Hach TSS sc, TSS W sc, TSS TITANIUM2 sc, TSS TITANIUM7 sc, TSS VARI scTSS XL sc User Manual

...
DOC023.52.90154
TSS sc
TSS W sc
TSS HT sc
TSS VARI sc
TSS XL sc TSS TITANIUM2 sc TSS TITANIUM7 sc
09/2018, Edition 3
3
Table of contents
Section1 Technical data ......................................................................................................................... 5
1.1 Dimensions ......................................................................................................................................... 7
Section2 General information................................................................................................................ 9
2.1 Safety information............................................................................................................................... 9
2.1.1 Hazard notices in this manual.................................................................................................... 9
2.1.2 Warning labels ........................................................................................................................... 9
2.2 Areas of application .......................................................................................................................... 10
2.2.1 TSS sc/TSS W sc: 0,001 to 9999 FNU; 0,001 to 500 g/L........................................................ 10
2.2.2 TSS HT sc: 0,001 to 9999 FNU; 0,001 to 500 g/L ................................................................... 10
2.2.3 TSS VARI sc: 0,001 9999 FNU; 0,001 up to 500 g/L............................................................... 10
2.2.4 TSS XL sc: 0,001 to 9999 FNU; 0,001 to 500 g/L.................................................................... 10
2.2.5 TSS TITANIUM2 sc/TSS TITANIUM7 sc: 0,001 to 9999 FNU; 0,001 to 500 g/L .................... 10
2.3 Measuring principle........................................................................................................................... 11
2.3.1 Turbidity according to DIN standards....................................................................................... 11
2.3.2 Measurement of solids according to plant-specific curves....................................................... 11
2.4 Handling............................................................................................................................................ 11
2.5 Scope of delivery .............................................................................................................................. 11
2.6 Function test ..................................................................................................................................... 11
Section3 Installation............................................................................................................................. 13
3.1 Installation overview of basin sensor ................................................................................................ 13
3.2 Pipe installation options for sensor installation................................................................................. 14
3.3 Connect the sensor cable ................................................................................................................. 15
Section4 Operation...............................................................................................................................17
4.1 User interface and navigation ........................................................................................................... 17
4.2 Sensor setup..................................................................................................................................... 17
4.3 Sensor data logger ........................................................................................................................... 17
4.4 Menu structure.................................................
................................................................................. 17
4.4.1 SENSOR STATUS................................................................................................................... 17
4.4.2 SENSOR setup........................................................................................................................ 18
4.5 CALIBRATE...................................................................................................................................... 20
4.5.1 Calibration of the TURBIDITY (TRB) parameter...................................................................... 21
4.5.1.1 Select the TURBIDITY (TRB) parameter ........................................................................ 21
4.5.1.2 FACTOR .........................................................................................................................21
4.5.1.3 OFFSET.......................................................................................................................... 21
4.5.1.4 1 to 3-point calibration..................................................................................................... 22
4.5.2 Calibration of the SOLID (TS) parameter................................................................................. 22
4.5.2.1 Select the SOLID (TS) parameter................................................................................... 22
4.5.2.2 FACTOR .........................................................................................................................23
4.5.2.3 1 to 3-point calibration..................................................................................................... 23
4.5.3 General information about calibration...................................................................................... 24
4.5.3.1 Clear recorded points...................................................................................................... 24
4.5.3.2 Clear a calibration point .................................................................................................. 24
4
Section5 Maintenance ..........................................................................................................................25
5.1 Maintenance schedule ......................................................................................................................25
5.2 List of wearing parts ..........................................................................................................................25
5.3 Clean the measurement windows .....................................................................................................25
5.4 Replace the wiper profile...................................................................................................................26
Section6 Troubleshooting....................................................................................................................27
6.1 Error messages.................................................................................................................................27
6.2 Warnings ...........................................................................................................................................27
Section7 Replacement parts and accessories ...................................................................................29
7.1 Replacement parts ............................................................................................................................29
7.2 Accessories.......................................................................................................................................29
Section8 Warranty and liability............................................................................................................31
AppendixA Modbus register ..................................................................................................................33
5
Section 1 Technical data
Subject to change.
