Hanna Instruments HI 720 User Manual

Instruction Manual
HI 720
Conductivity
Process Controller
with Inductive Probe
www.hannainst.com
1
TABLE OF CONTENTS
PRELIMINARY EXAMINATION ............................................................................5
MODEL IDENTIFICATION..................................................................................5
GENERAL DESCRIPTION & THEORY OF OPERATION ..........................................6
SPECIFICATIONS ..............................................................................................9
FUNCTIONAL DESCRIPTION ........................................................................... 10
INSTALLATION ................................................................................................12
OPERATIONAL GUIDE.................................................................................... 14
• SETUP MODE ..........................................................................................14
• CALIBRATION MODE ............................................................................... 34
• CONTROL MODE .................................................................................... 37
• IDLE MODE ............................................................................................. 43
• LAST CALIBRATION DATA VIEWING MODE ................................................. 43
• DIAGNOSTIC MODE ............................................................................... 44
TEMPERATURE COMPENSATION ..................................................................... 46
CONCENTRATION CURVES ............................................................................ 48
HOLD MODE ................................................................................................ 50
IN-LINE CLEANING ........................................................................................ 52
COMMUNICATION ........................................................................................54
• PC COMMUNICATION ............................................................................ 54
• SHORT MESSAGING SERVICE (SMS) .......................................................... 67
• MODEM CONNECTION ..........................................................................71
ERRORS - FAULT CONDITIONS ....................................................................... 73
ALARM - ERROR CONFIGURATION ................................................................. 74
SELFTEST PROCEDURES ..................................................................................77
ACCESSORIES ................................................................................................ 80
2
3
Dear Customer,
Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instrument. It will provide you with the necessary information for correct use of the instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com or see the back cover for our worldwide contact list. This instrument is in compliance with the CE directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any notice­able damage, notify your Dealer or the nearest Hanna Customer Service Center immediately.
Note Save all packing materials until you are sure that the instrument
functions correctly. Any damaged or defective items must be returned in their original packing materials together with the supplied acces­sories.
WARRANTY
All Hanna Instruments meters are guaranteed for two years (sensors, electrodes and probes for six months) against defects in workmanship and
materials when used for their intended purpose and maintained according to instructions.
This warranty is limited to repair or replacement free of charge. Damage due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instru­ment. If under warranty, report the model number, date of purchase, serial num­ber and the nature of the failure.
If the repair is not covered by the warranty, you will be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Customer Service department and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is properly packaged for complete protection.
MODEL IDENTIFICATION
The models HI 720 XYZ- The meaning of the last letters is according to the following scheme:
αα
α are conductivity controllers.
αα
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance notice.
4
5
GENERAL DESCRIPTION & THEORY OF OPERATION
This instrument allows conductivity measurements without any electrical contact between electrodes and process fluid.
The measurement is based on inductive coupling of two toroidal transformers by the liquid.
The instrument supplies a high frequency, reference voltage to the “Drive Coil”, and a strong magnetic field is generated in the toroid.
The liquid passes through the hole in the toroid and can be considered as one­turn secondary winding. The magnetic field induces a voltage in this liquid wind­ing, the current induced in the flow is proportional to this voltage, and the con­ductance of the liquid one-turn winding is according to the Ohm’s law.
The conductance is proportional to the specific conductivity and a constant fac­tor determined by the sensor geometry and installation.
The liquid also passes through the second toroid and therefore the liquid turn can be considered as a primary winding of the second toroidal transformer. The current in the liquid will create a magnetic field in the second toroid, and the induced current can be measured as an output.
The output current of this “Receive Coil” is therefore proportional to the specific conductivity of process liquid.
For an inductive cell, the cell constant is defined as the measured conductivity, obtained by making a loop through the sensor with a resistor R, multiplied by that R value.
The cell constant depends only on the sensor geometry. However, when the probe is immersed in a liquid, the induced current in the solution is affected by the piping or any other container where the probe is inserted. This effect is negligible when there is an area of at list 3 cm of liquid around the cell.
