Hach-Lange PHC30103 User Manual

User Manual
Refillable pH Probe: Model PHC30101 or PHC30103

Safety information

Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.

Specifications

Note: Specifications are subject to change without notice.
Specifications Details
DOC022.52
.80032
Probe type Digital combination probe with refillable double junction reference and built-in
temperature sensor
pH range pH 0 to 14
pH resolution User-selectable stabilization time and resolution—Fast: 0.1, Fast: 0.01,
Medium: 0.01, Slow: 0.01 or Slow: 0.001
Sodium (alkalinity) error -0.6 pH at pH 12.6 in 1 M NaOH
Slope -59 mV/pH (90 to 110% at 25 °C per Nernstian theoretical value)
Temperature accuracy ±0.3 °C (±0.54 ° F)
Operating temperature range 0 to 50 °C (32 to 122 °F)
Storage temperature range 5 to 40 °C (41 to 104 °F)
Junction Ceramic
Reference type Ag/AgCl
Minimum sample depth 15 mm (0.59 in.)
Dimensions Diameter: 12 mm (0.47 in.)
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Cable connection M12 digital output and connector compatible with HQd meters

Product overview

The PHC301 series probe is a pH refillable combination probe with a built-in temperature sensor (Figure 1). The PHC30101 or PHC30103 probe is available with a 1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory use. The pH refillable combination probe measures pH in wastewater, drinking water and general aquous applications. The probe is not suitable for use with organic solvents. A 28-mL bottle of reference electrolyte filling solution (3 M KCl solution saturated with AgCl) is included with the probe.
1
Figure 1 Probe overview
1 Probe soaker bottle 4 Protective tape and filling-hole
2 Glass bulb, reference junction and temperature sensor 5 1 or 3 meter (3.28 or 9.84 ft) cable
3 Filling-hole cap 6 Probe soaker bottle holder

Preparation for use

Prepare the probe for use before calibration or sample measurement.
1. Turn the probe soaker bottle cap counter-clockwise to loosen the cap.
2. Remove the soaker bottle from the probe.
3. Rinse the reference junctions
remove the viscous storage/filling solution completely. Blot dry with a lint-free cloth.
4. Remove the protective tape from the filling hole before initial use (refer to Figure 1 on page 2). Dispose of the protective tape.
5. Add filling solution to the probe as necessary (refer to Fill the probe on page 10). The filling solution must be above the standard solution or sample level during measurement (Figure 2).
6. Make sure that the filling hole is open during measurement for the proper flow of the filling solution.
Note: For the best stabilization time, condition the probe for several minutes in the sample before use.
and glass bulb with deionized water thoroughly to
Figure 2 Measurement method
1 Container 3 Filling solution level
2 Probe body 4 Standard solution or sample level
2

Calibration

Before calibration:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
It is not necessary to recalibrate when moving a calibrated probe from one HQd meter to another if the additional meter is configured to use the same calibration options.
To view the current calibration, push , select View Probe Data, then select View Current Calibration.
If any two probes are connected, push the UP or DOWN arrow to change to the single display mode in order to show the Calibrate option.
Prepare the probe for use (refer to Preparation for use on page 2).
Calibration notes:
pH buffers can be used in any order. Use buffers that are two pH units apart.
Additional standard sets along with the minimum number of calibration points can be selected in the Calibration Options. For a two point calibration, it is recommended that two buffers be selected: one with a pH above and one with a pH below the expected sample pH. For a one point calibration, select the buffer nearest to the expected sample pH.
The calibration is recorded in the probe and the data log. The calibration is also sent to a PC, printer or flash memory stick if connected.
Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed.
If a calibration error occurs, refer to Troubleshooting on page 10.
1. Connect the probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn on the meter.
Calibration procedure:
2. Push Calibrate. The
display shows the buffers that are necessary for calibration.
3. Prepare the fresh buffers in separate beakers or appropriate containers.
4. Rinse the probe with deionized water. Blot dry with a lint­free cloth.
5. Put the probe in the pH buffer solution and stir gently. Make sure that the reference junctions are completely submerged. Shake the probe from side to side in the standard solution to refresh the reference junction.
3
6. Push Read.
Stir gently. The display will show "Stabilizing" and a progress bar as the probe stabilizes in the standard. The display shows the buffer that has just been read and shows the temperature corrected pH value when the reading is stable.
7. Repeat steps 4
- 6 until the minimum number of calibration points specified in the current method have been collected.
8. Push Done to view the calibration summary. The display will not show Done until the minimum number of calibration points have been collected.
9. Push Store to accept the calibration and go back to measurement mode.

Sample measurement

Before measurement:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
If complete traceability is necessary, enter a sample ID and operator ID before measurement. Refer to the HQd meter manual for more information.
Regular calibration is required for the best measurement accuracy (refer to Calibration on page 3).
Prepare the probe for use (refer to Preparation for use on page 2).
Measurement notes:
Data is automatically stored in the data log when Press to Read or Interval is selected in the Measurement Mode. When Continuous is selected, data will only be stored when Store is selected.
Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed.
If a measurement error occurs, refer to Troubleshooting on page 10.
Measurement procedure:
4
1. Connect the
probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn the meter on.
2. For the best stabilization time, condition the probe in the sample for at least 10 minutes before the initial sample measurement.
3. Rinse the probe with deionized water and then with the sample. Blot dry with a lint-free cloth.
4. Put the probe in the sample and stir gently. Make sure that the reference junctions are completely submerged. Do not put the probe on the bottom or sides of the container. Shake the probe from side to side in the sample to refresh the reference junction.
5. Push Read. The display will show "Stabilizing" and a progress bar as the probe stabilizes in the sample. The display will show the lock icon when the reading stabilizes.
6. Repeat steps 2-4 for additional measurements.
7. When measurements are done, store the probe (refer to
Storage
on page
10).

Run a check standard

The run check standard feature validates instrument performance between sample measurements. Use the run check standard feature for periodic or user-defined interval measurements of a traceable standard solution. Set the criteria for check standards from the PHC301
Note: Access control must be off or a valid password must be entered before any of the check standard method options can be changed.
1. Push
2. Select Run Check Standard.
Note: Select the correct probe if two probes are connected to the meter.
3. Get the standard solution shown on the display.
4. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
5. Put the probe in the standard solution until the temperature sensor is completely
submerged. Move the probe up or down or gently tap on the beaker to remove air bubbles from the probe.
6. Push Read. The display will show "Stabilizing" and a progress bar as the reading stabilizes. The display shows the value of the check standard and either Check Standard Passed or Check Standard Failed.
Settings menu.
. The Full Access Options menu is shown.
5
Loading...
+ 9 hidden pages