Hach-Lange ISENa38103 User Manual

Page 1
User Manual
Sodium (Na+) Ion Selective Probe: Model ISENa38101 or ISENa38103

Safety information

Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.

Specifications

Note: Specifications are subject to change without notice.
DOC022.52
.80027
Specifications Details
Probe type Digital, combination ion selective probe with a double junction, refillable
junction reference and a built-in temperature sensor
Electrode resistance < 800 Mohms at 25 °C (77 °F)
Range 0.023 mg/L (1x10-6 M) to 23,000 mg/L (1 M) Na
Linear region 1.4 mg/L to 23,000 mg/L
Slope -59 mV/pNa (90 to 110% at 25 °C (77 °F) per Nernstian theoretical value)
Operating temperature range 0 to 50 °C (32 to 122 °F)
Storage temperature range 5 to 40 °C (41 to 104 °F)
Junction Double ceramic porous pin
Reference type Ag/AgCl
Fill solution 0.02M NH4Cl
Response time in linear region 1 to 2 minutes in sample concentration > 1.4 mg/L
Minimum sample volume 25 mL
Dimensions Diameter: 12 mm (0.47 in.)
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Cable connection M12 digital output and connector compatible with HQd meters
+

Product overview

The ISENa381 series probe is a refillable, combination sodium probe with a built-in temperature sensor (Figure 1). The ISENa38101 or ISENa38103 probe is available with a 1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory use. The probe measures absolute mV values in wastewater, drinking water and general applications. The probe measures sodium concentration in water samples. A 59 mL bottle of 0.02 M NH4Cl Electrode Filling Solution is included with the probe.
1
Page 2
Figure 1 Probe overview
1 Probe soaker bottle 4 Protective tape and filing-hole
2 Reference junction, glass bulb and temperature sensor 5 1 or 3 meter (3.28 or 9.84 ft) cable
3 Filling-hole cap 6 Probe soaker bottle holder

Preparation for use

Prepare the probe for use before calibration or sample measurement.
1. Turn the probe soaker bottle cap counter-clockwise to loosen the cap.
2. Remove the soaker bottle from the probe.
3. Rinse the reference junction and glass bulb with Ionic Strength Adjustor (ISA) rinse
solution (refer to Prepare the ISA rinse solution on page 11 cloth.
4. Remove the protective tape from the filling hole before initial use (refer to Figure 1 on page 2). Dispose of the protective tape.
5. Add filling solution to the probe as necessary (refer to Fill the probe on page 11). The filling solution must be above the standard solution or sample level during measurement (Figure 2).
6. Make sure that the filling hole is open during measurement for the proper flow of the filling solution.
Note: After long-term storage, the probe must be conditioned in 25 mL of 100 mg/L Na standard solution with the contents of one Sodium ISA powder pillow for at least 30 minutes prior to use. If probe stabilization is slow after storage, condition the probe for up to one hour in 25 mL of 100 mg/L Na+ standard solution with the contents of one Sodium ISA pillow.
). Blot dry with a lint-free
+
2
Page 3
Figure 2 Measurement method
1 Container 3 Filling solution level
2 Probe body 4 Standard solution or sample level