Otherwise, it is necessary to multiply measurements by the installation factor:
Conductivity = (cell constant)(installation factor)/(measured resistance)
The installation factor is < 1 for conductive piping/containers, and > 1 for non­conductive piping/containers.
Since this type of sensor has no electrodes, common problems due to the elec­trode contact, such as polarization and contamination, will not affect the perfor­mance of our electrodeless sensor.
The probe working life is extremely long and the sensor maintenance will be an exceptional operation.
The controllers are equipped with a graphic display, easy to understand as your cellular phone. Simple messages guide the users through all operations and parameter setting.
6
7
The main features of the HI 720 controller series include:
• Measurement and control of conductivity or concentration
• Concentration can be measured as usual TDS (fixed ratio) or through custom conductivity/temperature/concentration curves
• Customizing temperature coefficient table and NaCl temperature compensa­tion according to IEC 746-3, in addition to the standard linear compensation
• Auto-ranging
• Display reading offset adjustment for temperature
• Temperature level alarm
• Cleaning in place activated through two cleaning commands, or triggered by a variety of events
• Hold management, including a digital input to enter the hold mode through an external trigger
• Conductivity probe check
• Digital transmitter input
• Pt100 or Pt1000 temperature sensor with automatic recognition and dam­age test
• Calibration time-out
• Logging of the last 100 error, configuration and cleaning events
• Alarm fault current (3.6 mA or 22 mA)
• SMS messages sending
• Alarm configuration can be customized: different errors can lead different actions (alarm relay activation, fault current, hold mode, automatic cleaning, SMS message)
• RS485 communication with additional capabilities, such as error log file down­loading and cleaning commands
• Daily programmable control timing
• Diagnostic features
• Password protection
SPECIFICATIONS
Range 0 to 2000 mS/cm (auto-ranging)
-30 to 130°C / -22 to 266°F
Resolution 1 μS/cm (0 to 1999 μS/cm)
0.01 mS/cm (2.00 to 19.99 mS/cm)
0.1 mS/cm (20.0 to 199.9 mS/cm) 1 mS/cm (200 to 2000 mS/cm)
0.1°C / 0.2°F
Accuracy ±2% f.s. (conductivity) / ±0.5°C / ±1°F Typical EMC Deviation ±2% f.s. (conductivity) / ±0.5°C / ±1°F Temp. Compensation Automatic or manual, -30 to 130°C Temperature Probe 3-wire or 2-wire Pt100 or Pt1000 sensor
with automatic recognition and damage test
Digital Input Digital Transmitter, Hold & Advanced Cleaning inputs Digital Output 1 digital insulated contact closed upon Hold mode Analog Output 1 or 2 independent outputs;
0-22 mA (configuring as 0-20 mA or 4-20 mA)
Digital Serial Output RS485 Dosing Relay 1, 2, 3 or 4 electromechanical relays SPDT;
5A-250 Vac, 5A-30 Vdc (resistive load);
fuse protected: 5A, 250 V fuse
Alarm Relay 1 electromechanical relay SPDT;
5A-250 Vac, 5A-30 Vdc (resistive load);
fuse protected: 5A, 250 V fuse
Installation Category II Power Supply 24 Vdc/ac, or
115 Vac or 230 Vac or 100 Vac ±10%, 50/60 Hz;
fuse protected: 400 mA, 250 V fast fuse
Power Consumption 10 VA Max Oscillation Frequency 8 MHz Environment 0 to 50°C (32 to 122°F); RH max 85% non-condensing Enclosure Single case 1/2 DIN Weight Approx. 1.6 kg (3.5 lb.)
8
9
FUNCTIONAL DESCRIPTION
1. Graphic display (128 x 64 dots)
2. DIAG key, to enter/exit diagnostic mode; to change the conductivity or
concentration range while in setup or calibration mode
3. SETUP key, to enter/exit setup mode
4. CAL DATA key, to enter/exit last calibration data viewing mode
5. CAL key, to enter/exit calibration mode
6. key, to increase current digit by 1, shift to next option, or move to next record while in diagnostic mode
7. key, to decrease current digit by 1, shift to previous option, or move to previous record while in diagnostic mode
8. key, to cyclically change temperature compensation option for conduc­tivity readings: compensation enabled (temperature is continuously dis­played) or compensation disabled (actual reading). Only displayed values are affected by this key; control and logging are not affected.
