Hanna Instruments HI 120 User Manual

Instruction Manual
ON
Calibration Check
LOG
CLR
CFM
CAL
pH / mV Meter
HI 122
HI 120 & HI 122
pH/mV/Temperature
Bench Meters
with Calibration Check
www.hannainst.com
1
Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instruments. This manual will provide you with the necessary information for correct use
of the instruments, as well as a precise idea of their versatility. If you need additional technical information, do not hesitate to e-mail us at
tech@hannainst.com or turn to the back cover for our worldwide contact list. These instruments are in compliance with directives.
WARRANTY
HI 120 and HI 122 are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Electrodes and probes are guaranteed for six months. This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packed for complete protection.
TABLE OF CONTENTS
WARRANTY ..................................................................................................... 2
PRELIMINARY EXAMINATION .............................................................................. 3
GENERAL DESCRIPTION .................................................................................... 3
FUNCTIONAL DESCRIPTION HI 120 .................................................................... 4
FUNCTIONAL DESCRIPTION HI 122 .................................................................... 5
HI 120 AND HI 122 SPECIFICATIONS ........................................................... 10
OPERATIONAL GUIDE ...................................................................................... 11
pH CALIBRATION ............................................................................................ 15
ENHANCED CALIBRATION MESSAGES ................................................................ 20
ELECTRODE CONDITION & ELECTRODE RESPONSE TIME ...................................... 21
RELATIVE mV CALIBRATION ............................................................................ 22
GOOD LABORATORY PRACTICE (GLP) ................................................................ 23
SETUP .......................................................................................................... 27
LOGGING ...................................................................................................... 31
AutoHOLD ..................................................................................................... 35
PRINTING (HI 122 only) ................................................................................ 36
TEMPERATURE CALIBRATION (for technical personnel only) .................................. 40
mV CALIBRATION (for technical personnel only) .................................................. 41
PC INTERFACE ............................................................................................... 42
pH BUFFER TEMPERATURE DEPENDENCE ......................................................... 44
ELECTRODE CONDITIONING & MAINTENANCE ..................................................... 45
PRINTER MAINTENANCE (HI 122 only) ............................................................ 48
TROUBLESHOOTING GUIDE ............................................................................. 49
TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS ..................................... 51
ACCESSORIES ................................................................................................ 52
2
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer or the nearest Hanna Service Center.
Each instrument is supplied with:
HI 1131P Glass-body Combination pH Electrode
HI 7662-T Temperature probe
pH 4.01 & 7.01 Buffer solutions, 20 mL each
HI 7071S Electrolyte solution
• 5 paper rolls (HI 122 only)
• 12VDC Power Adapter
• Instruction Manual
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the original packing with the supplied accessories.
GENERAL DESCRIPTION
The Hanna HI 120 and HI 122 are professional bench meters for pH, ORP (Oxidation Reduction Potential) and Temperature measurements with Calibration Check. Relative mV feature is also provided.
Calibration Check performs a set of diagnostic tests during calibration using the history of electrode slope and offset to detect problems that can cause loss of accuracy.
Calibration Check Features are:
• Enhanced Calibration Messages
• Electrode Condition on LCD Display
• Electrode Response Time on LCD Display
• Calibration Alarm Time-Out
• Out of Calibration Range
Other features include:
• Up to five-point calibration with seven memorized buffers (1.68, 4.01,
6.86, 7.01, 9.18, 10.01 and 12.45 pH) and two custom buffers.
• pH calibration using pH buffers with 0.001 resolution.
• pH reading with manual or automatic temperature compensation.
• Up to 50 samples for Log on demand mode and up to 1000 samples for AutoLOG mode.
• Two selectable alarm limits.
• User selectable AutoLOG modes.
• Printing feature in four selectable languages (HI 122 only).
• Large easy-to-read LCD which shows the pH or mV and Temperature simultaneously, together with graphic symbols.
• AutoHOLD feature to freeze first stable reading on the LCD.
• GLP feature to view last calibration data for pH and Relative mV.
• PC interface.
