Hanna Instruments HI 93114 User Manual

Page 1
Instruction Manual
HI 93114
Turbidity & Chlorine
Measurements
www.hannainst.com
Page 2
Dear Customer, Thank you for choosing a HANNA instruments
product. Please read this instruction manual carefully
before using the instrument. This manual will provide you with the necessary
information for the correct use of the instrument, as well as a precise idea of its versatility.
For any additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with the directives.
TABLE OF CONTENTS
Preliminary Examination ............................. 3
General Description.................................... 4
Principle of Operation ................................. 6
Functional Description ................................ 9
Specifications ........................................... 12
Operational Guide ..................................... 13
Calibration................................................. 19
Diagnostic mode ....................................... 27
PRELIMINARY EXAMINATION
®
Remove the instrument from the packing ma­terial and examine it to make sure that no damage has occurred during shipping. If there is any damage, immediately notify your dealer.
HI 93114 is supplied complete with:
Glass cuvet with cap
Batteries (4 x 1.5V AA)
Instruction manual
An optional starter kit HI 731327 is also avail­able, complete with:
2 glass cuvets with caps
Calibration standards HI 93102-0 (AMCO-
AEPA-1 @0 NTU* solution, 30 mL bottle) and HI 93102-20 (AMCO-AEPA-1 @20 NTU* solution, 30 mL bottle)
• HI 93703-50 cuvet cleaning solution
Tissue for wiping cuvets
Rugged carrying case
Note: Save all packing material until you are
sure that the instrument functions cor­rectly. Any defective item must be re­turned in its original packaging with the supplied accessories.
Logging ..................................................... 28
User-selectable Shutdown......................... 30
Battery Replacement ................................ 31
Diagnostic Codes...................................... 32
Accessories.............................................. 33
Warranty ................................................... 34
CE Declaration of Conformity .................... 35
2
* 1 NTU (Nephelometric Turbidity Unit) = 1 FTU
(Formazine Turbidity Unit)
3
Page 3
GENERAL DESCRIPTION
With the hand held HI 93114, three important parameters in water quality analysis are at your fingertips. Free and total chlorine measure­ments and US EPA compliant turbidity mea­surements are performed quickly and reliably with this lab-grade microprocessor-based field instrument.
HI 93114 is particularly useful for potable water measurements and municipalities.
With HI 93114 measurements can be per­formed using either concentration or relative absorbance of the sample depending on the requirements of the application.
HI 93114 can be calibrated using the factory preprogrammed calibration points or can be custom calibrated when in colorimetric mode.
This rugged instrument has been engineered to offer all the sophistication and reliability of benchtop laboratory instrumentation in the size of a handheld meter.
HI 93114 complies with GLP standards (Good Laboratory Practice), that is:
• When switched on, the LCD displays all segments (display check)
• Battery status is monitored during every measurement cycle warning the user if the batteries become weak. In addition, the meter will turn itself off before low voltage causes erroneous readings
• It utilizes a real time clock and recalls calibration data such as date, time and calibration values
The large dual-level display can show the measured parameter on the main part, while the lower level indicates current mode (e.g. “F CL” for free chlorine or “TR” for turbidity). Additional symbols show low battery condi­tion, logging mode, date, time, etc.
A pure green LED is utilized as light source for both turbidimetric and colorimetric mea­surements. A silicon photocell receives the transmitted light from colorimetric channel, while another photocell detects the scattered light from the turbidimetric (nephelometric) channel.
In order to measure chlorine parameters, it is necessary to do the zero with the blank sample, and then add 1 packet of reagent. After placing the cuvet back in the meter and pressing READ, the measurement is shown directly on the LCD.
The instrument also features an automatic shut-off with user-selectable delay at 10, 20, 30, 40, 50 or 60 minutes.
To facilitate field tests, the meter provides a logging mode for storing up to 25 measure­ments along with time and date information for retrieval at a later date.
