Hersteller HAMEG Instruments GmbH KONFORMITÄTSERKLÄRUNG
Manufacturer Industriestraße 6 DECLARATION OF CONFORMITY
Fabricant D-63533 Mainhausen DECLARATION DE CONFORMITE
Die HAMEG Instruments GmbH bescheinigt die Konformität für das Produkt
The HAMEG Instruments GmbH herewith declares conformity of the product
HAMEG Instruments GmbH déclare la conformite du produit
Bezeichnung / Product name / Designation:
Oszilloskop
Oscilloscope
Oscilloscope
Typ / Type / Type: HM1008
mit / with / avec: –
Optionen / Options / Options: –
mit den folgenden Bestimmungen / with applicable regulations / avec les
directives suivantes
EMV Richtlinie 89/336/EWG ergänzt durch 91/263/EWG, 92/31/EWG
EMC Directive 89/336/EEC amended by 91/263/EWG, 92/31/EEC
Directive EMC 89/336/CEE amendée par 91/263/EWG, 92/31/CEE
Niederspannungsrichtlinie 73/23/EWG ergänzt durch 93/68/EWG
Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC
Directive des equipements basse tension 73/23/CEE amendée par 93/68/CEE
HAMEG instruments fulfi ll the regulations of the EMC directive. The
conformity test made by HAMEG is based on the actual generic- and
product standards. In cases where different limit values are applicable,
HAMEG applies the severer standard. For emission the limits for
residential, commercial and light industry are applied. Regarding the
immunity (susceptibility) the limits for industrial environment have
been used.
The measuring- and data lines of the instrument have much infl uence
on emmission and immunity and therefore on meeting the acceptance
limits. For different applications the lines and/or cables used may
be different. For measurement operation the following hints and
conditions regarding emission and immunity should be observed:
1. Data cables
For the connection between instruments resp. their interfaces and
external devices, (computer, printer etc.) suffi ciently screened cables
must be used. Without a special instruction in the manual for a reduced
cable length, the maximum cable length of a dataline must be less than
3 meters and not be used outside buildings. If an interface has several
connectors only one connector must have a connection to a cable.
Basically interconnections must have a double screening. For IEEE-bus
purposes the double screened cables HZ72S and HZ72L from HAMEG
are suitable.
2. Signal cables
Basically test leads for signal interconnection between test point and
instrument should be as short as possible. Without instruction in the
manual for a shorter length, signal lines must be less than 3 meters
and not be used outside buildings.
Sicherheit / Safety / Sécurité: EN 61010-1:2001 (IEC 61010-1:2001)
Überspannungskategorie / Overvoltage category / Catégorie de surtension: II
Verschmutzungsgrad / Degree of pollution / Degré de pollution: 2
EN 61000-3-2/A14 Oberschwingungsströme / Harmonic current emissions /
Émissions de courant harmonique:
Klasse / Class / Classe D.
EN 61000-3-3 Spannungsschwankungen u. Flicker / Voltage fl uctuations and fl icker /
Fluctuations de tension et du fl icker.
Datum /Date /Date
24. 02. 2005
Unterschrift / Signature / Signatur
Manuel Roth
Manager
This will not cause damage or put the instrument out of operation. Small
deviations of the measuring value (reading) exceeding the instruments
specifi cations may result from such conditions in individual cases.
4. RF immunity of oscilloscopes.
4.1 Electromagnetic RF fi eld
The infl uence of electric and magnetic RF fi elds may become visible
(e.g. RF superimposed), if the fi eld intensity is high. In most cases
the coupling into the oscilloscope takes place via the device under
test, mains/line supply, test leads, control cables and/or radiation.
The device under test as well as the oscilloscope may be effected by
such fi elds.
Although the interior of the oscilloscope is screened by the cabinet,
direct radiation can occur via the CRT gap. As the bandwidth of
each amplifi er stage is higher than the total –3dB bandwidth of the
oscilloscope, the infl uence RF fi elds of even higher frequencies may
be noticeable.
4.2 Electrical fast transients / electrostatic discharge
Electrical fast transient signals (burst) may be coupled into the
oscilloscope directly via the mains/line supply, or indirectly via test
leads and/or control cables. Due to the high trigger and input sensitivity
of the oscilloscopes, such normally high signals may effect the trigger
unit and/or may become visible on the CRT, which is unavoidable.
These effects can also be caused by direct or indirect electrostatic
discharge.
HAMEG Instruments GmbH
Signal lines must screened (coaxial cable - RG58/U). A proper ground
connection is required. In combination with signal generators double
screened cables (RG223/U, RG214/U) must be used.
3. Infl uence on measuring instruments
Under the presence of strong high frequency electric or magnetic fi elds,
even with careful setup of the measuring equipment an infl uence of
such signals is unavoidable.
2
Subject to change without notice
Contents
General information regarding the CE marking 2
100 MHz CombiScope HM1008 4
Specifi cations 5
Important hints 6
List of symbols used: 6
Positioning the instrument 6
Safety 6
Proper operation 6
CAT I 6
Environment of use. 6
Environmental conditions 7
Warranty and repair 7
Maintenance 7
Line voltage 7
Description of the controls 8
Basic signal measurement 10
Signals which can be measured 10
Amplitude of signals 10
Values of a sine wave signal 10
DC and ac components of an input signal 11
Timing relationships 11
Connection of signals 11
AUTOSET 19
Component tester 19
CombiScope 21
DSO Operation 22
DSO operating modes 22
Memory resolution 22
Memory depth 23
Horizontal resolution with X magnifi er 23
Maximum signal frequency in DSO mode 23
Display of aliases 23
Vertical amplifi er operating modes 23
Data transfer 23
HO710: RS-232 Interface. Remote control 24
Selection of Baud rate 24
Data transmission 24
Loading of new fi rmware 24
General information concerning MENU 25
Controls and Readout 26
First time operation and initial adjustments 12
Trace rotation TR 12
Probe adjustment and use 12
1 kHz adjustment 12
1 MHz adjustment 13
Operating modes of the vertical amplifi er 13
XY operation 14
Phase measurements with Lissajous fi gures 14
Measurement of phase differences in dual
channel Yt mode 14
Measurement of amplitude modulation 15
Triggering and time base 15
Automatic peak triggering (MODE menu) 15
Normal trigger mode (See menu MODE) 16
Slope selection (Menu FILTER) 16
Trigger coupling (Menu: FILTER) 16
Video (tv triggering) 16
Frame sync pulse triggering 17
Line sync pulse triggering 17
LINE trigger 17
Alternate trigger 17
External triggering 17
Indication of triggered operation (TRIG’D LED) 17
Hold-off time adjustment 17
Time base B (2nd time base). Delaying,
Delayed Sweep. Analog mode. 18
Alternate sweep 18
Subject to change without notice
3
HM1008
1 GSa/s Real Time Sampling, 10 GSa/s Random Sampling
1 MPt memory per channel allows Memory oom
up to 50,000:1
Two Channels 1mV – 20 V/cm
8-Bit Low Noise Flash A/D Converters
Pre/Post Trigger -100 % to +400 %
Time Base 50 s/cm – 5 ns/cm
Acquisition modes: Single Event, Refresh, Average, Envelope,
Roll, Peak-Detect
Inputs CH 1, 2:
Input Impedance:1 MΩ II 15 pF
Coupling:DC, AC, GND (ground)
Max. Input Voltage:400 V (DC + peak AC)
Y Delay Line (analog):70 ns
Measuring Circuits:Measuring Category I
Anal
og mode only:
Auxiliary input:
Function (selectable):Extern Trigger, Z (unblank)
Coupling:AC, DC
Max. input voltage:100 V (DC +peak AC)
Triggering
Analog and Digital Mode
Automatic (Peak to Peak):
Min. signal height:5mm
Frequency range:10Hz - 200 MHz
Level control range:from Peak- to Peak+
Normal (without peak):Slope/Video
Min. signal height:5mm
Frequency range:0 - 200MHz
Level control range:–10 cm to +10cm
Operating modes:Slope/Video
Slope:positive, negative, both
Sources:CH 1, CH 2, alt. CH 1/2 (≥ 8mm), Line, Ext.
Coupling:AC: 10 Hz-200MHz
DC: 0 -200 MHz
HF: 30 kHz–200 MHz
LF: 0-5kHz
Noise Rej. switchable
Video:pos./neg. Sync. Impulse
Standards:525 Line/60Hz Systems
625 Line/50Hz Systems
Field:even/odd /both
Line:all/line number selectable
Source:CH 1, CH 2, Ext.
Indicator for trigger action: LED
External Trigger via:Auxiliary Input (0.3Vpp, 100 MHz)
Coupling:AC, DC
Max. input voltage:100 V (DC +peak AC)
Digital mode
Pre/Post Trigger:-100 % to +400% related to complete memory
Analog mode
2nd Trigger
Min. signal height:5mm
Frequency range:0 - 200MHz
Coupling:DC
Level control range:–10 cm to +10cm
Horizontal Deflection
Analog mode
Operating modes:A, ALT (alternating A/B), B
Time base A:0.5s/cm - 50 ns/cm (1-2-5 sequence)
Time base B:20 ms/cm – 50 ns/cm (1-2-5 sequence)
Accuracy A and B: ±3%
X Magnification x10:to 5 ns/cm
Accuracy: ±5%
Variable time base A/B: cont. 1:2.5
Hold Off time:var. 1:10 (LED-Indication)
Bandwidth X-Amplifier: 0 - 3 MHz (-3dB)
X Y phase shift ‹ 3°:‹ 220kHz
Digit
al mode
Time base range (1-2-5 sequence)
Refresh Mode:20 ms/cm - 5 ns/cm
with Peak Detect:20ms/cm – 2 ms/cm (min. Pulse Width 10 ns)
Roll Mode:50s/cm – 50 ms/cm
Accuracy time base
Time base:50 ppm
Display:±1%
MEMORY ZOOM: max. 50,000:1
Bandwidth X-Amplifier:0 - 100 MHz (-3dB)
X Y phase shift ‹ 3°:‹ 100 MHz
Analog mode:Frequency, Period, Vdc, Vpp, Vp+, Vp-
also in digital mode:V
rms
, V
avg
Cursor Measurements:
Analog mode:Δt, 1/Δt (f), tr, ΔV, V to GND, ratio X, ratio Y
plus in digital mode:V
pp
, Vp+, Vp-, V
avg
, V
rms
, pulse count
Resolution Readout/Cursor: 1000 x 2000Pts, Signals: 250 x 2000
Interfaces (plug-in):RS-232 (HO710)
Optional:IEEE-488, Ethernet, Dual-Interface
RS-232/USB
Mathematic functions
Number of Formula Sets:5 with 5 formulas each
Sources:CH 1, CH 2, Math 1-Math 5
Targets:5 math. memories, Math 1-5
Functions:ADD, SUB, 1/X, ABS, MUL, DIV, SQ, POS,
NEG, INV
Display:max. 2 math. memories (Math 1-5)
Display
CRT:D14-375GH
Display area (with graticule): 8cm x 10cm
Acceleration voltage:approx. 14 kV
General Information
Component tester
Test voltage:approx. 7V
rms
(open circuit), approx. 50 Hz
Test current:max. 7 mA
rms
(short circuit)
Reference Potential :Ground (safety earth)
Probe ADJ Output:1 kHz/1 MHz square wave signal 0.2 V
pp
(tr ‹ 4 ns)
Trace rotation:electronic
Line voltage:105 – 253 V, 50/60 Hz ±10 %, CAT II
Power consumption:42 Watt at 230V, 50Hz
Protective system:Safety class I (EN61010-1)
Weight:5.6 kg
Cabinet (W x H x D):285 x 125 x 380mm
Ambient temperature:0° C ...+40° C
Accessories supplied: Line cord, Operating manual, 2 Probes 10:1 with at tenuation ID, Windows Software for control and data transfer
Optional accessories: HO720 Dual-Inter face RS-232/USB, HO730 Dual-Inter face Ethernet/ USB, HO740 Interface IEEE-488 (GPIB), HZ70 Opto-Interface (with
optical fi ber cable)
Subject to change without notice
5
Important hints
Important hints
Please check the instrument for mechanical damage or loose
parts immediately after unpacking. In case of damage we advise
to contact the sender. Do not operate.
List of symbols used:
Consult the manual High voltage
Important note Ground
Positioning the instrument
As can be seen from the fi gures, the handle can be set into
different positions:
A = carrying
B = handle removal and horizontal carrying
C = horizontal operating
D and E = operating at different angles
F = handle removal
T = shipping (handle unlocked)
B
C
B
T
A
C
D
F
E
D
E
A
PUkT
PUOGkT
PUkT
ANALOG
PUOPFGkT
PUkT
HGOFFD
PUOPFGkT
PUOPFGkT
HGOPFFD
DIGITAL
MIXED SIGNAL
COMBISCOPE
PUkT
HM1508
PUOPFGkT
PUOPFGkT
PFGkT
PUkT
1 GSa · 1MB
PUkT
150 MHz
PUOPFGkT PUOPFGkT
PUOPFGkT
VOLTS/DIVV
HGOPFFD
B
HAMEG
C O M B I S C O P E
PUk PUk
PUkT
PUkT
PUOPFGkT
PUOPFGkT
PUkT
PUOPF
PUOPF
PUkT
PUOPF
HGOPFFD
PUkT
PUkT
VOLTS/DIVV
VOLTS/DIVV
HGOPFFD
HGOPFFD
PUkT
PUOPFGkT
PUkT
HGOPFFD
PUkT
PUOPFGkT
PUk PUk
PUk PUk
PUOPFGkT
PGkT
PUkT PUkT
PUOPFGkT
PUOPFGkT
PUOPF PUOPF
PUOPFGkT
INPUTS
PUkTKl
PUkTKl
15pF
15pF
max
max
400 Vp
400 Vp
PUkT
Attention!
When changing the handle position, the instrument
must be placed so that it can not fall (e.g. placed
on a table). Then the handle locking knobs must be
simultaneously pulled outwards and rotated to the
required position. Without pulling the locking knobs
they will latch in into the next locking position.
Handle mounting/dismounting
The handle can be removed by pulling it out further, depending
on the instrument model in position B or F.
Safety
The instrument fulfi ls the VDE 0411 part 1 regulations for
electrical measuring, control and laboratory instruments and
was manufactured and tested accordingly. It left the factory in
perfect safe condition. Hence it also corresponds to European
Standard EN 61010-1 resp. International Standard IEC 1010-1.
In order to maintain this condition and to ensure safe operation
the user is required to observe the warnings and other directions
for use in this manual. Housing, chassis as well as all measuring terminals are connected to safety ground of the mains.
All accessible metal parts were tested against the mains with
2200 V
The oscilloscope may only be operated from mains outlets with
a safety ground connector. The plug has to be installed prior to
connecting any signals. It is prohibited to separate the safety
ground connection.
Most electron tubes generate X-rays; the ion dose rate of this instrument remains well below the 36 pA/kg permitted by law.
. The instrument conforms to safety class I.
DC
T
T
Safe operation may be endangered if any of the following
was noticed:
– in case of visible damage.
– in case loose parts were noticed
– if it does not function any more.
– after prolonged storage under unfavourable conditions (e.g.
like in the open or in moist atmosphere).
– after any improper transport (e.g. insuffi cient packing not
conforming to the minimum standards of post, rail or transport fi rm)
Proper operation
Please note: This instrument is only destined for use by personnel well instructed and familiar with the dangers of electrical
measurements.
For safety reasons the oscilloscope may only be operated from
mains outlets with safety ground connector. It is prohibited
to separate the safety ground connection. The plug must be
inserted prior to connecting any signals.
CAT I
In case safe operation may not be guaranteed do not use the
instrument any more and lock it away in a secure place.
6
Subject to change without notice
This oscilloscope is destined for measurements in circuits not
connected to the mains or only indirectly. Direct measurements,
Important hints
i.e. with a galvanic connection to circuits corresponding to the
categories II, III, or IV are prohibited!
The measuring circuits are considered not connected to the
mains if a suitable isolation transformer fulfi lling safety class
II is used. Measurements on the mains are also possible if
suitable probes like current probes are used which fulfi l the
safety class II. The measurement category of such probes must
be checked and observed.
Measurement categories
The measurement categories were derived corresponding to
the distance from the power station and the transients to be
expected hence. Transients are short, very fast voltage or current excursions which may be periodic or not.
Measurement CAT IV:
Measurements close to the power station, e.g. on electricity
meters
Measurement CAT III:
Measurements in the interior of buildings (power distribution
installations, mains outlets, motors which are permanently
installed).
Measurement CAT II:
Measurements in circuits directly connected to the mains
(household appliances, power tools etc).
Environment of use.
The oscilloscope is destined for operation in industrial, business,
manufacturing, and living sites.
Environmental conditions
check all operating modes and fulfi lment of specifi cations. The
latter is performed with test equipment traceable to national
measurement standards.
Statutory warranty regulations apply in the country where the
HAMEG product was purchased. In case of complaints please
contact the dealer who supplied your HAMEG product.
Maintenance
Clean the outer shell using a dust brush in regular intervals.
Dirt can be removed from housing, handle, all metal and plastic
parts using a cloth moistened with water and 1 % detergent.
Greasy dirt may be removed with benzene (petroleum ether) or
alcohol, there after wipe the surfaces with a dry cloth. Plastic
parts should be treated with an antistatic solution destined
for such parts. No fl uid may enter the instrument. Do not use
other cleansing agents as they may adversely affect the plastic
or lacquered surfaces.
Line voltage
The instrument has a wide range power supply from 105 to 253
V, 50 or 60 Hz ±10%. There is hence no line voltage selector.
The line fuse is accessible on the rear panel and part of the line
input connector. Prior to exchanging a fuse the line cord must
be pulled out. Exchange is only allowed if the fuse holder is
undamaged, it can be taken out using a screwdriver put into the
slot. The fuse can be pushed out of its holder and exchanged.
The holder with the new fuse can then be pushed back in place
against the spring. It is prohibited to ”repair“ blown fuses or to
bridge the fuse. Any damages incurred by such measures will
void the warranty.
Operating ambient temperature: 0 to + 40 degrees C. During
transport or storage the temperature may be –20 to +55 degrees C.
Please note that after exposure to such temperatures or in case
of condensation proper time must be allowed until the instrument has reached the permissible range of 0 to + 40 degrees
resp. until the condensation has evaporated before it may be
turned on! Ordinarily this will be the case after 2 hours. The
oscilloscope is destined for use in clean and dry environments.
Do not operate in dusty or chemically aggressive atmosphere
or if there is danger of explosion.
The operating position may be any, however, suffi cient ventilation must be ensured (convection cooling). Prolonged operation
requires the horizontal or inclined position.
Do not obstruct the ventilation holes!
Specifi cations are valid after a 20 minute warm-up period
between 15 and 30 degr. C. Specifi cations without tolerances
are average values.
Warranty and repair
HAMEG instruments are subjected to a rigorous quality control.
Prior to shipment each instrument will be burnt in for 10 hours.
Intermittent operation will produce nearly all early failures.
After burn in, a fi nal functional and quality test is performed to
Type of fuse:
Size 5 x 20 mm; 250V~, C;
IEC 127, Bl. III; DIN 41 662
(or DIN 41 571, Bl. 3).
Cut off: slow blow (T) 0,8A.
Subject to change without notice
7
Front Panel Elements – Brief Description
Front Panel Elements – Brief Description
The fi gures indicate the page for complete discriptions
POWER (pushbutton switch) 26
Turns scope on and off.
INTENS (knob) 26
Intensity for trace- and readout brightness, focus and trace
rotation control.
FOCUS, TRACE, MENU (pushbutton switch) 26
Calls the Intensity Knob menu to be displayed and enables
the change of different settings by aid of the INTENS knob.
See item 2.
REM (pushbutton switch) 26
Switches the displayed menu, the remote mode (REM lit)
off.
ANALOG/DIGITAL (pushbutton switch) 27
Switches between analog (green) and digital mode (blue).
STOP / RUN (pushbutton switch) 27
RUN: Signal data acquisition enabled.
STOP: Signal data acquisition disabled. The result of the last
acquisition is displayed.
MATH (pushbutton switch) 27
Calls mathematical function menu if digital mode is pre-
sent.
ACQUIRE (pushbutton switch) 28
Calls the signal capture and display mode menu in digital
mode.
SAVE/RECALL (pushbutton switch) 29
Offers access to the reference signal (digital mode only) and
the instrument settings memory.
SETTINGS (pushbutton switch) 30
Opens menu for language and miscellaneous function; in
digital mode also signal display mode.
AUTOSET (pushbutton switch) 30
Enables appropriate, signal related, automatic instrument
settings.
in the chapter CONTROLS AND READOUT
▼
VOLTS/DIV-SCALE-VAR (knob) 32
Channel 2 Y defl ection coeffi cient, Y variabel and Y scaling
setting.
