Hach-Lange LCK 303, LCK 304, LCK 305 User Manual [en, de, es, fr, it]

DOC027.98.00789
POCKET COLORIMETERTM II -
Ammonium/Amonio/Ammoniaca
LCK 303/304/305
APPLICATION INSTRUCTION
PROGRAMMIERANLEITUNG
INSTRUCCIONES DE APLICACIÓN
ISTRUZIONI APPLICATIVE
GEBRUIKSAANWIJZING
INSTRUKTIONSHANDBOK
10/2010, Edition 2A
10/2010, Ausgabe 2A
10/2010, édition 2A
10/2010, Edición 2A
10/2010, edizione 2A
10/2010, Editie 2A
10/2010, Utgåva 2A
© HACH LANGE, 2010. All rights reserved. Printed in Germany
2
Table of contents
Section 1 General information .......................................................................5
1.1 Safety information .......................................................................................5
1.1.1 Use of hazard information ................................................................5
1.2 General product information .......................................................................5
Section 2 Installation ........................................................................................7
2.1 Packing list ..................................................................................................7
2.2 Insert the adapter ........................................................................................7
2.3 Replace the instrument cap ........................................................................7
Section 3 System start up ...............................................................................9
3.1 Instrument keys and display .......................................................................9
3.1.1 Key-Functions ...................................................................................9
3.2 Enter calibration data of HACH-LANGE Cuvette Tests ..............................9
3.2.1 Overview ...........................................................................................9
3.2.2 Data table .......................................................................................10
3.3 Programming procedure ........................................................................... 11
3.3.1 LCK 303 Ammonium-Nitrogen NH
3.3.2 LCK 304 Ammonium-Nitrogen NH4-N ............................................ 12
3.3.3 LCK 305 Ammonium-Nitrogen NH
3.4 Check entered data .................................................................................. 14
3.5 Overwrite stored calibration data ..............................................................15
3.5.1 Retrieve the factory calibration .......................................................15
Section 4 Operation ........................................................................................ 17
4.1 Analyse Ammonium-Nitrogen/Ammonium
with HACH-LANGE Cuvette Tests ....................... ................................17
4.2 Quality assurance ............................................ .........................................17
4.3 Evaluate HACH-LANGE Ammonium-Nitrogen/
Ammonium Cuvette Tests ...................................................................17
Section 5 Replacement Parts and Accessories ....................................19
5.1 Replacement Parts ...................................................................................19
5.2 Accessories ..............................................................................................19
Section 6 How to order ..................................................................................21
-N ............................................ 11
4
-N ............................................ 13
4
3
Table of contents
4
Section 1 General information
1.1 Safety information
Please read this entire application instruction and the POCKET COLORIMETERTM II instruction manual before unpacking, setting up, or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
To ensure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this manual.
1.1.1 Use of hazard information
DANGER Indicates a potentially or imminently hazardo us situation which, if not avoided, could result in death or serious injury.
Important Note: Information that requires special emphasis.
Note: Information that supplements points in the main text.
1.2 General product information
This application instruction describes how to evaluate HACH-LANGE Cuvette Tests on a single-wavelength POCKET COLORIMETER II.
5
General information
6
Section 2 Installation
Important Note: This application instruction is only valid in conjunction with the POCKET COLORIMETERTM II instruction manual and the applied LCK working procedure. Please read this entire application instruction and the entire POCKET COLORIMETER II instruction manual bef or e un packing, setting up or operating the instrument.
2.1 Packing list
POCKET COLORIMETER II 655 nm, Cat. No. 5870065
Instrument cap with HACH-LANGE label, Cat. No. LZV 759
Cuvette adapter 1 cm, Cat. No. 5954600
Blank value cuvettes, Cat. No. LCW 919
Application instruction DOC027.98.00789
2.2 Insert the adapter
Remove the instrument cap and carefully position the cuvette adapter into the cell holder with the arrow pointing downward. Fix it by pressing it down.
2.3 Replace the instrument cap
Use the instrument cap with the HACH-LANGE label when evaluating HACH-LANGE cuvettes.
