HAMEG HM1008 User Manual

100 MHz CombiScope
HM1008
Manual
English
®
Hersteller HAMEG Instruments GmbH KONFORMITÄTSERKLÄRUNG Manufacturer Industriestraße 6 DECLARATION OF CONFORMITY Fabricant D-63533 Mainhausen DECLARATION DE CONFORMITE
Die HAMEG Instruments GmbH bescheinigt die Konformität für das Produkt The HAMEG Instruments GmbH herewith declares conformity of the product HAMEG Instruments GmbH déclare la conformite du produit
Bezeichnung / Product name / Designation: Oszilloskop Oscilloscope Oscilloscope
Typ / Type / Type: HM1008
mit / with / avec:
Optionen / Options / Options:
mit den folgenden Bestimmungen / with applicable regulations / avec les directives suivantes
EMV Richtlinie 89/336/EWG ergänzt durch 91/263/EWG, 92/31/EWG EMC Directive 89/336/EEC amended by 91/263/EWG, 92/31/EEC Directive EMC 89/336/CEE amendée par 91/263/EWG, 92/31/CEE
Niederspannungsrichtlinie 73/23/EWG ergänzt durch 93/68/EWG Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC Directive des equipements basse tension 73/23/CEE amendée par 93/68/CEE
Angewendete harmonisierte Normen / Harmonized standards applied / Normes harmonisées utilisées:
General information regarding the CE marking
The measuring- and data lines of the instrument have much infl uence on emmission and immunity and therefore on meeting the acceptance limits. For different applications the lines and/or cables used may be different. For measurement operation the following hints and conditions regarding emission and immunity should be observed:
1. Data cables
For the connection between instruments resp. their interfaces and external devices, (computer, printer etc.) suffi ciently screened cables must be used. Without a special instruction in the manual for a reduced cable length, the maximum cable length of a dataline must be less than 3 meters and not be used outside buildings. If an interface has several connectors only one connector must have a connection to a cable.
Basically interconnections must have a double screening. For IEEE-bus purposes the double screened cables HZ72S and HZ72L from HAMEG are suitable.
2. Signal cables
Basically test leads for signal interconnection between test point and instrument should be as short as possible. Without instruction in the manual for a shorter length, signal lines must be less than 3 meters and not be used outside buildings.
Sicherheit / Safety / Sécurité: EN 61010-1:2001 (IEC 61010-1:2001) Überspannungskategorie / Overvoltage category / Catégorie de surtension: II Verschmutzungsgrad / Degree of pollution / Degré de pollution: 2
Elektromagnetische Verträglichkeit / Electromagnetic compatibility / Compatibilité électromagnétique
EN 61326-1/A1 Störaussendung / Radiation / Emission: Tabelle / table / tableau 4; Klasse / Class / Classe B.
Störfestigkeit / Immunity / Imunitée: Tabelle / table / tableau A1.
EN 61000-3-2/A14 Oberschwingungsströme / Harmonic current emissions / Émissions de courant harmonique: Klasse / Class / Classe D.
EN 61000-3-3 Spannungsschwankungen u. Flicker / Voltage fl uctuations and fl icker / Fluctuations de tension et du fl icker.
Datum /Date /Date
24. 02. 2005 Unterschrift / Signature / Signatur
Manuel Roth Manager
This will not cause damage or put the instrument out of operation. Small deviations of the measuring value (reading) exceeding the instruments specifi cations may result from such conditions in individual cases.
4. RF immunity of oscilloscopes.
4.1 Electromagnetic RF fi eld
The infl uence of electric and magnetic RF fi elds may become visible (e.g. RF superimposed), if the fi eld intensity is high. In most cases the coupling into the oscilloscope takes place via the device under test, mains/line supply, test leads, control cables and/or radiation. The device under test as well as the oscilloscope may be effected by such fi elds.
Although the interior of the oscilloscope is screened by the cabinet, direct radiation can occur via the CRT gap. As the bandwidth of each amplifi er stage is higher than the total –3dB bandwidth of the oscilloscope, the infl uence RF fi elds of even higher frequencies may be noticeable.
4.2 Electrical fast transients / electrostatic discharge Electrical fast transient signals (burst) may be coupled into the oscilloscope directly via the mains/line supply, or indirectly via test leads and/or control cables. Due to the high trigger and input sensitivity of the oscilloscopes, such normally high signals may effect the trigger unit and/or may become visible on the CRT, which is unavoidable. These effects can also be caused by direct or indirect electrostatic discharge.
HAMEG Instruments GmbH
Signal lines must screened (coaxial cable - RG58/U). A proper ground connection is required. In combination with signal generators double screened cables (RG223/U, RG214/U) must be used.
3. Infl uence on measuring instruments
Under the presence of strong high frequency electric or magnetic fi elds, even with careful setup of the measuring equipment an infl uence of such signals is unavoidable.
2
Subject to change without notice
Contents
General information regarding the CE marking 2
100 MHz CombiScope HM1008 4
Specifi cations 5
Important hints 6
List of symbols used: 6 Positioning the instrument 6 Safety 6 Proper operation 6 CAT I 6 Environment of use. 6 Environmental conditions 7 Warranty and repair 7 Maintenance 7 Line voltage 7
Description of the controls 8
Basic signal measurement 10
Signals which can be measured 10 Amplitude of signals 10 Values of a sine wave signal 10 DC and ac components of an input signal 11 Timing relationships 11 Connection of signals 11
AUTOSET 19
Component tester 19
CombiScope 21
DSO Operation 22 DSO operating modes 22 Memory resolution 22 Memory depth 23 Horizontal resolution with X magnifi er 23 Maximum signal frequency in DSO mode 23 Display of aliases 23 Vertical amplifi er operating modes 23
Data transfer 23
HO710: RS-232 Interface. Remote control 24 Selection of Baud rate 24 Data transmission 24 Loading of new fi rmware 24
General information concerning MENU 25
Controls and Readout 26
First time operation and initial adjustments 12
Trace rotation TR 12 Probe adjustment and use 12 1 kHz adjustment 12 1 MHz adjustment 13
Operating modes of the vertical amplifi er 13
XY operation 14 Phase measurements with Lissajous fi gures 14 Measurement of phase differences in dual
channel Yt mode 14
Measurement of amplitude modulation 15
Triggering and time base 15
Automatic peak triggering (MODE menu) 15 Normal trigger mode (See menu MODE) 16 Slope selection (Menu FILTER) 16 Trigger coupling (Menu: FILTER) 16 Video (tv triggering) 16 Frame sync pulse triggering 17 Line sync pulse triggering 17 LINE trigger 17 Alternate trigger 17 External triggering 17 Indication of triggered operation (TRIG’D LED) 17 Hold-off time adjustment 17 Time base B (2nd time base). Delaying,
Delayed Sweep. Analog mode. 18
Alternate sweep 18
Subject to change without notice
3
HM1008
1 GSa/s Real Time Sampling, 10 GSa/s Random Sampling
1 MPt memory per channel allows Memory oom up to 50,000:1
Two Channels 1mV – 20 V/cm
8-Bit Low Noise Flash A/D Converters
Pre/Post Trigger -100 % to +400 %
Time Base 50 s/cm – 5 ns/cm
Acquisition modes: Single Event, Refresh, Average, Envelope, Roll, Peak-Detect
RS-232 Interface, optional: USB/RS-232, IEEE-488, Ethernet/USB
Signal display: Yt and XY; Interpolation: Sinx/x, Pulse, Dot Join (linear)
Analog Mode: see HM1000-2
100 MHz CombiScope
®
HM1008
Cursor measurement choices in digital mode
Digital Mode: TV field and zoomed display of one selected line
Either PAL or NTSC: Line triggering with line counter
4
Subject to change without notice
Technische Daten
100 MHz CombiScope®HM1008
Valid at 23 °C after a 30 minute warm-up period
Vertical Deflection
Channels:
Analog: 2 Digital: 2
Operating Modes:
Analog: CH 1 or CH 2 separate, DUAL (CH 1 and
CH 2 alternate or chopped), Addition
Digital: CH 1 or CH 2 separate, DUAL (CH 1 and
CH 2), Addition
X in XY-Mode: CH 1 Invert: CH 1, CH 2 Bandwidth (-3 dB): 2 x 0 - 100 MHz Rise time: ‹ 3.5 ns Overshoot: max. 1 % Bandwith limiting (selectable): about 20 MHz (5 mV/cm - 20V/cm) Deflection Coefficients(CH 1, 2):14 calibrated steps
1mV – 2mV/cm (10MHz) ±5% (0 - 10 MHz (-3dB)) 5 mV – 20 V/cm ±3% (1-2-5 sequence) variable (uncalibrated): › 2.5 :1 to › 50V/cm
Inputs CH 1, 2: Input Impedance: 1 MΩ II 15 pF Coupling: DC, AC, GND (ground) Max. Input Voltage: 400 V (DC + peak AC) Y Delay Line (analog): 70 ns Measuring Circuits: Measuring Category I Anal
og mode only:
Auxiliary input:
Function (selectable): Extern Trigger, Z (unblank) Coupling: AC, DC Max. input voltage: 100 V (DC +peak AC)
Triggering
Analog and Digital Mode Automatic (Peak to Peak):
Min. signal height: 5mm Frequency range: 10Hz - 200 MHz Level control range: from Peak- to Peak+
Normal (without peak): Slope/Video
Min. signal height: 5mm Frequency range: 0 - 200MHz
Level control range: –10 cm to +10cm Operating modes: Slope/Video Slope: positive, negative, both Sources: CH 1, CH 2, alt. CH 1/2 (8mm), Line, Ext. Coupling: AC: 10 Hz-200MHz
DC: 0 -200 MHz HF: 30 kHz–200 MHz LF: 0-5kHz
Noise Rej. switchable
Video: pos./neg. Sync. Impulse
Standards: 525 Line/60Hz Systems
625 Line/50Hz Systems
Field: even/odd /both
Line: all/line number selectable
Source: CH 1, CH 2, Ext. Indicator for trigger action: LED External Trigger via: Auxiliary Input (0.3Vpp, 100 MHz) Coupling: AC, DC Max. input voltage: 100 V (DC +peak AC) Digital mode Pre/Post Trigger: -100 % to +400% related to complete memory Analog mode 2nd Trigger
Min. signal height: 5mm
Frequency range: 0 - 200MHz
Coupling: DC
Level control range: –10 cm to +10cm
Horizontal Deflection
Analog mode
Operating modes: A, ALT (alternating A/B), B
Time base A: 0.5s/cm - 50 ns/cm (1-2-5 sequence)
Time base B: 20 ms/cm – 50 ns/cm (1-2-5 sequence)
Accuracy A and B: ±3%
X Magnification x10: to 5 ns/cm
Accuracy: ±5% Variable time base A/B: cont. 1:2.5 Hold Off time: var. 1:10 (LED-Indication) Bandwidth X-Amplifier: 0 - 3 MHz (-3dB)
X Y phase shift ‹ 3°: ‹ 220kHz
Digit
al mode Time base range (1-2-5 sequence) Refresh Mode: 20 ms/cm - 5 ns/cm with Peak Detect: 20ms/cm – 2 ms/cm (min. Pulse Width 10 ns) Roll Mode: 50s/cm – 50 ms/cm
Accuracy time base
Time base: 50 ppm Display: ±1%
MEMORY ZOOM: max. 50,000:1 Bandwidth X-Amplifier: 0 - 100 MHz (-3dB) X Y phase shift ‹ 3°: ‹ 100 MHz
Digital Storage
Sampling rate (real time): 2x 500 MSa/s, 1 GSa/s interleaved Sampling rate (random sampling): 10GSa/s Bandwidth: 2 x 0 - 100 MHz (random) Memory: 1 M-Samples per channel Operating modes: Refresh, Average, Envelope/
Roll: Free Run/Triggered, Peak-Detect
Resolution (vertical): 8 Bit (25 Pts/cm) Resolution (horizontal):
Yt: 11 Bit (200 Pts/cm) XY: 8 Bit (25 Pts /cm)
Interpolation: Sinx/x, Dot Join (linear) Delay: 1 Million x 1/Sampling Rate to
4 Million x 1/Sampling Rate
Display refresh rate: max.170 /s at 1 MPts Display: Dots (acquired points only), Vectors (partly
interpolated), optimal (complete memory weighting and vectors)
Reference Memories: 9 with 2 kPts each (for recorded signals)
Display: 2 signals of 9 (free selectable)
Operation/Measuring/Interfaces
Operation: Menu (multilingual), Autoset,
help functions (multilingual)
Save/Recall (instrument parameter settings): 9 Signal display: max. 4 traces
analog: CH 1, 2 (Time Base A) in combination with
CH 1, 2 (Time Base B)
digital: CH1,2 and ZOOM or Reference or
Mathematics)
Frequency counter:
6 digit resolution: ›1 MHz – 200MHz 5 digit resolution: 0.5 Hz – 1MHz Accuracy: 50ppm
Auto Measurements:
Analog mode: Frequency, Period, Vdc, Vpp, Vp+, Vp- also in digital mode: V
rms
, V
avg
Cursor Measurements:
Analog mode: Δt, 1/Δt (f), tr, ΔV, V to GND, ratio X, ratio Y plus in digital mode: V
pp
, Vp+, Vp-, V
avg
, V
rms
, pulse count
Resolution Readout/Cursor: 1000 x 2000Pts, Signals: 250 x 2000 Interfaces (plug-in): RS-232 (HO710) Optional: IEEE-488, Ethernet, Dual-Interface
RS-232/USB
Mathematic functions
Number of Formula Sets: 5 with 5 formulas each Sources: CH 1, CH 2, Math 1-Math 5 Targets: 5 math. memories, Math 1-5 Functions: ADD, SUB, 1/X, ABS, MUL, DIV, SQ, POS,
NEG, INV
Display: max. 2 math. memories (Math 1-5)
Display
CRT: D14-375GH Display area (with graticule): 8cm x 10cm Acceleration voltage: approx. 14 kV
General Information
Component tester
Test voltage: approx. 7V
rms
(open circuit), approx. 50 Hz
Test current: max. 7 mA
rms
(short circuit)
Reference Potential : Ground (safety earth)
Probe ADJ Output: 1 kHz/1 MHz square wave signal 0.2 V
pp
(tr ‹ 4 ns)
Trace rotation: electronic Line voltage: 105 – 253 V, 50/60 Hz ±10 %, CAT II Power consumption: 42 Watt at 230V, 50Hz Protective system: Safety class I (EN61010-1) Weight: 5.6 kg Cabinet (W x H x D): 285 x 125 x 380mm Ambient temperature: 0° C ...+40° C
Accessories supplied: Line cord, Operating manual, 2 Probes 10:1 with at tenuation ID, Windows Software for control and data transfer Optional accessories: HO720 Dual-Inter face RS-232/USB, HO730 Dual-Inter face Ethernet/ USB, HO740 Interface IEEE-488 (GPIB), HZ70 Opto-Interface (with
optical fi ber cable)
Subject to change without notice
5
Important hints
Important hints
Please check the instrument for mechanical damage or loose parts immediately after unpacking. In case of damage we advise to contact the sender. Do not operate.
List of symbols used:
Consult the manual High voltage
Important note Ground
Positioning the instrument
As can be seen from the fi gures, the handle can be set into different positions:
A = carrying B = handle removal and horizontal carrying C = horizontal operating D and E = operating at different angles F = handle removal T = shipping (handle unlocked)
B
C
B
T
A
C
D
F
E
D
E
A
PUkT
PUOGkT
PUkT
ANALOG
PUOPFGkT
PUkT
HGOFFD
PUOPFGkT
PUOPFGkT
HGOPFFD
DIGITAL
MIXED SIGNAL
COMBISCOPE
PUkT
HM1508
PUOPFGkT
PUOPFGkT
PFGkT
PUkT
1 GSa · 1MB
PUkT
150 MHz
PUOPFGkT PUOPFGkT
PUOPFGkT
VOLTS/DIVV HGOPFFD
B
HAMEG
C O M B I S C O P E
PUk PUk
PUkT
PUkT
PUOPFGkT
PUOPFGkT
PUkT
PUOPF
PUOPF
PUkT
PUOPF
HGOPFFD
PUkT
PUkT
VOLTS/DIVV
VOLTS/DIVV
HGOPFFD
HGOPFFD
PUkT
PUOPFGkT
PUkT
HGOPFFD
PUkT
PUOPFGkT
PUk PUk
PUk PUk
PUOPFGkT
PGkT
PUkT PUkT
PUOPFGkT
PUOPFGkT
PUOPF PUOPF
PUOPFGkT
INPUTS
PUkTKl
PUkTKl
15pF
15pF
max
max
400 Vp
400 Vp
PUkT
Attention!
When changing the handle position, the instrument
must be placed so that it can not fall (e.g. placed on a table). Then the handle locking knobs must be simultaneously pulled outwards and rotated to the required position. Without pulling the locking knobs they will latch in into the next locking position.
Handle mounting/dismounting
The handle can be removed by pulling it out further, depending on the instrument model in position B or F.
Safety
The instrument fulfi ls the VDE 0411 part 1 regulations for electrical measuring, control and laboratory instruments and was manufactured and tested accordingly. It left the factory in perfect safe condition. Hence it also corresponds to European Standard EN 61010-1 resp. International Standard IEC 1010-1. In order to maintain this condition and to ensure safe operation the user is required to observe the warnings and other directions for use in this manual. Housing, chassis as well as all measu­ring terminals are connected to safety ground of the mains. All accessible metal parts were tested against the mains with 2200 V
The oscilloscope may only be operated from mains outlets with a safety ground connector. The plug has to be installed prior to connecting any signals. It is prohibited to separate the safety ground connection.
Most electron tubes generate X-rays; the ion dose rate of this in­strument remains well below the 36 pA/kg permitted by law.
. The instrument conforms to safety class I.
DC
T
T
Safe operation may be endangered if any of the following was noticed:
– in case of visible damage. – in case loose parts were noticed – if it does not function any more. – after prolonged storage under unfavourable conditions (e.g.
like in the open or in moist atmosphere).
– after any improper transport (e.g. insuffi cient packing not
conforming to the minimum standards of post, rail or trans­port fi rm)
Proper operation
Please note: This instrument is only destined for use by person­nel well instructed and familiar with the dangers of electrical measurements.
For safety reasons the oscilloscope may only be operated from mains outlets with safety ground connector. It is prohibited to separate the safety ground connection. The plug must be inserted prior to connecting any signals.
CAT I
In case safe operation may not be guaranteed do not use the instrument any more and lock it away in a secure place.
6
Subject to change without notice
This oscilloscope is destined for measurements in circuits not connected to the mains or only indirectly. Direct measurements,
Important hints
i.e. with a galvanic connection to circuits corresponding to the categories II, III, or IV are prohibited!
The measuring circuits are considered not connected to the mains if a suitable isolation transformer fulfi lling safety class II is used. Measurements on the mains are also possible if suitable probes like current probes are used which fulfi l the safety class II. The measurement category of such probes must be checked and observed.
Measurement categories
The measurement categories were derived corresponding to the distance from the power station and the transients to be expected hence. Transients are short, very fast voltage or cur­rent excursions which may be periodic or not.
Measurement CAT IV: Measurements close to the power station, e.g. on electricity meters
Measurement CAT III: Measurements in the interior of buildings (power distribution installations, mains outlets, motors which are permanently installed).
Measurement CAT II: Measurements in circuits directly connected to the mains (household appliances, power tools etc).
Environment of use.
The oscilloscope is destined for operation in industrial, business, manufacturing, and living sites.
Environmental conditions
check all operating modes and fulfi lment of specifi cations. The latter is performed with test equipment traceable to national measurement standards.
Statutory warranty regulations apply in the country where the HAMEG product was purchased. In case of complaints please contact the dealer who supplied your HAMEG product.
Maintenance
Clean the outer shell using a dust brush in regular intervals. Dirt can be removed from housing, handle, all metal and plastic parts using a cloth moistened with water and 1 % detergent. Greasy dirt may be removed with benzene (petroleum ether) or alcohol, there after wipe the surfaces with a dry cloth. Plastic parts should be treated with an antistatic solution destined for such parts. No fl uid may enter the instrument. Do not use other cleansing agents as they may adversely affect the plastic or lacquered surfaces.
Line voltage
The instrument has a wide range power supply from 105 to 253 V, 50 or 60 Hz ±10%. There is hence no line voltage selector.
The line fuse is accessible on the rear panel and part of the line input connector. Prior to exchanging a fuse the line cord must be pulled out. Exchange is only allowed if the fuse holder is undamaged, it can be taken out using a screwdriver put into the slot. The fuse can be pushed out of its holder and exchanged.
The holder with the new fuse can then be pushed back in place against the spring. It is prohibited to ”repair“ blown fuses or to bridge the fuse. Any damages incurred by such measures will void the warranty.
