Class I Division 2, Groups A–D, T4 / Class I, Zone 2 Group 2C, T4
Note: This product does not fulfill the requirements of the 94/9/EC Directive
(ATEX Directive).
hazardous location.
Note: This product does not fulfill the requirements of the 94/9/EC Directive
(ATEX Directive).
–
, SO4 2–, S2–, PO4 3–, Cl–, Anion Active
3
English 3
SpecificationDetails
Calibration/verificationAir calibration: One point, 100% water-saturated air
Sample calibration: Comparison with standard instrument
Probe immersion depth and pressure limitsPressure Limits at 34 m (112 ft.), 345 kPa (50 psi) maximum;
Sensor cable10 m (30 ft) integral cable with quick disconnect plug (all sensor
Probe weight1.0 kg (2 lb, 3 oz)
Probe dimensionsDiameter x length: 49.3 x 255.7 mm (1.9 x 10.1 in.)
Power requirements12 VDC, 0.25 A, 3 W
WarrantyProbe: 3 years against manufacturing defects
accuracy may not be maintained at this depth
types)
Up to 100 m possible with extension cables (non-Class I, Division
2 sensor types only)
Up to 1000 m with junction box (non-Class I, Division 2 sensor
types only)
Sensor cap: 2 years against manufacturing defects
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious
injury.
Indicates a potentially hazardous situation that may result in minor or moderate injury.
4 English
W A R N I N G
C A U T I O N
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special
emphasis.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument
could occur if not observed. A symbol on the instrument is referenced in the manual with a
precautionary statement.
This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential
injury. If on the instrument, refer to the instruction manual for operation or safety information.
This symbol indicates the presence of a light source that may have the potential to cause minor eye
injury. Obey all messages that follow this symbol to avoid potential eye injury.
This symbol indicates the presence of devices sensitive to Electro-static Discharge (ESD) and
indicated that care must be taken to prevent damage with the equipment.
Electrical equipment marked with this symbol may not be disposed of in European public disposal
systems after 12 August of 2005. In conformity with European local and national regulations (EU
Directive 2002/98/EC), European electrical equipment users must now return old or end-of-life
equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return endof-life equipment, producer-supplied electrical accessories, and all auxillary items for proper disposal.
Product overview
D A N G E R
Chemical or biological hazards. If this instrument is used to monitor a treatment process and/or
chemical feed system for which there are regulatory limits and monitoring requirements related to
public health, public safety, food or beverage manufacture or processing, it is the responsibility of the
user of this instrument to know and abide by any applicable regulation and to have sufficient and
appropriate mechanisms in place for compliance with applicable regulations in the event of malfunction
of the instrument.
This sensor is designed to work with a controller for data collection and operation. The sensor can be
used with several controllers. Refer to the controller-specific user manual for more information.
The primary applications for this sensor are municipal and industrial wastewater applications. LDO
sensor technology does not consume oxygen, and can measure DO concentration in low or no-flow
applications. Refer to Figure 1.
Make sure that all components shown in Figure 2 have been received. If any items are missing or
damaged, contact the manufacturer or a sales representative immediately. Refer to Figure 2.
Figure 2 Sensor component list
1 LDO sensor
1
Included user manual is not shown.
1
2 Calibration bags (2x)
Installation
Validate the sensor type
D A N G E R
Explosion hazard. Connect only peripheral components that are clearly marked as certified for Class 1,
Division 2 Hazardous Locations.
N O T I C E
The hazardous location certified version of this product does not fulfill the requirements of the 94/9/EC Directive
(ATEX Directive).
1. Go to the connector end of the cable.
2. Read the label on the connector end of the cable. For hazardous location certified sensors, the
label will show "Rated: Class 1 Division 2".
6 English
3. Examine the connector.
•
Hazardous location certified sensors have a safety lock connector. Refer to Figure 3
on page 7.
• Sensors that are not certified for hazardous locations have a quick-connect connector, without
a safety lock.
Connect the sensor in a hazardous location
D A N G E R
Explosion hazard. This equipment is suitable for use in non-hazardous locations or Class 1, Division 2,
Groups A, B, C, D Hazardous Locations with specified sensors and options when installed per the
Hazardous Location Installation Control Drawing. Always refer to the Control Drawing and applicable
electrical code regulations for proper installation instructions.
Explosion hazard. Do not connect or disconnect electrical components or circuits to the equipment
unless power has been removed or the area is known to be non-hazardous.
Use only a hazardous location certified sensor and cable lock in hazardous locations. The hazardous location
certified version of this product does not fulfill the requirements of the 94/9/EC Directive (ATEX Directive).
For more information, refer to Validate the sensor type on page
1. Remove the connector cap from the controller. Keep the connector cap to seal the connector
opening when the sensor is removed.
2. Connect the sensor to the controller. Refer to the controller manual for more information.
3. Close the safety lock over the connector.
4. To remove the connector safety lock, use a small flat screwdriver. Refer to Figure 3.
D A N G E R
N O T I C E
6 .
Figure 3 Connector safety lock
Connect the sensor in a non-hazardous location
Refer to Figure 4 to connect an LDO sensor to an sc controller. Refer to the specific sc controller
manual for hard-wiring instructions.
Figure 4 Connect the LDO sensor (non-hazardous location sensor shown)
English7
After the sensor is attached, scan for the sensor. Refer to Install the sensor
on page 8.
Install the sensor
There are two options to install the sensor:
• Connect the sensor while power to the controller is off. The controller will look for and install new
sensors when it is turned on.
• Connect the sensor while power to the controller is on. Use the Scan Devices command to install
the new sensor:
OptionDescription
sc200 controllerGo to MENU>TEST/MAINT>SCAN DEVICE
sc100 controllerGo to MENU>TEST/MAINT>SCAN SENSORS
sc1000 controller Go to MENU>SYSTEM SETUP>DEVICE MANAGEMENT>SCANNING FOR NEW
DEVICES
Refer to Connect the sensor in a non-hazardous location on page 7 for digital sensor connection.
Sensor installation options
The installation and accessory options available for the LDO sensor are supplied with installation
instructions in the hardware kit. Figure 5 shows several installation options. To order installation
hardware, refer to Replacement parts and accessories on page 17
Figure 5 Installation options
.
1 Rail mount4 Chain mount
2 Float mount5 Union mount
3 Air blast system mount
8 English
Operation
User navigation
Refer to the controller documentation for keypad description and navigation information.
Configure the sensor
Use the Configure menu to enter identification information for the sensor and to change options for
data handling and storage.
For information about sensor installation, refer to Install the sensor on page
Make sure that all of the Configuration menu values are correct for the application.
1. Go to MENU>SENSOR SETUP>[Select Sensor]>CONFIGURE.
2. Select an option, ENTER. The list of available options is shown in the table below.
OptionDescription
EDIT NAMEChanges the name that corresponds to the sensor on the top of the measure screen. The
name is limited to 10 characters in any combination of letters, numbers, spaces or
punctuation.
SET UNITSTEMP–Sets the temperature units to °C (default) or °F.
MEASURE–Set the measurement units in mg/L, ppm or % .
ALT/PRESS–Set the altitude in m or ft, or set the atmospheric pressure units in mmHg or
torr. (Default value = 0 ft)
ALT/PRESSEnter the value of the altitude or atmospheric pressure. This value must be accurate to
complete % saturation measurements and calibration in air. (Default = 0 ft).
