Key points. Hauptteile.
Puntos clave. Points clé.
Punti chiave. Pontos chaves.
Protective cap.
1
Schutzkappe
Tapón protector.
Bouchon protecteur.
Tappo protettore.
Tampa protetora.
S8 connector.
2
S8-Stecker
Conector S8.
Connecteur S8.
Connettore S8.
Conector S8.
PG 13.5 thread.
3
Montagegewinde PG 13,5
Rosca de sujeción PG 13,5.
Filetage PG 13,5.
Filettatura PG 13,5.
Rosca de fixação PG 13,5.
Reference element.
4
Referenzelement
Elemento de referencia.
Élément de référence.
Elemento di riferimento.
Elemento de referência.
Internal diaphragm.
5
Internes Diaphragma
Diafragma: interno.
Diaphragme interne.
Diaframma interno.
Diafragma interno.
Gel electrolyte.
6
Elektrolytgel
Electrolito gel.
Électrolyte gel.
Elettrolita gel.
Eletrólito gel.
Diaphragm: ceramic.
7
Diaphragma: Keramik
Diafragma: cerámico.
Diaphragme: céramic.
Diaframma: ceramico.
Diafragma: cerâmico.
Gold wire.
8
Golddraht.
Alambre de oro.
Fil en or.
Filo d’oro.
Fio de ouro.
Storage protector.
9
Lagerungsschutz
Protector de almacenamiento.
Tube protecteur de stockage.
Protettore per la conservazione.
Protetor para armazenamento.
(EU Directive 2002/96/EC)
3
Metallic electrode 53 54
Introduction
The 53 54 is a gold low maintenance
electrode for clean waters
SAFETY NOTE
Before use with hazardous substances, find out about the relevant hazards, protective measures, and what
to do in an emergency.
Electrode preparation
• Remove the storage protector
Installation
In immersion, insertion, flotation housings, etc. with PG 13,5 thread.
See Housing instruction manual.
In some cases the electrode is installed,
with the help of sockets and adapters,
directly onto pipes.
Cable. They are connected to coaxial
cables with an AS9 connector.
Calibration
See pH-meter instruction manual.
Measurements. Recommendations
• During measurement the samples must
be slightly stirred.
• The sample must cover the electrode’s
diaphragm
• After measurement, rinse the electrode
with distilled water. Dry it with soft tissue.
• Between measurements the electrode
should be immersed in aqueous solution. The most appropriate place is the
storage protector with some electrolyte.
Maintenance
• Diaphragm cleaning: use the diaphragm
cleaning solution, code LZW9600.
• Gold ring cleaning: can be ribbed with soft
abrasive, for example tooth paste.
Troubleshooting
... the pH-meter always measures the
same value.
Short circuit in the electrode, cable or
connector.
Substitute the electrode.
... fortuitous reading.
Broken cable.
Substitute the electrode.
... very slow response.
Dirt on the diaphragm and/or the metalic ring.
Reference system worn out.
Clean, regenerate or replace the
electrode.
Storage
The connector should be protected from
humidity by the protective cap
Place the electrode in the storage protector
with KCl 3M.
Life of an electrode
The life expectancy of the electrode could
be indefinite when the suitable maintenance is applied.
.
.
.
4
GB
Specifications
Measuring range ± 2000 mV
Indicator metal gold
Operating temp. 0 to 80 °C
Max. pressure 6 bar
Reference
element
Diaphragm ceramic
Electrolyte gel
Body material glass
Thread PG 13,5
Connector S8
Min. immersion
depth
Applications
Acid samples containing Fe or Cr.
Limitations
Low-conductivity solutions.
Viscous or dirty solutions able to clog quickly
the diaphragm.
encapsulated
Ag/AgCl crystals
17 mm
CERTIFICATE OF QUALITY
The attached electrode 53 54 has passed
the quality test and follows the next specifications:
• Deviation < ± 10 mV in redox standard.
• Response time in redox standard < 20 s.
5