7
Installation
8
Section 3 System start up
1 absorbance 21 absorbance 2
655 nm
3.1 Instrument keys and display
Figure 1 Instrument keys and display
1 POWER/BACKLIGHT 6 Measuring range indicator (2) 2 ZERO/SCROLL (BLUE key)7Menu indicator 3MENU 8Calibration adjusted indicator 4 Numeric display 9 Battery low indicator 5 Measuring range indicator (1) 10 READ/ENTER (GREEN key)
3.1.1 Key-Functions
ZERO/SCROLL (BLUE key) : In measurement mode, sets the instrument to zero. In
menu mode, scrolls through menu options. Also scrolls numbers when entering or editing a value.
READ/ENTER (GREEN key) : In measurement mode, initiali ze s sample measurement. In menu mode, selects a menu option. When entering numbers, moves one space to the right. Confirm the function when the entry is complete.
3.2 Enter calibration data of HACH-LANGE Cuvette Tests
3.2.1 Overview
Each POCKET COLORIMETER II has two channels in which measurements can be made. Each channel (measuring range) will accept user-defined calibration data.
9
System start up
This application instruction contains the calibration data of the HACH-LANGE Cuvette Tests for Ammonium-Nitrogen/Ammonium:
LCK 303: 2-47 mg/l NH
-N or 2.5-60 mg/l NH
4
4
LCK 304: 0.015-2 mg/l NH4-N or 0.02-2.5 mg/l NH4
LCK 305: 1-12 mg/l NH
-N or 1.2-15 mg/l NH
4
4
Select the calibration data of maximum two measuring ranges which will be the suitable one.
Each programming procedure has only to be followed once prior to the first test evaluation.
For detailed information please read the POCKET COLORIMETER II instruction manual.
3.2.2 Data table
3.2.2.1 Evaluation as Ammonium-Nitrogen NH4-N Table 1 Evaluation as Ammonium-Nitrogen NH4-N
LCK 303 LCK 304 LCK 305 RES (resolution) S0 (standard 0) A0 (absorbance 0) S1 (standard 1) A1 (absorbance 1)
--.-- -.--- --.--
02.00 0.015 00.99
0.113 0.044 0.203
47.00 2.000 12.00
1.862 1.832 1.871
3.2.2.2 Evaluation as Ammonium NH T able 2 Evaluation as Ammonium NH
LCK 303 LCK 304 LCK 305 RES (resolution) S0 (standard 0) A0 (absorbance 0) S1 (standard 1) A1 (absorbance 1)
--.-- -.--- --.--
02.50 0.020 01.30
0.110 0.047 0.198
60.00 2.500 15.00
1.845 1.783 1.788
4
4
10
System start up
3.3 Programming procedure
Note: Each programming procedure has only to be followed once prior to the first test evaluation.
Important Note: The following programming procedures are valid for the evaluation form Ammonium-Nitrogen NH4-N. To evaluate as Ammonium NH4 please choose the data mentioned in table 3.2.2.2 Evaluation as Ammonium NH4 on page 10.
Note: After programming the calibration data, it could be helpful to mark the channel(s) 1 and/or 2 accordingly with a label.
3.3.1 LCK 303 Ammonium-Nitrogen NH4-N
1. Press POWER to switch on the instrument.
2. Press MENU. "SEL" is displayed. A flashing arrow indicates the current
measuring range (1 or 2).
3. Select measuring range 1 or 2 for LCK 303 (2-47 mg/l NH4-N).
Note: Press GREEN key to toggle between measuring ranges.
4. Press MENU for about five seconds until "USER" followed by "CAL" is displayed.
5. Press BLUE key until "Edit" is displayed. Press GREEN key. " RES" is displayed.
6. Press GREEN key. "-.---" is displayed.
7. Press BLUE key until "--.--" is displayed. Press GREEN key. "S0" is displayed.
8. Press GREEN key until "A0" is displayed.
9. Press GREEN key until "S1" is displayed.
10. Press GREEN key. Enter "47.00" by using BLUE key to change value and GREEN key to move to the next space. Press GREEN key until "A1" is
displayed.
