Instruction Manual
BL 932700-0
BL 932700-1
Panel-Mounted
ORP Indicators &
Controllers
WARRANTY
These instruments are warranted for two years against defects in workmanship and materials when used for their intended purpose and
maintained according to instructions. Probes are warranted for six months.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from
whom you purchased the instrument. If under warranty, report the model
number, date of purchase, serial number and the nature of the failure. If
the repair is not covered by the warranty, you will be notified of the
charges incurred. If the instrument is to be returned to Hanna Instruments,
first obtain a Returned Goods Authorization Number from the Customer
Service department and then send it with shipment costs prepaid. When
shipping any instrument, make sure it is properly packaged for complete
protection.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment
in which they are used. Operation of these instruments in residential areas could cause
unacceptable interferences to radio and TV equipment. Avoid touching the electrode sensor at
all times. During operation, ESD wrist straps should be worn to avoid possible damage to
the electrode by electrostatic discharges. Any variation introduced by the user to the supplied
equipment may degrade the instrument’s EMC performance. To avoid electrical shock, do
not use these instruments when voltages at the measurement surface exceed 24 Vac or 60
Vdc. To avoid damages or burns, do not perform any measurement in microwave ovens.
Dear Customer,
Thank you for choosing a Hanna product.
This manual will provide you with the necessary information
for the correct operation of the meter. Please read it carefully
before using the instrument.
If you need additional technical information, do not hesitate
to e-mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
Each meter is supplied with:
• Mounting brackets
• Instruction manual
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
GENERAL DESCRIPTION
BL932700-0 and BL932700-1 are panel-mounted ORP
indicators and controllers designed for simplicity of use in a
wide range of industrial applications.
Connections and wiring to electrode, power supply, contacts
and external recorders are made via the terminal blocks on
the rear panel. The meters are equipped with a BNC socket
and accept input from conventional ORP electrodes.
Other features include: 4-20 mA output, overtime control
system, selection of dosing direction (Rdx/Oxd), contact for
external control and disabling of dosing action, multi-colour
LED for indicating if the meter is in measurement/dosing/
alarm condition, possibility to set (Off-Auto-On switch) dosing
action mode.
Two models are available:
• BL 932700-0 powered at 12 Vdc
• BL 932700-1 powered at 115 or 230 Vac
SPECIFICATIONS
Range ±1000 mV
Resolution 1 mV
Accuracy (@ 20°C/68°F) ±5 mV
Typical EMC Deviation ±5 mV / ±0.10 mA
Calibration Manual, through CAL trimmer
Dosing Contact Maximum 2A (fuse protected),
250 Vac, 30 Vdc
Dosing Selection Rdx or Oxd, selectable on the back panel
Contact Open=Reductant dosage=Relay ON if Measure > Setpoint
Contact Close=Oxidant dosage=Relay ON if Measure < Setpoint
Setpoint Adjustable, from -1000 to 1000 mV
Overtime Adjustable, typically from 5 to approx. 30 minutes
Recorder Output 4 to 20 mA, accuracy ±0.20 mA,
500 Ω maximum load
Power Consumption 10 VA
Installation Category II
Power supply: External (fuse protected)
BL932700-0 12 Vdc
BL932700-1 115/230 Vac ; 50/60Hz
Dimensions 83 x 53 x 99 mm (3.3x2.1x3.9”)
ASSEMBLING VIEW
ACCESSORIES
ORP SOLUTIONS
HI 7020M/L ORP test solution (200-275 mV), 230/500 mL
HI 7021M/L ORP test solution (240 mV), 230/500 mL
HI 7022M/L ORP test solution (470 mV), 230/500 mL
HI 7091M/L Pretreatment Reducing Solution, 230/500 mL
HI 7092M/L Pretreatment Oxidizing Solution, 230/500 mL
OTHER SOLUTIONS
HI 70300M/L Electrode storage solution, 230/500 mL bottle
HI 7061M/L Electrode cleaning solution for general pur-
poses, 230/500 mL bottle
ORP ELECTRODES
HI 3214P/2 BNC-type, platinum, double-junction, plastic-
body with 2m (6.6’) cable
HI 2003/5 Heavy-duty, BNC-type, platinum, double junc-
tion, plastic-body with 5m (16.5’) cable
HI 2012/5 Heavy-duty, BNC-type, platinum, double junc-
tion, plastic-body with 5m (16.5’) cable
EXTENSION CABLES FOR SCREW-TYPE ELECTRODES,
SCREW TO BNC CONNECTOR
HI 7855/5 Extension cable 5 m (16.5') long
HI 7855/10 Extension cable 10 m (33') long
OTHER ACCESSORIES
BL PUMPS Dosing Pumps with flow rate from 1.5 to
20LPH
HI 6050 Submersible electrode holder, 60 cm (24”)
HI 6051 Submersible electrode holder, 110 cm (43”)
HI 6054B Electrode holder for in-line applications
HI 6054T Electrode holder for in-line applications
HI 710005 12 Vdc power adapter, US plug
HI 710006 12 Vdc power adapter, European plug
HI 710012 12 Vdc power adapter, Australian plug
HI 710013 12 Vdc power adapter, South-African plug
HI 710014 12 Vdc power adapter, UK plug
HI 731326 Calibration screwdriver (20 pcs)
HI 740146 Mounting brackets
HI 7871 Level Controller (min and max)
HI 7873 Level Controller (min, max and overflow)
FUNCTIONAL DESCRIPTION
Front panel
1. Liquid Crystal Display
2. Switch for selecting dosing mode:
• OFF = dosing disabled
• Auto = automatic dosage, depending on setpoint value
and dosing selection
• ON = dosing always active
3. “MEAS” key to set the instrument to measurement mode
4. “SET” key to display and set the setpoint value
5. “SET” trimmer to adjust the setpoint value (within ±1000mV)
6. “CAL” trimmer
7. 3-colour LED indicator:
• Green = meter in measurement mode
• Orange/Yellow = dosing in progress
• Red, blinking = indicates an alarm condition
Rear panel
1. BNC plug for ORP electrode
2. Connection for electrode reference
3. Connection for potential Matching Pin
4. Rdx/Oxd dosage selection terminal:
• contact open = reductant selection
• contact closed = oxidant selection
5. 4-20 mA output terminal for recorder connection
6. Power supply terminal:
• for BL932700-0 model: 12 Vdc adapter
• for BL932700-1 model: 115 Vac or 230 Vac option
7. This contact acts as a switch for driving the dosing system (e.g.
dosing pump)
8. External control and disabling of dosing system
9. Jumper for enabling (jumper in) or disabling (jumper
removed) the overtime control
10. Trimmer for overtime setting (typically from 5 to 30 minutes)
All external cables connected to the rear panel
should end with cable lugs.
A circuit breaker (rated 6A max.) must be
connected in close proximity to the equipment, and in a position easy to reach by the
operator, for disconnection of the instrument
and of all the devices connected to the relays.
OPERATIONS
REAR PANEL CONNECTIONS
Terminals #1, #2 and #3: Electrode
• Connect an ORP electrode to the meter BNC plug (#1).
• To benefit from the differential input, connect the proper
electrode wire (if available) or a cable with a potential
matching pin (grounding bar) to the relevant terminal
(#3) on the rear panel.
Note: When the Matching Pin can not be immersed together
with the electrode in the solution, disable the differential
input by shorting terminals #3 (Matching Pin) and
#2 (Electrode Reference) with a jumper wire.
Terminals #4: Dosing selection
• For Rdx dosage, leave the circuit open.
• For Oxd dosage, short the terminals with a jumper wire.
Terminals #5: 4-20 mA Output
• These output terminals are used for connecting a recorder.
The output is from 4 to 20 mA and is proportional to the
measured ORP value.
Terminals #6: Power Supply
• Model BL932700-0: connect the 2 wires of a 12 Vdc
power adapter to the terminals +12 Vdc and GND.
• Model BL932700-1: connect a 3-wire power cable to the
terminals while paying attention to the correct earth (PE),
line (L) and neutral (N1 for 115 V or N2 for 230 V) contacts.
Terminals #7: Dosing Contact
• This contact drives the dosing system, accordingly to the
selected setpoint and dosing direction:
• if “Rdx” dosage is set, the relay is ON and dosing
activated if measured value is higher than setpoint;
• if “Oxd” dosage is set, the relay is ON and dosing
activated if measured value is lower than setpoint.
Note: The setpoint has a typical hysteresis value comparable
to the meter accuracy.
Terminals #8: External Disabling Contact
• This is a normally open contact and can be used for
example for connecting a level controller.
• When the contact is closed, any dosing action stops, the
LED indicator on the front panel will blink Red and the
LCD will show the “HALT” warning message.
Note: If the OFF/Auto/ON switch is in ON position, the
dosing action will not stop, even if the external
disablig contact is closed. Dosage will proceed, the LED
indicator will lit Orange/Yellow and the LCD will show
the “HALT” warning message.
Overtime system: jumper (#9) and trimmer (#10)
• This system allows the user to set a maximum dosing
period, by adjusting the rear trimmer from 5 (min) to
approx. 30 (max) minutes.
• When the set time is exceeded, any dosing action stops,
the LED indicator on the front panel will blink Red and the
LCD will show the “TIMEOUT” warning message. To exit
the overtime condition, set the OFF/Auto/ON switch to
“OFF” position, and then to “Auto” again.
• For disabling the overtime feature, simply remove the
jumper on the rear panel.
Note: The overtime system works only if the OFF/Auto/ON
switch is in “Auto” position.
OPERATING THE METER
Before proceeding make sure that:
• the setpoint value has been properly adjusted;
• all rear panel wiring and selections are correct;
• the Auto/OFF/ON switch is in the desired position.
Install or immerse the electrode in the solution to be monitored, then press the “MEAS” key (if necessary).
The LCD will show the ORP (mV) value. The LED indicator will
light up Green when the meter is in measurement mode and
dosing is not active, while will light up Orange/Yellow for
signaling that a dosing action is in progress.
CALIBRATION
This meter is factory calibrated. Anyway, it is possible to check
the calibration as follows:
• ensure the meter is in measurement mode;
• immerse electrode and Matching Pin (if used) in one of the
available ORP test solutions (see “Accessories”);
• shake briefly and wait for reading to stabilize;
• if necessary, adjust reading through the “CAL” trimmer.
SETPOINT
Press the “SET” key: the display will show the default or
previously adjusted value, together with the “SET” indication.
Using a small screwdriver adjust the “SET” trimmer until the
desired setpoint value is displayed.
After 1 minute the meter automatically returns to the normal
mode; or press the “MEAS” key.