Hach Company certifies this instrument was tested thoroughly,
inspected and found to meet its published specifications when it
was shipped from the factory.
The sension™4 Laboratory pH/ISE Mete r has been teste d and is
certified as indicated to the following instrumentation standards:
External Power Supplies Only:
115 VAC Supply: UL listed and CSA certified or
230 VAC Supply: CE marked per 73/23/EEC, VDE listed
Instrument tested with e xt er nal 230V, 50 Hz power supply.
Per 89/336/EEC EMC: EN 61326:1998 (Electrical Equipmen t
for measurement, control, and laboratory use— EMC
requirements) Supporting test records by Hach Company,
certified compliance by Hach Company.
Standards include:
IEC 1000-4-2: 1995 (EN 61000-4-2:1995) Electro-Static
Discharge Im munity (Criteria B)
IEC 1000-4-3: 1995 (EN 61000-4-3:1996) Radiated RF
Electro-Magnetic Fields (Criteria B)
IEC 1000-4-4: 1995 (EN 61000-4-4:1995) Electrical Fast
Transients/Burst (Criteria B)
IEC 1000-4-5: 1995 (EN 61000-4-5:1995) Surge (Criteria B)
IEC 1000-4-6: 1996 (EN 61000-4-6:1996) Conducted
Disturbance Induced by RF Fields (Criteria A)
IEC 1000-4-11: 1994 (EN 61000-4-11:1994) Voltage Dips,
Interruptions and Variations (Criteria B)
ENV 50204:1996 Radiated Electro-Magnetic Field from
Digital Telephones (Criteria B)
5
CERTIFICATION, continued
Emissions:
Instrument tested with external 230V, 50 Hz power supply.
Per 89/336/EEC EMC: EN 61326:1998 (Electrical Equipment
for measurement, control, and laboratory use— EMC
requirements). Class B emission limits. Supporting test records
by Hewlett Packard Hardware Test Center, Ft. Collins, CO
(A2LA #0905-01), certified compliance by Hach Company.
Standards include:
EN 61000-3-2 Harmonic Disturbances Caused by
Electrical Equipment
EN 61000-3-3 Voltage Fluctuations (Flicker) Disturbances
Caused by Electrical Equipment
Additional Standards include:
EN 55011 (CISPR 11) Emissions, Class B Limi ts
Additional Emissions Standard/s include:
CANADIAN INTERFERENCE-CAUSING EQUIPMENT
REGULATION, IECS-003: Class A emission limits.
Supporting test records by Hewlett Packard Hardware Test
Center, Ft. Collins, CO (A2LA #0905-01), certified compliance
by Hach Company.
This Class A digital apparatus meets all requirements of the
Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les
exigences du Règlement sur le matériel brouilleur du Canada.
FCC PART 15, Class “A” Limits:
Supporting test records by Hewlett Packard Hardware Test
Center, Ft. Collins, CO (A2LA #0905-01), certified compliance
by Hach Company.
This dev ice co mplies with P art 15 of the FCC Rules. Operat ion i s
subject to the following two conditions:
6
CERTIFICATION, continued
(1) This device may not cause harmful interference, and (2) this
device must accept any interference received, including
interference that may cause undesired operation.
Changes or modifications to this unit not expressly approved by
the party responsible for compliance could void the user's
authority to operate the equipment.
This equipment has been tested and found to comply with the
limits for a Class A digita l de vic e, pursua nt to P art 15 of th e FCC
Rules. These l i mits are designed to pr o vide reasonable prote cti on
against ha rmful interferenc e when the equipment is operated in a
commercial environment. This equipment generates, uses, and
can radiate radio frequency energy and, if not installed and used
in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this
equipment in a resid ential are a is likely to cause harmful
interference, in which case the user will be required to correct the
interference at his own expense. The following techniques of
reducing the interference problems are applied easily.
1.Disconnect the external power supply from sens
Laboratory pH meter to verify that the meter is not the source
of interference.
2.Move the meter away from the device receiving
the interference.
3.Reposition the receiving antenna for the device receiving
the interference.
