Hanna Instruments HI8936A, HI8931B, HI8936B, HI8936C, HI8936D Instruction Manual

...
Instruction Manual
HI8931A/B/C/D HI8936A/B/C/D
HI8936AL/BL/CL/DL
HI943500A/B/C/D
Conductivity
Process Instrumentation
This Instrument is in Compliance with the CE Directives
WWW.HANNAINST.COM
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual care­fully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. These instruments are in compliance with the directives EN 50081-1 and EN 50082-
1.
TABLE OF CONTENTS
Preliminary Examination........................... 3
General Description .................................. 3
Functional Description HI8931/HI 943500 . 5
Functional Description HI8936 ................. 9
Conductivity Probes ................................ 12
Specifications HI8931 & HI943500 ......... 14
Specifications HI8936 ............................. 15
Connections ............................................ 16
Operational Guide ................................... 24
Calibration Procedure of
HI8931 & HI8936 with HI7635................ 30
Calibration Procedure of
HI8931 & HI8936 with HI7638................ 34
Calibration Procedure of
HI943500 with HI7638 ............................ 38
Conductivity Versus
Temperature Chart ................................. 41
Diagnostic Tests ..................................... 42
Installation Examples .............................. 44
Probe Maintenance and Cleaning .......... 46
Accessories ............................................. 47
Warranty ................................................. 50
CE Declaration of Conformity ................ 51
ISO 9000 Certified Company since 1992
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer.
Note: Save all packing materials until you
are sure that the instrument functions correctly. All defective items must be returned in the original packing ma­terials together with the supplied ac­cessories.
GENERAL DESCRIPTION
The HI 8931 and HI 943500 are panel mounted conductivity meters and control­lers designed for simplicity of use in a wide range of industrial process applications.
The instruments are designed with a stan­dard DIN panel mount with membrane key­pads on the front, a large LCD and with auto-diagnostic functions.
Probes, power supply, contacts and record­ers are connected on the rear panel through screw terminals.
Using HI 8931 in conjunction with a 4-20 mA output transmitter HI 8936 series (or HI 8936L series, with LCD display) will assure you of a strong, interference-free signal at distances up to 300 meters (1000').
For in-line applications the probe should be HI 7635 and for tanks HI 7638 with external threads. These probes with built-in NTC sen­sor, provide temperature compensated con­ductivity measurements. The probe cable length is 3 meters (10').
32
HI 943500 features a direct connection up to 20 m (67'), without intermediate amplifi­ers, to the conductivity probe HI 7638 with 7-pin DIN connector and automatic tempera­ture compensation.
Four models with different measurement ranges are available to suit individual appli­cations:
HI8931A / HI943500A
HI8936A / HI8936AL
from 0.0 to 199.9 mS/cm
HI8931B / HI943500B
HI 8936B / HI8936BL
from 0.00 to 19.99 mS/cm
FUNCTIONAL DESCRIPTION
HI 8931 & HI 943500
HI8931C / HI 943500C
HI 8936C / HI8936CL
from 0 to 1999 µS/cm
HI8931D / HI 943500D
HI 8936D / HI8936DL
from 0.0 to 199.9 µS/cm
Other features include: recorder output in 0 to 20 mA or 4 to 20 mA configuration; LED indicators (for HI8931 and HI943500) which identify whether the controller is in opera­tion mode or selection mode.
Each instrument is supplied with a plastic front cover and two mounting brackets. Power cables are excluded.
Transmitters and conductivity probes are supplied separately.
KEYPAD
MEASURE To read the measurement values
and to enable the diagnostic tests
∆∆
AL To display the set tolerance of
∆∆
the alarm
SET To set the working point TEST SLOPE Diagnostic function TEST OFFSET Diagnostic function
When each key is pressed the LED is lighted indicating to the user that the function is in operation.
TRIMMERS
OFFSET For offset calibration SLOPE For slope calibration
∆∆
AL To set the alarm tolerance
∆∆
COARSE To coarsely adjust the working point SET FINE To finely adjust the working point
LEDS
SET ON To show that the dosage is active ALARM To show that the controller is in
alarm mode.
