Process water monitoring (condensates, cleaning cycles, heat
(IP65)
< 30 seconds< 30 seconds< 30 seconds< 30 seconds
150 °C150 °C150 °C150 °C
-1
Pure water production
exchangers)
Glass polyester
Chemical processes
Demineralized and softened water
0.1 μS—2 mS.cm
(IP65)
-1
Glass polyester
(IP65)
1 μS—20 mS.cm
Monitoring of
ultrapure water in
pharmaceutical and
food industries
Suitable for CIP-SIP
processes
Glass polyester
(IP65)
-1
-1
English 3
Specification8315 8316 8317 8394
Maximum pressure25 bars25 bars25 bars25 bars
Sample connection¾ inch NPT¾ inch NPT¾ inch NPT
Tri-Clamp 1½ or
2 inch
Inductive conductivity sensors
Specification8398 series8332
ApplicationCorrosive and fouling applicationsMeasurement in harsh conditions
MaterialPEEKPVDF
Cell constant K2.35 (cm-1)10 (cm-1)
Accuracy± 2 %± 2 %
Measurement range0—2000 mS/cm100 μS/cm—10 S/cm
Conductivity response< 1 second< 1 second
Temperature response< 2 minutes< 2 minutes
Maximum pressure18 bars7.5 bars
Maximum temperature140 °C120 °C
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.
Use of hazard information
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious
injury.
Indicates a potentially hazardous situation that may result in minor or moderate injury.
4 English
D A N G E R
W A R N I N G
C A U T I O N
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special
emphasis.
Precautionary labels
Read all labels and tags attached to the product. Personal injury or damage to the product could
occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary
statement.
This symbol, when noted on a product, indicates a potential hazard which could cause serious
personal injury and/or death. The user should reference this instruction manual for operation and/or
safety information.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock
and/or electrocution exists and indicates that only individuals qualified to work with hazardous
voltages should open the enclosure or remove the barrier.
This symbol, when noted on the product, indicates the presence of devices sensitive to electrostatic
discharge and indicates that care must be taken to prevent damage to them.
Electrical equipment marked with this symbol may not be disposed of in European public disposal
systems. In conformity with European local and national regulations, European electrical equipment
users must now return old or end-of-life equipment to the manufacturer for disposal at no charge to
the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Products marked with this symbol indicates that the product contains toxic or hazardous substances
or elements. The number inside the symbol indicates the environmental protection use period in
years.
Product overview
The module, when installed in an sc200 or a Polymetron 9500 controller, allows an analog sensor to
connect to the controller.
Installation
W A R N I N G
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
Connect the sensor to the module
Potential electrocution hazard. Always disconnect power to the controller when making electrical connections.
Electrocution hazard. High voltage wiring for the controller is conducted behind the high voltage barrier in the
controller enclosure. The barrier must remain in place except when installing modules, or when a qualified
installation technician is wiring for power, relays or analog and network cards.
Potential instrument damage. Delicate internal electronic components can be damaged by static electricity,
resulting in degraded performance or eventual failure.
To install the module and connect the sensor, refer to the illustrated steps that follow and the wiring
table.
Be sure to connect all sensor ground/shield wires to the controller enclosure grounding screws.
W A R N I N G
W A R N I N G
N O T I C E
English
5
Note: If the sensor cable is not long enough to reach the controller, an interconnect cable (model 150727,10000)
and junction box (model 08335=A=6000) should be used to extend the distance.
Table 1 Conductivity sensor wiring
TerminalDescriptionSensor
8398 831x and 8394
1 Inner electrodeWhite (white cable)White
2 Inner electrodeBlack (white cable)—
3 Ground—Green
4 ———
5 ———
6 ———
7 Temp -GreenBlack
8 ———
9 ———
10 Temp +YellowYellow
11 Outer electrodeBrown (brown cable)Red
12 Outer electrodeBlack (brown cable)—
In Step 3 of the following sequence, set the switch as follows:
Switch SettingDescription
Setting 1 Contacting conductivity sensor
Setting 2 Inductive conductivity sensor
6 English
English 7
8 English
Operation
User navigation
Refer to the User interface and navigation section of the controller documentation for keypad
description and navigation information.
Sensor configuration
Contacting conductivity sensor configuration
Use the CONFIGURE menu to enter identification information for the sensor and to change options
for data handling and storage.
1. Push the menu key and select SENSOR SETUP>[Select Sensor]>CONFIGURE.
2. Select an option and push enter. To enter numbers, characters or punctuation, push and hold the
up or down arrow keys. Push the right arrow key to advance to the next space.
OptionDescription
EDIT NAMEChanges the name that corresponds to the sensor on the top of the measure screen.
SENSOR S/NAllows the user to enter the serial number of the sensor, limited to 16 characters in any
SELECT MEASURE Changes the measured parameter to CONDUCTIVITY (default), TDS (total dissolved
DISPLAY FORMAT Changes the number of decimal places that are shown on the measure screen. When
MEAS UNITSChanges the units for the selected measurement—select the unit from the list
TEMP UNITSSets the temperature units to °C (default) or °F.
T-COMPENSATION Adds a temperature-dependent correction to the measured value:
The name is limited to 16 characters in any combination of letters, numbers, spaces or
punctuation. Only the first 12 characters are displayed on the controller.
combination of letters, numbers, spaces or punctuation.
solids), SALINITY or RESISTIVITY. All other configured settings are reset to the
default values.
Note: If SALINITY is selected, the measurement unit is defined as ppt (parts per
thousand) and cannot be changed.
set to auto, the number of decimal places changes automatically with changes in the
measured value.
available.