4.Try combinations of the above.
ion4
7
8
SAFETY PRECAUTIONS
Section 2.2 on page 21
Section 2.3 on page 21
Section
2.4 on page 23
Section 5.1 on page 45
9
10
SPECIFICATIONS
Specification subject to change without notice.
pH mode
Range-2.00 to 19.99
Resolution0.001/0.01/0.1 (selectable)
Slope range45 to 65 mV/decade at 25 °C
ISE mode
Range0.000 to 19900
ResolutionAuto-ranging
Millivolt mode
Range-2000 to 2000 mV
Resolution0.1 mV
Accuracy±0.2 mV or ±0.05% of reading,
Temperature mode
Range-10.0 to 110 °C (can also display °F)
Resolution0.1 °C
Accuracy±0.3 °C from 0–70 °C
whichever is greater
±1.0 °C from 70–110 °C
Display: Custom LCD
Inputs: 2 BNC; two 5-pin Hach pH /temperature or Hach
temperature probe; 2 pin-tip; (one combination input type for
each of th e two channels or two half-cells per chan nel).
Outputs: Two-way RS232
Power Re quirements: 6–9 VDC provided by 120 V 50/60 Hz or
230 V, 50 Hz external power supply or a customer-provided
supply with 50 mA output, 5.5-mm power plug with a 2.5 mm
center post opening.
Installation Cate gory: II (for external power supplies)
Input Impedance: >10
12
ohms
Instrument Drift: <40 µV/°C
Input Bias Current: <±1 picoamp at 25 °C; <±4 picoamp over
full range
11
SPECIFICATIONS, continued
Envi ronmental Requirements: 0 to 50 °C at 85%
non-condensing relative humidity. Enclosure is water resistant,
chemical resistant, dust proof.
Dimensions: 25.4 x 15 x 8.37 (10.15 x 6 x 3.35 in.)
12
OPERATION
DANGER
Handling chemical samples, standards, and reagents can be dangerous. Review the necessary
Material Safety Data Sheets and become familiar with all safety procedures before handling
any chemicals.
DANGER
La manipulation des échantillons chimiques, étalons et réactifs peut être dangereuse. Lire les Fiches
de Données de Sécurité des Produits (FDSP) et se f amiliariser avec toutes les procédures de sécurité
avant de manipuler tous les produit s chimiques.
PELIGRO
La manipulación de muestras química s, estándares y reactivo s puede ser peligrosa. Revise las fichas
de seguridad de materiales y familiarícese con los procedimientos de seguridad antes de manipular
productos químicos.
Das Arbeiten mit chemischen Proben, Standards und Reagenzien ist mit Gefahren verbunden.
Es wird dem Benu tzer di eser Pr oduk te empf ohlen , sic h v or der Arbei t mit s ich eren Verfahrensw eisen
und dem richtigen Gebrauch der Chemikalien vertraut zu machen und alle entsprechenden
Materialsicherheitsdatenblätter aufmerksam zu lesen.
A manipulação de amostras, padrões e reagentes químic os pode ser perigosa. Reveja a fo l ha dos
dados de segurança do material e familiarize-se com todos os procedimentos de segurança antes
de manipular quaisquer produtos químicos.
GEFAHR
PERIGO
13
14
SECTION 1INTRODUCTION
Hach provides pH and ISE meters for applications from pH
measurements to accurate Ion Sel ect ive Electrode work. This
manual describes the operation and use of the Hach sens
Portable p H/ISE/mV/Temperature Meter (see Figure 1).
This meter features a custom digital LCD display whic h
simultaneously displays temperature and measurement results.
This meter has all the features of a simple pH/ISE meter plus:
•two measurement channels
•a millivolt mode
•standard additions program
•automatic buffer recognition when in the pH mode
•pH averaging
•calibration review
ion™4
•datalogging
•bi-directional RS232 communication for printing results to a
computer or print er, or controlling the meter from a PC.
The meter is designed to be maintenanc e- fr ee. If the met er get s
dirty, wipe the surface with a damp cloth. Use a cotton-tipped
applicator to clean or dry the connectors if they get wet.
1.1Unpacking the Instrument
Remove the instrument and accessories from the shipping
container and inspect each item for any damage that may have
occurred during shipping. Verify that all ite ms listed on the
packing slip are included. If any items are missing or damaged,
contact Hach Customer Service, Loveland, Colorado for
instructions. Hach’s toll free phone number for customers within
the United States is 800-227-4224. For customers outside the
United States, contact the Hach office or distributor serving you.
1.2Standard Accessories
•AC to DC wall adapter
•Instrument Manual
•Include electrode and related accessories (covered in the
electrode manual).
15
SECTION 1, continued
1.3Keypad Description
Fi gure 1 illustrate s the meter’s keypad. The description and
function of each key are given in Table 1.
