Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol on the instrument is referenced in the
manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European
electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to
the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life
equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Specifications
Note: Specifications are subject to change without notice.
SpecificationsDetails
DOC022.52
.80029
Probe typeDigital combination probe with a non-refillable reference junction and a built-
in temperature sensor
Range0.018 mg/L (10-6 M) to 9,000 mg/L (0.5 M) as Ammonium
Sample pH rangepH 0 to 8.5
Linear region0.9 mg/L to 9,000 mg/L as Ammonium
Slope-58 mV/decade (90 to 110% at 25 °C (77 °F) in linear range per Nernstian
theoretical value)
Operating temperature range0 to 50 °C (32 to 122 °F)
Storage temperature range5 to 35 °C (41 to 95 °F)
JunctionDouble junction (ceramic porous pin and annular porous Teflon®)
Reference typeAg/AgCl
Fill solutiondriTEK Gel (non-refillable)
Response time in linear region< 60 seconds (application dependent)
Minimum sample volume25 mL
Minimum immersion depth25.4 mm (1 in.)
DimensionsDiameter: 12 mm (0.47 in.)
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Cable connectionM12 digital output and connector compatible with HQd meters
Product overview
The ISENH418101 or ISENH418103
built-in temperature sensor (Figure 1). The probe is available with a 1 or 3 m (3.28 or
9.84 ft) cable and is intended for laboratory use. The probe measures ammonium
concentration in water samples.
probe is a combination ammonium probe with a
1
Page 2
Figure 1 Probe overview
1 Sensor protection cap3 Sensing element
2 Reference junction4 1 or 3 m (3.28 or 9.84 ft) cable
Preparation for use
Prepare the probe for use before calibration or sample measurement.
1. Remove the sensor protection cap from the probe.
2. Rinse the probe with deionized water. Blot dry with a lint-free cloth. Do not touch the
tip of the probe.
Note: Prior to use, the probe must be conditioned for at least five minutes in 25 mL of the
lowest concentration standard solution in the calibration set.
If probe stabilization is slow after storage, condition the probe for up to one hour in 25 mL of
the lowest concentration standard solution in the calibration set.
Calibration
Before calibration:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
It is not necessary to recalibrate when moving a calibrated probe from one HQd meter to another if the additional meter is
configured to use the same calibration options.
Default calibration standard set for ISENH4181 probe requires 10 and 100 mg/L Ammonium standard solutions. A new
method can be made if custom calibration or measurement settings are needed. Refer to Advanced operation on page 7
for a list of additional calibration sets.
To view the current calibration, push , select View Probe Data, then select View Current Calibration.
If any two probes are connected, push the UP or DOWN arrow to change to the single display mode in order to show the
Calibrate option.
Prepare the probe for use (refer to Preparation for use on page 2
Calibration notes:
Stir the standards and samples at a slow and steady rate to prevent the formation of
•
a vortex.
•Additional standard sets along with the minimum number of calibration points can be
selected on the Calibration Options menu.
•Push Skip to omit a standard from the calibration routine. The display will not show
Skip until the minimum number of standards is met.
•Begin with the lowest concentration during calibration. This reduces carry-over
contamination to give the best results.
•Note the temperatures of the standards during calibration. Keep temperatures
between calibration standards within ±2 ºC for optimal results.
).
2
Page 3
•The calibration is recorded in the electrode and the data log. The calibration is also
sent to a PC, printer or flash memory stick if connected.
Air bubbles under the sensor tip when submerged can cause slow response or error
•
in measurement. If bubbles are present, gently shake the probe until bubbles are
removed.
•If a calibration error occurs, refer to Troubleshooting on page 10.
Calibration procedure:
1. Connect the
probe to the
meter. Make sure
that the cable
locking nut is
securely
connected to the
meter. Turn the
meter on.
6. Add a stir bar
and put the probe
in the first
standard solution
in the set. Do not
put the probe on
the bottom or
sides of the
container.
