Key points. Hauptteile.
Puntos clave. Points clé.
Punti chiave. Pontos chaves.
Fixed cable (standard 3m).
1
Festes Kabel (Standard 3 m).
Cable fijo (estándard 3m).
Câble fixe (standard 3m).
Cavo fisso (standard 3m).
Cabo fixo (padrão 3m).
PG 13.5 thread.
2
Montagegewinde PG 13,5.
Rosca de sujeción PG 13,5.
Filetage PG 13,5.
Filettatura PG 13,5.
Rosca de fixação PG 13,5.
Reference element.
3
Referenzelement.
Elemento de referencia.
Élément de référence.
Elemento di riferimento.
Elemento de referência.
Gel electrolyte.
4
Elektrolytgel.
Electrolito gel.
Électrolyte gel.
Elettrolita gel.
Eletrólito em gel.
Diaphragm: ceramic.
5
Diaphragma: Keramik.
Diafragma: cerámico.
Diaphragme: céramique.
Diaframma: ceramico.
Diafragma: cerâmico.
Sensitive glass membrane.
6
Sensitive Glasmembran.
Membrana sensible.
Membrane sensible.
Membrana sensibile.
Membrana sensível de vidro.
Storage protector.
7
Lagerungsschutz.
Protector de almacenamiento.
Tube protecteur de stockage.
Protettore per la conservazione.
Protetor para armazenamento.
(EU Directive 2002/96/EC)
3
pH electrode 53 38
Introduction
The 53 38 is a pH electrode for general
use. Low maintenance electrode, as it
does not require monitoring of electrolyte level.
SAFETY NOTE
Before use with hazardous substances, find out about the relevant hazards, protective measures, and what
to do in an emergency.
Electrode preparation
• Remove the storage protector
• Verify that there are not air bubbles
inside the membrane
removed by shaking the electrode as a
clinical thermometer.
Installation
In immersion, insertion, flotation housings,
etc. with PG 13,5 thread.
See Housing instruction manual.
In some cases the electrode is installed,
with the help of sockets and adapters,
directly onto pipes.
Calibration
See pH-meter instruction manual.
Measurements. Recommendations
See pH-meter instruction manual.
Ensure that the electrode’s specifications
are adapted to the process.
Sample must cover the diaphragm
Maintenance
• Deposits forming on the electrodes,
caused by continual exposure of the sensors to a dirty or contaminated medium, is
one of the most frequent causes of problems when measuring pH. When possible
it’s recommended to use a housing with
automatic sensor cleaning system.
• When an electrode gets contaminated,
the simple rinsing with distilled water is
not suffi cient. The most suitable cleaning solution is one which acts in a more
selective way over the contamination
(look for regeneration solutions).
• Re-hydration. A “dry” electrode re-
covers by immersing it in diluted HCl
solution for several hours. It is recommended to read the information supplied with the electrode’s regenerating
solutions.
Troubleshooting
... the meter always shows the same pH
value.
Short circuit in the electrode or cable.
Replace the electrode.
... shifting of pH values.
Reference system contamination.
Replace the electrode.
... unstable reading.
Diaphragm blockage.
Clean, regenerate, replace the electrode.
... impossible to perform calibration.
.
. They can be
.
4
GB
... very slow response.
Dirt on the diaphragm or the membrane.
De-hydrated membrane.
Electrode ageing or wear.
Clean, regenerate, replace the electrode.
Storage
Place the electrode in the storage protector with KCl 3M.
Life of an electrode
The average life expectancy of a pH process electrode oscillates between 6 or 12
months, depending on the working conditions (in standard applications).
Electrodes exposed to high temperatures
do not last as long. Very alkaline mediums
also reduces the life span of electrodes.
If an electrode has too short life span,
less than three months, ask us about it.
Probably there is a more suitable electrode for the application.
Specifications
Measuring range 0 to 14 pH
Operating temp. 0 to 80 °C
Max. pressure 2 bar
Reference element Ag/AgCl wire
Diaphragm ceramic
Electrolyte gel
Body material
Thread PG 13,5
Connector —
Min. immersion
depth
interior glass /
ext. polycarbonate
17 mm
Applications
Swimming pools, drinking waters.
Aqueous samples in general.
Limitations
Low conductivity, dirty or viscous solutions.
Products with colloids or solids in suspension. Samples containing substances which
react with the silver ion.
CERTIFICATE OF QUALITY
The attached electrode 53 38 has passed
the quality test and follows the next specifications:
• Asymmetry potential < ± 15 mV.
• Sensitivity, pH 4 to 7 (at 25 ˚C) > 98%.
• Response time, pH 4 to 7 < 20 s.
5