Hach-Lange AMTAX User Manual

DOC023.52.00025

AMTAX sc, AMTAX indoor sc

USER MANUAL
04/2013, Edition 7A
AMTAX sc, AMTAX indoor sc
© HACH-LANGE GmbH, 2006–2010; 2012, 2013. All rights reserved. Printed in Germany. sd/sk

Table of Contents

Section 2 General Information ................................................................................................9
2.1 Safety information ........................................................................................................................ 9
2.1.1 Use of hazard information ................................................................................................... 9
2.1.2 Precautionary labels ........................................................................................................... 9
2.1.3 Change instrument labels .................................................................................................10
2.2 Product overview ....................................................................................................................... 10
Section 3 Installation .............................................................................................................13
3.1 Basic installation overview ......................................................................................................... 13
3.2 Unpack the instrument ............................................................................................................... 14
3.3 Mechanical installation ............................................................................................................... 14
3.3.1 Mount the instrument ........................................................................................................ 14
3.3.1.1 Wall mount ............................................................................................................... 14
3.4 Initial instrument setup ............................................................................................................... 17
3.4.1 Open the enclosure ........................................................................................................... 17
3.4.2 Remove the shipping transport locks ................................................................................ 19
3.4.3 Installation of the collecting tray ........................................................................................ 21
3.4.4 Connect the humidity sensor ............................................................................................ 22
3.4.5 Determine the appropriate installation option ................................................................... 23
3.5 Electrical installation .................................................................................................................. 24
3.5.1 Electrostatic Discharge (ESD) Considerations ................................................................. 25
3.5.2 Enclosure breakouts ......................................................................................................... 25
3.5.3 Insert tubing and/or cables ................................................................................................ 26
3.5.4 Connect the Filter Probe to the analyzer .......................................................................... 26
3.5.5 Connect the optional heated drain .................................................................................... 27
3.6 Installation of reagents ............................................................................................................... 29
3.7 Gas-sensitive electrode ............................................................................................................. 31
3.7.1 Installation of the electrode and electrolyte ....................................................................... 31
3.7.1.1 Fill the electrode with electrolyte .............................................................................. 31
3.8 Supply power to the analyzer ..................................................................................................... 33
3.9 Connect the data network .......................................................................................................... 34
Section 4 System Start Up .....................................................................................................35
4.1 Initializing the instrument ........................................................................................................... 35
Section 5 Operation ...............................................................................................................37
5.1 Sensor diagnostics menu ........................................................................................................... 37
5.2 Sensor setup menu .................................................................................................................... 37
5.2.1 System setup menu .......................................................................................................... 42
5.3 Calibration process .................................................................................................................... 42
5.4 Cleaning process ....................................................................................................................... 42
5.5 Measurement process ............................................................................................................... 43
Section 6 Maintenance ...........................................................................................................45
6.1 General maintenance ................................................................................................................. 45
6.1.1 Clean the analyzer ............................................................................................................ 45
6.1.1.1 Cleaning interval ......................................................................................................46
6.1.2 Replace the fan filter ......................................................................................................... 46
6.1.3 Fuse replacement ............................................................................................................. 47
6.2 Reagent replacement ................................................................................................................. 47
6.3 Routine maintenance schedule .................................................................................................. 48
6.4 Scheduled maintenance ............................................................................................................ 49
3
Table of Contents
6.5 Replace the membrane cap, electrolyte and electrode ..............................................................49
6.6 Validation (Analytical quality assurance) ....................................................................................52
6.7 Shut the analyzer down ..............................................................................................................54
6.7.1 Shut the analyzer down for an extended period ................................................................54
6.8 Modify from single channel to dual channel ...............................................................................55
Section 7 Troubleshooting ....................................................................................................57
7.1 Troubleshooting the controller ....................................................................................................57
7.2 Troubleshooting the analyzer .....................................................................................................57
7.2.1 LED status .........................................................................................................................57
7.2.2 Error messages .................................................................................................................57
7.2.3 Warnings ...........................................................................................................................60
7.3 Troubleshooting the Electrode ...................................................................................................62
Section 8 Replacement Parts and Accessorie ....................................................................65
8.1 Standards and reagents .............................................................................................................65
8.2 Analyzer accessories .................................................................................................................65
8.3 Mounting hardware and accessories ..........................................................................................66
8.4 Replacement parts .....................................................................................................................66
Section 9 Contact Information .............................................................................................73
Section 10 Warranty and liability ..........................................................................................75
Appendix A Plumbing and Connection Options .................................................................77
A.1 Safety information ......................................................................................................................77
A.1.1 Electrostatic Discharge (ESD) Considerations ..................................................................77
A.2 Connect a 2-parameter option ...................................................................................................78
A.2.1 Remove the T-fitting ..........................................................................................................79
A.3 Drain line considerations ............................................................................................................79
A.4 Tubing considerations ................................................................................................................79
A.5 Option 1 plumbing and connections ...........................................................................................80
A.6 Option 2 plumbing and cable connections .................................................................................82
A.7 Option 3 plumbing and connections ...........................................................................................84
A.8 Option 4 plumbing and connections ...........................................................................................86
A.9 Option 5 plumbing and connections ...........................................................................................88
A.10 Option 6 plumbing and connections .........................................................................................90
A.11 Option 7 plumbing and connections .........................................................................................92
A.12 Option 8a plumbing and connections .......................................................................................94
A.13 Option 8b plumbing and connections .......................................................................................96
A.14 Option 9a plumbing and connections .......................................................................................98
A.15 Option 9b plumbing and connections .....................................................................................100
A.16 Option 10a plumbing and connections ...................................................................................102
A.17 Option 10b plumbing and connections ...................................................................................104
A.18 Option 11a plumbing and connections ...................................................................................106
A.19 Option 11b plumbing and connections ...................................................................................108
Appendix B Fieldbus Communications .............................................................................111
B.1 Fieldbus control ........................................................................................................................111
B.2 Remote controlled measurement series ..................................................................................111
B.3 External trigger contact, Control by external signal .................................................................112
B.4 Modbus register information ....................................................................................................112
4

