Hach ULTRATURB sc User Manual

Page 1
DOC023.53.03231
ULTRATURB sc
basic/plus/seawater
User Manual
07/2017, Edition 7
© HACH Company, 2007–2010, 2017. All rights reserved. Printed in Germany
Page 2
Page 3
Table of contents
Section 1 Specifications .............................................................................................................................................. 5
1.1 Dimensions................................................................................................................................................................ 6
Section 2 General information ..................................................................................................................................... 7
2.1 Safety information...................................................................................................................................................... 7
2.1.1 Use of hazard information ................................................................................................................................ 7
2.1.2 Precautionary labels ........................................................................................................................................ 7
2.2 Applications ............................................................................................................................................................... 8
2.3 Measuring principle ................................................................................................................................................... 8
2.4 Handling .................................................................................................................................................................... 8
2.5 Items supplied ........................................................................................................................................................... 9
2.6 Function check .......................................................................................................................................................... 9
Section 3 Installation....................................................................................................................................................11
3.1 Assembling .............................................................................................................................................................. 11
3.1.1 Connectors..................................................................................................................................................... 12
3.2 Instrument layout ..................................................................................................................................................... 13
3.3 Connecting sensor cable ......................................................................................................................................... 14
Section 4 Operation..................................................................................................................................................... 15
4.1 Operating the sc controller ...................................................................................................................................... 15
4.2 Sensor setup ........................................................................................................................................................... 15
4.3 Sensor data logger .................................................................................................................................................. 15
4.4 Menu structure......................................................................................................................................................... 15
4.4.1 SENSOR DIAG .............................................................................................................................................. 15
4.4.2 SENSOR SETUP ........................................................................................................................................... 16
4.5 Calibration with standard solution............................................................................................................................ 18
4.6 Verify using dry standard CVM................................................................................................................................ 19
4.7 Zero point setting..................................................................................................................................................... 19
Section 5 Maintenance ................................................................................................................................................ 21
5.1 Maintenance schedule............................................................................................................................................. 21
5.2 Cleaning measuring chamber.................................................................................................................................. 21
5.3 Replacing wiper profiles (only plus and seawater version)...................................................................................... 22
5.4 Replacing desiccant ................................................................................................................................................ 23
5.5 Monitoring test equipment ....................................................................................................................................... 24
5.5.1 Preparing formazine solution in accordance with DIN EN ISO 7027 ............................................................. 24
Section 6 Faults, causes, rectification ...................................................................................................................... 27
6.1 Error messages ....................................................................................................................................................... 27
6.2 Warnings ................................................................................................................................................................. 27
Section 7 Replacement parts and accessories ...................................................................................................... 29
7.1 Sensor options......................................................................................................................................................... 29
7.2 Replacement parts .................................................................................................................................................. 29
7.3 Accessories ............................................................................................................................................................. 30
Section 8 Limited warranty ......................................................................................................................................... 31
Section 9 Contact ....................................................................................................................................................... 33
Appendix A ModBUS Register Information ............................................................................................................. 35
Page 4
Table of contents
Page 5
Section 1 Specifications
Specifications are subject to change without notice.
Components
Microprocessor-controlled turbidity bypass sensor ULTRATURB sc for very low to medium turbidities with comprehensive self-diagnostics
Measuring technique 90° infrared pulse scattered light technique in accordance with DIN EN ISO 7027
Measuring range
0.0001–1000 FNU (TE/F, NTU, FTU) can be programmed as required (0.0001–250 EBC = 2500 ppm SiO2)
Resolution 0.0001–0.9999 / 1.00–9.99 / 10.0–99.9 / 100–1000 FNU
Precision ±0.008 FNU or ±1 % of the measured value (0–10 FNU)
Reproducibility ±0.003 FNU or ±0.5 % of the measured value (0–2 FNU)
Technique variation coefficient 1 % in accordance with DIN 38402
Response time 1–60 s (can be programmed as required)
Air bubble compensation Physical-mathematical
Calibration
Permanently set in the factory (Validation using formazine, StablCal, dry standard CVM)
Sample flow rate Min. 0.2 L/min, max. 1 L/min, max. 6 bar (at 20 °C (at 68 °F))
Sample temperature Max. 50 °C (Max. 122 °F)
Salt content of the probe (only for seawater)
Tested up to 65 g/L
Ambient temperature +2 °C to +40 °C (+ 36 °F to +104 °F)
Sample connection Tubing (ID 13 mm) or fixed connection (PVC system component pipework)
Automatic cleaning of the measuring chamber (only for plus and seawater)
Automatic wiper cleaning, time controlled and as required
Materials
Measuring window: quartz Measuring chamber: Noryl GFN2 Wiper axle: stainless steel 1.4571 Wiperarm (seawater): titanium alloy Wiper profile: silicone
Housing
Enclosure rating: IP 65 Plastic housing ASA
Inspection interval
Once a year, service contract on request with warranty extension of five years; please request a quote
Dimensions (W × H × D) 250 × 240 × 110 mm
Mass Approx. 1.5 kg
User maintenance
basic: 2 h / month, typical plus and seawater: 0.5 h / month, typical
Page 6
Specifications
1.1 Dimensions
Fig. 1 ULTRATURB sc dimensions
105.15 mm [4.14 in]
221 mm
[8.7 in]
216 mm
[8.5 in]
233,4 mm
[9.19 in]
111 mm
[4.37 in]
85 mm
[3.35 in]
Ø 5.4 mm
[0.21 in]
Ø 5.4 mm
[0.21 in]
Page 7
Section 2 General information
2.1 Safety information
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
To ensure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this manual.
