Specifications on page 3Operation on page 11
General information on page 4Maintenance on page 25
Startup on page 7Troubleshooting on page 26
User interface and navigation
on page 9
Replacement parts on page 27
Specifications
Specifications are subject to change without notice.
SpecificationDetails
Dimensions (W x D x H) 6.1 x 3.2 x 15.2 cm (2.4 x 1.25 x 6 in.)
EnclosureIP67, waterproof at 1 m (3.3 ft) for 30 minutes (battery
Light sourceLight emitting diode (LED)
DetectorSilicon photodiode
DisplayLCD with backlight
Weight0.2 kg (0.43 lb)
Pollution degree2
Installation categoryI
Protection class3
Power requirements4 AAA batteries; approximate life of 2000 tests (use of
Operating environment 0 to 50 °C (32 to 122 °F), 0 to 90% relative humidity
Storage temperature–20 to 55 °C (–7.6 to 131 °F)
Photometric precision± 0.0015 Abs
WavelengthFixed wavelength ±2 nm, different for each model
Filter bandwidth15 nm
Absorbance range0 to 2.5 Abs
compartment not included). Keep out of direct
sunlight.
backlight decreases this number)
Rechargeable batteries are not recommended.
non-condensing
English 3
SpecificationDetails
Sample cell path length 1 cm (5–10 mL), 25 mm (10 mL)
Data storageLast 10 measurements
CertificationsCE mark
Warranty2 years
General information
In no event will the manufacturer be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or
omission in this manual. The manufacturer reserves the right to make
changes in this manual and the products it describes at any time,
without notice or obligation. Revised editions are found on the
manufacturer’s website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or
misuse of this product including, without limitation, direct, incidental and
consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical
application risks and install appropriate mechanisms to protect processes
during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or
operating this equipment. Pay attention to all danger and caution
statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not
impaired. Do not use or install this equipment in any manner other than
that specified in this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided,
will result in death or serious injury.
4 English
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided,
could result in death or serious injury.
Indicates a potentially hazardous situation that may result in minor or moderate
injury.
Indicates a situation which, if not avoided, may cause damage to the
instrument. Information that requires special emphasis.
C A U T I O N
N O T I C E
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed. A symbol on the
instrument is referenced in the manual with a precautionary statement.
This symbol, if noted on the instrument, references the instruction
manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed
of in European domestic or public disposal systems. Return old or
end-of-life equipment to the manufacturer for disposal at no charge
to the user.
Certification
Canadian Radio Interference-Causing Equipment Regulation,
IECS-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian
Interference-Causing Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de
la réglementation canadienne sur les équipements provoquant des
interférences.
FCC Part 15, Class "A" Limits
English5
Supporting test records reside with the manufacturer. The device
complies with Part 15 of the FCC Rules. Operation is subject to the
following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including
interference that may cause undesired operation.
Changes or modifications to this equipment not expressly approved by
the party responsible for compliance could void the user's authority to
operate the equipment. This equipment has been tested and found to
comply with the limits for a Class A digital device, pursuant to Part
15 of the FCC rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated
in a commercial environment. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference, in which case
the user will be required to correct the interference at their expense.
The following techniques can be used to reduce interference problems:
1. Move the equipment away from the device receiving the
interference.
2. Reposition the receiving antenna for the device receiving the
interference.
3. Try combinations of the above.
Product overview
The single wavelength Pocket Colorimeter II instruments are portable
filter photometers used for testing water, treated waters, wastewater,
estuary and seawater. Refer to Figure 1. The single wavelength
models are configured at the factory to measure at a specific
wavelength.
The single wavelength models have two channels in which
measurements can be made. Until a user-prepared calibration curve is
entered, the single wavelength instruments show only a direct readout
of absorbance. To measure concentration, enter a user-prepared
calibration curve. Refer to User-entered calibration on page 21.
6 English
Figure 1 Instrument overview
1 Instrument cap3 Display
2 Cell holder4 Keypad
Startup
Install the batteries
W A R N I N G
Explosion hazard. Incorrect battery installation can cause the release
of explosive gases. Be sure that the batteries are of the same
approved chemical type and are inserted in the correct orientation.
Do not mix new and used batteries.
