CE declaration of conformity ............................. 30
Remove the instrument from the packing material
®
and examine it carefully to make sure that no
damage has occurred during shipping. If there is
any noticeable damage, immediately notify your
dealer.
Each meter is supplied with batteries, rugged
carrying case and instructions
Note: Save all packing materials until you are sure
that the instrument functions correctly. All
defective items must be returned in the original
packing materials together with the supplied
accessories.
GENERAL DESCRIPTION
HANNA instruments® has designed these two rugged
waterproof salinity and sodium meters for regular use
in the lab, as well as in harsh processing
environments.
HI 931100 is an ion-selective meter that uses a sodium
electrode to read the salt (NaCl) content of a solution.
HI 931100 autoranges (4 scales available) from sample
to sample allowing an extremely wide range of
measurements without recalibration.
HI 931101 employs the FC 300B combination
sodium electrode to provide sodium readings from 15.0
mg/L to 60 g/L.
The calibration is automatic at 2 points, the first is at
2.3 g/L while the second can be either at 0.23 g/L (low
range) or at 23.0 g/L (high range).
A separate temperature probe, HI 7662, provides temperature readings from 0 to 80ºC.
32
SPECIFICATIONS HI 931100FUNCTIONAL DESCRIPTION HI 931100
Range
0.150 to 1.500 / 1.50 to 15.00 / 15.0 to 150.0
150 to 300 g/L NaCl
0.0 to 80.0°C
Resolution
0.001 / 0.01 / 0.1 / 1 g/L NaCl
0.1°C
Accuracy±5% of reading (NaCl)
(@20°C)±0.5°C
Typical EMC Deviation
±2% FS (NaCl) / ±0.5°C
CalibrationAutomatic, 1 or 2 point
at 3.00 g/L (HI 7083) and
0.30 g/L (HI 7085) or 30.0 g/L (HI 7081)
Temperature Compensation
25°C (77°F)
ElectrodeFC300B sodium electrode
with BNC and 1 m (3.3) cable (not included)
Temperature Probe
HI 7662 (not included)
Input Impedance1012 Ohm
Battery Type / Life
4 x 1.5V AA / approx. 200 hours of use
Environment0 to 50°C (32 to 122°F);
RH max 100%
Dimensions196 x 80 x 60 mm
(7.7 x 3.1 x 2.4")
Weight500 g (1.1 lb.)
1. Battery compartment
2. BNC connector for electrode
3. Temperature probe socket
4. LCD (Liquid Crystal Display)
5. ON/OFF key, to turn the meter on or off
6. CAL key, to enter/exit calibration mode
7. CFM key, to confirm calibration
8. BUF key, to select calibration buffer value
54
1. Battery compartment
2. BNC connector for electrode
3. Temperature probe socket
4. LCD (Liquid Crystal Display)
5. ON/OFF key, to turn the meter on or off
6. CAL key, to enter/exit calibration mode
7. MODE key, to select Na or pNa range
8. CFM key, to confirm calibration
9. BUF key, to select calibration buffer value
SPECIFICATIONS HI 931101FUNCTIONAL DESCRIPTION HI 931101
Range0.00 to 3.00 pNa
15.0 to 150.0 mg/L Na
0.150 to 1.500 / 1.50 to 15.00 / 15.0 to 60.0 g/L Na
0.0 to 80.0°C
Resolution0.01 pNa / 0.1 mg/L Na
0.001 / 0.01 / 0.1 / 1 g/L Na
0.1°C
Accuracy±0.05 pNa
(@20°C)±5% of reading (Na)
±0.5°C
Typical EMC Deviation
±2% FS (pNa and Na) / ±0.5°C
CalibrationAutomatic, 1 or 2 point
at 2.3 g/L (HI 7080) and
0.23 g/L (HI 7087) or 23.0 g/L (HI 7086)
Temperature Compensation
25°C (77°F)
ElectrodeFC300B sodium electrode
with BNC an 1 m (3.3) cable (not included)
Temperature Probe
HI 7662 (not included)
Input Impedance1012 Ohm
Battery Type / Life
4 x 1.5V AA / approx. 200 hours of use
Environment0 to 50°C (32 to 122°F);
RH max 100%
Dimensions196 x 80 x 60 mm
(7.7 x 3.1 x 2.4")
Weight500 g (1.1 lb.)
