Measuring Ranges: 0.05 to 20 mg/L; 1 to 100 mg/L and 10 to 1000 mg/L NH
< 5 minutes
Adjustable measuring interval5 to 120 minutes
–N Measuring Range)
4
-N Measuring Range)
4
–N Measuring Range)
4
–N Measuring Range)
4
–N Measuring Range)
4
–N Measuring Range)
4
–N Measuring Range)
4
–N Measuring Range)
4
–N
4
–N
4
Power supply
Power supply with power cable only using the sc1000 controller
(analyzer, Filter Probe sc and drain tubing: 115 V versions or 230 V versions)
Data transmissionData transmission with data cable on the sc1000 controller
Electrical power consumption500 VA
Electrical fuse protection
Via sc1000 controller
Maximum of 2 analysis instruments for each sc1000 controller.
OutputsRelay, current outputs, bus interface via sc1000 controller
Operating temperature
Storage temperature
AMTAX sc: –20 to 45 °C (–4 to 113 °F); 95 % relative humidity, non-condensing
AMTAX indoor sc: 5 to 40 °C (41 to 104 °F); 95 % relative humidity, non-condensing
–20 to 60 °C (–4 to 140 °F); 95 % relative humidity, non-condensing
4 to 55 °C (39 to 131 °F); 95 % relative humidity, non-condensing (electrode)
Sample temperature4 to 40 °C (39 to 104 °F)
Sample pressureWith continuous sample preparation –30 mbar to +50 mbar at overflow vessel
Sample flowRange: 1.0 L/h–20.0 L/h
Sample qualityUltra filtrated or
comparable
Sample levelLevel of liquid in basin with filtration probe must be below analyzer
5
Specifications
Permissible pH value of the sample 5 to 9
Permissible hardness range<= 50 °dH 8.95 mMol/L
Permissible chlorid range<= 1000 mg/L Cl
–
Dimensions (Figure 1, page 7,
Figure 1, page 7)
Data and power cable lengths2 m (80 in.) (from edge of enclosure)
Weight
CertificationsCE conform. Listed to UL and CSA safety standards by TUV.
AMTAX sc: (W × H × D) 540 × 720 × 390 mm (21,25 × 28,35 × 15,35 in.)
AMTAX indoor sc: (W × H × D) 540 × 720 × 370 mm (21,25 × 28,35 × 14.5 in.)
AMTAX sc: Approx. 31 kg, without Filter Probe sc and without chemicals
AMTAX indoor sc: Approx. 29 kg, without Filter Probe sc and without chemicals
6
Specifications
Figure 1 Instrument dimensions AMTAX sc
7
Specifications
[1.57in]
850 mm
[33.15 in]
720 mm
[28.35 in]
540 mm
[21.26 in]
370 mm
[14.57 in]
Figure 2 Instrument dimensions AMTAX indoor sc
8
Section 2General Information
2.1Safety information
Please read this entire manual before unpacking, setting up or
operating this equipment. Pay attention to all danger and caution
statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not
impaired, do not use or install this equipment in any manner other
than that specified in this manual.
2.1.1 Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation
which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation
which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in
minor or moderate injury.
2.1.2 Precautionary labels
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC),
European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at
no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories and all auxiliary items for proper disposal.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or
electrocution exists.
Important Note: Indicates a situation which, if not avoided, may
cause damage to the instrument. Information that requires special
emphasis.
Note: Information that supplements points in the main text.
Read all labels and tags attached to the instrument. Personal injury
or damage to the instrument could occur if not observed. A symbol,
if noted on the instrument, will be included with a danger or caution
statement in the manual.
This symbol, if noted on the product, indicates the need for protective eye wear.
This symbol, when noted on the product, identifies the location of the connection for Protective Earth (ground).
This symbol, when noted on the product, identifies the location of a fuse or current limiting device.
This symbol, when noted on the product, indicated that the marked item can be hot and should not be touched
without care.
9
General Information
18-32 kg (39.7-70.5 lbs)
This symbol, when noted on the product, identifies a risk of chemical harm and indicates that only individuals
qualified and trained to work with chemicals should handle chemicals or perform maintenance on chemical delivery
systems associated with the equipment.
This symbol, when noted on the product, indicated the presence of devices sensitive to Electro-static Discharge
(ESD) and indicated that care must be taken to prevent damage with the equipment.
When carrying or transporting the instrument/instrument components and if the total weight is more than 18 kg,
make sure that suitable lifting equipment is used and/or that the instrument/instrument components are carried by
2 people.
2.1.3 Change instrument labels
2.2Product overview
Several safety labels (3 in the analytical section) are applied to the
instrument. If necessary, apply the correct language label over the
existing safety labels.