Section 1 Specifications

Specifications are subject to change without notice.
Enclosure rating
AMTAX sc: IP55 AMTAX indoor sc: IP54
Enclosure material ASA/PC UV-resistant
Measuring method GSE (Gas-Sensitive Electrode)
–N
4
–N
4
–N
4
–N
4
Measuring range
0.02 to 5.0 mg/L NH
0.05 to 20 mg/L NH 1 to 100 mg/L NH 10 to 1000 mg/L NH
0.02 mg/L NH4–N (0.02 to 5.0 mg/L NH
Detection limit
0.05 mg/L NH 1 mg/L NH 10 mg/L NH
–N (0.05 to 20 mg/L NH4–N Measuring Range)
4
–N (1 to 100 mg/L NH4–N Measuring Range)
4
–N (10 to 1000 mg/L NH4–N Measuring Range)
4
1 mg/L: 3 % + 0.02 mg/L >1 mg/L: 5 % + 0.02 (0.02 to 5.0 mg/L NH
Measuring accuracy (with standard solution)
3 % + 0.05 mg/L (0.05 to 20 mg/L NH 3 % + 1.0 mg/L (1 to 100 mg/L NH
4.5 % + 10 mg/L (10 to 1000 mg/L NH –N (0.02 to 5.0 mg/L NH4–N Measuring Range)
4
Repeatability (with standard solution)
3 % + 0.02 mg/L NH 2 % + 0.05 mg/L (0.05 to 20 mg/L NH 2 % + 1.0 mg/L (1 to 100 mg/L NH 2 % + 10 mg/L (10 to 1000 mg/L NH Measuring Range: 0.02 to 5 mg/L NH
–N: 3 measurements (shortest 15 minutes)
4
Response time (90 %)
0.02 to 0.2 mg/L NH
0.2 to 5 mg/L NH4–N: 1 measurement (5 minutes)
Measuring Ranges: 0.05 to 20 mg/L; 1 to 100 mg/L and 10 to 1000 mg/L NH < 5 minutes
Adjustable measuring interval 5 to 120 minutes
–N Measuring Range)
4
-N Measuring Range)
4
–N Measuring Range)
4
–N Measuring Range)
4
–N Measuring Range)
4
–N Measuring Range)
4
–N Measuring Range)
4
–N Measuring Range)
4
–N
4
–N
4
Power supply
Power supply with power cable only using the sc1000 controller (analyzer, Filter Probe sc and drain tubing: 115 V versions or 230 V versions)
Data transmission Data transmission with data cable on the sc1000 controller
Electrical power consumption 500 VA
Electrical fuse protection
Via sc1000 controller Maximum of 2 analysis instruments for each sc1000 controller.
Outputs Relay, current outputs, bus interface via sc1000 controller
Operating temperature
Storage temperature
AMTAX sc: –20 to 45 °C (–4 to 113 °F); 95 % relative humidity, non-condensing AMTAX indoor sc: 5 to 40 °C (41 to 104 °F); 95 % relative humidity, non-condensing
–20 to 60 °C (–4 to 140 °F); 95 % relative humidity, non-condensing 4 to 55 °C (39 to 131 °F); 95 % relative humidity, non-condensing (electrode)
Sample temperature 4 to 40 °C (39 to 104 °F)
Sample pressure With continuous sample preparation –30 mbar to +50 mbar at overflow vessel
Sample flow Range: 1.0 L/h–20.0 L/h
Sample quality Ultra filtrated or
comparable
Sample level Level of liquid in basin with filtration probe must be below analyzer
5
Specifications
Permissible pH value of the sample 5 to 9
Permissible hardness range <= 50 °dH 8.95 mMol/L
Permissible chlorid range <= 1000 mg/L Cl
Dimensions (Figure 1, page 7,
Figure 1, page 7)
Data and power cable lengths 2 m (80 in.) (from edge of enclosure)
Weight
Certifications CE conform. Listed to UL and CSA safety standards by TUV.
AMTAX sc: (W × H × D) 540 × 720 × 390 mm (21,25 × 28,35 × 15,35 in.) AMTAX indoor sc: (W × H × D) 540 × 720 × 370 mm (21,25 × 28,35 × 14.5 in.)
AMTAX sc: Approx. 31 kg, without Filter Probe sc and without chemicals AMTAX indoor sc: Approx. 29 kg, without Filter Probe sc and without chemicals
6
Specifications
Figure 1 Instrument dimensions AMTAX sc
7
Specifications
[1.57 in]
850 mm
[33.15 in]
720 mm
[28.35 in]
540 mm
[21.26 in]
370 mm
[14.57 in]
Figure 2 Instrument dimensions AMTAX indoor sc
8

