Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using this instrument. This
manual will provide you with the necessary information for correct use of this
instrument, as well as a precise idea of its versatility.
Table of
Contents
Included ...............................................................................................................................................3
Keypad Function .......................................................................................................................................................................7
Guide to Indicators ..................................................................................................................................................................8
Setting Up edge® .................................................................................................................................................................... 9
Setting Up edge® ..................................................................................................................................................................10
General Setup ...................................................................................................................................................................11-12
Logging Function ...........................................................................................................................................................13-15
Viewing Logged Data ................................................................................................................................................... 15-17
PC & Storage Interface .................................................................................................................................................18-19
Basic vs Standard pH Mode ...............................................................................................................................................20
pH Meter Congurations .............................................................................................................................................21-27
pH GLP Information ......................................................................................................................................................29-30
Basic vs Standard EC Mode ...............................................................................................................................................33
EC Meter Conguration ............................................................................................................................................... 33-36
EC/TDS GLP Information ............................................................................................................................................39-40
NaCl % GLP Information .............................................................................................................................................41-42
Dissolved Oxygen Meter Conguration ...............................................................................................................48-49
Dissolved Oxygen GLP Information ........................................................................................................................51-52
Remove the instrument and accessories from the packaging and verify damage
has not occurred during shipping. Remove protective lm from meter. Notify
your nearest Hanna Customer Service Center if damage is observed.
Each instrument is supplied with:
edge®
Bench cradle
Wall cradle
Electrode holder
USB cable
5 Vdc Power Adapter
Instruction Manual
Quality Certicate
Model Specic Components include:
HI 2020 (pH)HI 2030 (EC)HI 2040 (DO)
Included
HI 11310: Digital pH Electrode
with integrated temperature
sensor
pH 4.01, 7.01 & 10.01 Buer
Sachets
HI 700601 General Purpose
Cleaning Solution
HI 763100: Digital 4 ring
conductivity probe with
integrated temperature
sensor
1413 µS/cm and 12.88 mS/cm
Conductivity Standard
Sachets
HI 764080: Digital
polarographic dissolved
oxygen probe with integrated
temperature sensor
HI 7041S Electrolyte
for DO probe
2 DO membrane caps for
HI 764080 DO probe
2 replacement O-rings for
membrane caps
Note: Save all packing material until you are sure that the instrument works
correctly. Any defective item must be returned in its original packing.
Before using this product, make sure that it is entirely suitable for your specic
application and for the environment in which it is used.
Operation of this instrument may cause interference to other electronic
equipment, requiring the operator to take steps to correct interference. Any
variation introduced by the user to the supplied equipment may degrade the
instrument’s EMC performance.
To avoid damages or burns, do not put the instrument in microwave ovens.
For your and the instrument’s safety, do not use or store the instrument in
hazardous environments.
Safety
Measures
www.GlobalTestSupply.com
3
Description
Product
Diagram
edge® enables the user to make fast, accurate measurements of commonly
measured laboratory parameters using one of the Hanna edge® digital sensors
for pH, Conductivity or Dissolved Oxygen. Each digital sensors has a unique
serial number and the parameter that is measures is automatically identied
by the meter. Once connected to the meter, the sensor(s) are ready to measure
their parameter along with temperature.
The user interface permits you to adapt edge® to your exact measurement
requirements. The intuitive design simplies conguration, calibration,
measurement, data logging and transfer of data to a USB thumb drive
or computer. edge® also oers a basic operation mode that streamlines
measurement conguration and is useful for many routine applications. (Every
feature and measurement detail is designed to give you an edge in measurement
technology.)
edge® is versatile in many ways. The slim meter and probe can be used as
a portable device (using its rechargeable battery) or used in its bench or wall
cradles (that also power the meter) as a line-powered laboratory instrument.
• Sleek, clean, intuitive design
• Internal clock and date
• Adjustable resolution for pH and EC measurements
• Auto parameter recognition
• Dedicated GLP key
• GLP data included with logged data
• Basic mode for simplied operation
• Simplied data transfer to a PC
• Up to 8 hour battery life when used as a portable device
Side & Back
Diagram
4
View
www.GlobalTestSupply.com
Product
Diagram
1. Liquid Crystal Display (LCD)
2. Capacitive Touch Keypad
3. 3 mm jack input for edge® digital
probes
4. Top mounted ON/OFF button
www.GlobalTestSupply.com
5. Micro USB device connection for
power or PC interface
6. Standard USB host connection
for data transfer to a USB
thumb-drive
Diagram5
Probe
Diagram
Probes
pH Electrode
• Probes process signal directly for noise free measurements
• Auto sensor recognition
• Store calibration specic data from the last calibration
• Are built with materials suitable for use in chemical analysis
• Have integrated temperature measurement
• Incorporate a 3 mm jack termination
• Unique serial ID in every probe for traceability
Diagram
6
EC ElectrodeDO Electrode
www.GlobalTestSupply.com
Keypad
Function
1. CAL/MODIFY - Used to enter
and exit calibration mode. In
SETUP, used to initiate changes of a
conguration setting.
2. GLP/CFM - Used to display GLP
calibration information. In SETUP,
used to conrm change made. In
calibration, used to accept calibration
points.
3. RANGE/u - Used to select
measurement range. In SETUP, used
to move to right in pick list. In log RCL,
used to view GLP data for a data point.
4. SETUP/CLR - Used to enter/exit
SETUP mode. During calibration, used
to clear previous calibration data. In
log RCL, used to clear log records.
www.GlobalTestSupply.com
5. / - Used to scroll through
SETUP menu. Used to change
selection when modifying a
parameter in SETUP.
6. RCL (Recall) - Used to view
logged records or view % log memory
used.
7. LOG - Used to log data by
manual log on demand or manual log
on stability or to start/stop interval
logging.
Note: You can increase/
decrease the speed to
change the value of a
parameter.
Proceed as follows:
Press and hold down the
p or q key, then slide the
nger toward the double
apex to increase the speed
that a value changes.
Diagram7
Guide to
Indicators
1. Mode tags
2. Conrm tag
3. USB connection status
4. pH electrode diagnostics
5. Probes symbol
6. Battery symbol
7. Arrow tags, displayed when they
are available
8. pH calibration buers used
The third line of the LCD (9) is a dedicated message line. During measurement
the user may use the pq arrows to select desired message. Options include
date, time, calibration data, battery charge or no message. If a measurement
error or log status change occurs during measurement, the third line will display
a pertinent message.
Diagram
8
9. Third LCD line, message area
10. Labels
11. Second LCD line, temperature
measurement
12. Temperature units
13. Temp. Compensation status
14. Measurement line
15. Measurement units
16. Stability Indicator
www.GlobalTestSupply.com
The main operating modes of edge® are setup, calibration, measurement,
data logging, and data export. Follow this general outline of steps to get you
started. The following topics are expanded upon in the sections that follow in
this manual.
1. Familiarize yourself with the design features of this unique meter.
2. Decide how the meter will be used and set up the wall or bench cradle in a
clean area near line power.
3. Turn edge® on using the ON/OFF button located on the top of the meter.
4. Plug in the probe required for measurement.
5. SETUP the measurement parameters required for the measurement you
will be making.
6. Calibrate the sensor/probe.
You are now ready for measurements.
Bench Cradle Setup
Insert electrode holder arm into the
post on the pivoting base.
Connect the probe connector to the
socket located at the bottom of the
instrument.
Setting Up
edge®
Slide edge® into the cradle while
positioning the probe cable behind the
cradle. Put the probe/sensor into the
electrode holder and secure cable in clips.
Connect the power adapter cable to the
rear socket of the bench cradle. Connect
the other end to the power adapter and
plug into line power. Verify the battery
icon indicates charging.
Wall Cradle Setup
Choose suitable wall location. (Use 2.5 mm
or US #3 bit). Fasten the wall cradle using
the provided screws. Snap cover over
screw heads.
www.GlobalTestSupply.com
Setup/Installation9
Setting Up
edge®
Connect the power adapter cable to the
bottom socket of the wall cradle. Connect
the other end to the power adapter and
plug into line power.
Connect the 3 mm probe jack to the socket
located at the bottom of edge®.
Slide edge® into the wall cradle. Verify the
battery icon indicates charging.
Power Connection
Alternatively to using the cradle for
power, edge® can be powered by micro
USB socket at the top. Plug the 5 VDC
adapter into the power supply socket or by
connecting directly to a PC.
Note: edge® is supplied with a rechargeable
battery inside, which provides about 8
hours of continuous use. Whenever edge®
i s connected to the power adapter or to a PC,
the battery is charging.
Electrode
& Probe
Connections
Setup/Installation
10
Connect the 3 mm probe jack to the probe
input located on the bottom of edge®.
Make sure the probe is completely
connected. If the probe is recognized,
“CONNECTING” message is displayed along
with sensor model.
If the probe is not connected or not
recognized, “NO PROBE” message is
displayed.
www.GlobalTestSupply.com
The following General Setup options are displayed regardless of the sensor
being used. These settings remain when switching to another probe type or
when no probe is attached. Options are tabulated in the table below with choices
and default values. Options are accessed by pressing SETUP. Loop through the
options by using the pq arrows. To modify a setting, press MODIFY. The option
may be modied by using u, p and q keys. Press CFM to conrm the change. To
exit SETUP press SETUP.
OptionDescriptionChoicesDefault
*Only seen
when cable
connection
between micro
USB and PC is
made.
Log
Set Calibration
Expiration
Warning
Probe Specic
Set Date
Select if PC is being
used for charging
battery (and meter
will be used for
logging) or if Data
will be exported to
the PC.
Select log type to be
used from 3 types of
logging:
Manual log on
demand
Manual log on
stability (3 types
of stability criteria
available)
Timed interval lot
logging
Meter will indicate
“CAL DUE” when
set time in this
parameter has been
exceeded.
Parameters that are specic to a measurement type are inserted here in the
SETUP list.
Press MODIFY key
to Set current date,
displayed in ISO
format. Press CFM to
save changes.
Press MODIFY key
to Set current time,
displayed in ISO
format. Press CFM to
save changes.
Used to save battery
life by automatically
turning o when no
key press is detected
for time set and
meter is not in active
logging or calibration
mode.
www.GlobalTestSupply.com
24hr:MM:SS
Time
5, 10, 30, 60 Min
or O
Set timeAvailable
10 MINAvailable
Setup/Installation11
General
Setup
OptionDescriptionChoicesDefault
If enabled, a short
audible tone is
Sound
Temperature
Unit
LCD Contrast
Flash Format*
Only seen when
log errors are
present.
Message
Transition
Reset Cong
To Default
Instrument
Firmware/
Probe
Firmware
produced for key
stroke or calibration
conrmation and
a longer tone for
wrong key.
Select degree
Celsius or Fahrenheit
scale for displayed
and logged
temperatures.
Permits modication
of the display
contrast for various
lighting conditions.
Permits formatting
the ash drive.
User may choose
how messages are
displayed on third
LCD line of display.
Press the MODIFY key and CFM (when prompted) to reset
parameters.
Displays rmware
version of meter.
Using the u
key switches to
Probe rmware (if
connected) and
diagnostic made for
troubleshooting.
On or OOnAvailable
ºC or ºFºCAvailable
1 to 83Available
On or OOFFAvailable
Word scroll
messages or
letter scroll
messages
View only
Letter scroll
messages
Current
rmware
version.
Basic mode
(pH and EC)
Available
Available:
RESETS with
Basic Mode
OFF.
Available
Setup/Installation
12
User ID and Serial
Meter ID/
Meter SN/
Probe SN
CSV le
separator
Number of meter
and probe (if
connected). Use
u to change
between the three
parameters.
Used to separate
columns in the CSV
le..
Meter ID is user
selectable
Comma (,) or
Semicolon (;)
0000/ Serial
Number
CommaAvailable
*Note: Options that are seen under special conditions only.
www.GlobalTestSupply.com
Available
edge® oers a basic operation mode that streamlines measurement
conguration for pH and EC measurements and is useful for many routine
applications. Basic pH SETUP reduces parameter selection to the basic set.