• NONE—Temperature compensation not required
• USP—Set the alarm level for the standard USP definition table
• ULTRA PURE WATER—Not available for TDS. Set the compensation type
according to the sample characteristics—Select NaCl , HCl, AMMONIA or ULTRA
PURE WATER
• USER—Select BUILT IN LINEAR, LINEAR or TEMP TABLE:
• BUILT IN LINEAR—Use the pre-defined linear table (slope defined as 2.0%/°C,
reference temperature as 25 °C)
• LINEAR—Set the slope and reference temperature parameters if different from
the built-in parameters
• TEMP TABLE—Set the temperature and multiplication factor points (refer to
Configure user tables on page 11)
• NATURAL WATER—Not available for TDS
CONFIG TDSTDS only—changes the factor that is used to convert conductivity to TDS: NaCl
CABLE PARAMSets the sensor cable parameters to improve measurement accuracy when the sensor
(0.49 ppm/µS) or CUSTOM (enter factor between 0.01 and 99.99 ppm/µS).
cable is extended or shortened from the standard 5 m. Enter the cable length,
resistance and capacitance.
English 9
OptionDescription
TEMP ELEMENTSets the temperature element to PT100 or PT1000 for automatic temperature
FILTERSets a time constant to increase signal stability. The time constant calculates the
LOG SETUPSets the time interval for data storage in the data log—5, 30 seconds, 1, 2, 5, 10,
RESET DEFAULTS Sets the configuration menu to the default settings. All sensor information is lost.
compensation. If no element is used, the type can be set to MANUAL and a value for
temperature compensation can be entered.
average value during a specified time—0 (no effect) to 60 seconds (average of signal
value for 60 seconds). The filter increases the time for the sensor signal to respond to
actual changes in the process.
15 (default), 30, 60 minutes.
Inductive conductivity sensor configuration
Use the CONFIGURE menu to enter identification information for the sensor and to change options
for data handling and storage.
1. Push the menu key and select SENSOR SETUP>[Select Sensor]>CONFIGURE.
2. Select an option and push enter. To enter numbers, characters or punctuation, push and hold the
up or down arrow keys. Push the right arrow key to advance to the next space.
OptionDescription
EDIT NAMEChanges the name that corresponds to the sensor on the top of the measure screen.
SENSOR S/NAllows the user to enter the serial number of the sensor, limited to 16 characters in any
SELECT MEASURE Changes the measured parameter to CONDUCTIVITY (default), RESISTIVITY,
DISPLAY FORMAT Changes the number of decimal places that are shown on the measure screen. When
MEAS UNITSChanges the units for the selected measurement—select the unit from the list
TEMP UNITSSets the temperature units to °C (default) or °F.
T-COMPENSATION Adds a temperature-dependent correction to the measured value:
The name is limited to 16 characters in any combination of letters, numbers, spaces or
punctuation. Only the first 12 characters are displayed on the controller.
combination of letters, numbers, spaces or punctuation.
SALINITY, TDS (total dissolved solids) or CONCENTRATION. All other configured
settings are reset to the default values.
Note: If SALINITY is selected, the measurement unit is defined as ppt (parts per
thousand) and cannot be changed.
set to auto, the number of decimal places changes automatically with changes in the
measured value.
available.
• LINEAR—Set the slope and reference temperature parameters
• NATURAL WATER—Not available for TDS or CONCENTRATION
• TEMP TABLE—Set the temperature and multiplication factor points (refer to
Configure user tables on page 11)
• NONE—Temperature compensation not required
CONFIG TDSTDS only—changes the factor that is used to convert conductivity to TDS: NaCl
(0.49 ppm/µS) or CUSTOM (enter factor between 0.01 and 99.99 ppm/µS).
10 English
OptionDescription
CONFIG CONCCONCENTRATION only—sets the type of concentration table to use.
• BUILT-IN—Set the CHEMICAL that is measured:
• H2SO4 0-30%
• HNO3 0-28%
• HCl 0-18%
• NaOH 0-16%
Note: Although the range is shown as 0-16%, the tables can be more accurately
defined as 0-10% for all temperatures.
• NaCl 0-25%
• H3PO4 0-40%
• USER TABLE—Set the conductivity and concentration% points (refer to Configure
user tables on page 11)
TEMP ELEMENTSets the temperature element to PT100 or PT1000 for automatic temperature
FILTERSets a time constant to increase signal stability. The time constant calculates the
LOG SETUPSets the time interval for data storage in the data log—5, 30 seconds, 1, 2, 5, 10,
RESET DEFAULTS Sets the configuration menu to the default settings. All sensor information is lost.
compensation. If no element is used, the type can be set to MANUAL and a value for
temperature compensation can be entered.
average value during a specified time—0 (no effect) to 60 seconds (average of signal
value for 60 seconds). The filter increases the time for the sensor signal to respond to
actual changes in the process.
15 (default), 30, 60 minutes.
Configure user tables
Depending on the type of sensor and the type of measurement required, user-defined tables can be
selected during sensor configuration. The following explains how to configure a user-defined table.
Concentration tables
By default, this table contains only 2 points (P1 and P2) that define the minimum and maximum
limits. When P1 and P2 have been defined, additional points can be inserted between them. As an
example to define the minimum conductivity and concentration as zero, and the maximum
conductivity as 750 with a maximum concentration of 30. proceed as follows:
1. From the USER TABLE option select P1>EDIT POINT.
2. Enter the minimum conductivity (X) as 0.0 and push enter.
3. Enter the minimum concentration% (Y) as 0.0 and push enter.
4. Select SAVE CHANGES? and push enter.
5. From the USER TABLE option select P2>EDIT POINT.
6. Enter the maximum conductivity (X) as 750.0 and push enter.
7. Enter the maximum concentration% (Y) as 30.0 and push enter.
8. Select SAVE CHANGES? and push enter.
The following table has been defined:
ConductivityConcentration%
P1 0 0
P2 750 30
To insert an additional point with a conductivity of 250 and a concentration of 15, select the point with
a conductivity value immediately above the value to insert. In this example select P2. The new
conductivity value of 250 will then become P2 and the existing P2 will become P3 as follows:
English
11
1. From the USER TABLE option select P2>INSERT POINT.
2. Enter the conductivity (X) as 250.0 and push enter.
3. Enter the concentration% (Y) as 15.0 and push enter.
4. Select SAVE CHANGES? and push enter.
The table will now be defined as:
ConductivityConcentration%
P1 0 0
P2 250 15
P3 750 30
Additional points can be inserted up to a maximum of P10.