11. Press GREEN key. Enter "1.862" by using BLUE key to change value and GREEN key to move to the next space. Press GREEN key until "Add" is
displayed.
12. Press MENU until "Edit" is displayed. Press GREEN key until „S0“ is displayed.
13. Press GREEN key. Enter "02.00" by using BLUE key to change value and GREEN key to move to the next space. Press GREEN key until "A0" is
displayed.
14. Press GREEN key. Enter "0.113" by using BLUE key to change value and GREEN key to move to the next space. Press GREEN key until „S1“ is
displayed.
15. Press MENU until "Edit" is displayed to check the entered data. Press GREEN key. „RES“ is displayed.
11
System start up
Press BLUE key to scroll through the entries. Compare the entries with the data shown in the table below.
Table 3 Calibration data of LCK 303
RES --.--
S0 2.00 A0 0.113 S1 47.00 A1 1.862
16. After comparing the entries press MENU two times to enter the measuring mode. See 4.3 Evaluate HACH-LANGE Ammonium-Nitrogen/ Ammonium
Cuvette Tests on page 17.
Note: In case of any mistake retrieve the factory calibration first. See 3.5.1 Retrieve the factory
calibration on page 15. After retrieving the factory calibration, start the programming procedure
again. If the instrument is shut-off or loses power during data entry, all edits will be lost. Automatic shut-off is 60 minutes.
3.3.2 LCK 304 Ammonium-Nitrogen NH4-N
1. Press POWER to switch on the instrument.
2. Press MENU. "SEL" is displayed. A flashing arrow indicates the current
measuring range (1 or 2).
3. Select measuring range 1 or 2 for LCK 304 (0.015-2 mg/l NH4-N).
Note: Press GREEN key to toggle between measuring ranges.
4. Press MENU for about five seconds until "USER" followed by "CAL" is displayed.
5. Press BLUE key until "Edit" is displayed. Press GREEN key. " RES" is displayed.
6. Press GREEN key. "-.---" is displayed.
7. Press GREEN key. "S0" is displayed.
8. Press GREEN key until "A0" is displayed.
9. Press GREEN key until "S1" is displayed.
10. Press GREEN key. Enter "2.000" by using BLUE key to change value and GREEN key to move to the next space. Press GREEN key until "A1" is
displayed.
11. Press GREEN key. Enter "1.832" by using BLUE key to change value and GREEN key to move to the next space. Press GREEN key until "Add" is
displayed.
12. Press MENU until "Edit" is displayed. Press GREEN key until „S0“ is displayed.
13. Press GREEN key. Enter "0.015" by using BLUE key to change value and GREEN key to move to the next space. Press GREEN key until "A0" is
displayed.
12
System start up
14. Press GREEN key. Enter "0.044" by using BLUE key to change value and GREEN key to move to the next space. Press GREEN key until „S1“ is
displayed.
15. Press MENU until "Edit" is displayed to check the entered data. Press GREEN key. „RES“ is displayed. Press BLUE key to scroll through the entries. Compare the entries with the data shown in the table below.
Table 4 Calibration data of LCK 304
RES -.---
S0 0.015 A0 0.044 S1 2.000 A1 1.832
16. After comparing the entries press MENU two times to enter the measuring mode. See 4.3 Evaluate HACH-LANGE Ammonium-Nitrogen/ Ammonium
Cuvette Tests on page 17.
Note: In case of any mistake retrieve the factory calibration first. See 3.5.1 Retrieve the factory
calibration on page 15. After retrieving the factory calibration, start the programming procedure
again. If the instrument is shut-off or loses power during data entry, all edits will be lost. Automatic shut-off is 60 minutes.
3.3.3 LCK 305 Ammonium-Nitrogen NH4-N
1. Press POWER to switch on the instrument.
2. Press MENU. "SEL" is displayed. A flashing arrow indicates the current
measuring range (1 or 2).