2. In three
separate beakers
or appropriate
containers,
prepare
Ammonium
standard solutions
(minimum 25 mL
volume).
7. Put the beaker
on an
electromagnetic
stirrer and stir at a
moderate rate.
Check for air
bubbles and
remove them if
necessary.
3. Add the
contents of one
Ammonium ionic
strength
adjustment (ISA)
liquid pillow per
25 mL to each
standard.
8. Push Read.
The display will
highlight the
standard value
and proceed to
the next standard
value. The display
will show
"Stabilizing" and a
progress bar as
the reading
stabilizes. The
display shows the
standard value
when the reading
is stable.
4. Push
Calibrate. The
display shows the
current standard
value that is to be
read from the
standard solution
set.
9. Repeat steps
5-8 for the other
Ammonium
standard solutions
in the set.
5. Rinse the
probe with
deionized water.
Blot dry with a lintfree cloth. Do not
touch the tip of the
probe.
10. Push Done to
view the
calibration
summary. The
display will not
show Done until
the minimum
number of
calibration points
have been
collected.
3
Page 4
11. Push Store to
accept the
calibration and
return to the
measurement
mode.
Measurement—direct method
Before measurement:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
If complete traceability is required, enter a sample ID and operator ID before measurement. Refer to the HQd meter manual
for more information.
Regular calibration is required for the best measurement accuracy (refer to Calibration on page 2).
Prepare the probe for use (refer to Preparation for use on page 2).
Measurement notes:
•
Stir the standards and samples at a slow and steady rate to prevent the formation of
a vortex.
•Stabilization times with smaller concentration changes generally will be longer and
can be minimized by proper stirring and conditioning. Experiment to determine the
proper stir rate if necessary.
•The integrated temperature sensor and HQd meter software do not compensate for
differences in temperature between calibration standards and samples. Measurement
stabilization is not dependent on temperature stabilization. Temperatures of
calibration standards and samples should be kept within ±2 °C of each other for
optimal results.
•Data is automatically stored in the data log when Press to Read or Interval is
selected in the Measurement Mode. When Continuous is selected, data will only be
stored when Store is selected.
•Between measurements, rinse the probe with deionized water. Blot dry with a lint-free
cloth. Do not touch the tip of the probe.
•Air bubbles under the sensor tip when submerged can cause slow response or error
in measurement. If bubbles are present, gently shake the probe until bubbles are
removed.
•If a measurement error occurs, refer to Troubleshooting on page 10.
Measurement procedure:
4
Page 5
1. Connect the
probe to the
meter. Make sure
that the cable
locking nut is
securely
connected to the
meter. Turn the
meter on.
2. Prepare a
minimum of 25 mL
of the sample(s) in
beakers or
appropriate
containers. Add
the contents of
one Ammonium
ionic strength
adjustment (ISA)
liquid pillow per
25 mL to each
sample.
3. Rinse the
probe with
deionized water.
Blot dry with a lintfree cloth. Do not
touch the tip of the
probe.
4. Add a stir bar
and put the probe
in the sample. Do
not put the probe
on the bottom or
sides of the
container.
5. Put the beaker
on an
electromagnetic
stirrer and stir at a
moderate rate.
Check for air
bubbles and
remove them if
necessary.
6. Push Read.
The display will
show "Stabilizing"
and a progress
bar as the probe
stabilizes in the
sample. The
display will show
the lock icon when
the reading
stabilizes.
7. Repeat steps 2
- 6 for additional
measurements.
Low-level measurements
Use the following techniques for measurements at low concentrations
•Clean the probe regularly as specified in Maintenance on page 10.
•Soak the probe in the lowest concentration standard solution for up to 1 hour before
calibration and measurement.
•Set the stability criteria to a low value (refer to Change measurement options
on page 7).
•Stir the standards and samples at a slow and steady rate to prevent the formation of
a vortex.
