HACH LANGE Pocket Colorimeter 2 User Manual

DOC022.98.80452
Pocket Colorimeter II Single
Wavelength
04/2014, Edition 1
Bedienungsanleitung
Manual do utilizador Uživatelská příručka
Gebruikershandleiding
Ръководство на потребителя
Használati útmutató
Руководство пользователя
Navodila za uporabo
Korisnički priručnik
Εγχειρίδιο λειτουργίας
User Manual
Manuale dell'utente Manuel d'utilisation
Brugsanvisning
Instrukcja obsługi
Bruksanvisning
Käyttäjän käsikirja
Manual de utilizare
Kullanım Kılavuzu
Návod na obsluhu
English..................................................................................................3
Deutsch...............................................................................................29
Italiano.................................................................................................58
Français..............................................................................................86
Español.............................................................................................115
Português..........................................................................................143
Čeština..............................................................................................171
Dansk................................................................................................197
Nederlands........................................................................................223
Polski.................................................................................................252
Svenska............................................................................................279
Suomi................................................................................................305
български.........................................................................................331
Magyar..............................................................................................361
Română.............................................................................................388
Русский.............................................................................................415
Türkçe...............................................................................................445
Slovenský jazyk.................................................................................471
Slovenski...........................................................................................498
Hrvatski.............................................................................................525
Ελληνικά............................................................................................551
2

Table of contents

Specifications on page 3 Operation on page 11 General information on page 4 Maintenance on page 25 Startup on page 7 Troubleshooting on page 26 User interface and navigation
on page 9
Replacement parts on page 27

Specifications

Specifications are subject to change without notice.
Specification Details
Dimensions (W x D x H) 6.1 x 3.2 x 15.2 cm (2.4 x 1.25 x 6 in.) Enclosure IP67, waterproof at 1 m (3.3 ft) for 30 minutes (battery
Light source Light emitting diode (LED) Detector Silicon photodiode Display LCD with backlight Weight 0.2 kg (0.43 lb) Pollution degree 2 Installation category I Protection class 3 Power requirements 4 AAA batteries; approximate life of 2000 tests (use of
Operating environment 0 to 50 °C (32 to 122 °F), 0 to 90% relative humidity
Storage temperature –20 to 55 °C (–7.6 to 131 °F) Photometric precision ± 0.0015 Abs Wavelength Fixed wavelength ±2 nm, different for each model Filter bandwidth 15 nm Absorbance range 0 to 2.5 Abs
compartment not included). Keep out of direct sunlight.
backlight decreases this number) Rechargeable batteries are not recommended.
non-condensing
English 3
Specification Details
Sample cell path length 1 cm (5–10 mL), 25 mm (10 mL) Data storage Last 10 measurements Certifications CE mark Warranty 2 years

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website.

Safety information

N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
4 English
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation that may result in minor or moderate injury.
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.
C A U T I O N
N O T I C E
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
Certification Canadian Radio Interference-Causing Equipment Regulation,
IECS-003, Class A:
Supporting test records reside with the manufacturer. This Class A digital apparatus meets all requirements of the Canadian
Interference-Causing Equipment Regulations. Cet appareil numérique de classe A répond à toutes les exigences de
la réglementation canadienne sur les équipements provoquant des interférences.
FCC Part 15, Class "A" Limits
English 5
Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including
interference that may cause undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:
1. Move the equipment away from the device receiving the
interference.
2. Reposition the receiving antenna for the device receiving the
interference.
3. Try combinations of the above.

Product overview

The single wavelength Pocket Colorimeter II instruments are portable filter photometers used for testing water, treated waters, wastewater, estuary and seawater. Refer to Figure 1. The single wavelength models are configured at the factory to measure at a specific wavelength.
The single wavelength models have two channels in which measurements can be made. Until a user-prepared calibration curve is entered, the single wavelength instruments show only a direct readout of absorbance. To measure concentration, enter a user-prepared calibration curve. Refer to User-entered calibration on page 21.
6 English
Figure 1 Instrument overview
1 Instrument cap 3 Display 2 Cell holder 4 Keypad