Measurement
Measurement method
Combined multiple-beam alternating light technique with IR diode system and beam focus
Turbidity (TRB)
2-channel 90° scattered light measurement in accordance with DIN/EN 27027/ISO7027, wavelength = 860 nm Additional measurement value verification through eight-channel multiple-angle measurement
Solid matter (TS)
Modified absorption measurement: Eight-channel multiple-angle measurement, wavelength = 860 nm
Air-bubble compensation Software-based
Measurement value compensation
Software-based (process-adaptable)
Measuring range
Turbidity (TRB) 0,001 to 9999 FNU
Solid matter (TS) 0,001 to 500 g/L
Measurement accuracy Turbidity (TRB)
Up to 1000 FNU/NTU: < 5 % of measurement value ± 0,01 FNU/NTU
Reproducibility
Turbidity (TRB) < 3 %
Solid matter (TS) < 4 %
Response time 1 s <
T90 < 300 s (adjustable)
Calibration
Turbidity (TRB) Calibrated before shipping
Solid matter (TS) To be calibrated on site by the customer
Zero point Calibrated permanently before shipping
Environmental conditions
Pressure range
TSS sc: <
10 bar or < 100 m < 145 PSI
TSS W sc: <
6 bar or < 60 m < 87 PSI
TSS HT sc: <
10 bar or < 100 m < 145 PSI
TSS VARI sc: <
16 bar or < 160 m < 232 PSI
TSS XL sc: <
16 bar or < 160 m < 232 PSI
TSS TITANIUM2 sc: <
10 bar or < 100 m < 145 PSI
TSS TITANIUM7 sc: <
10 bar or < 100 m < 145 PSI
Flow speed Max. 3 m/s (any air bubbles created affect the measurement)
Ambient temperature
TSS sc: 0 to 60 °C, briefly 80 °C 32 to 140 °F, briefly 176 °F TSS W sc: 0 to 50 °C, briefly 70 °C 32 to 122 °F, briefly 158 °F TSS HT sc: 0 to 90 °C, briefly 95 °C 32 to 194 °F, briefly 203 °F TSS VARI sc: 0 to 80 °C, briefly 95 °C 32 to 176 °F, briefly 203 °F TSS XL sc: 0 to 80 °C, briefly 95 °C 32 to 176 °F, briefly 203 °F TSS TITANIUM2 sc: 0 to 60 °C, briefly 80 °C 32 to 140 °F, briefly 176 °F TSS TITANIUM7 sc: 0 to 60 °C, briefly 80 °C 32 to 140 °F, briefly 176 °F
Distance Sensor – wall/floor
Solid matter (TS) > 10 cm, turbidity (TRB) > 50 cm
6
Technical data
Equipment properties
Dimensions
Basin sensor: Ø × L 40 mm × 330 mm (1,57 in × 13 in) Installation sensor (TriClamp): Ø × L 40 mm × 332 mm (1,57 in × 13 in) TSS VARI sc, TSS XL sc: Ø × L 40 mm × 232 mm (1,57 in × 9,13 in)
Materials
Parts in contact with medium (For TITANIUM as stipulated in order specification)
Head: stainless steel DIN 1.4460 Sleeve, shaft, shank: stainless steel DIN 1.4571 Sapphire glass Gaskets: FKM, optional FFKM
(type HT on request)
Wipers (optional): PA (GF), TPV
TSS sc TSS W sc TSS XL sc TSS VARI sc
Sensor connection cable (permanently connected), Semoflex (PUR): 1 AWG 22/12 V DC twisted cable pair, 1 AWG 24/data twisted cable pair, shared cable screen
TSS HT sc TSS TITANIUM sc
Sensor connection cable (permanently connected), Teflon (PTFE): 1 AWG 22/12 V DC twisted cable pair, 1 AWG 22/data twisted cable pair, shared cable shield
Cable gland
TSS sc, TSS W sc, TSS HT sc, TSS VARI sc, TSS XL sc: Stainless steel 1.4305
TSS TITANIUM2 sc: Grade 2 titanium TSS TITANIUM7 sc: Grade 7 titanium
Mass
Basin sensor, installation sensor (TriClamp): Approximately 1,6 kg TSS VARI sc, TSS XL sc: Approximately 1,5 kg
Cable length 10 m (32,81 ft), max. 100 m (328 ft) with extension cable
Other
Inspection interval Upon request once per year, service contract with warranty extension to 5 years
Maintenance requirements 1 hour/month, typically
Compliance CE compliant
7
Technical data
1.1 Dimensions
Figure 1 Dimensions
1 Basin sensor 3 TSS XL sc
2 Installation sensor (TriClamp) 4 TSS VARI sc
8
Technical data
9
Section 2 General information
2.1 Safety information
Please read this entire manual before unpacking, setting up, or operating this equipment. Pay attention to all hazard and warning notices. Failure to do so could result in serious injury to the operator or damage to the equipment.
To prevent damage to or impairment of the device's protection equipment, the device may only be used or installed as described in this manual.
2.1.1 Hazard notices in this manual
Note: Information that supplements points in the main text.
2.1.2 Warning labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed.
NOTICE
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.
DANGER
Explosion hazard. This product is not suitable for use in hazardous areas.
DANGER
Indicates a potentially or imminently hazardous situation that, if not avoided, can result in death or serious injury.
WARNING
Indicates a potentially or imminently dangerous situation that, if it is not avoided, can lead to death or to serious injuries.
CAUTION
Indicates a possible dangerous situation that can have minor or moderate injuries as the result.
NOTICE
Indicates a situation that, if it is not avoided, can lead to damage to the device. Information that requires special emphasis.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
This symbol may be found on an enclosure or barrier within the product and indicates a risk of electric shock and/or death by electrocution.
10
General information
2.2 Areas of application
Figure 2 Overview
2.2.1 TSS sc/TSS W sc: 0,001 to 9999 FNU; 0,001 to 500 g/L
Extremely accurate turbidity and solid sensors made from stainless steel for color-independent measurement of highly concentrated sludges.
This sensor is available in a basin version (with/without wiper) or an installation version (TriClamp) (with/without wiper) (refer to 1, 2, 3 and 4 in Figure 2 Overview).