Section 2 General Information

2.1 Safety information

Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this manual.

2.1.1 Use of hazard information

DANGER Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION Indicates a potentially hazardous situation that may result in minor or moderate injury.

2.1.2 Precautionary labels

This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories and all auxiliary items for proper disposal.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or electrocution exists.
Important Note: Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.
Note: Information that supplements points in the main text.
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol, if noted on the instrument, will be included with a danger or caution statement in the manual.
This symbol, if noted on the product, indicates the need for protective eye wear.
This symbol, when noted on the product, identifies the location of the connection for Protective Earth (ground).
This symbol, when noted on the product, identifies the location of a fuse or current limiting device.
This symbol, when noted on the product, indicated that the marked item can be hot and should not be touched without care.
9
General Information
18-32 kg (39.7-70.5 lbs)
This symbol, when noted on the product, identifies a risk of chemical harm and indicates that only individuals qualified and trained to work with chemicals should handle chemicals or perform maintenance on chemical delivery systems associated with the equipment.
This symbol, when noted on the product, indicated the presence of devices sensitive to Electro-static Discharge (ESD) and indicated that care must be taken to prevent damage with the equipment.
When carrying or transporting the instrument/instrument components and if the total weight is more than 18 kg, make sure that suitable lifting equipment is used and/or that the instrument/instrument components are carried by 2 people.

2.1.3 Change instrument labels

2.2 Product overview

Several safety labels (3 in the analytical section) are applied to the instrument. If necessary, apply the correct language label over the existing safety labels.
The AMTAX sc (Figure 3, Figure 4) measures the ammonium ions present in treated aqueous solutions (waste water, process water and surface water). The measured value is displayed in mg/L NH
-N on the controller. The AMTAX sc must be used in
4
combination with the sc1000 controller. The sc1000 controller is used to configure, power and output the measured values.
Conversion formula: NH
-N : NH
4
+
= 1 : 1.288
4
The AMTAX sc can operate using single or dual channel modes. Operation with the Filter Probe sc is single channel only. The sc analyzer can be converted from a single channel operation to dual channel operation. Contact the manufacturer for more information.
10
General Information
Dual channel operation is only possible with continuous sample preparation, e.g. FILTRAX or Ultrafiltration. Sample preparation and filtration must be provided before installing the analysis instrument.
1 LED for operating state. Refer to
Table 9, page 57 for more
information.
Figure 3 AMTAX sc enclosure
2 Door lock 3 Rating plate with model number,
serial number, voltage and frequency information and power consumption information
1 LED for operating state. Refer to
Table 9, page 57 for more
information.
Figure 4 AMTAX indoor sc enclosure
2 Door lock 3 Rating plate with model number,
serial number, voltage and frequency information and power consumption information
11
General Information
12

