Hach Company certifies this instrument was tested thoroughly,
inspected and found to meet its published specifications when it
was shipped from the factory.
The sension™4 Laboratory pH/ISE Mete r has been teste d and is
certified as indicated to the following instrumentation standards:
External Power Supplies Only:
115 VAC Supply: UL listed and CSA certified or
230 VAC Supply: CE marked per 73/23/EEC, VDE listed
Instrument tested with e xt er nal 230V, 50 Hz power supply.
Per 89/336/EEC EMC: EN 61326:1998 (Electrical Equipmen t
for measurement, control, and laboratory use— EMC
requirements) Supporting test records by Hach Company,
certified compliance by Hach Company.
Standards include:
IEC 1000-4-2: 1995 (EN 61000-4-2:1995) Electro-Static
Discharge Im munity (Criteria B)
IEC 1000-4-3: 1995 (EN 61000-4-3:1996) Radiated RF
Electro-Magnetic Fields (Criteria B)
IEC 1000-4-4: 1995 (EN 61000-4-4:1995) Electrical Fast
Transients/Burst (Criteria B)
IEC 1000-4-5: 1995 (EN 61000-4-5:1995) Surge (Criteria B)
IEC 1000-4-6: 1996 (EN 61000-4-6:1996) Conducted
Disturbance Induced by RF Fields (Criteria A)
IEC 1000-4-11: 1994 (EN 61000-4-11:1994) Voltage Dips,
Interruptions and Variations (Criteria B)
ENV 50204:1996 Radiated Electro-Magnetic Field from
Digital Telephones (Criteria B)
5
Page 6
CERTIFICATION, continued
Emissions:
Instrument tested with external 230V, 50 Hz power supply.
Per 89/336/EEC EMC: EN 61326:1998 (Electrical Equipment
for measurement, control, and laboratory use— EMC
requirements). Class B emission limits. Supporting test records
by Hewlett Packard Hardware Test Center, Ft. Collins, CO
(A2LA #0905-01), certified compliance by Hach Company.
Standards include:
EN 61000-3-2 Harmonic Disturbances Caused by
Electrical Equipment
EN 61000-3-3 Voltage Fluctuations (Flicker) Disturbances
Caused by Electrical Equipment
Additional Standards include:
EN 55011 (CISPR 11) Emissions, Class B Limi ts
Additional Emissions Standard/s include:
CANADIAN INTERFERENCE-CAUSING EQUIPMENT
REGULATION, IECS-003: Class A emission limits.
Supporting test records by Hewlett Packard Hardware Test
Center, Ft. Collins, CO (A2LA #0905-01), certified compliance
by Hach Company.
This Class A digital apparatus meets all requirements of the
Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les
exigences du Règlement sur le matériel brouilleur du Canada.
FCC PART 15, Class “A” Limits:
Supporting test records by Hewlett Packard Hardware Test
Center, Ft. Collins, CO (A2LA #0905-01), certified compliance
by Hach Company.
This dev ice co mplies with P art 15 of the FCC Rules. Operat ion i s
subject to the following two conditions:
6
Page 7
CERTIFICATION, continued
(1) This device may not cause harmful interference, and (2) this
device must accept any interference received, including
interference that may cause undesired operation.
Changes or modifications to this unit not expressly approved by
the party responsible for compliance could void the user's
authority to operate the equipment.
This equipment has been tested and found to comply with the
limits for a Class A digita l de vic e, pursua nt to P art 15 of th e FCC
Rules. These l i mits are designed to pr o vide reasonable prote cti on
against ha rmful interferenc e when the equipment is operated in a
commercial environment. This equipment generates, uses, and
can radiate radio frequency energy and, if not installed and used
in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this
equipment in a resid ential are a is likely to cause harmful
interference, in which case the user will be required to correct the
interference at his own expense. The following techniques of
reducing the interference problems are applied easily.
1.Disconnect the external power supply from sens
Laboratory pH meter to verify that the meter is not the source
of interference.
2.Move the meter away from the device receiving
the interference.
3.Reposition the receiving antenna for the device receiving
the interference.
4.Try combinations of the above.
ion4
7
Page 8
8
Page 9
SAFETY PRECAUTIONS
Section 2.2 on page 21
Section 2.3 on page 21
Section
2.4 on page 23
Section 5.1 on page 45
9
Page 10
10
Page 11
SPECIFICATIONS
Specification subject to change without notice.
pH mode
Range-2.00 to 19.99
Resolution0.001/0.01/0.1 (selectable)
Slope range45 to 65 mV/decade at 25 °C
ISE mode
Range0.000 to 19900
ResolutionAuto-ranging
Millivolt mode
Range-2000 to 2000 mV
Resolution0.1 mV
Accuracy±0.2 mV or ±0.05% of reading,
Temperature mode
Range-10.0 to 110 °C (can also display °F)
Resolution0.1 °C
Accuracy±0.3 °C from 0–70 °C
whichever is greater
±1.0 °C from 70–110 °C
Display: Custom LCD
Inputs: 2 BNC; two 5-pin Hach pH /temperature or Hach
temperature probe; 2 pin-tip; (one combination input type for
each of th e two channels or two half-cells per chan nel).
Outputs: Two-way RS232
Power Re quirements: 6–9 VDC provided by 120 V 50/60 Hz or
230 V, 50 Hz external power supply or a customer-provided
supply with 50 mA output, 5.5-mm power plug with a 2.5 mm
center post opening.
Installation Cate gory: II (for external power supplies)
Input Impedance: >10
12
ohms
Instrument Drift: <40 µV/°C
Input Bias Current: <±1 picoamp at 25 °C; <±4 picoamp over
full range
11
Page 12
SPECIFICATIONS, continued
Envi ronmental Requirements: 0 to 50 °C at 85%
non-condensing relative humidity. Enclosure is water resistant,
chemical resistant, dust proof.
Dimensions: 25.4 x 15 x 8.37 (10.15 x 6 x 3.35 in.)
12
Page 13
OPERATION
DANGER
Handling chemical samples, standards, and reagents can be dangerous. Review the necessary
Material Safety Data Sheets and become familiar with all safety procedures before handling
any chemicals.
DANGER
La manipulation des échantillons chimiques, étalons et réactifs peut être dangereuse. Lire les Fiches
de Données de Sécurité des Produits (FDSP) et se f amiliariser avec toutes les procédures de sécurité
avant de manipuler tous les produit s chimiques.
PELIGRO
La manipulación de muestras química s, estándares y reactivo s puede ser peligrosa. Revise las fichas
de seguridad de materiales y familiarícese con los procedimientos de seguridad antes de manipular
productos químicos.
Das Arbeiten mit chemischen Proben, Standards und Reagenzien ist mit Gefahren verbunden.
Es wird dem Benu tzer di eser Pr oduk te empf ohlen , sic h v or der Arbei t mit s ich eren Verfahrensw eisen
und dem richtigen Gebrauch der Chemikalien vertraut zu machen und alle entsprechenden
Materialsicherheitsdatenblätter aufmerksam zu lesen.
A manipulação de amostras, padrões e reagentes químic os pode ser perigosa. Reveja a fo l ha dos
dados de segurança do material e familiarize-se com todos os procedimentos de segurança antes
de manipular quaisquer produtos químicos.
