Hanna Instruments HI 931100 User Manual

Instruction Manual
HI 931 100 • HI 931101
W aterproof Instrument s
for Salinity and Sodium Content Measurements
www.hannainst.com
1
Dear Customer, Thank you for choosing a HANNA instruments
product. Please read this instruction manual carefully before
using the instrument. If you need additional technical information, do not
hesitate to e-mail us at tech@hannainst.com These instruments are in compliance with the
directives.
TABLE OF CONTENTS
Preliminary Examination...................................... 3
General Description ............................................. 3
Functional Description HI931100 ......................... 4
Specifications HI931100 ...................................... 5
Functional Description HI931101 ......................... 6
Specifications HI 931101 ..................................... 7
Operational Guide................................................ 8
Na Calibration (HI 931101)................................. 12
NaCl Calibration (HI 931100) ............................. 15
Special Applications & Additional Info ............... 18
Temperature Calibration ..................................... 21
FC 300B Conditioning & Maintenance............... 23
Battery Replacement ......................................... 26
Accessories....................................................... 27
Warranty ............................................................ 29
CE declaration of conformity ............................. 30
Remove the instrument from the packing material
®
and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, immediately notify your dealer.
Each meter is supplied with batteries, rugged carrying case and instructions
Note: Save all packing materials until you are sure
that the instrument functions correctly. All defective items must be returned in the original packing materials together with the supplied accessories.
GENERAL DESCRIPTION
HANNA instruments® has designed these two rugged waterproof salinity and sodium meters for regular use in the lab, as well as in harsh processing environments. HI 931100 is an ion-selective meter that uses a sodium electrode to read the salt (NaCl) content of a solution. HI 931100 autoranges (4 scales available) from sample to sample allowing an extremely wide range of measurements without recalibration. HI 931101 employs the FC 300B combination sodium electrode to provide sodium readings from 15.0 mg/L to 60 g/L. The calibration is automatic at 2 points, the first is at
2.3 g/L while the second can be either at 0.23 g/L (low range) or at 23.0 g/L (high range). A separate temperature probe, HI 7662, provides tem­perature readings from 0 to 80ºC.
32
SPECIFICATIONS HI 931100FUNCTIONAL DESCRIPTION HI 931100
Range
0.150 to 1.500 / 1.50 to 15.00 / 15.0 to 150.0 150 to 300 g/L NaCl
0.0 to 80.0°C
Resolution
0.001 / 0.01 / 0.1 / 1 g/L NaCl
0.1°C
Accuracy ±5% of reading (NaCl) (@20°C) ±0.5°C
Typical EMC Deviation
±2% FS (NaCl) / ±0.5°C
Calibration Automatic, 1 or 2 point
at 3.00 g/L (HI 7083) and
0.30 g/L (HI 7085) or 30.0 g/L (HI 7081)
Temperature Compensation
25°C (77°F)
Electrode FC300B sodium electrode
with BNC and 1 m (3.3) cable (not included)
Temperature Probe
HI 7662 (not included)
Input Impedance 1012 Ohm Battery Type / Life
4 x 1.5V AA / approx. 200 hours of use
Environment 0 to 50°C (32 to 122°F);
RH max 100%
Dimensions 196 x 80 x 60 mm
(7.7 x 3.1 x 2.4")
Weight 500 g (1.1 lb.)