Section 3 Installation

DANGER Only qualified personnel should conduct the tasks described in this section of the manual.
CAUTION The enclosure may tip forwards if it has not been fixed in place. Only open the enclosure if the enclosure is properly mounted.
DANGER Potential danger in the event of contact with chemical/biological materials. Handling chemical samples, standards and reagents can be dangerous. Familiarize yourself with the necessary safety procedur es and t he corr ect handling of the chemicals before the work and r ead and fo llow all relevant safety data sheets.
Normal operation of this instrument may involve the use of hazardous chemicals or biologically harmful samples.
Observe all cautionary information printed on the original solution containers and safety data sheet prior to their use.

3.1 Basic installation overview

Dispose of all consumed solutions in accordance with national regulations and laws.
Select the type of protective equipment suitable to the concentration and quantity of the dangerous material at the respective work place.
1. Unpack the instrument (section 3.2).
2. Mount the instrument (section 3.3, page 14).
3. Remove any transport locks (section 3.4.2, page 19).
4. Install the collecting tray and the humidity sensor (section 3.4.3,
page 21 and section 3.4.4, page 22).
5. Determine the appropriate installation option (section 3.4.5,
page 23).
6. Mount the Filter Probe sc or Filtrax, if necessary. Refer to the appropriate manual for more information.
7. Connect the Filter Probe sc or Filtrax to the AMTAX sc, if necessary. Refer to section 3.5.3, page 26 for the Filter Probe sc. Refer to the Filtrax manual for more information.
8. Connect the Drain Heating Connection, if necessary.
9. Make all plumbing connections (Appendix A Plumbing and
Connection Options, page 77).
10. Install the reagents and prepare the Electrode and Electrolyte (section 3.6, page 29 and section 3.7.1, page 31).
11. Connect the AMTAX sc to the sc1000 controller to supply power to the system (section 3.8, page 33).
12. Connect the data network (section 3.9, page 34).
13
Installation

3.2 Unpack the instrument

3.3 Mechanical installation

CAUTION Pay attention to the weight (approximately 31 kg) of the instrument. Do not try to carry the instrument without assistance. Use only suitable lifting tackle for transport.
Open the shipping container while it is on its end and then slide the analyzer out of the cardboard. The items supplied will vary depending on the order. Standard items supplied for a minimal configuration include:
AMTAX sc and user manual
Collecting tray
Initial set of reagent and cleaning solutions
2 standard solutions and electrolyte / membrane caps
Fastening bracket and angle bracket
Accessories for tubing and flow-through variant
Plug set

3.3.1 Mount the instrument

Select a suitable place to install the instrument. Plan the mechanical installation before positioning posts or drilling holes. Refer to Figure 1, page 7, Figure 2, page 8 for instrument dimensions.
Make sure that the fastening has sufficient load bearing capacity (approximately 160 kg). The wall plugs must be selected and approved to suit the properties of the wall.
Plan cable and tubing routes to avoid sharp bends and tripping hazards.
When connecting two analyzers (e. g. for measuring two parameters with a FILTRAX or Ultrafiltration), plan where the instruments are to be installed and consider the length of the heated drain tubing (2 m).
The AMTAX sc can be mounted in three different ways:
Wall Mount (section 3.3.1.1),
Rail Mount; refer to the instruction sheet supplied with the
Rail Mounting Hardware.
Stand Mount; refer to the instruction sheet supplied with the Rail Mounting Hardware.
3.3.1.1 Wall mount
14
Refer to Figure 5, Figure 6 and following instructions to mount the analyzer to a wall.
1. Align and install the fastening bracket to the wall.
2. Attach the angle bracket to the instrument using the
supplied screws.
3. Slide the bottom of the enclosure onto the fastening bracket.
Installation
4. Attach the enclosure to the fastening bracket.
5. Attach the angle bracket on the enclosure to the wall.
Figure 5 Bracket dimensions for wall mounting
15
Installation
Figure 6 Wall mounting the analyzer
1 Socket head cap screw, M5 X 8 (2X) 4 Screw, customer supplied 2 Washer, M5 (4X) 5 Socket head cap screw, M5 X 40 (2X) 3 Angle bracket 6 Fastening bracket
16

