Hanna Instruments HI83325 Instruction Manual

Page 1
HI83325
Nutrient Analysis Photometer
INSTRUCTION MANUAL
Page 2
Dear Customer,
Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for correct use of the
instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e‑mail us at
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
Page 3
1. PRELIMINARY EXAMINATION ............................................................................................
2. SAFETY MEASURES...................... ....................................................................................
3. SPECIFICATIONS ..............................................................................................................
4. DESCRIPTION..... .............................................................................................................
4.1. GENERAL DESCRIPTION.............................................................................................
4.2. PRECISION AND ACCURACY........................................................................................
4.3. FUNCTIONAL DESCRIPTION........................................................................................
4.4. PRINCIPLE OF OPERATION.........................................................................................
4.5. OPTICAL SYSTEM......................................................................................................
5. GENERAL OPERATIONS......... ............................................................................................
5.1. POWER CONNECTION AND BATTERY MANAGEMENT......................................................
5.2. GENERAL SETUP........................................................................................................
5.3. USING HANNA DIGITAL ELECTRODES...........................................................................
5.4. MODE SELECTION.....................................................................................................
5.5. LOGGING DATA.........................................................................................................
5.6. ADDING SAMPLE / USER NAMES TO LOG DATA............................................................
5.7. DATA MANAGEMENT..................................................................................................
5.8. CONTEXTUAL HELP....................................................................................................
6. NUTRIENT SAMPLES PREPARATION GUIDE......... ................................................................
6.1. INTRODUCTION TO PLANT NUTRIENTS........................................................................
6.2. PREPARING NUTRIENT SAMPLES FOR ANALYSIS..........................................................
6.3. PROCEDURE FOR DILUTION FACTOR: 5.......................................................................
6.4. PROCEDURE FOR DILUTION FACTOR: 10.....................................................................
6.5. PROCEDURE FOR DILUTION FACTOR: 50.....................................................................
6.6. REMOVING TURBIDITY AND COLOR.............................................................................
7. PHOTOMETER MODE......... ...............................................................................................
7.1. METHOD SELECTION.................................................................................................
7.2. COLLECTING AND MEASURING SAMPLES AND REAGENTS..............................................
7.2.1. PROPER USE OF SYRINGE...................................................................................
7.2.2. PROPER USE OF DROPPER...................................................................................
7.2.3. PROPER USE OF POWDER PACKET.........................................................................
7.3. CUVETTE PREPARATION.............................................................................................
7.4. TIMERS AND MEASUREMENT FUNCTIONS...................................................................
7.5. CHEMICAL FORMULA / UNIT CONVERSIONS................................................................
7.6. METER VALIDATION / CAL CHECK................................................................................
7.7. ABSORBANCE MEASUREMENTS..................................................................................
TABLE OF CONTENTS
5
5
6
7
7
7
8
9
10
11
11
11
14
14
15
15
16
17
17
17
20
22
23
25
27
28
28
29
29
29
29
30
31
32
32
33
3
Page 4
8. PROBE MODE................... ...............................................................................................
8.1. pH CALIBRATION......................................................................................................
8.2. pH CALIBRATION MESSAGES......................................................................................
8.3. pH MEASUREMENT...................................................................................................
8.4. pH MEASUREMENT MESSAGES / WARNINGS...............................................................
8.5. pH GLP.....................................................................................................................
8.6. pH ELECTRODE CONDITIONING AND MAINTENANCE.....................................................
9. METHOD PROCEDURES........ ............................................................................................
9.1. AMMONIA LOW RANGE..............................................................................................
TABLE OF CONTENTS
9.2. AMMONIA MEDIUM RANGE........................................................................................
9.3. AMMONIA HIGH RANGE.............................................................................................
9.4. CALCIUM..................................................................................................................
9.5. MAGNESIUM............................................................................................................
9.6. NITRATE...................................................................................................................
9.7. PHOSPHATE HIGH RANGE..........................................................................................
9.8. POTASSIUM.............................................................................................................
9.9. SULFATE..................................................................................................................
10. ERROR DESCRIPTIONS......... ..........................................................................................
11. STANDARD METHODS......... ...........................................................................................
12. ACCESSORIES................................................................................................................
12.1. REAGENT SETS.......................................................................................................
12.2. pH ELECTRODES.....................................................................................................
12.3. pH SOLUTIONS.......................................................................................................
12.4. OTHER ACCESSORIES..............................................................................................
13. ABBREVIATIONS............................................................................................................
RECOMMENDATIONS FOR USERS..........................................................................................
WARRANTY.........................................................................................................................
34
34
36
36
38
39
40
42
42
45
48
51
54
57
60
63
66
68
69
69
69
70
71
72
73
74
75
4
Page 5
1. PRELIMINARY EXAMINATION
Remove the instrument and accessories from the packaging and examine it carefully to make sure that no damage has occurred during shipping. Notify your nearest Hanna Customer Service Center if damage is observed.
Each HI83325 is supplied in carrying case with:
• Sample Cuvette and Cap (4 pcs.)
• Cloth for Wiping Cuvettes
Scissors
USB Cable
• 5 Vdc Power Adapter
• Instruction Manual
• Quality Certificate
• 100 mL plastic graduated beaker with cap
• 170 mL plastic graduated beaker
Note: Save all packing material until you are sure that the instrument works correctly.
Any damaged or defective item must be returned in its original packing material with the supplied accessories.
• 3 mL plastic pipette
• 5 mL graduated syringe
• 60 mL graduated syringe
• graduated cylinder
• spoon
• funnel
• filter paper
• demineralizer bottle for 10 L of water
• activated carbon for 50 tests
PRELIMINARY EXAMINATION
2. SAFETY MEASURES
•The chemicals contained in the reagent kits may be hazardous if improperly handled.
•Read the Safety Data Sheets (SDS) before performing tests.
• Safety equipment: Wear suitable eye protection and clothing when required, and follow instructions carefully.
• Reagent spills: If a reagent spill occurs, wipe up immediately and rinse with plenty of water. If reagent contacts skin, rinse the affected area thoroughly with water. Avoid breathing released vapors.
• Waste disposal: for proper disposal of reagent kits and reacted samples, contact a licensed waste disposal provider.
SAFETY MEASURESS
5
Page 6
3. SPECIFICATIONS
Measurement Channels
SPECIFICATIONS
Absorbance
pH
Temperature
Additional Specifications
3 x optical channels
1 x digital electrode channel (pH measurement) Range 0.000 to 4.000 Abs Resolution 0.001 Abs Accuracy ±0.003 Abs (at 1.000 Abs) Light Source light emitting diode Bandpass Filter Bandwidth 8 nm
Bandpass Filter Wavelength Accuracy ±1.0 nm
Light Detector silicon photocell Cuvette Types round, 24.6 mm diameter Number of Methods 12 Range ‑2.00 to 16.00 pH (± 1000.0 mV)* Resolution 0.01 pH (0.1 mV) Accuracy ±0.01 pH (±0.2 mV) (@ 25 °C / 77 °F) Temperature Compensation ATC (‑5.0 to 100.0 °C; 23.0 to 212.0 °F)*
Calibration
Electrode Intelligent pH / temperature electrode Range ‑20.0 to 120.0°C (‑4.0 to 248.0 °F) Resolution 0.1 °C (0.1 °F) Accuracy ±0.5 °C (±0.9 °F) (@ 25 °C / 77 °F) Logging 1000 readings (mixed photometer and electrode) Display 128 x 64 pixel B/W LCD with backlight USB‑A (Host) Functions mass‑storage host USB‑B (Device) Functions power input, mass‑storage device
Battery Life
Power Supply
Environment
2 points, eligible from 5 available buffers (4.01,
6.86, 7.01, 9.18, 10.01 pH)
> 500 photometer measurements, or 50 hours of
continuous pH measurement
5 Vdc USB 2.0 power adapter/type micro‑B connector
3.7 Vdc Li‑polymer rechargeable battery,
non‑serviceable
0 to 50 °C (32 to 122 °F);
0 to 95% RH, non‑serviceable
Dimensions 206 x 177 x 97 mm (8.1 x 7.0 x 3.8") Weight 1.0 kg (2.2 lbs.)
*Limits will be reduced to actual probe/sensor limits.
6
Page 7
4. DESCRIPTION
4.1.GENERAL DESCRIPTION HI83325 multiparameter photometer is compact and versatile meter with two measurement
modes: Absorbance and pH/ mV. Absorbance mode include CAL Check feature and 12 different methods that cover a wide variety of applications, making it ideal for both benchtop and portable operation.
• Digital electrode input for pH measurements
• Certified CAL Check cuvettes to confirm meter functionality
• Dual purpose micro‑USB flash drive
• Li‑polymer rechargeable battery
• Auto‑off
• Absorbance mode
• User and sample name entry
• GLP features
4.2.PRECISION AND ACCURACY
Precision is how closely repeated measurements are to one another. Precision is usually expressed as standard deviation (SD). Accuracy is defined as the closeness of a test result to the true value. Although good precision suggests good accuracy, precise results can be inaccurate. The figure explains these definitions. For each method, the accuracy is expressed in the related measurement section.
