4.3 Menu structure.................................................................................................................................. 21
6.3 Wear parts ........................................................................................................................................ 37
Section7 Replacement parts and accessories ...................................................................................... 39
Pollution Degree
Protection Class
Installation Category
Maximum Altitude
2
1
II
2000 m (6.562 ft.)
Protection classIP20
InstallationDIN rail EN 50022 35 × 15.0
1
Pentium is a registered trademark of the Intel Corporation.
2
Microsoft Windows is a brand name for operating systems of the Microsoft Corporation.
8
Section2General information
2.1Safety information
Please read the entire manual carefully before unpacking, assembling or operating the
instrument. Pay attention to all hazard and warning notices. Failure to do so could result in
serious injury to the operator or damage to the instrument.
To prevent damage to or impairment of the device's protection equipment, the device may
only be used or installed as described in this manual.
2.1.1Hazard notices in this manual
Indicates a potentially or imminently hazardous situation that, if not avoided, can result in death or
serious injury.
Indicates a potentially or imminently dangerous situation that, if it is not avoided, can lead to
death or to serious injuries.
Indicates a possible dangerous situation that can have minor or moderate injuries as the result.
DANGER
WARNING
CAUTION
Indicates a situation that, if it is not avoided, can lead to damage to the device. Information that
requires special emphasis.
Note: Information that supplements points in the main text.
2.1.2Warning labels
Observe all labels and tags attached to the instrument. Failure to do so may result in
personal injury or damage to the instrument.
This symbol may be attached to the device and refers to operation and/or safety notes in the user manual.
This symbol may be found on an enclosure or barrier within the product and indicates a risk of electric shock
and/or death by electrocution.
Electrical equipment marked with this symbol may no longer be disposed of in unsorted domestic or industrial
waste in Europe after August 12, 2005. In conformity with the provisions in force (EU Directive 2002/96/EC),
consumers in the EU must return old electrical devices to the manufacturer for disposal from this date, at no
charge to the consumer.
Note: You obtain instructions on the correct disposal of all (marked and not marked) electrical products that
were supplied or manufactured by Hach-Lange at your relevant Hach-Lange sales office.
NOTICE
9
General information
2.2Areas of application
The RTC113 ST-Module (Real Time Controller, Sludge Thickening-Module) is an open
and closed-loop control unit for universal applications. It can be used by mechanical
sludge thickening devices, for example belt thickeners or drum thickeners in wastewater
treatment plants.
The RTC113 ST-Module
•Optimizes polymer consumption and
•Uniformly manages the concentration of solids in the thick sludge
1-channelOpen/closed-loop controller for one thickener
2-channelOpen/closed-loop controller for two thickeners
The use of an RTC Module does not release the operator from the duty of care to the system. No
guarantees as to the functionality or operational safety of the system.
In particular, the operator must make sure that instruments connected to the RTC
open/closed-loop controller are always fully functional.
To make sure these instruments supply correct, reliable measurement values, regular
maintenance work (for example, cleaning of the sensor and laboratory comparative
measurements) is essential! (Refer to the user manual for the relevant instrument.)
Table1 Versions of the RTC113 ST-Module
NOTICE
2.3Scope of delivery
The combination of pre-assembled components supplied by the manufacturer does not represent
a standalone functional unit. In accordance with EU guidelines, this combination of
pre-assembled components is not supplied with a CE mark, and there is no EU declaration of
conformity for the combination.
However, the conformity of the combination of components with the guidelines can be proved
through technical measurements.
Each RTC Module is supplied with:
•SUB-D connector (9 pin)
•user manual
•ferrite core
Check that the order is complete. All listed components must be present. If anything is
missing or damaged, contact the manufacturer or distributor immediately.
NOTICE
10
2.4Instrument overview
Figure1 Base module RTC 24 V version .
General information
1PE (protective earth)5sc 1000 connection: RS485 (CX1010-N031)
224 V6Battery compartment
30 V7CPU base module, consisting of Ethernet port with
battery compartment (CX1010-N000), CPU module with
CF card (CX1010-0021) and passive aeration element.
4Automatic circuit breaker (ON/OFF switch for item 7 and
8 without fuse function).
Note: All components are pre-wired.
8Power supply module, consisting of bus coupler
(CX1 100-0002) and terminal module 24V.
11
General information
Abbildung 2 Base module RTC 100-240 V version
1L(+)7Automatic circuit breaker (ON/OFF switch for item 10
and 11 without fuse function).
2N(–)8sc 1000 connection: RS485 (CX1010-N041)
3Input AC 100–240 V / Input DC 95 V–250 V9Battery compartment
4PE (protective earth)10 CPU base module, consisting of Ethernet port with
battery compartment (CX1010-N000), CPU module with
CF card (CX1010-0021) and passive aeration element.
524 V transformer (Specification section 3.1.1, page 17)11 Power supply module, consisting of bus coupler
6Output DC 24 V, 0,75 A
(CX1100-0002) and terminal module 24V.
Note: All components are pre-wired.
12
Figure3 Design of the analog and digital input and output modules
2
1
General information
1Input- or Output- Module or Bus Termination Module
analog or digital
Note: The number of LEDs indicates the number of channels.
2.5Theory of operation
2.5.1Theory of operation of the RTC Module
The RTC113 RTC Module outputs analog (0/4–20 mA) and digital (0/24 V) signals for the
polymer dosing rate or the feed flow rate of mechanical sludge thickening devices. Digital
fieldbus signals from sc1000 communication cards can also be used.