•Use a dilute ionic strength adjustor (ISA) solution for calibration and measurements:
8. When
measurements
are done, store
the probe (refer to
Storage
on page
10).
(<1 mg/L NH
4
+
–N).
1. Dissolve the contents of one ionic strength adjustor powder pillow in 50 mL of
deionized water.
2. Add 5 mL of this solution to every 25 mL of standard or sample that is used.
Note: The ionic strength adjuster can be omitted only when all of the following conditions are
true:
•The sample does not contain interferences.
5
Page 6
Interferences
•The sample pH is in the range that is given in the specifications.
Omission of the ISA is accepted by the regulatory reporting agency (if the measurement is
•
for regulatory reporting).
The sensing element responds to ammonium as well as other ions. Typically, probe
response to another ion increases the potential, and causes a positive error. The
response to other ions can be semi-quantitatively determined through the Nikolsky
equation, an extended Nernst equation:
E = Eº + (RT/(zF))ln[aNa + KNax × ax]
Where
•ax—the activity of the interfering ion
•KNax—the selectivity coefficient for the interfering ion relative to ammonium
For the ammonium ISE, the major interferences include perchlorate (ClO
-
(ClO
) and iodide (I-). These ions must be avoided. Other interferences are listed in
3
-
), chlorate
4
Table 1. To eliminate most of these interferences, add the Ammonium ISA to adjust the
sample pH.
The selectivity coefficient is the approximate apparent increase in the measured
concentration caused by one unit of the interfering ion (e.g., 1 unit of K+ raises the
ammonium concentration by 0.1). The approximate selectivity coefficients for some ions
with the IntelliCAL® Ammonium ISE are shown in Table 1.
Table 1 Interferences
InterferenceSelectivity coefficient
Potassium (K+)0.1 (interferes more)
Sodium (Na+)0.002
Magnesium (Mg2+)0.00002
Lithium (Li+)0.00003
Calcium (Ca2+)0.00006 (interferes less)
Run a check standard
The run check standard feature validates instrument performance between sample
measurements. Use the run check standard feature for periodic or user-defined interval
measurements of a traceable standard solution. Set the criteria for check standards from
the ISENH4181
Note: Access control must be off or a valid password must be entered before any of the check
standard method options can be changed.
1. Push
2. Select Run Check Standard.
Note: Select the correct probe if two probes are connected to the meter.
3. Prepare the standard solution shown on the display. Add one powder pillow per
25 mL of standard solution.
4. Put the probe in the standard solution and push Read. The display will show
"Stabilizing" and a progress bar as the reading stabilizes. The display shows the
value of the check standard and either Check Standard Passed or Check Standard
Failed.
5. If the display shows Check Standard Passed, the check standard measurement is
within the accepted limits set by the administrative user. Select Done to continue with
the sample measurement.
6. If the display shows Check Standard Failed, the measurement is outside of
accepted limits set by the administrative user and a recalibration is recommended. If
Settings menu.
. The Full Access Options menu is shown.
6
Page 7
the acceptance criteria is set to Cal Expires on Failure: Yes, the display shows the
calibration icon and a question mark until the probe is recalibrated. To correct the
probe calibration and status indicator, calibrate the probe (refer to Calibration
on page
2).
Advanced operation
Parameter-specific settings can be changed through the Full Access Options menu.
Details about menu navigation, available options and how to change them are given in
the screens, tables and procedures throughout this section.
The settings that can be changed are shown in Table 2.
Table 2 Parameter-specific settings
SettingOptions
Measurement Options
Calibration Options
Check Standard Options
Change measurement options
Methods are groups of factory-set or user-defined settings relevant to specific
applications. If the meter is set to a factory-set method and the Modify Current Settings
option is chosen, a prompt for a new name is shown after the changes are entered. The
settings are saved with this name to distinguish them from the factory-set methods, which
cannot be changed. A saved method can be used instead of multiple adjustments to the
individual settings. Changes made to a user-defined method are automatically saved with
the existing name. Multiple methods can be saved for the same probe on each meter.