DESCRIPTION
7
Page 8
4.3. FUNCTIONAL DESCRIPTION
DESCRIPTION
1) Splash‑proof keypad
2) Liquid Crystal Display (LCD)
3) Indexing mark
4) Protective port covers
5) Light‑blocking cover panel
6) Cuvette holder
7) ON/OFF power button
8) 3.5 mm TRRS (jack) input for digital electrodes
9) Standard USB host connector for data transfer to a USB flash drive
10) Micro‑USB device connector for power or PC interface
8
Page 9
Keypad Description
The keypad contains 12 direct keys and 3 functional keys with the following functions:
Press the functional keys to perform the function displayed above them on the LCD.
Press to access the list of photometer methods.
Press to move up in a menu or a help screen, to increment a set value, or to access second level functions.
Press to toggle between photometer and pH (electrode) mode.
Press to move left in a menu or to decrement a set value.
Press to move down in a menu or a help screen, to decrement a set value, or to access second level functions.
Press to move right in a menu or to increment a set value.
Press to access the setup screen.
Press to log the current reading.
Press to review saved logs.
DESCRIPTION
Press to exit the current screen.
Press to display the help screen.
ON/OFF power button.
4.4. PRINCIPLE OF OPERATION
Absorption of light is a typical phenomenon of interaction between electromagnetic radiation and matter. When a light beam crosses a substance, some of the radiation may be absorbed by atoms, molecules or crystal lattices. If pure absorption occurs, the fraction of light absorbed depends both on the optical path length through the matter and on the physical‑chemical characteristics of the substance according to the Lambert‑Beer Law:
‑log I/I
= el c d
o
or
c d
A = e
l
9
Page 10
= intensity of incident light beam
I
o
I = intensity of light beam after absorption e
l
= molar extinction coefficient at wavelength l c = molar concentration of the substance d = optical path through the substance
DESCRIPTION
Therefore, the concentration "c" can be calculated from the absorbance of the substance as the other factors are constant. Photometric chemical analysis is based on specific chemical reactions between a sample and reagent to produce a light‑absorbing compound.
4.5. OPTICAL SYSTEM
10
Instrument Block Diagram
The internal reference system (reference detector) of the HI83325 photometer compensates for any drifts due to power fluctuations or ambient temperature changes, providing a stable source of light for your blank (zero) measurement and sample measurement. LED light sources offer superior performance compared to tungsten lamps. LEDs have a much higher luminous efficiency, providing more light while using less power. They also produce little heat, which could otherwise affect electronic stability. LEDs are available in a wide array of wavelengths, whereas tungsten lamps have poor blue/violet light output. Improved optical filters ensure greater wavelength accuracy and allow a brighter, stronger signal to be received. The end result is higher measurement stability and less wavelength error. A focusing lens collects all of the light that exits the cuvette, eliminating errors from cuvette imperfections and scratches, eliminating the need to index the cuvette.
Page 11
5. GENERAL OPERATIONS
5.1. POWER CONNECTION AND BATTERY MANAGEMENT
The meter can be powered from an AC/DC adapter (included) or from the built‑in rechargeable battery. The meter will perform an auto‑diagnostic test when it is first powered on. During this test, the HANNA® logo will appear on the LCD. After 5 seconds, if the test was successful, the last method used will appear on the display. The battery icon on the LCD will indicate the battery status:
‑ battery is charging from external adapter ‑ battery fully charged (meter connected to AC/DC adapter)
‑ battery capacity (no external adapter) ‑ battery near 0% (no external adapter)
‑ battery at 0% (no external adapter)
GENERAL OPERATIONS
To conserve battery, the meter will turn off automatically after 15 minutes of inactivity (30 minutes before a READ measurement). If a photometer measurement is on the screen, an auto‑log is created before shutdown.
5.2. GENERAL SETUP
Press SETUP key to enter in Setup menu, highlight desired option using and press Select.
CAL Check (Photometer Only)
Press Select to enter the CAL Check screen. The date, time and values for the last CAL Check are displayed on the screen. To start a new CAL Check press Check key and follow the prompts on the screen.
11
Page 12
Temperature Unit (pH Only)
Option: °C or °F
Press the functional key to select the desired temperature unit.
Backlight
Values: 0 to 8
Press the Modify key to access the backlight intensity. Use the functional keys or the keys to increase or decrease the value.
GENERAL OPERATIONS
Press the Accept key to confirm or ESC to return to the Setup menu without saving the new value.
Contrast
Values: 0 to 20
Press the Modify key to change the display’s contrast. Use the functional keys or the keys to increase or decrease the value. Press the Accept key to confirm the value or ESC to return to the Setup menu without saving the new value.
12
Date / Time
Press the Modify key to change the date/time. Press the functional keys or the keys to highlight the value to be modified (year, month, day, hour, minute or second). Use the keys to change the value. Press the Accept key to confirm or ESC to return to the Setup without saving the new date or time.
Time Format
Option: AM/PM or 24-hour Press the functional key to select the desired time format.
Page 13
Date Format
Press the Modify key to change the Date Format. Use the keys to select the desired format. Press the Select key to confirm or ESC to return to the Setup menu without saving the new format.
Decimal Separator
Option: Comma ( , ) or Period ( . ) Press the functional key to select the desired decimal separator. The decimal separator is used on the measurement screen and CSV files.
Language
Press the Modify key to change the Language. Use the keys to select the desired language. Press Select to choose one of the 7 languages installed.
GENERAL OPERATIONS
Beeper
Option: Enable or Disable
When enabled, a short beep is heard every time a key is pressed. A long beep alert sounds when the pressed key is not active or an error is detected. Press the functional key to enable/disable the beeper.
Instrument ID
Option: 0 to 999999
This option is used to set the instrument’s ID (identification number). Press the Modify key to access the instrument ID screen. Use the functional keys or the keys to highlight the digit to be modified. Press the keys in order to set the desired value. Press the Accept key to confirm the value or ESC to return to the Setup menu without saving the new value.
13
Page 14
Meter Information
Press the Select key to view the model, serial number, firmware version and selected language. Press ESC to return to the Setup menu.
GENERAL OPERATIONS
Probe Information (pH mode only)
Press the Select key to view model number, serial number and firmware version for the connected probe. Press ESC to return to the Setup menu.
5.3. USING HANNA DIGITAL ELECTRODES
The HI83325 can be used to perform direct pH measurements by connecting a HANNA® digital pH electrode with a 3.5 mm TRRS connector. To begin taking probe measurements, connect the electrode to the 3.5 mm port marked with “EXT PROBE” located at the rear of the meter. If the meter is in “Photometer Mode”, set the meter to “Probe Mode” by pressing the MODE key.
14
5.4. MODE SELECTION
The HI83325 has two operational modes: Photometer Mode and Probe Mode. Photometer Mode enables on‑demand measurement of a cuvette using the integrated optical system. Photometric‑related functions, such as Method selection, Zero, Read, and Timers are available in this mode.
Probe Mode enables continuous measurement using a Hanna Digital Electrode connected to the
3.5 mm port. Probe‑related functions, such as calibration and GLP, are available in this mode. To switch between Photometer Mode and Probe Mode, use the
Note: The active mode cannot be switched while in menus, such as Setup, Recall, Method, etc.
button.
Page 15
5.5. LOGGING DATA
The instrument features a data log function to help you keep track of all your analysis. The data log can hold 1000 individual measurements. Storing, viewing and deleting the data is possible using the LOG and RECALL keys. Storing data: You can store only a valid measurement. Press LOG and the last valid measurement will be stored with date and time stamp.
5.6. ADDING SAMPLE / USER NAMES TO LOG DATA
A sample ID and user ID can be added to the saved log. Use the  keys to highlight the Sample ID or User ID then press Modify.
Text Entry
Sample ID and User ID care entered using the alphanumeric multi‑tapping keypad.
GENERAL OPERATIONS
Enter one character at a time by pressing the key with the assigned character repeatedly until the desired character is highlighted. For reference, a list of the characters available for the current key will be shown under the text box. The character will be entered after a two‑second delay or after another key is pressed.
Once all characters have been entered, press Accept to use the displayed text.
15
Page 16
The following functions are available during Text Entry:
Accept: Press to accept the current displayed text.
Arrow: Press to delete the last character.
Clear: Press to delete all characters.
Press to discard all changes and return to the previous screen.
GENERAL OPERATIONS
5.7. DATA MANAGEMENT
Viewing and deleting: You can view, export and delete the data by pressing the RECALL key. Use the  keys to scroll through the saved logs. Press Info to view additional information about the selected log.
Data Export:
Log data can be exported to a USB flash drive or to a PC. To access Data Export functions, press Recall then Export.
16
Use the keys to select the desired export location. For export to USB Flash Drive, insert the USB Flash Drive into the dedicated port at the back of the meter labeled HOST USB, then follow the on‑screen prompts.
For export to PC, connect the meter to a PC using the supplied micro‑USB cable. Insert the cable into the port at the back of the meter labeled PC PWR. Follow the on‑screen prompts. When the meter says PC connected, use a file manager (such as Windows Explorer or Mac Finder) to move the file from the meter to the PC. The meter will appear as a removable disk.
Page 17
Log data is exported as a single file containing all logged photometer and probe data. The file name is: “HI83325.csv”. The CSV file (Comma‑Separated Values) may be opened with a text editor or spreadsheet application.