2.5.2Input signals
The most important input signals are:
•Sludge influent TSS concentration (concentration of solids)
•Feed flow rate of the machine-controlled sludge thickening
2.5.3Parameters for configuration
•Thickened sludge TSS concentration (optional)
•Status of the thickened sludge pump (on/off)
The most important parameters for configuration are:
2LED area with installed LEDs or free LED installation
spaces
•The required specific polymer dosing [g polymer/kg TSS]
•The target TSS concentration in thickened sludge
Note: In a closed-loop circuit, TSS measurement is required in thickened sludge.
13
General information
2.5.4Operating modes
The RTC113 ST-Module can be operated as a combined open-loop and closed-loop
controller. Several variants can be configured.
1. Configuration of a specified polymer rate [L/h] with a specified feed flow rate [m
2. Configuration of a specified specific polymer dosing rate [g polymer/kg TSS]. One of
the following settings is adjusted:
a. The polymer flow rate according to the TSS concentration and the feed flow rate
•Based on the actual feed flow rate [L/h] and TSS concentration [g/L] in the feed
Or:
b. The feed flow rate according to the specified polymer dosing rate and the
3
/h].
(Figure4).
flow, the polymer dosing rate [L/h] is calculated for the required specific dosing
rate.
measured TSS concentration of the influent (Figure5).
•Based on the measurement value of the TSS concentration from the influent [g/L]
and the configurable specified polymer dosing rate [L/h], the feed flow rate [m
is calculated such that it corresponds to the pre-defined specific polymer dosing
rate [g/kg].
3. Both variants 2a and 2b can be combined with the closed-loop controller described
below:
a. Closed-loop control of the TSS concentration in the thickened sludge
•The specific polymer dosing rate is adjusted according to the difference between
the target and actual TSS concentration in the outlet.
3
/h]
14
Figure4 Adjustment of the polymer dosing rate to the feed flow rate
General information
1Static thickener8Pump for open-loop control of the polymer dosing rate
2Measurement of the feed flow rate 9Pump for the feed flow rate
3Open-loop control of the polymer dosing rate (feed flow
rate measurement value)
4TSS measurement from the influent 11 TSS measurement in the thickened sludge outlet
5Open-loop control of the polymer dosing rate (influent
Figure5 Adjustment of the feed flow rate to the specified polymer dosing rate
1Static thickener8Pump for polymer dosing: constant here
2Influent9Mechanical sludge thickener
3TSS measurement from the influent10 TSS measurement in the thickened sludge outlet
4Open-loop control for the feed flow rate11 Closed-loop control for the feed flow rate
5RTC113 ST-Module12 Thickened sludge pump
6Pump for open-loop control of the feed flow rate 13 Digester
7Polymer supply
16
Section3Installation
Only qualified experts may perform the tasks described in this section of the manual, while
adhering to all locally valid safety regulations.
Always lay cables and hoses so that they are straight and do not pose a tripping hazard.
Before the power supply is switched on, refer to the instructions in the relevant manuals.
3.1Installation of the RTC Module
Only install the RTC Module on a DIN rail. The module must be attached horizontally, with
at least 30 mm (1.2 in.) space at the top and bottom to make sure that the passive
aeration element can function correctly.
When used indoors, the RTC Module must be installed in a control cabinet. When used
outdoors, the RTC Module requires a suitable enclosure that follows the technical
specifications.
DANGER
CAUTION
CAUTION
The RTC Module is only operated via the sc1000 controller (see the user manual for the
sc1000 controller).
Note: The software version of the sc1000 controller must be V3.14 or above.
3.1.1Supply voltage of the RTC Module
Table2 Supply voltage of the RTC Module
Voltage24 V DC (–15 % / +20 %), max. 25 W
Recommended fuse C2
With 110–230 V option110–230 VAC, 50-60 Hz, approximately 25 VA
Note: An external deactivation switch is recommended for all installations.
3.2Connection of process measurement instruments for the TSS concentration
of solids
The measurement signals of the sc sensors for the measurement of the concentration of
solids (e. g. SOLITAX sc) are provided to the RTC113 ST-Module via the RTC
communication card (YAB117) in the sc1000.
3.2.1Power supply of the sc sensors and the sc1000 controller
See operating instructions of the respective sc sensors and the sc1000 controller.
3.3sc1000 controller connection
Connect the SUB-D plug supplied to a dual-core, sheathed data cable (signal or bus
cable). For additional information regarding the data cable connection, refer to the
enclosed assembly instructions.
17
Installation
3.4Connection to the automation unit on the plant side
The one-channel and two-channel versions of the ST-Module are equipped with various
modules that must be connected to the plant automation system.
•The feed flow rate must be provided to the ST-Module as a 0/4 to 20 mA signal.
•The polymer flow rate must be provided (on both versions) to the ST-Module as a
0/4 to 20 mA signal.
•The status signal of the thickened sludge pump (on/off) must be a digital input signal
(24 V/0 V).
•The polymer pump can be operated in pulse/pause mode (PWM).
•The status signals and fault indications are output as 0 V/24 V signals.
•Measurement errors are shown 5 minutes after the error occurs. In the event of a new
startup (return of power supply), the unit is set back to ON (24 V) after approximately
1 minute and 40 seconds if there are no measurement errors.
•In the event of a new startup (return of power supply), the RTC operating signal is set
back to ON (24 V) after approximately 1 minute and 25 seconds.
