MCP ASSEMBLY
RESPONSE TIME (0.5ns/div)
OUTPUT VOLTAGE (5mV/div)
RISE TIME: 293ps
FALL TIME: 539ps
FWHM : 455ps
SUPPLY VOLTAGE (MCP): -1520V
FOR HIGH RESOLUTION TOF-MS
PRELIMINARY DATA
MAY 1998
High-speed response MCP assembly ideal for high resolution TOF-MS detector
Incorporates a two-stage MCP with 4
The F4655-13 MCP assembly is an ideal detector for high resolution TOF-MS (time-of-flight mass spectroscopy) because of a
two-stage MCP with 4µm channel diameter. Our own advanced
technology in optimizing MCP design allows obtaining an excellent output waveform with negligible ringing even when detecting very high-speed signals. The dimensions are exactly the same
as the F4655-10 MCP assembly, ensuring a compact configuration and easy handling.
F4655-13
µµ
µm channel diameter
µµ
FEATURES
Excellent time response
ΩΩ
50
Ω impedance matching
ΩΩ
Compact and lightweight
SPECIFICATIONS
GENERAL
Parameter Value Unit
Assembly Outer Diameter 38.0 mm
Effective Diameter 14.5 mm
Maximum Height 31.9 mm
MCP Channel Diameter 4
Number of MCP Stages 2 -
MAXIMUM R ATINGS (A bso lute v al u es )
MCP Supply Voltage (In-Out) 2 kV
MCP-Out to Anode Voltage 0.5 kV
Electric Potential at Each
Terminal
Operating Vacuum Condition
Baking Temperature
Baking Time
Baking Vacuum Condition
∗
Baking cannot be performed.
CHARACTERI STI CS ( at 1. 3×10-4 Pa (1×10-6 Torr), Ta=+25°C)
Gain at 2.0kV
Plate Resistance per MCP 10 to 100
Dark Count at 2.0kV 5 cps/cm
.
Subject to local technical requirements and regulations, availability of products included in this promotional material may v ary. Please consult with our sales office.
Information furnished by HAMAMA TSU is believed to be reliable. Ho w e v er, no responsibility is assumed for possible inaccuracies or omissions. Specifications are subject
to change without notice. No patent rights are granted to any of the circuits described herein. ©1998 Hamamatsu Photonics K.K.
MCP-In: -2.5
MCP-Out: -0.5
Anode: GND
-4
-
∗
-
∗
- ∗
1×10
(5×10-6)
6
6.7×10
m
µ
kV
Pa (Torr)
°C
h
Pa (Torr)
-
MΩ
TMCPF0088
APPLICATION
TOF-MS (Time-of-flight mass spectr ometer)
Figure 1: Typical Output Wavef orm
TMCPB0079EA
2
MCP ASSEMBLY FOR HIGH RESOLUTION TOF-MS F4655-13
Figure 2: Dimensional Outlines (Unit: mm)
MCP-IN ELECTRODE
(SUS-NM)
MCP-OUT ELECTRODE
(SUS-NM)
2- 2.2
(P.C.D. 26)
3
30°
30°
3-M1.6 SCREWS
(P.C.D. 21.5)
PROTECTION ELECTRODE
FOR CHARGING-UP
(SAME POTENTIAL AS SUBSTRATE)
(SUS-NM)
EFFECTIVE AREA
14.5
3-CAPACITORS
3- 3.5
(P.C.D. 32)
186 38
MCP (2 STAGE)
3
0.5
10.3
13.8
SUBSTRATE
(SUS 304)
INSULATOR (Al2O3)
ANODE
(SUS316L)
BNC-R
3
18.1
TMCPA0044EA
HAMAMATSU PHOTONICS K.K., Electron Tube Center
314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193 Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205
U.S.A.
: Hamamatsu Corporation: 360 Foothill Road, P.O.Box 6910, Bridgewater, N.J. 08807-0910, U.S.A., Telephone:(1)908-231-0960, Fax:(1)908-231-1218
Germany
: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone:(49)8152-375-0, Fax:(49)8152-2658
France
: Hamamatsu Photonics France: S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone:(33)1 69 53 71 00, Fax:(33)1 69 53 71 10
United Kingdom
North Europe
Italy
: Hamamatsu Photonics UK Limited: Lough Point, 2 Gladbeck Way, Windmill Hill, Enfield, Middlesex EN2 7JA, United Kingdom, Telephone:(44)181-367-3560, Fax:(44)181-367-6384
: Hamamatsu Photonics Norden AB: FŠršgatan 7, S-164 40 Kista, Sweden, Telephone:(46)8-703-29-50, Fax:(46)8-750-58-95
: Hamamatsu Photonics Italia S.R.L.: Via della Moia, 1/E 20020 Arese, (Milano), Italy ,Telephone: (39)2 935 81 733 Fax: (39)2 935 81 741
TMCP1021E01
MAY 1998 SI
Printed in Japan (500)