Hanna Instruments HI 93102 User Manual

Instruction Manual
HI 93102
Multiparameter
Turbidity &
Ion Specific Meter
www.hannainst.com
Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully
before using the instrument. This manual will provide you with the necessary
information for correct use of the instrument, as well as a precise idea of its versatility. If you need more technical information, do not hesi­tate to e-mail us at tech@hannainst.com.
This instrument is in compliance with direc­tives.
TABLE OF CONTENTS
General Description .................................... 4
Principle of Operation .................................. 6
Functional Description................................. 9
PRELIMINARY EXAMINATION
HI 93102 is supplied complete with:
Glass cuvet with cap
Batteries (4 x 1.5V AA)
Instruction manual
The HI 731327 complete maintenance kit is also available, including:
2 glass cuvets
Primary calibration standards: HI 93102-0 AMCO-AEPA-1 0 NTU* cali-
bration solution, 30 mL HI 93102-20 AMCO-AEPA-1 20 NTU* cali-
bration solution, 30 mL
• HI 93703-50 cleaning solution, 230 mL
Tissue for wiping cuvets
Rugged carrying case
Specifications ........................................... 13
Operational Guide ..................................... 14
Cali bra tio n................................................. 22
Diagnostic Mode ....................................... 3 0
Logging with C 102.................................... 31
User-selectable Auto-off ............................ 33
Battery Replacement ................................ 34
Diagnostic Codes...................................... 35
Accessories.............................................. 36
Warranty ................................................... 38
CE Declaration of Conformity .................... 39
2
Note: Save all packing material until you are
sure that the instrument functions cor­rectly. Any defective item must be re­turned in its original packaging with the supplied accessories.
* 1 NTU (Nephelometric Turbidity Unit) = 1 FTU (For-
mazine Turbidity Unit)
3
GENERAL DESCRIPTION
The Hanna HI 93102 is a portable micropro­cessor driven, multiparameter, turbidity and ion specific meter. It measures Free & Total Chlorine, Cyanuric Acid, pH, Iodine, Bromine, Low Range Iron and Turbidity.
In the colorimetric mode, the user can select either factory preprogrammed calibration set­tings or calibrate the meter using customized calibration values based on the concentration or relative absorbance of the sample. Calibra­tion data is also stored in a non-volatile EEPROM.
In the turbidity mode, periodic recalibration of the meter with primary standards according to regulatory requirements or personal experi­ence is suggested. Turbidity ranges are 0.00-
9.99 NTU and 10.0-50.0 NTU. HI 93102 complies with G.L.P. Standards
(Good Laboratory Practice), that is:
• When switched on, the LCD displays all segments (display check).
• Battery status is monitored during every measurement cycle warning the user if the batteries become weak.
In addition, HI 93102 will turn itself off before low voltage causes erroneous read­ings.
• It utilizes a real time clock and recalls calibration data such as date, time and calibration values.
There are eight keys for the different operationals modes. The large Liquid Crystal Display is dual-level: the upper level has four digits and can display the measured param­eter in hundredths. The lower level has three characters and indicates current mode (e.g. F CL for free chlorine or TR for turbidity). Different LCD segments indicate low battery, logging mode, date, time, etc.
A pure green LED has been utilized as a light source for both turbidimetric and colorimetric measurements. A silicon photocell is used to receive transmitted light from colorimetric chan­nel while another photocell receives scattered light from the turbidimetric (nephelometric) channel.
In order to measure colorimetric parameters, all the operator has to do is zero the blank sample and then add 1 packet of reagent (for Bromine, Chlorine, Cyanuric Acid, Iodine and Low Range Iron) or 0.2 mL of Phenol Red (for pH). After placing the cuvet back in the meter and pressing READ, the measurements are shown directly on the LCD.
The instrument operates with four AA batter­ies and may be operator-programmed to turn itself off automatically after 10, 20, 30, 40, 50 or 60 minutes of inactivity.