5.8. CONTEXTUAL HELP HI83325 offers an interactive contextual help mode that assists the user at any time.
To access the help screen press HELP. The instrument will display additional information related to the current screen. To read all the available information, scroll the text using the keys. To exit help mode press ESC key and the meter will return to the previous screen.
GENERAL OPERATIONS
6. NUTRIENT SAMPLES PREPARATION GUIDE
6.1. INTRODUCTION TO PLANT NUTRIENTS
The three elements that are mostly needed by the plants are nitrogen (N), phosphorus (P) and potassium (K). They are called the macronutrients while other elements, needed by plants in smaller amounts, are called microelements. In hydroponics, plants need a balanced nutritive solution, composed of macro and microelements.
Shortage or excess of even only one nutritive element may cause an imbalance in plant physiology and in the absorption of the other nutrients. Nutrients shortages may result in irregular plant growth, low resistance to diseases, scarce production both in quantity and quality, while nutrients excess may cause waste of fertilizer, pollution of the groundwater and the possible accumulation of dangerous substances in the crops produced.
NITROGEN
Nitrogen (N) is mostly absorbed by plants as nitrates (NO ammonium (NH
+
). In hydroponics, an adequate ratio between the two forms is generally used in
4
) and, in smaller amount, in the form of
3
nutritive solutions.
NUTRIENT SAMPLES PREPARATION GUIDE
17
Page 18
PRESENT IN proteins, enzymes, chlorophyll, hormones, vitamins, DNA and RNA
• is fundamental for plants in phase of growth
• promotes lengthening of trunks and sprouts
ACTION
SHORTAGE EFFECTS
EXCESS EFFECT
NUTRIENT SAMPLES PREPARATION GUIDE
PHOSPHORUS
• increases the production of foliage
• helps to absorb other nutrients (in particular phosphorus)
• assists a bigger production for both size and number of fruits
• slower growth
• smaller leaves
• yellowing of leaves
• smaller fruits
• premature ripening
• reduction in resistance to diseases and atmospheric agents
• increase of water demand (caused by an excessive production of leaves)
• bad quality of fruits
• delayed ripening
• reduction in potassium absorption
Phosphorus (P) has an important role in many fundamental biochemical and physiological
3
processes. Plants take up phosphorus in the form of phosphate ion (PO
PRESENT IN DNA and RNA, ATP, ADP
• stimulates the roots growth
• stimulates blooming
ACTION
SHORTAGE EFFECTS
• stimulates fecundation and ripeness
• strengthens the plant tissues
• is necessary in the formation of seeds
• delayed ripening
• slower growth
• small leaves
• decrease of production (smaller fruits and difficult seeds formation)
• reduction of root system
).
4
18
EXCESS EFFECT
• premature ripening
• excess of fruit‑setting
• negative effects on the absorption of some microelements such as iron, zinc, boron and copper
Page 19
POTASSIUM
Potassium (K) is essential in proteic synthesis. The problem of lack of potassium is quite frequent in calcareous soils.
+
Potassium is absorbed as K
.
NUTRIENT SAMPLES PREPARATION GUIDE
PRESENT IN
ACTION
SHORTAGE EFFECTS
EXCESS EFFECT
tissues responsible for the growth of plants (primary and secondary meristems), embryos and cell vacuole
• improves the quality of fruits and flowers
• gives more resistance both to frost and to diseases caused by fungi (increases the cuticular thickness)
• regulates the cellular turgidity (helps to regulate the osmotic processes and increases the resistance to dryness)
• regulates the stomatic opening and closing (it means a strong influence on transpiration and photosynthesis)
• slower growth
• smaller fruits, less colored and less preserved
• increase of transpiration
• less resistance to the cold
• reduced absorption of calcium and magnesium
• increase of water consumption
• increase of the substrate salinity
IRRIGATION WATER
In agricultural areas it is quite common to find altered values in the chemical composition of irrigation waters. The problem concerns mostly the high nitrate concentration, usually determined by excessive fertilization or irrational liquid manure spreading. The analysis of irrigation waters allows us to find out which are the substances present in major or minor quantity and to organize an advantageous fertilization plan.
For example, if the quaninty of water utilized for crop cultivation is 250 mm/ha (=2500000 L/ha) and the nitrate (NO
) concentration is 150 mg/L (34 mg/L as nitrate‑nitrogen NO3‑N), soil receives
3
85 kg/ha of nitrogen. In choosing type and fertilizer to be used, it is important to consider this information, in order not to waste fertilizer nor to induce soil pollution.
19
Page 20
NUTRIENTS SOLUTIONS
The nutrients requirements of the plant are determined by the type of plant, its age and the environmental conditions. The control of chemical composition of nutrients solutions given to the plants is an operation that allows a correct preparation of the fertilizer. In analyzing the solution it is typically necessary to perform a dilution, depending on the concentration of substances.
A dilution factor of 5 usually covers the analysis of residual solution in recycling systems. The nutritive elements are differently absorbed by the plants, hence the nutrient solution loses substances, becomes impoverished and must be enriched.
A dilution factor of 10 normally corresponds to the typical values of nutrients solutions. It is therefore possible to verify that the solution given to the plants contains the correct quantities of nutritive substances.
6.2. PREPARING NUTRIENT SAMPLES FOR ANALYSIS
Nutrient samples need proper preparation before they can be analyzed by photometric methods. The three most common problems are:
1. High concentration (samples contain too much nutrient for the analysis method)
NUTRIENT SAMPLES PREPARATION GUIDE
2. Turbidity (samples appear cloudy or hazy)
3. Color (samples have a colored tint from soil or impurities) High nutrient concentration is overcome by dilution of the sample by a known amount with
demineralized water. This is most often encountered when measuring the macro‑nutrients: ammonia, nitrate, phosphorus, and potassium. The following sections explain procedures for diluting samples by factors of 5, 10, and 50. The table below recommends the dilution procedure and the method to use based on the estimated nutrient concentration:
20
Page 21
Recommended Procedures/Dilutions According to Nutrient Concentration:
NUTRIENT SAMPLES PREPARATION GUIDE
Parameter
Ammonia
Nitrate
Phosphorus
Potassium
Estimated Concentration
< 2.5 ppm NH
2.5 ‑ 9 ppm NH
9 ‑ 100 ppm NH
< 25 ppm NO
25 ‑ 130 ppm NO
130 ‑ 300 ppm NO
‑N
3
‑N
3
‑N
3
‑N
3
‑N
3
‑N
3
< 9 ppm P
3
(< 27 ppm PO
9 ‑ 45 ppm P (27 ‑ 135 ppm PO
45 ‑ 100 ppm P (135 ‑ 300 ppm PO
)
4
3
)
4
3
)
4
< 18 ppm K
18 ‑ 90 ppm K
90 ‑ 180 ppm K
180 ‑ 1000 ppm K
Dilution Factor
No dilution
No dilution
No dilution
No dilution
5
10
No dilution
5
10
No dilution
5
10
50
Method Selection
Ammonia LR
Ammonia MR
Ammonia HR
Nitrate
Nitrate
Nitrate
Phosphate HR
Phosphate HR
Phosphate HR
Potassium
Potassium
Potassium
Potassium
Typical Usage
Irrigation Water
Irrigation Water
Recycled Nutrient Solution Fresh Nutrient Solution
Irrigation Water
Recycled Nutrient Solution
Fresh Nutrient Solution
Irrigation Water
Recycled Nutrient Solution
Fresh Nutrient Solution
Irrigation Water
Recycled Nutrient Solution
Fresh Nutrient Solution
Fresh Nutrient Solution
The concentration of the micro‑nutrients (calcium, magnesium, sulfate) is low enough in most samples that dilution is typically not required. If necessary, a dilution procedure can be used for these parameters as well.
21
Page 22
6.3. PROCEDURE FOR DILUTION FACTOR: 5
Note: For a more accurate dilution, use laboratory-grade glass pipettes and volumetric flasks.
• Use the graduate cylinder to measure exactly 20 mL of sample
• Remove the cap and fill the Demineralizer Bottle with tap water.
NUTRIENT SAMPLES PREPARATION GUIDE
• Replace the cap and shake gently for at least 2 minutes.
22
• Open the upper part of the Demineralizer Bottle cap and gently squirt the demineralized water into the cylinder, up to the 100 mL mark.
Note: The ion exchange resin contained in the Demineralizer Bottle is provided with an indicator substance. The indicator will change from green to blue when the resin has been exhausted and needs to be replaced.
Page 23
• Pour the solution in the large 170 mL beaker, replace the cap and invert several times to mix.
• If the solution contains some turbidity or color, follow the procedure in 6.6. REMOVING TURBIDITY AND COLOR.
6.4. PROCEDURE FOR DILUTION FACTOR: 10
Note: For a more accurate dilution, use laboratory-grade glass pipettes and volumetric flasks.
• Add 10 mL of sample to the graduated cylinder using the 5 mL syringe (twice).
Note: To measure exactly 5 mL of sample with the syringe, push the plunger completely into the syringe and insert the tip into the sample. Pull the plunger out until the lower edge of the seal is on the 5 mL mark of the syringe.
NUTRIENT SAMPLES PREPARATION GUIDE
23
Page 24
• Remove the cap and fill the Demineralizer Bottle with tap water.
• Replace the cap and shake gently for at least 2 minutes.
• Open the upper part of the Demineralizer Bottle cap and squirt gently the demineralized water into the cylinder, up to the 100 mL mark.
