Hach-Lange PHC10130 User Manual

DOC022.52
User Manual
Conductivity Probe: Model CDC40101, CDC40103, CDC40105, CDC40110, CDC40115 or CDC40130

Safety information

Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.

Specifications

Note: Specifications are subject to change without notice.
.80022
Specifications Details
Probe type Graphite, 4-pole conductivity probe
Conductivity range 0.01 μS/cm to 200.0 mS/cm
Cell constant 0.40 cm-1 ±10%
Conductivity resolution 0.0 to 19.99 μS/cm: 0.01 μS/cm
20.0 to 199.9 μS/cm: 0.1 μS/cm
200 to 1999 μS/cm: 1
2.00 to 19.99 mS/cm: 0.01 mS/cm
20.0 to 200.0 mS/cm: 0.1 mS/cm
Conductivity accuracy ±0.5% of reading
TDS (total dissolved solids) range 0 to 50,000 mg/L as NaCl
TDS resolution 0.0 to 19.99 mg/L: 0.01 mg/L
200 to 1999 mg/L: 1 mg/L
2.00 to 19.99 g/L: 0.01 g/L
20.0 to 50.0 g/L: 0.1 g/L
TDS accuracy ±0.5% of reading
Sample salinity range 0 to 42 (ppt) (‰)
Salinity resolution 0.01 parts per thousand (ppt) (‰)
μS/cm
Salinity accuracy ±1 parts per thousand (ppt) (‰)
Temperature accuracy ±0.3 °C (±0.54 °F)
Operating temperature range -10 to 110 °C (14 to 230 °F)
Storage temperature range 5 to 40 °C (41 to 104 °F)
Minimum sample depth 45 mm (1.77 in.)
Dimensions (standard) Diameter: 15 mm (0.59 in.)
Length: 184 mm (7.24 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
1
Specifications Details
Dimensions (rugged) Diameter: 46 mm (1.81 in.)
Length: 223 mm (8.73 in.)
Cable length: 5, 10, 15 or 30 m (16.40, 32.81, 49.21 or 98.42 ft)
Cable connection M12 digital output and connector compatible with HQd meters

Product overview

The CDC401 series probe is a graphite, 4-pole conductivity probe (Figure 1). The CDC40101 or CDC40103 standard conductivity probe is available with a 1 or 3 m (3.28 or
9.84 ft) cable and is intended for laboratory use. The CDC40105, CDC40110, CDC40115 or CDC40130 rugged conductivity probe is available with a 5, 10, 15 or 30 m (16.40, 32.81, 49.21 or 98.42 ft) cable. The probe measures electrical conductivity, salinity, resistivity or total dissolved solids (TDS) in wastewater, drinking water and general applications.
Figure 1 Probe overview
1 Temperature sensor and 4-pole graphite design
conductivity cell
2 Standard probe (1 or 3 meter cable) 4 Rugged probe (5, 10, 15, or 30 meter cable)
3 Shroud (rugged model)

Calibration

Before calibration:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
It is not necessary to recalibrate when moving a calibrated probe from one HQd meter to another if the additional meter is configured to use the same calibration options.
To view the current calibration, push , select View Probe Data, then select View Current Calibration.
If any two probes are connected, push the UP or DOWN arrow to change to the single display mode in order to show the Calibrate option.
If a rugged probe, remove the shroud from the probe (refer to Remove the shroud
Calibration notes:
Do not touch the tip of the probe.
Additional conductivity standards can be selected in the Calibration Options menu.
The cell constant is derived from the calibration standard.
Do not dilute conductivity standards and samples.
on page 8).
2
The meter will automatically correct the calibration measurement to the selected reference temperature (20 or 25 °C) using the default NaCl-based, non-linear temperature coefficient. Settings can be changed in the CDC401 Calibration Options menu.
The calibration is recorded in the probe and the data log. The calibration is also sent
• to a PC, printer or flash memory stick if connected.
Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed.
If a calibration error occurs, refer to Troubleshooting on page 9.
Calibration procedure:
1. Connect the
probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn on the meter.
6. Push Read. Stir gently. The display will show "Stabilizing" and a progress bar as the probe stabilizes in the standard. The display shows the standard solution value that has just been read and shows the temperature corrected value when the reading is stable.
2. Push Calibrate. The
display shows the conductivity standard solution that is necessary for calibration.
7. Push Done to view the calibration summary.
3. Add fresh conductivity standard solution to a beaker or an appropriate container.
8. Push Store to accept the calibration and return to the measurement mode. If a rugged probe, install the shroud on the probe (refer to
Install the shroud
on page
9).
4. Rinse the probe with deionized water. Blot dry with a lint­free cloth.
5. Put the probe in the standard solution and stir gently. Make sure that the temperature sensor is completely submerged.
3
Loading...
+ 7 hidden pages