Hach-Lange SI628 C User Manual [en, de, es, fr, it]

DOC023.98.90242
Transmitters si628 C
Manual
Bedienungsanleitung
Manual
Manuel d’utilisateur
Manuale di istruzioni
05/2010, Edition 1
05/2010, Edición 1
05/2010, Edition 1
05/2010, Edizione 1
© HACH LANGE GmbH, 2010, All rights reserved. Printed in Spain
si628 C_HACH LANGE_Ed. 0510.indd IND:1si628 C_HACH LANGE_Ed. 0510.indd IND:1 20/5/10 11:08:3920/5/10 11:08:39
si628 C_HACH LANGE_Ed. 0510.indd IND:2si628 C_HACH LANGE_Ed. 0510.indd IND:2 20/5/10 11:08:4020/5/10 11:08:40
ENGLISHDEUTSCHESPAÑOLFRANÇAISITALIANOCONTACT
si628 C_HACH LANGE_Ed. 0510.indd IND:3si628 C_HACH LANGE_Ed. 0510.indd IND:3 20/5/10 11:08:4020/5/10 11:08:40
si628 C_HACH LANGE_Ed. 0510.indd IND:4si628 C_HACH LANGE_Ed. 0510.indd IND:4 20/5/10 11:08:4020/5/10 11:08:40
Table of contents
1. Technical Specifications . . . . . . . . . . . . . . . . . . . 3
1.1. Dimensions
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2. General information
2.1. Safety information
2.1.1. Use of hazard information
2.2. General product information
2.3. Items supplied
3. Installation
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1. si628 C (wall mount)
3.1.1. Assembly
3.1.2. Connections
3.2. si628 C (panel mount)
3.2.1. Assembly
3.2.2. Connections
3.3. Cell cables
3.3.1. Cells with MP-5 connector
3.3.2. Cells with connector S8
3.4. Protective wiring of switching contacts
4. Operation
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.1. Description
4.2. Setting up
4.3. Quick guide
4.4. Programming
4.5. Calibration.
4.5.1. Recognized standards
4.6. Measurement
4.7. Access codes
. . . . . . . . . . . . . . . . . . . . . . 5
. . . . . . . . . . . . . . . . . . . . . . 5
. . . . . . . . 5
. . . . . . . . . . . . . 6
. . . . . . . . . . . . . . . . . . . . . . . . . . 6
. . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . . . . . . . 8
. . . . . . . . . . . . . . . 10
. . . . . . . . . . . . . . . . . . . . . . . . 10
. . . . . . . . . . . . . . . . . . . . . 11
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
. . . . . . . 12
. . . . . . . . . 13
. . . . 13
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
. . . . . . . . . . . . . . . . . . . . . . . . . . . 15
. . . . . . . . . . . . . . . . . . . . . . . . . . 16
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
. . . . . . . . . . . . 18
. . . . . . . . . . . . . . . . . . . . . . . . . . 19
. . . . . . . . . . . . . . . . . . . . . . . . . . 20
ENGLISH
5. Maintenance
5.1. Cleaning
5.2. Fuse replacement
6. Error messages
7. Replacement parts
8. Warranty and liability
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
. . . . . . . . . . . . . . . . . . . . . . . 21
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
. . . . . . . . . . . . . . . . . . . . . . . . 23
. . . . . . . . . . . . . . . . . . . . 24
Contact information
11
si628 C_HACH LANGE_Ed. 0510.indd ANG:1si628 C_HACH LANGE_Ed. 0510.indd ANG:1 20/5/10 11:08:4020/5/10 11:08:40
si628 C_HACH LANGE_Ed. 0510.indd ANG:2si628 C_HACH LANGE_Ed. 0510.indd ANG:2 20/5/10 11:08:4020/5/10 11:08:40
1. Technical specifications
Measuring ranges
Conductivity * 0.01 µS/cm to 199.9 mS/cm * with C=0.1 cm -1 cell °C –20.0 a 150.0
Measuring error
Conductivity 1% (±1 digit) Temperature 0.5°C (±1 digit)
Reproducibilidad
Conductivity ±0.1% (±1 digit) Temperature ±0.1°C (±1 digit)
Automatic temperature compensation
With temperature probe Pt 1000 or by keypad
Temperature coefficient, TC
Fixed at 2 % / °C
Reference temperature, RT
Fixed at 25 °C
Accepted cells
2 pole-cell. Constant between 0.05 and 50 cm-1
Recognised standards
147 µS/cm, 1413 µS/cm and 12.88 mS/cm (25°C)
Calibration possibilities
With 1 or 2 standards Access code: 100
Accepted calibration, 2 standards
Diference <30 % between the obtained cell constants
Control mode
ON / OFF. A relay programmable as maximum or minimum Activation relays lag 2 seconds Alarm relay programmable between 1 and 99 ss, min or hours
Language
English, German, Spanish, French, Italian, Portuguese
Key pad
LCD, alphanumeric, backlit, 2 lines of 16 characters
Inputs and outputs
2 pole cell, terminal strip Pt 1000 ATC, terminal strip Analogic output, galvanically isolated:
• 4 to 20 mA for measurement (R max.= 500 Ω)
• 21 mA or the last mA value measured for the instru­ment in hold
• 22 mA for the alarm
Relés
1 limit relay, with no potential. NO contact. 1 alarm relay, with no potential. NC Contact. Maximum load: en CA < 24 V / < 3 A / < 72 VA
Maximum cable length
25 m
Power supply
230 or 24 VAC ±10 %, 45-65 Hz, (115 VAC under demand) Consumption 4 VA Protection: Class II Overvoltage category: II
EC Directives
Electrical security according to 2006/95/EC Electromagnetic compatibility according to 2004/108/EC
Environmental conditions
Working temperature: 0 to 50 °C (32 to 122 °F) Storage temperature: –20 to 65 °C (–4 to 149 °F) Relative humidity < 80 % (non condensing) Max. height 2.000 m. at 230 V; 3.000 at 24 V
Enclosure
si628 C (wall mount): Thermoplastic material, protection IP55. Flame rating HB si628 C (panel mount): Noryl, protection IP54. Flame rating FV-1
Weight
si628 C (wall mount): 700 g si628 C (panel mount): 600 g
Dimensions
si628 C (wall mount): 105 x 170 x 100 mm (4 x 6.7 x 3.9 in.) si628 C (panel mount): 96 x 96 x 100 mm (3.8 x 3.8 x 3.9 in.)
Warranty
Instruments: 2 years
Specifications are subject to change without notice.
ENGLISH
3
si628 C_HACH LANGE_Ed. 0510.indd ANG:3si628 C_HACH LANGE_Ed. 0510.indd ANG:3 20/5/10 11:08:4020/5/10 11:08:40
1. Technical Specifications
1.1. Dimensions
si628 C (wall mount)
105 mm
4.1 in
6.7 in
170 mm
45 mm
1.8 in
50 mm
2 in
si628 C (panel mount)
90 mm
3.5 in
3.5 in
90 mm
3.8 in
96 mm
80 mm
3.1 in
96 mm
3.8 in
4
si628 C_HACH LANGE_Ed. 0510.indd ANG:4si628 C_HACH LANGE_Ed. 0510.indd ANG:4 20/5/10 11:08:4020/5/10 11:08:40
7 mm
0.3 in
2. General information
As a result of constant improvements to our products sometimes differences may exist between this manual and the instructions supplied with the instrument.
ENGLISH
2.1. Safety information
Please, read carefully this information before installing and using the instrument ! Pay attention to all danger and caution statements.
2.1.1. Use of hazard information DANGER
Indicates a potentially or imminently hazardous situation that, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation that, if not avoided, may result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that, if not avoided, may result in minor or moderate injury.
Important note
Indicates a situation that, if not avoided, could lead to damage to the instrument. Important infor­mation that requires special emphasis.
Note
Information that supplements points in the main text.
Precautionary labels
Read carefully all labels and tags attached to the instrument.
This symbol references the instruction manual for operation or safety information.
This symbol indicates that a risk of electrical shock or electrocution exists.
This symbol indicates the connection for Protective Earth.
This symbol indicates ground connection.
This symbol identifies a fuse.
Electrical equipment marked with this symbol may not be disposed of in European public disposal sys­tems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no charge to the user. Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
5
si628 C_HACH LANGE_Ed. 0510.indd ANG:5si628 C_HACH LANGE_Ed. 0510.indd ANG:5 20/5/10 11:08:4020/5/10 11:08:40
2. General information
DANGER
• Remember that the voltage across accessible parts of the open instrument, may be dangerous to life. Do not open the instrument when it is connected to the power supply, there are areas whose voltage could cause death.
• Always disconnect power to the instrument before it is opened.
• Repair or adjustment of an opened instrument under voltage shall be carried out only by a qualified technician who is aware of the hazards involved.
• Never work in an environment subject to explosion hazards. The housing of the instrument is not gas tight.
• The Company will not be responsible for any physical damage caused by non authorised work or manipulation.
• It is the responsibility of whoever uses this instrument to consult and establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
• Before connecting the instrument to a power supply unit, make sure that the mains voltage lies within the range:
230 or 24 V AC ±10%, 45-65 Hz (standard
version).