SALINITYEnter the salinity value. Salinity range: 0.00 to 250.00 parts per thousand (‰). Refer to
Enter a salinity correction value on page 10 for more information. (Default value = 0)
SIGNAL AVERAGESet the time interval to average signal in seconds
CLEAN INTRVLSet the time interval for manual sensor cleaning in days (Default value = 0 days. A value
of 0 days disables the clean interval.)
RESET CLN INTRVL Set the time interval to the last saved clean interval
LOG SETUPSets the time interval for data storage in the data log—0.5, 1, 2, 5, 10, 15 (default), 30,
60 minutes.
SET DEFAULTSRestores the configurable default values for the sensor. Does not change the setting for
slope or offset.
8.
Enter the atmospheric pressure value
The factory setting for atmospheric (air) pressure is 0 ft, or sea level. To change the default value,
use the steps in this procedure. The adjustment for air pressure is entered as either elevation or as
pressure units (preferred).
Note: Accurate air pressure is critical for saturated air calibration (Calibration with air on page 11
absolute pressure, not adjusted. If the absolute air pressure is not known, use the correct elevation for the location.
1. Go to MENU>SENSOR SETUP>[Select Sensor]>CONFIGURE>SET UNITS>AIR PRESS/ALT
UNITS.
2. Select one of the unit options listed:
OptionDescription
ftFeet—unit of measure for elevation
mMeters—metric unit of measure for elevation
). Use only
English 9
OptionDescription
mmHgMillimeters of mercury—metric unit of measure for absolute air pressure
torrUnit of measure for absolute air pressure
3. Confirm the selection. The value entry screen will show the selected units.
4. Enter the value, then confirm.
Enter a salinity correction value
Dissolved oxygen measurements in saline samples can show an apparent DO value that is very
different from the actual DO value. To correct for the influence of dissolved salts in a sample, enter a
salinity correction factor.
Note: If the presence or amount of salinity in the process is unknown, consult with the treatment facility engineering
staff.
1. Use a conductivity meter to measure the conductivity of the sample in mS/cm at a reference
temperature of 20 °C (68 °F).
2. Use Table 1 to estimate the salinity correction factor in parts per thousand (‰) saturation.
Note: The chloride ion concentration, in g/kg is equal to the chlorinity of the sample. Salinity is calculated with
the formula: Salinity = 1.80655 × chlorinity.
Salinity can be calculated with the relationship in section 2520 B of Standard Methods for the
Examination of Water and Wastewater.*,
3. Go to MENU>SENSOR SETUP>[Select Sensor]>CONFIGURE>SALINITY.
4. Enter the salinity correction factor and confirm.
Table 1 Salinity saturation (‰) per conductivity value (mS/cm)
mS/cm‰mS/cm‰mS/cm‰mS/cm‰
5 3 16 10 27 18 38 27
6 4 17 11 28 19 39 28
7 4 18 12 29 20 40 29
8 5 19 13 30 21 42 30
9 6 20 13 31 22 44 32
10 6 21 14 32 22 46 33
11 7 22 15 33 23 48 35
12 8 23 15 34 24 50 37
13 8 24 17 35 25 52 38
14 9 25 17 36 25 54 40
15 10 26 18 37 26
*
Standard Methods for the Examiniation of Water and Wastewater, 20th Edition. Editors Lenore
S. Clesceri, Arnold E. Greenberg and Andrew D. Eaton, p. 2-48-2-29 (1998). The relationship
between Chlorinity and Oxygen Solubitity is provided in the same reference in 4500-O:I p. 4-131.
10 English
Configure linear output on the controller
Linear outputs send probe data back to the facility PLC, SCADA or other data collection system.
1. Go to the controller output setup menu.
Option Description
sc200Go to MENU>SETTINGS>sc200 SETUP>OUTPUT SETUP>[Select Output]>SET FUNCTION.
sc100Go to MENU>SYSTEM SETUP>OUTPUT SETUP>[Select Output]>SET FUNCTION.
sc1000 Go to MENU>SYSTEM SETUP>OUTPUT SETUP>[Select Output]>SET FUNCTION.
2. Set the function for the controller.
OptionDescription
sc200LINEAR
sc100LINEAR CONTROL (Default value)
sc1000LINEAR CONTROL (Default value)
Modbus registers
A list of Modbus registers is available for network communication. Refer to www.hach.com or
www.hach-lange.com for more information.
Calibration for measurements
The sensor is calibrated to specification at the factory. The manufacturer does not recommend
calibration unless periodically required by regulatory agencies. If calibration is required, let the sensor
come to equilibrium with the process before calibration. Do not calibrate the sensor at setup.
Table 2 shows options for calibration.
Table 2 Calibration options
OptionDescription
AIR CALRecommended calibration method. This calibration modifies the slope.
SAMPLE CALCalibration by comparison with a hand-held DO meter . This calibration modifies the offset.
RESET DFLT CAL Resets the calibration gain (slope) and offset to the factory default: default gain=1.0; default
offset=0.0
Calibration with air
User notes:
•
Make sure that calibration bag has water inside.
• Make sure that the seal between the calibration bag and the sensor body is tight.
• Make sure that the sensor is dry when it is calibrated.
• Make sure the air pressure/elevation setting is accurate for the calibration location.
• Allow enough time for the sensor temperature to stabilize to the temperature of the calibration bag
location. A large difference in temperature between the process and the calibration location can
take up to 15 minutes to stabilize.
1. Remove the sensor from the process. Use a wet cloth to clean the sensor.
2. Put the entire sensor in a calibration bag with 25-50 mL of water. Make sure that the sensor cap
is not in contact with the water inside the calibration bag and that no water drops are on the
sensor cap (Figure 6).
3. Use a rubber band, tie or hand to create a tight seal around the sensor body.
English 11
4. Let the instrument stabilize for 15 minutes before calibration. Keep the calibration bag out of
direct sunlight during stablization.
5. Make sure that the current absolute air pressure or elevation is configured correctly. Refer to
Enter the atmospheric pressure value
Note: The manufacturer recommends the use of absolute or actual air pressure as a best practice.
on page 9.
6. Go to MENU>SENSOR SETUP>[Select Sensor]>CALIBRATE>AIR CAL.
7. Select the option for the output signal during calibration:
OptionDescription
ActiveThe instrument sends the current measured output value during the calibration procedure.
HoldThe sensor output value is held at the current measured value during the calibration procedure.
Transfer A preset value is sent during calibration. Refer to the controller user manual to change the preset
value.
8. The controller will show "Move the probe to bag". Allow the value to stabilize. Push ENTER to
accept the stable value. Alternately, let the calibration continue until the display shows
"Complete".
9. When the sensor is calibrated, put the sensor into the process. Push ENTER.
Figure 6 Air calibration procedure
If the value does not stabilize, the display will show "Unable to Calibrate" followed by an error
message. Table 3 shows the error message and resolution for calibration problems.
Table 3 Air calibration error messages
MessageDescriptionResolution
Cal fail, gain high The calculated gain value is too high.Repeat the calibration.
Cal fail, gain lowThe calculated gain value is too low.Repeat the calibration.
