Sodium (Na+) Ion Selective Probe: Model ISENa38101 or
ISENa38103
Safety information
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol on the instrument is referenced in the
manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European
electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to
the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life
equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Specifications
Note: Specifications are subject to change without notice.
DOC022.52
.80027
SpecificationsDetails
Probe typeDigital, combination ion selective probe with a double junction, refillable
junction reference and a built-in temperature sensor
Electrode resistance< 800 Mohms at 25 °C (77 °F)
Range0.023 mg/L (1x10-6 M) to 23,000 mg/L (1 M) Na
Linear region1.4 mg/L to 23,000 mg/L
Slope-59 mV/pNa (90 to 110% at 25 °C (77 °F) per Nernstian theoretical value)
Operating temperature range0 to 50 °C (32 to 122 °F)
Storage temperature range5 to 40 °C (41 to 104 °F)
JunctionDouble ceramic porous pin
Reference typeAg/AgCl
Fill solution0.02M NH4Cl
Response time in linear region1 to 2 minutes in sample concentration > 1.4 mg/L
Minimum sample volume25 mL
DimensionsDiameter: 12 mm (0.47 in.)
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Cable connectionM12 digital output and connector compatible with HQd meters
+
Product overview
The ISENa381 series probe is a refillable, combination sodium probe with a built-in
temperature sensor (Figure 1). The ISENa38101 or ISENa38103 probe is available with a
1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory use. The probe measures
absolute mV values in wastewater, drinking water and general applications. The probe
measures sodium concentration in water samples. A 59 mL bottle of 0.02 M NH4Cl
Electrode Filling Solution is included with the probe.
1
Page 2
Figure 1 Probe overview
1 Probe soaker bottle4 Protective tape and filing-hole
2 Reference junction, glass bulb and temperature sensor5 1 or 3 meter (3.28 or 9.84 ft) cable
3 Filling-hole cap6 Probe soaker bottle holder
Preparation for use
Prepare the probe for use before calibration or sample measurement.
1. Turn the probe soaker bottle cap counter-clockwise to loosen the cap.
2. Remove the soaker bottle from the probe.
3. Rinse the reference junction and glass bulb with Ionic Strength Adjustor (ISA) rinse
solution (refer to Prepare the ISA rinse solution on page 11
cloth.
4. Remove the protective tape from the filling hole before initial use (refer to Figure 1
on page 2). Dispose of the protective tape.
5. Add filling solution to the probe as necessary (refer to Fill the probe on page 11).
The filling solution must be above the standard solution or sample level during
measurement (Figure 2).
6. Make sure that the filling hole is open during measurement for the proper flow of the
filling solution.
Note: After long-term storage, the probe must be conditioned in 25 mL of 100 mg/L Na
standard solution with the contents of one Sodium ISA powder pillow for at least 30 minutes
prior to use. If probe stabilization is slow after storage, condition the probe for up to one hour in
25 mL of 100 mg/L Na+ standard solution with the contents of one Sodium ISA pillow.
). Blot dry with a lint-free
+
2
Page 3
Figure 2 Measurement method
1 Container3 Filling solution level
2 Probe body4 Standard solution or sample level
Calibration
Before calibration:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
It is not necessary to recalibrate when moving a calibrated probe from one HQd meter to another if the additional meter is
configured to use the same calibration options.
To view the current calibration, push , select View Probe Data, then select View Current Calibration.
If any two probes are connected, push the UP or DOWN arrow to change to the single display mode in order to show the
Calibrate option.
Prepare the probe for use (refer to Preparation for use on page 2
).
Calibration notes:
Stir the standards and samples at a slow and steady rate to prevent the formation of
•
a vortex.
•Additional standard sets along with the minimum number of calibration points can be
selected on the Calibration Options menu.
•Push Skip to omit a standard from the calibration routine. The display will not show
Skip until the minimum number of standards is met.
•Begin with the lowest concentration during calibration. This reduces carry-over
contamination to give the best results.
•For measurements below 1 ppm (10-5 M), use plastic lab-ware.
•Note the temperatures of the standards during calibration. Keep temperatures
between calibration standards within ±2 ºC for optimal results.
•The calibration is recorded in the electrode and the data log. The calibration is also
sent to a PC, printer or flash memory stick if connected.
•Air bubbles under the sensor tip when submerged can cause slow response or error
in measurement. If bubbles are present, gently shake the probe until bubbles are
removed.
