Access code for programming, system settings, key lock
We are glad that you have chosen this product.
When developing this device, we combined approved technique with innovative detail solutions.
Greatest importance has also been attached to the design of an apparatus which provides easy
access for connection and maintenance works.
Safety remarks
Disconnect the mains power connections bef ore carrying out any connection, maintenance or
repair works.
Before putting into service
- Please read these operating instructions and special remarks thoroughly before putting the
device into service.
- Familiarize yourself with the safety and operating requirements in order to safeguard
personnel and equipment.
- The apparatus has been subjected to exhaustive quality inspections before dispatch.
- Any service and maintenance works required may only be carried out by trained service
personnel.
- Observe the local safety regulations and the rules for prevention of accidents.
- Observe the rules concerning the handling of hazardous substances.
- Do only use original spare parts or spare parts authorised by the manufacturer.
- No liability or claims under guarantee will be accepted in respect of any modifications or
conversions of the apparatus, other than those carried out by us or by persons authorised by
us or for which we have given express permission (in a written form)! This also refers to any
damage due to incorrect operation and/or improper use of the equipment.
Permissible application
The permissible application of the water sampler consists in the extraction of liquid, aqueous
substances, temperature range 0°C to 40°C. T he device must only be installed in areas free from
explosion hazard and must only be used for sampling of non-explosive substances. Please refer to
the technical specifications in chapter “Specifications”. Other applications are not allowed!
Permissible operating and ambient conditions
- Sample extraction of liquid aqueous substances, temperature range: 0°C to 40°C.
- The sampler is designed for operation in non-hazardous areas (no explosion risk).
- The sampler can be operated at ambient temperatures from -20°C to +43°C.
- Sampling from pressurised lines is not possible without optional accessories.
- The device is weather-proof and suitable for outside operation.
Page 4 Manual BÜHLER BU4010ff
General Information
Redelivery for repair
Before returning any appliances to us, please pay attention to the following to avoid unnecessary cost
and repair delays:
- Redeliveries only after agreement by the supplier
- All appliances or parts returned to us for repair must be cleaned and free of any hazardous
substances whatsoever (acids, alkaline solutions, solvents, etc.) as in Germany official
regulations regarding waste stipulate that persons in possession of hazardous waste are
responsible for its disposal and that, at the same time, employers are responsible for
protecting their employees against hazardous materials.
- Any cleaning or disposal required to be carried out by us will be charged for accordingly.
All packing materials can be disposed of as usual.
Packing materials are: cardboard, wood, PS and PE.
If the packing is returned free of cost, we will take care of its disposal.
Page 5 Manual BÜHLER BU4010ff
General Information
Safety information
Please read this entire manual before unpacking, setting up
or operating this equipment. Pay attention to all danger and
caution statements. Failure to do so could result in serious
injury to the operator or damage to the equipment.
To ensure that the protection provided by this equipment is
not impaired, do not use or install this equipment in any
manner other than that specified in this manual.
Use of Hazard Information
DANGER
Indicates a potentially or imminently hazardous situation
which, if not avoided, could result in death or ser ious injury.
CAUTION
Indicates a potentially hazardous situation that may result in
minor or moderate injury.
Important Note: Information that requires special
emphasis.
Note: Information that supplements points in the main text.
Precautionary labels
Read all labels and tags attached to the instrument.
Personal injury or damage to the instrument could occur if
not observed.
This symbol, if noted on the instrument, references the instruction manual for operation
and/or safety information.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of
electrical shock and/or electrocution exists.
This symbol, when noted on the product, identifies the location of the connection for
This symbol, if noted on the product, indicates the need for protective eye wear.
Protective Earth (ground).
This symbol, when noted on the product, identifies the location of a fuse or current
limiting device.
Electrical equipment marked with this symbol may not be disposed of in European public
disposal systems after 12 August 2005. In conformity with European local and national
regulations (EU Directive 2002/96/EC), European electrical equipment users must now
return old or end-of life equipment to the Producer for disposal at no charge to the user.
Note: For all electrical products (marked or unmarked) which are supplied or produced
Page 6 Manual BÜHLER BU4010ff
by Hach-Lange, please contact the local Hach-Lange sales office for instructions for
proper disposal.
Installation
4010
Device designation
Housing Stainless steel with 40 mm insulation,
Thermostatic control Self-contained cooling / heating with 4 settings, no-frost.