Startup

Install the batteries

W A R N I N G
Explosion hazard. Incorrect battery installation can cause the release of explosive gases. Be sure that the batteries are of the same approved chemical type and are inserted in the correct orientation. Do not mix new and used batteries.
Install the batteries as shown in Figure 2.
English 7
Figure 2 Install the batteries
8 English

Install the cap cord

Attach the cap cord to prevent loss of the instrument cap. Refer to
Figure 3.
Figure 3 Install the cap cord

User interface and navigation

Keypad description

Figure 4 shows the keypad and gives the key functions.
English 9
Figure 4 Keypad
1 Power/Backlight key: Sets the
power to on and off. Push and hold for 1 second to set the backlight to on or off.
2 Zero/Scroll key: Sets the
instrument to zero, scrolls through menu options and numbers
3 Menu key: Enters and goes out of
menu mode.
4 Read/Enter key: Starts a sample
measurement, selects a menu option, moves the cursor to the next digit

Display description

Figure 5 shows the values and icons shown on the display.
10 English
Figure 5 Display
1 Numeric display: Measured value
or menu options
2 Range icon: Selected range or
parameter
3 Range value: Range(s) or
parameters
4 Menu icon: The instrument is in
menu mode.
5 Calibration adjusted icon: A user-
entered calibration curve was entered.
6 Low battery icon: Battery level is
10%. Flashes when the battery level is too low to complete measurements.

Operation

Configure the instrument

1. Push .
2. Push to scroll through the menu options. Push to select an
option.
Option Description SEL Sets the measurement range or parameter. Push to toggle
between the measurement ranges or parameters.
00:00 Sets the time in 24-hour format (hh:mm). Push to change the
time. Push to change the first digit, then to go to the next digit.
English 11
Option Description rCL Shows the last 10 measurements recorded. Push to show the
recorded measurements (01—most recent measurement, 10— oldest measurement). Push to scroll through the measurements. To select a measurement by number, push to select the number and then . Push to go out of this option.
SCA Not applicable to the single wavelength models.
3. Push to go back to measurement mode.

