Hach TU5300 sc, TU5400 sc Basic User Manual

Page 1
DOC023.52.90477
TU5300 sc/TU5400 sc
03/2017, Edition 4
Basic User Manual
Page 2
Page 3
Table of contents
Specifications on page 3 Calibration on page 21
Installation on page 10 Maintenance on page 21
User navigation on page 20 Troubleshooting on page 27
Operation on page 20
Additional information
An expanded user manual is available on the manufacturer's website. Videos on how to install, operate and do maintenance and troubleshooting on the TU5300 sc and the TU5400 sc turbidimeters are available on the TU5 Series Turbidimeters playlist at
http://www.youtube.com/user/hachcompany.
Specifications
Specifications are subject to change without notice.
Specification Details
Measurement method Nephelometry with scattered light collected at a 90-degree angle to the incident light
and 360 degrees around the sample vial
Primary compliance method
DIN EN ISO 7027
Enclosure Material: ASA Luran S 777K / RAL7000, TPE RESIN Elastocon® STK40,
Thermoplastic Elastomer TPS-SEBS (60 Shore) and stainless steel
Dimensions (W x D x H) 268 x 249 x 190 mm (10.6 x 9.8 x 7.5 in.)
Weight Instrument with the process head: 2.7 kg (6.0 lb); Instrument with the optional
automatic cleaning module: 5.0 kg (11.0 lb)
Power requirements 12 VDC, 14 VA supplied by the SC controller
Protection class III
Pollution degree 2
Installation category II
Mounting Indoor on a wall
Operating temperature 0 to 50 °C (32 to 122 °F)
Storage temperature –40 to 60 °C (–40 to 140 °F)
Humidity 5 to 95% relative humidity, non-condensing
Sensor cable length 1.6 m (5.25 ft)
1
Laser Class 1 laser product: Contains a non user-serviceable class 1 laser.
Optical light source 850 nm, maximum 0.55 mW
Fittings Sample inlet and outlet: ¼-in. OD tubing
Tubing requirements Polyethylene, polyamide or polyurethane tubing. Calibrated ¼ in. OD,
+0.03 or –0.1 mm (+0.001 or –0.004 in.)
1
Extension cables are available. The maximum sensor cable length is 50 m (164 ft) when no accessories are connected to the instrument.
English 3
Page 4
Specification Details
Measurement units TU5300 sc: NTU, FNU, TE/F, EBC or FTU; TU5400 sc: NTU, mNTU2, FNU, mFNU,
TE/F, EBC, FTU or mFTU.
Range 0 to 1000 NTU, FNU, TE/F and FTU; 0 to 250 EBC
Method detection limit TU5300 sc: 0.002 FNU at 25 °C (77 °F) ( < 0.025 FNU range); TU5400 sc:
0.0001 FNU at 25 °C (77 °F) ( < 0.025 FNU range)
Response time T90 < 30 seconds at 100 mL/min
Accuracy ± 2% or ± 0.01 FNU (the larger value) from 0 to 40 FNU
± 10% of reading from 40 to 1000 FNU based on Formazin primary standard at 25 °C (77 °F)
Linearity Better than 1% for 0 to 40 NTU based on Formazin primary standard at 25 °C
(77 °F).
Repeatability TU5300 sc: 0.002 FNU or 1% (the larger value) at 25 °C (77 °F) ( > 0.025 FNU
range); TU5400 sc: 0.0006 FNU or 1% (the larger value) at 25 °C (77 °F) ( >
0.025 FNU range)
Stray light < 0.01 FNU
Resolution 0.0001 FNU (0.0001 to 0.9999/1.000 to 9.999/10.00 to 99.99/100.0 to 1000 FNU)
Default: TU5300 sc: 0.001 FNU and TU5400 sc: 0.0001 FNU
Air bubble compensation Physical, mathematical
Sample requirements Temperature: 2 to 60 °C (35.6 to 140 °F)
Conductivity: 3000 µS/cm maximum at 25 °C (77 °F)
Flow rate3: 100 to 1000 mL/min; optimal flow rate: 200 to 500 mL/min
Pressure: 6 bar (87 psi) maximum compared to air, 2 to 40 °C (35.6 to 104 °F) sample; 3 bar (43.5 psi) maximum compared to air, 40 to 60 °C (104 to 140 °F) sample
Calibration options StablCal® or Formazin: 1-point calibration (20 FNU) for 0 to 40 FNU measurement
range, 2-point calibration (20 and 600 FNU) for 0 to 1000 FNU (full) measurement range or 2- to 6-point custom calibration for a measurement range of 0 FNU to the highest calibration point.