AUTO / CURSOR MEASURE (pushbutton switch) 33
Calls menus and submenus for automatic and cursor sup-
ported measurement.
LEVEL A/B (knob) 34
Trigger level control for time base A and B.
MODE (pushbutton switch) 34
Calls selectable trigger modes.
FILTER (pushbutton switch) 35
Calls selectable trigger fi lter (coupling) and trigger slope
menu.
SOURCE (pushbutton switch) 35
Calls trigger source menu.
TRIG’d (LED) 36
Lit on condition that time base is triggered.
NORM (LED) 36
Lit on condition that NORMAL or SINGLE triggering is pre-
sent.
HOLD OFF (LED) 36
Lit if a hold off time > 0% is chosen in time base menu (HOR
pushbutton
X-POS / DELAY (pushbutton switch) 36
Calls and indicates the actual function of the HORIZONTAL
knob
HORIZONTAL (knob) 37
Changes the X position resp. in digital mode the delay time
(Pre- resp. Post-Trigger).
TIME/DIV-SCALE-VAR (knob) 37
Time base A and B defl ection coeffi cient, time base variable
and scaling control.
MAG (pushbutton switch) 37
10 fold expansion in X direction in Yt mode, with simulta-
neous change of the defl ection coeffi cient display in the
readout.
).
, (X-POS = dark).
HELP (pushbutton switch) 30
Switches help texts regarding controls and menus on and
off.
POSITION 1 (knob) 30
Controls position of actual present functions: Signal (cur-
rent, reference or mathematics), Cursor and ZOOM (digital).
POSITION 2 (knob) 31
Controls position of actual present functions: Signal (current,
reference or mathematics) Cursor and ZOOM (digital).
CH 1/2-CURSOR-CH3/4-MA/REF-ZOOM (pushbutton) 32
Calls the menu and indicates the current function of
POSITION 1 and 2 controls.
VOLTS/DIV-SCALE-VAR (knob) 32
Channel 1 Y defl ection coeffi cient, Y variabel and Y scaling
setting.
8
Subject to change without notice
HOR / VAR (pushbutton switch) 38
Calls ZOOM function (digital) and analog time base A and
B, time base variable and hold off control.
CH1 (pushbutton switch) 39
Calls channel 1 menu with input coupling, inverting, probe
and Y variable control.
VERT/XY (pushbutton switch) 39
Calls vertical mode selection, addition, XY mode and band-
width limiter.
CH2 (pushbutton switch) 41
Calls channel 1 menu with input coupling, inverting, probe
and Y variable control.
CH1 (BNC-socket) 41
Channel 1 signal input and input for horizontal defl ection in
XY mode.
Front Panel Elements – Brief Description
POWER
POWER
15
13
14
17
16
18
123
INTENS
POWER
!
EXIT MENU
REMOTE OFF
POSITION 1POSITION 2
VOLTS / DIV
SCALE · VAR
20 V1 mV20 V1 mV
CH 1
VAR
X-INP
!
CAT I
FOCUS
TRACE
MENU
REM
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
INPUTS
1MΩII15pF
max
400 Vp
4
ANALOG
DIGITAL
56 789 1011 12
ANALOG
DIGITAL
MATH
RECALL
OSCILLOSCOPE
HM1008
·
1 MB
1 GSa
100 MHz
VOLTS / DIV
SCALE · VAR
CH 2HORMAG
RUNACQUIRE SETTINGSHELP
STOP
LEVEL A/B
TRIGGER
MODE
FILTER
SOURCE
AUX
X-POS
DELAY
TRIG ’d
NORM
HOLD OFF
VAR
AUXILIARY INPUT
TRIGGER
EXTERN
!
CAT I
Z-INPUT
SAVE/
AUTOSET
HORIZONTAL
TIME / DIV
SCALE · VAR
50s5ns
VAR
x10
1MΩ II
15pF
max
100 Vp
19
26
27
20
23
21
24
28
22
25
29
30
3431
32
33
35
MEMORY
COMBISCOPE
oom
CH2 (BNC-socket) 41
Channel 2 signal input.
AUX (pushbutton switch) 41
Calls AUXILIARY INPUT menu with intensity modulation (Z)
and external triggering selectable.
AUXILIARY INPUT (BNC-socket) 41
Input for external trigger or intensity (Z) modulation si-
gnal.
36
37
COMPONENT
TESTER
40
PROBE
ADJ
3839
PROBE / COMPONENT (pushbutton switch) 42
Calls COMPONENT TESTER mode settings and frequency
selection of PROBE ADJ signal.
COMPONENT TESTER (2 sockets with 4 mm Ø) 42
Connectors for test leads of the Component Tester. Left
socket is galvanically connected with protective earth.
PROBE / ADJ (socket) 42
Square wave signal output for frequency compensation of
x10 probes.
Subject to change without notice
9
Basic signal measurement
Basic signal measurement
Signals which can be measured
The following description pertains as well to analog as to
DSO operation. The different specifi cations in both operating
modes should be kept in mind.
The oscilloscope HM1008 can display all repetitive signals
with a fundamental repetition frequency of at least 100 MHz.
The frequency response is 0 to 100 MHz (-3 dB). The vertical
amplifi ers will not distort signals by overshoots, undershoots,
ringing etc.
Simple electrical signals like sine waves from line frequency
ripple to hf will be displayed without problems. However,
when measuring sine waves, the amplitudes will be displayed
with an error increasing with frequency. At 80 MHz the amplitude error will be around –10 %. As the bandwidths of individual instruments will show a certain spread (the 100 MHz
are a guaranteed minimum) the actual measurement error
for sine waves cannot be exactly determined.
Pulse signals contain harmonics of their fundamental frequency which must be represented, so the maximum useful
repetition frequency of nonsinusoidal signals is much lower
than 100 MHz. The criterion is the relationship between the
rise times of the signal and the scope; the scope’s rise time
should be <1/3 of the signal’s rise time if a faithful reproduction without too much rounding of the signal shape is to
be preserved.
The display of a mixture of signals is especially diffi cult
if it contains no single frequency with a higher amplitude
than those of the other ones as the scope’s trigger system
normally reacts to a certain amplitude. This is e.g. typical of
burst signals. Display of such signals may require using the
HOLD-OFF control.
Composite video signals may be displayed easily as the instrument has a tv sync separator.
The maximum sweep speed of 5 ns/cm allows suffi cient time
resolution, e.g. a 100 MHz sine wave will be displayed one
period per 2 cm.
The vertical amplifi er inputs may be dc or ac coupled. Use dc
coupling only if necessary and preferably with a probe.
Low frequency signals when ac coupled will show tilt (ac low
frequency – 3 dB point is 1.6 Hz), so if possible use dc coupling.
Using a probe with 10:1 or higher attenuation will lower the
–3 dB point by the probe factor. If a probe cannot be used
due to the loss of sensitivity dc coupling the scope and an
external large capacitor may help which, of course, must have
a suffi cient dc rating. Care must be taken, however, when
charging and discharging a large capacitor.
Amplitude of signals
In contrast to the general use of rms values in electrical engineering oscilloscopes are calibrated in Vpp as that is what is
displayed.
Derive rms from V
: divide by 2.84. Derive Vpp from rms: mul-
pp
tiply by 2.84.
Values of a sine wave signal
V
p
V
rms
V
mom
V
pp
V
= rms value
rms
V
= pp – value
PP
V
= momentary value, depends on time vs. period.
mom
The minimum signal for a one cm display is 1 mV
ded 1 mV/cm was selected and the variable is in the calibrated
position.
The available sensitivities are given in mV
pp
allow to indicate the amplitudes of the signals immediately on
the readout as the attenuation of probes is automatically taken
into account. Even if the probe attenuation was selected manually this will be overridden if the scope identifi es a probe with
an identifi cation contact as different. The readout will always
give the true amplitude.
It is important that the variable be in its calibrated position. The
sensitivity may be continuously decreased by using the variable
(see Controls and Readout). Each intermediate value between
the calibrated positions 1–2–5 may be selected. Without using
a probe thus a maximum of 400 V
may be displayed (20 V/div
PP
x 8 cm screen x 2.5 variable).
Amplitudes may be directly read off the screen by measuring
the height and multiplying by the V/div. setting.
Please note: Without a probe the maximum permis-
sible voltage at the inputs must not exceed 400 V
irrespective of polarity.
In case of signals with a dc content the peak value DC + AC
peak must not exceed + or – 400 V
. Pure AC of up to 800 VPP
P
is permissible.
If probes are used their possibly higher ratings are
only usable if the scope is dc coupled.
±5 % provi-
pp
or Vpp. The cursors
p
Dc coupling is preferable with all signals of varying duty cycle, otherwise the display will move up and down depending
on the duty cycle. Of course, pure dc can only be measured
with dc coupling.
The readout will show which coupling was chosen: = stands
for DC, ~ stands for AC.
10
Subject to change without notice
In case of measuring dc with a probe while the scope input is
ac coupled the capacitor in the scope input will see the input
dc voltage as it is in series with the internal 1 MΩ resistor.
This means that the maximum dc voltage (or DC + peak AC) is
that of the scope input, i.e. 400 V
! With signals which contain
P
dc and ac the dc content will stress the input capacitor while
the ac content will be divided depending on the ac impedance
of the capacitor. It may be assumed that this is negligible for
frequencies >40 Hz.
Considering the foregoing you may measure dc signals of up
to 400 V or pure ac signals of up to 800 V
with a HZ200 probe.
PP
Probes with higher attenuation like HZ53 100:1 allow to measure
dc up to 1200 V and pure ac of up to 2400 V
. (Please note the
PP
derating for higher frequencies, consult the HZ53 manual).
Stressing a 10:1 probe beyond its ratings will risk destruction of
the capacitor bridging the input resistor with possible ensuing
damage of the scope input!
In case the residual ripple of a high voltage is to be measured a
high voltage capacitor may be inserted in front of a 10:1 probe, it
will take most of the voltage as the value of the probe’s internal
capacitor is very low, 22 to 68 nF will be suffi cient.
If the input selector is switched to Ground the reference trace
on the screen may be positioned at graticule center or elsewhere.
DC and ac components of an input signal
voltage
peak
AC
DC
DC + AC
DC
AC
peak
= 400 V
max
The dashed curve shows an ac signal symmetrical to zero.
If there is a dc component the peak value will be dc + ac peak.
Timing relationships
The repetition frequency of a signal is equal to the number of
periods per second. Depending on the TIME/DIV setting one or
more periods or part of a period of the signal may be displayed.
The time base settings will be indicated on the readout in s/cm,
ms/cm, μs/cm and ns/cm. Also the cursors may be used to
measure the frequency or the period.
If portions of the signal are to be measured use delayed sweep
(analog mode) or zoom (DSO mode) or the magnifi er x 10. Use
the HORIZONTAL positioning control to shift the portion to be
zoomed into the screen center.
Pulse signals are characterized by their rise and fall times
which are measured between the 10 % and 90 % portions. The
following example uses the internal graticule of the crt, but also
the cursors may be used for measurement.
Measurement:
– Adjust the rising portion of the signal to 5 cm.
– Position the rising portion symmetrically to the graticule
centre line, using both Y and X positioning controls.
– Notice the intersections of the signal with the 10 and 90 %
lines and project these points to the centre line in order to
read the time difference.
Basic signal measurement
100%
90%
5 cm
10%
0%
t
tot
In the example it was 1.6 cm at 5 ns/cm equals 8 ns rise time.
When measuring very short rise times coming close to the scope
rise time it is necessary to subtract the scope’s (and if used the
probe’s) rise times geometrically from the rise time as seen on
the screen. The true signal rise time will become:
2
2
ta= t
t
is the rise time seen, t
tot
– t
tot
(3.5 ns with the HM1008), t
2
– t
osc
t
is the scope’s own rise time
osc
is the rise time of the probe, e.g.
t
2 ns. If the signal’s rise time is > 34 ns, the rise times of scope
and probe may be neglected.
ta= 82 - 3.52 - 22 = 6.9 ns
For the measurement of rise times it is not necessary to proceed
as outlined above. Rise times may be measured anywhere on
the screen. It is mandatory that the rising portion of the signal
be measured in full and that the 10 to 90 % are observed. In
case of signals with over- or undershoot the 0 and 100 % levels
are those of the horizontal portions of the signal, i.e. the overresp. undershoots must be disregarded for rise and fall time
measurements. Also, glitches will be disregarded. If signals
are very distorted, however, rise and fall time measurements
may be of no value.
For most amplifi ers, even if their pulse behaviour is far from
ideal, the following relationship holds:
350 350
t
=
——
a
B t
B =
——
a
tr/ns = 350/Bandwidth/MHz
Connection of signals
In most cases pressing the AUTOSET button will yield a satisfactory display (see AUTOSET). The following relates to special
cases where manual settings will be advisable. For a description
of controls refer to ”Controls and Readout“.
Take care when connecting unknown signals to the
inputs!
It is recommended to use probes whenever possible. Without
a probe start with the attenuator set to its 20 V/cm position.
If the trace disappears the signal amplitude may be too large
overdriving the vertical amplifi er or/and its dc content may be
too high. Reduce the sensitivity until the trace will reappear
onscreen. If calibrated measurements are desired it will be
necessary to use a probe if the signal becomes >160 Vp. Check
the probe specifi cations in order to avoid overstressing. If the
time base is set too fast the trace may become invisible, then
reduce the time base speed.
If no probe is used at least screened cable should be used,
such as HZ32 or HZ34. However, this is only advisable for low
Subject to change without notice
11
First time operation and initial adjustments
impedance sources or low frequencies (<50 kHz). With high
frequencies impedance matching will be necessary.
Nonsinusoidal signals require impedance matching, at both
ends preferably. At the scope input a feed through – 50 Ω-termination will be required. HAMEG offers a HZ22 termination. If
proper terminations are not used sizeable pulse aberrations will
result. Also sine wave signals of > 100 kHz should be properly
terminated. Most generators control signal amplitudes only if
correctly terminated.
HZ22 may only be used up to 7 V
or 20 VPP i.e. 1 W.
rms
For probes terminations are neither required nor allowed, they
would ruin the signal.
Probes feature very low loads at fairly low frequencies: 10 MΩ in
parallel to a few pF, valid up to several hundred kHz. However,
the input impedance diminishes with rising frequency to quite
low values. This has to be borne in mind as probes are, e.g.,
entirely unsuitable to measure signals across high impedance
high frequency circuits such as bandfi lters etc.! Here only FET
probes can be used. Use of a probe as a rule will also protect
the scope input due to the high probe series resistance (9 MΩ).
As probes cannot be calibrated exactly enough during manufacturing individual calibration with the scope input used is
mandatory! (See Probe Calibration).
Passive probes will, as a rule, decrease the scope bandwidth
resp. increase the rise time. We recommend to use HZ200 probes in order to make maximum use of the combined bandwidth.
HZ200 features 2 additional hf compensation adjustments.
First time operation and initial adjustments
Prior to fi rst time operation the connection between the instrument and safety ground must be ensured, hence the plug must
be inserted fi rst.
Use the red pushbutton POWER to turn the scope on. Several
displays will light up. The scope will then assume the set-up,
which was selected before it was turned off. If no trace and
no readout are visible after approximately 20 sec, push the
AUTOSET button.
As soon as the trace becomes visible select an average intensity with INTENS, then select FOCUS and adjust it, then select
TRACE ROTATION and adjust for a horizontal trace.
With respect to crt life use only as much intensity as necessary
and convenient under given ambient light conditions, if unused
turn the intensity fully off rather than turning the scope off and
on too much, this is detrimental to the life of the crt heater.
Do not allow a stationary point to stay, it might burn the crt
phosphor.
With unknown signals start with the lowest sensitivity 20 V/cm,
connect the input cables to the scope and then to the measuring object which should be deenergized in the beginning. Then
turn the measuring object on. If the trace disappears, push
AUTOSET.
Whenever the DC content is > 400 V
coupling must be used in
DC
order to prevent overstressing the scope input capacitor. This is
especially important if a 100:1 probe is used as this is specifi ed
for 1200 V
+ peak AC.
DC
AC coupling of low frequency signals may produce tilt.
If the dc content of a signal must be blocked it is possible to
insert a capacitor of proper size and voltage rating in front of the
probe, a typical application would be a ripple measurement.
When measuring small voltages the selection of the ground
connection is of vital importance. It should be as close to voltage
take-off point as possible, otherwise ground currents may deteriorate the measurement. The ground connections of probes
are especially critical, they should be as short as possible and
of large size.
If a probe is to be connected to a BNC connector use
a probe tip to BNC adapter.
If ripple or other interference is visible, especially at high sensitivity, one possible reason may be multiple grounding. The
scope itself and most other equipment are connected to safety
ground, so ground loops may exist. Also, most instruments will
have capacitors between line and safety ground installed which
conduct current from the live wire into the safety ground.
Trace rotation TR
The crt has an internal graticule. In order to adjust the defl ected
beam with respect to this graticule the Trace Rotation control
is provided. Select the function Trace Rotation and adjust for a
trace which is exactly parallel to the graticule.
Probe adjustment and use
In order to ensure proper matching of the probe used to the
scope input impedance the scope contains a calibrator with
short rise time and an amplitude of 0.2 V
± 1 %, equivalent to
pp
4 cm at 5 mV/cm when using 10:1 probes.
The inner diameter of the calibrator connector is 4.9 mm and
standardized for series F probes. Using this special connector is the only way to connect a probe to a fast signal source
minimizing signal and ground lead lengths and to ensure true
displays of pulse signals.
1 kHz adjustment
This basic adjustment will ensure that the capacitive attenuation
equals the resistive attenuation thus rendering the attenuation
of the probe independent of frequency. 1:1 probes can not be
adjusted and need no such adjustment anyway.
12
Subject to change without notice
incorrect correct incorrect
Operating modes of the vertical amplifier
Prior to adjustment make sure that the trace rotation adjustment was performed.
Connect the 10:1 probe to the input. Use dc coupling. Set
the VOLTS/DIV to 5 mV/cm and TIME/DIV to 0.2 ms/cm, both
calibrated. Insert the probe tip into the calibrator connector
PROBE ADJ.
You should see 2 signal periods. Adjust the compensation capacitor (see the probe manual for the location) until the square
wave tops are exactly parallel to the graticule lines (see picture
1 kHz). The signal height should be 4 cm ±1.6 mm (3% oscilloscope and 1% probe tolerance). The rising and falling portions
of the square wave will be invisible.
1 MHz adjustment
The HAMEG probes feature additional adjustments in the
compensation box which allow to optimise their hf behaviour.
This adjustment is a precondition for achieving the maximum
bandwidth with probe and a minimum of pulse aberrations.
This adjustment requires a calibrator with a short rise time (typ.
4 ns) and a 50 Ω output, a frequency of 1 MHz, an amplitude
of 0.2 V
requirements.
Connect the probe to the scope input to which it is to be adjusted.
Select the PROBE ADJ. signal 1 MHz. Select dc coupling and
5 mV/cm with VOLTS/DIV. and 0.1 μs/cm with TIME/DIV., both
calibrated. Insert the probe tip into the calibrator output connector. The screen should show the signal, rise and fall times will
be visible. Watch the rising portion and the top left pulse corner,
consult the manual for the location of the adjustments.
. The PROBE ADJ. output of the scope fulfi ls these
PP
Operating modes of the vertical amplifi er
The controls most important for the vertical amplifi er are:
VERT/XY
containing the operating modes and the parameters of the
individual channels.
Changing the operating mode is described in the chapter:
”Controls and Readout“.
Remark: Any reference to ”both channels“ always refers to
channels 1 and 2.
Usually oscilloscopes are used in the Yt mode. In analog mode
the amplitude of the measuring signal will defl ect the trace
vertically while a time base will defl ect it from left to right.
The vertical amplifi ers offer these modes:
– One signal only with CH1.
– One signal only with CH2.
– Two signals with channels 1 and 2 (DUAL trace mode)
In DUAL mode both channels are operative. In analog mode
the method of signal display is governed by the time base (see
also ”Controls and Readout“). channel switching may either
take place after each sweep (alternate) or during sweeps with
a high frequency (chopped).
The normal choice is alternate, however, at slow time base settings the channel switching will become visible and disturbing,
when this occurs select the chopped mode in order to achieve
a stable quiet display.
, CH1 , CH2 . They give access to the menus
incorrect correct incorrect
The criteria for a correct adjustment are:
– short rise time, steep slope.
– clean top left corner with minimum over- or undershoot,
fl at top.
After adjustment check the amplitude which should be the
same as with 1 kHz.
