Hach-Lange ORBISPHERE 51x User Manual

Page 1
DOC024.52.93002
ORBISPHERE Model 51x Analyzer
User Manual
05/2013, Edition 11
Page 2
Page 3
Table of Contents
1.1 Disclaimer ....................................................................................................................................5
1.2 Contact information...................................................................................................................... 5
1.3 Safety information ........................................................................................................................ 5
1.3.1 Use of hazard information...................................................................................................5
1.3.2 Safety precautions .............................................................................................................. 6
1.3.3 Service and repairs .............................................................................................................6
1.3.4 Precautionary labels............................................................................................................ 7
1.3.5 Intended use of this equipment...........................................................................................7
1.4 Product recycling information.......................................................................................................8
1.5 Product disposal ........................................................................................................................ 10
1.6 Restriction of hazardous substances (RoHS)............................................................................ 11
Section 2 Specifications.................................................................................................................. 13
2.1 Technical specifications .............................................................................................................13
2.2 Hardware description ................................................................................................................. 15
2.3 Model identification system........................................................................................................ 15
2.4 Security level table..................................................................................................................... 18
2.5 Default parameters ....................................................................................................................19
Section 3 Installation........................................................................................................................21
3.1 Unpacking .................................................................................................................................. 21
3.2 Installation check list .................................................................................................................. 22
3.3 Wall mount and pipe mount instruments.................................................................................... 23
3.3.1 Instrument dimensions......................................................................................................23
3.3.2 Wall mounting ...................................................................................................................24
3.3.3 Pipe mounting ................................................................................................................... 24
3.3.4 Connection panel (bottom of instrument).......................................................................... 25
3.4 Panel mount instrument ............................................................................................................. 27
3.4.1 Instrument dimensions......................................................................................................27
3.4.2 Mounting ...........................................................................................................................28
3.4.3 Connection panel (bottom of instrument).......................................................................... 29
3.5 Portable table instrument ........................................................................................................... 30
3.5.1 Instrument dimensions......................................................................................................30
3.5.2 Mounting ...........................................................................................................................31
3.5.3 Connection panel (bottom of instrument).......................................................................... 31
3.6 Connectors assembly instructions ............................................................................................. 32
3.6.1 Cable gland wiring instructions .........................................................................................32
3.6.2 Ethernet connector (table version).................................................................................... 33
3.6.3 USB-B client adapter cable...............................................................................................34
3.7 Connection to mains power supply ............................................................................................ 34
3.7.1 Power supply unit (portable instrument)............................................................................34
3.7.2 Power supply connection (low voltage instruments) ......................................................... 34
3.7.3 Power supply connection (high voltage instruments)........................................................ 35
3.8 Connections to electronic boards...............................................................................................36
3.8.1 Sensor cable ..................................................................................................................... 36
3.8.2 Electronic boards connectors............................................................................................ 36
3.8.3 Main board connections....................................................................................................37
3.8.4 Measurement board..........................................................................................................38
3.9 Measurement alarm relays ........................................................................................................ 39
3.10 Sensor installation....................................................................................................................40
3.10.1 EC Sensors.....................................................................................................................40
3.10.2 TC Sensors .....................................................................................................................40
Section 4 User Interface..................................................................................................................41
4.1 Instrument .................................................................................................................................. 41
4.2 Touch screen .............................................................................................................................41
1
Page 4
Table of Contents
4.2.1 Function keys on the header bar.......................................................................................42
4.2.2 Special function keys on portable instruments ..................................................................43
4.2.3 Menu navigation ................................................................................................................43
4.2.4 Rolling list ..........................................................................................................................43
4.2.5 Virtual keyboard ................................................................................................................44
4.2.6 Identification and authorization level .................................................................................44
4.2.7 Warning windows ..............................................................................................................44
4.3 Main menu structure...................................................................................................................45
Section 5 View Menu ........................................................................................................................47
5.1 Selection of the view style..........................................................................................................48
5.2 Configuration of the view styles..................................................................................................49
Section 6 Measurement Menu........................................................................................................51
6.1 Instrument configuration.............................................................................................................52
6.2 Measurement configuration........................................................................................................53
Section 7 Calibration menu.............................................................................................................59
7.1 Definitions...................................................................................................................................61
7.2 EC gas sensor calibration ..........................................................................................................61
7.2.1 O
7.2.2 O
7.2.3 H
7.3 TC gas sensor calibration...........................................................................................................63
7.4 Calibration errors (EC and TC sensors).....................................................................................64
7.5 Interference calibration (EC sensor)...........................................................................................64
7.6 Interference calibration (TC sensor)...........................................................................................65
7.7 Barometric pressure calibration..................................................................................................65
7.8 External pressure calibration (optional sensor)..........................................................................65
7.9 Calibration reports......................................................................................................................66
sensor calibration.........................................................................................................62
2
sensor calibration.........................................................................................................62
3
sensor calibration .........................................................................................................63
2
Section 8 Inputs/Outputs Menu.....................................................................................................67
8.1 Configure snooze .......................................................................................................................68
8.2 View inputs/outputs ....................................................................................................................68
8.3 Relays ........................................................................................................................................68
8.4 Analog outputs ...........................................................................................................................69
8.5 Analog output characteristics .....................................................................................................73
Section 9 Communication Menu....................................................................................................77
9.1 RS-485 simple mode configuration ............................................................................................78
9.1.1 Data available....................................................................................................................79
9.1.2 Example of use..................................................................................................................82
9.2 PROFIBUS-DP communication (optional)..................................................................................83
9.2.1 Installation .........................................................................................................................83
9.2.2 Input/Output data...............................................................................................................84
9.3 USB-A port (host).......................................................................................................................87
9.4 HTTP/TCP-IP .............................................................................................................................88
9.4.1 Overview ...........................................................................................................................88
9.4.2 PC interface.......................................................................................................................88
9.5 Data file transfer through the USB-B port (client).......................................................................90
9.5.1 PC software installation.....................................................................................................90
9.5.2 Microsoft synchronization software ...................................................................................90
9.5.3 Upload report files .............................................................................................................90
Section 10 Security Menu................................................................................................................93
10.1 Access rights management......................................................................................................93
10.2 Configure security ....................................................................................................................94
10.3 User management....................................................................................................................94
2
Page 5
Table of Contents
10.4 User action log file ................................................................................................................... 94
Section 1 1 Products Menu.............................................................................................................. 95
Section 12 Global Configuration Menu........................................................................................ 97
Section 13 Services menu .............................................................................................................. 99
13.1 Sensor diagnostics.................................................................................................................101
13.2 Language selection................................................................................................................ 102
13.3 Clock ...................................................................................................................................... 102
13.4 Screen....................................................................................................................................102
13.5 Buzzer.................................................................................................................................... 102
13.6 Boards info............................................................................................................................. 102
13.7 Batteries................................................................................................................................. 103
13.8 Software download ................................................................................................................103
13.9 End application ......................................................................................................................103
13.10 Channel activation (multi channel versions) ........................................................................103
Section 14 Maintenance and Troubleshooting........................................................................ 105
14.1 Instrument maintenance ........................................................................................................ 105
14.2 Troubleshooting .....................................................................................................................105
14.3 List of events and alarms ....................................................................................................... 106
14.4 Storage, handling and transportation.....................................................................................107
Section 15 Part Lists...................................................................................................................... 109
15.1 Accessories............................................................................................................................109
15.2 Spare parts ............................................................................................................................110
Section 16 Glossary .......................................................................................................................111
16.1 Gas units................................................................................................................................ 111
16.2 Generic terms and definitions ................................................................................................ 111
3
Page 6
Table of Contents
4
Page 7

Section 1 General Information

1.1 Disclaimer

The information in this manual has been carefully checked and is believed to be accurate. However, Hach Lange assumes no responsibility for any inaccuracies that may be contained in this manual. In no event will Hach Lange be liable for direct, indirect, special, incidental, or consequential damages resulting from any defect or omission in this manual, even if advised of the possibility of such damages. In the interest of continued product development, Hach Lange reserves the right to make improvements in this manual and the products it describes at any time, without notice or obligation.
Copyright © 2006-2013 by Hach Lange. All rights reserved. No part of the contents of this manual may be reproduced or transmitted in any form or by any means without the written permission of Hach Lange.

1.2 Contact information

Manufacturing site:
HACH LANGE Sàrl 6, route de Compois 1222 Vésenaz SWITZERLAND Tel. +41 22 594 6400 Fax +41 22 594 6499

1.3 Safety information

Please read the entire manual before unpacking, setting up, or operating this analyzer.
Pay particular attention to all warning and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
To ensure the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that which is specified in this manual.

1.3.1 Use of hazard information

Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
European HQ:
HACH LANGE GmbH Willstätterstraße 11 40549 Düsseldorf GERMANY Tel. +49 211 52 880 Fax +49 211 52 88143
DANGER
WARNING
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially or imminently hazardous situation that may result in minor or moderate injury.
NOTICE
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.
5
Page 8
General Information

1.3.2 Safety precautions

Always remove power from the controller before performing maintenance activities.
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical connections.
Potential Electrocution Hazard. If this equipment is used outdoors or in potentially wet locations a Ground Fault Interrupt device must be used for connecting the equipment to its main power source.
DANGER
WARNING
WARNING
WARNING
Potential Electrocution Hazard. Connect only safety low voltage < 33 VAC RMS.
CAUTION
Personal Injury Hazard. Only qualified personnel should conduct the tasks described in this manual.
NOTICE
Install the device in a location and position that gives easy access to the disconnect device and its operation.
NOTICE
Potential Instrument Damage. Delicate internal electronic components can be damaged by static electricity, resulting in degraded performance or eventual failure.
The power cord plug connection is also used as a main power switch.
The instrument must be connected to an electrical system which complies with applicable local regulations.
All the cables connected to the instrument must be fire resistant, type UL94V-1
The operator must read and understand this manual before using the instrument .
The instrument will not be used as a safety device. It does not provide a security function in a hazardous process.

1.3.3 Service and repairs

None of the analyzer’s components can be serviced by the user. Only personnel from Hach Lange or its approved representative(s) is (are) authorized to attempt repairs to the system and only components formally approved by the manufacturer should be used. Any attempt at repairing the analyzer in contravention of these principles could cause damage to the analyzer and corporal injury to the person carrying out the repair. It renders the warranty null and void and could compromise the correct working of the analyzer and the electrical integrity or the CE compliance of the analyzer.
6
Page 9

1.3.4 Precautionary labels

Read all labels and tags attached to the analyzer. Personal injury or damage to the analyzer could occur if not observed.
This symbol, when noted on a product, indicates a potential hazard which could cause serious personal injury and/or death. The user should reference this instruction manual for operation and/or safety information.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or electrocution exists and indicates that only individuals qualified to work with hazardous voltages should open the enclosure or remove the barrier.
This symbol, when noted on the product, indicates that the marked item can be hot and should not be touched without care.
This symbol, when noted on the product, indicates the presence of devices sensitive to electrostatic discharge and indicates that care must be taken to prevent damage to them.
This symbol, when noted on the product, identifies a risk of chemical harm and indicates that only individuals qualified and trained to work with chemicals should handle chemicals or perform maintenance on chemical delivery systems associated with the equipment.
General Information
This symbol, if noted on the product, indicates the need for protective eye wear.
This symbol, when noted on the product, identifies the location of the connection for protective earth (ground).
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems. In conformity with European local and national regulations, European electrical equipment users must now return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
Products marked with this symbol indicates that the product contains toxic or hazardous substances or elements. The number inside the symbol indicates the environmental protection use period in years.

1.3.5 Intended use of this equipment

This high accuracy ORBISPHERE instrument is designed for gas measurement using electrochemical (EC) or thermal conductivity (TC) sensors. The instrument can be used for process and laboratory analysis, in applications such as beverage, life sciences, power generation, and the electronics industry.
ORBISPHERE 51x analyzers are available as portable, wall or pipe mount, and rack mount versions. According to the configuration, the ORBISPHERE 51x has provision for up to three patented gas phase (or dissolved gas) ORBISPHERE electrochemical and/or thermal conductivity sensors.
Note: A "Normal sensor" or "Smart Sensor" can be connected to the measurement board. A "Smart sensor" is a sensor with a non-volatile memory which allows storage of parameters (calibration coefficient, dates, etc.). When a smart sensor is connected, these parameters are read by the instrument software. The sensor can be calibrated in the lab and installed on site afterwards.
NOTICE
Use of the instrument outside of the environmental conditions described in Technical
specifications on page 13 may cause damage to the instrument but without endangering the
user.
7
Page 10
General Information

1.4 Product recycling information

ENGLISH
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment
manufacturer or supplier for instructions on how to return end-of-life equipment for proper disposal.
DEUTSCH
Elektrogeräte, die mit diesem Symbol gekennzeichnet sind, dürfen in Europa nach dem 12. August 2005 nicht mehr über die öffentliche Abfallentsorgung entsorgt werden. In Übereinstimmung mit lokalen und nationalen europäischen Bestimmungen (EU-Richtlinie 2002/96/EC), müssen Benutzer von Elektrogeräten in Europa ab diesem Zeitpunkt alte bzw. zu verschrottende Geräte zur Entsorgung kostenfrei an den Hersteller zurückgeben.
Hinweis: Bitte wenden Sie sich an den Hersteller bzw. an den Händler, von dem Sie das Gerät bezogen haben, um Informationen zur Rückgabe des Altgeräts zur ordnungsgemäßen Entsorgung zu erhalten.
FRANCAIS
A partir du 12 août 2005, il est interdit de mettre au rebut le matériel électrique marqué de ce symbole par les voies habituelles de déchetterie publique. Conformément à la réglementation européenne (directive UE 2002/96/EC), les utilisateurs de matériel électrique en Europe doivent désormais retourner le matériel usé ou périmé au fabricant pour élimination, sans frais pour l'utilisateur.
Remarque: Veuillez vous adresser au fabricant ou au fournisseur du matériel pour les instructions de retour du matériel usé ou périmé aux fins d'élimination conforme.
ITALIANO
Le apparecchiature elettriche con apposto questo simbolo non possono essere smaltite nelle discariche pubbliche europee successivamente al 12 agosto 2005. In conformità alle normative europee locali e nazionali (Direttiva UE 2002/96/EC), gli utilizzatori europei di apparecchiature elettriche devono restituire al produttore le apparecchiature vecchie o a fine vita per lo smaltimento senza alcun costo a carico dell’utilizzatore.
Nota: Per conoscere le modalità di restituzione delle apparecchiature a fine vita da riciclare, contattare il produttore o il fornitore dell’apparecchiatura per un corretto smaltimento.
DANSK
Elektriske apparater, der er mærket med dette symbol, må ikke bortskaffes i europæiske offentlige affaldssystemer efter den 12. august 2005. I henhold til europæiske lokale og nationale regler (EU-direktiv 2002/96/EF) skal europæiske brugere af elektriske apparater nu returnere gamle eller udtjente apparater til producenten med henblik på bortskaffelse uden omkostninger for brugeren.
Bemærk: I forbindelse med returnering til genbrug skal du kontakte producenten eller leverandøren af apparatet for at få instruktioner om, hvordan udtjente apparater bortskaffes korrekt.
8
Page 11
General Information
SVENSKA
Elektronikutrustning som är märkt med denna symbol kanske inte kan lämnas in på europeiska offentliga sopstationer efter 2005-08-12. Enligt europeiska lokala och nationella föreskrifter (EU-direktiv 2002/96/EC) måste användare av elektronikutrustning i Europa nu återlämna gammal eller utrangerad utrustning till tillverkaren för kassering utan kostnad för användaren.
Obs! Om du ska återlämna utrustning för återvinning ska du kontakta tillverkaren av utrustningen eller återförsäljaren för att få anvisningar om hur du återlämnar kasserad utrustning för att den ska bortskaffas på rätt sätt.
ESPANOL
A partir del 12 de agosto de 2005, los equipos eléctricos que lleven este símbolo no deberán ser desechados en los puntos limpios europeos. De conformidad con las normativas europeas locales y nacionales (Directiva de la UE 2002/96/EC), a partir de esa fecha, los usuarios europeos de equipos eléctricos deberán devolver los equipos usados u obsoletos al fabricante de los mismos para su reciclado, sin coste alguno para el usuario.
Nota: Sírvase ponerse en contacto con el fabricante o proveedor de los equipos para solicitar instrucciones sobre cómo devolver los equipos obsoletos para su correcto reciclado.
NEDERLANDS
Elektrische apparatuur die is voorzien van dit symbool mag na 12 augustus 2005 niet meer worden afgevoerd naar Europese openbare afvalsystemen. Conform Europese lokale en nationale wetgegeving (EU-richtlijn 2002/96/EC) dienen gebruikers van elektrische apparaten voortaan hun oude of afgedankte apparatuur kosteloos voor recycling of vernietiging naar de producent terug te brengen.
Nota: Als u apparatuur voor recycling terugbrengt, moet u contact opnemen met de producent of leverancier voor instructies voor het terugbrengen van de afgedankte apparatuur voor een juiste verwerking.
POLSKI
Sprzęt elektryczny oznaczony takim symbolem nie może być likwidowany w europejskich systemach utylizacji po dniu 12 sierpnia 2005. Zgodnie z europejskimi, lokalnymi i państwowymi przepisami prawa (Dyrektywa Unii Europejskiej 2002/96/EC), użytkownicy sprzętu elektrycznego w Europie muszą obecie przekazywać Producentowi stary sprzęt lub sprzęt po okresie użytkowania do bezpłatnej utylizacji.
Uwaga: Aby przekazać sprzęt do recyklingu, należy zwrócić się do producenta lub dostawcy sprzętu w celu uzyskania instrukcji dotyczących procedur przekazywania do utylizacji sprzętu po okresie użytkownia.
PORTUGUES
Qualquer equipamento eléctrico que ostente este símbolo não poderá ser eliminado através dos sistemas públicos europeus de tratamento de resíduos sólidos a partir de 12 de Agosto de 2005. De acordo com as normas locais e europeias (Directiva Europeia 2002/96/EC), os utilizadores europeus de equipamentos eléctricos deverão agora devolver os seus equipamentos velhos ou em fim de vida ao produtor para o respectivo tratamento sem quaisquer custos para o utilizador.
Nota: No que toca à devolução para reciclagem, por favor, contacte o produtor ou fornecedor do equipamento para instruções de devolução de equipamento em fim de vida para a sua correcta eliminação.
9
Page 12
General Information