Temperature in sample compartment: 4°C (setting range: 0,0-9,9°C)
Control Microprocessor control, foil keyboard, graphical display (128*64 Pixel),
back lit
Data memory Non-volatile data memory: storage of sampling and malfunction data
like sample extractions, bottle changes, messages, external signals
Programming 12 user programs (can be edited freely)
Program start options Immediately, at a certain time, by an external signal
Program stop options End of sampling program after one program run, continuous operation
or x-runs
Pause mode Interruption of program at any time
Overfilling protection Adjustable from 1–999 samples/bottle
Interval setting 1 min. to 99 h 59 min. in steps of 1 minute
Pulse setting 1 to 9999 pulses/sample
Manual sample extraction Possible at any time without interrupting the current program run
Program protection Up to 5 years after voltage loss
Interface Mini-USB, RS422/485, RS 232
Communication Optional: TCP/IP RJ45 WEB board, 4-32GB SD/SDHC memorycard
Optional: Modbus or DP PROFIBUS connection
Languages Multi-language, selectable
Signal inputs • 2x analogue: 0/4-20 mA,
• 8x digital (flow, event, 1x free programmable)
Optional: expansion 4x digital, 3 free programmable and
8x Analogue Inputs 0-20mA / 0-10 V
Pulse-length min. 60 ms u. switch-level 7-24V, max. working
restistance 500 Ohm, signal-length max. 30 m
Signal outputs / status
messages
Sampling method -Vacuum system 20-350 ml, plastic vessel
Single sample volume
accuracy
Suction height Max. 8 m (at 1013 hPa and stagnant medium)
Pumping speed >0,5 m/s at a suction height up to at least 7,8 m (at 1013 hPa); pump
Suction hose PVC, L=7,5 m, ID=12 mm (ID 16 mm at flow-proportional option)
Sampling modes Time-related, flow-dependent, flow-proportional (option), event-related,
Bottle variants Plastic
Overall dimensions (Hxwxd)
Max. 8 digital outputs, 1x collective message
Optional: expansion 8x digital, 5 are free programmable
for bottle variants: 4x20 L, 4x25 L, 24x2,9 L, 24x2 L, 36x1 L
Page 7 Manual BÜHLER BU4010ff
Installation
Device designatio
n 4210
Device designation
4410
*) with opened top
Weight 100 kg with composite container; higher weight when using several
bottles and/or glass bottles
Power supply 230 V 50Hz
optional 115 V 50/60Hz
Power requirement Approx. 350 VA (with cooling)
Ambient temperature -20 - +43°C
Sample temperature 0 – 40°C
Standards Device meets ISO 5667 standard
Wetted materials PVC, Silicone, PS, PE, EPDM (optional: glass (Duran50) metering
vessel, sinker weight SS304)
(like 4010, with the following modifications)
Device type Fixed site sampler for sampling of faecal matter
Rinsing water inlet
Rinsing water outlet
Sampling method Vacuum system 20-500 ml
Sampling modes Time-related, flow-dependent, event-related, manual sample extraction
Bottle variants Plastic
Sample distribution Direct distribution with additional position for rinsing water. When using
Overall dimensions 1.290 (1.890*) x 690 x 645 mm
Weight 105 kg with composite container; higher weight when using several
Union nut R 3/4" (max. 2 bar)
Hose connector 25 mm
1 x 25 L, 4 x 14 L, 4 x 20 L
12 x 1 L, 23 x 1,0 L
Glass
12 x 1,0 L, 23 x 1,0 L
the rinsing function, the customer has to make sure that the rinsing
water can drain off freely
for bottle variants: 1x25 L, 4x14 L, 12x1 L
1400 (2015*) x 930 x 850 mm
for bottle variants: 4x20 L, 23x1 L
*) with opened top
bottles and/or glass bottles
(like 4010, with the following modifications)
Device type Self-emptying fixed site sampler for continuous operation without the
need of personnel (monitoring applications)
Emptying/rinsing Bottles are emptied, rinsed and re-filled on the same bottle position.
The rinsing head is installed above the bottle.
Bottle emptying for sample
collection
Hose inlet Hose connection at dosing unit (top)
Hose outlet Hose connection at dosing unit
Rinsing water inlet Union nut R 3/4" (max. 2 bar)
Rinsing water outlet Union nut R 1 1/4". When using the rinsing function, the customer has
Bottle variants Plastic
Overall dimensions (Hxwxd)
According to the device version: manual bottle emptying at 12 and 24-
bottle version or menu-driven emptying at 2 and 4-bottle version
Union nut R 3/4", hose ID 12 mm
Silicone hose 12x2, is only slipped over
to make sure that the rinsing water can drain off freely
2 x 10 L, 4 x 5 L, 4 x 10 L, 24 x 2 L (PPH)
Glass
12 x 1,6 L, 24 x 1,0 L, 24 x 2 L
• 1.290 (1930*) x 690 x 645 mm
Page 8 Manual BÜHLER BU4010ff
Installation
Device designation
6010
double
-
housing
Device designation
3010
for bottle variants: 2 x 10 L, 4 x 5 L, 12 x 1,6 L
• 1400 (2175*) x 800 x 850 mm
for bottle variants: 4 x 10 L, 24 x 1 L, 24 x 2 L
*) with opened top
Weight 115 kg with bottle option 2x10 L PE; higher weight when using other
bottle or glass bottle options
(like 4010, with the following modifications)
Device type
Measuring pot inlet Hose nozzle ½“, hose ID 12 mm
Measuring pot outlet Hose nozzle ID 40 mm. The customer has to make sure that the water
Bottle variants Plastic
Overall dimensions (Hxwxd)
Weight Approx. 250 kg with composite container; higher weight when using
Power requirement Approx. 450 VA (with cooling + SC1000 controller), higher requirement
Device type Compact device in plastic housing, especially suitable for high ambient
Suction height Max. 7,5 m (at 1013 hPa and stagnant medium)
Pumping speed >0,5 m/s at suction height up to at least 6 m (at 1013h Pa); pump
Suction hose PVC, L=7,5 m, ID=10 mm (ID 16 mm at flow-proportional option).
