The piston pipette is a handy, general-purpose microliter
pipette for transferring exact volumes of liquid. It operates on
the principle of air expulsion, making use of disposable tips.
The volume that is to be pipetted is shown in a clearly visible
display on the pipette handle. All piston pipettes are fitted with
a mechanical tip ejector system.
1.1Pipette tips
We recommend the use of original HACH pipette tips when
working with the piston pipette. The pipette tips are made of
polypropylene and can be autoclaved at 121 °C/250 °F.
14
Section 2Using the pipette
2.1Setting the volume
(only for variable piston pipettes)
The pipette volume is set with the push button at the end of the
handle. Turn the button anticlockwise to increase the volume
and clockwise to decrease it. Make sure that the volume
setting is correct (clicks into place) and the volume shown in
the display is completely visible.
Figure 2 Setting the volume
Important Note: The given precision and accuracy only apply
for the volume range specified on the pipette. Do not set the
volume to a value outside of this range. Do not use force to
turn the push button to a volume that is outside of the specified
range. This could damage the pipette.
2.2Tip ejection
Each piston pipette is fitted with a tip ejection system to
prevent possible contamination. Press thumb down on ejector
button and remove tip from tip cone.
15
Using the pipette
2.3Pipetting techniques
The pipette is operated by means of the push button at the
end of the handle.
Maximum accuracy can be achieved by carrying out the
following steps:
1. Move the piston up or down slowly and evenly. (especially
when pipetting highly viscous liquids)
2. Do not allow the push button to spring back into its original
position.
3. Use light pressure to fit a new tip to the cone, turning the
tip slightly as you do so.
4. Make sure that the tip is firmly in position on the tip cone
and there are no foreign bodies between the tip and the
cone.
5. Wet the tips with the liquid by repeatedly filling and
emptying them.
6. Hold the pipette almost vertical (no more than 10° from the
vertical) while liquid is being drawn into the tip.
7. The temperature of the pipette and the tip should match
that of the liquid.
16
Section 3Standard technique
1. Press the push button to the first pressure point.
2. Dip the pipette tip under the surface of the liquid (2–3 mm)
and slowly release the push button. Take the tip out of the
liquid and touch it against the inside wall of the vessel to
remove excess liquid.
3. Place the tip against the inside wall of the vessel and
press lightly and evenly on the push button until the first
pressure point is reached. Hold the pipette in this position.
After about one second, press the push button until the
second pressure point is reached. This completely
empties the tip.
4. Allow the push button to return to its rest position.
Figure 3 Standard technique
aRest positioncSecond pressure point
bFirst pressure point
Important Note: After use change pipette tip!
17
Section 4Daily check
Check the pipette daily for dust and external soiling before
starting work or after finishing work. Pay special attention to
the tip cone. Clean the pipette with 70% alcohol. Do not use
any other solvents to clean the pipette.
18
Section 5Maintenance
If the pipette is used each day, maintenance should be carried
out at least twice each year. Maintenance can be carried out
in your laboratory. Carry out the steps described below:
5.1Opening the pipettes (0.2–1.0 mL)
1. Press the ejector button.
2. Insert the tip of the opening tool into the opening at the
end of the ejector button.
3. Pull out the ejector button and the tip ejector piston.
Figure 4 Opening the pipettes
19
Maintenance
4. Remove the tip cone by turning it anticlockwise with the
enclosed tool.
5. Pull out the piston.
6. Remove the O-rings from the tip cone.
7. Clean the piston, the piston spring and the O-rings with a
clean, dry cloth.
8. Check the cylinder for foreign substances. Remember that
the cylinder does not have to be greased.
9. Grease the cleaned parts lightly with the accompanying
grease.
10. Reassemble the pipette.
20
Maintenance
5.2Opening the pipette (1–5 mL)
1. Remove ejector button and tip ejector as described above
(for 0.2–1.0 mL pipettes) section 5.1 on page 19.
2. Remove cover.
3. Remove cone and cylinder with the help of the tool by
pressing the two projections together. While pressing the
projections, push the tool towards the cone.
Figure 5 Opening the pipette
1Service tool to open the pipette
4. Take apart, clean and grease as described above (for
0.2–1.0 mL pipettes) section 5.1 on page 19.
5. Reassemble in reverse sequence.
21
Section 6Calibration
The pipette is calibrated in the factory with distilled water at
22 °C/72 °F. No recalibration is needed for routine use. The
pipette can, however, be recalibrated for other temperatures
and solutions of different viscosities. An appropriately
sensitive analytical balance, a small beaker and distilled water
are needed to carry out the calibration.
Important Note: If either the handle or the piston is replaced,
the pipette must be recalibrated. If the O-rings or the tip cone
are replaced, the calibration should be checked.
6.1Checking the calibration
1. Set the pipette volume as follows in Table 1 Checking the
2. Use light pressure to place a tip on the tip cone, turning
the tip slightly.
3. Pipette distilled water into previously weighed beaker at
least five times. Determine each weight in mg to the first
decimal place and compare this with the weight shown
above. If the weight is not within this range, recalibrate the
pipette.
23
Section 7Recalibration
The calibration is carried out with the supplied calibration key.
1. Place the calibration key in the holes of the calibration
groove at the top end of the handle.
2. Increase the volume by turning in clockwise direction or
reduce it by turning in anticlockwise direction.
Figure 6 Recalibration
1Calibration key
3. Afterwards check the volume as described above
Section 6 Calibration on page 22.
Important Note: The pipette can only be calibrated within the
limits of the volume range for which it is meant to be used.
24
Section 8Troubleshooting
The following table is intended to help you to identify and
eliminate errors:
Table 2 Troubleshooting
ErrorPossible causeResponse
Leakage
Inaccurate
pipetting
Tip not attached
properly
Foreign bodies
between cone and
tip
Foreign bodies
between O-ring and
piston
Insufficient grease
on the O-rings and
the piston
Defective O-ringsReplace O-rings
Incorrect operation
Tip not firmly
attached
Calibration has
changed, e.g. due to
faulty operation
Fix tip firmly in place
Clean the cone and
attach a new tip
Clean and grease
the O-rings and the
piston in the cylinder.
Use the supplied
grease
Grease adequately
Use pipette as
described in the
manual
Attach tip firmly
Recalibrate
25
Section 9Warranty, liability
and complaints
The manufacturer warrants that the product supplied is free of
material and manufacturing defects and undertakes the
obligation to repair or replace any defective parts at zero cost.
The warranty period for these pipettes is 24 months.
With the exclusion of the further claims, the supplier is liable
for defects including the lack of assured properties as follows:
all those parts that can be demonstrated to have become
unusable or that can only be used with significant limitations
due to a situation present prior to the transfer of risk, in
particular due to incorrect design, poor materials or
inadequate finish will be improved or replaced, at the
supplier's discretion. The identification of such defects must be
notified to the supplier in writing without delay, however at the
latest 7 days after the identification of the fault. If the customer
fails to notify the supplier, the product is considered approved
despite the defect. Further liability for any direct or indirect
damages is not accepted.
If specific maintenance work defined by the supplier is to be
performed within the warranty period by the customer
(maintenance) and these requirements are not met, claims for
damages due to the failure to comply with the requirements
are rendered void.
Any further claims, in particular claims for consequential
damages cannot be made.
Consumables and damage caused by improper handling, poor
installation or incorrect use are excluded from this clause.