4
5
Page 4
LED
CUVET MICROPROCESSOR
LIGHT
PRINCIPLE OF OPERATION
Turbidity Mode
HI 93114 has been designed to perform mea­surements according to the USEPA’s 180.1 method and the Standard Method 2130B.
The instrument functions by passing a beam of light through a vial containing the sample being measured.
The light source is a Pure Green LED to ensure that any interference caused by a colored samples is minimized.
A sensor, positioned at 90° with respect to the direction of light, detects the amount of light scattered by the undissolved particles present in the sample. The microprocessor converts such readings into NTU* values.
Colorimetric Mode
The color of every object we see is deter­mined by a process of absorption and emis­sion of the electromagnetic radiation (light) of its molecules.
Colorimetric analysis is based on the principle that specific compounds react with others to form a color, the intensity of which is proportional to the concentration of the substance being measured.
EMITTED LIGHT
Block diagram of an ion specific measurement
DETECTOR
When a substance is exposed to a beam of light intensity Io, a portion of the radiation is absorbed by the substance's molecules and a radiation of intensity I, lower than Io, is emitted.
The quantity of radiation absorbed is given by the Lambert-Beer Law:
log Io/I = ελ c d
NTU units are equal to FTU units. However, there are other known measurement units for turbidity, namely the Jackson Turbidity Unit (JTU) based on the old method of Jackson's candle, and Silica Unit (mg / L of SiO2). The conversion table between these measurement units is shown below:
JTU NTU/FTU SiO2 (mg/L)
JTU 1 19 2.5
NTU/FTU 0.053 1 0.13
SiO2 (mg/L) 0.4 7.5 1
* 1 NTU = 1 FTU
6
Where log Io/I = Absorbance (A)
ε
= molar extinction coefficient of the
λ
substance at wavelength λ
c = molar concentration of the
substance
d = optical distance light travels
through the sample
Since other factors are known, the concen­tration "c" can be calculated from the color intensity of the substance determined by the emitted radiation I.
An LED (Light Emitting Diode) emits radiation at a relatively narrow spectrum, supplying the system with the intensity Io.
7
Page 5
A substance absorbs a color complimentary to the color it emits. For example, a substance appears yellow because it absorbs blue light. As a result, the Hanna meters use LED’s with specific wavelengths to measure samples.
The optical distance (d) is measured by the internal diameter of the cuvet containing the sample.
The photoelectric cell collects the radiation I that is not absorbed by the sample and converts it into an electric current.
The microprocessor converts the value into the desired measuring unit and displays it on the LCD.
The measurement process is done in two phases: setting the meter to zero and actual measurement.
The cuvet is an optical element and hence has an important role in the measurement process. Both the measurement and the calibration cuvets must be optically identical to provide the same measurement conditions.
It is also important that the surface of the cuvet is clean and free from scratches or dents, in order to avoid measurement interference due to unwanted reflection and absorption of light.
It is recommended that wherever possible the cuvet walls are not touched by the operator.
Furthermore, in order to maintain the same conditions during the zeroing and the measuring phases, it is necessary to close the cuvets to prevent any contamination.
FUNCTIONAL DESCRIPTION
1) Primary LCD: When the meter is switched on, the LCD
shows all segments for a few seconds. It then displays four dashes to indicate
“ready to measure”. In “Read” and “Zero” mode, “SIP” is shown
to indicate “Sample In Progress”. The upper level also indicates the concen-
tration or turbidity value of the sample, as well as different diagnostic modes, such as “-BA-” for low battery.
8
9
Page 6
2) Secondary LCD: The three-digit lower level shows the cur-
rent measurement mode (“F CL”, “tCL”, or “tr”), and diagnostic or calibration modes, such as “d11”, “2 Fn”, “5 c1”.
3) DATE: Indicates that the main LCD
is showing the current date, the date of last calibration or the date of logged measure­ment in memory.
4) TIME: Indicates that the main LCD
is showing the current time, the time of last calibration or the time of logged measure­ment in memory.