The meter limits calibration to 5 standard pH buers; 6.86, 7.01, 4.01, 9.18
and 10.01 buers. All pH measurements will display, log and export with 0.01
pH resolution. Interval logging is also eliminated. Manual and Manual medium
stability log on demand are still functional. pH CONDITION, RESPONSE graphs
are not displayed, giving the display a “Basic” measurement screen with just pH
data and temperature. Basic EC reduces specic EC SETUP parameters to 3. The
meter defaults settings to common parameters and auto ranges measurements
automatically. The Basic EC meter may be used for conductivity and TDS
measurements (salinity is not available). Interval logging is also eliminated.
Manual and Manual medium stability Log on demand are still functional.
Note
:
• If powering edge® through the micro USB connector to a PC, a SETUP
option will require the choice “LOG ON EDGE” or “EXPORT TO PC“.
• 1000 log records can be stored into edge® memory. This memory is shared
between all measurement types (pH, EC, DO) and all logging types (Manual,
Manual Stability, Interval logs).
• The maximum number of records for an Interval lot is 600 records (provided
log space is available).
• A record is a stored reading and a lot is a group of records.
Each time an Interval log is initiated, a new lot is
created. The maximum number of Interval lots that
may be stored is 100. If a 101
LOTS” will be displayed. Some lots will need to be
deleted. The lot numbering is up to 999 and restarts if
all lot logs are deleted. All log on demand and stability
log on demand are stored in a single lot. The maximum
number of records that may be stored in a Manual or
Stability lot is 200 records.
st
lot is attempted, “MAX
Basic
Mode
Logging
Function
If the log memory is full during any logging session,
the “LOG FULL“ message will be displayed on the
third LCD line for a brief moment and logging will
cease. The display will return to the measurement
screen.
Logging type is congured in SETUP.
Types of Logging
Interval logging: A continuous log recorded using a user-selected timed period.
(This is not available in Basic mode).
Manual log on demand: Readings are logged each time LOG key is pressed. All
of the records are stored in a single Manual lot for the measurement type. New
records made on dierent days are stored in the same Manual lot.
Manual Stability log on demand: A log on demand that is made each time LOG
key is pressed and the stability criteria is reached. Stability criteria may be set to
Fast, Medium or Accurate settings.
Setup/Installation13
www.GlobalTestSupply.com
Logging
Function
In Setup mode, choose log parameter, press MODIFY then use the u arrow to
select between Interval, Manual, or Stability. When Interval is displayed, use the
p and q arrows to select the setting for the timed interval. When Stability is
displayed, use the p and q to select the measurement stability setting.
A complete set of GLP information including date, time, range selection,
temperature reading, calibration information and probe serial number is stored
with each log made.
Interval Logging
Select Interval and sampling period in the SETUP menu (Not available in Basic
mode). To start Interval logging, press the LOG key while the instrument is in
measurement mode.
A “PLEASE WAIT” message will be displayed followed
by the number of free spaces. During active interval
logging, lot information is displayed on the third
LCD line. The line indicates in which lot the data will
be placed and keeps track of the number of logged
records taken. The “LOG“ tag is on continuously during
active logging.
Pressing u during an interval log will display the
number of logs available.
Setup/Installation
14
Pressing the LOG key again will stop the Interval
logging session. The “LOG STOPPED“ message will be
displayed for a few seconds.
If a sensor failure occurs during interval logging, the
message “OUT OF SPEC.” will alternate with logging
information.
Manual Logging
Select Manual in the SETUP menu. To initiate a
Manual log, press the LOG key while the instrument
is in measurement. The “PLEASE WAIT” screen will
be displayed briey followed by a screen indicating
the measurement has been saved and then a screen
indicating the log record number.
The “LOG” tag will be displayed on all 3 screens.
“PLEASE WAIT”
“SAVED” with the log record number
“FREE” with the number of free spaces available
www.GlobalTestSupply.com
Stability Logging
Select Stability and choose measurement stability
criteria in the SETUP menu. Only Stability Medium
is available in Basic mode. To initiate the Stability
log, press the LOG key while the instrument is in
measurement.
The “PLEASE WAIT” screen will be displayed briey
followed by a screen showing the stability tag, “LOG”
tag and a “WAITING” message. The log can be stopped
while the “WAITING” message is displayed by pressing
LOG again.
When the stability selected criteria has been met,
a “SAVED” message will be displayed followed by a
screen indicating how much log space is available. The
“LOG” tag will be displayed on all 4 screens.
“PLEASE WAIT”
“WAITING”
“SAVED” with the log record number
“FREE” with the number of free spaces available
Logging
Function
All log records stored on edge® may be viewed on the meter by pressing the RCL
key. The log records are grouped according to measurement (pH, EC, DO).
The parameter displayed rst is based on the probe/
sensor connected. The display also indicates the
percentage of log memory used. Press CFM to display
those logs.
If no sensor or probe is connected, use the u arrow
to select the desired measurement type. Press CFM to
display those logs.
Once a parameter is selected by pressing CFM, use the
pq keys to chose the type of logging records to view.
Choices are:
• Manual log on demand lot,
• Manual log on stability lot,
• Individual Interval logging lots.
Setup/Installation15
www.GlobalTestSupply.com
Viewing
Logged
Data
Viewing
Logged
Data
If no data was logged for the selected measurement
range, the instrument displays the following
messages (ex. for pH range):
“NO MANUAL LOGS”
“NO STABILITY LOGS”
Press CFM to enter inside lot information to view recorded data.
Use the ARROW keys to toggle between dierent records.
Use u to display GLP data including calibration information, date, time, etc.
Press CLR then CFM when deleting records or lots.
Press RCL to exit the logging type.
Press RCL to exit the parameter selection screen.
Press RCL to return to the measurement screen.
Delete Logging Type/Lot
Press RCL and select the parameter log.
Use the pq keys to select the Manual/Stability
records or Interval lots to delete. Press CLR. The
instrument will display “CLEAR MANUAL” for Manual
Records, “CLEAR STAB” for Stability Records.
For Interval lots, the message “CLEAR”, followed by the
selected lot will be displayed with “CFM” tag blinking.
Press the pq keys to select a dierent lot. Press CFM.
The instrument will display “PLEASE WAIT”.
Setup/Installation
16
“CLEAR DONE” is displayed for a few seconds after the
selected Interval lot is deleted.
Delete Records (Manual and Stability log on demand)
To delete individual records (Manual and Stability logs only), enter Manual
(Stability) log by pressing CFM when Manual (Stability) is displayed. Use the
ARROW keys to select the record to be deleted and then press CLR.
The instrument will display “CLEAR REC.“ and record
number along with “CFM” tag blinking. Use the ARROW
keys to select another record if necessary.
Press CFM. The instrument will display “PLEASE WAIT” and then “CLEAR
DONE” message. When individual logs are deleted within saved MANUAL or
STABILITY logs, the logs will renumber, lling in the deleted data but staying in
chronological order.
www.GlobalTestSupply.com
To delete all records of the MANUAL (STABILITY) log, proceed as described on
page 16 for LOTS.
Select the Manual (Stability) lot and press CLR.
The ”CLEAR” message will be displayed along with
”MANUAL” or “STABILITY” and CFM tag blinking on
the LCD. Press the CFM key to conrm the deleting of
the selected lot (MANUAL or STABILITY) or all records.
Press CLR to exit without deleting.
The lot number is used to identify particular sets of data. The lot numbers
are allocated successively until 100, even if some lots were deleted. The total
number of lots that can be saved is 100. If some are deleted (for example 1-50),
fty additional logs may be stored. These will be numbered 101-150. The lots are
allocated successively (provided available memory space) until 999 is reached.
After this, it is necessary to delete all the LOT logs to start over the numbering.
Delete All
All pH logs, (or all EC, all DO logs) may be deleted in a single clear. This function
will delete all MANUAL, STABILITY and INTERVAL logs for the measurement type
selected.
Press the RCL key. The pH, EC, or DO type will be
blinking. Use u to select desired measurement
parameter log data to delete.
While the measurement type is blinking and message
states “LOG RECALL”, press CLR.
“CLEAR ALL” and measurement type will be displayed
with “CFM” tag blinking. Press CFM.
Viewing
Logged
Data
“PLEASE WAIT” and the percent cleared will be
displayed until completed. The procedure can be
repeated for the other measurement modes.
Note: If CLR is pressed in error, press CLR again to exit
without deleting.
www.GlobalTestSupply.com
Setup/Installation17
PC & Storage
Interface
Logged data on edge® can be transferred from the meter to a USB ash drive
by using the log recall function. The minimum requirement for the drive is USB
2.0. Select the pH, EC or DO record you wish to export and follow the simple
steps below.
Connect USB ash drive to the USB port, located on
the top of the meter. Press the RCL key and select
the parameter to view (pH, EC, or DO). Press the CFM
key. Select Manual, Stability, or interval lots by using
the pq keys. Press the LOG key (not CFM). The “USB
HOST” tag should come on.
“PLEASE WAIT” message appears followed by
“EXPORT”. Press CFM to export the selected record or
lot. If CFM is not pressed in 10 seconds, the USB host
will become inactive.
The meter will display the percentage of export.
The export percentage should go to 100%. Remove
USB ash drive.
Setup/Installation
18
If the selected le is already saved on the ash drive,
edge® will ask for conrmation of overwriting the
existing le. The message “OVERWRITE” and “CFM”
tag will blink. Press CFM for overwriting the existing
le or CAL to exit without exporting.
After exporting the display will return to the
selected le. Press the RCL key twice to return to
measurements.
Note: Do not remove USB ash drive during an active export transfer.
www.GlobalTestSupply.com
Logged data on the edge® can be transferred from the meter to a PC by
following these simple directions. Suitable operating systems include Windows
(Xp minimum), OS X or Linux.
1. Connect edge® to the PC using the supplied micro USB cable.
2. Turn on edge®.
3. Press SETUP and select “LOG ON EDGE”
4. Press MODIFY then use pq keys to change to “EXPORT TO USB”
5. Press CFM and the USB/PC Tag is displayed.
6. Press SETUP to exit
The PC should detect the USB as a removable drive. Open the drive to view the
stored les. Log les are formatted as Comma separated values (*.CSV) and can
be opened with any text editor or spreadsheet program. (Note: Field separator
may be set as comma or semicolon depending upon region preferences, see
Setup.)
Note:
• Western Europe (ISO-88859-1) character set and English language are
suggested settings.
• Other les may be visible depending upon computer settings. All les
stored will appear in this folder.
• Adjust Font or column width appropriately. Adjust the decimal places if the
pH was logged with 0.001 resolution.
• All conductivity measurements will be displayed as μS/cm.
Interval logs are designated as pH, EC or DO Lots. ie. PHLOT001, ECLOT002,
DOLOT003.
PC & Storage
Interface
The Manual Lots are PHLOTMAN, ECLOTMAN, and DOLOTMAN.
The Stability Lots are PHLOTSTA, ECLOTSTA, and DOLOTSTA. All stability logs,
regardless of stability setting, are located in the same stability le for that
measurement.
Click on the desired log to view data.
Note:
• If “ºC !” appears in log data, the electrode/probe was used beyond it’s
operation specications and the data is not considered reliable.
• If “ºC !!” appears in log data, the temperature sensor within the probe or
electrode is broken and the device should be replaced. Logged data should
not be considered reliable.
Setup/Installation19
www.GlobalTestSupply.com
Basic vs
Standard
pH Mode
For optimum pH measurements, follow these steps:
1. Understand the benets and features of Standard and Basic Operation.
2. Set up edge® meter by conguring preferences.
3. Calibration
4. Measurement
The “Standard” pH operation includes up to a 5-point buer calibration, use
of custom buers, choice of displaying 0.001 pH resolution, use of the full
diagnostic features of Calibration Check™ (that includes buers used, probe
condition, response time and messages indicating a contaminated buer or pH
sensor requiring maintenance during calibration). Additional Sensor Check™
indicators are available if using HI 11311 or HI 12301 pH electrodes. These
include a continuous diagnostic to detect if the electrode is broken and at the
time of calibration, a diagnostic to indicate the reference junction has been
compromised or fouled by sample contamination. The Standard pH operation
also includes full logging capability including Interval, Manual log on demand
and Manual log on stability.