1.5 Product disposal

Note: The following only applies to European customers.
Hach Lange is committed to ensuring that the risk of any environmental damage or pollution caused by any of its products is minimized as far as possible. The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC) that came into force on August 13 2005 aims to reduce the waste arising from electrical and electronic equipment; and improve the environmental performance of all those involved in the life cycle of electrical and electronic equipment.
In conformity with European local and national regulations (EU Directive 2002/96/EC stated above), electrical equipment marked with the above symbol may not be disposed of in European public disposal systems after 12 August 2005.
Hach Lange will offer to take back (free of charge to the customer) any old, unserviceable or redundant analyzers and systems which carry the above symbol, and which were originally supplied by Hach Lange. Hach Lange will then be responsible for the disposal of this equipment.
In addition, Hach Lange will offer to take back (at cost to the customer) any old, unserviceable or redundant analyzers and systems which do not carry the above symbol, but which were originally supplied by Hach Lange. Hach Lange will then be responsible for the disposal of this equipment.
Should you wish to arrange for the disposal of any piece of equipment originally supplied by Hach Lange, please contact your supplier or our After Sales Service department in Geneva for instructions on how to return this equipment for proper disposal.
10
Page 13

1.6 Restriction of hazardous substances (RoHS)

The European Union RoHS Directive and subsequent regulations introduced in member states and other countries limits the use of six hazardous substances used in the manufacturing of electrical and electronic equipment.
Currently, monitoring and control instruments do not fall within the scope of the RoHS Directive, however Hach Lange has taken the decision to adopt the recommendations in the Directive as the target for all future product design and component purchasing.
This product is compliant with the European Union RoHS Directive.
Note: The following only applies to exports of this product into the People’s Republic of China.
General Information
Locking system X
Spacer X
External connectors X
O:
X:
11
Page 14
General Information
12
Page 15

Section 2 Specifications

Specifications are subject to change without notice.

2.1 Technical specifications

OPERATING CONDITIONS
-5°C to +50°C - 1 channel instrument
Operating temperature limits
Storage temperature limits -20°C to +70°C Operating humidity limits 0 to 95% non condensing relative humidity Operating altitude From 0 to 2,000 m. (6,550 ft.) above sea level
-5°C to +45°C - 2 channel instrument
-5°C to +40°C - 3 channel instrument
EN61326-1:2006 Directive 2004/108/EEC
EMC requirements
CE compliance EN61010-1: 2010 Directive 2006/95/EEC Safety rating ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1
Enclosure ratings
Wall and panel mount
Portable
Analog current output version on the measurement board(s)
Analog voltage output version on the measurement board(s)
Note: The wall mount instrument is a Class A product. In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures.
IP 65 Totally protected against dust. Protected against low pressure jets of water from all directions.
Note: For table models, IP 65 is applicable only when a cable is connected to the ethernet socket.
NEMA4X (wall mount only) Totally protected against dust. Protected against pressure jets of water from all directions.
POWER SUPPLY
Universal 100 VAC to 240 VAC @ 50/60Hz - 40VA 10 to 30 VDC - 30W
5V supplied through an external power supply 100-240 VAC @ 50/60Hz - 1A
Optional: Battery pack with 4 hour autonomy (charges in about 6 hours)
Battery charge time can increase at T > 30°C Battery charge level indicated on display
ANALOG OUTPUTS
4-20 mA (default) or 0-20 mA (configuration with software) 3 configurable outputs Maximum load: 500 ohm Sensitivity: 20µA Accuracy: ± 0.5% (between operating temperature limits) 0- 5 V output (hardware option) 3 configurable outputs Minimum load: 10 KOhm Sensitivity: 5 mV Accuracy: ± 0.5% (between operating temperature limits)
13
Page 16
Specifications
DIGITAL OUTPUTS
Three alarm relays per measurement board 1A-30 VAC or 0.5A-50 VDC on a resistance load Configurable to Normally Open [NO] or Normally Closed [NC] contacts by changing the
Measurement alarm relays on the measurement board(s)
System alarm relay on the main board. One system board with one system relay per instrument
Thermal cut off Prevents ageing of sensors when exposed to high temperatures
Options
Wall and pipe mount (H x D x W)
Panel mount: Face (housing) (H x D x W)
Portable (H x D x W)
jumper positions.
WARNING
Potential Electrocution Hazard. Connect only safety low voltage < 33 VAC RMS.
One "instrument system alarm" relay per instrument 1A-30 VAC or 0.5A-50 VDC on a resistance load Normally closed [NC] (NO relay also available) when instrument is turned on. Opens when
a system alarm is detected, and when it does not receive any signal.
WARNING
Potential Electrocution Hazard. Connect only safety low voltage < 33 VAC RMS.
GENERAL
COMMUNICA TION
RS-485 or PROFIBUS-DP (optional) USB client USB host Ethernet 10/100 Base-T
SIZE AND WEIGHT
236.5 x 160 x 250 mm - weight 3.8 kg
9.31 x 6.30 x 9.84 ins. - weight 8.38 lbs 156 (123) x 250 x 220 (214) mm - weight 2.9 kg
6.14 (4.84) x 9.84 x 8.86 (8.43) ins. - weight 6.39 lbs 225 x 250 x 219 mm - weight 3.8 kg
8.86 x 9.84 x 8.62 ins. - weight 8.38 lbs
14
Page 17

2.2 Hardware description

The instrument hardware is made of one main board, up to 3 measurement boards - one for each measurement channel (= one per sensor), a power supply and a battery pack as options.
The main board includes the controls for power, display, the touch screen, the barometric sensor, the alarms, and communication ports. The measurement board(s) performs measurements and executes commands from the main board. It holds the "Analog outputs" and "Relays" that send information to external systems, and "Digital Inputs" that receive information from external systems such as a “Hold” input.
A hardware watchdog is activated at program start up, to check that the system is not frozen (i.e. infinite loop, system crash, etc.). If the watchdog is not refreshed by the software every minute, the measurement display, the relays and the analog outputs are frozen for up to 2 minutes. Then the reset shuts down the instrument for 10 seconds and the start-up procedure is performed. At the same time all the hardware (sensor, measurement board) are reset.

2.3 Model identification system

The analyzer identification number and the instrument serial number are located on the label on the back panel, and can be found on order confirmation and invoice papers.
The different models available are described in the following two matrices. Depending on the channel configuration, the first three digits of the model number will be either 510, 511 or 512 as illustrated below.
Specifications
510 model number matrix
510
A A A B B B C C C D D D
Channel 1
Channel 2
Channel 3
510/
Oxygen EC sensor Hydrogen EC sensor Ozone EC sensor Carbon dioxide TC sensor (max. 2 channels)
W Wall P Panel T Portable
1 100-240 VAC (for W, P and T) 2 10-30 VDC (for W and P) 3 100-240 VAC & battery (for T only)
C 0/4-20 mA V 0-5 V
0 RS-485 1 PROFIBUS-DP / RS-485
P External pressure 0 None
0 Standard K Korean language
/ 00
15
Page 18
Specifications
511 model number matrix
511
A A A D D D E E E F F F
Channel 1
Channel 2
Channel 3
511/
512 model number matrix
512
512/
1
These models are available with oxygen, nitrogen and hydrogen combinations.
The valid combinations are AE (oxygen EC and nitrogen TC), AF (oxygen EC and hydrogen TC), and EF (nitrogen TC and hydrogen TC).
1
For gas combinations, see explanation below
Channel 1
Channel 2
0/ 00
Oxygen EC sensor Carbon dioxide TC sensor (max. 2 channels) Nitrogen TC sensor (max. 2 channels) Hydrogen TC sensor (max. 2 channels)
W Wall P Panel T Portable
1 100-240 VAC (for W, P and T) 2 10-30 VDC (for W and P) 3 100-240 VAC & battery (for T only)
C 0/4-20 mA V 0-5 V
0 RS-485 1 PROFIBUS-DP / RS-485
P External pressure 0 None
0 Standard K Korean language
/ 00
W Wall P Panel T Portable
1 100-240 VAC (for W, P and T) 2 10-30 VDC (for W and P) 3 100-240 VAC & battery (for T only)
C 0/4-20 mA V 0-5 V
0 RS-485 1 PROFIBUS-DP / RS-485
P External pressure 0 None
0 Standard K Korean language
16
Page 19
Specifications
Example 1: 510 / ACD / W1C00 000
Analyzer model 510 multi channel
One oxygen EC sensor, one ozone EC sensor and one carbon dioxide TC sensor
Wall mounted
100-240 VAC
0/4 20 mA analog output
RS-485
No external pressure sensor
Standard software (English, French, German, Italian, Spanish, Russian, Japanese, Chinese)
Example 2: 511 / AE0 / P1C1P 00K
Analyzer model 511 multi channel
One oxygen EC sensor and one nitrogen TC sensor
Panel mounted
100-240 VAC
0/4 20 mA analog output
PROFIBUS-DP/RS-485
External pressure sensor
Korean language software
Example 3: 512 / EF0 / W1C0P 000
Analyzer model 512 multi channel
One nitrogen TC sensor and one hydrogen TC sensor
Wall mounted
100-240 VAC
0/4 20 mA analog output
RS-485
External pressure sensor
Standard software (English, French, German, Italian, Spanish, Russian, Japanese, Chinese)
17
Page 20
Specifications
Main 01234View 01234
iew
XXXXX
Numeric
XXXXX
Measurement
XXXXX
Graphi
c
XXXXX
Calibration X X X Statistical XXXXX Inputs/outputs
XXX
Event
XXXXX
Communication
X
X
Diagnostic
XXXXX
Securit
y
X
X
Configure
X
X
Products
XXX
Global config
X
X
Measurement 01234
Services
X
X
Start stop
XXXXX
(portable instr. only)
Config instrument
X
X
Config chan 1
X
X
Gas senso
r
01234
Config chan 2
X
X
Calibration chan 1
XXX
Config chan 3
X
X
Calibration chan 2
XXX
Measurement file
X
X
Calibration chan 3
XXX
erification chan 1
XXX
Calibration 01234
erification chan 2
XXX
Gas senso
r
XXX
erification chan 3
XXX
Interferences
XXX
Config. chan 1
X
X
Baromete
r
X
X
Config. chan 2
X
X
Ext pressure
X
X
Config. chan 3
X
X
Amplifiers
X
Reports
XXX
Inputs/Outputs 01234 Reports 0 1234
Snooze
X
X
Sensor chan 1
XXX
iew
XXX
Sensor chan 2
XXX
Relay
s
X
X
Sensor chan 3
XXX
Analog outputs
X
X
Interf. chan 1
XXX
Interf. chan 2
XXX
Communication 01234 Interf. chan 3
XXX
RS 485
X
X
Baromete
r
X
X
RS 485 simple
X
X
Ext pressure
X
X
Profibus DP
X
X
USB-
AXX
HTTP / TCPIP
X
X
Securit
y
01234
Configuration
X
Access table
X
iew log file
X
X
Products 01234
Select product
XXX
Modify produc
t
X
X
Global config 01234
Save config
X
X
Select config
X
X
Services 01234
Diagnostic
X
X
Language
X
X
Clock
X
X
Screen
X
X
Buzze
r
X
X
Boards info
X
X
Batteries
X
X
Soft download
X
More
X
Security level : 0 1 2 3 4
vailable
Optional
Not available today
Diagram Legend:

2.4 Security level table

A cross means that the user who has this user security level can access this function or setting (see User management on page 94).
Note: When not shown, the sub-levels carry the same security level as the level above.
V
18
Note: For the USB-A option in the Communication Menu, level 4 access is required to import the access table data.
V
V
A
XXX
XX XX
V V V
Page 21

2.5 Default parameters

The table below indicates the factory default configurations. The instrument has these settings when started for the first time.
Parameter Default settings Customer settings
Security Disabled Measurement
Measurement mode Continuous Data filter Disabled Sample phase Liquid Units ppm-ppb Display resolution XX.XX Storage mode Rolling buffer
Sensor Membrane 2956A Temp unit °C Pressure unit bar Calibration
Mode in air (O Hold Enabled
Analog outputs
Range 4-20 mA (0-5V) Output Gas measurement Extended mode Disabled Characteristics Linear mode
Alarm relays Disabled Thermal cutoff Enabled
Thermal cutoff temp 65°C
Calibration timer Disabled Service timer Disabled Buzzer
Screen tap Enabled Alarm sound Disabled
Display
Minigraph Enabled Temperature Disabled
, O3)
2
Specifications
19
Page 22
Specifications
20
Page 23

Section 3 Installation

This section provides necessary information to install and connect the analyzer. The installation of the analyzer should be performed in accordance with relevant local regulations.
Electrocution Hazard. Do not connect AC power to a 5 VDC powered model.
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical connections.
Potential Electrocution Hazard. If this equipment is used outdoors or in potentially wet locations a Ground Fault Interrupt device must be used for connecting the equipment to its main power source.
DANGER
WARNING
WARNING
WARNING
Personal Injury Hazard. Only qualified personnel should conduct the tasks described in this section of the manual.
Install the device in a location and position that gives easy access to the disconnect device and its operation.

3.1 Unpacking

Remove carefully the instrument and its accessories from the box and packing material, referring to the packing list included to confirm that everything has been delivered.
Potential Electrocution Hazard. A protective earth (PE) ground connection is required for both 100-240 VAC and 5 VDC wiring applications. Failure to connect a good PE ground connection can result in shock hazards and poor performance due to electromagnetic interferences. ALWAYS connect a good PE ground to the controller terminal.
CAUTION
NOTICE
NOTICE
Potential Instrument Damage. Delicate internal electronic components can be damaged by static electricity, resulting in degraded performance or eventual failure.
Please visually inspect the instrument for shipping damage. If anything is missing or damaged, contact the manufacturer or your dealer immediately.
You may want to retain the box and other packing material in case later you need to ship the instrument (see Storage, handling and transportation on page 107). Please dispose safely and ecologically of the box and packing material (if not stored for future use).
Please read through this manual thoroughly before carrying out the installation.
21
Page 24
Installation

3.2 Installation check list

To complete the installation, proceed to the following actions:
1. Follow the installation instructions in this section carefully.
2. On completion, verify once more that everything is properly connected.
3. If using the instrument with a TC sensor it is very important to ensure the purge gas supply
has been connected properly and has been turned on.
4. Turn the instrument on.
5. Set the language.
6. Set the security levels, users ID’s and passwords.
7. Configure the instrument and measurement channels.
8. If not using an external pressure sensor make sure the option is disabled (see Advanced
configuration - EC sensor on page 55).
9. Perform a barometric sensor calibration.
10. Perform an external pressure calibration (optional).
11. Perform the gas sensor(s) calibration.
12. Perform any interference calibrations.
13. Set the alarm thresholds.
The instrument should now be ready for operation. If a problem should arise, please refer initially to Troubleshooting on page 105. If the difficulty cannot be overcome, please contact your Hach Lange representative who will be happy to assist you.
22
Page 25

3.3 Wall mount and pipe mount instruments

3.3.1 Instrument dimensions

Installation
Figure 1 Wall and pipe mount instrument dimensions (in millimeters)
23
Page 26
Installation

3.3.2 Wall mounting

Attach the U bracket provided to the wall with two screws (not provided).
Tilt the instrument slightly backwards to align the bracket pins and the insertion slots, and slide the instrument onto the bracket as shown.
Insert the 2 locking screws with washers through the side slots.
Adjust instrument angle for better screen vision, and lock both side screws.

