Hach sc100 1720E Instrument Manual

Page 1
Catalog Number 60100-18
Hach sc100 1720E Analysis System
Instrument Manual
11/03 2ed
© Hach Company, 2003. All rights reserved. Printed in the U.S.A. ds/dp
Page 2
Table of Contents
Section 2 General Information ..................................................................................................................................... 6
2.1 Safety Information ..................................................................................................................................................... 6
2.2 General Product Information ..................................................................................................................................... 6
2.3 Theory of Operation................................................................................................................................................... 7
Section 3 Installation..................................................................................................................................................... 8
3.1 Mechanical Installation .............................................................................................................................................. 8
3.1.1 Controller Dimension Illustrations .................................................................................................................... 9
3.1.2 Using the Optional Sun Shield ....................................................................................................................... 11
3.1.3 Mounting the Controller.................................................................................................................................. 12
3.2 Electrical Installation................................................................................................................................................ 14
3.2.1 Installation in Conduit..................................................................................................................................... 14
3.2.3 Wiring for Power at the Controller.................................................................................................................. 15
3.3 Alarms and Relays .................................................................................................................................................. 18
3.3.1 Connecting the Relays ................................................................................................................................... 18
3.3.2 Connecting the Analog Outputs ..................................................................................................................... 18
3.4 Connecting/Wiring the Sensor Cable ...................................................................................................................... 19
3.5 Connecting the Optional Digital Output ................................................................................................................... 21
3.6 Turbidimeter Installation Information ....................................................................................................................... 24
3.6.1 Mounting the Turbidimeter Body.................................................................................................................... 24
3.6.2 Installing the Head Assembly......................................................................................................................... 24
3.7 Installing a Sample Line .......................................................................................................................................... 24
3.8 Sample Connections ............................................................................................................................................... 25
Section 4 Operation..................................................................................................................................................... 26
4.1 Using the Keypad .................................................................................................................................................... 26
4.2 Controller Display Features ..................................................................................................................................... 27
4.2.1 Important Key Presses................................................................................................................................... 27
4.3 Instrument Setup ..................................................................................................................................................... 28
4.3.1 Software Text Abbreviation Conventions....................................................................................................... 28
4.3.2 Adjusting Display Contrast............................................................................................................................. 28
4.3.3 Specifying the Displayed Language............................................................................................................... 29
4.3.4 Setting the Time and Date ............................................................................................................................. 29
4.4 Changing the Sensor Name .................................................................................................................................... 30
4.4.1 Setting Up System Security ........................................................................................................................... 31
4.5 Output Options ........................................................................................................................................................ 32
4.5.1 Output Setup Menu (from System Setup)...................................................................................................... 32
4.5.2 Hold/Transfer Outputs.................................................................................................................................... 33
4.5.3 Release Outputs ............................................................................................................................................ 33
4.6 Relay Options .......................................................................................................................................................... 34
4.6.1 Relay Setup Menu (from System Setup) ....................................................................................................... 34
4.7 Data and Event Logging Options............................................................................................................................. 36
4.7.1 Data Logging Options .................................................................................................................................... 36
4.8 Digital Network Options ........................................................................................................................................... 36
4.9 Menu Structure ........................................................................................................................................................ 36
4.9.1 Sensor Diagnostics Menu .............................................................................................................................. 36
4.9.2 Sensor Setup Menu ....................................................................................................................................... 37
4.9.3 System Setup Menu....................................................................................................................................... 37
4.9.4 Test/Maint Menu ............................................................................................................................................ 38
2
Page 3
Table of Contents
Section 5 System Startup ........................................................................................................................................... 40
5.1 General Operation ................................................................................................................................................... 40
5.2 Starting Sample Flow .............................................................................................................................................. 40
5.3 Sensor Calibration and Verification ......................................................................................................................... 40
5.3.1 User-prepared Calibration.............................................................................................................................. 41
5.3.2 Calibration with StablCal
5.4 Instrument Verification............................................................................................................................................. 43
5.4.1 Dry Verification............................................................................................................................................... 44
5.4.2 Wet Verification .............................................................................................................................................. 45
5.5 Calibration and Verification History ......................................................................................................................... 46
6.1 Maintenance Schedule ............................................................................................................................................ 47
6.2 Removing a Sensor from the System...................................................................................................................... 47
6.3 Reinstalling a Sensor on the System....................................................................................................................... 47
6.4 Cleaning .................................................................................................................................................................. 48
6.4.1 Cleaning the Controller .................................................................................................................................. 48
6.4.2 Cleaning the Photocell Window ..................................................................................................................... 48
6.4.3 Cleaning the Turbidimeter Body and Bubble Trap......................................................................................... 48
6.4.4 Replacing the Lamp Assembly ...................................................................................................................... 49
6.5 Fuse Replacement .................................................................................................................................................. 51
® ...............................................................................................................................................................42
Section 7 Troubleshooting.......................................................................................................................................... 52
7.1 Error Codes ............................................................................................................................................................. 52
7.2 Warnings ................................................................................................................................................................. 52
7.3 Event Codes ............................................................................................................................................................ 54
Section 8 Replacement Parts and Accessories........................................................................................................ 55
9.1 Information Required ............................................................................................................................................... 57
9.2 International Customers .......................................................................................................................................... 57
9.3 Technical and Customer Service (U.S.A. only) ....................................................................................................... 57
Section 10 Repair Service........................................................................................................................................... 58
Section 11 Limited Warranty ...................................................................................................................................... 59
Section 10 Compliance Information .......................................................................................................................... 60
Appendix A ModBUS Register Information .............................................................................................................. 62
Index ............................................................................................................................................................................. 70
3
Page 4
Section 1 Specifications
Specifications are subject to change without notice.