Cal fail, unstableThe value did not stabilize in the maximum allowed calibration time.Repeat the calibration.
Sample CAL - calibration by comparison
This calibration method uses an alternate sensor attached to a hand-held meter.
1. Put the alternate sensor into the process. Put the second sensor as close as possible to the first
sensor.
2. Wait for the DO value to stabilize .
3. On the controller for the first sensor, go to MENU>SENSOR SETUP>[Select
Sensor]>CALIBRATE>SAMPLE CAL.
12 English
4. Select the option for the output signal during calibration:
OptionDescription
ActiveThe instrument sends the current measured output value during the calibration procedure.
HoldThe sensor output value is held at the current measured value during the calibration procedure.
Transfer A preset value is sent during calibration. Refer to the controller user manual to change the preset
value.
5. The controller will show:
•
"Press ENTER when stabilized"
• The current dissolved oxygen measurement
• The current temperature measurement
6. When the measurement is stable, push ENTER. The display will show an entry screen.
Note: The measurement will usually stabilize in 2 to 3 minutes.
If the value does not stabilize, the display will show "Unable to Calibrate" followed by an error
message. Table 4 shows the error message and resolution for calibration problems.
Table 4 Sample cal error messages
MessageDescriptionResolution
Cal fail, offset high The calculated offset value is too high.Repeat the calibration.
Cal fail, offset lowThe calculated offset value is too low.Repeat the calibration.
Cal fail, unstableThe value did not stabilize in the maximum allowed calibration time. Repeat the calibration.
Exit the calibration procedure
1. During calibration, push the BACK key. Three options are shown:
OptionDescription
ABORTStop the calibration. A new calibration must start from the beginning.
BACK TO CAL Return to the current calibration.
LEAVEExit the calibration temporarily. Access to other menus is allowed while the calibration
2. Select one of the options. Confirm.
continues in the background. A calibration for a second sensor (if present) can be started.
To return to the calibration, push the MENU key and select Sensor Setup, [Select Sensor].
Reset calibration defaults
Calibration settings can be reset to the factory defaults. Gain and offset values are set to 1.0 and 0.0,
respectively.
1. Go to MENU>SENSOR SETUP>[Select Sensor]>CALIBRATE>RESET CAL DEFLT.
2. The display will show a confirmation message. Confirm to reset the sensor to the factory default
calibration curve.
Maintenance
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
English 13
D A N G E R
Explosion hazard. Do not connect or disconnect electrical components or circuits to the equipment
unless power has been switched off or the area is known to be non-hazardous.
D A N G E R
Explosion hazard. Substitution of components may impair suitability for Class 1, Division 2. Do not
replace any component unless power has been switched off and the area is known to be nonhazardous.
The hazardous location certified version of this product does not fulfill the requirements of the 94/9/EC Directive
(ATEX Directive).
N O T I C E
Maintenance schedule
The maintenance schedule shows minimum intervals for regular maintenance tasks. Perform
maintenance tasks more frequently for applications that cause electrode fouling.
Note: Do not disassemble the probe for maintenance or cleaning.
Maintenance taskRecommended minimum frequency
Clean the sensor90 days
Inspect the sensor for damage90 days
Calibrate the sensorAs recommended by regulatory agencies
Clean the sensor
Clean the exterior of the sensor with a soft, wet cloth.
Note: If the sensor cap must be removed for cleaning, do not expose the interior of the cap to direct sunlight for
extended periods of time.
Set or change the clean interval
Application conditions may need shorter or longer durations between manual sensor cleanings. The
default clean interval is 0 days. To change the interval, refer to the steps in this procedure.
1. Go to MENU>SENSOR SETUP>[Select Sensor]>CONFIGURE>CLEAN INTRVL.
2. Change the shown value as needed. Confirm the change.
•
To turn off the clean interval, set the value to '0'.
Change the sensor cap
W A R N I N G
Potential explosion hazard. The sensor setup cap is not rated for hazardous location use.
Replacement sensor caps and setup caps are shipped with installation instructions. Refer to the
included instructions to change the cap.
For best performance and accuracy, replace the sensor cap:
Every two years
•
• When routine inspection shows significant erosion of the sensor cap
14 English
Troubleshooting
Diagnostic and test menu
The diagnostic and test menu shows current and historical information about the LDO sensor.
To access the diagnostic and test menu, go to MENU>SENSOR SETUP>[Select
Sensor]>DIAG/TEST.
Refer to Table 5.
Table 5 DIAG/TEST menu
OptionDescription
SENSOR INFOSOFTWARE VERS—Shows the installed software version
BOOT VERSION—Shows the installed boot version
DRIVER VERS—Shows the installed software driver version
LOT CODEShows the sensor cap manufacturing lot
SERIAL NUMBERSensor serial number
GAIN CORRAdjust the calibration gain value.
Range: 0.50 to 2.00
OFFSET CORRAdjust the calibration offset value (mg/L or ppm).
Range: –3.00 to +3.00
PHASE DIAGShows the phase for total, red and blue wavelengths. Updates once per second.
AMPL DIAGShows the amplitude for red and blue wavelengths. Updates once per second.
DAYS TO CLEANShows the number of days until the next scheduled manual cleaning.
SENSOR LIFEShows the number of days until the next scheduled sensor cap replacement
Error list
If an error occurs, the reading on the measurement screen flashes. Output behavior is determined by
controller settings. Refer to the controller manual for details.
To show the current sensor errors, go to MENU>DIAGNOSTICS>[Select Sensor]>ERROR LIST.
Refer to Table 6.
Table 6 Error list for the LDO sensor
ErrorPossible causeResolution
RED AMPL LOW (Value is below 0.01)
OR
BLUE AMPL LOW (Value is below 0.01)
The sensor cap is not installed,
or is not installed correctly.
The light path is blocked in the
sensor cap.
The sensor is not operating
correctly.
Remove the sensor cap and install it
again.
Inspect the inside of the sensor cap
and lens.
Make sure that the LED is flashing.
Contact the manufacturer.
Warning list
When the warning icon flashes (sc100 and sc200) or when the screen turns yellow (sc1000), a
message is shown on the bottom of the measurement screen. On the sc1000, the screen turns
English15
yellow to show a warning. To show the current sensor warnings, go to MENU>
DIAGNOSTICS>[Select Sensor]>WARNING LIST. Refer to Table 7.
Table 7 Sensor warning list
WarningDefinitionResolution
EE SETUP ERRStorage is corrupt. The values have been set
EE RSRVD ERR
TEMP < 0 CThe process temperature is below 0 °C
TEMP > 50 CThe process temperature is above 50 °C
RED AMPL LOWValue falls below 0.03 Refer to Table 6 on page 15 .
RED AMPL HIGHValue is greater than 0.35 Call technical support.
BLUE AMPL LOWValue is below 0.03 Refer to Table 6 on page 15.
BLUE AMPL HIGH Value is greater than 0.35 Call technical support.
CAP CODE FAULT The sensor cap code has become corrupt.
to the factory default.
(32 °F)
(120 °F)
The code has been reset automatically to the
default cap and lot codes.
Contact technical support.
Increase the process temperature or stop
use until the process temperature is in the
sensor specification range.
Decrease the process temperature or stop
use until the process temperature is in the
sensor specification range.
