Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol on the instrument is referenced in the
manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal
systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
Specifications
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.
C A U T I O N
Personal injury hazard. Glass components can break. Handle with care to prevent cuts.
Specifications are subject to change without notice.
SpecificationsDetails
Probe typeDigital combination glass pH probe with refillable electrolyte and built-in temperature sensor
Body materialGlass
pH rangepH 0 to 14
pH resolutionUser-selectable stabilization time and resolution—Fast: 0.1, Fast: 0.01, Medium: 0.01, Slow:
0.01 or Slow: 0.001
pH accuracy±0.01 pH
Sodium (alkalinity) error<0.3 pH at pH 13 in 0.1 M NaOH
Slope–59 mV/pH (85 to 115% at 25 °C (77 °F) per Nernstian theoretical value)
Isopotential pointpH 6.00
Temperature compensation 30 kΩ NTC thermistor
Temperature accuracy±0.3 °C (±0.54 ° F)
Operating temperature–10 to 100 °C (14 to 212 °F)
Storage temperature5 to 40 °C (41 to 104 °F)
Junction typeAnnular porous ring
Reference typeRed Rod
1
ElectrolyteRefillable, saturated KCl
1
SpecificationsDetails
Storage solutionSaturated KCl
Minimum sample depth14 mm (0.55 in.)
DimensionsBody diameter: 12 mm (0.47 in.)
Body length: 103 mm (4.1 in.)
Total length: 200 mm (7.87 in.)
Cable length: 1 m (3.28 ft)
Cable connectionM12 digital output and connector compatible with HQd meters
1
Radiometer Analytical’s Red Rod technology: Encapsulation of reference elements with a red protection tube.
Product overview
The probe is a digital pH combination electrode with refillable electrolyte, a glass body
and an integrated temperature sensor. The probe is for laboratory use and measures pH
in general aqueous samples.
Note: Do not use the probe with organic solvents, emulsions, samples with high solids content or
high alkaline sample types.
A 59-mL bottle of reference electrolyte filling solution (saturated KCl solution) is included
with the probe. Refer to Figure 1.
Figure 1 Probe overview
1 Probe soaker bottle5 Probe soaker bottle holder
2 Filling-hole cap6 Reference junction
3 Protective tape and filling-hole7 Glass bulb and temperature sensor
4 Cable, 1 meter (3.28 ft)
Preparation for use
Prepare the probe before a calibration or a sample measurement.
1. Turn the probe soaker bottle cap counter-clockwise to loosen the cap.
2. Remove the soaker bottle from the probe.
3. Fully rinse the reference junction and the glass bulb with deionized water to remove
all of the storage solution or possible salt particles. Dry with a no-lint cloth.
4. Remove the protective tape from the filling hole before initial use (refer to Product
overview on page 2). Discard the protective tape.
2
5. Add the filling solution to the probe as necessary (refer to Fill the probe
on page 11). The filling solution must be above the standard solution or the sample
level during measurement or calibration (Figure 2).
6. For Red Rod probes, make sure that there is approximately 0.5 cm (0.2 in.) of KCl
crystals in the filling solution in the probe. Refer to Table 1. For measurements at
temperatures higher than 60 °C (140 °F), add KCl crystals through the filling hole.
Refer to Table 1 to calculate the quantity of KCl crystals to add. The quantity of
necessary KCl crystal increases linearly from 60 to 100 °C.
7. For Red Rod probes, make sure that the KCl crystals move freely through the liquid
when the probe is inverted. If necessary, shake the probe to loosen the crystals.
8. Make sure that the filling hole is open during measurement for the correct flow of the
filling solution.
Note: For the best stabilization time, condition the probe for 2 hours in pH buffer 4.01 before use, if
the probe was not used for 3 weeks or more. In low ionic strength (LIS) samples, condition the
probe for a maximum of 5 minutes in the sample before use. In all other cases a condition before
use is not necessary.
Figure 2 Filling solution level
1 KCl crystals
2 Filling solution level4 Standard solution or sample level
1
Red Rod probes only
1
3 Filling solution
Table 1 KCl crystal level for Red Rod pH probes
TemperatureAmount of KCl crystalsAddTemperatureAmount of KCl crystalsAdd
60 °C (158 °F)0.5 g (0.5 cm)0 g90 °C (194 °F)0.875 g0.375 g
70 °C (158 °F)0.625 g0.125 g100 °C (212 °F)1.0 g (1.0 cm)0.5 g
80 °C (176 °F)0.750 g0.25 g
Calibration
Before calibration:
The probe must have the correct service-life time stamp. Set the date and the time in the meter before the probe is attached.