76
OPERATIONAL GUIDE
FC 300B
HI 7662N
E
INITIAL PREPARATION
Each meter is supplied complete with four 1.5V AA batteries.
Remove the back cover, unwrap
the batteries and install them
while paying attention to their
polarity.
PROBE CONNECTION
Connect the FC300B sodium
electrode to the BNC socket on
the top of the instrument.
Connect the HI 7662 temperature probe to the RCA socket.
SODIUM
ELECTRODE
TEMPERATUR
PROBE
For accurate readings, calibration should be performed (or verified) every day if measuring the same
type of samples (e.g. cheese), or every time the
sample type is changed (e.g. cheese, meat, sea
water, etc.).
The recommended working temperature is around
20-25°C (68-77°F). For greater accuracy, perform
the calibration with standard solutions at a temperature as close as possible to the temperature of the
sample.
TAKING SODIUM MEASUREMENTS
Remove the protective cap from the
FC300B sodium electrode and, if
possible, rinse it with some of the
sample to be measured.
The temperature probe can be
used independently to take temperature measurements.
If the probe is not plugged in,
the secondary display shows
NaCl
gm
"25°C" with the "°C" indicator
blinking.
To switch the instrument on,
press and hold the ON/OFF key
for a fraction of a second.
These meters have a built-in protection against electromagnetic interferences and the holding of the
keys is to ensure that the commands are not mistaken for erroneous signals.
In order to take accurate measurements, make
sure the instrument is calibrated before use
(see "Calibration" sections for details).
Shift downwards the rubber sleeve
(see picture) to have a properly working junction.
°C
Immerse the tip of the FC300B electrode (at least 2.5 cm/1") into the
sample to be tested while making
sure that the electrode junction is
completely immersed, and without
touching the bottom of the beaker.
Wait for approximately 10 to 15 minutes to allow
for the electrode to adjust and stabilize, i.e. until
no drifting is observed for at least a couple of
minutes.
During measurement, if possible,
the sample should be gently
stirred (100 rpm).
MAGNETIC STIRRER
HI 190M
ON
POWER
OFF MAX
SPEED
98
TAKING pNa MEASUREMENTS (HI 931101 only)
Press the MODE key during measurement to convert the value displayed by the HI 931101 to pNa.
Note that
pNa = -log a
Na
where aNa is the activity of ion Na.
°C
MODE
Press the MODE key again to return to g/L measurements.
TEMPERATURE MEASUREMENTS
Connect the temperature probe to
the meter and turn on.
Dip the HI 7662 temperature probe
into the sample. Allow the reading
to stabilize.
If the electrode is used rarely (once a month or
less), store it dry. Before using it again, let it
condition overnight in HI 7081 (30.0 g/L NaCl) or HI
7086 (23 g/L Na 1M)
To avoid very long response time, do not store
or rinse the electrode with deionized water. Do
not rub the electrode.
Note: The temperature is always
displayed on the secondary
NaCl
gm
LCD.
AFTER USE
Turn the meter off by pressing the
ON/OFF key again.
For faster and accurate measurements, the electrode
should be stored in a solution with a Na/NaCl concentration similar to the sample to be measured.
After use, shift back upwards the
rubber sleeve (see picture).
°C
1110
Na CALIBRATION (HI 931101)
Calibration can be performed at 1 (offset only) or 2
points.
For best accuracy, a 2 point procedure is recommended.
PREPARATION
• Pour small quantities of
HI 7080
HI7080 (2.3 g/L Na) solution
and HI 7087 (0.23 g/L Na) or
HI 7080
HI 7087
HI 7086 (23 g/L Na) into two
beakers.