3. Select measuring range 1 or 2 for LCK 305 (1-12 mg/l NH
Note: Press GREEN key to toggle between measuring ranges.
4. Press MENU for about five seconds until "USER" followed by "CAL" is displayed.
5. Press BLUE key until "Edit" is displayed. Press GREEN key. " RES" is displayed.
6. Press GREEN key. "-.---" is displayed.
7. Press BLUE key until "--.--" is displayed. Press GREEN key until "S0" is
displayed.
8. Press GREEN key until "A0" is displayed.
9. Press GREEN key until "S1" is displayed.
10. Press GREEN key. Enter "12.00" by using BLUE key to change value and GREEN key to move to the next space. Press GREEN key until "A1" is
displayed.
11. Press GREEN key. Enter "1.871" by using BLUE key to change value and GREEN key to move to the next space. Press GREEN key until "Add" is
displayed.
4
-N).
13
System start up
12. Press MENU until "Edit" is displayed. Press GREEN key until „S0“ is displayed.
13. Press GREEN key. Enter "00.99" by using BLUE key to change value and GREEN key to move to the next space. Press GREEN key until "A0" is
displayed.
14. Press GREEN key. Enter "0.203" by using BLUE key to change value and GREEN key to move to the next space. Press GREEN key until „S1“ is
displayed.
15. Press MENU until "Edit" is displayed to check the entered data. Press GREEN key. „RES“ is displayed. Press BLUE key to scroll through the entries. Compare the entries with the data shown in the table below.
Table 5 Calibration data of LCK 305
RES --.--
S0 0.99 A0 0.203 S1 12.00 A1 1.871
16. After comparing the entries press MENU two times to enter the measuring mode. See 4.3 Evaluate HACH-LANGE Ammonium-Nitrogen/ Ammonium
Cuvette Tests on page 17.
Note: In case of any mistake retrieve the factory calibration first. See 3.5.1 Retrieve the factory
calibration on page 15. After retrieving the factory calibration, start the programming procedure
again. If the instrument is shut-off or loses power during data entry, all edits will be lost. Automatic shut-off is 60 minutes.
3.4 Check entered data
To avoid erroneous measurements it is recommended to check the data after completing the programme procedure.
1. Press MENU for about five seconds until "USER" followed by "CAL" is displayed.
2. Press BLUE key to scroll to "Edit". Press GREEN key.
3. "RES" is displayed. Press BLUE key to scroll through the entries. Compare the
entries with the figures in the 3.2.2.1 Evaluation as Ammonium-Nitrogen NH4-N
on page 10.
In case a value was not stored correctly, retrieve the factory calibration first. See 3.5.1
Retrieve the factory calibration on page 15. After retrieving the factory calibration,
start the programming procedure again.
14
System start up
3.5 Overwrite stored calibration data
With the POCKET COLORIMETER II it is possible to overwrite stored calibration data. Before overwriting calibration data it is recommended to retrieve the factory calibration first. After retrieving the factory calibration, you can program new calibration data. In case of any mistake retrieve the factory calibration first.
3.5.1 Retrieve the factory calibration
1. Press MENU. "SEL" is displayed. A flashing arrow indicates the current
measuring range (1 or 2). Press GREEN key to toggle between measur ing ranges. Select measuring range retrieving the factory calibration.
1. Press MENU for about five seconds until "USER" followed by "CAL" is displayed.
2. Press BLUE key to scroll to "dFL". Press GREEN key.
15
System start up
16
Section 4 Operation
4.1 Analyse Ammonium-Nitrogen/Ammonium with
HACH-LANGE Cuvette Tests
DANGER Handling chemical samples, standards and reagents can be dangerous. Users of this product are advised to familiarize themselves with safety procedures and the correct use of chemicals, and to carefully read all relevant material safety data sheets.
To analyse Ammonium or Ammonium-Nitrogen follow the working procedure enclosed in the according Cuvette Test LCK 303, LCK 304 or LCK 305.
4.2 Quality assurance
The measurement results must be subjected to plausibility checks (dilute and/or spike the sample).