9. CFM key, to confirm current choice
10. Contact LEDs, each turns on when the corresponding relay is energized
11. Status (red) and alarm (green) LEDs
10
1. Connections for conductivity probe
2. Connections for Pt100/Pt1000 temperature sensor
3. Digital Transmitter input
4. Hold input
5. Advanced Cleaning input (optional)
6. RS485 output terminal
7. Hold output
8. Analog output #1
9. Analog output #2 (optional)
10. Power supply input
11. Relay #3 - for Advanced Cleaning feature (optional)
12. Relay #4 - for Advanced Cleaning feature (optional)
13. Relay #1 - first dosing terminal
14. Relay #2 - second dosing terminal (optional)
15. Alarm relay
11
INSTALLATION
Mechanical Dimensions
Electrical Connections
Power Input: connect a 3-wire power cable to the line (L), earth
(PE) and neutral (N) terminal connections. Power: 100 Vac-120 mA / 115 Vac-100 mA / 230 Vac-50 mA. Line Contact: 400 mA fuse inside. PE must be connected to ground; leakage current 1 mA.
Probe Connections (#1 at page 9): connect the Tx and Rx coaxial cables to
the relevant terminals, as indicated. Connect the probe cable shield to the
“Shield” terminal.
• Pt100/Pt1000 Terminals (#2 at page 9): these contacts connect the Pt100/
Pt1000 sensor for automatic temperature compensation.
If the conductivity probe features a built-in Pt100/Pt1000 sensor, connect the
related wires to pins 1, 2 and 3; pin 4 will not be used.
It is also possible to use a separate Pt100/Pt1000 temperature probe. In the
case of shielded wire, connect the shield to pin 4. In the case of a 2-wire sensor connect the Pt100/Pt1000 to pins 1 and 3, and short pins 2 and 3 with a jumper wire.
If the Pt100/Pt1000 probe has more than 2 wires, connect the two wires of
one end to pins 2 and 3 (pin 2 is an auxiliary input to compensate for the cable resistance) and one wire from the other end to pin 1. Leave the fourth wire unconnected, if present.
Note The instrument automatically recognizes the sensor type (Pt100 or
Pt1000).
Analog Outputs: when using shielded cable, if the shield is not connected at
the other end of the cable, then connect it to the “+” terminal, otherwise leave it floating.
Note All cables connected to rear panel should end with cable lugs.
Note A circuit breaker (rated 10 A max.) must be connected in close prox-
imity to the equipment, and in a position easy to reach by the opera­tor, for disconnection of the instrument and all devices connected to the relays.
12
13
OPERATIONAL GUIDE
The HI 720 process controller can operate in six main modes:
• Setup mode
• Calibration mode
• Control mode
• Idle mode
• Last calibration data viewing mode
• Diagnostic mode (active errors & event log file scrolling) All operating modes are described in the following sections.
SETUP MODE
The Setup mode allows the user to set all needed characteristics of the meter. To enter the mode, press the SETUP key and enter the password when the device is in idle or control mode. If the correct password is not entered, the user can view the setup parameters (except passwords, phone numbers and PIN code), but not modify them (and the device remains in control mode).
Entering the Password
• Press the SETUP key to enter the setup mode.
The display will show “0000” (default value),
with the first digit highlighted.
• Enter the first digit of the password by using the
up & down arrow keys.
• Move to the next digit with the key and enter the desired value as described
above. Continue for the last two digits. When the whole password has been
inserted, press CFM to confirm.
Entering Setup Items
• After the password has been confirmed, the dis-
play will show the name of the first setup group
(General Setting).
• Using the arrow keys it is possible to cycle
through all setup groups, while pressing SETUP
will exit the mode, and pressing CFM will con-
firm the desired selection.