54
THE REAR PANEL OF HI 8931A/B/C/D
THE REAR PANEL OF HI 943500A/B/C/D
1. ALARM CONT Terminals for connec­tion to an external alarm device
2. SET CONT Terminals (Max. 2A, 220V) for connection to a dosing pump
3.4. +20V and COMM Terminals for con­nection to a conductivity transmitter
HI8936
5. SET SELECT Terminal for reverse con­trol operation
6. 0-20/4-20mA Terminals for connection to a recorder
7. Fuse Holder
8. Power Supply
9. IN/OUT TRANSMITTER Terminals for connection to a conductivity transmitter.
Unplug the instrument from power supply before the replacement of the fuse.
1. Conductivity Probe Connector (7-pin DIN)
2. mA OUTPUT Terminals for connection to a recorder
3. SET SELECT Terminal for reverse con­trol operation
4. SET Terminals for connection to a dos­ing pump
5. ALARM Terminals for connection to an external alarm device
6. Power Supply
7. Fuse Holder.
Unplug the instrument from power supply before the replacement of the fuse.
76
MECHANICAL DIMENSIONS
OF HI8931 AND HI 943500
Front view of the panel-mounted unit
FUNCTIONAL DESCRIPTION
HI 8936
69mm
2.71"
72mm
2.83"
141mm
5.55"
144mm
5.67"
These dimensions show the cutout size for the installation.
Side view of the panel-mounted unit
144mm
5.67"
0.25/4mm
0.01/0.160"
ADJUSTABLE LOCATION BRACKET
HI 8936A
HI8936B HI8936C HI8936D
HI 8936AL
HI8936BL HI8936CL HI8936DL
135mm
5.31"
190mm MIN
7.50"
Adjustable location brackets (supplied with the meter) allow the indicator to slide into the cutout and will hold the unit securely in place. 190 mm (7.50") is the minimum amount of room required to install the indicator with the cables connected.
1. Back Cover
2. Top Plastic Cover
3. LCD Display (for HI8936AL, HI8936BL, HI8936CL, HI8936DL only)
4. Screws for fastening the top cover.
98
SIDE VIEW
HI 8936A HI8936B HI8936C HI8936D
HI 8936AL HI8936BL HI8936CL HI8936DL
MECHANICAL DIMENSIONS
OF HI8936A, HI 8936B, HI8936C,
HI8936D
165mm
6.50"
100mm
3.94"
HI 8614
4 x 4.3mm
90mm
3.54"
4 x 0.17"
155mm
6.10"
MECHANICAL DIMENSIONS
OF HI8936AL, HI 8936BL, HI8936CL,
HI8936DL
110mm
4.33"
1. Top Plastic Cover
2. Wire cable glands.
90mm
3.54"
24mm
0.94"
45mm
1.77"
4 x 4.3mm
4 x 0.17"
165mm
6.50" 126mm
5.00"
58mm
2.28"
110mm
100mm
4.33"
3.94"
HI 8614L
155mm
6.10"
1110
CONDUCTIVITY PROBES
HI 7635 In-line Conductivity Probe
HI7635 is a one piece, molded conductivity
probe with pipe threads (1" NPT) at both ends.
This allows the probe to attach to an in-line system, and to be used in conjunction with the HI8936 conductivity transmitter.
The HI7635 uses a 4-ring potentiometric measuring method. This method is highly accurate and requires very little maintenance.
The construction of the housing is rugged, fiber-reinforced polypropylene.
The maximum working pressure of this unit is 5 BAR (72.5psi).
Do not use in systems where the tempera­ture exceeds 80°C (176°F).
HI7635
HI7638 Tank Conductivity Probe
HI7638 conductivity probe combines the
proven 4-ring potentiometric method of mea­suring conductivity with the platinum sensor and stainless steel external thread.
This method incorporates a series of four platinum rings into the probe shaft and is highly accurate requiring very little mainte­nance.
The removable cover is made of Ultem which resists the harmful effect of most chemicals and can be unscrewed for quick and simple maintenance.
This probe can withstand temperatures of up to 120°C (248°F) and pressure of up to 5 bar (72.5 psi).
This probe is supplied complete with a 7-pin DIN connector.
