KCl + glycerine
2
Key points. Hauptteile. Puntos clave. Points clé. Punti chiave. Pontos chaves.
MP-4 connector (IP 67).
1
MP-4-Stecker (IP 67).
Conector MP-4 (IP 67)
Connecteur MP-4 (IP 67).
Connettore MP-4 (IP 67).
Conector MP-4 (IP 67).
PG 13.5 thread.
2
Montagegewinde PG 13,5.
Rosca de sujeción PG 13,5.
Filetage PG 13,5.
Filettatura PG 13,5.
Rosca de fixação PG 13,5.
Reference element.
3
Referenzelement.
Elemento de referencia.
Élément de référence.
Elemento di riferimento.
Elemento de referência.
Internal diaphragm.
4
Internes Diaphragma.
Diafragma: interno.
Diaphragme interne.
Diaframma interno.
Diafragma interno.
Solid electrolyte.
5
Festelektrolyt.
Electrolito sólido.
Électrolyte solide.
Elettrolita solido.
Eletrólito sólido.
Pt 1000 temperature sensor.
6
Temperatursensor Pt 1000.
Sensor de temperatura, Pt 1000.
Capteur de température, Pt 1000.
Sensore di temperatura, Pt 1000.
Sensor de temperatura, Pt 1000.
Diaphragm: open circular.
7
Diaphragma: offener Ring.
Diafragma: abierto circular.
Diaphragme: ouvert circulaire.
Diaframma: aperto circolare.
Diafragma: aberto circular.
Sensitive glass membrane.
8
Sensitive Glasmembran.
Membrana sensible.
Membrane sensible.
Membrana sensibile.
Membrana sensível de vidro.
Storage protector.
9
Lagerungsschutz.
Protector de almacenamiento.
Tube protecteur de stockage.
Protettore per la conservazione.
Protetor para armazenamento.
(EU Directive 2002/96/EC)
3
pH electrode 53 37 T
Introduction
The 53 37 T is a low maintenance pH electrode for waste water.
SAFETY NOTE Before use with hazardous substanc­es, find out about the relevant haz­ards, protective measures, and what to do in an emergency.
Electrode preparation
• Remove the storage protector
• Verify that there are not air bubbles
inside the membrane removed by shaking the electrode as a clinical thermometer.
Installation
In immersion, insertion, flotation hous­ings, etc. with PG 13,5 thread.
See Housing instruction manual.
This electrode is not recommended for in­stallation on pipe. Cable. They are connected to coaxial ca­bles with an MP-4 connector.
Calibration
See pH-meter instruction manual.
Measurements. Recommendations
See pH-meter instruction manual. Ensure that the electrode’s specifications are adapted to the process. Sample must cover the diaphragm
Maintenance
• Deposits forming on the electrodes,
caused by continual exposure of the sen­sors to a dirty or contaminated medium, is one of the most frequent causes of prob­lems when measuring pH. When possible it’s recommended to use a housing with automatic sensor cleaning system.
• When an electrode gets contaminated,
the simple rinsing with distilled water is not suffi cient. The most suitable clean­ing solution is one which acts in a more selective way over the contamination.
Do not use for cleaning solutions with pH<2. They provoke rapid alteration of the solid electrolyte.
Troubleshooting
... the meter always shows the same pH
value. Short circuit in the electrode or cable.
Substitute the cable or the elec­trode.
... shifting of pH values.
Reference system contamination.
Replace the electrode
... unstable reading.
Reference system contamination.
Clean, regenerate, replace the elec­trode.
... impossible to perform calibration. ... very slow response.
Dirt on the membrane. De-hydrated membrane. Electrode ageing or wear.
Clean, regenerate, replace the elec­trode.
.
. They can be
.
4
GB
Storage
Place the electrode in the storage protec­tor with KCl + glycerine.
Life of an electrode
The average life expectancy of a pH proc­ess electrode oscillates between 6 or 12 months, depending on the working con­ditions. Electrodes exposed to high temperatures do not last as long. Very alkaline mediums also reduces the life span of electrodes.
If an electrode has too short life span, less than three months, ask us about it. Probably there is a more suitable elec­trode for the application.
Specifications
Measuring range 2 to 14 pH Operating temp. 0 to 80 °C Max. pressure 2 bar Reference element cartridge Diaphragm open circular Electrolyte solid Body material glass Thread PG 13,5 Connector MP-4 Temperature sensor Pt 1000 Min. immersion depth 12 mm
Applications
Dirty water, with mud, waste water, etc.
Limitations
Solutions with pH < 2, very clean water, distilled and de-mineralised water. Temperature > 80ºC. Pressure > 2 bar.
CERTIFICATE OF QUALITY
The attached electrode 53 37 T has passed the quality test and follows the next specifications:
• Asymmetry potential < ± 15 mV.
• Sensitivity, pH 4 to 7 (at 25 ˚C) > 98%.
• Response time, pH 4 to 7 < 20 s.
• Temperature error
0.5 ºC.