Table3 Connections for the 1-channel RTC113 ST-Module
ModuleNameConnection SignalFunction
2x digital inputKL1002
4x digital output
2x analog outputKL4012
1x analog inputKL30111(+) - 2(-)0/4 to 20 mAInput of the feed flow rate
1x analog inputKL30111(+) - 2(-)0/4 to 20 mAInput of the polymer flow rate
Bus terminationKL9010Bus termination
1
KL2134
1Input + 24 V/0 VStatus of thickened sludge pump (on/off)
2Source + 24 V24 V for relay
1+24 V/0 VPolymer pump on/off (24 V/0 V)
5+24V/0V
4+24 V/0 VInput signals OK (24 V), input signal faulty (0 V)
8+24 V/0 VRTC operational (24 V), RTC faulty (0 V)
1(+) - 3(-)0/4 to 20 mAOutput of the polymer pump flow rate
5(+) - 7(-)0/4 to 20 mAOutput of the feed flow rate
Closed-loop control of the feed flow rate
active/inactive (24 V/0 V)
1
Ground to connection 3 and 7 or to the supply voltage connections
Table4 Connections for the 2-channel RTC113 ST-Module
ModuleNameConnection SignalChannel Function
1Input + 24 V/0 V1Status of thickened sludge pump (on/off)
4x digital inputKL1002
2Source + 24 V124 V for relay
5Input + 24 V/0 V2Status of thickened sludge pump (on/off)
6Source + 24 V224 V for relay
18
Table4 Connections for the 2-channel RTC113 ST-Module
ModuleNameConnection SignalChannel Function
1+24 V/0 V1Polymer pump on/off (24 V/0 V)
5+24V/0V1
2+24 V/0 V1Input signals OK (24 V), input signal faulty (0 V)
8x digital
1
output
2x analog
output
2x analog
output
1x analog input KL30111(+) - 2(-)0/4 to 20 mA1Input of the feed flow rate
1x analog input KL30111(+) - 2(-)0/4 to 20 mA1Input of the polymer flow rate
1x analog input KL30111(+) - 2(-)0/4 to 20 mA2Input of the feed flow rate
1x analog input KL30111(+) - 2(-)0/4 to 20 mA2Input of the polymer flow rate
Bus termination KL9010Bus termination
4+24 V/0 V2Input signals OK (24 V), input signal faulty (0 V)
8+24 V/0 V2RTC operational (24 V), RTC faulty (0 V)
1(+) - 3(-)0/4 to 20 mA1Output of the polymer pump flow rate
5(+) - 7(-)0/4 to 20 mA1Output of the feed flow rate
1(+) - 3(-)0/4 to 20 mA2Output of the polymer pump flow rate
5(+) - 7(-)0/4 to 20 mA2Output of the feed flow rate
Closed-loop control of the feed flow rate
active/inactive (24 V/0 V)
Closed-loop control of the feed flow rate
active/inactive (24 V/0 V)
Installation
1
Ground to connection 3 and 7 or to the supply voltage connections
19
Installation
20
Section4Parameterization and operation
4.1Operating the sc controller
The RTC Module can only be operated via the sc1000 controller
in conjunction with the RTC communication card. Before the RTC
Module is used, the user must be familiar with the functionality of
the sc1000 controller. Learn how to navigate through the menu
and perform the relevant functions.
4.2sc1000 setup
1. Open the MAIN MENU.
2. Select
3. Select
4. Select
RTC MODULES / PROGNOSYS and confirm.
RTC MODULES and confirm.
RTC and confirm.
Possible warning messages:
MODBUS ADDRESS, PROBE SERVICE
Note: Refer to Section6 Troubleshooting, page37 for a list of all
possible error and warning messages together with a description of all
necessary countermeasures to be taken.
4.4Configuration of RTC113 ST-Module parameters on the sc1000 controller
The following menu items are in the SC1000 SETUP menu.
4.4.1RTC113 ST-Module open and closed-loop controller
RTC MODULES / PROGNOSYS
RTC MODULES
RTC
CONFIGURE
SELECT SENSOR
Select the sensors installed for the open/closed-loop controller
(refer to section4.5, page26).
21
Parameterization and operation
4.4.1RTC113 ST-Module open and closed-loop controller (Continued)
RTC MODULES / PROGNOSYS
RTC MODULES
RTC
PRESELECT PROG.
CHANNEL 1
Based on the feed flow rate [m³/h] and measured TSS
POLYMER DOSING
CONTROL
FEED FLOW CONTROL
CLOSED LOOP EFFLUENT
CONTROL
CHANNEL 2As per channel one
concentration [g/L] from the influent, the polymer dosing rate
[L/h] is calculated such that it corresponds to the target specific
polymer dosing rate [g/kg].
Based on the measured TSS concentration [g/L] and a fixed
polymer dosing rate [L/h], the feed flow [m
that it corresponds to the specific polymer dosing rate [g/kg].
If activated, the specific polymer dosing rate FACTOR
POLYMER DOSING is adjusted based on the difference
between the target and actual TSS concentration in the
thickened sludge.
The change in the specific dosing rate affects the polymer
dosing rate [L/h] in the POLYMER DOSING CONTROL module
and affects the feed flow rate in the FEED FLOW CONTROL
module.
3
/h] is calculated such
Activation/
deactivation
Activation/
deactivation
Activation/
deactivation
22
Parameterization and operation
4.4.1RTC113 ST-Module open and closed-loop controller (Continued)
RTC MODULES / PROGNOSYS
RTC MODULES
RTC
CONTROL PARAMETER
CHANNEL 1
FACTOR POLYMER
DOSING
POLYMER CONCENTRATION
MANUAL POLYMER
DOSING
MANUAL FEED FLOW
MAX DECREASE CLOSED
L
MAX INCREASE CLOSED L
SET-POINT TSS
P GAIN TSS
INTEGRAL TIME TSS
DERIVATIVE TIME TSS
CHANNEL 2As per channel one
Required specific polymer dosing [g/kg]. This parameter
determines how many grams of polymer per kilogram of TSS
are fed by the machine.