115 V AC ±10%, 45-65 Hz (under demand).
Important notes
• Read carefully the manual of the instrument.
• Avoid direct sunlight over the instrument.
• Exclude the following environmental influences:
- vibrations
- atmospheric humidity higher than 80 %
- corrosive gases present
- temperature < 0 °C, or > 50 °C (< 32 °F, or >
122 °F).
- powerful electric and magnetic fields.
• Only use original accessories and spare parts.
• Have the instrument serviced only by the Company Service.
2.2. General product information
Proper use
The si628 C is used for conductivity and temperature measurement in food industry, waste water, in the environment, etc.
2.3. Items supplied
si628 C (wall mount)
• Instrument (with strips for connections).
• User manual.
6
si628 C_HACH LANGE_Ed. 0510.indd ANG:6si628 C_HACH LANGE_Ed. 0510.indd ANG:6 20/5/10 11:08:4120/5/10 11:08:41
si628 C (panel mount)
• Instrument (with strips for connections).
• Clamps to fix to the panel.
• User manual.
3. Installation
3.1. si628 C (wall mount)
DANGER
Only qualified personnel should conduct the tasks described in this section of the manual.
3.1.1. Assembly Wall mounting
Open the case by unscrewing the 4 A screws.
25 mm
1 in
50 mm
2 in
A
50 mm
2 in
ENGLISH
115 mm
4.5 in
25 mm
1 in
Front view Side view Rear view
For wall mounting of the instruments, use the B drill holes indicated on the scheme.
Pipe or column mounting
Use the pipe adaptor, PN LZU9120.99. Install the pipe adaptor on the rear part of the instrument using both C drill holes as indicated on the figure. Fit the unit with two clamps to the column.
C
C
B
Cable tie
Rear view Pipe adaptor Instrument mounted
7
si628 C_HACH LANGE_Ed. 0510.indd ANG:7si628 C_HACH LANGE_Ed. 0510.indd ANG:7 20/5/10 11:08:4120/5/10 11:08:41
3. Installation
Details
To open the transparent front door, press with two fingers the part marked with (
) in the scheme.
Cable glands.
On the lower part of the instrument there are 5 cable glands, 2 PG 7 and 3 PG 9.
PG 7 (2 units)
PG 9 (3 units)
3.1.2. Connections
When making any wiring connections to the instrument, any warnings and notes found throughout the manual must be adhered to. For more safety information see “Safety information” on page 5.
DANGER
Always disconnect power to the instrument when making any electrical connections.
System functions
1, 2 Voltage 230 or 24 VAC ± 10%. (115 VAC under demand).
3 Ground connections, see
notes (
page 9). 4, 5 Current output 4-20 mA (EC). 6, 7 Input for temperature probe. 8 Cell cable shielding. 9 to 12 Input for 2-pole cell. 13, 14 Limit relay contacts. 15, 16 Alarm relay contacts.
DANGER
High-voltage wiring for the instrument is protected by a protective casing. The protective casing
Important
must remain in place unless a qualified installation technician is installing wiring for power.
8
si628 C_HACH LANGE_Ed. 0510.indd ANG:8si628 C_HACH LANGE_Ed. 0510.indd ANG:8 20/5/10 11:08:4220/5/10 11:08:42
3. Installation
Terminal assignment (interior panel)
Important notes:
• The si628 C does not require ground con­nection (terminals 3).
• Connect the ground only when there could
VCA
WHEN IT IS ON, IT INDICATES A FUSED FUSE.
4-20 mA
Pt 1000
CELL
L
A
be interference problems.
ENGLISH
DANGER
Relays:
- Limit the current up to 3 A by using one fuse.
- Do not connect cables with voltage higher than 24 V.
9
si628 C_HACH LANGE_Ed. 0510.indd ANG:9si628 C_HACH LANGE_Ed. 0510.indd ANG:9 20/5/10 11:08:4220/5/10 11:08:42
3. Installation
3.2. si628 C (panel mount)
DANGER
Only qualified personnel should conduct the tasks described in this section of the manual.
3.2.1. Assembly
For the si628 C mounting a drill hole is required according to the diagram.
92 mm
3.7 in
3.7 in
92 mm
Drill hole
Panel
After putting the instrument in the panel, it must be fitted using two clamps as indicated in the figure.
21 mm
0.8 in
80 mm
3.1 in
7 mm
0.3 in
Plug-in screw connector
10
si628 C_HACH LANGE_Ed. 0510.indd ANG:10si628 C_HACH LANGE_Ed. 0510.indd ANG:10 20/5/10 11:08:4320/5/10 11:08:43
Side view
Clamp to fix to the panel
3. Installation
When making any wiring connections to the instrument, any warnings and notes found throughout the manual must be adhered to. For more safety information see “Safety information” on page 5.