Calibration

Before calibration:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
It is not necessary to recalibrate when moving a calibrated probe from one HQd meter to another if the additional meter is configured to use the same calibration options.
To view the current calibration, push , select View Probe Data, then select View Current Calibration.
If any two probes are connected, push the UP or DOWN arrow to change to the single display mode in order to show the Calibrate option.
Prepare the probe for use (refer to Preparation for use on page 2
).
Calibration notes:
Stir the standards and samples at a slow and steady rate to prevent the formation of
• a vortex.
Additional standard sets along with the minimum number of calibration points can be selected on the Calibration Options menu.
Push Skip to omit a standard from the calibration routine. The display will not show Skip until the minimum number of standards is met.
Begin with the lowest concentration during calibration. This reduces carry-over contamination to give the best results.
For measurements below 1 ppm (10-5 M), use plastic lab-ware.
Note the temperatures of the standards during calibration. Keep temperatures between calibration standards within ±2 ºC for optimal results.
The calibration is recorded in the electrode and the data log. The calibration is also sent to a PC, printer or flash memory stick if connected.
Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed.
If a calibration error occurs, refer to Troubleshooting on page 12.
3
Page 4
Calibration procedure:
1. Connect the
probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn the meter on.
6. Add a stir bar and put the probe in the first standard solution in the set. Make sure that the reference junctions are
fully immersed in the solution. Do not put the probe on the bottom or sides of the container. Shake the probe from side to side in the sample to refresh the reference junction.
2. In three separate beakers or appropriate containers, prepare 10, 100 and 1000 mg/L Na
+
standard solution or use appropriate standards that are above and below your application range as the standard set.
7. Put the beaker on an electromagnetic stirrer and stir at a moderate rate. Check for air bubbles and remove them if necessary.
3. Add the contents of one Sodium ionic strength adjustment (ISA) powder pillow per 25 mL to each standard.
8. Push Read. The display will highlight the standard value and proceed to the next standard value. The display will show "Stabilizing" and a progress bar as the reading stabilizes. The display shows the standard value when the reading is stable.
4. Push Calibrate. The
display shows the current standard value that is to be read from the standard solution set.
9. Repeat steps 5-8 for the other Na+) standard solutions in the set.
5. Rinse the probe with ISA rinse solution (refer to Prepare
the ISA rinse solution
on page 11). Never use deionized water.. Blot dry with a lint­free cloth. Do not touch the tip of the probe.
10. Push Done to view the calibration summary. The display will not show Done until the minimum number of calibration points have been collected.
4
Page 5
11. Push Store to
accept the calibration and return to the measurement mode.

Measurement—direct method

Before measurement:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
If complete traceability is required, enter a sample ID and operator ID before measurement. Refer to the HQd meter manual for more information.
Regular calibration is required for the best measurement accuracy (refer to Calibration on page 3).
Prepare the probe for use (refer to Preparation for use on page 2).
Measurement notes:
Stir the standards and samples at a slow and steady rate to prevent the formation of a vortex.
Stabilization times with smaller concentration changes generally will be longer and can be minimized by proper stirring and conditioning. Experiment to determine the proper stir rate if necessary.
Note the temperatures of the samples during measurement. Temperatures of calibration standards and samples should be kept within ±2 °C of each other for optimal results.
Data is automatically stored in the data log when Press to Read or Interval is selected in the Measurement Mode. When Continuous is selected, data will only be stored when Store is selected.
Between measurements, rinse the probe with deionized water. Blot dry with a lint-free cloth. For faster stabilization between measurements, put the probe in 100 mg/L sodium standard (containing Sodium ISA) that is similar in concentration to the samples to be analyzed.
Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed.
If a measurement error occurs, refer to Troubleshooting on page 12.
5
Page 6
Measurement procedure:
1. Connect the
probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn the meter on.
6. Push Read. The display will show "Stabilizing" and a progress bar as the probe stabilizes in the sample. The display will show the lock icon when the reading stabilizes.
2. Prepare a minimum of 25 mL of the sample(s) in beakers or appropriate containers. Add the contents of one Sodium ionic strength adjustment (ISA) powder pillow per 25 mL to each sample.
7. Repeat steps 2 - 6 for additional measurements.
3. Rinse the probe with ISA rinse solution (refer to Prepare
the ISA rinse solution
on page 11). Never use deionized water.. Blot dry with a lint­free cloth. Do not touch the tip of the probe.
8. When measurements are done, store the probe (refer to
Storage
on page
12).
4. Add a stir bar and put the probe in the sample. Make sure that the reference junctions are fully immersed in the solution. Do not put the probe on the bottom or sides of the container. Shake the probe from side to side in the sample to refresh the reference junction.
5. Put the beaker on an electromagnetic stirrer and stir at a moderate rate. Check for air bubbles and remove them if necessary.

Run a check standard

The run check standard feature validates instrument performance between sample measurements. Use the run check standard feature for periodic or user-defined interval measurements of a traceable standard solution. Set the criteria for check standards from the ISENa381 Settings menu.
Note: Access control must be off or a valid password must be entered before any of the check standard method options can be changed.
1. Push
2. Select Run Check Standard.
Note: Select the correct probe if two probes are connected to the meter.
3. Prepare the standard solution shown on the display. Add one powder pillow per 25 mL of standard solution.
6
. The Full Access Options menu is shown.
Page 7