•Units
•Significant digits
•Auto stabilization
•Stability criteria
•Upper and lower range limits
•Standard set
•Calibration units
•Minimum calibration points
•Slope limit
•Calibration reminder
•Standard
•Check standard reminder
•Acceptance criteria
1. Make sure a probe is connected to the meter.
2. Push and select ISENH4181 Settings.
3. Select Modify Current Settings.
4. Select Measurement Options and update the settings:
OptionDescription
Chemical FormSets the concentration value—NH
4
+
, NH
+
-NH3 or NH
4
+
-N.
4
7
Page 8
OptionDescription
UnitsSets the preferred unit for ISE measurements—mg/L (default), µg/L, g/L,
g/kg, mol/L, mmol/L, mol/kg, %, ppm or ppb.
Note: The mV units are shown when the detailed display is selected.
Significant Digits Sets the significant digits shown—2, 3 (default) or 4.
Auto Stabilization Sets auto stabilization—on (default) or off.
The default stability drift rate is 1.0 mV/min.
Stability CriteriaWhen Auto Stabilization is off, sets the stability criteria—0.1 to
9.9 mV/min.
Lower stability criteria will require longer stabilization times, but the
•
measurement will be more precise.
•Higher stability criteria will require shorter stabilization times, but the
measurements may be less precise.
Measurement
Limits
5. If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.
6. Push EXIT until the meter returns to the measurement mode.
Change calibration options
1. Make sure a probe is connected to the meter.
2. Push
3. Select Modify Current Settings.
4. Select Calibration Options and update the settings:
OptionDescription
Std SetSets the temperature compensated standard sets that are used for
Sets the measurement limits—Lower limit (default: 0.018 mg/L) or Upper
limit (default: 9,000 mg/L).
The measurement limits can be set to match the acceptable values for
the sample. When the measurement is above the upper limit setting or
below the lower limit setting, the meter shows an "Out of limits"
message. This message is an alert to a potential problem with the
process conditions.
and select ISENH4181 Settings.
calibration—
1, 10 or 100 mg/L as NH
•
•10 or 100 mg/L as NH
•0.1, 1 or 10 mg/L as NH
•0.1, 1, 10, 100 or 1000 mg/L as NH
Standard set values are shown on the Calibration Options screen.
Custom standard sets are characterized at 25 °C (77 °F). Custom
standard values are not temperature compensated. Select the Custom
buffer to make a custom standard. Up to five standard values can be
made .
Note: Only the minimum calibration points must be measured for Done to
be shown on the calibration screen.
+
–N
4
+
–N
4
+
–N
4
+
–N
4
Chemical Form Sets the chemical form.
Calibration
Units
Sets the preferred unit for ISE Calibration—mg/L (default), µg/L (available
only for custom calibration set), g/L, g/kg, mol/L, mmol/L, mol/kg, %, ppm
or ppb.
8
Page 9
OptionDescription
Std Set ValuesWhen Std Set is set to Custom, sets the standard set values.
Up to five standard values can be made. Each standard value can include
a standard set value, Custom or No Standard.
Minimum Cal
Points
Slope LimitSets the slope limit—1 to 30% (acceptable slope criteria, default = 15%).
Sets the minimum number of calibration points necessary before a
calibration can be completed—2 or 3.
The slope must fall within set limits for successful calibration.
5. Select Calibration Reminder and update the settings:
OptionDescription
Reminder
Repeat
ExpiresCalibration expires after the selected time—Immediately, Reminder + 30 min
Meter will make an audible sound when a calibration is due and repeat the
sound at the selected interval—Off (default), 2 h, 4 h, 8 h, 2 d, 5 d or 7 d.
(default), Reminder + 1 h, Reminder + 2 h or Continue Reading.
Note: The meter cannot be used to read samples after calibration has
expired unless Continue Reading is selected.