GEFAHR
PERIGO
13
Page 14
14
Page 15
SECTION 1INTRODUCTION
Hach provides pH and ISE meters for applications from pH
measurements to accurate Ion Sel ect ive Electrode work. This
manual describes the operation and use of the Hach sens
Portable p H/ISE/mV/Temperature Meter (see Figure 1).
This meter features a custom digital LCD display whic h
simultaneously displays temperature and measurement results.
This meter has all the features of a simple pH/ISE meter plus:
•two measurement channels
•a millivolt mode
•standard additions program
•automatic buffer recognition when in the pH mode
•pH averaging
•calibration review
ion™4
•datalogging
•bi-directional RS232 communication for printing results to a
computer or print er, or controlling the meter from a PC.
The meter is designed to be maintenanc e- fr ee. If the met er get s
dirty, wipe the surface with a damp cloth. Use a cotton-tipped
applicator to clean or dry the connectors if they get wet.
1.1Unpacking the Instrument
Remove the instrument and accessories from the shipping
container and inspect each item for any damage that may have
occurred during shipping. Verify that all ite ms listed on the
packing slip are included. If any items are missing or damaged,
contact Hach Customer Service, Loveland, Colorado for
instructions. Hach’s toll free phone number for customers within
the United States is 800-227-4224. For customers outside the
United States, contact the Hach office or distributor serving you.
1.2Standard Accessories
•AC to DC wall adapter
•Instrument Manual
•Include electrode and related accessories (covered in the
electrode manual).
15
Page 16
SECTION 1, continued
1.3Keypad Description
Fi gure 1 illustrate s the meter’s keypad. The description and
function of each key are given in Table 1.
Figure 1Hach sension4 Meter
16
Page 17
SECTION 1, continued
Table 1 Keys and Description
KeyDescription
Exit/Power On-OffTur ns the instrument on; turns it off from a Reading mode.
Acts as a NO or Cancel key when the question mark icon is flashing.
In Setup mode, backs up one step toward the Reading mode.
Performs the following and returns to the most recent Reading mode:
• Exits the Store or Recall mode
• Aborts a calibration
• Exits a calibration review
Arrow KeysScrolls between op tions in Setup mode.
Scrolls through data points in Store and Recall modes.
Scrolls between the option to print or erase one data point and all data points.
Changes the default temperature when a temperature probe is not in use.
READ/ENTER
Key
Recall KeyRecalls stored sample data (from Reading mode only).
Store KeyStores the current (displayed) measurement (from Reading mode only).
Erase KeyErases recalled data points.
ISE/mV KeyToggles between ISE concentration value and mV in Reading, Calibr ation, and Cal
pH/mV KeyToggles between pH value and mV value in Reading, Calibration, and Cal
Std AddnMoves meter into program to use for standard additions.
Print KeySends current or recalled data to a printer or a computer via the RS232 port.
ChannelToggles between channel 1 and 2. Will not toggle in Calibration, Cal Review, and
Accepts numerical input.
Acts as a “YES” answer when the question mark is flashing.
Allows user to edit a setup when the setup number is flashing.
Accepts the current Setup option when that option is flashing.
Initiates a new measurement when the meter has stabilized in the Display Lock
Enabled mode.
Review modes.
Review modes.
Recall modes.
Time KeyIn Reading mode, allows user to view the current time and date. In Recall Data
mode, it toggles between the time and date of the stored measurement.
Cal KeyEnters Calibration mode (from Reading mode only)
Review KeyEnters Calibration Review mode (from Reading mode only)
Setup/CE K e yEnters Setup mo de (from Rea ding mode only) or clears a numeric entry when the
keypad icon is displayed.
17
Page 18
SECTION 1, continued
1.4Display Fields and Icons
The display has two screens. The upper screen displays
measurements or standard values, operation mode, slope,
sample/default temperature, pH or mV units, error codes, and
indicates if the meter reading is stable. The lower screen displays
keys that are active.
Fi gure 2 shows the icons and screens displayed by the meter and
Table 2 describes each element. All icons on the display will be
displayed if the power key is held down for several seconds.
Figure 2sension4 Display
18
Page 19
SECTION 1, continued
Table 2 Main Display Elements
Item No.Description
1Indicates meter is in Standard Addition mode.
2 (Cal
Review)
3Used with
4Indicates the meter is measuring/reviewing a sample (sample # is displayed to the right).
5Indicates data is being sent to a printer/computer.
6Indicates recalled data that is currently displayed is being erased.
7Indicates meter is in Setup mode.
8Indicates all data points are being printed or erased.
9Numerical field tha t dis pl ays Setup, Sampl e . an d Sta nda rd numbers when those words are
10Flashing
11Indicates the meter is measuring/reviewing a standard (standard # is displayed above).
12Numerical field that displays the values of standards and samples, slope, pH or mV.
13Indicates measurement units (pH, mV, mg/L, µg/L, ppm, ppb, M).
14When
Indicates meter is in Calibration Review mode.
? to ask if user wants to store calibration, standard value, or sample value
displayed with the number. If
Standard 1.
Standard and 1 are displayed, the meter is measuring
? and CAL indicate calibr ati on is nece ssary. Also a prompt to pres s the ENTER or
EXIT key.
Default is displaye d, the meter is using the def ault tempera ture valu e to calculate the
temperature correction for the pH value.
15Temperature units (choice of °C or °F).
16Indicates value displayed in small numerical field (item 20) is in millivolts.
17Indicates channel 2 is in use.
18Indicates an inactive key has been pressed and that function is not allowed.
19Indicates ENTER key is active.
20Numerical field that displays temperature value.
21Indicates arrow keys are active.
22Indicates the date is being set or displayed.
23Indicates EXIT key is active.
24Faulty probe connection or incorrect probe attached. Usually displayed with an error code.
25Indicates numeric part of the keys is active.
26Indicates channel 1 is in use.
27NA
28Display Lock icon. Indicates reading is locked after stability is achieved.
29In setup mode, indicates whether Display Lock setting is
30Indicates a meter function problem.
On or Off.
19
Page 20
SECTION 1, continued
Table 2 Main Display Elements (Continued)
Item No.Description
31Indicates the displayed number is the electrode slope.
32Indicates Time is being set or displayed.
33When on or flashing,
disappears, the reading is stable.
34Indicates meter is in recall mode and the data displayed is stored data.
35 (CAL)Indicates meter is in Calibration mode. If the
Stabilizing... indicates signal from sample is not yet stable. When it
? is flashing, calibration is necessary.
1.5Audible Signals
The meter will beep under certain conditions:
•a non-functional key press is made (one beep)
•when display lock is enabled and measurement stability is
reached during reading mode (three beeps)
•any time measuremen t stability is reached during calibration
mode, regardless of the Display Lock setting
•to signal an error condition (two beeps)
•if the number entry and a press of the ENTER key causes two
beeps, the meter will automatically return to the beginning of
number entry.
20
Page 21
SECTION 2INSTRUMENT SETUP
2.1Instrument Description
This sension™4 Lab o ra tor y pH/IS E/m V meter is des ign ed for
laboratory use and operates on 115/230 VAC power. Two
channels are available for electrode connection.
The meter measures from -2.0 to 19.99 pH units and the sample
temperature. Displayed pH values are temperature corrected
using the measured sample temperature or a default temperature
setting. To display mV and pH values, press the
toggle between the units.