1. Battery compartment
2. BNC connector for electrode
3. Temperature probe socket
4. LCD (Liquid Crystal Display)
5. ON/OFF key, to turn the meter on or off
6. CAL key, to enter/exit calibration mode
7. CFM key, to confirm calibration
8. BUF key, to select calibration buffer value
54
1. Battery compartment
2. BNC connector for electrode
3. Temperature probe socket
4. LCD (Liquid Crystal Display)
5. ON/OFF key, to turn the meter on or off
6. CAL key, to enter/exit calibration mode
7. MODE key, to select Na or pNa range
8. CFM key, to confirm calibration
9. BUF key, to select calibration buffer value
SPECIFICATIONS HI 931101FUNCTIONAL DESCRIPTION HI 931101
Range 0.00 to 3.00 pNa
15.0 to 150.0 mg/L Na
0.150 to 1.500 / 1.50 to 15.00 / 15.0 to 60.0 g/L Na
0.0 to 80.0°C
Resolution 0.01 pNa / 0.1 mg/L Na
0.001 / 0.01 / 0.1 / 1 g/L Na
0.1°C
Accuracy ±0.05 pNa (@20°C) ±5% of reading (Na)
±0.5°C
Typical EMC Deviation
±2% FS (pNa and Na) / ±0.5°C
Calibration Automatic, 1 or 2 point
at 2.3 g/L (HI 7080) and
0.23 g/L (HI 7087) or 23.0 g/L (HI 7086)
Temperature Compensation
25°C (77°F)
Electrode FC300B sodium electrode
with BNC an 1 m (3.3) cable (not included)
Temperature Probe
HI 7662 (not included)
Input Impedance 1012 Ohm Battery Type / Life
4 x 1.5V AA / approx. 200 hours of use
Environment 0 to 50°C (32 to 122°F);
RH max 100%
Dimensions 196 x 80 x 60 mm
(7.7 x 3.1 x 2.4")
Weight 500 g (1.1 lb.)
76
OPERATIONAL GUIDE
FC 300B
HI 7662N
E
INITIAL PREPARATION
Each meter is supplied com­plete with four 1.5V AA batter­ies.
Remove the back cover, unwrap the batteries and install them while paying attention to their polarity.
PROBE CONNECTION
Connect the FC300B sodium electrode to the BNC socket on the top of the instrument.
Connect the HI 7662 tempera­ture probe to the RCA socket.
SODIUM
ELECTRODE
TEMPERATUR PROBE
For accurate readings, calibration should be per­formed (or verified) every day if measuring the same type of samples (e.g. cheese), or every time the sample type is changed (e.g. cheese, meat, sea water, etc.).
The recommended working temperature is around 20-25°C (68-77°F). For greater accuracy, perform the calibration with standard solutions at a tempera­ture as close as possible to the temperature of the sample.
TAKING SODIUM MEASUREMENTS
Remove the protective cap from the FC300B sodium electrode and, if possible, rinse it with some of the sample to be measured.
The temperature probe can be used independently to take tem­perature measurements.
If the probe is not plugged in, the secondary display shows
NaCl
gm
"25°C" with the "°C" indicator blinking.
To switch the instrument on, press and hold the ON/OFF key for a fraction of a second.
These meters have a built-in protection against elec­tromagnetic interferences and the holding of the keys is to ensure that the commands are not mis­taken for erroneous signals.
In order to take accurate measurements, make sure the instrument is calibrated before use (see "Calibration" sections for details).
Shift downwards the rubber sleeve (see picture) to have a properly work­ing junction.
°C
Immerse the tip of the FC300B elec­trode (at least 2.5 cm/1") into the sample to be tested while making sure that the electrode junction is completely immersed, and without touching the bottom of the beaker.
Wait for approximately 10 to 15 minutes to allow for the electrode to adjust and stabilize, i.e. until no drifting is observed for at least a couple of minutes.
During measurement, if possible, the sample should be gently stirred (100 rpm).
MAGNETIC STIRRER HI 190M
ON
POWER
OFF MAX
SPEED
98
TAKING pNa MEASUREMENTS (HI 931101 only)
Press the MODE key during measurement to con­vert the value displayed by the HI 931101 to pNa. Note that
pNa = -log a
Na
where aNa is the activity of ion Na.
°C
MODE
Press the MODE key again to return to g/L mea­surements.
TEMPERATURE MEASUREMENTS
Connect the temperature probe to the meter and turn on.
Dip the HI 7662 temperature probe into the sample. Allow the reading to stabilize.