NUTRIENT SAMPLES PREPARATION GUIDE
• Pour the solution in the large 170 mL beaker, replace the cap and invert several times to mix.
24
• If the solution contains some turbidity or color, follow the procedure in 6.6. REMOVING TURBIDITY AND COLOR.
Page 25
6.5. PROCEDURE FOR DILUTION FACTOR: 50
Note: For a more accurate dilution, use laboratory-grade glass pipettes and volumetric flasks.
• Add 10 mL of sample to the graduated cylinder using the 5 mL syringe (twice).
Note: To measure exactly 5 mL of sample with the syringe, push the plunger completely into the syringe and insert the tip into the sample. Pull the plunger out until the lower edge of the seal is on the 5 mL mark of the syringe.
NUTRIENT SAMPLES PREPARATION GUIDE
• Remove the cap and fill the Demineralizer Bottle with tap water.
• Replace the cap and shake gently for at least 2 minutes.
25
Page 26
• Open the upper part of the Demineralizer Bottle cap and squirt gently the demineralized water into the cylinder, up to the 100 mL mark.
• Pour the solution in the large 170 mL beaker, replace the cap and invert several times to mix.
• Clean and dry the graduated cylinder, then pour 20 mL of the diluted solution from the
NUTRIENT SAMPLES PREPARATION GUIDE
170‑mL beaker to the graduated cylinder.
• Open the upper part of the Demineralizer Bottle cap and squirt gently the demineralized water into the cylinder, up to the 100 mL mark.
26
• Clean and dry the large 170 mL beaker, then pour the solution from the graduated cylinder to the large 170 mL beaker, replace the cap and invert several times to mix.
• If the solution contains some turbidity or color, follow the procedure in 6.6. REMOVING TURBIDITY AND COLOR.
Page 27
6.6. REMOVING TURBIDITY AND COLOR
Turbidity and color in samples will adversely affect the nutrient analysis. This procedure removes turbidity and color.
Note: Perform any necessary dilutions before attempting to remove turbidity or color.
1. If the sample is extremely turbid, pour the sample into the large 170‑mL beaker. Allow the sample to stand in the beaker until most of the solid particles have settled. Then, use the pipette to transfer the particle‑clear supernatant solution to the 100‑mL graduated cylinder. Discard sample containing visible particles. Repeat the process until you have filled the graduated cylinder to the 100‑mL line. Clean the 170‑mL beaker with demineralized water and dry it before using it again.
2. Pour 100 mL of sample into the large 170‑mL beaker.
3. Add 1 powder packet of Activated Carbon.
4. Mix well using the spoon and then wait for 5 minutes.
NUTRIENT SAMPLES PREPARATION GUIDE
5. Fold a filter disc twice as shown in the figure. Separate one side from the other three to form a cone. Insert the folded filter disc in the funnel.
27
Page 28
6. Filter the treated sample into an empty beaker. The sample is now ready.
Note: Filter at least 40 mL of solution if all four methods will be tested. If the solution is still turbid or colored, treat it again with a packet of active carbon. After use, throw the filter disc away and wash the syringe and the filter assembly well. Always use a new disc for another sample.
7. PHOTOMETER MODE
7.1. METHOD SELECTION
In order to select the desired method press the METHOD key and a screen with the available methods will appear. Press the  keys to highlight the desired method. Press Select.
PHOTOMETER MODE
After the desired method is selected, follow the procedure described in the related section. Before performing a method read all the instructions carefully.
28
Page 29
7.2. COLLECTING AND MEASURING SAMPLES AND REAGENTS
7.2.1. PROPER USE OF SYRINGE
(a) Push the plunger completely into the syringe and insert the tip
into the solution.
(b) Pull the plunger up until the lower edge of the seal is exactly on
the mark for the desired volume.
(c) Take out the syringe and clean the outside of the syringe tip, be
sure that no drops are hanging on the tip of the syringe. Then, keeping the syringe in vertical position above the cuvette, push the plunger down into the syringe, the desired volume has been delivered into the cuvette.
7.2.2. PROPER USE OF DROPPER
(a) For reproducible results, tap the dropper on the table several times and wipe the outside of the
tip with a cloth.
(b) Always keep the dropper bottle in a vertical position while dosing the reagent.
PHOTOMETER MODE
7.2.3. PROPER USE OF POWDER PACKET
(a) Use scissors to open the powder packet (b) Push the edges of the packet to form a spout (c) Pour out the content of the packet.
29
Page 30
7.3. CUVETTE PREPARATION
Proper mixing is very important for reproducibility of the measurements. The proper mixing technique for each method is listed in the method procedure. (a) Invert the cuvette a couple of times or for a specified time: hold the cuvette in the vertical
position. Turn the cuvette upside‑down and wait for all of the solution to flow to the cap end, then return the cuvette to the upright vertical position and wait for all of the solution to flow to the cuvette bottom. This is one inversion. The correct speed for this mixing technique is 10‑15 complete inversions in 30 seconds.
This mixing technique is indicated with “invert to mix” and the following icon:
PHOTOMETER MODE
(b) Shaking the cuvette, moving the cuvette up and down. The movement may be gentle or
vigorous. This mixing method is indicated with “shake gently” or “shake vigorously”, and one of the following icons:
30
shake gently shake vigorously
In order to avoid reagent leaking and to obtain more accurate measurements, close the cuvette first with the supplied HDPE plastic stopper black cap. Whenever the cuvette is placed into the measurement holder, it must be dry outside and free of fingerprints, oil or dirt. Wipe it thoroughly with HI731318 or a lint‑free cloth prior to insertion. Shaking the cuvette can generate bubbles in the sample, causing higher readings. To obtain accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette.
and then the
Page 31
Do not let the reacted sample stand too long after reagent is added. For best accuracy, respect the timings described in each specific method.
It is possible to take multiple readings in a row, but it is recommended to take a new zero reading for each sample and to use the same cuvette for zeroing and measurement when possible. Discard the sample immediately after the reading is taken, or the glass might become permanently stained. All the reaction times reported in this manual are at 25 °C (77 °F). In general, the reaction time should be increased for temperatures lower than 20 °C (68 °F), and decreased for temperatures higher than 25 °C (77 °F).
Interference
In the method measurement section the most common interferences that may be present in a typical water sample have been reported. It is possible that a particular application could introduce other compounds that will also interfere.
7.4. TIMERS AND MEASUREMENT FUNCTIONS
Each method requires a different preparation procedure, reaction times, sample preparations, etc. If a timer or timers are necessary for proper sample preparation, the Timer key will be available.
To use a reaction timer, press the Timer key.
The default timer will start immediately. To stop and reset the timer, press Stop.
PHOTOMETER MODE
If the selected method requires more than one timer, the meter will automatically select each timer in the appropriate order. To bypass the default order, you may press the desired key to activate a different timer (only while the current timer is stopped). Press Continue to start the active timer.
For some methods, the timer is only necessary after a Zero measurement has been performed. In this case, the timer key will only be available after the Zero measurement has been performed.
If the method requires a Zero or Read measurement after a timer has expired, the meter will automatically perform the appropriate action. Follow the instructions in the Method Procedure.
To perform a Zero or Read measurement, insert the appropriate prepared cuvette, then press the Zero or Read key. A Zero measurement must be conducted before Read measurements.
31
Page 32
7.5. CHEMICAL FORMULA / UNIT CONVERSION
Chemical formula/unit conversion factors are pre‑programmed into the instrument and are method specific. In order to view the displayed result in the desired chemical formula press keys to access the second level function and then press the Chem Frm key to toggle between the available chemical formulas for the selected method.
PHOTOMETER MODE
7.6. METER VALIDATION / CAL CHECK
WARNING: Do not validate the meter with standard solutions other than the HANNA Standards. For accurate validation results, please perform tests at room temperature (18 to 25 °C;
64.5 to 77.0 °F).
Validation of the HI83325 involves absorbance measurements of certified HANNA Standards (see “Accessories”). The “CAL Check” screen guides the user through the measurement of each CAL Check Standard and applies the factory calibration corrections to each measurement. The
HI83325 stores the results of the most recent CAL Check measurements which may be viewed on
the “CAL Check” screen. Compare these results with the values printed on the Certificate provided with each HANNA® CAL Check Standards kit.
To perform a validation:
®
CAL Check
®
CAL Check
32
1. Press Setup button.
2. Highlight CAL Check, then press Select.
3. Follow the prompts on the screen. The meter will prompt to
®
measure each cuvette provided in the HANNA
CAL Check
Standards kit. To abort the process at any time, press ESC button.
Page 33
4. Press ESC to return in Setup menu.
7.7. ABSORBANCE MEASUREMENTS
Raw absorbance measurements may be performed on the HI83325 for personal or diagnostic purposes. For example, you may monitor the stability of a reagent blank by occasionally measuring its absorbance versus deionized water.
To measure the raw absorbance of a prepared sample:
1. Enable “Photometer Mode” if necessary by pressing the MODE key.
2. Press the METHOD key.
3. Highlight the appropriate Absorbance method (according to the wavelength to be used), then
press Select. Absorbance methods are located at the bottom of the method list.