Bottle variants Plastic
Sampling method Vacuum system 20-350 ml with plastic metering vessel
Ambient temperature -20°C to + 50°C (tropicalised)
Overall dimensions (Hxwxd)
Weight Approx. 70 kg with composite container, higher weight when using
Fixed site sampler in
in the measuring pot can drain off freely
1 x 25 L, 1 x 50 L, 2 x 10 L
4 x 6,0 L, 4 x 10 L, 4 x 14 L,
12 x 2,9 L, 24 x 1,0 L
Glass
12 x 2,0 L, 24 x 1,0 L
1.690 (2290*) x 1280 x 645 mm
*) with opened top
several bottles and/or glass bottles
when using the optional spiral pump
(like 4010, with the following modifications)
temperatures. Material: PE with 50 mm insulation / Styrosun / PC
(GF10)
capacity can be adjusted electronically
Max. hose length 30 m
1 x 25 L, 4 x 14 L, 12 x 2,9 L, 24 x 1,0 L
Glass
12 x 2,0 L, 24 x 1,0 L
1.100 (1.640*) x 760 x 745 mm
*) with opened top
several bottles and/or glass bottles
(measuring station)
Page 9 Manual BÜHLER BU4010ff
Installation
Characteristics: sampler 4010
The device consists of the following component parts:
1. Protective top (can be opened)
2. Upper part of housing
3. Back lit LC-display
4. Keyboard
5. Lower part of housing (cooled
sample compartment)
6. Motor-driven distributor
7. Distributor plate
8. Sample bottles
9. Carrying tray for bottles
10. Fixing bar
11. Fixing bar
12. Dosing unit
13. Pinch valve
14. Type plate
Remark: The illustrated sampler corresponds to a standard version with 12 bottles and
distributor plate.
According to the sampler version, some components such as dosing unit or
distributor/bottles can be different!
Remark: Please read the entire manual before unpacking, installing and operating this device.
Observe all danger and safety regulations. Non-observance can lead to serious
personal injury or to damage to the instrument.
DANGER The workings described in this chapter of the manual must only be carried our by
qualified personnel.
Page 10 Manual BÜHLER BU4010ff
Installation
Characteristics: sampler 4210
Compared to the standard samplers, the special sampler for sampling of faecal matter is equipped
with a rinsing unit in the metering vessel and an additional outlet. Furthermore the sample is filled
directly into the bottles. The section from the metering vessel to the distributor outlet is rinsed.
Method of operation: The metering vessel is rinsed automatically after each sample extraction. To do this, the distributor arm first moves on the discharge position and
then back on the bottle. Thus a sample contamination is avoided.
The picture shows the metering vessel with additional rinsing water
connection.
The rinsing head (arrow) can be seen after having removed the
metering vessel.
The picture shows the distributor unit:
1. Detailed view of discharge position when rinsing.
2. Distributor arm with direct dosing.
3. Sample bottles.
4. Bottle carrying tray.
Page 11 Manual BÜHLER BU4010ff
Operation
Characteristics: sampler 4410
1. Protective top (can be
opened)
2. Upper part of housing
3. Back lit LC-display
4. Keyboard
5. Lower part of housing
(cooled sample compart ment)
6. Type plate
7. Glass sample bottles
8. Distributor drive
9. Fixing bar
10. Dosing unit
11. Pinch valve
12. Combined dosing and
rinsing head
13. Automatic discharge valve
14. Swivelling sample discharge tube
Remark: The illustrated sampler corresponds to a standard version with 24 bottles.
According to the sampler version, some components such as dosing unit or
distributor/bottles can be different!