For accurate readings, use for the second point
the HI 7087 solution if you are going to measure
samples with low sodium content (below 2.3 g/
L), or the HI 7086 standard if you are going to
measure samples with high sodium content (over
2.3 g/L).
• For accurate calibration use
two beakers for each solution, the first one for rinsing
RINSECALIBRATION
HI 7080HI 7080
the electrode, the second one
for calibration.
If possible, use plastic beakers to minimize any
EMC interferences.
• Turn the meter on by pressing the ON/OFF key.
ON/OFF
• Remove the protective cap and rinse the electrode with some of the calibration solution you
are going to use first.
•Shift downwards the rubber
sleeve for a properly working
junction (see picture).
SINGLE POINT CALIBRATION (OFFSET)
• Dip the FC300B sodium elec-
trode into the HI7080 (2.3g/L
Na) solution.
Note: Immerse the electrode
approximately 2.5 cm
(1") into the solution.
• Press the CAL key. The "CAL" and "BUF" indicators will be displayed with the "BUF" and the
"" symbols blinking.
CAL
• As soon as the standard solution is recognized, only the
"" symbol continues
blinking to indicate that the
electrode is stabilizing. The
required standard solution being value is displayed on the
primary LCD.
• As soon as the electrode stabilized, the "CON" symbol appears on the display.
Press the CFM key to save
the calibration point. For best
results, wait for a couple of
minutes before confirming.
• HI 931101 will display "0.230
CFM
gm".
gm
CAL
BUF
°C
1312
• For exiting the calibration mode
and save the single point calibration, press the CAL key. For
better accuracy however, it is
recommended to perform a 2
point calibration.
TWO POINT CALIBRATION
CAL
Wait for a couple of minutes
and then press the CFM key.
CFM
• The two point calibration procedure is now completed and the meter automatically returns to the
normal measurement mode.
• After the first point is confirmed,
immerse the FC300B electrode
into the second standard solution: HI 7087 at 0.23 g/L Na or
HI 7086 at 23 g/L Na.
Note: The electrode should be
immersed approximately
2.5 cm (1") into the solution.
•Select the calibration value by
pressing the BUF key. Make
sure that the selected value corresponds to the solution being
used for the second calibration
point.
• If the reading is not close to the
selected calibration standard, the
"BUF" indicator will continue to
blink.
Otherwise, if the reading is close
to the selected calibration value,
the meter will recognize the standard solution and only the ""
symbol will blink, to indicate that
the reading is stabilizing.
BUF
NaCl CALIBRATION (HI 931100)
Calibration can be performed at 1 (offset only) or 2
points.
For best accuracy, a two point procedure is recommended.
PREPARATION
• Pour small quantities of
HI7083 (3.00 g/L NaCl) solution and HI 7085 (0.30 g/L) orHI 7081 (30 g/L) into two clean
HI 7083
HI 7083
HI 7085
beakers.
For more accurate readings use for the second
poit the HI 7085 solution if you are going to
measure samples with low sodium chloride content (< 3 g/L), or the HI 7081 solution for samples
with high sodium chloride content (> 3 g/L).
• For an accurate calibration, use two beakers for
each solution, the first one for
rinsing the electrode, the second one for calibration.
If possible use plastic beakers to minimize any
EMC interferences.
• Turn the meter on by pressing the ON/OFF key.
•Once stabilized "CON" symbol
will appear on the display.
•Remove the protective cap and rinse the electrode with some of the calibration solution you
are going to use first.
•Shift downwards the rubber
sleeve for a properly working
junction (see picture).
1514
SINGLE POINT CALIBRATION (OFFSET)
• Dip the FC300B sodium electrode into the HI7083 (3.00 g/L
NaCl) solution.
Note: Immerse the electrode
approx. 2.5 cm (1") into
the solution.
•Press the CAL key. The "CAL" and "BUF" indicators will be displayed with the "BUF" and the
"" symbols blinking.