Verification options Glass verification rod (solid secondary standard) ≤ 0.1 NTU, StablCal or Formazin
Verification (RFID or Link2SC®)
Verification of the measurement value by comparison of the process and lab measurements with RFID or Link2SC.
Certifications CE compliant; US FDA accession number: 1420492-xxx. This product complies with
IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50. Australian RCM.
Warranty 1 year (EU: 2 years)
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website.
2
1 mNTU = 0.001 NTU
3
For the best results, operate the instrument at a flow rate of 200 mL/min when the maximum particle size is 20 µm. For larger particles (150 µm maximum), the best flow rate is 350 to 500 mL/min.
4 English
Page 5
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
C A U T I O N
Indicates a potentially hazardous situation that may result in minor or moderate injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
This symbol indicates the need for protective eye wear.
This symbol indicates a laser device is used in the equipment.
This symbol indicates that the marked item can be hot and should not be touched without care.
English 5
Page 6
This symbol identifies a risk of chemical harm and indicates that only individuals qualified and trained to work with chemicals should handle chemicals or perform maintenance on chemical delivery systems associated with the equipment.
This symbol indicates radio waves.
Class 1 laser product
D A N G E R
Personal injury hazard. Never remove covers from the instrument. This is a laser-based instrument and the user risks injury if exposed to the laser.
Class 1 laser product, IEC60825-1:2014, 850 nm, maximum 0.55 mW
Location: Rear of the instrument.
Conforms to U.S. regulations 21 CFR 1040.10 and 1040.11 in accordance with Laser Notice No. 50.
Location: Rear of the instrument.
This instrument is a Class 1 Laser product. There is invisible laser radiation when the instrument is defective and when the instrument lid is open. This product complies with EN 61010-1, "Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use" and with IEC/EN 60825-1, "Safety of Laser Products" and with 21 CFR 1040.10 in accordance with Laser Notice No. 50. Refer to the labels on the instrument that supply laser information.
RFID module
Instruments with the optional RFID module receive and transmit information and data. The RFID module operates with a frequency of 13.56 MHz.
RFID technology is a radio application. Radio applications are subject to national conditions of authorization. The use of instruments with the optional RFID module is currently permitted in the regions that follow:
EU (European Union) countries, EFTA (European Free Trade Association) countries, Turkey, Serbia, Macedonia, Australia, Canada, US, Chile, Ecuador, Venezuela, Mexico, Brazil, South Africa, India, Singapore, Argentina, Columbia, Peru and Panama
The use of instruments with the optional RFID module outside of the above-mentioned regions can violate national laws. The manufacturer reserves the right also to get authorization in other countries. In case of doubt, contact the manufacturer.
Safety information for RFID modules
W A R N I N G
Multiple hazards. Do not disassemble the instrument for maintenance. If the internal components must be cleaned or repaired, contact the manufacturer.
6 English
Page 7
W A R N I N G
Electromagnetic radiation hazard. Do not use the instrument in dangerous environments.
N O T I C E
This instrument is sensitive to electromagnetic and electromechanical interference. These interferences can have an effect on the analysis performance of this instrument. Do not put this instrument near equipment that can cause interference.
Obey the safety information that follows to operate the instrument in accordance with local, regional and national requirements.
• Do not operate the instrument in hospitals and equivalent establishments or near medical equipment, such as pace makers or hearing aids.
• Do not operate the instrument near highly flammable substances, such as fuels, highly flammable chemicals and explosives.
• Do not operate the instrument near combustible gases, vapors or dust.
• Keep the instrument away from strong vibration or shock.
• The instrument can cause interference in immediate proximity to televisions, radios and computers.