Interferences

4. Put the probe in the standard solution and push Read. The display will show
"Stabilizing" and a progress bar as the reading stabilizes. The display shows the value of the check standard and either Check Standard Passed or Check Standard Failed.
5. If the display shows Check Standard Passed, the check standard measurement is within the accepted limits set by the administrative user. Select Done to continue with the sample measurement.
6. If the display shows Check Standard Failed, the measurement is outside of accepted limits set by the administrative user and a recalibration is recommended. If the acceptance criteria is set to Cal Expires on Failure: Yes, the display shows the calibration icon and a question mark until the probe is recalibrated. To correct the probe calibration and status indicator, calibrate the probe (refer to Calibration on page
3).
The glass membrane responds to sodium as well as other ions. Typically, probe response to another ion increases the potential, and causes a positive error. The response to other ions can be semi-quantitatively determined through the Nikolsky equation, an extended Nernst equation:
E = E° + (RT/(zF)) ln [aNa + KNax × ax]
Where:
ax = the activity of the interfering ion
KNax = the selectivity coefficient for the interfering ion relative to sodium
If the probe is exposed to high levels of interferences, soak the probe in 1 M sodium chloride to help remove the absorbed ions from the glass membrane. The major interferences are silver and hydrogen ions. Hydrogen ion concentration is decreased by the ISA, which raises the pH.
If the samples are highly acidic, or have a high buffer capacity, check that the sample pH is above 9 after adding ISA. If necessary, add ammonium hydroxide (NH4OH) to the calibration standards and samples in equal proportions to raise the pH. The ammonium hydroxide (NH4OH) will not affect the measurement due to the low selectivity coefficient of NH
The ions that interfere in sodium determinations are given for molar concentrations of all ions. The smaller the value of the selectivity coefficient, the lower the interference. Approximate values of selectivity constants (K) are ordered from highest to lowest in
Table 1.
Interference Selectivity coefficient
Ag+ (>1000) H+ (20) - reduced by ISA addition
Li+ (0.01) K+ (0.001)
Ti+ (0.0002)

Advanced operation

Parameter-specific settings can be changed through the Full Access Options menu. Details about menu navigation, available options and how to change them are given in the screens, tables and procedures throughout this section.
4
+
ion.
Table 1 Interferences
7
Page 8
The settings that can be changed are shown in Table 2.
Table 2 Parameter-specific settings
Setting Options
Measurement Options
Calibration Options
Check Standard Options

Change measurement options

Methods are groups of default or user-defined settings relevant to specific applications. If the meter is set to the default method and the Modify Current Settings option is chosen, a prompt for a new name is shown after the changes are entered. The settings are saved with this name to distinguish them from the default method settings, which cannot be changed. A saved method can be used instead of multiple adjustments to the individual settings. Changes made to a user-defined method are automatically saved with the existing name. Multiple methods can be saved for the same probe on each meter.
Units Significant digits
Auto stabilization
Stability criteria
Upper and lower range limits
Standard set
Calibration units
Minimum calibration points
Slope limit
Calibration reminder
Standard
Check standard reminder
Acceptance criteria
1. Make sure a probe is connected to the meter.
2. Push and select ISENa381 Settings.
3. Select Modify Current Settings.
4. Select Measurement Options and update the settings:
Option Description
Units Sets the preferred unit for ISE measurements—mg/L (default), µg/L, g/L,
g/kg, mol/L, mmol/L, mol/kg, %, ppm or ppb.
Note: The mV units are shown when the detailed display is selected.
Significant Digits Sets the significant digits shown—2, 3 (default) or 4.
Auto Stabilization
Sets auto stabilization—on (default) or off.
The default stability drift rate is 1.0 mV/min.
8
Page 9
Option Description
Stability Criteria When Auto Stabilization is off, sets the stability criteria—0.1 to
9.9 mV/min.
Lower stability criteria will require longer stabilization times, but the
• measurement will be more precise.
Higher stability criteria will require shorter stabilization times, but the measurements may be less precise.
Measurement Limits
5. If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method name.
6. Push EXIT until the meter returns to the measurement mode.