CAL
• As soon as the standard solution is recognized, only the
"" symbol will continue
blinking to indicate that the electrode is stabilizing.
• To quit the calibration and save
the first point, press the CAL
key.
For better accuracy however, it is recommended
that a two-point calibration is performed.
TWO POINT CALIBRATION
• After the first point is confirmed,
immerse the FC300B electrode
into the second solution:
HI7085 at 0.3 g/L NaCl or
HI7081 at 30 g/L NaCl.
Note: The electrode should be
immersed approximately
2.5 cm (1") into the solution.
• Select the second point value
by pressing the BUF key. Make
sure that the selected value corresponds to the one being used
for the second calibration point.
• If the reading is not close to
the selected value, the "BUF"
symbol will blink.
CAL
BUF
The value of the standard solution being used will be displayed
on the primary LCD.
• As soon as the electrode is stabilized, "CON" symbol lights up.
For best results, wait for a couple
of minutes before pressing the
CFM key.
• HI 931100 will then display "30.0
gm".
Otherwise, if the reading is
close to the selected value, the
meter will recognize the standard solution. Only the ""
symbol will blink, to indicate
that the reading is stabilizing.
•Once stabilized "CON" symbol
will appear on the display.
CFM
For best results, wait for a
couple of minutes and press
NaCl
CAL
BUF
gm
°C
the CFM key.
• The two point calibration is now
completed and the meter re-
CFM
turns to the normal mode.
1716
N
SPECIAL APPLICATIONS &
ADDITIONAL INFORMATION
HOW TO MEASURE SODIUM CHLORIDE IN
CHEESE AND MEAT (HI 931100)
1. Grind a sample of cheese or meat.
MAGNETIC STIRRER
HI 190M
FILTERED
SOLUTIO
ON
POWER
OFF MAX
SPEED
2. Pour a quantity (X g) in a glass beaker. Add
(X) x 10 mL of deionized water (note: ignoring
water content of cheese and meat).
3. Cover the beaker to prevent any spilling during
stirring, place the beaker on a hot plate stirrer
and agitate the sample at 40-50°C (104-122°F)
for 15-20 minutes.
4. Let the sample cool to the ambient temperature,
then filter it.
5. Dip the FC 300B sodium electrode into the fil-
tered sample. The tip of the electrode should be
immersed for approximately 2.5 cm/1" (i.e. the
electrode junction must be completely immersed)
without touching the bottom of the beaker. During measurement, the sample should be gently
stirred (100 rpm).
6. Wait for the reading to stabilize, i.e. no drift
should be observed for at least a couple of
minutes. Note the measured value.
7. Since the dilution rate is 1/10, the reading in the
g/L range has to be divided by 100 to get the
actual reading of NaCl (per gram of cheese or
meat), or the value can be considered directly
expressed in % of NaCl.
Note: A response time of 10 to 15 minutes for the
first measurement of cheese or meat, is the
correct expected time for the electrode to
stabilize. Next measurements of cheese or
meat samples will require shorter response
time (less than 2 minutes). Do not rinse the
sodium electrode between subsequent measurements since contamination is negligible.
IONIC STRENGTH ADJUSTER (ISA)
When the NaCl concentration of the sample is lower
than 5 g/L, it is recommend to use the HI 7090 ISA
solution. In such case, calibration should be performed with standard solution containing 2% of ISA.
ISA (Ionic Strength Adjuster) = 50% of NH4Cl 4M +
50% of NH4OH 4M (use 2 mL of ISA every 100 mL
of sample).
1918
TEMPERATURE
The best working temperature is around 20-25°C
(68-77°F).
TEMPERATURE CALIBRATION
(for Technical Personnel only)
For greatest accuracy, it is recommended to calibrate the meter with standard solution at a temperature as close as possible to the temperature of the
sample (±1°C/±2°F).