• Once a group is selected, the display shows the
first parameter of the group together with its
current value.
14
The user can choose the desired parameter with the up & down arrow keys,
and confirm the selection by pressing CFM.
Note For some groups, it is necessary to enter several sublevels before
choosing a parameter value. Follow the displayed messages to pro­ceed, confirm or exit each screen.
• If a fixed set of values is available for the se­lected item, use the arrow keys to switch be­tween them. Otherwise, if a numeric value has to be entered, use the arrow keys to change the value of the highlighted digit and cycle through the number’s digits.
• Once a value is set, press the CFM key to con­firm, and the instrument will turn to the previ­ous screen.
Note After modifying a value in the concentration curve triplets or in
the conductivity/temperature couples for the temperature com-
pensation algorithm, then it is not possible to exit without saving. To return to the previous value, the number must be entered again.
• If a wrong value is confirmed, the “Invalid value!” or “Invalid choice!” message advises the user, the value is not accepted and the instrument will not switch to the next item until a correct value will be confirmed.
• When the last item of the group is reached, by pressing the CFM key again the display will show the group name. The group can then be changed by press­ing the up & down arrow keys.
Note If no activity is performed for about 5 minutes after entering the
setup mode, the mode is automatically exited and the instrument returns to the previous mode.
15
This table lists all available setup items together with their code, short descrip­tion, valid values and default setting.
Setup Item Description Valid Values Default
GENERAL SETTINGS
G.00 Measurement mode Conductivity, Conductivity
(see note 8) Concentration, TDS
G.01 Conductivity range Auto range, Auto range
1999 μS/cm,
19.99 mS/cm,
199.9 mS/cm, 2000 mS/cm
G.02 TDS range Auto range, Auto range
1000 ppm, 10.00 ppt,
100.0 ppt, 1000 ppt
G.03 Concentration unit Auto (ppm/ppt), Auto
(see note 31) ppm, ppt, % (ppm/ppt)
G.04 Concentration XXXX, XXX.x, XX.xx
format XX.xx, X.xxx G.05 TDS factor 0.00 to 1.00 0.50 G.06 Moving average 1 to 30 1
length (see note 30) G.10 Factory ID 0000 to 9999 0000 G.11 Process ID alias 00 to 99 00
RS485 address G.13 Language English, Italiano English G.14 LCD contrast 0 to 10 6 G.15 LCD backlight On, Off On G.98 Calibration 0000 to 9999 0000
password (see note 2) G.99 General password 0000 to 9999 0000
(see note 2)
TEMPERATURE
b.01 Compensation mode Automatic (ATC), ATC
(see notes 7, 20) Manual (MTC) b.02 Start manual temp. -30.0 to 130.0°C 25.0°C
(see notes 7, 20)
Setup Item Description Valid Values Default
TEMPERATURE
b.03 Temperature measure °C, °F °C
unit (see note 32)
b.10 Temp. compensation Linear, Linear
algorithm (note 28) NaCl
(IEC 746, table BII),
User: user defined table
b.11 Reference temp. 20°C, 25°C 25°C
(see note 28)
b.12 Temp. coefficient 0.00 to 20.00 % 2.00 %
(see notes 28, 29)