HI 93102 and all accessories such as sample vials, reagent pillows, primary standards, can be easily stored in the optional carrying case.
To facilitate field tests, the meter provides a logging mode. In this mode, the user can store up to twenty five time-tagged measure­ments in RAM and scroll the memory at any time.
4
5
PRINCIPLE OF OPERATION
Turbidity Mode
surements according to the USEPA’s 180.1 method and the Standard Method 2130B.
The instrument functions by passing a beam of light through a vial containing the sample being measured.
The light source is a Pure Green LED to ensure that any interference caused by a colored samples is minimized.
A sensor, positioned at 90° with respect to the direction of light, detects the amount of light scattered by the undissolved particles present in the sample. The microprocessor converts such readings into NTU* values.
Colorimetric Mode
The color of every object we see is deter­mined by a process of absorption and emis­sion of the electromagnetic radiation (light) of its molecules.
Colorimetric analysis is based on the principle that specific compounds react with others to form a color, the intensity of which is proportional to the concentration of the substance being measured.
LED
EMITTED LIGHT
CUVET MICROPROCESSOR
Block diagram of an ion specific measurement
LIGHT
DETECTOR
When a substance is exposed to a beam of light intensity Io, a portion of the radiation is absorbed by the substance's molecules and a radiation of intensity I, lower than Io, is emitted.
The quantity of radiation absorbed is given by the Lambert-Beer Law:
log Io/I = ελ c d
NTU units are equal to FTU units. However, there are other known measurement units for turbidity, namely the Jackson Turbidity Unit (JTU) based on the old method of Jackson's candle, and Silica Unit (mg / L of SiO2). The conversion table between these measurement units is shown below:
JTU NTU/FTU SiO2 (mg/L)
JTU 1 19 2.5
NTU/FTU 0.053 1 0.13
SiO2 (mg/L) 0.4 7.5 1
6
* 1 NTU = 1 FTU
Where log Io/I = Absorbance (A)
ελ= molar extinction coefficient of the
substance at wavelength λ
c = molar concentration of the
substance
d = optical distance light travels
through the sample
Since other factors are known, the concen­tration "c" can be calculated from the color intensity of the substance determined by the emitted radiation I.
An LED (Light Emitting Diode) emits radiation at a relatively narrow spectrum, supplying the system with the intensity Io.
7
A substance absorbs a color complimentary to the color it emits. For example, a substance appears yellow because it absorbs blue light. As a result, the Hanna meters use LED’s with specific wavelengths to measure samples.
The optical distance (d) is measured by the internal diameter of the cuvet containing the sample.
The photoelectric cell collects the radiation I that is not absorbed by the sample and converts it into an electric current.
The microprocessor converts the value into the desired measuring unit and displays it on the LCD.
The measurement process is done in two phases: setting the meter to zero and actual measurement.
The cuvet is an optical element and hence has an important role in the measurement process. Both the measurement and the calibration cuvets must be optically identical to provide the same measurement conditions.
It is also important that the surface of the cuvet is clean and free from scratches or dents, in order to avoid measurement interference due to unwanted reflection and absorption of light.
It is recommended that wherever possible the cuvet walls are not touched by the operator.
Furthermore, in order to maintain the same conditions during the zeroing and the measuring phases, it is necessary to close the cuvets to prevent any contamination.
8
FUNCTIONAL DESCRIPTION
1) Primary LCD: The four-digit LCD shows
all segments for several seconds when the meter is switched on. It then dis­plays four dashes to indi­cate “ready to measure”. It is also the area where the date, time and value of last calibration are shown. In “Read” and “Zero” mode, “SIP” is shown to indicate “Sample
9
In Progress”. The upper level also indicates the concentration or turbidity of the sample, as well as different diagnostic modes, such as “-BA-” for low bat­tery.
2) Secondary LCD: The three-digit LCD shows
the current mode of mea­surement that is “F CL”, “tCL”, “CY”, “PH”, “Id”, “Br”, “L FE”, “tr”, and di­agnostic or calibration modes, such as “d11”, “2 Fn”, “5c1”.
