HAMAMATSU L8121-01 Datasheet

150 kV
NEW
MICROFOCUS X-RAY SOURCE
L8121-01
FEATURES APPLICATIONS
Focal Spot Size: 5 µm (at 4 W)
Providing 5 µm focal spot size, for the sealed-off type X-ray source with built-in power supply. It gives a clear clair­voyance image despite of higher magnification ratio.
Easy Handling
Convenient handling (with no high voltage wiring) since the air cooled X-ray source and the high voltage power supply are in one package.
Continuous Operation
It allows continuous maximum output, 150 kV and 500 µA, operation.
Built-in External Control Function
It is automatically operable from external control since it has a built-in RS-232C interface.
X-ray Non-destructive Test
• Printed Circuits • Electronic Components
• Semiconductor Devices • Ceramics
• Soldered Parts • Rubbers
• Combinated Materials • Plastics
• New materials • Die casting
Multilayer Printed Circuit Board X-ray Inspection•Drilling Instrument
Industrial On-line Process Control and Measurement
Industrial X-ray Computer Tomography System
SPECIFICATIONS
GENERAL
Parameter
Target Voltage Target Current Maximum Output Power
“S” Small Spot Mode
Focal Spot Size
Beam Angle Minimum Distance Focus/Object Environmental X-ray Dose Leakage Operation
“M” Middle Spot Mode “L” Large Spot Mode
Value/Description
40 to 150
0 to 500
75
7 (10 W isowatts), 5 (4 W)
20 (30 W isowatts)
50 (75 W)
43 17 20
Continuous
Unit
kV µA
W
µm µm µm
degree
mm
mSv/h Max.
X-RAY TUBE UNIT
Parameter Description
X-ray Tube Cooling Method Window Material Target Material Cathode Material Window Position High Voltage Power Supply Operation /Storage Temperature Operation/Storage Humidity Weight
NOTE: a
Without moisture condensation.
+10 °C to +40 °C / 0 °C to +50 °C
Sealed Type
Forced Air Cooling
Beryllium (200 µmt)
Tungsten
Dispenser Cathode
End-window
Built-in Type
85 % RH Max.
Approx. 17 kg
a
X-RAY CONTROL UNIT
Parameter Description
Function
Protection External Control Operation /Storage Temperature Operation/Storage Humidity Power Consumption Inter Voltage
Installation Method
Weight
Target Voltage/Current Pre-set, Auto Warm-up
Safety Interlock
RS-232C
0 to +40 °C
85 % RH Max.
Less than 240 W
24 Vdc
Designed for both bench-top
and rack mounting
Approx. 5 kg
a
Loading...