2.2.2 TSS HT sc: 0,001 to 9999 FNU; 0,001 to 500 g/L
Extremely accurate turbidity and solid sensors made from stainless steel for color-independent measurement of highly concentrated sludges. Working temperature up to 90 °C (194 °F); up to 95 °C (203 °F) briefly.
This sensor is available in a basin version (without wiper) or an installation version (TriClamp) (without wiper) (refer to 1 and 2 in Figure 2 Overview).
2.2.3 TSS VARI sc: 0,001 9999 FNU; 0,001 up to 500 g/L
Extremely accurate turbidity and solid sensors made from stainless steel for color-independent measurement of highly concentrated sludges. This sensor connects to VARIVENT® piping systems (without wiper) (refer to 6 in Figure 2 Overview).
2.2.4 TSS XL sc: 0,001 to 9999 FNU; 0,001 to 500 g/L
Extremely accurate turbidity and solid sensors made from stainless steel for color-independent measurement of highly concentrated sludges. This sensor connects to TriClamp piping systems. (without wiper) (refer to 5 in Figure 2 Overview).
2.2.5 TSS TITANIUM2 sc/TSS TITANIUM7 sc: 0,001 to 9999 FNU; 0,001 to 500 g/L
Extremely accurate turbidity and solid sensors made from TITANIUM GRADE 2/TITANIUM GRADE 7 for color-independent measurement of highly concentrated sludges. This sensor has been specially developed for use in aggressive media and is
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
1 Basin sensor 4 Installation sensor (TriClamp) with wiper
2 Installation sensor (TriClamp) 5 TSS XL sc
3 Basin sensor with wiper 6 TSS Vari sc
11
General information
available in the basin or installation version (TriClamp) (without wiper) (refer to 1 and 2 in
Figure 2 Overview).
2.3 Measuring principle
2.3.1 Turbidity according to DIN standards
Turbidity is measured in accordance with DIN standard EN 27027 (ISO 7027) and is calibrated by the manufacturer. Measurement is exceptionally simple and accurate.
2.3.2 Measurement of solids according to plant-specific curves
Software-based optimization routines enable extremely precise simulation of medium-specific calibration curves with few calibration points. Usually, a single calibration point is sufficient.
Up to three calibration points can be defined for a highly fluctuating medium. The combined multiple-beam alternating light technique records solids in the medium with even greater accuracy.
2.4 Handling
Do not subject the sensor to any hard mechanical impacts.
2.5 Scope of delivery
TSS sc sensor
Protective cap for sensor tip (depending on model)
Test log
User manual
TSS sc wiper kit for 5 changes including screws and screwdriver (LZY634, optional)
2.6 Function test
After unpacking and checking for any transport damage, perform a brief function check.
1. Connect the sensor to the sc controller (refer to 3.3, page 15).
2. Switch on the power supply to the sc controller.
The display is activated and the sensor enters measurement mode.
Note: The measurement value that is shown in air is not relevant.
3. If no warning or error messages are shown, the function check is complete.
12
General information
13
Section 3 Installation
Note: Depending on the area of application, the sensor may have to be installed with additional optional accessories.
3.1 Installation overview of basin sensor
Note: This system must be installed by qualified personnel.
Figure 3 Installation example with optional accessories
DANGER
Explosion hazard. The TSS sc sensors are not suitable for use in hazardous locations.
CAUTION
Personal injury hazard. The installation of this system must only be carried out by qualified experts in accordance with all local safety regulations.
14
Installation
3.2 Pipe installation options for sensor installation
Figure 4 Pipe installation options with optional accessories
1 TSS sc TriClamp with retractable ball valve fitting
(maximum alternating pressure 1,5 bar; max. operating pressure 6 bar) LZU300.99.00000
1
1
x= identifier for the nominal diameter of the pipe
2
y= identifier for the material selection of the associated connecting flange
4 TSS XL sc with XL LZU304.99.100x0
1 measuring tube
2 TSS sc Inline, TSS W sc Inline, TSS HT sc Inline with
LZY630.00.1y000
2
safety installation fitting (max.
operating pressure 6
bar)
5 TSS VARI sc with VARIVENT LZU304.99.000x0
1
measuring tube
3 TSS sc TriClamp with LZU302.99.000x0
1 welding
connector
15
Installation
3.3 Connect the sensor cable
1. Unscrew the protective caps from the controller socket and the cable plug and retain
them.
2. Pay attention to the guide in the plug and push the plug into the socket.
3. Tighten the nut by hand.
Note: Extension cables are available in various lengths (refer to Section 7 Replacement parts and
accessories).
Maximum cable length 100 m (328 ft).
Figure 5 Connect the sensor plug to the controller
CAUTION
Personal injury hazard. Always lay cables and hoses so that they are straight and do not pose a tripping hazard.
16
Installation
Figure 6 Pin configuration
Number Description Standard cable, cable color Teflon cable, cable color
1 +12 VDC Brown Pink
2 Earth Black Gray
3 Data (+) Blue Brown
4 Data (–) White White
5 Shield Shield (gray) Shield (gray)
6 Guide
6
1 2
5
3
4
17
Section 4 Operation
4.1 User interface and navigation
The sensor can be operated with all sc controllers. Refer to the controller documentation for keypad description and navigation information.