5) LOBAT: Blinks to warn the user of low
battery voltage.
6) LOG: Blinks to indicate that the
meter is in the scroll mode viewing the logged data.
If fixed, it indicates that the meter is in the log mode and every reading will be stored in memory.
7) ON/OFF key:Turns the meter on and off.
8) ZERO/ key: In ion specific mode, it ze-
ros the sample. In calibration and diagnostic
modes, it works as ENTER. In turbidity mode, it is not
used.
9) READ/ key: Takes concentration/tur-
bidity measurement of the sample.
In diagnostic or calibration mode, shifts the flashing digit to the right.
10
10) CAL key: If pressed during calibration, the calibration procedure will be aborted and the last cali­bration data will be retrieved.
If pressed together with the ALT key for less then 3 sec­onds, the diagnostic mode will be entered.
If pressed together with the ALT key again, the meter will quit diagnostic mode.
If pressed for more then 3 seconds, a blinking “CAL” will appear and the calibration procedure is entered.
11) GLP/Abs key: In ion specific mode, it will toggle concentration/absor­bance readings on main LCD.
In turbidity mode, GLP data (date, time and the 2 calibra­tion values) will be shown.
If pressed in time/date setup mode, the meter will quit cur­rent mode without making any change.
12) key Scrolls upwards through the parameters to be measured.
In calibration and diagnostic modes, increments the blink­ing digit by one.
If pressed together with ALT while the meter is in logging mode, the upper LCD will show the stored data (date, time, value).
13) ALT key: Activates alternative functions.
14) key: Scrolls downwards through the parameters to be measured.
11
Page 7
In calibration and diagnostic modes, decreases the blink­ing digit by one.
If pressed together with ALT while the meter is in logging mode, the upper LCD will show the current lot number.
SPECIFICATIONS
HI 93114
Range Turbidity 0.00 to 50.0 NTU*
Resolution
Accuracy
Free & Total Cl
Light Source Pure green LED Light Source Life Life of the instrument Light Detector Two silicon photocells Battery Type 4 x 1.5V AA alkaline Battery Life 60 hours or 1000 measurements Auto-off Selectable after
Environment 0 to 50°C (32 to 122°F);
Dimensions 220 x 82 x 66 mm
Weight 510 g (1.1 lb.)
Free Cl
Free Cl
2 2
2 2
2
Total Cl
Turbidity 0.01 and 0.1 NTU*
Total Cl
Turbidity ±0.5 NTU* or ±5% (whichever greater)
0.00 to 2.50 mg/L
0.00 to 3.50 mg/L
0.01 mg/L
0.01 mg/L
±0.03 mg/L or
±3% (whichever greater)
10, 20, 30, 40, 50 or 60 minutes
RH max 95% non-condensing
(8.7 x 3.2 x 2.6")
* 1 NTU = 1 FTU
12
OPERATIONAL GUIDE
SET CURRENT TIME/DATE
To set or change the current time, turn on the meter. After initialization routine, the LCD will show:
Press and hold the ALT and GLP keys to­gether. Display will show the current date in MM.DD format (e.g. August 28 is shown as
08.28).
+
Release the keys. The month digits will blink. Make the necessary adjustments with the and keys. To skip to the day digits, press the key.
After the adjustments, press the key. The unit will store the newly set month-day data in its EEPROM and will show the current time by a 24 hour clock HH.MM format, e.g. 2:28 pm is:
Similarly, make the necessary adjustments as described above and press . The newly set up “month - day - hour - minute” data will be stored in memory.
13
Page 8
CURRENT TIME/DATE RECALL
To recall current TIME/DATE press and hold the ALT and READ keys together. The cur­rent time and an intermittent “TIME” will be displayed.
+
Release the keys. Press and hold the ALT and READ keys
again and the meter will show the current date together with an intermittent “DATE”.