If the electrode is used rarely (once a month or less), store it dry. Before using it again, let it condition overnight in HI 7081 (30.0 g/L NaCl) or HI
7086 (23 g/L Na 1M) To avoid very long response time, do not store
or rinse the electrode with deionized water. Do not rub the electrode.
Note: The temperature is always
displayed on the secondary
NaCl
gm
LCD.
AFTER USE
Turn the meter off by pressing the ON/OFF key again.
For faster and accurate measurements, the electrode should be stored in a solution with a Na/NaCl concen­tration similar to the sample to be measured.
After use, shift back upwards the rubber sleeve (see picture).
°C
1110
Na CALIBRATION (HI 931101)
Calibration can be performed at 1 (offset only) or 2 points.
For best accuracy, a 2 point procedure is recom­mended.
PREPARATION
• Pour small quantities of
HI 7080
HI7080 (2.3 g/L Na) solution and HI 7087 (0.23 g/L Na) or
HI 7080
HI 7087
HI 7086 (23 g/L Na) into two beakers.
For accurate readings, use for the second point the HI 7087 solution if you are going to measure samples with low sodium content (below 2.3 g/ L), or the HI 7086 standard if you are going to measure samples with high sodium content (over
2.3 g/L).
• For accurate calibration use two beakers for each solu­tion, the first one for rinsing
RINSE CALIBRATION
HI 7080 HI 7080
the electrode, the second one for calibration.
If possible, use plastic beakers to minimize any EMC interferences.
• Turn the meter on by press­ing the ON/OFF key.
ON/OFF
• Remove the protective cap and rinse the elec­trode with some of the calibration solution you are going to use first.
Shift downwards the rubber sleeve for a properly working junction (see picture).
SINGLE POINT CALIBRATION (OFFSET)
• Dip the FC300B sodium elec- trode into the HI7080 (2.3g/L Na) solution.
Note: Immerse the electrode
approximately 2.5 cm (1") into the solution.
• Press the CAL key. The "CAL" and "BUF" indi­cators will be displayed with the "BUF" and the " " symbols blinking.
CAL
• As soon as the standard so­lution is recognized, only the " " symbol continues blinking to indicate that the electrode is stabilizing. The required standard solution be­ing value is displayed on the primary LCD.
• As soon as the electrode sta­bilized, the "CON" symbol ap­pears on the display.
Press the CFM key to save the calibration point. For best results, wait for a couple of minutes before confirming.
HI 931101 will display "0.230
CFM
gm".
gm
CAL
BUF
°C
1312
• For exiting the calibration mode and save the single point cali­bration, press the CAL key. For better accuracy however, it is recommended to perform a 2 point calibration.
TWO POINT CALIBRATION
CAL
Wait for a couple of minutes and then press the CFM key.
CFM
• The two point calibration procedure is now com­pleted and the meter automatically returns to the normal measurement mode.
• After the first point is confirmed, immerse the FC300B electrode into the second standard solu­tion: HI 7087 at 0.23 g/L Na or
HI 7086 at 23 g/L Na. Note: The electrode should be
immersed approximately
2.5 cm (1") into the solu­tion.
Select the calibration value by pressing the BUF key. Make sure that the selected value cor­responds to the solution being used for the second calibration point.
• If the reading is not close to the selected calibration standard, the "BUF" indicator will continue to blink.
Otherwise, if the reading is close to the selected calibration value, the meter will recognize the stan­dard solution and only the " " symbol will blink, to indicate that the reading is stabilizing.
BUF
NaCl CALIBRATION (HI 931100)
Calibration can be performed at 1 (offset only) or 2 points.
For best accuracy, a two point procedure is recom­mended.
PREPARATION
• Pour small quantities of HI7083 (3.00 g/L NaCl) solu­tion and HI 7085 (0.30 g/L) or HI 7081 (30 g/L) into two clean
HI 7083
HI 7083
HI 7085
beakers. For more accurate readings use for the second
poit the HI 7085 solution if you are going to measure samples with low sodium chloride con­tent (< 3 g/L), or the HI 7081 solution for samples with high sodium chloride content (> 3 g/L).