Note: The conductivity value taken into account is temperature compensated according to the temperature
compensation option configured.
Temperature tables
The process is the same as described for the concentration tables, but the X value refers to
temperature and the Y value is a factor used to multiply against the non-compensated conductivity
value to obtain a temperature compensated conductivity value.
For example, if the temperature table was configured as follows:
TemperatureValue
P1 0 1
P2 20 1.1
P3 40 1.2
This can be displayed graphically as:
This shows that a temperature of 10° corresponds to a factor of 1.05. Therefore, a non-compensated
conductivity value of 10 mS/cm at 10° will give a temperature compensated value of 10.5 mS/cm (i.e.
10 mS/cm multiplied by a factor of 1.05).
Calibrate the sensor
About sensor calibration
The sensor characteristics slowly shift over time and cause the sensor to lose accuracy. The sensor
must be calibrated regularly to maintain accuracy. The calibration frequency varies with the
application and is best determined by experience.
12
English
Contacting conductivity sensors: Use air (zero calibration) and the process sample to define the
calibration curve. When the process sample is used, the reference value must be determined with a
certified secondary verification instrument.
Inductive conductivity sensors: Use air (zero calibration) and a reference solution or process
sample to define the calibration curve. A reference solution calibration is recommended for best
accuracy. When the process sample is used, the reference value must be determined with a certified
secondary verification instrument.
Calibration limits
A calibration will fail if the value is ± 10% of the target value.
Cell constant
Before making a calibration make sure the sensor cell parameters are correct.
1. Push the menu key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
2. If the pass code is enabled in the security menu for the controller, enter the pass code.
3. Select CELL CONSTANT and push enter.
4. Contacting conductivity sensors: Select the cell K range for the sensor (0.01, 0.1 or 1.0) then
enter the actual K value as printed on the label attached to the sensor.
Inductive conductivity sensors: Enter the actual K value as printed on the label attached to the
sensor.
Zero calibration procedure
Use the zero calibration procedure to define the unique zero point of the sensor.
1. Remove the sensor from the process. Wipe with a clean towel to make sure the sensor is dry.
2. Push the menu key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
3. If the pass code is enabled in the security menu for the controller, enter the pass code.
4. Select ZERO CAL and push enter.
5. Select the option for the output signal during calibration:
OptionDescription
ACTIVEThe instrument sends the current measured output value during the calibration procedure.
HOLDThe sensor output value is held at the current measured value during the calibration procedure.
TRANSFER A preset output value is sent during calibration. Refer to the controller user manual to change
6. Place the sensor in air, push enter.
7. Review the calibration result:
• PASS—the sensor is calibrated and ready to measure samples.
• FAIL—the calibration is outside of accepted limits. Clean the sensor and retry. Refer to
Troubleshooting on page 18 for more information.
8. If the calibration passed, push enter to continue.
9. If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID. Refer
to Change calibration options on page 17.
10. On the NEW SENSOR screen, select whether the sensor is new:
Option Description
YESThe sensor was not calibrated previously with this controller. The days of operation and previous
NOThe sensor was calibrated previously with this controller.
the preset value.
calibration curves for the sensor are reset.
English 13
11. Return the sensor to the process and push enter. The output signal returns to the active state
and the measured sample value is shown on the measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.
Calibration with the process sample
The sensor can remain in the process sample, or a portion of the process sample can be removed
for calibration.
1. Push the menu key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
2. If the pass code is enabled in the security menu for the controller, enter the pass code.
3. Select SAMPLE CAL and push enter.
4. Select the option for the output signal during calibration:
OptionDescription
ACTIVEThe instrument sends the current measured output value during the calibration procedure.
HOLDThe sensor output value is held at the current measured value during the calibration procedure.
TRANSFER A preset output value is sent during calibration. Refer to the controller user manual to change
5. With the sensor in the process sample, push enter. The measured value is shown. Wait for thevalue to stabilize and push enter.
6. Measure the conductivity value with a certified secondary verification instrument. Use the arrowkeys to enter this value if different from the value displayed and push enter.
7. Review the calibration result:
• PASS—the sensor is calibrated and ready to measure samples.
• FAIL—the calibration is outside of accepted limits. Clean the sensor and retry. Refer to
Troubleshooting on page 18 for more information.
8. If the calibration passed, push enter to continue.
9. If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID. Refer
to Change calibration options on page 17.
10. On the NEW SENSOR screen, select whether the sensor is new:
Option Description
YESThe sensor was not calibrated previously with this controller. The days of operation and previous
NOThe sensor was calibrated previously with this controller.
11. Return the sensor to the process and push enter. The output signal returns to the active state
and the measured sample value is shown on the measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.
the preset value.
calibration curves for the sensor are reset.
Calibration with a reference solution (inductive conductivity sensors)
Calibration adjusts the sensor reading to match the value of a reference solution. Use a reference
solution that is at the same value or higher than the expected measurement readings.
Note: If the sensor is calibrated for the first time, be sure to complete the zero calibration first.
1. Thoroughly rinse the clean sensor in deionized water.
2. Push the menu key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
3. If the pass code is enabled in the security menu for the controller, enter the pass code.
4. Select SOLN CAL and push enter.
14
English
5. Select the option for the output signal during calibration:
OptionDescription
ACTIVEThe instrument sends the current measured output value during the calibration procedure.
HOLDThe sensor output value is held at the current measured value during the calibration procedure.
TRANSFER A preset output value is sent during calibration. Refer to the controller user manual to change
the preset value.
6. Put the sensor in the reference solution. Make sure that the sensing area is fully immersed in the
solution. Stir the sensor to remove bubbles.
7. Wait for the sensor and solution temperature to equalize. This can take 30 minutes or more if the
temperature difference between the process and reference solution is significant.