3
FUNCTIONAL DESCRIPTION
NUMReso
RelmVmVpH
AutoLOG
SETGLPAutoHOLD
ON OFF
Front Panel
Rear Panel
1
2
3
4
Primary LCD
Secondary LCD
Electrode Condition
5
6
7
8
9
Electrode Response
RS232POWERTEMPpH/ORPPIN
Calibration Check
RCL
LOG
MODE
CLR
CFMCAL
pH / mV Meter
HI 120
HI 120
1) Left Keyboard
2) ON switch
3) OFF switch
4) Right Keyboard
5) Pin input socket
6) BNC electrode connector
7) Temperature probe socket
8) Power adapter socket
9) RS232 serial communication connector
4
FUNCTIONAL DESCRIPTION
ON OFF
NUMReso
RelmVmVpH
AutoLOG
PaperPrintAutoHOLD
1
2
3
4
Front Panel
Primary LCD
Secondary LCD
Electrode Condition
Electrode Response
Rear Panel
567
8
9
10
Paper Roll
RS232POWERTEMPpH/ORPPIN
Calibration Check
LOG
CLR
CFMCAL
pH / mV Meter
HI 122
HI 122
1) Left Keyboard
2) ON switch
3) OFF switch
4) Right Keyboard
5) Pin input socket
6) BNC electrode connector
7) Temperature probe socket
8) Power adapter socket
9) RS232 serial communication connector
10) Printer
5
HI 120
1 2
3
4
5
10 11 12
7
6
9
8
NUMResolution
RelmVmVpH
AutoLOG
SETGLPAutoHOLD
KEYBOARD ON THE LEFT
Shortcuts to alternate functions
1) AutoHOLD key, to freeze the first stable reading on the LCD.
2) GLP key, to display Good Laboratory Practice Information.
3) SET key, to enter/exit SETUP mode.
4) key, to toggle between parameters while in RECALL or SETUP
modifying mode (backwards), to toggle between absolute mV and temperature while in Relative mV mode and between pH buffer and temperature while in pH calibration mode.
5) AutoLOG key, to start/stop AutoLOG mode.
6) key, to toggle between parameters while in RECALL or SETUP
modifying mode (forwards), to toggle between absolute mV and temperature while in Relative mV mode and between pH buffer and temperature while in pH calibration mode.
7) pH key, to select pH range.
8) mV key, to select mV range.
9) RelmV key, to select Relative mV range.
11) Resolution key, to select pH resolution.
Numerical keys
10) key, to change sign.
12) NUM key, to activate the numerical keys (0 to 9).
6
HI 120
1
9
8
7
6
5
4
3
2
Calibration Check
RCL
LOG
MODE
CLR
CFMCAL
pH / mV Meter
HI 120
MAIN KEYBOARD ON THE RIGHT
1) CAL key, to enter and exit/escape calibration mode.
SET key (second function), to enter/exit SETUP mode.
2) CFM key, to confirm different values.
GLP key (second function), to display Good Laboratory Practice Information.
3) CLR key, to clear calibration or logged data.
4) key, to manually increase temperature value or other parameters.
AutoHOLD key (second function), to freeze the first stable reading on the LCD.
5) key, to manually decrease temperature value or other parameters.
AutoLOG key (second function), to start/stop AutoLOG mode.
6) MODE key, to select the measurement unit or to switch focused data.
7) 2nd key, to select second key function.
8) LOG key, to store measured data.
9) RCL key, to enter/exit view logged data mode.
Note: To select second key function, press 2nd and then the desired key.
The “2nd” tag will appear on the LCD until the desired key is pressed. To leave second key function selection, press 2nd again.
7
HI 122
1 2
3
4
5
10 11 12
7
6
9
8
NUMResolution
RelmVmVpH
AutoLOG
PaperPrintAutoHOLD
KEYBOARD ON THE LEFT
Shortcuts to alternate functions
1) AutoHOLD key, to freeze the first stable reading on the LCD.
2) Print key, to obtain a printout or to cancel printing.
3) Paper key, to pull out the paper.
4) key, to toggle between parameters while in RECALL or SETUP
modifying mode (backwards), to toggle between absolute mV and temperature while in Relative mV mode and between pH buffer and temperature while in pH calibration mode.
5) AutoLOG key, to start/stop AutoLOG mode.
6)
key, to toggle between parameters while in RECALL or SETUP modifying mode (forwards), to toggle between absolute mV and temperature while in Relative mV mode and between pH buffer and temperature while in pH calibration mode.
7) pH key, to select pH range.