In the ISE mode, the meter measures from 0 to 19900 with the
highest resolution as ten thousandths (0.0001). To display mV
and concentrati on values, press the
the units.
2.2Power Connections
A 115 or 230 AC to DC converter is required to power the meter.
Plug the power jack of the adapte r into the meter power conn ector
(see Figure 3). Then plug the adapter into the wall.
ISE/mV key to togg le between
pH/mV key to
2.3pH/ISE and Temperature Probe Connections
2.3.1pH/ISE Probe Connections
Electrodes may be simult aneousl y atta ched to t he 5-pin a nd BNC
connectors of the same channel as long as they are not in contact
with the s ame solution. To select either of t he connectors for
measurement, go to
Five-pin Connec tors
Attach pH electrodes with 5-pin connectors to the meter input by
lining th e pins up with the holes in the m eter port (see Figure 3).
Push towards the instrument.
BNC Connectors
For pH or ISE probes with BNC connectors, slide the connector
into the input. Push towards the instrument and turn the metal
sleeve clockwise to lock into position. Do not use a pH probe
with tem perature sensor to measure temperature while using a n
ISE probe.
Setup 1 and choose one.
21
Page 22
SECTION 2, continued
Pin-type Connectors
When using half-cells, conne ct reference electrodes with p in tip
connectors by pushing the connector straight into the center
reference input.
Note: If using a combination electrode with a BNC or 5-pin connector,
the reference pin-tip jack is not used.
Figure 3sension4 Power, RS232, and Probe Connections
2.3.2Temperature Probe Connection
Hach pH electrodes with the 5-pin connector have the
temperature sensing unit included in the electrode probe and
require only the 5-pin connector. If using a pH electrode with
BNC connector, connect the Hach Temperature Probe
(Cat. No. 51980-00) to the 5-pin connector on the meter
(see Figure 3). The user can also measure the temperature
manually and enter the value as the default temperature on the
meter (see Section 2.6 on page 23). Do not use a pH probe with
temperature sensor to measure temperature while using an
ISE probe.
22
Page 23
SECTION 2, continued
2.4Printer and Computer Connections
The meter can send data to a computer or printer via the 9-pin
serial port on the meter (see Figure 3). The printer cable and computer cable are d ifferent. The printer cable is a 9-pin to
25-pin cable and the computer cable is a 9-pin to 9-pin cable. Be
sure to use the correct cable.
The meter can print to serial printers without an adapter. For
parallel printers, a converter and cable adapter are required. The
Citizen PN60 printer (Cat. No. 26687-00) requires a special
Citizen adapter (supplied with the printer). Pressing the
key will send the currently displayed data to the printer or the
user can enable the autoprint function (see 3.1.10 on page 30).
The data may be eit her a curr ent measurement or recalled data.
T o commun icate with a computer , conn ect the 9-pin seria l port on
the meter to a 9-pin serial port of the computer using a 9-pin to
9-pin cable. Pressing the
displayed data to the computer. The data may be either a current
measurement or recalled data. To send commands from a
computer, see the Table 6 on page 51.
PRINT key will send the currently
PRINT
2.5Turning the Meter On
After plugging the correct p ower adapter into the wall, turn the
instrument on using the
side of the keypad). Press the key once to power the instrument
up. The display will show the software version number, then
move to the Reading mode.
2.6Temperature Measurement
The meter displays temperatur e in the ran ge of -10.0 to +110 °C
simultaneously with sample results. If a temperature probe is
properly connected, actual temperature measurements are
displayed in the temperature/mV field.
The meter requires a temperature to calculate temperaturecorrected pH readin gs. The meter uses tempera ture data from one
of three sources:
•The temperature sensor in the sample
•The factory default setting (20 °C)
I/O/EXIT key (located on the upper left
•A user-entered default setting
23
Page 24
SECTION 2, continued
If a temperature probe is connected properly, the meter will
display the current sample temperature and will not allow the
temperature to be manually set.
To manually set the temperature
When a temperature probe is not used to supply the temperature
for pH temperature compensation, th e default te mperature
(20 °C) will be used. In either case,
Default will be displayed
above the temperature.
To change the defaul t temperature, first put the meter in Readi ng
or Calibration mode . Pr ess the arrow keys to increase or decr eas e
the displayed default temperature value to the desired
temperature. The value adjusts in 0.1 °C (0.1 to 0.2 °F)
increments. To scroll rapidly, hold the arrow key down.
Note: Acceptable temperatures range from -10 to 110 °C.
To change the units of temperature measurement, see Section
3.1.3 on page 26.
2.7Millivolt Measurements
The meter can be used to measure absolute millivolts (mV). To
display a c urrent mi lli v olt reading, press t he
The mV value is displayed with
key again to return to pH or ISE measurement units.
Absolute millivolts are displayed with 0.1 mV resolution in the
range of -2000.0 to +2000.0. The millivolt mode is useful when
measuring oxidation-reduction potential, performing
potentiometri c titrat ions, or p reparing c alibration curves . Detailed
instructions for any Hach electrode are given in the electrode
instruction manual. T itr ation inst ructions are i ncluded in the Hach
ORP Electrode Instruction Manual, or in standard analytical
chemistry texts.
pH/mV or ISE/mV key .
mV in the units field. Press the
24
Page 25
SECTION 3INSTRUMENT OPE RATION
3.1Setup Menu
The sension™4 Laboratory pH/ISE meter has a setup menu for
each channel which allows the analyst to choose options for
connector choice, display lock, temperature units, display
resolution, automatic buffer recognition, time, date and others
(see Table 3). Each channel can have separate setup options
except date and time.
To access the Setup menu, press the
SETUP key. The arrow icons
will be displayed, indicating that additional options are available
within the menu. Press the up or down arrow key to scroll to the
desired option, then press
EXIT until the meter retu rns to Reading mode.
ENTER. To exit the setup menus, press
When using the Setup menu, the screen will display a number in
the upper right numerical field, indicating which option is being
changed. Table 3 describes these options.
Table 3 sensION4 Setup Options
Setup
Number
1Use BNC or 5-pin connector
2Display lock (On or Off)
3Temperature units (°C or °F)
4pH measurement resolution (0.0, 0.00, or 0.000)
5Buffer auto recognition (6.86 or 7.00)
Setup Description
3.1.1Choosing the Probe Connector
6Stability criteria
7Time of day (24-hour clock)
8Date (mm/dd)
9Year (4 digits)
10Print interval settings
This setup ensures the potential from the appropriate electrode is
detected b y th e met er. Do not use both a pH electrode and an ISE
electrode from the same channel in the same solution. Both can
be connected if they are in separate solutions, but only the
electrode on the selected connector (BNC or 5-pi n) will be activ e.
25
Page 26
SECTION 3, continued
1.From the reading mode, press SETUP.
2.The
Setup icon and the number 1 (flashing) will be displayed.
Change the connec tor c hoice b y pressi ng
the selection between
3.When the desired option is selected, press
the reading mode.
3.1.2Turning Display Lock Off and On
Setup 2 is the Display Lock option. This feature stops
measurement reading fluctuation on the display once a stable
reading is reached. The default setting is Off.
When this feature is not used, th e measurement value may
continue to fluctuate.