It is important to fi rst adjust 1 kHz, then 1 MHz. It may be necessary to check the 1 kHz adjustment again.
Please note that the calibrator signals are not calibrated with
respect to frequency and thus must not be used to check the
time base accuracy, also their duty cycle may differ from 1:1.The
probe adjustment is completed if the pulse tops are horizontal
and the amplitude calibration is correct.
In DSO mode no channel switching is necessary as each input
has its own A/D converter, signal acquisition is simultaneous.
In ADD mode the two channels 1 and 2 are algebraically added (±CH1 ±CH2). With + polarity the channel is normal, with
– polarity inverted. If + Ch1 and – CH2 are selected the difference
will be displayed or vice versa.
Same polarity input signals:
Both channels not inverted: = sum
Both channels inverted: = sum
Only one channel inverted: = difference
Opposite polarity input signals:
Both channels not inverted: = difference
Both channels inverted: = difference
One channel inverted: = sum.
Please note that in ADD mode both position controls will be
operative. The INVERT function will not affect positioning.
Often the difference of two signals is to be measured at signal
take-offs which are both at a high common mode potential.
While this one typical application of the difference mode one
important precaution has to be borne in mind: The oscilloscope vertical amplifi ers are two separate amplifi ers and do not
constitute a true difference amplifi er with as well a high CM
rejection as a high permissible CM range! Therefore please
observe the following rule: Always look at the two signals in
the one channel only or the dual modes and make sure that
they are within the permissible input signal range; this is the
case if they can be displayed in these modes. Only then switch
to ADD. If this precaution is disregarded grossly false displays
Subject to change without notice
13
Operating modes of the vertical amplifier
may result as the input range of one or both amplifi ers may
be exceeded.
Another precondition for obtaining true displays is the use of
two identical probes at both inputs. But note that normal probe
tolerances (percent) will cause the CM rejection to be expected
to be rather moderate. In order to obtain the best possible results proceed as follows: First adjust both probes as carefully
as possible, then select the same sensitivity at both inputs and
then connect both probes to the output of a pulse generator
with suffi cient amplitude to yield a good display. Readjust one
(!) of the probe adjustment capacitors for a minimum of overor undershoot. As there is no adjustment provided with which
the resistors can be matched a residual pulse signal will be
unavoidable.
When making difference measurements it is good practice
to fi rst connect the ground cables of the probes to the object
prior to connecting the probe tips. There may be high potentials
between the object and the scope. If a probe tip is connected
fi rst there is danger of overstressing the probe or/and the scope
inputs! Never perform difference measurements without both
probe ground cables connected.
XY operation
This mode is accessed by VERT/XY > XY. In analog mode
the time will be turned off. The channel 1 signal will defl ect in X
direction (X-INP. = horizontal input), hence the input attenuators,
the variable and the POSITION 1 control will be operative. The
HORIZONTAL control will also remain functional.
Channel 2 will defl ect in Y direction.
Please note:
– As the trigonometric functions are
periodic limit the calculation to
angles <90 degrees. This is where
this function is most useful.
– Do not use too high frequencies,
because, as explained above, the
two amplifi ers are not identical,
their phase difference increases with frequency. The spec
gives the frequency at which the phase difference will stay
<3 degrees.
– The display will not show which of the two frequencies does
lead or lag. Use a CR combination in front of the input of the
frequency tested. As the input has a 1 MΩ resistor it will be
suffi cient to insert a suitable capacitor in series. If the ellipse
increases with the C compared to the C short-circuited the
test signal will lead and vice versa. This is only valid <90
degrees. Hence C should be large and just create a barely
visible change.
If in XY mode one or both signals disappear, only a line or a point
will appear, mostly very bright. In case of only a point there is
danger of phosphor burn, so turn the intensity down immediately; if only a line is shown the danger of burn will increase the
shorter the line is. Phosphor burn is permanent.
a
sin ϕ =
b
a
cos ϕ = 1 – (—
b
a
ϕ = arc sin
b
—
2
)
—
Measurement of phase differences in dual channel
Yt mode
Please note: Do not use ”alternate trigger“ because the time
differences shown are arbitrary and depend only on the respective signal shapes! Make it a rule to use alternate trigger only
in rare special cases.
The x 10 magnifi er will be inoperative in XY mode. Please note
the differences in the Y and X bandwidths, the X amplifi er has
a lower – 3 dB frequency than the Y amplifi er. Consequently
the phase difference between X and Y will increase with frequency.
In XY mode the X signal (CH1 = X-INP). can not be inverted.
The XY mode may generate Lissajous fi gures which simplify
some measuring tasks and make others possible:
– Comparison of two signals of different frequency or adju-
stment of one frequency until it is equal to the other resp.
becomes synchronized.
– This is also possible for multiples or fractions of one of the
frequencies.
Phase measurements with Lissajous fi gures
The following pictures show two sine waves of equal amplitude
and frequency but differing phase.
ab
0° 35° 90° 180°
Calculation of the phase angle between the X- and Y-signals (after reading a and b off the screen) is possible using the following
formulas and a pocket calculator with trigonometric functions.
This calculation is independent of the signal amplitudes:
The best method of measuring time or phase differences is using
the dual channel Yt mode. Of course, only times may be read off
the screen, the phase must then be calculated as the frequency
is known. This is a much more accurate and convenient method
as the full bandwidth of the scope is used, and both amplifi ers
are almost identical. Trigger the time base from the signal
which shall be the reference. It is necessary to position both
traces without signal exactly on the graticule center (POSITION
1 and 2). The variables and trigger level controls may be used,
this will not infl uence the time difference measurement. For
best accuracy display only one period at high amplitude und
observe the zero crossings. One period equals 360 degrees.
It may be advantageous to use ac coupling if there is an offset
in the signals.
t = horizontal spacing of the
zero transitions in div
T= horizontal spacing for one
period in div
In this example t = 3 cm and T = 10 cm, the phase difference in
degrees will result from:
5 3
ϕ° =
—
T 10
or in angular units:
t 3
arc ϕ° =
T 10
· 360° = — · 360° = 108°
—
· 2π = — · 2π = 1,885 rad
14
Subject to change without notice
Triggering and time base
Very small phase differences with moderately high frequencies
may yield better results with Lissajous fi gures.
However, in order to get higher precision it is possible to switch
to higher sensitivities – after accurately positioning at graticule
centre – thus overdriving the inputs resulting in sharper zero
crossings. Also, it is possible to use half a period over the full
10 cm. As the time base is quite accurate increasing the time
base speed after adjusting for e.g. one period = 10 cm and
positioning the fi rst crossing on the fi rst graticule line will also
give better resolution.
Measurement of amplitude modulation
Please note: Use this only in analog mode because in DSO mode
alias displays may void the measurement! For the display of
low modulation frequencies a slow time base (TIME/DIV) has
to be selected in order to display one full period of the modulating signal. As the sampling frequency of any DSO must be
reduced at slow time bases it may become too low for a true
representation.
The momentary amplitude at time t of a hf carrier frequency
modulated by a sinusoidal low frequency is given by:
u = UT · sinΩt + 0,5 m · UT · cos (Ω - ω) t - 0,5 m · UT · cos (Ω - ω) t
where: UT = amplitude of the unmodulated carrier
Ω = 2πF = angular carrier frequencyω = 2πf = modulation angular frequencym = modulation degree (≤1
In addition to the carrier a lower side band F – f and an upper
side band F + f will be generated by the modulation.
U
T
v100%)
Set the scope controls as follows in order to display the picture
2 signal:
CH1 only, 20 mV/cm, AC
TIME/DIV: 0.2 ms/cm
Triggering: NORMAL, AC, internal.
Use the time base variable or external triggering.
Reading a and b off the screen the modulation degree will
result:
a – b a – b
m =
——
a + b a + b
(1 + m) and b = UT (1 – m)
a = U
T
bzw. m =
—— · 100 [%]
When measuring the modulation degree the amplitude and time
variables can be used without any infl uence on the result.
Triggering and time base
The most important controls and displays for these functions
are to be found in the shaded TRIGGER area, they are described
in „Controls and Readout“.-
In YT mode the signal will defl ect the trace vertically while the
time will defl ect it horizontally, the speed can be selected.
In general periodic voltage signals are displayed with a periodically repeating time base. In order to have a stable display
successive periods must trigger the time base at exactly the
same time position of the signal (amplitude and slope).
0,5 m · U
T
0,5 m · U
T
F – f F F + f
Picture 1: Amplitudes and frequencies with AM (m = 50 %) of
the spectra
As long as the frequencies involved remain within the scope’s
bandwidth the amplitude-modulated hf can be displayed. Preferably the time base is adjusted so that several signal periods
will be displayed. Triggering is best done from the modulation
frequency. Sometimes a stable displayed can be achieved by
twiddling with the time base variable.
m · U
T
U
T
ba
Pure DC can not trigger the time base, a voltage
change is necessary.
Triggering may be internal from any of the input signals or
externally from a time-related signal.
For triggering a minimum signal amplitude is required which
can be determined with a sine wave signal. With internal triggering the trigger take-off within the vertical amplifi ers is directly
following the attenuators. The minimum amplitude is specifi ed
in mm on the screen. Thus it is not necessary to give a minimum
voltage for each setting of the attenuator.
For external triggering the appropriate input connector is used,
the amplitude necessary there is given in V
. The voltage for
pp
triggering may be much higher than the minimum, however, it
should be limited to 20 times the minimum. Please note that
for good triggering the voltage resp. signal height should be a
good deal above the minimum. The scope features two trigger
modes to be described in the following:
Automatic peak triggering (MODE menu)
Consult the chapters MODE > AUTO, LEVEL A/B , FILTER
and SOURCE in ”Controls and Readout“. Using AUTOSET
this trigger mode will be automatically selected. With DCcoupling and with alternate trigger this mode will be left while
the automatic triggering will remain.
Picture 2: Amplitude modulated HF. F = 1 MHz, f = 1 kHz,
m = 50 %, U
= 28,3 mV
T
rms
Automatic triggering causes a new time base start after the end
of the foregoing and after the hold-off time has elapsed even
Subject to change without notice
15
Triggering and time base
without any input signal. Thus there is always a visible trace in
analog mode, and in DSO mode the trace will also be shown.
The position of the trace(s) without any signal is then given by
the settings of the POSITION controls.
As long as there is a signal scope operation will not need more
than a correct amplitude and time base setting. With signals
< 20 Hz their period is longer than the time the auto trigger
circuit will wait for a new trigger, consequently the auto trigger
circuit will start the time base then irrespective of the signal so
that the display will not be triggered and free run, quite independent of the signal’s amplitude which may be much larger
than the minimum.
Also in auto peak trigger mode the trigger level control is active.
Its range will be automatically adjusted to coincide with the
signal’s peak-to-peak amplitude, hence the name. The trigger
point will thus become almost independent of signal amplitude.
This means that even if the signal is decreased the trigger will
follow, the display will not loose trigger. As an example: the
duty cycle of a square wave may change between 1:1 and 100:1
without loosing the trigger.
Depending on the signal the LEVEL A/B control may have to be
set to one of its extreme positions.
The simplicity of this mode recommends it for most uncomplicated signals. It is also preferable for unknown signals.
negative portion of a signal. This is valid in automatic and
normal modes.
Trigger coupling (Menu: FILTER)
Consult chapters: MODE > AUTO, LEVEL A/B , FILTER
and SOURCE
coupling will be used unless ac coupling was selected before.
The frequency responses in the diverse trigger modes may be
found in the specifi cations.
With internal dc coupling with or without LF fi lter use normal
triggering and the level control. The trigger coupling selected
will determine the frequency response of the trigger channel.
AC:
This is the standard mode. Below and above the fall-off of the
frequency response more trigger signal will be necessary.
DC:
With direct coupling there is no lower frequency limit, so this
is used with very slowly varying signals. Use normal triggering
and the level control. This coupling is also indicated if the signal
varies in its duty cycle.
HF:
A high pass is inserted in the trigger channel, thus blocking low
frequency interference like fl icker noise etc.
in ”Controls and Readout“. In AUTOSET DC
This trigger mode is independent of the trigger source and
usable as well for internal as external triggering. But the signal
must be > 20 Hz.
Normal trigger mode (See menu MODE)
Consult the chapters: MODE > AUTO, LEVEL A/B , FILTER
and SOURCE in ”Controls and Readout“. Information
about how to trigger very diffi cult signals can be found in the
HOR menu
VAR, HOLD-OFF time setting, and time base B operation are
explained.
With normal triggering and suitable trigger level setting triggering may be chosen on any point of the signal slope. Here, the
range of the trigger level control depends on the trigger signal
amplitude. With signals <1 cm care is necessary.
In normal mode triggering there will be no trace visible in the
absence of a signal or when the signal is below the minimum
trigger amplitude requirement!
Normal triggering will function even with complicated signals. If
a mixture of signals is displayed triggering will require repetition
of amplitudes to which the level can be set. This may require
special care in adjustment.
where the functions time base fi ne adjustment
Slope selection (Menu FILTER)
After entering FILTER the trigger slope may be selected using
the function keys. See also ”Controls and Readout“. AUTOSET
will not change the slope.
Noise Reject:
This trigger coupling mode or fi lter is a low pass suppressing
high frequencies. This is useful in order to eliminate hf interference of low frequency signals. This fi lter may be used in
combination with dc or ac coupling, in the latter case very low
frequencies will also be attenuated.
LF:
This is also a low pass fi lter with a still lower cut-off frequency
than above which also can be combined with dc or ac coupling.
Selecting this fi lter may be more advantageous than using dc
coupling in order to suppress noise producing jitter or double
images. Above the pass band the necessary trigger signal will
rise. Together with ac coupling there will also result a low
frequency cut-off.
Video (tv triggering)
Selecting MODE > Video will activate the TV sync separator
built-in. It separates the sync pulses from the picture content
and enables thus stable triggering independent of the changing
video content.
Composite video signals may be positive or negative. The
sync pulses will only be properly extracted if the polarity
is right. The definition of polarity is as follows: if the video
is above the sync it is positive, otherwise it is negative. The
polarity can be selected after selecting FILTER. If the polarity is wrong the display will be unstable resp. not triggered
at all as triggering will then initiated by the video content.
With internal triggering a minimum signal height of 5 mm
is necessary.
Positive or negative slope may be selected in auto or normal
trigger modes. Also, a setting ”both“ may be selected which will
cause a trigger irrespective of the polarity of the next slope.
Rising slope means that a signal comes from a negative potential and rises towards a positive one. This is independent
of the vertical position. A positive slope may exist also in the
16
Subject to change without notice
The PAL sync signal consists of line and frame signals which
differ in duration. Pulse duration is 5 μs in 64 μs intervals. Frame
sync pulses consist of several pulses each 28 μs repeating each
half frame in 20 ms intervals.
Both sync pulses differ hence as well in duration as in their
repetition intervals. Triggering is possible with both.
Triggering and time base
Frame sync pulse triggering
Remark:
Using frame sync triggering in dual trace chopped mode may
result in interference, then the dual trace alternate mode
should be chosen. It may also be necessary to turn the readout off.
In order to achieve frame sync pulse triggering call MODE,
select video signal triggering and then FILTER to select frame
triggering. It may be selected further whether ”all“, ”only even“
or ”only odd“ half frames shall trigger. Of course, the correct tv
standard must be selected fi rst of all (625/50 or 525/60).
The time base setting should be adapted, with 2 ms/cm a complete half frame will be displayed. Frame sync pulses consist
of several pulses with a half line rep rate.
Line sync pulse triggering
In order to choose line snyc triggering call MODE and select
VIDEO, enter FILTER, make sure that the correct video standard
is selected (625/50 or 525/60) and select Line.
If ALL was selected each line sync pulse will trigger. It is also
possible to select a line number ”LINE No.“.
In order to display single lines a time base setting of TIME/DIV.
= 10 μs/cm is recommended, this will show 1½ lines. In general
the composite video signal contains a high dc component which
can be removed by ac coupling, provided the picture is steady.
Use the POSITION control to keep the display within the screen.
If the video content changes like with a regular tv program only
dc coupling is useful, otherwise the vertical position would
continuously move.
The sync separator is also operative with external triggering.
Consult the specifi cations for the permissible range of trigger
voltage. The correct slope must be chosen as the external
trigger may have a different polarity from the composite video.
In case of doubt display the external trigger signal.
LINE trigger
Consult SOURCE in ”Controls and Readout“ for specifi c
information.
If the readout shows Tr:Line the trigger signal will be internally
taken from the line (50 or 60 Hz).
Alternate trigger
This mode is selected with SOURCE > Alt. 1/2. The readout will display Tr:alt, but no more the trigger point symbol
indicating level and time position. Instead an arrow pointing
upwards will indicate the trigger time position if this lies within
the screen area.
This trigger mode is to be used with greatest care and should be
an exception rather than the rule, because the time relationships
visible on the screen are completely meaningless, they depend
only on the shape of the signals and the trigger level!
In this mode the trigger source will be switched together with
the channel switching, so that when CH1 is displayed in the
dual channel alternate mode the trigger is taken from CH1
and when CH2 is displayed the trigger is taken from CH2. This
way two uncorrelated signals can be displayed together. If this
mode is inadvertently chosen the time relationships between
the signals will also be lost when both signals are correlated!
(Except for the special case that both happen to be square waves
with extremely fast rise times). Of course, this trigger mode is
only possible in the dual channel alternate mode and also not
with external or line trigger. Ac coupling is recommended for
most cases.
External triggering
In analog mode this trigger mode may be selected with SOURCE
> Extern. In DSO mode it is only possible if channels 3 and
4 are turned off. The readout will display Tr:ext. CH4
the input for the external trigger, all internal sources will be
disconnected. In this mode the trigger point symbol (level and
time position) will not be displayed, only the trigger time position
will be indicated. External triggering requires a signal of 0.3 to
3 V
, synchronous to the vertical input signal(s).
PP
Triggering will also be possible within limits with multiples or
fractions of the vertical input signal frequency. As the trigger
signal may have any polarity it may happen that the vertical
input signal will start with a negative slope in spite of having
selected positive slope; slope selection refers now to the external trigger.
will be
Indication of triggered operation (TRIG’D LED)
Refer item in ”Controls and Readout“. The LED labelled
TRIG’D indicates triggered operation provided:
– Suffi cient amplitude of the internal or external trigger signal.
– The trigger point symbol is not above or below the signal.
This trigger signal is independent of the scope input signals and
is recommended for all signals synchronous with the line. Within
limits this will also be true for multiples or fractions of the line
frequency. As the trigger signal is taken off internally there is
no minimum signal height on the screen for a stable display.
Hence even very small voltages like ripple or line frequency
interference can be measured.
Please note that with line triggering the polarity switching will
select either the positive or negative half period of the line, not
the slope. The trigger level control will move the trigger point
over most of a half wave.
Line frequency interference may be checked using a search
coil which preferably should have a high number of turns and
a shielded cable. Insert a 100 Ω resistor between the center
conductor and the BNC connector. If possible the coil should
be shielded without creating a shorted winding.
If these conditions are met the trigger comparator will output
triggers to start the time base and to turn on the trigger indication. The trigger indicator is helpful for setting the trigger
up, especially with low frequency signals (use normal trigger)
resp. very short pulses.
The trigger indication will store and display triggers for 100 ms.
With signals of very low rep rate the indicator will fl ash accordingly. If more than one signal period is shown on the screen
the indicator will fl ash each period.
Hold-off time adjustment
Consult ”Controls and Readout“ HOR > Hold-off time for
specifi c information.
After the time base defl ected the trace from left to right the trace
will be blanked so the retrace is invisible. The next sweep will,
however, not immediately start. Time is required to perform
Subject to change without notice
17
Triggering and time base
internal switching, so the next start is delayed for the so called
hold-off time, irrespective of the presence of triggers. The
hold-off time can be extended from its minimum by a factor of
10:1. Manipulation of the hold-off time and thus of the time for
a complete sweep period from start to start can be useful e.g.
when data packets are to be displayed. It may seem that such
signals can not be triggered. The reason is that the possible
start of a new sweep does not conincide with the start of a data
packet, it may start anywhere, even before a data packet. By
varying the hold-off time a stable display will be achieved by
setting it just so that the hold-off ends before the start of a data
packet. This is also handy with burst signals or non-periodic
pulse trains.
A signal may be corrupted by noise or hf interference so a double
display will appear. Sometimes varying the trigger level can not
prevent the double display but will only affect the apparent time
relationship between two signals. Here the variable hold-off
time will help to arrive at a single display.
Sometimes a double display will appear when a pulse signal
contains pulses of slightly differing height requiring delicate
trigger level adjustment. Also here increasing the hold-off
time will help.