•If a calibration error occurs, refer to Troubleshooting on page 12.
3
Page 4
Calibration procedure:
1. Connect the
probe to the
meter. Make sure
that the cable
locking nut is
securely
connected to the
meter. Turn the
meter on.
6. Add a stir bar
and put the probe
in the first
standard solution
in the set. Make
sure that the
reference
junctions are
fully
immersed in the
solution. Do not
put the probe on
the bottom or
sides of the
container. Shake
the probe from
side to side in the
sample to refresh
the reference
junction.
2. In three
separate beakers
or appropriate
containers,
prepare 10,
100 and
1000 mg/L Na
+
standard solution
or use appropriate
standards that are
above and below
your application
range as the
standard set.
7. Put the beaker
on an
electromagnetic
stirrer and stir at a
moderate rate.
Check for air
bubbles and
remove them if
necessary.
3. Add the
contents of one
Sodium ionic
strength
adjustment (ISA)
powder pillow per
25 mL to each
standard.
8. Push Read.
The display will
highlight the
standard value
and proceed to
the next standard
value. The display
will show
"Stabilizing" and a
progress bar as
the reading
stabilizes. The
display shows the
standard value
when the reading
is stable.
4. Push
Calibrate. The
display shows the
current standard
value that is to be
read from the
standard solution
set.
9. Repeat steps
5-8 for the other
Na+) standard
solutions in the
set.
5. Rinse the
probe with ISA
rinse solution
(refer to Prepare
the ISA rinse
solution
on page 11).
Never use
deionized water..
Blot dry with a lintfree cloth. Do not
touch the tip of the
probe.
10. Push Done to
view the
calibration
summary. The
display will not
show Done until
the minimum
number of
calibration points
have been
collected.
4
Page 5
11. Push Store to
accept the
calibration and
return to the
measurement
mode.
Measurement—direct method
Before measurement:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
If complete traceability is required, enter a sample ID and operator ID before measurement. Refer to the HQd meter manual
for more information.
Regular calibration is required for the best measurement accuracy (refer to Calibration on page 3).
Prepare the probe for use (refer to Preparation for use on page 2).
Measurement notes:
•
Stir the standards and samples at a slow and steady rate to prevent the formation of
a vortex.
•Stabilization times with smaller concentration changes generally will be longer and
can be minimized by proper stirring and conditioning. Experiment to determine the
proper stir rate if necessary.
•Note the temperatures of the samples during measurement. Temperatures of
calibration standards and samples should be kept within ±2 °C of each other for
optimal results.
•Data is automatically stored in the data log when Press to Read or Interval is
selected in the Measurement Mode. When Continuous is selected, data will only be
stored when Store is selected.
•Between measurements, rinse the probe with deionized water. Blot dry with a lint-free
cloth. For faster stabilization between measurements, put the probe in 100 mg/L
sodium standard (containing Sodium ISA) that is similar in concentration to the
samples to be analyzed.
•Air bubbles under the sensor tip when submerged can cause slow response or error
in measurement. If bubbles are present, gently shake the probe until bubbles are
removed.
•If a measurement error occurs, refer to Troubleshooting on page 12.
5
Page 6
Measurement procedure:
1. Connect the
probe to the
meter. Make sure
that the cable
locking nut is
securely
connected to the
meter. Turn the
meter on.
6. Push Read.
The display will
show "Stabilizing"
and a progress
bar as the probe
stabilizes in the
sample. The
display will show
the lock icon when
the reading
stabilizes.
2. Prepare a
minimum of 25 mL
of the sample(s) in
beakers or
appropriate
containers. Add
the contents of
one Sodium ionic
strength
adjustment (ISA)
powder pillow per
25 mL to each
sample.
7. Repeat steps
2 - 6 for additional
measurements.
3. Rinse the
probe with ISA
rinse solution
(refer to Prepare
the ISA rinse
solution
on page 11).
Never use
deionized water..
Blot dry with a lintfree cloth. Do not
touch the tip of the
probe.
8. When
measurements
are done, store
the probe (refer to
Storage
on page
12).
4. Add a stir bar
and put the probe
in the sample.
Make sure that the
reference
junctions are fully
immersed in the
solution. Do not
put the probe on
the bottom or
sides of the
container. Shake
the probe from
side to side in the
sample to refresh
the reference
junction.
5. Put the beaker
on an
electromagnetic
stirrer and stir at a
moderate rate.
Check for air
bubbles and
remove them if
necessary.