DANGER
Always disconnect power to the instrument when making any electrical connections.
3.2.2. Connections System functions
1 Ground connection, see Important
notes (page 12).
2, 3 Power supply 230 or 24 VAC ± 10%. (115 VAC under demand). 4, 5 Alarm relay contacts. 6, 7 Limit relay contacts. 8, 9 Current output 4-20 mA (EC). 10 to 13 Input for 2-pole cell. 14 Cell cable shielding. 15, 16 Input for temperature probe.
DANGER
High-voltage wiring for the instrument is protected by a protective casing. The protective casing must remain in place unless a qualified installation technician is installing wiring for power.
ENGLISH
11
si628 C_HACH LANGE_Ed. 0510.indd ANG:11si628 C_HACH LANGE_Ed. 0510.indd ANG:11 20/5/10 11:08:4320/5/10 11:08:43
3. Installation
Terminal assignment (rear panel)
Notas importantes:
• El si628 C
does not require ground connec-
tion (terminal 1).
• Connect the ground only when there could
CRISON INSTRUMENTS, S.A.
E-08328 ALELLA - Barcelona
Made in Spain
CELL4-20 mA
Transmitter si628 C
S/N:
IP 54
be interference problems.
DANGER
RED
VCA
L 1 L 2
Relays:
- Limit the current up to 3 A by using
one fuse.
- Do not connect cables with voltage higher than 24 V.
3.3. Cell cables
Compensation of the cable length
To minimize the effect of the cable resistance in a conductivity measurement, the cable length is compensa­ted by the 4-wire system. With the Pt 1000 temperature sensor, the cable resistance has little importance. However, in order to reduce the resistance of the conductor to a half, 2 cables have been connected to the ends of the sensor.
3.3.1. Cells with MP-5 connector
MP-5 connector, 5 contacts
TERMINALS
si628 C
(wall mount)
616 White 616 Brown 715 Pink 7 15 Yellow 814 Shield
913 Green 10 12 Blue 11 11 Gray 12 10 Red
si628 C
(panel mount)
COLOUR
12
si628 C_HACH LANGE_Ed. 0510.indd ANG:12si628 C_HACH LANGE_Ed. 0510.indd ANG:12 20/5/10 11:08:4320/5/10 11:08:43
3. Installation
3.3.2. Cells with connector S8
TERMINALS
si628 C
(wall mount)
9 13 Middle 10 12 Middle 11 11 Shield 12 10 Shield
si628 C
(panel mount)
COLOUR
ENGLISH
Important note:
Fit a jumper between terminals 9 and 10 (12 and 13 in panel mount instrument) and another between the terminals 11 and 12 (10 and 11 in panel mount instrument).
3.4. Protective wiring of switching contacts
Relay contacts are subjected to electrical erosion. Especially with inductive and capacitive loads, the service life of the contacts will be reduced. For suppression of sparks and arcing, components such as RC combinations, nonlinear resistors, series resistors and diodes should be used.
Typical protective wirings
AC applications with inductive load
DC applications with inductive load
AC/DC with capacitive load
RC
Load
RC
Contact relay
Typical RC at 230 VAC: Capacitor 0.1 µF/630 V Resistor 100 Ω/1 W
Diode 1N4007
Maximum relays load: AC < 24 V / < 3 A / < 72 VA.
si628 C_HACH LANGE_Ed. 0510.indd ANG:13si628 C_HACH LANGE_Ed. 0510.indd ANG:13 20/5/10 11:08:4320/5/10 11:08:43
Load
R
R, for example:
5 Ω / 1 W at 24 V / 0.4 A
13
4. Operation
4.1. Description
Display
si628 C have an alphanumeric display that explains itself. One of the most significant screens is des­cribed below as an example:
Measured value.
Measuring units.
Keys
MEASURING 23.2 °C 538 µS/cm @25 °C
ENTER.
• Horizontal movements to move forward in the “diagram”.
• Acceptance of numerical values. ESCAPE.
• Horizontal movements to move back in the “diagram”.
• Vertical movements to move back in the “diagram”.
• To erase numerical values.
• Modification of numerical values.
• Cursor “>” movements on the screen.
Sample temperature Resolution: 1 °C manual selection.
0.1 °C measuring with Pt 1000 Reference temperature
4.2. Setting up
Connect the si628 C to mains. At power up the first time only the following screens will appear:
si628_C
V1.0
LANGUAGE >Español
Italiano
The instrument is already measuring conductivity, with the following configuration:
• Temperature coefficient, CT, fixed at 2% / °C.
