Read the entire instruction sheet carefully before unpacking, setting up or operating this
device. Pay attention to all danger and warning statements. Failure to do so could result in
serious injury to the operator or damage to the device.
To make sure that the protection provided by this device is not impaired, do not use or
install this device in any manner other than that specified in this instruction sheet.
Danger notes in this instruction sheet
Indicates a potentially or imminently hazardous situation that, if not avoided, can result in death or
serious injury.
Indicates a potentially or imminently dangerous situation that, if it is not avoided, can lead to
death or to serious injuries.
Indicates a possible dangerous situation that can have minor or moderate injuries as the result.
DANGER
WARNING
CAUTION
If shown on the device, this symbol references the user manual for operation and/or for safety information.
This symbol may be shown on a product enclosure or barrier and indicates a risk of electric shock and/or a risk of
death by electrocution.
Electrical devices marked with this symbol may no longer be disposed of in unsorted domestic or industrial waste in
Europe after August 12, 2005. In conformity with the provisions in force (EU Directive 2002/96/EC), consumers in
the EU must return old electrical devices to the manufacturer for disposal from this date. This is free for the
consumer.
Note: Contact the manufacturer or supplier for instructions on how to return end-of-life devices,
manufacturer-supplied electrical accessories, and all auxiliary items for proper disposal.
Introduction
NOTICE
Indicates a situation that, if it is not avoided, can lead to damage to the device. Information that
requires special emphasis.
Warning labels
Read all labels and tags attached to the device. Failure to do so could result in personal
injury or damage to the device. Symbols shown on the device are noted in this instruction
sheet with a danger or caution statement.
In this instruction sheet, the assembly and installation of the power connection box
(catalog number LQV155.99.0000x) is described.
The power connection box serves to connect sc analyzers to a power supply. It offers a
power connection, fuses and power sockets for two sc analyzers.
DANGER
Risk of electric shock. The procedures described in this instruction sheet may only be performed
by qualified personnel.
3
The parts included in the scope of delivery for the power connection box are listed in
Figure 1, page 4. Under some circumstances, not all supplied parts are required.
Therefore the only parts listed are those mentioned in this instruction sheet.
The dimensions of the installation are shown in Figure 2, page 4.
Note: Important note: Leave underneath the power connection box (i.e. 50–100 mm) for
cables and connectors.
Figure 1 Parts included in the scope of delivery for the power connection box (LQV155.99.0000x)
1Power connection box 5Nut, M6 (4x) (LZQ058)
2Assembly kit for the rail or pole assembly (LZQ059)6Washer for M6 (4x) (LZQ058)
3Clamping plate for LZQ060 (LZQ059)7Hexagons socket screw, M6 x 100 mm (4x) (LZQ058)
4Screw set for LZQ058 (LZQ059)
Figure 2 Dimensions
1Lid screw (4x)3Lid
2Screw hole (4x)
4
Assembly
Risk of electric shock.
This device may only be connected to a single-phase power supply at 100–240 V~, which is
protected with a 15 A fuse or a current breaker.
Risk of electric shock.
The wire connection of the power supply must always contain a 2-pin disconnect switch available
on site. The disconnect switch must meet the applicable regulations. This disconnect switch must
be installed in the immediate vicinity of the device and be easily accessible; it must also be
labeled as a disconnect switch. Before opening the power supply cover, disconnect the system
electrically via the disconnect switch.
Installation location
The power connection box can be installed on a wall, a vertical pole or on horizontal rails.
During the installation of the power connection box, read the user manuals for the sensor and for
the controller. Pay particular attention to all safety instructions and danger notes.
DANGER
DANGER
NOTICE
Leave sufficient space underneath the power connection box (i.e. 50–100 mm) for cables and
connectors.
Note: Always make sure that all parts of the system can be accessed without hindrances when the
installation location is selected.
Wall assembly
1. Loosen the four lid screws and remove the lid (refer to Figure 2, page 4).
2. Secure the housing with the four screws provided to the wall (refer to Figure 1,
page 4 and Figure 2, page 4). The dimensions required for assembly can be found in
Figure 3, page 5.
Figure 3 Dimensions
NOTICE
5
Rail or pole assembly
Note: Make sure that the diameter of the rail or of the pole is between 35 and 55 mm.
Note: Instructions on how to secure the housing with the clamping plate can be found in Figure 4,
page 6 or Figure 5, page 6.
Figure 4 Assembly on a horizontal rail
1Washer for M64Rail
2Nut5Power connection box
3Clamping plate6Hexagon socket screw, M6 x 100 mm
Figure 5 Pole assembly
1Washer for M64Rail
2Nut5Power connection box
3Clamping plate6Hexagon socket screw, M6 x 100 mm
6
Electrical connections
Risk of electric shock. For installation in wet areas or areas in which wetness can occur, a
residual-current circuit breaker must always be installed with this socket. For assembly in outdoor
areas, overvoltage protection is required. For fixed wire connections (cable conduit), a clearly
labeled 2-pin disconnect switch must be installed on site next to the socket, in accordance with
the applicable regulations. For the wire connection to the on-site disconnect switch, disconnect
the power supply to the socket.