Complete the sensor setup cap procedure.
If no setup cap is available for the sensor
cap, call technical support.
Event list
The Event list keeps a log of changes to how data is recorded by the sensor. To show sensor events,
go to MENU>DIAGNOSTICS>[Select Sensor]>EVENT LIST.
Refer to Table 8.
Table 8 Event list for the sensor
EventDescription
ALT/PRESSURE UNIT CHANGEAtmospheric pressure or altitude units have changed.
ALT/PRESSURE CHANGEThe value for altitude or atmospheric pressure has changed.
TEMP UNIT CHANGEThe units for temperature have changed.
MEAS UNIT CHANGEA new unit of measurement has changed.
SALINITY CHANGEThe value for salinity has changed.
SET DEFAULTSensor settings have been reset to the default values.
SENSOR SETUP CHANGEThe sensor setup has changed.
CLEAN INTERVAL TIMER CHANGEThe time between sensor cleaning has changed.
SENSOR CAP LIFE TIMER CHANGEThe time between sensor cap replacements has changed.
16 English
Replacement parts and accessories
Replacement items
DescriptionItem no. (US / EU)
LDO Probe, with one sensor cap and 2 calibration bags9020000 / LXV416.99.20001
LDO Probe for hazardous locations, with one sensor cap and 2 calibration bags9020300 / —
Sensor cap, replacement (includes the sensor setup cap, which is not rated for
use in Class 1, Division 2 hazardous locations)
Accessories
DescriptionItem no. (US / EU)
Sensor cable lock for hazardous locations6139900 / —
High output air blast cleaning system, 115 V (not rated for use in hazardous
locations)
High output air blast cleaning system, 230 V (not ATEX rated for use in
hazardous locations)
Calibration bag (1x)5796600 / 5796600
Cable, sensor extension, non-hazardous location, 7.7 m (25 ft)
Hardware kit for pipe mount (PVC)9253000 / LZY714.99.21810
Hardware kit for float mount (PVC)9253100 / LZX914.99.42200
Hardware kit for air blast mount9253500 / LZY812
Hardware kit for chain mount (stainless steel)— / LZX914.99.11200
Hardware kit for union mount9257000 / 9257000
HQd meter with LDO rugged probe (not rated for use in hazardous locations)8505200 / HQ40D.99.310.000
1
15 m (49 ft) and 30 m (98 ft) also available
1
9021100 / 9021150
6860000 / 6860003.99.0001
6860100 / 6860103.99.0001
US: 5796000, 7.7 m (25 ft)
EU: LZX849, 10 m (33 ft)
English 17
Inhaltsverzeichnis
Technische Daten auf Seite 18Betrieb auf Seite 24
Allgemeine Informationen auf Seite 19Wartung auf Seite 30
Produktübersicht auf Seite 20Fehlersuche und Behebung auf Seite 31
Installation auf Seite 21Ersatzteile und Zubehör auf Seite 33
Kalibrierung für Messungen auf Seite 27
Technische Daten
Änderungen vorbehalten.
Technische DatenDetails
Benetzte Materialien (Sensor)CPVC, Sensorende und Kabelende
Polyurethan, Umspritzung am Kabelende und Kabelmantel
Edelstahl 316, Gehäuse und Schrauben
Viton, O-Ring
Noryl, Mutter am Kabelende
IP-KlassifikationIP68
Benetzte Materialien (Sensorkappe)Acryl
Messbereich (gelöster Sauerstoff)0 bis 20 ppm (0 bis 20 mg/l)
Klasse 1, Unterteilung 2, Gruppen A-D, T4/Klasse 1, Zone 2 Gruppe
2C, T4
Hinweis: Dieses Produkt erfüllt nicht die Anforderungen der Richtlinie 94/9/EG
(ATEX-Richtlinie).
Verwendung an gefährlichen Standorten.
Hinweis: Dieses Produkt erfüllt nicht die Anforderungen der Richtlinie 94/9/EG
(ATEX-Richtlinie).
–
, SO4 2–, S2–, PO4 3–, Cl–, anionische Aktiv-
3
18 Deutsch
Technische DatenDetails
Kalibrierung/PrüfungLuftkalibrierung: Ein Punkt, 100 % wassergesättigte Luft
In-Line-Kalibrierung: Vergleich mit genormtem Messgerät
Eintauchtiefe des Sensors und
Druckgrenzen
Sensorkabel10 m fest eingebautes Kabel mit Schnelltrennstecker (alle
Sensorgewicht1,0 kg
Sensor-DimensionenDurchmesser x Länge: 49,53 x 255,27 mm
Stromversorgung12 V Gleichstrom, 0,25 A, 3 W
GarantieSensor: 3 Jahre bei Standardapplikationen
Druckgrenzen bei 34 m, max. 345 kPa (50 psi); Messgenauigkeit bei
dieser Tiefe nicht gewährleistet
Sensortypen)
Mit Verlängerungskabeln sind bis zu 100 m möglich (nur Sensortypen,
die nicht der Klasse 1, Unterteilung 2 entsprechen)
Mit Verteilerkasten sind bis zu 1000 m möglich (nur Sensortypen, die
nicht der Klasse 1, Unterteilung 2 entsprechen)
Sensorkappe: 2 Jahre bei Standardapplikationen
Allgemeine Informationen
Der Hersteller ist nicht verantwortlich für direkte, indirekte, versehentliche oder Folgeschäden, die
aus Fehlern oder Unterlassungen in diesem Handbuch entstanden. Der Hersteller behält sich
jederzeit und ohne vorherige Ankündigung oder Verpflichtung das Recht auf Verbesserungen an
diesem Handbuch und den hierin beschriebenen Produkten vor. Überarbeitete Ausgaben der
Bedienungsanleitung sind auf der Hersteller-Webseite erhältlich.
Sicherheitshinweise
H I N W E I S
Der Hersteller ist nicht für Schäden verantwortlich, die durch Fehlanwendung oder Missbrauch dieses Produkts
entstehen, einschließlich, aber ohne Beschränkung auf direkte, zufällige oder Folgeschäden, und lehnt jegliche
Haftung im gesetzlich zulässigen Umfang ab. Der Benutzer ist selbst dafür verantwortlich, schwerwiegende
Anwendungsrisiken zu erkennen und erforderliche Maßnahmen durchzuführen, um die Prozesse im Fall von
möglichen Gerätefehlern zu schützen.
Bitte lesen Sie dieses Handbuch komplett durch, bevor Sie dieses Gerät auspacken, aufstellen oder
bedienen. Beachten Sie alle Gefahren- und Warnhinweise. Nichtbeachtung kann zu schweren
Verletzungen des Bedieners oder Schäden am Gerät führen.
Stellen Sie sicher, dass die durch dieses Messgerät bereitgestellte Sicherheit nicht beeinträchtigt
wird. Verwenden bzw. installieren Sie das Messsystem nur wie in diesem Handbuch beschrieben.
Bedeutung von Gefahrenhinweisen
G E F A H R
Kennzeichnet eine mögliche oder drohende Gefahrensituation, die, wenn sie nicht vermieden wird, zum Tod oder
zu schweren Verletzungen führt.
W A R N H I N W E I S
Kennzeichnet eine mögliche oder drohende Gefahrensituation, die, wenn sie nicht vermieden wird, zum Tod oder
zu schweren Verletzungen führen kann.