• The warranty does not cover improper use or wear.
FCC conformance for RFID
This instrument may contain a registered radio frequency identification device (RFID). Refer to
Table 1 for the Federal Communications Commission (FCC) registration information.
Table 1 Registration information
Parameter Value
FCC identification number (FCC ID) YCB-ZBA987
IC 5879A-ZBA987
Frequency 13.56 MHz
Certification
Canadian Radio Interference-Causing Equipment Regulation, IECS-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including interference that may cause
undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a
English
7
Page 8
residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:
1. Disconnect the equipment from its power source to verify that it is or is not the source of the interference.
2. If the equipment is connected to the same outlet as the device experiencing interference, connect the equipment to a different outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the interference.
5. Try combinations of the above.
Product overview
D A N G E R
Chemical or biological hazards. If this instrument is used to monitor a treatment process and/or chemical feed system for which there are regulatory limits and monitoring requirements related to public health, public safety, food or beverage manufacture or processing, it is the responsibility of the user of this instrument to know and abide by any applicable regulation and to have sufficient and appropriate mechanisms in place for compliance with applicable regulations in the event of malfunction of the instrument.
The TU5300 sc and the TU5400 sc turbidimeters are used with an SC controller to measure low­range turbidity mostly in finished drinking water applications. Refer to Figure 1.
The TU5300 sc and the TU5400 sc turbidimeters measure scattered light at an angle of 90° in a 360° radius around the axis of the incident light beam.
An optional RFID module and an automatic system check option are available4. The RFID module is shown in Figure 1. The RFID module lets process and laboratory turbidity measurements be easily compared. A description of the automatic system check option is given in the expanded user manual on the manufacturer's website.
PROGNOSYS predictive diagnostic software is available for the TU5300 sc and TU5400 sc turbidimeters. To use PROGNOSYS, connect the turbidimeter to an SC controller with PROGNOSYS.
Videos on how to install, operate and do maintenance and troubleshooting on the TU5300 sc and the TU5400 sc turbidimeters are available on the TU5 Series Turbidimeters playlist at
http://www.youtube.com/user/hachcompany.
For the accessories, refer to the expanded user manual on the manufacturer's website.
4
The RFID module and automatic system check option is only available at the time of purchase.
8 English
Page 9
Figure 1 Product overview
1 Programmable button 8 Process head
2 Status indicator light
5
9 Overflow drain
3 RFID module indicator (optional) 10 Process head (open)
4 Cleaning lid screws (3x) 11 Process head (closed)
5 Cleaning lid 12 Channels for cables
6 Sample inlet 13 Extension connector for accessories
7 Sample outlet 14 Sensor cable
Product components
Make sure that all components have been received. Refer to Figure 2. If any items are missing or damaged, contact the manufacturer or a sales representative immediately.
5
Shows the instrument status. Refer to the expanded user manual on the manufacturer's website for more information.
English 9
Page 10
Figure 2 Product components
1 TU5300 sc or TU5400 sc 6 Cleaning lid screws and washers for hot water
applications
2 Wall mount bracket (two tubing clips on bracket) 7 Vial replacement tool
3 Tubing clips 8 Flow regulator
4 Tubing clip screws, 2.2 x 6 mm 9 Service bracket
5 Mounting screws, 4 x 16 mm 10 Desiccant cartridge
Installation
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
Installation guidelines
N O T I C E
Make sure that there is a floor drain near the instrument. Examine the instrument daily for leaks.
This instrument is rated for an altitude of 3100 m (10,710 ft) maximum. Use of this instrument at an altitude higher than 3100 m can slightly increase the potential for the electrical insulation to break down, which can result in an electric shock hazard. The manufacturer recommends that users with concerns contact technical support.
Installation overview
Figure 3 shows the installation overview with no accessories and the clearances necessary. Refer to
the expanded manual on the manufacturer's website for the system overview with all of the accessories.
10
English
Page 11
Figure 3 Installation overview with no accessories
1 Service bracket 5 Sample inlet
2 Process head 6 Sample outlet
3 SC controller 7 TU5300 sc or TU5400 sc
4 Flow regulator
Wall mount
Install the instrument on a wall in a vertical position. Install the instrument so that it is level.