2.1.1 Use of hazard information
DANGER Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION Indicates a potentially hazardous situation that may result in minor or moderate injury.
Important note: Information that requires special emphasis.
Note: Information that supplements points in the main text.
2.1.2 Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol, if noted on the instrument, will be included with a danger or caution statement in the manual.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories and all auxiliary items for proper disposal.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or electrocution exists.
Page 8
General information
2.2 Applications
ULTRATURB sc bypass sensors are innovative precision turbidity measuring instruments developed using the latest technical findings. ULTRATURB sc bypass sensos are developed primary for waste and drinking water applications. The very robust seawater version is developed especially for seawater applications with high salt concentrations (e.g. drinking water abstraction from seawater, fish farming or seawater aquariums). Turbidities in the range of FNU (NTU) 0.0001
1000 are measured and displayed
alphanumerically using a controller. This high resolution facilitates precision differentiation, even for ultra clear liquids.
A pulsed, long-life IR radiation source (LED) ensures that costs are minimised.
ULTRATURB sc turbidity sensors are designed in accordance with DIN EN ISO 7027 and are permanently calibrated before leaving the factory. All key data is set to practical standard values.
The instruments are immediately ready for use after connection to the power supply and the supply of sample water. The measuring range and all data outputs are modified to suit requirements using menus on the controller.
All optical and electronic assemblies are installed in housings that are physically very strong and proof against water jets.
ULTRATURB sc turbidity sensors are available in the classic basic version, the convenient plus version and the robust seawater version.
The measuring chambers of the ULTRATURB plus sc and of the ULTRATURB seawater sc also have automatic wiper cleaning that reliably prevents soiling of the optical systems as it starts to build up and significantly reduces the user maintenance. Cleaning is performed at an interval set to suit local requirements. The cleaning intervals are adjusted to suit the expected level of soiling during commissioning. The user can straightforwardly change the interval at a later date. The number of cleaning operations already performed can be checked on the SENSOR SETUP menu using the COUNTER command.
2.3 Measuring principle
The nephelometric scattered light technique is a method for the determination of the very low to medium turbidities in liquids that is comparable worldwide.
Using this method, the light scattered sideways by the turbidity particles is measured over an angle of 90°. The low acquisition limit for this method makes it possible to even reliably differentiate turbidity particles in distilled water. This technique is defined in DIN EN ISO 7027. ULTRATURB sc sensors are designed in accordance with the requirements in this standard.
2.4 Handling
The sensor contains high-quality optical and electronic assemblies. For this reason attention is to be paid to ensuring that the sensor is not subjected to any hard mechanical knocks. There are no components that can be maintained by the user inside the sensor or the display unit – except the manual cleaning of the measuring chamber and the replacement of the wiper profiles and the desiccant.
Page 9
General information
2.5 Items supplied
ULTRATURB sc sensor
Connecting cable (length as per order)
User Manual
Factory test certificate
Accessory set LZP816
Wiper set (for 4 changes) LZV275
(only plus and seawater version)
2.6 Function check
After unpacking, all components should be checked for any transport damage and a short function check performed prior to installation.
For this purpose the sensor is connected to the display unit and the display unit plugged into the mains. Shortly after the display unit is plugged in, the display is activated and the instrument switches to the display of measurements. Here the measured value taken in air is meaningless.
Note: It is only possible to check the operative zero point using very high purity water!
If no messages appear in the lower part of the display, the function check is then complete.
Page 10
10
General information
Page 11
11
Section 3 Installation
DANGER Installation may only be carried out by qualified experts in accordance with all local safety regulations.
3.1 Assembling
Select an appropriate installation location for the instrument to guarantee
safe installation
safe operation
problem-free maintenance
Plan how to lay cables and hoses and their path in advance. Lay the hoses, data cables and power cables without any bends. Only original replacement and accessory parts recommended by the manufacturer may be used.
Make sure the mount has a sufficient bearing capacity. The dowels must be selected and authorized according to the condition of the wall. The manufacturer shall accept no liability if the instrument is installed incorrectly.
Install the instrument in a horizontal position.
Fig. 2 Assembling
Page 12
12
Installation
3.1.1 Connectors
Important Note: The instrument can be damaged if the connectors and/or the union nuts are attached too tightly. Only hand-tighten the union nuts for the hose mounts. If necessary, hold the connectors with a tool (SW 22).
The measuring instrument can be integrated into the measurement and analysis process using tubing (ID 13 mm) or a fixed connection (PVC system component pipework, compatible with 1" union nut).
3.1.1.1 Selection of diaphragm plate
With samples that tend to emit gases, bubbles may form in the measuring chamber. This causes extremely strongly fluctuating measurement values. In the event of strongly fluctuating measurement values, replace the diaphragm plate in the drain connector.
Note: Observe the flow quantity and flow direction of the sample provided in the technical data.
Table 1 : Connect o r s
External thread Pipe thread in accordance with DIN ISO 228
large G1A small G½A
Table 2: Preselect the diaphragm plate
Flow quantity/pressure Diaphragm plate
large large hole
medium medium hole
small small hole
Page 13
Installation
13
3.2 Instrument layout
Fig. 3 Instrument layout ULTRATURB sc
* Included in the accessory set (connection) LZP816
1. Screw top, HRR195 8. Diaphragm plate* (if necessary) HRS185: 1.2 mm HRS186: 2 mm HRS187: 3 mm
2. Measuring chamber sealing plug, HAD087 9. Tubing nipple* (Ø 13 mm), HXA072
3. Clutch wheel, BVK001 10. Union nut* 1“, ERM033
4. Plug socket, HRB151 11. Drain
5. Sealing ring (pre-assembled), EZD166 12. Feed
6. Union (pre-assembled), HXA087 13. Sensor cable socket
7. O-ring*, EZD114 14. Wiper holder
(only plus version: casting compound brown, BHH211) (only seawater version: casting compound black/blue, BHH233)
Page 14
14
Installation
3.3 Connecting sensor cable
CAUTION Always lay cables and hoses so that they do not pose a tripping risk.