Operating ambient temperature: 0 to + 40 degrees C. During transport or storage the temperature may be –20 to +55 de­grees C.
Please note that after exposure to such temperatures or in case of condensation proper time must be allowed until the instru­ment has reached the permissible range of 0 to + 40 degrees resp. until the condensation has evaporated before it may be turned on! Ordinarily this will be the case after 2 hours. The oscilloscope is destined for use in clean and dry environments. Do not operate in dusty or chemically aggressive atmosphere or if there is danger of explosion.
The operating position may be any, however, suffi cient ventila­tion must be ensured (convection cooling). Prolonged operation requires the horizontal or inclined position.
Do not obstruct the ventilation holes!
Specifi cations are valid after a 20 minute warm-up period between 15 and 30 degr. C. Specifi cations without tolerances are average values.
Warranty and repair
HAMEG instruments are subjected to a rigorous quality control. Prior to shipment each instrument will be burnt in for 10 hours. Intermittent operation will produce nearly all early failures. After burn in, a fi nal functional and quality test is performed to
Type of fuse:
Size 5 x 20 mm; 250V~, C; IEC 127, Bl. III; DIN 41 662 (or DIN 41 571, Bl. 3). Cut off: slow blow (T) 0,8A.
Subject to change without notice
7
Front Panel Elements – Brief Description
Front Panel Elements – Brief Description
The fi gures indicate the page for complete discriptions
POWER (pushbutton switch) 26
Turns scope on and off.
INTENS (knob) 26
Intensity for trace- and readout brightness, focus and trace
rotation control.
FOCUS, TRACE, MENU (pushbutton switch) 26
Calls the Intensity Knob menu to be displayed and enables
the change of different settings by aid of the INTENS knob. See item 2.
REM (pushbutton switch) 26
Switches the displayed menu, the remote mode (REM lit)
off.
ANALOG/DIGITAL (pushbutton switch) 27
Switches between analog (green) and digital mode (blue).
STOP / RUN (pushbutton switch) 27 RUN: Signal data acquisition enabled. STOP: Signal data acquisition disabled. The result of the last
acquisition is displayed.
MATH (pushbutton switch) 27 Calls mathematical function menu if digital mode is pre-
sent.
ACQUIRE (pushbutton switch) 28 Calls the signal capture and display mode menu in digital
mode.
SAVE/RECALL (pushbutton switch) 29 Offers access to the reference signal (digital mode only) and
the instrument settings memory.
SETTINGS (pushbutton switch) 30 Opens menu for language and miscellaneous function; in
digital mode also signal display mode.
AUTOSET (pushbutton switch) 30 Enables appropriate, signal related, automatic instrument
settings.
in the chapter CONTROLS AND READOUT
VOLTS/DIV-SCALE-VAR (knob) 32
Channel 2 Y defl ection coeffi cient, Y variabel and Y scaling
setting.
AUTO / CURSOR MEASURE (pushbutton switch) 33
Calls menus and submenus for automatic and cursor sup-
ported measurement.
LEVEL A/B (knob) 34
Trigger level control for time base A and B.
MODE (pushbutton switch) 34
Calls selectable trigger modes.
FILTER (pushbutton switch) 35
Calls selectable trigger fi lter (coupling) and trigger slope
menu.
SOURCE (pushbutton switch) 35
Calls trigger source menu.
TRIG’d (LED) 36
Lit on condition that time base is triggered.
NORM (LED) 36
Lit on condition that NORMAL or SINGLE triggering is pre-
sent.
HOLD OFF (LED) 36
Lit if a hold off time > 0% is chosen in time base menu (HOR
pushbutton
X-POS / DELAY (pushbutton switch) 36
Calls and indicates the actual function of the HORIZONTAL
knob
HORIZONTAL (knob) 37
Changes the X position resp. in digital mode the delay time
(Pre- resp. Post-Trigger).
TIME/DIV-SCALE-VAR (knob) 37
Time base A and B defl ection coeffi cient, time base variable
and scaling control.
MAG (pushbutton switch) 37
10 fold expansion in X direction in Yt mode, with simulta-
neous change of the defl ection coeffi cient display in the readout.
).
, (X-POS = dark).
HELP (pushbutton switch) 30 Switches help texts regarding controls and menus on and
off.
POSITION 1 (knob) 30 Controls position of actual present functions: Signal (cur-
rent, reference or mathematics), Cursor and ZOOM (digi­tal).
POSITION 2 (knob) 31 Controls position of actual present functions: Signal (current,
reference or mathematics) Cursor and ZOOM (digital).
CH 1/2-CURSOR-CH3/4-MA/REF-ZOOM (pushbutton) 32 Calls the menu and indicates the current function of POSITION 1 and 2 controls.
VOLTS/DIV-SCALE-VAR (knob) 32
Channel 1 Y defl ection coeffi cient, Y variabel and Y scaling setting.
8
Subject to change without notice
HOR / VAR (pushbutton switch) 38
Calls ZOOM function (digital) and analog time base A and
B, time base variable and hold off control.
CH1 (pushbutton switch) 39
Calls channel 1 menu with input coupling, inverting, probe
and Y variable control.
VERT/XY (pushbutton switch) 39
Calls vertical mode selection, addition, XY mode and band-
width limiter.
CH2 (pushbutton switch) 41
Calls channel 1 menu with input coupling, inverting, probe
and Y variable control.
CH1 (BNC-socket) 41
Channel 1 signal input and input for horizontal defl ection in
XY mode.
Front Panel Elements – Brief Description
POWER
POWER
15
13
14
17
16
18
1 2 3
INTENS
POWER
!
EXIT MENU
REMOTE OFF
POSITION 1 POSITION 2
VOLTS / DIV
SCALE · VAR
20 V 1 mV 20 V 1 mV
CH 1
VAR
X-INP
!
CAT I
FOCUS TRACE
MENU
REM
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
INPUTS
1MΩII15pF
max
400 Vp
4
ANALOG
DIGITAL
5 6 7 8 9 10 11 12
ANALOG
DIGITAL
MATH
RECALL
OSCILLOSCOPE
HM1008
·
1 MB
1 GSa
100 MHz
VOLTS / DIV
SCALE · VAR
CH 2 HOR MAG
RUN ACQUIRE SETTINGS HELP
STOP
LEVEL A/B
TRIGGER
MODE
FILTER
SOURCE
AUX
X-POS
DELAY
TRIG ’d
NORM
HOLD OFF
VAR
AUXILIARY INPUT
TRIGGER EXTERN
!
CAT I
Z-INPUT
SAVE/
AUTOSET
HORIZONTAL
TIME / DIV
SCALE · VAR
50s 5ns
VAR
x10
1MΩ II
15pF
max
100 Vp
19
26 27
20
23
21
24
28
22
25
29
30
3431
32
33
35
MEMORY
COMBISCOPE
oom
CH2 (BNC-socket) 41
Channel 2 signal input.
AUX (pushbutton switch) 41
Calls AUXILIARY INPUT menu with intensity modulation (Z)
and external triggering selectable.
AUXILIARY INPUT (BNC-socket) 41
Input for external trigger or intensity (Z) modulation si-
gnal.
36
37
COMPONENT
TESTER
40
PROBE
ADJ
3839
PROBE / COMPONENT (pushbutton switch) 42
Calls COMPONENT TESTER mode settings and frequency
selection of PROBE ADJ signal.
COMPONENT TESTER (2 sockets with 4 mm Ø) 42
Connectors for test leads of the Component Tester. Left
socket is galvanically connected with protective earth.
PROBE / ADJ (socket) 42
Square wave signal output for frequency compensation of
x10 probes.
Subject to change without notice
9
Basic signal measurement
Basic signal measurement
Signals which can be measured
The following description pertains as well to analog as to DSO operation. The different specifi cations in both operating modes should be kept in mind.
The oscilloscope HM1008 can display all repetitive signals with a fundamental repetition frequency of at least 100 MHz. The frequency response is 0 to 100 MHz (-3 dB). The vertical amplifi ers will not distort signals by overshoots, undershoots, ringing etc.
Simple electrical signals like sine waves from line frequency ripple to hf will be displayed without problems. However, when measuring sine waves, the amplitudes will be displayed with an error increasing with frequency. At 80 MHz the am­plitude error will be around –10 %. As the bandwidths of in­dividual instruments will show a certain spread (the 100 MHz are a guaranteed minimum) the actual measurement error for sine waves cannot be exactly determined.
Pulse signals contain harmonics of their fundamental fre­quency which must be represented, so the maximum useful repetition frequency of nonsinusoidal signals is much lower than 100 MHz. The criterion is the relationship between the rise times of the signal and the scope; the scope’s rise time should be <1/3 of the signal’s rise time if a faithful repro­duction without too much rounding of the signal shape is to be preserved.
The display of a mixture of signals is especially diffi cult if it contains no single frequency with a higher amplitude than those of the other ones as the scope’s trigger system normally reacts to a certain amplitude. This is e.g. typical of burst signals. Display of such signals may require using the HOLD-OFF control.
Composite video signals may be displayed easily as the in­strument has a tv sync separator.
The maximum sweep speed of 5 ns/cm allows suffi cient time resolution, e.g. a 100 MHz sine wave will be displayed one period per 2 cm.
The vertical amplifi er inputs may be dc or ac coupled. Use dc coupling only if necessary and preferably with a probe.
Low frequency signals when ac coupled will show tilt (ac low frequency – 3 dB point is 1.6 Hz), so if possible use dc coupling. Using a probe with 10:1 or higher attenuation will lower the –3 dB point by the probe factor. If a probe cannot be used due to the loss of sensitivity dc coupling the scope and an external large capacitor may help which, of course, must have a suffi cient dc rating. Care must be taken, however, when charging and discharging a large capacitor.
Amplitude of signals
In contrast to the general use of rms values in electrical engi­neering oscilloscopes are calibrated in Vpp as that is what is displayed.
Derive rms from V
: divide by 2.84. Derive Vpp from rms: mul-
pp
tiply by 2.84.
Values of a sine wave signal
V
p
V
rms
V
mom
V
pp
V
= rms value
rms
V
= pp – value
PP
V
= momentary value, depends on time vs. period.
mom
The minimum signal for a one cm display is 1 mV ded 1 mV/cm was selected and the variable is in the calibrated position.
The available sensitivities are given in mV
pp
allow to indicate the amplitudes of the signals immediately on the readout as the attenuation of probes is automatically taken into account. Even if the probe attenuation was selected manu­ally this will be overridden if the scope identifi es a probe with an identifi cation contact as different. The readout will always give the true amplitude.
It is important that the variable be in its calibrated position. The sensitivity may be continuously decreased by using the variable (see Controls and Readout). Each intermediate value between the calibrated positions 1–2–5 may be selected. Without using a probe thus a maximum of 400 V
may be displayed (20 V/div
PP
x 8 cm screen x 2.5 variable).
Amplitudes may be directly read off the screen by measuring the height and multiplying by the V/div. setting.
Please note: Without a probe the maximum permis-
sible voltage at the inputs must not exceed 400 V irrespective of polarity.
In case of signals with a dc content the peak value DC + AC peak must not exceed + or – 400 V
. Pure AC of up to 800 VPP
P
is permissible.
If probes are used their possibly higher ratings are
only usable if the scope is dc coupled.
±5 % provi-
pp
or Vpp. The cursors
p
Dc coupling is preferable with all signals of varying duty cyc­le, otherwise the display will move up and down depending on the duty cycle. Of course, pure dc can only be measured with dc coupling.
The readout will show which coupling was chosen: = stands for DC, ~ stands for AC.
10
Subject to change without notice
In case of measuring dc with a probe while the scope input is ac coupled the capacitor in the scope input will see the input dc voltage as it is in series with the internal 1 MΩ resistor. This means that the maximum dc voltage (or DC + peak AC) is that of the scope input, i.e. 400 V
! With signals which contain
P
dc and ac the dc content will stress the input capacitor while the ac content will be divided depending on the ac impedance
of the capacitor. It may be assumed that this is negligible for frequencies >40 Hz.
Considering the foregoing you may measure dc signals of up to 400 V or pure ac signals of up to 800 V
with a HZ200 probe.
PP
Probes with higher attenuation like HZ53 100:1 allow to measure dc up to 1200 V and pure ac of up to 2400 V
. (Please note the
PP
derating for higher frequencies, consult the HZ53 manual). Stressing a 10:1 probe beyond its ratings will risk destruction of the capacitor bridging the input resistor with possible ensuing damage of the scope input!
In case the residual ripple of a high voltage is to be measured a high voltage capacitor may be inserted in front of a 10:1 probe, it will take most of the voltage as the value of the probe’s internal capacitor is very low, 22 to 68 nF will be suffi cient.
If the input selector is switched to Ground the reference trace on the screen may be positioned at graticule center or else­where.
DC and ac components of an input signal
voltage
peak
AC
DC
DC + AC
DC
AC
peak
= 400 V
max
The dashed curve shows an ac signal symmetrical to zero. If there is a dc component the peak value will be dc + ac peak.
Timing relationships
The repetition frequency of a signal is equal to the number of periods per second. Depending on the TIME/DIV setting one or more periods or part of a period of the signal may be displayed. The time base settings will be indicated on the readout in s/cm, ms/cm, μs/cm and ns/cm. Also the cursors may be used to measure the frequency or the period.
If portions of the signal are to be measured use delayed sweep (analog mode) or zoom (DSO mode) or the magnifi er x 10. Use the HORIZONTAL positioning control to shift the portion to be zoomed into the screen center.
Pulse signals are characterized by their rise and fall times which are measured between the 10 % and 90 % portions. The following example uses the internal graticule of the crt, but also the cursors may be used for measurement.
Measurement:
– Adjust the rising portion of the signal to 5 cm.
– Position the rising portion symmetrically to the graticule
centre line, using both Y and X positioning controls.
– Notice the intersections of the signal with the 10 and 90 %
lines and project these points to the centre line in order to read the time difference.
Basic signal measurement
100%
90%
5 cm
10%
0%
t
tot
In the example it was 1.6 cm at 5 ns/cm equals 8 ns rise time.
When measuring very short rise times coming close to the scope rise time it is necessary to subtract the scope’s (and if used the probe’s) rise times geometrically from the rise time as seen on the screen. The true signal rise time will become:
2
2
ta= t
t
is the rise time seen, t
tot
– t
tot
(3.5 ns with the HM1008), t
2
– t
osc
t
is the scope’s own rise time
osc
is the rise time of the probe, e.g.
t
2 ns. If the signal’s rise time is > 34 ns, the rise times of scope and probe may be neglected.
ta= 82 - 3.52 - 22 = 6.9 ns
For the measurement of rise times it is not necessary to proceed as outlined above. Rise times may be measured anywhere on the screen. It is mandatory that the rising portion of the signal be measured in full and that the 10 to 90 % are observed. In case of signals with over- or undershoot the 0 and 100 % levels are those of the horizontal portions of the signal, i.e. the over­resp. undershoots must be disregarded for rise and fall time measurements. Also, glitches will be disregarded. If signals are very distorted, however, rise and fall time measurements may be of no value.
For most amplifi ers, even if their pulse behaviour is far from ideal, the following relationship holds:
350 350 t
=
——
a
B t
B =
——
a
tr/ns = 350/Bandwidth/MHz
Connection of signals
In most cases pressing the AUTOSET button will yield a satis­factory display (see AUTOSET). The following relates to special cases where manual settings will be advisable. For a description of controls refer to ”Controls and Readout“.
Take care when connecting unknown signals to the
inputs!
It is recommended to use probes whenever possible. Without a probe start with the attenuator set to its 20 V/cm position. If the trace disappears the signal amplitude may be too large overdriving the vertical amplifi er or/and its dc content may be too high. Reduce the sensitivity until the trace will reappear onscreen. If calibrated measurements are desired it will be necessary to use a probe if the signal becomes >160 Vp. Check the probe specifi cations in order to avoid overstressing. If the time base is set too fast the trace may become invisible, then reduce the time base speed. If no probe is used at least screened cable should be used, such as HZ32 or HZ34. However, this is only advisable for low
Subject to change without notice
11
First time operation and initial adjustments
impedance sources or low frequencies (<50 kHz). With high frequencies impedance matching will be necessary.
Nonsinusoidal signals require impedance matching, at both ends preferably. At the scope input a feed through – 50 Ω-ter­mination will be required. HAMEG offers a HZ22 termination. If proper terminations are not used sizeable pulse aberrations will result. Also sine wave signals of > 100 kHz should be properly terminated. Most generators control signal amplitudes only if correctly terminated.
HZ22 may only be used up to 7 V
or 20 VPP i.e. 1 W.
rms
For probes terminations are neither required nor allowed, they would ruin the signal.
Probes feature very low loads at fairly low frequencies: 10 MΩ in parallel to a few pF, valid up to several hundred kHz. However, the input impedance diminishes with rising frequency to quite low values. This has to be borne in mind as probes are, e.g., entirely unsuitable to measure signals across high impedance high frequency circuits such as bandfi lters etc.! Here only FET probes can be used. Use of a probe as a rule will also protect the scope input due to the high probe series resistance (9 MΩ). As probes cannot be calibrated exactly enough during ma­nufacturing individual calibration with the scope input used is mandatory! (See Probe Calibration).
Passive probes will, as a rule, decrease the scope bandwidth resp. increase the rise time. We recommend to use HZ200 pro­bes in order to make maximum use of the combined bandwidth. HZ200 features 2 additional hf compensation adjustments.
First time operation and initial adjustments
Prior to fi rst time operation the connection between the instru­ment and safety ground must be ensured, hence the plug must be inserted fi rst.
Use the red pushbutton POWER to turn the scope on. Several displays will light up. The scope will then assume the set-up, which was selected before it was turned off. If no trace and no readout are visible after approximately 20 sec, push the AUTOSET button.
As soon as the trace becomes visible select an average inten­sity with INTENS, then select FOCUS and adjust it, then select TRACE ROTATION and adjust for a horizontal trace.
With respect to crt life use only as much intensity as necessary and convenient under given ambient light conditions, if unused turn the intensity fully off rather than turning the scope off and on too much, this is detrimental to the life of the crt heater. Do not allow a stationary point to stay, it might burn the crt phosphor.
With unknown signals start with the lowest sensitivity 20 V/cm, connect the input cables to the scope and then to the measu­ring object which should be deenergized in the beginning. Then turn the measuring object on. If the trace disappears, push AUTOSET.
Whenever the DC content is > 400 V
coupling must be used in
DC
order to prevent overstressing the scope input capacitor. This is especially important if a 100:1 probe is used as this is specifi ed for 1200 V
+ peak AC.
DC
AC coupling of low frequency signals may produce tilt.
If the dc content of a signal must be blocked it is possible to insert a capacitor of proper size and voltage rating in front of the probe, a typical application would be a ripple measurement.
When measuring small voltages the selection of the ground connection is of vital importance. It should be as close to voltage take-off point as possible, otherwise ground currents may de­teriorate the measurement. The ground connections of probes are especially critical, they should be as short as possible and of large size.
If a probe is to be connected to a BNC connector use
a probe tip to BNC adapter.
If ripple or other interference is visible, especially at high sen­sitivity, one possible reason may be multiple grounding. The scope itself and most other equipment are connected to safety ground, so ground loops may exist. Also, most instruments will have capacitors between line and safety ground installed which conduct current from the live wire into the safety ground.
Trace rotation TR
The crt has an internal graticule. In order to adjust the defl ected beam with respect to this graticule the Trace Rotation control is provided. Select the function Trace Rotation and adjust for a trace which is exactly parallel to the graticule.
Probe adjustment and use
In order to ensure proper matching of the probe used to the scope input impedance the scope contains a calibrator with short rise time and an amplitude of 0.2 V
± 1 %, equivalent to
pp
4 cm at 5 mV/cm when using 10:1 probes.
The inner diameter of the calibrator connector is 4.9 mm and standardized for series F probes. Using this special connec­tor is the only way to connect a probe to a fast signal source minimizing signal and ground lead lengths and to ensure true displays of pulse signals.
1 kHz adjustment
This basic adjustment will ensure that the capacitive attenuation equals the resistive attenuation thus rendering the attenuation of the probe independent of frequency. 1:1 probes can not be adjusted and need no such adjustment anyway.
12
Subject to change without notice
incorrect correct incorrect
Operating modes of the vertical amplifier
Prior to adjustment make sure that the trace rotation adjust­ment was performed.
Connect the 10:1 probe to the input. Use dc coupling. Set the VOLTS/DIV to 5 mV/cm and TIME/DIV to 0.2 ms/cm, both calibrated. Insert the probe tip into the calibrator connector PROBE ADJ.
You should see 2 signal periods. Adjust the compensation ca­pacitor (see the probe manual for the location) until the square wave tops are exactly parallel to the graticule lines (see picture 1 kHz). The signal height should be 4 cm ±1.6 mm (3% oscillo­scope and 1% probe tolerance). The rising and falling portions of the square wave will be invisible.