Install the batteries as shown in Figure 2.
English 7
Figure 2 Install the batteries
8 English
Install the cap cord
Attach the cap cord to prevent loss of the instrument cap. Refer to
Figure 3.
Figure 3 Install the cap cord
User interface and navigation
Keypad description
Figure 4 shows the keypad and gives the key functions.
English 9
Figure 4 Keypad
1 Power/Backlight key: Sets the
power to on and off. Push and hold
for 1 second to set the backlight to
on or off.
2 Zero/Scroll key: Sets the
instrument to zero, scrolls through
menu options and numbers
3 Menu key: Enters and goes out of
menu mode.
4 Read/Enter key: Starts a sample
measurement, selects a menu
option, moves the cursor to the
next digit
Display description
Figure 5 shows the values and icons shown on the display.
10 English
Figure 5 Display
1 Numeric display: Measured value
or menu options
2 Range icon: Selected range or
parameter
3 Range value: Range(s) or
parameters
4 Menu icon: The instrument is in
menu mode.
5 Calibration adjusted icon: A user-
entered calibration curve was
entered.
6 Low battery icon: Battery level is
10%. Flashes when the battery
level is too low to complete
measurements.
Operation
Configure the instrument
1. Push .
2. Push to scroll through the menu options. Push to select an
option.
Option Description
SELSets the measurement range or parameter. Push to toggle
between the measurement ranges or parameters.
00:00 Sets the time in 24-hour format (hh:mm). Push to change the
time. Push to change the first digit, then to go to the next
digit.
English 11
Option Description
rCLShows the last 10 measurements recorded. Push to show the
recorded measurements (01—most recent measurement, 10—
oldest measurement). Push to scroll through the measurements.
To select a measurement by number, push to select the number
and then . Push to go out of this option.
SCANot applicable to the single wavelength models.
3. Push to go back to measurement mode.
Measurement
Basic colorimetry
Colorimetry measures the amount of color in a clear medium, such as
a liquid, to identify the quantity of a particular substance (the analyte) in
the liquid. Typically, the concentration of the analyte is proportional to
the intensity of the color in the clear medium (solution). In most
methods, a darker color indicates a higher analyte concentration.
Absorbance (Abs) at a specific wavelength is typically used to measure
the amount of light absorbed by the solution. Absorbance (Abs) is
calculated as:
Abs = –log T or Abs = –log (IT/IO)
Where:
T = transmittance
IT = intensity of the light transmitted through the sample
IO = intensity of the light that enters the sample
Some substances, such as dyes and different metal ions, have
inherent color and can be measured without any additions. In most
cases, a chemical reaction between an indicator and the analyte is
necessary to get a colored product that can be measured.
Once the relationship between the amount of color (measured as
absorbance) and a known concentration of a sample is identified, the
instrument can be used to measure concentrations of unknown
samples. A user-entered calibration curve is used to measure the
sample concentration.
To identify the amount of color in a sample, the instrument measures
the amount of light the solution absorbs. The absorption of light is
dependent on the wavelength of the light and the color of the solution.
12English
The combination of an LED light source and an interference filter sets
the measurement wavelength.
Select the best wavelength
The single wavelength instruments each have a different LED and
interference filter to measure at a specific wavelength.
The wavelength (color) of light used is typically selected so that it has a
maximum absorption, but other wavelengths can be selected to
minimize interferences or other factors. For the best results, select the
instrument wavelength with knowledge about the absorbance spectra
of the species of interest, as well as the spectra of other colored
species that could be in the sample. Figure 6 shows a typical
absorption spectrum.
Refer to Table 1 to select the best instrument wavelengths to use for
testing. Do not use this table for samples that have more than one
absorption region that adds to the visible color. For example, a green
solution can have a yellow and a blue absorption peak. One or the
other peak can be used for measurements if both have different
analyte concentration. Other samples can look brown because there
are several spectra that add to the visible color.
English 13
Figure 6 Select the best wavelength – sample spectrum
The measurement range of the instrument is 0 to approximately
1.50 Abs, but can be used up to a measurement range of 2.5 Abs if the
chemistry method supports that range.
If sample absorbances are more than 1.50 Abs:
1. Dilute the sample or use smaller sample cells for the best linearity
and accuracy.