Regular checks with corresponding standard solutions (see Section 5 Replacement
Parts and Accessories on page 19) ensure that your measurement system is
functioning properly and is being correctly operated.
4.3 Evaluate HACH-LANGE Ammonium-Nitrogen/
Ammonium Cuvette Tests
Note: Make sure that the instrument is programmed with the corresponding calibration data, see 3.4 Check entered data on page 14.
1. Fill DI water into an empty cuvette (LCW 919) up to the lower edge of the barcode label.
2. Close the cuvette with the rubber stopper. This is the blank cuvette.
3. Thoroughly clean the outside of the cuvette.
4. Press POWER to switch on the instrument.
5. Make sure that the correct measuring range is selected!
Otherwise press MENU. "SEL" is displayed. A flashing arrow indicates the current measuring range (1 or 2). Press GREEN key to select measuring range. Press MENU to get back to the measuring mode.
6. Insert the blank cuvette into the cell holder equipped with the cuvette adapter 1cm.
Note: Make sure that the cuvette adapter is positioned correctly.
7. Place the instrument cap with the HACH-LANGE label over the cell holder to cover the cuvette.
8. Press BLUE key. "0.00" (LCK 303/305) or "0.000“ (LCK 304) is displayed.
9. Remove the instrument cap and insert the sample cuvette.
10. Place the instrument cap over the cell holder to cover the cuvette.
11. Press GREEN key.
17
Operation
12. The result is displayed in mg/l Ammonium-Nitrogen/Ammonium.
Note: In case the Ammonium-Nitrogen/Ammonium concentration is out of the selected measuring range the value for the lowest or highest concentration is displayed and flashes. Repeat the test using the appropriate LCK Cuvette Test measuring range or dilute your sample.
Note: A new blank measurement should be performed prior to each individual test series.
18
Section 5 Replacement Parts and
Accessories
5.1 Replacement Parts
Description Quantity Cat. Number
POCKET COLORIMETERTM II 655 nm 1 5870065 Instrument cap with HACH-LANGE label 1 LZV 759 Cuvette adapter 1 cm 1 5954600 Blank value cuvettes 1 pk/5 pc LCW 919
5.2 Accessories
Description Quantity Cat. Number
Ammonium Cuvette Test (2-47 mg/l NH4-N) 1 pk/25 pc LCK 303 Ammonium Cuvette Test (0.015-2 mg/l NH Ammonium Cuvette Test (1-12 mg/l NH4-N) 1 pk/25 pc LCK 305 Standard and Ring Test solution for LCK 304 1 LCA 700 Standard and Ring Test solution for LCK 303 1 LCA 703 Standard and Ring Test solution for LCK 305 1 LCA 704 Pipette 0.2-1 ml 1 BBP 078 Pipette tips for BBP 078 1 pk/100 pc BBP 079 Pipette 1-5 ml 1 BBP 065 Pipette tips for BBP 065 1 pk/75 pc BBP 068 Cuvette Stand for HACH-LANGE cuvettes 1 LYW 915
-N) 1 pk/25 pc LCK 304
4
19
Replacement Parts and Accessories
20
Section 6 How to order
Orders/repair service
Please contact your representative:
HACH-LANGE GmbH
Willstätterstraße 11 D-40549 Düsseldorf Tel. +49 (0)2 11 52 88-0 Fax +49 (0)2 11 52 88-143 info@hach-lange.de www.hach-lange.com
HACH LANGE LTD
Pacific Way Salford GB-Manchester, M50 1DL Tel. +44 (0)161 872 14 87 Fax +44 (0)161 848 73 24 info@hach-lange.co.uk www.hach-lange.co.uk
HACH LANGE LTD
Unit 1, Chestnut Road Western Industrial Estate IRL-Dublin 12 Tel. +353(0)1 46 02 5 22 Fax +353(0)1 4 50 93 37 info@hach-lange.ie www.hach-lange.ie
Information Required
Customer account number (if available) Billing address
Your name and phone number Shipping ad dress
Purchase order number Catalog number
Brief description or model number Quantity
21
How to order
22
Inhalt
Abschnitt 1 Allgemeine Informationen ......................................................25
1.1 Sicherheitsinformationen .......................................................................... 25
1.1.1 Bedeutung der Hinweise ................................................................25
1.2 Allgemeine Produktinformationen ............... .. ... ... ......................................25
Abschnitt 2 Installation ................................................................................... 27
2.1 Verpackungsliste ......................................................................................27