Na CONCENTRATION WITH HI 931100
It is always possible to determine the Na concentration by using the formula:
XNa = X
• (23 / 58.4)
NaCl
where
XNa = concentration of Na
and
X
= concentration of NaCl
NaCl
INITIAL PREPARATION
Prepare a 0°C/32°F bath with ice and water and
another one at 50°C/122°F using hot water at a
known temperature. Place insulation material around
the containers to minimize temperature changes.
Use a
Checktemp
or another accurate thermometer
with a resolution of 0.1° as reference.
0 °C
(32 °F)
°C
50 °C
(122 °F)
°C
Note: A blinking "°C" indicates
that the temperature probe
NaCl
is not connected.
PROCEDURE
• Connect the HI 7662 temperature probe to the
RCA socket on the top of the instrument.
•Turn the instrument off.
gm
°C
• To enter the temperature calibration mode, press
the CAL and ON/OFF keys simultaneously.
• Immerse the temperature probe
°C
into the 0°C/32°F bath together
with the reference thermometer.
0°C
(32°F)
•When the reading stabilizes,
press the CAL key. The
"0.0°C", "CAL" and "BUF"
symbols will be displayed. The
CAL
BUF
°C
"" symbol will blink for a
few seconds.
2120
• When the input value becomes stable, the ""
symbol disappears and
CON
°C
"CON" will start blinking.
Press the CFM key to confirm the calibration
value.
°C
• Immerse the temperature
probe into the 50°C (122°F)
bath.
50°C
(122°F)
•Press the CAL key to display the "0.0°C" and then
press the BUF key to select "50.0°C". "CAL" and
CAL
BUF
°C
"BUF" symbols. ""
symbol will blink for a few
seconds.
• When the input value becomes stable, the ""
symbol disappears and
CON
°C
"CON" will start blinking.
• Press the CFM key to confirm the calibration
value.
• The calibration is now complete.
• Turn the meter off and on again to return to the
normal mode.
FC300B CONDITIONING & MAINTENANCE
PREPARATION
Remove the protective cap.
DO NOT BE ALARMED IF ANY
SALT DEPOSITS ARE PRESENT.
This is normal with sodium electrodes
and they will disappear when rinsed
with water.
During transport tiny bubbles of air
may have formed inside the glass
bulb. The electrode cannot function
properly under these conditions.
These bubbles can be removed by
"shaking down" the electrode as you
would do with a glass thermometer.
If the bulb and/or junction
are dry, soak the electrode in HI 7081 (NaCl
30.0 g/L) solution overnight before use with
HI931100. Soak it in
HI7086 (Na 23 g/L) solu-
tion overnight before use
with HI931101.
If the electrolyte solution
is more than 1cm (½")
below the filling hole, add
HI 8093 1M KCl+AgCl
electrolyte solution.
Shift downwards the rubber sleeve during measurements and calibration.
SODIUM MEASUREMENT
Rinse the sodium electrode tip with
some of the sample to be tested,
then immerse the tip (approx. 2.5
cm /1") in the sample and stir gently.
2322
The sensitive bulb should not touch the bottom of
the beaker.
The Hanna HI 76405 electrode
holder can be used for this
purpose.
During measurement, the
sample should be gently
stirred (approx. 100 rpm). The
Hanna HI190M stirrer can be
used for this purpose.
MAGNETIC STIRRER
HI 190M
ON
POWER
OFF MAX
SPEED
STORAGE
When used frequently, the electrode should be stored
in a solution with a Na or NaCl concentration similar
to the sample to be measured. For longer periods of
storage, the electrode should be sored with a few
drops of appropriate solution in the protective cap.
For frequent use in different samples (1 or 2 times
per week), store the electrode in HI7081 (30.0 g/L
NaCl) solution if used for NaCl measurements with
HI 931100.
Store the electrode in HI 7080 (2.3 g/L Na) solution
if used for Na measurements with HI931101.
For occasional use (once a month or less) store the
electrode dry and let it conditioning overnight in
HI7081 (30.0 g/L NaCl) or HI7086 (23 g/L Na)
solution before use.