Install with the wall mount bracket
Refer to the illustrated steps that follow to install the instrument on a wall with the wall mount bracket. The mounting hardware to install the wall mount bracket on a wall is supplied by the user.
If a 1720D, 1720E, or FT660 instrument is replaced, remove the instrument from the wall. Then do steps 2 to 4 of the illustrated steps that follow to install the instrument on the existing hardware.
Note: When the accessories are used, the installation location of the tubing clips is different. Refer to the documentation supplied with the accessories for tubing clip installation.
English
11
Page 12
12 English
Page 13
Install directly on a wall
As an alternative, refer to the illustrated steps that follow to install the instrument directly on a wall. The mounting hardware is supplied by the user. Remove the thin, plastic film from the mounting holes on the back of the instrument.
Install the desiccant cartridge
N O T I C E
Make sure that the desiccant cartridge is installed or damage to the instrument will occur.
For initial installation, complete the steps below. For replacement, refer to the documentation supplied with the desiccant cartridge.
1. Look at the install by date on the packaging. Refer to Figure 4. Do not use if the current date is past the install by date.
2. Make sure that the indicator on the new desiccant cartridge is blue. Refer to Figure 4.
3. Install the new desiccant cartridge. Refer to the illustrated steps that follow.
English
13
Page 14
Figure 4 Examine the desiccant cartridge
1 Install by date (mm.yyyy = month and year) 2 Indicator (blue = not expired, pink = expired)
14 English
Page 15
English 15
Page 16
Replace the cleaning lid screws
N O T I C E
Do not overtighten the screws or breakage will occur. Hand tighten the screws.
If the sample temperature is 40 to 60 °C (104 to 140 °F), the cleaning lid screws will become hot. To prevent burns, replace the standard cleaning lid screws with the cleaning lid screws and washers for hot water. Refer to Figure 1 on page 9 for the location of the cleaning lid screws.
Install the service bracket
The service bracket holds the process head (or the optional automatic cleaning module) when it is not installed on the instrument.
Refer to Installation overview on page 10 to install the service bracket the correct distance from the instrument. Refer to the illustrated steps that follow to install the service bracket.
Install the flow sensor (optional)
The optional flow sensor identifies if the sample flow is within specifications. A warning shows on the controller display and the status indicator light when a no flow, low flow or high flow warning occurs.
Install the optional flow sensor. Refer to the documentation supplied with the optional flow sensor.
Install the automatic cleaning module (optional)
The automatic cleaning module cleans the inside of the process vial at a selected time interval. Install the optional automatic cleaning module. Refer to the documentation supplied with the automatic cleaning module.
Connect to an SC controller
C A U T I O N
Personal injury hazard. Do not look into the vial compartment when the instrument is connected to power.
16 English
Page 17
1. Install the latest software version on the SC controller before the instrument is connected to the SC controller. Get the latest software version from www.hach.com.
Refer to the software installation instructions supplied in the box or supplied in the software download for the SC controller.
2. Remove power to the SC controller.
3. Connect the sensor cable to the quick-connect fitting of the SC controller. Refer to Figure 5. Keep
the connector cap for later use.
4. Supply power to the SC controller.
The SC controller looks for the instrument.
5. When the SC controller finds the instrument, push enter.
On the main screen, the controller shows the turbidity value measured by the turbidimeter.
Figure 5 Connect the sensor cable to the SC controller
Plumbing
Plumb the instrument
W A R N I N G
Explosion hazard. Make sure that the drain tube is free of all obstructions. If the drain tube has a blockage or is pinched or bent, high pressure can build up in the instrument.
W A R N I N G
Personal injury hazard. The sample line contains water under high water pressure that can burn skin if hot. Qualified personnel must remove the water pressure and wear personal protective equipment during this procedure.
N O T I C E
Do not let water get in the vial compartment or instrument damage will occur. Before the process head is installed on the instrument, make sure that there are no water leaks. Make sure that all tubing is fully seated. Make sure that the vial nut is tight.