3.3.3 Pipe mounting

Figure 2 Wall mount bracket
Assemble the pipe mount bracket to the U-bracket, using the two screws provided
Attach this assembly to the pipe using two clamps (not provided) as shown on the left
The rest of the procedure is similar to the wall mount version, pictured above.
Slide the instrument onto the bracket.
Insert the 2 locking screws with washers through the side slots.
24
Figure 3 Pipe mount diagram
Adjust the instrument angle for better screen vision, and lock both side screws.
Page 27

3.3.4 Connection panel (bottom of instrument)

Cable protection shield
The cable protection located at the bottom of the instrument (shown in place in Figure 4 below) must be lowered, and eventually removed, to get access to the bottom of the instrument.
Pull the plastic protection firmly towards you, then lower it as shown in Figure 5 below. To remove it completely, push it back against the wall. Proceed in reverse order to install the cable protection. Make sure the 4 pins are securely in place.
Front panel door
A square key is provided to open the instrument front panel locks. The two locks are located on the right side of the instrument on the top and bottom panels (bottom panel indicated with the number 12 in Figure 6 on the next page).
The front panel can be easily pivoted to the left as shown below. To retain the instrument waterproof tightness, make sure the seal is clean and in good condition before closing the front panel.
Installation
Figure 4 Cable protection shield in place and
front door closed
Figure 5 Cable protection shield lowered and
front door open
25
Page 28
Installation
Figure 6 Wall and pipe mount connection panel
1. Power cable. The type of connection will
vary depending on the instrument specification (see Connection to mains
power supply on page 34)
2. USB-B client 4 pin connector. Use the
adapter cable (see USB-B client adapter
cable on page 34)
3. Ethernet cable gland
4. Sensor channel 3 cable gland (option)
5. Sensor channel 2 cable gland (option)
6. Sensor channel 1 cable gland
7. USB-A host connector for mass storage
device
8. External pressure sensor cable gland
(option)
9. Input/Output 3 cable gland (option)
10. Input/Output 2 cable gland (option)
11. Input/Output 1 cable gland
12. Keylock (bottom panel)
26
Page 29

3.4 Panel mount instrument

3.4.1 Instrument dimensions

Installation
Figure 7 Panel mount instrument dimensions (in millimeters)
27
Page 30
Installation

3.4.2 Mounting

1. Cut an opening in the panel to accommodate
the bracket frame provided (this is the same size as previous generations of ORBISPHERE type 3600 instruments).
Figure 9 Opening dimensions
Figure 8 Panel mount bracket frame
2. Install the provided frame in the opening
3. Fold the 6 tabs over the panel lips, using
adjustable joint pliers.
4. Slide the instrument in the bracket frame. The
instrument should go over the four “T” pins. Rotate the 4 fast locking screws on both sides of the front panel and slide it in.
5. Rotate the 4 fast locking screws 1/4 turn twice in
the lock direction as indicated on the side of the front panel. This locks the instrument in place on the four “T” pins.
6. To access the connections inside the
instrument, remove the instrument housing (six screws on the back panel, and slide the housing back out)
7. Pass the cables through the housing, then
through the cable gland (if applicable) and then perform the connections as detailed below.
Note: Do not forget to pass the cable through the housing before passing it through the cable gland on the back panel.
28
Page 31
Alternative instrument mounting procedure
When it is not convenient to work from the back of the panel, the instrument can be connected before fitting in the panel.
1. Install the panel support frame in the panel opening
2. Slip the cables through the panel opening
3. Remove the instrument cover
4. Slip the cables through the instrument cover
5. Slip the cables through the instrument back panel cable glands
6. Connect the cables to the instrument electronic boards
7. Tighten the cable glands
8. Reinstall the instrument cover
9. Install the instrument in the panel opening

3.4.3 Connection panel (bottom of instrument)

Installation
Figure 10 Panel mount connection panel
1. Power cable. The type of connection will
vary depending on the instrument specification (see Connection to mains
power supply on page 34)
2. Ethernet cable gland
3. USB-A host connector for mass storage
device
4. USB-B client 4 pin connector. Use the
adapter cable (see USB-B client adapter
cable on page 34)
5. Input/Output 1 cable gland
6. Sensor channel 1 cable gland
7. External pressure sensor cable gland
(option)
8. Input/Output 2 cable gland (option)
9. Input/Output 3 cable gland (option)
10. Sensor channel 3 cable gland (option)
11. Sensor channel 2 cable gland (option)
29
Page 32
Installation

3.5 Portable table instrument

3.5.1 Instrument dimensions

30
Figure 11 Portable instrument dimensions (in millimeters)
Page 33

3.5.2 Mounting

Install the instrument on a laboratory table. Place it on a clean flat surface in a safe area. Locate the instrument convenient to the power source connectors, supplies and the PC.
When carrying the table instrument, grip the top handle firmly, and gently place it on the working table. Avoid putting the instrument on the floor. The portable instrument provides two folded legs hidden in the lower frame. Pull them out to modify the display angle to adapt to user positioning (as illustrated in Figure 44 on page 107).

3.5.3 Connection panel (bottom of instrument)

Installation
Figure 12 Table version
1. Sensor channel 1 - Lemo 10 connector
2. Sensor channel 2 - Lemo 10 connector
(option)
3. Sensor channel 3 - Lemo 10 connector
(option)
4. Ethernet - Harting RJ Industrial
waterproof connector (see Ethernet
connector (table version) on page 33). An
adapter cable is available as an option.
5. Power connection for the external power
supply unit
6. On/Off power switch
7. USB-A host connector for mass storage
device
8. External pressure sensor (option). Lemo
4 connector
9. USB-B client 4 pin connector. Use the
adapter cable (see USB-B client adapter
cable on page 34).
10. Input/Output cable gland (option)
31
Page 34
Installation

3.6 Connectors assembly instructions

Potential Electrocution Hazard. In order to maintain the NEMA/IP environmental ratings of the enclosure, use only conduit fittings and cable glands rated for at least NEMA 4X/IP65 to route cables into the instrument.

3.6.1 Cable gland wiring instructions

A waterproof cable gland is provided each time a cable must be connected inside the instrument. The nickel-plated brass cable glands are EMC-types, designed so that the cable shields attach directly to the instrument housing as a ground. Typical cable wiring instructions are detailed below.
WARNING
Gland parts (washers not shown):
1. Nut
2. Rubber gasket (seal)
3. Gland fitting with O-ring (attached to
instrument housing)
1. Unscrew the cable gland nut. Inside, the assembly is composed of a rubber gasket, and two metal washers. Note that the ethernet gland on panel and wall mount instruments does not have washers and the gasket is cut.
2. If wiring a sensor cable, the cable has already been prepared so simply remove the piece of plastic protection from the exposed shielding.
For other cables, strip off external insulation as required, and 25 mm of shielding. Strip the wires about 8 mm from their ends (see illustration below).
3. Pass the cable through the nut, the rubber gasket, and the two washers
4. Pinch the shield so that its entire circumference is pressed between the two washers and
pass the cable into the housing, blocking the cable gland
32
NOTICE
It is vitally important to ensure the shielding is pinched and secured between the two washers to ensure the shielding attaches directly to the instrument housing as a ground. Failure to do this could cause damage to the instrument, and for sensor cables give incorrect readings.
5. Reattach and tighten the cable gland nut
6. Attach the wires to the corresponding terminal block connections
Page 35

3.6.2 Ethernet connector (table version)

The Harting RJ Industrial® IP 67 Push Pull connector can be locked and unlocked using one hand and minimal force. Only a few steps are necessary in order to quickly and reliably attach an Industrial Ethernet cable to the Harting RJ Industrial® connector. Only a Harting cable should be used.
Installation
1. Push the cable gland and housing over the cable
2. Strip the skin on 24 mm and the shielding screen on 13
mm
3. Prepare the individual wires for insertion into the splicing
element, according to the PROFInet® color code (see
PROFInet® color code on page 34)
4. Insert the wires into the splicing element up to the end of
the wire chambers
5. Push the splicing element onto the RJ 45 data module
and engage
6. Put the splicing element and the RJ 45 data module into
the IDC assembly tool
7. Press data module and element together with the aid of
the 1 DC assembly tool
8. Remove the terminated data module from the assembly
tool
9. Put on the upper shielding shell and press it over the
cable screen
10. Put on lower shielding shell and lock it with the upper
shell with an audible "click"
11. Push housing over the installed data module and lock it
with an audible "click"
12. Tighten cable gland
33
Page 36
Installation
Signal Function Wire color
RJ 45
pin N°
TD+ Transmission Data + Yellow 1
TD - Transmission Data - Orange 2
RD+ Receiver Data + White 3
RD - Receiver Data - Blue 6
PROFInet® color code

3.6.3 USB-B client adapter cable

This supplied cable is needed to connect the instrument to a PC.
Connect to the instrument, and connect the USB connector to the USB outlet on the user’s PC.
Figure 13 USB-B adapter cable

3.7 Connection to mains power supply

3.7.1 Power supply unit (portable instrument)

For portable instruments, an external power supply unit is supplied as standard. Connect the power supply unit to the table instrument with the FIXCON® connector provided. Plug the unit into the mains socket.
Note: Power supply unit input requirements: 100-240 VAC @ 50/60Hz - 1A

3.7.2 Power supply connection (low voltage instruments)

For low voltage instruments (10-30 VDC), connection to the mains power supply is with a 7-pin BINDER connector (supplied).
Connect the power cable to the connector as follows:
Figure 14 Wall mount BINDER connector
34
Pin Connections:
2) Power 10-30 VDC
5) Ground
7) Earth
1+ 3 + 4 + 6) Unused
Page 37

3.7.3 Power supply connection (high voltage instruments)

High voltage instruments (100-240 VAC) have a 4-pin male connector pre-wired internally with a male BINDER connector ready for mains connection. A compatible female connector is supplied with the instrument.
If this female connector was supplied with a mains power plug already pre-attached (cable part numbers 33031, 33032, 33033 and 33034) then the female connector can be plugged directly into the instrument power connector. The two connectors are grooved to avoid an incorrect fitting. Tighten the female connector to the instrument power connector finger-tight.
If no power cable was ordered with the equipment, a mains power plug must be connected to the supplied female connector as described in the following procedure.
CAUTION
Personal Injury Hazard. Only qualified personnel should conduct the tasks described in this section of the manual.
User-supplied power cable specifications:
3-wire (live, neutral and earth)
cable
wire selection ≥ 1mm
ø 7mm; 9.5mm
2
, AWG18; 2.5mm2, AWG14
Installation
WARNING
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical connections.
Prepare the user-supplied power cable as follows:
Strip off 23 mm (0.9 ins.) of shielding from the power cable. Cut back the live and neutral wires to 15 mm (0.6 ins.) in length but leave the earth wire as is. Then strip off a small amount of external insulation from the three wires as required:
Wire the female connector as follows:
1. Take the narrow end of the connector (4) in one hand and the main body (2) in the other
and unscrew the two. Pull away the cable clamp (3) and unscrew the end plug (1) to reveal the four parts that make up the connector.
2. Loosen the screws on the cable clamp (3) to allow enough room to pass the power cable
through.
35
Page 38
Installation
3. Pass the power cable through the end plug (1), the main body (2), and the cable clamp (3),
and then connect the three wires (live, neutral and earth) to the connector (4) as follows:
1) Live (brown)
2) Neutral (blue)
3) Not used
Earth (green and yellow)
4. Slide the cable clamp (3) back onto the connector (4) and tighten the screws on the clamp to secure the cable.
5. Screw the two parts (4) and (2) back together.
6. Secure the power cable by screwing the end plug (1) back in place.
7. The female connector can now be plugged directly into the instrument power connector. The
two connectors are grooved to avoid an incorrect fitting. Tighten the female connector to the instrument power connector finger-tight.

3.8 Connections to electronic boards

Potential Instrument Damage. Delicate internal electronic components can be damaged by static electricity, resulting in degraded performance or eventual failure.
Note: Any loose connection wires should be bundled tightly together with the use of nylon cable ties.

3.8.1 Sensor cable

An ORBISPHERE cable is needed to connect the sensor(s) to the instrument. The table instruments have a Lemo 10 connector where a sensor cable has to be connected. The other versions have a cable gland for cable passage, and the cable must be permanently connected to the corresponding measuring board connector.
Therefore the table instruments require a standard sensor cable, while the other versions require a sensor cable with free wires on the instrument end. The free wires are connected to the connector J8 on the corresponding measuring board, as detailed later in this chapter.
Note: The numbers and earth symbol are stamped on the end of the connector. Ensure it is connected correctly.
NOTICE
Table instrument with LEMO 10 socket on the back panel
All other versions with a cable gland for sensor cable passage
Adapter to connect a N°32505 cable (connector on both sides) to the instruments without a Lemo 10 socket on the back panel.

3.8.2 Electronic boards connectors

Connectors P8 on the main board, and connectors J7 and J8 on the measurement board(s) are made of two parts. Push down carefully the black levers on either side of the connector and pull it out securely. Perform all connections with these connectors unplugged. Once finished, attach the connectors to the boards by pushing them firmly in place (levers up).
Instrument Sensor cable
10 wire shielded Part N° 32505.mm (connector on both sides)
10 wire shielded Part N° 32501.mm (connector on sensor side only)
Part N° 32517.mm
36
Page 39

3.8.3 Main board connections

Figure 15 Main board
Installation
Figure 16 Connector P8
Connector P8:
1. RS-485 (signal A)
2. RS-485 (signal B)
3. PROFIBUS-DP (GND)
4. PROFIBUS-DP (+ 5 V)
5. PROFIBUS-DP (signal -)
6. PROFIBUS-DP (signal +)
7. PROFIBUS-DP (signal RTS) Connector P3
Ethernet RJ 45. Connect the wall and panel mount instruments to the local network by passing an ethernet cable through the ethernet cable gland (location illustrated in Figure 6 on page 26 for the wall mount and Figure 10 on page 29 for the panel mount). Connect to the P3 connector illustrated above.
Note: For portable instruments, the ethernet connection is located on the back panel (see Figure 12 on
page 31). A waterproof Harting RJ industrial socket is provided to fit to a client network cable. Accessory
adapter and cable are available as an option.
8. Not used
9. Not used
10. Not used
11. System alarm relay (N.O.)
12. System alarm relay (N.C.)
13. System alarm relay (Common)
37
Page 40
Installation

3.8.4 Measurement board

The different measurement boards for the EC and TC sensors are illustrated in Figure 17 and
Figure 18 below. The type of board is easily identified by the color of the J8 connector. For EC
boards this connector is colored orange, and for TC boards it is colored black.
It is extremely important that sensors are connected to the correct measurement board. Connecting a TC sensor to an EC measurement board (and vice versa) will cause irreparable damage to the measurement board.
Note: The colors indicated are the wire colors in the sensor cable.
NOTICE
Figure 17 EC Measurement board
Figure 19 Connector J7
Connector J7 (inputs & outputs)
Measurement alarms relays
1. Common
2. Output relay 1
3. Output relay 2
4. Output relay 3
Note: Relays N.O. or N.C. depends on the jumper position on the relays (see Measurement alarm
relays on page 39).
Figure 18 TC Measurement board
Figure 20 Connector J8
Analog current (or voltage) outputs
5. GND
6. Output 1
7. Output 2
8. Output 3
Digital inputs
9. EC sensor: Not used
TC sensor: Hold input (short circuit to pin
12)
38
10. Not used
11. Not used
Page 41
Installation
Analog inputs (when ext. press. se n so r is directly connected)
12. EC sensor: Not used
TC sensor: GND
13. Green: Input ext. press. sensor P+
14. White: Input ext. press. sensor P-
15. Red: Output ext. press. sensor +
16. Black: Ground (GND)
Note: On multi channel systems, the external pressure sensor must be wired to the channel 1 measurement board, but the signal is used to compensate all channels.
Analog inputs (when ext. press. sensor extension, Part N° 32548.xx, is used)
12. EC sensor: Not used
TC sensor: GND
13. Green: Input ext. press. sensor P+
14. Yellow: Input ext. press. sensor P-
15. White: Output ext. press. sensor +
16. Brown: Ground (GND)
Connector J8 (sensor)
Note: Remember, this connector is colored orange for EC sensors and black for TC sensors.
A1100 EC sensor 31xxx EC sensor
1)
Guard electrode
2)
RS485A+
3)
Thermistor A
4)
Anode electrode
5)
RS485B-
6)
Thermistor B
7)
GND
8)
+ 5V
9)
Cathode electrode
10)
Not used
Guard electrode Not used Thermistor A Anode electrode Not used Thermistor B Not used Not used Cathode electrode Not used
31xxxS smart EC
sensor
Guard electrode I2C-SCL Thermistor A Anode electrode I2C-SDA Thermistor B GND + 5V Cathode electrode Not used
TC sensor
GND for power V2 signal Solenoid Relay coil +12V power +24V power V3 signal GND for signal
-5V power Temperature
Cable
sensor
wall/panel
mount
Yellow
Pink
Grey
Red
Purple
White Black
Green
Blue
Brown
LEMO 10
sensor
portable
Brown
Red
Orange
Yellow Green
Blue
Purple
Grey White Black
Note: When changing the type of sensor (e.g. from a 31xxx EC sensor to a 31xxxS smart EC sensor) please contact your local Hach Lange representative.