Figure 1Hach sension4 Meter
16
SECTION 1, continued
Table 1 Keys and Description
KeyDescription
Exit/Power On-OffTur ns the instrument on; turns it off from a Reading mode.
Acts as a NO or Cancel key when the question mark icon is flashing.
In Setup mode, backs up one step toward the Reading mode.
Performs the following and returns to the most recent Reading mode:
• Exits the Store or Recall mode
• Aborts a calibration
• Exits a calibration review
Arrow KeysScrolls between op tions in Setup mode.
Scrolls through data points in Store and Recall modes.
Scrolls between the option to print or erase one data point and all data points.
Changes the default temperature when a temperature probe is not in use.
READ/ENTER
Key
Recall KeyRecalls stored sample data (from Reading mode only).
Store KeyStores the current (displayed) measurement (from Reading mode only).
Erase KeyErases recalled data points.
ISE/mV KeyToggles between ISE concentration value and mV in Reading, Calibr ation, and Cal
pH/mV KeyToggles between pH value and mV value in Reading, Calibration, and Cal
Std AddnMoves meter into program to use for standard additions.
Print KeySends current or recalled data to a printer or a computer via the RS232 port.
ChannelToggles between channel 1 and 2. Will not toggle in Calibration, Cal Review, and
Accepts numerical input.
Acts as a “YES” answer when the question mark is flashing.
Allows user to edit a setup when the setup number is flashing.
Accepts the current Setup option when that option is flashing.
Initiates a new measurement when the meter has stabilized in the Display Lock
Enabled mode.
Review modes.
Review modes.
Recall modes.
Time KeyIn Reading mode, allows user to view the current time and date. In Recall Data
mode, it toggles between the time and date of the stored measurement.
Cal KeyEnters Calibration mode (from Reading mode only)
Review KeyEnters Calibration Review mode (from Reading mode only)
Setup/CE K e yEnters Setup mo de (from Rea ding mode only) or clears a numeric entry when the
keypad icon is displayed.
17
SECTION 1, continued
1.4Display Fields and Icons
The display has two screens. The upper screen displays
measurements or standard values, operation mode, slope,
sample/default temperature, pH or mV units, error codes, and
indicates if the meter reading is stable. The lower screen displays
keys that are active.
Fi gure 2 shows the icons and screens displayed by the meter and
Table 2 describes each element. All icons on the display will be
displayed if the power key is held down for several seconds.
Figure 2sension4 Display
18
SECTION 1, continued
Table 2 Main Display Elements
Item No.Description
1Indicates meter is in Standard Addition mode.
2 (Cal
Review)
3Used with
4Indicates the meter is measuring/reviewing a sample (sample # is displayed to the right).
5Indicates data is being sent to a printer/computer.
6Indicates recalled data that is currently displayed is being erased.
7Indicates meter is in Setup mode.
8Indicates all data points are being printed or erased.
9Numerical field tha t dis pl ays Setup, Sampl e . an d Sta nda rd numbers when those words are
10Flashing
11Indicates the meter is measuring/reviewing a standard (standard # is displayed above).
12Numerical field that displays the values of standards and samples, slope, pH or mV.
13Indicates measurement units (pH, mV, mg/L, µg/L, ppm, ppb, M).
14When
Indicates meter is in Calibration Review mode.
? to ask if user wants to store calibration, standard value, or sample value
displayed with the number. If
Standard 1.
Standard and 1 are displayed, the meter is measuring
? and CAL indicate calibr ati on is nece ssary. Also a prompt to pres s the ENTER or
EXIT key.
Default is displaye d, the meter is using the def ault tempera ture valu e to calculate the
temperature correction for the pH value.
15Temperature units (choice of °C or °F).
16Indicates value displayed in small numerical field (item 20) is in millivolts.
17Indicates channel 2 is in use.
18Indicates an inactive key has been pressed and that function is not allowed.
19Indicates ENTER key is active.
20Numerical field that displays temperature value.
21Indicates arrow keys are active.
22Indicates the date is being set or displayed.
23Indicates EXIT key is active.
24Faulty probe connection or incorrect probe attached. Usually displayed with an error code.
25Indicates numeric part of the keys is active.
26Indicates channel 1 is in use.
27NA
28Display Lock icon. Indicates reading is locked after stability is achieved.
29In setup mode, indicates whether Display Lock setting is
30Indicates a meter function problem.
On or Off.
19
Loading...
+ 43 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.