4.2 Sensor setup
When the sensor is connected for the first time, the sensor serial number is displayed as the name of the sensor. To change the sensor name:
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CONFIGURE and confirm.
5. Select EDIT and confirm.
6. Edit the name and confirm to return to the CONFIGURE menu.
Complete sensor configuration in the same manner, with the following menu options selected:
MEAS UNITS
PARAMETERS
CLEAN. INTERVAL
RESPONSE TIME
LOGGER INTERVAL
7. Go back to the MAIN MENU or the measurement mode display.
4.3 Sensor data logger
There is a data log and event log available for each sensor. The data memory is used to store measurement data at preset intervals; the event memory stores events such as configuration changes, alarms and warning conditions. Both logs can be exported to CSV format (refer to the controller manual).
4.4 Menu structure
Note: Refer to Section 6 Troubleshooting for a list of all possible error and warning messages together with a description of all necessary countermeasures to be taken.
4.4.1 SENSOR STATUS
SELECT SENSOR (if there is more than one sensor)
ERROR LIST
Possible error messages: MEAS OVERRANGE, CAL. INSUFF. +/–, ZERO, CAL REQUIRED, EE RSRVD ERR, ERROR PROBE, LED FAILURE
WARNING LIST Possible warning messages: REPLACE PROFILE, TEST/MAINT, GASKET
18
Operation
4.4.2 SENSOR setup
SELECT SENSOR (if there is more than one sensor)
WIPE Triggers a wiping operation
CALIBRATE (turbidity)
SET OUTMODE Behavior of the outputs during calibration and zero point adjustment
HOLD
ACTIVE
SET TRANSFER
SELECTION
SENSOR MEASURE Current, uncorrected measurement value
FACTOR
Can be set from 0,10 to 10,00; a detailed description is provided in section 4.5
CALIBRATE
OFFSET
Can be set from -100 to +100, a detailed description is provided in section 4.5
CALIBRATE
CALIBRATE
MEMORY
POINT 1 Calibration point 1 is recorded
POINT 2 Calibration point 2 is recorded
POINT 3 Calibration point 3 is recorded
CLEAR MEMORY Clears the recorded values for all points.
POINT 1 Current calibration for point 1
POINT 2 Current calibration for point 2
POINT 3 Current calibration for point 3
SET CAL DEFLT Security prompt, reset to default calibration
CALIBRATE (TS content)
SET OUTMODE Behavior of the outputs during calibration and zero point adjustment
HOLD
ACTIVE
SET TRANSFER
SELECTION
SENSOR MEASURE Current, uncorrected measurement value
FACTOR Can be set from 0,10 to 10,00; detailed description in section 4.5 CALIBRATE
CALIBRATE
MEMORY
POINT 1 Calibration point 1 is recorded
POINT 2 Calibration point 2 is recorded
POINT 3 Calibration point 3 is recorded
CLEAR MEMORY Clears the recorded value for all points.
POINT 1 Current calibration for point 1
POINT 2 Current calibration for point 2
POINT 3 Current calibration for point 3
SET CAL DEFLT Security prompt, all calibration points are cleared
19
Operation
CONFIGURE
EDIT NAME
Name can include up to 16 characters, FACTORY CONFIG: device number
MEAS UNITS
TRB: (FNU, EBC, TE/F, NTU, FTU) TS: (mg/L, g/L, ppm, %) FACTORY CONFIG: FNU
PARAMETERS TRB, TS, FACTORY CONFIG: TRB
CLEAN. INTERVAL 15 min, 30 min, 1 h, 4 h, 12 h, 1 day, 3 days, 7 days, FACTORY CONFIG: 4 h
RESPONSE TIME 1 to·300 s, DEFAULT CONFIG: 60 s
LOGGER INTERVAL
10 s, 30 s, 1 min, 2 min, 3 min, 4 min, 5 min, 6 min, 10 min, 15 min, 30 min, FACTORY CONFIG: 10 min
SET DEFAULTS Security prompt, reset to default configuration for all menu options listed above.
4.4.2 SENSOR setup
SELECT SENSOR (if there is more than one sensor)
20
Operation
4.5 CALIBRATE
Note: Turbidity measurement has been calibrated by the manufacturer — it does not need to be calibrated again.
Note: It is imperative to calibrate for solid matter measurement (refer to section 4.5.2 Calibration of
the SOLID (TS) parameter).
The zero point for turbidity and solid matter measurement has been set in the sensors by the manufacturer.