HI 7638
®
142mm
5.60"
73mm
2.87"
52mm
2.04"
1" NPT
1312
SPECIFICATIONS HI 8931 & HI 943500
SPECIFICATIONS HI 8936
Range HI8931A - HI943500A HI8931B - HI943500B HI8931C - HI943500C HI8931D - HI943500D
Accuracy (@20°C/68°F)
Typical EMC Deviation
4-20mA INPUT from Transmitter HI8931A HI8931B HI8931C HI8931D HI943500
Conductivity Probe
Calibration
Temperature Compensation HI8931 HI943500
Readout
Recorder Output 1 Set Point Relay
1 Alarm Relay
Power Supply
Panel Cutout Environment Weight Enclosure
0.0 to 199.9 mS/cm
0.00 to 19.99 mS/cm 0 to 1999 µS/cm
0.0 to 199.9 µS/cm ±2% of Full Scale
excluding probe error
±2.5% of Full Scale
±0.4 mA
HI8936 A or AL (not included) HI8936 B or BL (not included) HI8936 C or CL (not included) HI8936 D or DL (not included)
------
HI7635 for in-line applications or
HI7638 for tank (not included)
Manual 2-point through offset
and slope trimmers
See transmitter HI8936
Automatic from 0 to 60°C with
2% as temperature coefficient(β)
4-digit LCD plus graphic
symbols
0 to 20 mA or 4 to 20 mA (isolated)
Isolated, 2A, Max. 240V,
resistive load, 1,000,000 strokes
Isolated, 2A, Max. 240V,
resistive load, 1,000,000 strokes
110/115V or 220/240V
(user selectable); 50/60 Hz
141 x 69 mm (5.6 x 2.7")
-10 to 50°C (14 to 122°F) 1 Kg (2.2 lb.)
DIN 43 700 144x72 mm
(5.7x2.8") in black anodized
aluminum. Front and back with
shockproof ABS plastic and a
transparent cover.
Range HI8936A & AL HI8936B & BL HI8936C & CL HI8936D & DL
Accuracy (@20°C/68°F)
Typical EMC Deviation
Conductivity Probe
Calibration
Temperature Compensation
Output
Power Supply HI8936A/B/C/D HI8936AL/BL/CL/DL
Protection Environment
Dimensions
Weight
0.0 to 199.9 mS/cm
0.00 to 19.99 mS/cm 0 to 1999 µS/cm
0.0 to 199.9 µS/cm ±2% of Full Scale
excluding probe error
±2% of Full Scale
±0.4 mA
HI7635 for in-line applications or
HI7638 for tank
(not included)
Manual 2-point through Offset
and Slope trimmers
Fixed or automatic from 0 to
50°C (32 to 122°F) with 2% as
temperature coefficient (β)
4 to 20 mA not-isolated
Max. 500 ohm
12 to 30 VDC 17 to 36 VDC
IP 65
0 to 50°C (32 to 122°F)
95% RH max
165 x 110 x 90 mm (LxWxH)
(6.5 x 4.3 x 3.5")
1 Kg (2.2 lb.)
1514
CONNECTIONS
REAR CONNECTIONS FOR HI8931
• Power Connection Terminals
(4-screw terminal strip) for connection to a 3­wire power cable ac­cording to the indicated voltage (115V or 230V).
• Set Select
These contacts permit the activation of a Set Contact relay when the measured value is lower (connected termi­nals) or higher (open terminals) than the user's set value. See also page 27.
• Alarm Contacts
• IN/OUT Transmitter
2 wires of the 4-core signal cable from the conductivity transmitter (HI8936) have to be connected to these ter­minals and the other 2 wires to the "+20V" and "COMM" paying care­ful attention to the po­larity.
+20V supply is the regulated DC supply required for the operation of the external Conductivity Transmitter HI8936.
• Set Contacts
Dosing pumps or other control equipment may be connected to the "SET CONT" (Max. 2A, 220V) terminals. These contacts act only as a "dry" switch allowing electrical continuity, not as a power supply.
During normal opera­tion these terminals re­main open.
If the measured con­ductivity level is not within the tolerance of the set value, the alarm contact is closed. These contacts act only as a switch. See also page 28.
• +4-20 mA
These are the output terminals for connec-
tion to a recorder or other control equip­ment.
The output current varies from 4 to 20 mA and is proportional to the measured con­ductivity value.
1716