1. Unscrew the protective caps from the controller socket and the cable plug and retain them.
2. Pay attention to the guide in the plug and slide the plug into the socket.
3. Tighten the nut by hand.
Note: Extension cables are available in various lengths (refer to Section 7, Replacement
parts and accessories, page 29).
Maximum cable length 100 m (328 ft).
Fig. 4 Connection of the sensor plug to the controller
Fig. 5 Sensor connector pin assignment
Number Description Cable colour
1 +12 VDC brown 2 Ground black 3Data (+) blue 4 Data (–) white 5 Screen Screen (grey) 6 Notch
6
1 2
5
3
4
Page 15
15
Section 4 Operation
4.1 Operating the sc controller
The sensor can be operated with all sc controllers. Prior to using the sensor, familiarise yourself with the principle of operation of your controller. Learn how to navigate in the menus and run appropriate functions.
4.2 Sensor setup
When you connect the sensor for the first time, the serial number of the sensor is displayed as the sensor name. You can change the sensor name as follows:
1. Open the MAIN MENU.
2. Choose SENSOR SETUP and accept.
3. Choose the related sensor and accept.
4. Choose CONFIGURE and accept.
5. Choose EDITED NAME and accept.
6. Edit the name and accept to return to the SENSOR SETUP menu.
In the same way complete your system configuration using the following commands:
MEAS UNITS
CLEAN. INTERVAL
RESPONSE TIME
LOGGER INTERVAL
RESOLUTION
SET DEFAULTS
4.3 Sensor data logger
A data memory and event memory per sensor are available via the sc controller. While measured data are saved in the data memory at stipulated intervals, the event memory collects numerous events such as configuration changes, alarms and warning conditions. Both the data memory and the event memory can be read out in CSV format. For information on how you can download the data, please see the controller manual.
4.4 Menu structure
4.4.1 SENSOR DIAG
Note: In Section 7, page 29 you will find, along with a list of all error messages and warnings, also a description of all the actions necessary.
ERROR LIST
Possible error messages: WIPER POS, LED C., GAIN TOO HIGH
WARNING LIST
Possible warnings: PROFILE COUNTER, MOIST, USER CAL INTERVAL
Page 16
16
Operation
4.4.2 SENSOR SETUP
WIPE Initiates a wiping action
VERIFY
Verification using CVM module
Verification with dry standard
CALIBRATE
STANDARD
Selection as per CAL. CONFIG OUTPUT MODE
Calibration with standard solution
Procedure during calibration, menu based
OFFSET
Selection as per CAL. CONFIG OUTPUT MODE
Zero point setting Zero point setting, menu based
CAL. FAKTORS
FAC. STANDARD 0.50 to 2.00
OFFSET -0.100 to +0.100 TRBFNU
CAL. CONFIG
OUTPUT MODE
HOLD ACTIVE
TRANSFER CHOICE
Behaviour of the outputs during calibration or zero point setting
CAL. INTERVAL Adjustable from 0 to 365 days
SET CAL DEFLT Reset to the default calibration
CONFIGURE
EDITED NAME
SET DEFAULTS Instrument number
Up to 16-character name possible
MEAS UNITS
mg/L, FNU, NTU, TE/F, EBC
SET DEFAULTS FNU
CLEAN. INTERVAL
SET DEFAULTS 12 h
10 min, 20 min, 30 min, 2 h, 6 h, 12 h, 10:00
RESPONSE TIME
SET DEFAULTS 15 s
0 to 60 s
LOGGER INTERVAL
SET DEFAULTS 10 min
1-30 min
RESOLUTION
SET DEFAULTS
0.xxx
0.xxx or 0.xxxx (<1 FNU)
SET DEFAULTS Confirmation prompt
Reset to default configuration for all menu commands listed above.
Page 17
Operation
17
TEST/MAINT
PROBE INFO
ULTRATURBsc Instrument name EDITED NAME SERIAL NUMBER RANGE 0.001 ... 1000 FNU MODEL NUMBER Item no. sensor SOFTWARE VERS Sensor software DRIVER VERS
CAL. DATA
CAL. DATE Date of the last calibration FACTOR Default setting 1.00
OFFSET
Default setting
0.000 TRBFNU
COUNTER
TOTAL TIME PROFILE MOTOR
MAINT. PROC.
CLN. MEAS CHAMB
OUTPUT MODE information
Procedure during cleaning
REPLACE PROFILE
OUTPUT MODE information
Procedure during wiper change
WIPE RESOLUTION Initiates a wiping action
SIGNALS
AVER.: Average value S. VAL.: Individual measured value M: Measured level R: Reference level Q: Quotient M/R MOIST Relative humidity in %
OUTPUT MODE
ACTIVE HOLD TRANSFER
CHOICE
Behaviour of the instrument outputs in the MAINT. PROC. menu
Page 18
18
Operation
4.5 Calibration with standard solution
1. Open the MAIN MENU.
2. Choose SENSOR SETUP and accept.
3. Choose the related sensor and accept.
4. Choose CALIBRATE and accept.
5. Choose STANDARD and accept.
6. Shut-off the feed and accept TURN OFF SAMPLE INLET.
7. Drain the measuring chamber using the lower feed union. Accept DRAIN
MEAS. CHAMBER.