3.9 Measurement alarm relays

Figure 21 Output relays
The three output relays are located on the measurement board (see also Figure 17 and Figure 18 on page 38).
They can be individually configured to Normally Open (NO) or to Normally Closed (NC) by physically moving the jumper on each relay. The picture example on the left (EC measurement board):
Upper relay is set to NC
Middle relay is set to NO
Lower relay is shown with no jumper, to show the 3 pins.
Note: For all measurement boards, J4 is relay 1, J5 is relay 2 and J6
is relay 3
39
Page 42
Installation

3.10 Sensor installation

3.10.1 EC Sensors

For EC sensor installation, servicing, and maintenance ensure you follow the instructions in the Sensor Installation and Maintenance manual that was supplied with the instrument.

3.10.2 TC Sensors

For TC sensor installation, servicing and maintenance ensure you follow the instructions in the TC Sensor Installation and Maintenance manual that was supplied with the instrument. Pay particular attention to the installation and connection of the purge gas supply.
Do not place the TC sensor into a liquid sample until a constant supply of dry purge gas has been connected, as liquid could condense inside the measuring chamber and cause damage to the thermal conductor chip.
To ensure the continuation of purge gas while the sensor is in contact with the sample, it is highly recommended to use a backup purge gas cylinder with an automatic changeover valve that activates when the first cylinder is empty.
NOTICE
The use of an ORBISPHERE Model 29089 gas regulator (or similar) is also recommended to deliver a constant, pressure regulated supply of dry purge gas to the sensor, filtered to 40 µm.
In addition, and to prevent any damage to the sensor electronics, the use of a purge safety backup unit (ORBISPHERE Model 32605) is highly recommended to ensure the supply of purge gas remains uninterrupted to the sensor in the event of a mains power outage.
The above ORBISPHERE accessories are explained in more detail in the TC Sensor Installation and Maintenance manual.
40
Page 43

Section 4 User Interface

4.1 Instrument

The instrument front panel provides these user interfaces:
A touch screen acting as display, touch pad and keyboard. Contrast can be adjusted.
A LED, showing when the instrument is on.
A buzzer which sounds each time the screen is touched, and when an event alarm is set.
Turning instrument On and Off
The instrument portable versions have a power switch located on the back panel. There is no power switch for the wall or panel instruments. The mains must be disconnected to turn the instrument off. The LED indicates when the instrument is on.
Measurement window
Note: The ORBISPHERE 51x can be ordered with one to three channels.
The main (numeric) measurement window continuously displays:
One to three sensor numeric values
One to three measured sensor trends (for the last 1 minute to last hour)
One to three measured sensor data alarm limits and other events
One temperature value

4.2 Touch screen

The user interface on the front panel is a 320x240 pixels display with touch screen. To make navigation user friendly, the interface software is Windows CE based, providing easy selection through menus.
Figure 22 Numeric view
Touching some items on the display calls a related function, similar to a shortcut.
All the measurement, configuration, calibration and "standard service" routines can be called by pressing buttons and menu bars on screen.
Measurement displays show one measurement per sensor in use (up to 3 lines for a 3 channel ORBISPHERE 51x).
The display can be configured to only show a single sensor measurement, or to show a parameterized graphic representation of the last measurements.
41
Page 44
User Interface

4.2.1 Function keys on the header bar

Shortcut to the user login window. Pressing this button for more than 2 seconds calls the ID and password window (see Identification and
authorization level on page 44).
Closed padlock indicates that the touch screen is locked.
Open padlock indicates that the instrument is in view mode only,
When a user is logged in, this box show the authorization level of
This icon is used for adjusting the display brightness to improve visibility. It is available all the time to any user, regardless of the user security level. This icon is a shortcut to the backlight adjustment window. See Screen on page 102
Short cut to the data storage window. Number shows the number of measurement currently stored in volatile memory.
but no user is logged in (level 0).
this user as 1, 2, 3 or 4 (4 being the highest, see User
management on page 94).
normal - snooze
No storage
Store once: When the buffer is full (10,000 positions), the
recording of measurement stops.
Rolling buffer: When the buffer is full, the latest measurement set
replaces the oldest one (first-in, first-out)
In the event of an alarm, the “snooze” button stops the instrument buzzer and returns all the relays in the instrument to their normal state during the "snooze time". The icon indicates if the alarms is on “snooze” or not. This "snooze" is configurable (see Configure
snooze on page 68).
Current date and time. This is also a shortcut to the date and time setting window.
Call the contextual menu. This menu is in the header bar and its content is related to the view displayed.
42
Opens the main menu page for easy navigation through all available menus.
Page 45

4.2.2 Special function keys on portable instruments

For the portable instruments, the state of the battery is indicated at the bottom of the measurement view. Pressing the battery icon calls the “Batteries” state window.
The green sample mode start/stop button starts the measurement process when in sample mode. Pressing the button again manually interrupts the process and an “aborted” message is displayed in the numeric view.
See Sample mode measurements (portable instrument) on page 52.
For the portable instruments, the state of the battery is indicated at the bottom of the measurement view. Pressing the battery icon calls the “Batteries” state window.
100% 75% 50% 25% empty

4.2.3 Menu navigation

User Interface
Figure 23 Main menu window

4.2.4 Rolling list

Pressing the “menu” button in the header bar calls the main menu. The display is made of three columns:
The left column is the menus, or submenus (greyed out
options are not available)
The center column shows a tree view of actual position
inside the menu structure
The right column has the generic controls detailed below.
Return to previous menu (one step back)
Jump directly to main menu
Close the menu and go back to measurement view display. If the menu button is pressed again, the menu returns to its previous state (tree structure is saved)
Help topics concerning current menu
Figure 24 Rolling list
example
For convenience, selection through a possible large list of items has been designed with a rolling list, like in this example. Use the up and down arrow to navigate, or select directly one item and press OK.
43
Page 46
User Interface

4.2.5 Virtual keyboard

When a text box (alphanumeric field) has to be edited and is pressed, a virtual keyboard appears on screen. It can be used as a PC keyboard (pressing CAP gives access to special keys).
Once values have been entered, press the Enter key to confirm and exit the virtual keyboard.
Figure 25 Virtual editing
keyboard
During the editing, the edited field name is displayed, along with units where applicable.

4.2.6 Identification and authorization level

Once the access rights have been set, (see User
management on page 94) it is necessary to log in as an
authorized user to get access to the instrument functionalities and settings.
Press the closed padlock for two seconds to open the identification window. The user identification and password must be entered to access functionalities authorized by the security level of the given user (5 levels available. See User
management on page 94).
Figure 26 Identification
window
Note: To get to level 0, press the unlock button and OK, without entering any ID or password.
For security, when the session inactivity delay period has expired (adjustable, via Configure security on page 94), the user is logged off automatically.

4.2.7 Warning windows

44
At various stages, a warning message may be displayed to request confirmation from the operator that his last action(s) must really be saved or cleared, or that there is a problem that did not enable the requested action, such as during instrument calibration (example shown left).
Figure 27 Warnings
Page 47

4.3 Main menu structure

This is the structure of the main menu which is used to control every functionality of the instrument. These submenus are detailed in the following sections of this Operator Manual.
User Interface
Figure 28 Main menu structure
45
Page 48
User Interface
46
Page 49

Section 5 View Menu

Figure 29 View menu
47
Page 50
View Menu

5.1 Selection of the view style

Numeric view
This is the default view. Display shows the numeric measurement value identified for each gas measurement channel available, a graphic showing measurement value evolution during the set time frame, and sample temperature.
This display can be configured to suit individual conditions and convenience.
Diagnostic view
The diagnostic view contains useful information for troubleshooting purposes. The amount of information displayed depends on the gas being measured and the channel configuration.
Statistic view
This feature offers statistical data that matches with Total Quality management tools. Statistics is a tool to better analyze how a process behave. The 51x statistics window gives some useful information.
The statistics are calculated from the data in the measurement file. The values are updated each time a new value is added to this file. Therefore the changes made in the configuration window are considered only once a new value is added.
Cp process capability
Cp is an index used to assess the width of the process spread in comparison to the width of the specification. It is calculated by dividing the allowable spread by the actual spread.
A Cp of one indicates that the width of the process and the width of the specification are the
same.
A Cp of less than one indicates that the process spread is greater than the specification. This
means that some of the data lies outside the specification.
A Cp of greater than one indicates that the process spread is less than the width of the
specification. Potentially this means that the process can fit inside the specification limits
CPk process variability
Cpk takes into account the center of the data relative to the specifications, as well as the variation in the process.
A Cpk value of one indicates that the tail of the distribution and the specification are an equal
distance from the overall average.
A Cpk of less than one means that some of the data is beyond the specification limit.
A Cpk greater than one indicates that the data is within the specification.
The larger the Cpk, the more central and within specification the data.
CPm process repeatability
Capability index that takes into account variation between the process average and the target. If the process average and the target are the same value, Cpm will be the same as Cpk. If the average drifts from the target, Cpm will be less than Cpk.
Skewness
negative
An asymmetric frequency distribution is skewed to the left if the lower tail is longer than the upper tail, and skewed to the right if the upper tail is longer than the lower tail. Distributions of positive-valued random variables values are often skewed right.
48
positive
Figure 30 Skewness
Page 51
Kurtosis
Kurtosis is a parameter that describes the shape of a random variable’s probability distribution.
The graphs on the left illustrate the notion of kurtosis. The lower curve has higher kurtosis than the upper curve. It is more peaked at the center, and it has fatter tails
Figure 31 Kurtosis

5.2 Configuration of the view styles

Numeric view configuration
List of parameters that can be adjusted to customize the numeric view display:
Display temperature: no, channel x
Display channel 1, 2, 3: yes/no
Mini graph confi guration button, per channel:
Select the information displayed on the grid of the minigraph
Autoranging: Set automatically the upper and lower bound to get a correct range to enable
easier reading of the measurement values.
Display time base: yes/no
Display mini graph: yes/no
Upper bound: Adjust graph upper limit
Lower boun d: Adjust graph lower limit
Time base: Adjust graph time span
Grid button: Set up the graph to display the x or y axes, the grid, or the thresholds (see below)
Auto Scale update: Automatically set the graph upper and lower bounds to best fit the actual values
displayed.
Clean button: Clear the slope displayed. The slope restarts from the left side.
View Menu
Diagnostic view configuration Select the channel to be displayed in the diagnostic window, and press OK.
Statistic view configuration
On the left side of the screen, select the channel to be displayed in the statistic window, and press OK. To configure the statistic view for a channel, select the channel number in the right side of the window, to
take you directly to the configuration window.
Scope:
Number of values: Statistic calculation range (from 10 to 10,000 values). Number of values taken in
consideration in the log file since the last value stored. The recorded values with alarms are not considered for calculation, but are part of the log file.
Histogram:
Upper limit: Select High or High High alarm value, or a custom value.
•Lower limit: Select Low or Low Low alarm value, or a custom value. Capability:
Upper limit: Select High or High High alarm value, or a custom value.
•Lower limit: Select Low or Low Low alarm value, or a custom value.
49
Page 52
View Menu
50
Page 53

Section 6 Measurement Menu

Figure 32 Measurement menu
51
Page 54
Measurement Menu

6.1 Instrument configuration

TPO or TPA calculation (portable instrument)
The TPO (Total Package Oxygen) and TPA (Total Package Air) features are available on the portable instrument for an EC sensor measuring oxygen (refer also to the table in Generic terms and definitions
on page 111 for an explanation of these two terms).
To initiate these options, the instrument must be configured in “Sample mode” (see TPO and TPA
parameters (portable instrument)below) and the TPO or TPA calculation must be enabled (see Advanced configuration - EC sensor on page 55).
The operator also needs to ensure the package is shaken for about 5 minutes before measuring, and to know the total volume of the package and the total volume of the contents of the package.
Continuous mode or sample mode (portable instrument)
Continuous mode is typically used for process measurement, whereas sample mode is aimed at lab measurements of small volume individual samples such as cans, bottles, etc. The measurement mode is a parameter defined for the entire instrument, not for each channel.
Continuous mode cycle
Every 2 sec. measurements are refreshed on the display
The relays and the analog outputs are updated
Measurements are continuously stored in memory (volatile and non volatile memory) according to
individual settings
Sample mode cycle (portable instruments)
When the user wants to measure a sample:
He prepares the sample
He starts the measurement
Measurement is stopped when all active channels have reached their "stop criteria" (defined in the
channel configuration screen - see details in Stop criteria configuration (portable instrument) on
page 53). The measurement value is stored according to individual settings. If the channel is
configured for TPO or TPA calculation the user must enter the respective parameters (see TPO and
TPA parameters (portable instrument) on page 53)
The user can then start a new measurement Continuous or sample mode selection
Select Measurement mode:
Continuous mode for on line process
Sample mode for lab sample analysis (portable instruments only)
Selection of units for barometric, external pressure and temperature Sample mode measurements (portable instrument)
When the numeric view is selected, the normal measurement window is displayed. Press the function key (on the header bar) to initiate the sample measurement.
The text "In progress" and the measurement value are displayed sequentially. The measurement process stops when the stop criteria selected are fulfilled. See Stop criteria configuration (portable
instrument) on page 53
The measurement process is stopped when:
The "stop criteria" is fulfilled, typically when the gas concentration reaches the set threshold
The function key has been pressed
The "Maximum time" criteria is reached
An error has occurred (e.g. sensor out)
When the sample measurement stops because the stop criteria are fulfilled, the gas concentration and temperature are no longer refreshed. They indicate the measurement when the stop criteria was reached. If the channel was configured for TPO or TPA calculation, enter the parameters (see TPO and
TPA parameters (portable instrument) on page 53).
Note: The mini graph is continuously refreshed.
If the sample mode is stopped for another reason (user abort, max. time, or measurement error), the “aborted” message is displayed. The reason for the stop is explained when the icon is pressed (as with any event).
52
Page 55
Measurement Menu
TPO and TPA parameters (portable instrument)
The parameters required for TPO and TPA calculation are the same.
Overflow volume: Total package size
Net content volume: Volume of liquid in the package
Press the Compute button to calculate the TPO or TPA. The parameters can be changed and the value recalculated if necessary.
TPO values are displayed in ppm, TPA values in mL. To store the measurement, press the OK button.
Stop criteria configuration (portable instrument) Press the Stop criteria button.
Note: The parameters available for configuration depend on the type of stop criteria being defined.
Available in “Sample measurement mode” (see Instrument configuration on page 52), this setting allows configuration of the stop criteria for each channel:
Above threshold: The stop criteria is met when the gas concentration is greater than the parameter
entered in "Threshold”
•Below threshold: The stop criteria is met when the gas concentration is lower than the parameter
entered in "Threshold”
Stability: The stop criteria is met when the variation of the gas concentration is smaller than the
parameter entered in "Variation"
•Time: The stop criteria is met when the elapsed time reaches the “Max. time” parameter.
Note: The number of samples taken into account to compute the "Variation" can be adjusted using the parameter "Depth".
The parameter "Max. time" is the maximum time allowed to reach the target. If “Time” is not the type
of stop criteria, then when this delay has elapsed, the measurement stops and the message “aborted” is displayed.
The time filter allows you to filter the stop criteria. The sample mode is stopped when the stop criteria
is fulfilled for a time greater than the parameter "Time Filter". As an example, if criteria is set to "Above threshold" and “Time Filter” is set to 10 sec. the
measurement stops when the gas concentration is above the "Threshold" for more than 10 seconds.