+
TURBIDITY MEASUREMENTS
Fill the vial with the sample. The surface of the vial should be clean and scratch free.
Turn the meter on. After the initialization rou­tine, the LCD will show:
Press READ. The meter will intermittently display “SIP” on the upper level of the LCD.
After a few seconds the display will show the turbidity value, e.g. 5.34 NTU:
COLORIMETRIC MEASUREMENTS
Turn the meter on. After the initialization rou­tine, the LCD will display:
Use the and keys to choose the desired pa­rameter:
F CL Free Chlorine t CL Total Chlorine
Use the and keys to set the lower level of the LCD to turbidity (tr).
Insert the sample into the cuvet holder and ensure that the notch on the cap is positioned securely into the groove.
14
Measuring Chlorine
Fill the vial with the sample (blank). The surface of the vial should be clean and scratch free.
Insert the blank sample into the cuvet holder and ensure that the notch on the cap is positioned securely into the groove. Press ZERO.
15
Page 9
The meter will show “SIP” for a few seconds and then a zero indication:
For the Free or Total Chlorine add the con­tent of its respective packet:
1 x
Replace the cap and shake the cuvet. For best results wait 2½ mins. for Total Chlorine. Insert the reacted sample into the cuvet holder and ensure that the notch on the cap is posi­tioned securely into the groove. Press READ.
The main LCD will show four dashes and “d 00” is displayed on the lower part, with the sec­ond “0” flashing:
Using the , and keys change the lower row of the display to show “d 11”:
Press the key. This key is also used as a toggle between USER (USR) and FACTORY (FCT) programmed functions. Select the USER mode (if necessary by pressing the key more than once):
Press ALT and CAL until four dashes are displayed on the upper display and “# fn” are shown in the lower part of the LCD.
The meter will show “SIP” for a few seconds and then the concentration:
Measurements in user-customized mode
Note: The meter must be calibrated for this
purpose. Follow the 2 point procedure on page 26 before proceeding.
Turn the meter on, and press both ALT and CAL buttons.
+
16
Using the and keys select the number from 0 to 1 where the appropriate calibration data have been stored. Add the appropriate reagent into the blank sample cuvet. Shake and allow a few sec­onds for color to develop.
17
Page 10
Insert the reacted sample into the cuvet holder and make sure that the notch on the cap is positioned securely into the groove. Press READ.
The meter will first show “SIP” for a few seconds and afterwards the sample concen­tration:
CALIBRATION
TURBIDITY CALIBRATION
The meter should be properly calibrated with a standard prepared as described in USEPA method 180.1.
Span calibration
To calibrate the span of the meter, fill the cuvet with the primary tur­bidity formazine standard of 50NTU.
Inspect and clean thoroughly the surface of the vial.
Shake the standard vigorously for a few seconds and wait a few minutes for the bubbles to disap­pear.
Turn the meter on and press both ALT and CAL momentarily.
+
The LCD will show four dashes on the upper and “d 00” on the lower part of the LCD with the second “0” blinking.
18
Using the , and keys change the lower row of the display to show “d 21”.
Insert the previously prepared 50 NTU stan­dard into the cuvet holder and make sure that the notch on the cap is positioned se­curely into the groove.
Press .
19
Page 11
The display will blink “-Lt-” for several sec­onds, indicating that the LED is being adjusted for the turbidimetric channel.
Afterwards, a sequence of numbers between
-511 to 512 will appear on the upper part of the LCD indicating different levels of LED light intensity.
In approximately one minute, the adjustment will be made and the calibration data stored in the non-volatile memory.
The display will show four dashes again indicating the end of the span cali­bration procedure.
Press the ALT and CAL keys together again to leave the diagnostic mode.
Two-point customized calibration
To enter the turbidity calibration mode, the meter should be in “tur­bidity” mode.
Use the and keys to set the lower level of the LCD to “tr”.
To enter the calibration mode, press and hold the ALT and CAL keys together for at least three seconds. The upper display will start flashing “CAL” for approximately 3 seconds.