1.From the reading mode, press
2.The
Setup icon and the number 1 (flashing) will be displayed.
Press the up arrow once so the Setup number is 2.
ENTER; this toggles
BNC and 5 pin.
EXIT to return to
SETUP.
3.Change the Display Lock status by pressing ENTER; this
toggles the Display Lock between off and on. When the
Display Lock is disabled, the Display Lock icon and
displayed. When this feature is enabled, the Display Lock
icon is displayed when the sta bilizatio n criteria are met.
4.When the desired option is selected, press
the reading mode.
3.1.3Selecting Temperature Units
1.From the reading mode, press SETUP.
2.The
3.Change the temperature unit by pressing
Setup icon and the number 1 (flashing) will be displayed.
Press the up arrow twice so the Setup number is
toggles the temperature units between °C and °F. The default
is °C.
Off are
EXIT to return to
3.
ENTER; this key
4.When the desired option is selected, press
EXIT to return to
the reading mode.
26
Page 27
SECTION 3, continued
3.1.4Selecting Measurement Resolution
The meter can display pH measurement values to tenths (0.0),
hundredths (0.00) or thousandths (0.000). The default is
hundredths. The choice of resolution affects the stability criteria.
1.From the reading mode, press SETUP.
2.The
Setup icon and the number 1 (flashing)will be displayed.
Press the up arrow until the Setup number is
3.Change the resolution by pressing
between the three resolution options.
4.When the desired option is selected, press EXIT to return to
the reading mode.
3.1.5Selecting Auto Buffer Recognition
The sension4 Meter is designed to auto-recognize and calibrate
on 1.68, 4.01, 6.86 or 7.00, 10.01, and 12.45 pH buffers.
The only selection option for pH buffer auto recognition is
choosing 6.86 or 7.00. The default is pH 7.00. To change
this option:
1.From the reading mode, press
2.The
Setup icon and the number 1 (flashing)will be displayed.
Press the up arrow until the Setup number is
4.
ENTER; this toggles
SETUP.
5.
3.1.6Selecting Stability Criteria
3.Change the buffer value by pressing
ENTER; this toggles the
between the choices 6.86 and 7.00.
4.When the desired option is selected, press
EXIT to return to
the reading mode.
The meter is designed to recognize a stable ISE reading based on
the rate of drift in mV/min (criteria are automatically selected for
pH measu rement based on resolution). If the elec trode voltage is
drifting s l ower than the chosen mV stability criteria, the disp lay
will show a stabilized r eading. If the electrode voltage is drifting
faster than the chosen mV stability criteria, the display will
continue to display
Stabilizing... and measuremen t values
will fluctuate.
27
Page 28
SECTION 3, continued
The default stabilit y drift rate is 0.500 mV/min. Lower stability
criteria (i.e., 0.300 mV/min) will require longer stabilization
times, but the measur emen t will be more precise . Higher st abili ty
criteria will require shorter stabilization times, but measurements
may be less precise.
To change this option:
3.1.7Setting the Time
1.From the reading mode, press
SETUP.
2.The Setup icon and the number 1 (flashing)will be displayed.
Press the up arrow until the Setup number is 6.
3.Change the stability criteria by pressing ENTER. The display
will show _ _ _ _. The numbered keypad will become active.
4.Enter the desired mV value using the keypad. Press ENTER
when entry is complete.
5.If a number entry error occurs, start over by pressing
SETUP/CE.
6.When the desired value has been accepted , press EXIT to
return to the Reading mode.
1.From the reading mode, press SETUP.
2.The
3.Change the time by pressing
4.Press the desired number key for the left most digit. The
Setup icon and the number 1 (flashing) will be displayed.
Press the up arrow until the Setup number is
ENTER. The main display will
change to
_ _:_ _, with the left place holder flashing. The
7.
numerical keypad will become active.
meter uses a 24-hour military clock (12 a.m. = 00:00; 1 p.m.
= 13:00). All four digit places must have a number. If the
left-mos t digit(s) is not necessary, use zero for t he value
(i.e., 08:15 for 8:15 a.m. or 00:30 for 12:30 a.m.).
28
Page 29
SECTION 3, continued
5.Once a number ke y is presse d, the ne xt digit plac e holder wil l
flash. Continue to enter the desired numbers until all four
places have a value. Press
6.If a number entry error occurs, start over by pressing
SETUP/CE.
7.When the desired value has been accepted , press EXIT to
return to the Reading mode.
3.1.8Setting the Month and Day
1.From the reading mode, press SETUP.
ENTER.
2.The
3.Change the date b y pressin g
Setup icon and the number 1 (flashing)will be displayed.
Press the up arrow until the Setup number is
will show
change to
Date and a datenear the bottom of the display.
ENTER. The numeric display will
_ _/_ _, with the left place holder fl ashing . The dat e
8. The display
format has two places for the month on the left side of the
slash and two place s for the day of the month on th e right side
of the slas h.
4.The numerical keypad will become active. Press the desired
number key for the left most digit. All four digit places must
have a number. If the left- most digit is not nec essary, use zero
for the value (i.e., 02/06 is February 6).
5.Once a number ke y is presse d, the ne xt digit plac e holder wil l
flash. Continue to enter the desired numbers until all four
places have a value. Press
ENTER.
6.If an number entry error occurs, start over by pressing
SETUP/CE.
7.When the desired value has been accepted , press
return to the Reading mode.
29
EXIT to
Page 30
SECTION 3, continued
3.1.9Setting the Year
1.From the reading mode, press SETUP.
2.The
Press the up arrow until the Setup number is
will show
3.Change the year by pressing
change to
numerical keypad will become active.
4.Press the desired number key for the left most digit. Once a
number key is pressed, the next digit place holder will flash.
Continue to enter the desired numbers until all four places
have a value. Press
5.If a number entry error occurs, start over by pressing
SETUP/CE.
6.When the desired value has been accepted , press
return to the Reading mode.
3.1.10 Setting the Print Interval
The default for this option is Off. To transfer data when this
option is disabled, press the PRINT key. When this option is
enabled, the meter will automatically send data at the selected
interval to a computer or printer via the RS232 port.
Setup icon and the number 1 (flashing) will be displayed.
9. The display
Date and a yearin the main display.
ENTER. The main display will
_ _ _ _, with the left place holder flashing. The
ENTER to accept the value.
EXIT to
1.From the reading mode, press
2.The
3.To change the print interval, press ENTER.
4.Use the arrow keys to scroll through the options:
5.Press ENTER to select the desired option. Press EXIT to leave
SETUP.
Setup icon and the number 1 (flashing) will be displayed.
Press the up arrow until the Setup number is
10. The display
will show the current setting.
Off
30 sec
0:01 min (1 minute), 0:05 min., 0:20 min.
1:00 hr, 2:00 hr, 6:00 hr
the setup.
30
Page 31
SECTION 3, continued
3.2Calibrating the Meter
The sension4 Meter is designed to auto-recognize and calibrate
on 1.68, 4.01, 6.86 or 7.00, 10.00, and 12.45 pH buffers.
Calibrating with buffers that have pH values other than these
requires a modified procedure. Each channel requires
independent calibration.
Hach recommends a daily t wo-poi nt calib ration us ing b uf fer s that
bracket the sample pH. This will verify the electrode is working
properly and allow the slope value to be stored.