8. Connect the calibration syringe to the feed union (see Figure) and add the calibration standard. Accept POUR STD INTO MEAS. CHAMBER.
9. Accept PRESS ENTER WHEN STABLE x.xxx TRBFNU.
10. Enter the concentration of the standard solution. Accept CALIBRATE
(x.xxx TRBFNU).
11. Remove the calibration syringe and accept REMOVE CALIBRATION SYRINGE.
12. Re-connect the feed and accept CONNECT SAMPLE INLET.
13. Open the feed and accept CAL READY OPEN INLET.
14. Accept OUTPUT ACTIVE.
Page 19
Operation
19
4.6 Verify using dry standard CVM
1. Open the MAIN MENU.
2. Choose SENSOR SETUP and accept.
3. Choose the related sensor and accept.
4. Choose VERIFY and accept.
5. Choose STANDARD and accept.
6. Shut-off the feed and accept TURN OFF SAMPLE INLET.
7. Drain the measuring chamber via the lower feed union and accept DRAIN
MEAS. CHAMBER.
8. Open the screw top and remove the sealing plug and wiper holder.
9. Carefully dry the measuring chamber.
10. Clean the CVM module with a cloth ( included in the accessory setof the CVM
modul).
11. Guide the CVM module into the measuring chamber. The arrow mark must point downwards! Ensure that the spring-loaded ball engages by turning slightly in the sample feed opening on the measuring chamber, see Figure). Accept INSERT CVM DRY STANDARD.
12. Accept x.xxx TRBFNU.
13. Remove the CVM module and accept REMOVE CVM MODULE .
14. Re-fit the sealing plug and wiper holder and accept REPLACE WIPER
HOLDER AND CAP.
15. Screw back on the screw top and accept TIGHTEN CAP.
16. Re-connect the feed and accept CONNECT SAMPLE INLET.
17. Open the feed and accept VER. READY OPEN INLET.
18. Accept OUTPUT ACTIVE.
4.7 Zero point setting
1. Open the MAIN MENU.
2. Choose SENSOR SETUP and accept.
3. Choose the related sensor and accept.
4. Choose CALIBRATE and accept.
5. Choose OFFSET and accept.
6. Shut off the feed and drain the measuring chamber. Connect a membrane
filter (LZV325) to the measuring chamber feed. Open the feed and accept POUR 0 STD INTO MEAS. CHAMBER.
7. Accept PRESS ENTER WHEN STABLE x.xxx TRBFNU.
8. Set the zero point and accept CALIBRATE (x.xxx TRBFNU).
9. Accept CAL READY OUTPUT ACTIVE.
Page 20
20
Operation
Page 21
21
Section 5 Maintenance
It has been possible to reduce the scope of the maintenance to be performed by the user to only a few actions. Clearly summarised in a table and described in detail in the following sections, these actions can be quickly and straightforwardly performed by qualified personnel.
5.1 Maintenance schedule
5.2 Cleaning measuring chamber
The cleanliness of the measuring chamber is crucial for correct measured results. This cylindrical optical measuring chamber has a rotating wiper with three wiper profiles (only on plus and seawater version); these prevent normal soiling of the optical system as it starts to build up. With very stubborn deposits, manual cleaning with a suitable cleaning agent (e.g. citric acid) is recommended.
CAUTION Observe safety regulations and wear safety clothing!
Safety glasses
Gloves
Overalls
1. Open the MAIN MENU.
2. Choose SENSOR SETUP and accept.
3. Choose the related sensor and accept.
4. Choose TEST/MAINT and accept.
5. Choose MAINT. PROC and accept.
6. Choose CLN. MEAS CHAMB and accept.
7. Shut-off the feed and accept TURN OFF SAMPLE INLET.
8. Drain the measuring chamber via the lower feed union and accept DRAIN
MEAS. CHAMBER.
Maintenance task Interval
Clean measuring chamber:
Dependent on the substances contained
in the water
Replace wiper profile (only plus and seawater version)
After 1200 cycles
Replace desiccant Every 2 years
Monitor test equipment (CVM dry standard)
Every 2 years
(factory test with certification)
Check zero point
Dependent on the substances contained
in the water
Check gradient At least once a year
Page 22
22
Maintenance
9. Remove the screw top and the sealing plug for the measuring chamber and
accept OPEN MEAS. CHAMBER.
10. Remove the wiper holder and accept REMOVE WIPER HOLDER (only for plus and for seawater version).
11. Clean the measuring chamber and accept CLN. MEAS CHAMB.
12. Re-fit the wiper holder and accept REPLACE WIPER HOLDER (only for plus
and seawater version).
13. Seal the measuring chamber again with sealing plug and screw top and accept CLOSE MEAS. CHAMBER.
14. Open the sample feed and accept TURN OFF SAMPLE INLET. An automatic wipe is performed.
5.3 Replacing wiper profiles (only plus and seawater version)
The life of the wiper profiles is, on the one hand, dependent on the number of cleaning actions performed, and on the other hand is also dependent on the type of deposits to be removed. Thus the life of the wiper profiles varies from case to case.
The wiper profiles supplied with the instrument are sufficient to the cover requirements for an average year.