Kennzeichnet eine mögliche Gefahrensituation, die zu geringeren oder moderaten Verletzungen führen kann.
V O R S I C H T
Deutsch 19
H I N W E I S
Kennzeichnet eine Situation, die, wenn sie nicht vermieden wird, das Gerät beschädigen kann. Informationen, die
besonders beachtet werden müssen.
Warnhinweise
Lesen Sie alle am Gerät angebrachten Aufkleber und Hinweise. Nichtbeachtung kann Verletzungen
oder Beschädigungen des Geräts zur Folge haben. Im Handbuch werden auf die am Gerät
angebrachten Symbole in Form von Warnhinweisen verwiesen.
Dies ist das Sicherheits-Warnsymbol. Befolgen Sie alle Sicherheitshinweise im Zusammenhang mit
diesem Symbol, um Verletzungen zu vermeiden. Wenn es am Gerät angebracht ist, beachten Sie die
Betriebs- oder Sicherheitsinformationen im Handbuch.
Dieses Symbol kennzeichnet das Vorhandensein einer Lichtquelle, die kleinere Augenverletzungen
verursachen kann. Befolgen Sie zur Vermeidung potentieller Augenverletzungen alle Mitteilungen, die
diesem Symbol nachgestellt sind.
Dieses Symbol kennzeichnet das Vorhandensein von Geräten, die empfindlich auf elektrostatische
Entladung (ESD) reagieren und zeigt an, dass Vorsicht erforderlich ist, um Schäden an diesem Gerät
zu vermeiden.
Elektrogeräte, die mit diesem Symbol gekennzeichnet sind, dürfen ab 12. August 2005 nicht in
öffentlichen europäischen Abfallsystemen entsorgt werden. Benutzer von Elektrogeräten müssen in
Europa in Einklang mit lokalen und nationalen europäischen Regelungen (EU-Richtlinie 2002/98/EG)
Altgeräte kostenfrei dem Hersteller zur Entsorgung zurückgeben.
Hinweis: Für die Rückgabe von Altgeräten, Zubehör und Zusatzausstattungen für eine Entsorgung/Recycling wenden
Sie sich bitte an den Gerätehersteller oder Lieferanten, der Ihnen genaue Anweisungen dazu geben wird.
Produktübersicht
G E F A H R
Chemische und biologische Risiken. Wird das Gerät dazu verwendet, ein Verfahren und/oder ein
chemisches Zufuhrsystem zu überwachen, für das vorgeschriebene Grenzwerte und
Überwachungsvorschriften im Bereich der öffentlichen Sicherheit, der Gesundheit oder im Bereich der
Lebensmittel- oder Getränkeherstellung bestimmt wurden, so unterliegt es der Verantwortung des
Benutzers des Geräts, alle solche Bestimmungen zu kennen und diese einzuhalten und für
ausreichende und entsprechende Vorsorgemaßnahmen zur Einhaltung der für den Fall einer
Fehlfunktion des Geräts bestehenden Bestimmung zu sorgen.
Dieser Sensor wird an einen Controller angeschlossen und dient zur Aufnahme von Messdaten. Der
Sensor kann zusammen an verschiedenen Controllern verwendet werden. Weitere Informationen
finden Sie in der controllerspezifischen Bedienungsanleitung.
Die primären Anwendungen für diesen Sensor sind Abwasseranwendungen in Kommunen und in der
Industrie. Die LDO-Sensortechnologie verbraucht keinen Sauerstoff und kann die SauerstoffKonzentration in Anwendungen mit geringem oder keinem Durchfluss messen. Siehe Abbildung 1.
Vergewissern Sie sich, dass Sie alle in Abbildung 2 dargestellten Komponenten erhalten haben.
Wenn irgendwelche Positionen fehlen oder beschädigt sind, kontaktieren Sie bitte den Hersteller
oder Verkäufer. Siehe Abbildung 2.
Abbildung 2 Liste der Sensorkomponenten
1 LDO-Sensor
1
Das enthaltene Benutzerhandbuch ist nicht dargestellt.
1
2 Kalibrierungsbeutel (2 x)
Installation
Überprüfen des Sensortyps
G E F A H R
Explosionsgefahr. Verbinden Sie nur periphere Komponenten, die eindeutig für die Verwendung in
gefährlichen Umgebungen nach Klasse 1, Abschnitt 2 zugelassen sind.
H I N W E I S
Die für gefährliche Standorte zertifizierte Version dieses Produkts erfüllt nicht die Anforderungen der
Richtlinie 94/9/EG (ATEX-Richtlinie).
1. Nehmen Sie das Steckerende des Kabels.
Deutsch 21
2. Lesen Sie das Etikett am Steckerende des Kabels. Bei Sensoren, die für gefährliche Standorte
zertifiziert sind, steht Folgendes auf dem Etikett: "Einstufung: Klasse 1, Unterteilung 2".
3. Untersuchen Sie den Anschluss.
•
Für gefährliche Standorte zertifizierte Sensoren verfügen über einen Anschluss mit
Sicherheitsschloss. Siehe Abbildung 3 auf Seite 22.
• Sensoren, die nicht für die Verwendung an gefährlichen Standorten zertifiziert sind, verfügen
über einen Schnellanschlussstecker ohne Sicherheitsschloss.
Anschließen des Sensors an gefährlichen Standorten
G E F A H R
Explosionsgefahr. Mit bestimmten Sensoren und wenn es gemäß der Kontrollzeichnung zur Installation
an gefährlichen Standorten installiert wurde, eignet sich dieses Gerät zur Verwendung an
ungefährlichen Standorten oder an gefährlichen Standorten der Klasse I, Unterteilung 2, Gruppen A, B,
C, D. Ziehen Sie für ordnungsgemäße Installationsanweisungen immer die Kontrollzeichnung und die
geltenden elektrischen Sicherheitsvorschriften heran.
Explosionsgefahr. Elektrische Komponenten oder Schaltungen nur bei getrennter Stromversorgung
oder in ungefährlichen Bereichen anschließen oder abtrennen.
Verwenden Sie an gefährlichen Standorten nur Sensoren und Sicherungskabel, die für gefährliche Standorte
zertifiziert sind. Die für gefährliche Standorte zertifizierte Version dieses Produkts erfüllt nicht die Anforderungen
der Richtlinie 94/9/EG (ATEX-Richtlinie).
Für weitere Informationen siehe Überprüfen des Sensortyps auf Seite 21.
1. Nehmen Sie die Steckerabdeckung vom Controller ab. Behalten Sie die Steckerabdeckung, um
die Steckeröffnung abzudichten, wenn der Sensor entnommen wird.
2. Schließen Sie den Sensor an den Controller an. Weitere Informationen finden Sie im
Benutzerhandbuch des Controllers.
3. Schließen Sie das Sicherheitsschloss über dem Anschluss.
4. Entfernen Sie das Sicherheitsschloss des Anschlusses mit einem kleinen flachen
Schraubendreher. Siehe Abbildung 3.
Abbildung 3 Anschluss-Sicherheitsschloss
G E F A H R
H I N W E I S
22 Deutsch
Anschließen des Sensors an ungefährlichen Standorten
Weitere Informationen zum Anschließen eines LDO-Sensors an einen SC-Controller finden Sie in
Abbildung 4. Anweisungen für die Festverdrahtung finden Sie in dem Handbuch zum jeweiligen SC-
Controller.