Change calibration options

1. Make sure a probe is connected to the meter.
2. Push
3. Select Modify Current Settings.
4. Select Calibration Options and update the settings:
Option Description
Std Set Sets the temperature compensated standard sets that are used for
Sets the measurement limits—Lower limit (default: 0.023 mg/L) or Upper limit (default: 14.00 mg/L).
The measurement limits can be set to match the acceptable values for the sample. When the measurement is above the upper limit setting or below the lower limit setting, the meter shows an "Out of limits" message. This message is an alert to a potential problem with the process conditions.
and select ISENa381 Settings.
calibration—
10, 100 or 1000 mg/L
100 or 1000 mg/L
Standard set values are shown on the Calibration Options screen.
Custom standard sets are characterized at 25 °C (77 °F). Custom standard values are not temperature compensated. Select the Custom buffer to make a custom standard. Up to five standard values can be made (refer to Table 3).
Note: Only the minimum calibration points must be measured for Done to be shown on the calibration screen.
Chemical Form Sets the chemical form.
Calibration Units
Std Set Values When Std Set is set to Custom, sets the standard set values (refer to
Minimum Cal Points
Slope Limit Sets the slope limit—1 to 15% (acceptable slope criteria, default = 10%).
Sets the preferred unit for ISE Calibration—mg/L (default), µg/L (available only for custom calibration set), g/L, g/kg, mol/L, mmol/L, mol/kg, %, ppm or ppb.
Table 3).
Up to five standard values can be made. Each standard value can include a standard set value, Custom or No Standard.
Sets the minimum number of calibration points necessary before a calibration can be completed—2 or 3.
The slope must fall within set limits for successful calibration.
9
Page 10
5. Select Calibration Reminder and update the settings:
Option Description
Reminder Repeat
Expires Calibration expires after the selected time—Immediately, Reminder + 30 min
Meter will make an audible sound when a calibration is due and repeat the sound at the selected interval—Off (default), 2 h, 4 h, 8 h, 2 d, 5 d or 7 d.
(default), Reminder + 1 h, Reminder + 2 h or Continue Reading.
Note: The meter cannot be used to read samples after calibration has expired unless Continue Reading is selected.
6. If prompted, enter a name for the new method settings. Additional changes made to the settings of an existing method are automatically saved with the same method name.
7. Push EXIT until the meter returns to the measurement mode.
Table 3 Custom buffer sets
Buffer set values Option Description
Std1
Std2
Std3
Std4
Std5
10.0 mg/L Pre-set temperature compensated standard values.
100.0 mg/L
1000.0 mg/L
2299.0 mg/L (0.1 M NaCl)
Custom Custom standard value. Custom standard values are not temperature
No standard Standard is undefined when this option is selected.

Change check standard options

1. Make sure a probe is connected to the meter.
2. Push and select ISENa381 Settings.
3. Select Modify Current Settings.
4. Select Check Standards Options and update the settings:
Option Description
Standard Sets the check standard—10, 100, 1000 mg/L, 0.1 M NaCl or Custom.
Standard Units When Standard is set to Custom, sets the preferred unit for ISE check
Standard Value When Standard is set to Custom, enter the standard value using the
compensated.
The standard value is shown on the Check Standards Options screen.
standard—mg/L (default), µg/L, g/L, g/kg, mol/L, mmol/L, mol/kg, %, ppm or ppb.
up/down arrow keys.
10
5. Select Check Standard Reminder and update the settings:
Option Description
Reminder Sets the check standard reminder—On or Off (default).
The meter automatically shows the check standard screen if Reminder is On.
Allow Defer Allows the postponement of check standard reminders—Yes or No.
Measurement of the check standard can be delayed if Allow Defer is set to Yes.
Page 11

Maintenance

Clean the probe

6. Select Acceptance Criteria and update the settings:
Option Description
Acceptance Limits Sets the tolerance limits for check standard—1% to 20%.
Cal Expires on Failure Recalibration required if check standard fails—Yes or No.
The calibration expires if the check standard fails and Cal Expires is set to Yes.
7. If prompted, enter a name for the new method settings. Additional changes made to the settings of an existing method are automatically saved with the same method name.
8. Push EXIT
Clean the probe when:
Drifting/inaccurate readings occur as a result of contamination on the glass sensor or the probe being left dry for extended periods of time.
Slow stabilization time occurs as a result of contamination on the glass sensor.
The slope is out of range as a result of contamination on the glass sensor.
For general contaminants:
until the meter returns to the measurement mode.
1. Rinse the probe with ISA rinse solution and blot dry with a lint-free cloth.
2. Soak the glass bulb for 12 to 16 hours in Hach Electrode Cleaning Solution.
3. Rinse or soak the probe for 1 minute in 25 mL of 100 mg/L sodium standard (that
contains Sodium ISA).
4. Soak the probe in pH 4 buffer for up to 20 minutes, then rinse with deionized water.
5. Calibrate and test the probe again using the measurement procedures.

