0.10
0.18
0.29
0.40
40
ACCESSORIES
pH CALIBRATION SOLUTIONS:
HI77400PpH 4.01 & 7.01 Buffer Solution,
20 mL sachet, 5 pcs each
HI 7004MpH 4.01 Buffer Solution, 230 mL
HI 7004LpH 4.01 Buffer Solution, 460 mL
HI 7006MpH 6.86 Buffer Solution, 230 mL
HI 7006LpH 6.86 Buffer Solution, 460 mL
HI77700PpH 7.01 Buffer Sol. 10x20 mL
HI 7007MpH 7.01 Buffer Solution, 230 mL
HI 7007LpH 7.01 Buffer Solution, 460 mL
HI 7009MpH 9.18 Buffer Solution, 230 mL
HI 7009LpH 9.18 Buffer Solution, 460 mL
HI770710P pH 7.01 & 10.01 Buffer Solution,
20 mL sachet, 5 pcs each
HI 7010MpH 10.01 Buffer Solution, 230 mL
HI 7010LpH 10.01 Buffer Sol., 460 mL
pH CALIBRATION SOLUTIONS IN FDA
APPROVED BOTTLE:
HI 8004LpH 4.01 Buffer Solution, 460 mL
HI 8006LpH 6.86 Buffer Solution, 460 mL
HI 8007LpH 7.01 Buffer Solution, 460 mL
HI 8009LpH 9.18 Buffer Solution, 460 mL
HI 8010LpH 10.01 Buffer Solution, 460 mL
ELECTRODE STORAGE SOLUTIONS:
HI 70300MStorage Solution, 230 mL
HI 70300LStorage Solution, 460 mL
ELECTRODE STORAGE SOLUTIONS IN
FDA APPROVED BOTTLE:
HI 80300MStorage Solution, 230 mL
HI 80300LStorage Solution, 460 mL
ELECTRODE CLEANING SOLUTIONS:
HI 7061MGeneral Cleaning Sol., 230 mL
HI 7061LGeneral Cleaning Sol., 460 mL
HI 7073MProtein Cleaning Sol., 230 mL
41
HI 7073LProtein Cleaning Sol., 460 mL
HI 7074MInorganic Cleaning Sol., 230 mL
HI 7074LInorganic Cleaning Sol., 460 mL
HI 7077MOil & Fat Cleaning Sol., 230 mL
HI 7077LOil & Fat Cleaning Sol., 460 mL
ELECTRODE CLEANING SOLUTIONS IN
FDA APPROVED BOTTLE:
HI 8061MGeneral Cleaning Solution, 230 mL
HI 8061LGeneral Cleaning Solution, 460 mL
HI 8073MProtein Cleaning Solution, 230 mL
HI 8073LProtein Cleaning Solution, 230 mL
HI 8077MOil & Fat Cleaning Solution, 230 mL
HI 8077LOil & Fat Cleaning Solution, 460 mL
REFILLING ELECTROLYTE SOLUTIONS:
HI 70713.5M KCl + AgCl Electrolyte,
4x50 mL, for single junction electrodes
HI 70721M KNO3 Electrolyte, 4x50 mL
HI 70823.5M KCl Electrolyte, 4x50 mL,
for double junction electrodes
REFILLING ELECTROLYTE SOLUTIONS IN
FDA APPROVED BOTTLE:
HI 80713.5M KCl + AgCl Electrolyte,
4x50 mL, for single junction electrodes
HI 80721M KNO3 Electrolyte, 4x50 mL
HI 80823.5M KCl Electrolyte, 4x50 mL,
Glass-body, single half-cell pH electrode.
Use: general purpose.
9.5mm DIA
0.37"
12 mm
0.5"
HI 2110
120 mm
"
HI 5311
Glass-body, Ag/AgCl reference half-cell electrode, double junction, refillable with 4mm plug
with 1m (3.3') cable. Use: general purpose with
wide temperature range.
12 mm
0.5"
HI 5311
120 mm
4.7"
Kynar® is registered Trademark of "Pennwalt Corp."
46
HI 5412
Glass-body, single Calomel reference halfcell electrode, refillable with 4mm plug with 1m
(3.3') cable. Use: general purpose with constant
temperature range.
Plastic-body (Ultem®), gel-filled, combination
platinum ORP electrode. Use: general purpose.
12 mm
0.5"
HI 3230
120 mm
"S"VERSION
4.7"
HI4430B / HI 4410S
Plastic-body (Ultem®), gel-filled, combination
gold ORP electrode. Use: general purpose.
12 mm
0.5"
HI 4430
120 mm
"S"VERSION
Ultem® is registered Trademark of "General Electrics Co."
4.7"
47
EXTENSION CABLES FOR SCREW-TYPE
ELECTRODES ONLY (SCREW TO BNC
CONNECTOR):
HI 7855 SERIES CABLE CONNECTORS
CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC
CONNECT TO
SCREW TYPE
ELECTRODES
CONNECT TO THE
BNC SOCKET
OF THE METER
HI7855/1Extension cable 1m (3.3') long
HI7855/3Extension cable 3m (9.9') long
HI7855/5Extension cable 5m (16.5') long
HI7855/10Extension cable 10m (33') long
HI7855/15Extension cable 15m (49.5') long
POWER UNITS:
HI 710005Voltage adapter from 110 VAC
to 12 VDC
HI 710006Voltage adapter from 220 VAC
to 12 VDC
48
OTHER ACCESSORIES:
ChecktempCPocket-size thermometer with
penetration probe and 0.1°C
resolution (range -50.0 to
150.0°C)
HI 76405Electrode holder
HI 7669/2WTemperature probe with 1 m
(3.3') screened cable
HI 8427pH and ORP electrode simula-
tor with 1 m (3.3') coaxial cable
ending in female BNC connectors (HI 7858/1)
HI 92000Windows® compatible applica-
tion software
HI 920010PC connection cable
HI 931001pH and ORP electrode simula-
tor with display with 1 m (3.3')
coaxial cable ending in female
BNC connectors (HI 7858/1)
Windows® is registered Trademark of "Microsoft Co."