6. If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.
7. Push EXIT until the meter returns to the measurement mode.
Change check standard options
1. Make sure a probe is connected to the meter.
2. Push and select ISENH4181 Settings.
3. Select Modify Current Settings.
4. Select Check Standards Options and update the settings:
OptionDescription
StandardSets the check standard—10, 100, 1000 mg/L or Custom.
Standard Units When Standard is set to Custom, sets the preferred unit for ISE check
Standard Value When Standard is set to Custom, enter the standard value using the
5. Select Check Standard Reminder and update the settings:
OptionDescription
ReminderSets the check standard reminder—On or Off (default).
Allow Defer Allows the postponement of check standard reminders—Yes or No.
The standard value is shown on the Check Standards Options screen.
The meter automatically shows the check standard screen if Reminder is On.
Measurement of the check standard can be delayed if Allow Defer is set to Yes.
6. Select Acceptance Criteria and update the settings:
OptionDescription
Acceptance LimitsSets the tolerance limits for check standard—1% to 20%.
Cal Expires on Failure Recalibration required if check standard fails—Yes or No.
The calibration expires if the check standard fails and Cal Expires is
set to Yes.
9
Page 10
Maintenance
Clean the probe
Storage
7. If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.
8. Push EXIT until the meter returns to the measurement mode.
Clean the probe when:
Drifting/inaccurate readings occur as a result of contamination on the sensing
•
element or improper storage conditions.
•Slow response time occurs as a result of contamination on the sensing element.
•The slope is out of range as a result of contamination on the sensing element.
For general contaminants, complete the following steps.
1. Rinse the probe with deionized water. Blot dry with a lint-free cloth. Do not touch the
tip of the probe.
2. If harsh contaminants are attached to the probe, polish the probe tip with a soft cloth
or cotton swab to remove the contaminants.
3. Soak for 30 minutes in 100 mg/L Ammonium standard solution.
The probe can be stored dry. To protect the sensing element, rinse with DI water and blot
dry with a lint-free cloth. Install the sensor protection cap.
Note: The probe must be conditioned after long-term storage. Refer to Preparation for use
on page 2.
Troubleshooting
Message or symptomPossible causeAction
Software not updatedTo download the most current version of the
software, refer to the applicable product page on the
manufacturer's website.
Probe not supported
HQd meter does not support
IntelliCAL® probe
Probe not connected properlyDisconnect, then connect the probe. Tighten the
Software not updatedTo download the most current version of the
Connect a probe or probe
requires service
Large number of methods stored on
probe.
Refer to the HQd Series meter manual for specific
instructions for the meter model.
Contact a Technical Support Representative.
locking nut.
software, refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual.
Continue to let probe connect. Do not disconnect
probe.
mV reading is the same for all
solutions
10
Damaged probeMake sure connectivity with another probe or meter
to confirm isolated issue with probe. Contact a
Technical Support Representative.
Electrical issueContact a Technical Support Representative.
Page 11
Message or symptomPossible causeAction
Dirty sensing elementClean the probe (refer to Clean the probe
on page
10).
Slow response time
Slope out of range (refer to
Check probe response
on page 12)
Low sample temperature or
temperature difference between
samples
Bubbles trapped under sensor tipGently shake the probe until bubbles are removed
pH is incorrectVerify the pH is < 8.5 after each ISA addition.
Ionic strength adjustor (ISA) not
used
Insufficient conditioningCondition for at least 5 minutes in a 100 mg/L
Damaged probeContact a Technical Support Representative.
Incorrect standardsCalibrate using freshly prepared standards.
Dirty sensing elementClean the probe and recalibrate.
Bubbles trapped under sensor tipGently shake the probe until bubbles are removed
Dirty sensing elementClean the probe (refer to Clean the probe
Clogged referenceRinse reference junction with deionized water
Check the sample temperature. The lower the
temperature or the greater the difference of
temperatures between samples, traditionally the
longer the response time.
from under sensor tip.