Table 1 Turbidimeter Specifications
Nephelometric light scatter at 90 degrees relative to the incident light beam. The incident light
Method of Detection
Range 0.001–100 nephelometric turbidity units (NTU)
beam is composed of a tungsten filament light source with a color temperature between 2200 and 3000K.
Accuracy
Linearity
Resolution (Displayed) 0.0001 NTU up to 9.9999 NTU; 0.001 NTU from 10.000 to 99.999 NTU
Repeatability Better than ±1.0% of reading or ±0.002 NTU, whichever is greater for each range.
Response Time
Sample Flow Required 200 to 750 mL/minute (3.2 to 11.9 gal/hour)
Sensor Storage Temperature
Operating Temperature 0 to 50 °C (32–122 °F) for single sensor system, 0 to 40 °C (32–104 °F) for two sensor system
± 2% of reading or ± 0.015 NTU (whichever is greater) from 0 to 40 NTU; ± 5% of reading from 40 to 100 NTU
Better than 1% 0–40 NTU on formazin. Allows for accurate calibration at high turbidity values. Temperature dependent ±2 °C.
For a full-scale step change, initial response in 1 minute, 15 seconds. Varies with flow rate, see the table below. The response time is also dependent on the signal averaging time, which is user selectable.
Flow Rate
% Step Change
750 500 250
10 1¼ minutes 1½ minutes 2½ minutes 50 2 minutes 2½ minutes 6 minutes 90 3½ minutes 3½ minutes 9 minutes 99 4 minutes 5 minutes 12 minutes
–20 to 60 °C (–4 to 140 °F); 95% relative humidity, non-condensing.
Sample Temperature Range 0 to 50 °C (32–122 °F)
Operating Humidity 5 to 95% non-condensing
Power Requirements 12 V dc ±5%, 12.5 watts maximum
Sample Inlet Fitting ¼-inch NPT female. ¼-inch compression fitting (supplied)
Drain Fitting ½-inch NPT female, ½-inch hose barb (supplied)
Signal Average (Filter) Time no averaging, 6, 30, 60, and 90 seconds, user selectable. Default is 30 seconds.
Sensor Dimensions Turbidimeter body and cap: 25.4 x 30.5 x 40.6 cm (10 x 12 x 16 inches)
Sensor Cable Length 2 m (6.6 ft); Optional 7.62 m (25 ft) extension cable. Maximum cable length is 9.62 m (31.6 ft).
Sensor Cable Rating
Mounting Options Wall; floor stand
Shipping Weight 1720E Turbidimeter and Controller: 6.31 kg (13.5 lb); 1720E Turbidimeter only: 4.71 kg (10 lb)
Cable: 105 °C, 300 V, PVC jacket Wires: 22 AWG, PVC jacket
4
Page 5
Table 1 Turbidimeter Specifications (continued)
1. StablCal
Calibration Methods
2. Formazin – user-prepared primary or wet calibration of the instrument
3. Multi-sensor calibration – Performed with a specialized calibration procedure for up to eight
sensors on a single set of fresh StablCal
1. StablCal
Verification (Wet) Method
range of measurement. For regulatory verification, standards of 0.1. 0.3. 0.5 and 1.0 NTU are available.
2. Formazin – fresh user-prepared standard
1. ICE-PIC™ Verification Module with factory-set values of 20.0 or 1.0 ±25%. Unique value is
Verification (Dry) Method
assigned when dry verification is done immediately after calibration and is used as pass/fail criteria for subsequent verifications.
1. Mandatory before calibration
Recommended Cleaning Intervals
2. Optional before verification
3. Mandatory upon verification failure
Languages English (default), German, Spanish, Nederlands
Installation Environment Indoor
Primary Compliance Method USEPA 180.1; Hach Method 8195; ASTM D 6698; Standard Methods 2130B
®
(stabilized formazin) – primary or wet calibration of the instrument
®
standards.