Polymer concentration [g/L] fed via the polymer pump.g/L
The RTC outputs the polymer flow rate [L/h] if
•FEED FLOW CONTROL is activated
•No open-loop control mode (see above) is activated
•The TSS measurement from the influent reports an error, or
•The flow measurement from the influent reports an error.
The RTC outputs the feed flow rate [m³/h] if
•POLYMER DOSING CONTROL is activated
•No open-loop control mode (see above) is activated
•The TSS measurement from the influent reports an error, or
•The flow measurement from the influent reports an error
This value defines the maximum decrease of the specific
polymer dosing rate FACTOR POLYMER DOSING [g/kg] if
CLOSED LOOP EFFLUENT CONTROL is selected.
This value defines the maximum increase of the specific
polymer dosing rate FACTOR POLYMER DOSING [g/kg] if
CLOSED LOOP EFFLUENT CONTROL is selected.
Required setpoint of the TSS concentration in the thickened
sludge.
Note: This parameter is only considered if CLOSED LOOP
EFFLUENT CONTROL is activated.
Proportional gain for the PID closed-loop controller for the TSS
concentration in the thickened sludge.
Note: P GAIN TSS [L/g] is divided by 100 before it is multiplied
by the deviation of the actual TSS concentration from the
required TSS setpoint.
Integral time for the PID closed-loop controller for the TSS
concentration in the thickened sludge.
Note: INTEGRAL TIME TSS is set to "0" to deactivate the
integral part of the PI open-loop controller.
Derivative time for the PID closed-loop controller for the TSS
concentration in the thickened sludge.
g/kg
L/h
m³/h
g/kg
g/kg
g/L
L/g
min
min
23
Parameterization and operation
4.4.1RTC113 ST-Module open and closed-loop controller (Continued)
RTC MODULES / PROGNOSYS
RTC MODULES
RTC
INPUT/OUTPUT LIMITS
CHANNEL 1
3
FEED FLOW LOW
FEED FLOW HIGH
FEED FLOW SMOOTHING
LIMIT TSS IN LOW
LIMIT MAX TSS IN HIGH
TSS IN SMOOTHING
LIMIT TSS OUT LOW
LIMIT TSS OUT HIGH
TSS OUT SMOOTHING
Feed flow rate input signals below this value [m
this value (to avoid low flow peaks).
Feed flow rate input signals above this value [m
this value (to avoid high flow peaks).
Feed flow measurement values are smoothed in line with this
parameter.
TSS measurement values from the influent that are below this
value [g/L] are set to this value (to avoid low peaks).
TSS measurement values from the influent that are above this
value [g/L] are set to this value (to avoid high peaks).
The TSS measurement values from the influent are smoothed in
line with this parameter.
The TSS values of the thickened sludge that are below this
value [g/L] are set to this value (to avoid low peaks).
The TSS values of the thickened sludge that are above this
value [g/L] are set to this value (to avoid high peaks).
The TSS measurement values from the effluent are smoothed in
line with this parameter.
When FEED FLOW CONTROL is activated, measurement
POLYMER DOSING
MINIMUM
values for the polymer dosing rate that are below this value
[m³/h] are set to this value (to avoid low peaks in the dosing
flow).
Any RTC calculation above this value [g/L] is set to this value
and delivered to the polymer pump.
POLYMER DOSING
MAXIMUM
When FEED FLOW CONTROL is activated, measurement
values for the polymer dosing rate that are above this value
[m³/h] are set to this value (to avoid high peaks in the dosing
flow).
CHANNEL 2As per channel one
INPUTS
CHANNEL 1
MIN FEED FLOW
MAX FEED FLOW
0/4...20mA
MIN POLYMER FLOW
MAX POLYMER FLOW
0/4...20mA
Minimum flow rate [m³/h] from the influent in accordance with
the 0/4 mA measurement signal.
Maximum flow rate [m³/h] from the influent in accordance with
the 20 mA measurement signal.
Transfer range of 0/4 to 20 mA current loop (as set in connected
flow measuring instrument).
Minimum polymer dosing in [L/h] in accordance with the 0/4 mA
measurement signal.
Maximum polymer dosing in [L/h] in accordance with the 20 mA
measurement signal.
Transfer range of 0/4 to 20 mA current loop (as set on
connected flow measuring instrument).
CHANNEL 2As per channel one
/h] are set to
3
/h] are set to
3
m
3
m
min
g/L
g/L
min
g/L
g/L
min
L/h
L/h
m³/h
m³/h
L/h
L/h
/h
/h
24
Parameterization and operation
4.4.1RTC113 ST-Module open and closed-loop controller (Continued)
RTC MODULES / PROGNOSYS
RTC MODULES
RTC
OUTPUTS
CHANNEL 1
MIN FEED FLOWMinimum feed flow rate [m³/h] in accordance with 0/4 mA.m³/h
MAX FEED FLOWMaximum feed flow rate [m³/h] in accordance with 20 mA.m³/h
0/4...20mA
MIN POLYMER FLOW
MAX POLYMER FLOW
0/4...20mA
CONTROL CYCLE
MIN RUNTIME
CHANNEL 2As per channel one
MODBUS
ADDRESS
DATA ORDER
DATALOG INTRVLIndicates the interval in which the data is saved in the log file. [min]
SET DEFAULTSRestores the factory settings.
MAINTENANCE
RTC DATA
RTC MEASUREMEN
RTC ACTUAT VAR
DIAG/TEST
EEPROMHardware test
RTC COMM TOCommunication time-out
RTC CRCCommunication check sum
MODBUS ADDRESS
Transfer range of 0/4 to 20 mA current loop (as set on
connected flow measuring instrument).