+
+
If the CAL key is not pressed, the upper dis­play will show “----”, in­dicating that calibration mode was not entered. In which case, hold down the ALT and CAL keys together for 3 seconds to restart the procedure.
After entering the cali­bration mode, the dis­play will show the first point of the previous calibration. The most significant digit will also be blinking.
Using the , , keys, set the turbidity of first calibration standard (from 0.00 to 50.0 NTU).
Insert the standard vial into the cuvet holder and ensure that the notch is positioned se­curely into the groove. Press the key. The display will indicate Sample In Progress (SIP).
After the first calibra­tion point is memo­rized, the LCD will indi­cate the second point of the previous calibration with the most sig­nificant digit blinking.
Using the , , keys, set the turbidity of the second calibration standard (from 0.00 to 50.0 NTU). Insert the standard into the cuvet holder and make sure that the notch is positioned securely into the groove. Press the key. The display will indicate “SIP” again.
To confirm entry into the calibration mode, press the CAL key again while “CAL” is blinking.
20
21
Page 12
After the second calibration point is memo­rized, the unit will store the calibration data together with time and date in the EEPROM while intermittently indicating “CAL” and “Stor” for several seconds.
Subsequently, the up­per display will show “----”, indicating that the meter is calibrated and ready to measure turbidity of an unknown sample.
By pressing CAL during calibration, user can quit the calibration mode at any time without changing the previously stored calibration data.
COLORIMETRIC CALIBRATION
Zero calibration
To calibrate the span of the meter, fill the cuvet with a clean deionized water sample. Inspect and thoroughly clean the surface of the vial.
Turn the meter on and press both ALT and CAL momentarily.
+
The display will show four dashes and “d 00”. The second “0” will blink to allow the user to make a selection.
22
Using , , keys set the lower part to “d 31”.
Insert the previously prepared deionized wa­ter standard into the cuvet holder and make sure that the notch on the cap is positioned securely into the groove. Press the key.
The display will blink “-Lc-” for several sec­onds, indicating that adjustment of the LED for the colorimetric measurements is in progress.
After this, a sequence of numbers between
-511 to 512 will appear on the upper LCD
indicating the different levels of LED light intensity. In approximately one minute, the adjustment will be made and the calibration data will be stored in the non-volatile memory.
Display will show four dashes again indicating the end of the zero cali­bration procedure.
Press the ALT and CAL keys again to leave the diagnostic mode.
Two-point customized calibration
To enter the colorimetric calibration mode, press momentarily both the ALT and CAL keys. The LCD will show four dashes and “d00”.
+
+
23
Page 13
Using the and keys, set the lower part of the LCD to “d 11” and then press the key.
The key is used as a toggle function in this mode and allows the user to select be­tween USER (USR) or FACTORY (FCT) pro­grammed functions.
Select the USER mode and press ALT and CAL keys together to leave the diagnostic mode. The display of the meter will indicate four dashes together with a number 0 or 1 on the lower part of the LCD.
+
Press and hold the ALT and CAL keys to­gether for at least three seconds. The upper display will start flashing “CAL” for approxi­mately three seconds.
+
To confirm entry into the calibration mode, press the CAL key again while “CAL” is still blinking.
If the CAL key is not pressed, the upper row of the display will show “----”, indicating that the calibration mode was not entered. In which case, hold down the ALT and CAL keys together for 3 seconds to restart the procedure.
After entering the cali­bration mode, the dis­play will show the first point of the previous calibration with the most significant digit blinking.
Using the , and keys, choose the concentration of the first calibration standard.
Insert the known standard into the cuvet holder and make sure that the notch is positioned securely into the groove. Press the key. The display will indicate sample in progress.
After the first calibra­tion point is memo­rized, the LCD will indi­cate the second point of the previous calibration with the most sig­nificant digit blinking.
Using the , , keys, choose the known concentration of the second calibration stan­dard.