Whenever the hold-off time was increased it should reset to its
minimum for other measurements, otherwise the brightness
will suffer as the sweep rep rate will not be maximum. The
following pictures demonstrate the function of the hold-off:
heavy parts are displayed
period
signal
be defl ected vertically by the input signal. In fact the input signal
does continuously defl ect the trace vertically, but this will be
only visible during the unblanking time. This is, by the way, one
marked difference to DSO operation where the input signal
is only measured during the acquisition time, for most of the
time the DSO will not see the signal. Also, in analog mode the
signal itself will be seen on the screen in real time, whereas
a DSO can only show a reconstruction of the signal acquired
some time later.
In analog mode thus the display will always start on the left. Let
us assume one period of a signal is displayed at a convenient
time base setting. Increasing the sweep speed with TIME/DIV.
will expand the display from the start, so that parts of the signal
will disappear from the screen. It is thus possible to expand
the beginning of the signal period and show fi ne detail, but it
is impossible to show such fi ne detail for ”later“ parts of the
signal.
The x10 Magnifi er (MAG x10) may be used to expand the display
and the horizontal positioning control can shift any part of the
display into the centre, but the factor of 10 is fi xed.
The solution requires a second time base, called time base B.
In this mode time base A is called the delaying sweep and
time base B the delayed sweep. The signal is fi rst displayed
by TB A alone. Then TB B is also turned on which is the mode
”A intensifi ed by B“. TB B should always be set to a higher sweep
rate than A, thus its sweep duration will be also shorter than
that of A. The TB A sweep sawtooth is compared to a voltage
which can be varied such that TB A functions as a precision
time delay generator. Depending on the amplitude of the comparison voltage a signal is generated anywhere between sweep
start and end.
sweep
Fig. 1
adjusting
HOLD OFF time
Fig. 2
Fig. 1: Display with minimum hold-off time (basic setting).
Double image, no stable display.
Fig. 2: By increasing the hold-off a stable display is achieved.
Time base B (2nd time base). Delaying, Delayed
Sweep. Analog mode
Consult ”Controls and Readout“ HOR and TIME/DIV. for
specifi c information.
As was described in ”Triggering and time base“ a trigger will
start the time base. While waiting for a trigger – after runout
of the hold-off time – the trace will remain blanked. A trigger
will cause trace unblanking and the sweep ramp which defl ects
the trace from left to right with the speed set with TIME/DIV.
At the end of the sweep the trace will be blanked again and
retrace to the start position. During a sweep the trace will also
In one of two operating modes this signal will start TB B immediately. The TB A display will be intensifi ed for the duration of
TB B, so that one sees which portion of the signal is covered by
TB B, By varying the comparison voltage the start of TB B can
be moved over the whole signal as it is displayed by TB A. Then
the mode is switched to TB B. The signal portion thus selected is
now displayed by TB B. This is called „B delayed by A“. Portions
of the signal can thus be expanded enormously, however, the
higher the speed of TB B the darker the display will become as
the rep rate will remain that of the accepted signal triggers while
the duration of TB B is reduced with increasing speed.
In cases where there is jitter the TB B can be switched to wait
for a trigger rather than starting immediately. When a trigger
arrives TB B will start on it. The jitter is removed, however, the
effect is also, that the TB B start now can be only from signal
period to signal period, no continuous adjustment is possible
in this mode.
Alternate sweep
In this mode the signal is displayed twice, with both time bases.
An artifi cial Y offset can be added in order to separate the two
displays on the screen. The operation is analogous to Y dual
trace alternate mode, i.e., the signal is alternately displayed by
both time bases, not simultaneously which is not possible with
a single gun crt. What was said above about how TB B can be
started holds also here.
18
Subject to change without notice
Component Tester
AUTOSET
For specific information consult ”Controls and Readout“
AUTOSET
The following description is valid for both analog and DSO
modes. AUTOSET does not change from analog to DSO mode
or vice versa. If in DSO mode the modes ”Roll“, ”Envelope“ or
”Average“ (ACQUIRE) are present or the trigger mode „Single“
(MODE) is selected, theses modes will be switched off as AUTOSET always switches to ”Refresh“ acquistion. The signal to
be displayed must meet the amplitude and frequency requirements of automatic triggering, to enable a useful automatic
instrument setting.
All controls except for the POWER switch are electronically
scanned, all functions can also be controlled by the microcomputer, i.e. also via the interfaces.
This is a precondition for AUTOSET as this function must be able
to control all functions independent of control settings. AUTOSET will always switch to YT mode, but preserve the previous
selection of CH1, CH2 or dual trace; ADD or XY modes will be
switched to dual trace Yt.
Automatic setting of the vertical sensitivities and the time base
will present a display within 6 cm height (4 cm per signal in dual
trace) and about 2 signal periods. This is true for signals not
differing too much from a 1:1 duty cycle. For signals containing
several frequencies like video signals the display may be any.
.
Component Tester
Specifi c information can be found in ”Controls and Readout“ under COMPONENT/PROBE
The scope has a built-in component tester. The test object is
connected with 4 mm banana plugs. In this mode the Y amplifi ers
and the time base are turned off. Only individual components
may be tested, i.e. they must not be part of a circuit, if voltages
are to be applied to the BNC connectors. If the components
are part of a circuit this must be deenergized and disconnected
from safety ground. Except for the two test leads there may
be no further connection between scope and component. (See
”Tests within a circuit“). As described in section ”Safety“ all
ground connections of the scope are connected to safety ground
including those of the component tester. As long as individual
components are tested this is of no consequence.
The display can only be affected by the controls contained in
the FOCUS/TRACE menu: A-Int., Focus, Trace rotation, HORIZONTAL position.
If components are to be tested which are parts of a circuit or
an instrument those circuits resp. instruments must fi rst be
deenergized. If they are connected to the mains they must be
unplugged. This will prevent a connection between scope and
circuit via the safety ground which may affect the measurement.
Do not test charged capacitors.
and COMPONENT TESTER .
Initiating the AUTOSET function will set the following operating
conditions:
– last selection of ac or dc coupling
– internal triggering
– automatic triggering
– trigger level set to the center of its range
– calibrated Y sensitivities
– calibrated time base
– AC or DC trigger coupling unmodifi ed
– HF trigger coupling switched to DC
– LF or Noise Reject fi lters left
– X magnifi er switched off
– Y and X positioning automatic
Please note:
For pulse signals with duty cycles approaching 400:1 no automatic signal display will be possible.
In such cases switch to normal trigger mode and set the trigger
position about 5 mm above the centre. If the trigger LED will then
light up a trigger is generated and the time base is operating. In
order to obtain a visible display it may be necessary to change
the time base and V/DIV settings. Depending on the duty cycle
and the frequency the signal may still remain invisible. This
applies only to analog mode. In DSO mode the trace is always
of equal brightness because not the signal is shown but a low
frequency construction of it, also, there is no information in the
trace intensity.
The principle of the test is very simple: a sine wave generator
within the scope generates a 50 Hz ±10 % voltage which is
applied to a series connection of the test object and a resistor
within the scope. The sine wave proper defl ects in X direction,
the voltage across the resistor which is proportional to the test
current defl ects in Y direction.
If the object contains neither capacitors nor inductors, there will
be no phase shift between voltage and current, so a straight
line will show up which will be more or less slanted, depending
on the value of the object’s resistance, covering appr. 20 Ω to
4.7 kΩ. If there is a short the trace will be vertical, i.e. (almost)
no voltage produces already high current. A horizontal line will
thus indicate an open, there is only voltage but no current.
Capacitors or inductors will create ellipses. The impedance
may be calculated from the ellipse’s geometric dimensions.
Capacitors of appr. 0.1μ to 1000 μF will be indicated.
– An ellipse with its longer axis horizontal indicates a high
impedance (low capacitance or high inductance)
– An ellipse with its longer axis vertical will indicate a low
impedance (high capacitance or low inductance)
– A slanted ellipse will indicate a lossy capacitor or induc-
tor.
Semiconductors will show their diode characteristics, however,
only 20 V
ristics can only be displayed up to 10 V
are available, so the forward and reverse characte-
pp
in each direction. The
p
test is a two-terminal test, hence it is not possible to measure
e.g. the current gain of a transistor. One can only test B-C, B-E,
and C-E. The test current is only a few mA, so the test will not
harm ordinary semiconductors. (Sensitive devices like delicate
HF transistors etc. should not be tested). The limitation to 10
Subject to change without notice
19
Component Tester
Vp with bipolar transistors will suffi ce mostly as usual defects
will show up.
The best method to verify whether a component is defective is
the comparison to a good one. If the lettering of a component
is not legible at least it is possible to see whether it is a npn or
pnp transistor or which end of a diode is the cathode.
Please note that reversing the test leads will also invert the
picture, i.e. turn it 180 degrees.
In most cases, e.g. with service and repair, it will be suffi cient
to receive a good/bad result (open, short). With MOS components the usual precautions are to be observed, but note, that
except for a possible short MOSFETs and JFETs can not be
suffi ciently tested. Indications to be expected depend strongly
on the kind of FET:
– With depletion type MOSFETs and all JFETs the channel
will conduct if prior to testing the gate was connected to the
source. The Rdson will be shown. As this can be very low it
may look like a plain short although the part is good!
– With enhancement type MOSFETs an open will be seen in
all directions, as the threshold voltage G – S is not available.
With power MOSFETs the antiparallel diode S – D can be
seen.
Tests of components within circuits are possible in many cases
but less indicative because other components may be in parallel.
But also here the comparison with a good circuit might help. As
both circuits must be deenergized it is only necessary to switch
the test leads back and forth between both in order to localize
a defective spot. Sometimes like with stereo amplifi ers, pushpull circuits, bridge circuits there is a comparison circuit right
on the same board. In cases of doubt one component lead can
be unsoldered, the other one should then be connected to the
ground lead. This is labelled with a ground symbol. The pictures
show some practical examples:
20
Subject to change without notice
CombiScope
CombiScope
®
HAMEG oscilloscopes are either analog or they are CombiScopes, i.e. they contain a complete analog scope and the additional
hardware and software so sample and digitize the signals. The
®
HM1008 is a 100 MHz 1 GS/s CombiScope
®
With a HAMEG CombiScope
the user is always sure: he needs
.
only to switch from DSO to analog in order to see the true signal.
This is especially important when a signal is to be documented
in DSO mode. The user of a pure DSO needs to know the signal
better than the scope!
The advantages of DSO operation are:
– Capture and storage of single events
– There will be no fl icker with very low frequency signals
– Fast signals with a low rep rate resp. low duty cycle can be
displayed at high intensity
– Due to the storage of all signals they may be easily docu-
mented and processed.
– Due to the down conversion of high frequencies into the kHz
area the very expensive analog crts are replaced by cheap
computer monitor tubes or lcd displays. Also, the bulk part
of the components may be the same as used in computers,
thus inexpensive.
The disadvantages of DSO operation are:
– An analog scope displays the signal itself in real time. In
a DSO the signal is not displayed but only a low frequency
reconstruction of the signal. The limitations and problems
of sampling operation as well as those of analog/digital
conversion hold. The display can not be in real time as, after
capturing a signal, the DSO must take time out to perform
calculations the result of which will then be displayed much
later.
– Therefore the capture rate of ordinary DSOs is orders of
magnitude lower than that of any analog scope. Hence a
DSO is least suited to catch rare events.
– There is no information in the trace, the trace is always of
equal intensity. Thus valuable information (so called Z-axis)
is lost. Also the fast slopes of a pulse which are invisible on
an analog scope will be of the same intensity as the slower
parts of the signal, this is a gross misrepresentation. The
reason is that DSOs ordinarily do not show only the sampled
points but they interpolate by drawing a continuous trace.
– The vertical resolution is mostly only 8 bits. In an analog
scope there is no loss of fi ne detail by digitizing. Even if the
trace is not very crisp details can be seen in it.
– Due to the sampling and the lack of a low pass fi lter in the
input frequencies above half the sampling frequency will
cause so called aliases, i.e. low frequency ghost signals.
Sampling is practically the same as frequency conversion
resp. multiplication, it creates sum and difference frequencies, beat frequencies which may be orders of magnitude
lower than the signal frequency and give grossly erroneous
results. In practice, therefore, only frequencies 1/10 or less
of the sampling frequency can be reliably displayed. The
meaning of the Nyquist theorem is mostly misunderstood:
if the sampling frequency is only twice the signal frequency
there will only two points be displayed on the screen: any
number of signal shapes may be drawn which fi t through
these two points. The Nyquist theorem contains a hidden
knowledge that the signal is a sine wave. It is easily understood that, in order to depict an unknown signal shape one
needs at least 1 or 2 points per centimeter; in other words:
the useful signal frequency is only 1/10 to 1/20 at best.
– An analog scope has a frequency response which follows
closely the Gaussian curve, this means in practice that also
frequencies far beyond the –3 dB frequency will be shown,
reduced in amplitude, but they will be shown. This not only
preserves fi ne detail of a signal but it allows also to see, e.g.,
very high frequency wild oscillations in a circuit. This is not
the case after sampling because all frequencies beyond half
the sampling frequency will be ”folded“ back into the lower
frequency band.
– Due to limited memory depth the maximum sampling rate
must be reduced with each DSO when the time base is
set to slow sweep speeds, it may be reduced from GS/s to
KS/s! Most users are not aware of this trap, they think
that if they bought a DSO with 100 MHz bandwidth and
1 GS/s they are safe when measuring kHz-signals. But
such low frequency signals will be distorted and aliases
displayed.
Please note: This list of disadvantages is by far incomplete! It
scratches but the surface.
There are 3 methods of sampling:
1. Real time sampling:
Here the Nyquist theorem must be observed, but, as mentioned, in practice the signal frequency is far less than 1/10 the
sampling frequency. Consequently, with a 1 GS/s rate signals
with up to 100 MHz can be adequately reconstructed. Obviously,
this is the only mode for single event capturing.
2. Equivalent time sampling:
This is the normal operating mode for all sampling scopes.
(Sampling scopes are very old, they are still the fastest scopes
with bandwidths > 50 GHz because they have no input amplifi er. Sampling scopes are far superior to DSOs because their Y
resolution is identical to that of an analog scope). In this mode
consecutive periods of the signal are sampled, each period
contributes but one sample. The signal period is thus scanned
and very many periods are necessary in order to achieve one
full screen display. This way a very high „effective“ sampling
rate is achieved, this method exchanges bandwidth for time. In
a sampling scope a very accurate display is created which is,
as far as the shape is concerned, almost as good as that of an
analog scope. In a DSO, however, the sample points are 8 bit a/d
converted, losing resolution. The bandwidth achieved is given
alone by the hf properties of the input and the minimum realizable duration of the sampling pulse, so 14 GHz at a sensitivity of
2 mV/cm and 50 Ω were standard in the 60s. In a DSO, however,
which should be used like an analog scope, a high impedance
(1 MΩ) wide range (e.g. 1 mV/cm to 20 V/cm) attenuator must
be included and also an input amplifi er. This is why a DSO can
not reach the bandwidths of sampling scopes. Equivalent time
sampling suffers fully from the problems of aliasing. As it requires the (not necessarily periodic) repetition of the signal in
invariant shape for e.g. millions of periods it is unsuitable for
the capture of single events. Equivalent time sampling can not
display the rising portion of a signal without a delay line.
3. Random sampling:
Random sampling is also very old (1952) and also no invention
of DSOs. It is similar to equivalent time sampling in that it
Subject to change without notice
21
CombiScope
requires a multitude of signal repetitions with invariant shape
in order to reconstruct it once on the screen. Therefore also
in this mode a very high ”effective“ sampling rate is achieved.
However, the samples are not taken step for step along the signal but randomly distributed over the signal period. An analog
computer is used to estimate the arrival of the next trigger, and
the time base is already started when it arrives. This has two
enormous advantages:
st
The rising portion of the signal can be shown without the
1
need for a delay line which would severely limit the achievable bandwidth.
nd
2
Due to the randomness of the samples alias signals will be
broken up.
The foregoing explains why it is HAMEG policy to offer CombiScopes rather than pure DSOs which combine the best of both
worlds although the cost of such an instrument is markedly
higher than that of a pure DSO, take alone the complicated high
frequency crt. It is the low cost of manufacturing which causes
the drive towards DSOs.
DSO Operation
The 100 MHz scope has 2 8bit a/d converters of the fl ash type
which is the best there is. The maximum sample rate of each
is 500 MS/s which is the rate available in dual channel mode
for the capture of single events. The maximum sampling rate
in all other operating modes is 1 GS/s.
Higher effective (!) sampling rates are possible as explained
above in equivalent and random sampling modes. As very many
signals repetitions are needed to reconstruct the signal once
any changes in signal shape such as noise will show up.
The reconstructed signal may be displayed either by showing
only the sampled points or with interpolation between them by
drawing straight lines.
5. XY display of signals which were previously captured in Yt
mode and protected against overwriting by STOP:
XY: readout ”XY“
Signal capture is triggered in SINGLE, REFRESH, ENVELOPE,
and AVERAGE modes and untriggered in ROLL and XY modes.
The normal (Refresh) mode is similar to the operating mode of
an analog scope. Triggering will cause signal acquisition and
display from left to right. After the next acquisition the display
will be replaced by the new information. If automatic triggering
was selected there will be a reference trace in the absence of a
signal the position of which is dependent on the vertical position
control setting. Signals with a repetition rate lower than the rep
rate of the automatic triggering can not properly trigger so the
resulting display will be untriggered.
In contrast to an analog scope the last display will remain on the
screen if the signal disappears in normal trigger mode.
In SINGLE mode the signal will be acquired only once. Acquisition can start if STOP (RUN key) is not illuminated (if necessary
press RUN until STOP extinguishes). The next trigger received
will cause the single acquisition. After this STOP will light up
and the trigger mode will be automatically switched to normal
dc coupled if auto was selected.
The trigger symbol on the screen allows to directly see resp.
determine the voltage level desired for triggering in the normal
mode, the voltage follows from the position and the VOLTS/CM
selected. The ground reference will be indicated by a ground
symbol in the screen centre.
After selecting SINGLE the trigger level symbol may be positioned using the LEVEL control. If e.g. the symbol is 2 cm above
the ground reference symbol the trigger level will be 2 cm x
Volts/cm (x probe factor if any).
Example: 2 cm x 1 V/cm x 10 (probe) = + 20 V.
The signals stored in DSO mode can be read via an interface and
documented. See the chapter ”Data Transfer“ for details.
DSO operating modes
In DSO mode the following operating modes are available:
1. Menu: ACQUIRE: Repetitive triggered signal acquisition and
display in usual Yt representation.
REFRESH: readout shows ”rfr“ (real time sampling) or
Random sampling: readout ”RS:xGSa“.
The operating mode may be further subdivided:
Envelope: readout ”env“
Average: readout ”avg:x“
(x may be a number 2.. 512)
2. ROLL mode, untriggered continuous signal acquisition,
display will „roll“ over the screen from left to right in usual
Yt mode:
Roll: readout ”rol“
3. Single sweep, triggered (menu: Trigger MODE) signal capture
in usual Yt mode:
Single: readout ”sgl“
4. Untriggered continuous signal capture, display in XY mode
(Menu: trigger MODE):
XY: readout ”XY“
Memory resolution
Vertical resolution:
The 8 bit a/d converters have a resolution of 256 possible vertical positions. The screen display has a resolution of 25 points
per cm. This is advantageous for display, documentation and
post-processing.
There may be some difference between the display on screen
and documentation, e.g. on a printer, this results from various
tolerances in the analog circuitry involved. The trace positions
are defi ned by:
Median horizontal line: 10000000b 80h 128d
Top line: 11100100b E4h 228d
Bottom line: 00011100b 1Ch 28d
In contrast to an analog display with its theoretically infi nite
resolution this is limited to 25 points per cm in DSO mode. If
there is any noise superimposed on the signal this may cause
frequent change of the lowest bit and thus jumping of the trace
in vertical direction.
Horizontal resolution:
A maximum of 4 simultaneous signal displays may be shown
on the screen. Each signal display will consist of 2048 points
(bytes). 2000 points will be distributed over 10 cm. The resolution
is thus 200 points per cm. Please note that this a 4 to 8 times
improvement over customary VGA (50 points per div) or LCD (25
points per div.) DSO displays.
22
Subject to change without notice
Data transfer
Memory depth
1 GS/s means that one million samples will be taken of the signal
and stored. With normal triggering and time base settings of
>20 ms/cm there will be 500,000 samples.
The screen display is calculated from the whole memory contents. Within the menu Settings->Display several display modes
may be selected:
Dots: the sampling points only are displayed.
Vectors: interpolation (sin x/x) or dot join is used to gene
rate a continuous trace.
Optimal: In this mode all samples are used to calculate
the display. This way the display of aliases is
less likely.