2. If a smaller sample cell such as the 1-cm (10-mL) cell is used,
complete the calibration with the smaller sample cells.
Note: Absorbance increases with the increase of the sample cell pathlength.
Use a sample cell with a shorter pathlength to measure solutions with a
darker colored.
3. Monitor the calibration curve to identify the measurement range for
a specific test.
The measurement range is the concentration range in which the
deviation from linearity is within acceptable limits.
Calibration curve
Calibration curves should ideally intersect the zero intercept for
absorbance. The zero intercept is the zero concentration point on the
calibration graph. When there is no analyte in the sample, the
absorbance will be zero.
A non-zero intercept (a positive or negative absorbance measurement
at zero concentration) can occur for many reasons. Factors that can
cause a non-zero intercept include the reagent blank, pH, temperature,
interfering species or turbidity differences between the zeroing solution
(blank) and the sample.
To adjust for a non-zero intercept caused by the reagent blank,
measure the absorbance of the prepared reagent blank and then
subtract it from the measured absorbance of the prepared sample. In
English15
an aqueous sample, add the reagents to deionized water to prepare
the reagent blank. The prepared reagent blank includes only the
amount of color that is added to the deionized water by the reagent and
not the analyte. The prepared sample includes the amount of color that
is added by the reagent and the analyte.
For some chemistries, the intensity of the color decreases as the
analyte concentration increases. These chemistries are referred to as
bleaching chemistries because the measured sample is lighter in color
than the reagent blank that was used to zero the instrument. This
instrument is able to measure bleaching (or negative) absorbance
chemistries directly. Set the instrument zero with the reagent blank (the
most highly colored solution) and then read the sample or bleached
color directly.
Single wavelength procedure
Before starting
Always measure solutions in sample cells or AccuVac® Ampules. Do not put
the instrument in the sample or pour the sample into the cell holder.
Make sure that the sample cells are clean and there are no scratches where
the light passes through them.
Make sure that there are no fingerprints or liquid on the external surface of the
sample cells or AccuVac® Ampules. Wipe with a lint-free cloth.
Rinse the sample cell and cap with the sample three times before the sample
cell is filled.
Always insert the sample cell in the correct and consistent orientation so that
the results are more repeatable and precise. Refer to Figure 7.
Install the instrument cap over the cell holder before ZERO or READ is
pushed. Refer to Figure 8.
Measure the volume of the liquid reagent accurately. Use a pipet if possible.
If the test result is over-range, dilute a fresh sample with a known volume of
deionized water and repeat the test. Multiply the result by the dilution factor.
When the test is completed, immediately empty and rinse the prepared sample
cell. Rinse the sample cell and cap three times.
16English
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used.
Use the recommended personal protective equipment.
Dispose of reacted solutions according to local, state and federal regulations.
Refer to the Safety Data Sheets for disposal information for unused reagents.
Refer to the environmental, health and safety staff for your facility and/or local
regulatory agencies for further disposal information.
Figure 7 Sample cell orientation
1 Orientation mark2 Sample cell, 25-mm (10 mL)
English 17
Figure 8 Install the instrument cap over the cell holder
Sample collection
• Collect samples in clean glass or plastic bottles.
• Rinse the sample bottle several times with the sample to be
collected.
• Analyze the samples as soon as possible for best results.
• Homogenize samples that contain solids to get a representative
sample.
18English
• Filter samples that are turbid with filter paper and a funnel.
Reagent solution procedure
1. Select the range that
has a saved user
calibration. Refer to
Configure the
instrument on page 11.
Note: To enter a user
calibration, refer to
User-entered calibration
on page 21.
4. Insert the blank into
the cell holder in the
correct orientation.
Refer to Figure 7
on page 17.
2. Prepare the blank:
Fill the sample cell with
10 mL of the blank
solution (typically
sample).
5. Install the instrument
cap over the cell holder.
3. Clean the blank
sample cell.
6. Push ZERO. The
display shows “0.000”,
or the degree of
resolution that was
previously selected.
English 19
7. Remove the sample
cell from the cell holder.
8. Prepare the
sample: Fill a second
sample cell with 10 mL
of sample.
9. Add the reagent to
the second sample cell.
Wait the specified
reaction time for full
color development if
applicable.
