Hach-Lange PHC10115 User Manual

DOC022.52
User Manual
Gel-filled pH Probe: Model PHC10101, PHC10103, PHC10105, PHC10110, PHC10115 or PHC10130

Safety information

Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.

Specifications

Note: Specifications are subject to change without notice.
.80023
Specifications Details
Probe type Digital combination non-refillable, gel-filled probe with double junction
reference and built-in temperature sensor
pH range pH 2 to 14
pH resolution User-selectable stabilization time and resolution—Fast: 0.1, Fast: 0.01,
Medium: 0.01, Slow: 0.01 or Slow: 0.001
Slope -59 mV/pH (90 to 110% at 25 °C (77 °F) per Nernstian theoretical value)
Operating temperature range 0 to 50 °C (32 to 122 °F)
Storage temperature range 5 to 40 °C (41 to 104 °F)
Junction Open
Reference type Ag/AgCl
Sodium (alkalinity) error -0.6 pH at pH 12.6 in 1 M NaOH
Temperature accuracy ±0.3 °C (±0.54 °F)
Minimum sample depth 20 mm (0.79 in.)
Dimensions (standard) Diameter: 12 mm (0.47 in.)
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Dimensions (rugged) Diameter: 46 mm (1.81 in.)
Length: 223 mm (8.73 in.)
Cable length: 5, 10, 15 or 30 m (16.40, 32.81, 49.21 or 98.42 ft)
Cable connection M12 digital output and connector compatible with HQd meters

Product overview

The PHC101 series probe is a non-refillable, gel-filled combination pH probe with a built­in temperature sensor (Figure 1). The standard PHC10101 or PHC10103 probe comes with a 1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory use. The rugged PHC10105, PHC10110, PHC10115 or PHC10130 comes with a 5, 10, 15 or 30 m (16.40,
1
32.81, 49.21 or 98.42 ft) cable and is intended for field use. The probe provides pH measurement in wastewater, drinking water and general aqueous applications. The probe is not suitable for use with organic solvents or samples with pH less than 2.
Figure 1 Probe overview
1 Shroud (rugged model) 6 Locking ring (rugged model)
2 Probe storage cap 7 Rugged probe (5, 10, 15 or 30 meter cable)
3 Probe soaker bottle 8 Standard probe (1 or 3 meter cable)
4 Glass bulb and temperature sensor 9 Probe storage cap or soaker bottle holder
5 Reference junctions and protective tape

Preparation for use

To prepare the probe for initial use:
1. If a rugged probe, remove the shroud and probe storage cap (refer to Remove the
shroud on page 10
2. If a standard probe, turn the probe soaker bottle cap counter-clockwise to loosen the cap. Remove the soaker bottle from the probe.
3. Remove the protective tape from the reference junctions (refer to Figure 1 on page 2). Dispose of the protective tape.
4. If the probe will be used immediately, prepare the probe for calibration or sample measurement.
5. If the probe will not be used immediately, store the probe (refer to Storage on page 10).
To prepare the probe for calibration or sample measurement:
1. If a rugged probe, remove the probe storage cap.
2. If a standard probe, turn the probe soaker bottle cap counter-clockwise to loosen the
cap. Remove the soaker bottle from the probe.
3. Rinse the reference junctions and glass bulb thoroughly with deionized water to remove the 3 M KCl solution completely. Blot dry with a lint-free cloth.
4. For the best stabilization time, condition the probe for several minutes in the sample before use.
5. If a rugged probe, make sure that the shroud is installed before field use (refer to
Install the shroud on page 10).
).
2
Note: Damage to the sensing elements can occur if the shroud is not installed during field use. Damage under these conditions is not covered by the product warranty.

Calibration

Before calibration:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
It is not necessary to recalibrate when moving a calibrated probe from one HQd meter to another if the additional meter is configured to use the same calibration options.
To view the current calibration, push , select View Probe Data, then select View Current Calibration.
If any two probes are connected, push the UP or DOWN arrow to change to the single display mode in order to show the Calibrate option.
Prepare the probe for use (refer to Preparation for use on page 2).
If a rugged probe, remove the shroud from the probe (refer to Remove the shroud on page 10).
Calibration notes:
pH buffers can be used in any order. Use buffers that are two pH units apart.
Additional standard sets along with the minimum number of calibration points can be selected in the Calibration Options. For a two point calibration, it is recommended that two buffers be selected: one with a pH above and one with a pH below the expected sample pH. For a one point calibration, select the buffer nearest to the expected sample pH.
The calibration is recorded in the probe and the data log. The calibration is also sent to a PC, printer or flash memory stick if connected.
Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed.
If a calibration error occurs, refer to Troubleshooting on page 10.
1. Connect the probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn on the meter.
Calibration procedure:
2. Push Calibrate. The
display shows the buffers that are necessary for calibration.
3. Prepare the fresh buffers in separate beakers or appropriate containers.
4. Rinse the probe with deionized water. Blot dry with a lint­free cloth.
5. Put the probe in the pH buffer solution and stir gently. Make sure that the reference junctions are completely submerged. Shake the probe from side to side in the standard solution to refresh the reference junction.
3
6. Push Read.
Stir gently. The display will show "Stabilizing" and a progress bar as the probe stabilizes in the standard. The display shows the buffer that has just been read and shows the temperature corrected pH value when the reading is stable.
7. Repeat steps 4 - 6 until the minimum number of calibration points specified in the current method have been collected.
8. Push Done to view the calibration summary. The display will not show Done until the minimum number of calibration points have been collected.
9. Push Store to accept the calibration and go back to measurement mode. If a rugged probe, install the shroud on the probe (refer to
Install the shroud
on page 10).

Sample measurement

Before measurement:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
If complete traceability is necessary, enter a sample ID and operator ID before measurement. Refer to the HQd meter manual for more information.
Regular calibration is required for the best measurement accuracy (refer to Calibration on page 3).
Prepare the probe for use (refer to Preparation for use on page 2).
To deploy a rugged probe at a distance, toss the probe body with a gentle underhand throw. Do not swing the probe by the cable as this may cause injury to the user, will cause severe strain on the cable and will shorten the service life of the probe.
Measurement notes:
Data is automatically stored in the data log when Press to Read or Interval is selected in the Measurement Mode. When Continuous is selected, data will only be stored when Store is selected.
Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed.
If a measurement error occurs, refer to Troubleshooting on page 10.
Measurement procedure:
4
Loading...
+ 8 hidden pages