TEST/MAINT
PROBE INFO
SENSOR NAME Device name
EDITED NAME
SERIAL NUMBER
TURBIDITY 0,001 to 9999 FNU
SOLID 0,001 to 500 g/L
MODEL NUMBER Item no. Sensor
CODE VERS Sensor software
PROFILE
PROFILE COUNTER Counter 20,000 backwards
RESET CONFIG MANUAL RESET, security prompt
COUNTERS
MANUAL RESET. PRESS ENTER: security prompt TEST/MAINT: COUNTER X DAYS BACKWARDS, GASKET (GASK.): COUNTER X DAYS BACKWARDS, TOTAL: OPERATING HOURS COUNTER, MOTOR: WIPE CYCLE COUNTER
INTERVAL Default for maintenance counter
SERVICE
WIPE
SIGNALS Explanation: refer to service manual
S5E1
S5E3
S6E1
S6E3
S5E2
S5E4
S6E2
S6E4
SET OUTMODE Equipment output behavior in the SERVICE menu
HOLD
ACTIVE
SET TRANSFER
SELECTION
Service access
4.4.2 SENSOR setup
SELECT SENSOR (if there is more than one sensor)
21
Operation
Installation conditions in the pipes can cause interfering ground reflection when measuring turbidity, which in turn may cause the zero point to shift. Compensate for this effect with an offset correction (section 4.5.1.3 OFFSET). If there are deviations that are unrelated to the factors described above between the measurement values shown and the laboratory results, the slope of the calibration curve can be adjusted using a factor (refer to section 4.5.1 Calibration of the TURBIDITY (TRB) parameter).
At least a 1-point calibration must be carried out for a solid matter measurement. In difficult application conditions, a 2-point or 3-point calibration may be necessary (refer to section 4.5.2 Calibration of the SOLID (TS) parameter).
4.5.1 Calibration of the TURBIDITY (TRB) parameter
Before the sensor can be calibrated to the TURBIDITY (TRB) parameter, the parameter must be selected.
4.5.1.1 Select the TURBIDITY (TRB) parameter
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CONFIGURE and confirm.
5. Select PARAMETERS and confirm.
6. Select the TRB parameter and confirm.
7. Go back to the MAIN MENU or the measurement mode display.
4.5.1.2 FACTOR
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CALIBRATE and confirm.
5. Select FACTOR and confirm.
6. Set the desired factor and confirm.
7. Go back to the MAIN MENU or the measurement mode display.
4.5.1.3 OFFSET
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CALIBRATE and confirm.
5. Press OFFSET and confirm.
6. Set the desired offset and confirm.
7. Go back to the MAIN MENU or the measurement mode display.
22
Operation
4.5.1.4 1 to 3-point calibration
Note: The turbidity measurement has been calibrated by the manufacturer.
Note: Before the sensor can be calibrated to the TRB parameter, the parameter must be selected
(refer to
4.5.1.1 Select the TURBIDITY (TRB) parameter).
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CALIBRATE and confirm.
5. Select CALIBRATE and confirm.
6. Select MEMORY and confirm.
7. Select POINT... (point 1, 2 or 3) and confirm.
Once the calibration point has been recorded by the probe, a mark "<<" is shown after the point or points that has/have been recorded for approximately
3 seconds.
Note: If the Calibrate menu is closed and then opened again before the calibration is complete, the "<<" mark is shown again. This indicates that the calibration for this point/these points has not yet been completed. The old calibration values are still being used.
8. Select the recorded POINT and confirm.
9. Enter the laboratory comparison value and confirm.
To record more calibration points, repeat steps 6 to 9.
10. Go back to the MAIN MENU or the measurement mode display.
The instrument automatically sorts the saved calibration points according to the size of the calibration values, irrespective of the sequence in which the calibration points were recorded.
Point 1 is always assigned to the smallest calibration value.
Point 2 is assigned to the next smallest calibration value.
Point 3 is assigned to the largest calibration value.
The value calculated in the laboratory can be corrected at any time by overwriting it.
4.5.2 Calibration of the SOLID (TS) parameter
Before the probe can be calibrated to the SOLID (TS) parameter, the parameter must be selected.
4.5.2.1 Select the SOLID (TS) parameter
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CONFIGURE and confirm.
5. Select PARAMETERS and confirm.
6. Select the parameter TS and confirm.
7. Go back to the MAIN MENU or the measurement mode display.
23
Operation
4.5.2.2 FACTOR
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CALIBRATE and confirm.
5. Select FACTOR and confirm.
6. Set the desired factor and confirm.
7. Go back to the MAIN MENU or the measurement mode display.
4.5.2.3 1 to 3-point calibration
Note: It is imperative to calibrate for solid measurement (refer to section 4.5.2 Calibration of the
SOLID (TS) parameter).
Note: Before the sensor can be calibrated to the TS parameter, the parameter must be selected (refer to
4.5.2.1 Select the SOLID (TS) parameter).
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CALIBRATE and confirm.
5. Select CALIBRATE and confirm.
6. Select MEMORY and confirm.
7. Select POINT... (point 1, 2 or 3) and confirm.
Note: Points 2 and 3 are not shown unless point 1 or points 1 and 2 have already been recorded.
Note: This comparison is made from a grab sample and not a known standard.
Once the calibration point has been recorded by the probe, a mark "<<" is shown after the point or points that has/have been recorded for approximately
3 seconds.
Note: If the Calibrate menu is closed and then opened again before the calibration is complete, the "<<" mark is shown again. This indicates that the calibration for this point/these points has not yet been completed. The old calibration values are still being used.
8. Remove a sample and determine the solid matter content in the laboratory.
9. Select the recorded POINT and confirm.
10. Enter the laboratory comparison value and confirm.
To record more calibration points, repeat steps 6 to 10.
11. Go back to the MAIN MENU or the measurement mode display.
The instrument automatically sorts the saved calibration points according to the size of the calibration values, irrespective of the sequence in which the calibration points were recorded.