• Cell constant C=1 cm
• Reference temperature, RT, fixed at 25°C.
• Limit relay is activated at 1000 µS/cm, and desactivated at 800 µS/cm.
14
si628 C_HACH LANGE_Ed. 0510.indd ANG:14si628 C_HACH LANGE_Ed. 0510.indd ANG:14 20/5/10 11:08:4420/5/10 11:08:44
Possibility to select other language with keys.
Français
English
Deutsch
Português
-1
.
MEASURE S/cm
• Alarm relay is activated after 1minute with the reading out of limit.
• Output current 4 mA = 200 µS/cm, 20 mA = 2000 µS/cm.
To modify the default configuration (except from the temperature Coefficient and the reference tem­perature) see page 16.
MEASURING 22.3°C
µS/cm
545
@25°
4. Operation
4.3. Quick guide
This page shows the instrument’s operation by blocks. It is intended to provide a quick idea about all the instrument’s capabilities and how to gain access to them
MEASURING 22.3°C
545
µS/cm
@25°
MEASURE S/cm
ENGLISH
Security access code.
It limits the access to calibration or programming of the instrument
CÓDIGO 000
Vary the underlined digit with
and accept with .
Authorised code = 100
by non authorised people. When the “Code” screen has been shown for more than 30 sec. the instrument will automatically return to MEASURE.
Instrument in hold state. When being in the grey zone, the
instrument is in hold state:
• Contact relays are opened.
• According to the configuration, the current output is frozen at 21 mA or at the mA corresponding to the last measured value.
To quit the hold state it is necessary to return to MEASURE. The current output 4 to 20 will activate in 10 sec.
CALIBRATE
CELL
LIMIT
ALARM
OUTPUT 4-20 MA
HOLD
LANGUAGE
See “calibration... “
( page 17)
See “programming...”
( page 16)
15
si628 C_HACH LANGE_Ed. 0510.indd ANG:15si628 C_HACH LANGE_Ed. 0510.indd ANG:15 20/5/10 11:08:4520/5/10 11:08:45
4. Operation
4.4. Programming
The instrument is supplied with the default configuration specified in Setting up, page 14. To modify the default configuration and adapt it to specific applications follow the diagram below.
MEASURING 22.3°C
µS/cm
545
@25°
CODE 000
Enter code 100.
Select the nominal value of cell constant to be used. If the exact value is known, it can be directly entered, avoiding the calibration procedure.
Selection of the functioning values of the corresponding relays. Input “ON” (closed contact). Output “OFF” (open contact). Selectable values between
0.0 µS and 200 mS.
Select the time before the alarm activation. Selectable between 1 to 99 min or hours. 0= deactivated.
Select the conductivity values corresponding to the Current Output (4 mA and 20 mA). Selectable values between
0.0 µS and 200 mS.
CALIBRATE
CELL
LIMIT (*)
ALARM
OUTPUT 4-20 mA
Note: When programming, the instrument is in hold state:
• The contact relays are open.
• The current output is frozen at 21 mA. To quit the hold state it is necessary to return to MEASURE.
CONST:
1.000 cm
Vary the underlined digit with
and accept with
–1
ON - - - - - - OFF
1000 µS 800 µS
Vary the underlined digit with
and accept with
ALARM: 01 min
4 mA 20 mA
pH: 0.0 2.0 mS
Vary with
STORING DATA
Selection of the current output when the instrument is in Hold state: fixed at 21 mA or at the mA corresponding to the last measured value.
Selection of the working language: English, German, Spanish, French, Italian, Portuguese.
(*) LIMIT and OUTPUT 4-20 mA: Allows the introduction of five digits, placing the decimal point you want.
16
si628 C_HACH LANGE_Ed. 0510.indd ANG:16si628 C_HACH LANGE_Ed. 0510.indd ANG:16 20/5/10 11:08:4520/5/10 11:08:45
HOLD
LANGUAGE
MEASURE S/cm
> 21 mA last
Vary with
LANGUAGE > Español
Italiano
MEASURING 22.3°C
µS/cm
545
@25°
4. Operation
4.5. Calibration
With the calibration the instrument adapts to the cell
MEASURING 22.3°C
545
and its variations with time. Therefore a periodical cali­bration is necessary. Calibration can be carried out with 1 or 2 standards, to be selected between: 147 µS/cm, 1413 µS/cm or
12.88 mS/cm. If it appears an anomaly during the
calibration process an error message will be displayed (see page 22). If the cell constant value is exactly known, it can be manually entered (see page 16) avoiding the cali­bration procedure.
starts calibration with the standard 147 µS/cm. selects the next standard, 1413 µS/cm, to perform
calibration.