With fixed wiring, a disconnecting device (local interruption) must be integrated into the supply
line. The disconnecting device must meet the applicable standards and regulations. It must be
installed near the device, be able to be reached easily by the operator and labeled as a
disconnecting device.
If the connection is established using a mains connection cable that is permanently connected to
the power supply, the plug of the mains connection cable can serve as local interruption.
Electrical dangers and fire hazards. Use only the supplied power cable.
Only qualified experts may perform the tasks described in this section of the manual, while
adhering to all locally valid safety regulations.
DANGER
DANGER
WARNING
NOTICE
Use only earthed sockets for the connection of this device to the power supply.
If you are not sure if the sockets are earthed, have this checked by a qualified electrician.
The power plug serves in addition to the power supply to isolate the device quickly from the
mains where necessary.
This is recommended for long-term storage and can prevent potential dangers in the event of a
fault.
Therefore make sure that the sockets to which the device is connected are easy to reach by each
user at all times.
DANGER
The products of the manufacturer for use outdoors have a comprehensive protection against the
penetration of moisture and dust. If these products are connected to a mains socket of the main
power line via a cable with plug instead of fixed-installation wire connections, the plug and socket
are considerably more susceptible to the penetration of moisture and dust.
It lies in the responsibility of the operator to protect plug and socket in a way that the connection
has an adequate protection against the penetration of moisture and dust and fulfils the local
safety conditions.
When the device is used outdoors, it should be connected only via a suitable socket that has at
least protection rating IP44 (protection against spray water from all directions).
The power connection box is available with and without a pre-assembled power cable. If
no pre-assembled power cable is provided, the power connection box can be connected
to the mains power via a fixed wire connection or a power cable with suitable strain relief.
The wire connection is established via the same connection terminals for a fixed wire
connection and a cable wire connection.
For the wire connection, proceed as follows (refer to Figure 6, page 9):
1. Make sure that the housing is well secured.
2. Remove the four screws and the socket lid (component 8).
7
3. Unscrew the nut of the strain relief (component 5).
4. Remove the cap of the strain relief (component 4).
5. Strip off the insulation of all three conductors of the power cable (phase, neutral
conductor and earth conductor) over a length of 8 mm. The connection terminals are
designed for solid or stranded conductors (18 – 16 AWG).
6. Feed the cables through the strain relief or cable guide. If strain relief is used, feed
the power cable through the strain relief nut (component 5), the cap (component 4)
and the cable gland (component 3).
7. If strain relief is used, insert the cap (component 4) into the cable gland (component
3) and screw the nut (component 5) onto the cable gland. Do not tighten the nut.
NOTICE
Leave the bridges between the terminals 1 and 2 as well as 3 and 4. Both the conductors of the
power cable and the bridges must be connected to the corresponding terminals.
8. Open the fuse holder of terminal 1 — to do this, push the fuse holder down (refer to
Figure 7, page 9).
9. Connect the phase to terminal 1 (refer to Figure 8, page 10) and tighten the screw.
10. Insert the fuse holder again.
11. Open the fuse holder of terminal 4.
12. Connect the neutral conductor to terminal 4 and tighten the screw.
13. Insert the fuse holder again.
DANGER
Risk of electric shock.
A good ground connection to this product is required to make sure that this and all other
connected devices fulfill the safety standards and safety provisions.
NOTICE
The terminal block for the ground connection has several terminals. The outer screws (top and
bottom screw) are intended for the front terminals.
14. Connect the front ground cable to terminal 5 and tighten the top screw.
15. For wire connections with a power cable, tighten the nut of the cable gland
(component 3 in Figure 6, page 9), so that the cable sits firmly in the strain relief. Do
not tighten the inner conductors of the power cable too tightly under tension (instead
leave a small loop for service work).
16. Position the socket lid again and tighten the four screws.
8
Figure 6 Cable connections
1Cable guide6Power socket
2Sealing ring
3Cable glandObserve the output voltage at the power sockets.
The output voltage supplied by the power connection box to
the power sockets corresponds to the country-specific mains
voltage to which the power connection box is connected.
Never connect consumers with a lower input voltage to the
power connection box, if the power connection box is
operated at a higher mains voltage.