1 MHz adjustment
The HAMEG probes feature additional adjustments in the compensation box which allow to optimise their hf behaviour. This adjustment is a precondition for achieving the maximum bandwidth with probe and a minimum of pulse aberrations.
This adjustment requires a calibrator with a short rise time (typ. 4 ns) and a 50 Ω output, a frequency of 1 MHz, an amplitude of 0.2 V requirements.
Connect the probe to the scope input to which it is to be adjusted. Select the PROBE ADJ. signal 1 MHz. Select dc coupling and 5 mV/cm with VOLTS/DIV. and 0.1 μs/cm with TIME/DIV., both calibrated. Insert the probe tip into the calibrator output connec­tor. The screen should show the signal, rise and fall times will be visible. Watch the rising portion and the top left pulse corner, consult the manual for the location of the adjustments.
. The PROBE ADJ. output of the scope fulfi ls these
PP
Operating modes of the vertical amplifi er
The controls most important for the vertical amplifi er are: VERT/XY containing the operating modes and the parameters of the individual channels.
Changing the operating mode is described in the chapter: ”Controls and Readout“.
Remark: Any reference to ”both channels“ always refers to channels 1 and 2.
Usually oscilloscopes are used in the Yt mode. In analog mode the amplitude of the measuring signal will defl ect the trace vertically while a time base will defl ect it from left to right.
The vertical amplifi ers offer these modes:
– One signal only with CH1. – One signal only with CH2. – Two signals with channels 1 and 2 (DUAL trace mode)
In DUAL mode both channels are operative. In analog mode the method of signal display is governed by the time base (see also ”Controls and Readout“). channel switching may either take place after each sweep (alternate) or during sweeps with a high frequency (chopped).
The normal choice is alternate, however, at slow time base set­tings the channel switching will become visible and disturbing, when this occurs select the chopped mode in order to achieve a stable quiet display.
, CH1 , CH2 . They give access to the menus
incorrect correct incorrect
The criteria for a correct adjustment are:
– short rise time, steep slope.
– clean top left corner with minimum over- or undershoot,
fl at top.
After adjustment check the amplitude which should be the same as with 1 kHz.
It is important to fi rst adjust 1 kHz, then 1 MHz. It may be ne­cessary to check the 1 kHz adjustment again.
Please note that the calibrator signals are not calibrated with respect to frequency and thus must not be used to check the time base accuracy, also their duty cycle may differ from 1:1.The probe adjustment is completed if the pulse tops are horizontal and the amplitude calibration is correct.
In DSO mode no channel switching is necessary as each input has its own A/D converter, signal acquisition is simultaneous.
In ADD mode the two channels 1 and 2 are algebraically ad­ded (±CH1 ±CH2). With + polarity the channel is normal, with – polarity inverted. If + Ch1 and – CH2 are selected the difference will be displayed or vice versa.
Same polarity input signals:
Both channels not inverted: = sum Both channels inverted: = sum Only one channel inverted: = difference
Opposite polarity input signals:
Both channels not inverted: = difference Both channels inverted: = difference One channel inverted: = sum.
Please note that in ADD mode both position controls will be operative. The INVERT function will not affect positioning.
Often the difference of two signals is to be measured at signal take-offs which are both at a high common mode potential. While this one typical application of the difference mode one important precaution has to be borne in mind: The oscillos­cope vertical amplifi ers are two separate amplifi ers and do not constitute a true difference amplifi er with as well a high CM rejection as a high permissible CM range! Therefore please observe the following rule: Always look at the two signals in the one channel only or the dual modes and make sure that they are within the permissible input signal range; this is the case if they can be displayed in these modes. Only then switch to ADD. If this precaution is disregarded grossly false displays
Subject to change without notice
13
Operating modes of the vertical amplifier
may result as the input range of one or both amplifi ers may be exceeded.
Another precondition for obtaining true displays is the use of two identical probes at both inputs. But note that normal probe tolerances (percent) will cause the CM rejection to be expected to be rather moderate. In order to obtain the best possible re­sults proceed as follows: First adjust both probes as carefully as possible, then select the same sensitivity at both inputs and then connect both probes to the output of a pulse generator with suffi cient amplitude to yield a good display. Readjust one (!) of the probe adjustment capacitors for a minimum of over­or undershoot. As there is no adjustment provided with which the resistors can be matched a residual pulse signal will be unavoidable.
When making difference measurements it is good practice to fi rst connect the ground cables of the probes to the object prior to connecting the probe tips. There may be high potentials between the object and the scope. If a probe tip is connected fi rst there is danger of overstressing the probe or/and the scope inputs! Never perform difference measurements without both probe ground cables connected.
XY operation
This mode is accessed by VERT/XY > XY. In analog mode the time will be turned off. The channel 1 signal will defl ect in X direction (X-INP. = horizontal input), hence the input attenuators, the variable and the POSITION 1 control will be operative. The HORIZONTAL control will also remain functional.
Channel 2 will defl ect in Y direction.
Please note:
– As the trigonometric functions are
periodic limit the calculation to angles <90 degrees. This is where this function is most useful.
– Do not use too high frequencies,
because, as explained above, the two amplifi ers are not identical, their phase difference increases with frequency. The spec gives the frequency at which the phase difference will stay <3 degrees.
– The display will not show which of the two frequencies does
lead or lag. Use a CR combination in front of the input of the frequency tested. As the input has a 1 MΩ resistor it will be suffi cient to insert a suitable capacitor in series. If the ellipse increases with the C compared to the C short-circuited the test signal will lead and vice versa. This is only valid <90 degrees. Hence C should be large and just create a barely visible change.
If in XY mode one or both signals disappear, only a line or a point will appear, mostly very bright. In case of only a point there is danger of phosphor burn, so turn the intensity down immedia­tely; if only a line is shown the danger of burn will increase the shorter the line is. Phosphor burn is permanent.
a sin ϕ = b
a cos ϕ = 1 – (— b
a ϕ = arc sin b
2
)
Measurement of phase differences in dual channel Yt mode
Please note: Do not use ”alternate trigger“ because the time
differences shown are arbitrary and depend only on the respec­tive signal shapes! Make it a rule to use alternate trigger only in rare special cases.
The x 10 magnifi er will be inoperative in XY mode. Please note the differences in the Y and X bandwidths, the X amplifi er has a lower – 3 dB frequency than the Y amplifi er. Consequently the phase difference between X and Y will increase with fre­quency.
In XY mode the X signal (CH1 = X-INP). can not be inverted.
The XY mode may generate Lissajous fi gures which simplify some measuring tasks and make others possible:
– Comparison of two signals of different frequency or adju-
stment of one frequency until it is equal to the other resp. becomes synchronized.
– This is also possible for multiples or fractions of one of the
frequencies.
Phase measurements with Lissajous fi gures
The following pictures show two sine waves of equal amplitude and frequency but differing phase.
ab
0° 35° 90° 180°
Calculation of the phase angle between the X- and Y-signals (af­ter reading a and b off the screen) is possible using the following formulas and a pocket calculator with trigonometric functions. This calculation is independent of the signal amplitudes:
The best method of measuring time or phase differences is using the dual channel Yt mode. Of course, only times may be read off the screen, the phase must then be calculated as the frequency is known. This is a much more accurate and convenient method as the full bandwidth of the scope is used, and both amplifi ers are almost identical. Trigger the time base from the signal which shall be the reference. It is necessary to position both traces without signal exactly on the graticule center (POSITION 1 and 2). The variables and trigger level controls may be used, this will not infl uence the time difference measurement. For best accuracy display only one period at high amplitude und observe the zero crossings. One period equals 360 degrees. It may be advantageous to use ac coupling if there is an offset in the signals.
t = horizontal spacing of the
zero transitions in div
T= horizontal spacing for one
period in div
In this example t = 3 cm and T = 10 cm, the phase difference in degrees will result from:
5 3 ϕ° =
T 10
or in angular units:
t 3 arc ϕ° =
T 10
· 360° = — · 360° = 108°
· 2π = — · 2π = 1,885 rad
14
Subject to change without notice
Triggering and time base
Very small phase differences with moderately high frequencies may yield better results with Lissajous fi gures.
However, in order to get higher precision it is possible to switch to higher sensitivities – after accurately positioning at graticule centre – thus overdriving the inputs resulting in sharper zero crossings. Also, it is possible to use half a period over the full 10 cm. As the time base is quite accurate increasing the time base speed after adjusting for e.g. one period = 10 cm and positioning the fi rst crossing on the fi rst graticule line will also give better resolution.
Measurement of amplitude modulation
Please note: Use this only in analog mode because in DSO mode alias displays may void the measurement! For the display of low modulation frequencies a slow time base (TIME/DIV) has to be selected in order to display one full period of the modu­lating signal. As the sampling frequency of any DSO must be reduced at slow time bases it may become too low for a true representation.
The momentary amplitude at time t of a hf carrier frequency modulated by a sinusoidal low frequency is given by:
u = UT · sinΩt + 0,5 m · UT · cos (Ω - ω) t - 0,5 m · UT · cos (Ω - ω) t
where: UT = amplitude of the unmodulated carrier Ω = 2πF = angular carrier frequency ω = 2πf = modulation angular frequency m = modulation degree (1
In addition to the carrier a lower side band F – f and an upper side band F + f will be generated by the modulation.
U
T
v100%)
Set the scope controls as follows in order to display the picture 2 signal:
CH1 only, 20 mV/cm, AC TIME/DIV: 0.2 ms/cm Triggering: NORMAL, AC, internal. Use the time base variable or external triggering.
Reading a and b off the screen the modulation degree will result:
a – b a – b m =
——
a + b a + b
(1 + m) and b = UT (1 – m)
a = U
T
bzw. m =
—— · 100 [%]
When measuring the modulation degree the amplitude and time variables can be used without any infl uence on the result.
Triggering and time base
The most important controls and displays for these functions are to be found in the shaded TRIGGER area, they are described in „Controls and Readout“.-
In YT mode the signal will defl ect the trace vertically while the time will defl ect it horizontally, the speed can be selected. In general periodic voltage signals are displayed with a peri­odically repeating time base. In order to have a stable display successive periods must trigger the time base at exactly the same time position of the signal (amplitude and slope).
0,5 m · U
T
0,5 m · U
T
F – f F F + f
Picture 1: Amplitudes and frequencies with AM (m = 50 %) of the spectra
As long as the frequencies involved remain within the scope’s bandwidth the amplitude-modulated hf can be displayed. Pre­ferably the time base is adjusted so that several signal periods will be displayed. Triggering is best done from the modulation frequency. Sometimes a stable displayed can be achieved by twiddling with the time base variable.
m · U
T
U
T
ba
Pure DC can not trigger the time base, a voltage
change is necessary.
Triggering may be internal from any of the input signals or externally from a time-related signal.
For triggering a minimum signal amplitude is required which can be determined with a sine wave signal. With internal trigge­ring the trigger take-off within the vertical amplifi ers is directly following the attenuators. The minimum amplitude is specifi ed in mm on the screen. Thus it is not necessary to give a minimum voltage for each setting of the attenuator.
For external triggering the appropriate input connector is used, the amplitude necessary there is given in V
. The voltage for
pp
triggering may be much higher than the minimum, however, it should be limited to 20 times the minimum. Please note that for good triggering the voltage resp. signal height should be a good deal above the minimum. The scope features two trigger modes to be described in the following:
Automatic peak triggering (MODE menu)
Consult the chapters MODE > AUTO, LEVEL A/B , FILTER
and SOURCE in ”Controls and Readout“. Using AUTOSET this trigger mode will be automatically selected. With DC­coupling and with alternate trigger this mode will be left while the automatic triggering will remain.
Picture 2: Amplitude modulated HF. F = 1 MHz, f = 1 kHz, m = 50 %, U
= 28,3 mV
T
rms
Automatic triggering causes a new time base start after the end of the foregoing and after the hold-off time has elapsed even
Subject to change without notice
15
Triggering and time base
without any input signal. Thus there is always a visible trace in analog mode, and in DSO mode the trace will also be shown. The position of the trace(s) without any signal is then given by the settings of the POSITION controls.
As long as there is a signal scope operation will not need more than a correct amplitude and time base setting. With signals < 20 Hz their period is longer than the time the auto trigger circuit will wait for a new trigger, consequently the auto trigger circuit will start the time base then irrespective of the signal so that the display will not be triggered and free run, quite inde­pendent of the signal’s amplitude which may be much larger than the minimum.
Also in auto peak trigger mode the trigger level control is active. Its range will be automatically adjusted to coincide with the signal’s peak-to-peak amplitude, hence the name. The trigger point will thus become almost independent of signal amplitude. This means that even if the signal is decreased the trigger will follow, the display will not loose trigger. As an example: the duty cycle of a square wave may change between 1:1 and 100:1 without loosing the trigger.
Depending on the signal the LEVEL A/B control may have to be set to one of its extreme positions.
The simplicity of this mode recommends it for most uncompli­cated signals. It is also preferable for unknown signals.
negative portion of a signal. This is valid in automatic and normal modes.
Trigger coupling (Menu: FILTER)
Consult chapters: MODE > AUTO, LEVEL A/B , FILTER and SOURCE coupling will be used unless ac coupling was selected before. The frequency responses in the diverse trigger modes may be found in the specifi cations.
With internal dc coupling with or without LF fi lter use normal triggering and the level control. The trigger coupling selected will determine the frequency response of the trigger channel.
AC:
This is the standard mode. Below and above the fall-off of the frequency response more trigger signal will be necessary.
DC:
With direct coupling there is no lower frequency limit, so this is used with very slowly varying signals. Use normal triggering and the level control. This coupling is also indicated if the signal varies in its duty cycle.
HF:
A high pass is inserted in the trigger channel, thus blocking low frequency interference like fl icker noise etc.
in ”Controls and Readout“. In AUTOSET DC
This trigger mode is independent of the trigger source and usable as well for internal as external triggering. But the signal must be > 20 Hz.
Normal trigger mode (See menu MODE)
Consult the chapters: MODE > AUTO, LEVEL A/B , FILTER
and SOURCE in ”Controls and Readout“. Information about how to trigger very diffi cult signals can be found in the HOR menu VAR, HOLD-OFF time setting, and time base B operation are explained.
With normal triggering and suitable trigger level setting trigge­ring may be chosen on any point of the signal slope. Here, the range of the trigger level control depends on the trigger signal amplitude. With signals <1 cm care is necessary.
In normal mode triggering there will be no trace visible in the absence of a signal or when the signal is below the minimum trigger amplitude requirement!
Normal triggering will function even with complicated signals. If a mixture of signals is displayed triggering will require repetition of amplitudes to which the level can be set. This may require special care in adjustment.
where the functions time base fi ne adjustment
Slope selection (Menu FILTER)
After entering FILTER the trigger slope may be selected using the function keys. See also ”Controls and Readout“. AUTOSET will not change the slope.
Noise Reject:
This trigger coupling mode or fi lter is a low pass suppressing high frequencies. This is useful in order to eliminate hf inter­ference of low frequency signals. This fi lter may be used in combination with dc or ac coupling, in the latter case very low frequencies will also be attenuated.
LF:
This is also a low pass fi lter with a still lower cut-off frequency than above which also can be combined with dc or ac coupling. Selecting this fi lter may be more advantageous than using dc coupling in order to suppress noise producing jitter or double images. Above the pass band the necessary trigger signal will rise. Together with ac coupling there will also result a low frequency cut-off.
Video (tv triggering)
Selecting MODE > Video will activate the TV sync separator built-in. It separates the sync pulses from the picture content and enables thus stable triggering independent of the changing video content.
Composite video signals may be positive or negative. The sync pulses will only be properly extracted if the polarity is right. The definition of polarity is as follows: if the video is above the sync it is positive, otherwise it is negative. The polarity can be selected after selecting FILTER. If the pola­rity is wrong the display will be unstable resp. not triggered at all as triggering will then initiated by the video content. With internal triggering a minimum signal height of 5 mm is necessary.
Positive or negative slope may be selected in auto or normal trigger modes. Also, a setting ”both“ may be selected which will cause a trigger irrespective of the polarity of the next slope.
Rising slope means that a signal comes from a negative po­tential and rises towards a positive one. This is independent of the vertical position. A positive slope may exist also in the
16
Subject to change without notice
The PAL sync signal consists of line and frame signals which differ in duration. Pulse duration is 5 μs in 64 μs intervals. Frame sync pulses consist of several pulses each 28 μs repeating each half frame in 20 ms intervals.
Both sync pulses differ hence as well in duration as in their repetition intervals. Triggering is possible with both.
Triggering and time base
Frame sync pulse triggering
Remark: Using frame sync triggering in dual trace chopped mode may result in interference, then the dual trace alternate mode should be chosen. It may also be necessary to turn the rea­dout off.
In order to achieve frame sync pulse triggering call MODE, select video signal triggering and then FILTER to select frame triggering. It may be selected further whether ”all“, ”only even“ or ”only odd“ half frames shall trigger. Of course, the correct tv standard must be selected fi rst of all (625/50 or 525/60).
The time base setting should be adapted, with 2 ms/cm a com­plete half frame will be displayed. Frame sync pulses consist of several pulses with a half line rep rate.
Line sync pulse triggering
In order to choose line snyc triggering call MODE and select VIDEO, enter FILTER, make sure that the correct video standard is selected (625/50 or 525/60) and select Line.
If ALL was selected each line sync pulse will trigger. It is also possible to select a line number ”LINE No.“.
In order to display single lines a time base setting of TIME/DIV. = 10 μs/cm is recommended, this will show 1½ lines. In general the composite video signal contains a high dc component which can be removed by ac coupling, provided the picture is steady. Use the POSITION control to keep the display within the screen. If the video content changes like with a regular tv program only dc coupling is useful, otherwise the vertical position would continuously move.
The sync separator is also operative with external triggering. Consult the specifi cations for the permissible range of trigger voltage. The correct slope must be chosen as the external trigger may have a different polarity from the composite video. In case of doubt display the external trigger signal.
LINE trigger
Consult SOURCE in ”Controls and Readout“ for specifi c information.
If the readout shows Tr:Line the trigger signal will be internally taken from the line (50 or 60 Hz).
Alternate trigger
This mode is selected with SOURCE > Alt. 1/2. The read­out will display Tr:alt, but no more the trigger point symbol indicating level and time position. Instead an arrow pointing upwards will indicate the trigger time position if this lies within the screen area.
This trigger mode is to be used with greatest care and should be an exception rather than the rule, because the time relationships visible on the screen are completely meaningless, they depend only on the shape of the signals and the trigger level!
In this mode the trigger source will be switched together with the channel switching, so that when CH1 is displayed in the dual channel alternate mode the trigger is taken from CH1 and when CH2 is displayed the trigger is taken from CH2. This way two uncorrelated signals can be displayed together. If this mode is inadvertently chosen the time relationships between the signals will also be lost when both signals are correlated! (Except for the special case that both happen to be square waves with extremely fast rise times). Of course, this trigger mode is only possible in the dual channel alternate mode and also not with external or line trigger. Ac coupling is recommended for most cases.
External triggering
In analog mode this trigger mode may be selected with SOURCE
> Extern. In DSO mode it is only possible if channels 3 and 4 are turned off. The readout will display Tr:ext. CH4 the input for the external trigger, all internal sources will be disconnected. In this mode the trigger point symbol (level and time position) will not be displayed, only the trigger time position will be indicated. External triggering requires a signal of 0.3 to 3 V
, synchronous to the vertical input signal(s).
PP
Triggering will also be possible within limits with multiples or fractions of the vertical input signal frequency. As the trigger signal may have any polarity it may happen that the vertical input signal will start with a negative slope in spite of having selected positive slope; slope selection refers now to the ex­ternal trigger.
will be
Indication of triggered operation (TRIG’D LED)
Refer item in ”Controls and Readout“. The LED labelled TRIG’D indicates triggered operation provided: – Suffi cient amplitude of the internal or external trigger signal. – The trigger point symbol is not above or below the signal.
This trigger signal is independent of the scope input signals and is recommended for all signals synchronous with the line. Within limits this will also be true for multiples or fractions of the line frequency. As the trigger signal is taken off internally there is no minimum signal height on the screen for a stable display. Hence even very small voltages like ripple or line frequency interference can be measured.
Please note that with line triggering the polarity switching will select either the positive or negative half period of the line, not the slope. The trigger level control will move the trigger point over most of a half wave.
Line frequency interference may be checked using a search coil which preferably should have a high number of turns and a shielded cable. Insert a 100 Ω resistor between the center conductor and the BNC connector. If possible the coil should be shielded without creating a shorted winding.