6.2 Measurement configuration

EC sensor
Sensor’s membrane number selection
•Medium: Liquid or gas phase.
Gas unit type: Partial, Fraction, Dissolved.
Gas unit: The list of available units depends on unit type selected above.
Note: This is the gas concentration measured by the EC sensor. When a composite unit is selected (e.g. ppm-ppb) the unit will change depending on the range of the value to display.
•Liquid: When medium is liquid, select water or a liquid with a different solubility (if available).
Display resolution: Maximum resolution depends on gas, membrane and unit. A maximum of 5
digits can be displayed. Decimals can be limited to 0, 1, 2 or 3 decimals for easier reading. That does not affect the actual resolution of data measured and stored, but only the data displayed.
Thermal cutoff: To protect the sensor, the thermal cutoff function allows for setting a sample high
temperature limit. If exceeded (during a Cleaning In Place cycle for example) the electrical signal to the sensor is cut off, the measurement session is suspended and the system displays a “HOT” alarm message. The system resumes when temperature drops to 90% of the specified cutoff temperature.
Thermal cut off options: Disabled / enabled.
Thermal cut off temperature: To be set according to conditions.
TC sensor
The measurement configuration for a TC sensor is the same as for an EC sensor with the addition of one extra selection criteria:
Purge gas: From the drop-down list, select the purge gas being used for the TC sensor.
53
Page 56
Measurement Menu
Measurement alarms configuration
Set the thresholds for the low/high concentration levels, according to the application. Each alarm type can be individually enabled or disabled without losing its settings.
These events can activate the relays and can be displayed.
•Low-low: 2nd stage for too low concentration
•Low: 1st stage for too low concentration
•High: 1st stage for too high concentration
High-high: 2nd stage for too high concentration
•Hysteresis: A percentage of the above concentration values. The hysteresis is used to prevent relay
Delay: The delay in seconds, before alarms go on whenever concentration values go above ”High
“flickering” when the measurement is just at the alarm levels. Set this to a minimum, but enough to eliminate flickering.
As an example, if the High Alarm is set to 40 units and the Hysteresis is set to 10%, then the High Alarm is activated once the measurement reaches 40 units, but only deactivated once the measurement drops below 36 units. With the Low Alarm the opposite is true, in that if the Low Alarm is set to 20 units and the Hysteresis set to 10%, then the Low Alarm is activated when the measurement drops below 20 units, and deactivated when the measurement rises above 22 units.
alarms” or below “Low alarms”. Set this to a minimum value, but enough to avoid alarms for non-representative peaks beyond the set level.
54
Figure 33 Alarms system diagram
Measurement filter configuration
The filters are aimed at “flattening” the measurement curve in situations where the process shows atypical peak values that could otherwise hamper the interpretation of measurement readings. The filter is applied on the last set of measurements each time a measurement is taken.
•Mean: Mathematical average of the last set (depth) of measurement values.
Median filter: Allows for eliminating atypical peak measurement values, and average the remaining
ones. The calculation sorts the last measurements set (depth) by values, then delete the highest and lowest values, and averages the remaining values (central depth).
Example for depth 7, central depth 5: Sorted values, both ends eliminated, the average of the center five is then 3.88.
0.7 - 1.1 - 4.0 - 4.3 - 4.4 - 5.6 - 7.0
Example for depth 5, central depth 3: Sorted values, both ends eliminated, the average of the center three is then 4.23.
1.1 - 4.0 - 4.3 - 4.4 - 5.6
Example for depth 8, central depth 4: Sorted values, both ends eliminated, the average of the center four is then 4.43.
0.7 - 1.1 - 4.0 - 4.3 - 4.4 - 5.0 - 5.6 - 7.0
Page 57
Measurement Menu
Advanced configuration - EC sensor
Enable pressure sensor: Check as appropriate. See External pressure calibration (optional sensor)
on page 65
Enable negative concentration: Check as appropriate. See O3 sensor calibration on page 62
TPO enable (portable instruments in “Sample mode ” measu rement only): Check if TPO
calculation is required.
TPA enable (portable instruments in “Sample mode” measurement only): Check if TPA
calculation is required.
If TPA is enabled, enter the TPA K coefficient if different from the displayed value. Advanced configuration - TC sensor
Enable ext. pressure sensor: Check as appropriate. See External pressure calibration (optional
sensor) on page 65
Enable negative concentration: Check as appropriate.
Continuous purge during thermal cut off: If thermal cutoff has been enabled (see Measurement
configuration on page 53), then check this box to ensure that a continuous purge of the TC sensor
takes place while the measurement session is suspended due to the thermal cutoff temperature value being exceeded.
Note: To manually set the TC sensor into a continuous purge mode, press the Continuous Purge button that is available from the Services - Diagnostic - Channel x - Amplifiers menu. See details in the section entitled Amplifiers (TC sensor only) on page 102.
Offset and slope corrections: Enable correction as appropriate.
If enabled, the correction values for offset and slope must be entered. These values cannot be
negative.
Liquid to gas fa ctor: Enable correction as appropriate. If checked, the percentage correction factor
must be entered. This value cannot be negative.
Note: If you believe you need to enable these corrections, it is advisable to contact a Hach Lange Service Representative first.
Interferences configuration
These options are available to take into account the influence of some components or gases in the sample during measurement. All available interference corrections are disabled by default.
The following interference corrections are available:
For oxygen measurements - CO
For hydrogen measurements - Temperature, H2O, He, O
For nitrogen measurements - Temperature, H2O, O2, H
, H2S, Chlorine, Salt, H
2
2
2
2
Note: If you believe you need to enable any of the above, it is advisable to contact a Hach Lange Service Representative first.
Interferences in oxygen measurement
In some applications, like in the beverage industry, there can be high concentrations of carbon dioxide in the sample. Hach Lange recommends using the CO2 interference option if a carbon dioxide concentration of over 1% in gas phase, or 15 ppm in dissolved phase is present.
In the petroleum industry, the detection of oxygen is sometimes hampered by significant concentrations of hydrogen sulfide in the sample. Hach Lange recommends using the H
S
2
interference option if the hydrogen sulfide concentration exceeds 0.15% in gas phase, or 5 ppm in dissolved phase. To operate the oxygen sensor in these conditions requires using a di fferent sensor and electrolyte. When using this mode your system will experience sensitivity loss of about 50 times higher than the minimum sensitivity for the membrane.
The instrument can also compensate for the presence of high hydrogen levels in the sample.
CO2 or H2S: Select CO
, H2S or All disabled.
2
Chlorinity/Salinity: Select Chlorinity, Salt or All disabled. For chlorinity or salt, it is required to
enter the actual concentration in the sample.
•H2: Select H
partial pressure of H
enabled or disabled. It is required to enter the interfering gas pressure, which is the
2
in the sample.
2
When enabled, the H2 interference must be calibrated. See Interference calibration (EC sensor) on
page 64. A warning pop-up window is displayed to remind the user of this action.
55
Page 58
Measurement Menu
Interferences in hydrogen measurement
The presence of oxygen in the sample can have an effect on the hydrogen sensor to varying degrees, depending on the membrane being used. The oxygen correction option enables the instrument to correct for this condition.
Temperature compensation: If there is a significant fluctuation in the sample temperature
Interfering H2O: This only applies to measurements taken in 100% humid gases or in traces of
Interfering O2: If the sample is known to contain oxygen, check this box.
The instrument can also compensate for the presence of helium in the sample.
Temperature compensation: As described above.
Interfering H2O: As described above
Interfering He constant: If helium is the interfering gas enable this option to allow the instrument
Interfering He: If the sample contains a known concentration of helium, check this box and enter
Interferences in nitrogen measurement
The presence of oxygen in the sample affects the nitrogen sensor. Oxygen produces about three times the effect of an equivalent amount of nitrogen in a TC sensor causing a comparably erroneous displayed value. For example, a properly calibrated nitrogen sensor, placed in air having 80% nitrogen and 20% oxygen would, uncorrected, display approximately 140% nitrogen (80% + 3*20%).
In the beverage application there is often little oxygen present, and this interference can be ignored. However, if this simplification cannot be made use this oxygen correction option to enable the instrument to correct for this condition.
Temperature compensation: If there is a significant fluctuation in the sample temperature
Interfering H2O: This only applies to measurements taken in 100% humid gases or in traces of
Interfering O2: If the sample is known to contain oxygen, check this box.
The instrument can also compensate for the presence of high hydrogen levels in the sample.
Temperature compensation: As described above
Interfering H2O: As described above
Interfering H2: If the sample is known to contain hydrogen, check this box.
enable temperature compensation to reduce the influence of these temperature fluctuations.
dissolved gas in water. It allows a correction to the slope and temperature coefficients calculated during calibration.
to correct for the presence of helium in the sample.
the concentration percentage in the box provided.
enable temperature compensation to reduce the influence of these temperature fluctuations.
dissolved gas in water. It allows a correction to the slope and temperature coefficients calculated during calibration.
56
Page 59
Measurement Menu
Measured data storage
There is one measurement file per channel which contains the data generated by the measurement cycle. The measurement files are updated in volatile memory, and regularly copied in non-volatile memory (file back-up). At start up, the measurement files in volatile memory are updated with the files from the non-volatile memory.
Note: When the measurement file is full, it is managed as a First in-First out buffer.
This dialog box allows adjustment of the parameters for recording and storing measurements. Two storage mode selections are available depending on the measurement mode selected (see Instrument
configuration on page 52).
Storage modes in continuous measurement mode:
No storage
•Store once: When the volatile memory is full (10,000 positions), the recording of measurement stops
Rolling buffer: When the volatile memory is full, the latest measurement set replaces the oldest one
continuously (first-in, first-out)
Storage modes in sample measurement mode:
Only final measurement: The measurement when the stop criteria are met is saved (one per
sample)
Only when sample mode started: The measurements when the sample mode is started are saved.
All measurements: All the measurements are stored continuously.
Note: Data stored in volatile memory are lost when instrument is off, non-volatile memory is permanent. In case of an accidental power off event, the instrument resumes measurement storage after the last measurement stored in flash.
RAM time (volatile memory): Delay in seconds between two recordings of measured data.
FLASH time (non-volatile memory): Delay in seconds between two data file transfers from volatile
memory into non-volatile memory. The last data file erases the previous one. This field is only available if the Auto save in flash box is checked.
Save in flash now: Press this button to store measurement data in flash (non-volatile memory)
immediately. After pressing this button, press OK to initiate the process. A warning screen appears informing you that the operation can take up to 30 seconds. Press Yes to continue with the process, or No to abort.
Auto save in flash: Check this box to save measurements in flash (non-volatile memory)
automatically. Measurements are saved at regular time intervals as defined in the FLASH time box.
Start logging measurement: Store once mode. Starts and stops the measurement recording
session. Measurement recording is stopped when the buffer is full.
Purge data: Clear all data for all channels in the volatile and non-volatile memories.
Open data 1, 2, or 3: Opens a table showing the measured values which are stored in the volatile
memory (RAM). Use the scroll bar at the right to move to another data range (the id range will be shown in the title bar). The page number being viewed and the total number of pages are shown at the bottom.
Use the keys at the bottom to move directly to the first page, previous page, next page or last page.
Note: If TPO or TPA calculation is enabled, a ”TPO data” or a ”TPA data” button will be available underneath the “Open data” button described above. Pressing this button will display the TPO or TPA calculated data in a screen similar to that for standard data.
57
Page 60
Measurement Menu
58
Page 61

Section 7 Calibration menu

Figure 34 Calibration menu
Note: The amplifiers calibration option is reserved for Hach Lange service technicians only, and is therefore not explained in this manual.
59
Page 62
Calibration menu
60
Figure 35 Calibration menu (cont. from previous page)
Page 63

7.1 Definitions

Definitions
We define 2 types of calibration:
the calibration of the gas to measure (e.g. O
the calibration of interferences (e.g. H2O for a CO2 sensor)
To calibrate the gas to measure (main gas), the user usually puts the sensor in the main gas without any interfering gas.
To calibrate the interferences, the user usually puts the sensor in the interfering gas without any of the main gas.
Calibrations can only be performed once the instrument has been installed, configured and each channel has been set up. You must also ensure that you have the correct access rights to access the calibration menu (minimum of level 2 for sensor calibrations and level 3 for barometric pressure calibration).
Select sensor calibration from the calibration menu, then select the channel to calibrate. There are two types of gas sensor calibration available, depending on the gas being measured and the
type of sensor being used:
1. In Air: For Oxygen and Ozone with an EC sensor.
2. Direct value: Any gas with either an EC or TC sensor. This calibration exposes the sensor to a gas
Note: All the calibration information for a smart EC Sensor is stored in the sensor’s memory. When a smart sensor is connected to the instrument for the first time, a calibration report (see example in
Calibration reports on page 66) is automatically generated giving the details of the last calibration.
If the sensor being calibrated is an EC sensor, follow the instructions in the next section entitled EC gas
sensor calibration. For TC sensors, refer to TC gas sensor calibration on page 63.
Calibration menu
for an O2 sensor)
2
with a known partial pressure, or a liquid sample with a known gas concentration.

7.2 EC gas sensor calibration

Calibration of the measured gas Start Before initiating a calibration process, the calibration parameters must be
Modify calibration parameters
set by pressing on the Modify button. The last calibration parameters are memorized, so this step can be ignored if the correct parameters are already set.
Note: When calibration is started, a calibration event is set, and the analog output and relay are put on hold (if selected below) to avoid unwanted alarms or potential process problems. This resumes when calibration is finished.
Calibration mode: 2 types available, depending on the gas being
measured:
Direct value: Any gas
•In Air (default): For O
•Medium: Select liquid or gas (direct calibration only)
Concentration unit type: Partial, fraction or dissolved (dissolved is
for calibration in a liquid only)
Concentration unit: The list of available units depends on unit type
selected above.
•Liquid: Select as appropriate, available when liquid has been selected in medium (above).
Enter the gas concentration according to the value in the calibration media, when direct value is used
Hold during calibration: On by default, this stops any output from the instrument during the calibration process to avoid sending invalid information to any connected device.
Interference enabled : If selected, this takes into account the influence of interferences (see settings below) during calibration. By default the same interference as during measurement is selected.
Press OK to start calibration
or O
2
3
61
Page 64
Calibration menu
Calibration results A calibration screen will be displayed showing current measurement data
Verification Similar to the calibration procedure, but for verification of the actual
which is continually refreshed. The value “% ideal current” is a percentage of the current against the ideal
current for the membrane type selected. If this percentage is not within the accepted range, an error message is displayed and the calibration process fails (see Calibration errors (EC and TC sensors) on page 64). A warning message can be displayed when this value is close to the boundaries, but when calibration can be accepted.
The message is first displayed in the result box. The dialog box with the error message or the warning is displayed when the finish button is pressed.
The value “% last calibration” shows the ratio between the current measurement and the previous sensor calibration.
The value “% variation” indicates the variation during the last 3 measurements, which is the stability of the measurements. A variation as low as possible is needed for a precise calibration.
The display shows the actual calibration parameters, and the actual readings (temperature, barometer, current).
Note: In case of calibration failure, consider replacing the membrane. See the sensor maintenance manual for details.
calibration values. The result of the measurements made during the verification is not stored and the actual calibration data is not modified.