It is not necessary to calibrate again if a calibrated probe is moved from one HQd meter to another. Make sure that the
additional meter is configured to use the same calibration options.
3
Before calibration:
To view the calibration data, push , then select View Probe Data>View Current Calibration.
If two probes are connected, push the UP or DOWN arrow to change to the single display mode to show the Calibrate option
(available only on HQ40d and HQ440d meters).
Prepare the probe for use (refer to Preparation for use on page 2).
Calibration notes:
•Use buffers that are two pH units apart.
•Select additional standard sets with the minimum number of calibration points in the
Calibration Options. For a two-point calibration, select: one buffer with a pH above
and one buffer with a pH below the sample pH. For a one-point calibration, select the
buffer nearest to the sample pH.
•The calibration is recorded in the probe and the data log. Send the calibration to a
PC, a printer or a flash memory stick if connected.
•Air bubbles under the sensor tip can cause a slow response or an error in the
measurement. If there are bubbles under the sensor tip, carefully shake the probe to
remove the bubbles.
•Make sure that the filling hole is open and above the sample solution level.
•Use a laboratory stirrer or shake the probe from side to side in the solution to refresh
the reference junction.
•The meter display shows "Stabilizing" and a progress bar as the probe stabilizes in
the standard. The display shows the buffer and the temperature corrected pH value
when the measurement is stable.
•The calibration and sample measurement conditions must be as similar as possible
(e.g., the temperature of the solution, the stir procedure, the stir rate and the position
of the probe).
•If a calibration error occurs, refer to Troubleshooting on page 12.
1. Connect the probe
to the meter. Make
sure that the cable
locking nut is correctly
connected to the
meter. Set the meter
to on.
Calibration procedure:
2. Push Calibrate.
The display shows the
buffers that are
necessary for
calibration.
3. Prepare the fresh
buffers in separate
beakers or applicable
containers.
4. Rinse the probe
with deionized water.
Dry with a no-lint
cloth.
5. Put the probe in
the pH buffer solution
and stir lightly. Make
sure to fully put the
reference junction into
the standard.
4
6. Push Read. Wait
until the reading
stabilizes.
7. Do steps 4–6
again to collect the
minimum number of
calibration points
specified in the
calibration options.
Sample requirements
8. Push Done to view
the calibration
summary.
Note: The display will
not show "Done" until
the minimum number
of calibration points is
collected.
9. Push Store to
accept the calibration
and go back to the
measurement mode.
N O T I C E
Some probes are not compatible with specific sample types. Probe damage can occur.
•Samples must be aqueous. The probe can also make measurements in partially
aqueous or some water-miscible solvents. Read the results with caution because the
full pH scale changes when the solvent system changes.
•Probes with a Red-Rod reference system are compatible with solutions that contain
silver complexing or binding agents (e.g., TRIS, proteins and sulfides), as they have
an encapsulated reference system.
•Proteins can collect on the sensing element. Make sure that the probe stays clean
when these types of samples are measured.
•Do not use the probe in solutions that are out of the temperature range of the probe.
Refer to Specifications on page 1.
•For low ionic strength (LIS) or high purity samples, use an LIS chamber or a flow cell
to prevent sample contamination.
Measurement
Before measurement:
The probe must have the correct service-life time stamp. Set the date and the time in the meter before the probe is attached.
If the complete traceability is necessary, enter a sample ID and an operator ID before the measurement. Refer to the HQd
meter manual for more information.
A regular calibration is necessary for the best measurement accuracy (refer to Calibration on page 3).
Prepare the probe for use (refer to Preparation for use on page 2).
Measurement notes:
•Data is automatically stored in the data log when Press to Read or Interval is
selected in the Measurement Mode. When Continuous is selected, data is only
stored when Store is pushed.
•Air bubbles under the sensor tip can cause a slow response or an error in the
measurement. If there are bubbles under the sensor tip, carefully shake the probe to
remove the bubbles.
•For the best stabilization time, condition the probe for several minutes in the sample
or in a solution comparable to the sample in terms of pH and ionic strength before the
initial sample measurement.
5
Loading...
+ 11 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.