4. Prepare the sample cuvette according to the method.
5. Insert a cuvette filled with deionized water, then press Zero.
6. Insert the prepared sample cuvette, then press Read.
PHOTOMETER MODE
WARNING: Never use Absorbance methods for validation using HANNA factory calibration corrections for CAL Check cuvettes are applied while in CAL Check mode only!
®
CAL Check cuvettes. The
33
Page 34
8. PROBE MODE
8.1. pH CALIBRATION
Press MODE to enter in pH/ mV measurement mode.
Press Calibrate to access electrode calibration functions.
PROBE MODE
Calibration Mode
While in pH Calibration Mode, the display will show the current pH reading, the current temperature reading, the current selected buffer, and the buffer number (“Buffer: 1” for the 1st buffer, “Buffer: 2” for the 2nd buffer).
The following functions are available in pH Calibration Mode:
Clear: Press to clear the current calibration from the probe.
34
Confirm: Press to accept the current calibration point. Only
available if the measurement is stable and within the limits for the selected buffer.
Press to cycle through the list of available buffers: 4.01, 6.86, 7.01, 9.18, 10.01 pH.
Press to exit calibration and return to pH Measurement Mode.
Page 35
Preparation
Pour small quantities of the buffer solutions into clean beakers. If possible, use plastic beakers to minimize any EMC interferences. For accurate calibration and to minimize cross‑contamination, use two beakers for each buffer solution: one for rinsing the electrode and one for calibration. If you are measuring in the acidic range, use pH 7.01 or 6.86 as the first buffer and pH 4.01 as the second buffer. If you are measuring in the alkaline range, use pH 7.01 or 6.86 as the first buffer and pH 10.01 or 9.18 as the second buffer.
Procedure
Calibration can be performed using one or two calibration buffers. For more accurate measurements, a two‑point calibration is recommended. Submerse the pH electrode approximately 3 cm (1¼”) into a buffer solution and stir gently. From the Probe Measurement screen, press the Calibrate key to begin the calibration process.
When the reading is stable and close to the selected buffer, the Confirm key will become available. Press Confirm to accept and store the calibration point.
The meter will now prompt for the second buffer (“Buffer: 2”). To use only a one‑point calibration, press to the probe and return to Measurement mode. To continue calibrating with a second buffer, rinse and submerse the pH electrode approximately 3 cm (1¼”) into the second buffer solution and stir gently. If necessary, press
to exit calibration mode at this time. The meter will store the calibration information
keys to select a different buffer value.
PROBE MODE
When the reading is stable and close to the selected buffer, the Confirm key will become available. Press Confirm to accept and store the second calibration point.
The meter will store the two‑point calibration information to the probe and return to Measurement mode. The list of calibrated buffers will appear at the bottom of the screen.
35
Page 36
8.2. pH CALIBRATION MESSAGES
Clean Probe:
The "Clean Probe" message indicates poor electrode performance (offset out of accepted window, or slope under the accepted lower limit). Often, cleaning the probe will improve
PROBE MODE
the pH electrode's response. See pH Electrode Conditioning and Maintenance for details. Repeat calibration after cleaning.
Check Probe & Buffer:
The “Check Probe & Buffer” message appears when there is a large difference between the pH measurement and the selected buffer value, or the electrode slope is outside of the accepted slope limit. You should check your probe and confirm the correct buffer selection. Cleaning may also improve this response.
Wrong Temperature:
The buffer temperature is too extreme for the selected buffer value.
8.3. pH MEASUREMENT
The HI83325 can be used to perform direct pH measurements by connecting a HANNA pH electrode with a 3.5 mm TRRS connector. To begin taking probe measurements, connect the electrode to the 3.5 mm port marked with EXT PROBE located at the rear of the meter. If the meter is in “Photometer Mode”, set the meter to “Probe Mode” by pressing the MODE key.
®
digital
36
While taking pH probe measurements, the following functions are available:
Calibrate: Press to access electrode calibration functions.
GLP: Press to review the last calibration information, including date/time, buffers used, slope,
and offset.
Range: Press to switch between “pH” units and “mV” units.
Page 37
Press to switch to Photometer mode.
Press to access the meter’s Setup menu.
Press to log the current measurement.
Press to review the meter’s log history.
Press to view contextual help information.
For high accuracy it is recommended to calibrate your electrode often. pH electrodes should be recalibrated at least once per week, but daily calibration is recommended. Always recalibrate after cleaning an electrode. See page 34 for more information on pH calibration.
To take pH measurements:
• Remove the protective cap and rinse the electrode with water.
• Collect some sample in a clean, dry beaker.
• Preferably, rinse the electrode with a small amount of sample. Discard the rinse.
• Submerse the electrode tip approximately 3 cm (1¼”) into the sample to be tested and stir the sample gently. Make sure the electrode junction is completely submersed.
• Allow time for the electrode to stabilize in the sample. When the reading is stable.
symbol disappears, your
PROBE MODE
If measurements are taken successively in different samples, it is recommended to rinse the electrodes thoroughly with deionized or distilled water and then with some of the next sample to prevent cross‑contamination.
pH measurements are affected by temperature. HANNA temperature sensor and automatically calculate corrected pH values. The measured temperature is displayed on the screen with the pH measurements.
®
Digital pH electrodes include a built‑in
37
Page 38
8.4. pH MEASUREMENT MESSAGES / WARNINGS
No Probe:
No probe is connected or the probe is broken.
Connecting:
PROBE MODE
The meter has detected a probe and is reading the probe configuration and calibration information.
Incompatible Probe:
The connected probe is not compatible with this device.
Incompatible Calibration:
The probe’s current calibration is not compatible with this meter. The calibration must be cleared to use this probe.
Exceeded Probe Range:
The pH and/or temperature measurement exceed the specifications of the probe. The affected measurement value(s) will be flashing.
Broken Temperature Sensor:
The temperature sensor inside the probe is broken. Temperature compensation will revert to a fixed value of 25 °C (77 °C).
38
Cal Due:
The probe has no calibration. See section Probe Calibration.
Page 39
8.5. pH GLP
Good Laboratory Practice (GLP) refers to a quality control function used to ensure uniformity and consistency of sensor calibrations and measurements. To view the GLP information, press the GLP key from the Probe Measurement screen.
The pH GLP screen displays the following information about the last pH calibration:
• Date and time of the last calibration
• List of buffers used in the last calibration
• Calculated slope and offset
• Press ESC to return in measurement mode.
PROBE MODE
39
Page 40
8.6. pH ELECTRODE CONDITIONING AND MAINTENANCE
PROBE MODE
Remove the protective cap of the pH electrode. DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal with electrodes. They will disappear when rinsed with water. During transport, tiny bubbles of air may form inside the glass bulb affecting proper functioning of the electrode. These bubbles can be removed by “shaking down” the electrode as you would do with a glass thermometer. If the bulb and/or junction is dry, soak the electrode in HI70300 or HI80300 storage solution for at least one hour.
40
For refillable electrodes: If the filling solution (electrolyte) is more than 2½ cm (1”) below the fill hole, add HI7082 or
HI8082 3.5M KCl Electrolyte Solution for double junction electrodes.
Unscrew the fill hole cover during measurements so the liquid reference junction maintains an outward flow of electrolyte.
Measurement
Rinse the electrode tip with distilled water. Submerse the tip 3 cm (1¼”) in the sample and stir gently for a few seconds. For a faster response and to avoid cross‑contamination of the samples, rinse the electrode tip with a few drops of the solution to be tested, before taking measurements.
Page 41
Storage Procedure
To minimize clogging and ensure a quick response time, the glass bulb and the junction should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of HI70300 or HI80300 Storage Solution or, in its absence, Filling Solution (HI7082 or HI8082 for double junction electrodes). Follow the preparation procedure before taking measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.
Periodic Maintenance
Inspect the electrode and the cable. The cable used for connection to the instrument must be intact and there must be no points of broken insulation on the cable or cracks on the electrode stem or bulb. Connectors must be perfectly clean and dry. If any scratches or cracks are present, replace the electrode. Rinse off any salt deposits with water. For refillable electrodes: Refill the reference chamber with fresh electrolyte (HI7082 or HI8082 for double junction electrodes). Allow the electrode to stand upright for 1 hour. Follow the Storage Procedure above.
Cleaning Procedure
Use diagnostic messages to aid pH electrode troubleshooting. Several cleaning solutions are available:
• General – Soak in Hanna HI7061 or HI8061 General Cleaning Solution for approximately ½ hour.
• Protein –Soak in Hanna HI7073 or HI8073 Protein Cleaning Solution for 15 minutes.
• Inorganic – Soak in Hanna HI7074 Inorganic Cleaning Solution for 15 minutes.
• Oil/grease – Rinse with Hanna HI7077 or HI8077 Oil and Fat Cleaning Solution.
PROBE MODE
Note: After performing any of the cleaning procedures, rinse the electrode thoroughly with distilled water, refill the reference chamber with fresh electrolyte (not necessary for gel- filled electrodes) and soak the electrode in HI70300 or HI80300 Storage Solution for at least 1 hour before taking measurements.
Temperature Correlation For pH Sensitive Glass
Verify the temperature range by reading the limits on electrodes cap. The pH electrode’s life also depends on the temperature that is used. If constantly cycled between two temperatures, the life of the electrode is drastically reduced.