Abbildung 4 Anschließen des LDO-Sensors (ungefährlicher Standort)
Nachdem der Sensor angebracht ist, führen Sie eine Abfrage danach durch. Siehe Installieren des
Sensors auf Seite 23
.
Installieren des Sensors
Es gibt zwei Optionen zum Installieren des Sensors.
• Schließen Sie den Sensor an, während der Controller ausgeschaltet ist. Der Controller sucht nach
neuen Sensoren und installiert diese, sobald er eingeschaltet wird.
• Schließen Sie den Sensor an, während der Controller eingeschaltet ist. Installieren Sie den neuen
Sensor mit Hilfe der Menüauswahl "Geräte suchen".
OptionenBeschreibung
sc200 ControllerRufen Sie MENÜ>SERVICE>GERÄTE SCANNEN auf.
sc100 ControllerRufen Sie MENÜ>SERVICE>SUCHE SENSOREN auf.
sc1000 Controller Rufen Sie MENÜ>SYSTEM-SETUP>GERÄTE VERWALTUNG>SUCHE NEUE GERÄTE
auf.
Informationen zum Anschließen digitaler Sensoren finden Sie unter Anschließen des Sensors an
ungefährlichen Standorten auf Seite 23.
Installationsoptionen für den Sensor
Die für den LDO-Sensor verfügbaren Installations- und Zubehöroptionen werden mit den
Installationsanweisungen im Hardwarekit bereitgestellt. In Abbildung 5 sind mehrere
Installationsoptionen dargestellt. Informationen zum Bestellen von Installations-Hardware finden Sie
unter Ersatzteile und Zubehör auf Seite 33.
Eine Beschreibung der Bedienung und Menüführung entnehmen Sie bitte der ControllerDokumentation.
Konfigurieren des Sensors
Verwenden Sie das Menü „Konfigurieren“ zur Eingabe von Information zur Sensoridentifikation sowie
zur Änderung von Optionen zur Datenverarbeitung und -speicherung.
Weitere Informationen zur Sensormontage finden Sie unter Installieren des Sensors auf Seite 23.
Vergewissern Sie sich, dass alle Werte des Menüs "Konfiguration" für die Anwendung korrekt sind.
1. Rufen Sie MENÜ>SENSOR SETUP>[Sensor auswählen]>KONFIGURIEREN auf.
2. Wählen Sie eine Option aus, und drücken Sie EINGABE. In der folgenden Tabelle wird eine Liste
der verfügbaren Optionen angezeigt.
24 Deutsch
OptionenBeschreibung
EDIT NAMEÄndert den Namen des Messorts für den Sensor oben in der Messungsanzeige. Der
EINH. FESTLEGENTEMP: Legt die Temperatureinheit fest (°C oder °F)
HÖHE/DRUCKGeben Sie den Wert für die Höhe oder den Atmosphärendruck ein. Dieser Wert
SALINITÄTGeben Sie den Wert des Salzgehalts ein. Bereich des Salzgehalts: 0,00 bis
SIGNAL MITTELWLegen Sie das Zeitintervall für den Mittelwert des Signals in Sekunden fest.
RNG-INTERVALLLegen Sie das Zeitintervall für die manuelle Sensorreinigung in Tagen fest
RNG-ITVL RÜCKSSetzt das Reinigungsintervall zurück.
LOG SETUP (Einrichten
der Protokollierung)
WERKS-KONFIG.Stellt die konfigurierbaren Standardwerte des Sensors wieder her. Ändert nicht die
Name darf maximal 10 Zeichen lang sein und Buchstaben, Zahlen, Leerzeichen und
Satzzeichen enthalten.
MASSEINHEIT: Legt die Maßeinheiten in mg/l, ppm oder % fest.
HÖHE/DRUCK: Legt die Höhe in m oder Fuß fest oder aber die Einheiten des
Atmosphärendrucks in mmHg oder Torr. (Standardwert = 0 m)
muss genau sein, um die Messung der Sättigung in % und die Kalibrierung in Luft
abschließen zu können. (Standardwert = 0 m)
250,00 Promille (‰). Weitere Informationen finden Sie unter Eingeben eines
Salzgehalt-Korrekturwerts auf Seite 26. (Standardwert = 0)
(Standardwert = 0 Tage. Mit einem Wert von 0 Tagen wird das Reinigungsintervall
deaktiviert.)
Legt das Zeitintervall für die Datenspeicherung im Datenprotokoll fest: 0,5, 1, 2, 5,
10, 15 (Standardwert), 30 oder 60 Minuten.
Einstellung für Steilheit oder Offset.
Eingeben des Atmosphärendrucks
Die Werkseinstellung für den atmosphärischen (Luft-)Druck entspricht 0 m bzw. Normalnull. Befolgen
Sie die Schritte in dieser Vorgehensweise, um den Standardwert zu ändern. Die Anpassung des
Luftdrucks wird entweder als Höhe oder in Druckeinheiten (empfohlen) eingegeben.
Hinweis: Für die Kalibrierung in gesättigter Luft (Kalibrierung mit Luft auf Seite 27
entscheidend. Verwenden Sie nur den absoluten Luftdruck, nicht den angepassten. Falls der absolute Luftdruck
nicht bekannt ist, verwenden Sie die dem Standort entsprechende Höhe.
) ist der genaue Luftdruck
1. Rufen Sie MENÜ>SENSOR SETUP>[Sensor auswählen]>KONFIGURIEREN>EINHEIT?
>HÖHE/DRUCK auf.
2. Wählen Sie eine der aufgeführten Einheitenoptionen aus:
OptionenBeschreibung
0,3 m ( ft)Fuß: Maßeinheit für Höhe
mMeter: metrische Maßeinheit für Höhe
mmHgMillimeter Quecksilber: metrische Maßeinheit für absoluten Luftdruck
TorrMaßeinheit für absoluten Luftdruck
3. Bestätigen Sie die Auswahl. Im Bildschirm für die Werteeingabe werden die ausgewählten
Einheiten angezeigt.
4. Geben Sie den Wert ein und bestätigen Sie diesen anschließend.
Deutsch 25
Eingeben eines Salzgehalt-Korrekturwerts
Bei Messungen von gelöstem Sauerstoff in salzigen Proben kann ein Sauerstoff-Wert angezeigt
werden, der sich stark von dem tatsächlichen Sauerstoff-Wert unterscheidet. Zur Korrektur des
Einflusses von gelösten Salzen in einer Probe geben Sie einen Salzgehalt-Korrekturwert ein.
Hinweis: Falls das Vorhandensein von Salz oder der Grad des Salzgehalts in dem Prozess unbekannt ist,
konsultieren Sie das technische Personal der Behandlungsanlage.
1. Mit einer Leitfähigkeitsmessung bestimmen Sie die Leitfähigkeit der Probe in mS/cm bei einer
Referenztemperatur von 20 °C.
2. Anhand von Tabelle 1 schätzen Sie den Korrekturfaktor für den Salzgehalt in Promille (‰) der
Sättigung.
