Hanna Instruments HI772 Instruction Manual

Page 1
Please examine this product carefully. Make sure that the instrument is not damaged. If any damage occurred during shipment, please contact your local Hanna Instruments Office. Each HI772 meter is supplied complete with:
• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 25 tests
• 1 mL syringe with tip
• 1.5V AAA Battery (1 pc.)
Light High: There is too much light to perform a measurement. Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement. Please check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0.0” indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you use the same cuvette for reference (zero) and measurement. Over Range: A flashing value of the maximum concentration indicates the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must
be replaced. Once this indication is displayed, normal operation of the instrument will be interrupted. Change the battery and restart the meter.
For more details about spare parts and accessories see “Accessories”.
HI772
Marine Alkalinity
INSTRUCTION MANUAL
Thank You
Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instrument.
For more information about Hanna Instruments and our products, visit www.hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail us at tech@hannainst.com
Preliminary Examination
Specifications
Range 0.0 to 20.0 dKH Resolution 0.1 d KH Accuracy ±0.3 dKH ±5% of reading @ 25 ˚C / 77 ˚F Light So urce light emit ting diode @ 610 nm Light Detector silicon photocell
Method
Colori metric method. The reac tion causes a dis tinctive range of
colors from yel low to green to gree nish blue to develop Environment 0 to 50 °C (32 to 122 °F); RH max 95% non -condensing Batt ery Type 1.5V AAA (1 pc.) Auto-off af ter 10 minutes of non-use Dimensions 86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”) Weight 64 g (2.3 oz)
Functional Description
Cover
Cuvet te with cap
Cuvette holder
LCD Liquid Crystal Display
Operation Button
Errors and Warnings
Page 2
Turn the meter on by pressing the button. All segments will be displayed. When the display shows “Add”, “C.1” with “Press” blinking, the meter is ready.
Fill the cuvette with 10 mL of unreacted sample and replace the cap. Place the cuvette into the meter and close the meter’s cap.
Press the button. When the display shows “Add”, “C.2” with “Press” blinking the meter is zeroed.
Remove the cuvette from the meter and unscrew the cap. Using a 1 mL syringe carefully add exactly
1.00 mL of HI772S Reagent to the sample. Replace the cap and gently invert 5 times. Place the cuvette back into the meter.
Note: Pay attention not to spill reagent otherwise full color development may be inhibited.
Press the button. The instrument directly displays the concentration of alkalinity in dKH. Alkalinity conversion: 1 dKH = 17.86 ppm CaCO
3
= 0.358 meq/L
The meter automatically turns off after 10 minutes.
10
mL
• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it must be dry outside, and completely free of fingerprints, oil and dirt. Wipe it thoroughly with HI731318 or a lint-free cloth prior to insertion.
• Shaking the cuvette can generate bubbles, causing higher readings. To obtain accurate measurements, remove bubbles by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand for too long after reagent is added, as accuracy will be affected.
• After the reading it is important to immediately discard the sample, otherwise the glass might become permanently stained.
HI772
To save the battery, the instrument shuts down after 10 minutes of non-use. One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the instrument will display “bAd” then ”bAt” for 1 second and then turns off. To restart the instrument, the battery must be replaced with a new one. To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.
• Remove the battery from its location and replace it with a new one, inserting the negative
end first.
• Insert the battery cover and replace the screw with a screwdriver.
IST772 02 /17
Measurement Procedure
HI772S
Tips for an Accurate Measurement
Battery Management
Reagent sets
HI77 2-26 Reagent for 25 M arine Alkalinit y tests, 1 sy ringe and 1 tip
Other Accessories
HI 772-11 Marine Alkalinity Certif ied Standard Kit HI731318 Cloth for w iping cuvett es (4 pcs.) HI73 1321 Glass c uvetes (4 pcs.) HI73 1225 Black Cuvette C ap for Checker HC (4 pc s.) HI74 0028 P 1.5V A AA batter ies (12 pcs.) HI740142P 1 mL gradua ted syringe (10 pcs.) HI93703-50 Cuvette cleaning s olution (230 mL)
Accessories
Hanna I nstrument s reserves the r ight to modif y the design, c onstructi on, or appeara nce of its prod ucts without advance notice.
Warranty
HI772 is warranted for a period of one year after date of purchase against defects in
workmanship and materials when used for their intended purpose and maintained according to instructions. This warrant y is limited to repair or replacement f ree of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service is required, contact your local Hanna Instruments Office. If under warrant y, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Return Goods Authorization number from the Technical Ser vice department and then send it with shipping costs p repaid. When shipping any ins trument, make sure it is properly packed for complete protection.
Recommendations for Users
Before using Hanna Instruments product s, make sure that they are entirely suitable for your specific application and for the environment in which they are used. Operation of these instruments may cause unacceptable interferences to other electronic equipment, thus requiring the operator to take all necessary steps to correct such interferences. Any variation introduced by the user to the supplied equipment may degrade the instrument ’s EMC performance. To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument safety do not use or store the instrument in hazardous environments.
Loading...