8. With the sensor in the reference solution, push enter.
9. Use the arrow keys to enter the reference solution temperature and push enter.
10. Use the arrow keys to enter the reference solution slope and push enter. The measured value is
shown. Wait for the value to stabilize.
11. Use the arrow keys to enter the reference value and push enter.
12. Review the calibration result:
• PASS—the sensor is calibrated and ready to measure samples. The new cell constant K value
is shown.
• FAIL—the calibration slope or offset is outside of accepted limits. Clean the sensor and retry
with a fresh reference solution. Refer to Troubleshooting on page 18 for more information.
13. If the calibration passed, push enter to continue.
14. If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID. Refer
to Change calibration options on page 17.
15. On the NEW SENSOR screen, select whether the sensor is new:
Option Description
YESThe sensor was not calibrated previously with this controller. The days of operation and previous
calibration curves for the sensor are reset.
NOThe sensor was calibrated previously with this controller.
16. Return the sensor to the process and push enter. The output signal returns to the active state
and the measured sample value is shown on the measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.
This type of calibration is sometimes referred to as an electrical calibration and is used to verify the
electrical connection rather than the sensor itself.
1. Push the menu key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
2. If the pass code is enabled in the security menu for the controller, enter the pass code.
3. Select RESISTOR CAL and push enter.
4. Select the option for the output signal during calibration:
OptionDescription
ACTIVEThe instrument sends the current measured output value during the calibration procedure.
HOLDThe sensor output value is held at the current measured value during the calibration procedure.
TRANSFER A preset output value is sent during calibration. Refer to the controller user manual to change
the preset value.
English 15
5. Disconnect the sensor from the sensor cable and push enter. Connect a 50 kΩ, 5 kΩ, 500 Ω, or
50 Ω NIST resistor (accuracy ±0.05%) on pins 2-3 of the sensor cable and push enter. Use the
arrow keys to enter the value of the resistor and push enter.
6. Review the calibration result:
• PASS—the calibration is successful.
• FAIL—the cell constant is outside of accepted limits. Clean the sensor and retry. Refer to
Troubleshooting on page 18 for more information.
7. If the calibration passed, push enter to continue.
8. If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID. Refer
to Change calibration options on page 17.
9. On the NEW SENSOR screen, select whether the sensor is new:
Option Description
YESThe sensor was not calibrated previously with this controller. The days of operation and previous
calibration curves for the sensor are reset.
NOThe sensor was calibrated previously with this controller.
10. Reconnect the sensor, return it to the process and push enter. The output signal returns to the
active state and the measured sample value is shown on the measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.
This type of calibration is sometimes referred to as an electrical calibration and is used to verify the
electrical connection rather than the sensor itself.
1. Push the menu key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
2. If the pass code is enabled in the security menu for the controller, enter the pass code.
3. Select RESISTOR CAL and push enter.
4. Select the option for the output signal during calibration:
OptionDescription
ACTIVEThe instrument sends the current measured output value during the calibration procedure.
HOLDThe sensor output value is held at the current measured value during the calibration procedure.
TRANSFER A preset output value is sent during calibration. Refer to the controller user manual to change
5.
the preset value.
Remove the sensor from the sample. Place a 5 kΩ, 500 Ω, 50 Ω, or 5 Ω resistor loop through the
sensor and push enter. Push enter when the displayed resistor value is stable. Use the arrow
keys to enter the actual value of the resistor and push enter.
6. Review the calibration result:
• PASS—the calibration is successful.
• FAIL—the calibration slope or offset is outside of accepted limits. Clean the sensor and retry.
Refer to Troubleshooting on page 18 for more information.
16
English
7. If the calibration passed, push enter to continue.
8. If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID. Refer
to Change calibration options on page 17.
9. On the NEW SENSOR screen, select whether the sensor is new:
Option Description
YESThe sensor was not calibrated previously with this controller. The days of operation and previous
calibration curves for the sensor are reset.
NOThe sensor was calibrated previously with this controller.
10. Reconnect the sensor, return it to the process and push enter. The output signal returns to the
active state and the measured sample value is shown on the measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.
Temperature calibration
The temperature sensor has been calibrated at the factory. However, it is recommended to always
calibrate the temperature sensor before calibrating the measurement sensor.
1. Put the sensor in a container of water that is at a known temperature. Measure the temperature of
the water with an accurate thermometer or independent instrument.
2. Push the menu key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
3. If the pass code is enabled in the security menu for the controller, enter the pass code.
4. Select 1 PT TEMP CAL and push enter.
5. The raw temperature value is displayed. Push enter.
6. Enter the correct value if different from that displayed and push enter.
7. Push enter to confirm the calibration.
Exit calibration procedure
If the back key is pushed during a calibration, the user can exit the calibration.
1. Push the back key during a calibration. Three options are shown:
OptionDescription
QUIT CALStop the calibration. A new calibration must start from the beginning.
BACK TO CAL Return to the calibration.
LEAVE CALExit the calibration temporarily. Access to other menus is allowed. To return to the
calibration, push the menu key and select SENSOR SETUP.
2. Use the arrow keys to select one of the options and push enter.
Change calibration options
The user can set a calibration reminder or include an operator ID with calibration data from this
menu.
1. Push the menu key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
2. If the pass code is enabled in the security menu for the controller, enter the pass code.
English
17
3. Select CAL OPTIONS and push enter.
4. Use the arrow keys to select an option and push enter.
OptionDescription
CAL REMINDER Sets a reminder for the next calibration in days, months or years—select the required
OP ID ON CALIncludes an operator ID with calibration data—YES or NO (default). The ID is entered
delay from the list.
during the calibration.
Troubleshooting
Intermittent data
During calibration, data is not sent to the datalog which will result in the datalog having areas where
the data is intermittent.
Test the contacting conductivity sensor
Figure 1 Sensor connector
1 Electrodes2 Pt100
1. Disconnect the sensor from the module.
2. Electrodes: Check the insulation between the two electrodes. There should be an infinite
resistance when the sensor is dry and exposed to air.