Hinweis: Die Chloridionenkonzentration in g/kg entspricht dem Chloridgehalt in der Probe. Der Salzgehalt wird
mit folgender Formel berechnet: Salzgehalt = 1,80655 x Chloridgehalt.
Der Salzgehalt kann anhand des Verhältnisses in Abschnitt 2520 B der Standardmethoden für
die Untersuchung von Wasser und Abwasser
3. Rufen Sie MENÜ>SENSOR SETUP>[Sensor auswählen]>KONFIGURIEREN>SALZGEHALT
auf.
4. Geben Sie den Salzgehalt-Korrekturwert ein und bestätigen Sie diesen.
Tabelle 1 Salzgehaltsättigung (‰) pro Leitfähigkeitswert (mS/cm)
mS/cm‰mS/cm‰mS/cm‰mS/cm‰
5 3 16 10 27 18 38 27
6 4 17 11 28 19 39 28
7 4 18 12 29 20 40 29
8 5 19 13 30 21 42 30
9 6 20 13 31 22 44 32
10 6 21 14 32 22 46 33
11 7 22 15 33 23 48 35
12 8 23 15 34 24 50 37
13 8 24 17 35 25 52 38
14 9 25 17 36 25 54 40
15 10 26 18 37 26
berechnet werden.*,
Konfigurieren der Stromausgänge am Controller
Mit Hilfe der Stromausgänge können die Messwerte an SPS, PLS oder andere
Datenerfassungssysteme der Anlage übermittelt werden.
1. Rufen Sie im Controller das Menü zum Einrichten des Ausgangs auf.
Optionen Beschreibung
sc200Rufen Sie MENÜ>SC200 EINSTELLUNGEN>STROMAUSGÄNGE>AUSGANG
*
Standardmethoden für die Untersuchung von Wasser und Abwasser, 20. Ausgabe. Herausgeber
Lenore S. Clesceri, Arnold E. Greenberg und Andrew D. Eaton, S. 2-48-2-29 (1998). Das
Verhältnis zwischen Chloridgehalt und Sauerstofflöslichkeit ist in derselben Referenz in 4500-O:I
S. 4-131 angegeben.
26Deutsch
WÄHLEN>FUNKTION EINSTELLEN auf.
Optionen Beschreibung
sc100Rufen Sie MENÜ>SYSTEM SETUP>STROMAUSGÄNGE>mA AUSGANG>WÄHLE
sc1000Rufen Sie MENÜ>SYSTEM SETUP>STROMAUSGÄNGE>mA AUSGANG>WÄHLE
AUSGANG>FUNKTION EINSTELLEN auf.
AUSGANG>FUNKTION EINSTELLEN auf.
2. Legen Sie die Funktion für den Controller fest.
OptionenBeschreibung
sc200LINEAR
sc100LINEARE STEUERUNG (Standardwert)
sc1000LINEARE STEUERUNG (Standardwert)
Modbus-Register
Für die Netzwerkkommunikation ist eine Liste der Modbus-Register verfügbar. Weitere Informationen
finden Sie unter www.hach.com oder www.hach-lange.com.
Kalibrierung für Messungen
Der Sensor wird im Werk gemäß den Spezifikationen kalibriert. Der Hersteller empfiehlt keine
Kalibrierung, es sei denn, dies wird regelmäßig von Kontrollorganen gefordert. Falls eine Kalibrierung
erforderlich ist, lassen Sie den Sensor mit dem Prozess ins Gleichgewicht kommen, bevor Sie eine
Kalibrierung durchführen. Kalibrieren Sie den Sensor nicht beim Setup.
In Tabelle 2 werden Optionen für die Kalibrierung angezeigt.
Tabelle 2 Kalibrieroptionen
OptionenBeschreibung
LUFTKAL.Empfohlene Kalibrierungsmethode. Durch diese Kalibrierung wird die Steilheit verändert.
IN-LINE KAL.Kalibrierung durch Vergleich mit einem tragbaren Sauerstoff-Messgerät. Durch diese
STDKAL RÜCKS Setzt den Stellfaktor der Kalibrierung (Steilheit) und den Offset auf den Werksstandard:
Vergewissern Sie sich, dass sich Wasser im Kalibrierungsbeutel befindet.
• Vergewissern Sie sich, dass die Dichtung zwischen dem Kalibrierungsbeutel und dem
Sensorgehäuse dicht ist.
• Vergewissern Sie sich, dass der Sensor beim Kalibrieren trocken ist.
• Vergewissern Sie sich, dass die Luftdruck-/Höheneinstellung für den Kalibrierungsstandort akkurat
ist.
• Lassen Sie dem Sensor genügend Zeit, die Temperatur des Standortes des Kalibrierungsbeutels
anzunehmen. Bei einem großen Temperaturunterschied zwischen dem Prozess und dem
Kalibrierungsstandort kann die Temperaturstabilisierung bis zu 15 Minuten in Anspruch nehmen.
1. Entnehmen Sie den Sensor aus dem Prozess. Reinigen Sie den Sensor mit einem feuchten
Tuch.
2. Geben Sie den gesamten Sensor in einen Kalibrierungsbeutel mit 25 bis 50 ml Wasser.
Vergewissern Sie sich, dass die Sensorkappe keinen Kontakt zu dem Wasser im
Kalibrierungsbeutel hat und dass sich keine Wassertropfen auf der Sensorkappe befinden
(Abbildung 6).
Deutsch 27
3. Dichten Sie mit Hilfe eines Gummibands, eines Bindfadens oder der Hand den Kalibrierbeutel
zum Sensor hin ab.
4. Lassen Sie dem Messgerät vor der Kalibrierung 15 Minuten Zeit, sich zu stabilisieren. Setzen Sie
den Kalibrierungsbeutel während der Kalibrierung nicht direktem Sonnenlicht aus.
5. Vergewissern Sie sich, dass der aktuelle absolute Luftdruck bzw. die Höhenangabe
ordnungsgemäß konfiguriert ist. Siehe Eingeben des Atmosphärendrucks auf Seite 25.
Hinweis: Der Hersteller empfiehlt die Verwendung des absoluten bzw. tatsächlichen Luftdrucks als empfohlene
Vorgehensweise.
6. Rufen Sie MENÜ>SENSOR SETUP>[Sensor auswählen]>KALIBRIEREN>LUFTKAL. auf.
7. Wählen Sie eine Option für das Ausgangssignal während der Kalibrierung:
OptionenBeschreibung
AktivDas Gerät übermittelt während der Kalibrierung den aktuellen Messwert.
HaltenDas Gerät hält den aktuellen Messwert während der Kalibrierung.
Ersatzwert Das Gerät übermittelt während der Kalibrierung einen vorher eingestellten Wert. Informationen
zur Einstellung des Ersatzwertes entnehmen Sie bitte der Bedienungsanleitung des Controllers.
8. Auf dem Controller wird "Fühler in Beutel stecken" angezeigt. Lassen Sie dem Wert Zeit, sich zu
stabilisieren. Drücken Sie ENTER, um den stabilen Wert zu akzeptieren. Sie können die
Kalibrierung auch weiterlaufen lassen, bis auf dem Display "Abgeschlossen" angezeigt wird.