41
Page 42
9. METHOD PROCEDURES
9.1. AMMONIA LOW RANGE
SPECIFICATIONS
Range 0.00 to 3.00 mg/L (as NH3‑N) Resolution 0.01 mg/L Accuracy ±0.04 mg/L ±4% of reading at 25 °C Light Source LED with narrow band interference filter @ 420 nm Method Adaptation of the ASTM Manual of Water and Environmental Technology, D1426 Nessler method.
AMMONIA LOW RANGE
REQUIRED REAGENTS Code Description Quantity
HI93700A‑0 Ammonia Low Range Reagent A 4 drops HI93700B‑0 Ammonia Low Range Reagent B 4 drops
REAGENT SETS
HI93700‑01 Reagents for 100 tests HI93700‑03 Reagents for 300 tests
For other accessories see page 72.
42
MEASUREMENT PROCEDURE
• Select the Ammonia LR method using the procedure described in the
Method Selection section (see page 28).
• Follow procedures in Section 6 to prepare the sample for analysis.
• Fill the cuvette with 10 mL of unreacted sample (up to the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
10 mL
Page 43
• Add 4 drops of HI93700A‑0 Ammonia Low Range Reagent A. Replace the cap and mix the solution.
• Add 4 drops of HI93700B‑0 Ammonia Low Range Reagent B. Replace the cap and mix the solution.
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and 30 seconds and press Read. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of ammonia nitrogen (NH
-N).
3
AMMONIA LOW RANGE
43
Page 44
• Diluted Samples: If the sample was diluted, multiply this result by the dilution factor to calculate the concentration of the original undiluted sample. For example, if the diluted sample yields a result of 2.74 mg/L after being diluted by a factor of 5, then the original sample concentration would be 2.74 x 5 = 13.7 mg/L.
• Press  or  to access the second level functions.
• Press the Chem Frm key to convert the result to mg/L of ammonia (NH
+
(NH
).
4
3
AMMONIA LOW RANGE
• Press  or  to return to the measurement screen.
INTERFERENCE
Interference may be caused by: Acetone Alcohols Aldehydes Glycine Hardness above 1 g/L Iron Organic chloramines Sulfide Various aliphatic and aromatic amines
) and ammonium
44
Page 45
9.2. AMMONIA MEDIUM RANGE
SPECIFICATIONS
Range 0.00 to 10.00 mg/L (as NH Resolution 0.01 mg/L Accuracy ±0.05 mg/L ±5% of reading at 25 °C Light Source LED with narrow band interference filter @ 420 nm Method Adaptation of the ASTM Manual of Water and Environmental Technology, D1426, Nessler method.
REQUIRED REAGENTS Code Description Quantity
HI93715A‑0 Ammonia Medium Range Reagent A 4 drops HI93715B‑0 Ammonia Medium Range Reagent B 4 drops
REAGENT SETS
HI93715‑01 Reagents for 100 tests HI93715‑03 Reagents for 300 tests
For other accessories see page 72.
MEASUREMENT PROCEDURE
• Select the Ammonia MR method using the procedure described in the Method Selection section (see page 28).
• Follow procedures in Section 6 to prepare the sample for analysis.
3
‑N)
10 mL
AMMONIA MEDIUM RANGE
• Fill the cuvette with 10 mL of unreacted sample (up to the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for measurement.
45
Page 46
• Remove the cuvette.
• Add 4 drops of HI93715A‑0 Ammonia Medium Range Reagent A. Replace the cap and mix the solution.
• Add 4 drops of HI93715B‑0 Ammonia Medium Range Reagent B. Replace the cap and mix the solution.
AMMONIA MEDIUM RANGE
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and 30 seconds and press Read. When the timer ends the meter will perform the reading. The instrument displays the results to mg/L of ammonia nitrogen (NH
-N).
3
46
Page 47
• Diluted Samples: If the sample was diluted, multiply this result by the dilution factor to calculate the concentration of the original undiluted sample. For example, if the diluted sample yields a result of 2.78 mg/L after being diluted by a factor of 5, then the original sample concentration would be 2.78 x 5 = 13.9 mg/L.
• Press or to access the second level functions.
AMMONIA MEDIUM RANGE
• Press the Chem Frm key to convert the result in mg/L of ammonia (NH
+
(NH
).
4
• Press  or  to return to the measurement screen.
INTERFERENCES
Interference may be caused by: Acetone Alcohols Aldehydes Glycine Hardness above 1 g/L Iron Organic chloramines Sulfide Various aliphatic and aromatic amines
) and ammonium
3
47
Page 48
9.3. AMMONIA HIGH RANGE
SPECIFICATIONS
Range 0.0 to 100.0 mg/L (as NH3‑N) Resolution 0.1 mg/L Accuracy ±0.5 mg/L ±5% of reading at 25 °C Light Source LED with narrow band interference filter @ 420 nm Method Adaptation of the ASTM Manual of Water and Environmental Technology, D1426, Nessler method.
REQUIRED REAGENTS
AMMONIA HIGH RANGE
Code Description Quantity
HI93733A‑0 Ammonia High Range Reagent A 4 drops HI93733B‑0 Ammonia High Range Reagent B 9 mL
REAGENT SETS
HI93733‑01 Reagents for 100 tests HI93733‑03 Reagents for 300 tests
For other accessories see page 72.
MEASUREMENT PROCEDURE
• Select the Ammonia HR method using the procedure described in the Method Selection section (see page 28).
48
• Follow procedures in Section 6 to prepare the sample for analysis.
• Add 1mL of unreacted sample to the cuvette using 1mL syringe.
• Use the pipette to fill the cuvette up to the 10 mL mark with HI93733B‑0 Ammonia High Range Reagent B. Replace the cap and mix the solution.
• Place the cuvette into the holder and close the lid.
1 mL
Page 49
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Add 4 drops of HI93733A‑0 Ammonia High Range Reagent A. Replace the cap and swirl the solution.
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and 30 seconds and press Read. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of ammonia nitrogen (NH
-N).
3
AMMONIA HIGH RANGE
49
Page 50
•Diluted Samples: If the sample was diluted, multiply this result by the dilution factor to calculate the concentration of the original undiluted sample. For example, if the diluted sample yields a result of 47.5 mg/L after being diluted by a factor of 5, then the original sample concentration would be
47.5 x 5 = 237.5 mg/L.
• Press  or  to access the second level functions.
•Press the Chem Frm key to convert the result to mg/L of ammonia (NH
+
(NH
).
4
AMMONIA HIGH RANGE
•Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by: Acetone Alcohols Aldehydes Glycine Hardness above 1 g/L Iron Organic chloramines Sulfide Various aliphatic and aromatic amines
) and ammonium
3
50
Page 51
9.4. CALCIUM
SPECIFICATIONS
Range 0 to 400 mg/L (as Ca2+) Resolution 1 mg/L Accuracy ±10 mg/L ±5% of reading at 25 °C Light Source LED with narrow band interference filter @ 466 nm Method Adaptation of the Oxalate method.
REQUIRED REAGENTS Code Description Quantity
Buffer Reagent 4 drops HI93752A‑Ca Calcium Reagent A 7 mL HI93752B‑Ca Calcium Reagent B 1 mL
REAGENT SETS
HI937521‑01 Reagents for 50 tests HI937521‑03 Reagents for 150 tests
For other accessories see page 72.
MEASUREMENT PROCEDURE
• Select the Calcium method using the procedure described in the Method Selection section (see page 28).
CALCIUM
• Follow procedures in Section 6 to prepare the sample for analysis.
• Add 3 mL of unreacted sample to the cuvette using the 5 mL syringe.
• Use the pipette to fill the cuvette up to the 10 mL mark with the HI93752A‑Ca Calcium Reagent A.
• Add 4 drops of Buffer Reagent.
3 mL of sample
51
Page 52
• Replace the cap and invert several times to mix.
CALCIUM
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Add 1 mL of HI93752B‑Ca Calcium Reagent B to the sample using the 1 mL syringe. Invert the cuvette 10 times to mix (about 15 seconds).
52
• Reinsert the cuvette into the instrument.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 5 minutes.
Page 53
• After waiting 5 minutes, invert again the cuvette 10 times to mix (about 15 seconds).
• Reinsert the cuvette into the instrument and close the lid.
CALCIUM
• Press Read to start the reading. The instrument displays the results in mg/L of calcium (Ca
• Diluted Samples: If the sample was diluted, multiply this result by the dilution factor to calculate the concentration of the original undiluted sample. For example, if the diluted sample yields a result of 360 mg/L after being diluted by a factor of 5, then the original sample concentration would be 360 x 5 = 1800 mg/L.
INTERFERENCES
Interferences may be caused by: Acidity (as CaCO
) above 1000 mg/L
3
Alkalinity (as CaCO3) above 1000 mg/L Magnesium (Mg2+) above 400 mg/L
2+
).
53
Page 54
9.5. MAGNESIUM
SPECIFICATIONS
Range 0 to 150 mg/L (as Mg2+) Resolution 1 mg/L Accuracy ±5 mg/L ±3% of reading at 25 °C
MAGNESIUM
Light Source LED with narrow band interference filter @ 466 nm Method Adaptation of the Calmagite method.
REQUIRED REAGENTS Code Description Quantity
HI93752A‑Mg Magnesium Reagent A 1 mL HI93752B‑Mg Magnesium Reagent B 9 mL
REAGENT SETS
HI937520‑01 Reagents for 50 tests HI937520‑03 Reagents for 150 tests
For other accessories see page 72.