10. Clean the prepared
sample cell.
20 English
11. Insert the prepared
sample into the cell
holder in the correct
orientation. Refer to
Figure 7 on page 17.
12. Install the
instrument cap over the
cell holder.
13. Push READ. The
display shows the
measurement results.
Show the recorded measurements
Refer to the "rCL" option in Configure the instrument on page 11.
User-entered calibration
This instrument accepts a user-prepared calibration curve. The
calibration curve can be from 0 to 2.5 absorbance. Make sure that the
calibration curve includes standard values that are less and more than
the range of interest.
The instrument range will be the same as the calibration range. For
example, when the standards that are used are 1.00, 2.00 and 4.00.
The instrument range is 1.00 to 4.00.
There are two options to enter a user calibration curve:
• Enter a calibration curve with standards—The standard solution
values are entered with the keypad and the absorbance values are
measured.
• Enter a calibration curve with the keypad—The standard solution
values and absorbance values are entered with the keypad.
Note: If the instrument is set to off or the instrument power is removed before a
user-entered calibration curve is completed, the calibration curve is not saved.
The instrument automatically switches off in user-entered calibration entry mode
after 60 minutes of no activity. User-entered calibrations are completed when the
user goes out of calibration (cal) mode or edit mode.
English 21
Enter a calibration curve with standards
Note: Deionized water can be used for the blank unless the sample is
significantly more turbid or has more color than deionized water.
1. Set the instrument to the range to calibrate. Refer to Configure the
instrument on page 11.
2. Prepare the blank and the reacted standard solution. Refer to the
test procedure. Let the color fully develop.
3. Set the instrument to zero.
a. Insert the blank sample cell in the cell holder.
b. Install the instrument cap over the cell holder.
c. Push . The display shows “- - - -”, then “0.000”.
d. Remove the instrument cap.
e. Remove the sample cell from the cell holder.
4. Push and hold until "USER" and then "CAL" shows, then push
.
Note: If "USER" and "CAL" do not show, the factory calibration cannot be
changed on the selected range.
5. When "RES" shows on the display, set the resolution.
a. Push . The resolution setting (decimal placement) shows.
b. To change the resolution, push , then . Push to save the
change.
c. To not change the resolution, push .
6. When "S0" shows on the display, push . Push to enter the
blank value, then push .
Note: Push to go to the next digit.
7. When “A0” shows on the display, measure the absorbance of the
blank.
a. Insert the blank sample cell in the cell holder.
b. Install the instrument cap over the cell holder.
c. Push . The display shows the absorbance value for "S0".
d. Remove the sample cell from the cell holder.
8. Push to show "S1".
22 English
9. When "S1" shows on the display, push . Push to enter the first
standard value, then push .
Note: Push to enter the next digit.
10. When "A1" shows on the display, measure the absorbance of the
reacted standard solution.
a. Insert the reacted standard sample cell in the cell holder.
b. Install the instrument cap over the cell holder.
c. Push . The display shows the absorbance value for "S1".
d. Remove the sample cell from the cell holder.
11. The calibration is completed with two calibration points. If additional
standards are necessary for calibration:
a. Push until “Add” shows, then push .
b. Do steps 9–10 again to enter more standards.
12. Push two times to go back to measurement mode.
Enter a calibration curve with the keypad
At least two data pairs are necessary to enter a user-prepared
calibration curve. A concentration value and the absorbance value for
the given concentration is necessary for each data pair. A maximum of
10 data pairs can be entered.
Note: This procedure can also be used to change the data pairs in a userentered calibration curve.
1. Set the instrument to the range to calibrate. Refer to Configure the
instrument on page 11.
2. Push and hold until "USER" and then "CAL" shows, then push
.
Note: If "USER" and "CAL" do not show, the factory calibration cannot be
changed on the selected range.
3. Push until "EDIT" shows, then push .
4. When "RES" shows on the display, set the resolution.
a. Push . The resolution setting (decimal placement) shows.
b. To change the resolution, push , then . Push to save the
change.
c. To not change the resolution, push .
English 23
5. When "S0" shows on the display, push . Push to enter the
concentration value of the first data pair, then push .
Note: Push to go to the next digit.