Minimum polymer pump delivery rate in accordance with
0/4 mA.
Maximum polymer pump delivery rate in accordance with
20 mA.
Transfer range of 0/4 to 20 mA current loop (as set on
connected flow measuring instrument).
Pulse/pause mode for the polymer pump open-loop control for
dosing rates beneath the minimum polymer flow rate (MIN
POLYMER FLOW). The on/off duration in pulse/pause mode
can be affected by the duration of the CONTROL CYCLE. For
example, with a CONTROL CYCLE of 100 seconds and a
dosing control value of 60 %, the polymer pump is regularly
switched on for 60 seconds and switched off for 40 seconds.
Short cycle times increase the switching frequency but enable
more precise adaptation to individual requirements. CONTROL
CYCLE should be divisible by MIN RUNTIME and produce a
whole number.
The minimum ON time in pulse/pause dosing mode. The pump
is not activated for periods shorter than this. The MIN RUNTIME
must be shorter than the duration of the CONTROL CYCLE.
Start address of an RTC within the Modbus network. Default
setting: 41–61
Specifies the register order within a double word. Presetting:
NORMAL
Specifies the value measured by the RTC, e. g. the influent
measurement.
Specifies the variable calculated by the RTC, e. g. whether the
aeration should be switched on or off.
Here, the address is displayed where the communication
actually takes place. Presetting: 41
L/h
L/h
s
s
25
Parameterization and operation
4.5Select sensors
Figure6 Select sensor
1. To select sensors and their sequence for the RTC Module,
press RTC > CONFIGURE > SELECT SENSOR.
1ENTER — Saves the setting and returns to the
CONFIGURE menu.
2CANCEL — Returns to the CONFIGURE menu without
saving.
3ADD — Adds a new sensor to the selection.
2. Press
A selection list of all subscribers to the sc1000 network
opens.
3. Press the required sensor for the RTC Module and confirm
by pressing
Sensors in black type are available for the RTC Module.
Sensors in red type are not available for the RTC Module.
Note: Sensors marked (p) are available for PROGNOSYS if these
sensors have been selected in conjunction with an RTC (refer to the
PROGNOSYS user manual).
4DELETE — Removes a sensor from the selection.
5UP/DOWN — Moves the sensors up or down.
ADD (Figure6, item 3).
ENTER below the selection list.
26
Parameterization and operation
4. The selected sensor is shown in the sensor list.
Press
ADD (Figure6, item 3) to open the selection list again.
5. Select the second sensor for the RTC Module and confirm
by pressing
Note: Previously selected sensors are shown in gray.
The selected sensors are shown in the sensor list.
ENTER below the selection list.
6. To sort the sensors in the order specified for the RTC
Module, press the sensor and use the arrow keys to move it
(Figure6, item 5).
Press
DELETE (Figure6, item 4) to remove an incorrect
sensor from the sensor list again.
7. Press ENTER (Figure6, item 1) to confirm the list once it is
finished.
27
Parameterization and operation
4.6PRESELECT PROG
4.6.1POLYMER DOSING CONTROL
4.6.2FEED FLOW CONTROL
Based on the measured feed flow rate [m³/h] and the measured
TSS concentration [g/L] from the influent, the polymer dosing rate
[L/h] is calculated such that the setpoint corresponds to the
specific polymer dosing rate [g/kg].
Note: This open-loop control mode can only be activated if FEED FLOW
CONTROL is deactivated.
Note: The polymer flow rate is controlled via the RTC.
Based on the measured TSS concentration [g/L] and the
specified polymer dosing rate [L/h], the feed flow rate is
calculated such that it corresponds with the specific polymer
dosing rate [g/kg] (FACTOR POLYMER DOSING).
Note: This open-loop control mode can only be activated if POLYMER
DOSING CONTROL is deactivated.
Note: The feed flow rate is controlled via the RTC.
4.6.3CLOSED LOOP EFFLUENT CONTROL
If activated, the specific polymer dosing rate FACTOR POLYMER
DOSING is adjusted based on the difference between the target
and actual TSS concentration in the thickened sludge.
If FEED FLOW CONTROL is activated, the TSS load fed with the
sludge thickening is adjusted based on the difference between
the target and actual TSS concentration in the filtrate.
Note: This closed-loop control can only be activated if POLYMER
DOSING CONTROL (section4.6.1) or FEED FLOW CONTROL
(section4.6.2) is activated.
4.7CONTROL PARAMETER
4.7.1FACTOR POLYMER DOSING
Required specific polymer dosing [g/kg]. This parameter
determines how many grams of polymer per kilogram of TSS are
fed by the machine.
4.7.2POLYMER CONCENTRATION
28
Polymer concentration [g/L] fed via the polymer pump.
4.7.3MANUAL POLYMER DOSING
4.7.4MANUAL FEED FLOW
4.7.5MAX DECREASE CLOSED L
Parameterization and operation
The RTC outputs the polymer dosing rate [L/h] if
•FEED FLOW CONTROL is activated
•No open-loop control mode (section4.6.1 to section4.6.3) is
activated
•The TSS measurement from the influent reports an error, or
•The flow measurement from the influent reports an error.
The RTC outputs the feed flow rate [m3/h] if
•POLYMER DOSING CONTROL is activated
•No open-loop control mode (section4.6.1 to section4.6.3) is
activated
•The TSS measurement at the inlet reports an error, or
•The flow measurement from the influent reports an error.
4.7.6MAX INCREASE CLOSED L
4.7.7SET-POINT TSS
4.7.8P GAIN TSS
This value defines the maximum decrease of the specific polymer
dosing rate FACTOR POLYMER DOSING [g/kg] if CLOSED
LOOP EFFLUENT CONTROL is selected.