MEASUREMENT PROCEDURE
• Select the Magnesium method using the procedure described in the Method Selection section (see page 28).
• Follow procedures in Section 6 to prepare the sample for analysis.
54
• Add 1 mL of HI93752A‑Mg Magnesium Reagent A to the cuvette using a 1 mL syringe and use the pipette to fill the cuvette up to the 10 mL mark with the HI93752B‑Mg Magnesium Reagent B.
• Replace the cap and invert several times to mix.
• Place the cuvette into the holder and close the lid.
Page 55
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for measurement.
0.5 mL
• Remove the cuvette.
of sample
• Add 0.5 mL of sample to the cuvette using the second 1 mL syringe.
• Replace the cap and invert several times to mix.
• Reinsert the cuvette into the instrument and close the lid.
MAGNESIUM
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait for 15 seconds and press Read. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of magnesium (Mg
2+
).
55
Page 56
• Diluted Samples: If the sample was diluted, multiply this result by the dilution factor
to calculate the concentration of the original undiluted sample. For example, if the diluted sample yields a result of 47 mg/L after being diluted by a factor of 5, then the original sample concentration would be 47 x 5 = 235 mg/L.
MAGNESIUM
INTERFERENCES
Interferences may be caused by: acidity (as CaCO3) above 1000 mg/L, alkalinity (as CaCO3) above 1000 mg/L, Calcium (Ca2+) above 200 mg/L, Iron must be absent, Aluminum must be absent, Copper must be absent.
56
Page 57
9.6. NITRATE
NITRATE
SPECIFICATIONS
Range 0.0 to 30.0 mg/L (as NO
3
‑ N) Resolution 0.1 mg/L Accuracy ±0.5 mg/L ±10% of reading at 25 °C Light Source LED with narrow band interference filter @ 525 nm Method Adaptation of the cadmium reduction method.
REQUIRED REAGENTS Code Description Quantity
HI93728‑0 Nitrate Reagent 1 packet
REAGENT SETS
HI93728‑01 Reagents for 100 tests HI93728‑03 Reagents for 300 tests
For other accessories see page 72.
MEASUREMENT PROCEDURE
• Select the Nitrate method using the procedure described in
the Method Selection section (see page 28).
• Follow procedures in Section 6 to prepare the sample for
analysis.
• Fill the cuvette with 10 mL of sample, (up to the mark),
and replace the cap.
10 mL
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
57
Page 58
NITRATE
• Remove the cuvette and add one packet of HI93728‑0 Nitrate Reagent.
• Replace the cap and shake vigorously up and down for exactly
10 seconds. Continue to mix by inverting the cuvette gently for 50 seconds, while taking care not to induce air bubbles. Powder will not completely dissolve. Time and method of shaking could sensitively affect the measurement.
Note: The method is technique-sensitive. See procedure on page 30 Cuvette Preparation for proper mixing technique.
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 4 minutes and 30 seconds and press Read. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of nitrate-nitrogen (NO
-N).
3
58
Page 59
• Diluted Samples: If the sample was diluted, multiply this result by the dilution factor to
calculate the concentration of the original undiluted sample. For example, if the diluted sample yields a result of 2.1 mg/L after being diluted by a factor of 5, then the original sample concentration would be 2.1 x 5 = 10.5 mg/L.
• Press  or  to access the second level functions.
NITRATE
• Press the Chem Frm key to convert the result to mg/L of nitrate (NO
•Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by: Ammonia and amines, as urea and primary aliphatic amines Chloride above 100 ppm Chlorine above 2 ppm Copper Iron(III) Strong oxidizing and reducing substances Sulfide must be absent
3
¯).
59
Page 60
9.7. PHOSPHATE HIGH RANGE
SPECIFICATIONS
3
Range 0.0 to 30.0 mg/L (as PO
)
4
0.0 to 9.8 mg/L (as P) Resolution 0.1 mg/L Accuracy ±1.0 mg/L ±4% of reading at 25 °C Light Source LED with narrow band interference filter @ 525 nm Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method.
REQUIRED REAGENTS
PHOSPHATE HIGH RANGE
Code Description Quantity
HI93717A‑0 Phosphate High Range Reagent A 10 drops HI93717B‑0 Phosphate High Range Reagent B 1 packet
REAGENT SETS
HI93717‑01 Reagents for 100 tests HI93717‑03 Reagents for 300 tests
For other accessories see page 72.
MEASUREMENT PROCEDURE
• Select the Phosphate HR method using the
procedure described in the Method Selection section (see page 28).
• Follow procedures in Section 6 to prepare the
10 mL
sample for analysis.
60
• Fill the cuvette with 10 mL of unreacted sample
(up to the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for measurement.
Page 61
• Add 10 drops of HI93717A‑0 Phosphate High Range Reagent A.
• Add one packet of HI93717B‑0 Phosphate HR Reagent
B to the cuvette. Replace the cap and shake gently until completely dissolved.
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 5 minutes and press Read. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of phosphate (PO
3
¯).
4
PHOSPHATE HIGH RANGE
• Diluted Samples: If the sample was diluted, multiply this result by the dilution factor to
calculate the concentration of the original undiluted sample. For example, if the diluted sample yields a result of 15.3 mg/L after being diluted by a factor of 5, then the original sample concentration would be 15.3 x 5 = 76.5 mg/L.
61
Page 62
• Press  or  to access the second level functions.
• Press the Chem Frm key to convert the result to mg/L of phosphorus (P) and phosphorus
pentoxide (P
• Press  or  to return to the measurement screen.
PHOSPHATE HIGH RANGE
2O5
).
INTERFERENCES
Sulfide Chloride above 150000 mg/L Calcium above 10000 mg/L as CaCO Magnesium above 40000 mg/L as CaCO
3
3
Ferrous iron above 100 mg/L
62
Page 63
9.8. POTASSIUM
SPECIFICATIONS
Range 0.0 to 20.0 mg/L (as K) Resolution 0.1 mg/L Accuracy ±3.0 mg/L ±7% of reading at 25 °C Light Source LED with narrow band interference filter @ 466 nm Method Adaptation of the Turbidimetric Tetraphenylborate method.
REQUIRED REAGENTS Code Description Quantity
HI93750A‑0 Potassium Reagent A 6 drops HI93750B‑0 Potassium Reagent B 1 packet
REAGENT SETS
HI93750‑01 Reagents for 100 tests HI93750‑03 Reagents for 300 tests
For other accessories see page 72.
MEASUREMENT PROCEDURE
• Select the Potassium method using the procedure described
in the Method Selection section (see page 28).
• Follow procedures in Section 6 to prepare the sample for analysis.
• Fill the cuvette with 10 mL of sample (up to the mark).
POTASSIUM
10 mL
• Add 6 drops of HI93750A‑0 Potassium Reagent A. Replace the cap and swirl the solution.
• Place the cuvette into the holder and close the lid.
63
Page 64
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
POTASSIUM
• Add one packet of HI93750B‑0 Potassium Reagent B
Replace the cap and shake gently for 1 minute.
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 2 minutes and press Read. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of potassium (K).
64
• Diluted Samples: If the sample was diluted, multiply this result by the dilution factor to
calculate the concentration of the original undiluted sample. For example, if the diluted sample yields a result of 3.5 mg/L after being diluted by a factor of 5, then the original sample concentration would be 3.5 x 5 = 17.5 mg/L.
Page 65
• Press  or  to access the second level functions.
•Press the Chem Frm key to convert the result to mg/L of potassium oxide (K
•Press or to return to the measurement screen.
INTERFERENCES
Interferences may be caused by: Ammonium above 10 ppm Calcium above 10000 ppm as CaCO
3
Chloride above 12000 ppm Magnesium above 8000 ppm as CaCO
3
Sodium above 8000 ppm
POTASSIUM
O).
2
65
Page 66
9.9. SULFATE
SPECIFICATIONS
2
Range 0 to 150 mg/L (as SO
SULFATE
Resolution 1 mg/L
)
4
Accuracy ±5 mg/L ±3% of reading at 25 °C Light Source LED with narrow band interference filter @ 466 nm Method Sulfate is precipitated with barium chloride crystals.
REQUIRED REAGENTS Code Description Quantity
HI93751‑0 Sulfate Reagent 1 packet
REAGENT SETS
HI93751‑01 Reagents for 100 tests HI93751‑03 Reagents for 300 tests
For other accessories see page 72.
MEASUREMENT PROCEDURE
• Select the Sulfate method using the procedure described in the
Method Selection section (see page 28).
• Follow procedures in Section 6 to prepare the sample for analysis.
• Fill a cuvette with 10 mL of unreacted sample (up to the mark) and
replace the cap.
10 mL
66
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
Page 67
• Add one packet of HI93751‑0 Sulfate Reagent.
• Replace the cap and invert gently for 1 minute (about
30 inversions).
• Reinsert the cuvette into the instrument and close the lid.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively,
wait for 5 minutes and press Read. When the timer ends the meter will perform the reading.
2-
The instrument displays the concentration in mg/L of sulfate (SO
).