6. When "A0" shows on the display, push . Push to enter the
absorbance value of the first data pair, then push . "S1" shows on
the display.
7. Do steps 5–6 again to enter the second data pair (S1 and A1).
8. The calibration is completed with two data pairs. If additional data
pairs are necessary for calibration:
a. When “Add” shows, push .
b. Do steps 5–6 again to enter more data pairs.
9. Push two times to go back to measurement mode.
Remove a calibration point
To remove a calibration point from a user-entered calibration curve:
1. Set the instrument to the range to calibrate. Refer to Configure the
instrument on page 11.
2. Push and hold until "USER" and then "CAL" shows.
Note: If "USER" and "CAL" do not show, the factory calibration cannot be
changed on the selected range.
3. Push until "EDIT" shows, then push .
Note: Calibration points can also be removed in calibration (CAL) mode.
4. Push until the calibration point to remove shows (i.e., S0 or S1),
then push .
5. Push until "dEL" shows, then push .
Note: The minimum number of data pairs is two. When only two data pairs
remain, no more data pairs can be removed.
6. Push two times to go back to measurement mode.
24 English
Remove the calibration curve
1. Set the instrument to the applicable range. Refer to Configure the
instrument on page 11.
2. Push and hold until "USER" and then "CAL" shows.
Note: If "USER" and "CAL" do not show, the factory calibration cannot be
changed on the selected range.
3. Push until "dFL" shows, then push .
Maintenance
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.
N O T I C E
Do not disassemble the instrument for maintenance. If the internal components
must be cleaned or repaired, contact the manufacturer.
Clean the instrument
Clean the exterior of the instrument with a moist cloth and a mild soap
solution and then wipe the instrument dry.
Clean the sample cells
Chemical exposure hazard. Obey laboratory safety procedures and
wear all of the personal protective equipment appropriate to the
chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
C A U T I O N
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in
accordance with local, regional and national regulations.
English25
Most laboratory detergents are used at recommended concentrations.
Neutral detergents, such as Liquinox, are safer to use when regular
cleaning is necessary. To decrease the cleaning times, increase the
temperature or use an ultrasonic bath. To complete the cleaning, rinse
a few times with deionized water and then let the sample cell air dry.
Sample cells may also be cleaned with acid, followed by a thorough
rinse with deionized water.
Note: Always use acid to clean sample cells that were used for low-level metal
tests.
Special cleaning methods are necessary for individual procedures.
When a brush is used to clean sample cells, take extra care to avoid
scratches on the interior surfaces of the sample cells.
Replace the batteries
Replace the batteries when the battery power level is low. Refer to
Install the batteries on page 7.
Troubleshooting
ErrorDescriptionSolution
E-0 No zeroIn user calibration mode, a standard solution
E-1 Ambient light error1There is ambient light in the cell holder. Make
E-2 LED error
E-6 Abs errorThe absorbance value is not correct or the
1
was measured before the instrument zero
was set. Measure a blank solution to set the
instrument to zero.
sure that the instrument cap is fully installed
over the cell holder.
The LED (light source) is out of regulation.
Replace the batteries. Make sure that the
LED in the cell holder comes on when or
is pushed.
user-entered calibration curve has fewer than
two points. Enter or measure the absorbance
value again.
26 English
ErrorDescriptionSolution
E-7 Standard value
E-9 Flash errorThe instrument is not able to save data.
Reading
flashes
1
2
error
The reading is more
or less than the
instrument range.
When an E-1 or E-2 error occurs on a measurement, the display shows
“_.__”. The decimal place depends on the chemistry. If the E-1 or E-2 error
occurs while the instrument is set to zero, set the instrument to zero again.
The flashing value will be 10% over the upper test range limit.
The standard solution concentration is equal
to another standard solution concentration
that is already entered in the user-entered
calibration curve. Enter the correct standard
concentration.
If the reading is less than the instrument
range, make sure that the instrument cap is
fully installed over the cell holder. Measure a
blank. If the blank reading is not zero, set the
instrument to zero again.
If the reading is more than the instrument
range, identify if there is a light blockage in
2
the cell holder. Dilute the sample. Do the test
again.
For factory-calibrated programs, the
maximum and minimum values always equal
the factory-calibrated values and cannot be
changed.