8) mV key, to select mV range.
9) RelmV key, to select Relative mV range.
11) Resolution key, to select pH resolution.
Numerical keys
10) key, to change sign.
12) NUM key, to activate the numerical keys (0 to 9).
8
HI 122
9
8
7
6
5
4
3
2
1
Calibration Check
LOG
CLR
CFMCAL
pH / mV Meter
HI 122
MAIN KEYBOARD ON THE RIGHT
1) CAL key, to enter and exit/escape calibration mode. SET key (second function), to enter/exit SETUP mode.
2) CFM key, to confirm different values. GLP key (second function), to display Good Laboratory Practice Information.
3) CLR key, to clear calibration or logged data.
4) key, to manually increase temperature value or other parameters. AutoHOLD key (second function), to freeze the first stable reading on
the LCD.
5) key, to manually decrease temperature value or other parameters. AutoLOG key (second function), to start/stop AutoLOG mode.
6) MODE key, to select the measurement unit or to switch focused data. Paper key (second function), to pull out the paper.
7) 2nd key, to select second key function.
8) LOG key, to store measured data.
9) RCL key, to enter/exit view logged data mode.
Print key (second function), to obtain a printout or to cancel printing.
Note: To select second key function, press 2nd and then the desired key.
The “2nd” tag will appear on the LCD until the desired key is pressed. To leave second key function selection, press 2nd again.
9
EGNAR
Hp00.61ot00.2
Hp000.61ot000.2
Vm9.999± Vm0002±
)Fº0.842ot0.4–(Cº0.021ot0.02–
NOITULOSER
Hp10.0
Hp100.0
Vm1.0
Vm1
)Fº1.0(Cº1.0
YCARUCCA
F°86/C°02@
Hp10.0±
Hp200.0±
)Vm9.996±(Vm2.0± )Vm9.999±(Vm5.0±
)Vm0002±(Vm1±
)Fº7.0±(Cº4.0±
)rorreeborpgnidulcxe(
egnartesffoVmevitaleR Vm0002±
noitarbilaCHp
,noitarbilactniop-evifotpU elbaliavasreffubdradnats7
0.7,68.6,10.4,86.1( ,)54.21,10.01,81.9,1 sreffubmotsuc2dna
noitasnepmocerutarepmeT
:morfcitamotuArolaunaM
)Fº0.842ot0.4–(Cº0.021ot0.02–
edortcelEHp P1311IH
eborperutarepmeT 2667IH
lavretnIgniggoL setunim081otsdnoces5
)ylno221IH(retnirP repapediwmm44,xirtamtoD
ecafretniCP 232SRdetalosi-otpo
ecnadepmitupnI 01
21
mho
ylppusrewoP retpadaCDV21
snoisnemiD )”3.3x0.8x0.11(mm48x302x082
thgieW )bl2.4(gK9.1:retnirphtiwtik;)bl1.4(gK8.1
tnemnorivnE
)Fº221–23(Cº050
gnisnednoc-nonHR%59.xam
ytnarraW sraey2
HI 120 AND HI 122
SPECIFICATIONS
10
OPERATIONAL GUIDE
POWER CONNECTION
Plug the 12 VDC adapter into the power supply socket. Notes: • These instruments use non volatile memory to retain the pH,
mV, temperature calibrations and all other settings, even when unplugged.
• Make sure a fuse protects the main line.
ELECTRODE AND PROBE CONNECTIONS
For HANNA P type pH or ORP electrodes (with internal reference) connect the electrode’s BNC to the socket on the back of the instrument (#6 on page 4 and 5) and the pin to the appropiate socket (#5 on page 4 and 5).
Notes: • Electrode condition and response information is displayed on the
bar graph gauges during the day the calibration is performed only if HANNA P type (PIN) electrodes are used.
• If the electrode is not recognized as a HANNA P type electrode, the bar graph gauges will blink (25 seconds OFF, 4 seconds ON, full bar graph).
For temperature measurements and automatic temperature compensation, connect the temperature probe to the appropriate socket.
INSTRUMENT START-UP
• Turn the instrument on by pressing the ON switch.
• All LCD tags are displayed and a beep is heard while the instrument performs a self test.
