Instruction Manual
HI 141 Series
NTC Thermologgers
& HI 141001 Infrared Transmitter
These Instruments are in
www.hannainst.com
Compliance with the CE Directives
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for correct operation. Please read it carefully before using the thermologger. If
you need additional technical information, do not hesitate to
e-mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully. If any damage has occurred during shipment,
immediately notify your Dealer or the nearest Hanna Customer Service Center.
Each thermologger comes complete with one 3.6V Lithium AA
battery and magnetic key.
Note: Conserve all packing material until the instrument has
been observed to function correctly because any defective item must be returned to the Dealer in its original
packing.
The main features of the HI 141 thermologgers include:
• One or two channels, with internal or external sensor;
• 16000 samples (for 1 channel models) or
8000 samples/channel (for 2 channels models);
• Logging interval from 1 second to 24 hours;
• Storing of temperature at logging interval, or min or
max temperature between logging intervals;
• Logging delay start up to 199 hours, and magnetic start
through magnetic key provided with the thermologger;
• Programmable high and low alarms;
• Non-volatile storage of logging parameters and data in
EEPROM;
• BEPS (Battery Error Prevention System);
• Security Password and lot serial number;
• Optional LCD with status, temperature and information
display;
• IP67 waterproof casing.
The expected life is about 4 years at 1 minute measuring and
logging intervals.
All HI 141 thermologgers are factory calibrated.
WARRANTY
HI 141 loggers are warranted for one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. This
warranty is limited to repair or replacement free of charge. Damages due to accident,
misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty,
report the model number, date of purchase, serial number and the nature of the failure. If
the repair is not covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods
Authorization Number from the Customer Service department and then send it with
shipment costs prepaid. When shipping any instrument, make sure it is properly packaged
for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent
of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction and appearance
of its products without advance notice.
Recommendations for users
Before using these products, make sure that they are entirely suitable for the environment in which they
are used.
Operation of these instruments in residential areas could cause unacceptable interferences to radio and TV
equipments, requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instruments’ EMC
performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.
ISTR141R1 06/02
GENERAL DESCRIPTION
HI 141 series is a family of temperature data loggers with
either one or two channels, and an optional LCD. The user
interacts with the logger, setting data acquisition parameters
or downloading logged data, through a RS232 serial port on
a Windows® PC. The HI 141000 Windows® application software supports the communication between the logger and the
PC through the HI 141001 Infrared Transmitter.
Several models are available, as shown in the following table:
The optional LCD is divided into 3 areas: the status on the
top, the temperature readout in the middle, and useful
information at the bottom.
The information shown on the bottom is user-selectable, and
can include the current number of samples, the samples
beyond alarm limits, the min and max temperature, and the
high and low alarm thresholds.
FUNCTIONAL DESCRIPTION
1. Hook*
2. Green LED
3. LCD*
4. Photo LED
5. Internal sensor*
* depending on model
The status of the thermologger is indicated by the green and
red LEDs, and optional LCD:
Green LED will light up when magnetic switch is activated by
the key. When the key is removed from the top of the meter,
the green LED switches off.
6. Connection for external sensor(s)*
7. Infrared LED
8. Red LED
9. Internal magnetic switch: touch
the top of the meter with magnetic
key for activating magnetic start
SPECIFICATIONS
Model Sensor(s) Range
HI 141A (H) T1 internal
HI 141B (H) T1 external
HI 141C (H)* T1 internal
HI 141D (H) T1 external
HI 141E (H) T1 internal
T2 external
HI 141F (H) T1 external
HI 141G (H)* T1 internal
HI 141J (H) T1 external
* Warning: T1 temperature range is limited by LCD characteristics.
Do not use the logger out of this range.
Note: For models with two external sensors, T1 and T2 channels are
indicated on the probes (T1 is red marked, T2 is white).
Resolution 0.1ºC (-40.0 to 100.0ºC); 0.2ºC (temp. >100.0ºC)
0.1ºF (-40.0 to 190.0ºF) ; 0.3ºF (temp. >190.0ºF)
Accuracy
EXTERNAL DIAMETER 86.5 mm (3.4”) 159 mm (6.3”)
HEIGHT 35 mm (1.4”) 60 mm (23.6”)
WEIGHT 150 g (5.54 oz.) 200 g (7.39 oz.)
T2 external
T2 external
T2 external
±0.5ºC (-40.0 to 0.0 and 70.0 to 100.0 ºC)
±0.4ºC (0.0 to 70.0 ºC)
±1.0ºC (temperature>100.0ºC)
±1.0ºF (-40.0 to 32.0 and 158.0 to 212.0ºF)
±0.8ºF (32.0 to 158.0 ºF)
±2.0ºF (temperature>212.0ºF)
-40.0 to 80.0ºC
-
-40.0 to 175.0ºF
-40.0 to 125.0ºC
-
-40.0 to 257.0ºF
-20.0 to 70.0ºC*
-4.0 to 158.0ºF*
-40.0 to 125.0ºC
-40.0 to 257.0ºF
-40.0 to 80.0ºC
-40.0 to 175.0ºF
-40.0 to 125.0ºC
-40.0 to 257.0ºF
-40.0 to 125.0ºC
-40.0 to 125.0ºC
-40.0 to 257.0ºF
-20.0 to 70.0ºC*
-4.0 to 158.0ºF*
-40.0 to 125.0ºC
-40.0 to 257.0ºF
-40.0 to 125.0ºC
-40.0 to 125.0ºC
-40.0 to 257.0ºF
HI 141 HI 141001
START UP
In order to communicate with the
HI 141 thermologger, the
HI 141000 application software
must be installed on your PC. This
is done by inserting the first floppy
disk in the driver and running
Setup.exe.
Connect the HI 141001 Infrared
Transmitter to your computer
through an available RS232 port.
Place the HI 141 on the transmitter, taking care to align the ribs
on the logger with the slots on the
transmitter
Run HI 141000 software and re-
fer to the on-line help for any
further information about setting
the parameters of the logger, acquiring logged data, etc.
BATTERY REPLACEMENT
Battery replacement must only take place in a non hazardous
area using the appropriate battery type (3.6V Lithium AA
battery).
When the battery needs to be replaced, simply remove the four screws
on the rear cover of the logger and
replace the battery with a new one
paying attention to the correct polarity. Replace the cover and tighten
the four screws.
When battery is replaced, both LEDs will turn on, and the Red
LED will then turn off followed by the Green LED. This indicates
that a proper reset of the logger has been done. Should this not
occur, please reinstall the battery.
For loggers with LCD, the LCD will also light up for about 1
second, showing all segments.
Note: Dispose of the Lithium (Li) battery according to local
regulations.