N O T I C E
Before the instrument is plumbed, make sure that the desiccant cartridge and vial are installed.
Items supplied by the user:
• Flow shutoff valve
English
17
Page 18
• Tubing
6
• Tubing cutter
1. Plumb the instrument. Refer to the illustrated steps that follow and Figure 6.
Note: To plumb the instrument with accessories, refer to the documentation supplied with the accessories.
2. If a service bracket is not installed near the instrument, do the steps that follow:
a. Do not do illustrated steps 1 and 2. b. Do illustrated steps 3 and 4 with the process head on the instrument. c. Before step 5, remove the process head from the instrument. d. Hold the process head so that it is not above the instrument or put the process head on the
optional process head holder.
e. Do illustrated steps 5 and 6. f. When there are no leaks from the process head or the flow regulator, install the process head
on the instrument (illustrated step 8).
Figure 6 Plumbing overview – no accessories
1 Service bracket 4 Flow regulator
2 Sample inlet 5 Flow shutoff valve
3 Sample outlet
6
Refer to Specifications on page 3 for the tubing requirements.
18 English
Page 19
English 19
Page 20
Set the flow rate
1. Measure the flow with the flow regulator fully open. Make sure that the flow is in the middle of the
flow specification. Refer to Specifications on page 3.
2. Slowly close the flow regulator until the flow decreases by 20 to 30%.
Note: The flow regulator causes back pressure in the tubing and decreases the quantity of bubbles that can form in the vial.
User navigation
Refer to the controller documentation for keypad description and navigation information.
Operation
Refer to the expanded user manual on the manufacturer's website to configure the instrument settings and to compare process and lab measurements.
20
English
Page 21
Calibration
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
The instrument is factory calibrated and the laser light source is stable. The manufacturer recommends that a calibration verification be done periodically to make sure that the system operates as intended. The manufacturer recommends calibration as local regulations require and after repairs or comprehensive maintenance work.
Use the optional calibration lid and a vial(s) with a StablCal standard or Formazin standard to calibrate the instrument. As an alternative, use a syringe and StablCal standard or Formazin standard to calibrate the instrument.
Refer to the expanded user manual on www.hach.com to calibrate the instrument and configure the calibration settings.
Verification
Use the optional calibration lid and a sealed-vial 10-NTU StablCal standard (or a StablCal 10 NTU standard and a syringe) to do a primary calibration verification. As an alternative, use the optional calibration lid and the optional glass verification rod (< 0.1 NTU) to do a secondary calibration verification.
Refer to the expanded user manual on www.hach.com to do a verification and configure the verification settings.
Maintenance
W A R N I N G
Burn hazard. Obey safe handling protocols during contact with hot liquids.
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
C A U T I O N
Personal injury hazard. Never remove covers from the instrument. This is a laser-based instrument and the user risks injury if exposed to the laser.
C A U T I O N
Personal injury hazard. Glass components can break. Handle with care to prevent cuts.
N O T I C E
Do not disassemble the instrument for maintenance. If the internal components must be cleaned or repaired, contact the manufacturer.
English 21
Page 22
N O T I C E
Stop the sample flow to the instrument and let the instrument become cool before maintenance is done.
To set the output behavior during maintenance, push menu and select SENSOR SETUP>TU5x00 sc>DIAG/TEST>MAINTENANCE>OUTPUT MODE.
Maintenance schedule
Table 2 shows the recommended schedule of maintenance tasks. Facility requirements and
operating conditions may increase the frequency of some tasks.
Table 2 Maintenance schedule
Task 1 to 3 months 1 to 2 years As necessary
Clean the vial on page 22
Note: The cleaning interval is dependent on the water quality.
X
Clean the vial compartment on page 24 X
Replace the vial on page 24 X
Replace the desiccant cartridge on page 26
Note: The replacement interval is dependent on the ambient humidity, ambient temperature and sample temperature.
X
7
Replace the tubing on page 27 X
Clean spills
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations.
1. Obey all facility safety protocols for spill control.
2. Discard the waste according to applicable regulations.
Clean the instrument
N O T I C E
Do not use solvents to clean the instrument.
