Hach-Lange ISENH318103 User Manual

User Manual
Ammonia Probe: Model ISENH318101 or ISENH318103

Safety information

Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.

Specifications

Note: Specifications are subject to change without notice.
Specifications Details
DOC022.52
.80107
Probe type Digital combination gas-sensing probe with a refillable outer body, double-
junction reference and a built-in temperature sensor
Range 0.01 mg/L (5x10-7 M) to 14,000 mg/L (1 M) as NH3-N
Sample pH range > pH 11 per Ammonia ISA
Linear region 0.5 mg/L to 14,000 mg/L as NH3-N
Slope 57 mV/decade (90 to 110% in linear range at 25 °C (77 °F) per Nernstian
theoretical value)
Operating temperature range 5 to 50 °C (41 to 122 °F)
Storage temperature range 5 to 35 °C (41 to 95 °F)
Junction Double junction (annular)
Reference type Ag/AgCl
Fill solution 3 M KCl gel (non-refillable), 0.1 M NH4Cl (outer body, refillable)
Membrane Replaceable Hach ISENH3181 Ammonia membrane module
Response time in linear region < 60 seconds (application dependent)
Minimum sample volume 15 mL
Minimum immersion depth 25.4 mm (1 in.)
Dimensions Diameter: 12 mm (0.47 in.)
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Cable connection M12 digital output and connector compatible with HQd meters

Product overview

The ISENH318101 or ISENH318103 probe is a digital combination gas-sensing electrode with a refillable outer body, double-junction reference and built-in temperature sensor (Figure 1). The probe is available with a 1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory use. The probe measures ammonia concentration in water samples.
1
Figure 1 Probe overview
1 Soaker bottle 5 1 or 3 m (3.28 or 9.84 ft) cable
2 Membrane module protector cap (3x) 6 Fill solution bottle
3 Membrane module (3x) 7 Soaker bottle lid
4 Probe body 8 Glass bulb with integrated temperature sensor

Preparation for use

Prepare the probe for use before calibration or sample measurement.
1. Twist and remove the soaker bottle from the lid to release the pressure.
2. Remove the soaker bottle lid from the probe.
3. Rinse the probe with deionized water. Blot dry with a lint-free cloth. Do not touch the
tip of the probe.
4. Get a single membrane module from the shipping package. Do not touch the membrane surface.
5. Add 12 drops (0.5 mL) of the Ammonia probe filling solution in the membrane module.
6. Install the membrane module on the probe and tighten. Do not spill the filling solution.

Calibration

Before calibration:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
It is not necessary to recalibrate when moving a calibrated probe from one HQd meter to another if the additional meter is configured to use the same calibration options.
Default calibration standard set for ISENH3181 probe requires 1, 10 and 100 mg/L Ammonia standard solutions. A new method can be made if custom calibration or measurement settings are needed. Refer to Advanced operation on page 7 for a list of additional calibration sets.
To view the current calibration, push , select View Probe Data, then select View Current Calibration.
If any two probes are connected, push the UP or DOWN arrow to change to the single display mode in order to show the Calibrate option.
Make sure the Ammonia membrane module is assembled with the correct amount of Ammonia fill solution.
Prepare the probe for use (refer to Preparation for use on page 2
).
2
Calibration notes:
Stir the standards and samples at a slow and steady rate to prevent the formation of
• a vortex.
Additional standard sets along with the minimum number of calibration points can be selected on the Calibration Options menu.
Push Skip to omit a standard from the calibration routine. The display will not show Skip until the minimum number of standards is met.
Begin with the lowest concentration during calibration. This reduces carry-over contamination to give the best results.
Note the temperatures of the standards during calibration. Keep temperatures between calibration standards within ±2 ºC for optimal results.
The calibration is recorded in the electrode and the data log. The calibration is also sent to a PC, printer or flash memory stick if connected.
Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed.
If a calibration error occurs, refer to Troubleshooting on page 11.
Calibration procedure:
1. Connect the
probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn the meter on.
2. In three separate beakers or appropriate containers, prepare Ammonia standard solutions (minimum 25 mL volume).
3. Add the contents of one Ammonia ionic strength adjustment (ISA) powder pillow per 25 mL to each standard.
4. Push Calibrate. The
display shows the current standard value that is to be read from the standard solution set.
5. Rinse the probe with deionized water. Blot dry with a lint­free cloth. Do not touch the tip of the probe.
3
6. Add a stir bar
and put the probe in the first standard solution in the set. Do not put the probe on the bottom or sides of the container.
11. Push Store to accept the calibration and return to the measurement mode.
7. Put the beaker on an electromagnetic stirrer and stir at a moderate rate. Check for air bubbles and remove them if necessary.
8. Push Read. The display will highlight the standard value and proceed to the next standard value. The display will show "Stabilizing" and a progress bar as the reading stabilizes. The display shows the standard value when the reading is stable.
9. Repeat steps 5-8 for the other Ammonia standard solutions in the set.
10. Push Done to view the calibration summary. The display will not show Done until the minimum number of calibration points have been collected.

Measurement—direct method

Before measurement:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
If complete traceability is required, enter a sample ID and operator ID before measurement. Refer to the HQd meter manual for more information.
Make sure the Ammonia membrane module is assembled with the correct amount of Ammonia fill solution.
Regular calibration is required for the best measurement accuracy (refer to Calibration on page 2).
Prepare the probe for use (refer to Preparation for use on page 2
Measurement notes:
Stir the standards and samples at a slow and steady rate to prevent the formation of a vortex.
Stabilization times with smaller concentration changes generally will be longer and can be minimized by proper stirring and conditioning. Experiment to determine the proper stir rate if necessary.
4
).
The integrated temperature sensor and HQd meter software do not compensate for differences in temperature between calibration standards and samples. Measurement stabilization is not dependent on temperature stabilization. Temperatures of calibration standards and samples should be kept within ±2 °C of each other for optimal results.
Data is automatically stored in the data log when Press to Read or Interval is
• selected in the Measurement Mode. When Continuous is selected, data will only be stored when Store is selected.
Between measurements, rinse the probe with deionized water. Blot dry with a lint-free cloth. Do not touch the tip of the probe.
Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed.
If a measurement error occurs, refer to Troubleshooting on page 11.
Measurement procedure:
1. Connect the
probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn the meter on.
6. Push Read. The display will show "Stabilizing" and a progress bar as the probe stabilizes in the sample. The display will show the lock icon when the reading stabilizes.
2. Prepare a minimum of 25 mL of the sample(s) in beakers or appropriate containers. Add the contents of one Ammonia ionic strength adjustment (ISA) powder pillow per 25 mL to each sample.
7. Repeat steps 2 - 6 for additional measurements.
3. Rinse the probe with deionized water. Blot dry with a lint­free cloth. Do not touch the tip of the probe.
8. When measurements are done, store the probe (refer to
Storage
on page
10).
4. Add a stir bar and put the probe in the sample. Do not put the probe on the bottom or sides of the container.
5. Put the beaker on an electromagnetic stirrer and stir at a moderate rate. Check for air bubbles and remove them if necessary.
5
Loading...
+ 9 hidden pages