Measurement

Basic colorimetry
Colorimetry measures the amount of color in a clear medium, such as a liquid, to identify the quantity of a particular substance (the analyte) in the liquid. Typically, the concentration of the analyte is proportional to the intensity of the color in the clear medium (solution). In most methods, a darker color indicates a higher analyte concentration.
Absorbance (Abs) at a specific wavelength is typically used to measure the amount of light absorbed by the solution. Absorbance (Abs) is calculated as:
Abs = –log T or Abs = –log (IT/IO) Where: T = transmittance IT = intensity of the light transmitted through the sample IO = intensity of the light that enters the sample Some substances, such as dyes and different metal ions, have
inherent color and can be measured without any additions. In most cases, a chemical reaction between an indicator and the analyte is necessary to get a colored product that can be measured.
Once the relationship between the amount of color (measured as absorbance) and a known concentration of a sample is identified, the instrument can be used to measure concentrations of unknown samples. A user-entered calibration curve is used to measure the sample concentration.
To identify the amount of color in a sample, the instrument measures the amount of light the solution absorbs. The absorption of light is dependent on the wavelength of the light and the color of the solution.
12 English
The combination of an LED light source and an interference filter sets the measurement wavelength.
Select the best wavelength
The single wavelength instruments each have a different LED and interference filter to measure at a specific wavelength.
The wavelength (color) of light used is typically selected so that it has a maximum absorption, but other wavelengths can be selected to minimize interferences or other factors. For the best results, select the instrument wavelength with knowledge about the absorbance spectra of the species of interest, as well as the spectra of other colored species that could be in the sample. Figure 6 shows a typical absorption spectrum.
Refer to Table 1 to select the best instrument wavelengths to use for testing. Do not use this table for samples that have more than one absorption region that adds to the visible color. For example, a green solution can have a yellow and a blue absorption peak. One or the other peak can be used for measurements if both have different analyte concentration. Other samples can look brown because there are several spectra that add to the visible color.
English 13
Figure 6 Select the best wavelength – sample spectrum
1 Absorbance 2 Wavelength (nm)
Table 1 Light wavelength and color
Sample color Light absorbed Wavelength (nm)
Yellow-green Violet 420
Yellow Violet-blue 450
Orange Blue 476
Orange-red Blue-green 500
Red Green 528
Red-violet Yellow-green 550
14 English
Table 1 Light wavelength and color (continued)
Sample color Light absorbed Wavelength (nm)
Blue Yellow 580 Green-blue Orange 600 Blue-green Red 655
Measurement range
The measurement range of the instrument is 0 to approximately
1.50 Abs, but can be used up to a measurement range of 2.5 Abs if the chemistry method supports that range.
If sample absorbances are more than 1.50 Abs:
1. Dilute the sample or use smaller sample cells for the best linearity
and accuracy.
2. If a smaller sample cell such as the 1-cm (10-mL) cell is used,
complete the calibration with the smaller sample cells.
Note: Absorbance increases with the increase of the sample cell pathlength. Use a sample cell with a shorter pathlength to measure solutions with a darker colored.
3. Monitor the calibration curve to identify the measurement range for
a specific test. The measurement range is the concentration range in which the
deviation from linearity is within acceptable limits.
Calibration curve
Calibration curves should ideally intersect the zero intercept for absorbance. The zero intercept is the zero concentration point on the calibration graph. When there is no analyte in the sample, the absorbance will be zero.
A non-zero intercept (a positive or negative absorbance measurement at zero concentration) can occur for many reasons. Factors that can cause a non-zero intercept include the reagent blank, pH, temperature, interfering species or turbidity differences between the zeroing solution (blank) and the sample.
To adjust for a non-zero intercept caused by the reagent blank, measure the absorbance of the prepared reagent blank and then subtract it from the measured absorbance of the prepared sample. In
English 15
an aqueous sample, add the reagents to deionized water to prepare the reagent blank. The prepared reagent blank includes only the amount of color that is added to the deionized water by the reagent and not the analyte. The prepared sample includes the amount of color that is added by the reagent and the analyte.
For some chemistries, the intensity of the color decreases as the analyte concentration increases. These chemistries are referred to as bleaching chemistries because the measured sample is lighter in color than the reagent blank that was used to zero the instrument. This instrument is able to measure bleaching (or negative) absorbance chemistries directly. Set the instrument zero with the reagent blank (the most highly colored solution) and then read the sample or bleached color directly.
Single wavelength procedure Before starting
Always measure solutions in sample cells or AccuVac® Ampules. Do not put the instrument in the sample or pour the sample into the cell holder.
Make sure that the sample cells are clean and there are no scratches where the light passes through them.
Make sure that there are no fingerprints or liquid on the external surface of the sample cells or AccuVac® Ampules. Wipe with a lint-free cloth.
Rinse the sample cell and cap with the sample three times before the sample cell is filled.
Always insert the sample cell in the correct and consistent orientation so that the results are more repeatable and precise. Refer to Figure 7.
Install the instrument cap over the cell holder before ZERO or READ is pushed. Refer to Figure 8.
Measure the volume of the liquid reagent accurately. Use a pipet if possible.
If the test result is over-range, dilute a fresh sample with a known volume of deionized water and repeat the test. Multiply the result by the dilution factor.
When the test is completed, immediately empty and rinse the prepared sample cell. Rinse the sample cell and cap three times.
16 English
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective equipment.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory agencies for further disposal information.
Figure 7 Sample cell orientation
1 Orientation mark 2 Sample cell, 25-mm (10 mL)
English 17
Figure 8 Install the instrument cap over the cell holder
Sample collection
• Collect samples in clean glass or plastic bottles.
• Rinse the sample bottle several times with the sample to be collected.
• Analyze the samples as soon as possible for best results.
• Homogenize samples that contain solids to get a representative sample.
18 English
• Filter samples that are turbid with filter paper and a funnel.
Reagent solution procedure
1. Select the range that
has a saved user calibration. Refer to
Configure the instrument on page 11.
Note: To enter a user calibration, refer to
User-entered calibration
on page 21.
4. Insert the blank into the cell holder in the correct orientation. Refer to Figure 7 on page 17.
2. Prepare the blank: Fill the sample cell with 10 mL of the blank solution (typically sample).
5. Install the instrument cap over the cell holder.
3. Clean the blank sample cell.
6. Push ZERO. The display shows “0.000”, or the degree of resolution that was previously selected.
English 19
7. Remove the sample
cell from the cell holder.
8. Prepare the sample: Fill a second
sample cell with 10 mL of sample.
9. Add the reagent to the second sample cell. Wait the specified reaction time for full color development if applicable.
10. Clean the prepared
sample cell.
20 English
11. Insert the prepared
sample into the cell holder in the correct orientation. Refer to
Figure 7 on page 17.
12. Install the instrument cap over the cell holder.
13. Push READ. The display shows the measurement results.