The AMTAX sc (Figure 3, Figure 4) measures the ammonium ions
present in treated aqueous solutions (waste water, process water
and surface water). The measured value is displayed in mg/L
NH
-N on the controller. The AMTAX sc must be used in
4
combination with the sc1000 controller. The sc1000 controller is
used to configure, power and output the measured values.
Conversion formula: NH
-N : NH
4
+
= 1 : 1.288
4
The AMTAX sc can operate using single or dual channel modes.
Operation with the Filter Probe sc is single channel only. The sc
analyzer can be converted from a single channel operation to dual
channel operation. Contact the manufacturer for more information.
10
General Information
Dual channel operation is only possible with continuous sample
preparation, e.g. FILTRAX or Ultrafiltration. Sample preparation and
filtration must be provided before installing the analysis instrument.
1LED for operating state. Refer to
Table 9, page 57 for more
information.
Figure 3 AMTAX sc enclosure
2Door lock3Rating plate with model number,
serial number, voltage and
frequency information and power
consumption information
1LED for operating state. Refer to
Table 9, page 57 for more
information.
Figure 4 AMTAX indoor sc enclosure
2Door lock3Rating plate with model number,
serial number, voltage and
frequency information and power
consumption information
11
General Information
12
Section 3Installation
DANGER
Only qualified personnel should conduct the tasks described
in this section of the manual.
CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.
DANGER
Potential danger in the event of contact with
chemical/biological materials. Handling chemical samples,
standards and reagents can be dangerous. Familiarize
yourself with the necessary safety procedur es and t he corr ect
handling of the chemicals before the work and r ead and fo llow
all relevant safety data sheets.
Normal operation of this instrument may involve the use of
hazardous chemicals or biologically harmful samples.
•Observe all cautionary information printed on the original
solution containers and safety data sheet prior to their
use.
3.1Basic installation overview
•Dispose of all consumed solutions in accordance with
national regulations and laws.
•Select the type of protective equipment suitable to the
concentration and quantity of the dangerous material at
the respective work place.
1. Unpack the instrument (section 3.2).
2. Mount the instrument (section 3.3, page 14).
3. Remove any transport locks (section 3.4.2, page 19).
4. Install the collecting tray and the humidity sensor (section 3.4.3,
page 21 and section 3.4.4, page 22).
5. Determine the appropriate installation option (section 3.4.5,
page 23).
6. Mount the Filter Probe sc or Filtrax, if necessary. Refer to the
appropriate manual for more information.
7. Connect the Filter Probe sc or Filtrax to the AMTAX sc, if
necessary. Refer to section 3.5.3, page 26 for the Filter Probe
sc. Refer to the Filtrax manual for more information.
8. Connect the Drain Heating Connection, if necessary.
9. Make all plumbing connections (Appendix A Plumbing and
Connection Options, page 77).
10. Install the reagents and prepare the Electrode and Electrolyte
(section 3.6, page 29 and section 3.7.1, page 31).
11. Connect the AMTAX sc to the sc1000 controller to supply
power to the system (section 3.8, page 33).
12. Connect the data network (section 3.9, page 34).
13
Installation
3.2Unpack the instrument
3.3Mechanical installation
CAUTION
Pay attention to the weight (approximately 31 kg) of the
instrument. Do not try to carry the instrument without
assistance. Use only suitable lifting tackle for transport.
Open the shipping container while it is on its end and then slide the
analyzer out of the cardboard. The items supplied will vary
depending on the order. Standard items supplied for a minimal
configuration include:
•AMTAX sc and user manual
•Collecting tray
•Initial set of reagent and cleaning solutions
2 standard solutions and electrolyte / membrane caps
•Fastening bracket and angle bracket
•Accessories for tubing and flow-through variant
•Plug set
3.3.1 Mount the instrument
Select a suitable place to install the instrument. Plan the
mechanical installation before positioning posts or drilling holes.
Refer to Figure 1, page 7, Figure 2, page 8 for instrument
dimensions.
Make sure that the fastening has sufficient load bearing capacity
(approximately 160 kg). The wall plugs must be selected and
approved to suit the properties of the wall.
Plan cable and tubing routes to avoid sharp bends and tripping
hazards.
When connecting two analyzers (e. g. for measuring two
parameters with a FILTRAX or Ultrafiltration), plan where the
instruments are to be installed and consider the length of the
heated drain tubing (2 m).
The AMTAX sc can be mounted in three different ways:
•Wall Mount (section 3.3.1.1),
•Rail Mount; refer to the instruction sheet supplied with the
Rail Mounting Hardware.
•Stand Mount; refer to the instruction sheet supplied with the
Rail Mounting Hardware.
3.3.1.1 Wall mount
14
Refer to Figure 5, Figure 6 and following instructions to mount the
analyzer to a wall.