Run a check standard
The run check standard feature validates instrument performance between sample
measurements. Use the run check standard feature for periodic or user-defined interval
measurements of a traceable standard solution. Set the criteria for check standards from
the ISENa381 Settings menu.
Note: Access control must be off or a valid password must be entered before any of the check
standard method options can be changed.
1. Push
2. Select Run Check Standard.
Note: Select the correct probe if two probes are connected to the meter.
3. Prepare the standard solution shown on the display. Add one powder pillow per
25 mL of standard solution.
6
. The Full Access Options menu is shown.
Page 7
Interferences
4. Put the probe in the standard solution and push Read. The display will show
"Stabilizing" and a progress bar as the reading stabilizes. The display shows the
value of the check standard and either Check Standard Passed or Check Standard
Failed.
5. If the display shows Check Standard Passed, the check standard measurement is
within the accepted limits set by the administrative user. Select Done to continue with
the sample measurement.
6. If the display shows Check Standard Failed, the measurement is outside of
accepted limits set by the administrative user and a recalibration is recommended. If
the acceptance criteria is set to Cal Expires on Failure: Yes, the display shows the
calibration icon and a question mark until the probe is recalibrated. To correct the
probe calibration and status indicator, calibrate the probe (refer to Calibration
on page
3).
The glass membrane responds to sodium as well as other ions. Typically, probe response
to another ion increases the potential, and causes a positive error. The response to other
ions can be semi-quantitatively determined through the Nikolsky equation, an extended
Nernst equation:
E = E° + (RT/(zF)) ln [aNa + KNax × ax]
Where:
•ax = the activity of the interfering ion
•KNax = the selectivity coefficient for the interfering ion relative to sodium
If the probe is exposed to high levels of interferences, soak the probe in 1 M sodium
chloride to help remove the absorbed ions from the glass membrane. The major
interferences are silver and hydrogen ions. Hydrogen ion concentration is decreased by
the ISA, which raises the pH.
If the samples are highly acidic, or have a high buffer capacity, check that the sample pH
is above 9 after adding ISA. If necessary, add ammonium hydroxide (NH4OH) to the
calibration standards and samples in equal proportions to raise the pH. The ammonium
hydroxide (NH4OH) will not affect the measurement due to the low selectivity coefficient
of NH
The ions that interfere in sodium determinations are given for molar concentrations of all
ions. The smaller the value of the selectivity coefficient, the lower the interference.
Approximate values of selectivity constants (K) are ordered from highest to lowest in
Table 1.
InterferenceSelectivity coefficient
Ag+ (>1000)H+ (20) - reduced by ISA addition
Li+ (0.01)K+ (0.001)
Ti+ (0.0002)—
Advanced operation
Parameter-specific settings can be changed through the Full Access Options menu.
Details about menu navigation, available options and how to change them are given in
the screens, tables and procedures throughout this section.
4
+
ion.
Table 1 Interferences
7
Page 8
The settings that can be changed are shown in Table 2.
Table 2 Parameter-specific settings
SettingOptions
Measurement Options
Calibration Options
Check Standard Options
Change measurement options
Methods are groups of default or user-defined settings relevant to specific applications. If
the meter is set to the default method and the Modify Current Settings option is chosen, a
prompt for a new name is shown after the changes are entered. The settings are saved
with this name to distinguish them from the default method settings, which cannot be
changed. A saved method can be used instead of multiple adjustments to the individual
settings. Changes made to a user-defined method are automatically saved with the
existing name. Multiple methods can be saved for the same probe on each meter.
•Units
Significant digits
•
•Auto stabilization
•Stability criteria
•Upper and lower range limits
•Standard set
•Calibration units
•Minimum calibration points
•Slope limit
•Calibration reminder
•Standard
•Check standard reminder
•Acceptance criteria
1. Make sure a probe is connected to the meter.
2. Push and select ISENa381 Settings.
3. Select Modify Current Settings.
4. Select Measurement Options and update the settings:
OptionDescription
UnitsSets the preferred unit for ISE measurements—mg/L (default), µg/L, g/L,
g/kg, mol/L, mmol/L, mol/kg, %, ppm or ppb.
Note: The mV units are shown when the detailed display is selected.
Significant Digits Sets the significant digits shown—2, 3 (default) or 4.
Auto
Stabilization
Sets auto stabilization—on (default) or off.
The default stability drift rate is 1.0 mV/min.