Point 1 is always assigned to the smallest calibration value.
Point 2 is assigned to the next smallest calibration value.
Point 3 is assigned to the largest calibration value.
24
Operation
The value calculated in the laboratory can be corrected at any time by overwriting it.
4.5.3 General information about calibration
4.5.3.1 Clear recorded points
Points that have been saved under MEMORY can be reset and cleared at any time.
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CALIBRATE and confirm.
5. Select CALIBRATE and confirm.
A mark "<<" is shown after the recorded point or points for approximately 3 seconds.
6. Select MEMORY and confirm.
7. Select CLEAR MEMORY and confirm.
The sensor will continue working with the old calibration values.
8. Go back to the MAIN MENU or the measurement mode display.
4.5.3.2 Clear a calibration point
An individual calibration point can be cleared at any time by entering the value 0.0 for the concentration.
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CALIBRATE and confirm.
5. Select CALIBRATE and confirm.
6. Select the POINT to be cleared and confirm.
7. Enter the value 0 and confirm.
8. Go back to the MAIN MENU or the measurement mode display.
25
Section 5 Maintenance
The cleanliness of the measurement windows in the sensor head is critical for the accuracy of the measurement results!
Check the measurement windows for dirt and the wiper profile for wear once a month.
5.1 Maintenance schedule
5.2 List of wearing parts
* When operated in line with manufacturer settings and used appropriately
5.3 Clean the measurement windows
Safety glasses
Gloves
Overalls
The measurement windows are made of sapphire glass. The measurement windows can be cleaned with any conventional cleaning agent and a soft cloth.
WARNING
Multiple hazards. Do not disassemble the instrument for maintenance or service. If the internal components must be cleaned or repaired, contact the manufacturer.
CAUTION
Personal injury hazard. Only qualified personnel should conduct the tasks described in this section of the manual.
NOTICE
The gaskets on the wiper shaft must be replaced yearly! Failure to replace the gaskets regularly, could result in moisture ingress into the sensor head and
lead to irreparable damage to the device.
Maintenance task Maintenance interval
Visual inspection Monthly
Check calibration Monthly (depending on environmental conditions)
Inspection Every six months (counter)
Replace wiper shaft gaskets Every year (counter)
Replace wiper profile As indicated by counter (20,000 cycles)
Number Designation Average service life*
1 Wiper sets 1 year (with normal sand loading)
1 Gasket set including wiper shaft 1 year
WARNING
Chemical hazard. Always follow appropriate safety procedures when chemicals are handled. Always wear all personal protective equipment appropriate to the chemicals used.
26
Maintenance
In the case of stubborn deposits, the use of 5% hydrochloric acid is recommended.
5.4 Replace the wiper profile
Note: The service life of the wiper profiles depends on the number of cleaning cycles carried out and the type of deposits to be removed.
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select TEST/MAINT and confirm.
5. Select PROFILE; replace the wiper profile as described in Figure 7.
Note: Make sure that the wiper is within the tolerance range shown.
6. Select RESET and confirm.
7. Confirm MANUAL RESET. ARE YOU SURE?
8. Go back to the MAIN MENU or the measurement mode display
Figure 7 Wiper replacement
.
1 Sensor 4 Screw, torque 15 Ncm
2 Wiper shaft 5 Screw driver
3 Wiper 6 Tolerance range for wiper replacement
27
Section 6 Troubleshooting
6.1 Error messages
Possible sensor errors are shown by the controller.
6.2 Warnings
Possible sensor warnings are shown by the controller.
Table 1 Error messages
Error displayed Cause Resolution
MEAS OVERRANGE
Measurement range exceeded, signals too small, probe can no longer measure this concentration
If error occurs more frequently, find another installation location
CAL. INSUFF. -- Calibration insufficient
Probe requires another calibration point in a lower concentration
CAL. INSUFF. + Calibration insufficient
Probe requires another calibration point in a higher concentration
ZERO
Calibration is too close to the zero point
Calibrate again with higher concentration
CAL REQUIRED No existing calibration Calibrate probe
EE RSRVD ERR Error in the probe electronics Call the manufacturer's customer service department
ERROR PROBE Error in the probe electronics Call the manufacturer's customer service department
LED FAILURE Faulty LED Call the manufacturer's customer service department
Table 2 Warnings
Warning shown Cause Resolution
REPLACE PROFILE Counter at zero Replace wiper profile, reset counter
TEST/MAINT Counter expired Call the manufacturer's customer service department
GASKET Counter expired Call the manufacturer's customer service department
28
Troubleshooting
29
Section 7 Replacement parts and accessories
7.1 Replacement parts
Description Cat. No
Wiper set (for five replacements with screws and screwdriver) LZY634
Wiper shaft maintenance kit (consisting of wiper, two-piece wiper shaft and gaskets) LZY635
Manual, xx = language code DOC023.xx.90154
7.2 Accessories