3. Pt100: Compare the resistance measured directly on the connector with the values below:
Temperature (°C)Resistance (Ω)
0 100.00
10 103.90
20 107.70
30 111.67
40 115.54
50 119.40
60 123.24
70 127.07
80 130.89
90 134.70
100 138.50
18 English
Test the inductive conductivity sensor
If there is a doubt concerning the operation of the sensor, disconnect it from the controller but leave it
in place in your sample process. Check the following resistance values across the terminals indicated
(refer also to the wiring table Table 1 on page 6):
the arrows keys to select a calibration and push enter
to view the details.
service-level passcode). All previous calibration data is
lost.
information about the electrode polarization, the cable
capacitance and the time before the next
measurement.
English 19
Table 2 Sensor DIAG/TEST menu (continued)
OptionDescription
SENSOR SIGNALSShows the current sensor signal information.
SENSOR DAYSShows the number of days that the sensor has been in
RESET SENSORResets the number of days that the sensor has been in
operation.
operation and resets all calibration data to defaults.
Error list
Errors may occur for various reasons. An error icon consists of an exclamation point within a circle.
When an error occurs, the error icon and the measurement screen flash alternately in the main
display. All outputs are held when specified in the controller menu. To view errors, push the menu
key and select DIAGNOSTICS. Then select the device to view any problems associated with that
device.
A list of possible errors is shown in Table 3.
Table 3 Error list for conductivity sensors
ErrorDescriptionResolution
ADC FAILUREThe analog to digital conversion
failed
SENSOR MISSINGThe sensor is missing or
disconnected
SENS OUT RANGEThe sensor signal is outside of the
accepted limits (2 S/cm)
Make sure that the sensor module is
fully inserted into the controller
connector. Replace the sensor
module.
Examine the wiring and connections
for the sensor and for the module.
Make sure that the terminal block is
fully inserted into the module.
Make sure that the display format is
set for the correct measurement
range.
Warning list
A warning icon consists of an exclamation point within a triangle. Warning icons appear on the right
of the main display below the measurement value. A warning does not affect the operation of menus,
relays and outputs. To view warnings, push the menu key and select DIAGNOSTICS. Then select
the device to view any problems associated with that device. The warning icon will no longer be
displayed once the problem has been corrected or acknowledged.
A list of possible warnings is shown in Table 4.
Table 4 Warning list for conductivity sensors
WarningDescriptionResolution
MEAS TOO HIGHThe measured value is > 2 S/cm,
1,000,000 ppm, 200% or 20,000 ppt
MEAS TOO LOWThe measured value is < 0 μS/cm,
0 ppm, 0% or 0 ppt
ZERO TOO HIGHThe zero calibration value is too
high
ZERO TOO LOWThe zero calibration value is too low
20English
Make sure that the display format is
set for the correct measurement
range
Make sure that the sensor is
configured for the correct cell
constant.
Make sure that the sensor is held in
air during zero calibration and is not
located near radio frequency or
electromagnetic interference. Make
sure that the cable is shielded by
metal conduit.
Table 4 Warning list for conductivity sensors (continued)
WarningDescriptionResolution
TEMP TOO HIGHThe measured temperature is >
200 °C
TEMP TOO LOWThe measured temperature is <
-20 °C
CAL OVERDUEThe Cal Reminder time has expired Calibrate the sensor.
NOT CALIBRATEDThe sensor has not been calibrated Calibrate the sensor.
REPLACE SENSORThe sensor has been in operation >
365 days
CAL IN PROGRESSA calibration was started but not
completed
OUTPUTS ON HOLDDuring calibration, the outputs were
set to hold for a selected time.
WRONG LINEAR TCThe user-defined linear temperature
compensation is out of range
WRONG TC TABLEThe user-defined temperature
compensation table is out of range
WRNG USER CONC TABLEThe concentration measurement is
outside of the range of the user
table
WRNG BLT-IN TEMP TABLEThe measured temperature is
outside of the range of the built-in
temperature compensation table
WRNG BLT-IN CONC TABLEThe concentration measurement is
outside of the range of the built-in
concentration table
Make sure that the sensor is
configured for the correct
temperature element.
Calibrate the sensor with a
reference solution and reset the
sensor days. Refer to Sensor
diagnostic and test menu
on page 19. If the calibration fails,
call technical support.
Return to calibration.
The outputs will become active after
the selected time period.
The value must be between 0 and
4%/°C; 0 to 200 °C.
The temperature is above or below
the temperature range defined by
the table.
Make sure that the user table is set
for the correct measurement range.
Make sure that the temperature
compensation is configured
correctly.
Make sure that the concentration
measurement is configured for the
correct chemical and range.
English 21
Spezifikationen
Die Spezifikationen können ohne Vorankündigung Änderungen unterliegen.
Der Hersteller ist nicht verantwortlich für direkte, indirekte, versehentliche oder Folgeschäden, die
aus Fehlern oder Unterlassungen in diesem Handbuch entstanden. Der Hersteller behält sich
jederzeit und ohne vorherige Ankündigung oder Verpflichtung das Recht auf Verbesserungen an
diesem Handbuch und den hierin beschriebenen Produkten vor. Überarbeitete Ausgaben der
Bedienungsanleitung sind auf der Hersteller-Webseite erhältlich.
Sicherheitshinweise
H I N W E I S
Der Hersteller ist nicht für Schäden verantwortlich, die durch Fehlanwendung oder Missbrauch dieses Produkts
entstehen, einschließlich, aber ohne Beschränkung auf direkte, zufällige oder Folgeschäden, und lehnt jegliche
Haftung im gesetzlich zulässigen Umfang ab. Der Benutzer ist selbst dafür verantwortlich, schwerwiegende
Anwendungsrisiken zu erkennen und erforderliche Maßnahmen durchzuführen, um die Prozesse im Fall von
möglichen Gerätefehlern zu schützen.
