Hach-Lange LZU5390 User Manual [en, de, es, fr, it]

53 90
EC
53 90
1
2
3
4
5
6
7
2
Key points. Hauptteile. Puntos clave. Points clé. Punti chiave. Pontos chaves.
Cablepress.
1
Kabelverschraubung.
Prensacables.
Presse-câble.
Pressacavi.
Prensa cabos.
Connection protector.
2
Steckerschutz.
Protector del conector.
Protecteur du connecteur.
Protettore connessione.
Protetor do conector.
MP-5 connector.
3
MP-5-Stecker.
Conector MP-5.
Connecteur MP-5.
Connettore MP-5.
Conector MP-5.
1/2” NPT stainless steel.
4
Rosca 1/2” NPT en acero inox
Filetage 1/2” NPT en acier inox.
Filettatura 1/2” NPT in acciaio inos.
Rosca 1/2” NPT em aço inox.
Titanium body, electrode 1.
5
Titangehäuse, Elektrode 1.
Cuerpo de titanio, electrodo 1.
Corps en titane, électrode 1.
Corpo in titanio, elettrodo 1.
Corpo em titânio, eletrodo 1.
Pt 1000 temperature sensor.
6
Temperatursensor Pt 1000.
Sensor de temperatura, Pt 1000.
Capteur de température, Pt 1000.
Sensore di temperatura, Pt 1000.
Sensor de temperatura, Pt1000.
Electrode 2, titanium.
7
Elektrode 2, Titan.
Electrodo 2, titanio
Electrode 2, titane.
Elettrodo 2, titanio.
Eletrodo 2, titânio.
(EU Directive 2002/96/EC)
3
Conductivity cell
Introduction
The 53 90 is a cell for pipes and reactors. They include an integrated Pt 1000 tem­perature sensor.
SAFETY NOTE Before use with hazardous substanc­es, find out about the relevant haz­ards, protective measures, and what to do in an emergency.
Cell preparation
Before use for the first time immerse the electrodes in ethanol for 15 s, and rinse with distilled water. Apply the same procedure for a cell not being in use for long time.
Installation
The cell should always be mounted in im­mersion or fl otation housings, with PG 13,5 thread.
See Housing instruction manual.
Cable (not included). Connected to mul-
tiple cables with MP-5 connector. Mounting. Directly onto pipes or reactors.
Calibration
Before measurement, calibrate with stan­dard solutions.
See User’s Manual of conductivity meter.
Measurements. Recommendations
• Both, calibration and measurement
must be performed under slight stirring of standards and samples. The stirring speed in the two cases should be iden­tical.
• The sample must cover the measuring
electrodes 7.
• Ensure that during measurement there
are not air bubbles over the measuring electrodes.
• After measurement, rinse the cell with
distilled water or with sample to be measured.
With time and use the titanium alters
its colour. This change does not affect the good functioning of the cell.
Maintenance
Keep the electrodes surface clean 7. If the thorough rinse with distilled water is not sufficient, suitable solvent for the elimination of sample rests can be used. This solvent must not produce chemical attack over the cell materials. Inorganic contamination: Immerse the cell in diluted hydrochloric acid or in chro­mic acid. Afterwards rinse thoroughly with distilled water. Organic contamination: Use suitable solvent (see cell’s materials), next utilize alcohol and distilled water as in Cell prepa­ration is described.
Storage
Store the cell dry.
4
53 90
GB
Life of the cell
The cell lifespan is unlimited if suitable maintenance is applied and the cell does not break.
Troubleshooting
... Reading 0 independently of the mea-
sured sample.
Broken cable.
Replace the cell.
... Unstable reading. ... Impossible calibration. Dirty measuring electrodes or air
bubbles on their surface.
Clean or shake slightly the cell.
... The obtained readings differ a lot from
the real values.
Verify that the conductivity is covered
by the measuring interval of the cell.
Re-calibrate using standards in good
conditions.
... slow cell response
Regenerate the cell.
Specifications
Approx. constant 0.15 cm Measuring range 0.1 µS/cm to 10 mS/cm Operating temp. 0 to 100 °C Max. pressure 6 bar (25 °C) Temp. sensor Pt 1000 Body material titanium, PVDF, s.steel Electrode material titanium Thread 1/2” NPT Connector MP-5
Min. immers. depth
-1
10 mm
Applications
Clean waters, low conductivity waters, etc.
CERTIFICATE OF QUALITY
The attached cell 53 90 has passed the quality test and fulfils the following speci­fications:
• EC error 5 %. between 0,1 µS/cm and 10 mS/cm.
• Temperature error 0.5 ºC.
5