• For an accurate calibration, use two beakers for each solution, the first one for rinsing the electrode, the sec­ond one for calibration.
If possible use plastic beakers to minimize any EMC interferences.
• Turn the meter on by press­ing the ON/OFF key.
Once stabilized "CON" symbol will appear on the display.
Remove the protective cap and rinse the elec­trode with some of the calibration solution you are going to use first.
Shift downwards the rubber sleeve for a properly working junction (see picture).
1514
SINGLE POINT CALIBRATION (OFFSET)
• Dip the FC300B sodium elec­trode into the HI7083 (3.00 g/L NaCl) solution.
Note: Immerse the electrode
approx. 2.5 cm (1") into the solution.
Press the CAL key. The "CAL" and "BUF" indi­cators will be displayed with the "BUF" and the " " symbols blinking.
CAL
• As soon as the standard solu­tion is recognized, only the " " symbol will continue blinking to indicate that the elec­trode is stabilizing.
• To quit the calibration and save the first point, press the CAL key.
For better accuracy however, it is recommended that a two-point calibration is performed.
TWO POINT CALIBRATION
• After the first point is confirmed, immerse the FC300B electrode into the second solution:
HI7085 at 0.3 g/L NaCl or HI7081 at 30 g/L NaCl.
Note: The electrode should be
immersed approximately
2.5 cm (1") into the so­lution.
• Select the second point value by pressing the BUF key. Make sure that the selected value cor­responds to the one being used for the second calibration point.
• If the reading is not close to the selected value, the "BUF" symbol will blink.
CAL
BUF
The value of the standard solu­tion being used will be displayed on the primary LCD.
• As soon as the electrode is sta­bilized, "CON" symbol lights up.
For best results, wait for a couple of minutes before pressing the CFM key.
HI 931100 will then display "30.0 gm".
Otherwise, if the reading is close to the selected value, the meter will recognize the stan­dard solution. Only the " " symbol will blink, to indicate that the reading is stabilizing.
Once stabilized "CON" symbol will appear on the display.
CFM
For best results, wait for a couple of minutes and press
NaCl
CAL
BUF
gm
°C
the CFM key.
• The two point calibration is now completed and the meter re-
CFM
turns to the normal mode.
1716
N
SPECIAL APPLICATIONS &
ADDITIONAL INFORMATION
HOW TO MEASURE SODIUM CHLORIDE IN CHEESE AND MEAT (HI 931100)
1. Grind a sample of cheese or meat.
MAGNETIC STIRRER HI 190M
FILTERED
SOLUTIO
ON
POWER
OFF MAX
SPEED
2. Pour a quantity (X g) in a glass beaker. Add (X) x 10 mL of deionized water (note: ignoring water content of cheese and meat).
3. Cover the beaker to prevent any spilling during stirring, place the beaker on a hot plate stirrer and agitate the sample at 40-50°C (104-122°F) for 15-20 minutes.
4. Let the sample cool to the ambient temperature, then filter it.
5. Dip the FC 300B sodium electrode into the fil- tered sample. The tip of the electrode should be immersed for approximately 2.5 cm/1" (i.e. the electrode junction must be completely immersed) without touching the bottom of the beaker. Dur­ing measurement, the sample should be gently stirred (100 rpm).
6. Wait for the reading to stabilize, i.e. no drift should be observed for at least a couple of minutes. Note the measured value.
7. Since the dilution rate is 1/10, the reading in the g/L range has to be divided by 100 to get the actual reading of NaCl (per gram of cheese or meat), or the value can be considered directly expressed in % of NaCl.
Note: A response time of 10 to 15 minutes for the
first measurement of cheese or meat, is the correct expected time for the electrode to stabilize. Next measurements of cheese or meat samples will require shorter response time (less than 2 minutes). Do not rinse the sodium electrode between subsequent mea­surements since contamination is negligible.