“Basic” pH operation provides a simplied SETUP menu; there are no decisions
to make regarding the pH measurement itself. The meter will display 0.01 pH
resolution and permit a 3-point buer calibration from the following pH buers;
4.01, 6.86, 7.01, 9.18 or 10.01. Calibration Check™ and Sensor Check™ features
are limited to messages during calibration. Calibration reminders are also not
available. The GLP will still provide oset, slope, buers used and a calibration
date. The Basic pH operation includes Manual log on demand and Manual log on
stability (medium setting).
: When changing from Standard to Basic operation in SETUP, previous
Note
calibration data will be cleared. A prompt will force the user to facilitate this.
Operational Guide
20
Major dierences between Standard and Basic modes are shown below.
StandardBasic*
Calibration
Diagnostics
Log types
Recommended pH electrodes:
5 points including 2 custom
buers
Cal Check™ Feature
Sensor Check™ Feature
Error messages
GLP
Manual Log on demand
Manual Log on stability
(Fast, Medium, Accurate)
Interval Logging
HI 11310, HI 12300
HI 11311, HI 12301
HI 10530, HI 10430
3 points
Basic error messages
GLP basic
Manual Log on demand
Manual Log on stability
(Medium)
HI 11310*
HI 12300
* All Sensors work in this mode, but diagnostic and all buers will not be
available.
www.GlobalTestSupply.com
edge’s pH meter operation is congured by using the SETUP key with a pH probe
connected to the meter. The parameter-specic options will be seen inserted
into the menu. If Basic mode is “On”, the pH parameter list will not be displayed.
See Basic mode for a description of this operation before choosing how to
SETUP the meter.
OptionDescriptionChoices DefaultBasic mode
When “On”, a limited set of
Basic Mode
Information
parameters and calibration
buers are available for
use.
Visual indication of buers
used, Probe Condition
and Response times are
determined and displayed
when calibration uses 7 and
4 and/or 10 pH buers.
O or
On
O or
On
OAvailable
Buers used, probe
On
conditions, and
response times are
not displayed.
pH Meter
Congurations
First Custom
Buer
Second
Custom Buer
First
Calibration
Point
Resolution
Set Out Of
Calibration
Range
When “On”, it permits the
user to enter a custom pH
buer value to use during
electrode calibration.
When “On”, it permits the
user to enter a custom pH
buer value to use during
electrode calibration.
Allows the user to choose
how the rst point in
calibration will be made.
Allows the user to select
between 0.01 and 0.001 pH
resolution.
When set to “On”, a
measurement that is
outside the calibrated
range (buers used) will
trigger a warning message.
O or
value
O or
value
Oset
or point
0.01 or
0.001 pH0.01 pH
O or
On
ONot Available
ONot Available
Oset
On
Not Available
(automatically uses
Oset).
Not Available
(automatically uses
0.01 pH resolution).
Not Available, No
Error messages
displayed.
Calibration in Standard Mode
pH operation in standard mode oers full function of edge®. This includes
seven standard buers and two custom ones. Five pH buers may be used for
calibration.
The instrument should be recalibrated whenever:
• High accuracy and sensor verication are required.
• The pH electrode is replaced.
• At least once a week.
• After testing aggressive chemicals.
• If ”CAL DUE” is displayed on the third LCD line.
pH
Calibration
Every time you calibrate the instrument use fresh buers and perform electrode
maintenance as required. It is advised to choose calibration buers that bracket
the sample pH.
Operational Guide21
www.GlobalTestSupply.com
pH
Calibration
Preparation
Pour small quantities of the buer solutions into clean beakers. If possible,
use plastic to minimize any EMC interferences. For accurate calibration and to
minimize cross-contamination, use two beakers for each buer solution; one
for rinsing the electrode and one for calibration. If you are measuring in the
acidic range, use pH 7.01 or 6.86 as the rst buer and pH 4.01 (or 3.00*) as the
second buer. If you are measuring in the alkaline range, use pH 7.01 or 6.86 as
rst buer and pH 10.01 or 9.18 as the second buer.
* pH 3.00 is only visible when using specic pH electrodes and will replace 4.01
pH buer.
Procedure
Calibration can be performed using up to ve calibration buers. For accurate
measurements, a minimum of a three-point calibration is recommended. The
calibration buer can be selected from the calibration buer list that includes
the custom buers and the standard buers:
The custom buers allow the user to calibrate in a buer solution dierent from
a standard one. Two custom buers can be set in SETUP menu. See page 23 for
more information about using custom buers.
The instrument will automatically skip custom buers which are in a ±0.2 pH
window of an already calibrated buer.
Submerse the pH electrode approximately 3 cm (1¼”) into a buer solution and
stir gently. Press CAL to enter calibration.
Operational Guide
22
The “CAL” tag will appear and the “7.01” buer will be
displayed on the third LCD line. If necessary, press the
ARROW keys to select a dierent buer value. The “ ”
along with “STIR” tag will be displayed and “WAIT” will
blink on the LCD until the reading is stable.
When the reading is stable and close to the selected
buer, “CFM” tag will blink. Press CFM to conrm
calibration.
After conrming the rst point, the calibrated value
will be displayed on the rst LCD line and the second
expected buer value on the third LCD line (i.e. pH
4.01).
www.GlobalTestSupply.com
Rinse and submerse the pH electrode approximately 3 cm (1¼”) into the second
buer solution and stir gently. If necessary, press the ARROW keys to select a
dierent buer value.
The “
” along with “STIR” tag will be displayed and “WAIT” will blink on the LCD
until the reading is stable.
When the reading is stable and close to the selected buer, “CFM” tag will blink.
Press CFM to conrm calibration.
Repeat procedure with additional pH buers. A total of ve pH buers can be
utilized.
After conrming the last desired buer calibration
points, press CAL (or if all ve buer values were
calibrated) the instrument will automatically display
“SAVING” as it stores information. It will then return to
normal measurement mode.
Each time a buer is conrmed, the new calibration
data replaces the old data for the corresponding
buer or for any buer in the proximity of ±0.2 pH.
If current buer has no previous data stored and the
calibration is not full (ve buers), the current buer
is added to the existing calibration. If the existing
calibration is full, the instrument asks which buer to
replace.
Press the ARROW keys to select another buer to be replaced.
Press CFM to conrm the buer that will be replaced.
Press CAL to leave calibration without replacing.
pH
Calibration
Note: If the replaced buer is outside the ±0.2 pH window of the calibrated
buers, it is possible to select this buer during the next calibration.
Working With Custom Buers
If a custom buer was set in SETUP menu, it can be selected during calibration
by pressing the ARROW keys. The “C1” or “C2” tag will be displayed once selected.
Press u if you want to modify the custom buer
value. The buer value will start blinking.
Use the ARROW keys to change the buer value.
After 5 seconds, the buer value is updated. Press u
if you want to change it again.
Operational Guide23
www.GlobalTestSupply.com
pH
Calibration
Note: Custom buer values can be adjusted ±1.00 pH around the set
value during calibration. When a custom buer is displayed, the “C1” or “C2”
tag is displayed.
First Calibration Point
When performing a new calibration, or adding to an existing one, the user has
a choice of how the rst new calibration point will be treated in reference to
the existing calibration point. This is selected in SETUP by the option FIRST
CALIBRATION POINT. The two SETUP selectable options are “POINT” or “OFFSET”.
Point: A buer value can be recalibrated and added to the previous calibration
set. The electrode slope of the other calibration points will be reevaluated with
the recalibrated buer value.
Oset: The new buer calibration point can create a constant oset to all
existing pH calibration data (existing calibration must have a minimum of two
pH buers)
.
Recalibrating a pH sensor or adding to an existing calibration is simple and
follows the PROCEDURE outlined on page 22.
Press CAL. Place sensor in desired buer and select buer from choices. When
sensor has equilibrated, the CFM tag will turn on and blink. Press the CFM key.
Press CAL to escape the calibration. Alternately continue calibrating in
additional buers. The latest calibration point will be added to the existing data.
GLP will reect the latest calibration data. Older calibration buers will be seen
as blinking buers.
Note: Each time a buer is conrmed, the new
calibration data replaces the old data for the
corresponding buer or for any buer in the proximity
of ±0.2 pH. If the current buer has no previous data
stored and the calibration has not used ve buers,
the current buer is added to the existing calibration.
If the existing calibration is full, the instrument asks
which buer to replace.
Operational Guide
24
Note: When using Standard mode, the user can choose if they want the display
to show the CONDITION and RESPONSE gauges on the display. These are part of
the Cal Check™ system and are selected in SETUP by the option INFORMATION.
The choice is ON or OFF.
Electrode Condition and Electrode Response Time
edge® pH Calibration Check™ feature will assess electrode condition and
response time during each calibration and display it for the rest of the day.
The condition gauge shows the electrode’s condition that is based on the oset
and slope characteristics of the pH electrode at the time of calibration. The
response gauge is a function of the stabilization time between the rst and
second calibration buers, when calibration is performed between a pair of
4.01, 7.01, or 10.01 buers. These gauges reect the electrode’s performance
and should be expected to slowly decrease over the life of the electrode .
www.GlobalTestSupply.com
If the instrument is not calibrated the calibration history has been deleted, or it
has been calibrated only at one point, the electrode condition and the electrode
response gauges will be empty.
For a continuous display of the electrode’s condition and response, daily
calibration is necessary. This information can also be viewed in the GLP data.
Junction Condition (HI 11311 and HI 12301 Only)
edge’s pH Sensor Check™ feature assess the health of the pH electrode’s
reference junction during each calibration. The junction gauge may be viewed
directly in GLP but will also blink a warning on the display if the junction is
compromised (not 100%). Should this happen, the junction condition will
appear on the display with the junction blinking. The Junction Condition is a
function of the electrode’s reference impedance which should be kept low. If the
reference junction becomes fouled from a precipitate or coating, the impedance
will rise and cause the pH measurement to drift. This diagnostic feature serves
as a warning to clean the sensor.
pH
Calibration
Calibration in Basic Mode
Procedure
Basic mode operation permits up to three-point buer calibration.
For accurate measurements, at least a two-point calibration is recommended.
However, a single point calibration can also be used.
The calibration buers can be selected from the calibration buer list that
includes the standard buers, pH 4.01, 6.86, 7.01, 9.18 and 10.01.
Three-Point Calibration
Submerse the pH electrode approximately 3 cm (1¼”) into a buer solution and
stir gently. Press CAL. The “CAL” tag will appear and the “7.01” buer will be
displayed on the third LCD line. If necessary, press the
ARROW keys to select a dierent buer value.
The “ ” along with “STIR” tag will be displayed and
“WAIT” will blink on the LCD until the reading is stable.
Operational Guide25
www.GlobalTestSupply.com
pH
Calibration
When the reading is stable and close to the selected
buer, the “CFM” tag will blink. Press CFM to conrm
calibration.
After conrming the rst calibration point, the
calibrated value will be displayed on the rst LCD line
and the second expected buer value on the third LCD
line. (i.e. pH 4.01)
Rinse and submerse the pH electrode approximately
3 cm (1¼”) into the second buer solution and stir
gently.
If necessary, press the ARROW keys to select a dierent buer value.
The “
” along with “STIR” tag will be displayed and “WAIT” will blink on the LCD
until the reading is stable. When the reading is stable and close to the selected
buer, the “CFM” tag will blink. Press CFM to conrm calibration.
The calibrated value is then displayed on the rst LCD line and the third expected
buer value on the third LCD line.
After the second calibration point is conrmed, rinse and submerse the pH
electrode approximately 3 cm (1¼”) into the last buer solution and stir gently.
If necessary, press the ARROW keys to select a dierent buer value.