Note: NEVER STORE OR RINSE THE ELECTRODE
IN DISTILLED OR DEIONIZED WATER.
PERIODIC MAINTENANCE
Inspect electrode and cable. The cable used for
connection to the meter must be intact and there
must be no points of broken insulation on the cable
or cracks on the electrode stem or bulb. Connectors must be perfectly clean and dry.
If any scratches or cracks are present, replace the
electrode.
Rinse off any salt deposits with water.
Refill the electrode with fresh electrolyte (HI 8093)
and let it stand upright for 1 hour.
Follow the Storage Procedure above.
CLEANING PROCEDURE
GeneralSoak in Hanna HI 7061 general clean-
ing solution for approximately 1 hour.
Removal of films, dirt or deposits on the membrane/
junction:
-
Protein
Soak in Hanna HI 7073 protein cleaning solution for 15 minutes.
-
Inorganic
Soak in Hanna HI 7074 inorganic cleaning solution for 15 minutes.
-
Oil/grease
Rinse with Hanna HI 7077 or Oil & Fat
cleaning solution.
IMPORTANT: After performing any of the cleaning procedures rinse the electrode thoroughly
with a solution containing 2% ISA, refill (if
necessary) the reference chamber with fresh
electrolyte (HI8093) and soak the electrode in
a solution very close to the next sample to
be measured for at least 1 hour before taking
measurements.
TROUBLESHOOTING
Evaluate the performance of your electrode based
on the following.
• Noise (readings fluctuate up and down) could be
due to:
- Clogged/Dirty Junction: refer to the Cleaning
Procedure above.
- Loss of shielding due to low electrolyte level:
refill with HI 8093 solution
• Dry Membrane/Junction: soak it overnight in
HI7081 (30.0 g/L NaCl) or HI7086 (23 g/L Na)
solution before use.
• Drifting: soak the electrode tip in warm HI8093
solution for one hour. Refill with fresh HI 8093
electrolyte solution.
• Low Slope: refer to the Cleaning Procedure.
• No Slope: check the electrode for cracks in
glass stem or bulb and, if necessary, replace
the electrode.
• Slow Response/Excessive Drift: soak the tip
in HI 7061 solution for 30 minutes, rinse thor-
oughly in a solution with 2% of ISA, and then
follow the Cleaning Procedure.
2524
BATTERY REPLACEMENT
ACCESSORIES
If the batteries become weak the
LCD will display the "LOW BAT"
indication.
LOW BAT
Battery replacement must only take place in a safe
area and using the battery type specified in this
instruction manual.
To replace batteries, remove the two screws on the
rear cover and replace all four 1.5V AA batteries
with new ones, while paying attention to the correct
polarity. Replace the cover and tighten the two
screws.