If these conditions are met the trigger comparator will output triggers to start the time base and to turn on the trigger indi­cation. The trigger indicator is helpful for setting the trigger up, especially with low frequency signals (use normal trigger) resp. very short pulses. The trigger indication will store and display triggers for 100 ms. With signals of very low rep rate the indicator will fl ash accor­dingly. If more than one signal period is shown on the screen the indicator will fl ash each period.
Hold-off time adjustment
Consult ”Controls and Readout“ HOR > Hold-off time for specifi c information.
After the time base defl ected the trace from left to right the trace will be blanked so the retrace is invisible. The next sweep will, however, not immediately start. Time is required to perform
Subject to change without notice
17
Triggering and time base
internal switching, so the next start is delayed for the so called hold-off time, irrespective of the presence of triggers. The hold-off time can be extended from its minimum by a factor of 10:1. Manipulation of the hold-off time and thus of the time for a complete sweep period from start to start can be useful e.g. when data packets are to be displayed. It may seem that such signals can not be triggered. The reason is that the possible start of a new sweep does not conincide with the start of a data packet, it may start anywhere, even before a data packet. By varying the hold-off time a stable display will be achieved by setting it just so that the hold-off ends before the start of a data packet. This is also handy with burst signals or non-periodic pulse trains.
A signal may be corrupted by noise or hf interference so a double display will appear. Sometimes varying the trigger level can not prevent the double display but will only affect the apparent time relationship between two signals. Here the variable hold-off time will help to arrive at a single display.
Sometimes a double display will appear when a pulse signal contains pulses of slightly differing height requiring delicate trigger level adjustment. Also here increasing the hold-off time will help.
Whenever the hold-off time was increased it should reset to its minimum for other measurements, otherwise the brightness will suffer as the sweep rep rate will not be maximum. The following pictures demonstrate the function of the hold-off:
heavy parts are displayed
period
signal
be defl ected vertically by the input signal. In fact the input signal does continuously defl ect the trace vertically, but this will be only visible during the unblanking time. This is, by the way, one marked difference to DSO operation where the input signal is only measured during the acquisition time, for most of the time the DSO will not see the signal. Also, in analog mode the signal itself will be seen on the screen in real time, whereas a DSO can only show a reconstruction of the signal acquired some time later.
In analog mode thus the display will always start on the left. Let us assume one period of a signal is displayed at a convenient time base setting. Increasing the sweep speed with TIME/DIV. will expand the display from the start, so that parts of the signal will disappear from the screen. It is thus possible to expand the beginning of the signal period and show fi ne detail, but it is impossible to show such fi ne detail for ”later“ parts of the signal.
The x10 Magnifi er (MAG x10) may be used to expand the display and the horizontal positioning control can shift any part of the display into the centre, but the factor of 10 is fi xed.
The solution requires a second time base, called time base B.
In this mode time base A is called the delaying sweep and time base B the delayed sweep. The signal is fi rst displayed by TB A alone. Then TB B is also turned on which is the mode ”A intensifi ed by B“. TB B should always be set to a higher sweep rate than A, thus its sweep duration will be also shorter than that of A. The TB A sweep sawtooth is compared to a voltage which can be varied such that TB A functions as a precision time delay generator. Depending on the amplitude of the com­parison voltage a signal is generated anywhere between sweep start and end.
sweep
Fig. 1
adjusting
HOLD OFF time
Fig. 2
Fig. 1: Display with minimum hold-off time (basic setting). Double image, no stable display.
Fig. 2: By increasing the hold-off a stable display is achie­ved.
Time base B (2nd time base). Delaying, Delayed Sweep. Analog mode
Consult ”Controls and Readout“ HOR and TIME/DIV. for specifi c information.
As was described in ”Triggering and time base“ a trigger will start the time base. While waiting for a trigger – after runout of the hold-off time – the trace will remain blanked. A trigger will cause trace unblanking and the sweep ramp which defl ects the trace from left to right with the speed set with TIME/DIV. At the end of the sweep the trace will be blanked again and retrace to the start position. During a sweep the trace will also
In one of two operating modes this signal will start TB B imme­diately. The TB A display will be intensifi ed for the duration of TB B, so that one sees which portion of the signal is covered by TB B, By varying the comparison voltage the start of TB B can be moved over the whole signal as it is displayed by TB A. Then the mode is switched to TB B. The signal portion thus selected is now displayed by TB B. This is called „B delayed by A“. Portions of the signal can thus be expanded enormously, however, the higher the speed of TB B the darker the display will become as the rep rate will remain that of the accepted signal triggers while the duration of TB B is reduced with increasing speed.
In cases where there is jitter the TB B can be switched to wait for a trigger rather than starting immediately. When a trigger arrives TB B will start on it. The jitter is removed, however, the effect is also, that the TB B start now can be only from signal period to signal period, no continuous adjustment is possible in this mode.
Alternate sweep
In this mode the signal is displayed twice, with both time bases. An artifi cial Y offset can be added in order to separate the two displays on the screen. The operation is analogous to Y dual trace alternate mode, i.e., the signal is alternately displayed by both time bases, not simultaneously which is not possible with a single gun crt. What was said above about how TB B can be started holds also here.
18
Subject to change without notice
Component Tester
AUTOSET
For specific information consult ”Controls and Readout“ AUTOSET
The following description is valid for both analog and DSO modes. AUTOSET does not change from analog to DSO mode or vice versa. If in DSO mode the modes ”Roll“, ”Envelope“ or ”Average“ (ACQUIRE) are present or the trigger mode „Single“ (MODE) is selected, theses modes will be switched off as AU­TOSET always switches to ”Refresh“ acquistion. The signal to be displayed must meet the amplitude and frequency require­ments of automatic triggering, to enable a useful automatic instrument setting.
All controls except for the POWER switch are electronically scanned, all functions can also be controlled by the microcom­puter, i.e. also via the interfaces.
This is a precondition for AUTOSET as this function must be able to control all functions independent of control settings. AUTO­SET will always switch to YT mode, but preserve the previous selection of CH1, CH2 or dual trace; ADD or XY modes will be switched to dual trace Yt.
Automatic setting of the vertical sensitivities and the time base will present a display within 6 cm height (4 cm per signal in dual trace) and about 2 signal periods. This is true for signals not differing too much from a 1:1 duty cycle. For signals containing several frequencies like video signals the display may be any.
.
Component Tester
Specifi c information can be found in ”Controls and Readout“ un­der COMPONENT/PROBE
The scope has a built-in component tester. The test object is connected with 4 mm banana plugs. In this mode the Y amplifi ers and the time base are turned off. Only individual components may be tested, i.e. they must not be part of a circuit, if voltages are to be applied to the BNC connectors. If the components are part of a circuit this must be deenergized and disconnected from safety ground. Except for the two test leads there may be no further connection between scope and component. (See ”Tests within a circuit“). As described in section ”Safety“ all ground connections of the scope are connected to safety ground including those of the component tester. As long as individual components are tested this is of no consequence.
The display can only be affected by the controls contained in the FOCUS/TRACE menu: A-Int., Focus, Trace rotation, HORI­ZONTAL position.
If components are to be tested which are parts of a circuit or an instrument those circuits resp. instruments must fi rst be deenergized. If they are connected to the mains they must be unplugged. This will prevent a connection between scope and circuit via the safety ground which may affect the measure­ment.
Do not test charged capacitors.
and COMPONENT TESTER .
Initiating the AUTOSET function will set the following operating conditions: – last selection of ac or dc coupling – internal triggering – automatic triggering – trigger level set to the center of its range – calibrated Y sensitivities – calibrated time base – AC or DC trigger coupling unmodifi ed – HF trigger coupling switched to DC – LF or Noise Reject fi lters left – X magnifi er switched off – Y and X positioning automatic
Please note: For pulse signals with duty cycles approaching 400:1 no au­tomatic signal display will be possible.
In such cases switch to normal trigger mode and set the trigger position about 5 mm above the centre. If the trigger LED will then light up a trigger is generated and the time base is operating. In order to obtain a visible display it may be necessary to change the time base and V/DIV settings. Depending on the duty cycle and the frequency the signal may still remain invisible. This applies only to analog mode. In DSO mode the trace is always of equal brightness because not the signal is shown but a low frequency construction of it, also, there is no information in the trace intensity.
The principle of the test is very simple: a sine wave generator within the scope generates a 50 Hz ±10 % voltage which is applied to a series connection of the test object and a resistor within the scope. The sine wave proper defl ects in X direction, the voltage across the resistor which is proportional to the test current defl ects in Y direction.
If the object contains neither capacitors nor inductors, there will be no phase shift between voltage and current, so a straight line will show up which will be more or less slanted, depending on the value of the object’s resistance, covering appr. 20 Ω to
4.7 kΩ. If there is a short the trace will be vertical, i.e. (almost) no voltage produces already high current. A horizontal line will thus indicate an open, there is only voltage but no current.
Capacitors or inductors will create ellipses. The impedance may be calculated from the ellipse’s geometric dimensions. Capacitors of appr. 0.1μ to 1000 μF will be indicated.
– An ellipse with its longer axis horizontal indicates a high
impedance (low capacitance or high inductance)
– An ellipse with its longer axis vertical will indicate a low
impedance (high capacitance or low inductance)
– A slanted ellipse will indicate a lossy capacitor or induc-
tor.
Semiconductors will show their diode characteristics, however, only 20 V ristics can only be displayed up to 10 V
are available, so the forward and reverse characte-
pp
in each direction. The
p
test is a two-terminal test, hence it is not possible to measure e.g. the current gain of a transistor. One can only test B-C, B-E, and C-E. The test current is only a few mA, so the test will not harm ordinary semiconductors. (Sensitive devices like delicate HF transistors etc. should not be tested). The limitation to 10
Subject to change without notice
19
Component Tester
Vp with bipolar transistors will suffi ce mostly as usual defects will show up.
The best method to verify whether a component is defective is the comparison to a good one. If the lettering of a component is not legible at least it is possible to see whether it is a npn or pnp transistor or which end of a diode is the cathode.
Please note that reversing the test leads will also invert the picture, i.e. turn it 180 degrees.
In most cases, e.g. with service and repair, it will be suffi cient to receive a good/bad result (open, short). With MOS compo­nents the usual precautions are to be observed, but note, that except for a possible short MOSFETs and JFETs can not be suffi ciently tested. Indications to be expected depend strongly on the kind of FET:
– With depletion type MOSFETs and all JFETs the channel
will conduct if prior to testing the gate was connected to the source. The Rdson will be shown. As this can be very low it may look like a plain short although the part is good!
– With enhancement type MOSFETs an open will be seen in
all directions, as the threshold voltage G – S is not available. With power MOSFETs the antiparallel diode S – D can be seen.
Tests of components within circuits are possible in many cases but less indicative because other components may be in parallel. But also here the comparison with a good circuit might help. As both circuits must be deenergized it is only necessary to switch the test leads back and forth between both in order to localize a defective spot. Sometimes like with stereo amplifi ers, push­pull circuits, bridge circuits there is a comparison circuit right on the same board. In cases of doubt one component lead can be unsoldered, the other one should then be connected to the ground lead. This is labelled with a ground symbol. The pictures show some practical examples:
20
Subject to change without notice
CombiScope
CombiScope
®
HAMEG oscilloscopes are either analog or they are CombiSco­pes, i.e. they contain a complete analog scope and the additional hardware and software so sample and digitize the signals. The
®
HM1008 is a 100 MHz 1 GS/s CombiScope
®
With a HAMEG CombiScope
the user is always sure: he needs
.
only to switch from DSO to analog in order to see the true signal. This is especially important when a signal is to be documented in DSO mode. The user of a pure DSO needs to know the signal better than the scope!
The advantages of DSO operation are:
– Capture and storage of single events – There will be no fl icker with very low frequency signals – Fast signals with a low rep rate resp. low duty cycle can be
displayed at high intensity
– Due to the storage of all signals they may be easily docu-
mented and processed.
– Due to the down conversion of high frequencies into the kHz
area the very expensive analog crts are replaced by cheap computer monitor tubes or lcd displays. Also, the bulk part of the components may be the same as used in computers, thus inexpensive.
The disadvantages of DSO operation are:
– An analog scope displays the signal itself in real time. In
a DSO the signal is not displayed but only a low frequency reconstruction of the signal. The limitations and problems of sampling operation as well as those of analog/digital conversion hold. The display can not be in real time as, after capturing a signal, the DSO must take time out to perform calculations the result of which will then be displayed much later.
– Therefore the capture rate of ordinary DSOs is orders of
magnitude lower than that of any analog scope. Hence a DSO is least suited to catch rare events.
– There is no information in the trace, the trace is always of
equal intensity. Thus valuable information (so called Z-axis) is lost. Also the fast slopes of a pulse which are invisible on an analog scope will be of the same intensity as the slower parts of the signal, this is a gross misrepresentation. The reason is that DSOs ordinarily do not show only the sampled points but they interpolate by drawing a continuous trace.
– The vertical resolution is mostly only 8 bits. In an analog
scope there is no loss of fi ne detail by digitizing. Even if the trace is not very crisp details can be seen in it.
– Due to the sampling and the lack of a low pass fi lter in the
input frequencies above half the sampling frequency will cause so called aliases, i.e. low frequency ghost signals. Sampling is practically the same as frequency conversion resp. multiplication, it creates sum and difference frequen­cies, beat frequencies which may be orders of magnitude lower than the signal frequency and give grossly erroneous results. In practice, therefore, only frequencies 1/10 or less of the sampling frequency can be reliably displayed. The meaning of the Nyquist theorem is mostly misunderstood: if the sampling frequency is only twice the signal frequency there will only two points be displayed on the screen: any
number of signal shapes may be drawn which fi t through these two points. The Nyquist theorem contains a hidden knowledge that the signal is a sine wave. It is easily under­stood that, in order to depict an unknown signal shape one needs at least 1 or 2 points per centimeter; in other words: the useful signal frequency is only 1/10 to 1/20 at best.
– An analog scope has a frequency response which follows
closely the Gaussian curve, this means in practice that also frequencies far beyond the –3 dB frequency will be shown, reduced in amplitude, but they will be shown. This not only preserves fi ne detail of a signal but it allows also to see, e.g., very high frequency wild oscillations in a circuit. This is not the case after sampling because all frequencies beyond half the sampling frequency will be ”folded“ back into the lower frequency band.
– Due to limited memory depth the maximum sampling rate
must be reduced with each DSO when the time base is set to slow sweep speeds, it may be reduced from GS/s to KS/s! Most users are not aware of this trap, they think that if they bought a DSO with 100 MHz bandwidth and 1 GS/s they are safe when measuring kHz-signals. But such low frequency signals will be distorted and aliases displayed.
Please note: This list of disadvantages is by far incomplete! It scratches but the surface.
There are 3 methods of sampling:
1. Real time sampling: Here the Nyquist theorem must be observed, but, as mentio­ned, in practice the signal frequency is far less than 1/10 the sampling frequency. Consequently, with a 1 GS/s rate signals with up to 100 MHz can be adequately reconstructed. Obviously, this is the only mode for single event capturing.
2. Equivalent time sampling: This is the normal operating mode for all sampling scopes. (Sampling scopes are very old, they are still the fastest scopes with bandwidths > 50 GHz because they have no input ampli­fi er. Sampling scopes are far superior to DSOs because their Y resolution is identical to that of an analog scope). In this mode consecutive periods of the signal are sampled, each period contributes but one sample. The signal period is thus scanned and very many periods are necessary in order to achieve one full screen display. This way a very high „effective“ sampling rate is achieved, this method exchanges bandwidth for time. In a sampling scope a very accurate display is created which is, as far as the shape is concerned, almost as good as that of an analog scope. In a DSO, however, the sample points are 8 bit a/d converted, losing resolution. The bandwidth achieved is given alone by the hf properties of the input and the minimum realiz­able duration of the sampling pulse, so 14 GHz at a sensitivity of 2 mV/cm and 50 Ω were standard in the 60s. In a DSO, however, which should be used like an analog scope, a high impedance (1 MΩ) wide range (e.g. 1 mV/cm to 20 V/cm) attenuator must be included and also an input amplifi er. This is why a DSO can not reach the bandwidths of sampling scopes. Equivalent time sampling suffers fully from the problems of aliasing. As it re­quires the (not necessarily periodic) repetition of the signal in invariant shape for e.g. millions of periods it is unsuitable for the capture of single events. Equivalent time sampling can not display the rising portion of a signal without a delay line.
3. Random sampling: Random sampling is also very old (1952) and also no invention of DSOs. It is similar to equivalent time sampling in that it
Subject to change without notice
21
CombiScope
requires a multitude of signal repetitions with invariant shape in order to reconstruct it once on the screen. Therefore also in this mode a very high ”effective“ sampling rate is achieved. However, the samples are not taken step for step along the sig­nal but randomly distributed over the signal period. An analog computer is used to estimate the arrival of the next trigger, and the time base is already started when it arrives. This has two enormous advantages:
st
The rising portion of the signal can be shown without the
1
need for a delay line which would severely limit the achie­vable bandwidth.
nd
2
Due to the randomness of the samples alias signals will be
broken up.
The foregoing explains why it is HAMEG policy to offer Combi­Scopes rather than pure DSOs which combine the best of both worlds although the cost of such an instrument is markedly higher than that of a pure DSO, take alone the complicated high frequency crt. It is the low cost of manufacturing which causes the drive towards DSOs.
DSO Operation
The 100 MHz scope has 2 8bit a/d converters of the fl ash type which is the best there is. The maximum sample rate of each is 500 MS/s which is the rate available in dual channel mode for the capture of single events. The maximum sampling rate in all other operating modes is 1 GS/s.
Higher effective (!) sampling rates are possible as explained above in equivalent and random sampling modes. As very many signals repetitions are needed to reconstruct the signal once any changes in signal shape such as noise will show up.
The reconstructed signal may be displayed either by showing only the sampled points or with interpolation between them by drawing straight lines.
5. XY display of signals which were previously captured in Yt mode and protected against overwriting by STOP:
XY: readout ”XY“
Signal capture is triggered in SINGLE, REFRESH, ENVELOPE, and AVERAGE modes and untriggered in ROLL and XY modes.
The normal (Refresh) mode is similar to the operating mode of an analog scope. Triggering will cause signal acquisition and display from left to right. After the next acquisition the display will be replaced by the new information. If automatic triggering was selected there will be a reference trace in the absence of a signal the position of which is dependent on the vertical position control setting. Signals with a repetition rate lower than the rep rate of the automatic triggering can not properly trigger so the resulting display will be untriggered.
In contrast to an analog scope the last display will remain on the screen if the signal disappears in normal trigger mode.
In SINGLE mode the signal will be acquired only once. Acquisi­tion can start if STOP (RUN key) is not illuminated (if necessary press RUN until STOP extinguishes). The next trigger received will cause the single acquisition. After this STOP will light up and the trigger mode will be automatically switched to normal dc coupled if auto was selected. The trigger symbol on the screen allows to directly see resp. determine the voltage level desired for triggering in the normal mode, the voltage follows from the position and the VOLTS/CM selected. The ground reference will be indicated by a ground symbol in the screen centre.
After selecting SINGLE the trigger level symbol may be positio­ned using the LEVEL control. If e.g. the symbol is 2 cm above the ground reference symbol the trigger level will be 2 cm x Volts/cm (x probe factor if any). Example: 2 cm x 1 V/cm x 10 (probe) = + 20 V.
The signals stored in DSO mode can be read via an interface and documented. See the chapter ”Data Transfer“ for details.
DSO operating modes
In DSO mode the following operating modes are available:
1. Menu: ACQUIRE: Repetitive triggered signal acquisition and display in usual Yt representation.
REFRESH: readout shows ”rfr“ (real time sampling) or Random sampling: readout ”RS:xGSa“.
The operating mode may be further subdivided:
Envelope: readout ”env“ Average: readout ”avg:x“ (x may be a number 2.. 512)
2. ROLL mode, untriggered continuous signal acquisition, display will „roll“ over the screen from left to right in usual Yt mode:
Roll: readout ”rol“
3. Single sweep, triggered (menu: Trigger MODE) signal capture in usual Yt mode:
Single: readout ”sgl“
4. Untriggered continuous signal capture, display in XY mode (Menu: trigger MODE):
XY: readout ”XY“
Memory resolution
Vertical resolution:
The 8 bit a/d converters have a resolution of 256 possible ver­tical positions. The screen display has a resolution of 25 points per cm. This is advantageous for display, documentation and post-processing.
There may be some difference between the display on screen and documentation, e.g. on a printer, this results from various tolerances in the analog circuitry involved. The trace positions are defi ned by:
Median horizontal line: 10000000b 80h 128d Top line: 11100100b E4h 228d Bottom line: 00011100b 1Ch 28d
In contrast to an analog display with its theoretically infi nite resolution this is limited to 25 points per cm in DSO mode. If there is any noise superimposed on the signal this may cause frequent change of the lowest bit and thus jumping of the trace in vertical direction.