Bitte lesen Sie dieses Handbuch komplett durch, bevor Sie dieses Gerät auspacken, aufstellen oder
bedienen. Beachten Sie alle Gefahren- und Warnhinweise. Nichtbeachtung kann zu schweren
Verletzungen des Bedieners oder Schäden am Gerät führen.
Stellen Sie sicher, dass die durch dieses Messgerät bereitgestellte Sicherheit nicht beeinträchtigt
wird. Verwenden bzw. installieren Sie das Messsystem nur wie in diesem Handbuch beschrieben.
Bedeutung von Gefahrenhinweisen
Kennzeichnet eine mögliche oder drohende Gefahrensituation, die, wenn sie nicht vermieden wird, zum Tod oder
zu schweren Verletzungen führt.
Kennzeichnet eine mögliche oder drohende Gefahrensituation, die, wenn sie nicht vermieden wird, zum Tod oder
zu schweren Verletzungen führen kann.
G E F A H R
W A R N U N G
Deutsch 23
V O R S I C H T
Kennzeichnet eine mögliche Gefahrensituation, die zu geringeren oder moderaten Verletzungen führen kann.
Kennzeichnet eine Situation, die, wenn sie nicht vermieden wird, das Gerät beschädigen kann. Informationen, die
besonders beachtet werden müssen.
H I N W E I S
Aufkleber mit Vorsichtshinweisen
Bitte lesen Sie alle auf dem Produkt angebrachten Etiketten und Hinweise. Die Nichtbeachtung kann
zu Verletzungen an Personen oder einer Beschädigung des Produkts führen. Im Handbuch werden
auf die am Gerät angebrachten Symbole in Form von Warnhinweisen verwiesen.
Dieses Symbol auf einem Produkt zeigt eine potenzielle Gefahr an, die zu ernsthaften Verletzungen
und/oder zum Tod führen kann. Der Benutzer soll dieses Handbuch bei der Bedienung des Geräts
und/oder für Sicherheitsinformationen verwenden.
Dieses Symbol auf einer Verkleidung oder Schranke des Produkts weist auf die Gefahr von
Stromschlägen hin und macht darauf aufmerksam, dass ausschließlich für die Arbeit mit
gefährlichen Spannungen qualifiziertes Personal die Verkleidung öffnen oder die Schranke
entfernen darf.
Dieses Symbol auf dem Produkt weist auf das Vorhandensein von Bauteilen hin, die durch
elektrostatische Entladungen gestört werden können und macht darauf aufmerksam, dass mit
Vorsicht vorgegangen werden muss, um Schäden an diesen Bauteilen zu vermeiden.
Elektrische Geräte, die dieses Symbol aufweisen, dürfen in Europa nicht als Haushaltsabfall
entsorgt werden. Den lokalen und nationalen europäischen Bestimmungen gemäß müssen
Benutzer von Elektrogeräten diese nun zur für den Benutzer kostenlosen Entsorgung an den
Hersteller zurückgeben.
Hinweis: Mit der Wiederverwertung, der stofflichen Verwertung oder anderen Formen der Verwertung von
Altgeräten leisten Sie einen wichtigen Beitrag zum Schutz unserer Umwelt.
Produkte, die mit diesem Symbol gekennzeichnet sind, enthalten toxische oder gefährliche
Substanzen oder Elemente. Die Ziffer in diesem Symbol gibt den Umweltschutzzeitraum in Jahren
an.
Produktübersicht
Nach seiner Installation in einen sc200 oder Polymetron 9500 Controller ermöglicht das Modul den
Anschluss eines analogen Sensors an den Controller.
Installation
W A R N U N G
Mehrere Gefahren. Die in diesem Abschnitt des Handbuchs beschriebenen Eingriffe dürfen ausschließlich von
qualifiziertem Personal vorgenommen werden.
Sensor an das Instrument anschließen
Potenzielle Stromschlaggefahr! Stellen Sie stets die Spannungsversorgung am Controller ab, wenn elektrische
Anschlüsse durchgeführt werden.
Gefahr von Stromschlägen. Die Hochspannungsleitungen für den Controller verlaufen hinter der
Hochspannungssperre im Controllergehäuse. Die Schutzabdeckung darf nicht entfernt werden, außer beim
Anschluss von Spannungsversorgung, Relais, oder Analog- und Netzwerkkarten durch einen qualifizierten
Techniker.
24 Deutsch
W A R N U N G
W A R N U N G
H I N W E I S
Möglicher Geräteschaden. Empfindliche interne elektronische Bauteile können durch statische Elektrizität
beschädigt werden, wobei dann das Gerät mit verminderter Leistung funktioniert oder schließlich ganz ausfällt.
Um das Modul einzubauen und den Sensor anzuschließen, siehe nachfolgende Zeichnungen und
Anschlusstabelle.
Stellen Sie sicher, alle Erdleiter/Beilauflitzen des Sensors mit den Erdungsschrauben des
Controllergehäuses verbunden zu haben.
Hinweis: Wenn das Sensorkabel nicht bis zum Controller reicht, ist ein Verbindungskabel (Modell 150727,10000)
mit Anschlusskasten (Modell 08335=A=6000) erforderlich, um die Entfernung zu überbrücken.
Tabelle 1 Verkabelung des Leitfähigkeitssensors
KontaktBeschreibungSensor
8398 831x und 8394
1 Innere ElektrodeWeiß (weißes Kabel)Weiß
2 Innere ElektrodeSchwarz (weißes Kabel)—
3 Masse—Grün
4 ———
5 ———
6 ———
7 Temp -GrünSchwarz
8 ———
9 ———
10 Temp +GelbGelb
11 Äußere ElektrodeBraun (braunes Kabel)Rot
12 Äußere ElektrodeSchwarz (braunes Kabel)—
In Schritt 3 der folgenden Sequenz stellen Sie den Schalter wie folgt ein:
SchaltereinstellungBeschreibung
Einstellung 1 Kontakt-Leitfähigkeitssensor
Einstellung 2 Induktiver Leitfähigkeitssensor
Deutsch 25
26 Deutsch
Deutsch 27
Betrieb
Benutzernavigation
Siehe Abschnitt Benutzerschnittstelle und Navigation der Anleitung zum Controller bzgl. der
Tastaturbeschreibung und Navigationsinformationen.