Show the recorded measurements

Refer to the "rCL" option in Configure the instrument on page 11.

User-entered calibration

This instrument accepts a user-prepared calibration curve. The calibration curve can be from 0 to 2.5 absorbance. Make sure that the calibration curve includes standard values that are less and more than the range of interest.
The instrument range will be the same as the calibration range. For example, when the standards that are used are 1.00, 2.00 and 4.00. The instrument range is 1.00 to 4.00.
There are two options to enter a user calibration curve:
Enter a calibration curve with standards—The standard solution
values are entered with the keypad and the absorbance values are measured.
Enter a calibration curve with the keypad—The standard solution
values and absorbance values are entered with the keypad.
Note: If the instrument is set to off or the instrument power is removed before a user-entered calibration curve is completed, the calibration curve is not saved. The instrument automatically switches off in user-entered calibration entry mode after 60 minutes of no activity. User-entered calibrations are completed when the user goes out of calibration (cal) mode or edit mode.
English 21
Enter a calibration curve with standards
Note: Deionized water can be used for the blank unless the sample is significantly more turbid or has more color than deionized water.
1. Set the instrument to the range to calibrate. Refer to Configure the
instrument on page 11.
2. Prepare the blank and the reacted standard solution. Refer to the
test procedure. Let the color fully develop.
3. Set the instrument to zero.
a. Insert the blank sample cell in the cell holder. b. Install the instrument cap over the cell holder. c. Push . The display shows “- - - -”, then “0.000”. d. Remove the instrument cap. e. Remove the sample cell from the cell holder.
4. Push and hold until "USER" and then "CAL" shows, then push
.
Note: If "USER" and "CAL" do not show, the factory calibration cannot be changed on the selected range.
5. When "RES" shows on the display, set the resolution.
a. Push . The resolution setting (decimal placement) shows. b. To change the resolution, push , then . Push to save the
change.
c. To not change the resolution, push .
6. When "S0" shows on the display, push . Push to enter the
blank value, then push .
Note: Push to go to the next digit.
7. When “A0” shows on the display, measure the absorbance of the
blank.
a. Insert the blank sample cell in the cell holder. b. Install the instrument cap over the cell holder. c. Push . The display shows the absorbance value for "S0". d. Remove the sample cell from the cell holder.
8. Push to show "S1".
22 English
9. When "S1" shows on the display, push . Push to enter the first
standard value, then push .
Note: Push to enter the next digit.
10. When "A1" shows on the display, measure the absorbance of the
reacted standard solution.
a. Insert the reacted standard sample cell in the cell holder. b. Install the instrument cap over the cell holder. c. Push . The display shows the absorbance value for "S1". d. Remove the sample cell from the cell holder.
11. The calibration is completed with two calibration points. If additional
standards are necessary for calibration:
a. Push until “Add” shows, then push . b. Do steps 910 again to enter more standards.
12. Push two times to go back to measurement mode.
Enter a calibration curve with the keypad
At least two data pairs are necessary to enter a user-prepared calibration curve. A concentration value and the absorbance value for the given concentration is necessary for each data pair. A maximum of 10 data pairs can be entered.
Note: This procedure can also be used to change the data pairs in a user­entered calibration curve.
1. Set the instrument to the range to calibrate. Refer to Configure the
instrument on page 11.
2. Push and hold until "USER" and then "CAL" shows, then push
.
Note: If "USER" and "CAL" do not show, the factory calibration cannot be changed on the selected range.
3. Push until "EDIT" shows, then push .
4. When "RES" shows on the display, set the resolution.
a. Push . The resolution setting (decimal placement) shows. b. To change the resolution, push , then . Push to save the
change.
c. To not change the resolution, push .
English 23
5. When "S0" shows on the display, push . Push to enter the
concentration value of the first data pair, then push .
Note: Push to go to the next digit.
6. When "A0" shows on the display, push . Push to enter the
absorbance value of the first data pair, then push . "S1" shows on the display.
7. Do steps 56 again to enter the second data pair (S1 and A1).
8. The calibration is completed with two data pairs. If additional data
pairs are necessary for calibration:
a. When “Add” shows, push . b. Do steps 56 again to enter more data pairs.
9. Push two times to go back to measurement mode.
Remove a calibration point
To remove a calibration point from a user-entered calibration curve:
1. Set the instrument to the range to calibrate. Refer to Configure the
instrument on page 11.
2. Push and hold until "USER" and then "CAL" shows.
Note: If "USER" and "CAL" do not show, the factory calibration cannot be changed on the selected range.
3. Push until "EDIT" shows, then push .
Note: Calibration points can also be removed in calibration (CAL) mode.
4. Push until the calibration point to remove shows (i.e., S0 or S1),
then push .
5. Push until "dEL" shows, then push .
Note: The minimum number of data pairs is two. When only two data pairs remain, no more data pairs can be removed.
6. Push two times to go back to measurement mode.
24 English
Remove the calibration curve
1. Set the instrument to the applicable range. Refer to Configure the
instrument on page 11.
2. Push and hold until "USER" and then "CAL" shows.
Note: If "USER" and "CAL" do not show, the factory calibration cannot be changed on the selected range.
3. Push until "dFL" shows, then push .