1. Open the MAIN MENU.
2. Choose SENSOR SETUP and accept.
3. Choose the related sensor and accept.
4. Choose TEST/MAINT and accept.
5. Choose MAINT. PROC and accept.
6. Choose REPLACE PROFILE and accept.
7. Shut-off the feed and accept TURN OFF SAMPLE INLET.
8. Drain the measuring chamber via the lower feed union and accept DRAIN
MEAS. CHAMBER.
9. Remove the screw top and the sealing plug for the measuring chamber and accept OPEN MEAS. CHAMBER.
10. Remove the wiper holder and accept REMOVE WIPER HOLDER.
11. Clean the measuring chamber and accept CLN. MEAS CHAMB.
12. Renew the wiper profiles and accept REPLACE PROFILE.
13. Re-fit the wiper holder and accept REPLACE WIPER HOLDER.
14. Seal the measuring chamber again with sealing plug and screw top and
accept CLOSE MEAS. CHAMBER.
15. Open the sample feed and accept TURN OFF SAMPLE INLET. An automatic wipe is performed.
Page 23
Maintenance
23
5.4 Replacing desiccant
The ULTRATURB sc turbidity sensor is continuously subject to a wet environment. To provide protection against the moisture, there are 2 generously sized desiccant bags close to the optical and electronic components. To maintain this protection, the manufacturer recommends replacement or regeneration every 2 years.
To replace the desiccant, open the housing and replace the two desiccant bags, 4 units (see Figure 6, item 1) and 0.5 units (see Figure 6, item 2) as per the figure.
Figure 6 Replacing desiccant
1. Desiccant bag with 4 units, LZX304 2. Desiccant bag with 0.5 units, LZX303
1
2
Page 24
24
Maintenance
5.5 Monitoring test equipment
As part of a quality assurance system for test equipment monitoring, it is necessary to check the instrument calibration; this can be performed using a formazine standard series in accordance with DIN EN ISO 7027.
You can either make up the formazine solution yourself as per the instructions that follow 5.5.1 Preparing formazine solution in accordance with DIN EN ISO 7027 or procure it from the manufacturer (LCW 813 or Stabl Cal).
1. To check the instrument calibration, first thoroughly rinse out the measuring chamber using distilled water.
2. Prepare a turbidity series using formazine reference solutions as per DIN EN ISO 7027; the measuring series should correspond to the measuring range selected. For example, in the measuring range FNU 0–5.0 chose reference solutions with FNU 0.5–1.0–2.0–3.0–4.0.
Note: If there are any variations in the measurements, it is recommendable to measure a new standard preparation prior to changing the instrument calibration! Experience has shown that the variation in the formazine preparation can be greater than the possible drift in the ULTRATURB instrument.
When checking the instruments as per EN 450901 or GLP we therefore recommend recording the date in an instrument log and checking for any long­term drift. Checks on the operation of the ULTRATURB can in practice also be made by means of parallel measurements with turbidity photometers designed in accordance with DIN and ISO.
5.5.1 Preparing formazine solution in accordance with DIN EN ISO 7027
Chemicals
Chemicals of "analysis" purity are used. The chemicals must be stored in hard glass bottles.
Water
Place a membrane filter, pore size 0.1 µm (for bacteriological
experiments), for 1 hour in 100 mL of distilled water.
Filter 250 mL water through this filter and discard the water.
Then filter 500 mL of distilled water twice through the same filter and use
this water for making up the standard solution.
Formazine (C
2H4N2
) parent solution
CAUTION Hydrazine sulphate is toxic and possibly carcinogenic
Dissolve 10.0 g of hexamethylentetramine (C
6H12N4
) in water and top up
to 100 mL (solution A).
Dissolve 1 g of hydrazine sulphate (N
2H6SO4
) in water and top up to
100 mL (solution B).
Mix 5 mL of solution A with 5 mL of solution B.
After 24 hours at 25° C (±3 °C) (77 °F (± 5.4 °F)) top up this solution with
water to 100 mL.
Page 25
Maintenance
25
The turbidity of this parent solution is 400, stated in Formazine Attenuation Units (FAU) or Formazine Nephelometric Units (FNU). This solution can be stored in a dark place at 25 °C (±3 °C) (77 °F (± 5.4 °F)) for around 4 weeks.
Formazine reference solution
Dilute the parent solution with the aid of pipettes and measuring flasks to produce reference solutions in the required range. These solutions can only be stored for a very limited period.
Page 26
26
Maintenance
Page 27
27
Section 6 Faults, causes, rectification
6.1 Error messages
Possible sensor errors are displayed by the controller.
6.2 Warnings
Possible warning messages are displayed by the controller.
Table 3: Error messages
Error displayed Cause Rectification
NONE Correct operation WIPER POS. Wiper not is correct starting position Initiate wiper function, call service LED C. LED faulty Call service
GAIN TOO HIGH
Incorrect adjustment Incorrect calibration standard Measuring chamber soiled LED faulty
Clean measuring chamber, perform calibration, call service
Table 4: Warnings
Error displayed Cause Rectification
NONE Correct operation PROFILE COUNTER Counter elapsed Change wiper profiles MOIST Moisture > 10 % relative humidity Replace desiccant CAL. INTERVAL Counter elapsed Perform calibration
Page 28
28
Faults, causes, rectification
Page 29
29
Section 7 Replacement parts and accessories
7.1 Sensor options
Description Cat. no.