The scope acquires with as high a sampling rate as is possible
thus preventing to a large extent the production of alias signals.
It is always possible to zoom through the memory in order to
look at details, and, thanks to the deep memory, signal details
may be shown which remain invisible with shorter memory
DSOs.
Example:
This scope will sample with 1 GS/s in single channel mode down
to a time base setting of 100 us/cm. This equals 100,000 points
per cm. In MEMORY ZOOM signals of 150 MHz can still be seen.
Down to 100 us/cm hence aliases are not to be expected due
to the bandwidth limit of 150 MHz and the critical frequency
being > 500 MHz.
DSOs with a shorter memory like e.g. 10 K will only present
1000 points per cm which is equivalent to a sampling rate of
10 MHz, thus signals > 5 MHz will cause aliases , far below the
scope bandwidth. A deep memory is one of the most important
criterion of a DSO.
apparently untriggered changing displays, or may look like AM
modulated signals. If an alias is suspected change the signal
frequency or the time base or both. If aliases remain undetected
grossly erroneous results will be obtained which includes also
grossly (maybe orders of magnitude) false displays of signal
parameters like rise time etc.! Always watch for a stepped
display or print-out: this indicates an insuffi cient sampling rate
and consequently a false display. With an insuffi cient sampling
rate e.g. fast, short pulses may be completely ignored.
The best method to detect any false DSO display is to switch
to analog mode. In analog mode false displays are absolutely
impossible! An analog scope can at worst round the edges of
very fast signals.
Vertical amplifi er operating modes
In principle, in DSO mode there are the same modes available
as in analog mode, i.e.:
– CH1 only
– CH2 only
– CH1 and CH2 in dual trace mode Yt or XY
– Sum
– Difference
The main differences of DSO mode are:
– In dual channel mode both channels resp. signals are a/d
converted simultaneously.
No alternate or chopped channel switching.
– No fl ickering display even with low frequency signals as
the signals are stored and continuously displayed from the
memory with a suffi ciently high rep rate.
– Trace intensity is always the same. This is an advantage and
a disadvantage.
Horizontal resolution with X magnifi er
In principle, with a 10 x magnifi ed sweep, the resolution should
be reduced to 20 points per cm. However, the resolution remains at 200 points per cm as the information necessary will
be calculated from the memory. The magnifi ed portion may be
selected with the X-POS control. The fastest time base will be
5 ns/cm allowing a 2 cm per period display of 100 MHz.
Maximum signal frequency in DSO mode
The highest signal repetition frequency which still can be displayed well can not be exactly given. This is dependent as well
on the signal shape as on its amplitude displayed.
While it is fairly easy to recognize a square wave it requires at
least 10 samples per period to distinguish a sine wave from a
triangle. In other words: in practice, signals may still be recognized if their repetition frequency is <1/10 of the sampling
frequency. For a well defi ned display, however, many more than
10 points per cm are necessary.
Display of aliases
As explained the maximum sampling rate must be reduced for
slow time base settings. This may cause aliases. If e.g. a sine
wave is sampled only with one sample per period and if it should
be synchronous with the sampling frequency a horizontal line
will be shown as each time the same signal point is sampled. An
alias may also take the form of a signal of much lower frequency
(beat frequency between signal and sampling frequencies),
All so called Z axis (trace intensity) information is lost. In analog
mode the intensity depends on the signal rep rate resp. the
speed, thus mixed or unstable signals can be differentiated by
their respective trace intensity. Fast slopes of low frequency
signals are invisible in analog mode, in DSO mode they will be
shown as bright as the other signal portions.
Data transfer
Please note: Interface modules may only be exchanged after
the instrument was turned off. During operation the opening
of the interface must be covered.
There is an opening on the rear panel into which various interface modules can be inserted. Instruments are delivered with
RS-232 installed. (HO710)
The interface allows to either remotely control the scope or read
its settings. In DSO mode also the digitized and stored signals
may be retrieved. Interface cables must be shielded and must
not reach a length of 3 m or more.
Subject to change without notice
23
Data transfer
HO710: RS-232 Interface. Remote control
Safety hint:
All interface connections are galvanically connected
!
to the scope.
Measurements at high potentials are prohibited and endanger
the scope, the interface and all gear connected to the interface.
If the safety rules are disregarded any damage to HAMEG
products will void the warranty. Neither will HAMEG take any
responsibility for damages to people or gear of other make.
Description:
The RS-232 interface on the rear panel has the usual 9 pole
SubD connector. Via this bidirectional interface the scope can
be controlled remotely or its settings may be transferred. In
DSO mode also the digitized and stored signals can be read out.
The connection to a pc requires a 9 pole screened cable (1:1) of
3 m maximum length. The pinout is as follows:
Pin
no. Function
2 Tx date from scope to external device
3 Rx data from external device to scope
7 CTS ready to transmit
8 RTS ready to receive
5 ground (scope is connected to safety ground, safety class I)
9 + 5 V, max. 400 mA
The maximum signal on Tx, Rx, RTS and CTS is ±12 V. The
RS-232 interface parameters are:
N-8-2 no parity, 8 bits data, 2 stop bits (RTS/CTS hardware
protocol)
Selection of Baud rate
Baud rate setting is automatic. Range: 110 to 115200 , no parity,
8 bits data, 2 stop bits.
The fi rst SPACE CR (20 hex, OD hex) character sent after POWER
UP will set the Baud rate. This will remain set until POWER
DOWN or until a RM=0 command was sent. If activated previously also the front panel LOCAL (auto range) key may be used.
After release of remote (REM key (4) dark) data transfer can
only be reinitiated if a SPACE CR character was sent..
If the scope can not recognize a SPACE CR as the fi rst character TxD will be pulled low after 0.2 ms which causes a frame
error.
If the scope did recognize a SPACE CR und adjusted its Baud
rate accordingly it will answer with RETURN CODE „O CR LF“.
The front panel controls will be deactivated. The time between
Remote Off and Remote ON must be greater than:
tmin = 2 x (1/Baud rate) + 60 us.
Data transmission
After a successful Baud rate recognition and setting the scope
will be in the remote control mode and waits for commands. In
this mode no manual operation is possible. In order to return
to manual control a command must be sent from the PC or the
REMOTE OFF key depressed.
Loading of new fi rmware
Under www.hameg.de the most recent fi rmware is available
for downloading.
24
Subject to change without notice
CH I: 500 mV
POWER
General information concerning MENU
POWER
Pushbutton
CH I MENU
AC/DC/50 Ω
GND
50 Ω / 1 MΩ
INVERT
ON / OFF
VARIABLE
ON / OFF
PROBE
1 : 1 / 10 / 100
Menu Title
6 Function Pushbuttons (blue)
Menu
Intensity Knop Symbol
Arrow Keys
Indicator for Submenu
COMBISCOPE
MEMORY
oom
COMPONENT
TESTER
General information concerning MENU
Menu and HELP displays
Whenever a pushbutton is depressed the corresponding menu
will be displayed except for: EXIT MENU/REMOTE OFF
ANALOG / DIGITAL
. The menus offer options which may then be selected with
the blue pushbuttons. All pushbuttons are on/off.
Exiting a menu:
st
1
The user may set a time after which the menu will be left.
(SETTINGS
nd
2
Before the time set as described has elapsed a menu can
be exited by pressing EXIT MENU.
rd
3
Only manually if the function ”Man“ was selected .
th
4
Pressing the same pushbutton again.
th
5
Pressing another pushbutton.
, AUTOSET , RUN/STOP and MAG x10
pushbutton > , Misc > Menu OFF).
PROBE
ADJ
On Off
State indication by intensifi ed
display
Please note:
During the display of help texts and menus in full
size no signal display is possible.
Remarks
In operation all relevant measuring parameters will be shown
in the readout, provided the readout was activated and its in-
,
tensity is suffi cient.
The front panel LEDs add to the operating comfort and give
more information. In the end positions of the control knob an
acoustical signal will sound.
Apart from the POWER
scanned and stored. This allows to control the instrument from
stored information. Some controls and menus are only operative
in DSO mode or change their meaning in this mode. Explanations
are given with the warning: ”Only in DSO mode.“
pushbutton all control elements are
If a menu shows a knob symbol
knob
. It allows to change settings. Also arrows may be
this pertains to the INTENS
shown which point to available submenus.
In some modes some pushbuttons or knob operations are
meaningless and will hence not cause a menu display.
Please note:
If a menu is shown some other information dis-
played in the readout may disappear, this will
reappear immediately upon leaving the menu.
Each menu is assisted by HELP texts, which can be called by
pressing the HELP
readout. If HELP was called and the INTENS knob
explanation of the actual INTENS knob
and which will be also displayed by the
moved an
function will be given.
HELP will be left by pressing the pushbutton again.
Subject to change without notice
25
Controls and Readout
POWER
15
13
14
17
16
18
123
INTENS
POWER
!
EXIT MENU
REMOTE OFF
POSITION 1POSITION 2
VOLTS / DIV
SCALE · VAR
20 V1 mV20 V1 mV
CH 1
VAR
X-INP
!
CAT I
FOCUS
TRACE
MENU
REM
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
INPUTS
1MΩII15pF
max
400 Vp
4
ANALOG
DIGITAL
OSCILLOSCOPE
HM1008
·
1 GSa
100 MHz
VOLTS / DIV
SCALE · VAR
CH 2HORMAG
VAR
56 789 1011 12
1 MB
ANALOG
DIGITAL
RUNACQUIRE SETTINGSHELP
STOP
MODE
FILTER
SOURCE
AUX
!
CAT I
MATH
LEVEL A/B
TRIGGER
TRIG ’d
NORM
HOLD OFF
SAVE/
RECALL
X-POS
DELAY
AUXILIARY INPUT
TRIGGER
EXTERN
Z-INPUT
AUTOSET
HORIZONTAL
TIME / DIV
SCALE · VAR
50s5ns
VAR
x10
1MΩ II
15pF
max
100 Vp
19
26
27
20
23
21
24
28
22
25
29
30
3431
32
33
35
Controls and Readout
POWER
Mains switch with symbols I = ON and = OFF.
After turning the scope on and after the warm-up time of the
crt heater has elapsed the HAMEG logo, the instrument type
and the version number are displayed. If prior to switching off
the function ”Quick Start“ was selected (SETTINGS
> Misc)
the above will not be displayed. Then the scope will assume the
settings which existed before switching off.
INTENS knob
This knob controls various functions:
2.1 Trace intensity if the FOCUS/TRACE/MENU
does not light the knob symbol
. Turn left for decreasing and
right for increasing.
pushbutton
36
37
FOCUS TRACE MENU
If the knob symbol will light after pressing this pushbutton
the Int.Knob menu will be shown.
Depending on the operating mode the menu contains:
A-Int.: Intensity of the signal as displayed by time base A
B-Int.: Intensity of the signal as displayed by time base B
(analog mode only)
Zoom Int. Intensity of the signal as displayed by ZOOM
(digital mode only)
RO-Int.: Readout intensity
Focus: Focus for signal and readout
Readout
On Off: Turning the readout off will eliminate interference
of the readout with the signal(s). The knob symbol
will blink as long as the readout is off. Only menus
and help texts will be shown.
Trace rotation: Trace rotation (see TR)
After turn-on the readout will always be on.
EXIT MENU/REMOTE OFF (REM)
This pushbutton has two functions:
2.2 If the knob symbol XXX of the pushbutton
is illuminated
the control will act for those functions displayed in the menu,
which were activated.
26
Subject to change without notice
4.1 If a menu is displayed pressing this pushbutton will turn it off.
On condition a submenu is displayed, pressing this pushbutton
switches back to the menu.
Controls and Readout
4.2 Under remote control this pushbutton will be illuminated.
After pressing the pushbutton, control is returned to the front
panel.
ANALOG/DIGITAL
Switches between analog (green) and DSO modes (blue). The
operating mode (Yt or XY) will not be affected. If component
test mode was selected, possible only in analog mode, and the
scope is switched to digital the operating mode last used when
in DSO mode will be reinstalled.
Y parameters will not be changed by the switching. Time base
modes will be changed due to the different operation of the
time bases in both modes. After any switching time base A will
be selected. The time base speeds selected will not be affected
unless they are not available any more, then the maximum
value will be chosen.
RUN/STOP
This pushbutton has several functions:
6.1Analog mode, single sweep
The RUN/STOP pushbutton is used to set the time base for
a single sweep. Press MODE
menu. By pressing RUN/STOP the scope will wait for the next
trigger indicated by STOP blinking. After a signal triggered the
time base and the sweep is over the scope will stop, indicated
by STOP continuously illuminated. Each time the pushbutton
is depressed the scope will be activated anew for one trigger
resp. one sweep.
6.2 DSO mode, single capture
In order to prepare the scope for a single capture press MODE
and select ”Single“ in the menu presented. By pressing
RUN/STOP the scope will start to continuously acquire and store
so that later also the signal preceding the occurrence of the
trigger (Pre-trigger) will be available. STOP will blink in order to
signal that as yet no trigger was recognized. Upon receipt of a
trigger the acquisition will be completed so the signal following
the trigger (Post-trigger) will later be available, too. STOP will
then remain illuminated. The signal display on the screen will
be frozen. Pressing RUN/STOP again will prepare for another
acquisition as described. Each new acquisition will overwrite
the preceding one.
and activate ”Single“ in this
All entries and settings will be automatically stored upon leaving
the Mathematics menu or turning the scope off. Measurement
results will be lost after turn-off.
”Mathematics“ offers:
7.1 Equations set
Using the INTENS knob
5 sets of formulas can be selected for
editing. This way 5 user-defi ned formula sets may be created.
Each set of formulas consists of 5 lines with one equation
each, designated MA1 to MA5. An equation may occupy one or
more lines. In this case it has to be kept in mind that the lines
of equations are processed as a stack, i.e. starting with MA1 =
st
1
line to MA5 = 5th line.
Please note:
That formula set is valid which is shown prior to
leaving the MATH menu.
7.2 Edit
”Edit“ opens the ”Mathematics Edit“ submenu.
7.2.1 Equation
5 equations may be selected with the INTENS knob
. Each
equation consists of the name of a result (e.g. MA5), the
= sign, the function (e.g. ADD) and (fi rst operand, second operand). Remark: The second operand will not be displayed with
all functions.
7.2.2 Function
By using the INTENS knob
, the following functions may be
selected:
ADD: Operand 1 + operand 2
SUB: Operand 1 – operand 2
MUL: Operand 1 times operand 2.
DIV: Operand 1 divided by operand 2.
SQ: Operand 1 squared.
INV: Changes sign of operand 1.
1/: Calculates 1/operand 1 (reciprocal value).
ABS: Takes absolute value of operand 1 (removes sign)
POS: Only values of operand 1 > 0 are displayed, < 0 will
not be displayed.
NEG: Only values of operand 1 < 0 are displayed, > 0 will not be displayed.
In order to leave this mode enter the MODE menu and select
”Auto“ or ”Normal“.
6.2.1 DSO mode, ending or interrupting a capture.
Any capture in progress may be stopped by pressing RUN/
STOP.
6.2.2 DSO mode, reset function.
Pressing RUN/STOP twice in ”env“ (Envelope) or ”avg“ (Average)
mode, where the signal display is the result of more than one
signal data acquisition, causes the previous signal capture
results to be deleted and a new start of the signal capture.
MATH
Only available in DSO mode.
This pushbutton calls the ”Mathematics“ menu and the ”Formula editor“. With the ”Mathematics“ menu stored signals may be
mathematically processed, either directly or with reference to
other stored signals. The results may be graphically displayed
on the screen and determined by the cursor functions.
7.2.3 Operand 1
The INTENS knob
allows the selection of these signals as
operands:
CH1: CH1 signal.
CH2: CH2 signal.
RE1: Signal from reference memory 1
RE2: Signal from reference memory 2
MA1: Result of equation named MA1.
MA2: Result of equation named MA2
MA3: Result of equation named MA3.
MA4: Result of equation named MA4
MA5: Result of equation named MA5.
The next step after MA5 causes the display EDIT. Calling EDIT
opens a sub submenu described under item 7.2.5.
7.2.4 Operand 2:
The INTENS knob
allows the selection of the same operands
as mentioned under item 7.2.3 (Operand 1) if ADD, SUB, MUL
or DIV function is chosen.
The next step after ”MA5“ causes the display EDIT. Calling EDIT
opens a sub submenu described under item 7.2.5.
Subject to change without notice
27
Controls and Readout
POWER
FOCUS
TRACE
MENU
REM
4
ANALOG
DIGITAL
OSCILLOSCOPE
HM1008
1 GSa
100 MHz
123
INTENS
POWER
!
EXIT MENU
REMOTE OFF
7.2.5 Edit > Constant Edit
In the CW position of the INTENS knob
an additional item
”Edit“ and an arrow symbol is displayed. Pressing the associated
function button opens the sub submenu ”Constant Edit“. This
enables to choose a Number, its ”Dec.Point“ (decimal point) and
its ”Prefi x“ of the ”Unit“, all with the INTENS knob
. The unit
need not be selected it is only shown to assist the memory.
7.3 Display
Attention!
The display of mathematic signal(s) automatically
switches previously displayed reference signals off
and vice versa.
The function ”Display“ is available twice in the menu and may
be switched on or off in any combination. This allows to display:
no result, one result, the result of two equations as signals. The
displays will come forward upon leaving the ”Mathematics“
menu. Also the designation of the equation (e.g. MA2) will be
shown. Select the equations to be displayed with the INTENS
knob
.
The mathematics signal is automatically scaled, this is independent of the graticule, of Y and time base parameters, the
scale will not be shown. Hence the measurement of the signal
amplitudes must be performed using the CURSOR (V to GND)
after the ”reference“ (e.g. MA2) of the CURSOR to the ”mathematics signal“ and its scale was established (AUTO/CURSOR
MEASURE
pushbutton > Cursors > reference > e.g. MA2).
The readout may then display e.g.: ”V(MA2): 900 mV“.
Divisions by zero will be ignored and an error message displayed.
7.4 Units
Each function ”Display“ will be associated with a function ”Unit“
which can be selected with INTENS
and will be attached to
the result.
ACQUIRE
Available only in DSO mode.
This pushbutton opens the menu ACQUIRE which offers these
modes:
8.1 Refresh capture/display.
In this mode repetitive signals may be captured and displayed
much like in analog mode. The readout will show ”rfr“. Signal
acquisition may be stopped or started with the RUN/STOP pushbutton. The STOP pushbutton will be illuminated in stop.
A trigger will start a new acquisition which will overwrite the
display of the former. The display will remain on screen until
the next acquisition. This mode is available over the full time
base range (50 s/cm to 5 ns/cm).
567891011 12
· 1 MB
ANALOG
DIGITAL
RUNACQUIRE SETTINGSHELP
STOP
MATH
SAVE/
RECALL
AUTOSET
Please note:
After changing the time base acquisition and display
will always start fi rst at the trigger position which is
undelayed (Readout: ”Tt:0s“) on the screen centre.
With the 2
nd
acquisition the display will start at the
screen left. In most cases this is meaningless, but
the scope may seem not to react at slow time base
settings combined with long Posttrigger times.
In the utmost left position of the trigger point selec-
ted with the HORIZONTAL control
the readout
will indicate ”Tt:1.85ks“ for a time base setting of
50 s/cm. This means that 1,600 seconds must elapse until the trace will become visible at the screen
left, after another 250 s it will have reached the
screen centre (1,600 s + 250 s = 1.85 ks).
8.2 Envelope capture/display
Envelope is a special mode in refresh mode, the readout will
show ”env“. Also in this mode there must be suffi cient signal
for triggering.
In contrast to the refresh mode the results of several captures
will be examined and the maxima and minima stored, the envelope of the signal will then be displayed if it changes in amplitude
or/and frequency. Also any jitter will be shown.
Also in this mode pressing RUN/STOP
will stop the acquisition, indicated by STOP illuminated. Af ter pressing the pushbutton again the formerly stored signals will be erased and the envelope calculation star ts anew. In order to prevent an accidental
turning on of this mode operating any control which infl uences
the signal display will automatically switch envelope off.
Because this mode requires many signal repetitions and acquisitions it is not compatible with single sweep/acquisition. AUTO
or normal trigger modes must be selected.
8.3 Average mode capture/display
Also this is a special mode within the refresh mode. Also here
signal repetitions are needed.
The weighting of each acquisition can be selected with ”Average“ in the menu, any number between 2 and 512 may be
chosen using the INTENS knob
. The readout will show e.g.
”avg#512“.
The higher the number of acquisitions averaged the lower the
contribution of a single acquisition will be and the longer the
averaging will take. Averaging is a means to increase the accuracy inspite of the 8 bit converters, it is an exchange of time
against accuracy. Noise will be reduced by averaging.