Temperature Compensation Table
b.31 Number of couples 2 to 10 2 b.32 Edit/view couple 1 to b.31 item value 1 b.33 User table: actual 0 to f.s. see note 33
conductivity value for the couple selected in b.32
b.34 User table: -30.0 to 130.0°C see note 33
temperature value for the couple selected in b.32
Temperature Level Alarm (see Note 36)
b.41 Max temperature -30.0 to 130.0°C 130.0°C b.42 Min temperature -30.0 to 130.0°C -30.0°C
Reading Offset
b.50 Actual reading measurement -10.0°C measurement
(for ATC only, to measurement +10°C (see note 15) see notes 15, 20)
b.51 Temperature offset -10.0°C to +10.0°C 0.0°C
(ATC only, notes 15, 20)
CONCENTRATION
d.00 Concentration curve 1 to 4 1
in use
Curve 1/2/3/4 Table (see note 34)
d.01 Number of triplets 1 to 25 1 d.02 Edit/view triplet 1 to d.01 value 1
16
17
Setup Item Description Valid Values Default
Curve 1/2/3/4 Table (see note 34)
d.03 Not compensated 0 to f.s. 0 μS/cm
conductivity value for
the triplet selected in d.02 d.04 Temperature value for -30.0 to 130.0°C 0.0°C
the d.02 triplet d..05 Concentration value 0 to f.s 0 ppm
for the d.02 triplet
CONTROL
C.00 Control enable On, Off Off
Setpoint 1
C.10 Setpoint 1 mode Disabled, On/Off low
(M1) On/Off high,
(see note 1) On/Off low,
PID high (if available), PID low (if available)
C.11 Setpoint 1 value 0 to f.s. 500 μS/cm
(S1) (see note 1) C.12 Setpoint 1 0 to f.s. 20 μS/cm
hysteresis (H1) (see note 1) C.13 Setpoint 1 Up to f.s and 0 20 μS/cm
deviation (D1) (see note 1) C.14 Setpoint 1 reset time 0.1 to 999.9 minutes 999.9 min. C.15 Setpoint 1 rate time 0.0 to 999.9 minutes 0.0 min.
Setpoint 2
C.20 Setpoint 2 mode Disabled, On/Off low
(M2) On/Off high,
(see note 1) On/off low,
PID high (if available), PID low (if available)
C.21 Setpoint 2 value 0 to f.s. 1500 μS/cm
(S2) (see note 1) C.22 Setpoint 2 0 to f.s. 20 μS/cm
hysteresis (H2) (see note 1) C.23 Setpoint 2 Up to f.s and 0 20 μS/cm
deviation (D2) (see note 1)
18
Setup Item Description Valid Values Default
Setpoint 2
C.24 Setpoint 2 reset time 0.1 to 999.9 minutes 999.9 min. C.25 Setpoint 2 rate time 0.0 to 999.9 minutes 0.0 min.
Alarms
C.30 Low alarm value 0 to f.s 100 μS/cm
(LA) (see note 18) (see note 1)
C.31 High alarm value 0 to f.s 1900 μS/cm
(HA) (see note 18) (see note 1)
C.32 Max relay ON time 1 to 60 minutes 60 min.
(then an alarm is generated)
C.33 Alarm mask time 00:00 to 30:00 min. 00:30
(see notes 18, 27)
C.34 Alarm hysteresis 0 to f.s. 20 μS/cm
(AH) (see note 18)
Programmable Control
C.41 Hold time start (note 6) 00:00 to 23:59 00:00 C.42 Hold time stop (note 6) 00:00 to 23:59 00:00 C.51 Monday hold mode Off: disabled, Off
On: enabled
C.52 Tuesday hold mode Off: disabled, Off
On: enabled
C.53 Wednesday hold Off: disabled, Off
mode On: enabled
C.54 Thursday hold Off: disabled, Off
mode On: enabled
C.55 Friday hold mode Off: disabled, Off
On: enabled
C.56 Saturday hold mode Off: disabled, Off
On: enabled
C.57 Sunday hold mode Off: disabled, Off
On: enabled
Other Parameters
C.60 Proportional control 01:00 to 30:00 min. 05:00 min.