2.2 Küvettenadapter einsetzen .......................................................................27
2.3 Austausch der Instrumentenkappe ........................................................... 27
Abschnitt 3 Inbetriebnahme .........................................................................29
3.1 Gerätetasten und Display ............................................................ .............29
3.1.1 Tasten-Funktionen ..........................................................................29
3.2 Eingabe der Kalibrierdaten für HACH-LANGE Küvetten-Tests ................30
3.2.1 Überblick .........................................................................................30
3.2.2 Datentabelle ...................................................................................30
3.3 Programmierung ....................................................................................... 31
3.3.1 LCK 303 Ammonium-Stickstoff NH
3.3.2 LCK 304 Ammonium-Stickstoff NH4-N ........................................... 32
3.3.3 LCK 305 Ammonium-Stickstoff NH
3.4 Dateneingabe kontrollieren ....................................................................... 35
3.5 Überschreiben gespeicherter Kalibrierdaten .............................................35
3.5.1 Grundeinstellung wieder herstellen ................................................35
Abschnitt 4 Betrieb ..........................................................................................37
4.1 Analyse von Ammonium/Ammonium-Stickstoff mit HACH-LANGE
Küvetten-Tests .....................................................................................37
4.2 Qualitätssicherung .................................................................................... 37
4.3 Auswertung der HACH-LANGE Ammonium/Ammonium-Stickstoff
Küvetten-Tests ....................................................................................37
Abschnitt 5 Ersatzteile und Zubehör .........................................................39
5.1 Ersatzteile ................................................................................................. 39
5.2 Zubehör ....................................................................................................39
Abschnitt 6 Bestellhinweis ............................................................................41
-N ........................................... 31
4
-N ........................................... 33
4
23
Inhalt
24
Abschnitt 1Allgemeine Informationen
1.1 Sicherheitsinformationen
Bevor das Gerät ausgepackt, aufgestellt oder in Betrieb genommen wird, empfehlen wir dringend, die Bedienungsanleitung des POCKET COLORIMETER zu lesen. Beachten Sie besonders alle mit GEFAHR und VORSICHT überschriebenen Hinweise. Ein Versäumnis in dieser Richtung könnte zu einer ernsthaften Verletzung des Benutzers oder Schäden am Gerät führen.
Damit die im Gerät vorgesehenen Schutzvorrichtungen nicht beeinträchtigt werden, darf das Gerät auf keinen Fall anders installiert oder benutzt werden, als es in dieser Bedienungsanleitung angegeben ist.
1.1.1 Bedeutung der Hinweise
GEFAHR Weist auf eine potenzielle oder bevorstehende Gefahrensituation hin, deren Nichtbeachtung zu ernsthaften Verletzungen oder sogar zum Tod führen kann.
Wichtiger Hinweis: Diese Information muss un bedingt beachtet werden.
Hinweis: Diese Informationen ergänzen Punkte im Haupttext.
1.2 Allgemeine Produktinformationen
Diese Programmieranleitung beschreibt die Auswertung von HACH-LANGE Küvetten-Tests auf einem POCKET COLORIMETER II mit definierter Wellenlänge.
TM
II sorgfältig
25
Allgemeine Informationen
26
Abschnitt 2Installation
Wichtiger Hinweis: Diese Programmieranleitung ist nur in Verbindung mit der Bedienungsanleitung des POCKET COLORIMETERTM II und den LCK Arbeitsvorschriften gültig. Bevor das Gerät ausgepackt, aufgestellt oder in Betrieb genommen wird, empfehlen wir dringend, diese Programmieranleitung und die Bedienungsanleitung des POCKET COLORIMETER II sorgfältig zu lesen.