Sensorkonfigurierung
Kontakt-Leitfähigkeitssensor konfigurieren
Verwenden Sie das Konfigurationsmenü CONFIGURE, um eine Identifizierungsinformation für den
Sensor einzugeben und die Optionen für die Datenverwaltung und -speicherung zu ändern.
1. Die Menü-Taste drücken und SENSOR SETUP>[Select Sensor]>CONFIGURE
2. Wählen Sie eine Option aus und drücken Sie dann ENTER. Zur Eingabe der Zahlen, Zeichen
oder Satzzeichen die Pfeiltasten nach oben oder nach unten drücken und halten. Mit der
rechten Pfeiltaste zum nächsten Feld gehen.
OptionBeschreibung
EDIT NAMEDefiniert den dem Sensor zugewiesenen Namen, der oben in der Messanzeige
SENSOR-S/NErmöglicht dem Benutzer die Eingabe einer Sensor-Seriennummer. Die Nummer kann
SELECT MEASURE Stellt für den Messparameter die Optionen CONDUCTIVITY (Leitfähigkeit = Standard),
DISPLAY FORMAT Ändert die Anzahl der Dezimalstellen, die auf der Messanzeige dargestellt werden. Bei
MEAS UNITSÄndert die Einheiten für die gewählte Messung - wählen Sie eine Einheit aus der
TEMP UNITSLegt für die Temperatureinheit die Einstellung °C (Standard) oder °F fest.
T-COMPENSATION Addiert eine temperaturabhängige Korrektur zu dem gemessenen Wert.
erscheint. Der Name kann maximal 16 Zeichen bestehen und Buchstaben, Zahlen,
Leerzeichen und Satzzeichen enthalten. Nur die ersten 12 Zeichen werden auf dem
Controller angezeigt.
maximal aus 16 Zeichen bestehen und Buchstaben, Zahlen, Leerzeichen und
Satzzeichen enthalten.
TDS (vollständig gelöste Feststoffe), SALINITY (Salzgehalt) oder RESISTIVITY
(Resistivität) ein. Alle anderen konfigurierten Einstellungen werden auf die
Standardwerte zurückgestellt.
Hinweis: Bei der Auswahl der Option SALINITY (Salzgehalt) wird für die Messeinheit
automatisch ppt (parts per thousand) eingestellt. Diese Einstellung kann nicht
verändert werden.
der Einstellung "Auto" wird die Anzahl der Dezimalstellen automatisch mit den
Änderungen der gemessenen Werte geändert.
vorhandenen Liste aus.
• NONE—Temperaturkompensation nicht erforderlich
• USP—Einstellen des Alarmniveaus für die USP-Standarddefinitionstabelle
• ULTRA PURE WATER (Ultrareines Wasser) - Nicht für TDS verfügbar. Stellt die
Kompensierungsart in Abhängigkeit von den Probeneigenschaften ein—Wählen Sie
NaCl , HCl, AMMONIA oder ULTRA PURE WATER
• USER—BUILT IN LINEAR, LINEAR oder TEMP TABLE wählen:
• BUILT IN LINEAR—Die vordefinierte lineare Wertetabelle verwenden (Steigung
2,0%/°C, Referenztemperatur 25 °C)
• LINEAR—Einstellen der Steigungs- und Referenztemperaturparameter weicht
von den integrierten Parametern ab
• TEMP TABLE—Definiert die Temperatur und die Punkte für die
Multipliaktionsfaktoren (siehe Benutzertabellen konfigurieren auf Seite 30).
• NATURAL WATER—Nicht für TDS verfügbar
28 Deutsch
OptionBeschreibung
CONFIG TDSNur TDS—ändert den Faktor, der zur Umwandlung der Leitfähigkeit zu TDS verwendet
CABLE PARAMEinstellen der Sensorkabelparameter, um die Messgenauigkeit zu verbessern, wenn
TEMP ELEMENTStellt für das Temperaturelement PT100 oder PT1000 für die automatische
FILTERDefiniert eine Zeitkonstante zur Verbesserung der Signalstabilität. Die Zeitkonstante
LOG SETUPLegt das Zeitintervall für die Datenspeicherung im Datenprotokoll fest— 5 oder
RESET DEFAULTS Setzt das Konfigurationsmenü auf die Standardeinstellungen zurück. Alle
wird: NaCl (0,49 ppm/µS) oder CUSTOM (Faktor zwischen 0,01 und 99,99 ppm/µS
eingeben).
das Sensorkabel über die 5 m Standardlänge hinaus verlängert oder gekürzt wird.
Geben Sie die Kabellänge, den Widerstand und die Leitfähigkeit ein.
Temperaturkompensation ein. Wenn kein Element verwendet wird, kann hier die
Einstellung MANUAL eingestellt werden, um einen Wert für die
Temperaturkompensation einzugegeben.
berechnet den Durchschnittswert innerhalb eines festgelegten Zeitraums -0 (No Effect)
bis 60 Sekunden (Durchschnitt des Signalwerts für 60 Sekunden). Der Filter erhöht die
Ansprechzeit des Sensorsignals auf aktuelle Prozessänderungen.
30 Sekunden und 1, 2, 5, 10, 15 (Standard), 30 oder 60 Minuten.
Sensorinformationen gehen verloren.
Induktiven Leitfähfigkeitssensor konfigurieren
Verwenden Sie das Konfigurationsmenü CONFIGURE, um eine Identifizierungsinformation für den
Sensor einzugeben und die Optionen für die Datenverwaltung und -speicherung zu ändern.