3.2.1Performing a Calibration Using pH 1.68, 4, 6.86, 7, 10, and 12.45 Buffers
1.Select the desired channel. If using a probe without a
temperature sensor, see Sections 2.3.2 on page 22 and 2.6 on
page 23 for information about obtaining and using a
default temperature.
2.Prepare p H buffers that bracket the sa mple pH ac cording to
the electrode instr uction man ual. the m eter can be calibrated
with two to five autorecognized buffers.
Note: Use a 6.86 or 7.0 pH buffer for the mid-range buffer. To view or
change the setting for the mid-range buffer see Section 3.1.5.
Note: The pH values for the buffers are given for 25 °C. If the sample
temperature i s n ot 25 °C, the pH val ues d isp layed fo r the buffers wil l
reflect the correct pH value for the sample temperature.
3.Press I/O/EXIT to turn the instrument on. Select the desired
channel. If necessary, press the
reading mode (
pH will be displayed).
4.From the pH reading mode, press
appear in the upper display area, along with
PH/mV key to get into t he pH
CAL. CAL and ? will
Standard and 1.
5.Place the pH electrode in one of the buffers.
6.Press
READ/ENTER. The temperature and pH values will be
updated until a stable reading is reached.
Note: If the meter is measu ring in pH mo de, it automati cally m ov es to the
next calibration step when stabilization is reached (indicated by
three beeps). If measuring in mV mode, the three beeps will still
sound when the stabilization occurs, but you must press
ENTER to
31
Page 32
SECTION 3, continued
accept the reading. This lets the operator control the acceptance
point of the buffer.
7.When the reading has been accepted, the standard number
will change to
2.
8.Remove the probe from the first buffer and rinse with
deionized water. Place the probe in the second buffer.
9.Press
READ/ENTER. The temperature and pH values will be
updated until a stable reading is reached.
10. When the reading has been accepted, the standard number
will change to
11. Repeat steps 8 and 9 for the third, fourth, or fifth buffer or
press EXIT. The slope value and the Store and ? icons w ill
appear . Verify the slope v alue is within the range s specif ied in
the electrode manual.
12. To save the calibration and return to the reading mode, press
ENTER. To exit the calibration without saving it and return to
the reading mode, press
3.2.2Calibrating With Other Buffers
1.Select the desired channel. If using a probe without a
temperature sensor, see Sections 2.3.2 on page 22 and 2.6 on
page 23 for information about obtaining and using a
default temperature.
3.
EXIT.
2.Prepare two pH buffers that bracket the sample according to
3.Turn the instrument on. Select the desired channel. If
4.From the reading mode, press
the electr ode instruction man ual.
necessary, press the
mode (
pH will be displayed).
appear in the lower left part of the display.
PH/mV key to get into the pH reading
CAL. Functional keys will
CAL and ? will
appear in the upper display area, along with
The numeric keypad will become active.
32
Standard and 1.
Page 33
SECTION 3, continued
5.Place the pH electrode in a buffer (starting with the lowest pH
makes it ea sy to keep track).
6.Enter the pH value of the buffer using the number keys and
press
ENTER. A flashing underscore (__) indicates wh ere the
next number will be placed. It is not necessary that all four
places have a number entered in them.
7.The temperature and pH value will be updated until a stable
reading is reached.
8.When the reading has stabilized, the standard number will
change to
2. If measuring in the mV mode, press ENTER
to continue.
9.Rinse the electrode and place it in the next buffer.
10. Enter the pH value of the buffer using the number keys as
describe d above. Press ENTER.
11. When the reading has stabilized, the standard number will
change to 3. Repeat steps 9–10 for other calibration buffers.
12. Press
appear . Verify the slope v alue is within the ra nges specif ied in
the electrode manual.
13. To save the calibration and return to the reading mode, press
ENTER. To exit the calibration without saving it a nd return to
the reading mode, press EXIT.
3.2.3One-Point Calibration
The one-point calibration is used to adjust the calibration offset
and cannot be done unless a prior calibration using at least two
points is being used for measurements.
1.From the reading mode, press
will be displayed.
2.Place the electrode in a pH calibration buffer.
EXIT. The slope v al ue and the Store and ? icons will
CAL. Cal, 1, ?, and Standard
3.Press
READ/ENTER. The temperature and pH values will be
continuously updated until a stable reading is reached.
33
Page 34
SECTION 3, continued
4.When the reading has stabilized, the standard number will
change to
2.
3.3ISE Calibration
5.Press
and
EXIT. The adjusted offset value, the slope, and the Store
? icons will appear. Verify the offset value (isopotential
point) is within the range specified in the electrode manual.
6.To save the adjusted calibration and return to the reading
mode, press ENTER. To exit the calibration without saving it
and return to the reading mode, press
EXIT.
Calculation for ISE values does not require temperature
compensation, so a temperature probe is not necessary for ISE
testing. However, a temperature probe will still function properly
and display the correct sample temperature. Use the 5-pin
temperature probe available from Hach, Cat No. 51980-00.
1.Prepare standards according to the electrode instruction
manual or the instructions for the standard. Use two to five
standards for calibration. Pour 25 mL of each standard into a
50-mL beaker.
2.Turn the instrument on. Select the desired channel and the
BNC connector (Setup 1). If necessary, press the
ISE/mV key
to get into the ISE reading mode (the last concentration units
entered
3.Press
of the display.
will be displayed).
CAL. Functional keys will appear in the lower left p art
CAL, ?, and flashing units will appear.
4.To change the units, use the arrow keys to scroll to the
desired choice (g/L, mg/L, µg/L, ppm, ppb, M). Press
ENTER.
5.Add ionic strength adjustor to 25 mL of the least
concentrated sta nda rd.
Standard, and 1 will appear in the upper display. If the meter
6.
has been calibrated, the last value for standard 1 will appear.
The numeric keypad will become active.
7.Place the ISE electrode in the standard with the lowest value.
While measuring, stir the solution slowly and steadily.
34
Page 35
SECTION 3, continued
8.To enter or change th e standar d’s concentration value, use the
number keys, then press
ENTER. A flashing underscore (__)
indicates where the next number will be placed. It is not
necessary that al l four places have a num ber ente red in them.
To retain the standard concentration value, just press ENTER.
Note: Once a ne w v alue is ente red, the meter w ill assume the calibrati on
values have changed and will no longer display values for other
standards from the previous calibration.
9.The concentration value selected will be displayed until a
stable reading is reached. mV values may be observed during
calibration by pressing the
Note: If the meter is measuring in ISE mode, it automatically moves to
the next calibration step when stabilization is reached (indicated by
three beeps). If measuring in mV mode, the three beeps will still
sound when the stabilization occurs, but you must press
accept the reading. This lets the operator control the acceptance
point of t he standard.
ISE/mV key.
ENTER to
10. When the reading has stabilized, the standard number will
change to 2. The previous value for standa rd 2 will be
displayed if a new value for Standard 1 has not been entered.
11. Add ISA to 25 mL of the standard with the next highest
concentration.
12. Rinse the electrode and place it in the standard with the next
highest concentration. While measuring, stir the solution
slowly and st eadily.
13. Enter or change the standard’s concentration value using the
number keys as described above. Press
14. To retain the standard concentration value, just press
ENTER.
ENTER.