Insert the standard into the cuvet holder and make sure that the notch is positioned se­curely into the groove. Press the key. The display will blink “SIP” again.
After the second calibration point is memo­rized, the unit will store the calibration data, time and date in the EEPROM while intermit­tently indicating “CAL” and “Stor” for several seconds.
24
25
Page 14
Subsequently, the up­per display will show “----”, indicating that the meter is calibrated and ready to measure the concentration of an un­known sample.
By pressing the CAL key dur­ing calibration, the calibration mode can be exited at any time without changing the previously stored calibration data.
DIAGNOSTIC MODE
HI 93114 facilitates operations by providing a
diagnostic mode. In this mode, user can set or verify different parameters necessary to ensure optimum performance of the meter. To enter the diagnostic mode, turn the meter on and momentarily press ALT and CAL to­gether. The display will show four dashes to­gether with “d 00”:
+
Using the , and keys, select the re­quired diagnostic mode and press the key. The meter will execute one of the following user-diagnostic functions:
10 Customizes automatic shutdown 11 Selects User or Factory functions 12 Turns logging mode on or off 21 Calibrates span in turbidimetric mode 31 Calibrates span in colorimetric mode
40 Clears the logged memory
26
The following diagnostic modes are reserved for authorized service technicians:
00 Shows the Blank level in colorimetric mode 01 Shows the Sample level in colorimetric mode 02 Shows the Dark level in colorimetric mode 05 Shows the Ground voltage 06 Shows 5V on-board level 07 Shows battery voltage level 08 Shows 1.23V reference voltage level 09 Shows -5V on-board level 99 Shows software version number
27
Page 15
To quit diagnostic mode, press the ALT and CAL keys together again.
LOGGING
HI 93114 allows user to log 25 time/day-tagged
measurements. User can easily turn the log­ging mode on and off, review the logged memory, review the current lot number and clean the buffer (memory). HI 93114 also reminds the user if its memory is full.
TURNING THE LOG MODE ON OR OFF
Enter the diagnostic mode by pressing ALT and CAL together.
Select mode 12 and press the
key.
The display will show the current (vacant) lot together with “LOG” if the log mode is on. Otherwise it will show “----” if the log mode is off.
+
+
of the display to remind the user that every time a measurement is taken, the value is stored in the next available lot number.
REVIEWING THE LOGGED BUFFER
To review the memo­rized values, press to-
gether ALT and . The meter will scroll all the data in the buffer
showing the lot number, value, date and time. e.g. The first recorded reading in the buffer is
lot # 0, 0.35 mg/L of Free Chlorine, memorized on 23rd August at 3:34 pm;
The second logged data relates to lot # 1, 1.35 mg/L of the customized param­eter, logged on 23rd August at 3:55 pm.
First lot Second lot
+
By pressing the key, the meter toggles between the Log on and off positions. If the log on mode is selected, every time a mea­surement is taken (READ pressed) the rel­evant values will be stored in the current (vacant) lot number.
To quit diagnostic mode, press ALT and CAL together again.
The LCD will then show “CAL” and “Stor” for a few seconds. If the log on mode was selected “LOG” will appear on the bottom left hand side
28
+
CLEARING THE MEMORY
After all the buffer (memory) is taken up, the LCD will blink “Full”.
29
Page 16
To clear the buffer, press ALT and CAL.
Select mode 40. Press the key. The display will show the “Cln” message, indicating that memory is being cleaned.
+
The meter will store new settings in its non­volatile memory and the display will flash “CAL” and “Stor” alternately for several sec­onds.
The lot number will be reset to 00 automatically.
REVIEWING THE CURRENT LOT NUMBER
To check the current (vacant) lot number, while in log mode, press ALT and to­gether.
USER-SELECTABLE SHUTDOWN
With HI 93114, the users can customize the shutdown time to save power.