Replacement parts
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause
personal injury, damage to the instrument or equipment malfunction.
The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact
the appropriate distributor or refer to the company website for contact
information.
English 27
Replacement parts
DescriptionQuantityItem no.
AAA batteries, alkaline4/pkg4674300
Cap cord1 5955900
Instrument cap1 5954800
Sample cell, 25 mm (10 mL), with caps6/pkg2427606
Sample cell, 1 cm (10 mL), with caps2/pkg4864302
28 English
Inhaltsverzeichnis
Technische Daten auf Seite 29Betrieb auf Seite 37
Allgemeine Informationen
auf Seite 30
Inbetriebnahme auf Seite 34Fehlerbehebung auf Seite 55
Benutzerschnittstelle und Navigation
auf Seite 35
Wartung auf Seite 53
Ersatzteile auf Seite 56
Technische Daten
Änderungen vorbehalten.
Technische DatenDetails
Abmessungen (B x T x H)6,1 x 3,2 x 15,2 cm (2,4 x 1,25 x 6 Zoll)
GehäuseIP67, bei 1 m (3,3 Fuß) Tiefe für 30 Minuten
LichtquelleLeuchtdiode (LED)
DetektorSilizium-Photodiode
DisplayLCD mit Hintergrundbeleuchtung
Gewicht0,2 kg (0,43 Pfund)
Verschmutzungsgrad2
EinbaukategorieI
Schutzklasse3
StromversorgungAAA-Batterien; ungefähre Lebensdauer
Betriebsumgebungstemperatur 0 bis 50 °C (32 bis 122 °F); 0 bis 90 % relative
Lagerungstemperatur-20 bis 55 °C (-7,6 bis 131 °F)
Photometrische Präzision±0,0015 Abs
WellenlängeFeste Wellenlänge ±2 nm, je nach Modell
wasserdicht (außer Batteriefach). Aus direktem
Sonnenlicht fernhalten.
2000 Tests (weniger bei Verwendung der
Hintergrundbeleuchtung)
Wiederaufladbare Batterien werden nicht
empfohlen.
Luftfeuchtigkeit, nicht kondensierend
unterschiedlich
Deutsch 29
Technische DatenDetails
Filterbandbreite15 nm
Absorbanzbereich0 bis 2,5 Abs
Küvettenschichtdicke1 cm (5 – 10 ml), 25 mm (10 ml)
DatenspeicherLetzte 10 Messungen
ZertifizierungenCE-Zeichen
Garantie2 Jahre
Allgemeine Informationen
Der Hersteller ist nicht verantwortlich für direkte, indirekte,
versehentliche oder Folgeschäden, die aus Fehlern oder
Unterlassungen in diesem Handbuch entstanden. Der Hersteller behält
sich jederzeit und ohne vorherige Ankündigung oder Verpflichtung das
Recht auf Verbesserungen an diesem Handbuch und den hierin
beschriebenen Produkten vor. Überarbeitete Ausgaben der
Bedienungsanleitung sind auf der Hersteller-Webseite erhältlich.
Sicherheitshinweise
H I N W E I S
Der Hersteller ist nicht für Schäden verantwortlich, die durch Fehlanwendung
oder Missbrauch dieses Produkts entstehen, einschließlich, aber ohne
Beschränkung auf direkte, zufällige oder Folgeschäden, und lehnt jegliche
Haftung im gesetzlich zulässigen Umfang ab. Der Benutzer ist selbst dafür
verantwortlich, schwerwiegende Anwendungsrisiken zu erkennen und
erforderliche Maßnahmen durchzuführen, um die Prozesse im Fall von
möglichen Gerätefehlern zu schützen.
Bitte lesen Sie dieses Handbuch komplett durch, bevor Sie dieses
Gerät auspacken, aufstellen oder bedienen. Beachten Sie alle
Gefahren- und Warnhinweise. Nichtbeachtung kann zu schweren
Verletzungen des Bedieners oder Schäden am Gerät führen.
Stellen Sie sicher, dass die durch dieses Messgerät bereitgestellte
Sicherheit nicht beeinträchtigt wird. Verwenden bzw. installieren Sie
das Messsystem nur wie in diesem Handbuch beschrieben.
30 Deutsch
Loading...
+ 552 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.