15. When the reading has stabilized, the standard number will
change to 3. Repeat steps 10–12 for each st andard . The me ter
will accept up to seven ISE calibration points and five pH
calibration points.
16. After the last standard is measured, press
? icons will appear.
35
EXIT. The Store and
Page 36
SECTION 3, continued
17. To save the calibration and return to the reading mode, press
ENTER. To exit the calibration without saving it and return to
the reading mode, press EXIT.
18. After the calibration is stored, the meter is immediately ready
to begin measuring samples. See Section 3.6 on page 38.
3.4Reviewing Calibrations
pH calibrations result in a linear calibration curve, so only one
slope v alue and i ts corre lation co ef fi cient (r
Review mode. ISE calibration result in a linear, piecewise
calibration for each pair of standards (see Figure 4). Therefore,
the meter will di splay a sl ope for ea ch “calibration” between each
pair of standards. The first slope displayed is for Standards 1 and
2. The second slope displayed is for standards 2 and 3, and so
forth. Use the arrow keys to scroll to these values.
The mV values for the standards can be viewed by pressing the
ISE/mV or pH/mV key. This toggles between concentration and
mV values.
Figure 4pH and ISE Calibration Curves
2
) are displayed in Cal
pH
36
Page 37
SECTION 3, continued
1.Select the desired channel. From the pH or ISE reading
mode, press the
REVIEW key. T o re vie w th e pH calibr ation, be
sure the meter is in pH mode. To review the ISE calibration,
be sure the met er is in ISE rea ding mode . View mV value s b y
pressing the
pH/mV or ISE/mV key.
2.The meter will display the time the ca libration was stored.
Press the up arrow key once.
3.The display will show the standard number, standard
concentration and temperature. Press the up arrow once .
4.The meter will continue to scroll through the standard
information wit h each pre ss of the u p arro w k ey. When all the
standards have been displayed, press the up arrow key again.
5.For pH calibrations, The meter will display the slope and
offset of the calib rat ion c urv e. Pres s the up ar row again to see
2
the r
value for the curve. Skip to step 8 if reviewing a
pH calibration.
For ISE cali brati ons, the mete r will displa y the sl ope v al ue of
the curve between standards 1 and 2. Press th e up arr ow o nce.
6.The meter will display the slope value for the curve between
standards 2 and 3.
7.Each press of the up arrow will display the remaining slope
values until all the slope values have been displayed.
8.To exit Cal Re view mode, press
EXIT. To review any st andard
or slope information again, press the down arrow.
37
Page 38
SECTION 3, continued
3.5Measuring Sample pH
See instructions in the electrode manual for more information and
specific steps for using the electrode.
1.Place the electrode in the sample. Press READ/ENTER.
Stabilizing... will be displayed, along with the sample
temperature and the pH or mV reading. These values may
fluctuate until the system is stable.
2.When the reading is stable,
Stabilizing... will disappear. If the
Display Lock is enabled, the display will “lock in” on the pH
or mV and sample temperature. If the Display Lock is off,
Stabilizing... will still disappear, but the display will show the
current reading and temperature and the values may fluctuate.
3.Record or store the pH or mV value. See SECTION 4 on
page 41 to store or recall data.
4.Remove the electrode from the sample, rinse with deionized
water and place the electrode in the next sample. Repeat
steps 1–3 for each sample.
5.When measurements are complete, press the
I/O/EXIT key to
turn the meter off. Rinse the electrode with deionized water
and blot dry. Replace the protective cap on the electrode. For
proper electrode storage, see the electrode manual for
specific instructions.
3.6Measuring Samples Using ISEs
This is a general guide for ISE measurements. Most
measurements require reagent addition. See instructions in the
electrode manual for more information and specific steps.
1.After calibration, rinse the electrode and blot dry. If
2.When the reading is stable
necessary, add ISA to 25 mL of sample. Place the electrode( s)
into the sam ple.
Stabilizing... will disappear. If the
Display Lock is enabled, the display will show the
concentration or mV and sample temperature when a stable
reading is reached. If Display Lock is off,
Stabilizing... will
disappear, but the current readings may fluctuate.
38
Page 39
SECTION 3, continued
3.Record or store the concentration or mV value. See
SECTION 4 on page 41 to store o r recall data.
4.Remove the electrode from the sample, rinse with deionized
water, and blot dry. Place the electrode in the next prepared
sample. Repeat steps 1–3 for each sample.
5.When measurements are complete, press the
turn the meter off. Rinse the electrode with deionized water
and blot dry. Replace the protective cap on the electrode. For
proper electrode storage, see the electrode manual for
specific inst ructions.
3.7Measuring Samples Using Standard Additions
This method should only be used within the linear range of the
sensing electrode (see the electrode manual). In this mode, the
user may also set the stabilization criteria for the channel used.
The slope of the electrode being used must be known.
1.If necessary, add ISA to the sample. Rinse the electrode and
blot dry. Place the electrode in the sample.
2.Select the desired channel. From the Reading mode, press
STD ADDN.
3.The display will show
units will be flashing. Use the arrow keys to scroll to the
desired units, then press
STANDARD ADDITIONS, and the
ENTER.
I/O/EXIT key to
4.The display will show the slope that was used for the last
standard addition measurement. If desired, enter another
value using the numerical keypad. To enter a negative slope,
enter the number first, then press the
– sign should appear in front of the slope. Press
5.The display will show
Sample, mL, a flashin g ?, and a value
REVIEW/. ± key. The
ENTER.
for the sample v olu me .
6.To change the sample volume, use the numerical keypad.
When entry is comple te or no change is needed, pr ess
ENTER.
7.The meter will measure the sample for a b aseline voltage.
Stabilizing... will appear until the mV read ing is stable.
39
Page 40
SECTION 3, continued
8.After stabilizing, the display will show Standard, a flashing ?,
mL, and a value for the standard volume.
9.To change the standard volume, use the numerical keypad.
When entry is c ompl ete or no change is needed, pre ss ENTER.
10. The display will show
Standard, a flashing ?, units, and a
value for the standard concentration.
11. To change the standard concentration, use the numerical
keypad. Immediately add the selected volume of the selected
standard to the sample. Press
ENTER. Do not press ENTER
until the standard has been added to the sample.
12. The meter will return to Reading mode. The + icon in front of
Standard indicates that standard has been added to
the sample.
13. During measurement t he meter wi ll display a mV value.
After stabilization, the meter will calculate and display the
value of the original sample.
14. If desired, press
recalled
STANDARD ADDTIONS will appear in the display.
STORE to store this data . When this data is
40
Page 41
SECTION 4STORING AND RECALLING DATA
4.1Storing pH/ISE Measurements
The sension™4 meter can store up t o 99 mea sur ement readings.
Data must be stored to recall it later for review, downloading, or
printing. Although the meter display will only show the
temperature, data location, and pH/IS E value, the following
information is stored (and can be downloaded or printed) for
each sample:
•storage location•mV reading
•software ver sion•sample tempera tur e
•date•pH or ISE value
•time
•channel
The new data is saved in the next available memory l ocation,
numbered from 1 to 99. If no memory locations higher than the
current one are available, the meter will “wrap around” and
choose the next av aila ble locati on. The user also has the option of
choosing the storage location.
To store data:
1.Press
the next av ailable location).