1. Align and install the fastening bracket to the wall.
2. Attach the angle bracket to the instrument using the
supplied screws.
3. Slide the bottom of the enclosure onto the fastening bracket.
Installation
4. Attach the enclosure to the fastening bracket.
5. Attach the angle bracket on the enclosure to the wall.
Figure 5 Bracket dimensions for wall mounting
15
Installation
Figure 6 Wall mounting the analyzer
1Socket head cap screw, M5 X 8 (2X)4Screw, customer supplied
2Washer, M5 (4X)5Socket head cap screw, M5 X 40 (2X)
3Angle bracket6Fastening bracket
16
3.4Initial instrument setup
3.4.1 Open the enclosure
Installation
DANGER
To reduce the risk of electrica l shock, mak e sure tha t no water
can enter the enclosure or come into contact with circuit
boards.
CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.
2. Open the side latches and release the door catch.
3. Open the door and secure the door using the hook or
completely remove the door.
Figure 7 Open the enclosure AMTAX sc
1Latches3Pocket for user manual
2Door hook4Lock with key
17
Installation
Figure 8 Open the enclosure AMTAX indoor sc
1Latches3Lock with key
2Door hook
18
3.4.2 Remove the shipping transport locks
Prior to system start-up, the shipping transport locks must be
removed from the sc analyzer.
CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.
Important note: The electrode insulation for the measuring unit is
NOT a transport lock. Do NOT remove the cover on the measuring
unit.
1. Open the enclosure door and secure with the door hook.
2. Remove the transport lock on the analyzer panel (Figure 9).
Note: If the instrument is operated with a Filter Probe sc, it is equipped
with an internal compressor.
3. Remove the cable tie and pull out the compressor transport
lock to the left (Figure 10).
Note: Keep the transport locks for transporting and storage.
Installation
1Transport lock
Figure 9 Remove the analyzer panel transport locks
19
Installation
Figure 10 Compressor transport lock removal
1Compressor transport lock4Protective cover for compressor
2Cable tie5Fan locking screw
3Compressor 6Fan
1
The compressor, compressor transport lock and cable tie only apply to sc analyzers that operate using the Filter Probe sc.
1
20
3.4.3 Installation of the collecting tray
CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.
1. Open the enclosure door and secure with the door hook.
2. Slide the collecting tray into the bottom of the enclosure
Installation
(Figure 11).
1Collecting tray
Figure 11 Installation of the collecting tray
21
Installation
3.4.4 Connect the humidity sensor
CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.
1. Remove power from the instrument.
2. Open the enclosure door and secure with the door hook.
3. Connect the humidity sensor wires to the terminal screws on
the collecting tray (Figure 12).
Figure 12 Connect the humidity sensor
1Collecting tray2Humidity sensor connections
22
3.4.5 Determine the appropriate installation option
Before connecting tubing or cables, determine the option number
that corresponds the system configuration. Refer to Table 1. Based
on the option number, determine the sealing plug that will be used
to seal the enclosure openings, refer Table 2.
When the option number is determined, refer to Appendix A
Plumbing and Connection Options, page 77 for installation
1Sealing plug type 1
2Sealing plug type 2
3Sealing plug type 3
3.5Electrical installation
DANGER
High voltage wiring connections are present under the
protective cover. The protective cover must remain in place
unless a qualified installation technician is installing wiring for
the Filter Probe sc or the heated drain.
See Figure 14 for protective cover removal.
Figure 14 Remove the protective covers
1Protective cover for AC mains circuits (Back view)
2Protective cover for main PCB
24
3.5.1Electrostatic Discharge (ESD) Considerations
Important Note: To minimize hazards an d ESD risks, maintenance
procedures not requiring power to the analyzer should be
performed with power removed.
Delicate internal electronic components can be damaged by static
electricity, resulting in degraded instrument performance or
eventual failure.
The manufacturer recommends taking the following steps to
prevent ESD damage to the instrument:
•Before touching any instrument electronic components (such
as printed circuit cards and the components on them)
discharge static electricity. This can be accomplished by
touching an earth-grounded metal surface such as the chassis
of an instrument or pipe or a metal conduit.
•To reduce static build-up, avoid excessive movement.
Transport static-sensitive components in anti-static containers
or packaging.
•To discharge static electricity and keep it discharged, wear a
wrist strap connected by a wire to earth ground.
Installation
3.5.2 Enclosure breakouts
•Handle all static-sensitive components in a static-safe area. If
possible, use anti-static floor pads and work bench pads.
There are four main enclosure openings to insert tubing and cables
(Figure 15).
1Sample feed or refer to
Table 1, page 23 for
tubing options.
Figure 15 Enclosure breakouts
2Refer to Table 1,
page 23 for tubing
options.
3Refer to Table 1,
page 23 for tubing
options.