• The instrument displays then the date on the primary LCD and the time on the secondary LCD, along with the “Remove protective cap” and “Unscrew electrode refilling cap” messages alternatively blinking. These messages alert the user to follow displayed instructions in order to take proper measurements and to improve electrode response.
• The instrument automatically defaults to pH or mV measurement mode, if a HANNA P type pH or ORP electrode is detected.
• If no HANNA P type electrode is detected, the instrument starts in the same range as it was at power off.
11
pH MEASUREMENTS
Make sure the instrument has been calibrated before taking pH measurements.
• Press MODE to enter pH mode.
Note: To change pH resolution, press MODE again or
simply Resolution from the left keyboard.
• Submerge the electrode tip and the temperature probe approximately 4 cm (1½”) into the sample to be tested. Allow time for the electrode to stabilize.
• The pH is displayed on the primary LCD and the temperature on the secondary LCD.
OUT CAL RANGE feature warns the user if the current reading is out of the calibrated area. The calibrated area is that part of the pH range in which the calibration point assures an accurate reading. If the reading is taken out of the calibration area, the “OUT CAL RANGE” message will blink. The calibrated area is calculated in according with the pH resolution used during the reading. To avoid having this message, the calibration points have to be well distributed in the desired measurement range.
If measurements are taken successively in different samples, it is recommended to rinse the electrode thoroughly with deionized water or tap water and then with some of the next sample to prevent cross-contamination and to condition the electrode before immersing it into the sample solution.
The pH reading is affected by temperature. In order to measure the pH accurately, the temperature effect must be compensated for. To use the Automatic Temperature Compensation feature, connect and submerge the HI 7662-T temperature probe into the sample as close as possible to the electrode and wait for a few seconds.
If the temperature of the sample is known, manual temperature compensation can be performed by disconnecting the temperature probe.
The display will then show the default temperature of 25 ºC (77 ºF) or the last temperature reading with the “ºC” (or “ºF”) tag blinking. The temperature can be adjusted with the ARROW keys or the numerical keypad (from –20.0 ºC to 120.0 ºC or from –4.0 ºF to 248.0 ºF).
12
Press NUM to change the temperature value with the numerical keys. The “2nd” tag will blink. Press CLR if you want to delete digits of the displayed value. The remaining digits will shift to right. Introduce the desired value. If the value is out of temperature range, a long beep will be heard. Press NUM to confirm the new value or CAL to escape without changing the temperature.
mV/ORP MEASUREMENTS
Oxidation-reduction potential (REDOX) measurements provide the quantification of the oxidizing or reducing power of the tested sample.
To correctly perform a redox measurement, the surface of the ORP electrode must be clean and smooth.
• Press MODE or simply mV from the left keyboard to enter mV range.
• Submerge the tip of the ORP electrode (4 cm/1½”) into the sample to be tested and allow a few seconds for the reading to stabilize.
• The instruments display the mV reading on the primary LCD and the temperature on the secondary LCD.
• If the reading is out of range, the closest full-scale value will be displayed blinking on the primary LCD.
RELATIVE mV MEASUREMENTS
To enter Relative mV mode, press MODE or simply RelmV from the left keyboard. The relative mV reading will be displayed on the primary LCD and the current temperature value on the secondary LCD.
Note: Press or from the left keypad to toggle between temperature
and absolute mV reading on the secondary LCD.
13
The Relative mV reading is equal to the difference between the absolute mV input value and relative mV offset established in the relative mV calibration.
TEMPERATURE MEASUREMENTS
Connect the HI 7662-T temperature probe to the appropri- ate socket. Immerse the temperature probe into the sample and allow the reading on the secondary LCD to stabilize.
Note: The temperature can be displayed in Celsius degrees (ºC) or in
Fahrenheit degrees (ºF) (see SETUP for details, page 27).
14
pH CALIBRATION
Calibrate the instrument often, especially if high accuracy is required. The instrument should be recalibrated:
• Whenever the pH electrode is replaced.
• At least once a week.
• After testing aggressive chemicals.
• If “CAL DUE” tags are blinking during measurement.
• If “OUT CAL RANGE” message blinks during pH measurement (the measurement range is not covered by current calibration).
PREPARATION
Pour small quantities of the buffer solutions into clean beakers. If possible, use plastic or glass beakers to minimize any EMC interferences.
For accurate calibration and to minimize cross-contamination, use two beakers for each buffer solution. One for rinsing the electrode and one for calibration.