7.2.1 O2 sensor calibration

O2 sensor calibration
The O2 sensor needs to be calibrated after each sensor service. Wait at least 30 minutes after mounting a new membrane before recalibrating. The sensor is in contact with either:
Air at atmospheric pressure (In Air) at known concentration (Direct value). The gas can be dissolved or not.
O
2
In air calibration This calibration procedure places the O2 sensor in water-saturated air, to
Direct calibration This procedure calibrates the oxygen sensor against a liquid sample
provide a known oxygen reference against which to calibrate. Dry the sensor thoroughly, before placing the sensor storage cap under tap
water. Shake off any excess water, but leave a few drops inside the cap. Verify that the screw-on protection cap is in place on the sensor head. If you use a Dacron mesh inside the protection cap, make sure it is dry before attempting to calibrate. Then, loosely place the storage cap back on the sensor, holding it in place with a few turns of its collar.
Set the calibration parameters accordingly (see Modify calibration
parameters on page 61), and press calibrate.
containing a known level of dissolved O The instrument displays the sensitivity of the sensor as a percentage of the
sensitivity determined when calibration was last performed. Set the calibration parameters accordingly (see Modify calibration
parameters on page 61), and press calibrate.
flowing through the sample line.
2

7.2.2 O3 sensor calibration

O3 sensor calibration
The sensor is either in contact with:
Air at atmospheric pressure (In Air)
O
The procedure is the same as for the O sensor measures O the sensor behaves in O the O
To facilitate the follow up after an O
62
at known concentration (Direct Value). The gas can be dissolved or not.
3
measurement takes a long time to stabilize.
3
sensor. In the case of the In air calibration, the
during calibration. The O3 coefficient is deduced taking into account how
2
. As a different voltage is used at the anode to measure O2 and O3,
2
2
in air calibration, negative values can be displayed.
3
Page 65

7.2.3 H2 sensor calibration

Calibration menu
WARNING
Potential Fire and Explosion Hazard. Handle H
H2 sensor calibration
The recommended method is at known concentration (Direct Value). The gas can be pure H mixture of H in the calibration parameter window. The sensor is in contact with the calibration gas in gas phase at atmospheric pressure.
Make sure the H
with an inert gas (e.g. a mixture of H2/N2). The known concentration is entered by the user
2
concentration used for calibration is within the acceptable range for the
2
membrane. Hach Lange recommends the following calibration gas/inert gas combinations for calibrating with the following membranes:
Membrane Recommended calibration gas

7.3 TC gas sensor calibration

Calibration of the measured gas Start Before initiating a calibration process, the calibration parameters must be
Modify calibration parameters
gas with great care!
2
or a
2
NOTICE
2956A 1% H 2952A 10% H2 / 90% N 2995A 100% H
29015A 100% H
set by pressing on the Modify button. The last calibration parameters are memorized, so this step can be ignored if the correct parameters are already set.
Similarly, if only the calibration value has changed, then this can be updated directly instead of pressing the Modify button.
Note: When calibration is started, a calibration event is set, and the analog output and relay are put on hold (if selected below) to avoid unwanted alarms or potential process problems. This resumes when calibration is finished.
Gas Phase: Select liquid or gas (direct calibration only)
Gas unit type: Partial, fraction or dissolved (dissolved is for
calibration in a liquid only)
•Gas unit: The list of available units depends on unit type selected
above.
•Liquid: Select as appropriate.
Enter the gas concentration according to the value in the calibration
media.
Hold during calibration: On by default, this stops any output from the
instrument during the calibration process to avoid sending invalid information to any connected device.
Automatic calibration stop: If selected, when the stability criteria is
reached, the calibration process stops automatically.
Interferences enabled: If selected, this takes into account the influence
of interferences during calibration. By default the same interferences as during measurement are selected.
Press OK to start calibration
/ 99% N
2
2
2
2
2
63
Page 66
Calibration menu
Calibration results A calibration screen will be displayed showing current measurement data
Verification Similar to the calibration procedure, but for verification of the actual
which is continually refreshed. The value “% ideal current” is a percentage of the current against the ideal
current for the membrane type selected. If this percentage is not within the accepted range, an error message is displayed and the calibration process fails (see Calibration errors (EC and TC sensors) on page 64). A warning message can be displayed when this value is close to the boundaries, but when calibration can be accepted.
The message is first displayed in the result box. The dialog box with the error message or the warning is displayed when the finish button is pressed.
The value “% last calibration” shows the ratio between the current measurement and the previous sensor calibration.
The value “% variation” indicates the variation during the last 3 measurements, which is the stability of the measurements. A variation as low as possible is needed for a precise calibration.
The display shows the actual calibration parameters, and the actual readings (temperature, barometer, current).
Note: In case of calibration failure, consider replacing the membrane. See the sensor maintenance manual for details.
calibration values. The result of the measurements made during the verification is not stored and the actual calibration data is not modified.

7.4 Calibration errors (EC and TC sensors)

Calibration errors (EC and TC sensors)
The calibration is not possible in the following circumstances:
When the "ratio ideal current" is greater than 170% or smaller than 30%
When the sensor cannot measure (thermal cut off, sensor out, etc.)
When the "ratio ideal current" is greater than 150% or smaller than 50%, a warning is displayed but
the calibration is valid.

7.5 Interference calibration (EC sensor)

Interference calibration (EC sensor)
The hydrogen interference calibration is available for EC sensors measuring O required when H lower part of the screen displays the actual measurement.
Enter in the upper box the purity of the hydrogen that is used for this calibration. A reasonably pure
(e.g. 99.8% or better) hydrogen source and an accurate pressure gauge are needed.
Hold during calibration: Selected by default, this holds the measurement sequence to avoid storing
invalid values.
Instructions Using a flow chamber, expose the sensor to the H
interference has been enabled (see Interferences configuration on page 55). The
2
barometric pressure that is shown on the display. Press Start to initiate the
interference calibration.
H
2
concentrations, and is
2
source at the same
2
64
Page 67

7.6 Interference calibration (TC sensor)

Interference calibration (TC sensor)
This calibration is available for TC sensors and is required when one of the interferences has been enabled (see Interferences configuration on page 55). The lower part of the screen displays the actual measurement.
Select from the interfering gas to be calibrated from those available, or press the Change
configuration button to reconfigure and go back to the process described in Interferences
configuration on page 55
Hold during calibration: Selected by default, this holds the measurement sequence to avoid storing
invalid values during calibration.
Instructions For interfering oxygen on measured hydrogen or nitrogen, expose both
sensors to ambient air, or a 100% oxygen source, and press Start to initiate the calibration.
For interfering hydrogen on measured oxygen or nitrogen, expose both sensors to a 100% hydrogen source and press Start to initiate the calibration.

7.7 Barometric pressure calibration

Calibration menu
Barometric pressure calibration
Upper box shows the barometric pressure measured by the instrument. Using a precision certified barometer, measure barometric pressure in the location where the measuring
instrument is used. Compare the values, if values are the same press Cancel, otherwise enter the new barometric value in the lower box and Validate the new setting.
On completion a calibration report is created.
Note: The barometric sensor has been factory calibrated.

7.8 External pressure calibration (optional sensor)

External pressure calibration
Two calibration methods for the external pressure sensor can be selected:
Two point calibration (recommended)
One point calibration
By default the two point calibration is selected.
Two point calibration (recommended)
One point calibration Connect a certified absolute pressure gauge to the sample line.
Connect a certified absolute pressure gauge to the sample line, and use a certified precision barometer.
The two point calibration starts with barometric pressure for the lower
point. Expose the external pressure sensor to the atmosphere. Enter the barometric pressure read on the barometer in the upper
window and validate P1.
Expose the external pressure sensor to line pressure, making sure it is
exposed to the same pressure as the certified absolute pressure gauge. Enter the absolute pressure value read on the certified absolute
pressure gauge in the lower box, and validate P2.
Note: Any pressure can be used for P1 and P2, but for an accurate calibration P1 and P2 should be as different as possible.
Expose the external pressure sensor to the line pressure, making sure it
is exposed to the same pressure as the certified absolute pressure gauge.
Enter the absolute pressure value read on the certified absolute pressure gauge in lower box, and validate P1.
Note: Any pressure can be used for P1, but for one point it should be as close as possible to the sample pressure
65
Page 68
Calibration menu

7.9 Calibration reports

Calibration reports
Once a calibration is completed (for a gas or pressure sensor), the calibration report is updated with the new details. The calibration report contains data for the last 50 calibrations.
For full details on the data displayed for all the different type of calibration reports, refer to the examples listed under Data available on page 79.
Each calibration record will contain parameters useful for traceability. For instance, it will contain:
the user name and ID
the date and time
the calibration coefficient
all the measurements which influence the calibration (temperature, barometric pressure, current,
etc.)
Calibration reports are generated after a sensor calibration. In addition, once a smart sensor is connected to the instrument for the first time, a calibration report is automatically generated giving details of the last calibration of that sensor.
For interfering gas calibration reports, you will be asked to select the interfering gas from those available.
66
Page 69

Section 8 Inputs/Outputs Menu

Figure 36 Inputs/Outputs menu
67
Page 70
Inputs/Outputs Menu

8.1 Configure snooze

Configure snooze
In the event of an alarm, the “snooze” button stops the instrument buzzer and returns all the relays in the instrument to their normal state during a "snooze time".
Enter the snooze time in seconds and press OK.

8.2 View inputs/outputs

View inputs/outputs
For each channel, this view option displays the state of the 3 alarm relays (on or off), and the analog output current (or voltage, depending on the instrument version) value for each.

8.3 Relays

Relays
There are three measurement alarm relays per channel and one system alarm relay for the instrument. These relays are configurable as either standard or custom events through the instrument menu.
An alarm relay can be activated or deactivated
When the alarm is OFF, it is activated,
When the alarm is ON, it is deactivated
All the relays are activated as soon as the instrument is ON (but alarms are OFF). When the instrument is OFF, the relays are deactivated, thus in this state, all alarms are ON. The logic "Relay deactivated = Alarm ON" has been chosen for this safety reason.
When the main board does not communicate with the measurement board for more than 30 seconds, the measurement board switches all the alarm relays and the analog output to the alarm state.
Relays configuration The three relays (per channel) can be triggered by several standard events,
Test channel relays The three measurement alarm relays can be manually activated for testing
Test system relay Similarly, the system alarm relay can be manually activated for testing
Note: In both the Test Channel Relays and Test System Relay options above, once the tests have been completed press Cancel to exit the screen. At the same time, this will reset all relays (including the system relay) back to a status of “Not under test”.
or a combination of events (custom). The relays output can be used to turn on a beacon, horn or PLC (see Connections to electronic boards on
page 36)
Note: Relays can be set to Normally Open [NO] or Normally Closed [NC] by changing the jumper positions on the measurement board (see
Measurement alarm relays on page 39).
Select a standard event in the rolling list
If “Custom event” has been selected, it has to be configured by touching the configure button
Press on the text box to open the selection menu (rolling menu). Select
the events that must trigger the relay, and press OK.
Proceed in the same manner for other events that should trigger the relay.
purposes:
Select Relay On, Off or Not under test.
See note regarding the relays below. “Not under test” means the relay is in operating mode, and it will be triggered normally.
Note: A relay set to NO will close when activated (On), but a relay set to NC will open. See Measurement alarm relays on page 39
purpose.
Select Relay On, Off or Not under test.
68
Page 71

8.4 Analog outputs

Inputs/Outputs Menu
Figure 37 Analog outputs menu
69
Page 72
Inputs/Outputs Menu
Analog outputs
There are three analog outputs per channel. These outputs are configurable in terms of function, content, and behavior through the instrument menus. Analog outputs are used to output a voltage or a current which is a function (e.g. a linear characteristic) of a measurement: AOut = f (M). The analog outputs can be typically connected to a PLC. Knowing the function (f), the PLC can compute the value of the measurement.
Two types of instrument hardware are available:
measurement board with current output (I = 0-20 mA or 4-20 mA).
measurement board with voltage output (U = 0-5 V). Instrument
configuration
Standard event indication
Extended event indication
Select analog output range of current:
4-20 mA or 0-20 mA
The 4-20 mA range (recommended) allows for an extended event indication mode that can be selected and configured (default = standard mode)
Note: Features of the instrument with a voltage analog output are similar to the 0-20 mA features.
For some events (sensor out, purge failure, etc.) the actual measurement is not significant, but the PLC needs to know how the analog output behaves in these cases. Two "Event indication modes" are available:
Standard mode (default)
Extended mode
Refer to Table 1 Standard event indication below.
The "Extended event indication" mode is only available when the 4-20 mA output is selected. In this mode, the range between 0 mA and 4 mA is used to indicate selected events. The events are defined using the channel configuration option (see Channel configuration on page 71).
Note: The extended mode is not available for the voltage output versions of the instrument.
Table 1 Standard event indication
Analog output
Gas concentration
Temperature 20 mA 20 mA 5 V
External pressure
Event output range
0-20 mA 4-20 mA 0/5 V
20 mA 20 mA 5 V
20 mA 20 mA 5 V
Event
Channel out
Sensor out
Thermal cut-off
Interfering gas error
Channel out
Sensor out
Channel out
External pressure sensor out
70
Page 73
Inputs/Outputs Menu
Analog outputs (continued) Channel configuration For each of the three analog output channels, set the type of measurement
that will be transmitted through this channel, and the output characteristics.
Meas. type: Select between the type of measurements available in the
rolling list.
Characteristics: Select either Linear, Tri-linear or None (see Analog
output characteristics on page 73).
Limits: Press the configure button to adjust the analog output set points
for each output. Enter values in the appropriate text boxes. In Linear mode, only the ML and MH values can be adjusted. Tri linear mode allows all limits to be adjusted, and the None mode denies access to this screen.
The authorized user may define a maximum of 12 customized events for each of the three outputs on each channel and change the order of priority of all events.
Note: This only applies to Tri linear and Linear outputs. It is not available if the output characteristic is set to None.
Configure the events that should be signaled at the corresponding current shown in the right column.
Only one event signal at a time can be sent via the current output. As
there is a possibility to have several events at the same time, an order of priority must be set. This order has been set by default, but it can be modified to suit particular needs and conditions. Touch the priority number in the left column and edit it.
The shaded events in the list have preset outputs and only the priority
can be changed. The other events can be customized by the user. Touch a white text box to call up the rolling list. Select an event from this list and press OK. Then adjust the priority as required.
Note: When an event occurs, measurement information is superseded by the event information on the output.
The following table lists the default configuration. The first three events on the list are pre-set and only the priority can be changed:
Table 2 Extended event table
Priority Event I [mA]
Sensor out 0.25
0
External pressure sensor out 0.50
1
Interfering gas error 0.75
2 3 Custom Event 1 1.00 4 Custom Event 2 1.25 5 Custom Event 3 1.50 6 Custom Event 4 1.75 7 Custom Event 5 2.00 8 Custom Event 6 2.25
9 Custom Event 7 2.50 10 Custom Event 8 2.75 11 Custom Event 9 3.00 12 Custom Event 10 3.25 13 Custom Event 11 3.50 14 Custom Event 12 3.75
71
Page 74
Inputs/Outputs Menu
Analog outputs (continued) Calibration of the analog
output
Direct test Test to check the calibration of the analog outputs. A precision amperometer
Characteristics test This is a test for the correct operation of the peripherals connected to each
The calibration of the analog output is aimed at aligning the internally
calculated current to the real current output. This was performed at factory,
but could become necessary again because of electronic tolerances. A
precision amperometer (or voltmeter for the voltage versions) connected at
the corresponding analog output connection point is required. See
Measurement board on page 38.
Select the analog output channel to calibrate and press the Start button.
Measure with the amperometer the current value for point 1. It should be
below 4 mA
Edit point 1 and enter the same value as read on the amperometer, then press the Continue button.
Measure with the amperometer the current value for point 2. It should be above 20 mA.
Edit point 2 and enter the same value as read on the amperometer, before pressing the Continue button.
Calibration of the selected analog output channel is completed. Proceed in the same way for the other analog output channels available.
connected at the analog output connection point is required.
Select a value (4, 12, 20 mA available) for each channel and compare this value (+
A calibration is required if the value on the amperometer differs from the current selected +
Note: It is possible to test one analog output without interfering with the others. During the test, the other analog outputs will continue to indicate the measurement.
analog output, by verifying that the PLC computes the correct value. The analog output will send the current corresponding to the value entered
in the text boxes.
Type in a test value for each analog output, and check for the related action on the peripheral.
0.02 mA) with what the amperometer shows.
0.02 mA).
72
Page 75