2.1 Verpackungsliste
POCKET COLORIMETER II 655 nm, Kat. Nr. 5870065
Instrumentenkappe mit HACH-LANGE Logo, Kat. Nr. LZV 759
Küvettenadapter 1 cm, Kat. Nr. 5954600
Leerküvetten (1 Pk/5 St), Kat. Nr. LCW 919
Programmieranleitung DOC027.98.00789
2.2 Küvettenadapter einsetzen
Instrumentenkappe entfernen. Küvettenadapter vorsichtig mit der Pfeilmarkierung nach vorne in den Küvettenhalter einsetzen und einrasten lassen.
2.3 Austausch der Instrumentenkappe
Zur Auswertung von HACH-LANGE Küvetten-Test s die Instrumentenkappe mit dem HACH-LANGE Logo verwenden.
27
Installation
28
Abschnitt 3Inbetriebnahme
1 absorbance 21 absorbance 2
655 nm
3.1 Gerätetasten und Display
Abbildung 1 Gerätetasten und Display
1 POWER/BACKLIGHT
(Einschalten/Displaybeleuchtung)
2 ZERO/SCROLL (BLAUE Taste)
(Nulleinstellung/Blättern)
3MENU
(Auswahl des Betriebsmodus)
4 Numerische Anzeige 9 Batteriezustandsanzeige 5 Messbereichsanzeige (1) 10 READ/ENTER (GRÜNE Taste)
6 Messbereichsanzeige (2)
7 Menü-Anzeige
8 Anzeige „Kalibrierung anpassen“
(Mess- und Eingabe-Taste)
3.1.1 Tasten-Funktionen
ZERO/SCROLL (BLAUE Taste) : Im Messmodus setzt diese Taste das Gerät auf
Null zurück. Im Menümodus dient sie zum Durchblättern der Menüfunktionen. Bei der Eingabe von Zahlenwerten lässt sich damit die Zahl um jeweils eins erhöhen.
READ/ENTER (GRÜNE Taste) : Im Messmodus wird mit dieser Taste eine Messung ausgelöst. Im Menümodus dient sie zur Auswahl einer Me nüfunktion. Bei der Eingabe von Zahlenwerten setzt sie den Cursor um ein Feld nach rechts.
29
Inbetriebnahme
3.2 Eingabe der Kalibrierdaten für HACH-LANGE Küvetten-Tests
3.2.1 Überblick
Jedes POCKET COLORIMETER II hat zwei Kanäle mit denen Messungen durchgeführt werden können. Jeder Kanal (Messbereich) kann mit benutzerdefinierten Kalibierdaten programmiert werden.
Diese Programmieranleitung beinhaltet die Kalibrierdaten fü r HACH-LANGE Küvetten-Tests zur Bestimmung von Ammonium/Ammonium-Stickstoff:
LCK 303: 2-47 mg/l NH
LCK 304: 0.015-2 mg/l NH4-N oder 0.02-2.5 mg/l NH4
LCK 305: 1-12 mg/l NH4-N oder 1.2-15 mg/l NH
Maximal zwei Kalibrierdatensätze mit den passenden Messbereichsgrenzen auswählen.
Die Programmierung muss nur einmal vor der ersten Anwendung durchgeführt werden.
Für genaue Informationen die Bedienungsanleitung des POCKET COLORIMETER II lesen.
3.2.2 Datentabelle
3.2.2.1 Auswertung als Ammonium-Stickstoff NH4-N
Tabelle 1 Auswertung als Ammonium-Stickstoff NH
-N oder 2.5-60 mg/l NH
4
4
4
-N
4
LCK 303 LCK 304 LCK 305 RES (Auflösung) S0 (Standard 0) A0 (Extinktion 0) S1 (Standard 1) A1 (Extinktion 1)
30
--.-- -.--- --.--
02.00 0.015 00.99
0.113 0.044 0.203
47.00 2.000 12.00
1.862 1.832 1.871