2.Press
location number. To sto re the data i n another lo cation, use the
arrow keys to scrol l to anot her locati on or enter a locatio n
using the number keys. Press
Note: If al l memory loca tions are fu ll, th e mete r will prompt to o v erw rite a
data point by displaying
the data in that location with the current data. Press EXIT to retur n
to the previous screen without replacing the data.
STORE. The display will prompt Store Sample #? (# is
ENTER to store the measurement reading in that
•instrument model and serial
number
The question mark will be f lashi ng.
ENTER.
Erase Sampl e ##? Press ENTER to replace
3.The meter will store the reading and ret urn to Read ing mode.
41
Page 42
SECTION 4, continued
4.2Recalling Stored Data
1.To recall stored d ata, press the RECALL key while in the
Reading mode. The screen will display the most recently
saved or recalled measurement data.
2.Use the arrow keys to scroll to the desired storage location.
Press
location. The question mark will flash. Enter the number of
the desired storage location. Press
storage location or
RECALL again to allow number entry of a storage
ENTER to accept the
EXIT to escape.
3.To display the mV value of the reading, press the
ISE/mV key.
4.To review the time of the calibration, press
review the date of the calibration, press
EXIT to return the reading value.
5.When recalling is complete, press EXIT to return to the
Reading mode.
4.3Erasing Data
4.3.1Erasing Single Data Points
1.To erase data, it must be recalled first. Press the RECALL key
while in the Reading mode. See Section 4.2 for
more information.
2.When the desired data point is displayed, press
pH/mV or
TIME once. To
TIME twice. Press
ERASE.
3.The meter will displ ay
4.The meter will recall the most recently stored or recalled.
5.Repeat steps 2-3 for each data point that needs to be deleted.
Erase Sample # and a flashing ?. Press
ENTER to erase the data. T he data will disappear.
There are three options at this point:
•Press ERASE, then ENTER, to erase the displayed data.
•Press EXIT to exit Recall mode.
•Press an arrow key to scroll to other data points.
42
Page 43
SECTION 4, continued
4.3.2Erasing All Data Points
1.To erase data, it must be recalled first. Press the RECALL key
while in the Reading mode. See Section 4.2 for
more information.
2.When the data point is displayed, press
3.Press the up arrow. Th e instrument will show
with the ? (flashing). At this point the options are:
ERASE.
Erase and All
•Press EXIT to return to Recall mode without erasing.
•Press the down arrow to return to the single point
erase prompt.
•Press ENTER to erase all da ta and return to Reading
mode.
4.After all t he data is erased, the meter will return to the
Reading mode.
43
Page 44
44
Page 45
SECTION 5PRINTING/TRANSFER RING DATA
5.1Connecting to Printers/Computers
5.1.1Connecting with the RS232 Cable
The standard 9-p in RS232 connect or on the meter connects with a
9-pin D-sub connector. A suitable cable is listed under Optional Apparatus page 57.
The RS232 interfac e output is an 8-bit d ata w ord plus one stop bit
and no parity with a baud rate of 1200 . It can commu nicate wit h a
serial printer or a serial port on a computer.
5.1.2Connecting to a Printer
Connecting a ser ial pr inter to the meter requir es a 9-pin to 25-pin
RS232 cable. See Figure 5. The cable provides a direct link
between the instrument and the 25-pin connector used for the
serial port on most serial printers. Table 4 shows the proper pin
connections for 25-pin printer cables. Using cables that do not
match the pin information in the table may cause
undesirable operation.
Parallel printers require a serial-to-parallel adapter. This allows
use of printers that are normally used for IBM-compatible
applications.
The Citizen PN60 pr inter requires a sp ecial cabl e to fit into the
printer. This cable is shipped with the printer when ordered from
Hach Company.
Connect the RS232 cable to the meter by lining up the holes in
the cable connector with the pins of the meter serial port. Gently
and firmly push the cable into the meter. Then secure the
connection by screwing in the screws on either side of the cable
port. Connect the cable to the printer in the same manner. Once
the communication link is esta blished, pres s
the computer.
Note: For optimum performance and ESD protection, use a five-
conductor shielded cable. Use a metal shell for the printer or
computer terminal connector, and connect the shield of the cable to
the metal shell and the sleeve (signal ground) of the RS232 plug.
Follow the printer manufacturer’s instructions to configure
the printer for compat ibility with the meter.
Figure 5RS232 Cable Connector
PRINT to send data to
5.1.3Connecting to a Personal Computer
Connect the meter to a personal computer (PC) with the computer
interface cable (Cat. No. 48129-00) listed under Optional Accessories page 57. The cable pro vides a direct l ink betwee n the
meter and the 9-pin D connector used for the serial port on most
personal computers. If your computer has a 25-pin D connector,
use a 9-pin to 25-pin adapter (available at most computer
supply stores).
46
Page 47
SECTION 5, continued
Table 5 shows the proper pin connections for 9-pin computer
cables. Using cables that do not match the pin information in the
table may cause undesirable operation.
Table 5 Standard 9-pin to 9-pin Computer Cable
9-pin D Connector SocketComputer 9-pin D Connector, plug
Connect the RS232 cable to the meter by lining up the holes in
the cable connector with the pins of the meter serial port. Gently
and firmly pus h the cable into the meter. Then secure the
connection by screwing in the screws on either side of the cable
port (see Figure 5). Connect the cable to the computer in the
same manner.
To transfer data, th e commun ication p arameters (baud rate, data
bits and parity) of the meter and the computer must match. Once
the communication link is esta blished, pres s
PRINT to send data to
the computer.
Use a communicat ions software, such as HachLink™
(Cat. No. 49665-00) to collect data from the instrument.
HachLink is a Windows-based application that allows a personal
computer to capture data from several Hach instruments,
including the sens
ion electrochemi cal meter s. The captu red data
can be stored in a text file as a spread-sheet compatible format or
a free-format text. Data captured in the sp readsheet format is
easily transferred into most spreadsheet programs
(i.e., MS
®
Excel, Microsoft® Works, Lotus 1 23) f or graphi ng and
reporting.
To install and run HachLink™ Data Capture, the computer and
software must meet the following minimum requirements:
47
Page 48
SECTION 5, continued
•IBM PC/AT or compatible with a 386SX processor (16 MHz
or better)
•4 megabytes of RAM
•Hard disk drive with 2 megabytes or more of free space
•3-1/2 inch, 1.44 megabyte floppy disk drive
•VGA graphics with 640 x 480 or higher resolution, 16 or
more colors
•Mouse or other pointing device
•A 9-pin serial port (or 25-pin serial port with 9-pin adapter)
•Windows 3.1 or later
•DOS 3.3 or later
5.2Sending Data to Printers/Computers
5.2.1Printing Data at Intervals
The last option in the Setup menu lets the user choose a print
interval. When this option is enabled, the mete r will
automatically send data to a printer or computer at the selected
time interval. See Section 3.1.10 on page 30 for more information
on enabling this option
5.2.2Sending Currently Displayed Data
To print or transfer a current reading:
1.Wait until the display is stable. Press
Note: Data can be printed before the reading is stable. An asterisk will
be printed with the measurem ent data to indicate the stab ility crit eria
were not met.
PRINT.
2.The word PRINT will be briefly displayed, then the meter
3.The printout for data that is printed without being stored will
will return to Reading mode.
not have a storage location number.