Horizontal resolution:
A maximum of 4 simultaneous signal displays may be shown on the screen. Each signal display will consist of 2048 points (bytes). 2000 points will be distributed over 10 cm. The resolution is thus 200 points per cm. Please note that this a 4 to 8 times improvement over customary VGA (50 points per div) or LCD (25 points per div.) DSO displays.
22
Subject to change without notice
Data transfer
Memory depth
1 GS/s means that one million samples will be taken of the signal and stored. With normal triggering and time base settings of >20 ms/cm there will be 500,000 samples.
The screen display is calculated from the whole memory con­tents. Within the menu Settings->Display several display modes may be selected:
Dots: the sampling points only are displayed. Vectors: interpolation (sin x/x) or dot join is used to gene
rate a continuous trace.
Optimal: In this mode all samples are used to calculate
the display. This way the display of aliases is less likely.
The scope acquires with as high a sampling rate as is possible thus preventing to a large extent the production of alias signals. It is always possible to zoom through the memory in order to look at details, and, thanks to the deep memory, signal details may be shown which remain invisible with shorter memory DSOs.
Example:
This scope will sample with 1 GS/s in single channel mode down to a time base setting of 100 us/cm. This equals 100,000 points per cm. In MEMORY ZOOM signals of 150 MHz can still be seen. Down to 100 us/cm hence aliases are not to be expected due to the bandwidth limit of 150 MHz and the critical frequency being > 500 MHz.
DSOs with a shorter memory like e.g. 10 K will only present 1000 points per cm which is equivalent to a sampling rate of 10 MHz, thus signals > 5 MHz will cause aliases , far below the scope bandwidth. A deep memory is one of the most important criterion of a DSO.
apparently untriggered changing displays, or may look like AM modulated signals. If an alias is suspected change the signal frequency or the time base or both. If aliases remain undetected grossly erroneous results will be obtained which includes also grossly (maybe orders of magnitude) false displays of signal parameters like rise time etc.! Always watch for a stepped display or print-out: this indicates an insuffi cient sampling rate and consequently a false display. With an insuffi cient sampling rate e.g. fast, short pulses may be completely ignored.
The best method to detect any false DSO display is to switch to analog mode. In analog mode false displays are absolutely impossible! An analog scope can at worst round the edges of very fast signals.
Vertical amplifi er operating modes
In principle, in DSO mode there are the same modes available as in analog mode, i.e.:
– CH1 only – CH2 only – CH1 and CH2 in dual trace mode Yt or XY – Sum – Difference
The main differences of DSO mode are:
– In dual channel mode both channels resp. signals are a/d
converted simultaneously.
No alternate or chopped channel switching.
– No fl ickering display even with low frequency signals as
the signals are stored and continuously displayed from the memory with a suffi ciently high rep rate.
– Trace intensity is always the same. This is an advantage and
a disadvantage.
Horizontal resolution with X magnifi er
In principle, with a 10 x magnifi ed sweep, the resolution should be reduced to 20 points per cm. However, the resolution re­mains at 200 points per cm as the information necessary will be calculated from the memory. The magnifi ed portion may be selected with the X-POS control. The fastest time base will be 5 ns/cm allowing a 2 cm per period display of 100 MHz.
Maximum signal frequency in DSO mode
The highest signal repetition frequency which still can be dis­played well can not be exactly given. This is dependent as well on the signal shape as on its amplitude displayed.
While it is fairly easy to recognize a square wave it requires at least 10 samples per period to distinguish a sine wave from a triangle. In other words: in practice, signals may still be re­cognized if their repetition frequency is <1/10 of the sampling frequency. For a well defi ned display, however, many more than 10 points per cm are necessary.
Display of aliases
As explained the maximum sampling rate must be reduced for slow time base settings. This may cause aliases. If e.g. a sine wave is sampled only with one sample per period and if it should be synchronous with the sampling frequency a horizontal line will be shown as each time the same signal point is sampled. An alias may also take the form of a signal of much lower frequency (beat frequency between signal and sampling frequencies),
All so called Z axis (trace intensity) information is lost. In analog mode the intensity depends on the signal rep rate resp. the speed, thus mixed or unstable signals can be differentiated by their respective trace intensity. Fast slopes of low frequency signals are invisible in analog mode, in DSO mode they will be shown as bright as the other signal portions.
Data transfer
Please note: Interface modules may only be exchanged after the instrument was turned off. During operation the opening of the interface must be covered.
There is an opening on the rear panel into which various inter­face modules can be inserted. Instruments are delivered with RS-232 installed. (HO710)
The interface allows to either remotely control the scope or read its settings. In DSO mode also the digitized and stored signals may be retrieved. Interface cables must be shielded and must not reach a length of 3 m or more.
Subject to change without notice
23
Data transfer
HO710: RS-232 Interface. Remote control
Safety hint: All interface connections are galvanically connected
!
to the scope.
Measurements at high potentials are prohibited and endanger the scope, the interface and all gear connected to the inter­face.
If the safety rules are disregarded any damage to HAMEG products will void the warranty. Neither will HAMEG take any responsibility for damages to people or gear of other make.
Description:
The RS-232 interface on the rear panel has the usual 9 pole SubD connector. Via this bidirectional interface the scope can be controlled remotely or its settings may be transferred. In DSO mode also the digitized and stored signals can be read out. The connection to a pc requires a 9 pole screened cable (1:1) of 3 m maximum length. The pinout is as follows:
Pin no. Function
2 Tx date from scope to external device 3 Rx data from external device to scope 7 CTS ready to transmit 8 RTS ready to receive 5 ground (scope is connected to safety ground, safety class I) 9 + 5 V, max. 400 mA
The maximum signal on Tx, Rx, RTS and CTS is ±12 V. The RS-232 interface parameters are: N-8-2 no parity, 8 bits data, 2 stop bits (RTS/CTS hardware protocol)
Selection of Baud rate
Baud rate setting is automatic. Range: 110 to 115200 , no parity, 8 bits data, 2 stop bits.
The fi rst SPACE CR (20 hex, OD hex) character sent after POWER UP will set the Baud rate. This will remain set until POWER DOWN or until a RM=0 command was sent. If activated previous­ly also the front panel LOCAL (auto range) key may be used. After release of remote (REM key (4) dark) data transfer can only be reinitiated if a SPACE CR character was sent.. If the scope can not recognize a SPACE CR as the fi rst charac­ter TxD will be pulled low after 0.2 ms which causes a frame error. If the scope did recognize a SPACE CR und adjusted its Baud rate accordingly it will answer with RETURN CODE „O CR LF“. The front panel controls will be deactivated. The time between Remote Off and Remote ON must be greater than: tmin = 2 x (1/Baud rate) + 60 us.
Data transmission
After a successful Baud rate recognition and setting the scope will be in the remote control mode and waits for commands. In this mode no manual operation is possible. In order to return to manual control a command must be sent from the PC or the REMOTE OFF key depressed.
Loading of new fi rmware
Under www.hameg.de the most recent fi rmware is available for downloading.
24
Subject to change without notice
CH I: 500 mV
POWER
General information concerning MENU
POWER
Pushbutton
CH I MENU
AC/DC/50 Ω
GND
50 Ω / 1 MΩ
INVERT
ON / OFF
VARIABLE
ON / OFF
PROBE
1 : 1 / 10 / 100
Menu Title
6 Function Pushbuttons (blue)
Menu
Intensity Knop Symbol
Arrow Keys
Indicator for Submenu
COMBISCOPE
MEMORY
oom
COMPONENT
TESTER
General information concerning MENU
Menu and HELP displays
Whenever a pushbutton is depressed the corresponding menu will be displayed except for: EXIT MENU/REMOTE OFF ANALOG / DIGITAL
. The menus offer options which may then be selected with
the blue pushbuttons. All pushbuttons are on/off.
Exiting a menu:
st
1
The user may set a time after which the menu will be left.
(SETTINGS
nd
2
Before the time set as described has elapsed a menu can
be exited by pressing EXIT MENU.
rd
3
Only manually if the function ”Man“ was selected .
th
4
Pressing the same pushbutton again.
th
5
Pressing another pushbutton.
, AUTOSET , RUN/STOP and MAG x10
pushbutton > , Misc > Menu OFF).
PROBE
ADJ
On Off
State indication by intensifi ed display
Please note:
During the display of help texts and menus in full
size no signal display is possible.
Remarks
In operation all relevant measuring parameters will be shown in the readout, provided the readout was activated and its in-
,
tensity is suffi cient.
The front panel LEDs add to the operating comfort and give more information. In the end positions of the control knob an acoustical signal will sound.
Apart from the POWER scanned and stored. This allows to control the instrument from stored information. Some controls and menus are only operative in DSO mode or change their meaning in this mode. Explanations are given with the warning: ”Only in DSO mode.“
pushbutton all control elements are
If a menu shows a knob symbol knob
. It allows to change settings. Also arrows may be
this pertains to the INTENS
shown which point to available submenus.
In some modes some pushbuttons or knob operations are meaningless and will hence not cause a menu display.
Please note:
If a menu is shown some other information dis-
played in the readout may disappear, this will reappear immediately upon leaving the menu.
Each menu is assisted by HELP texts, which can be called by pressing the HELP readout. If HELP was called and the INTENS knob explanation of the actual INTENS knob
and which will be also displayed by the
moved an
function will be given.
HELP will be left by pressing the pushbutton again.
Subject to change without notice
25
Controls and Readout
POWER
15
13
14
17
16
18
1 2 3
INTENS
POWER
!
EXIT MENU
REMOTE OFF
POSITION 1 POSITION 2
VOLTS / DIV
SCALE · VAR
20 V 1 mV 20 V 1 mV
CH 1
VAR
X-INP
!
CAT I
FOCUS TRACE
MENU
REM
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
INPUTS
1MΩII15pF
max
400 Vp
4
ANALOG
DIGITAL
OSCILLOSCOPE
HM1008
·
1 GSa
100 MHz
VOLTS / DIV
SCALE · VAR
CH 2 HOR MAG
VAR
5 6 7 8 9 10 11 12
1 MB
ANALOG
DIGITAL
RUN ACQUIRE SETTINGS HELP
STOP
MODE
FILTER
SOURCE
AUX
!
CAT I
MATH
LEVEL A/B
TRIGGER
TRIG ’d
NORM
HOLD OFF
SAVE/
RECALL
X-POS
DELAY
AUXILIARY INPUT
TRIGGER EXTERN
Z-INPUT
AUTOSET
HORIZONTAL
TIME / DIV
SCALE · VAR
50s 5ns
VAR
x10
1MΩ II
15pF
max
100 Vp
19
26 27
20
23
21
24
28
22
25
29
30
3431
32
33
35
Controls and Readout
POWER
Mains switch with symbols I = ON and = OFF.
After turning the scope on and after the warm-up time of the crt heater has elapsed the HAMEG logo, the instrument type and the version number are displayed. If prior to switching off the function ”Quick Start“ was selected (SETTINGS
> Misc) the above will not be displayed. Then the scope will assume the settings which existed before switching off.
INTENS knob
This knob controls various functions:
2.1 Trace intensity if the FOCUS/TRACE/MENU does not light the knob symbol
. Turn left for decreasing and
right for increasing.
pushbutton
36
37
FOCUS TRACE MENU
If the knob symbol will light after pressing this pushbutton the Int.Knob menu will be shown.
Depending on the operating mode the menu contains:
A-Int.: Intensity of the signal as displayed by time base A B-Int.: Intensity of the signal as displayed by time base B (analog mode only) Zoom Int. Intensity of the signal as displayed by ZOOM
(digital mode only)
RO-Int.: Readout intensity Focus: Focus for signal and readout Readout On Off: Turning the readout off will eliminate interference
of the readout with the signal(s). The knob symbol will blink as long as the readout is off. Only menus and help texts will be shown. Trace rotation: Trace rotation (see TR) After turn-on the readout will always be on.
EXIT MENU/REMOTE OFF (REM)
This pushbutton has two functions:
2.2 If the knob symbol XXX of the pushbutton
is illuminated the control will act for those functions displayed in the menu, which were activated.
26
Subject to change without notice
4.1 If a menu is displayed pressing this pushbutton will turn it off. On condition a submenu is displayed, pressing this pushbutton switches back to the menu.
Controls and Readout
4.2 Under remote control this pushbutton will be illuminated.
After pressing the pushbutton, control is returned to the front panel.
ANALOG/DIGITAL
Switches between analog (green) and DSO modes (blue). The operating mode (Yt or XY) will not be affected. If component test mode was selected, possible only in analog mode, and the scope is switched to digital the operating mode last used when in DSO mode will be reinstalled.
Y parameters will not be changed by the switching. Time base modes will be changed due to the different operation of the time bases in both modes. After any switching time base A will be selected. The time base speeds selected will not be affected unless they are not available any more, then the maximum value will be chosen.
RUN/STOP
This pushbutton has several functions:
6.1 Analog mode, single sweep The RUN/STOP pushbutton is used to set the time base for a single sweep. Press MODE menu. By pressing RUN/STOP the scope will wait for the next trigger indicated by STOP blinking. After a signal triggered the time base and the sweep is over the scope will stop, indicated by STOP continuously illuminated. Each time the pushbutton is depressed the scope will be activated anew for one trigger resp. one sweep.
6.2 DSO mode, single capture
In order to prepare the scope for a single capture press MODE
and select ”Single“ in the menu presented. By pressing RUN/STOP the scope will start to continuously acquire and store so that later also the signal preceding the occurrence of the trigger (Pre-trigger) will be available. STOP will blink in order to signal that as yet no trigger was recognized. Upon receipt of a trigger the acquisition will be completed so the signal following the trigger (Post-trigger) will later be available, too. STOP will then remain illuminated. The signal display on the screen will be frozen. Pressing RUN/STOP again will prepare for another acquisition as described. Each new acquisition will overwrite the preceding one.
and activate ”Single“ in this
All entries and settings will be automatically stored upon leaving the Mathematics menu or turning the scope off. Measurement results will be lost after turn-off.
”Mathematics“ offers:
7.1 Equations set
Using the INTENS knob
5 sets of formulas can be selected for
editing. This way 5 user-defi ned formula sets may be created.
Each set of formulas consists of 5 lines with one equation each, designated MA1 to MA5. An equation may occupy one or more lines. In this case it has to be kept in mind that the lines of equations are processed as a stack, i.e. starting with MA1 =
st
1
line to MA5 = 5th line.
Please note:
That formula set is valid which is shown prior to
leaving the MATH menu.
7.2 Edit
”Edit“ opens the ”Mathematics Edit“ submenu.
7.2.1 Equation 5 equations may be selected with the INTENS knob
. Each equation consists of the name of a result (e.g. MA5), the = sign, the function (e.g. ADD) and (fi rst operand, second ope­rand). Remark: The second operand will not be displayed with all functions.
7.2.2 Function By using the INTENS knob
, the following functions may be
selected:
ADD: Operand 1 + operand 2 SUB: Operand 1 – operand 2 MUL: Operand 1 times operand 2. DIV: Operand 1 divided by operand 2. SQ: Operand 1 squared. INV: Changes sign of operand 1. 1/: Calculates 1/operand 1 (reciprocal value). ABS: Takes absolute value of operand 1 (removes sign) POS: Only values of operand 1 > 0 are displayed, < 0 will
not be displayed. NEG: Only values of operand 1 < 0 are displayed, > 0 will not be displayed.
In order to leave this mode enter the MODE menu and select ”Auto“ or ”Normal“.
6.2.1 DSO mode, ending or interrupting a capture. Any capture in progress may be stopped by pressing RUN/ STOP.
6.2.2 DSO mode, reset function. Pressing RUN/STOP twice in ”env“ (Envelope) or ”avg“ (Average) mode, where the signal display is the result of more than one signal data acquisition, causes the previous signal capture results to be deleted and a new start of the signal capture.
MATH
Only available in DSO mode.
This pushbutton calls the ”Mathematics“ menu and the ”Formu­la editor“. With the ”Mathematics“ menu stored signals may be mathematically processed, either directly or with reference to other stored signals. The results may be graphically displayed on the screen and determined by the cursor functions.
7.2.3 Operand 1 The INTENS knob
allows the selection of these signals as
operands:
CH1: CH1 signal. CH2: CH2 signal. RE1: Signal from reference memory 1 RE2: Signal from reference memory 2 MA1: Result of equation named MA1. MA2: Result of equation named MA2 MA3: Result of equation named MA3. MA4: Result of equation named MA4 MA5: Result of equation named MA5.
The next step after MA5 causes the display EDIT. Calling EDIT opens a sub submenu described under item 7.2.5.
7.2.4 Operand 2: The INTENS knob
allows the selection of the same operands as mentioned under item 7.2.3 (Operand 1) if ADD, SUB, MUL or DIV function is chosen. The next step after ”MA5“ causes the display EDIT. Calling EDIT opens a sub submenu described under item 7.2.5.
Subject to change without notice
27
Controls and Readout
POWER
FOCUS TRACE
MENU
REM
4
ANALOG
DIGITAL
OSCILLOSCOPE
HM1008
1 GSa
100 MHz
1 2 3
INTENS
POWER
!
EXIT MENU
REMOTE OFF
7.2.5 Edit > Constant Edit In the CW position of the INTENS knob
an additional item ”Edit“ and an arrow symbol is displayed. Pressing the associated function button opens the sub submenu ”Constant Edit“. This enables to choose a Number, its ”Dec.Point“ (decimal point) and its ”Prefi x“ of the ”Unit“, all with the INTENS knob
. The unit
need not be selected it is only shown to assist the memory.
7.3 Display
Attention!
The display of mathematic signal(s) automatically
switches previously displayed reference signals off and vice versa.
The function ”Display“ is available twice in the menu and may be switched on or off in any combination. This allows to display: no result, one result, the result of two equations as signals. The displays will come forward upon leaving the ”Mathematics“ menu. Also the designation of the equation (e.g. MA2) will be shown. Select the equations to be displayed with the INTENS knob
.
The mathematics signal is automatically scaled, this is inde­pendent of the graticule, of Y and time base parameters, the scale will not be shown. Hence the measurement of the signal amplitudes must be performed using the CURSOR (V to GND) after the ”reference“ (e.g. MA2) of the CURSOR to the ”mathe­matics signal“ and its scale was established (AUTO/CURSOR MEASURE
pushbutton > Cursors > reference > e.g. MA2).
The readout may then display e.g.: ”V(MA2): 900 mV“.
Divisions by zero will be ignored and an error message dis­played.
7.4 Units
Each function ”Display“ will be associated with a function ”Unit“ which can be selected with INTENS
and will be attached to
the result.
ACQUIRE
Available only in DSO mode.
This pushbutton opens the menu ACQUIRE which offers these modes:
8.1 Refresh capture/display.
In this mode repetitive signals may be captured and displayed much like in analog mode. The readout will show ”rfr“. Signal acquisition may be stopped or started with the RUN/STOP push­button. The STOP pushbutton will be illuminated in stop.
A trigger will start a new acquisition which will overwrite the display of the former. The display will remain on screen until the next acquisition. This mode is available over the full time base range (50 s/cm to 5 ns/cm).
5 6 7 8 9 10 11 12
· 1 MB
ANALOG
DIGITAL
RUN ACQUIRE SETTINGS HELP
STOP
MATH
SAVE/
RECALL
AUTOSET
Please note:
After changing the time base acquisition and display
will always start fi rst at the trigger position which is undelayed (Readout: ”Tt:0s“) on the screen centre. With the 2
nd
acquisition the display will start at the screen left. In most cases this is meaningless, but the scope may seem not to react at slow time base settings combined with long Posttrigger times.
In the utmost left position of the trigger point selec-
ted with the HORIZONTAL control
the readout will indicate ”Tt:1.85ks“ for a time base setting of 50 s/cm. This means that 1,600 seconds must elap­se until the trace will become visible at the screen left, after another 250 s it will have reached the screen centre (1,600 s + 250 s = 1.85 ks).
8.2 Envelope capture/display
Envelope is a special mode in refresh mode, the readout will show ”env“. Also in this mode there must be suffi cient signal for triggering.
In contrast to the refresh mode the results of several captures will be examined and the maxima and minima stored, the enve­lope of the signal will then be displayed if it changes in amplitude or/and frequency. Also any jitter will be shown.
Also in this mode pressing RUN/STOP
will stop the acquisi­tion, indicated by STOP illuminated. Af ter pressing the pushbut­ton again the formerly stored signals will be erased and the en­velope calculation star ts anew. In order to prevent an accidental turning on of this mode operating any control which infl uences the signal display will automatically switch envelope off.
Because this mode requires many signal repetitions and acqui­sitions it is not compatible with single sweep/acquisition. AUTO or normal trigger modes must be selected.
8.3 Average mode capture/display
Also this is a special mode within the refresh mode. Also here signal repetitions are needed.