The same holds as for envelope: the acquisition may be stopped
by pressing the RUN/STOP pushbutton, STOP will be illuminated. Pressing RUN/STOP again will restart. In order to prevent
28
Subject to change without notice
Controls and Readout
an inadvertent entering of this mode the operation of any control
will automatically cause this mode to be reset.
As repetitive acquisitions are needed for calculation of an average single sweep will not be compatible.
8.4 Roll mode capture/display
Roll mode means that the signal(s) will be continuously acquired
without the need for a trigger. Hence all controls, displays and
readouts for the trigger and ZOOM will be disabled. The readout
will show ”rol“.
The result of the last acquisition will be displayed at the right
hand edge of the graticule, all formerly acquired signals will
be shifted one address to the left. The result at the left hand
screen edge will be dropped. There is no waiting for a trigger
and thus the hold-off time is minimum. As in any other mode
the signal acquisition may be stopped and restarted any time
with the RUN/STOP pushbutton.
In roll mode the time base available is limited to 50 s/cm to
50 ms/cm. Faster time bases do not make sense as the signal
could not be observed any more.
In case the time base was set outside the limit cited it will be
automatically changed to the next value within the limits upon
entering this mode.
8.5 Peak Detect Auto Off
On Peak Detect Auto mode this mode of acquisition will be automatically selected if YT and time base speeds of 20 ms/cm to
2 ms/cm were chosen. This mode is only available with: Refresh,
Envelope, Average, and Single sweep. The readout will show PD
preceding the symbol for the operating mode selected.
At slow sweep speeds the sampling rate will be low, i.e. there
are relatively large gaps between samples. Glitches occurring
during those gaps will go by unnoticed. With Peak Detect, however, the highest sampling rate will be used such that glitches
will be caught. The samples will be looked at and the ones with
the highest values stored and displayed.
8.6 Random Auto Off
Provided single sweep was not selected Random Sampling will
be automatically selected beginning at a certain sweep speed.
The time base setting will be indicated in the readout, e.g. ”RS:
10 GSa“ (= Random sampling with 10 GS/s effective sampling
rate), the real time time base speed will be 5 ns/cm. Without
Random Sampling ”RS“ Real Time Sampling will be used with
a maximum sampling rate of 1 GS/s (one channel only) or
500 kSa/s (two channel mode).
9.1 Analog and DSO modes
Under ”Save/Recall“ the current instrument settings may be
saved or settings saved earlier recalled. There are 9 nonvolatile
memories available.
9.1.1 Saving the actual settings
Upon pressing the pushbutton the submenu ”Front Panel
Save“ will be accessed. A memory number will be offered
(1 to 9) which can be changed by the INTENS knob
. By pressing ”Save“ all settings will be saved in the memory location
the number of which was selected.
9.1.2 Recall the actual settings
In the submenu ”Front Panel Recall“ a memory number will
be displayed (1 to 9) which can be changed with the INTENS
knob
. Pressing the pushbutton ”Recall“ causes loading of
the settings stored.
9.2DSO mode
The menu options described in 9.1.1 and 9.1.2 are also available
in DSO mode. Additionally, the menu options ”Reference Save“
and ”Reference Display“ will be available. A reference is a signal
which was resp. can be stored away for later reference to it 9
nonvolatile memory locations are provided.
9.2.1 Reference Save
9.2.1.1 Source x
In the submenu the source can be selected with the INTENS
knob
. Signals from the logic inputs CH3 and CH4 can not be
stored as references.
9.2.1.2 Destination RE x
There are 9 memory locations available into which reference
signals from the source selected before can be stored. Use the
INTENS knob
for selection.
9.2.1.3 Save
Pressing ”Save“ will store the signal from the source selected
into the memory selected.
9.2.2 Reference Display
Attention!
The display of reference signals automatically swit-
ches previously displayed mathematic signal(s) off
and vice versa.
Random Sampling requires repetitive signals, each signal period will contribute one sample. At an effective sampling rate of
10 GS/s the time difference from sample to sample along the
signal period will be 0.1 ns. However, note that with random
sampling the samples are not taken in sequence along the signal period but randomly with respect to it. Random Sampling
allows to generate the 200 points per cm in X direction at the
fastest time base of 5 ns/cm.
Remark: 5 ns/cm are also available in other modes. In Real
Time Sampling mode and 1 GS/s on one channel each 1 ns a
sample is taken, hence at 5 ns/cm there are 5 points per cm.
The ”missing“ 195 points are generated by interpolation using
sin x/x.
SAVE/RECALL
This pushbutton will open up a menu. The number of choices
in this menu is dependent upon whether there is analog or
DSO mode.
9.2.2.1 RE x, On Off, associated settings
When in this submenu using the INTENS knob will allow
selection of 2 reference signals which can then be displayed
alongside with 2 input signals.
9.2.2.2 RE x
After calling this function the memory location can be selected
with the INTENS knob
. (RE 1 to 9)
9.2.2.3 On Off
With the pushbutton on/off control is possible. When the
contents of the reference memory are displayed, the memory
number is indicated with RE x (x = 1 to 9) at the right hand
screen side. Switching to ”on“ will produce another menu item
(”Assoc.Set“).
Please note:
If both reference displays are ”on“ and if both
memory locations are identical (e.g. RE1, RE1) the
signal will be displayed twice on the same spot.
Subject to change without notice
29
Controls and Readout
POWER
FOCUS
TRACE
MENU
REM
4
ANALOG
DIGITAL
OSCILLOSCOPE
HM1008
1 GSa
100 MHz
123
INTENS
POWER
!
EXIT MENU
REMOTE OFF
9.2.2.4 Associated settings
If the pushbutton labelled ”Assoc.Set.“ is depressed the readout
will show all oscilloscope settings which were loaded into the
reference together with the signal. The signal parameters may
be recalled.
SETTINGS
Pressing this pushbutton will open the SETTINGS menu which
offers the following submenus:
10.1 Language
In thus submenu the language can be selected: English, German, French are available for choice.
10.2 Misc (Miscellaneous)
10.2.1 Contr.Beep On Off
Switches the acoustical signal on or off, which informs about
CW or CCW positions of knobs.
567891011 12
· 1 MB
ANALOG
DIGITAL
RUNACQUIRE SETTINGSHELP
STOP
MATH
SAVE/
RECALL
AUTOSET
10.4.3 Optimum display
In this mode minimum and maximum signal values acquired
are taken into consideration and displayed in Vectors mode.
This makes sense as up to 1 MByte samples may be acquired,
but only up to 2 KBytes per channel can be displayed, hence
possibly min or max values may not be shown. Otherwise the
display is as described above.
AUTOSET
Choosing AUTOSET will cause an automatic instrument setting,
dependent upon the signal proper which selects positions, signal
amplitude and time base for a reasonable display. The choice
of analog or DSO mode will not be affected. In component test
mode (available only in analog mode), XY mode, or ADD automatically dual channel mode will be selected. If dual channel
or CH1 or CH2 were previously chosen this will remain.
The DSO modes Roll, Envelope or Average will be changed to
refresh mode.
10.2.2 Error Beep On Off
Will turn the acoustical error signal on or off.
10.2.3 Quick Start On Off
In off the HAMEG logo, the type and the version number will not
be shown, the instrument will be ready immediately.
10.2.4 Menu Off time
With the INTENS knob
the time of menu display may be de-
termined. EXIT MENU will terminate a menu on the spot.
In ”Man.“ mode the menu can be left:
– by pressing EXIT MENU.
– by pressing another pushbutton.
– by pressing the same pushbutton again with which the menu
was called.
10.3 Interface
This menu shows the interface parameters which can be
selected as usual.
10.4 Display
This submenu offers several modes of display:
10.4.1 Dots
In this mode the samples are shown as what they are, i.e. dots
(points). This representation is valuable for judging whether
enough samples were gathered in order to reconstruct a signal
suffi ciently.
10.4.2 Vectors
In this mode the sampling points are interconnected by drawing
straight lines. If there are only few samples sin x/x interpolation
is used to ”create“ intermediate points which are then joined
by straight lines.
AUTOSET will further set the intensity to an average value if it
was set too low. If a menu was opened it will be turned off by
AUTOSET. During the display of HELP texts AUTOSET is not
available.
HELP
Pressing the HELP pushbutton will turn the signal display off
and display the help text.
If a menu was opened the help text will refer to this menu resp.
to the special menu or submenu option selected. If a knob is
moved another help text referring to this knob will appear. Press
HELP again to deactivate the text.
POSITION 1 (knob)
This knob can assume various functions which depend upon the
operating mode, the functions selected with the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
selected.
13.1 Y position
13.1.1 Y position CH1 (Analog and DSO modes)
POSITION 1 will set the Y position of CH1 provided there is YT
mode and the pushbutton CH1/2–CURSOR–MA/REF–ZOOMpushbutton
is not illuminated.
13.1.2 REF (reference) signal position (DSO mode only)
The POSITION 1 control functions as the position control for
the signals stored in the reference memory if these conditions
are fulfi lled:
st
1
A reference signal must be on display (SAVE/RECALL
ber of memory location, select with INTENS) > On (with or
without associated settings).
nd
After pressing CH1/2–CURSOR–MA/REF–ZOOM-push-
2
button > Math./Ref. was selected, the pushbutton will
light up green.
13.1.3 Mathematics signal position (DSO mode only)
The POSITION 1 control will assume the function of position control for mathematics signals after the following procedure:
Press the MATH pushbutton > Display (upper display area),
select an equation with the INTENS knob
the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
(MA1.. MA5), press
, select
”Math./Ref.“ The pushbutton will light up green.
13.1.4 Y position of 2
nd
time base B (Analog mode)
The POSITION 1 control will assume the function of Y position
control of the signal displayed by time base B in alternate time
base mode after the following procedure. This is convenient in
order to be able to separate the displays of the (same) signal
with both time bases on the screen. Press the HOR
pushbutton > ”Search“. Press the CH1/2–CURSOR–MA/REF–ZOOMpushbutton
, select the function ”TB B“. The pushbutton will
light up green.
13.1.5 Y position for ZOOM (DSO mode.)
This is the equivalent of the former in DSO mode, where the
function is called ZOOM. The POSITION 1 control will assume
the function of Y position control of the zoomed signal after the
following procedure. Again, the intent is to be able to separate
the two displays of the (same) signal on the screen in alternate
time base mode. Press the HOR
pushbutton > ”Search“.
Press the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
select ”TB B“. The pushbutton will light up green.
select ”Cursors“ or ”Cur. Track“. The push-button will light
up in blue.
Please note:
The function ”Cur. Track“ is only available if two
cursors are indeed displayed, then both cursors
can be moved simultaneously (tracking) without a
change of their respective positions.
POSITION 2 (knob)
Also this control may assume diverse functions dependent on
the operating mode, the function selected via the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
vated.
14.1. Y position
14.1.1 Y position CH2 (Analog and DSO modes)
POSITION 2 will function as Y position control of CH2 in Yt mode
and if the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
not illuminated.
14.1.2 Y position of reference signals (DSO mode)
The POSITION 2 control will function as Y position control of
reference signals if the following conditions are fulfi lled:
– A reference signal must on display. (SAVE/RECALL push-
button
> Reference Display > (upper display area) Rex
(x = number of memory location, select with INTENS) > On
(with or without associated settings).
– Press CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
,
> Math./Ref. The pushbutton will light up green.
and the menu item acti-
is
13.2. X position in XY mode (Analog and DSO modes)
POSITION 1 will function as X position control of CH1 in XY mode
and provided the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
is not illuminated.
Note:
The HORIZONTAL control
will also be functional
in XY mode.
13.3. CURSOR position (Analog and DSO modes)
The POSITION 1 control will function as Y position control of
the cursors if the following conditions are met: The CURSOR
display must have been activated (AUTO/CURSOR-MEASURE
pushbutton
>Cursors >Cursors On pushbutton depressed).
Press the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
VOLTS / DIV
SCALE · VAR
15
13
14
17
16
18
POSITION 1POSITION 2
VOLTS / DIV
SCALE · VAR
20 V1 mV20 V1 mV
CH 1
VAR
14.1.3 Y position of mathematics signals (DSO mode)
The POSITION 2 control functions as Y position control for mathematics signals, if the following conditions are fulfi lled:
Press the MATH pushbutton
select an equation with the INTENS knob
> displays (upper display area),
(MA1.. MA5). Press
the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
Math./Ref. The pushbutton will light up green.
14.2 Y position of CH2 in XY mode (Analog and DSO modes.)
POSITION 2 will function as the Y position control of CH2 in XY
mode provided the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
is not illuminated.
14.3 Y position of cursors (Analog and DSO modes)
The POSITION 2 control will function as Y position control of
,
LEVEL A/B
TRIGGER
MODE
FILTER
SOURCE
CH 2HORMAG
VAR
AUX
X-POS
DELAY
TRIG ’d
NORM
HOLD OFF
HORIZONTAL
TIME / DIV
SCALE · VAR
50s5ns
VAR
x10
, select
19
26
27
20
23
21
24
28
22
25
29
30
Subject to change without notice
31
Controls and Readout
POWER
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
VOLTS / DIV
SCALE · VAR
CH 2HORMAG
VAR
15
13
14
17
16
18
POSITION 1POSITION 2
VOLTS / DIV
SCALE · VAR
20 V1 mV20 V1 mV
CH 1
VAR
the cursors if the following conditions are met: The cursors
must be activated (Press AUTO/CURSOR-MEASURE pushbutton
>Cursors >Cursors On), press the CH1/2–CURSOR–MA/
REF–ZOOM-pushbutton
select ”Cursors“ or ”Cur. Track“.
The pushbutton will light up blue.
Note:
The function Cur. Track (cursor tracking) is only
available if 2 cursors are on display. The cursors
will then be moved simultaneously (tracking) without changing their respective positions.
CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
19
26
27
20
23
21
24
28
22
25
29
30
LEVEL A/B
TRIGGER
MODE
FILTER
SOURCE
AUX
TRIG ’d
NORM
HOLD OFF
X-POS
DELAY
50s5ns
VAR
HORIZONTAL
TIME / DIV
SCALE · VAR
x10
”>“ to the sensitivity indication (”CH1>5mV..“) in order to show
that the sensitivity is now uncalibrated. The results of cursor
measurements will be fl agged accordingly.
In this mode the sensitivity can be changed with the VOLTS/
DIV–SCALE–VAR control knob from 1 mV/cm to > 20 V/cm.
16.3SCALE (DSO mode)
The display height of a mathematic signal resp. a reference signal can be changed after the function „Math./Ref.“
is selected in the „Pos./Scale“ menu. The latter is called
by pressing the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
.
This pushbutton calls a menu which allows to select the functions of the controls POSITION 1 and 2. With mathematics
functions this will also affect in some way the SCALE function
which is changed by the VOLTS/DIV controls.
The pushbutton will signal the function activated corresponding
to the front panel labelling:
dark: Y position CH1 and CH2.
blue: Y position of cursors.
green: Y position of:
– Mathematics signal(s)
– Reference signal(s)
– ZOOM or time base B display of signal(s)
VOLTS/DIV–SCALE–VAR knob
This knob is a double function CH1 control.
16.1 Selection of vertical sensitivity
Provided VAR on pushbutton CH1
sitivity will be calibrated. Turning the control CCW will decrease,
turning it CW will increase the sensitivity. 1 mV/cm to 20 V/cm
can be selected in a 1-2-5 sequence. The readout will display
the calibrated sensitivity (e.g. ”CH1: 5mV..“). Depending on the
sensitivity selection the signal will be shown with smaller or
greater amplitude.
Please note:
This sensitivity selection is always active, e.g. also,
if CH2 only was chosen. In that case CH1 may be
used as trigger source.
16.2 Variable control
Select this function with the CH1 pushbutton
VAR on the CH1 pushbutton will light up. The readout will add a
is not illuminated the sen-
>Variable On.
VOLTS/DIV–SCALE–VAR knob
This double function control belongs to CH2.
17.1 Selection of vertical sensitivity
The sensitivity will be calibrated provided VAR on the CH2
pushbutton
will decrease, turning it CW will increase the sensitivity. The
sensitivities can be selected from 1 mV/cm to 20 V/cm in a
1-2-5 sequence. The readout will show the sensitivity (e.g.
”CH2:5mV..“). Depending on the sensitivity the signal will be
displayed with smaller or greater amplitude.
Please note:
The sensitivity control is always active, i.e. also if
17.2 Variable control
The variable control is activated by pressing the CH2 pushbutton
up. The readout will add the ”>“ preceding the sensitivity
(e.g. ”CH2 >5mV..“) in order to indicate that the sensitivity is
uncalibrated. The results of cursor measurements will be
labelled accordingly.
In this mode the sensitivity may be changed continuously
with the VOLTS/DIV–SCALE–VAR control from 1 mV/cm to
> 20 V/cm.
17.3 SCALE (DSO mode)
The display height of a mathematic signal resp. a reference signal can be changed after the function „Math./Ref.“
is selected in the „Pos./Scale“ menu. The latter is called
by pressing the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
is not illuminated. Turning the control CCW
CH2 is not selected. CH2 may then still be used e.g.
as a trigger source.
>Variable On. VAR on the pushbutton will light
.
32
Subject to change without notice
Controls and Readout
AUTO/CURSOR MEASURE pushbutton
Pressing this pushbutton will open the menu ”Measurement“
which offers the submenus > ”Cursors“ and ”Auto“.
If the submenu Cursors was selected and a measuring mode
”Cursors On“ must be activated too, in order to generate the
cursor line(s) on the screen. The measurement result will be
shown in the readout!
Please note:
In order to move the cursors it is necessary to press
the pushbutton CH1/2–CURSOR–MA/REF–ZOOM
and look for the menu ”Pos./Scale“. In this menu
the selection of ”Cursors“ (long lines) or ”auxiliary
cursors“ (short lines) or other symbols will determine which cursor lines/symbols can be moved by
the POSITION 1 and 2 controls.
18.1 Cursors (Analog and DSO modes)
Depending on the operating mode (Yt or XY) this submenu will
offer various cursor measuring functions which will affect as
well the cursor lines as their position.
18.1.1 Cursors On Off
With ”Cursors On“ the CURSORS and the results of cursor
measurements will be displayed by the readout in the top right
corner of the screen. (e.g. ΔV(CH2):16.6 mV). If a variable control was activated, the readout will indicate this by replacing
the ”:“ by a ”>“.
18.1.4 Respect
It may be necessary to determine for which signal resp. channel
the CURSOR measurement shall be valid. This is signalled by
showing the INTENS knob
signal next to the channel number.
After selection of the signal source, the CURSOR lines must
then be positioned to the signal or portions of it displayed by
this channel.
18.2 Auto (Analog mode)
Depending on the operating mode this submenu offers various
automatic measurements of the trigger signal. The following
conditions must be met:
a) For frequency or period measurements suffi cient trigger
signals must be available. Use normal triggering for signals
< 20 Hz. Please note that signals of very low frequency may
require seconds to complete one measurement.
b) In order to measure DC or the DC content of a signal the
input channel as well as the trigger must be set to DC coupling.
Further notes:
– Due to the limited frequency response of the trigger channel
the accuracy will decrease with increasing frequency.
– The frequency responses of the vertical channel and the
trigger channel differ substantially, this may affect the signal
display.
18.1.2 Meas. Type
If this function is activated one of the measurement modes offered may be selected with the INTENS knob
. In most cases
the unit going with a mode will be shown also.
18.1.3 Unit
In the modes ”Ratio X“ and ”Ratio Y“ the INTENS knob symbol
will be shown in addition to a unit, this may then be used to
change the unit.
” rat ” (ratio), display of ratios
In this mode the ratios of duty cycles or amplitudes may be
determined with the CURSORS. The distance between the long
CURSOR lines is equal to 1.
” % ” (percent), display of percentages
The distance between the long CURSOR lines is equal to 100
%. The result will be determined by the distance of the short
auxiliary cursor line to the long reference line (lower resp. left),
if appropriate with a negative sign.
” ° ” (degree), measurement of degrees
The distance between the long CURSOR lines is equal to 360
degrees and must be exactly as long as a signal period. The
measurement result will be determined from the distance
between the reference line to the short auxiliary cursor line.
If appropriate with a negative sign. For further information
please consult ”Measurements of phase differences in dual
channel mode (Yt)“ in the section ”First time operation and
presettings“.
”π”
One period of a sine wave is equal to 2 π, hence the distance
between the two long CURSOR lines must be set to one period. If the distance between the reference line and the short
CURSOR line equals 1.5 periods, ”3 π“ will be displayed. If the
short cursor line is left of the reference line a negative sign
will be shown.
– When measuring very low frequency signals the display will
follow the signal.
