Sodium (Na+) Ion Selective Probe: Model ISENa38101 or
ISENa38103
Safety information
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol on the instrument is referenced in the
manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European
electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to
the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life
equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Specifications
Note: Specifications are subject to change without notice.
DOC022.52
.80027
SpecificationsDetails
Probe typeDigital, combination ion selective probe with a double junction, refillable
junction reference and a built-in temperature sensor
Electrode resistance< 800 Mohms at 25 °C (77 °F)
Range0.023 mg/L (1x10-6 M) to 23,000 mg/L (1 M) Na
Linear region1.4 mg/L to 23,000 mg/L
Slope-59 mV/pNa (90 to 110% at 25 °C (77 °F) per Nernstian theoretical value)
Operating temperature range0 to 50 °C (32 to 122 °F)
Storage temperature range5 to 40 °C (41 to 104 °F)
JunctionDouble ceramic porous pin
Reference typeAg/AgCl
Fill solution0.02M NH4Cl
Response time in linear region1 to 2 minutes in sample concentration > 1.4 mg/L
Minimum sample volume25 mL
DimensionsDiameter: 12 mm (0.47 in.)
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Cable connectionM12 digital output and connector compatible with HQd meters
+
Product overview
The ISENa381 series probe is a refillable, combination sodium probe with a built-in
temperature sensor (Figure 1). The ISENa38101 or ISENa38103 probe is available with a
1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory use. The probe measures
absolute mV values in wastewater, drinking water and general applications. The probe
measures sodium concentration in water samples. A 59 mL bottle of 0.02 M NH4Cl
Electrode Filling Solution is included with the probe.
1
Figure 1 Probe overview
1 Probe soaker bottle4 Protective tape and filing-hole
2 Reference junction, glass bulb and temperature sensor5 1 or 3 meter (3.28 or 9.84 ft) cable
3 Filling-hole cap6 Probe soaker bottle holder
Preparation for use
Prepare the probe for use before calibration or sample measurement.
1. Turn the probe soaker bottle cap counter-clockwise to loosen the cap.
2. Remove the soaker bottle from the probe.
3. Rinse the reference junction and glass bulb with Ionic Strength Adjustor (ISA) rinse
solution (refer to Prepare the ISA rinse solution on page 11
cloth.
4. Remove the protective tape from the filling hole before initial use (refer to Figure 1
on page 2). Dispose of the protective tape.
5. Add filling solution to the probe as necessary (refer to Fill the probe on page 11).
The filling solution must be above the standard solution or sample level during
measurement (Figure 2).
6. Make sure that the filling hole is open during measurement for the proper flow of the
filling solution.
Note: After long-term storage, the probe must be conditioned in 25 mL of 100 mg/L Na
standard solution with the contents of one Sodium ISA powder pillow for at least 30 minutes
prior to use. If probe stabilization is slow after storage, condition the probe for up to one hour in
25 mL of 100 mg/L Na+ standard solution with the contents of one Sodium ISA pillow.
). Blot dry with a lint-free
+
2
Figure 2 Measurement method
1 Container3 Filling solution level
2 Probe body4 Standard solution or sample level
Calibration
Before calibration:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
It is not necessary to recalibrate when moving a calibrated probe from one HQd meter to another if the additional meter is
configured to use the same calibration options.
To view the current calibration, push , select View Probe Data, then select View Current Calibration.
If any two probes are connected, push the UP or DOWN arrow to change to the single display mode in order to show the
Calibrate option.
Prepare the probe for use (refer to Preparation for use on page 2
).
Calibration notes:
Stir the standards and samples at a slow and steady rate to prevent the formation of
•
a vortex.
•Additional standard sets along with the minimum number of calibration points can be
selected on the Calibration Options menu.
•Push Skip to omit a standard from the calibration routine. The display will not show
Skip until the minimum number of standards is met.
•Begin with the lowest concentration during calibration. This reduces carry-over
contamination to give the best results.
•For measurements below 1 ppm (10-5 M), use plastic lab-ware.
•Note the temperatures of the standards during calibration. Keep temperatures
between calibration standards within ±2 ºC for optimal results.
•The calibration is recorded in the electrode and the data log. The calibration is also
sent to a PC, printer or flash memory stick if connected.
•Air bubbles under the sensor tip when submerged can cause slow response or error
in measurement. If bubbles are present, gently shake the probe until bubbles are
removed.
•If a calibration error occurs, refer to Troubleshooting on page 12.
3
Calibration procedure:
1. Connect the
probe to the
meter. Make sure
that the cable
locking nut is
securely
connected to the
meter. Turn the
meter on.
6. Add a stir bar
and put the probe
in the first
standard solution
in the set. Make
sure that the
reference
junctions are
fully
immersed in the
solution. Do not
put the probe on
the bottom or
sides of the
container. Shake
the probe from
side to side in the
sample to refresh
the reference
junction.
2. In three
separate beakers
or appropriate
containers,
prepare 10,
100 and
1000 mg/L Na
+
standard solution
or use appropriate
standards that are
above and below
your application
range as the
standard set.
7. Put the beaker
on an
electromagnetic
stirrer and stir at a
moderate rate.
Check for air
bubbles and
remove them if
necessary.
3. Add the
contents of one
Sodium ionic
strength
adjustment (ISA)
powder pillow per
25 mL to each
standard.
8. Push Read.
The display will
highlight the
standard value
and proceed to
the next standard
value. The display
will show
"Stabilizing" and a
progress bar as
the reading
stabilizes. The
display shows the
standard value
when the reading
is stable.
4. Push
Calibrate. The
display shows the
current standard
value that is to be
read from the
standard solution
set.
9. Repeat steps
5-8 for the other
Na+) standard
solutions in the
set.
5. Rinse the
probe with ISA
rinse solution
(refer to Prepare
the ISA rinse
solution
on page 11).
Never use
deionized water..
Blot dry with a lintfree cloth. Do not
touch the tip of the
probe.
10. Push Done to
view the
calibration
summary. The
display will not
show Done until
the minimum
number of
calibration points
have been
collected.
4
11. Push Store to
accept the
calibration and
return to the
measurement
mode.
Measurement—direct method
Before measurement:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
If complete traceability is required, enter a sample ID and operator ID before measurement. Refer to the HQd meter manual
for more information.
Regular calibration is required for the best measurement accuracy (refer to Calibration on page 3).
Prepare the probe for use (refer to Preparation for use on page 2).
Measurement notes:
•
Stir the standards and samples at a slow and steady rate to prevent the formation of
a vortex.
•Stabilization times with smaller concentration changes generally will be longer and
can be minimized by proper stirring and conditioning. Experiment to determine the
proper stir rate if necessary.
•Note the temperatures of the samples during measurement. Temperatures of
calibration standards and samples should be kept within ±2 °C of each other for
optimal results.
•Data is automatically stored in the data log when Press to Read or Interval is
selected in the Measurement Mode. When Continuous is selected, data will only be
stored when Store is selected.
•Between measurements, rinse the probe with deionized water. Blot dry with a lint-free
cloth. For faster stabilization between measurements, put the probe in 100 mg/L
sodium standard (containing Sodium ISA) that is similar in concentration to the
samples to be analyzed.
•Air bubbles under the sensor tip when submerged can cause slow response or error
in measurement. If bubbles are present, gently shake the probe until bubbles are
removed.
•If a measurement error occurs, refer to Troubleshooting on page 12.
5
Loading...
+ 11 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.