Add ISA to each sample and standard (one powder
pillow per 25 mL of solution).
standard solution with ISA.
from under sensor tip.
on page 10).
thoroughly and shake the probe downward to
remove any air bubbles. The air bubbles will not be
visible.
Improper storage conditionsClean or condition the probe and attempt another
Drifting/inaccurate readings
Out of rangeMeasurement value is outside of
Stabilization criteria not optimized
for the application
Magnetic stirrers may generate
sufficient heat to change solution
temperature.
Damaged probeContact a Technical Support Representative.
Electromagnetic Forces (EMF) from
items such as voltaic cells,
thermoelectric devices, electrical
generators, resistors and
transformers
Bubbles trapped under sensor tipGently shake the probe until bubbles are removed
range
calibration. To re-condition the probe and reference
junctions, allow the probe to soak in a conditioning
solution (that contains ISA) for at least 30 minutes
prior to use.
Adjust the stabilization criteria in the measurement
options menu.
Put a piece of insulating material between the stirrer
and beaker.
Do not use in areas where EMF is present.
from under sensor tip.
Make sure that the sample is within the range of the
probe.
11
Page 12
Message or symptomPossible causeAction
Out of limits
Temperature out of range
Below detection limit
Check standard value is outside of
limits set in the current method
Measurement value is outside of
measurement limits set in the
current method.
Calibration temperature value is
outside of range
Measured temperature is outside
the range of the electrode.
Check standard temperature value
is outside of range
Measurement is not quantifiable
with current saved calibration
(based on IUPAC-defined practical
detection limit).
Make sure that the standard is within the limits of the
current method.
Make another method that expands the acceptable
limits.
Make sure that the sample is within the limits of the
current method.
Make a new method with an expanded range.
Make sure that the sample temperature is within the
range of the probe.
Make sure that the temperature sensor is working
correctly.
Make sure that the standard temperature is within
the range of the probe.
Make sure that the temperature sensor is working
correctly.
Make sure that the check standard temperature is
within the range of the probe.
Perform a new calibration. Check that sample
concentration is bracketed between two standard
solution values (if within linear range).
Re-run calibration and measurement, optimizing
meter settings for slope acceptance and stabilization
criteria for expected sample concentration.
Re-run calibration and sample measurement with
the tips for low-level measurement.
Measurement value is outside of
range.
Make sure sample is within the range of the probe.
Check probe response
To make sure there is a probe response, measure the probe potential (in mV) of two
Ammonium Standard Solutions that are above and below the expected sample
concentration. For example, use 10 and 100 mg/L Ammonium Standard Solutions. The
two solutions should have potentials (difference in mV readings) that are 57 mV apart at
25 ºC (within the slope limits of the method is acceptable). Both solutions should be
above 1.8 mg/L Ammonium.
Check accuracy of sample reading
To make sure the sample measurement is accurate, add a spike of Ammonium Standard
Solution with the volumetric pipet. Refer to Table 3 and formulas to calculate the percent
of recovery.
Typically a percent of recovery of 100% ±5% is a good indication that the instrument,
technique and the sample do not contribute to measurement errors.
Table 3 Spike reference
Measured sample concentrationVolume of standard at addConcentration of standard
1 to 2 mg/L0.5 mL100 mg/L
3 to 6 mg/L1.0 mL100 mg/L
12
7 to 15 mg/L0.3 mL1000 mg/L
15 to 30 mg/L0.5 mL1000 mg/L
30 to 60 mg/L1.0 mL1000 mg/L
Page 13
Percent recovery
Use the following formula to calculate the percent recovery when the sample volume is
25 mL:
E = (C x V1) / V
2
R = (A / (E + S)) x 100
S = mg/L of Ammonium in sample (before spike)
•
•C = concentration of standard used for spiking (mg/L)
•V1 = spike volume (mL)
•V2 = spike volume (mL) + 25 mL sample volume
•E = expected concentration of spike (mg/L)
•R = percent recovery
•A = actual reading on meter after spike (mg/L Ammonium)