– When measuring pulse signals there may be an infl uence
of the duty cycle or the slope selected on the accuracy of
the result.
– The signal must remain within the graticule area, i.e. the
input amplifi ers must not be overdriven.
Please note:
Complex signals should be measured using the
CURSORs.
18.2.1 Auto On Off
If Auto is On the result of the automatic measurement will be
shown in the readout in the top right corner. (e.g. DC(Tr):100μV.
(Tr) points out that it refers to the trigger signal. Sometimes
a ”?“ will be displayed, this indicates that there is no or an
insuffi cient signal.
If a variable is activated und thus the sensitivity or time base
uncalibrated the ”:“ will be replaced by a ”>“ symbol.
18.2.2 Meas. Type (measurement type).
The measurement type can be selected with the INTENS knob
.
18.2.3 Respect
Tr indicates that the measurement is done with respect to the
trigger signal. If e.g. the CH1 signal is used for triggering, the
result will be with respect to that signal.
18.3 Auto (DSO mode)
Depending on the operating mode this submenu offers various
automatic measurements. The display in the ”reference“ fi eld
will signal to which source the result belongs. (Tr = trigger signal
or the signal of the channel shown.)
Subject to change without notice
33
Controls and Readout
POWER
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
VOLTS / DIV
SCALE · VAR
CH 2HORMAG
VAR
15
14
16
18
13
17
POSITION 1POSITION 2
VOLTS / DIV
SCALE · VAR
20 V1 mV20 V1 mV
CH 1
VAR
See the hints in 18.2 regarding period (trigger period) or frequency (trigger frequency) measurements.
For voltage measurements the result will be calculated from
the displayed signal data and the signal data selected with
”Respect“.
Please note:
Complex signals should be measured with CURSOR.
18.3.1 Auto On Off
If Auto is On the readout will display the result of the automatic
measurement in the top right corner. (e.g. DC(CH2):100μV).
(CHx) denotes the source channel, (Tr) the trigger signal.
Sometimes a ”?“ may be displayed, if there is no signal or an
insuffi cient one.
If a variable is activated and the sensitivity or time base thus
uncalibrated the ”:“ will be replaced by a ”>“ symbol.
18.3.2 Meas. Type (measurement type)
If this function is activated the mode can be selected with the
INTENS knob
, the display in the ”Respect“ fi eld will change
accordingly.
19
26
27
20
23
21
24
28
22
25
29
30
LEVEL A/B
TRIGGER
MODE
FILTER
SOURCE
AUX
TRIG ’d
NORM
HOLD OFF
X-POS
DELAY
VAR
HORIZONTAL
TIME / DIV
SCALE · VAR
50s5ns
x10
in those modes where there is no direct relationship between
trigger signal and trigger point.
In normal trigger mode the LEVEL control will move the trigger symbol anywhere. In automatic peak-to-peak detection
mode the level can only be selected between the signal’s peak
values.
The movement of the trigger symbol is vertical only. The range
of this symbol is limited in order to prevent that this symbol will
overwrite other readout information. As soon as the trigger symbol leaves the graticule its form will change, this change signals
in which direction the trigger symbol left the graticule.
Analog mode only: Depending on the time base mode the LEVEL
control will affect the time base A or B triggering. Press the
HOR pushbutton
in order to select the time base mode in
the ”Time base“ menu. In ”Search“ mode (alternate time base
mode) the last trigger level setting for time base A will remain
valid (graticule left) if time base B is switched to triggered
mode. (Menu ”Time base“: set B trigger to positive or negative
slope). Thereafter the LEVEL A/B control will control the time
base B trigger, a second trigger point symbol will be shown
and marked with ”B“.
In one channel only mode the active channel will be indicated.
In those modes where 2 channels are activated the INTENS
knob symbol
will be shown. After pressing the ”Reference“
pushbutton the reference channel may be selected with the
INTENS knob
.
18.3.3 Respect
Tr indicates that the measurement is done with respect to the
trigger signal. If e.g. the CH1 signal is used for triggering, the
result will be with respect to that signal.
In case of voltage measurement is done with respect to the
active channel. In those modes where several channels are
activated, the INTENS knob symbol
indicates that different
channels can be selected.
LEVEL A/B knob
The LEVEL control allows to set the trigger level, i.e. the voltage
resp. signal level, which will generate a trigger to start the time
base whenever the signal passes that level. In most Yt modes
the readout will show a symbol the vertical position of which
indicates the signal point which will trigger. The trigger symbol
will be ”parked“ on the second graticule line from the bottom
34
Subject to change without notice
MODE pushbutton
Pressing this pushbutton will open the ”Trigger“ menu, where
Auto, Normal, Single (sweep triggering) can be selected. Choosing ”Slope“ will allow to trigger on any signal shape. For video
signals select ”Video“ and press the FILTER pushbutton
order to fi nd a choice of special trigger modes for composite
video signals.
In XY mode the pushbuttons MODE
, FILTER and SOURCE
are disabled as there is no triggering in XY mode.
20.1 Auto (trigger)
Automatic triggering (Auto) is active if the NORM display
not illuminated. In ”Auto“ the analog time base resp. signal
capture (DSO mode) will be periodically started even if there is
no signal or when no triggers are generated because the settings are incorrect. Signals of < 20 Hz can not be triggered as
the automatic start will have occurred before the signal arrived.
Automatic triggering is possible with or without peak detection.
The LEVEL A/B
control will be active in both modes.
In peak detection mode the range of the level control is limited
to the peak-to-peak voltage of the signal. Without peak detection
in
is
Controls and Readout
any level can be set. If the trigger level is set such that no triggers are generated the automatic triggering will nevertheless
start the time base. The signal will thus remain visible but will
be untriggered.
Whether peak detection is active or not depends on the mode
and the settings in ”FILTER“ (trigger coupling). The mode active
will be shown by the behaviour of the trigger point symbol when
turning the LEVEL knob.
20.2. Normal (trigger)
If the NORM-LED
ted.
In normal trigger mode both the peak detection and the automatic time base start will be disabled. Hence if there is no
suffi cient trigger signal the screen will remain dark in analog
mode. In DSO mode signal capturing will also stop unless the
roll mode was selected.
In this mode there is no lower frequency limit for signals.
20.3 Single (sweep/capture)
In single sweep/capture mode the time base selected will accept
only one trigger for one sweep/capture after it was armed. The
NORM-LED will light up, Auto triggering is disabled.
For further information about the precise operation see RUN/
STOP pushbutton
lights up normal triggering was selec-
description.
FILTER pushbutton
After this pushbutton is depressed it will depend on the settings
chosen in MODE
In XY mode the pushbuttons: MODE
are disabled as XY displays can not be triggered.
21.1 Menu: Slope
The menu ”Edge“ will appear if ”Edge“ was selected in the
TRIGGER menu to be called with MODE
the FILTER
tion see ”Trigger coupling“ (Menu FILTER) under the heading
”Triggering and time bases“ and the instrument specifi cations.
The following settings are available:
21.1.1 Trig. Filter
– AC: The trigger signal is AC coupled via a large capacitor in
order to reach a low cut-off frequency.
Readout: ”Tr:Source, Slope, AC“
– DC: The trigger signal is DC coupled. No peak triggering is
possible.
Readout: ”Tr: Source, Slope, DC“
– HF: AC coupling with a small capacitor suppressing low
frequency signals. Hence the signal display and the trigger
signal derived are not any more identical, the trigger point
symbol will be ”parked“ in DSO mode and will not react to
the LEVEL A/B
symbol is switched off. As a combination of HF coupling and
LF or Noise Reject is not meaningful both menu options will
not be shown.
Readout: ”Tr:Source, Slope, HF“.
– LF: The trigger signal is sent through a low pass in order to
suppress high frequency components. As this will already
suppress hf the noise rejection mode will be set to OFF
automatically.
Readout: ”Tr:Source, Slope, AC or DC, LF“.
(Edge or Video) which menu will be offered.
, FILTER and SOURCE
pushbutton and after
pushbutton was depressed. For further informa-
control. In analog mode the trigger point
– Noise Reject: Noise rejection (reduction) means a reduced
trigger amplifi er bandwidth and consequently less trigger
signal noise.
Readout: ”Tr:Source, Slope, AC or DC, NR“.
21.1.2 Slope
”SLOPE“ determines whether the rising or falling portion of
a signal shall trigger, the level is set with the LEVEL A/B
control.
In BOTH MODE both slopes will trigger, this is also true in single
sweep mode. This allows e.g. the display of eye diagrams.
21.2 Menu: Video
In order to reach the menu VIDEO proceed as follows:
Press MODE
press the FILTER
found under VIDEO (tv signal triggering) in the chapter Triggering and time bases“ and in the instrument specifi ca-tions. The
following settings are available:
21.2.1 Frame, Line.
Depending on the setting chosen triggering will be on frame
or line sync pulses. The selection will also affect other menu
items.
Readout: ”Tr:Source, TV“.
21.2.1.1 Frame
– ALL: In this mode the sync pulses of each half frame can
trigger.
– Even: In this mode only the sync pulses of even half frames
can trigger.
– Odd: In this mode only the sync pulses of odd half frames
can trigger.
21.2.1.2 Line.
– All: In this mode all line sync pulses can trigger.
– Line No: The line number with its line pulse that is used for
triggering can be selected with the INTENS knob
– Line min: One pushbutton operation will be suffi cient to
switch back to the lowest possible line number.
21.2.2 Norm
The pushbutton allows the selection of the US standard of 525
lines and 60 Hz or the European standard with 625 lines and
50 Hz. With any change of standard the line number will be
automatically changed, too.
21.2.3 Polarity
Composite video signals may have both polarities. Selection of
the right polarity is vital as the scope should be triggered by the
sync pulses and not the video content.
Positive polarity is defi ned by the video content being more
positive than the sync signals and vice versa.
If the polarity was wrongly selected there will be no triggering
at all, an untriggered display or no signal capture.
to open the TRIGGER menu, select VIDEO, then
pushbutton. Further information can be
.
SOURCE pushbutton
Depressing this pushbutton will call various menus depending on the previously selected mode (MODE
EDGE, VIDEO, LOGIC. In XY mode the pushbuttons: MODE
FILTER
triggered.
In the TRIGGER SOURCE menu the source is selected from
which the trigger signal is to be taken. The options depend on
the actual mode of the scope.
, SOURCE are disabled as XY displays can not be
pushbutton):
,
Subject to change without notice
35
Controls and Readout
POWER
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
VOLTS / DIV
SCALE · VAR
CH 2HORMAG
VAR
15
14
16
18
13
17
POSITION 1POSITION 2
VOLTS / DIV
SCALE · VAR
20 V1 mV20 V1 mV
CH 1
VAR
22.1 Edge-/Video-Trigger
22.1.1 CH1
Conditions: Analog or DSO mode, EDGE or VIDEO selected.
CH1 will then be the trigger source, no matter whether it is
displayed or not. Readout: ”Tr:CH1, (Slope), Filter (TV)“.
22.1.2 CH2
Conditions: Analog or DSO mode, EDGE or VIDEO selected.
CH2 will then be the trigger source, no matter whether it is
displayed or not. Readout: ”Tr:CH2, (Slope), Filter (TV).“
22.1.3 Alt. 1/2
Conditions: Analog mode, EDGE-Triggering.
Alternate triggering with the signals from channels 1 and 2 as
described in the section ”Alternate Trigger“ of chapter ”Triggering
and time bases“. Please note that in this trigger mode the apparent time relationships between the two signals on the screen are
meaningless and misleading, the relative position of the two signals
depends only on their shape and the trigger level selected.
In dual channel mode (DUAL) alternate triggering is only possible in conjunction with alternate dual channel operation. If
previously dual trace chopped mode was selected (VERT/XY
pushbutton) > DUAL chop) it will be automatically changed to
alternate mode when alternate triggering is selected. After ”Alt.
1/2“ is turned off dual trace chopped mode may be selected
again. Readout: ”Tr:alt, Slope, Filter“.
22.1.4 External
In this mode the trigger signal comes from AUXILIARY INPUT
). Readout: ”Tr:alt, Slope, Filter“.
LEVEL A/B
TRIGGER
MODE
FILTER
SOURCE
AUX
DELAY
TRIG ’d
NORM
HOLD OFF
X-POS
50s5ns
VAR
HORIZONTAL
TIME / DIV
SCALE · VAR
x10
HOLD-OFF display (Analog mode only)
This display will light up if the hold-off time was set to > 0% in
order to indicate that the longer than minimum hold-off time
may cause a lower rep rate of the time base and thus a darker
display. Setting the hold-off time requires pressing the HOR
pushbutton
which calls the menu ”Time base“. Only the time
base A hold-off time may be changed.
See the section ”Hold-off time setting“ in the chapter ”Triggering
and time bases“.
X-POS DELAY pushbutton
This pushbutton allows to change the function of the
HORIZONTAL knob
26.1Analog mode
The pushbutton signals the actually selected function in accordance with the front panel lettering:
dark: X position control
green: Delay time control
26.1.1 X-POS
If the pushbutton is dark the HORIZONTAL-knob
as X position control, i.e. it moves the signal display horizontally.
The position control is especially useful when the magnifi er
(MAG. x 10
) is switched on. The magnifi er will magnify the
display 10 times around the screen centre, with the control the
portion of the signal to be studied can be shifted on-screen.
.
19
26
27
20
23
21
24
28
22
25
29
30
functions
22.1.5 AC Line
The trigger signal is taken from the line which feeds the scope.
See also the section ”Line triggering“ in the chapter ”Triggering
and time bases“. Readout: ”Tr:alt, Line, Slope“.
TRIG’d display (not in XY mode)
This LED will light up if the time base receives a trigger signal.
It depends upon the trigger signal whether the LED will just
blink or remain illuminated.
NORM display
This display will light up provided ”Auto“ triggering was not
selected. The mode can be selected in the ”Trigger“ menu called
by pressing (MODE
remain dark as long as there is no suffi cient trigger signal.
36
Subject to change without notice
). The light points out that the screen will
26.1.2 DELAY
In order to change the function of the HORIZONTAL-knob
to
DELAY proceed as follows:
Press the HOR
pushbutton which will present the TIME BASE
menu, select SEARCH or B ONLY, then the function of the knob
will be changed if the pushbutton is depressed. It will light up
to show that the knob is now the delay time control.
In SEARCH mode both traces (time base A and B) alternate.
Unlike the former time base A ONLY mode, a sector with higher intensity is visible on the A trace. This sector can be moved
continuously by the delay time control. The time between the
A trace start and the beginning of the intensifi ed sector is the
delay time. This information is also displayed in the readout (”Dt:
…“) and is an aid to fi nd the position of the intensifi ed sector
which may be very small. If time base B ONLY is chosen the
Controls and Readout
intensifi ed sector is no longer visible, but the DELAY function
still can be used.
Without activated B Trigger function, the B time base will be
started after the A time base delay time ”elapsed“.
26.2 DSO mode
The pushbutton will signal the actual function in accordance
with the front panel lettering:
dark: The knob functions as X position control.
green: The knob functions as delay time control.
26.2.1 X-POS
If the pushbutton is dark the HORIZONTAL
knob functions
as X position control of the trigger time, i.e. it moves the trigger
point symbol horizontally. This allows to display signal portions
before and after the trigger, called Pre-Trigger and Post-Trigger. If the trigger point symbol is located on the screen centre
the readout will show ”Tt:0s“, hence the trigger time indication
is always referred to the screen centre. Values with a positive
sign are Post-Trigger times, such with a negative sign PreTrigger times.
If the X-POS DELAY pushbutton is depressed the ”Hor.Knob“
menu will be called, it contains the following options:
st
1
Centre: Pressing the function pushbutton ”Center“ will set
the trigger time to the screen centre ”Tt:0s“ which is the
standard setting.
nd
2
Coarse On Off: changes the speed of the HORIZONTAL
knob.
26.2.2 DELAY
This pushbutton will be illuminated if the TIME BASE menu
was called with the HOR
B ONLY selected. The HORIZONTAL
pushbutton and SEARCH or
knob can then be used to
select a portion of the time base display which is to be displayed
expanded in time.
In SEARCH mode the normal and the expanded displays are
displayed simultaneously. The expanded portion of the signal
will be shown on the normal display as an intensifi ed sector.
The length of this sector is dependent upon the setting of the
nd
2
”Z“ ,time base which is shown in the readout as ”Z...“ and
is equal to the run time of the Z time base.
HORIZONTAL knob
The various functions of this knob depend on the operating mode
and are described under X-POS DELAY
pushbutton.
TIME/DIV.–SCALE–VAR knob
This knob is normally used as the time base speed selector, but
has also other functions dependent on the operating mode. In
XY mode this control is disabled.
28.1.2 Time base B time/cm selection
This function is active if in the ”Time base“ menu (HOR
push-button) SEARCH or B ONLY was selected and the option
”B variable On Off“ was set to Off.
Turning the control CCW will decrease, turning it CW will increase the time base speed. The speed can be selected between 20
ms/cm.. 50 ns/cm in a 1-2-5 sequence and will be calibrated.
The readout will show the speed (e.g. ”B:50ns“).
The time base B allows to display portions of the time base
A display on an expanded time base scale. This implies that
the speed of TB B must always be greater than that of TB A.
Therefore with the exception of 50 ns/cm TB B can not be set
to the same speed as TB A.
Further information is available in the section ”Time base B
nd
(2
time base/Delay/Triggering“ (Analog mode) in the chapter
”Triggering and time bases“.
28.1.3 Variable
The TIME/DIV–SCALE–VAR control may also be used to change
the time base speed continuously but uncalibrated. VAR will light
up on top of the HOR
pushbutton in order to warn that the
time base is uncalibrated and the knob has now that function.
In order to arrive at that function press HOR
which calls
the TIME BASE menu. Depending whether time base A or B is
selected either ”A variable On Off“ or ”B variable On Off“ will
be shown. The function pushbutton can then be used to select
On/Off.
In order to point out that the time base is now uncalibrated
the readout will replace ”:“ by ”>“ preceding the time/cm. (e.g.
”A>500ns“ and ”B>200ns“). Also the results of cursor time/period measurements will be marked that way.
28.2 DSO mode
28.2.1 ZOOM OFF (A time base time/cm selection)
Select the menu ”Zoom“ by pressing HOR
and then ”Off“ in
order to set the function of the knob TIME/DIV–SCALE–VAR to
time base A speed as in analog mode. If ”Zoom Off“ is active
always the whole memory will be displayed.
Turning the control CCW will decrease, turning it CW will
increase the time base speed. Depending on the signal capture/display the time base can be set from 50 s/cm to 5 ns/cm in
a 1-2-5 sequence (e.g. ”A:50ns“) and will be calibrated. There
is no variable function as in analog mode.
28.2.2 Search–Zoom only (Zoom time base speed selection)
One of the functions SEARCH or ZOOM ONLY may be selected in
the ZOOM menu after pressing HOR
. ”Zoom Off“ is equivalent
to time base A in analog mode. With ”Zoom only“ a portion of the
display in Zoom ”Off“ can be expanded over the whole screen.
This is possible because there is a very large memory for signal
capture/display. The ”Zoom Off“ display will present the whole
memory contents. With SEARCH both the ”Zoom Off“ and the
expanded ”Zoom only“ displays will be visible. With ”Zoom only“
only the expanded display will show up.
28.1 Analog mode
28.1.1 Time base A time/cm selection
This function is active if in the ”Time base“ menu (HOR
pushbutton) ”A only“ was selected and the option ”A variable
On Off“ was set to Off.
Turning the knob CCW will decrease, turning it CW will increase the time base speed. The time base speed may be chosen
between 500 ms/cm... 50 ns/cm in a 1-2-5 sequence and will be
calibrated. The readout will show the setting (e.g. ”A:50ns“).
The Zoom time base speed will be indicated in the readout ”Z:...“
and is calibrated. Turning the knob CCW will decrease, turning
it CW will increase the time base speed. This can be selected
from 20 ms/cm to 5 ns/cm in a 1-2-5 sequence. The maximum
expansion is 50,000 times (”A:10ms“ and ”Z:200ns“.)
MAG pushbutton
In analog mode only: pressing this pushbutton will turn on the
x 10 magnifi er. No menu will be shown.
Subject to change without notice
37
Controls and Readout
POWER
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
VOLTS / DIV
SCALE · VAR
CH 2HORMAG
VAR
15
14
16
18
13
17
POSITION 1POSITION 2
VOLTS / DIV
SCALE · VAR
20 V1 mV20 V1 mV
CH 1
VAR
If ”x10“ is illuminated on the MAG pushbutton the magnifi er
is activated. The adjusted time/cm will be shown in the top
left readout. Depending on the time base mode turning on the
magnifi er will have these effects:
29.1 ”Time base A only“
The display will be expanded around the screen centre by a
factor of 10, the time/cm adjusted accordingly.