3.4 Initial instrument setup

3.4.1 Open the enclosure

Installation
DANGER To reduce the risk of electrica l shock, mak e sure tha t no water can enter the enclosure or come into contact with circuit boards.
CAUTION The enclosure may tip forwards if it has not been fixed in place. Only open the enclosure if the enclosure is properly mounted.
1. Unlock the instrument (item 4, Figure 7, item 3, Figure 8).
2. Open the side latches and release the door catch.
3. Open the door and secure the door using the hook or
completely remove the door.
Figure 7 Open the enclosure AMTAX sc
1 Latches 3 Pocket for user manual 2 Door hook 4 Lock with key
17
Installation
Figure 8 Open the enclosure AMTAX indoor sc
1 Latches 3 Lock with key 2 Door hook
18

3.4.2 Remove the shipping transport locks

Prior to system start-up, the shipping transport locks must be removed from the sc analyzer.
CAUTION The enclosure may tip forwards if it has not been fixed in place. Only open the enclosure if the enclosure is properly mounted.
Important note: The electrode insulation for the measuring unit is
NOT a transport lock. Do NOT remove the cover on the measuring unit.
1. Open the enclosure door and secure with the door hook.
2. Remove the transport lock on the analyzer panel (Figure 9).
Note: If the instrument is operated with a Filter Probe sc, it is equipped with an internal compressor.
3. Remove the cable tie and pull out the compressor transport lock to the left (Figure 10).
Note: Keep the transport locks for transporting and storage.
Installation
1 Transport lock
Figure 9 Remove the analyzer panel transport locks
19
Installation
Figure 10 Compressor transport lock removal
1 Compressor transport lock 4 Protective cover for compressor 2 Cable tie 5 Fan locking screw 3 Compressor 6 Fan
1
The compressor, compressor transport lock and cable tie only apply to sc analyzers that operate using the Filter Probe sc.
1
20

3.4.3 Installation of the collecting tray

CAUTION The enclosure may tip forwards if it has not been fixed in place. Only open the enclosure if the enclosure is properly mounted.
1. Open the enclosure door and secure with the door hook.
2. Slide the collecting tray into the bottom of the enclosure
Installation
(Figure 11).
1 Collecting tray
Figure 11 Installation of the collecting tray
21
Installation

3.4.4 Connect the humidity sensor

CAUTION The enclosure may tip forwards if it has not been fixed in place. Only open the enclosure if the enclosure is properly mounted.
1. Remove power from the instrument.
2. Open the enclosure door and secure with the door hook.
3. Connect the humidity sensor wires to the terminal screws on
the collecting tray (Figure 12).
Figure 12 Connect the humidity sensor
1 Collecting tray 2 Humidity sensor connections
22

3.4.5 Determine the appropriate installation option

Before connecting tubing or cables, determine the option number that corresponds the system configuration. Refer to Table 1. Based on the option number, determine the sealing plug that will be used to seal the enclosure openings, refer Table 2.
When the option number is determined, refer to Appendix A
Plumbing and Connection Options, page 77 for installation
information.
Table 1 System configuration options
Installation
Location Filtration Drain
Filter Probe sc Any 1 1 1 1 A.5 , page 80 Filter Probe sc Heated 1 1 1 2 A.6 , page 82
OUTDOOR
INDOOR
1
For 2-parameter options, refer to Connect a 2-parameter option, page 78.
FILTRAX Heated 1 1 1 3 A.7 , page 84 FILTRAX 2 heated 2 1 2 4 A.8 , page 86 2 FILTRAX Heated 1 2 1 5 A.9 , page 88 2 FILTRAX 2 heated 2 2 2 6 A.10 , page 90 Filter Probe sc Unheated 1 1 1 7 A.11 , page 92
FILTRAX Unheated
2 FILTRAX Unheated
Continuous sample feed
2 continuous sample feeds
Unheated
Unheated
Number of
analyzers
11 18 aA.12 , page 94 21 28 bA.13 , page 96 12 19 aA.14 , page 98 22 29 bA.15 , page 100 11 110 aA.16 , page 102 22 210 bA.17 , page 104 12 111 aA.18 , page 106 22 211 bA.19 , page 108
Sample
lines
(Ch1, Ch2)
Number of
parameter
Option
1
Refer to the following section
#
for more information:
Table 2 Sealing plug types
Analysis instrument 1 Analysis instrument 2
Option
Opening 1 Opening 2 Opening 3 Opening 1 Opening 2 Opening 3
1 Plug 2 Plug 3 Plug 3
2 Plug 2 Plug 1 Plug 3
3 Plug 1 Plug 1 Plug 3
4 Plug 1 Plug 1 Plug 3 Plug 1 Plug 1 Plug 3
5 Plug 1 Plug 1 Plug 1
6 Plug 1 Plug 1 Plug 1 Plug 1 Plug 1 Plug 3
7 Plug 2 Plug 3 Plug 3
8 Plug 1 Plug 3 Plug 3 Plug 3 Plug 3 Plug 3
9 Plug 1 Plug 1 Plug 3 Plug 3 Plug 3 Plug 3
10 Plug 3 Plug 3 Plug 3 Plug 3 Plug 3 Plug 3
11 Plug 3 Plug 3 Plug 3 Plug 3 Plug 3 Plug 3
23
Installation
Figure 13 Sealing plug types
1 Sealing plug type 1 2 Sealing plug type 2 3 Sealing plug type 3