The instrument is maintenance free. Regular cleaning is not necessary for normal operation. If the exterior of the instrument becomes dirty, wipe the instrument surfaces with a clean, moist cloth.
Clean the vial
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
When the turbidity reading shows that there is contamination in the process vial or "VIAL CLARITY" shows on the controller display, clean the vial.
7
Two years or as identified by instrument notification.
22 English
Page 23
1. Push menu.
2. Select SENSOR SETUP>TU5x00 sc>DIAG/TEST>MAINTENANCE>VIAL CLEANING.
3. Complete the steps that show on the controller display. The date the vial was cleaned is
automatically saved after the last screen shows.
4. If the optional automatic cleaning module is installed, push menu and select SETUP>TU5x00 sc>START WIPE to start the automatic cleaning process.
5. If the optional automatic cleaning module is not installed, clean the vial with the optional vial wiper if available. Refer to Clean the vial with the optional wiper on page 24.
6. If the turbidity readings do not go back to the original values, do the illustrated steps that follow to clean the vial.
Note: Hold the output values of the SC controller as necessary before the illustrated steps are done. Refer to the SC controller documentation to hold the outputs.
English 23
Page 24
Clean the vial with the optional wiper
N O T I C E
Carefully remove most of the water in the vial. Carefully put the vial wiper into the process vial so that no water spills out.
Clean the process vial with the optional vial wiper as shown in the illustrated steps that follow.
Clean the vial compartment
Clean the vial compartment only when the compartment has contamination. Make sure that the tool to clean the vial compartment has a soft surface and does not damage the instrument. Table 3 shows the options on how to clean the vial compartment.
Table 3 Cleaning options
Contaminant Options
Dust Vial compartment wiper, micro fiber cloth, lint-free cloth
Liquid, oil Cloth, water and cleaning agent
Replace the vial
N O T I C E
Keep water out of the vial compartment or instrument damage will occur. Before the automatic cleaning module is installed on the instrument, make sure that there are no water leaks. Make sure that all tubing is fully seated. Make sure that the vial nut is tight.
N O T I C E
Hold the automatic cleaning module vertically when it is installed on the instrument or the vial can break. If the vial breaks, water will get in the vial compartment and instrument damage will occur.
24 English
Page 25
N O T I C E
Do not to touch or scratch the glass of the process vial. Contamination or scratches on the glass can cause measurement errors.
Note: Make sure that no particles fall into the vial compartment.
1. Push menu.
2. Select SENSOR SETUP>DIAG/TEST>MAINTENANCE>VIAL REPLACEMENT.
3. Complete the steps that show on the controller display. The date the vial was replaced is
automatically saved after the last screen shows.
Refer to the illustrated steps that follow to replace the vial. To protect the new vial from contamination, use the vial replacement tool to install the vial.
At illustrated step 3, put the process head on its side on a flat surface if a service bracket is not installed near the instrument.
English 25
Page 26
Replace the desiccant cartridge
The controller display will show when a desiccant cartridge replacement is due. Refer to the documentation included in the desiccant cartridge bag to replace the desiccant cartridge.
26
English
Page 27
Replace the tubing
Replace the tubing when the tubing has a blockage or has damage.
Turn the flow shutoff valve to stop flow to the instrument. Then refer to Plumb the instrument on page 17 to replace the tubing.
Troubleshooting
Refer to the expanded user manual on www.hach.com for troubleshooting information.
English 27
Page 28
28 English
Page 29
Page 30

HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Tel. (970) 669-3050 (800) 227-4224 (U.S.A. only) Fax (970) 669-2932 orders@hach.com www.hach.com
HACH LANGE GMBH
Willstätterstraße 11 D-40549 Düsseldorf, Germany Tel. +49 (0) 2 11 52 88-320 Fax +49 (0) 2 11 52 88-210 info-de@hach.com www.de.hach.com
HACH LANGE Sàrl
6, route de Compois 1222 Vésenaz SWITZERLAND Tel. +41 22 594 6400 Fax +41 22 594 6499
©
Hach Company/Hach Lange GmbH, 2015–2017.
All rights reserved. Printed in Germany.
Loading...