SPECIFICATIONS
Model Range Resolution Accuracy
HI 140A(H) -30.0 to 70.0°C 0.5°C ±1.5oC
-22.0 to 158.0°F 0.5°F ±3°F
HI 140B(H) -10.0 to 30.0°C 0.2°C ±0.5oC
14 to 86°F 0.4°F ±1°F
HI 140C(H) -30.0 to 10.0°C 0.2°C ±0.5°C
-22 to 50°F 0.4°F ±1°F
HI 140D(H) 20.0 to 60.0°C 0.2°C ±0.5°C
68 to 140°F 0.4°F ±1°F
HI 140E(H) -30.0 to -10.0°C 0.1°C ±0.3°C
-22 to 14°F 0.2°F ±0.6°F
HI 140F(H) 20.0 to 40.0°C 0.1°C ±0.3°C
68 to 104°F 0.2°F ±0.6°F
HI 140G(H) -5.0 to 15.0°C 0.1°C ±0.3°C
23 to 59°F 0.2°F ±0.6°F
HI 140H(H) 10 to 120°C 1°C ±2°C
50 to 248°F 2°F ±4°F
Part numbers ending with “H” are provided with a molded eye (hook)
for easy mounting. Models HI 140H & HI 140 HH are provided with
an external probe with 1.5 m (5’) cable.
START UP
In order to communicate with the
HI 140 thermologger, the HI 92140
application software must be installed
on your PC.
Connect the HI 90140 infrared transmitter to your computer through an
RS232 port.
Place the HI 140 on the transmitter
as shown in the picture.
Run the HI 92140 software and
refer to the on-line help for any
further information about setting the
parameters of the logger, acquire
logged data, etc.
HI 140 HI 90140
EXTERNAL DIAMETER 86.5 mm (3.4”) 159 mm (6.3”)
HEIGHT 35 mm (1.4”) 47.5 mm (1.9”)
WEIGHT 150 g (5.54 oz.) 150 g (5.54 oz.)
BATTERY REPLACEMENNT
Battery replacement must only take place in a non hazardous area using the appropriate battery type (1.5V AA).
When the batteries need to be
replaced, simply remove the screws
on the rear cover of the logger
and replace the three batteries
with new ones, while paying attention to the correct polarity.
After replacing the batteries, press the reset button on the
printed circuit board. The red LED blinks and then the
green LED stays on.
Reattach the cover and tighten the screws.
Note: Each time the batteries are replaced, all logger
parameters have to be set again by the user as soon
as possible to avoid draining the batteries.
WARRANTY
HI 140 loggers are warranted for one year against defects in workmanship and materials
when used for their intended purpose and maintained according to instructions. This warranty
is limited to repair or replacement free of charge. Damages due to accident, misuse,
tampering or lack of prescribed maintenance are not covered. If service is required, contact
the dealer from whom you purchased the instrument. If under warranty, report the model
number, date of purchase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number
from the Customer Service department and then send it with shipment costs prepaid. When
shipping any instrument, make sure it is properly packaged for complete protection.
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.
Recommendations for users
Before using these products, make sure
that they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential areas could cause unacceptable
interferences to radio and TV equipments,
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to
the supplied equipment may degrade the
instruments’ EMC performance.
To avoid damages or burns, do not perform any measurement in microwave
ovens.