48
Page 49
SECTION 5, continued
5.2.3Sending Recalled Data Points
1.To transfer data, it must be recalled first. Press the RECALL
key while in the Reading mode. See Section 4.2 on page 42
for more information.
2.When the desired sample data is displayed, press
3.The word
PRINT will be briefly displayed, then the meter
will return to Reading mode.
5.2.4Sending All Stored Data and the Average pH Value
1.To transfer all data, it must be recalled first. Press the
RECALL key while in the Reading mode. See Section 4.2 on
page 42 for mo re informa tion.
2.When a data poi nt i s d is played, press PRINT. The display will
show
Print Sample # ?.
3.Press the up arrow. Th e instrument will show
and All with the flashing ?. At this point the options are:
•Press EXIT to return to Reading mode without printing.
•Press the down arrow to return to the prompt for pr inting
single data points.
PRINT.
Print, Sample
•Press ENTER to pr int all d ata a nd the a v er age pH of the s tored
data (data that is printed but not stored will not be included).
The word
printed. Then the meter will return to the most recently stored
sample data. Press
arrow key to scroll to another data point.
PRINT will be displayed until all the data has been
* This line will not be included unless Print All is chosen. This calculation is included for some reporting
regulations, but is not just an average of the pH readings. The calculation for the pH average is:
n
pH
AVG
∑
i=1
-log
=
------------------------------- -
10
pH–i()
10
()
n
5.2.5PC Control of the Meter through the RS232 Interface
The meter can be controlled remotely using a PC. The
communication protocols are 1200 baud rate, 8-bit word, 1-stop
bit, no parity. A summary of the commands is shown in Table 6.
50
Page 51
SECTION 5, continued
Table 6 sension4 RS232 Commands
Key
Instrument IDIID
Exit KeyEXTOK
Up ArrowUPAOK
Down ArrowDOWOK
EnterENTOK
RecallRECOK
StoreSTOOK
EraseERAOK
ISE/mVISEOK
pH/mVPHMOK
Std AddnSTDOK
PrintPRIOK
ChannelCHAOK
Time/DateTIMOK
CalCALOK
Correcting problem conditions with the sension electrochemical
meters is limited to responding to the error message displayed.
Other problems must be handled b y a Hach technic ian at a service
center. Refer to Repair Service. Do not attempt to service the
meter as there are no field-serviceable parts. Opening the meter
case will void the warranty.
6.2Shorting Test
This test detects the mete r offset
1.Turn the meter on. Connect the shorting cap to the BNC
connector (see Figure 6). Make sure the channel with the
shorting cap is selected. Select BNC from Setup 1 choices.
2.Change the output to display in mV (press the ISE/mV or
pH/mV key to toggle between mV and concentration). After
the meter has stabilized, simultaneously press
ENTER. The display should show 0.0 mV. If it does not,
contact Hach Service. Repeat for the other channel.
CAL and
Figure 6Attaching Shorting Caps to the BNC Connector
53
Page 54
SECTION 6, continued
6.3Error Codes And Errors
Errors and error codes indicate a functional problem with the
meter and/or the electrode. Error codes numbers will appear in
the temperature field along with other icons. Table 7 describes
the possible error codes, errors, and some possible solutions to
eliminate the cause.
Table 7 Error Codes
Error CodeError Name & Display IconsDescription
Calibration slope is too high or low.
Ensure correct pH buffers are used. Be sure a pH
probe is connected to the meter. Disconnect any
probe connected to the BNC connector.
The value read will not work in the calibration
algorithm. Repeat calibration with new buffers
or standards.
Service.
E2
E3
E9
Cal slope error.
SLOPE will be displayed.
Cal Std Error.
ERROR, CAL and Standard
will be displayed.
Corrupt dataRecalled data had a bad checksum. Call Hach
ERROR and
6.4Meter Service Request Questionnaire
1.What is the complete l ot code of the meter and electrode?
2.On what date was the meter purchased?
3.How long has th e meter been in use?
4.What types of samples are being tested?
5.What is the temperature of the samples being tested?
6.How often is the meter being used?
7.How is the meter being stored bet w een use s?
8.If the meter has been in use for a while, what maintenance
9.Describe the suspected problem or failure of the meter.
10. Please have your meter, electrode, buffers/standards , and this
has been performed?
completed questionnaire near the phone before calling
technical support.
54
Page 55
GENERAL INFORMATION
At Hach Company, customer service is an
important part of ev e ry product we make.
With that in mind, we have compiled the
following informa t ion for your convenience.
Temperature Probe, 5-pin ........................................................................each .............51980-00
Thermometer, mercury, -20 to 110 °C.....................................................each .............20959-11
58
Page 59
HOW TO ORDER
Information Required
By Telephone:By Mail:
6:30 a.m. to 5:00 p.m. MSTHach Company
Monday through FridayP.O. Box 389
(800) 227-HACHLoveland, Colorado 80539-0389
(800-227-4224)U.S.A.
By FAX:Ordering informationby E-mail:
(970) 669-2932orders@hach.com
•Hach account number (if available)•Billing address
•Your name and phone number•Shipping address
•Purchase order number•Catalog number
•Brief description or model number•Quantity
Technical and Customer Service (U.S.A. only)
Hach Technical and Customer Service Department personnel
are eager to answer questions about our products and their use.
Specialists in analytical methods, they are happy to put their
talents to work for you. Call 1-800-227-4224 or E-mail
techhelp@hach.com.
International Customers
Hach maintains a worldwide network of dealers and distributors.
To locate the representative nearest you, send E-mail to
Hach warrants most products against defective materials or
workmanship for at least one year from the date of shipment;
longer warranties may apply to some items.
HACHWARRANTSTOTHE ORIGINAL BUYERTHAT HACH
PRODUCTSWILL CONFORMTO ANY EXPRESSWRITTEN
WARRANTY GIVEN BY HACHTOTHE BUYER. EXCEPT AS
EXPRESSLY SET FORTH INTHE PRECEDING SENTENCE,
HACH MAKES NOWARRANTY OF ANY KINDWHATSOEVER
WITH RESPECTTO ANY PRODUCTS.HACH EXPRESSLY
DISCLAIMS ANYWARRANTIES IMPLIED BY LAW,
INCLUDING BUT NOT BINDINGTO ANYWARRANTY
OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE.
LIMITATION OF REMEDIES:
Hach shall, at its option, replace
or repair nonconforming products or refund all amounts paid by
the buyer.
BREACH OFWARRANTY.
LIMITATION OF DAMAGES: IN NO EVENT SHALL HACH
BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL
DAMAGES OF ANY KIND FOR BREACH OF ANYWARRANTY,
NEGLIGENCE, ONTHE BASIS OF STRICT LIABILITY, OR
OTHERWISE.
THIS ISTHE EXCLUSIVE REMEDY FOR ANY
This warranty applies only to Hach products purchased and
delivered in the United States.
Catalog descriptions, pictures and specifications, although
accurate to the best of our knowledge, are not a guarantee
or warranty.
For a complete description of Hach Company’s warranty policy,
request a copy of our Terms and Conditions of Sale for U.S. Sales
from our Customer Service Department.
Hach warr ants the meter against defective materials or
workmanship thr ee years from the dat e of shipment. The Docking
Station is warranted for one year after the date of shipment.