4Cap7Lid
5Nut8Power connection box
Note
Figure 7 Fuse holder
1Fuse2Fuse holder
9
Figure 8 Circuit diagram (see inside of the lid)
1Phase, line (live)7Black for left power socket
2Connection to terminal 1 via bridge8Black for right power socket
3Connection to terminal 4 via bridge9White for left power socket
4Neutral conductor10 White for right power socket
5Protective ground (two terminals)11 Fuse, T 5 A H, 250 V (4x)
6Protective ground for left and right power socket12 Lid protective ground
Clean the power connection socket
Risk of electric shock.
Disconnect the power connection socket from the power supply before it is cleaned in order to
avoid electric shocks.
Wipe the outside of the housing with a damp cloth. Make sure that the connector is well
sealed with the sealing cap. Use a mild cleaning agent if required. Solvents must not be
used.
Replace the fuse
Risk of electric shock.
The procedures described in this section of the instruction sheet may only be performed by
qualified personnel.
Risk of electric shock.
An incorrect fuse can lead to injuries and damage. The fuse must always be replaced with a fuse
of the same type and with the same nominal values.
DANGER
DANGER
DANGER
10
1. Make sure that the power connection box is disconnected from the power supply.
2. Loosen the screws and remove the lid.
3. Open the fuse holder (there are two fuses for each analyzer connection: one for the
phase and another for the neutral conductor — refer to Figure 7, page 9) and replace
damaged fuses with fuses of the same type and with the same nominal values (T
5 A, time-lag, H, 250 V). Close the fuse holder (component 11, Figure 8, page 10).
4. Position the lid again and tighten the screws.
Connect the sc analyzer
Risk of electric shock.
If a power socket (component 4 in Figure 9, page 11) is not to be used, make sure that the power
socket is well sealed with the cap (component 2 in Figure 9, page 11).
1. Remove the sealing cap from the connector.
2. Connect the analyzer to the power socket.
3. Secure the connector of the analyzer to the power socket.
4. Connect the sealing caps of the connector (component 1) and of the power socket
(component 2) securely together.
Figure 9 Connections of the analyzer
DANGER
1Sealing cap for the analyzer connector3Analyzer connector
2Sealing cap for the power socket4Power socket of the analyzer
11
Technical data
Subject to change.
General information
Power supply 100–240 V~, + 10 %–15 %; 50/60 Hz; maximum 1000 VA
Installation categoryII
Pollution degreeII
FusesT 5 A (time-lag) H, 250 V (4x)
Dimensions
Weight1250 g
Housing protection ratingIP56 (with protective caps or connectors assembled)
CertificationsCE, cTUVus
(H x W x T) housing: 168 mm x 122 mm x 80 mm
Assembled: (free space for the cables) x width x (free space for the hardware)
Assembly
Cable gland
Environmental conditions
Ambient temperature–20 °C to +45 °C, 95 % relative humidity, non-condensing
Materials
EnclosureAluminum with powder coating
Wall, pole or rail (with supplied assembly kit LZQ059):
Ø 35 mm–55 mm
Strain relief: water resistant, 8–13 mm
Cable guide: 1,3 cm (0,5 inches)
Conductors: maximum 1,5 mm2 (AWG 16)
Replacement parts and accessories
DescriptionQuantityCatalog number
Clamping plate 1LZQ060
Cable guide1LZX981
Fuse set (4x, T 5 H/250 V)1LZY460
Assembly kit for rail or pole assembly1LZQ059
Screws set1LZY058
12
Warranty and liability
The manufacturer warrants that the product supplied is free of material and manufacturing
defects and undertakes the obligation to repair or replace any defective parts at zero cost.
The warranty period for instruments is 24 months. If a service contract is taken out within
6 months of purchase, the warranty period is extended to 60 months.
With the exclusion of the further claims, the supplier is liable for defects including the lack
of assured properties as follows: all those parts that, within the warranty period calculated
from the day of the transfer of risk, can be demonstrated to have become unusable or that
can only be used with significant limitations due to a situation present prior to the transfer
of risk, in particular due to incorrect design, poor materials or inadequate finish will be
improved or replaced, at the supplier's discretion. The identification of such defects must
be notified to the supplier in writing without delay, however at the latest 7 days after the
identification of the fault. If the customer fails to notify the supplier, the product is
considered approved despite the defect. Further liability for any direct or indirect damages
is not accepted.
If instrument-specific maintenance and servicing work defined by the supplier is to be
performed within the warranty period by the customer (maintenance) or by the supplier
(servicing) and these requirements are not met, claims for damages due to the failure to
comply with the requirements are rendered void.
Any further claims, in particular claims for consequential damages cannot be made.
Consumables and damage caused by improper handling, poor installation or incorrect use
are excluded from this clause.
The manufacturer process instruments are of proven reliability in many applications and
are therefore often used in automatic control loops to provide the most economical
possible operation of the related process.
To avoid or limit consequential damage, it is therefore recommended to design the control
loop such that a malfunction in an instrument results in an automatic change over to the
backup control system; this is the safest operating state for the environment and the
process.