This value defines the maximum increase of the specific polymer
dosing rate FACTOR POLYMER DOSING [g/kg] if CLOSED
LOOP EFFLUENT CONTROL is selected.
Required setpoint of the TSS concentration in the thickened
sludge.
Note: · This parameter is only considered if CLOSED LOOP EFFLUENT
CONTROL (section4.6.3) is activated.
Proportional gain for the PID closed-loop controller for the TSS
concentration in the thickened sludge.
Note: P GAIN TSS [L/g] is divided by 100 before it is multiplied by the
deviation of the actual TSS concentration from the required TSS
setpoint.
4.7.9INTEGRAL TIME TSS
Integral time for the PID closed-loop controller for the TSS
concentration in the thickened sludge.
Note: INTEGRAL TIME TSS is set to "0" to deactivate the integral part of
the PI open-loop controller.
29
Parameterization and operation
4.7.10 DERIVATIVE TIME TSS
4.8INPUT/OUTPUT LIMITS
4.8.1FEED FLOW LOW
4.8.2FEED FLOW HIGH
4.8.3FEED FLOW SMOOTHING
Derivative time for the PID closed-loop controller for the TSS
concentration in the thickened sludge.
Feed flow rate input signals below this value [m3/h] are set to this
value. This means that very low feed flow rates can be avoided.
Feed flow rate input signals above this value [m3/h] are set to this
value. This avoids load peaks.
Feed flow measurement values are smoothed in line with this
parameter.
SMOOTHING = 1: The signal for the flow rate measurement
is not smoothed.
4.8.4LIMIT TSS IN LOW
4.8.5LIMIT MAX TSS IN HIGH
4.8.6TSS IN SMOOTHING
SMOOTHING = 2: Smoothing is performed over 3 minutes.
SMOOTHING = 3: Smoothing is performed over 5 minutes.
SMOOTHING = 5: Smoothing is performed over 12 minutes.
SMOOTHING = 10: Smoothing is performed over
25 minutes.
Example:
With the setting SMOOTHING = 2, it takes 3 minutes for the
smoothed value to reach 95 % of the final value after an abrupt
change in the feed flow rate.
TSS measurement values from the influent that are below this
value [g/L] are set to this value (to avoid low peaks).
Measurement values from the influent that are above this value
[g/L] are set to this value (to avoid high peaks).
TSS measurement values from the influent are smoothed in line
with this parameter.
30
SMOOTHING = 1: The signal is not smoothed.
SMOOTHING = 2: Smoothing is performed over 3 minutes.
SMOOTHING = 3: Smoothing is performed over 5 minutes.
SMOOTHING = 5: Smoothing is performed over 12 minutes.
SMOOTHING = 10: Smoothing is performed over
25 minutes.
4.8.7LIMIT TSS OUT LOW
4.8.8LIMIT TSS OUT HIGH
Parameterization and operation
TSS measurement values for the thickened sludge that are below
this value [g/L] are set to this value (to avoid low peaks).
4.8.9TSS OUT SMOOTHING
4.8.10 POLYMER DOSING MINIMUM
TSS measurement values for the thickened sludge that are
above this value [m
TSS measurement values from the effluent are smoothed in line
with this parameter.
SMOOTHING = 1: The signal is not smoothed.
SMOOTHING = 2: Smoothing is performed over 3 minutes.
SMOOTHING = 3: Smoothing is performed over 5 minutes.
SMOOTHING = 5: Smoothing is performed over 12 minutes.
SMOOTHING = 10: Smoothing is performed over
25 minutes.
RTC calculations below this value [g/L] are set to this value and
transferred to the polymer pump.
Note: When FEED FLOW CONTROL is activated, measurement values
for the polymer dosing rate that are below this value [m
value (to avoid low peaks in the dosing flow).
3
/h] are set to this value (to avoid high peaks).
3
/h] are set to this
4.8.11 POLYMER DOSING MAXIMUM
4.9INPUTS
4.9.1MIN FEED FLOW
4.9.2MAX FEED FLOW
4.9.30/4...20 mA
RTC calculations above this value [g/L] are set to this value and
transferred to the polymer pump.
Note: When FEED FLOW CONTROL is activated, measurement values
for the polymer dosing rate that are above this value [m
value (to avoid high peaks in the dosing flow).
Minimum flow rate [m³/h] from the influent in accordance with the
0/4 mA measurement signal.
Maximum flow rate [m³/h] from the influent in accordance with the
20 mA measurement signal.
Transfer range of the 0/4 to 20 mA current loop (as set in
connected flow measuring instrument).
3
/h] are set to this
31
Parameterization and operation
4.9.4MIN POLYMER FLOW
4.9.5MAX POLYMER FLOW
4.9.60/4...20 mA
4.10 OUTPUTS
4.10.1 MIN FEED FLOW
4.10.2 MAX FEED FLOW
Minimum polymer dosing in [L/h] in accordance with the 0/4 mA
measurement signal.
Maximum polymer dosing in [L/h] in accordance with the 20 mA
measurement signal.
Transfer range of the 0/4 to 20 mA current loop (as set in
connected flow measuring instrument).
Minimum feed flow rate [m³/h] in accordance with 0/4 mA.
Maximum feed flow rate [m³/h] in accordance with 20 mA.
4.10.3 0/4...20 mA
4.10.4 MIN POLYMER FLOW
4.10.5 MAX POLYMER FLOW
4.10.6 0/4...20 mA
4.10.7 CONTROL CYCLE
Transfer range of 0/4 to 20 mA current loop (as set in connected
flow measuring instrument).