8.5 Analog output characteristics

Analog output characteristics “Linear” analog output The "Linear" output is the default setting for the analog output. It is
illustrated below (4-20 mA output is shown, 0-20 mA or 0-5 V settings are similar).
Inputs/Outputs Menu
The goal of this setting is to use all the points available on the slope from 4 mA to 20 mA to show the range of measurements that are usual in the measured process. Setting the output this way allows for the highest signal resolution for the actual conditions.
The downside is that any measures below the set range will have the same analog signal locked at 4 mA. Similarly, any measure over the set range will have the same analog signal locked at 20 mA. Settings must be made in balancing these aspects.
Settings
For the output, set ML and MH in the current measuring unit (e.g. °C for a temperature output). When a compound unit is selected the smallest unit will be used (e.g. ppb for a "ppm-ppb" compound unit).
These points should be set keeping in balance the following conditions (see illustration above):
The smaller is Range 1, the better is the analog signal resolution within
the expected range of measure.
In Range 0 the analog output only shows that measurement is below the
ML value. Similarly in Range 2 the analog output only shows that measurement is over the MH value.
The formula to compute the measurement knowing the current I (or voltage U) and the resolution R is given in the following table:
Linear Range Measurement M Resolution R
4-20 mA 20 > I > 4 M = ML + (MH - ML)
(I - 4) / 16 R = (MH-ML) / 808
*
0-20 mA 20 > I > 0 M = ML + (MH - ML) * I / 20 R = (MH-ML) / 1010
0 - 5 V 5 > U > 0 M = ML + (MH - ML)
U / 5 R = (MH-ML) / 1010
*
73
Page 76
Inputs/Outputs Menu
Analog output characteristics (continued) “Tri-linear” analog
output
The "Tri-linear" output brings benefits over the “Linear output” discussed before. It is illustrated below (4-20 mA output is shown, 0-20 mA or 0-5 V settings are similar).
Compared to the “Linear” mode, the expected range of measure is Range 2. A Range 1 and 3 are available to show the measures falling out of this Range 2, but normally at a lower resolution. Expected measurements for the measured process are supposed to be in Range 2 most of the time, and in Range 1 or 3 occasionally (problems, calibration, line stop, etc.). The benefits are:
The PLC can compute the measurement over a large range (1, 2 and 3).
The PLC can compute a higher resolution signal for the expected
measuring range (Range 2: MH > M > ML).
Carefully selecting the set points allows for an individual resolution for each range, so a different resolution can be applied to Range 1, 2 and 3, allowing to tailor the analog output to the actual conditions.
As before, the downside is that any measure below or over the Range 1, 2 and 3 will have the same signal locked at 4 mA and 20 mA respectively, but Range 1, 2 and 3 should cover a larger range than in the “Linear” mode. Settings must made in balancing these aspects.
Settings
For each output, set MLL, ML, MH, and MHH in the current measuring unit (e.g. °C for a temperature output). When a compound unit is selected, the smallest unit will be used (e.g. ppb for a "ppm-ppb" compound unit). Also set AOL (Analog Output Low) and AOH (High) in mA (or Volts).
These points should be set keeping in balance the following conditions (see illustration above):
The smaller is Range 2, the better is the analog signal resolution within the expected range of measure.
Size of Range 1 and 3 should be set to deliver an adequate level of resolution for the measures falling out of the expected range of measure.
In Range 0 the analog outputs only shows that measurement is below the MLL value. Similarly in Range 4 the analog output only shows that measurement is over the MHH value.
The formula to compute the measurement knowing the current or the voltage and the resolution R is given in the following table:
74
Page 77
Inputs/Outputs Menu
Tri-linear Range Measurement M Resolution R
1: AOL >
4-20 mA
0-20 mA
0-5 V
Analog output characteristics (continued) None This is the default value.
2: AOH > I > AOL M=ML+(MH-ML)*(I-AOL)/ (AOH-AOL) R=(MH-ML)*20/((AOH-AOL)*1010)
3: 20 > I > AOH M=MH+(MHH-MH)
1: AOL >
2: AOH >
3: 20 > I > AOH M=MH+(MHH-MH)*(I-AOH)/(20-AOH) R=(MHH-MH)*20/((20-AOH)*1010)
1: AOL >
2: AOH >
3: 5 > U > AOH M=MH+(MHH-MH)*(U-AOH) / (5-AOH) R=(MHH-MH)*5/((5-AOH)*1010)
I > 4 M=MLL+(ML-MLL)*(I-4)/(AOL-4) R=(ML-MLL)*20/((AOL-4)*1010)
I > 0 M=MLL+(ML-MLL)*I/AOL R=(ML-MLL)*20/(AOL*1010)
I > AOL M=ML+(MH-ML)*(I-AOL)/ (AOH-AOL) R=(MH-ML)*20/((AOH-AOL)*1010)
U > 0 M=MLL+(ML-MLL)*U/AOL R=(ML-MLL)*5/(AOL*1010)
U > AOL M=ML+(MH-ML)*(U-AOL)/ (AOH-AOL) R=(MH-ML)*5/((AOH-AOL)*1010)
Setting the analog output to “None” means that the output value will always be zero and importantly ensures that no current is emitted on that output, so reducing power consumption as well as reducing heat within the instrument.
(I-AOH) / (20-AOH) R=(MHH-MH)*20/((20-AOH)*1010)
*
75
Page 78
Inputs/Outputs Menu
76
Page 79

Section 9 Communication Menu

Figure 38 Communication menu
The external RS-485 port of the main board is directly connected to a RS-485 bus (single twisted pair). Optionally it can be connected to a fieldbus module (gateway).
The RS 485 menu allows to select between RS485 simple or Profibus DP communication protocol, depending on application.
Press on the text box to select either the RS-485 simple or the PROFIBUS-DP
communication protocol.
77
Page 80
Communication Menu

9.1 RS-485 simple mode configuration

This protocol allows the instrument to output data to an external device (PLC, SCADA, PC, etc.). The communication is unidirectional. The data are output on the RS-485 link as simple ASCII text. If for instance you use a PC, the data can be easily visualized and saved in a file using the "Hyperterminal" software.
To use this communications mode, on the instrument:
Select the RS-485 option from the Communication menu
Choose the protocol Simple (default configuration) and press on OK
Then select the RS-485 Simple option from the Communication menu:
"Baud rate", "Parity", "No of stop bits", "Byte size" Standard parameters of the RS-485
link.
"Enable" The measurements can be sent continuously (approximately every 2 sec.).
This field allows enabling or disabling this feature.
"Mode" This is the format of the measurements sent continuously (see Cyclic
measurements on page 79 of this document). In "Expert" mode, more data are sent.
These additional data can be useful for diagnostic purpose.
Note: In case of problem verify first that jumper J3 is not installed on the mother board (default configuration).
Send data
This dialog box is used to send text files to an external device. The possible files are the following:
Calibration reports
User actions log file
Instrument configuration
Measurements stored in the instrument memory.
The button "Stop Cyclic meas." allows to stop and to restart the cyclic transmission of measurements. It is advised to stop the cyclic transmission in order not to mix cyclic measurements and data of the file being transmitted. This button has the same effect as the "Enable" checkbox of the "Communication/RS-485 Simple/Configuration" window.
After stopping the cyclic measurements, select the "Calibration Reports", "User Actions", "Configuration", "Measurements" button to send the corresponding file, or the "All files" button to send all these files in one shot.
Note: The measurements file includes TPO or TPA measurements if these calculations are enabled.
Once the button is pressed, the file is sent immediately. The field "Current file status" shows “Sending” alongside the file transmission progress bar. On completion this changes to “Sent”.
78
Page 81

9.1.1 Data available

All individual data are separated by at least one tabulation character (ASCII code=0x09).
For the cyclic measurements, the data format is detailed. For the files, only one example for each file is given to explain the data format.
Cyclic measurements
1. If the option "Mode = standard" is chosen, the following message is sent for each active
channel:
CHn\t Gas\t Gas Unit\t Temperature\t Temperature Unit\t Barometric Pressure\t Barometric Pressure Unit\t\ Event\t\r\n
with:
\t ................................................the ASCII tab character: code=0x09
\r ................................................the ASCII Carriage Return character: code=0x0D
\n ..............................................the ASCII Line Feed character: code=0x0A
CHn ..........................................the 2 ASCII characters "CH" + the channel no (from 1 to 3).
Gas............................................the gas concentration.
Gas Unit ....................................the gas unit.
Temperature..............................the temperature.
Temperature Unit.......................the temperature unit.
Barometric Pressure..................the barometric pressure.
Barometric Pressure Unit ..........the barometric pressure unit.
Event .........................................the event bit mask in hexadecimal format.
Communication Menu
The values are not described here (see List of events and alarms on page 106).
Example of one measurement:
CH1 697.176 mbar 20.1 °C 0.982 bar C00
2. If the option "Mode = expert" is chosen, the following message is sent:
CHn\t Gas\t Gas Unit\t Temperature\t Temperature Unit\t BarometricPressure\t Barometric Pressure Unit\t
Event\t Current\t µA\t Partial pressure\t bar\t External Pressure\t External Pressure Unit\t Time\t Index\r\n
with:
Current ......................................The current measured in [microA].
Partial pressure .........................The partial pressure in [bar].
External Pressure .....................The external pressure.
External Pressure Unit...............The external pressure unit.
Time ..........................................The time of the measurement. Format "hh:mm:ss."
Index..........................................This is the index of the last measurement.
This number starts at 0 at power up of the program.
Example of one measurement:
CH1 697.173 mbar 20.1 °C 0.982 bar C00 80.056229 µA 0.697 bar 1.000 bar 12:59:42 5923
79
Page 82
Communication Menu
EC gas sensor calibration report example
Calibration report nb 1
Type . . . . . . . . . . . . . . . . . . . . .Sensor calibrated on this instrument
Date (yy.mm.dd - hh:mm) . . . . .05.02.17 - 18:40
Operator . . . . . . . . . . . . . . . . . .jp
Operator ID . . . . . . . . . . . . . . . .3
Calibration coefficient . . . . . . . .122.40 µA/bar
Ratio with ideal membrane . . . . 92.72 %
Ratio with last calibration . . . . .97.35 %
Stability . . . . . . . . . . . . . . . . . . .0.000 %
Calibration mode . . . . . . . . . . . .Direct value
Calibration value . . . . . . . . . . . .0.750000
Gas unit. . . . . . . . . . . . . . . . . . .bar
Gas phase. . . . . . . . . . . . . . . . .Liquid
Liquid. . . . . . . . . . . . . . . . . . . . .Water
Interferences . . . . . . . . . . . . . . .Disable
Temperature . . . . . . . . . . . . . . .20.1 °C
Barometric pressure . . . . . . . . .1.020 bar
External pressure . . . . . . . . . . .0.976 bar
Sensor current. . . . . . . . . . . . . .79.857 µA
Smart EC gas sensor connection calibration report example
Calibration report nb 1
Type . . . . . . . . . . . . . . . . . . . . .Sensor calibrated on another instrument
Date (yy.mm.dd - hh:mm) . . . . .05.01.17 - 15:30
Operator . . . . . . . . . . . . . . . . . . Armstrong
Calibration coefficient . . . . . . . .62.75 µA/bar
Connection date . . . . . . . . . . . .05.03.15 - 12:26
Serial number . . . . . . . . . . . . . .1
Membrane . . . . . . . . . . . . . . . . .2956A
TC gas sensor calibration report example
Calibration report nb 1
Date (yy.mm.dd - hh:mm) . . . . .06.02.18 - 15:20
Operator . . . . . . . . . . . . . . . . . .jp
Operator ID . . . . . . . . . . . . . . . .3
Calibration coefficient . . . . . . . .18.23 mV/bar
Ratio with ideal membrane . . . . 107.22 %
Ratio with last calibration . . . . .106.32 %
Stability . . . . . . . . . . . . . . . . . . .0.337 %
Calibration value . . . . . . . . . . . .965.000000
Gas unit. . . . . . . . . . . . . . . . . . .mbar
Gas phase. . . . . . . . . . . . . . . . .Gas
Liquid. . . . . . . . . . . . . . . . . . . . .Water
Interferences . . . . . . . . . . . . . . .Disable
Temperature . . . . . . . . . . . . . . .26.44 °C
Barometric pressure . . . . . . . . .0.962 bar
External pressure . . . . . . . . . . .0.976 bar
Sensor slope . . . . . . . . . . . . . . .18.797 mV/s
80
Barometric sensor calibration report example
Calibration report nb 1
Date (yy.mm.dd - hh:mm) . . . . .05.02.16 - 20:38
Operator . . . . . . . . . . . . . . . . . .jp
Operator ID . . . . . . . . . . . . . . . .3
Former barometric pressure . . .0.970 bar
New Barometric pressure . . . . .0.971 bar
Page 83
External pressure calibration report example
Calibration report nb 1
Date (yy.mm.dd - hh:mm) . . . . . 05.02.16 - 21:37
Operator . . . . . . . . . . . . . . . . . . Armstrong
Operator ID . . . . . . . . . . . . . . . .1007
Nb of cal. points . . . . . . . . . . . .1
New slope . . . . . . . . . . . . . . . . .45.454545 [bar/V]
New offset . . . . . . . . . . . . . . . . . 0.103926 [bar]
Former P1 value . . . . . . . . . . . . 0.973 bar
New P1 value . . . . . . . . . . . . . . 0.972 bar
Interfering gas calibration report example
Calibration report nb 1
Date (yy.mm.dd - hh:mm) . . . . . 05.02.17 - 20:03
Operator . . . . . . . . . . . . . . . . . . jp
Operator ID . . . . . . . . . . . . . . . .3
Calibration Coefficient. . . . . . . . 366.127661
H
gas purity . . . . . . . . . . . . . . .2.500000 %
2
Ratio with last calibration . . . . .99.990075 %
% of ideal current . . . . . . . . . . . 111.624287 %
Stability . . . . . . . . . . . . . . . . . . . 0.000000 %
Temperature . . . . . . . . . . . . . . . 20.1 °C
Barometric pressure . . . . . . . . .1.020 bar
External pressure . . . . . . . . . . . 0.977 bar
Sensor current . . . . . . . . . . . . . 7.990 µA
Communication Menu
User action log file example
The "User action log file" below contains 3 user actions.
Nr mm/dd hh:mm:ss User ID User Name Action ID Description
1 1/21 15:13:44 1007 Armstrong 140 Measurement config. 0 1/21 15:13:27 1007 Armstrong 132 Identification 2 1/21 15: 9:15 1007 Armstrong 132 Identification
Measurement file example
6 measurements are described below:
Nr mm/dd hh:mm:ss Gas Temp Mask Fluor. phi Barom Ext P. Index
[ppb] [°C] [°] [bar] [bar]
0 2/17 21:15:37 75.051 20.1 400 26.039 1.005 0.000 2271 1 2/17 21:15:27 75.043 20.1 400 26.045 1.005 0.000 2266 2 2/17 21:15:17 75.047 20.1 400 26.052 1.005 0.000 2261 3 2/17 21:14:57 75.044 20.1 400 26.041 1.005 0.000 2256 4 2/17 21:14:47 75.047 20.1 400 26.038 1.005 0.000 2251 5 2/17 21:14:37 75.050 20.1 400 26.054 1.005 0.000 2246
81
Page 84
Communication Menu
Configuration report example
The "Configuration" below is given for a one channel instrument.
INSTRUMENT CONFIGURATION
Measurement mode . . . . . . . . .Continuous mode
Pressure unit . . . . . . . . . . . . . . .[bar]
Temperature unit. . . . . . . . . . . .[°C]
Storage mode . . . . . . . . . . . . . .Rolling buffer
Storage RAM time. . . . . . . . . . .10 [s]
Storage FLASH time . . . . . . . . .3600 [s]
Channel 1
Membrane . . . . . . . . . . . . . . . . .2956A
Medium . . . . . . . . . . . . . . . . . . .Liquid
Gas unit. . . . . . . . . . . . . . . . . . .mbar
Liquid. . . . . . . . . . . . . . . . . . . . .Water
Resolution displayed . . . . . . . . .3
Thermal cut off . . . . . . . . . . . . .Enabled60.0 [°C]
Alarms
Low Low . . . . . . . . . . . . . . . . . .Disabled0.000000 [mbar]
Low . . . . . . . . . . . . . . . . . . . . . .Disabled0.100000 [mbar]
High. . . . . . . . . . . . . . . . . . . . . .Enabled5.000000 [mbar]
High High . . . . . . . . . . . . . . . . .Disabled1.100000 [mbar]
Hysteresis . . . . . . . . . . . . . . . . .1 [%]
Delay . . . . . . . . . . . . . . . . . . . . .0 [s]
Filter State. . . . . . . . . . . . . . . . .Disabled
Type . . . . . . . . . . . . . . . . . . . . .Median
Depth. . . . . . . . . . . . . . . . . . . . .5
Central depth. . . . . . . . . . . . . . .3
Interferences CO
or H2S . . . . . . . . . . . . . . . .All disabled
2
Chlorinity/Salinity. . . . . . . . . . . .All disabled19.000000 [g/l]
H
. . . . . . . . . . . . . . . . . . . . . . .Disabled0.100000 [bar]
2

9.1.2 Example of use

In this example we use:
One PC with a RS232 port.
One "RS-485<->RS232 converter"
Procedure:
1. Connect both RS-485 wires of the instrument to the "RS-485<->RS232 converter".
2. Connect the "RS-485<->RS232 converter" to the PC RS232 port using a standard cable
(RS232 DB9 straight cable).
On the PC:
1. Run "Hyperterminal" on the PC.
2. Configure the PC COM port used (e.g. COM2). Menu "File/Properties/Configure".
3. Configure the parameters "Baud rate", "Parity", "Nb of stop bits", "Byte size" (Menu
"File/Properties/Configure"). Use the same parameters for the instrument and the PC.
4. Configure the "Font = Courier 10" (Menu "View/Font").
5. Connect "Hyperterminal" (Menu "Call/Call").
6. Save the data received in the file of your choice (Menu "Transfer/Capture Text/Start").
On the instrument:
1. Use the menu "Communication/RS-485 Simple/Send files" and the button "All files". When
the transfer is finished, close the file with "Hyperterminal" (Menu "Transfer/Capture Text/Stop"). Now, all the reports are saved in a text file on your PC.
82
Page 85

9.2 PROFIBUS-DP communication (optional)

9.2.1 Installation

On the ORBISPHERE CD, there is an “Orbi3218.gsd” and an “Orbi3218.bmp” file available in the “Profibus DP” folder to help configure the PROFIBUS-DP. The GSD file contains the following elements:
The module Gateway Version >= 2.0 - n channel for receiving data from the instrument, and where n is 1, 2 or 3 relating to the number of channels in use.
Note: Gateway version >= 2.0 and user software version >= 2.15 are mandatory
The module Gateway Version < 2.0 for receiving data from an instrument equipped with a profibus gateway version < 2.0 or user software < 2.15. In this case, data from 3 channels are sent whatever the number of channels configured for the instrument.
Communication Menu
WARNING
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical connections.
CAUTION
Personal Injury Hazard. Only qualified personnel should conduct the tasks described in this section of the manual.
NOTICE
Potential Instrument Damage. Proper ESD (electrostatic discharge) protocols must be followed to prevent damage to the product. All fittings must be properly seated and tightened to prevent any water and dust ingress.
1. Install the PROFIBUS-DP module and the jumper J3 on the main board (location highlighted in the illustration below).
2. Select the menu "Configuration/RS-485" and choose "PROFIBUS-DP" as the protocol.
3. Select the menu "Configuration/PROFIBUS-DP", choose the slave address and restart the
instrument.
83
Page 86
Communication Menu