4
SULFATE
INTERFERENCES
Interferences may be caused by: Calcium (as CaCO
) above 20000 mg/L
3
Chloride (as Cl¯) above 40000 mg/L Magnesium (as MgCO3) above 10000 mg/L Silica (as SiO2) above 500 mg/L Color or suspended matter in large amounts will interfere: suspended matter should be removed by previous filtration. Organic matter in large amounts may impede the precipitation of barium sulfate.
67
Page 68
10. ERROR DESCRIPTIONS
The instrument shows clear warning messages when erroneous conditions appear and when measured values are outside the expected range. These messages are described below.
No Light: The light source is not functioning properly.
Light Leak: There is an excess amount of ambient light
ERROR DESCRIPTIONS
reaching the detector.
Inverted Cuvette: The sample and the zero cuvettes are inverted.
Light Low: The instrument cannot adjust the light level. Please check that the sample does not contain any debris.
Light High: There is too much light to perform a measurement. Please check the preparation of the zero cuvette.
68
Ambient temperature out of limit: The meter is too hot or too cold for an accurate measurement. Allow the meter to reach 10 °C to 40 °C (50 °F to 104 °F) before performing a measurement.
Ambient temperature changed: The temperature of the meter has changed significantly since the zero measurement has been performed. A zero measurement must be performed again.
Out of range: The measured value is outside the limits of the method.
Page 69
11. STANDARD METHODS Description Range Method
Ammonia LR 0.00 to 3.00 mg/L Nessler Ammonia MR 0.00 to 10.00 mg/L Nessler Ammonia HR 0.0 to 100.0 mg/L Nessler Calcium 0 to 400 mg/L Oxalate Magnesium 0 to 150 mg/L Calmagite Nitrate 0.0 to 30.0 mg/L Cadmium reduction Phosphate HR 0.0 to 30.0 mg/L Amino Acid Potassium 0.0 to 20.0 mg/L Tetraphenylborate
Sulfate 0 to 150 mg/L Barium Chloride
STANDARD METHODS
12. ACCESSORIES
12.1. REAGENT SETS Code Description
HI93700‑01 100 ammonia LR tests HI93700‑03 300 ammonia LR tests HI93715‑01 100 ammonia MR tests HI93715‑03 300 ammonia MR tests HI93717‑01 100 phosphate HR tests HI93717‑03 300 phosphate HR tests HI93728‑01 100 nitrate tests HI93728‑03 300 nitrate tests HI93733‑01 100 ammonia HR tests HI93733‑03 300 ammonia HR tests HI93750‑01 100 potassium tests HI93750‑03 300 potassium tests HI93751‑01 100 sulfate tests HI93751‑03 300 sulfate tests HI937520‑01 50 magnesium tests
ACCESSORIES
69
Page 70
Code Description
HI937520‑03 150 magnesium tests HI937521‑01 50 calcium fresh water tests HI937521‑03 150 calcium fresh water tests
ACCESSORIES
12.2 pH ELECTRODES Code Description
HI10530 Triple ceramic, double junction, low temperature glass, refillable pH electrode with conical tip and temperature sensor HI10430 Triple ceramic, double junction, high temperature glass, refillable pH electrode with temperature sensor HI11310 Glass body, double junction, refillable pH/temperature electrode HI11311 Glass body, double junction, refillable pH/temperature electrode with enhanced diagnostics HI12300 Plastic body, double junction, gel filled, non refillable pH/temperature electrode HI12301 Plastic body, double junction, gel filled, non refillable pH/temperature electrode with enhanced diagnostics HI10480 Glass body, double junction with temperature sensor for wine analysis FC2320 Double junction, open reference, non refillable, electrolyte viscolene, PVDF body with conical tip, pH/temperature electrode FC2100 Double junction, open reference, non refillable, electrolyte viscolene, glass body with conical tip, pH/temperature electrode FC2020 Double junction, open reference, non refillable, electrolyte viscolene, PVDF body with conical tip, pH/temperature electrode
70
Note: The enhanced diagnostics information are not displayed by meter.
Page 71
12.3 pH SOLUTIONS BUFFER SOLUTIONS Code Description
HI70004P pH 4.01 Buffer Sachets, 20 mL (25 pcs.) HI70007P pH 7.01 Buffer Sachets, 20 mL (25 pcs.) HI70010P pH 10.01 Buffer Sachets, 20 mL (25 pcs.) HI7001L pH 1.68 Buffer Solution, 500 mL HI7004L pH 4.01 Buffer Solution, 500 mL HI7006L pH 6.86 Buffer Solution, 500 mL HI7007L pH 7.01 Buffer Solution, 500 mL HI7009L pH 9.18 Buffer Solution, 500 mL HI7010L pH 10.01 Buffer Solution, 500 mL HI8004L pH 4.01 Buffer Solution in FDA approved bottle, 500 mL HI8006L pH 6.86 Buffer Solution in FDA approved bottle, 500 mL HI8007L pH 7.01 Buffer Solution in FDA approved bottle, 500 mL HI8009L pH 9.18 Buffer Solution in FDA approved bottle, 500 mL HI8010L pH 10.01 Buffer Solution in FDA approved bottle, 500 mL
ACCESSORIES
ELECTRODE STORAGE SOLUTIONS
HI70300L Storage Solution, 500 mL HI80300L Storage Solution in FDA approved bottle, 500 mL
ELECTRODE CLEANING SOLUTIONS
HI70000P Electrode Rinse Sachets, 20 mL (25 pcs.) HI7061L General Cleaning Solution, 500 mL HI7073L Protein Cleaning Solution, 500 mL HI7074L Inorganic Cleaning Solution, 500 mL HI7077L Oil & Fat Cleaning Solution, 500 mL HI8061L General Cleaning Solution in FDA approved bottle, 500 mL HI8073L Protein Cleaning Solution in FDA approved bottle, 500 mL HI8077L Oil & Fat Cleaning Solution in FDA approved bottle, 500 mL
71
Page 72
ELECTRODE REFILL ELECTROLYTE SOLUTIONS
HI7082 3.5M KCl Electrolyte, 4x30 mL, for double junction electrodes HI8082 3.5M KCl Electrolyte in FDA approved bottle, 4x30 mL, for double junction electrodes.
ACCESSORIES
12.4. OTHER ACCESSORIES Code Description
HI72083300 carrying case HI731318 cloth for wiping cuvettes (4 pcs.) HI731331 glass cuvettes (4 pcs.) HI731335N cap for cuvette (4 pcs.) HI731340 200 µL automatic pipette HI731341 1000 µL automatic pipette HI731342 2000 µL automatic pipette HI740034P cap for 100 mL beaker (10 pcs.) HI740036P 100 mL plastic beaker (10 pcs.) HI740038 60 mL glass bottle and stopper HI740142P 1 mL graduated syringe (10 pcs) HI740143 1 mL graduated syringe (6 pcs.) HI740144 pipette tip (6 pcs.) HI740157P plastic refilling pipette (20 pcs.) HI740220 25 mL graduated glass vial (2 pcs.) HI740223 170 mL plastic beaker HI740224 170 mL plastic beaker (12 pcs.) HI740225 60 mL graduated syringe HI740226 5 mL graduated syringe HI740227 filter assembly HI740228 filter discs (25 pcs.) HI 740229 100 mL graduated cylinder DEMI‑02 demineralizer HI75110/220E USB power adapter, European plug
72
Page 73
Code Description
HI75110/220U USB power adapter, USA plug HI76404A electrode holder HI83325‑11 CAL Check cuvette kit for HI83325 HI83300‑100 Sample preparation kit consisting of activated carbon for 50 tests,
demineralizer bottle for 10 L of water, 100 mL graduated beaker with cap, 170 mL graduated beaker with cap, 3 mL pipette, 60 mL syringe, 5 mL syringe, graduated cylinder, spoon, funnel, filter paper (25 pcs.).
HI920015 USB to micro USB cable connector HI93703‑50 cuvette cleaning solution (230 mL) HI93703‑55 activated carbon (50 pcs.)
ACCESSORIES
13. ABBREVIATIONS
EPA: US Environmental Protection Agency °C: degree Celsius °F: degree Fahrenheit µg/L: micrograms per liter (ppb) mg/L: milligrams per liter (ppm) g/L: grams per liter (ppt) mL: milliliter GLP good laboratory practice UHR ultra high range ULR ultra low range HR: high range MR: medium range LR: low range PAN: 1‑(2‑pyridylazo)‑2‑naphtol TPTZ: 2,4,6‑tri‑(2‑pyridyl)‑1,3,5‑triazine
ABBREVIATIONS
73
Page 74
Recommendations for Users
Before using this product make sure that they are entirely suitable for your specific application and for the environment in which they are used. Operation of these instruments may cause unacceptable interferences to other electronic equipment. Take all necessary steps to correct such interferences. Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance. To avoid damages or burns, do not put the instrument in a microwave oven. For yours and the instrument safety do not use or store the instrument in hazardous environments.
Page 75
Warranty
The HI83325 is warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered.
If service is required, contact your local Hanna Instruments Office. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization (RGA) number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packed for complete protection.
Hanna Instruments reserves the right to modify the design, construction or appearance of its products
without advance notice.
Page 76
World Headquarters
Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA www.hannainst.com
Local Office
Hanna Instruments USA 270 George Washington Highway Smithfield, RI 02917 Phone: 800.426.6287 Fax: 401.765.7575 e‑mail: tech@hannainst.com
MAN83325 01/17‑1
Printed in ROMANIA
Loading...