Minimum polymer pump delivery rate in accordance with 0/4 mA.
Maximum polymer pump delivery rate in accordance with 20 mA.
Transfer range of 0/4 to 20 mA current loop (as set in connected
flow measuring instrument).
Pulse/pause mode for the polymer pump open-loop control for
dosing rates beneath the minimum polymer flow rate (MIN
POLYMER FLOW). The on/off duration in pulse/pause mode can
be affected by the duration of the CONTROL CYCLE. For
example, with a CONTROL CYCLE of 100 seconds and a dosing
control value of 60 %, the polymer pump is switched on for
60 seconds and switched off for 40 seconds. Short cycle times
increase the switching frequency but enable more precise
adaptation to individual requirements.
32
Note: CONTROL CYCLE must be divisible by MIN RUNTIME and
produce a whole number.
Parameterization and operation
4.10.8 MIN RUNTIME
Minimum ON time in pulse/pause dosing mode. The pump is
activated for this runtime at the very least. The MIN RUNTIME
must be shorter than the duration of the CONTROL CYCLE.
4.11 Displayed measurement values and variables
The following measurement values and variables are shown on
the sc1000 display and transferred via fieldbus (refer to
Measurement 1Qin 1m
Measurement 2 Qavg 1m
Measurement 3 Qdos1 L/h Polymer flow rate
Measurement 4 TSin 1 g/L TSS concentration from the influent
Measurement 5 TSef 1 g/L TSS concentration from the effluent
Actuat var 6 Pdos1 L/h Polymer dosing
Actuat var 7 Fact 1 g/kg Specific polymer dosing
Actuat var 8 Feed 1 m
/h Flow rate from the influent
3
/h Average flow rate
3
/h Feed flow rate
RTC113 ST-Module, two-channel Parameter Unit Description
3
Measurement 1 Qin 1 m
Measurement 2 Qavg 1 m
Measurement 3 Qdos 1 L/h Polymer flow rate 1
Measurement 4 TSin 1 g/L TSS concentration from the influent 1
Measurement 5 TSef 1 g/L TSS concentration in the effluent 1
Measurement 6 Qin 2 m
Measurement 7 Qavg 2 m
Measurement 8 Qdos 2 L/h Polymer flow rate 2
Measurement 9 TSin 2 g/L TSS concentration from the influent 2
Measurement 10 TSef 2 g/L TSS concentration in the effluent 2
Actuat var 11 Pdos 1 L/h Polymer dosing 1
Actuat var 12 Fact 1 g/kg Specific polymer dosing 1
Actuat var 13 Feed 1 m
Actuat var 14 Pdos2 L/h Polymer dosing 2
Actuat var 15 Fact 2 g/kg Specific polymer dosing 2
Actuat var 16 Feed 2 m
/h Flow rate from the influent 1
3
/h Average flow rate
3
/h Flow rate from the influent 2
3
/h Average flow rate
3
/h Feed flow rate 1
3
/h Feed flow rate 2
33
Parameterization and operation
34
Section5Maintenance
Multiple hazards
Only qualified personnel must conduct the tasks described in this section of the manual.
5.1Maintenance schedule
IntervalMaintenance task
DANGER
Visual inspection
CF card2 years
Battery5 years
Application-specificCheck for contamination and corrosion
Replacement by manufacturer's service department (Section8,
page41)
Replacement by manufacturer's service department (Section8,
page41)
35
Maintenance
36
Section6Troubleshooting
6.1Error messages
Possible RTC errors are displayed by the sc controller.
Table5
Displayed errorsCauseResolution
Supply RTC with voltage
RTC MISSING
RTC CRC
CHECK CONFIG
RTC FAILURE
INFLOW1 NOT G.Influent measurement signal faultyTest sensor, check cable connections
INFLOW2 NOT G.Influent measurement signal faultyTest sensor, check cable connections
No communication between RTC and RTC
communication card
Interrupted communication between RTC and
RTC communication card
The sensor selection of the RTC was deleted
by deleting or selecting a new sc1000
participant.
Brief general read/write error on the CF card,
mostly caused by a brief interruption to the
power supply.
Test connection cable
Reset the sc1000 and the RTC (switch so it is
completely voltage free and switch back on)
Make sure +/- connections of the connector
cable between RTC and RTC communication
card in the sc1000 are installed correctly.
From MAIN MENU > RTC MODULES /
PROGNOSYS > RTC MODULES > RTC >
CONFIGURE > SELECT SENSOR, select the
correct sensor for the RTC again and confirm.
Acknowledge error. If this message is shown
frequently, eliminate the cause of the power
disruptions. If necessary, inform the service
team of the manufacturer (Section8, page41).
6.2Warnings
Possible RTC sensor warnings are displayed by the sc controller.
Table6
Displayed warningsCauseResolution
The RTC menu SET DEFAULTS was opened.
MODBUS ADDRESS
PROBE SERVICEA configured sensor is in service status.The sensor must exit service status.
This deleted the Modbus address of the RTC in
the sc1000.
Go to MAIN MENU > RTC MODULES /
PROGNOSYS > RTC MODULES > RTC >
CONFIGURE > MODBUS > ADDRESS and
set the correct MODBUS address.
6.3Wear parts
Table7
DesignationNumberService life
CF card, type RTC module1 piece
Battery1 piece
~2years
~5years
37
Troubleshooting
38
Section7Replacement parts and accessories
7.1Replacement parts
DescriptionCat. No
DIN rail NS 35/15, punched according to DIN EN 60715 TH35, made of galvanized steel.
Length: 35 cm (13.78 in.)