period (if available)
19
Setup Item Description Valid Values Default
Other Parameters
C.70 Hold mode end delay 00 to 99 seconds 00 sec C.80 On/Off control 00:00 to 30:00 min. 00:00 min
action delay (note 26)
OUTPUT Relays
O.01 Relay 1 mode Disabled, Control-
(see notes 16, 17) Control-setpoint 1, setpoint 1
Control-setpoint 2, Simple cleaning, Hold mode
O.02 Relay 2 mode Disabled, Control-
(see notes 16, 17) Control-setpoint 1, setpoint 2
Control-setpoint 2, Simple cleaning, Hold mode
O.03 Relay 3 mode Disabled, Disabled
(see notes 12, 16) Simple cleaning,
Advanced cleaning, Hold mode
O.04 Relay 4 mode Disabled, Disabled
(see notes 12, 16) Simple cleaning,
Advanced cleaning, Hold mode
O.05 Hold digital output Disabled or Hold mode
Hold mode
Analog Output 1
O.10 Analog output 1 Recorder or Recorder
mode (notes 1, 17) Control-setpoint 1
O.11 Analog output 1 0-20 mA or 4-20 mA
type 4-20 mA
O.12 Analog output 1 0 to f.s. 0 μS/cm
minimum value (O_VARMIN1 (O_VARMIN1) O_VARMAX1 -5% f.s. for
O_VARMAX1, O_VARMIN1 O_HOLD1 O_VARMAX1)
Setup Item Description Valid Values Default
Analog Output 1
O.13 Analog output 1 0 to f.s. 1999 μS/cm
maximum value (O_VARMIN1 (O_VARMAX1) O_VARMAX1 -5% f.s. for
O_VARMAX1, O_VARMIN1 O_HOLD1 O_VARMAX1)
O.14 Analog output 1 User selected value, Previous value
hold mode (note 13) Previous value
O.15 Analog output 1 0 to f.s. 1000 μS/cm
hold mode value (it must be within the (O_HOLD1) O_VARMIN1 to
O_VARMAX1 interval)
Analog Output 2
O.20 Analog output 2 Recorder (temperature), Recorder
mode (notes 1, 17) Control-setpoint 2
(conductivity/concentration /temperature)
O.21 Analog output 2 0-20 mA or 4-20 mA
type 4-20 mA
O.22 Analog output 2 -30.0 to 130.0°C 0.0°C
minimum value (O_VARMIN2 (O_VARMIN2) O_VARMAX2 -10°C,
O_VARMIN2 O_HOLD2 O_VARMAX2)
O.23 Analog output 2 -30.0 to 130.0°C 100.0°C
maximum value (O_VARMIN2 (O_VARMAX2) O_VARMAX2 -10°C,
O_VARMIN2 O_HOLD2 O_VARMAX2)
O.24 Analog output 2 User selected value, Previous value
hold mode (note 13) Previous value
O.25 Analog output 2 -30.0 to 130.0°C 25.0°C
hold mode value (it must be within the (O_HOLD2) O_VARMIN2 to
O_VARMAX2 interval)
20
21
Setup Item Description Valid Values Default
INPUT
I.00 Input selection Inductive probe or Inductive
(note 10) Digital transmitter probe
I.03 Digital transmitter 00 to 99 00
address (active if I.00=Digital transmitter)
I.04 Digital transmitter 1200, 2400, 4800, 19200 bps
baud rate 9600 or 19200 bps I.10 Calibration timeout 0 to 12 months 12 months I.11 Life check Disabled, Disabled
(see notes 9, 20) 1 hour, 2 hours,
or 4 hours
REAL TIME CLOCK (see Note 4)
r.00 Current day 01 to 31 from RTC r.01 Current month 01 to 12 from RTC r.02 Current year 2000 to 2099 from RTC r.03 Current time 00:00 to 23:59 from RTC
COMMUNICATION
P.00 Connection type PC, PC
HI 504900 GSM module, HI 504901 GSM supervisor, HI 504902 Modem
P.01 Baud rate 1200, 2400, 4800, 19200 bps
9600 or 19200 bps P.02 Call answer enable On, Off Off P.03 Modem country Insert the phone 000
code country code for the
HI 504902 modem
(see note 35) P.11 PIN code 0000 to 9999 0000 P.12 Phone number #1 + --------------------
(see note 21) (“Not set” means no number entered)
P.13 Phone number #2 + --------------------
(see note 21) (“Not set” means no number entered)
Setup Item Description Valid Values Default
COMMUNICATION
P.14 Remaining SMS 000 to 200, or 222 100
(see note 22) P.15 Repeated SMS 0 to 5 (see note 23) 2 P.16 SMS delay between 05 to 60 minutes 10 minutes
messages (see note 24) P.17 SIM card expiration 01 to 31 01
day (see note 25) P.18 SIM card expiration 01 to 31 01
month (see note 25) P.19 SIM card expiration 2000 to 2099 2010
year (see note 25)
CLEANING Simple Cleaning
L.00 Rinsing time 5 to 99 seconds 20 seconds L.01 Pause time 10 to 9999 minutes 1440 min.