The “ ” along with “STIR” tag will be displayed and “WAIT” will blink on the LCD
until the reading is stable.
Operational Guide
26
When the reading is stable and close to the selected buer, the “CFM” tag will
blink. Press CFM to conrm calibration.
At the end of calibration the instrument displays “SAVING”, stores the calibration
value and returns to normal measurement mode.
The calibration sequence may be reduced to two buer values or a single one.
Press CAL to return to measurement mode after the desired number of buers
have been calibrated.
Note:
• When performing a new calibration or adding to an existing calibration the
rst calibration point will be treated as an oset. See page 24 for details.
• Press CAL after the rst or second calibration point are conrmed and
the instrument will store the calibration data. Then it will return to
measurement mode.
• If the value measured by the instrument is not close to the selected buer,
“WRONG BUFFER” will blink. Check if the correct buer has been used or
selected.
www.GlobalTestSupply.com
• If the buer temperature exceeds the temperature limits of the buer,
“WRONG BUFFER TEMPERATURE” will be displayed.
• Press CLR after entering calibration to clear previous calibrations stored on
probe. “CLEAR CALIBRATION” message will be displayed. Press CFM. The
instrument will return to measurement mode displaying the “CAL DUE”
message.
• If measured value diers more than expected “CHECK ELECTRODE CHECK
BUFFER“ will be displayed. Take appropriate action, cleaning the probe if
necessary or refreshing the pH buer.
pH Buer Temperature Dependence
Calibration buers are aected by temperature. During calibration with standard
pH buers (not Custom), the instrument will display the pH buer value at 25 ºC,
however, it will use the appropriate value for that buer at the temperature of
the buer for the calibration. Immediately after exiting calibration, the buer
will read its value at the temperature of measurement.
The Calibration Check™ feature may ag diagnostic messages during a
calibration. As electrode aging is normally a slow process, substantial changes
from previous calibrations are likely due to a temporary problem with the probe
or buers that can be addressed easily. These messages are seen in Standard
and Basic modes.
Wrong Buer
This message appears when the dierence between
the pH reading and the value of the selected buer
is too great. If this error message is displayed, check
if you have selected the proper calibration buer and
have poured the desired buer.
Operational Guide27
www.GlobalTestSupply.com
Calibration
Messages
Wrong Old Points Inconsistency
“WRONG OLD POINTS INCONSISTENCY” is displayed if
the new calibration diers signicantly from the last
value of that probe in that buer. In this case it may
be best to clear the previous calibration and attempt a
new calibration with fresh buers.
To clear calibration information, press CAL then
press CLR. The “CLEAR CALIBRATION” message will
be displayed. Either press the CFM key and clear all
calibration information or press CAL to exit. The probe
may retain a single point calibration if rst point was
accepted.
Once calibration information is cleared, the message
“CAL DUE” will be displayed
Clean Electrode
“CLEAN ELECTRODE” indicates poor electrode
performance (oset out of accepted window, or
slope under the accepted lower limit). Often, cleaning
the probe will improve the pH electrodes response.
See pH Electrode Conditioning and Maintenance for
details. Repeat calibration after cleaning.
Check Electrode Check Buer
“CHECK ELECTRODE CHECK BUFFER” appears when
electrode slope exceeds the highest accepted slope
limit. You should check your probe and use fresh
buer. Cleaning may also improve this response.
Operational Guide
28
Bad Electrode
“BAD ELECTRODE” appears if the cleaning procedure
performed as a result of the above two messages is
found to be unsuccessful. In this case it is advised to
replace the probe.
Wrong Buer Temperature
“WRONG BUFFER TEMPERATURE” appears if the
temperature of the buer is outside the dened
buer temperature range. The calibration buers are
aected by temperature changes in a dened manner.
During calibration, the instrument will automatically
calibrate to the pH value corresponding to the measured temperature but
display it to the value at 25 ºC. Immediately after calibration, the buer should
read the value of the buer at the temperature of measurement.
Note: Temperature limits will be reduced to actual sensor specications.
www.GlobalTestSupply.com
Contaminated Buer
“CONTAMINATED BUFFER” appears in order to alert
that the buer could be contaminated. Refresh your
buer and continue the calibration procedure.
Broken Temperature Sensor
If the temperature sensor should malfunction or break at any time, a temperature
of “25.0 ºC” will blink on the second LCD line and the
message “BROKEN TEMPERATURE SENSOR” will
appear on the third LCD line after leaving calibration.
The calibration will have the compensation at 25 ºC.
Replace probe if this occurs.
Note: If this occurs during logging “25 ºC !” will appear in the CSV le.
Calibration
Messages
Good Laboratory Practice (GLP) refers to a quality control function used to
ensure uniformity of sensor calibrations and measurements. The dedicated
GLP key opens a le of the latest calibration information. Use the qp keys to
scroll the stored information. This includes the buers used, temperature of the
buer, time and date of the last calibration, the sensor serial number and the
calculated oset and percent slope. This information is available in Basic and
Standard Modes. This information is also included with every data log. Newest
calibration points report as a solid number, older calibration data (that is still
used) will be displayed blinking.
If calibration has not been performed, the instrument
displays a blinking “NO CAL” message.
The pH calibration oset and slope (the GLP slope is
the average of the calibration slopes; the percentage
is referenced to the ideal slope value at 25 ºC). The
condition and response indicators displayed are from
the last calibration.
Pressing the qp keys, the last calibration date (yyyy.
mm.dd) together with the current reading is displayed.
pH GLP
Information
www.GlobalTestSupply.com
Operational Guide29
pH GLP
Information
Note: If a custom buer was used in calibration, the “C1” and “C2” tags will be
displayed. If only the one custom buer is used in calibration, the tag will be “C1”
and the value will be displayed.
If disabled, “EXPIRATION WARNING DISABLED” is
displayed.
Or if enabled the number of days until the calibration
alarm “CAL DUE” will be displayed. (i.e. “CAL EXPIRES
IN 2 DAYS”)
The number of days since the calibration expired. (i.e.
“CAL EXPIRED 2 DAYS AGO”).
The probe serial number together with the current
reading.
If a buer is not from the last calibration, the buer
tag will be displayed blinking.
Operational Guide
30
In Standard Mode, Condition and Response gauges are visible on the day of
calibration (See Electrode Condition And Electrode Response Time page 24).
If congured in SETUP, a countdown message displays the number of days
remaining until a new Calibration is due.
If using a HI 11311 or HI 12301 sensor, the glass
impedance is continuously monitored, updated and
reported here and the Junction Condition Gauge is
fully visible.
www.GlobalTestSupply.com
When an edge® pH probe is connected, the instrument
will recognize it and a the cap removal message will be
displayed along with “PH”. Press any key to skip the
message. The instrument will enter measurement
mode. Make sure the electrode has been calibrated
before taking measurements.
Rinse the pH sensor with water and a sample if
possible. Submerse the electrode tip approximately 3 cm
(1¼”) into the sample to be tested and stir sample
gently. Allow time for the electrode to stabilize.
The pH is displayed on the rst LCD line and the temperature on the second LCD
line. Using the ARROWS, date, time, battery status and oset with slope can be
displayed on the third LCD line. If the reading is out of measurement range, the
closest full scale value will be displayed blinking on the rst LCD line.
If measurements are taken successively in dierent samples, it is recommended
to rinse the electrode thoroughly with deionized water or tap water and then
with some of the next sample to prevent cross-contamination. The pH reading is
aected by temperature. The temperature eect is compensated for using the
temperature sensor inside the probe. The resulting measurement is the actual
pH at the temperature of measurement.
Error Messages During Measurement
If the pH or temperature exceeds the limits of the
sensor, the message “ELECTRODE OUT OF SPEC”
will scroll on the third LCD line. The temperature will
continue to be displayed. If temperature exceeds
the meter specication of 120 ºC, then “120 ºC” will
blink on the display. If interval logging, the message
“OUT OF SPEC.” will alternate with the LOG specic
message. In both these cases the Log le will indicate
a “ºC!” next to the data.
pH
Measurement
In case the temperature sensor is damaged, “BROKEN
TEMPERATURE SENSOR” message will be displayed
with “25.0 ºC” blinking on the second LCD line. The Log
le will indicate “ºC!!” next to the data.
mV Reading Of The pH
The mV reading of the measured pH can be displayed
on the LCD by pressing the RANGE key.
www.GlobalTestSupply.com
Operational Guide31
pH
Measurement
Third Line Messages Displayed During Measurement
All the messages described in General Setup (page 11)
are displayed in pH range.
Temperature Sensor Problem (if there is one)
Cal Due or Oset and Slope Value
Time
Date
Battery or Charge Status
Logging Messages
Out of Calibration Range
Operational Guide
32
www.GlobalTestSupply.com
Steps To Optimize EC Measurement
Follow these steps to optimize measurement using an EC probe:
1. Determine what measurement you wish to make with the EC probe. (See
below)
2. Determine if Standard or Basic mode is best for your measurement.
3. Connect the Probe and congure your measurement using the SETUP
menu.
4. Calibrate the EC probe.
5. Take measurements using an EC probe.
Measurements Available With The EC Probe
The four-ring EC probe may be used for 3 dierent measurement applications
with edge®.
• It may be used for temperature compensated or absolute conductivity
measurements (with units of μS/cm or mS/cm).
• TDS (Total Dissolved Solids) measurements (a calculated measurement
derived from the ionized particles in a sample and the conductivity
measurement). This has measurement units of mg/L, ppm or g/L.
• Salinity (not available in basic mode): 3 dierent seawater salinity scales
are supported; Practical Salinity (PSU), the Natural Seawater Scales (g/L),
and the NaCl %.
The RANGE key is used to change from conductivity to TDS (and to Salinity
available in standard mode only).
Basic Versus Standard EC Mode
Standard operation mode allows complete conguration of all options for the
measurement of Conductivity, TDS or Salinity measurements. Measurement data
can be logged using Manual log on demand, Manual Log on Stability or Interval
logs. All logged data can be exported to a thumb drive or PC. Basic operation
mode is used for conductivity and TDS measurements. Common default values
are used for Cell Factor (C.F.), temperature compensation coecient and TDS
conversion factor. Measurement units are user-selectable for TDS. The meter
uses continuous auto-ranging to simplify the measurement experience.
Set up ParametersFully selectableDefault values used
Log types
Manual Log on demand
Manual Log on stability
(Fast, Medium, Accurate)
Interval Logging
Manual Log on demand
Manual Log on stability
(Medium)
EC (Electrolytic Conductivity) meter operation is congured using the SETUP
key with an EC probe connected to the meter. The EC-specic parameters will
be seen inserted into the menu. If Basic mode is “On”, the EC parameter list will
be simplied, limiting the options the user can change. See Basic mode for a
description of this operation before choosing how to SETUP the meter.
Operational Guide33
www.GlobalTestSupply.com
EC Meter
Conguration
EC Meter
Conguration
OptionDescriptionChoicesDefaultBasic mode
Basic ModeO or OnOAvailable
Temperature
Compensation
-1
)User may
C.F (cm
T.Coef. (%/ºC)This parameter
T.Ref. (ºC)The user may
TDS Conversion
Factor.
View T.Ref. or
T.Coef.
The user
may select
Automatic
Temperature
Compensation
or No TC to
congure
absolute
conductivity.
enter actual
Cell factor
value Manual
calibration. (see
page 35)
is related to the
solution being
measured at
temperatures
other than 20
or 25 ºC. It is
used to correct
measured
conductivity
to a reference
temperature by
applying a xed
factor for linear
compensation.
select either
20 ºC or 25 ºC
reference
temperature for
temperature
corrected
conductivity.
This factor
is used to
mathematically
convert
conductivity to a
TDS value.
The user may
choose whether
to display the
reference
temperature
(T.Ref) or the
Temperature
coecient on
the display
with the
measurement.
No TC or ATCATCNot available.
0.01 to 9.999
-1
cm
0.00 to 6.00
(%/ºC) Note:
Setting to 0.00
is the same as
using No TC.
20 ºC or 25 ºC 25 ºCAvailable
0.40 to 0.80 0.50Not Available.