8
Page 9
OptionDescription
Stability CriteriaWhen Auto Stabilization is off, sets the stability criteria—0.1 to
9.9 mV/min.
Lower stability criteria will require longer stabilization times, but the
•
measurement will be more precise.
•Higher stability criteria will require shorter stabilization times, but the
measurements may be less precise.
Measurement
Limits
5. If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.
6. Push EXIT until the meter returns to the measurement mode.
Change calibration options
1. Make sure a probe is connected to the meter.
2. Push
3. Select Modify Current Settings.
4. Select Calibration Options and update the settings:
OptionDescription
Std SetSets the temperature compensated standard sets that are used for
Sets the measurement limits—Lower limit (default: 0.023 mg/L) or Upper
limit (default: 14.00 mg/L).
The measurement limits can be set to match the acceptable values for
the sample. When the measurement is above the upper limit setting or
below the lower limit setting, the meter shows an "Out of limits" message.
This message is an alert to a potential problem with the process
conditions.
and select ISENa381 Settings.
calibration—
10, 100 or 1000 mg/L
•
•100 or 1000 mg/L
Standard set values are shown on the Calibration Options screen.
Custom standard sets are characterized at 25 °C (77 °F). Custom
standard values are not temperature compensated. Select the Custom
buffer to make a custom standard. Up to five standard values can be
made (refer to Table 3).
Note: Only the minimum calibration points must be measured for Done to
be shown on the calibration screen.
Chemical FormSets the chemical form.
Calibration
Units
Std Set ValuesWhen Std Set is set to Custom, sets the standard set values (refer to
Minimum Cal
Points
Slope LimitSets the slope limit—1 to 15% (acceptable slope criteria, default = 10%).
Sets the preferred unit for ISE Calibration—mg/L (default), µg/L (available
only for custom calibration set), g/L, g/kg, mol/L, mmol/L, mol/kg, %, ppm
or ppb.
Table 3).
Up to five standard values can be made. Each standard value can include
a standard set value, Custom or No Standard.
Sets the minimum number of calibration points necessary before a
calibration can be completed—2 or 3.
The slope must fall within set limits for successful calibration.
9
Page 10
5. Select Calibration Reminder and update the settings:
OptionDescription
Reminder
Repeat
ExpiresCalibration expires after the selected time—Immediately, Reminder + 30 min
Meter will make an audible sound when a calibration is due and repeat the
sound at the selected interval—Off (default), 2 h, 4 h, 8 h, 2 d, 5 d or 7 d.
(default), Reminder + 1 h, Reminder + 2 h or Continue Reading.
Note: The meter cannot be used to read samples after calibration has
expired unless Continue Reading is selected.
6. If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.
7. Push EXIT until the meter returns to the measurement mode.
Table 3 Custom buffer sets
Buffer set values OptionDescription
Std1
Std2
Std3
Std4
Std5
10.0 mg/LPre-set temperature compensated standard values.
100.0 mg/L
1000.0 mg/L
2299.0 mg/L (0.1 M NaCl)
CustomCustom standard value. Custom standard values are not temperature
No standardStandard is undefined when this option is selected.
Change check standard options
1. Make sure a probe is connected to the meter.
2. Push and select ISENa381 Settings.
3. Select Modify Current Settings.
4. Select Check Standards Options and update the settings:
OptionDescription
StandardSets the check standard—10, 100, 1000 mg/L, 0.1 M NaCl or Custom.
Standard Units When Standard is set to Custom, sets the preferred unit for ISE check
Standard Value When Standard is set to Custom, enter the standard value using the
compensated.
The standard value is shown on the Check Standards Options screen.
5. Select Check Standard Reminder and update the settings:
OptionDescription
ReminderSets the check standard reminder—On or Off (default).
The meter automatically shows the check standard screen if Reminder is On.
Allow Defer Allows the postponement of check standard reminders—Yes or No.
Measurement of the check standard can be delayed if Allow Defer is set to Yes.
Page 11
Maintenance
Clean the probe
6. Select Acceptance Criteria and update the settings:
OptionDescription
Acceptance LimitsSets the tolerance limits for check standard—1% to 20%.
Cal Expires on Failure Recalibration required if check standard fails—Yes or No.
The calibration expires if the check standard fails and Cal Expires is
set to Yes.
7. If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.
8. Push EXIT
Clean the probe when:
•Drifting/inaccurate readings occur as a result of contamination on the glass sensor or
the probe being left dry for extended periods of time.