ULTRATURB basic sc turbidity bypass sensor without cable LPV415.99.01001 ULTRATURB basic sc turbidity bypass sensor with 0.35 m (1.1 ft.) cable LPV415.99.11001 ULTRATURB basic sc turbidity bypass sensor with 1 m (3.3 ft.) cable LPV415.99.81001 ULTRATURB basic sc turbidity bypass sensor with 5 m (16.40 ft.) cable LPV415.99.21001 ULTRATURB basic sc turbidity bypass sensor with 10 m (32.81 ft.) cable LPV415.99.31001 ULTRATURB basic sc turbidity bypass sensor with 15 m (49.21 ft.) cable LPV415.99.41001 ULTRATURB basic sc turbidity bypass sensor with 20 m (65.62 ft.) cable LPV415.99.51001 ULTRATURB basic sc turbidity bypass sensor with 30 m (98.43 ft.) cable LPV415.99.61001 ULTRATURB basic sc turbidity bypass sensor with 50 m (164.04 ft.) cable LPV415.99.71001 ULTRATURB plus sc turbidity bypass sensor without cable LPV415.99.00001 ULTRATURB plus sc turbidity bypass sensor with 0.35 m (1.1 ft.) cable LPV415.99.10001 ULTRATURB plus sc turbidity bypass sensor with 1 m (3.3 ft.) cable LPV415.99.80001 ULTRATURB plus sc turbidity bypass sensor with 5 m (16.40 ft.) cable LPV415.99.20001 ULTRATURB plus sc turbidity bypass sensor with 10 m (32.81 ft.) cable LPV415.99.30001 ULTRATURB plus sc turbidity bypass sensor with 15 m (49.21 ft.) cable LPV415.99.40001 ULTRATURB plus sc turbidity bypass sensor with 20 m (65.62 ft.) cable LPV415.99.50001 ULTRATURB plus sc turbidity bypass sensor with 30 m (98.43 ft.) cable LPV415.99.60001 ULTRATURB plus sc turbidity bypass sensor with 50 m (164.04 ft.) cable LPV415.99.70001 ULTRATURB seawater sc turbidity bypass sensor without cable LPV415.99.02001 ULTRATURB seawater sc turbidity bypass sensor with 0.35 m (1.1 ft.) cable LPV415.99.12001 ULTRATURB seawater sc turbidity bypass sensor with 1 m (3.3 ft.) cable LPV415.99.82001 ULTRATURB seawater sc turbidity bypass sensor with 5 m (16.40 ft.) cable LPV415.99.22001 ULTRATURB seawater sc turbidity bypass sensor with 10 m (32.81 ft.) cable LPV415.99.32001 ULTRATURB seawater sc turbidity bypass sensor with 15 m (49.21 ft.) cable LPV415.99.42001 ULTRATURB seawater sc turbidity bypass sensor with 20 m (65.62 ft.) cable LPV415.99.52001 ULTRATURB seawater sc turbidity bypass sensor with 30 m (98.43 ft.) cable LPV415.99.62001 ULTRATURB seawater sc turbidity bypass sensor with 50 m (164.04 ft.) cable LPV415.99.72001
7.2 Replacement parts
Description Cat. no.
Set of wiper profiles (for 4 changes) LZV275 Desiccant bag 0.5 U LZX303 Desiccant bag 4 U LZX304 Accessory set (connection) LZP816
Page 30
30
Replacement parts and accessories
7.3 Accessories
Description Cat. no.
Extention cable 1 m (3.3 ft.) 6122400 Extention cable 5 m (16.40 ft.) LZX848 Extention cable 10 m (32.81 ft.) LZX849 Extention cable 15 m (49.21 ft.) LZX850 Extention cable 20 m (65.62 ft.) LZX851 Extention cable 30 m (98.43 ft.) LZX852 Formazine turbidity standard LCW813 Turbidity calibration set for liquid standard LZV451 Set of filters for zero point calibration (0.2 mm membrane filter incl. connecting material) LZV325
0.2 µm membrane filter (without accessories) EXF014 CVM Certified Verification Module (dry standard) approx. 0.6 FNU LZV414.00.00000 CVM Certified Verification Module (dry standard) approx. 1.5 FNU LZV414.00.10000 CVM Certified Verification Module (dry standard) approx. 6 FNU LZV414.00.20000 CVM Certified Verification Module (dry standard) approx. 15 FNU LZV414.00.30000 CVM Certified Verification Module (dry standard) approx. 25 FNU LZV414.00.40000
Page 31
31
Section 8 Limited warranty
Hach Company warrants its products to the original purchaser against any defects that are due to faulty material or workmanship for a period of one year from date of shipment unless otherwise noted in the product manual.
In the event that a defect is discovered during the warranty period, Hach Company agrees that, at its option, it will repair or replace the defective product or refund the purchase price excluding original shipping and handling charges. Any product repaired or replaced under this warranty will be warranted only for the remainder of the original product warranty period.
This warranty does not apply to consumable products such as chemical reagents; or consumable components of a product, such as, but not limited to, lamps and tubing.
Contact Hach Company or your distributor to initiate warranty support. Products may not be returned without authorization from Hach Company.
Limitations
This warranty does not cover:
Damage caused by acts of God, natural disaster, labor unrest, acts of war (declared or undeclared), terrorism, civil strife or acts of any governmental jurisdiction
Damage caused by misuse, neglect, accident or improper application or installation
Damage caused by any repair or attempted repair not authorized by Hach Company
Any product not used in accordance with the instructions furnished by Hach Company
Freight charges to return merchandise to Hach Company
Freight charges on expedited or express shipment of warranted parts or
product
Travel fees associated with on-site warranty repair
This warranty contains the sole express warranty made by Hach Company in connection with its products. All implied warranties, including without limitation, the warranties of merchantability and fitness for a particular purpose, are expressly disclaimed.