The weighting of each acquisition can be selected with ”Ave­rage“ in the menu, any number between 2 and 512 may be chosen using the INTENS knob
. The readout will show e.g.
”avg#512“.
The higher the number of acquisitions averaged the lower the contribution of a single acquisition will be and the longer the averaging will take. Averaging is a means to increase the ac­curacy inspite of the 8 bit converters, it is an exchange of time against accuracy. Noise will be reduced by averaging.
The same holds as for envelope: the acquisition may be stopped by pressing the RUN/STOP pushbutton, STOP will be illumina­ted. Pressing RUN/STOP again will restart. In order to prevent
28
Subject to change without notice
Controls and Readout
an inadvertent entering of this mode the operation of any control will automatically cause this mode to be reset.
As repetitive acquisitions are needed for calculation of an ave­rage single sweep will not be compatible.
8.4 Roll mode capture/display
Roll mode means that the signal(s) will be continuously acquired without the need for a trigger. Hence all controls, displays and readouts for the trigger and ZOOM will be disabled. The readout will show ”rol“. The result of the last acquisition will be displayed at the right hand edge of the graticule, all formerly acquired signals will be shifted one address to the left. The result at the left hand screen edge will be dropped. There is no waiting for a trigger and thus the hold-off time is minimum. As in any other mode the signal acquisition may be stopped and restarted any time with the RUN/STOP pushbutton. In roll mode the time base available is limited to 50 s/cm to 50 ms/cm. Faster time bases do not make sense as the signal could not be observed any more. In case the time base was set outside the limit cited it will be automatically changed to the next value within the limits upon entering this mode.
8.5 Peak Detect Auto Off
On Peak Detect Auto mode this mode of acquisition will be au­tomatically selected if YT and time base speeds of 20 ms/cm to 2 ms/cm were chosen. This mode is only available with: Refresh, Envelope, Average, and Single sweep. The readout will show PD preceding the symbol for the operating mode selected.
At slow sweep speeds the sampling rate will be low, i.e. there are relatively large gaps between samples. Glitches occurring during those gaps will go by unnoticed. With Peak Detect, how­ever, the highest sampling rate will be used such that glitches will be caught. The samples will be looked at and the ones with the highest values stored and displayed.
8.6 Random Auto Off
Provided single sweep was not selected Random Sampling will be automatically selected beginning at a certain sweep speed. The time base setting will be indicated in the readout, e.g. ”RS: 10 GSa“ (= Random sampling with 10 GS/s effective sampling rate), the real time time base speed will be 5 ns/cm. Without Random Sampling ”RS“ Real Time Sampling will be used with a maximum sampling rate of 1 GS/s (one channel only) or 500 kSa/s (two channel mode).
9.1 Analog and DSO modes
Under ”Save/Recall“ the current instrument settings may be saved or settings saved earlier recalled. There are 9 nonvolatile memories available.
9.1.1 Saving the actual settings Upon pressing the pushbutton the submenu ”Front Panel Save“ will be accessed. A memory number will be offered (1 to 9) which can be changed by the INTENS knob
. By pres­sing ”Save“ all settings will be saved in the memory location the number of which was selected.
9.1.2 Recall the actual settings In the submenu ”Front Panel Recall“ a memory number will be displayed (1 to 9) which can be changed with the INTENS knob
. Pressing the pushbutton ”Recall“ causes loading of
the settings stored.
9.2 DSO mode The menu options described in 9.1.1 and 9.1.2 are also available in DSO mode. Additionally, the menu options ”Reference Save“ and ”Reference Display“ will be available. A reference is a signal which was resp. can be stored away for later reference to it 9 nonvolatile memory locations are provided.
9.2.1 Reference Save
9.2.1.1 Source x
In the submenu the source can be selected with the INTENS knob
. Signals from the logic inputs CH3 and CH4 can not be
stored as references.
9.2.1.2 Destination RE x There are 9 memory locations available into which reference signals from the source selected before can be stored. Use the INTENS knob
for selection.
9.2.1.3 Save Pressing ”Save“ will store the signal from the source selected into the memory selected.
9.2.2 Reference Display
Attention!
The display of reference signals automatically swit-
ches previously displayed mathematic signal(s) off and vice versa.
Random Sampling requires repetitive signals, each signal peri­od will contribute one sample. At an effective sampling rate of 10 GS/s the time difference from sample to sample along the signal period will be 0.1 ns. However, note that with random sampling the samples are not taken in sequence along the sig­nal period but randomly with respect to it. Random Sampling allows to generate the 200 points per cm in X direction at the fastest time base of 5 ns/cm.
Remark: 5 ns/cm are also available in other modes. In Real Time Sampling mode and 1 GS/s on one channel each 1 ns a sample is taken, hence at 5 ns/cm there are 5 points per cm. The ”missing“ 195 points are generated by interpolation using sin x/x.
SAVE/RECALL
This pushbutton will open up a menu. The number of choices in this menu is dependent upon whether there is analog or DSO mode.
9.2.2.1 RE x, On Off, associated settings When in this submenu using the INTENS knob will allow selection of 2 reference signals which can then be displayed alongside with 2 input signals.
9.2.2.2 RE x After calling this function the memory location can be selected with the INTENS knob
. (RE 1 to 9)
9.2.2.3 On Off With the pushbutton on/off control is possible. When the contents of the reference memory are displayed, the memory number is indicated with RE x (x = 1 to 9) at the right hand screen side. Switching to ”on“ will produce another menu item (”Assoc.Set“).
Please note:
If both reference displays are ”on“ and if both
memory locations are identical (e.g. RE1, RE1) the signal will be displayed twice on the same spot.
Subject to change without notice
29
Controls and Readout
POWER
FOCUS TRACE
MENU
REM
4
ANALOG
DIGITAL
OSCILLOSCOPE
HM1008
1 GSa
100 MHz
1 2 3
INTENS
POWER
!
EXIT MENU
REMOTE OFF
9.2.2.4 Associated settings If the pushbutton labelled ”Assoc.Set.“ is depressed the readout will show all oscilloscope settings which were loaded into the reference together with the signal. The signal parameters may be recalled.
SETTINGS
Pressing this pushbutton will open the SETTINGS menu which offers the following submenus:
10.1 Language
In thus submenu the language can be selected: English, Ger­man, French are available for choice.
10.2 Misc (Miscellaneous)
10.2.1 Contr.Beep On Off
Switches the acoustical signal on or off, which informs about CW or CCW positions of knobs.
5 6 7 8 9 10 11 12
· 1 MB
ANALOG
DIGITAL
RUN ACQUIRE SETTINGS HELP
STOP
MATH
SAVE/
RECALL
AUTOSET
10.4.3 Optimum display In this mode minimum and maximum signal values acquired are taken into consideration and displayed in Vectors mode. This makes sense as up to 1 MByte samples may be acquired, but only up to 2 KBytes per channel can be displayed, hence possibly min or max values may not be shown. Otherwise the display is as described above.
AUTOSET
Choosing AUTOSET will cause an automatic instrument setting, dependent upon the signal proper which selects positions, signal amplitude and time base for a reasonable display. The choice of analog or DSO mode will not be affected. In component test mode (available only in analog mode), XY mode, or ADD auto­matically dual channel mode will be selected. If dual channel or CH1 or CH2 were previously chosen this will remain.
The DSO modes Roll, Envelope or Average will be changed to refresh mode.
10.2.2 Error Beep On Off Will turn the acoustical error signal on or off.
10.2.3 Quick Start On Off In off the HAMEG logo, the type and the version number will not be shown, the instrument will be ready immediately.
10.2.4 Menu Off time With the INTENS knob
the time of menu display may be de-
termined. EXIT MENU will terminate a menu on the spot.
In ”Man.“ mode the menu can be left:
– by pressing EXIT MENU. – by pressing another pushbutton. – by pressing the same pushbutton again with which the menu
was called.
10.3 Interface
This menu shows the interface parameters which can be selected as usual.
10.4 Display
This submenu offers several modes of display:
10.4.1 Dots In this mode the samples are shown as what they are, i.e. dots (points). This representation is valuable for judging whether enough samples were gathered in order to reconstruct a signal suffi ciently.
10.4.2 Vectors In this mode the sampling points are interconnected by drawing straight lines. If there are only few samples sin x/x interpolation is used to ”create“ intermediate points which are then joined by straight lines.
AUTOSET will further set the intensity to an average value if it was set too low. If a menu was opened it will be turned off by AUTOSET. During the display of HELP texts AUTOSET is not available.
HELP
Pressing the HELP pushbutton will turn the signal display off and display the help text.
If a menu was opened the help text will refer to this menu resp. to the special menu or submenu option selected. If a knob is moved another help text referring to this knob will appear. Press HELP again to deactivate the text.
POSITION 1 (knob)
This knob can assume various functions which depend upon the operating mode, the functions selected with the CH1/2–CUR­SOR–MA/REF–ZOOM-pushbutton selected.
13.1 Y position
13.1.1 Y position CH1 (Analog and DSO modes)
POSITION 1 will set the Y position of CH1 provided there is YT mode and the pushbutton CH1/2–CURSOR–MA/REF–ZOOM­pushbutton
is not illuminated.
13.1.2 REF (reference) signal position (DSO mode only) The POSITION 1 control functions as the position control for the signals stored in the reference memory if these conditions are fulfi lled:
st
1
A reference signal must be on display (SAVE/RECALL
> Reference Display > (upper display area) REx (x = num-
and the menu option
30
Subject to change without notice
Controls and Readout
POWER
ber of memory location, select with INTENS) > On (with or without associated settings).
nd
After pressing CH1/2–CURSOR–MA/REF–ZOOM-push-
2
button > Math./Ref. was selected, the pushbutton will light up green.
13.1.3 Mathematics signal position (DSO mode only) The POSITION 1 control will assume the function of position con­trol for mathematics signals after the following procedure: Press the MATH pushbutton > Display (upper display area), select an equation with the INTENS knob the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
(MA1.. MA5), press
, select
”Math./Ref.“ The pushbutton will light up green.
13.1.4 Y position of 2
nd
time base B (Analog mode) The POSITION 1 control will assume the function of Y position control of the signal displayed by time base B in alternate time base mode after the following procedure. This is convenient in order to be able to separate the displays of the (same) signal with both time bases on the screen. Press the HOR
push­button > ”Search“. Press the CH1/2–CURSOR–MA/REF–ZOOM­pushbutton
, select the function ”TB B“. The pushbutton will
light up green.
13.1.5 Y position for ZOOM (DSO mode.) This is the equivalent of the former in DSO mode, where the function is called ZOOM. The POSITION 1 control will assume the function of Y position control of the zoomed signal after the following procedure. Again, the intent is to be able to separate the two displays of the (same) signal on the screen in alternate time base mode. Press the HOR
pushbutton > ”Search“. Press the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton select ”TB B“. The pushbutton will light up green.
select ”Cursors“ or ”Cur. Track“. The push-button will light up in blue.
Please note:
The function ”Cur. Track“ is only available if two
cursors are indeed displayed, then both cursors can be moved simultaneously (tracking) without a change of their respective positions.
POSITION 2 (knob)
Also this control may assume diverse functions dependent on the operating mode, the function selected via the CH1/2–CUR­SOR–MA/REF–ZOOM-pushbutton vated.
14.1. Y position
14.1.1 Y position CH2 (Analog and DSO modes)
POSITION 2 will function as Y position control of CH2 in Yt mode and if the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton not illuminated.
14.1.2 Y position of reference signals (DSO mode) The POSITION 2 control will function as Y position control of reference signals if the following conditions are fulfi lled:
– A reference signal must on display. (SAVE/RECALL push-
button
> Reference Display > (upper display area) Rex (x = number of memory location, select with INTENS) > On (with or without associated settings).
– Press CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
,
> Math./Ref. The pushbutton will light up green.
and the menu item acti-
is
13.2. X position in XY mode (Analog and DSO modes)
POSITION 1 will function as X position control of CH1 in XY mode and provided the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
is not illuminated.
Note:
The HORIZONTAL control
will also be functional
in XY mode.
13.3. CURSOR position (Analog and DSO modes)
The POSITION 1 control will function as Y position control of the cursors if the following conditions are met: The CURSOR display must have been activated (AUTO/CURSOR-MEASURE pushbutton
>Cursors >Cursors On pushbutton depressed).
Press the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
VOLTS / DIV
SCALE · VAR
15
13
14
17
16
18
POSITION 1 POSITION 2
VOLTS / DIV
SCALE · VAR
20 V 1 mV 20 V 1 mV
CH 1
VAR
14.1.3 Y position of mathematics signals (DSO mode) The POSITION 2 control functions as Y position control for ma­thematics signals, if the following conditions are fulfi lled: Press the MATH pushbutton select an equation with the INTENS knob
> displays (upper display area),
(MA1.. MA5). Press the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton Math./Ref. The pushbutton will light up green.
14.2 Y position of CH2 in XY mode (Analog and DSO modes.) POSITION 2 will function as the Y position control of CH2 in XY mode provided the CH1/2–CURSOR–MA/REF–ZOOM-push­button
is not illuminated.
14.3 Y position of cursors (Analog and DSO modes) The POSITION 2 control will function as Y position control of
,
LEVEL A/B
TRIGGER
MODE
FILTER
SOURCE
CH 2 HOR MAG
VAR
AUX
X-POS
DELAY
TRIG ’d
NORM
HOLD OFF
HORIZONTAL
TIME / DIV
SCALE · VAR
50s 5ns
VAR
x10
, select
19
26 27
20
23
21
24
28
22
25
29
30
Subject to change without notice
31
Controls and Readout
POWER
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
VOLTS / DIV
SCALE · VAR
CH 2 HOR MAG
VAR
15
13
14
17
16
18
POSITION 1 POSITION 2
VOLTS / DIV
SCALE · VAR
20 V 1 mV 20 V 1 mV
CH 1
VAR
the cursors if the following conditions are met: The cursors must be activated (Press AUTO/CURSOR-MEASURE pushbutton
>Cursors >Cursors On), press the CH1/2–CURSOR–MA/
REF–ZOOM-pushbutton
select ”Cursors“ or ”Cur. Track“.
The pushbutton will light up blue.
Note:
The function Cur. Track (cursor tracking) is only
available if 2 cursors are on display. The cursors will then be moved simultaneously (tracking) wit­hout changing their respective positions.
CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
19
26 27
20
23
21
24
28
22
25
29 30
LEVEL A/B
TRIGGER
MODE
FILTER
SOURCE
AUX
TRIG ’d
NORM
HOLD OFF
X-POS
DELAY
50s 5ns
VAR
HORIZONTAL
TIME / DIV
SCALE · VAR
x10
>“ to the sensitivity indication (”CH1>5mV..“) in order to show that the sensitivity is now uncalibrated. The results of cursor measurements will be fl agged accordingly.
In this mode the sensitivity can be changed with the VOLTS/ DIV–SCALE–VAR control knob from 1 mV/cm to > 20 V/cm.
16.3 SCALE (DSO mode) The display height of a mathematic signal resp. a refe­rence signal can be changed after the function „Math./Ref.“ is selected in the „Pos./Scale“ menu. The latter is called by pressing the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
.
This pushbutton calls a menu which allows to select the func­tions of the controls POSITION 1 and 2. With mathematics functions this will also affect in some way the SCALE function which is changed by the VOLTS/DIV controls.
The pushbutton will signal the function activated corresponding to the front panel labelling:
dark: Y position CH1 and CH2. blue: Y position of cursors. green: Y position of:
– Mathematics signal(s) – Reference signal(s) – ZOOM or time base B display of signal(s)
VOLTS/DIV–SCALE–VAR knob
This knob is a double function CH1 control.
16.1 Selection of vertical sensitivity
Provided VAR on pushbutton CH1 sitivity will be calibrated. Turning the control CCW will decrease, turning it CW will increase the sensitivity. 1 mV/cm to 20 V/cm can be selected in a 1-2-5 sequence. The readout will display the calibrated sensitivity (e.g. ”CH1: 5mV..“). Depending on the sensitivity selection the signal will be shown with smaller or greater amplitude.
Please note:
This sensitivity selection is always active, e.g. also,
if CH2 only was chosen. In that case CH1 may be used as trigger source.
16.2 Variable control
Select this function with the CH1 pushbutton VAR on the CH1 pushbutton will light up. The readout will add a
is not illuminated the sen-
>Variable On.
VOLTS/DIV–SCALE–VAR knob
This double function control belongs to CH2.
17.1 Selection of vertical sensitivity
The sensitivity will be calibrated provided VAR on the CH2 pushbutton will decrease, turning it CW will increase the sensitivity. The sensitivities can be selected from 1 mV/cm to 20 V/cm in a 1-2-5 sequence. The readout will show the sensitivity (e.g. ”CH2:5mV..“). Depending on the sensitivity the signal will be displayed with smaller or greater amplitude.
Please note:
The sensitivity control is always active, i.e. also if
17.2 Variable control
The variable control is activated by pressing the CH2 push­button up. The readout will add the ”>“ preceding the sensitivity (e.g. ”CH2 >5mV..“) in order to indicate that the sensitivity is uncalibrated. The results of cursor measurements will be labelled accordingly.
In this mode the sensitivity may be changed continuously with the VOLTS/DIV–SCALE–VAR control from 1 mV/cm to > 20 V/cm.
17.3 SCALE (DSO mode)
The display height of a mathematic signal resp. a refe­rence signal can be changed after the function „Math./Ref.“ is selected in the „Pos./Scale“ menu. The latter is called by pressing the CH1/2–CURSOR–MA/REF–ZOOM-push­button
is not illuminated. Turning the control CCW
CH2 is not selected. CH2 may then still be used e.g. as a trigger source.
>Variable On. VAR on the pushbutton will light
.
32
Subject to change without notice
Controls and Readout
AUTO/CURSOR MEASURE pushbutton
Pressing this pushbutton will open the menu ”Measurement“ which offers the submenus > ”Cursors“ and ”Auto“.
If the submenu Cursors was selected and a measuring mode ”Cursors On“ must be activated too, in order to generate the cursor line(s) on the screen. The measurement result will be shown in the readout!
Please note:
In order to move the cursors it is necessary to press
the pushbutton CH1/2–CURSOR–MA/REF–ZOOM and look for the menu ”Pos./Scale“. In this menu the selection of ”Cursors“ (long lines) or ”auxiliary cursors“ (short lines) or other symbols will deter­mine which cursor lines/symbols can be moved by the POSITION 1 and 2 controls.
18.1 Cursors (Analog and DSO modes)
Depending on the operating mode (Yt or XY) this submenu will offer various cursor measuring functions which will affect as well the cursor lines as their position.
18.1.1 Cursors On Off With ”Cursors On“ the CURSORS and the results of cursor measurements will be displayed by the readout in the top right corner of the screen. (e.g. ΔV(CH2):16.6 mV). If a variable con­trol was activated, the readout will indicate this by replacing the ”:“ by a ”>“.
18.1.4 Respect It may be necessary to determine for which signal resp. channel the CURSOR measurement shall be valid. This is signalled by showing the INTENS knob
signal next to the channel number. After selection of the signal source, the CURSOR lines must then be positioned to the signal or portions of it displayed by this channel.
18.2 Auto (Analog mode) Depending on the operating mode this submenu offers various automatic measurements of the trigger signal. The following conditions must be met:
a) For frequency or period measurements suffi cient trigger
signals must be available. Use normal triggering for signals < 20 Hz. Please note that signals of very low frequency may require seconds to complete one measurement.
b) In order to measure DC or the DC content of a signal the
input channel as well as the trigger must be set to DC coup­ling.
Further notes:
– Due to the limited frequency response of the trigger channel
the accuracy will decrease with increasing frequency.
– The frequency responses of the vertical channel and the
trigger channel differ substantially, this may affect the signal display.
18.1.2 Meas. Type If this function is activated one of the measurement modes of­fered may be selected with the INTENS knob
. In most cases
the unit going with a mode will be shown also.
18.1.3 Unit In the modes ”Ratio X“ and ”Ratio Y“ the INTENS knob symbol
will be shown in addition to a unit, this may then be used to
change the unit.
” rat ” (ratio), display of ratios
In this mode the ratios of duty cycles or amplitudes may be determined with the CURSORS. The distance between the long CURSOR lines is equal to 1.
” % ” (percent), display of percentages
The distance between the long CURSOR lines is equal to 100 %. The result will be determined by the distance of the short auxiliary cursor line to the long reference line (lower resp. left), if appropriate with a negative sign.
” ° ” (degree), measurement of degrees
The distance between the long CURSOR lines is equal to 360 degrees and must be exactly as long as a signal period. The measurement result will be determined from the distance between the reference line to the short auxiliary cursor line. If appropriate with a negative sign. For further information please consult ”Measurements of phase differences in dual channel mode (Yt)“ in the section ”First time operation and presettings“.