IONIC STRENGTH ADJUSTER (ISA)
When the NaCl concentration of the sample is lower than 5 g/L, it is recommend to use the HI 7090 ISA solution. In such case, calibration should be per­formed with standard solution containing 2% of ISA.
ISA (Ionic Strength Adjuster) = 50% of NH4Cl 4M + 50% of NH4OH 4M (use 2 mL of ISA every 100 mL of sample).
1918
TEMPERATURE
The best working temperature is around 20-25°C (68-77°F).
TEMPERATURE CALIBRATION
(for Technical Personnel only)
For greatest accuracy, it is recommended to cali­brate the meter with standard solution at a tempera­ture as close as possible to the temperature of the sample (±1°C/±2°F).
Na CONCENTRATION WITH HI 931100
It is always possible to determine the Na concentra­tion by using the formula:
XNa = X
• (23 / 58.4)
NaCl
where
XNa = concentration of Na
and
X
= concentration of NaCl
NaCl
INITIAL PREPARATION
Prepare a 0°C/32°F bath with ice and water and another one at 50°C/122°F using hot water at a known temperature. Place insulation material around the containers to minimize temperature changes.
Use a
Checktemp
or another accurate thermometer
with a resolution of 0.1° as reference.
0 °C
(32 °F)
°C
50 °C
(122 °F)
°C
Note: A blinking "°C" indicates
that the temperature probe
NaCl
is not connected.
PROCEDURE
• Connect the HI 7662 temperature probe to the RCA socket on the top of the instrument.
Turn the instrument off.
gm
°C
To enter the temperature calibration mode, press the CAL and ON/OFF keys simultaneously.
Immerse the temperature probe
°C
into the 0°C/32°F bath together with the reference thermom­eter.
0°C
(32°F)
When the reading stabilizes, press the CAL key. The "0.0°C", "CAL" and "BUF" symbols will be displayed. The
CAL
BUF
°C
" " symbol will blink for a few seconds.
2120
When the input value be­comes stable, the " " symbol disappears and
CON
°C
"CON" will start blinking. Press the CFM key to confirm the calibration
value.
°C
Immerse the temperature probe into the 50°C (122°F) bath.
50°C
(122°F)
Press the CAL key to dis­play the "0.0°C" and then press the BUF key to se­lect "50.0°C". "CAL" and
CAL
BUF
°C
"BUF" symbols. " " symbol will blink for a few seconds.
When the input value be­comes stable, the " " symbol disappears and
CON
°C
"CON" will start blinking.
Press the CFM key to confirm the calibration value.
• The calibration is now complete.
• Turn the meter off and on again to return to the normal mode.
FC300B CONDITIONING & MAINTENANCE
PREPARATION
Remove the protective cap. DO NOT BE ALARMED IF ANY
SALT DEPOSITS ARE PRESENT. This is normal with sodium electrodes and they will disappear when rinsed with water.
During transport tiny bubbles of air may have formed inside the glass bulb. The electrode cannot function properly under these conditions. These bubbles can be removed by "shaking down" the electrode as you would do with a glass thermometer.
If the bulb and/or junction are dry, soak the elec­trode in HI 7081 (NaCl
30.0 g/L) solution over­night before use with
HI931100. Soak it in HI7086 (Na 23 g/L) solu-
tion overnight before use with HI931101.
If the electrolyte solution is more than 1cm (½") below the filling hole, add HI 8093 1M KCl+AgCl electrolyte solution.
Shift downwards the rub­ber sleeve during mea­surements and calibra­tion.
SODIUM MEASUREMENT
Rinse the sodium electrode tip with some of the sample to be tested, then immerse the tip (approx. 2.5 cm /1") in the sample and stir gen­tly.
2322
The sensitive bulb should not touch the bottom of the beaker.
The Hanna HI 76405 electrode holder can be used for this purpose.
During measurement, the sample should be gently stirred (approx. 100 rpm). The Hanna HI190M stirrer can be used for this purpose.