T.Ref.(ºC) or
T.Coef.(%/ºC)
1.000 cm
1.90 (%/ºC)
Close for natural
waters or salt
solutions.
T.Ref (ºC)T.Ref (ºC)
-1
ATC is
automatically
used.
Not available.
Automatically
determined
during
calibration.
Not available.
Automatically
set to 1.90%/ºC.
Automatically
set to 0.50.
automatically
displayed during
measurement.
Operational Guide
34
www.GlobalTestSupply.com
OptionDescriptionChoicesDefaultBasic mode
EC RangeIf AUTO is used,
TDS RangeIf AUTO is used,
TDS UnitSelect units of
EC Salinity Scale Three
edge® will
automatically
nd the correct
conductivity
range and
unit. If a xed
value is used,
measurements
made outside
the ranges are
considered
Out of Range,
but units will
remain xed
throughout the
measurement.
edge® will
automatically
nd the
correct TDS
measurement
range and
units. If a xed
value is used,
measurements
made outside
the ranges are
considered
Out of Range,
but units will
remain xed
throughout the
measurement.
measurement
for TDS
measurement.
measurement
scales are
available
for salinity
measurement
in Seawater.
(Practical
Salinity Scale,
Percent Scale
and Natural
Seawater Scale).
AUTO,
29.99 μS/cm,
299.9 μS/cm,
2999 μS/cm,
29.99 mS/cm,
200.0 mS/cm,
500.0 mS/cm
AUTO,
14.99 mg/L,
149.9 mg/L,
1499 mg/L,
14.99 g/L,
100.0 g/L,
400 g/L
mg/L or ppm ppmAvailable to
PSU, NaCl%, g/L NaCl%Not available.
AUTONot available but
AUTONot available but
measurement
autoranges as
needed.
measurement
autoranges as
needed.
select.
EC Meter
Conguration
Cell Factor (cm-1) Manual Calibration
This option may be used to perform a manual calibration in a custom standard.
1. Rinse the pro be with a aliquot of standard and shake o excess solution.
2. Place the probe in the standard. The sleeve holes must be covered with
solution.
3. Press SETUP and use the ARROW keys to get to C.F. (cm-1)
4. Press MODIFY.
5. Use the pq keys to change C.F. (cm-1) until the display reads the Custom
Standard value.
Operational Guide35
www.GlobalTestSupply.com
EC Meter
Conguration
6. Press CFM. The message “MANUAL CALIBRATION CLEARS PREVIOUS
CALIBRATIONS” will be displayed on the third line LCD. “CAL” and “CFM” tags
will blink. Press CFM to conrm the manual calibration.
Note: GLP will display “Manual” for a standard. Using this calibration technique
will erase any previous calibrations done in CAL. Log les will indicate “MANUAL”
as standard.
User Selectable EC/TDS Range (Not Available In Basic Mode)
The EC and TDS measurements may be congured in SETUP as AUTO (meaning
auto ranging, the measurement automatically nds the correct conductivity
or TDS unit and resolution), or it may be congured with a user-selected
measurement range and resolution (if you know ahead of time where your
samples will fall). If AUTO is selected, the meter will select the scale with
the highest possible resolution, but may change in the middle of a series of
measurements changing units and displayed resolution.
Note: The selected range is only active during measurements. Auto ranging is
used during calibration. If a xed range is selected and during measurement
goes beyond the range limits, the full scale value of that range will ash on the
display. All log data in the CSV les will be displayed in uS/cm.
EC/TDS
Calibration
General Guidelines
Calibrate the instrument frequently, especially if high accuracy is required.
The instrument should be recalibrated:
• Whenever the EC probe is replaced.
• At least once a week.
• After testing aggressive chemicals.
• If “CAL DUE” is displayed on the third LCD line.
Every time you calibrate the instrument use fresh standard and perform
electrode maintenance as required. It is recommended to choose a calibration
standard that is close to the sample.
Preparation
Pour small quantities of the standard solutions into a beaker. If possible, use a
plastic beaker to minimize any EMC interferences. For accurate calibration and to
minimize cross-contamination, use two beakers for the standard solution; one
for rinsing the probe and one for calibration.
Procedure
Select standard to be used for calibration. 0.00 μS (probe in air) may be used for
an oset calibration. This should be done rst. Hanna Conductivity standards
are available at 84 μS/cm, 1413 μS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm,
111.8 mS/cm and are used for the cell factor calibration.
Rinse the probe with calibration solution or deionized water. Shake o excess
solution. Submerse the probe in the calibration solution. The sleeve holes must
be completely submersed. If possible, center the probe in the beaker away from
the bottom or beaker walls.
Raise and lower the probe to rell the center cavity and tap the probe repeatedly
to remove any air bubbles that may be trapped inside the sleeve.
Operational Guide
36
www.GlobalTestSupply.com
Press CAL to enter calibration. The “CAL” tag and the
recognized standard value will appear on the third LCD
line. If necessary, press the ARROW keys to select a
dierent standard value. The “
“ along with “STIR” tag
will be displayed and “WAIT” will blink on the LCD until
the reading is stable.
When the reading is stable and close to the selected
standard, “CFM” tag will blink, the message “ SOL STD”
and the value will be displayed.
Press CFM to conrm calibration. The instrument
displays “SAVING”, stores the calibration values and
returns to measurement mode.
Conductivity vs Temperature Chart
The following table lists the temperature dependence of Hanna EC calibration
standards. edge® uses these values and their temperature coecients during
calibration.
EC/TDS
Calibration
ºCºF
03271507766448300654002760
54182208966553500741003180
1050933010206759600832003615
15591048011476865400925004063
1660.81072011737067200944004155
1762.61095011997168500963004245
1864.41119012257369800982004337
1966.211430125174713001002004429
206811670127876724001021004523
2169.811910130578740001040004617
2271.612150133279752001059004711
2373.412390135981765001079004805
2475.212640138682783001098004902
257712880141384800001118005000
2678.813130144086813001138005096
2780.613370146787830001157005190
2882.413620149489849001177005286
2984.213870152190863001197005383
308614120154892882001218005479
3187.814370157594900001239005575
HI7030
HI8030
(μS/cm)
HI7031
HI8031
(μS/cm)
HI7033
HI8033
(μS/cm)
HI7034
HI8034
(μS/cm)
HI7035
HI8035
(μS/cm)
HI7039
HI8039
(μS/cm)
Operational Guide37
www.GlobalTestSupply.com
NaCl %
Calibration
Preparation
Pour a small quantity of the calibration solution into a beaker. If possible, use a
plastic beaker to minimize any EMC interferences.
Before pressing CAL verify in SETUP:
• Basic mode is o.
• Salinity Scale is set to NaCl%.
In measurement mode use the RANGE key to select the Salinity measurement.
The NaCl% tag will be on. NaCl calibration is a single point calibration at 100.0%
NaCl. Use the HI 7037L calibration solution (sea water solution) as a 100% NaCl
calibration solution.
Procedure
Rinse the probe with some of the HI 7037L calibration solution or deionized
water. Shake o excess solution. Submerse the probe in the calibration solution.
The sleeve holes must be completely submersed. If possible, center the probe in
the beaker away from the bottom or beaker walls. Raise and lower the probe to
rell the center cavity and tap the probe repeatedly to remove any air bubbles
that may be trapped inside the sleeve.
Press CAL to enter calibration. The ” “, ”STIR” and “CAL”
tags will turn on and the rst LCD line will display the
percent NaCl reading, the second LCD line will display
the temperature and the third LCD line will display
”WAIT” message blinking until the reading is stable.
Calibration
Messages
Operational Guide
38
When the reading is stable and close to the standard,
“CFM” tag will blink and “SOLUTION STANDARD”
message will be displayed.
Press CFM to conrm calibration.
The Instrument displays “SAVING”, stores the calibration values and returns to
measurement mode.
Note: If a new EC calibration is performed, the NaCl calibration is automatically
cleared. A new NaCl calibration is required.
Wrong Standard
If the reading is too far from the expected value, the
message “WRONG STANDARD” will scroll. Calibration
cannot be conrmed. In this case check if the
calibration solution has been used correctly or clean
the probe by following the instructions for EC PROBE
MAINTENANCE (see page 56).
www.GlobalTestSupply.com
Wrong Standard Temperature
If the temperature is out of the 0.0 to 60.0 ºC range,
“WRONG STANDARD TEMPERATURE” message will be
displayed and the temperature value will blink.
Calibration
Messages
GLP is a set of functions that allows storage and retrieval of data regarding the
maintenance and status of the electrode. The following information can be
viewed on the third LCD line during measurement:
TEMPERATURE SENSOR problem (if there is one)
CAL DUE or CELL FACTOR
T.Coef. or T.Ref. (User selected)
TIME
DATE
RANGE
BATTERY or CHARGE STATUS
To view more information, enter GLP mode by pressing GLP key.
EC calibration data is stored automatically after a successful calibration.
To view the EC calibration data, press GLP when the instrument is in EC
measurement mode. The instrument will display the calibration standard and
the temperature of the calibrated standard. Use the pq keys to scroll through
the calibration data displayed on the third LCD line.
The cell factor in cm-1 determined from the calibration
with the current reading.
The calibration oset factor in μS/cm together with
the current reading.
EC/TDS GLP
Information
The Solution Standard and calibration temperature.
The temperature coecient used during calibration
with the current reading.
www.GlobalTestSupply.com
Operational Guide39
EC/TDS GLP
Information
The reference temperature together with the current
reading.
The time (hh:mm:ss) of the last calibration together
with the current reading.
The date (yyyy.mm.dd) of the last calibration together
with the current reading.
Calibration Expiration status together with the
current reading:
If disabled, “EXPIRATION WARNING DISABLED” is
displayed.
If enabled, the number of days until the calibration
alarm “CAL DUE” will be displayed. (i.e. “CAL EXPIRES
IN 2 DAYS”)
Operational Guide
40
If enabled, the number of days since the calibration
expired. (I.E. “CAL EXPIRED 2 DAYS AGO”).
The serial number of the probe together with the
current reading.
www.GlobalTestSupply.com
NaCl % Calibration Data in GLP
To view the NaCl% calibration data, press GLP when the instrument is in NaCl%
measurement mode. Use the ARROW keys to scroll through the calibration data.
The instrument will display the calibration temperature and solution.
The cell factor in cm-1 determined from the calibration
with the current reading.
The salinity coecient determined from the
calibration together with the current reading.
The Solution Standard together with the calibrated
temperature.
The time (hh:mm:ss) of the last calibration together
with the current reading.
NaCl % GLP
Information
The date (yyyy.mm.dd) of the last calibration together
with the current reading.
Calibration Expiration status together with the
current reading:
If disabled, “EXPIRATION WARNING DISABLED” is
displayed.
If enabled, the number of days until the calibration
alarm “CAL DUE” will be displayed. (i.e. “CAL EXPIRES
IN 2 DAYS”)
www.GlobalTestSupply.com
Operational Guide41
NaCl % GLP
Information
If enabled, the number of days since the calibration
expired. (I.E. “CAL EXPIRED 2 DAYS AGO”).
The serial number of the probe.
Note: Press GLP at any moment and the instrument will return to measurement
mode.
EC/TDS
Measurements
The RANGE key will change measurement from conductivity to TDS to Salinity.
Conductivity Measurements
Connect the conductivity probe to the instrument and
wait until probe parameters are loaded. The following
message is then displayed on the LCD: “PROBE
CONNECTED”.
Verify if the probe has been calibrated. Suspend the probe into the solution
to be tested. The sleeve holes must be completely submersed. Tap the probe
repeatedly to remove any air bubbles that may be trapped inside the sleeve.
The conductivity value will be displayed on the rst
LCD line, the temperature on the second LCD line and
calibration or range specic information on the third
LCD line.
To toggle between information displayed on the third LCD line, use the ARROW
keys.
If the reading is out of range when the range is set to
automatic, the full-scale value (200.0 mS/cm for ATC
mode or 500.0 mS/cm for absolute conductivity) will
be displayed blinking.