π One period of a sine wave is equal to 2 π, hence the distance between the two long CURSOR lines must be set to one pe­riod. If the distance between the reference line and the short CURSOR line equals 1.5 periods, ”3 π“ will be displayed. If the short cursor line is left of the reference line a negative sign will be shown.
– When measuring very low frequency signals the display will
follow the signal.
– When measuring pulse signals there may be an infl uence
of the duty cycle or the slope selected on the accuracy of the result.
– The signal must remain within the graticule area, i.e. the
input amplifi ers must not be overdriven.
Please note:
Complex signals should be measured using the
CURSORs.
18.2.1 Auto On Off
If Auto is On the result of the automatic measurement will be shown in the readout in the top right corner. (e.g. DC(Tr):100μV. (Tr) points out that it refers to the trigger signal. Sometimes a ”?“ will be displayed, this indicates that there is no or an insuffi cient signal.
If a variable is activated und thus the sensitivity or time base uncalibrated the ”:“ will be replaced by a ”>“ symbol.
18.2.2 Meas. Type (measurement type). The measurement type can be selected with the INTENS knob
.
18.2.3 Respect Tr indicates that the measurement is done with respect to the trigger signal. If e.g. the CH1 signal is used for triggering, the result will be with respect to that signal.
18.3 Auto (DSO mode) Depending on the operating mode this submenu offers various automatic measurements. The display in the ”reference“ fi eld will signal to which source the result belongs. (Tr = trigger signal or the signal of the channel shown.)
Subject to change without notice
33
Controls and Readout
POWER
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
VOLTS / DIV
SCALE · VAR
CH 2 HOR MAG
VAR
15
14
16
18
13
17
POSITION 1 POSITION 2
VOLTS / DIV
SCALE · VAR
20 V 1 mV 20 V 1 mV
CH 1
VAR
See the hints in 18.2 regarding period (trigger period) or fre­quency (trigger frequency) measurements.
For voltage measurements the result will be calculated from the displayed signal data and the signal data selected with ”Respect“.
Please note:
Complex signals should be measured with CURSOR.
18.3.1 Auto On Off
If Auto is On the readout will display the result of the automatic measurement in the top right corner. (e.g. DC(CH2):100μV). (CHx) denotes the source channel, (Tr) the trigger signal. Sometimes a ”?“ may be displayed, if there is no signal or an insuffi cient one.
If a variable is activated and the sensitivity or time base thus uncalibrated the ”:“ will be replaced by a ”>“ symbol.
18.3.2 Meas. Type (measurement type) If this function is activated the mode can be selected with the INTENS knob
, the display in the ”Respect“ fi eld will change
accordingly.
19
26 27
20
23
21
24
28
22
25
29 30
LEVEL A/B
TRIGGER
MODE
FILTER
SOURCE
AUX
TRIG ’d
NORM
HOLD OFF
X-POS
DELAY
VAR
HORIZONTAL
TIME / DIV
SCALE · VAR
50s 5ns
x10
in those modes where there is no direct relationship between trigger signal and trigger point.
In normal trigger mode the LEVEL control will move the trig­ger symbol anywhere. In automatic peak-to-peak detection mode the level can only be selected between the signal’s peak values.
The movement of the trigger symbol is vertical only. The range of this symbol is limited in order to prevent that this symbol will overwrite other readout information. As soon as the trigger sym­bol leaves the graticule its form will change, this change signals in which direction the trigger symbol left the graticule.
Analog mode only: Depending on the time base mode the LEVEL control will affect the time base A or B triggering. Press the HOR pushbutton
in order to select the time base mode in the ”Time base“ menu. In ”Search“ mode (alternate time base mode) the last trigger level setting for time base A will remain valid (graticule left) if time base B is switched to triggered mode. (Menu ”Time base“: set B trigger to positive or negative slope). Thereafter the LEVEL A/B control will control the time base B trigger, a second trigger point symbol will be shown and marked with ”B“.
In one channel only mode the active channel will be indicated. In those modes where 2 channels are activated the INTENS knob symbol
will be shown. After pressing the ”Reference“ pushbutton the reference channel may be selected with the INTENS knob
.
18.3.3 Respect Tr indicates that the measurement is done with respect to the trigger signal. If e.g. the CH1 signal is used for triggering, the result will be with respect to that signal.
In case of voltage measurement is done with respect to the active channel. In those modes where several channels are activated, the INTENS knob symbol
indicates that different
channels can be selected.
LEVEL A/B knob
The LEVEL control allows to set the trigger level, i.e. the voltage resp. signal level, which will generate a trigger to start the time base whenever the signal passes that level. In most Yt modes the readout will show a symbol the vertical position of which indicates the signal point which will trigger. The trigger symbol will be ”parked“ on the second graticule line from the bottom
34
Subject to change without notice
MODE pushbutton
Pressing this pushbutton will open the ”Trigger“ menu, where Auto, Normal, Single (sweep triggering) can be selected. Choo­sing ”Slope“ will allow to trigger on any signal shape. For video signals select ”Video“ and press the FILTER pushbutton order to fi nd a choice of special trigger modes for composite video signals.
In XY mode the pushbuttons MODE
, FILTER and SOURCE
are disabled as there is no triggering in XY mode.
20.1 Auto (trigger)
Automatic triggering (Auto) is active if the NORM display not illuminated. In ”Auto“ the analog time base resp. signal capture (DSO mode) will be periodically started even if there is no signal or when no triggers are generated because the set­tings are incorrect. Signals of < 20 Hz can not be triggered as the automatic start will have occurred before the signal arrived. Automatic triggering is possible with or without peak detection. The LEVEL A/B
control will be active in both modes.
In peak detection mode the range of the level control is limited to the peak-to-peak voltage of the signal. Without peak detection
in
is
Controls and Readout
any level can be set. If the trigger level is set such that no trig­gers are generated the automatic triggering will nevertheless start the time base. The signal will thus remain visible but will be untriggered.
Whether peak detection is active or not depends on the mode and the settings in ”FILTER“ (trigger coupling). The mode active will be shown by the behaviour of the trigger point symbol when turning the LEVEL knob.
20.2. Normal (trigger)
If the NORM-LED ted.
In normal trigger mode both the peak detection and the au­tomatic time base start will be disabled. Hence if there is no suffi cient trigger signal the screen will remain dark in analog mode. In DSO mode signal capturing will also stop unless the roll mode was selected. In this mode there is no lower frequency limit for signals.
20.3 Single (sweep/capture)
In single sweep/capture mode the time base selected will accept only one trigger for one sweep/capture after it was armed. The NORM-LED will light up, Auto triggering is disabled.
For further information about the precise operation see RUN/ STOP pushbutton
lights up normal triggering was selec-
description.
FILTER pushbutton
After this pushbutton is depressed it will depend on the settings chosen in MODE In XY mode the pushbuttons: MODE
are disabled as XY displays can not be triggered.
21.1 Menu: Slope
The menu ”Edge“ will appear if ”Edge“ was selected in the TRIGGER menu to be called with MODE the FILTER tion see ”Trigger coupling“ (Menu FILTER) under the heading ”Triggering and time bases“ and the instrument specifi cations. The following settings are available:
21.1.1 Trig. Filter – AC: The trigger signal is AC coupled via a large capacitor in
order to reach a low cut-off frequency.
Readout: ”Tr:Source, Slope, AC“
DC: The trigger signal is DC coupled. No peak triggering is
possible.
Readout: ”Tr: Source, Slope, DC“
HF: AC coupling with a small capacitor suppressing low
frequency signals. Hence the signal display and the trigger signal derived are not any more identical, the trigger point symbol will be ”parked“ in DSO mode and will not react to the LEVEL A/B symbol is switched off. As a combination of HF coupling and LF or Noise Reject is not meaningful both menu options will not be shown.
Readout: ”Tr:Source, Slope, HF“.
LF: The trigger signal is sent through a low pass in order to
suppress high frequency components. As this will already suppress hf the noise rejection mode will be set to OFF automatically.
Readout: ”Tr:Source, Slope, AC or DC, LF“.
(Edge or Video) which menu will be offered.
, FILTER and SOURCE
pushbutton and after
pushbutton was depressed. For further informa-
control. In analog mode the trigger point
Noise Reject: Noise rejection (reduction) means a reduced
trigger amplifi er bandwidth and consequently less trigger signal noise.
Readout: ”Tr:Source, Slope, AC or DC, NR“.
21.1.2 Slope ”SLOPE“ determines whether the rising or falling portion of a signal shall trigger, the level is set with the LEVEL A/B control.
In BOTH MODE both slopes will trigger, this is also true in single sweep mode. This allows e.g. the display of eye diagrams.
21.2 Menu: Video
In order to reach the menu VIDEO proceed as follows: Press MODE press the FILTER found under VIDEO (tv signal triggering) in the chapter Trigge­ring and time bases“ and in the instrument specifi ca-tions. The following settings are available:
21.2.1 Frame, Line. Depending on the setting chosen triggering will be on frame or line sync pulses. The selection will also affect other menu items.
Readout: ”Tr:Source, TV“.
21.2.1.1 Frame
ALL: In this mode the sync pulses of each half frame can
trigger.
Even: In this mode only the sync pulses of even half frames
can trigger.
Odd: In this mode only the sync pulses of odd half frames
can trigger.
21.2.1.2 Line. – All: In this mode all line sync pulses can trigger. – Line No: The line number with its line pulse that is used for
triggering can be selected with the INTENS knob
Line min: One pushbutton operation will be suffi cient to
switch back to the lowest possible line number.
21.2.2 Norm The pushbutton allows the selection of the US standard of 525 lines and 60 Hz or the European standard with 625 lines and 50 Hz. With any change of standard the line number will be automatically changed, too.
21.2.3 Polarity Composite video signals may have both polarities. Selection of the right polarity is vital as the scope should be triggered by the sync pulses and not the video content.
Positive polarity is defi ned by the video content being more positive than the sync signals and vice versa.
If the polarity was wrongly selected there will be no triggering at all, an untriggered display or no signal capture.
to open the TRIGGER menu, select VIDEO, then
pushbutton. Further information can be
.
SOURCE pushbutton
Depressing this pushbutton will call various menus depen­ding on the previously selected mode (MODE EDGE, VIDEO, LOGIC. In XY mode the pushbuttons: MODE FILTER triggered. In the TRIGGER SOURCE menu the source is selected from which the trigger signal is to be taken. The options depend on the actual mode of the scope.
, SOURCE are disabled as XY displays can not be
pushbutton):
,
Subject to change without notice
35
Controls and Readout
POWER
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
VOLTS / DIV
SCALE · VAR
CH 2 HOR MAG
VAR
15
14
16
18
13
17
POSITION 1 POSITION 2
VOLTS / DIV
SCALE · VAR
20 V 1 mV 20 V 1 mV
CH 1
VAR
22.1 Edge-/Video-Trigger
22.1.1 CH1
Conditions: Analog or DSO mode, EDGE or VIDEO selected. CH1 will then be the trigger source, no matter whether it is displayed or not. Readout: ”Tr:CH1, (Slope), Filter (TV)“.
22.1.2 CH2 Conditions: Analog or DSO mode, EDGE or VIDEO selected. CH2 will then be the trigger source, no matter whether it is displayed or not. Readout: ”Tr:CH2, (Slope), Filter (TV).“
22.1.3 Alt. 1/2 Conditions: Analog mode, EDGE-Triggering. Alternate triggering with the signals from channels 1 and 2 as described in the section ”Alternate Trigger“ of chapter ”Triggering and time bases“. Please note that in this trigger mode the appa­rent time relationships between the two signals on the screen are meaningless and misleading, the relative position of the two signals depends only on their shape and the trigger level selected.
In dual channel mode (DUAL) alternate triggering is only pos­sible in conjunction with alternate dual channel operation. If previously dual trace chopped mode was selected (VERT/XY pushbutton) > DUAL chop) it will be automatically changed to alternate mode when alternate triggering is selected. After ”Alt. 1/2“ is turned off dual trace chopped mode may be selected again. Readout: ”Tr:alt, Slope, Filter“.
22.1.4 External In this mode the trigger signal comes from AUXILIARY INPUT
). Readout: ”Tr:alt, Slope, Filter“.
LEVEL A/B
TRIGGER
MODE
FILTER
SOURCE
AUX
DELAY
TRIG ’d
NORM
HOLD OFF
X-POS
50s 5ns
VAR
HORIZONTAL
TIME / DIV
SCALE · VAR
x10
HOLD-OFF display (Analog mode only)
This display will light up if the hold-off time was set to > 0% in order to indicate that the longer than minimum hold-off time may cause a lower rep rate of the time base and thus a darker display. Setting the hold-off time requires pressing the HOR pushbutton
which calls the menu ”Time base“. Only the time
base A hold-off time may be changed.
See the section ”Hold-off time setting“ in the chapter ”Triggering and time bases“.
X-POS DELAY pushbutton
This pushbutton allows to change the function of the HORIZONTAL knob
26.1 Analog mode The pushbutton signals the actually selected function in accor­dance with the front panel lettering:
dark: X position control green: Delay time control
26.1.1 X-POS
If the pushbutton is dark the HORIZONTAL-knob as X position control, i.e. it moves the signal display horizon­tally.
The position control is especially useful when the magnifi er (MAG. x 10
) is switched on. The magnifi er will magnify the display 10 times around the screen centre, with the control the portion of the signal to be studied can be shifted on-screen.
.
19
26 27
20
23
21
24
28
22
25
29 30
functions
22.1.5 AC Line The trigger signal is taken from the line which feeds the scope. See also the section ”Line triggering“ in the chapter ”Triggering and time bases“. Readout: ”Tr:alt, Line, Slope“.
TRIG’d display (not in XY mode)
This LED will light up if the time base receives a trigger signal. It depends upon the trigger signal whether the LED will just blink or remain illuminated.
NORM display
This display will light up provided ”Auto“ triggering was not selected. The mode can be selected in the ”Trigger“ menu called by pressing (MODE remain dark as long as there is no suffi cient trigger signal.
36
Subject to change without notice
). The light points out that the screen will
26.1.2 DELAY In order to change the function of the HORIZONTAL-knob
to
DELAY proceed as follows:
Press the HOR
pushbutton which will present the TIME BASE menu, select SEARCH or B ONLY, then the function of the knob will be changed if the pushbutton is depressed. It will light up to show that the knob is now the delay time control.
In SEARCH mode both traces (time base A and B) alternate. Unlike the former time base A ONLY mode, a sector with high­er intensity is visible on the A trace. This sector can be moved continuously by the delay time control. The time between the A trace start and the beginning of the intensifi ed sector is the delay time. This information is also displayed in the readout (”Dt: …“) and is an aid to fi nd the position of the intensifi ed sector which may be very small. If time base B ONLY is chosen the
Controls and Readout
intensifi ed sector is no longer visible, but the DELAY function still can be used.
Without activated B Trigger function, the B time base will be started after the A time base delay time ”elapsed“.
26.2 DSO mode
The pushbutton will signal the actual function in accordance with the front panel lettering:
dark: The knob functions as X position control. green: The knob functions as delay time control.
26.2.1 X-POS
If the pushbutton is dark the HORIZONTAL
knob functions as X position control of the trigger time, i.e. it moves the trigger point symbol horizontally. This allows to display signal portions before and after the trigger, called Pre-Trigger and Post-Trig­ger. If the trigger point symbol is located on the screen centre the readout will show ”Tt:0s“, hence the trigger time indication is always referred to the screen centre. Values with a positive sign are Post-Trigger times, such with a negative sign Pre­Trigger times.
If the X-POS DELAY pushbutton is depressed the ”Hor.Knob“ menu will be called, it contains the following options:
st
1
Centre: Pressing the function pushbutton ”Center“ will set
the trigger time to the screen centre ”Tt:0s“ which is the standard setting.
nd
2
Coarse On Off: changes the speed of the HORIZONTAL
knob.
26.2.2 DELAY This pushbutton will be illuminated if the TIME BASE menu was called with the HOR B ONLY selected. The HORIZONTAL
pushbutton and SEARCH or
knob can then be used to select a portion of the time base display which is to be displayed expanded in time.
In SEARCH mode the normal and the expanded displays are displayed simultaneously. The expanded portion of the signal will be shown on the normal display as an intensifi ed sector. The length of this sector is dependent upon the setting of the
nd
2
”Z“ ,time base which is shown in the readout as ”Z...“ and
is equal to the run time of the Z time base.
HORIZONTAL knob
The various functions of this knob depend on the operating mode and are described under X-POS DELAY
pushbutton.
TIME/DIV.–SCALE–VAR knob
This knob is normally used as the time base speed selector, but has also other functions dependent on the operating mode. In XY mode this control is disabled.
28.1.2 Time base B time/cm selection This function is active if in the ”Time base“ menu (HOR push-button) SEARCH or B ONLY was selected and the option ”B variable On Off“ was set to Off.
Turning the control CCW will decrease, turning it CW will increa­se the time base speed. The speed can be selected between 20 ms/cm.. 50 ns/cm in a 1-2-5 sequence and will be calibrated. The readout will show the speed (e.g. ”B:50ns“). The time base B allows to display portions of the time base A display on an expanded time base scale. This implies that the speed of TB B must always be greater than that of TB A. Therefore with the exception of 50 ns/cm TB B can not be set to the same speed as TB A.
Further information is available in the section ”Time base B
nd
(2
time base/Delay/Triggering“ (Analog mode) in the chapter
”Triggering and time bases“.
28.1.3 Variable The TIME/DIV–SCALE–VAR control may also be used to change the time base speed continuously but uncalibrated. VAR will light up on top of the HOR
pushbutton in order to warn that the time base is uncalibrated and the knob has now that function. In order to arrive at that function press HOR
which calls the TIME BASE menu. Depending whether time base A or B is selected either ”A variable On Off“ or ”B variable On Off“ will be shown. The function pushbutton can then be used to select On/Off. In order to point out that the time base is now uncalibrated the readout will replace ”:“ by ”>“ preceding the time/cm. (e.g. ”A>500ns“ and ”B>200ns“). Also the results of cursor time/pe­riod measurements will be marked that way.
28.2 DSO mode
28.2.1 ZOOM OFF (A time base time/cm selection)
Select the menu ”Zoom“ by pressing HOR
and then ”Off“ in order to set the function of the knob TIME/DIV–SCALE–VAR to time base A speed as in analog mode. If ”Zoom Off“ is active always the whole memory will be displayed.
Turning the control CCW will decrease, turning it CW will increase the time base speed. Depending on the signal captu­re/display the time base can be set from 50 s/cm to 5 ns/cm in a 1-2-5 sequence (e.g. ”A:50ns“) and will be calibrated. There is no variable function as in analog mode.
28.2.2 Search–Zoom only (Zoom time base speed selection) One of the functions SEARCH or ZOOM ONLY may be selected in the ZOOM menu after pressing HOR
. ”Zoom Off“ is equivalent to time base A in analog mode. With ”Zoom only“ a portion of the display in Zoom ”Off“ can be expanded over the whole screen. This is possible because there is a very large memory for signal capture/display. The ”Zoom Off“ display will present the whole memory contents. With SEARCH both the ”Zoom Off“ and the expanded ”Zoom only“ displays will be visible. With ”Zoom only“ only the expanded display will show up.
28.1 Analog mode
28.1.1 Time base A time/cm selection
This function is active if in the ”Time base“ menu (HOR pushbutton) ”A only“ was selected and the option ”A variable On Off“ was set to Off.
Turning the knob CCW will decrease, turning it CW will increa­se the time base speed. The time base speed may be chosen between 500 ms/cm... 50 ns/cm in a 1-2-5 sequence and will be calibrated. The readout will show the setting (e.g. ”A:50ns“).
The Zoom time base speed will be indicated in the readout ”Z:...“ and is calibrated. Turning the knob CCW will decrease, turning it CW will increase the time base speed. This can be selected from 20 ms/cm to 5 ns/cm in a 1-2-5 sequence. The maximum expansion is 50,000 times (”A:10ms“ and ”Z:200ns“.)
MAG pushbutton
In analog mode only: pressing this pushbutton will turn on the x 10 magnifi er. No menu will be shown.
Subject to change without notice
37
Controls and Readout
POWER
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
VOLTS / DIV
SCALE · VAR
CH 2 HOR MAG
VAR
15
14
16
18
13
17
POSITION 1 POSITION 2
VOLTS / DIV
SCALE · VAR
20 V 1 mV 20 V 1 mV
CH 1
VAR
If ”x10“ is illuminated on the MAG pushbutton the magnifi er is activated. The adjusted time/cm will be shown in the top left readout. Depending on the time base mode turning on the magnifi er will have these effects:
29.1 ”Time base A only“
The display will be expanded around the screen centre by a factor of 10, the time/cm adjusted accordingly.