Theoretical value of the standard at the calibration temperature.
starts calibration with standard 1413 µS/cm. selects the next standard, 12.88 mS/cm, to perform
calibration.
23.2°C Standard 147 µS/cm
Standard 1413 µS/cm
µS/cm
@25°
CODE 000
CALIBRATE
CAL. 23.2°C
142.6 µS/cm
23.2°C
CAL. 23.2°C
1364 µS/cm
ENGLISH
Introduce the code
100.
If you have a cell without CAT, manually enter the temperature value.
Once the reading is stable, the display changes automatically.
Calibration can be carried out with 1 or 2 standards. If the calibration is with only one standard, press
P1, first standard used. C1, cell constant. If the calibration was with 2 standards, press
P2 and C2.
.
to display
23.2°C Standard 12.88 mS/cm
P1=147,0 µS/cm
C
=0,393 cm
1
-1
Press to perform calibration with this standard.
P2=1413,0 µS/cm
C
=0,397 cm
Note: In calibration the instrument is in Hold state:
• Contact relays are opened.
• Current output is fixed at 21 mA. To quit the Hold estate it is necessary to return to MEASURE.
si628 C_HACH LANGE_Ed. 0510.indd ANG:17si628 C_HACH LANGE_Ed. 0510.indd ANG:17 20/5/10 11:08:4620/5/10 11:08:46
2
to MEASURE
-1
17
4. Operation
4.5.1. Recognized standards
Table of values according to temperature.
Values stored in the si628 C.
°C µS/cm µS/cm mS/cm
15.0 119 1147 10.48
16.0 122 1173 10.72
17.0 125 1199 10.95
18.0 127 1225 11.19
19.0 130 1251 11.43
20.0 133 1278 11.67
21.0 136 1305 11.91
22.0 139 1332 12.15
23.0 142 1359 12.39
24.0 145 1386 12.64
25.0 147 1413 12.88
26.0 150 1440 13.13
27.0 153 1467 13.37
28.0 156 1494 13.62
29.0 159 1522 13.87
30.0 162 1549 14.12
18
si628 C_HACH LANGE_Ed. 0510.indd ANG:18si628 C_HACH LANGE_Ed. 0510.indd ANG:18 20/5/10 11:08:4720/5/10 11:08:47
4. Operation
4.6. Measure
Conductivity and temperature measurement
The value of the specific conductivity, the sample temperature in °C and the reference temperature appear simultaneously on the display. If the Pt 1000 temperature probe is not connected it is necessary to inform the instrument about the temperature of the solution using the arrow keys Conductivity cells incorporate a Pt 1000 temperature sensor.
Conductivity measurement
... with Pt 1000 probe ... without Pt 1000 probe
.
ENGLISH
MEASURING 22.3°C
µS/cm
545
Temperature
automatically
@25°
measured
MEASURING 22.3°C
µS/cm
545
@25°
CODE 000
MEASURING 25°C
µS/cm
545
temperature
@25°
Manually
entered
MEASURING 22°C
µS/cm
565
@25°
19
si628 C_HACH LANGE_Ed. 0510.indd ANG:19si628 C_HACH LANGE_Ed. 0510.indd ANG:19 20/5/10 11:08:4720/5/10 11:08:47
4. Operation
4.7. Access codes
The code 100 is for access to the calibration and programming of the instrument. There are codes with other functions:
Code Action
100 Main Menu.
401 Modification of the relay lag of the limits relays activation and deactivation.
Standard value: 2s. Programmable values: 0 to 99 s
413 Modification of the time for alarm activation when there is out of range reading.
Standard value: 60 s. Programmable values: 0 to 99 s/min/h
473 Modification of the delay time activation of the analogical output 4 to 20 mA after quitting Hold.
Standard value: 10 s. Programmable values: 0 to 99 s
497 Possibility to adjust the temperature for minimizing the effect of long cables.
20
si628 C_HACH LANGE_Ed. 0510.indd ANG:20si628 C_HACH LANGE_Ed. 0510.indd ANG:20 20/5/10 11:08:4720/5/10 11:08:47
5. Maintenance
5.1. Clean the measuring instrument and accessories
Important note: Never use cleaning such as turpentine, acetone or similar products to clean the instrument including the display and accessories. Only clean the housing and accessories using a soft, damp cloth. Mild soap solution may also be used. Dry the cleaned parts carefully with a soft cotton cloth.
CAUTION
When using chemicals or solvents, comply with the instructions of the producer and the general safety rules.