Transformer 90–240 V AC/24 V DC 0.75 A, module for DIN rail assembly LZH166
Terminal for 24 V connection without power supplyLZH167
Terminal for protective earthLZH168
SUB-D connectorLZH169
C2 circuit breakerLZH170
CPU base module with Ethernet port, passive ventilation element. (CX1010-0021) and
RS422/485 connection module (CX1010-N031)
Power supply module, consisting of a bus coupler and a 24 V terminal module (CX1100-0002)LZH172
Digital output module 24 V DC (4 outputs) (KL2134)LZH174
Analog output module (2 outputs) (KL4012)LZH176
Analog input module (1 input) (KL3011)LZH177
Digital input module 24 V DC (2 inputs) (KL1002) LZH204
Digital output module 24 V DC (8 outputs) (KL2408) LZH205
Bus termination module (KL9010)LZH178
RTC communication cardYAB117
CF card type RTC-ModuleLZY748-00
Villa 14 – Rue 2 Casa
Plaisance
Quartier Racine Extension
MA-Casablanca 20000
Tél. +212 (0)522 97 95 75
Fax +212 (0)522 36 89 34
info-maroc@hach-lange.com
www.hach-lange.ma
42
Section9Warranty and liability
The manufacturer warrants that the supplied product is free of material and manufacturing
defects, and undertakes to repair or to replace any defective parts without charge.
The warranty period is 24 months. If a maintenance contract is taken out within 6 months
of purchase, the warranty period is extended to 60 months.
With the exclusion of further claims, the supplier is liable for defects, including the lack of
assured properties, as follows: all parts that, within the warranty period calculated from the
day of the transfer of risk, can be demonstrated to have become unusable or that can only
be used with significant limitations owing to circumstances prior to transfer of risk, in
particular due to incorrect design, substandard materials or inadequate finish, shall be
repaired or replaced at the supplier's discretion. The identification of such defects must be
reported to the supplier in writing as soon as possible, but no later than 7 days after the
discovery of the fault. If the customer fails to notify the supplier, the product is considered
approved despite the defect. Further liability for indirect or direct damages is not accepted.
If device-specific maintenance- or inspection work prescribed by the supplier is to be
performed within the guarantee period by the customer (maintenance) or by the supplier
(inspection) and these requirements are not met, claims for damages that result from
non-observance of these requirements are void.
Further claims, in particular for consequential damages, cannot be made.
Wear and damage caused by improper handling, incorrect installation or non-designated
use are excluded from this clause.
The process instruments of the manufacturer have proven their reliability in many
applications and are therefore often used in automatic control loops to enable the most
economical and efficient operation of the relevant process.
To avoid or limit consequential damage, it is therefore recommended that the control loop
be designed such that an instrument malfunction results in an automatic changeover to
the backup control system. This guarantees the safest operating condition both for the
environment and the process.
43
Warranty and liability
44
AppendixAModbus address setting
The same slave address must be set for Modbus communication on the sc1000 controller
display and in the RTC module. Since 20 slave numbers are reserved for internal
purposes, the following numbers are available for assignment:
1, 21, 41, 61, 81, 101…
The start address 41 is preset at the factory.
NOTICE
If this address is to be or must be changed because, for example, it has already been allocated
for another RTC, the changes must be made both on the sc1000 controller and on the CF card of
the RTC module.
This can only be done by the manufacturer service department (Section8)!
45
Modbus address setting
46
AppendixBConfiguration of the network modules
B.1RTC113 ST-Module Profibus/Modbus telegram
Table8 RTC113 ST-Module, one-channel
RegisterParameterUnit Description
3
MEASUREMENT 1Qin 1m
MEASUREMENT 2 Qavg 1m
MEASUREMENT 3 Qdos1 L/h Polymer flow rate
MEASUREMENT 4 TSin 1 g/L TSS concentration in the inflow
MEASUREMENT 5 TSef 1 g/L TSS concentration in the outflow
ACTUAT VAR 6 Pdos1 L/h Polymer dosing
ACTUAT VAR 7 Fact 1 g/kg Specific polymer dosing
ACTUAT VAR 8 Feed 1 m
Table9 RTC113 ST-Module, two-channel
RegisterParameter Unit Description
MEASUREMENT 1 Qin 1 m
MEASUREMENT 2 Qavg 1 m
MEASUREMENT 3 Qdos 1 L/h Polymer flow rate 1
MEASUREMENT 4 TSin 1 g/L TSS concentration in inflow 1
MEASUREMENT 5 TSef 1 g/L TSS concentration in outflow 1
MEASUREMENT 6 Qin 2 m
MEASUREMENT 7 Qavg 2 m
MEASUREMENT 8 Qdos 2 L/h Polymer flow rate 2
MEASUREMENT 9 TSin 2 g/L TSS concentration in inflow 2
MEASUREMENT 10 TSef 2 g/L TSS concentration in outflow 2
ACTUAT VAR 11 Pdos 1 L/h Polymer dosing 1
ACTUAT VAR 12 Fact 1 g/kg Specific polymer dosing 1
ACTUAT VAR 13 Feed 1 m
ACTUAT VAR 14 Pdos2 L/h Polymer dosing 2
ACTUAT VAR 15 Fact 2 g/kg Specific polymer dosing 2
ACTUAT VAR 16 Feed 2 m
/h Flow rate in the inflow
3
/h Average flow rate
3
/h Feed flow rate
3
/h Flow rate in inflow 1
3
/h Average flow rate
3
/h Flow rate from the influent 2
3
/h Average flow rate
3
/h Feed flow rate 1
3
/h Feed flow rate 2
47
Configuration of the network modules
48
Index
Numerics
1-channel version ..................................................... 18
2-channel version ..................................................... 18