Na SOLUTIONS FOR HI931101
HI 7080M2.3 g/L Na solution, 230 mL bottle
HI 8080M2.3 g/L Na solution, 230 mL FDA bottle
HI 7080L2.3 g/L Na solution, 500 mL bottle
HI 8080L2.3 g/L Na solution, 500 mL FDA bottle
HI 7086M23 g/L Na solution, 230 mL bottle
HI 8086M23 g/L Na solution, 230 mL FDA bottle
HI 7086L23 g/L Na solution, 500 mL bottle
HI 8086L23 g/L Na solution, 500 mL FDA bottle
HI 7087M0.230 g/L Na solution, 230 mL bottle
HI 8087M0.230 g/L Na solution, 230 mL FDA bottle
HI 7087L0.230 g/L Na solution, 500 mL bottle
HI 8087L0.230 g/L Na solution, 500 mL FDA bottle
NaCl SOLUTIONS FOR HI 931100
HI 7081M30 g/L NaCl solution, 230 mL bottle
HI 8081M30 g/L NaCl solution, 230 mL FDA bottle
HI 7081L30 g/L NaCl solution, 500 mL bottle
HI 8081L30 g/L NaCl solution, 500 mL FDA bottle
HI 7083M3.0 g/L NaCl solution, 230 mL bottle
HI 8083M3.0 g/L NaCl solution, 230 mL FDA bottle
HI 7083L3.0 g/L NaCl solution, 500 mL bottle
HI 8083L3.0 g/L NaCl solution, 500 mL FDA bottle
HI 7085M0.3 g/L NaCl solution, 230 mL bottle
HI 8085M0.3 g/L NaCl solution, 230 mL FDA bottle
HI 7085L0.3 g/L NaCl solution, 500 mL bottle
HI 8085L0.3 g/L NaCl solution, 500 mL FDA bottle
HI 7090MISA solution, 230 mL bottle
HI 8090MISA solution, 230 mL FDA bottle
HI 7090LISA solution, 500 mL bottle
HI 8090LISA solution, 500 mL FDA bottle
ELECTRODE MAINTENANCE SOLUTIONS
HI 7061MGeneral cleaning solution, 230 mL bottle
HI 8061M General cleaning solution, 230 mL FDA
bottle
HI 7061LGeneral cleaning solution, 500 mL bottle
2726
HI 8061LGeneral cleaning so luti on, 500 mL FDA
bottle
HI 7073M Protein cleaning solution, 230 mL bottle
HI 8073M Protein cleaning solution, 230 mL FDA
bottle
HI 7073LProtein cleaning solution, 500 mL bottle
HI 8073LProtein cleaning solution, 500 mL FDA
bottle
HI 7074M Inorganic cleaning solution, 230 mL bottle
HI 7074LInorganic cleaning solution, 500 mL bottle
HI 7077M Oil & Fat cleaning solution, 230 mL bottle
HI 8077M Oil & Fat cleaning solution, 230 mL
FDA bottle
HI 7077LOil & Fat cleaning solution, 500 mL bottle
HI 8077LOil & Fat cleaning solution, 500 mL
FDA bottle
HI 80931M KCl + AgCl electrolyte solution, 4 x
50 mL bottle
OTHER ACCESSORIES
HI 98501
ChecktempC
electronic thermometer
with penetration, stainless steel probe
and 0.1°C resolution (range: -50.0 to
150.0°C)
FC 300BSodium electrode, glass body, single
junction, refillable, with BNC connector and 1 m (3.3') cable
HI 190M/DMini-stirrer 230 V (1 liter max stirring
capacity, speed range from 100 to
max 1000 rpm)
HI 190M/UMini-stirrer 115 V (1 liter max stirring
capacity, speed range from 100 to
max 1000 rpm)
HI 710031Rugged carrying case
HI 7662Stainless steel temperature probe with
1 m (3.3') cable
HI 76405Electrode holder
WARRANTY
All Hanna Instruments meters are warranted for
two years against defects in workmanship and mate-
rials when used for their intended purpose and maintained according to instructions.
Electrodes and probes are warranted for six
months.
This warranty is limited to repair or replacement free of
charge.
Damages due to accident, misuse, tampering or lack
of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you
purchased the instrument. If under warranty, report the
model number, date of purchase, serial number and
the nature of the failure.
If the repair is not covered by the warranty, you will be
notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization
Number from the Customer Service department and
then send it with shipment costs prepaid.
When shipping any instrument, make sure it is properly packaged for complete protection.
Hanna Instruments reserves the right to modify
the design, construction and appearance of its
products without advance notice.
2928
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely
suitable for the environment in which they are used.
Operation of these instruments in residential area could cause
unacceptable interferences to radio and TV equipments, requiring
the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment
may degrade the instruments' EMC performance.
To avoid electrical shock, do not use these instruments when
voltages at the measurement surface exceed 24 Vac or 60 Vdc.
To avoid damages or burns, do not perform any measurement in
microwave ovens.
3130
SALES & TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and Technical
offices, please see www.hannainst.com
MANNACLR2 08/05
32
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.