9.2.2 Input/Output data

The main board:
Writes the latest measurement data to the Profibus Input Buffer.
Checks if a command written by the Profibus Master must be executed (Profibus Output
Buffer). If a command is to be executed, the instrument executes it and writes the result (status, data, etc.) in the Profibus Input Buffer.
All numbers are coded in “Big Endian” format, and float values are coded according to IEEE Standards. The field types “Byte” and “Double Word” are unsigned.
Measurements
Measurements are formatted in the Profibus Input Buffer as follows:
Barometric pressure Input float 32 bits 0 Barometric pressure unit Input byte 8 bits 4 Channel 1 gas concentration Input float 32 bits 5 Channel 1 gas unit Input byte 8 bits 9 Channel 1 temperature Input float 32 bits 10 Channel 1 temperature unit Input byte 8 bits 14 Channel 1 external pressure Input float 32 bits 15 Channel 1 external pressure unit Input byte 8 bits 19 Channel 1 events Input double word 32 bits 20 Channel 1 measurement index Input double word 32 bits 24 Channel 2 gas concentration Input float 32 bits 28 Channel 2 gas unit Input byte 8 bits 32 Channel 2 temperature Input float 32 bits 33 Channel 2 temperature unit Input byte 8 bits 37 Channel 2 external pressure Input float 32 bits 38 Channel 2 external pressure unit Input byte 8 bits 42 Channel 2 events Input double word 32 bits 43 Channel 2 measurement index Input double word 32 bits 47 Channel 3 gas concentration Input float 32 bits 51 Channel 3 gas unit Input byte 8 bits 55 Channel 3 temperature Input float 32 bits 56 Channel 3 temperature unit Input byte 8 bits 60 Channel 3 external pressure Input float 32 bits 61 Channel 3 external pressure unit Input byte 8 bits 65 Channel 3 events Input double word 32 bits 66 Channel 3 measurement index Input double word 32 bits 70
Name Type Size Offset
84
Note: The table above shows information for a 3-channel instrument. This will be adjusted according to the number of channels in use.
Page 87
Communication Menu
The gas, temperature and barometric pressure unit values are coded as defined in the following tables:
Gas unit Value Temperature unit Value
bar 0 K 0 bar 0 mbar 1 °C 1 mbar 1 Pa 2 °F 2 psia 2 kPa 3 atm. 3 hPa 4 Pa 4 psia 5 kPa 5 atm. 6 hPa 6 mbar->bar 9 Pa->KPa 10 %Vbar 12 ppm Vbar 13 %Vext 14 ppm Vext 15 ppm Vbar->%Vbar 16 ppm Vext->%Vext 17 ppm 18 ppb 19 g/l 20 mg/l 21 µg/l 22 %O
2
%Air 24 Note: For the field “Event”, please see the column “Bit mask g/kg 25 V/V 26 %W 27 cc/kg 28 ml/l 29
23
value” in Table 4 on page 106.
Barometric
pressure unit
Value
Note: If the instrument stops sending measurement data to the module, then after 30 seconds the module sets the event mask to the value PROFIBUS-DP value not updated (0x80000000) bit mask.
85
Page 88
Communication Menu
Commands
The “Command Output Buffer” is formatted as follows:
Output command toggle (OCT) Output byte 8 bits 0 Output command ID (OCI) Output byte 8 bits 1 Output command data byte 1 (OCD1) Output byte 8 bits 2 Output command data byte 2 (OCD2) Output byte 8 bits 3 Output command data byte 3 (OCD3) Output byte 8 bits 4 Output command data byte 4 (OCD4) Output byte 8 bits 5
The “Command Input Buffer” is located just after the measurement data and is formatted as follows:
Input command toggle (ICT) Input byte 8 bits 74 Input command status (ICS) Input byte 8 bits 75 Input command data byte 1 (ICD1) Input byte 8 bits 76 Input command data byte 2 (ICD2) Input byte 8 bits 77 Input command data byte 3 (ICD3) Input byte 8 bits 78 Input command data byte 4 (ICD4) Input byte 8 bits 79
Name Type Size Offset
Name Type Size Offset
The following commands are available:
Change product
Activate sensor (valid for EC sensors only)
Change product command - output
Name Value Comment
OCT 1-2
OCI 1
Channel number:
OCD1 0
OCD2 0-99 Product number
OCD3 0-1
OCD4 Not used
0 = Channel 1 1 = Channel 2 2 = Channel 3
Erase measurement files: 0 = Never erase the measurement files. 1 = Erase measurement file if necessary (e.g. gas unit changes)
Change product command - input
Name Value Comment
ICT 1-2
0 = OK
ICS 0-3
ICD1 Not used ICD2 Not used ICD3 Not used ICD4 Not used
1 = Unknown command ID 2 = Invalid parameter (e.g. invalid channel no or product number) 3 = Execution failure
86
Page 89
Activate sensor command - output
Name Value Comment
OCT 1-2
OCI 2
Channel number:
OCD1 0
OCD2 0-1
OCD3 Not used OCD4 Not used
0 = Channel 1 1 = Channel 2 2 = Channel 3
Sensor activation: 0 = Deactivate the EC sensor 1 = Activate the EC sensor
Activate sensor command - input
Name Value Comment
ICT 1-2
0 = OK
ICS 0-3
ICD1 Not used ICD2 Not used ICD3 Not used ICD4 Not used
1 = Unknown command ID 2 = Invalid parameter (e.g. invalid channel no) 3 = Execution failure
Communication Menu

9.3 USB-A port (host)

This option allows the export or import of data from an external mass storage device. The device must first be connected to the instrument through the USB-A port.
Select one of the two import options (product list or access table) to import data from the storage device. This is useful for transferring these files to additional instruments without the need of having to re-enter the data individually on each instrument.
Note: The imported data will override any current settings on the instrument.
Select the export option to export data from the instrument to the storage device. For information regarding the uploaded files, refer to Uploaded Files on page 91.
For both import and export options, the progress bar is updated to give an indication of the progress of the selected option.
87
Page 90
Communication Menu

9.4 HTTP/TCP-IP

9.4.1 Overview

When activated this option downloads data from the instrument directly to a web page that can be accessed from a PC. To be able to use this option, the instrument must be connected to the network (specifically Connector P3 - see details in Main board connections on page 37) and the network must have a DHCP server installed.
Check the Enable server box to enable the web server communication link.
Enter the Server name for the instrument. This is free format text and should typically be
used to identify the instrument.
Check the Enable security box if you require a password to be entered on the PC to access
the web page.
If any of the details on the previous screen have been changed, a warning message will be displayed as illustrated left.
The changes must be confirmed, after which the instrument must be powered down and powered up again for the changes to take effect.
Note: The IP Diagnostics button at the bottom of the screen is for use by experienced IT personnel only to help resolve any communications problems.

9.4.2 PC interface

Once the server has been enabled and the interface information set up, access the information by launching an internet browser and typing “http://” followed by the server name that has been assigned to the instrument, in the address box as illustrated below:
If the enable security option has been checked on the instrument, you will be required to enter a username and password on your PC to gain access to the web page.
The username and password must be a valid username/password combination that has been set up on the instrument (see also User management on
page 94 on how to set up users on the instrument).
Domain information is not required.
88
Page 91
Communication Menu
Once a valid username/password combination has been entered, the initial web page will be displayed giving a list of options:
Note: If TPO or TPA calculations have been enabled for a portable instrument, then this data is accessible from an additional option in the above screen entitled “TPO/TPA Measurements stored / Channel n”, where n is the channel configured for TPO or TPA calculations.
Click on any of these options and the data will be displayed on the PC screen. The following shows an example of the screen when selecting the Current Measurement option:
89
Page 92
Communication Menu

9.5 Data file transfer through the USB-B port (client)

The USB port allows you to copy data files from the instrument to a PC using Microsoft synchronization software. The files in the instrument are coded in a binary format, so they must be converted to a readable format on your PC to make them suitable for viewing. This is done automatically using the software supplied on the ORBISPHERE CD (see PC software
installation below and Upload report files on page 90 for details).
Check that the instrument and PC are powered, then connect them with the USB cable that was supplied with the instrument. Follow the instructions below:

9.5.1 PC software installation

Insert the ORBISPHERE CD inside the PC drive. If the auto executable installation does not start, browse the CD with Windows Explorer and double click on the setup.exe file to start the installation. Follow the step by step instructions appearing on the screen.
Once the installation is complete, two icons are installed on the PC Desktop:
Orbisphere 410, 51x USB upload is used to upload and convert report files from the
instrument to the PC (see Upload report files on page 90).
Orbisphere 410, 51x Setup is software used by the Hach Lange after sales technicians to
upload new software versions. To avoid an accidental software modification, a key is required.

9.5.2 Microsoft synchronization software

Windows XP - A copy of the synchronization software is available on the ORBISPHERE CD. Inside the ActiveSync folder double click on ActiveSync.exe to install on the PC.
Windows Vista and Windows 7 - Download and install the Windows Mobile Device Center application which contains the synchronization software.
Once installed, the synchronization software starts automatically when the instrument is connected to the PC and an icon will appear in the taskbar at the bottom of the screen.
Note: By default the software proposes to set up a partnership with the instrument. This is not required, so make sure this is set to NO before continuing.

9.5.3 Upload report files

Double click on the Orbisphere 410, 51x USB upload icon on the PC desktop (created during the process described previously in PC software installation on page 90) to start the upload and convert process.
When the main screen is displayed, click on the Wizard button in the top left corner.
Check that the USB connection from the instrument to PC has been made, and that the synchronization software has been activated and the link recognized (the icon in the taskbar should be colored green). Click on Next.
Choose the directory where the files are to be stored. If the directory path does not exist it will be automatically created. Click on Next.
90
A warning message appears to warn that all files currently in the directory (if any exist) will be
deleted prior to upload. Click on Yes to continue with the process or No to abort. If Yes is selected, a progress bar of the upload is displayed.
Should any of the files be missing (e.g. a calibration file will be missing if no calibration has been performed), a warning message is displayed, but no action is required. Click on OK. Once the upload is complete, the files are converted and stored in the folder defined above.
Click on Finish to terminate the process, or on either of the other two buttons to open the folder or view the text files.
Page 93
Communication Menu
Uploaded Files
There are a number of files that are uploaded to the PC during this process. However, only the text files (with a .txt file extension) are in a readable format on the PC. Most document editors (Word, Notepad, etc.) can be used to open these files, as well as spreadsheet and other reporting tools (e.g. Excel).
There are four reports available:
Instrument Configuration
Calibration Reports
Measurements
User Actions
The reports show information for all channels (where applicable). Below is an example of the Instrument Configuration report as viewed using the Notepad utility.
91
Page 94
Communication Menu
92
Page 95

Section 10 Security Menu

Note: When the instrument is started for the very first time, security is disabled by default. It is highly recommended that each user be entered into the system and given appropriate access rights as soon as possible to avoid any unauthorized access. Details of this process are described in this section.

10.1 Access rights management

Access rights management
Each user has a unique ID and user password. The ID and password are used by the software to:
Allow or deny a user to perform specific actions.
To trace this action with his "ID" in a log file.
Once the ID and password are entered, the user is allowed to perform actions according to the "Access level" that has been attributed to his ID by the Manager. See Security level table on page 18
Level Typical rights Comments
0 View parameters, change views Press the unlock button and OK to access
1 + Start / Stop measurements
2 + Calibration
Figure 39 Security menu
Table 3 Access levels
3 + Modify parameters
+ Modify table "User ÅÆ Access level"
4
+ Enable/Disable "Access right" features
Note: At startup, all the menus are locked. The user has to identify himself to get access beyond the different views (see Function keys on the header bar on page 42)
There is at least one ID having the level 4
93
Page 96
Security Menu

10.2 Configure security

Configure security
This enables defining the users with their access level when the software starts for the first time. It is possible to configure several parameters related to confidentiality. This requires a user access level 4.
Note: Access rights are disabled by default.
Access rights: When enabled, it is required to log in as a registered user (see User management on
page 94) to access the menus. When disabled (default), all menu are access free, and the effect of
leaving the text box blank in user login window is that there will be no name recorded for the action in the log file.
Enter a maximum session time in minutes for improved confidentiality. The user is logged out
automatically when the set delay for inactivity is over.
User action logging: When enabled, every action from a logged on user is recorded in a user log file
for traceability.
Clear all user actions log file. Confirm to clear the log file. This functionality is aimed at clearing demo
or test logs for example. The log file is a rolling buffer recording the past 100 actions.

10.3 User management

User management
This window shows the list of registered users for the instrument. They are listed by name, ID, password and access level.
Note: The "User password" must be at least 4 characters long.
Pressing on an empty line, or pressing the Add button brings a window to add a new user. Name, ID, password and access level (from 1 to 4) must be entered.
Pressing on a registered user line brings a window for editing or deleting the user data in the list.
Note: The list can contain up to 99 users

10.4 User action log file

User action log file
Each time the user performs an important action, a record is written in the "User action log file". It is a rolling buffer which contains the last 1000 user actions. The "User Interface" will allow viewing this log file (Menu Security / View log file). This log file contains the following data:
line number
the action name
the user name and ID
the current date and time.
Note: Unsuccessful attempts to register are recorded in the log file without a user ID.
94
Page 97

Section 11 Products Menu

Figure 40 Products menu
95
Page 98
Products Menu
Products
This option allows users to save and/or use previously saved product configurations. A maximum of 100 different product configurations can be stored in the instrument. The basic measurement configuration (gas to analyze, gas unit, alarm limits, analog outputs, etc.) can be set up for a product and will be automatically used by the instrument when that product is selected.
Product configurations can be moved from instrument to instrument if required. However, as the sensor only analyzes oxygen, only products configured to analyze oxygen can be selected on this instrument. Products configured to analyze other gases can, however, be set up on this instrument and easily transferred to other 410 or 51x instruments analyzing gases other than oxygen.
For ease of use, where product configurations are identical or similar, a Copy facility exists on the modify product screen. This enables copying a stored configuration and storing it in one or more additional locations. Then use the modify product option to identify and/or modify the duplicate configurations.
Select product Note: If the PROFIBUS-DP communications protocol has been enabled,
products can be selected for analysis using that facility (see Input/Output
data on page 84 and specifically Change product command - output on page 86 for details).
Select the product (0-99) to be analyzed, or use the Next and Previous
buttons to scroll sequentially through the existing product list. Alternatively, use the Search facility to search for a product. Enter a full
or partial search criteria. If only one match is found, this product is automatically selected. If a number of products match the search criteria, then a list of matches will be displayed. Select a product directly from the list of matching products.
Press OK to select the product or Cancel to exit.
Modify product Select the product (index 0-99) to modify, or use the Next and Previous
buttons to scroll sequentially through the existing product list. Alternatively, use the Search facility to search for a product. Enter a full
or partial search criteria. If only one match is found, this product is automatically selected. If a number of products match the search criteria, then a list of matches will be displayed. Select a product directly from the list of matching products.
Select the gas to analyze (up to three can be selected) from the drop
down list.
After selecting a product and gas, press Configure to configure the
product. Configure the product as required (refer to Measurement
configuration on page 53 for additional information).
Press Analog out to configure the analog outputs. Configure the analog
output as required (refer to Channel configuration on page 71 for additional information).
Press OK to accept the configuration, or Cancel to exit.
96
Page 99

Section 12 Global Configuration Menu

Figure 41 Global configuration menu
Global configuration
The global configuration option allows users to save, and use previously saved, instrument configurations. A maximum of 10 configurations can be saved, with configuration 0 (zero) the instrument default.
Once all the instrument parameters have been set up, use this option to save the configuration. Selecting pre-defined configurations avoids the need to re-enter all the parameters when using the instrument for a different application.
Save Define the index (1-9) to save the current configuration. Use the Next
and Previous buttons to scroll sequentially through existing configurations, to overwrite an existing configuration or save as a new one.
Enter a name to define the current configuration.
Enter any comments to associate with this configuration.
Select • Select the configuration (index 0-9) to use on the instrument.
Confirmation will be required for the selected configuration. The instrument must then be restarted (powered off and then back on) in order for the new configuration to take effect.
97
Page 100
Global Configuration Menu
98
Loading...