Advanced Cleaning (see Note 19)
L.10 Pre-rinsing time 0 to 99 seconds 20 seconds L.11 Cleaning time 0 to 99 seconds 10 seconds L.12 Rinsing time 5 to 99 seconds 20 seconds L.13 Pause time 10 to 9999 minutes 1440 min. L.14 Minimum pause time 10 to L.13 minutes 10 minutes L.15 Cleaning trigger Timer only, Timer only
External only, Timer and external,
Timer masked by external L.16 Repeated cycles no. 0 to 10 0 L.17 Economy cycle no. 0 to 10 0
ERROR CONFIGURATION (see Note 5)
E.00 High alarm (00) Alarm relay -------------> On
22 mA fault current----> On
3.6 mA fault current---> Off
Auto cleaning ----------> Off
SMS sending -----------> Off
22
23
Setup Item Description Valid Values Default
ERROR CONFIGURATION
E.01 Low alarm (01) Alarm relay -------------> On
22 mA fault current----> Off
3.6 mA fault current---> On Auto cleaning ----------> Off SMS sending -----------> Off
E.02 Max relay ON time Alarm relay -------------> On
error (02) 22 mA fault current----> On
3.6 mA fault current---> Off Auto cleaning ----------> Off SMS sending -----------> Off
E.03 Life check error (03) Alarm relay -------------> On
(see note 9) 22 mA fault current----> On
3.6 mA fault current---> Off Hold mode --------------> On Auto cleaning ----------> Off SMS sending -----------> Off
E.10 Conductivity input Alarm relay -------------> On
overflow (10) 22 mA fault current----> On
3.6 mA fault current---> Off Hold mode --------------> Off Auto cleaning ----------> Off SMS sending -----------> Off
E.12 Calibration timeout Alarm relay -------------> Off
(12) 22 mA fault current----> Off
3.6 mA fault current---> Off SMS sending -----------> Off
E.20 Broken temperature Alarm relay -------------> On
probe (20) 22 mA fault current----> On
3.6 mA fault current---> Off Hold mode --------------> Off SMS sending -----------> Off
Setup Item Description Valid Values Default
ERROR CONFIGURATION
E.21 Temperature level Alarm relay -------------> On
error (21) 22 mA fault current----> On
3.6 mA fault current---> Off
SMS sending -----------> Off E.40 Digital transmitter Alarm relay -------------> On
error (40) 22 mA fault current----> On
3.6 mA fault current---> Off
Hold mode --------------> On
SMS sending -----------> Off E.50 GSM/Modem Alarm relay -------------> On
module error (50) 22 mA fault current----> On
3.6 mA fault current---> Off
E.60 Temp. compensation Alarm relay -------------> On
error (60) 22 mA fault current----> On
3.6 mA fault current---> Off
SMS sending -----------> Off E.61 Temperature out of Alarm relay -------------> On
concentration table 22 mA fault current----> On (61) 3.6 mA fault current---> Off
SMS sending -----------> Off E.62 Conductivity out of Alarm relay -------------> On
concentration table 22 mA fault current----> On (62) 3.6 mA fault current---> Off
Auto-cleaning ----------> Off
SMS sending -----------> Off E.63 Concentration out Alarm relay -------------> On
of user concentration 22 mA fault current----> On table (63) 3.6 mA fault current---> Off
SMS sending -----------> Off E.90 Power reset (90) Alarm relay -------------> Off
22 mA fault current----> On
3.6 mA fault current---> Off
SMS sending -----------> Off
24
25
Loading...
+ 30 hidden pages