Some states within the United States do not allow the disclaimer of implied warranties and if this is true in your state the above limitation may not apply to you. This warranty gives you specific rights, and you may also have other rights that vary from state to state.
This warranty constitutes the final, complete, and exclusive statement of warranty terms and no person is authorized to make any other warranties or representations on behalf of Hach Company.
Limitation of Remedies
The remedies of repair, replacement or refund of purchase price as stated above are the exclusive remedies for the breach of this warranty. On the basis of strict liability or under any other legal theory, in no event shall Hach Company be liable for any incidental or consequential damages of any kind for breach of warranty or negligence.
Page 32
32
Limited warranty
Page 33
33
Section 9 Contact
HACH Company World Headquarters
P.O. Box 389 Loveland, Colorado 80539-0389 U.S.A. Tel (800) 227-HACH (800) -227-4224 (U.S.A. only) Fax (970) 669-2932 orders@hach.com www.hach.com
Repair Service in the United States:
HACH Company Ames Service 100 Dayton Avenue Ames, Iowa 50010 Tel (800) 227-4224 (U.S.A. only) Fax (515) 232-3835
Repair Service in Canada:
Hach Sales & Service Canada Ltd. 1313 Border Street, Unit 34 Winnipeg, Manitoba R3H 0X4 Tel (800) 665-7635 (Canada only) Tel (204) 632-5598 Fax (204) 694-5134 canada@hach.com
Repair Service in Latin America, the Carib­bean, the Far East, Indian Subcontinent, Africa, Europe, or the Middle East:
Hach Company World Headquarters, P.O. Box 389 Loveland, Colorado, 80539-0389 U.S.A. Tel +001 (970) 669-3050 Fax +001 (970) 669-2932 intl@hach.com
HACH LANGE GMBH
Willstätterstraße 11 D-40549 Düsseldorf Tel. +49 (0)2 11 52 88-320 Fax +49 (0)2 11 52 88-210 info@hach-lange.de www.hach-lange.de
HACH LANGE LTD
Pacific Way Salford GB-Manchester, M50 1DL Tel. +44 (0)161 872 14 87 Fax +44 (0)161 848 73 24 info@hach-lange.co.uk www.hach-lange.co.uk
HACH LANGE LTD
Unit 1, Chestnut Road Western Industrial Estate IRL-Dublin 12 Tel. +353(0)1 46 02 5 22 Fax +353(0)1 4 50 93 37 info@hach-lange.ie www.hach-lange.ie
HACH LANGE GMBH
Hütteldorferstr. 299/Top 6 A-1140 Wien Tel. +43 (0)1 9 12 16 92 Fax +43 (0)1 9 12 16 92-99 info@hach-lange.at www.hach-lange.at
HACH LANGE
Rorschacherstrasse 30 a CH­9424 Rheineck Tel. +41 (0)71 886 91 11 Fax +41 (0)71 886 91 66 info@hach-lange.ch www.hach-lange.ch
HACH LANGE FRANCE S.A.S.
8, mail Barthélémy Thimonnier Lognes
F-77437 Marne-La-Vallée cedex 2 Tél. +33 (0)8 20 20 14 14 Fax +33 (0)1 69 67 34 99 info@hach-lange.fr www.hach-lange.fr
HACH LANGE SA
Motstraat 54 B-2800 Mechelen Tél. +32 (0)15 42 35 00 Fax +32 (0)15 41 61 20 info@hach-lange.be www.hach-lange.be
DR. LANGE NEDERLAND B.V.
Laan van Westroijen 2a NL-4003 AZ Tiel Tel. +31(0)344 63 11 30 Fax +31(0)344 63 11 50 info@hach-lange.nl www.hach-lange.nl
HACH LANGE APS
Åkandevej 21 DK-2700 Brønshøj Tel. +45 36 77 29 11 Fax +45 36 77 49 11 info@hach-lange.dk www.hach-lange.dk
HACH LANGE AB
Vinthundsvägen 159A SE-128 62 Sköndal Tel. +46 (0)8 7 98 05 00 Fax +46 (0)8 7 98 05 30 info@hach-lange.se www.hach-lange.se
HACH LANGE S.R.L.
Via Riccione, 14 I-20156 Milano Tel. +39 02 39 23 14-1 Fax +39 02 39 23 14-39 info@hach-lange.it www.hach-lange.it
HACH LANGE S.L.U.
Edif. Arteaga Centrum C/Larrauri, 1C- 2ª Pl. E-48160 Derio/Vizcaya Tel. +34 94 657 33 88 Fax +34 94 657 33 97 info@hach-lange.es www.hach-lange.es
HACH LANGE LDA
Av. do Forte nº8 Fracção M P-2790-072 Carnaxide Tel. +351 214 253 420 Fax +351 214 253 429 info@hach-lange.pt www.hach-lange.pt
HACH LANGE SP.ZO.O.
ul. Opolska 143 a PL-52-013 Wrocław Tel. +48 (0)71 342 10-83 Fax +48 (0)71 342 10-79 info@hach-lange.pl www.hach-lange.pl
HACH LANGE S.R.O.
Lešanská 2a/1176 CZ-141 00 Praha 4 Tel. +420 272 12 45 45 Fax +420 272 12 45 46 info@hach-lange.cz www.hach-lange.cz
HACH LANGE S.R.O.