Fill the probe

To add filling solution to the probe:
1. If the filling hole is closed, remove the filling-hole cap from the filling hole (refer to
Figure 1 on page 2).
2. Remove the cap from the tip of the filling solution bottle.
3. Hold the bottle so that the tip is down. Put the tip of the bottle in the filling hole.
4. Slowly squeeze the bottle and fill the probe completely.
Note: Fill the probe completely for the best performance.
5. If the probe will not be used immediately, store the probe (refer to Storage on page 12).
6. Keep the filling solution bottle and cap for later use.
Note: If the dispensing tip becomes clogged, remove the dispensing tip and soak the tip in warm water. Dry thoroughly and assemble.

Prepare the ISA rinse solution

To prepare the ISA rinse solution, add one Sodium Ionic Strength Adjustor (ISA) powder pillow (0.4 g) to every 25 mL of deionized water. Put the ISA rinse solution in a rinse bottle.

Prepare the storage solution

To prepare the storage solution, add one Sodium Ionic Strength Adjustor (ISA) powder pillow (0.4 g) to 25 mL of 100 mg/L Na+ standard solution.
11
Page 12

Storage

For the best probe performance, do not let the reference junction dry out.
Short-term storage
1. Put the filling-hole cap in the filling hole (Figure 1 on page 2).
2. Store the probe in a 25 mL of 100 mg/L Na+ standard solution with one Sodium Ionic
Strength Adjustor (ISA) powder pillow (0.4 g).
Long-term storage
1. Put the filling-hole cap in the filling hole (refer to Figure 1 on page 2).
2. Rinse the probe with ISA rinse solution (refer to Prepare the ISA rinse solution
on page 11). Never rinse with deionized water. Dry the probe with a lint-free cloth.
3. Fill the probe soaker bottle half full with 0.02 M NH4Cl Electrode Filling Solution..
4. Loosen the soaker bottle cap and put the soaker bottle on the probe.
5. Turn the soaker bottle cap clockwise to tighten the soaker bottle cap.
6. Make sure that the solution in the soaker bottle completely covers the glass bulb and
reference junction holes.
7. Before use after long-term storage, condition the probe in 25 mL of 100 mg/L Na+ standard solution with one Sodium ISA pillow (0.4 g) for at least 8 hours.

