Instruction Manual
HI 141 Series
NTC Thermologgers
& HI 141001 Infrared Transmitter
www.hannainst.com
WARRANTY
HI 141 loggers are warranted for one year against defects in workmanship and
materials when used for their intended purpose and maintained according to instructions. This warranty is limited to repair or replacement free of charge. Damages due to
accident, misuse, tampering or lack of prescribed maintenance are not covered. If
service is required, contact the dealer from whom you purchased the instrument. If
under warranty, report the model number, date of purchase, serial number and the
nature of the failure. If the repair is not covered by the warranty, you will be notified
of the charges incurred. If the instrument is to be returned to Hanna Instruments, first
obtain a Returned Goods Authorization Number from the Customer Service department
and then send it with shipment costs prepaid. When shipping any instrument, make sure
it is properly packaged for complete protection.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for correct
operation.
If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
Each thermologger comes complete with one 3.6V Lithium
AA battery, HI 740221 magnetic start key and instructions.
Note: Conserve all packing material until the instrument
has been observed to function correctly because any
defective item must be returned to the Dealer in its
original packing.
GENERAL DESCRIPTION
HI 141 series of temperature data loggers are provided with
either one or two channels, and an optional LCD. The user
interacts with the logger, setting data acquisition parameters
or downloading logged data, through a RS232 serial port on
a Windows® PC. The HI 141000 Windows® application software supports the communication between the logger and the
PC through the HI 141001 Infrared Transmitter.
The main features of the HI 141 thermologgers include:
• One or two channels, with internal or external sensor;
• 16000 samples (for 1 channel models) or
8000 samples/channel (for 2 channels models);
• Logging interval from 1 second to 24 hours;
• Storing of temperature at logging interval, or min or
max temperature between logging intervals;
• Logging delay start up to 199 hours, and magnetic start
through magnetic key provided with the thermologger;
• Programmable high and low alarms;
• Non-volatile storage of logging parameters and data in
EEPROM;
• BEPS (Battery Error Prevention System);
• Security Password and lot serial number;
• Optional LCD with status, temperature and information
display;
• IP67 waterproof casing.
The expected life is about 4 years at 1 minute measuring and
logging intervals.
All HI 141 thermologgers are factory calibrated.
The optional LCD is divided into 3 areas: the status on the top,
the temperature readout in the middle, and useful information
at the bottom.
All rights are reserved. Reproduction in whole or in part is prohibited without the written
consent of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
Recommendations for users
Before using these products, make sure that they are entirely suitable for the environment in which
they are used. Operation of these instruments in residential areas could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary steps to correct
interferences. Any variation introduced by the user to the supplied equipment may degrade the
instruments’ EMC performance. To avoid damages or burns, do not perform any measurement in
microwave ovens.
IST141R2 05/05
Several models are available, as shown in the following table:
The information shown on the bottom is user-selectable, and
can include the current number of samples, the samples
beyond alarm limits, the min and max temperature, and the
high and low alarm thresholds.
FUNCTIONAL DESCRIPTION
SPECIFICATIONS
START UP
1. Hook*
2. Green LED
3. LCD*
4. Photo LED
5. Internal sensor*
6. Connection for external sensor(s)*
7. Infrared LED
8. Red LED
9. Internal magnetic switch: touch
the top of the meter with magnetic
key for activating magnetic start
* depending on model
The status of the thermologger is indicated by the green and
red LEDs, and optional LCD:
Green LED will light up when magnetic switch is activated by
the key. When the key is removed from the top of the meter,
the green LED switches off.
Model Sensor(s) Range
HI 141A (H) T1 internal -40.0 to 80.0ºC / -40.0 to 175.0ºF
HI 141B (H) T1 external -40.0 to 125.0ºC / -40.0 to 257.0ºF
HI 141C (H)* T1 internal -20.0 to 70.0ºC / -4.0 to 158.0ºF
HI 141D (H) T1 external -40.0 to 125.0ºC / -40.0 to 257.0ºF
HI 141E (H) T1 internal -40.0 to 80.0ºC / -40.0 to 175.0ºF
T2 external -40.0 to 125.0ºC / -40.0 to 257.0ºF
HI 141F (H) T1 external -40.0 to 125.0ºC / -40.0 to 257.0ºF
T2 external -40.0 to 125.0ºC / -40.0 to 257.0ºF
HI 141G (H)* T1 internal -20.0 to 70.0ºC / -4.0 to 158.0ºF
T2 external -40.0 to 125.0ºC / -40.0 to 257.0ºF
HI 141J (H) T1 external -40.0 to 125.0ºC / -40.0 to 257.0ºF
T2 external -40.0 to 125.0ºC / -40.0 to 257.0ºF
* Warning: T1 temperature range is limited by LCD characteristics.
Do not use the logger out of this range.
Note: For models with two external sensors, T1 and T2 channels are
indicated on the probes (T1 is red marked, T2 is white).
Resolution 0.1ºC (-40.0 to 100.0ºC); 0.2ºC (temp. >100.0ºC)
0.1ºF (-40.0 to 190.0ºF) ; 0.3ºF (temp. >190.0ºF)
Accuracy
EXTERNAL DIAMETER 86.5 mm (3.4”) 159 mm (6.3”)
HEIGHT 35 mm (1.4”) 60 mm (23.6”)
WEIGHT 150 g (5.54 oz.) 200 g (7.39 oz.)
±0.5ºC (-40.0 to 0.0 and 70.0 to 100.0 ºC)
±0.4ºC (0.0 to 70.0 ºC)
±1.0ºC (temperature>100.0ºC)
±1.0ºF (-40.0 to 32.0 and 158.0 to 212.0ºF)
±0.8ºF (32.0 to 158.0 ºF)
±2.0ºF (temperature>212.0ºF)
HI 141 HI 141001
In order to communicate with the
HI 141 thermologger, the
HI 141000 application software
must be installed on your PC.
This is done by inserting the first
floppy disk in the driver and
running Setup.exe.
Connect the HI 141001 Infrared
Transmitter to your computer
through an available RS232 port.
Place the HI 141 on the transmitter, taking care to align the
ribs on the logger with the slots
on the transmitter
Run HI 141000 software and
refer to the on-line help for any
further information about setting
the parameters of the logger,
acquiring logged data, etc.
BATTERY REPLACEMENT
Battery replacement must only take place in a non hazardous
area using the appropriate battery type (3.6V Lithium AA
battery).
When the battery needs to be
replaced, simply remove the four
screws on the rear cover of the
logger and replace the battery
with a new one paying attention
to the correct polarity. Replace the
cover and tighten the four screws.
When battery is replaced, both LEDs will turn on, and the Red
LED will then turn off followed by the Green LED. This indicates
that a proper reset of the logger has been done. Should this
not occur, please reinstall the battery.
For loggers with LCD, the LCD will also light up for about 1
second, showing all segments.
Note: Dispose of the Lithium (Li) battery according to local
regulations.