KCl 3M
2
Key points. Hauptteile. Puntos clave. Points clé. Punti chiave. Pontos chaves.
MP-4 connector (IP 67).
1
MP-4-Stecker (IP 67).
Conector MP-4 (IP 67)
Connecteur MP-4 (IP 67).
Connettore MP-4 (IP 67).
Conector MP-4 (IP 67).
PG 13.5 thread.
2
Montagegewinde PG 13,5.
Rosca de sujeción PG 13,5.
Filetage PG 13,5.
Filettatura PG 13,5.
Rosca de fi xação PG 13,5.
Reference element.
3
Referenzelement.
Elemento de referencia.
Élément de référence.
Elemento di riferimento.
Elemento de referência.
Internal diaphragm.
4
Internes Diaphragma.
Diafragma: interno.
Diaphragme interne.
Diaframma interno.
Diafragma interno.
Gel electrolyte.
5
Elektrolytgel.
Electrolito gel.
Électrolyte gel.
Elettrolita gel.
Eletrólito em gel.
Pt 1000 temperature sensor.
6
Temperatursensor Pt 1000.
Sensor de temperatura, Pt 1000.
Capteur de température, Pt 1000.
Sensore di temperatura, Pt 1000.
Sensor de temperatura, Pt 1000.
Ring diaphragm of porous PTFE.
7
Diaphragma aus porösem PTFE.
Diafragma anular de PTFE poroso.
Diaphragme annulaire PTFE poreux.
Diaframma anulare di PTFE poroso.
Diagrama anelar de PTFE poroso.
Sensitive glass membrane.
8
Sensitive Glasmembran.
Membrana sensible.
Membrane sensible.
Membrana sensibile.
Membrana sensível de vidro.
Storage protector.
9
Lagerungsschutz.
Protector de almacenamiento.
Tube protecteur de stockage.
Protettore per la conservazione.
Protetor de armazenamento.
3
pH electrode 53 34 T
Introduction
The 53 34 T is a pH electrode with built-in Pt 1000 A.T.C. Low maintenance electrode, as it does not require monitoring of electrolyte levels.
SAFETY NOTE Before use with hazardous substanc­es, find out about the relevant haz­ards, protective measures, and what to do in an emergency.
Electrode preparation
• Remove the storage protector
• Verify that there are not air bubbles inside
the membrane by shaking the electrode as a clinical ther­mometer.
Installation
In immersion, insertion, flotation hous­ings, etc. with PG 13,5 thread.
See Housing instruction manual.
In some cases the electrode is installed, with the help of sockets and adapters, directly onto pipes. Cable. They are connected to coaxial ca­bles with an MP-4 connector.
Calibration
See pH-meter instruction manual.
Measurements. Recommendations
See pH-meter instruction manual. Ensure that the electrode’s specifications are adapted to the process. Sample must cover the diaphragm
Maintenance
• Deposits forming on the electrodes,
caused by continual exposure of the sen­sors to a dirty or contaminated medium, is one of the most frequent causes of prob­lems when measuring pH. When possible it’s recommended to use a housing with automatic sensor cleaning system.
• When an electrode gets contaminated,
the simple rinsing with distilled water is not suffi cient. The most suitable cleaning solution is one which acts in a more selec­tive way over the contamination (look for regeneration solutions).
Re-hydration. A “dry” electrode recovers
by immersing it in diluted HCl solution for several hours. It is recommended to read the information supplied with the elec­trode’s regenerating solutions.
. They can be removed
.
.
Troubleshooting
... the meter always shows the same pH
value. Short circuit in the electrode or cable. Substitute the cable or the electrode.
... shifting of pH values.
Reference system contamination.
Replace the electrode
... unstable reading.
Diaphragm blockage.
Clean, regenerate, replace the elec­trode.
4
GB
... impossible to perform calibration. ... very slow response.
Dirt on the diaphragm or the mem­brane. De-hydrated membrane. Electrode ageing or wear.
Clean, regenerate, replace the elec­trode.
Storage
Place the electrode in the storage protector
with KCl 3M.
Life of an electrode
The average life expectancy of a pH process electrode oscillates between 6 or 12 months, depending on the working conditions. Electrodes exposed to high temperatures do not last as long. Very alkaline mediums also reduces the life span of electrodes.
If an electrode has too short life span, less than three months, ask us about it. Probably there is a more suitable elec­trode for the application.
Specifications
Measuring range 0 to 14 pH Operating temp. 0 to 100 °C Max. pressure 5 bar Reference element cartridge
Diaphragm
Electrolyte gel Body material glass Thread PG 13,5 Connector MP-4 Temperature sensor Pt 1000 Min. immersion depth 12 mm
Applications
Waste water, galvanising industry, osmosis processes, demineralized water, dirty or viscous samples , etc.
ring of porous
PTFE
CERTIFICATE OF QUALITY
The attached electrode 53 34 T has passed the quality test and follows the next speci­fi cations:
• Asymmetry potential < ± 15 mV.
• Sensitivity, pH 4 to 7 (at 25 ˚C) > 98%.
• Response time, pH 4 to 7 < 20 s.
• Temperature error
0.5 ºC.
5