®
(stabilized formazin) – recommended for verification in the appropriate application
Specifications
Limit of Detection 0.0032 NTU (according to criteria specified by ISO 15839)
Table 2 Controller Specifications
Component Description
Controller Operating Temperature
Controller Storage Temperature –20 to 70 °C (–4 to 158 °F); 95% relative humidity, non-condensing
Enclosure NEMA 4X/IP66 metal enclosure with a corrosion-resistant finish
Power Requirements
Pollution Degree/Installation Category II; II
Outputs
Relays
Microprocessor-controlled measuring unit with measured value display, temperature display (for some parameters), and menu-driven system
–20 to 60 °C (–4 to 140 °F); 95% relative humidity, non-condensing with sensor load <7 W; –20 to 40 °C (–4 to 104 °F) with sensor load <25 W
100–230 V ac ±10%, 50/60 Hz; Power: 11 W with 7 W sensor load, 35 W with 25 W sensor load
Two (Analog outputs, each selectable for 0–20mA or 4–20 mA), maximum impedance 500 ohm. Output span programmable over any portion of the 0–100
1
NTU range. Optional digital network connection
. Infrared Data Acquisition (IrDA).
Three SPDT, user-configurable contacts rated 100–230 V ac, 5 Amp resistive maximum.
Controller Dimensions ½ DIN—144 x 144 x 150 mm (5.7 x 5.7 x 5.9 inches)
Controller Weight 1.6 kg (3.5 lb)
1. See Replacement Parts and Accessories on page 55.
5
Page 6
Section 2 General Information
2.1 Safety Information
Please read this entire manual before unpacking, setting up, or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
To ensure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this manual.
Use of Hazard Information DANGER: Indicates a potentially or imminently hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION: Indicates a potentially hazardous situation that may result in minor or moderate injury.
Note: Information that requires special emphasis.
Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or electrocution exists.
This symbol, if noted on the product, indicates the need for protective eye wear.
This symbol, when noted on the product, identifies the location of the connection for Protective Earth (ground).
2.2 General Product Information
The controller enclosure is NEMA4X/IP66-rated and has a corrosion-resistant finish designed to withstand corrosive environmental constituents such as salt spray and hydrogen sulfide. The controller display shows the current turbidity reading if connected to a single sensor, or two readings when two sensors are connected.
The 1720E Turbidimeter is a continuous-reading nephelometric turbidimeter designed for low-range turbidity monitoring. This process turbidimeter is capable of measuring turbidity from 0.001 to 100.0 NTU. Calibration is based on formazin, the primary turbidity reference standard adopted by the APHA Standard Methods for the Examination of Water and Wastewater and the U.S. Environmental Protection Agency (EPA) and on StablCal standard.
®
which is also recognized as a primary
6
Page 7
2.3 Theory of Operation
Figure 1 90 Degree Detector
General Information
The 1720E Turbidimeter measures turbidity by directing a strong beam of collimated light from the sensor head assembly down into the sample in the turbidimeter body. Light scattered at 90° relative to the center line of incident light by suspended particles in the sample is detected by the submerged photocell (see Figure 1).
The amount of light scattered is proportional to the turbidity of the sample. If the turbidity of the sample is negligible, little light will be scattered and detected by the photocell and the turbidity reading will be low. High turbidity, on the other hand, will cause a high level of light scattering and result in a high reading.
Sample enters the turbidimeter body and flows through the baffle network of the bubble trap. The flow allows bubbles to either cling to surfaces of the baffle system or rise to the surface and vent to atmosphere. After traveling through the bubble trap, sample enters the center column of the turbidimeter body, rises into the measuring chamber and spills over the weir into the drain port. A reading is taken once per second.
7
Page 8
Section 3 Installation
DANGER Only qualified personnel should conduct the installation tasks described in this section of the manual. The 1720E/sc100 product configuration is not intended for installation in hazardous locations.