9. Wenn der Sensor kalibriert ist, fügen Sie den Sensor in den Prozess ein. Drücken Sie ENTER.
Abbildung 6 Luftkalibrierung
Wenn sich der Wert nicht stabilisiert, wird in der Anzeige "Kann nicht kalibrieren" gefolgt von einer
Fehlermeldung angezeigt. In Tabelle 3 werden die Fehlermeldung und die Auflösung für
Kalibrierungsprobleme angezeigt.
Tabelle 3 Fehlermeldungen beim Kalibrieren mit Luft
MeldungBeschreibungLösung
Kal. n. erfolgr., Stellfaktor hochDer berechnete Verstärkungswert ist zu hoch.Kalibrierung wiederholen.
Kal. n. erfolgr., Stellfaktor niedrig Der berechnete Verstärkungswert ist zu niedrig.Kalibrierung wiederholen.
Kal. n. erfolgr., instabilDer Wert hat sich in der maximal zulässigen
Kalibrierungszeit nicht stabilisiert.
Kalibrierung wiederholen.
28 Deutsch
IN-LINE KAL: Kalibrierung durch Vergleich
Bei dieser Kalibrierung wird ein alternativer Sensor verwendet, der an ein tragbares Messgerät
angeschlossen ist.
1. Fügen Sie den alternativen Sensor in den Prozess ein. Fügen Sie den zweiten Sensor so nahe
wir möglich neben dem ersten Sensor ein.
2. Warten Sie, bis der Sauerstoff-Wert stabil ist.
3. Rufen Sie auf dem Controller für den ersten Sensor MENÜ>SENSOR SETUP>[Sensor
auswählen]>KALIBRIEREN>IN-LINE KAL. auf.
4. Wählen Sie eine Option für das Ausgangssignal während der Kalibrierung:
OptionenBeschreibung
AktivDas Gerät übermittelt während der Kalibrierung den aktuellen Messwert.
HaltenDas Gerät hält den aktuellen Messwert während der Kalibrierung.
Ersatzwert Das Gerät übermittelt während der Kalibrierung einen vorher eingestellten Wert. Informationen
5. Auf dem Controller wird Folgendes angezeigt:
•
"Drücke ENTER wenn stabil."
• Die aktuelle Messung des gelösten Sauerstoffs
• Die aktuelle Temperaturmessung
6. Wenn der Messwert stabil ist, drücken Sie ENTER. Auf dem Display wird ein Eingabebildschirm
angezeigt.
Hinweis: Die Messung stabilisiert sich in der Regel in 2 bis 3 Minuten.
Wenn sich der Wert nicht stabilisiert, wird in der Anzeige "Kann nicht kalibrieren" gefolgt von einer
Fehlermeldung angezeigt. In Tabelle 4 werden die Fehlermeldung und die Auflösung für
Kalibrierungsprobleme angezeigt.
MeldungBeschreibungLösung
Kal. n. erfolgr., Offset hochDer berechnete Offsetwert ist zu hoch.Kalibrierung wiederholen.
Kal. n. erfolgr., Offset niedrig Der berechnete Offsetwert ist zu niedrig.Kalibrierung wiederholen.
Kal. n. erfolgr., instabilDer Wert hat sich in der maximal zulässigen
zur Einstellung des Ersatzwertes entnehmen Sie bitte der Bedienungsanleitung des Controllers.
Tabelle 4 Fehlermeldungen bei der IN-LINE Kalibrierung
Kalibrierung wiederholen.
Kalibrierungszeit nicht stabilisiert.
Verlassen der Kalibrierungsprozedur
1. Drücken Sie während der Kalibrierung die Taste ZURÜCK. Es werden drei Optionen angezeigt:
OptionenBeschreibung
ABBRECHENStoppt die Kalibrierung. Eine neue Kalibrierung muss von Anfang an ausgeführt werden.
ZURÜCK ZU KAL Kehrt zur aktuellen Kalibrierung zurück.
VERLASSENVerlässt die Kalibrierung vorübergehend. Zugriff auf andere Menüs ist zulässig, während
2. Wählen Sie eine der Optionen aus. Bestätigen Sie die Auswahl.
die Kalibrierung im Hintergrund fortgesetzt wird. Sofern vorhanden, kann die Kalibrierung
eines zweiten Sensors gestartet werden. Um zur Kalibrierung zurückzukehren, drücken
Sie MENÜ und wählen Sie „Sensor-Setup“, [Sensor wählen].
Deutsch 29
Zurücksetzen der Kalibrierung auf die Standardwerte
Kalibrierungseinstellungen können auf die werkseitigen Standardwerte zurückgesetzt werden. Die
Werte für Stellfaktor und Offset werden auf 1,0 bzw. 0,0 gesetzt.
1. Rufen Sie MENÜ>SENSOR SETUP>[Sensor
auswählen]>KALIBRIEREN>ZURÜCKSETZEN>STDKAL RÜCKS auf.
2. Im Display wird eine Bestätigungsmeldung angezeigt. Bestätigen Sie, dass Sie den Sensor auf
die werkseitige Standardkalibrierungskurve zurücksetzen möchten.
Wartung
G E F A H R
Mehrere Gefahren. Nur qualifiziertes Personal sollte die in diesem Kapitel des Dokuments
beschriebenen Aufgaben durchführen.
G E F A H R
Explosionsgefahr. Elektrische Komponenten oder Schaltungen nur bei ausgeschalteter
Stromversorgung oder in ungefährlichen Bereichen anschließen oder abtrennen.
G E F A H R
Explosionsgefahr. Das Austauschen von Komponenten kann die Eignung für Klasse I, Unterteilung 2,
beeinträchtigen. Komponenten nur bei ausgeschalteter Stromversorgung und in ungefährlichen
Bereichen austauschen.
Die für gefährliche Standorte zertifizierte Version dieses Produkts erfüllt nicht die Anforderungen der
Richtlinie 94/9/EG (ATEX-Richtlinie).
H I N W E I S
Wartungsplan
Im Wartungsplan sind die Mindestintervalle für die regelmäßige Wartung angegeben. Führen Sie die
Wartungsaufgaben häufiger aus, wenn die Anwendungsbedingungen zum Zusetzen oder zu
Verunreinigungen der Elektrode führen.
Hinweis: Nehmen Sie das Gerät zur Wartung oder Reinigung nicht auseinander.
WartungsaufgabeEmpfohlene Häufigkeit
Reinigen des Sensors90 Tage
Sensor auf Schäden prüfen90 Tage
Kalibrieren des SensorsWie von Kontrollorganen empfohlen
Reinigen des Sensors
Reinigen Sie das Äußere des Sensors mit einem weichen, feuchten Tuch.
Hinweis: Wenn die Sensorkappe zum Reinigen entfernt werden muss, setzen Sie das Innere der Kappe nicht über
einen längeren Zeitraum direktem Sonnenlicht aus.
Festlegen oder Ändern des Reinigungsintervalls
Anwendungsbedingungen erfordern u. U. kürzere oder längere Zeitabstände zwischen manuellen
Sensorreinigungen. Das Standardreinigungsintervall beträgt 0 Tage. Befolgen Sie die Schritte in
dieser Vorgehensweise, um das Intervall zu ändern.
1. Rufen Sie MENÜ>SENSOR SETUP>[Sensor auswählen]>KONFIGURIEREN>RNG-INTERVALL
auf.
30 Deutsch
Loading...