Note: Temperature-compensation and absolute conductivity (NoTC) are
selected in the SETUP conguration.
Operational Guide
42
Automatic Temperature Compensation (ATC): The conductivity probe has
a built-in temperature sensor; the temperature value is used to automatically
compensate the EC/TDS reading. When this option is selected, “ATC” tag is
displayed. A temperature coecient for the sample must also be set.
www.GlobalTestSupply.com
No Temperature Compensation (No TC): The temperature value is displayed,
but not taken into account. When this option is selected, the “NoTC” tag will be
displayed. The reading displayed on the primary LCD is the uncompensated EC
or TDS value.
Note:
• The default compensation mode is ATC.
• The compensation is referenced to the selected reference temperature
(see SETUP for details).
If temperature compensation is selected, measurements are compensated
using the temperature coecient (default value 1.90 %/ºC, is recommended for
natural water samples).
To change the temperature coecient, enter the SETUP mode and select
“T.Coef.(%/ºC)” (see SETUP).
The current temperature coecient can be viewed on the third LCD line by
pressing the pq keys until it is displayed. The value is displayed along with
Cell Factor (the factor that is used to convert the measured conductance to
conductivity, based on the geometry of the cell). If the temperature reading is
out of range, the closest full scale value will be displayed with “ºC” blinking on
the second LCD line.
Error Messages During Measurement
If the EC temperature exceeds the limit of the probe, the message “PROBE OUT
OF SPEC” will scroll on the third LCD line. The temperature will continue to be
displayed blinking. If temperature exceeds the meter specication of 120 ºC,
then “120 ºC” will blink on the display. If interval logging, the message “OUT OF
SPEC.” will alternate with the Log specic messages in both of these cases. The
Log le will indicate a “ºC!” next to the data. In the
case the temperature sensor is damaged, “BROKEN
TEMPERATURE SENSOR” , “---” along with NoTC tag
will be displayed and the unit tag will blink on the LCD.
If logging when this occurs, the Log le will indicate
“ºC!!” next to the data. Absolute conductivity (NoTC)
will be marked with an “A” in the CSV les and dashes
will appear in the temperature eld.
EC/TDS
Measurements
TDS Measurements
Press the RANGE key. The instrument will switch
to TDS measuring range. The TDS reading will be
displayed on the rst LCD line and the temperature
reading on the second LCD line. If the reading is out of
range, the full-scale value (100.0 g/L for ATC mode or
400.0 g/L for uncompensated TDS) will be displayed
blinking along with a message on the display.
Salinity Measurements (Not Available In Basic Mode)
Press the RANGE key twice to switch from conductivity to the congured Salinity
scale. Verify the desired scale is congured in SETUP. The meter supports three
salinity scales: Practical Scale 1978, Percent Scale %, and Natural Sea Water
1966, [g/L]. (Information on the 3 scales follows)
Operational Guide43
www.GlobalTestSupply.com
Salinity
Measurements
Salinity
Measurements
Note:
• These are for determining salinity as they relate to general oceanographic
use.
• Practical salinity and the Natural Seawater require a conductivity
calibration.
• NaCl % requires a calibration in HI 70371 standard.
PSU - Practical Salinity Units
The practical salinity (S) of seawater relates the ratio of electrical conductivity
of a normal seawater sample at 15 ºC and 1 atmosphere to a potassium chloride
solution (KCl) with a mass of 32.4356 g/Kg water at the same temperature and
pressure.
Under these conditions the ratio is equal to 1, and S=35. The practical salinity
scale may be applied to values to through 42 PSU at temperatures between -2
to 35 ºC.
According to the denition, salinity of a sample in PSU (practical salinity units) is
calculated using the following formula:
Operational Guide
44
R
- ratio of sample conductivity to standard conductivity at Temp =(T)
T
CT(sample) - uncompensated conductivity at T ºC;
C(35,15) =42.914µS/cm - the corresponding conductivity of KCI solution
containing a mass of 32.4356 g KCl/1 Kg solution
rT - Temperature compensation polynomial
The NaCl % scale is an older salinity scale used for seawater salinity. In this scale
100% salinity is equivalent to roughly 10% solids. High percentages were made
by evaporation. To display NaCl in % units, enter SETUP and select NaCl% unit.
Press the RANGE key until “NaCl%” is displayed on the LCD. The instrument will
display the NaCl% reading on the rst LCD line and the temperature reading on
the second LCD line.
www.GlobalTestSupply.com
If the reading is out of range, the full-scale value
(400.0%) will be displayed blinking.
If this occurs during a log, a “!“ will be placed next to
the measurement unit. Data found in the CSV les
with a “!“ should not be considered reliable.
Natural Sea Water Scale
The Natural Sea Water Scale extends from 0 - 80.0 g/L. It determines salinity
based upon a conductivity ratio of sample to “standard seawater” at 15 °C.
Salinity
Measurements
Where R
equation:
S = - 0.08996 + 28.2929729R
- 10.67869R
is the conductivity ratio and salinity is dened by the following
15
3
+ 5.98624R
15
+ 12.80832R
15
4
-1.32311R
15
2
15
5
15
Note: The formula can be applied for temperatures
between 10 ºC and 31 ºC.
If the reading is out of the measurement range, the display will ash the highest
value possible and a warning message will be displayed. If this occurs during a
log a “!” will be placed next to the measurement unit. Data found in the CSV le
with a ”!” should not be considered reliable.
www.GlobalTestSupply.com
Operational Guide45
Dissolved
Oxygen
Setup
Steps To Optimize
1. Determine if Concentration or % Saturated measurements will be made.
2. Prepare the Dissolved Oxygen (DO) probe for measurement.
3. Connect the probe to the meter and congure the SETUP parameters.
4. Calibrate the DO sensor.
5. Take measurements using the DO sensor.
Measurements Available
Concentration measurements in water, and % oxygen-saturated measurements
are available using edge® together with HI 764080 DO probe. Algorithms
used for concentration measurements (units of ppm or mg/L) are based upon
the oxygen solubility in air-saturated fresh water. Compensation for salinity
and altitude are made by conguring SETUP options. Percent saturation
measurements are based upon the partial pressure of oxygen and are suitable
for measurement in samples other than air-saturated fresh water. It is advised
to check material compatibility of the probe with the sample.
DO Probe Preparation
CAUTION: Use care during servicing and use. The HI 764080 contains a glass
insulator. Do not drop or handle carelessly.
Probes from Hanna Instruments are shipped dry.
1. Carefully remove the cardboard shipping tube used to
protect the probe during shipping. Save the tube, should the
probe be stored dry again.
2. Open membrane package and remove one O-ring and one
membrane cap.
3. Rinse the membrane cap with a small amount of HI7041
electrolyte and discard.
4. Position o-ring in cap as indicated. Rell membrane cap 3/
full with electrolyte solution, ensure to cover the o-ring.
5. Holding the membrane cap by the top, tap the side walls
of the membrane cap to dislodge gas bubbles and force them
to rise to the surface. Do not tap on the membrane directly
as it may damage it.
6. With the probe facing down, slowly screw the cap
counterclockwise until completely tightened. Electrolyte will
overow.
7. Rinse outer body of the probe and inspect membrane for
entrapped gas bubbles. The cathode area should be free of
bubbles.
8. Connect the DO probe to edge® meter and turn meter on.
9. Allow probe conditioning (polarizing) function to occur.
During this process, the following message will be
displayed on the LCD, “DISSOLVED OXYGEN PROBE
CONDITIONING”.
The conditioning message will be displayed for about
60 seconds while the DO probe is conditioned. If the
probe was conditioned and a new conditioning is not
necessary, press any key to enter measurement mode.
4
Operational Guide
46
www.GlobalTestSupply.com
The probe is polarized with a xed voltage of approximately 800 mV between
the cathode and anode. Probe polarization is essential for stable measurements.
With the probe properly polarized, oxygen is continually consumed as it passes
through gas permeable PTFE membrane.
If polarization is interrupted, the electrolyte solution continues to be
enriched with oxygen until it reaches an equilibrium with the surrounding
solution. Whenever measurements are taken with a non-polarized probe, the
measurement will be drifty and inaccurate. The measurement will jump when
the probe is moved.
Note: When not in use and during polarization, use the protective transparent
cap.
Dissolved
Oxygen
Setup
Dissolved
Oxygen
Probe
Diagram
1. Strain Relief
2. Probe Cap
3. PEI Probe Body
4. Temperature Sensor
5. Threads for Membrane Cap
6. Ag/AgCl Anode and Reference
7. Glass Insulator
8. Platinum Cathode
9. O-Ring
10. Disposable Membrane Cap
11. Oxygen Permeable PTFE Membrane
12. Shipping Tube
www.GlobalTestSupply.com
Operational Guide47
Dissolved
Oxygen
Meter
Conguration
DO (Dissolved Oxygen) meter operation is congured using the SETUP key
with a DO probe connected to the meter. The parameter-specic options will
be seen inserted into the menu. There is no Basic mode for Dissolved Oxygen
measurements.
ParameterDescriptionChoicesDefault
Altitude (m)
Salinity (g/L)
DO Units
Concentration
measurements of
dissolved oxygen
change depending
on atmospheric
pressure. A
convenient way
to estimate
atmospheric
pressure eects
is by the related
parameter of
elevation (m) above
or below sea level.
Enter altitude in
meters closest to
the actual altitude
to ensure the most
accurate calibration
and concentration
measurement.
Dissolved oxygen
solubility decreases
if water contains
salts. Selecting this
factor as to be close
to your known salt
level, will improve
the accuracy of
DO concentration
calibration and
measurement.
Select preferred
measurement units
for DO concentration.
Temperature, altitude and salinity compensation are used for DO concentration
measurements (ppm or mg/L). When the water is colder, it can hold more
dissolved oxygen, when it is warmer it holds less oxygen. Compensation
for temperature-related solubility is done automatically using the built-in
temperature sensor within the DO probe and algorithms in edge®. When water
is measured at an altitude below sea level, oxygen solubility increases, but
above sea level the oxygen solubility decreases. To compensate for this during
calibration and measurement, the user must provide the approximate altitude
(in meters) in the SETUP menu. The settings are in 100 m increments; select
the value closest to the actual altitude. Some examples of altitudes around the
world follow:
LocationMeterFeet
Sebkha paki Tah, Morocco-55-180
Lake Frome, Australia-6-20
Netherlands, coastal providence-1 to -7-3 to -23
Lake Michigan, USA176577
Lake Geneva; France, Switzerland3721220
Denver, CO USA16095279
Mount Everest884829029
The solubility of oxygen in water is also inuenced by the amount of salt the
water contains. Seawater typically has a salinity of 35 g/L and the oxygen
solubility is 18 % less compared to fresh water at 25 °C. By entering the
approximate salinity value, the calibration and subsequent concentration
measurement will be compensated to display the correct oxygen concentration.
A 18 % error would result if the salinity value is not entered.
Dissolved
Oxygen
Meter
Conguration
Note: Salinity and Altitude have no eect on % oxygen solubility range.
When water is fresh, containing no sea water, the concentration of oxygen will
be at a maximum. The solubility of the oxygen dissolved in water is decreased
when water is brackish or seawater. The solubility of oxygen in water is
decreased when measurements are made at elevations above sea level.
Before proceeding with the calibration, make sure the probe is ready for
measurements (see page 46), i.e. the membrane cap is lled with electrolyte,
the probe is connected to the meter and properly polarized. For an accurate
calibration, it is recommended to wait at least 15 minutes to ensure conditioning
of the probe. Keep the protective cap on during polarization time and remove
it for calibration and measurements. Follow the calibration procedure. Calibrate
the probe frequently, especially if high accuracy is required. The probe can be
calibrated at 2 points: 100.0 % (slope calibration) and 0.0 % (zero calibration).
Initial Preparation
Prepare a fresh bottle of HI 7040 by following package directions. Use solution
within one month of preparation. Pour small quantities of HI 7040 Zero Oxygen
solution into a beaker. If used, remove the protective cap from the DO probe.