To change the shutdown time, enter the diagnos­tic mode by momentarily pressing ALT and CAL. Select mode 10 and press the key repeat­edly to set the desired shutdown time from 10 to 60 minutes with 10 minute increments, or disable the shutdown mode by choosing the OFF selection.
+
BATTERY REPLACEMENT
A “LOBAT” indication appears on the display when the batteries are weak and require re­placement. The instrument can still perform approximately 50 measurements.
A “-BA-” indication will appear when the bat­teries are too weak to ensure accurate mea­surements. This mes­sage appears for a few seconds and then the meter turns off. Batteries must be replaced.
Batteries should only be replaced in a safe area and using 1.5V AA alkaline type.
To replace the batter­ies, simply remove the two screws on the rear cover of the instrument and replace all four 1.5V AA batteries with new ones, while paying at­tention to their polarity.
After the selection is completed, exit the di­agnostic mode by press­ing ALT+CAL together.
30
+
31
Page 17
DIAGNOSTIC CODES
ACCESSORIES
LOBAT Weak batteries: change all batter-
ies as soon as possible.
-BA- Exhausted batteries: change all bat­teries immediately.
-LO- Low light level is received during the zeroing procedure: check the cuvet for scratches and ensure that the sample is not excessively turbid. Repeat the reading. If the problem persists, recalibrate the meter us­ing deionized water in the diagnos­tic mode “31” (see “Calibration”).
-CAP- High light intensity during last mea­surement: make sure that the cuvet is capped and placed properly in the holder, and that the ambient light does not reach the photode­tector. Repeat the measurement. If the problem persists, contact your dealer or the nearest HANNA office.
Er 1 Hardware error: repeat the measure-
ment. If the error message appears again, contact your dealer or the nearest HANNA office.
HI 731318 Tissue for wiping cuvets (4pcs) HI 731327 Maintenance kit: rugged
carrying case including HI93102-0 and HI93102-20 calibration solutions, HI 93703- 50 cuvet cleaning solution, tissue for wiping cuvets and 2 cuvets
HI 93102-0 AMCO-AEPA-1 calibration so-
lution @0 NTU* , 30 mL
HI 93102-20 AMCO-AEPA-1 calibration so-
lution @20 NTU*, 30 mL
HI 93701-01 Reagent kit for 100 free chlo-
rine tests
HI 93701-03 Reagent kit for 300 free chlo-
rine tests
HI 93703-50 Cuvet cleaning solution, 230
mL bottle
HI 93711-01 Reagent kit for 100 total chlo-
rine tests
HI 93711-03 Reagent kit for 300 total chlo-
rine tests
rnG Out of range: check the measuring
procedure and verify the concentra­tion of the sample to ensure that is not too high.
32
* 1 NTU = 1 FTU.
33
Page 18
WARRANTY
Hanna Instruments meters are guaranteed for two years against defects in workman-
ship and materials when used for their in­tended purpose and maintained according to instructions.
This warranty is limited to repair or replace­ment free of charge. Damages due to acci­dents, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom the instrument was purchased. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. First obtain a Returned Goods Autho­rization number from the Customer Service department and then return the instrument indicating the Authorization # with shipment costs prepaid.
If the repair is not covered by the warranty, you will be notified of the charges.
When shipping any instrument, make sure it is properly packaged for complete protection.
CE DECLARATION OF CONFORMITY
HI 93114
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
34
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment in which it is used.
Operation of this instrument in residential areas could cause unacceptable interference to radio and TV equipment, requiring the operator to take all neces­sary steps to correct the interference.
Any variation introduced by the user to the supplied equipment, may degrade the instrument's EMC performance.
To avoid electrical shocks, do not use this instrument when voltage at the measurement surface exceeds 24 Vac or 60 Vdc.
To avoid damage or burns, do not perform any measurement in microwave ovens.
35
Page 19
SALES & TECHNICAL SERVICE
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and Technical offices, please see www.hannainst.com
MANC114R2 09/05
Loading...