4Power and data cables
25
Installation
3.5.3 Insert tubing and/or cables
1Guide the tubing or cables through
the enclosure openings
(Figure 15).
2Push the plug from the top onto the
tubing or onto the cable.
3.5.4 Connect the Filter Probe to the analyzer
CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.
DANGER
Disconnect power from the sc analyzer at the sc1000 before
removing the protective covers in the an aly ze r.
Important Note: Make sure that for the use of a filter probe, the
level of the water in which the filtration probe is submersed, is
below the level of the analyzer.
1. Open the enclosure door and secure with the door hook.
2. Open the analyzer panel.
3. Remove the two screws from the protective cover and remove
the cover (item 1, Figure 14, page 24).
4. Connect the earth ground (green/yellow) wire (item 9,
Figure 16, page 28) from the Filter Probe sc to the ground
terminal (item 5, Figure 16, page 28).
3Pull down the plug with the tubing
or the cables. Seal any unused
entry with Plug #3.
26
5. Connect the power connector to the appropriate terminal
connection (items 4 and 11, Figure 16, page 28).
6. Remove the three screws securing the bottom panel cover.
(item 2, Figure 14, page 24). Remove the panel.
7. Connect the data connector (item 10, Figure 16, page 28) to
the main board (item 12, Figure 16, page 28).
8. Install all covers and panels.
9. Connect the white air tubing (item 8, Figure 16, page 28) from
the Filter Probe sc to the air tubing connection on the analyzer
(Figure 16, page 28).
10. For sample and drain line connections refer to A.4, page79.
3.5.5 Connect the optional heated drain
CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.
DANGER
Disconnect power from the sc analyzer at the sc1000 before
removing the protective covers in the analyzer.
Refer to Figure 16 and the following procedure to connect the
heated drain.
1. Open the enclosure door and secure if necessary.
2. Open the analyzer panel.
3. Remove the protective cover (Figure 14, page 24).
4. Connect the earth ground wire (green/yellow) to the ground
5. Connect the cables for the heated drain (item 6, Figure 16,
6. Connect the drain tube as described in the appropriate Option
Installation
wire terminal strip.
page 28) to the terminal block (item 3, Figure 16, page 28).
Configuration, refer to sectionA.5, page80 for more
information.
7. Place the drain tube to the appropriate drain or basin.
8. Install all covers and panels.
27
Installation
Figure 16 Connect the Filter Probe sc and optional heated drain
1Bottom panel cover8Filter Probe sc air tube (white)
2Protective cover9Filter Probe sc ground wire
3Heated drain (optional) power connector10 Filter Probe sc data cable connector
4Filter Probe sc power connector11 Filter Probe sc power cable connector
5Ground wire terminal strip12 Filter Probe sc data connector
6Heated drain power cable connector13 Remote control input (15–30 V DC) (Refer to
7Heated drain ground wire
section B.3 , page 112)
28
3.6Installation of reagents
Installation
DANGER
Potential danger in the event of contact with
chemical/biological materials. Handling chemical samples,
standards and reagents can be dangerous. Familiarize
yourself with the necessary safety procedur es and t he corr ect
handling of the chemicals before the work and r ead and fo llow
all relevant safety data sheets.
Normal operation of this instrument may involve the use of
hazardous chemicals or biologically harmful samples.
•Observe all cautionary information printed on the original
solution containers and safety data sheet prior to their
use.
•Dispose of all consumed solutions in accordance with
national regulations and laws.
•Select the type of protective equipment suitable to the
concentration and quantity of the dangerous material at
the respective work place.
CAUTION
Avoid unnecessary contact with sample flows of unknown
concentration. This can produce hazards due to trace
chemicals, radiation or biological effects.
CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.
Important Note: Always lay the drain tubing so that there is a
continuous fall (minimum 3°), the outlet is clear (not pressurized)
and the drain tubing is not longer than 2 meters. For more
information refer to Appendix A, page 77.
Important Note: Incorrect use of the reagents can damage the
instrument. Carefully read the labels on the containers to make
sure that there are no mistakes.
The reagents and chemicals supplied are ready to use. The
reagents need to be placed in the analysis instrument and tubings
connected. Refer to Table 3 to determine the correct standards.
1. Place the reagent containers in the instrument (Figure 17).
2. Insert the tubing in the reagent containers.
3. Screw the reagents to the supplied caps.
Figure 17 Chemicals and reagents in the AMTAX sc (Refer to Table 3, page 30 for more information)
1Electrolyte solution replacement bottles7Sample line
2Electrode assembly8Reagent
3Fastener for membrane cap9Standard solution: High standard
4Electrode panel connector10 Standard solution: Low standard
5Connection to cuvette11 Cleaning solution
6Drain
30
Loading...
+ 88 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.