Maintenance

C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
N O T I C E
Do not disassemble the instrument for maintenance. If the internal components must be cleaned or repaired, contact the manufacturer.

Clean the instrument

Clean the exterior of the instrument with a moist cloth and a mild soap solution and then wipe the instrument dry.

Clean the sample cells

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
C A U T I O N
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations.
English 25
Most laboratory detergents are used at recommended concentrations. Neutral detergents, such as Liquinox, are safer to use when regular cleaning is necessary. To decrease the cleaning times, increase the temperature or use an ultrasonic bath. To complete the cleaning, rinse a few times with deionized water and then let the sample cell air dry. Sample cells may also be cleaned with acid, followed by a thorough rinse with deionized water.
Note: Always use acid to clean sample cells that were used for low-level metal tests.
Special cleaning methods are necessary for individual procedures. When a brush is used to clean sample cells, take extra care to avoid scratches on the interior surfaces of the sample cells.

Replace the batteries

Replace the batteries when the battery power level is low. Refer to
Install the batteries on page 7.

Troubleshooting

Error Description Solution
E-0 No zero In user calibration mode, a standard solution
E-1 Ambient light error1There is ambient light in the cell holder. Make
E-2 LED error
E-6 Abs error The absorbance value is not correct or the
1
was measured before the instrument zero was set. Measure a blank solution to set the instrument to zero.
sure that the instrument cap is fully installed over the cell holder.
The LED (light source) is out of regulation. Replace the batteries. Make sure that the
LED in the cell holder comes on when or
is pushed.
user-entered calibration curve has fewer than two points. Enter or measure the absorbance value again.
26 English
Error Description Solution
E-7 Standard value
E-9 Flash error The instrument is not able to save data.
Reading flashes
1
2
error
The reading is more or less than the instrument range.
When an E-1 or E-2 error occurs on a measurement, the display shows “_.__”. The decimal place depends on the chemistry. If the E-1 or E-2 error occurs while the instrument is set to zero, set the instrument to zero again. The flashing value will be 10% over the upper test range limit.
The standard solution concentration is equal to another standard solution concentration that is already entered in the user-entered calibration curve. Enter the correct standard concentration.
If the reading is less than the instrument range, make sure that the instrument cap is fully installed over the cell holder. Measure a blank. If the blank reading is not zero, set the instrument to zero again.
If the reading is more than the instrument range, identify if there is a light blockage in
2
the cell holder. Dilute the sample. Do the test again.
For factory-calibrated programs, the maximum and minimum values always equal the factory-calibrated values and cannot be changed.