Description Cat. No
Silicone gasket for TriClamp fitting LZY653
PTFE gasket for TriClamp fitting LZY654
FKM Gasket for TriClamp fitting LZY655
Two-piece clip with thumb screw for TriClamp fitting LZY656
Three-piece clip with thumb screw for TriClamp fitting (for use with PTFE gasket) LZY657
Cable extension kit (5 m/16,40 ft) LZX848
Cable extension kit (10 m/32,81 ft) LZX849
Cable extension kit (15 m/49,21 ft) LZX850
Cable extension kit (20 m/65,62 ft) LZX851
Cable extension kit (30 m/98,43 ft) LZX852
Cable extension kit (50 m/164,04 ft) LZX853
Sensor bracket including 90° adapter LZX414.00.10000
Consisting of:
Base ATS 010
Mounting attachment HPL061
Holding clamp (2×) LZX200
Assembly pipe 2 m BRO075
HS small parts set LZX416
1,8 m extension pipe LZY414
1,0 m extension pipe LZY413
Second attachment point (including holding clamp) LZX456
90° sensor adapter AHA034
Set of small parts for securing sensor LZX417
90° base ATS0 11
Retractable ball valve fitting for all TSS sc TriClamp sensors (except TITANIUM, VARI & XL) LZU300.99.00000
DN65 measuring tube for TSS VARI sc LZU304.99.00010
DN80 measuring tube for TSS VARI sc LZU304.99.00020
DN100 measuring tube for TSS VARI sc LZU304.99.00030
DN125 measuring tube for TSS VARI sc LZU304.99.00040
DN65 measuring tube for TSS XL sc LZU304.99.10010
DN80 measuring tube for TSS XL sc LZU304.99.10020
DN100 measuring tube for TSS XL sc LZU304.99.10030
DN125 measuring tube for TSS XL sc LZU304.99.10040
DN150 measuring tube for TSS XL sc LZU304.99.10050
DN200 measuring tube for TSS XL sc LZU304.99.10060
DN250 measuring tube for TSS XL sc LZU304.99.10070
30
Replacement parts and accessories
Unprocessed welding connector for all TSS sc TriClamp sensors (except VARI & XL) LZU302.99.00000
DN65 welding connector for all TSS sc TriClamp sensors (except VARI & XL) LZU302.99.00010
DN80 welding connector for all TSS sc TriClamp sensors (except VARI & XL) LZU302.99.00020
DN100 welding connector for all TSS sc TriClamp sensors (except VARI & XL) LZU302.99.00030
DN125 welding connector for all TSS sc TriClamp sensors (except VARI & XL) LZU302.99.00040
DN150 welding connector for all TSS sc TriClamp sensors (except VARI & XL) LZU302.99.00050
DN200 welding connector for all TSS sc TriClamp sensors (except VARI & XL) LZU302.99.00060
DN250 welding connector for all TSS sc TriClamp sensors (except VARI & XL) LZU302.99.00070
Unprocessed welding connector for TSS XL sc LZU302.99.10000
DN65 welding connector for TSS XL sc LZU302.99.10010
DN80 welding connector for TSS XL sc LZU302.99.10020
DN100 welding connector for TSS XL sc LZU302.99.10030
DN125 welding connector for TSS XL sc LZU302.99.10040
DN150 welding connector for TSS XL sc LZU302.99.10050
DN200 welding connector for TSS XL sc LZU302.99.10060
DN250 welding connector for TSS XL sc LZU302.99.10070
Welding connector for all TSS sc TriClamp sensors (except VARI & XL) LZU303.99.00000
6-bar safety installation fitting with stainless steel flange for TSS sc Inline, TSS W sc Inline and TSS HT sc Inline
LZY630.00.10000
6-bar safety installation fitting with carbon steel flange for TSS sc Inline, TSS W sc Inline and TSS HT sc Inline
LZY630.00.11000
6-bar safety installation fitting without flange for TSS sc Inline, TSS W sc Inline and TSS HT sc Inline
LZY630.00.12000
7.2 Accessories
Description Cat. No
31
Section 8 Warranty and liability
The manufacturer warrants that the supplied product is free of material and manufacturing defects, and undertakes to repair or to replace any defective parts without charge.
The warranty period is 24 months. If a maintenance contract is taken out within 6 months of purchase, the warranty period is extended to 60 months.
With the exclusion of further claims, the supplier is liable for defects, including the lack of assured properties, as follows: all parts that, within the warranty period calculated from the day of the transfer of risk, can be demonstrated to have become unusable or that can only be used with significant limitations owing to circumstances prior to transfer of risk, in particular due to incorrect design, substandard materials or inadequate finish, shall be repaired or replaced at the supplier's discretion. The identification of such defects must be reported to the supplier in writing as soon as possible, but no later than 7 days after the discovery of the fault. If the customer fails to notify the supplier, the product is considered approved despite the defect. Further liability for indirect or direct damages is not accepted.
If device-specific maintenance- or inspection work prescribed by the supplier is to be performed within the guarantee period by the customer (maintenance) or by the supplier (inspection) and these requirements are not met, claims for damages that result from non-observance of these requirements are void.
Further claims, in particular for consequential damages, cannot be made.
Wear and damage caused by improper handling, incorrect installation or non-designated use are excluded from this clause.
The process instruments of the manufacturer have proven their reliability in many applications and are therefore often used in automatic control loops to enable the most economical and efficient operation of the relevant process.