5.2. Fuse replacement (only in si628 C wall mount)
DANGER
Always disconnect power to the instrument when replacing the fuse
Unscrew for fuse replacement.
ENGLISH
VCA
WHEN IT IS ON, IT INDICATES A FUSED FUSE.
4-20 mA
Pt 1000
CELL
L
A
21
si628 C_HACH LANGE_Ed. 0510.indd ANG:21si628 C_HACH LANGE_Ed. 0510.indd ANG:21 20/5/10 11:08:4720/5/10 11:08:47
6. Error messages
During calibration
On display Problem Possible causes
Error C
Unstable reading
> 30 %
Deviation of the cell constant value higher than 30 % of its nominal value.
Unstable reading for more than 60 s.
Cell “worn out”. Error in the selection of the cell. Verify programming.
Defective or dirty cell. Standard from the refrigerator.
more than 1 min.
Same standard was used for both calibration points. Press ESC to quit calibration.
The standard used is not among the 3 in memory: 147 µS/cm, 1413 µS/cm or 12.88 mS/cm. at 25 °C. Press ESC to quit calibration.
SAME STANDARD
UNKOWN
STANDARD
°C OUT OF RANGE
1st and 2nd buffer have the same nominal value.
Incorrect buffer solution.
Standards temperature. Standard temperature > 35 °C or < 10 °C.
Other messages
On display Problem Possible causes
OUT OF R. 22.3°C
Measured value out of range.
- - - - -- mS/cm @25°
Conductivity reading > 200 mS/cm. After 1 minute, the current output will remain fixed at 22 mA.
°C OUT OF RANGE
Measured temperature value out of limits.
13.37 mS/cm @25°
LIMIT 22.3 °C 835 µS/cm @25°
ALARM 22.3°C
Measured value out of the programmed limits L1 and/or L2.
Activation of the alarm relay
835 µS/cm @25°C
22
si628 C_HACH LANGE_Ed. 0510.indd ANG:22si628 C_HACH LANGE_Ed. 0510.indd ANG:22 20/5/10 11:08:4820/5/10 11:08:48
Pt 1000 probe broken or temp. < –20 or >150°C. After 1 minute the current output will remain fixed at 22 mA.
Verify the programmed limit values. Verify process evolution.
The conductivity value is off limits for longer than the programmed delay of alarm relay activation. The current output will be set at 22 mA, but the limit will continue to relay on.
7. Replacement parts
P/N Description
LZU1001.99 0.1 A, fuse for si628 C 230 and 115 VCA (only wall mount). LZU1002.99 0.5 A, fuse for si628 C 24 VCA (only wall mount). LZU1004.99 Clamps to fix si628 C (only panel mount). LZU9120.99 Pipe adaptor for si628 C (only wall mount). LZU1003.99 Coaxial cable with AS9 connector, 3 m. LZU1005.99 Coaxial cable with AS9 connector, 5 m. LZU1010.99 Coaxial cable with AS9 connector, 10 m. LZU1015.99 Coaxial cable with AS9 connector, 15 m. LZU1020.99 Coaxial cable with AS9 connector, 20 m. LZU1025.99 Coaxial cable with AS9 connector, 25 m. LZU9044.99 LZU9045.99 LZU9046.99 LZU9047.99 LZU9251.99
LZU1199.99
LZW9700.99 Conductivity standard solution 147 µS/cm, with certificate of analysis, 250 ml flask. LZW9710.99 Conductivity standard solution 1413 µS/cm, with certificate of analysis, 250 ml flask. LZW9720.99 Conductivity standard solution 12.88 mS/cm, with certificate of analysis, 250 ml flask. LZW9730.99 Conductivity standard solution 80.4 mS/cm, with certificate of analysis, 250 ml flask. LZW9731.99 Conductivity standard solution 80.4 mS/cm, with certificate of analysis, 1000 ml flask. LZW9740.99 Conductivity standard solution 111.8 mS/cm, with certificate of analysis, 250 ml flask. LZW9741.99 Conductivity standard solution 111.8 mS/cm, with certificate of analysis, 1000 ml flask.
For cells please ask for specific brochure.
Multiple cable with MP-5 connector, 3 m. Multiple cable with MP-5 connector, 5 m. Multiple cable with MP-5 connector, 10 m. Multiple cable with MP-5 connector, 15 m. Multiple cable with MP-5 connector, 20 m.
MP-5 connector, for multiple cables connectable to conductivity cells, temperature compensators and chlorine galvanic cells.