3.5 Electrical installation

DANGER High voltage wiring connections are present under the protective cover. The protective cover must remain in place unless a qualified installation technician is installing wiring for the Filter Probe sc or the heated drain.
See Figure 14 for protective cover removal.
Figure 14 Remove the protective covers
1 Protective cover for AC mains circuits (Back view) 2 Protective cover for main PCB
24

3.5.1 Electrostatic Discharge (ESD) Considerations

Important Note: To minimize hazards an d ESD risks, maintenance procedures not requiring power to the analyzer should be performed with power removed.
Delicate internal electronic components can be damaged by static electricity, resulting in degraded instrument performance or eventual failure.
The manufacturer recommends taking the following steps to prevent ESD damage to the instrument:
Before touching any instrument electronic components (such as printed circuit cards and the components on them) discharge static electricity. This can be accomplished by touching an earth-grounded metal surface such as the chassis of an instrument or pipe or a metal conduit.
To reduce static build-up, avoid excessive movement. Transport static-sensitive components in anti-static containers or packaging.
To discharge static electricity and keep it discharged, wear a wrist strap connected by a wire to earth ground.
Installation

3.5.2 Enclosure breakouts

Handle all static-sensitive components in a static-safe area. If
possible, use anti-static floor pads and work bench pads.
There are four main enclosure openings to insert tubing and cables (Figure 15).
1 Sample feed or refer to
Table 1, page 23 for
tubing options.
Figure 15 Enclosure breakouts
2 Refer to Table 1,
page 23 for tubing
options.
3 Refer to Table 1,
page 23 for tubing
options.
4 Power and data cables
25
Installation

3.5.3 Insert tubing and/or cables

1 Guide the tubing or cables through
the enclosure openings (Figure 15).
2 Push the plug from the top onto the
tubing or onto the cable.

3.5.4 Connect the Filter Probe to the analyzer

CAUTION The enclosure may tip forwards if it has not been fixed in place. Only open the enclosure if the enclosure is properly mounted.
DANGER Disconnect power from the sc analyzer at the sc1000 before removing the protective covers in the an aly ze r.
Important Note: Make sure that for the use of a filter probe, the
level of the water in which the filtration probe is submersed, is below the level of the analyzer.
1. Open the enclosure door and secure with the door hook.
2. Open the analyzer panel.
3. Remove the two screws from the protective cover and remove
the cover (item 1, Figure 14, page 24).
4. Connect the earth ground (green/yellow) wire (item 9,
Figure 16, page 28) from the Filter Probe sc to the ground
terminal (item 5, Figure 16, page 28).
3 Pull down the plug with the tubing
or the cables. Seal any unused entry with Plug #3.
26
5. Connect the power connector to the appropriate terminal
connection (items 4 and 11, Figure 16, page 28).
6. Remove the three screws securing the bottom panel cover. (item 2, Figure 14, page 24). Remove the panel.
7. Connect the data connector (item 10, Figure 16, page 28) to the main board (item 12, Figure 16, page 28).
8. Install all covers and panels.
9. Connect the white air tubing (item 8, Figure 16, page 28) from
the Filter Probe sc to the air tubing connection on the analyzer (Figure 16, page 28).
10. For sample and drain line connections refer to A.4, page79.