Roľnícka 21 SK-831 07 Bratislava – Vaj nory Tel. +421 (0)2 4820 9091 Fax +421 (0)2 4820 9093 info@hach-lange.sk www.hach-lange.sk
HACH LANGE KFT.
Vöröskereszt utca. 8-10. H-1222 Budapest XXII. ker.
Tel. +36 (06)1 225 7783 Fax +36 (06)1 225 7784 info@hach-lange.hu www.hach-lange.hu
HACH LANGE S.R.L.
Str. Căminului nr. 3 Sector 2 RO-021741 Bucureşti Tel. +40 (0) 21 205 30 03 Fax +40 (0) 21 205 30 17 info@hach-lange.ro www.hach-lange.ro
HACH LANGE
8, Kr. Sarafov str. BG-1164 Sofia Tel. +359 (0)2 963 44 54 Fax +359 (0)2 866 15 26 info@hach-lange.bg www.hach-lange.bg
HACH LANGE SU ANALİZ SİSTEMLERİ LTD.ŞTİ.
Hilal Mah. 75. Sokak Arman Plaza No: 9/A TR-06550 Çankaya/ANKARA Tel. +90 (0)312 440 98 98 Fax +90 (0)312 442 11 01 bilgi@hach-lange.com.tr www.hach-lange.com.tr
Page 34
34
Contact
HACH LANGE D.O.O.
Fajfarjeva 15 SI-1230 Domžale Tel. +386 (0)59 051 000 Fax +386 (0)59 051 010 info@hach-lange.si www.hach-lange.si
ΗΑCH LANGE E.Π.Ε.
Αυλίδος 27 GR-115 27 Αθήνα Τηλ . +30 210 7777038 Fax +30 210 7777976 info@hach-lange.gr www.hach-lange.gr
HACH LANGE E.P.E.
27, Avlidos str GR-115 27 Athens Tel. +30 210 7777038 Fax +30 210 7777976 info@hach-lange.gr www.hach-lange.gr
HACH LANGE D.O.O.
Ivana Severa bb 42 000 Varaždin Tel. +385 (0) 42 305 086 Fax +385 (0) 42 305 087 info@hach-lange.hr www.hach-lange.hr
HACH LANGE MAROC SAR­LAU
Villa 14 – Rue 2 Casa Plai­sance Quartier Racine Extension MA-Casablanca 20000 Tél. +212 (0)522 97 95 75 Fax +212 (0)522 36 89 34 info-maroc@hach-lange.com www.hach-lange.ma
Page 35
35
Appendix A ModBUS Register Information
Table 5: Sensor ModBUS Registers
Group Name Register # Data Type Length R/W Description
measurement 40001 Float 2 R diplayed measurement value unit 40003 Unsigned Integer 1 R/W unit parameter 40004 Unsigned Integer 1 R/W parameter reserved 40005 Unsigned Integer 1 R/W reserved correction 40006 Float 2 R/W correction offset 40008 Float 2 R/W offset reserved 40010 Unsigned Integer 1 R/W reserved cleaning_interval 40011 Unsigned Integer 1 R/W cleaning interval led_current 40012 Unsigned Integer 1 R/W LED current wiper state 40013 Unsigned Integer 1 R/W wiper state resp time 40014 Unsigned Integer 1 R/W response time drv_struct_ver 40015 Unsigned Integer 1 R driver structure version drv_firmw_ver 40016 Unsigned Integer 1 R driver firmware version drv_cont_ver 40017 Unsigned Integer 1 R driver content version location 40018 String 8 R/W location profile 40026 Integer 2 R profile counter motor_cycles 40028 Integer 2 R motor cycles operating_hours 40030 Integer 2 R operating hours service_counter 40032 Integer 2 R service counter profile reset val 40034 Integer 2 R/W profile reset val service reset val 40036 Integer 2 R/W service reset val des_measurement 40038 Float 2 R/W desired measurement value meas_single_value 40040 Float 2 R measurement single value M 40042 Float 2 R m R 40044 Float 2 R r ouotient 40046 Float 2 R quotient humidity_main 40048 Float 2 R humidity - main cal_date 40050 Time 2 R calibration time and date user_cal_date 40052 Time 2 R user calibration time and date fac_meas_0 40054 Float 2 R factor meas. 0 fac_meas_1 40056 Float 2 R factor meas. 1 fac_meas_2 40058 Float 2 R factor meas. 2 fac_meas_3 40060 Float 2 R factor meas. 3 fac_ref_0 40062 Float 2 R factor ref 0 fac_ref_1 40064 Float 2 R factor ref 1 fac_quo_q 40066 Float 2 R factor quotient q dist_q 40068 Float 2 R distortion light q fac_quo_m 40070 Float 2 R factor quotient m dist_m 40072 Float 2 R distortion light m prg_vers 40074 String 4 R program version ser_no 40078 Integer 2 R serial number process 40080 Unsigned Integer 1 R/W process register menu 40081 Unsigned Integer 1 R menu state
Page 36
36
ModBUS Register Information
user_cal_int 40082 Integer 1 R/W user cal int cal_out_cfg 40083 Integer 1 R/W cal. Output mode meas_mode 40084 Integer 1 R/W measure mode meas_val_format 40085 Integer 2 R measurement format meas_prec 40087 Integer 1 R/W measurement precesion logger_int 40088 Integer 1 R/W logger interval service output 40089 Integer 1 R/W service output mode
Table 5: Sensor ModBUS Registers (continued)
Group Name Register # Data Type Length R/W Description
Page 37
ModBUS Register Information
37
Page 38
38
ModBUS Register Information
Loading...