Page 12 Manual BÜHLER BU4010ff
Operation
Characteristics: sampler 6010
1. Measuring unit with
transducers and display
2. Control
3. Dosing unit
4. Distributor
5. Cable feed through
(opt.: left side)
6. Main switch
7. Display of protection
against dry running
8. Water inlet and outlet
connection (opt.: left side)
9. Pump switch I/O
10. Terminal box
11. Pump (eccentric spiral
pump)
12. Water inlet and outlet of
measuring pot
13. Measuring pot
14. Fixture of electrodes/
measuring probes
Page 13 Manual BÜHLER BU4010ff
Operation
Schematic illustration
Water supply / water drain
1. Measuring transducer with
display
2. Sampler control
3. Dosing system
4. Feed tube of dosing system
5. Distributor
6. Bottles with tray
7. Measuring pot drain pipe
8. Pump (e.g. eccentric spiral
pump)
9. Outlet valve (manual/
automatic) for cleaning of
measuring pot
10. Supply valve for measuring
pot
11. Measuring pot with pressureless
overflow
12. Fixture of electrodes/
measuring probes
Overall dimensions: 1690 x 1280 x 645 mm (hxwxd)
Page 14 Manual BÜHLER BU4010ff
Operation
Characteristics: sampler 3010
1. Protective top (can be opened)
2. Upper par t of housing
3. Back-lit LC-display with
keyboard
4. MiniUSB interface
5. Lower part of housing (cooled
compartment)
6. Motor-driven distributor
7. Distributor plate
8. Sample bottles
9. Bottle tray
10. Fixing bar
(illus. with option: mobile version)
11. Fixing bar
12. Dosing unit
13. Pinch valve
14. Type plate
Remark: The illustrated sampler corresponds to a standard version with 12 bottles and
distributor plate.
According to the sampler version, some components such as dosing unit or
distributor/bottles can be different!
Page 15 Manual BÜHLER BU4010ff
Operation
Dimensions of devices
Page 16 Manual BÜHLER BU4010ff
Operation
Page 17 Manual BÜHLER BU4010ff
Operation
Installing the apparatus
- The workings described in this chapter of the manual must only be carried out by qualified
personnel.
- We recommend to install the apparatus as near as possible to the point of extraction. If installed
outside, the device should be fixed on a solid flat base (e.g. plinth). The fixing bars do already
have the necessary fixing holes.
- If the device is equipped with castors (mobile version), please make sure that the two lock-type
castors are locked during operation.
- For better deaeration / ventilation install the apparatus with a distance of approx. 10 cm to the wall.
- The hose has to be laid with a continuous fall from the apparatus to the point of extraction, without
fail. Lower lying points can lead to deposits in the hose which may freeze up in winter.
- To fix the suction hose, we recommend to use the extraction unit available as accessory.
- Fix the hose in such a way that the suction aperture lies in the direction of flow, so that coarse
matter and fibres cannot be forced into the suction aperture.
Remark: To open the top cover both front screws (left + right)
have to be removed. Then the top cover can be lifted
up to the point where the retention bars snap-in. To
close the cover again, the two retention bars have to
be lifted slightly.
After having closed the cover, the
two screws (left + right) have to be screwed in again.
Mains connectionThe device is equipped with a power supply lead. The connecting data can be found in chapter “Technical Data”.
Connect/disconnect The device is connected/disconnected by means of the mains plug (except
devices with main switch).
Hose routing The apertures for the suction hose are on the left and on the right at the front.
Signal inputs The signal inputs (e.g. for flow meter) are located on the „fold-out“ plate and
can be connected according to the plan of terminal connections shown there.
MessagesThe relays required for messages (up to 4) can be fixed on the top hat rail. For
details please see the circuit diagram
Remark:
The apparatus must be cleaned regularly in accordance with the degree of contamination present. In
view of the quality of samples, we recommend to clean thoroughly especially the wetted parts like
dosing unit, electrodes, distributor, bottles and inlet hose. Failure to do so could result in damage or
destruction to the equipment, device thatare not covered by warranty.
Page 18 Manual BÜHLER BU4010ff
Operation
Replacement of fuses
1. Main fuse of device
(is installed on fold-out plate)
Fine-wire fuse 230V/T8A 5x20
To replace the fuse, pull out fuse holder,
open the cover and remove the fuse.
2. Fuse protection of CPU board
by an automatic fuse. The fuse is
reactivated by disconnecting the device
from the mains.
Troubleshooting
If a fault develops in your sampler, we can only provide prompt assistance if you give us the type and
serial number of the apparatus concerned. You will find these data on the type plate.
The more precise your description of the fault is, the better our fault diagnosis will be.
Symptom Possible cause Action
Sampler does not
function at all
Cooling system does
not function
- Mains connection
- Fuse defective
- Power supply
- Cooling system leaking
-Check mains cable
-Check all fuses and replace them
if necessary
-Check mains connection of
cooling machine
-Contact service department
Page 19 Manual BÜHLER BU4010ff