ENGLISH
23
si628 C_HACH LANGE_Ed. 0510.indd ANG:23si628 C_HACH LANGE_Ed. 0510.indd ANG:23 20/5/10 11:08:4820/5/10 11:08:48
8. Warranty and liability
The manufacturer warrants that the product supplied is free of material and manufacturing defects and undertakes the obligation to repair or replace any defective parts at zero cost.
The warranty period for instruments is 24 months. If a service contract is taken out within 6 months of pur­chase, the warranty period is extended to 60 months.
With the exclusion of the further claims, the supplier is liable for defects including the lack of assured prop­erties as follows: all those parts that, within the warranty period calculated from the day of the transfer of risk, can be demonstrated to have become unusable or that can only be used with significant limitations due to a situation present prior to the transfer of risk, in particular due to incorrect design, poor materials or inadequate finish will be improved or replaced, at the supplier’s discretion. The identification of such defects must be notified to the supplier in writing without delay, however at the latest 7 days after the identification of the fault. If the customer fails to notify the supplier, the product is considered approved despite the defect. Further liability for any direct or indirect damages is not accepted.
If instrument-specific maintenance and servicing work defined by the supplier is to be performed within the warranty period by the customer (maintenance) or by the supplier (servicing) and these requirements are not met, claims for damages due to the failure to comply with the requirements are rendered void.
Any further claims, in particular claims for consequential damages cannot be made.
Consumables and damage caused by improper handling, poor installation or incorrect use are excluded from this clause.
The manufacturer process instruments are of proven reliability in many applications and are therefore often used in automatic control loops to provide the most economical possible operation of the related process.
To avoid or limit consequential damage, it is therefore recommended to design the control loop such that a malfunction in an instrument results in an automatic change over to the backup control system; this is the safest operating state for the environment and the process.
24
si628 C_HACH LANGE_Ed. 0510.indd ANG:24si628 C_HACH LANGE_Ed. 0510.indd ANG:24 20/5/10 11:08:4820/5/10 11:08:48
Inhalt
1. Technische Daten . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1. Abmessungen
2. Allgemeine Informationen . . . . . . . . . . . . 5
2.1. Sicherheitshinweise
2.1.1. Bedeutung von Gefahrenhinweisen
2.2. Allgemeine Produktinformationen
2.3. Inhalt
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
. . . . . . . . . . . . . . . . . . . . . . . . . 4
. . . . . . . . . . . . . . . . . . . . . 5
5
. . . . . . . . . 6
3. Montage und Installation
3.1. si628 C (Wandmontage)
3.1.1. Montage
3.1.2. Anschlüsse
3.2. si628 C (Schalttafelmontage)
3.2.1. Montage
3.2.2. Anschlüsse
3.3. Zellenkabel
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
. . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . . . . . . . . . 8
. . . . . . . . 10
. . . . . . . . . . . . . . . . . . . . . . . . 10
. . . . . . . . . . . . . . . . . . . . . . . 11
3.4. Einbau von Schutzvorrichtungen für die Relaiskontakte
4. Bedienung und Funktion
4.1. Beschreibung
4.2. Inbetriebnahme
4.3. Kurzanleitung
4.4. Programmierung
4.5. Kalibrierung
4.5.1. Geeignete Standards
4.6. Messung
4.7. Zugangscodes
5. Wartung
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.1. Reinigung des Geräts und des Zubehörs
. . . . . . . . . . . . . . . . . . . . . . . . . . 13
. . . . . . . . . . . . . . . . . . 14
. . . . . . . . . . . . . . . . . . . . . . . . . . 14
. . . . . . . . . . . . . . . . . . . . . . . . 14
. . . . . . . . . . . . . . . . . . . . . . . . . . 15
. . . . . . . . . . . . . . . . . . . . . . . . 16
. . . . . . . . . . . . . . . . . . . . . . . . . . . 17
. . . . . . . . . . . . . 18
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
. . . . . . . . . . . . . . . . . . . . . . . . . 20
. . 21
5.2. Austausch der Sicherung . . . . . . . . . . . . . . . . 21
DEUTSCH
6. Fehlermeldungen
7. Ersatzteile
8. Gewährleistung und Haftung
. . . . . . . . . . . . . . . . . . . . . . . . . 22
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
. . . . . . . . . . . . 24
Contact information
1
si628 C_HACH LANGE_Ed. 0510.indd ALE:1si628 C_HACH LANGE_Ed. 0510.indd ALE:1 20/5/10 11:08:4820/5/10 11:08:48
si628 C_HACH LANGE_Ed. 0510.indd ALE:2si628 C_HACH LANGE_Ed. 0510.indd ALE:2 20/5/10 11:08:4820/5/10 11:08:48
Loading...
+ 97 hidden pages