MAGNETIC STIRRER HI 190M
ON
POWER
OFF MAX
SPEED
STORAGE
When used frequently, the electrode should be stored in a solution with a Na or NaCl concentration similar to the sample to be measured. For longer periods of storage, the electrode should be sored with a few drops of appropriate solution in the protective cap.
For frequent use in different samples (1 or 2 times per week), store the electrode in HI7081 (30.0 g/L NaCl) solution if used for NaCl measurements with HI 931100.
Store the electrode in HI 7080 (2.3 g/L Na) solution if used for Na measurements with HI931101.
For occasional use (once a month or less) store the electrode dry and let it conditioning overnight in HI7081 (30.0 g/L NaCl) or HI7086 (23 g/L Na) solution before use.
Note: NEVER STORE OR RINSE THE ELECTRODE
IN DISTILLED OR DEIONIZED WATER.
PERIODIC MAINTENANCE
Inspect electrode and cable. The cable used for connection to the meter must be intact and there must be no points of broken insulation on the cable or cracks on the electrode stem or bulb. Connec­tors must be perfectly clean and dry.
If any scratches or cracks are present, replace the electrode.
Rinse off any salt deposits with water. Refill the electrode with fresh electrolyte (HI 8093)
and let it stand upright for 1 hour. Follow the Storage Procedure above.
CLEANING PROCEDURE
General Soak in Hanna HI 7061 general clean-
ing solution for approximately 1 hour.
Removal of films, dirt or deposits on the membrane/ junction:
-
Protein
Soak in Hanna HI 7073 protein clean­ing solution for 15 minutes.
-
Inorganic
Soak in Hanna HI 7074 inorganic clean­ing solution for 15 minutes.
-
Oil/grease
Rinse with Hanna HI 7077 or Oil & Fat cleaning solution.
IMPORTANT: After performing any of the clean­ing procedures rinse the electrode thoroughly with a solution containing 2% ISA, refill (if necessary) the reference chamber with fresh electrolyte (HI8093) and soak the electrode in
a solution very close to the next sample to be measured for at least 1 hour before taking
measurements.
TROUBLESHOOTING
Evaluate the performance of your electrode based on the following.
• Noise (readings fluctuate up and down) could be due to:
- Clogged/Dirty Junction: refer to the Cleaning
Procedure above.
- Loss of shielding due to low electrolyte level:
refill with HI 8093 solution
• Dry Membrane/Junction: soak it overnight in HI7081 (30.0 g/L NaCl) or HI7086 (23 g/L Na)
solution before use.
• Drifting: soak the electrode tip in warm HI8093 solution for one hour. Refill with fresh HI 8093 electrolyte solution.
• Low Slope: refer to the Cleaning Procedure.
• No Slope: check the electrode for cracks in
glass stem or bulb and, if necessary, replace the electrode.
• Slow Response/Excessive Drift: soak the tip in HI 7061 solution for 30 minutes, rinse thor- oughly in a solution with 2% of ISA, and then follow the Cleaning Procedure.
2524
BATTERY REPLACEMENT
ACCESSORIES
If the batteries become weak the LCD will display the "LOW BAT" indication.
LOW BAT
Battery replacement must only take place in a safe area and using the battery type specified in this instruction manual.
To replace batteries, remove the two screws on the rear cover and replace all four 1.5V AA batteries with new ones, while paying attention to the correct polarity. Replace the cover and tighten the two screws.