3) DATE: Indicates that the upper
level of LCD is showing the current date, the date of last calibration or the date of logged measure­ment in memory.
4) TIME: Indicates that the upper
level of LCD is showing the current time, the time of last calibration or the time of logged measure­ment in memory.
5) LOBAT: Blinking segment warns
user of low battery volt­age.
6) LOG: If intermittent, it indicates
that the user is in the scroll mode viewing the logged measurements. If fixed, it indicates that the meter is in the log mode and every reading taken will be stored in memory.
7) ON/OFF key: Turns the meter on and off.
8) ZERO / key: In ion specific (colorimet­ric) mode, it zeros the sample. In calibration and diagnostic modes, it func­tions as ENTER (not used in turbidity mode).
9) READ/ key: Takes the measuremnt of concentration/turbidity of the sample which is shown on the LCD. In diagnostic or calibration mode, shifts the flashing digit to the right.
10) CAL key: If pressed during calibra- tion, the calibration proce­dure will be aborted and the last calibration data will be reinstated. If pressed together with the ALT key for less then 3 seconds, the diagnostic mode will be entered. If pressed to­gether with the ALT key again, the meter will quit diagnostic mode. If pressed for more then 3 seconds, an intermittent “CAL” prompt will appear on the upper LCD level and the calibration proce­dure is entered.
11) GLP/Abs key: In ion specific mode, it will toggle concentration/ab­sorbance readings on the upper LCD. In turbidity mode, date, time and the two calibration values of the current mode will be
10
11
shown. If pressed in time/date setup mode, the meter will quit current mode without making any changes to current time/ date.
12) key Scrolls upwards through the parameters to be mea­sured. In calibration/diag­nostic mode, increments the blinking digit by one. If pressed together with ALT while the meter is in logging mode, the upper LCD will show the data (date/time/value) in the memory.
13) ALT key: Alternative functions.
14) key: Scrolls downwards through the parameters to be mea­sured. In calibration/diag­nostic mode, decreases the blinking digit by one. If pressed together with ALT while the meter is in logging mode, the upper LCD will show the current lot number.
SPECIFICATIONS
HI 93102
Range Turbidity 0.00 to 50.0 NTU*
Bromine 0.00 to 8.00 mg/L
Free Cl
2
Total Cl
2
CYS 0 to 80 mg/L
Iodine 0.0 to 12.5 mg/L
LR Iron 0.00 to 1.00 mg/L
pH 5.9 to 8.5
Resolution
Turbidity 0.01 and 0.1 NTU*
Bromine 0.01 mg/L
Free Cl
2
Total Cl
2
CYS 1 mg/L
Iodine 0.1 mg/L
LR Iron 0.01 mg/L
pH 0.1
Accuracy
Turbidity ±0.5 NTU* or ±5% (whichever greater)
Bromine ±0.08 mg/L; ±3%
Free Cl
2
Total Cl
2
CYS ±1 mg/L; ±15%
Iodine ±0.1 mg/L; ±5%
LR Iron ±0.02 mg/L; ±3%
pH 0.1 pH Light Source Pure Green LED Light Source Life Life of the instrument Light Detector Two Silicon Photocells Power Source 4 x 1.5V AA alkaline batteries Battery Life 60 hours or 1000 measurements Auto-off Selectable after 10, 20, 30,
Environment 0 to 50°C (32 to 122°F)
Dimensions 220 x 82 x 66 mm
Weight 510 g (1.1 lb.)
0.00 to 2.50 mg/L
0.00 to 3.50 mg/L
0.01 mg/L
0.01 mg/L
±0.03 mg/L; ±3% ±0.03 mg/L; ±3%
40, 50 or 60 min. of non-use
RH max 95% (non-condensing)
(8.7 x 3.2 x 2.6")
12
13
* 1 NTU = 1 FTU
Loading...
+ 14 hidden pages