3.5.5 Connect the optional heated drain

CAUTION The enclosure may tip forwards if it has not been fixed in place. Only open the enclosure if the enclosure is properly mounted.
DANGER Disconnect power from the sc analyzer at the sc1000 before removing the protective covers in the analyzer.
Refer to Figure 16 and the following procedure to connect the heated drain.
1. Open the enclosure door and secure if necessary.
2. Open the analyzer panel.
3. Remove the protective cover (Figure 14, page 24).
4. Connect the earth ground wire (green/yellow) to the ground
5. Connect the cables for the heated drain (item 6, Figure 16,
6. Connect the drain tube as described in the appropriate Option
Installation
wire terminal strip.
page 28) to the terminal block (item 3, Figure 16, page 28).
Configuration, refer to sectionA.5, page80 for more information.
7. Place the drain tube to the appropriate drain or basin.
8. Install all covers and panels.
27
Installation
Figure 16 Connect the Filter Probe sc and optional heated drain
1 Bottom panel cover 8 Filter Probe sc air tube (white) 2 Protective cover 9 Filter Probe sc ground wire 3 Heated drain (optional) power connector 10 Filter Probe sc data cable connector 4 Filter Probe sc power connector 11 Filter Probe sc power cable connector 5 Ground wire terminal strip 12 Filter Probe sc data connector 6 Heated drain power cable connector 13 Remote control input (15–30 V DC) (Refer to 7 Heated drain ground wire
section B.3 , page 112)
28

3.6 Installation of reagents

Installation
DANGER Potential danger in the event of contact with chemical/biological materials. Handling chemical samples, standards and reagents can be dangerous. Familiarize yourself with the necessary safety procedur es and t he corr ect handling of the chemicals before the work and r ead and fo llow all relevant safety data sheets.
Normal operation of this instrument may involve the use of hazardous chemicals or biologically harmful samples.
Observe all cautionary information printed on the original solution containers and safety data sheet prior to their use.
Dispose of all consumed solutions in accordance with national regulations and laws.
Select the type of protective equipment suitable to the concentration and quantity of the dangerous material at the respective work place.
CAUTION Avoid unnecessary contact with sample flows of unknown concentration. This can produce hazards due to trace chemicals, radiation or biological effects.
CAUTION The enclosure may tip forwards if it has not been fixed in place. Only open the enclosure if the enclosure is properly mounted.
Important Note: Always lay the drain tubing so that there is a
continuous fall (minimum 3°), the outlet is clear (not pressurized) and the drain tubing is not longer than 2 meters. For more information refer to Appendix A, page 77.
Important Note: Incorrect use of the reagents can damage the instrument. Carefully read the labels on the containers to make sure that there are no mistakes.
The reagents and chemicals supplied are ready to use. The reagents need to be placed in the analysis instrument and tubings connected. Refer to Table 3 to determine the correct standards.
29
Installation
Table 3 Reagents and measuring ranges
Reagent
CAL 1: Standard 1
(low) CAL 2:
Standard 2 (high)
Reagent orange BCF1009 28944-52 BCF1009 28944-52 BCF1009 28944-52 BCF1009 28944-52 Cleaning
solution
Lid color
EU US
transparent
light blue BCF1149 25147-54 BCF1011
gray LCW867 28942-46 LCW867 28942-46 LCW867 28942-46 LCW867 28942-46
gray
Measuring range 1
0.02–5 mg/L
EU US EU US EU US EU US
BCF1148 25146-54 BCF1010
Measuring range 2
0.05–20 mg/L
28941-54
(1 mg/L)
28943-54 (10 mg/L)
Measuring range 3
1–100 mg/L
BCF1020
BCF1021
28943-54 (10 mg/L)
58958-54 (50 mg/L)
Measuring range 4
10–1000 mg/L
BCF1012
BCF1013
28258-54 (50 mg/L)
28259-54
(500 mg/L)
1. Place the reagent containers in the instrument (Figure 17).
2. Insert the tubing in the reagent containers.
3. Screw the reagents to the supplied caps.
Figure 17 Chemicals and reagents in the AMTAX sc (Refer to Table 3, page 30 for more information)
1 Electrolyte solution replacement bottles 7 Sample line 2 Electrode assembly 8 Reagent 3 Fastener for membrane cap 9 Standard solution: High standard 4 Electrode panel connector 10 Standard solution: Low standard 5 Connection to cuvette 11 Cleaning solution 6 Drain
30
Loading...
+ 88 hidden pages