Na SOLUTIONS FOR HI931101
HI 7080M 2.3 g/L Na solution, 230 mL bottle HI 8080M 2.3 g/L Na solution, 230 mL FDA bottle HI 7080L 2.3 g/L Na solution, 500 mL bottle HI 8080L 2.3 g/L Na solution, 500 mL FDA bottle HI 7086M 23 g/L Na solution, 230 mL bottle HI 8086M 23 g/L Na solution, 230 mL FDA bottle HI 7086L 23 g/L Na solution, 500 mL bottle HI 8086L 23 g/L Na solution, 500 mL FDA bottle HI 7087M 0.230 g/L Na solution, 230 mL bottle HI 8087M 0.230 g/L Na solution, 230 mL FDA bottle HI 7087L 0.230 g/L Na solution, 500 mL bottle HI 8087L 0.230 g/L Na solution, 500 mL FDA bottle
NaCl SOLUTIONS FOR HI 931100
HI 7081M 30 g/L NaCl solution, 230 mL bottle HI 8081M 30 g/L NaCl solution, 230 mL FDA bottle HI 7081L 30 g/L NaCl solution, 500 mL bottle HI 8081L 30 g/L NaCl solution, 500 mL FDA bottle HI 7083M 3.0 g/L NaCl solution, 230 mL bottle HI 8083M 3.0 g/L NaCl solution, 230 mL FDA bottle HI 7083L 3.0 g/L NaCl solution, 500 mL bottle HI 8083L 3.0 g/L NaCl solution, 500 mL FDA bottle HI 7085M 0.3 g/L NaCl solution, 230 mL bottle HI 8085M 0.3 g/L NaCl solution, 230 mL FDA bottle HI 7085L 0.3 g/L NaCl solution, 500 mL bottle HI 8085L 0.3 g/L NaCl solution, 500 mL FDA bottle HI 7090M ISA solution, 230 mL bottle HI 8090M ISA solution, 230 mL FDA bottle HI 7090L ISA solution, 500 mL bottle HI 8090L ISA solution, 500 mL FDA bottle
ELECTRODE MAINTENANCE SOLUTIONS
HI 7061M General cleaning solution, 230 mL bottle HI 8061M General cleaning solution, 230 mL FDA
bottle
HI 7061L General cleaning solution, 500 mL bottle
2726
HI 8061L General cleaning so luti on, 500 mL FDA
bottle
HI 7073M Protein cleaning solution, 230 mL bottle HI 8073M Protein cleaning solution, 230 mL FDA
bottle
HI 7073L Protein cleaning solution, 500 mL bottle HI 8073L Protein cleaning solution, 500 mL FDA
bottle
HI 7074M Inorganic cleaning solution, 230 mL bottle HI 7074L Inorganic cleaning solution, 500 mL bottle HI 7077M Oil & Fat cleaning solution, 230 mL bottle HI 8077M Oil & Fat cleaning solution, 230 mL
FDA bottle
HI 7077L Oil & Fat cleaning solution, 500 mL bottle HI 8077L Oil & Fat cleaning solution, 500 mL
FDA bottle
HI 8093 1M KCl + AgCl electrolyte solution, 4 x
50 mL bottle
OTHER ACCESSORIES
HI 98501
ChecktempC
electronic thermometer with penetration, stainless steel probe and 0.1°C resolution (range: -50.0 to
150.0°C)
FC 300B Sodium electrode, glass body, single
junction, refillable, with BNC connec­tor and 1 m (3.3') cable
HI 190M/D Mini-stirrer 230 V (1 liter max stirring
capacity, speed range from 100 to max 1000 rpm)
HI 190M/U Mini-stirrer 115 V (1 liter max stirring
capacity, speed range from 100 to max 1000 rpm)
HI 710031 Rugged carrying case HI 7662 Stainless steel temperature probe with
1 m (3.3') cable
HI 76405 Electrode holder
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workmanship and mate-
rials when used for their intended purpose and main­tained according to instructions.
Electrodes and probes are warranted for six months.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure.
If the repair is not covered by the warranty, you will be notified of the charges incurred.
If the instrument is to be returned to Hanna Instru­ments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid.
When shipping any instrument, make sure it is prop­erly packaged for complete protection.
Hanna Instruments reserves the right to modify
the design, construction and appearance of its
products without advance notice.
2928
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24 Vac or 60 Vdc.
To avoid damages or burns, do not perform any measurement in microwave ovens.
3130
SALES & TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and Technical offices, please see www.hannainst.com
MANNACLR2 08/05
32
Loading...