Replacement parts

W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the instrument or equipment malfunction. The replacement parts in this section are approved by the manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information.
English 27
Replacement parts
Description Quantity Item no.
AAA batteries, alkaline 4/pkg 4674300
Cap cord 1 5955900
Instrument cap 1 5954800
Sample cell, 25 mm (10 mL), with caps 6/pkg 2427606
Sample cell, 1 cm (10 mL), with caps 2/pkg 4864302
28 English

Inhaltsverzeichnis

Technische Daten auf Seite 29 Betrieb auf Seite 37 Allgemeine Informationen
auf Seite 30
Inbetriebnahme auf Seite 34 Fehlerbehebung auf Seite 55 Benutzerschnittstelle und Navigation
auf Seite 35
Wartung auf Seite 53
Ersatzteile auf Seite 56

Technische Daten

Änderungen vorbehalten.
Technische Daten Details
Abmessungen (B x T x H) 6,1 x 3,2 x 15,2 cm (2,4 x 1,25 x 6 Zoll) Gehäuse IP67, bei 1 m (3,3 Fuß) Tiefe für 30 Minuten
Lichtquelle Leuchtdiode (LED) Detektor Silizium-Photodiode Display LCD mit Hintergrundbeleuchtung Gewicht 0,2 kg (0,43 Pfund) Verschmutzungsgrad 2 Einbaukategorie I Schutzklasse 3 Stromversorgung AAA-Batterien; ungefähre Lebensdauer
Betriebsumgebungstemperatur 0 bis 50 °C (32 bis 122 °F); 0 bis 90 % relative
Lagerungstemperatur -20 bis 55 °C (-7,6 bis 131 °F) Photometrische Präzision ±0,0015 Abs Wellenlänge Feste Wellenlänge ±2 nm, je nach Modell
wasserdicht (außer Batteriefach). Aus direktem Sonnenlicht fernhalten.
2000 Tests (weniger bei Verwendung der Hintergrundbeleuchtung)
Wiederaufladbare Batterien werden nicht empfohlen.
Luftfeuchtigkeit, nicht kondensierend
unterschiedlich
Deutsch 29
Technische Daten Details
Filterbandbreite 15 nm Absorbanzbereich 0 bis 2,5 Abs Küvettenschichtdicke 1 cm (5 – 10 ml), 25 mm (10 ml) Datenspeicher Letzte 10 Messungen Zertifizierungen CE-Zeichen Garantie 2 Jahre

Allgemeine Informationen

Der Hersteller ist nicht verantwortlich für direkte, indirekte, versehentliche oder Folgeschäden, die aus Fehlern oder Unterlassungen in diesem Handbuch entstanden. Der Hersteller behält sich jederzeit und ohne vorherige Ankündigung oder Verpflichtung das Recht auf Verbesserungen an diesem Handbuch und den hierin beschriebenen Produkten vor. Überarbeitete Ausgaben der Bedienungsanleitung sind auf der Hersteller-Webseite erhältlich.

Sicherheitshinweise

H I N W E I S
Der Hersteller ist nicht für Schäden verantwortlich, die durch Fehlanwendung oder Missbrauch dieses Produkts entstehen, einschließlich, aber ohne Beschränkung auf direkte, zufällige oder Folgeschäden, und lehnt jegliche Haftung im gesetzlich zulässigen Umfang ab. Der Benutzer ist selbst dafür verantwortlich, schwerwiegende Anwendungsrisiken zu erkennen und erforderliche Maßnahmen durchzuführen, um die Prozesse im Fall von möglichen Gerätefehlern zu schützen.
Bitte lesen Sie dieses Handbuch komplett durch, bevor Sie dieses Gerät auspacken, aufstellen oder bedienen. Beachten Sie alle Gefahren- und Warnhinweise. Nichtbeachtung kann zu schweren Verletzungen des Bedieners oder Schäden am Gerät führen.
Stellen Sie sicher, dass die durch dieses Messgerät bereitgestellte Sicherheit nicht beeinträchtigt wird. Verwenden bzw. installieren Sie das Messsystem nur wie in diesem Handbuch beschrieben.
30 Deutsch
Loading...
+ 552 hidden pages