Operational Guide49
www.GlobalTestSupply.com
Dissolved
Oxygen
Calibration
Dissolved
Oxygen
Calibration
100% Saturated Calibration
It is suggested to perform the slope calibration in water-saturated air.
Pour water into a small beaker.
Rinse the polarized probe with clean water.
Dry the probe tip and allow a few seconds for the LCD reading to stabilize (probe
in air).
Suspend probe with membrane just over the beaker of water.
Press CAL.
The “ ” tag will be displayed along with “WAIT” blinking
on the LCD until the reading is stable.
When the reading is stable and is within the limits,
“CFM” tag starts blinking. Press CFM to conrm the
100.0 % DO calibration.
Press CAL to leave calibration after the rst point. The instrument will display
“SAVING” message and it will return to measurement mode memorizing the
slope calibration data
Zero Calibration
Press CAL or continue with calibration after conrming
rst point. The meter will display “WAIT” and “0.0 %”
will be displayed in lower right corner.
Operational Guide
50
Submerse the probe membrane and temperature
sensor into the beaker containing HI 7040 Zero
Oxygen solution and stir gently for 2-3 minutes, the
reading will go down.
When the reading is stable and stops decreasing,
“CFM” tag starts blinking. Press CFM to conrm the
0.0 % DO calibration.
The instrument will display “SAVING” message and it
will return to measurement mode. Rinse probe tip o
in water before taking measurements in samples.
www.GlobalTestSupply.com
If the reading is outside limits, “WRONG STANDARD“
message will be displayed.
If the temperature is out of range (0.0 - 50.0 ºC) during
calibration, then the message “WRONG STANDARD
TEMPERATURE” will be displayed and temperature
value will blink.
Dissolved
Oxygen
Calibration
Messages
GLP refers to a quality control function used to ensure uniformity of probe
calibrations and measurements. The dedicated GLP key opens a le of the latest
calibration information. Use the qp keys to scroll the stored information. This
includes the standards used, temperature of the standard, Altitude and Salinity
factors, time and date of the last calibration, the expired calibration information
and the probe serial number. This information is also included with all logged
data.
Last DO Calibration Data
The last DO calibration data is stored automatically
after a successful calibration. To view the DO
calibration data, press GLP when the instrument is in
measurement mode.
The instrument will display:
The calibration standard and the calibration
temperature:
0.0 % will be displayed if the instrument was
calibrated at this point.
100 % calibration point, if instrument was calibrated
in water-saturated air.
The altitude and salinity setting at the moment of
calibration together with the current reading.
Dissolved
Oxygen
GLP
Information
The time of day that the calibration was performed
together with the current reading.
www.GlobalTestSupply.com
Operational Guide51
Dissolved
Oxygen
GLP
Information
The date of the calibration together with the current
reading.
Calibration Expiration status together with the
current reading: If disabled, “EXPIRATION WARNING
DISABLED” is displayed.
If enabled, the number of days until the calibration
alarm “CAL DUE” will be displayed. (i.e. “CAL EXPIRES
IN 2 DAYS”)
If enabled, the number of days the calibration has
expired (i.e. “CAL EXPIRED 2 DAYS AGO”).
The probe serial number together with the current
reading.
Dissolved
Oxygen
Measurements
Operational Guide
52
Make sure that the probe is polarized, calibrated and the protective cap has been
removed.
Rinse probe.
Submerse the probe in the sample to be tested, make sure temperature probe is
also immersed. Allow reading to stabilize.
Note: The sample should be stirred when taking a reading.
The Dissolved Oxygen value (in %) is displayed on the
rst LCD line and the temperature on the secondary
LCD line.
Press RANGE to change the reading from % to ppm
(mg/L) and vice versa.
www.GlobalTestSupply.com
Oxygen is consumed during the measurement. For accurate DO measurements,
water movement of 0.3 m/s is suggested. This is to ensure that the oxygendepleted membrane surface is constantly measuring a representative sample.
The use of a magnetic stirrer is recommended. The probe has a built-in
temperature sensor. Make sure it is also in sample. The measured temperature
is indicated on the second LCD line as shown on page 52. Allow the probe to
reach thermal equilibrium before taking any measurement. This can take several
minutes. The greater the dierence between the temperature at which the
probe was stored and the temperature of the sample, the longer the time will
be.
If the DO temperature exceeds the limits of the
probe, the message “PROBE OUT OF SPEC” will scroll
on the third LCD line and LCD will display dashes. If
the temperature exceeds the meter specication
120 ºC, then “120 ºC” will blink on the display. If interval
logging, the message “OUT OF SPEC.” will alternate
with the Log specic messages in both of these cases.
The Log le will indicate “ºC!” next to the data. In the
case the temperature sensor is damaged, “BROKEN
TEMPERATURE SENSOR” will be displayed and the
temperature will display “----” along with the unit
tag blinking on the second LCD line. The Log le will
indicate “ºC!!” next to the data.
Dissolved
Oxygen
Measurements
www.GlobalTestSupply.com
Operational Guide53
pH Probe
Maintenance
Remove the protective cap of the pH electrode.
DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT.
This is normal with electrodes. They will disappear when rinsed with water.
54
During transport, tiny bubbles of air may form inside the glass bulb aecting
proper functioning of the electrode. These bubbles can be removed by “shaking
down” the electrode as you would do with a glass thermometer. If the bulb and/
or junction is dry, soak the electrode in HI 70300 or HI 80300 storage solution
for at least one hour.
For rellable electrodes:
If the lling solution (electrolyte) is more than 2½ cm (1”) below the ll hole, add
HI 7082 or HI 8082 3.5M KCl Electrolyte Solution for double junction electrodes.
Unscrew the ll hole cover during measurements so the liquid reference junction
maintains an outward ow of electrolyte.
Measurement
Rinse the electrode tip with distilled water. Submerse the tip 3 cm (1¼”) in the
sample and stir gently for a few seconds.
For a faster response and to avoid cross-contamination of the samples, rinse
the electrode tip with a few drops of the solution to be tested, before taking
measurements.
Storage Procedure
To minimize clogging and ensure a quick response time, the glass bulb and the
junction should be kept moist and not allowed to dry out.
Maintenance
www.GlobalTestSupply.com
Replace the solution in the protective cap with a few drops of HI 70300 or
HI 80300 Storage Solution or, in its absence, Filling Solution (HI 7082 or
HI 8082 for double junction electrodes).
Follow the preparation procedure before taking measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.
Periodic Maintenance
Inspect the electrode and the cable. The cable used for connection to the
instrument must be intact and there must be no points of broken insulation on
the cable or cracks on the electrode stem or bulb. Connectors must be perfectly
clean and dry. If any scratches or cracks are present, replace the electrode. Rinse
o any salt deposits with water.
For rellable electrodes: Rell the reference chamber with fresh electrolyte
(HI 7082 or HI 8082 for double junction electrodes). Allow the electrode to
stand upright for 1 hour.
Follow the Storage Procedure above.
Cleaning Procedure
Use diagnostic messages to aid pH electrode troubleshooting. Several cleaning
solutions are available:
• General – Soak in Hanna HI 7061 or HI 8061 General Cleaning Solution
for approximately ½ hour.
• Protein –Soak in Hanna HI 7073 or HI 8073 Protein Cleaning Solution
for 15 minutes.
• Inorganic – Soak in Hanna HI 7074 Inorganic Cleaning Solution for 15
minutes.
• Oil/grease – Rinse with Hanna HI 7077 or HI 8077 Oil and Fat Cleaning
Solution.
pH Probe
Maintenance
Note: After performing any of the cleaning procedures, rinse the electrode
thoroughly with distilled water, rell the reference chamber with fresh
electrolyte (not necessary for gel-lled electrodes) and soak the electrode
in HI 70300 or HI 80300 Storage Solution for at least 1 hour before taking
measurements.
Maintenance55
www.GlobalTestSupply.com
EC Probe
Maintenance
Rinse the probe with clean water after measurements. If a more thorough
cleaning is required, remove the probe sleeve and clean the probe with a cloth
or a nonabrasive detergent. Make sure to reinsert the sleeve onto the probe
properly and in the right direction. After cleaning the probe, recalibrate the
instrument.
The insulator used to support the platinum rings is made of glass. Use extreme
caution when handling this probe.
56
Maintenance
www.GlobalTestSupply.com
The DO probe body is made of PEI.
A temperature sensor provides temperature measurements of the sample. Use
the protective cap when the probe is not in use. To replace the membrane or
rell with electrolyte, proceed as follows:
For a new probe, remove the protective shipping tube by gently
twisting and pulling it o the body of the probe (see g. 1).
If the membrane was previously installed, unscrew the membrane
cap by turning it clockwise (see g. 2).
The new membrane cap should be rinsed with electrolyte solution.
Rell with clean electrolyte solution.
Gently tap the sides of the membrane cap to ensure that no air
bubbles remain trapped. Do not tap the bottom directly with your
nger, as this will damage the membrane.
Make sure that the rubber O-ring sits properly inside the membrane
cap. With the sensor facing down, slowly screw the membrane cap
clockwise. Some electrolyte will overow.
Dissolved
Oxygen
Probe
Maintenance
The platinum cathode (DO Probe Diagram, page 47) should always be bright
and untarnished. If it is tarnished or stained, the cathode should be cleaned.
You can use a clean lint-free cardboard or cloth. Rub the cathode very gently
side to side 4-5 times. This will be enough to polish and remove any stains
without damaging the platinum tip. Afterwards, rinse the probe with deionized
or distilled water and install a new membrane cap using fresh electrolyte and
follow DO Probe Preparation page 46.
Important
In order to have accurate and stable measurements, it is important that the
membrane surface is in perfect condition. This semipermeable membrane
isolates the sensor elements from the environment but allows oxygen to
enter. If any dirt is observed on the membrane, rinse carefully with distilled or
deionized water. If imperfections still exist, or any damage is evident (such as
wrinkles or tears-holes), the membrane should be replaced. Make sure that the
O-Ring sits properly in the membrane cap.
Maintenance57
www.GlobalTestSupply.com
Troubleshooting
Guide
SymptomsProblemsSolution
Slow response/excessive
drift.
Dirty pH electrode.
Clean the electrode and then soak
the tip in HI 7061 or HI 8061 for 30
minutes.
Readings uctuate up
and down (noise).
The meter does not
accept the buer/
standard solution for
calibration.
If the display shows: “pH”
and “-2.00” or “16.00”
blinking.
The display shows EC,
TDS or Salinity readings
blinking.
EC: EC probe sleeve not
properly inserted; air
bubbles inside sleeve.
DO: DO probe electrolyte
contains entrapped gas.
pH: Dirty electrode or
contaminated buer.
EC: The EC probe is
defective.
Out of range in the pH
scale.
Out of range in EC, TDS or
Salinity scale.
Out of range in DO scale.
Clean the electrode. Rell with fresh
solution (for rellable electrodes only).
Check cable and connectors.
Reinstall the sleeve. Tap the probe to
remove air bubbles. Move to center
of beaker. Verify top hole in sleeve is
covered with solution.
Remove cap. Rell, tap and reinstall.
Follow the cleaning procedure. If still
no results, replace the electrode.
Replace buer.
Follow the cleaning procedure. If still
no results replace the probe. Verify
correct standard is selected.
A) Verify that the shipping cap has
been removed.
B) Make sure the pH sample is in the
specied range.
C) Check electrolyte level and general
state of the electrode.
Verify the plastic shipping spacer has
been removed from probe. Recalibrate
the probe. Make sure the solution is in
specied range. Make sure the range is
not locked. (Select Auto range)
Verify area of cathode is free of
bubbles inside cap. Verify solution
movement past membrane. Remove
cap, inspect and clean if necessary.
Install new cap, fresh electrolyte with
no bubbles, permit longer polarization.
Stir or increase ow rate.
58
Maintenance
www.GlobalTestSupply.com
SymptomsProblemsSolution
A) Verify the shipping cap has been
removed.