To avoid or limit consequential damage, it is therefore recommended that the control loop be designed such that an instrument malfunction results in an automatic changeover to the backup control system. This guarantees the safest operating condition both for the environment and the process.
32
Warranty and liability
33
AppendixA Modbus register
Table 3 Sensor Modbus registers
Tag name Group name Register Data type Length R/W Description
TURBIDITY FNU Measurement 40001 Float 2 R Turbidity in FNU
TURBIDITY NTU Measurement 40001 Float 2 R Turbidity in NTU
TURBIDITY TEF Measurement 40001 Float 2 R Turbidity in TEF
TURBIDITY FTU Measurement 40001 Float 2 R Turbidity in FTU
TURBIDITY EBC Measurement 40003 Float 2 R Turbidity in EBC
SOLID mg/L Measurement 40005 Float 2 R Solid in mg/L
SOLID ppm Measurement 40005 Float 2 R Solid in ppm
SOLID g/L Measurement 40007 Float 2 R Solid in g/L
SOLID % Measurement 40009 Float 2 R Solid in percent
Reserved Reserved 40011 Unsigned integer 1 R Spar e
SET PARAMETER Configuration 40012 Unsigned integer 1 R/W Parameter
UnitTM Unit 40013 Unsigned integer 1 R/W Turbidity unit
UnitDS Unit 40014 Unsigned integer 1 R/W Solid unit
OFFSET Calibration 40015 Float 2 R/W Turbidity offset
Factor TRB Calibration 40017 Float 2 R/W Turbidity factor
Factor TS Calibration 40019 Float 2 R/W Solid factor
Reserved Reserved 40021 Unsigned integer 1 R Reserved
RESPONSE TIME Configuration 40022 Unsigned integer 1 R/W Response time
CLEAN. INTERVAL Configuration 40023 Unsigned integer 1 R/W Cleaning interval
LOGGER INTERVAL Configuration 40024 Unsigned integer 1 R/W Logger interval
Outputmodekal Service 40025 Unsigned integer 1 R/W Output mode "Calibrate"
Outputmodesrv Service 40026 Unsigned integer 1 R/W Output mode "Service"
EDITED NAME Configuration 40027 String 8 R/W Measurement location
PROFILE COUNTER Configuration 40035 Unsigned integer 1 R/W Profile counter
SERIAL NUMBER Configuration 40036 String 6 R/W Serial number
CAL. DATE Configuration 40042 Time2 2 R Date of factory calibration
TURBIDITY Calibration 40044 Float 2 R
Turbidity sensor measurement value
SOLID Calibration 40046 Float 2 R Solid sensor measurement value
PROGRAM Maintenance 40048 Float 2 R Application version
BOOTPROGR. Maintenance 40050 Float 2 R Bootloader version
STRUCTURE Maintenance 40052 Unsigned integer 1 R Structure driver version
FIRMWARE Maintenance 40053 Unsigned integer 1 R Register driver version
CONTENT Maintenance 40054 Unsigned integer 1 R Firmware driver version
FormatMinFNU Configuration 40055 Float 2 R Turbidity lower limit in FNU
FormatMaxFNU Configuration 40057 Float 2 R Turbidity upper limit in FNU
FormatMinEBC Configuration 40059 Float 2 R Turbidity lower limit in EBC
FormatMaxEBC Configuration 40061 Float 2 R Turbidity upper limit in EBC
FormatMinGL Configuration 40063 Float 2 R Solid lower limit in g/L
FormatMaxGL Configuration 40065 Float 2 R Solid upper limit in g/L
FormatMinMGL Configuration 40067 Float 2 R Solid lower limit in mg/L
FormatMaxMGL Configuration 40069 Float 2 R Solid upper limit in mg/L
FormatMinPR Configuration 40071 Float 2 R Solid upper limit in percent
34
FormatMaxPR Configuration 40073 Float 2 R Solid upper limit in percent
S5E1 Maintenance 40075 Float 2 R Signal LED S5E1
S5E3 Maintenance 40077 Float 2 R Signal LED S5E3
S6E1 Maintenance 40079 Float 2 R Signal LED S6E1
S6E3 Maintenance 40081 Float 2 R Signal LED S6E3
S5E2 Maintenance 40083 Float 2 R Signal LED S5E2
S5E4 Maintenance 40085 Float 2 R Signal LED S5E4
S6E2 Maintenance 40087 Float 2 R Signal LED S6E2
S6E4 Maintenance 40089 Float 2 R Signal LED S6E4
Table 3 Sensor Modbus registers(Continued)
Tag name Group name Register Data type Length R/W Description
HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Tel. (970) 669-3050 (800) 227-4224 (U.S.A. only) Fax (970) 669-2932 orders@hach.com www.hach.com
HACH LANGE GMBH
Willstätterstraße 11 D-40549 Düsseldorf, Germany Tel. +49 (0) 2 11 52 88-320 Fax +49 (0) 2 11 52 88-210 info-de@hach.com www.de.hach.com
HACH LANGE Sàrl
6, route de Compois 1222 Vésenaz SWITZERLAND Tel. +41 22 594 6400 Fax +41 22 594 6499
© Hach Company/Hach Lange GmbH, 2009, 2012, 2018
. All rights reserved. Printed in Germany 09/2018, Edition 3
Loading...