29.2 ”Search“ (A and B times bases alternated)
The time base A speed will not be affected. The time base B
speed will be increased by a factor of 10, hence the time base
B display will be expanded 10-fold in X direction.
29.3 ”B only“:
The time base B speed will be increased by a factor of 10 and
the display hence expanded 10-fold in X direction.
HOR pushbutton
This pushbutton will open the ”Time base“ menu the contents
of which depends on the operating mode selected.
30.1 Analog mode
The following modes are available:
30.1.1 A only
Only the time base A will be operative. The readout will thus
only show ”A...“ in the top left corner. The TIME/DIV-SCALEVAR knob will set the time base A speed. With the MAG x10
pushbutton the display may be expanded around the screen
centre, increasing the speed by 10. If the mode is changed from
time base A to ”Search“ or ”B only“ all settings of time base A
remain intact including triggering.
30.1.2 Search
This mode implies alternate time base operation. The readout
will show the speeds of both time bases (”A...“ and ”B...“). The
TIME/DIV-SCALE-VAR knob will set the time base B speed.
In alternate time base mode part of the time base A display will
be intensifi ed. The horizontal position of the intensifi ed portion
may be shifted using the HORIZONTAL
function was set to ”Delay“, this is the case if the X-POS DELAY
pushbutton is illuminated. The length of the intensifi ed sector
is determined by the speed of time base B and is equal to the
run time of B. Hence this intensifi ed portion of A will be spread
over the full screen as displayed by B and thus expanded. The
Y position of the signal is the same when displayed with either
knob, provided its
19
LEVEL A/B
X-POS
HORIZONTAL
26
27
TRIG ’d
NORM
HOLD OFF
DELAY
VAR
TIME / DIV
SCALE · VAR
50s5ns
x10
20
23
21
24
28
22
25
29
TRIGGER
MODE
FILTER
SOURCE
AUX
30
A or B. This means, however, that both displays will be written
over each other.
In order to separate them on the screen for better readability
an artifi cial Y offset may be added to the time base B display.
Press the CH1/2-CURSOR-MA/REF-ZOOM
pushbutton
which calls the ”Pos./Scale“ menu. Press the function pushbutton ”TB B“, this will cause the POSITION 1 knob to act as
the position control for the time base B display. (Trace separation.) See also 13.1.4 Y position 2
nd
time base. This function
makes sense only in ”Search“ and is unavailable in any other.
Also in ”Search“ the 10 x magnifi er is available by pressing
the pushbutton MAG x10
. The magnifi er will affect solely
time base B.
30.1.3 ”B only“:
In this mode only time base B will be displayed, the readout
will thus only show the time base B speed in the top left corner
(”B..“). The TIME/DIV-SCALE-VAR knob will set the time base
B speed. The 10x magnifi er is available by pressing MAG x10
and will expand the display around the screen centre.
30.1.4 B trigger –
Edge
In this mode time base B will not start immediately after the
delay time set elapsed, but it will be only set ready waiting
for a signal trigger. This has the advantage that any jitter is
removed, but the delay time adjustment will now only have
the effect that the time base B display will jump from signal
period to period. In this setting a positive slope will trigger.
The (trigger) LEVEL A/B
knob will set the trigger level for
B. Only normal triggering and DC-coupling are possible. All
parameters of time base A remain stored and preserved. (LEVEL, auto or normal, Slope, coupling). In addition to the delay
time (”Dt:...“) also the B trigger parameters are shown in the
readout: ”BTr:slope, DC“. In SEARCH mode the trigger point
symbol will be preceded by ”B“. As mentioned changing the
delay time will not cause a continuous move of the intensifi ed
portion of the time base A display and the time base B display,
but jumps from signal period to period.
If the trigger level symbol of time base B is shifted outside the
signal representation by time base A there will be no triggering
of time base B any more and thus no time base B display. The
same holds in time base B only mode.
30.1.5 B trigger –
Edge
Except for the negative edge the function is identical to the one
described above (30.1.4).
38
Subject to change without notice
Controls and Readout
30.1.6 B trigger – OFF
Time base B will be started upon the end of the delay time set.
The delay time can be changed continuously in this mode which
can be watched on the intensifi ed sector of the time base A
display. The disadvantage here is that with very long delay times
jitter of the time base B display may crop up.
As time base B is not operated in the signal triggered mode
the controls for time base B trigger will be disabled resp. are
only for time base A.
30.1.7 A variable – On Off
If ”On“ was selected the TIME/DIV-SCALE-VAR knob
function as variable control for the time base A speed. Only in
time base A only mode this option will be available in the menu.
For a full description see ”28.1.3 Variable“.
30.1.8 B variable – On Off
If ”On“ was selected the TIME/DIV-SCALE-VAR knob
function as the time base B variable control. For a full description see ”28.1.3 Variable“.
30.1.9 Holdoff …%
In this mode the hold-off time may be selected from 0 to 100 %
with the INTENS knob
after a sweep before a new one can start and decrease thus the
repetition rate which may darken the display. This is indicated
by the HOLD OFF-LED
valid for time base A.
Further information can be found in the section ”Hold-off adjustment“ in the chapter ”Triggering and time bases“.
30.2. DSO mode
In the ”Zoom“ menu the following time base functions are
available:
30.2.1 Off
In ”Zoom Off“ condition only time base A is active. The readout
will thus only show ”A...“ in the top left corner. The speed can
be set with the TIME/DIV-SCALE-VAR knob
30.2.2 Search
Part of the time base A display will be intensifi ed and this
portion will also be displayed expanded over the full screen.
With the HORIZONTAL knob
expanded display can be shifted provided that the pushbutton
X-POS DELAY
length of the intensifi ed sector is determined by the speed of
the Z time base.
The Y position of both displays is identical so they are written one
over the other. In order to separate them for better readability
an artifi cial Y offset may be added to the Z time base display.
Press the pushbutton CH1/2-CURSOR-MA/REF-ZOOM
call the menu ”Pos./Scale“.
Then press the function pushbutton ZOOM. Now the POSITION
1 knob will function as Y position control for time base Z. As this
only makes sense in SEARCH mode it is unavailable in others.
30.2.3 Zoom only
Only the Z time base will be displayed. The readout will hence
only show ”Z...“ in the top left corner. The TIME/DIV-SCALE-VAR
knob affects only time base B.
is illuminated which means ”Delay“. The
. Values > 0 extend the waiting time
lighting up. The hold-off time is only
.
the intensifi ed sector and the
will
will
to
31.1. AC DC
Pressing the pushbutton will switch from AC to DC or vice versa.
The mode selected will be shown in the readout following the
sensitivity setting: ~ is for AC and = is for DC.
31.1.1 DC coupling
The signal will be directly coupled, from the BNC connector via
the attenuator to the vertical amplifi er. The input resistance is
1 MΩ in all positions of the attenuator.
31.1.2 AC coupling
A capacitor is inserted between the BNC connector and the
attenuator, blocking the DC content of the signal and creating a
low frequency cut-off at approx. 2 Hz. This will affect the shape
and amplitude of signals with low frequency content. If the DC
content of the signal changes or the duty cycle of pulses the
capacitor will charge or discharge, this will cause a momentary
Y shift of the display.
31.2 Ground On Off
The pushbutton will either connect the amplifi er to the signal
or to ground. If set to Ground the readout will show a ground
symbol following the sensitivity setting, at the same place where
formerly the coupling was indicated. In the Ground position and
with automatic triggering a trace will be visible, this is handy
for setting the Y position of it e.g. to the screen centre without
disconnecting the signal. The readout will show a symbol (
0 V which will be close to the vertical centre line of the graticule,
it is the zero reference for any measurements. After switching
back to the signal its amplitude can now be determined with
respect to the formerly set zero reference.
31.3 Invert On Off (unavailable in analog XY mode)
This pushbutton will alternate between not inverted or inverted
of the CH1 signal. The readout will indicate an inverted display
by placing a bar above the CH1. The trigger signal derived from
Ch1 will not be affected by an inversion.
31.4 Probe submenu
Pressing the pushbutton will open the ”CH1 probe“ submenu.
31.4.1 *1 - *10 - *100 - *1000
A selection of probe attenuation factors from 1 to 1,000 is possible. The value selected will be automatically taken into account
regarding the sensitivity indicated and measurements.
31.4.2 Auto
If Auto is selected HAMEG probes with automatic probe identifi cation connectors will be recognized by the scope and the
appropriate factor taken into account. The factor recognized
will be indicated following ”auto“.
Probes without that special connector will cause the display
”auto *1“ and treated as 1:1 probes.
31.5 Variable On Off
If activated VAR on the CH1 pushbutton
The readout will replace ”:“ by ”>“ (e.g. ”CH1>5mV“.) and this
also in any cursor measurements pointing out that the sensitivity
is uncalibrated.
The VOLTS/DIV-SCALE-VAR knob
which can change the sensitivity continuously between 1 mV/cm
to > 20 V/cm.
will be illuminated.
will function as variable
) for
CH1 pushbutton
This pushbutton opens the CH1 menu which contains the following options referring to CH1
resp. to the signal on CH1.
VERT/XY pushbutton
This pushbutton switches the ”vertical“ menu on/off. This menu
allows to select the operating modes of the vertical amplifi ers.
Subject to change without notice
39
Controls and Readout
POWER
CH 1/2
CURSOR
MA/REF
ZOOM
AUTO/
CURSOR
MEASURE
VERT/XY
VOLTS / DIV
SCALE · VAR
CH 2HORMAG
VAR
15
14
16
18
13
17
POSITION 1POSITION 2
VOLTS / DIV
SCALE · VAR
20 V1 mV20 V1 mV
CH 1
VAR
32.1CH1
If CH1 is selected only CH1 will be turned on, the mode is Yt. Also
the readout will only display the parameters of CH1. (sensitivity,
inverted/not inverted, coupling.)
Although CH2 will not appear in the readout it may be used e.g.
as a trigger input. Its controls are active but are not shown.
32.2CH2
If CH2 is selected only CH2 will be active, it is Yt mode, and only
its parameters will be shown in the readout.
Although CH1 will not appear in the readout it may be used e.g.
as a trigger input. Its controls are active but are not shown.
32.3.1 DUAL trace alt./ chop
In dual trace mode both channels are turned on and the parameters of both are shown in the readout. Between the sensitivity indications there is an indication whether alternate ”alt.“ or chopped
”chp“ mode is active. Normally, the mode will be automatically set
by the time base speed selection, but it may be directly set using
the function pushbutton. For time base speeds of 500 ms/cm to
500 μs/cm chopped will be used, from 200 μs/cm to 50 ns/cm
alternate. This refers to unmagnifi ed time bases.
Alternate is the preferred mode, at any time one channel is displayed for a full sweep, after each sweep the other channel has
its turn. At slow sweep speeds this will cause annoying fl icker,
at still slower ones the channel switching becomes visible. Here,
the chopped mode steps in, both channels are switched at some
high frequency so they are both visible at any sweep speed.
This is, however, not appropriate for fast sweep speeds as the
switching may become visible and may interfere with the proper
signal display.
32.3.2 DUAL (DSO mode).
In DSO mode there is an a/d converter for each channel so both
are measured simultaneously. Hence no channel switching is
necessary and no information pertaining to it is shown.
32.4 ADD
In ADD mode the signals of both channels are algebraically
added and displayed as one sum signal. The Y position can be
changed with both position controls. If one channel is inverted
the difference will be displayed. Only one ”0 V“ symbol will
be shown in the readout. The ADD mode will be indicated by
placing a ”+“ symbol between the sensitivity indications of both
channels. In DSO mode ”1+2“ will be displayed at the end of
the trace.
Please note that the results of cursor measurements in this
mode will only be correct if the sensitivities of both channels are
identical, otherwise the readout will show ”CH1<>CH2“.
19
LEVEL A/B
X-POS
HORIZONTAL
26
27
TRIG ’d
NORM
HOLD OFF
DELAY
VAR
TIME / DIV
SCALE · VAR
50s5ns
x10
20
23
21
24
28
22
25
29
TRIGGER
MODE
FILTER
SOURCE
AUX
30
Automatic voltage measurements can not be performed in ADD
mode. The readout will show ”n/a“ = not available.
As the trigger signals are taken off the inputs and not from the
added signal there is no true reference for the trigger point
symbol, the symbol will thus be switched off in analog mode.
However, the LEVEL A/B control
is active.
In DSO mode a trigger time symbol is displayed one line above
the lowest graticule line to indicate the trigger time position
along the signal and thus can only be moved horizontally.
32.5 XY
In this mode CH1 will move the trace in X direction, hence the
readout will show ”CHX..“, CH2 will move the trace in Y direction,
hence ”CHY...“ will be shown rather than ”CH2...“.
As the time bases are not involved in XY no time base related
information will be shown. Also the trigger circuits are disabled
so no trigger information is shown, either.
The magnifi er MAG x10
is disabled. The ”0-Volt“ symbols
will be shown as triangles at the right hand graticule and above
the sensitivities.
Both the HORIZONTAL
or the POSITION 1 knobs will
move the trace horizontally. The Y position is controlled by the
POSITION 2 knob.
32.5.1 Analog mode
The CH1 signal can not be inverted, there is hence no menu item
in the CH1 menu (CH1 pushbutton
VAR
knob is disabled.
). The TIME/DIV-SCALE-
Please note that the bandwidths and phase differences in XY
analog and DSO modes differ considerably so there may be
changes in the signal display when switching the mode.
32.5.2 DSO mode
The readout will indicate the sampling rate with which the a/d
converters digitise the input signals. The appropriate sampling
rate must be set depending on the signals and can be selected
with the TIME/DIV-SCALE-VAR knob, although the time bases
are disabled. With high sampling rates there may be gaps in
Lissajous representations. With too low sampling rates the
display may not allow any more to determine the frequency
relationship of the signals.
It is recommended to fi rst look at the signals in DUAL mode
and to set the sampling rate such that at least one signal period
40
Subject to change without notice
Controls and Readout
POWER
CH 1
VAR
X-INP
!
CAT I
3431
VERT/XY
INPUTS
1MΩII15pF
max
400 Vp
32
CH 2HORMAG
VAR
35
33
will be displayed. Then XY should be selected. In XY DSO mode
both channels may be inverted.
32.6 Bandwidth Full/20 MHz
This pushbutton will select full or 20 MHz bandwidth.
Full: Full bandwidth will be the one given in the specifi cations.
20 MHz: Provided measuring modes allow full bandwidth
(i.e. ≥ 5 mV/cm) this can be reduced to 20 MHz (–3 dB) in order
to attenuate high frequency noise e.g. The readout will show
BWL = bandwidth limited. The bandwidth limitation affects both
channels and pertains to analog and DSO modes.
In XY DSO mode the limitation is equal to Yt mode. In XY analog
mode the limitation affects only CH2.
CH2 pushbutton
This pushbutton opens the CH2 menu which offers the following
options:
33.1 AC DC
The pushbutton will alternate between AC and DC coupling.
The readout shows a ”~“ or ”=“ symbol behind the sensitivity
indication.
33.1.1 DC coupling
The signal will be directly coupled to the input amplifi er via
the BNC connector
resistance of the scope is a constant 1 MΩ irrespective of the
sensitivity selected.
33.1.2 AC coupling
A capacitor is inserted between the BNC connector and the attenuator, thus the DC content of the signal is blocked and a high
pass with a lower cut-off frequency of approx. 2 Hz is created.
Low frequency signals will thus be more or less differentiated,
hence their shape and amplitude affected.
If the DC content of the signal changes, e.g. the duty cycle of
pulses, the capacitor must charge or discharge. This will cause
a momentary Y shift of the display.
33.2 Ground (GND) On Off
The pushbutton will alternate between switching the amplifi er
input to the signal or to ground.
and the input attenuator. The input
AUX
VAR
AUXILIARY INPUT
TRIGGER
EXTERN
!
CAT I
3637
Z-INPUT
x10
1MΩ II
15pF
max
100 Vp
29
30
Referred to the trace position 0 V a DC voltage may be measured
after the input was returned to the signal.
33.3 Invert On Off
The pushbutton will alternate between not inverted and inverted.
An inverted signal will be indicated in the readout by bar above
the channel symbol. The trigger signal taken from an input will
not be affected.
33.4 Probe menu
This pushbutton opens the ”CH2 probe“ submenu.
33.4.1 *1 - *10 - *100 - *1000
A selection of 4 factors can be made, the factor chosen will be
automatically taken into account for all displays and measurements.
33.4.2 auto
If ”auto“ was selected HAMEG probes with special probe identifi cation connectors will be automatically identifi ed and the
appropriate factor taken into account. The probe factor identifi ed
will be shown behind ”auto“.
Probes without that special connector will be treated as 1:1
probes (display ”auto*1“), for those the factor must be manually set.
33.5 Variable On Off
If the variable is on VAR on the pushbutton will be illuminated
and indicates that the sensitivity is now uncalibrated, the readout
will show ”>“ instead of ”:“ (e.g. ”CH2>5 mV“) The results of
cursor measurements will be identifi ed accordingly.
The VOLTS/DIV-SCALE-VAR knob
of CH2 has now the
function of variable, the sensitivity can be varied between
1 mV/cm to > 20 V/cm.
INPUT CH1 – BNC connector
This is the CH1 signal input connector. In Yt mode it is a Y input,
in XY mode it is the X signal input. The connector housing is
connected to the instrument housing and thus to safety ground.
The ring around the connector is the probe identifi cation contact,
no voltage may be applied here.
If the signal is disconnected resp. the amplifi er input connected
to ground the readout will show a ground symbol behind the
sensitivity indication. In automatic trigger mode the trace will
be visible in a reference position which can be used as a 0 V
ground reference. The readout will show a symbol ( ) for 0 V
which will be close to the vertical centre line of the graticule, it
is the zero reference for any measurements.
INPUT CH2 – BNC connector
This is the CH2 signal input connector. It is a Y input in Yt and
XY mode. The connector housing is connected to the instrument housing and thus to safety ground. The ring around the
connector is the probe identifi cation contact, no voltage may
be applied here.
Subject to change without notice
41
Controls and Readout
POWER
POWER
CH 1
VAR
X-INP
!
CAT I
3431
VERT/XY
INPUTS
1MΩII15pF
max
400 Vp
32
CH 2HORMAG
VAR
33
35
3637
AUX pushbutton
This pushbutton belongs to the AUXILIARY INPUT . The opening of the menu depends upon the actual operating mode.
36.1 Analog mode
36.1.1 AUXILIARY INPUT
setting external triggering press SOURCE
Source“ menu, then „External“.
36.1.2 If „External“ triggering was not selected the menu „Z
Input“ will open up. If „Off“ is chosen AUXILIARY INPUT
no function. If „On“ is chosen it will function as Z input i.e. intensity modulation input. This input is destined for TTL signals,
a voltage of >1 V
will turn off the trace.
p
is the external trigger input. For
, select the „Trig.
has
AUX
VAR
x10
29
30
AUXILIARY INPUT
!
CAT I
TRIGGER
EXTERN
Z-INPUT
1MΩ II
15pF
max
100 Vp
PROBE ADJ. – connector
A square wave signal of 0.2 Vpp is available for the adjustment
of 10 : 1 probes. The frequency can be selected by pressing
the pushbutton PROBE ADJ. and calling the menu ”Utilities“.
Further information may be found in the section ”Probe adjustment and use“ in the chapter ”Operation and pre-settings“.
PROBE ADJ – pushbutton
This pushbutton produces the menu ”Utilities“, two options
are available:
39.1 COMP. Tester On Off.
If ”On“ is chosen analog mode is selected automatically. A trace
and a readout display ”Component Tester“ will appear.
36.2 DSO mode
36.2.1 In DSO mode the AUX pushbutton is deactivated and the
AUXILIARY INPUT
can only serve as an input for external
trigger signals.
AUXILIARY INPUT (BNC connector)
In DSO mode this connector can be used as external trigger
input.
In analog mode it can serve as external trigger or Z-axis (intensity modulation) input.
The connector housing is connected to the instrument housing
and thus to safety ground. Although the ring around the connector has no function, no voltage may be applied here.
CH I: 500 mV
In this mode the 4 mm connectors labelled ”COMPONENT
TESTER“ become the measuring input. See also the chapter
”Component Tester“.
Choosing ”Off“ will return the instrument to all former settings.
39.2 Calibrator 1kHz 1MHz
Depending on the setting the square wave signal frequency at
the PROBE ADJ – socket is 1 kHz or 1 MHz.
COMPONENT TESTER - (sockets)
Both 4 mm diameter sockets serve as a two pole input for
component test. Further information can be found under section
Component Tester.