If the display shows:
“mV” and “-1000” or
“1000” blinking.
The meter does not
measure temperature.
“----” is displayed on
second LCD line.
The meter fails to
calibrate NaCl.
The meter fails to
calibrate or gives faulty
readings.
At startup the meter
displays all LCD tags
permanently.
CAL “Prod” message at
startup.
Out of range in the mV
scale.
Broken temperature
sensor.
Incorrect EC calibration.
Broken pH electrode.Replace the electrode.
One of the keys is stuck.
Instrument was not
factory calibrated or lost
factory calibration.
B) Make sure the sample pH is within
specied range.
C) Verify electrolyte level in pH sensor
is topped o.
D) Verify no bubbles inside pH
membrane.
Replace the probe.
Recalibrate the meter in EC range. Set
cell constant to 1.
Check the keyboard or contact your
local Hanna Oce.
Contact Hanna Technical Support for
help.
Troubleshooting
Guide
www.GlobalTestSupply.com
Maintenance59
Specications
pH
Specications
Specications
EC
Range
Resolution
Accuracy
@ 25 °C / 77 °F
pH Calibration
Temperature
compensation
pH Electrode
Log feature
Range
Resolution
Accuracy
@ 25 °C / 77 °F
pHTemperature
-2.00 to 16.0 0 pH
-2.000 to 16.000 pH*
±1000.0 mV
0.01 pH
0.001 pH*
0.1 mV
±0.01pH
±0.002 pH*
±0.2 mV
Autom atic, up to 3 p oints (5 po ints*) calib ration, 5 s tandard ( 7
stan dard*) buf fers ava ilable (1.68 *, 4.01 or 3.00, 6 .86, 7.01, 9.18, 10.0 1,
12.45*) and 2 c ustom buf fers*
Autom atic -5 to 100º C ( 23 to 212º F) (us ing integr al temper ature sen sor)
HI 11310 In telligent p H/tempe rature el ectrod e (include d)
Up to 1000 * records or ganized in :
Log on de mand (Max . 200 logs)
Log on s tabilit y (Max. 200 l ogs)
Interval logging*
-20.0 to 120 .0 ºC (-4.0 to 248.0 °F)**
0.1 ºC
±0.5 ºC
ECTDSSalinity
0.00 to 29.9 9 μS/cm,
3.00 to 29 .99 μS/cm,
30.0 to 29 9.9 μS/cm,
300 to 2999 μS/cm,
3.00 to 29 .99 mS/cm,
30.0 to 20 0.0 mS/cm,
up to 500. 0 mS/cm,
absolute
conductivity ‡
0.01 μS/cm,
00.1 μS/cm,
0.01 μS/cm,
0.01 mS /cm,
00.1 mS/cm
±1% of readin g
(±0.05 μS/cm or 1
digit, whichever is
greater)
0.00 to 14.99 pp m (mg/L),
15.0 to 149.9 ppm (mg /L),
150. to 1499. ppm (mg/L),
1.50 to 14.99 g /L,
15.0 to 100. 0 g/L,
up to 400 .0 g/L abso lute
TDS ‡ (wit h 0.80 fact or)
0.01 ppm,
0.1 ppm,
1 ppm,
0.01 g/L ,
0.1 g/L
±1% of readin g
(±0.03 ppm or 1 dig it,
whichever is greater)
0.0 to 400.0 % NaCl *,
2.00 to 42.0 0 PSU *,
0.0 to 80.0 g/ L *
0. 1% Na Cl ,
0.01 P SU,
0.01 g/L
±1% of readin g
60
Specications
EC Calibration
Salin ity
Calib ration %Na Cl
(onl y)
www.GlobalTestSupply.com
1 Cell Fac tor calib ration;
6 stan dards ava ilable: 84, 1413 μS/cm , 5.00, 12.88, 80.0, 1 11.8 mS/cm,
1 point o ffset : 0.00 μS/cm
Single p oint with H I 7037L Standa rd
Temperature
compensation
Conductivity
temperature
coefficient
TDS factor
EC Prob e
Log feature
ECTDSSalinity
Autom atic -5 to 100 ºC ( 23 to 212 ºF)
NoTC - can be se lected to m easure
absolute conductivity.
0.00 to 6.0 0% / ºC (for EC and T DS only). Def ault value i s 1.90% / ºC
0.40 to 0.8 0 (default val ue is 0.50)
HI 763100
Up to 1000 * (40 0) records or ganized in : Manual log o n demand (Ma x. 200
logs), Man ual log on st ability ( Max. 200 lo gs), Interv al logging * ( Max. 600
logs; 100 lots)
Specications
DO
0.00 to 45.00 p pm (mg/L)
Range
Resolution
Accuracy
@ 25 °C / 77 °F
DO Calibration
Altitude Compensation
Resolution
Salin ity Compe nsation
Resolution
Temperature
Compensation
DO Pro be
Log Feature
0.0 to 300.0 %
0.0 to 50.0 ºC ( 32.0 to 122.0 °F )**
0.01 ppm (mg /L)
0. 1%
0.1 °C
±1.5% of re ading or ±1 dig it
±0.5 ºC
One or t wo points a t 0% (HI 7040) an d 100% (water s aturat ed
air)
-500 to 4,0 00 m (-1640 to 1 3120’)
100 m (328 ’)
0 to 40 g/ L
1 g/L
0.0 to 50.0 ºC
(32.0 to 12 2 ºF)
HI 764080
Up to 1000 r ecords or ganized in M anual Log on D emand
(Max. 2 00 logs)
Manual L og on Stabil ity (Max . 200 logs)
Inter val Logg ing 100 lots ,
Max. 6 00 logs/l ot
Additional Specications
PC Int erface
Storage Interface
Power Supply
Environment
Dimensions
Weight
*Standard Mode Only
** Temperature limits will be reduced to actual probe/sensor limits.
‡Absolute conductivity (or TDS) is the conductivity (or TDS) value without temperature
compensation.
Micro US B
USB
5 VDC Ada pter (incl uded)
0-50 ºC ( 32-122 ºF) M ax 95% RH non- condensi ng
202 x 140 x 12 mm ( 7.9 x 5.5 z 0.5”)
250g (8.8 2 oz)
www.GlobalTestSupply.com
Dissolved
Oxygen
Specications
Specications61
Accessories
Electrodes/Probes
HI 10530
HI 10430
HI 11310Glass body, double junction, rellable pH/temperature electrode
HI 11311
HI 12300
HI 12301
HI 763100EC/temperature probe
HI 764080DO/temperature probe
Triple ceramic, single junction, low temperature glass, rellable pH
electrode with conical tip and temperature sensor
Single ceramic,double junction, high temperature glass, rellable pH
electrode with temperature sensor
Glass body, double junction, rellable pH/temperature electrode with
enhanced diagnostics
Plastic body, double junction, gel lled, non rellable pH/temperature
electrode
Plastic body, double junction, gel lled, non rellable pH/temperature
electrode with enhanced diagnostics
pH
Buer Solutions
HI 70004PpH 4.01 Buer Sachets, 20 mL (25 pcs.)
HI 70007PpH 7.01 Buer Sachets, 20 mL (25 pcs.)
HI 70010PpH 10.01 Buer Sachets, 20 mL (25 pcs.)
HI 7001LpH 1.68 Buer Solution, 500 mL
HI 7004LpH 4.01 Buer Solution, 500 mL
HI 7006LpH 6.86 Buer Solution, 500 mL
HI 7007LpH 7.01 Buer Solution, 500 mL
HI 7009LpH 9.18 Buer Solution, 500 mL
HI 7010LpH 10.01 Buer Solution, 500 mL
HI 8004LpH 4.01 Buer Solution in FDA approved bottle, 500 mL
HI 8006LpH 6.86 Buer Solution in FDA approved bottle, 500 mL
HI 8007LpH 7.01 Buer Solution in FDA approved bottle, 500 mL
HI 8009LpH 9.18 Buer Solution in FDA approved bottle, 500 mL
HI 8010LpH 10.01 Buer Solution in FDA approved bottle, 500 mL
ELECTRODE STORAGE SOLUTIONS
HI 70300LStorage Solution, 500 mL
HI 80300LStorage Solution in FDA approved bottle, 500 mL
ELECTRODE CLEANING SOLUTIONS
HI 70000PElectrode Rinse Sachets, 20 mL (25 pcs.)
HI 7061LGeneral Cleaning Solution, 500 mL
HI 7073LProtein Cleaning Solution, 500 mL
HI 7074LInorganic Cleaning Solution, 500 mL
HI 7077LOil & Fat Cleaning Solution, 500 mL
HI 8061LGeneral Cleaning Solution in FDA approved bottle, 500 mL
HI 8073LProtein Cleaning Solution in FDA approved bottle, 500 mL
HI 8077LOil & Fat Cleaning Solution in FDA approved bottle, 500 mL
ELECTRODE REFILL ELECTROLYTE SOLUTIONS
HI 70823.5M KCl Electrolyte, 4x30 mL, for double junction electrodes
HI 8082
3.5M KCl Electrolyte in FDA approved bottle, 4x30 mL, for double junction
electrodes.
62
Accessories
www.GlobalTestSupply.com
EC
Conductivity Solutions
HI 70030P12880 μS/cm, 20 mL sachets (25 pcs.)
HI 70031P1413 μS/cm, 20 mL sachets (25 pcs.)
HI 70039P5000 μS/cm, 20 mL sachets (25 pcs.)
HI 7030M12880 μS/cm, 230 mL bottle
HI 7031M1413 μS/cm, 230 mL bottle
HI 7033M84 μS/cm, 230 mL bottle
HI 7034M80000 μS/cm, 230 mL bottle
HI 7035M111800 μS/cm, 230 mL bottle
HI 7039M5000 μS/cm), 230 mL bottle
HI 7030L12880 μS/cm, 500 mL bottle
HI 7031L1413 μS/cm, 500 mL bottle
HI 7033L84 μS/cm, 500 mL bottle
HI 7034L80000 μS/cm, 500 mL bottle
HI 7035L111800 μS/cm, 500 mL bottle
HI 7039L5000 μS/cm, 500 mL bottle
HI 7037L100% NaCl sea water standard solution, 500 mL bottle
HI 8030L12880 μS/cm, 500 mL FDA approved bottle
HI 8031L1413 μS/cm, 500 mL FDA approved bottle
HI 8033L84 μS/cm, 500 mL FDA approved bottle
HI 8034L80000 μS/cm, 500 mL FDA approved bottle
HI 8035L111800 μS/cm, 500 mL FDA approved bottle
HI 8039L5000 μS/cm, 500 mL FDA approved bottle
DO
HI 7040LZero Oxygen Solution, 500 mL
HI 7041SRelling Electrolyte Solution, 30 mL
HI 764080Spare DO probe
HI 764080A/P5 spare membranes
Accessories
Other Accessories
HI 75110/220UVoltage adapter from 115 Vac to 5 Vdc (USA plug)
HI 75110/220EVoltage adapter from 230 Vac to 5 Vdc (European plug)
HI 76404BElectrode holder
HI 2000WCWall cradle
HI 2000BCBench cradle
www.GlobalTestSupply.com
Accessories
63
Warranty
edge® is guaranteed for two years against defects in workmanship and
materials when used for its intended purpose and maintained according
to instructions. Electrodes and probes are guaranteed for six months. This
warranty is limited to repair or replacement free of charge. Damage due to
accidents, misuse, tampering or lack of prescribed maintenance is not covered. If
service is required, contact your local Hanna Oce. If under warranty, report the
model number, date of purchase, serial number and the nature of the problem.
If the repair is not covered by the warranty, you will be notied of the charges
incurred. If the instrument is to be returned to Hanna Instruments, rst obtain a
Returned Goods Authorization number from the Technical Service department
and then send it with shipping costs prepaid. When shipping any instrument,
make sure it is properly packed for complete protection.
64
Warranty
Hanna Instruments reserves the right to modify the design, construction or appearance of its
products without advance notice.
www.GlobalTestSupply.com
MANEDGE 05/15
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.