29.2 ”Search“ (A and B times bases alternated)
The time base A speed will not be affected. The time base B speed will be increased by a factor of 10, hence the time base B display will be expanded 10-fold in X direction.
29.3 ”B only“:
The time base B speed will be increased by a factor of 10 and the display hence expanded 10-fold in X direction.
HOR pushbutton
This pushbutton will open the ”Time base“ menu the contents of which depends on the operating mode selected.
30.1 Analog mode
The following modes are available:
30.1.1 A only Only the time base A will be operative. The readout will thus only show ”A...“ in the top left corner. The TIME/DIV-SCALE­VAR knob will set the time base A speed. With the MAG x10 pushbutton the display may be expanded around the screen centre, increasing the speed by 10. If the mode is changed from time base A to ”Search“ or ”B only“ all settings of time base A remain intact including triggering.
30.1.2 Search This mode implies alternate time base operation. The readout will show the speeds of both time bases (”A...“ and ”B...“). The TIME/DIV-SCALE-VAR knob will set the time base B speed.
In alternate time base mode part of the time base A display will be intensifi ed. The horizontal position of the intensifi ed portion may be shifted using the HORIZONTAL function was set to ”Delay“, this is the case if the X-POS DELAY
pushbutton is illuminated. The length of the intensifi ed sector is determined by the speed of time base B and is equal to the run time of B. Hence this intensifi ed portion of A will be spread over the full screen as displayed by B and thus expanded. The Y position of the signal is the same when displayed with either
knob, provided its
19
LEVEL A/B
X-POS
HORIZONTAL
26 27
TRIG ’d
NORM
HOLD OFF
DELAY
VAR
TIME / DIV
SCALE · VAR
50s 5ns
x10
20
23
21
24
28
22
25
29
TRIGGER
MODE
FILTER
SOURCE
AUX
30
A or B. This means, however, that both displays will be written over each other.
In order to separate them on the screen for better readability an artifi cial Y offset may be added to the time base B display. Press the CH1/2-CURSOR-MA/REF-ZOOM
pushbutton which calls the ”Pos./Scale“ menu. Press the function push­button ”TB B“, this will cause the POSITION 1 knob to act as the position control for the time base B display. (Trace sepa­ration.) See also 13.1.4 Y position 2
nd
time base. This function makes sense only in ”Search“ and is unavailable in any other. Also in ”Search“ the 10 x magnifi er is available by pressing the pushbutton MAG x10
. The magnifi er will affect solely
time base B.
30.1.3 ”B only“: In this mode only time base B will be displayed, the readout will thus only show the time base B speed in the top left corner (”B..“). The TIME/DIV-SCALE-VAR knob will set the time base B speed. The 10x magnifi er is available by pressing MAG x10 and will expand the display around the screen centre.
30.1.4 B trigger –
Edge In this mode time base B will not start immediately after the delay time set elapsed, but it will be only set ready waiting for a signal trigger. This has the advantage that any jitter is removed, but the delay time adjustment will now only have the effect that the time base B display will jump from signal period to period. In this setting a positive slope will trigger.
The (trigger) LEVEL A/B
knob will set the trigger level for B. Only normal triggering and DC-coupling are possible. All parameters of time base A remain stored and preserved. (LE­VEL, auto or normal, Slope, coupling). In addition to the delay time (”Dt:...“) also the B trigger parameters are shown in the readout: ”BTr:slope, DC“. In SEARCH mode the trigger point symbol will be preceded by ”B“. As mentioned changing the delay time will not cause a continuous move of the intensifi ed portion of the time base A display and the time base B display, but jumps from signal period to period.
If the trigger level symbol of time base B is shifted outside the signal representation by time base A there will be no triggering of time base B any more and thus no time base B display. The same holds in time base B only mode.
30.1.5 B trigger –
Edge Except for the negative edge the function is identical to the one described above (30.1.4).
38
Subject to change without notice
Controls and Readout
30.1.6 B trigger – OFF
Time base B will be started upon the end of the delay time set. The delay time can be changed continuously in this mode which can be watched on the intensifi ed sector of the time base A display. The disadvantage here is that with very long delay times jitter of the time base B display may crop up.
As time base B is not operated in the signal triggered mode the controls for time base B trigger will be disabled resp. are only for time base A.
30.1.7 A variable – On Off If ”On“ was selected the TIME/DIV-SCALE-VAR knob function as variable control for the time base A speed. Only in time base A only mode this option will be available in the menu. For a full description see ”28.1.3 Variable“.
30.1.8 B variable – On Off If ”On“ was selected the TIME/DIV-SCALE-VAR knob function as the time base B variable control. For a full descrip­tion see ”28.1.3 Variable“.
30.1.9 Holdoff …% In this mode the hold-off time may be selected from 0 to 100 % with the INTENS knob after a sweep before a new one can start and decrease thus the repetition rate which may darken the display. This is indicated by the HOLD OFF-LED valid for time base A.
Further information can be found in the section ”Hold-off adjust­ment“ in the chapter ”Triggering and time bases“.
30.2. DSO mode
In the ”Zoom“ menu the following time base functions are available:
30.2.1 Off In ”Zoom Off“ condition only time base A is active. The readout will thus only show ”A...“ in the top left corner. The speed can be set with the TIME/DIV-SCALE-VAR knob
30.2.2 Search Part of the time base A display will be intensifi ed and this portion will also be displayed expanded over the full screen. With the HORIZONTAL knob expanded display can be shifted provided that the pushbutton X-POS DELAY length of the intensifi ed sector is determined by the speed of the Z time base. The Y position of both displays is identical so they are written one over the other. In order to separate them for better readability an artifi cial Y offset may be added to the Z time base display. Press the pushbutton CH1/2-CURSOR-MA/REF-ZOOM call the menu ”Pos./Scale“.
Then press the function pushbutton ZOOM. Now the POSITION 1 knob will function as Y position control for time base Z. As this only makes sense in SEARCH mode it is unavailable in others.
30.2.3 Zoom only Only the Z time base will be displayed. The readout will hence only show ”Z...“ in the top left corner. The TIME/DIV-SCALE-VAR knob affects only time base B.
is illuminated which means ”Delay“. The
. Values > 0 extend the waiting time
lighting up. The hold-off time is only
.
the intensifi ed sector and the
will
will
to
31.1. AC DC
Pressing the pushbutton will switch from AC to DC or vice versa. The mode selected will be shown in the readout following the sensitivity setting: ~ is for AC and = is for DC.
31.1.1 DC coupling The signal will be directly coupled, from the BNC connector via the attenuator to the vertical amplifi er. The input resistance is 1 MΩ in all positions of the attenuator.
31.1.2 AC coupling A capacitor is inserted between the BNC connector and the attenuator, blocking the DC content of the signal and creating a low frequency cut-off at approx. 2 Hz. This will affect the shape and amplitude of signals with low frequency content. If the DC content of the signal changes or the duty cycle of pulses the capacitor will charge or discharge, this will cause a momentary Y shift of the display.
31.2 Ground On Off
The pushbutton will either connect the amplifi er to the signal or to ground. If set to Ground the readout will show a ground symbol following the sensitivity setting, at the same place where formerly the coupling was indicated. In the Ground position and with automatic triggering a trace will be visible, this is handy for setting the Y position of it e.g. to the screen centre without disconnecting the signal. The readout will show a symbol ( 0 V which will be close to the vertical centre line of the graticule, it is the zero reference for any measurements. After switching back to the signal its amplitude can now be determined with respect to the formerly set zero reference.
31.3 Invert On Off (unavailable in analog XY mode) This pushbutton will alternate between not inverted or inverted of the CH1 signal. The readout will indicate an inverted display by placing a bar above the CH1. The trigger signal derived from Ch1 will not be affected by an inversion.
31.4 Probe submenu
Pressing the pushbutton will open the ”CH1 probe“ submenu.
31.4.1 *1 - *10 - *100 - *1000 A selection of probe attenuation factors from 1 to 1,000 is pos­sible. The value selected will be automatically taken into account regarding the sensitivity indicated and measurements.
31.4.2 Auto If Auto is selected HAMEG probes with automatic probe iden­tifi cation connectors will be recognized by the scope and the appropriate factor taken into account. The factor recognized will be indicated following ”auto“.
Probes without that special connector will cause the display ”auto *1“ and treated as 1:1 probes.
31.5 Variable On Off
If activated VAR on the CH1 pushbutton The readout will replace ”:“ by ”>“ (e.g. ”CH1>5mV“.) and this also in any cursor measurements pointing out that the sensitivity is uncalibrated.
The VOLTS/DIV-SCALE-VAR knob which can change the sensitivity continuously between 1 mV/cm to > 20 V/cm.
will be illuminated.
will function as variable
) for
CH1 pushbutton
This pushbutton opens the CH1 menu which contains the follo­wing options referring to CH1
resp. to the signal on CH1.
VERT/XY pushbutton
This pushbutton switches the ”vertical“ menu on/off. This menu allows to select the operating modes of the vertical amplifi ers.
Subject to change without notice
39
Controls and Readout
POWER
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
VOLTS / DIV
SCALE · VAR
CH 2 HOR MAG
VAR
15
14
16
18
13
17
POSITION 1 POSITION 2
VOLTS / DIV
SCALE · VAR
20 V 1 mV 20 V 1 mV
CH 1
VAR
32.1 CH1 If CH1 is selected only CH1 will be turned on, the mode is Yt. Also the readout will only display the parameters of CH1. (sensitivity, inverted/not inverted, coupling.) Although CH2 will not appear in the readout it may be used e.g. as a trigger input. Its controls are active but are not shown.
32.2 CH2 If CH2 is selected only CH2 will be active, it is Yt mode, and only its parameters will be shown in the readout. Although CH1 will not appear in the readout it may be used e.g. as a trigger input. Its controls are active but are not shown.
32.3.1 DUAL trace alt./ chop In dual trace mode both channels are turned on and the parame­ters of both are shown in the readout. Between the sensitivity indi­cations there is an indication whether alternate ”alt.“ or chopped ”chp“ mode is active. Normally, the mode will be automatically set by the time base speed selection, but it may be directly set using the function pushbutton. For time base speeds of 500 ms/cm to 500 μs/cm chopped will be used, from 200 μs/cm to 50 ns/cm alternate. This refers to unmagnifi ed time bases. Alternate is the preferred mode, at any time one channel is dis­played for a full sweep, after each sweep the other channel has its turn. At slow sweep speeds this will cause annoying fl icker, at still slower ones the channel switching becomes visible. Here, the chopped mode steps in, both channels are switched at some high frequency so they are both visible at any sweep speed. This is, however, not appropriate for fast sweep speeds as the switching may become visible and may interfere with the proper signal display.
32.3.2 DUAL (DSO mode). In DSO mode there is an a/d converter for each channel so both are measured simultaneously. Hence no channel switching is necessary and no information pertaining to it is shown.
32.4 ADD
In ADD mode the signals of both channels are algebraically added and displayed as one sum signal. The Y position can be changed with both position controls. If one channel is inverted the difference will be displayed. Only one ”0 V“ symbol will be shown in the readout. The ADD mode will be indicated by placing a ”+“ symbol between the sensitivity indications of both channels. In DSO mode ”1+2“ will be displayed at the end of the trace. Please note that the results of cursor measurements in this mode will only be correct if the sensitivities of both channels are identical, otherwise the readout will show ”CH1<>CH2“.
19
LEVEL A/B
X-POS
HORIZONTAL
26 27
TRIG ’d
NORM
HOLD OFF
DELAY
VAR
TIME / DIV
SCALE · VAR
50s 5ns
x10
20
23
21
24
28
22
25
29
TRIGGER
MODE
FILTER
SOURCE
AUX
30
Automatic voltage measurements can not be performed in ADD mode. The readout will show ”n/a“ = not available.
As the trigger signals are taken off the inputs and not from the added signal there is no true reference for the trigger point symbol, the symbol will thus be switched off in analog mode. However, the LEVEL A/B control
is active.
In DSO mode a trigger time symbol is displayed one line above the lowest graticule line to indicate the trigger time position along the signal and thus can only be moved horizontally.
32.5 XY
In this mode CH1 will move the trace in X direction, hence the readout will show ”CHX..“, CH2 will move the trace in Y direction, hence ”CHY...“ will be shown rather than ”CH2...“.
As the time bases are not involved in XY no time base related information will be shown. Also the trigger circuits are disabled so no trigger information is shown, either.
The magnifi er MAG x10
is disabled. The ”0-Volt“ symbols will be shown as triangles at the right hand graticule and above the sensitivities.
Both the HORIZONTAL
or the POSITION 1 knobs will move the trace horizontally. The Y position is controlled by the POSITION 2 knob.
32.5.1 Analog mode The CH1 signal can not be inverted, there is hence no menu item in the CH1 menu (CH1 pushbutton VAR
knob is disabled.
). The TIME/DIV-SCALE-
Please note that the bandwidths and phase differences in XY analog and DSO modes differ considerably so there may be changes in the signal display when switching the mode.
32.5.2 DSO mode The readout will indicate the sampling rate with which the a/d converters digitise the input signals. The appropriate sampling rate must be set depending on the signals and can be selected with the TIME/DIV-SCALE-VAR knob, although the time bases are disabled. With high sampling rates there may be gaps in Lissajous representations. With too low sampling rates the display may not allow any more to determine the frequency relationship of the signals. It is recommended to fi rst look at the signals in DUAL mode and to set the sampling rate such that at least one signal period
40
Subject to change without notice
Controls and Readout
POWER
CH 1
VAR
X-INP
!
CAT I
3431
VERT/XY
INPUTS
1MΩII15pF
max
400 Vp
32
CH 2 HOR MAG
VAR
35
33
will be displayed. Then XY should be selected. In XY DSO mode both channels may be inverted.
32.6 Bandwidth Full/20 MHz
This pushbutton will select full or 20 MHz bandwidth.
Full: Full bandwidth will be the one given in the specifi cations.
20 MHz: Provided measuring modes allow full bandwidth
(i.e. 5 mV/cm) this can be reduced to 20 MHz (–3 dB) in order to attenuate high frequency noise e.g. The readout will show BWL = bandwidth limited. The bandwidth limitation affects both channels and pertains to analog and DSO modes. In XY DSO mode the limitation is equal to Yt mode. In XY analog mode the limitation affects only CH2.
CH2 pushbutton
This pushbutton opens the CH2 menu which offers the following options:
33.1 AC DC
The pushbutton will alternate between AC and DC coupling. The readout shows a ”~“ or ”=“ symbol behind the sensitivity indication.
33.1.1 DC coupling The signal will be directly coupled to the input amplifi er via the BNC connector resistance of the scope is a constant 1 MΩ irrespective of the sensitivity selected.
33.1.2 AC coupling A capacitor is inserted between the BNC connector and the at­tenuator, thus the DC content of the signal is blocked and a high pass with a lower cut-off frequency of approx. 2 Hz is created. Low frequency signals will thus be more or less differentiated, hence their shape and amplitude affected. If the DC content of the signal changes, e.g. the duty cycle of pulses, the capacitor must charge or discharge. This will cause a momentary Y shift of the display.
33.2 Ground (GND) On Off
The pushbutton will alternate between switching the amplifi er input to the signal or to ground.
and the input attenuator. The input
AUX
VAR
AUXILIARY INPUT
TRIGGER EXTERN
!
CAT I
36 37
Z-INPUT
x10
1MΩ II
15pF
max
100 Vp
29 30
Referred to the trace position 0 V a DC voltage may be measured after the input was returned to the signal.
33.3 Invert On Off
The pushbutton will alternate between not inverted and inverted. An inverted signal will be indicated in the readout by bar above the channel symbol. The trigger signal taken from an input will not be affected.
33.4 Probe menu
This pushbutton opens the ”CH2 probe“ submenu.
33.4.1 *1 - *10 - *100 - *1000 A selection of 4 factors can be made, the factor chosen will be automatically taken into account for all displays and measu­rements.
33.4.2 auto If ”auto“ was selected HAMEG probes with special probe iden­tifi cation connectors will be automatically identifi ed and the appropriate factor taken into account. The probe factor identifi ed will be shown behind ”auto“.
Probes without that special connector will be treated as 1:1 probes (display ”auto*1“), for those the factor must be manu­ally set.
33.5 Variable On Off
If the variable is on VAR on the pushbutton will be illuminated and indicates that the sensitivity is now uncalibrated, the readout will show ”>“ instead of ”:“ (e.g. ”CH2>5 mV“) The results of cursor measurements will be identifi ed accordingly.
The VOLTS/DIV-SCALE-VAR knob
of CH2 has now the function of variable, the sensitivity can be varied between 1 mV/cm to > 20 V/cm.
INPUT CH1 – BNC connector
This is the CH1 signal input connector. In Yt mode it is a Y input, in XY mode it is the X signal input. The connector housing is connected to the instrument housing and thus to safety ground. The ring around the connector is the probe identifi cation contact, no voltage may be applied here.
If the signal is disconnected resp. the amplifi er input connected to ground the readout will show a ground symbol behind the sensitivity indication. In automatic trigger mode the trace will be visible in a reference position which can be used as a 0 V ground reference. The readout will show a symbol ( ) for 0 V which will be close to the vertical centre line of the graticule, it is the zero reference for any measurements.
INPUT CH2 – BNC connector
This is the CH2 signal input connector. It is a Y input in Yt and XY mode. The connector housing is connected to the instru­ment housing and thus to safety ground. The ring around the connector is the probe identifi cation contact, no voltage may be applied here.
Subject to change without notice
41
Controls and Readout
POWER
POWER
CH 1
VAR
X-INP
!
CAT I
3431
VERT/XY
INPUTS
1MΩII15pF
max
400 Vp
32
CH 2 HOR MAG
VAR
33
35
36 37
AUX pushbutton
This pushbutton belongs to the AUXILIARY INPUT . The ope­ning of the menu depends upon the actual operating mode.
36.1 Analog mode
36.1.1 AUXILIARY INPUT
setting external triggering press SOURCE Source“ menu, then „External“.
36.1.2 If „External“ triggering was not selected the menu „Z Input“ will open up. If „Off“ is chosen AUXILIARY INPUT no function. If „On“ is chosen it will function as Z input i.e. in­tensity modulation input. This input is destined for TTL signals, a voltage of >1 V
will turn off the trace.
p
is the external trigger input. For
, select the „Trig.
has
AUX
VAR
x10
29 30
AUXILIARY INPUT
!
CAT I
TRIGGER EXTERN
Z-INPUT
1MΩ II
15pF
max
100 Vp
PROBE ADJ. – connector
A square wave signal of 0.2 Vpp is available for the adjustment of 10 : 1 probes. The frequency can be selected by pressing the pushbutton PROBE ADJ. and calling the menu ”Utilities“. Further information may be found in the section ”Probe adjust­ment and use“ in the chapter ”Operation and pre-settings“.
PROBE ADJ – pushbutton
This pushbutton produces the menu ”Utilities“, two options are available:
39.1 COMP. Tester On Off.
If ”On“ is chosen analog mode is selected automatically. A trace and a readout display ”Component Tester“ will appear.
36.2 DSO mode
36.2.1 In DSO mode the AUX pushbutton is deactivated and the
AUXILIARY INPUT
can only serve as an input for external
trigger signals.
AUXILIARY INPUT (BNC connector)
In DSO mode this connector can be used as external trigger input. In analog mode it can serve as external trigger or Z-axis (in­tensity modulation) input.
The connector housing is connected to the instrument housing and thus to safety ground. Although the ring around the connec­tor has no function, no voltage may be applied here.
CH I: 500 mV
In this mode the 4 mm connectors labelled ”COMPONENT TESTER“ become the measuring input. See also the chapter ”Component Tester“.
Choosing ”Off“ will return the instrument to all former set­tings.
39.2 Calibrator 1kHz 1MHz
Depending on the setting the square wave signal frequency at the PROBE ADJ – socket is 1 kHz or 1 MHz.
COMPONENT TESTER - (sockets)
Both 4 mm diameter sockets serve as a two pole input for component test. Further information can be found under section Component Tester.
1 : 1 / 10 / 100
COMBISCOPE
42
Subject to change without notice
MEMORY
oom
COMPONENT
TESTER
40
39
PROBE
ADJ
38
Controls and Readout
Subject to change without notice
43
Oscilloscopes
Spectrum Analyzer
Power Supplies
Modularsystem
Serie s 8000
Programmable Instruments
Series 8100
authorized dealer
41-1008- 00E1
www.hameg.com
Subject to change without notice 41-1008-00E1 / 15-03-2007-gw HAMEG Instruments GmbH © HAMEG Instruments GmbH Industriestraße 6 A Rohde & Schwarz company D-63533 Mainhausen ® registered trademark Tel +49 (0) 61 82 800-0
DQS-Certifi cation: DIN EN ISO 9001:2000 Fax +49 (0) 61 82 800-100 Reg.-Nr.: 071040 QM sales@hameg.de
Loading...