•Slow stabilization time occurs as a result of contamination on the glass sensor.
•The slope is out of range as a result of contamination on the glass sensor.
For general contaminants:
until the meter returns to the measurement mode.
1. Rinse the probe with ISA rinse solution and blot dry with a lint-free cloth.
2. Soak the glass bulb for 12 to 16 hours in Hach Electrode Cleaning Solution.
3. Rinse or soak the probe for 1 minute in 25 mL of 100 mg/L sodium standard (that
contains Sodium ISA).
4. Soak the probe in pH 4 buffer for up to 20 minutes, then rinse with deionized water.
5. Calibrate and test the probe again using the measurement procedures.
Fill the probe
To add filling solution to the probe:
1. If the filling hole is closed, remove the filling-hole cap from the filling hole (refer to
Figure 1 on page 2).
2. Remove the cap from the tip of the filling solution bottle.
3. Hold the bottle so that the tip is down. Put the tip of the bottle in the filling hole.
4. Slowly squeeze the bottle and fill the probe completely.
Note: Fill the probe completely for the best performance.
5. If the probe will not be used immediately, store the probe (refer to Storage
on page 12).
6. Keep the filling solution bottle and cap for later use.
Note: If the dispensing tip becomes clogged, remove the dispensing tip and soak the tip in
warm water. Dry thoroughly and assemble.
Prepare the ISA rinse solution
To prepare the ISA rinse solution, add one Sodium Ionic Strength Adjustor (ISA) powder
pillow (0.4 g) to every 25 mL of deionized water. Put the ISA rinse solution in a rinse
bottle.
Prepare the storage solution
To prepare the storage solution, add one Sodium Ionic Strength Adjustor (ISA) powder
pillow (0.4 g) to 25 mL of 100 mg/L Na+ standard solution.
11
Page 12
Storage
For the best probe performance, do not let the reference junction dry out.
Short-term storage
1. Put the filling-hole cap in the filling hole (Figure 1 on page 2).
2. Store the probe in a 25 mL of 100 mg/L Na+ standard solution with one Sodium Ionic
Strength Adjustor (ISA) powder pillow (0.4 g).
Long-term storage
1. Put the filling-hole cap in the filling hole (refer to Figure 1 on page 2).
2. Rinse the probe with ISA rinse solution (refer to Prepare the ISA rinse solution
on page 11). Never rinse with deionized water. Dry the probe with a lint-free cloth.
3. Fill the probe soaker bottle half full with 0.02 M NH4Cl Electrode Filling Solution..
4. Loosen the soaker bottle cap and put the soaker bottle on the probe.
5. Turn the soaker bottle cap clockwise to tighten the soaker bottle cap.
6. Make sure that the solution in the soaker bottle completely covers the glass bulb and
reference junction holes.
7. Before use after long-term storage, condition the probe in 25 mL of 100 mg/L Na+
standard solution with one Sodium ISA pillow (0.4 g) for at least 8 hours.
Troubleshooting
Message or symptomPossible causeAction
Software not updatedTo download the most current version of the
software, refer to the applicable product page on
the manufacturer's website.
Probe not supported
HQd meter does not support
IntelliCAL® probe
Probe not connected properlyDisconnect, then connect the probe. Tighten the
Software not updatedTo download the most current version of the
Connect a probe or probe
requires service
mV reading is the same for all
solutions
Standard not recognized errorSoaker bottle not removedRemove the soaker bottle.
Large number of methods stored on
probe
Damaged probeMake sure that there is connectivity with another
Soaker bottle not removedRemove the soaker bottle.
Electrical issueContact a Technical Support Representative.
Refer to the HQd Series meter manual for
specific instructions for the meter model.
Contact a Technical Support Representative.
locking nut.
software, refer to the applicable product page on
the manufacturer's website.
Refer to the HQd Series meter manual.
Continue to let probe connect. Do not disconnect
probe.
probe or meter to confirm isolated issue with
probe. Contact a Technical Support
Representative.
12
Incorrect or contaminated standard
solution
Contaminated filling solutionReplace the filling solution.
Use fresh standard solution as specified in the
method.
Page 13
Message or symptomPossible causeAction
Tape not removed from the filling holeRemove the tape.
Contaminated glass sensorClean the probe (refer to Clean the probe
on page
Contaminated filling solutionDrain and replace the filling solution with fresh
solution.
11).