1. Die Menü-Taste drücken und SENSOR SETUP>[Select Sensor]>CONFIGURE
2. Wählen Sie eine Option aus und drücken Sie dann ENTER. Zur Eingabe der Zahlen, Zeichen
oder Satzzeichen die Pfeiltasten nach oben oder nach unten drücken und halten. Mit der
rechten Pfeiltaste zum nächsten Feld gehen.
OptionBeschreibung
EDIT NAMEDefiniert den dem Sensor zugewiesenen Namen, der oben in der Messanzeige
SENSOR-S/NErmöglicht dem Benutzer die Eingabe einer Sensor-Seriennummer. Die Nummer kann
SELECT MEASURE Stellt für den Messparameter die Optionen CONDUCTIVITY (Leitfähigkeit = Standard),
DISPLAY FORMAT Ändert die Anzahl der Dezimalstellen, die auf der Messanzeige dargestellt werden. Bei
MEAS UNITSÄndert die Einheiten für die gewählte Messung - wählen Sie eine Einheit aus der
TEMP UNITSLegt für die Temperatureinheit die Einstellung °C (Standard) oder °F fest.
erscheint. Der Name kann maximal 16 Zeichen bestehen und Buchstaben, Zahlen,
Leerzeichen und Satzzeichen enthalten. Nur die ersten 12 Zeichen werden auf dem
Controller angezeigt.
maximal aus 16 Zeichen bestehen und Buchstaben, Zahlen, Leerzeichen und
Satzzeichen enthalten.
RESISTIVITY (Resistivität), SALINITY (Salzgehalt), TDS (vollständig gelöste
Feststoffe) oder CONCENTRATION (Konzentration) ein. Alle anderen konfigurierten
Einstellungen werden auf die Standardwerte zurückgestellt.
Hinweis: Bei der Auswahl der Option SALINITY (Salzgehalt) wird für die Messeinheit
automatisch ppt (parts per thousand) eingestellt. Diese Einstellung kann nicht
verändert werden.
der Einstellung "Auto" wird die Anzahl der Dezimalstellen automatisch mit den
Änderungen der gemessenen Werte geändert.
vorhandenen Liste aus.
Deutsch 29
OptionBeschreibung
T-COMPENSATION Addiert eine temperaturabhängige Korrektur zu dem gemessenen Wert.
• LINEAR—Einstellen der Steigungs- und Referenztemperaturparameter
• NATURAL WATER—Nicht für TDS oder CONCENTRATION verfügbar
• TEMP TABLE—Definiert die Temperatur und die Punkte für die
Multipliaktionsfaktoren (siehe Benutzertabellen konfigurieren auf Seite 30).
• NONE—Temperaturkompensation nicht erforderlich
CONFIG TDSNur TDS—ändert den Faktor, der zur Umwandlung der Leitfähigkeit zu TDS verwendet
CONFIG CONCNur CONCENTRATION —Definiert die zu verwendende Konzentrationstabelle.
TEMP ELEMENTStellt für das Temperaturelement PT100 oder PT1000 für die automatische
FILTERDefiniert eine Zeitkonstante zur Verbesserung der Signalstabilität. Die Zeitkonstante
LOG SETUPLegt das Zeitintervall für die Datenspeicherung im Datenprotokoll fest— 5 oder
RESET DEFAULTS Setzt das Konfigurationsmenü auf die Standardeinstellungen zurück. Alle
wird: NaCl (0,49 ppm/µS) oder CUSTOM (Faktor zwischen 0,01 und 99,99 ppm/µS
eingeben).
• BUILT-IN—Definiert die CHEMICAL (Chemikalie), die gemessen wird:
• H2SO4 0-30%
• HNO3 0-28%
• HCl 0-18%
• NaOH 0-16%
Hinweis: Auch wenn hier ein Bereich von 0-16% angezeigt wird, können die
Tabellen auch genauer für 0-10% für alle Temperaturen definiert werden.
• NaCl 0-25%
• H3PO4 0-40%
• USER TABLE—Definiert die Punkte für die Leitfähigkeit und die Konzentration
(siehe Benutzertabellen konfigurieren auf Seite 30).
Temperaturkompensation ein. Wenn kein Element verwendet wird, kann hier die
Einstellung MANUAL eingestellt werden, um einen Wert für die
Temperaturkompensation einzugegeben.
berechnet den Durchschnittswert innerhalb eines festgelegten Zeitraums -0 (No Effect)
bis 60 Sekunden (Durchschnitt des Signalwerts für 60 Sekunden). Der Filter erhöht die
Ansprechzeit des Sensorsignals auf aktuelle Prozessänderungen.
30 Sekunden und 1, 2, 5, 10, 15 (Standard), 30 oder 60 Minuten.
Sensorinformationen gehen verloren.
Benutzertabellen konfigurieren
Je nach erforderlichem Sensor- und Messungstyp können während der Sensorkonfigurierung
benutzerdefinierte Tabellen ausgewählt werden. Im Folgenden wird erklärt , wie eine
benutzerdefinierte Tabelle konfiguriert wird.
Konzentrationstabellen
Standardmäßig enthält diese Tabelle nur zwei Punkte (P1 und P2), die den oberen und unteren
Grenzwert definieren. Nachdem P1 und P2 definiert wurden, können zusätzliche Punkte im
Zwischenbereich eingegeben werden. Um z. B. für die Leitfähigkeit und die Konzentration den
Mindestwert null und für die Leitfähigkeit den Höchstwert 750 bei einer maximalen Konzentration von
30 zu definieren, verfahren Sie wie folgt:
1. Wählen Sie in der Option USER TABLE (Benutzerdefinierte Tabelle) die Funktion P1>EDIT
POINT (Punkt bearbeiten).
2. Geben Sie einen Mindestleitfähigkeitswert (X) in dem Format 0.0 ein und bestätigen Sie mit
Enter.
3. Geben Sie einen Mindestkonzentrationswert in Prozent (Y) in dem Format 0.0 ein und bestätigenSie mit Enter.
30
Deutsch
Loading...
+ 198 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.