If you are measuring in the acidic range, use pH 7.01 or 6.86 as first buffer and pH 4.01 or 1.68 as second buffer. If you are measuring in the alkaline range, use pH 7.01 or 6.86 as first buffer and pH 10.01/9.18 or 12.45 as second buffer.
For extended range measurements (acidic and alkaline), perform a five-point calibration by selecting five of the available buffers.
PROCEDURE
If 0.001 pH resolution is selected, each selected standard buffer value can be updated in according with the value on the production lot certificate at 25 ºC (77 ºF). Press 2nd then SET key when a standard pH buffer with
0.001 resolution is selected. The buffer value will start blinking and it can
be changed with the ARROW keys in a ±0.020 pH window. Calibration has a choice of 7 memorized buffers: pH 1.68, 4.01, 6.86,
7.01, 9.18, 10.01, 12.45 and 2 custom buffers.
The custom buffers are a special option that allows the user to calibrate in a buffer solution different from a standard one. Up to two custom buffers can be set in SETUP menu (see page 27). When selected during calibration, the “CUSTOM C1” or “CUSTOM C2” tags are displayed on the LCD and the custom buffer value can be changed in a ±1.0 pH window, around the set value.
For accurate measurements it is recommended to perform a five-point calibration. However, at least a two-point calibration is suggested.
The instruments will automatically skip the buffers used during calibration and the buffers which are in a ±0.2 pH window, around one of the calibrated buffers.
15
All new calibrations will override existing stored calibration data, in a ±0.2
CAL
pH window, at these calibration points. The slopes adjacent to the calibration points will be reevaluated.
If the new calibration point has no correspondence in the existing stored calibration data, it is added to it if this is not full, or the instrument will ask which buffer will be replaced by the current buffer.
If at least a two-point calibration has been performed and an offset correction of the electrode is desired, keeping unchanged the existing slopes, perform a one-point calibration with “OFFS” option selected in SETUP menu. If “Pnt” option is selected, the slopes adjacent to the calibration points will be reevaluated.
FIVE-POINT CALIBRATION
• Immerse the pH electrode and the temperature probe approximately 4 cm (1½”) into a buffer solution of your choice (pH 1.68, 4.01, 6.86, 7.01,
9.18, 10.01, 12.45, custom buffer 1 or 2, if these were set) and stir gently. The temperature probe should be close to the pH electrode.
• Press CAL. The “CLEAR CAL IF NEW ELECTRODE” message will be displayed blinking on the LCD for a few seconds if the instrument was calibrated before and calibration was not cleared.
Press CLR if you are using a new electrode or want to clear calibration history, or wait a few seconds to continue.
Press CAL again or the ARROW keys to skip this message. Note: It is very important to clear calibration history when a new
electrode is used because most errors and warning messages that appear during calibration depend on calibration history.
• The instruments will display the measured pH on the primary LCD and the “7.01” buffer on the secondary LCD, together with “CAL” and “Cal Point 1” tags and “7.01” tag blinking.
• If necessary, press the ARROW keys to select a different buffer value.
16
• The “ ” tag will blink on the LCD until the reading is stable.
CFM
CFM
CFM
• When the reading is stable and close to the selected buffer, “CFM” tag blinks.
• Press CFM to confirm calibration.
• The calibrated value is then displayed on the primary LCD and the secondary LCD will display the second expected buffer value, together with “CAL”, “Cal Point 2” and “7.01” tags and the corresponding buffer tag blinking.
• After the first calibration point is confirmed, immerse the pH electrode and the temperature probe approximately 4 cm (1½”) into the second buffer solution and stir gently. The temperature probe should be close to the pH electrode.
• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, “CFM” tag blinks.
• Press CFM to confirm calibration.
• The calibrated value is then displayed on the primary LCD and the secondary LCD will display the third expected buffer value.
• After the second calibration point is confirmed, immerse the pH electrode and the temperature probe approximately 4 cm (1½”) into the third buffer solution and stir gently. The temperature probe should be close to the pH electrode.
• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, “CFM” tag blinks.
• Press CFM to confirm calibration.
• The calibrated value is then displayed on the primary LCD and the secondary LCD will display the fourth expected buffer value.
17
Loading...
+ 39 hidden pages