Slow stabilization time
Slope out of range (refer to
Check probe response
on page 14)
Filling-hole cap is closedOpen the filling-hole cap while in use.
Low sample temperature or
temperature difference between
samples
Air bubbles around inner reference
electrode
pH is too lowMake sure the pH is greater than 9 after each
Ionic strength adjustor (ISA) not usedAdd ISA to each sample and standard.
Insufficient conditioningCondition for at least 8 hours in storage solution.
Damaged probeContact a Technical Support Representative.
Incorrect standardsCalibrate using freshly prepared standards.
Contaminated glass sensorClean the probe and recalibrate.
Air bubbles around inner reference
electrode
Contaminated glass bulbClean the probe (refer to Clean the probe
Clogged referenceRinse the reference junction with ISA rinse
Check the sample temperature. The lower the
temperature or the greater the difference of
temperatures between samples, the longer the
response time.
Gently tap the probe with hand or shake the
probe downward to remove any air bubbles.
ISA addition.
Gently tap the probe with hand or shake the
probe downward to remove any air bubbles.
on page 11).
solution thoroughly and shake the probe
downward to remove any air bubbles.
Drifting/inaccurate readings
Improper storage
conditions/dehydrated glass bulb
Stabilization criteria not optimized for
the application
Magnetic stirrers may generate
sufficient heat to change solution
temperature.
Filling-hole cap is closedOpen the filling-hole cap while in use.
Damaged probeContact a Technical Support Representative.
Electromagnetic Forces (EMF) such as
voltaic cells, thermoelectric devices,
electrical generators, resistors and
transformers
Colloidal and/or particulates in the
filling solution
Air bubbles around inner reference
electrode
Clean or condition the probe and attempt another
calibration. The probe may not function correctly
if the probe has been left dry for extended
periods of time. To re-condition the glass sensor
and reference junctions, allow the probe to soak
in storage solution (that contains Sodium ISA) for
at least 8 hours before use.
Adjust the stabilization criteria in the
Measurement Options menu.
Put a piece of insulating material between the
stirrer and beaker.
Do not use in areas where EMF is present.
Replace the filling solution, calibrate and retest.
Gently tap the probe with hand or shake the
probe downward to remove any air bubbles.
13
Page 14
Message or symptomPossible causeAction
Out of rangeMeasurement value is outside of range Make sure that the sample is within the range of
the probe.
Out of limits
Temperature out of range
Check standard value is outside of
limits set in the current method
Measurement value is outside of
measurement limits set in the current
method.
Calibration temperature value is
outside of range
Measured temperature is outside the
range of the probe
Check standard temperature value is
outside of range
Make sure that the standard is within the limits of
the current method.
Make another method that expands the
acceptable limits.
Make sure that the sample is within the limits of
the current method.
Make a new method with an expanded range.
Make sure that the sample temperature is within
the range of the probe.
Make sure that the temperature sensor is
working correctly.
Make sure that the standard temperature is
within the range of the probe.
Make sure that the temperature sensor is
working correctly.
Make sure that the check standard temperature
is within the range of the probe.
Check probe response
To check the probe response, measure the probe potential (in mV) of two Sodium
Standard Solutions one decade apart in concentration that are above and below the
expected sample concentration. For example, use 10 and 100 mg/L Sodium Standard
Solutions. The two solutions should have potentials (difference in mV readings) that are
59 mV (±3 mV) apart at 25 °C. Both solutions should be above 1.4 mg/L Na+.
Check accuracy of sample reading
To check the accuracy of the sample reading, add a spike of Standard Sodium Solution
with the volumetric pipet. Refer to Table 4
and formulas to calculate the percent of
recovery.
Typically a percent of recovery of 100% ±5% is a good indication that the instrument,
technique and the sample do not contribute to measurement errors.
Table 4 Spike reference
Measured sample concentrationVolume of standard at addConcentration of standard
1 to 2 mg/L0.5 mL100 mg/L
3 to 6 mg/L1.0 mL100 mg/L
7 to 15 mg/L0.3 mL1000 mg/L
15 to 30 mg/L0.5 mL1000 mg/L
30 to 60 mg/L1.0 mL1000 mg/L
Percent recovery
Use the following formulas to calculate the percent recovery when the sample volume is
25 mL.
E = (C x V1) / V
2
R = (A / (E + S)) x 100
Where:
14
S = mg/L of Na+ in sample (before spike)
•
•C = concentration of standard used for spiking (mg/L)