49
ELECTRODE APPLICATION
REFERENCE GUIDE
ApplicationElectrodes
1. AquariumHI 1332B, HI1911B, HI 1312S
2. Bath-waterHI 1910B, HI 1130B, HI 1110S
3. BeerHI 1131B, HI 1111S
4. BreadHI2031B, FC 200B, HI 2020S, FC 200S
5. CheeseFC 200B, FC 200S
6. Dairy productsFC 911B, FC 100B
7. Dirty waterHI 1910B, HI 1912B
8.EmulsionsHI1053B, HI 1050S
9. EnvironmentHI1230B, HI 1210S
10.FlasksHI1331B, HI1310S
11. Food industry general useFC 911B, FC 100B
12. FruitFC 200B, FC220B, FC 200S
13. Fruit juices, organicFC 210B
14. Galvanizing waste solutionHI 1130B, HI 1912B, HI 1110S
15. Heavy-duty applicationsHI 1135B, HI 1115S
16. High purity waterHI1053B, HI 1050S
17. HorticultureHI 1053B, FC 200B, HI1050S, FC 200S
18. In-line applicationsHI1134B, HI 1135B, HI2114B, HI 2910B,
19. Laboratory general useHI 1131B, HI 1230B, HI 1332B, HI1330B
20. LeatherHI1413B, HI 1410S
21. Lemon juiceFC 100B
22. M eatFC200B, HI 2031B, FC 200S, HI2020S
23. Micro plate sampling of less than HI 1083B, HI 1080S
100 mL
24. Milk and YogurtFC 210B
25.PaintsHI1053B, HI 1050S
26. PaperHI1413B, HI 1410S
27. Photographic chemicalsHI 1230B, HI 1210S
28. Quality controlHI 1332B, HI 1312S
29.SausagesFC200B, HI 2031B, FC 200S, HI 2020S
30. Semi-solid productsHI 2031B, HI 2020S
31.SkinHI1413B, HI 1410S
32. Soil samplesHI1230B, HI 1210S
33.SolventsHI1043B, HI 1040S
34. Strong acidHI 1043B, HI 1040S
35. Submersion applicationHI 1130B, HI 1110S
36. Surface measurementsHI 1413B, HI 1410S
37. Swimming poolH 1130B, HI 2114B, HI 2910B
38. Titrations with constant temperature H I 1131B, HI 1111S
range
39. Titrations with wide temperature HI 1131B, H I 1111S
range
40. Very high humidityFC 911B, HI 1912B, HI 1911B
41. Vials and test tubeH I 1330B, HI 1310S
42. Wine processingFC 220B
HI 1114S, HI1115S
HI1111S, H I 1210S, HI 1312S, HI1310S
B = BNC connection plug
S = Screw-type connection plug
50
WARRANTY
All Hanna Instruments meters are warranted
for two years against defects in workmanship
and materials when used for their intended
purpose and maintained according to the instructions.
The electrodes and probes are warrantedfor a period of six months.
Damages due to accidents, misuse, tampering
or lack of prescribed maintenance are not covered. This warranty is limited to repair or replacement free of charge.
If service is required, contact the dealer from
whom you purchased the instrument. If under
warranty, report the model number, date of
purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified of the charge for repair
or replacement. If the instrument is to be returned to Hanna Instruments, obtain a Return
Goods Authorization No. (RGA) from the Customer Service Department first and than send it
with shipment cost prepaid. When shipping any
instrument, make sure it is properly packaged
for complete protection.
To validate your warranty, fill out and return the
enclosed warranty card within 14 days from the
date of purchase.
All rights are reserved. Reproduction in whole
or in part is prohibited without the written
consent of the copyright owner.
Hanna Instruments reserves the right to modify
the design, construction and appearance of its
products without advance notice.
51
CE DECLARATION OF CONFORMITYCE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITYCE DECLARATION OF CONFORMITY
DECLARATION OF CONFORMITY
We
Hanna Instruments Srl
V.le delle industrie 12
35010 Ronchi di Villafranca (PD)
ITALY
herewith certify that the bench pH meters
HI 931402 HI 931410
have been tested and found to be in compliance with the following regulations:
IEC 801-2Electrostatic Discharge
IEC 801-3RF Radiated
IEC 801-4Fast Transient
EN 55022Radiated, Class B
Date of Issue: 01-02-1996
D.Volpato - Engineering Manager
On behalf of
Hanna Instruments S.r.l.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment
in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences
to radio and TV equipments, requiring the operator to take all necessary steps to correct
interferences.
The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid
touching this glass bulb at all times.
During operation, ESD wrist straps should be worn to avoid possible damage to the
electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instrument's
EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement
surface exceed 24VAC or 60VDC.
To avoid damages or burns, do not perform any measurement in microwave ovens.
53
http://www.hannainst.com
PRINTED IN
ITALY
MANPHBN3R2
03/96
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.