Troubleshooting

Message or symptom Possible cause Action
Software not updated To download the most current version of the
software, refer to the applicable product page on the manufacturer's website.
Probe not supported
HQd meter does not support IntelliCAL® probe
Probe not connected properly Disconnect, then connect the probe. Tighten the
Software not updated To download the most current version of the
Connect a probe or probe requires service
mV reading is the same for all solutions
Standard not recognized error Soaker bottle not removed Remove the soaker bottle.
Large number of methods stored on probe
Damaged probe Make sure that there is connectivity with another
Soaker bottle not removed Remove the soaker bottle.
Electrical issue Contact a Technical Support Representative.
Refer to the HQd Series meter manual for specific instructions for the meter model.
Contact a Technical Support Representative.
locking nut.
software, refer to the applicable product page on the manufacturer's website.
Refer to the HQd Series meter manual.
Continue to let probe connect. Do not disconnect probe.
probe or meter to confirm isolated issue with probe. Contact a Technical Support Representative.
12
Incorrect or contaminated standard solution
Contaminated filling solution Replace the filling solution.
Use fresh standard solution as specified in the method.
Page 13
Message or symptom Possible cause Action
Tape not removed from the filling hole Remove the tape.
Contaminated glass sensor Clean the probe (refer to Clean the probe
on page
Contaminated filling solution Drain and replace the filling solution with fresh
solution.
11).
Slow stabilization time
Slope out of range (refer to
Check probe response
on page 14)
Filling-hole cap is closed Open the filling-hole cap while in use.
Low sample temperature or temperature difference between samples
Air bubbles around inner reference electrode
pH is too low Make sure the pH is greater than 9 after each
Ionic strength adjustor (ISA) not used Add ISA to each sample and standard.
Insufficient conditioning Condition for at least 8 hours in storage solution.
Damaged probe Contact a Technical Support Representative.
Incorrect standards Calibrate using freshly prepared standards.
Contaminated glass sensor Clean the probe and recalibrate.
Air bubbles around inner reference electrode
Contaminated glass bulb Clean the probe (refer to Clean the probe
Clogged reference Rinse the reference junction with ISA rinse
Check the sample temperature. The lower the temperature or the greater the difference of temperatures between samples, the longer the response time.
Gently tap the probe with hand or shake the probe downward to remove any air bubbles.
ISA addition.
Gently tap the probe with hand or shake the probe downward to remove any air bubbles.
on page 11).
solution thoroughly and shake the probe downward to remove any air bubbles.
Drifting/inaccurate readings
Improper storage conditions/dehydrated glass bulb
Stabilization criteria not optimized for the application
Magnetic stirrers may generate sufficient heat to change solution temperature.
Filling-hole cap is closed Open the filling-hole cap while in use.
Damaged probe Contact a Technical Support Representative.
Electromagnetic Forces (EMF) such as voltaic cells, thermoelectric devices, electrical generators, resistors and transformers
Colloidal and/or particulates in the filling solution
Air bubbles around inner reference electrode
Clean or condition the probe and attempt another calibration. The probe may not function correctly if the probe has been left dry for extended periods of time. To re-condition the glass sensor and reference junctions, allow the probe to soak in storage solution (that contains Sodium ISA) for at least 8 hours before use.
Adjust the stabilization criteria in the Measurement Options menu.
Put a piece of insulating material between the stirrer and beaker.
Do not use in areas where EMF is present.
Replace the filling solution, calibrate and retest.
Gently tap the probe with hand or shake the probe downward to remove any air bubbles.
13
Page 14
Message or symptom Possible cause Action
Out of range Measurement value is outside of range Make sure that the sample is within the range of
the probe.
Out of limits
Temperature out of range
Check standard value is outside of limits set in the current method
Measurement value is outside of measurement limits set in the current method.
Calibration temperature value is outside of range
Measured temperature is outside the range of the probe
Check standard temperature value is outside of range
Make sure that the standard is within the limits of the current method.
Make another method that expands the acceptable limits.
Make sure that the sample is within the limits of the current method.
Make a new method with an expanded range.
Make sure that the sample temperature is within the range of the probe.
Make sure that the temperature sensor is working correctly.
Make sure that the standard temperature is within the range of the probe.
Make sure that the temperature sensor is working correctly.
Make sure that the check standard temperature is within the range of the probe.
Check probe response
To check the probe response, measure the probe potential (in mV) of two Sodium Standard Solutions one decade apart in concentration that are above and below the expected sample concentration. For example, use 10 and 100 mg/L Sodium Standard Solutions. The two solutions should have potentials (difference in mV readings) that are 59 mV (±3 mV) apart at 25 °C. Both solutions should be above 1.4 mg/L Na+.
Check accuracy of sample reading
To check the accuracy of the sample reading, add a spike of Standard Sodium Solution with the volumetric pipet. Refer to Table 4
and formulas to calculate the percent of
recovery.
Typically a percent of recovery of 100% ±5% is a good indication that the instrument, technique and the sample do not contribute to measurement errors.
Table 4 Spike reference
Measured sample concentration Volume of standard at add Concentration of standard
1 to 2 mg/L 0.5 mL 100 mg/L
3 to 6 mg/L 1.0 mL 100 mg/L
7 to 15 mg/L 0.3 mL 1000 mg/L
15 to 30 mg/L 0.5 mL 1000 mg/L
30 to 60 mg/L 1.0 mL 1000 mg/L
Percent recovery
Use the following formulas to calculate the percent recovery when the sample volume is 25 mL.
E = (C x V1) / V
2
R = (A / (E + S)) x 100
Where:
14
S = mg/L of Na+ in sample (before spike)
C = concentration of standard used for spiking (mg/L)
Page 15
V1 = spike volume (mL) V2 = spike volume (mL) + 25 mL sample volume
E = expected concentration of spike (mg/L)
R = percent recovery
A = actual reading on meter after spike (mg/L Na+)
15
Page 16
HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Tel. (970) 669-3050 (800) 227-4224 (U.S.A. only) Fax (970) 669-2932 orders@hach.com www.hach.com
HACH LANGE GMBH
Willstätterstraße 11 D-40549 Düsseldorf, Germany Tel. +49 (0) 2 11 52 88-320 Fax +49 (0) 2 11 52 88-210 info@hach-lange.de www.hach-lange.de
HACH LANGE Sàrl
6, route de Compois 1222 Vésenaz SWITZERLAND Tel. +41 22 594 6400 Fax +41 22 594 6499
©
Hach Company/Hach Lange GmbH, 2010, 2013. All rights reserved. Printed in U.S.A.
05/2013, Edition 3
Loading...