Figure 2 Controller Mounting Components
1
sc100
1. Controller 6. Lock washer, ¼-inch I.D. (4), Cat. No. 8H1336
2. Mounting foot for panel mounting (2),
Cat. No. 1000B4F3222
3. Bracket for panel and pipe mounting, Cat. No. 1000C4F3217-101
4. Gasket for panel mounting, rubber, Cat. No. 1000A4F3249-101
5. Hex nut, M6 (4), Cat. No. 5867300 10. Pan head screws (4), M6 x 1.0 x 150 mm, Cat. No. 5867600
3
2
4
7. Flat washer, ¼-inch I.D. (4), Cat. No. 8H1346
8. Pan head screws (4), M6 x 1.0 x 20 mm, Cat. No. 58674-00
9. Pan head screws (4), M6 x 1.0 x 100 mm, Cat. No. 5867500
5
6
7
8
9
10
Table 3 Customer-supplied Items
Item
14-AWG wire for electrical power connections in conduit or if allowed by local electrical codes, 115 or 230 V ac power cord plus a NEMA 4X-rated strain relief
High-quality, shielded instrumentation cable for connecting the analog outputs plus a NEMA 4X-rated strain relief Mounting hardware for the sensor Sun shield for mounting configurations where the sun strikes the front of the display (available from the manufacturer, order
separately). See Figure 8 on page 11. Common hand tools
3.1 Mechanical Installation
Install in an environment that is protected from corrosive fluids. The sensor is adversely affected by ClO corrosive liquids or gasses.
. Install the sensor in an area well ventilated from any
2
8
Page 9
3.1.1 Controller Dimension Illustrations
Figure 3 Controller Dimensions
144.0 mm (5.67 inches) 150.0 mm (5.91 inches)
sc100
Installation
144.0 mm (5.67 inches)
Figure 4 Controller Mounting Dimensions
144.02 mm (5.67 inches)
72.01 mm
(2.84 inches)
72.01 mm
(2.84 inches)
144.02 mm
(5.67 inches)
40.14 mm
(1.58 inches)
80.27 mm
(3.16 inches)
40.14 mm (1.58 inches)
M6 x 1.0
80.27 mm
(3.16 inches)
9
Page 10
Installation
Figure 5 Panel Mount Cut-out Dimensions
75 mm 2.955 inches
(ref only)
144 mm 5.67 inches
(ref only)
144 mm
5.67 inches (ref only)
72 mm
2.835 inches (ref only)
66.67 mm 2.625 inches
Figure 6 Conduit Hole Dimensions
75.07 mm 2.955 inches
28.57 mm 1.125 inches
135 mm
5.31 inches
67.4 mm
2.65 inches
133 mm 5.25 inches
144 mm 5.67 inches
68.96 mm 2.715 inches
28.57 mm 1.125 inches
10
150 mm
5.91 inches
99.31 mm
3.91 inches
50.8 mm
2.00 inches
6.35 mm 0.25 inch
15.24 mm 0.60 inch
25.4 mm 1.00 inches25.4 mm 1.00 inches
5.00 inches
(ref only)
127 mm
Page 11
3.1.2 Using the Optional Sun Shield
The optional sun shield was designed to increase the readability of the display by screening it from direct sunlight. See Replacement Parts and Accessories on
page 55 for ordering information.
Figure 7 Sun Shield Kit Components
Installation
12
3
4
6
5
7
8
1. Sun shield 5. Pipe mounting brackets (2), includes items 6 and 7, Cat. No. 9H1079
5
2. Pan head screws, M6 x 1.0 x 12 mm (6),
Cat. No. 200-1025
3. Lock washers, ¼-inch I.D. (2), Cat. No. 8H1336 7. Square nut,
4. Hex nuts, M6 x 1.0 (2), Cat. No. 5867300 8. Uni-strut, 27 cm (10.5 inch) length, Cat. No. 276F1227
6. Hex/slotted head screw,
number 5)
5
/16-inch (supplied with item number 5)
/16-inch x 1.0-inch (supplied with item
Figure 8 Mounting the Controller in the Sun Shield
1
2
3
5
2
6
1. Sun shield 4. Hole pattern for mounting controller 7. Slide mounting brackets into the
2. Uni-strut (rotate 90° as required) 5. Pipe (vertical or horizontal as required)
3. Pan head screw, lock washer (2 each) 6. Hex/slotted head screw and square nut
4
7
uni-strut as shown. Place the mounting brackets around the pipe and fasten the hardware.
11
Page 12
Installation
3.1.3 Mounting the Controller
Attach the controller to a rail or wall or mount it in a panel. Supplied mounting hardware is shown in Figure 9, Figure 10, and Figure 11.
Figure 9 Vertical or Horizontal Pipe Mounting the Controller
1
1. Controller 4. Flat washer, ¼-inch I.D. (4), Cat. No. 8H1346
2. Pipe (vertical or horizontal) 5. Hex nut, M6 (4), Cat. No. 5867300
3. Bracket, pipe mounting, Cat. No. 1000C4F3217-101 6. Pan head screw, M6 x 1.0 x 100 mm (4), Cat. No. 5867400
2
3
4
5
6
Figure 10 Wall Mounting the Controller
1
2
3
4
5
1. Controller 4. Pan head screw, M6 x 1.0 x 20 mm (4), Cat. No. 5867400
2. Bracket, Cat. No. 1000C4F3217-101 5. Customer-supplied hardware for wall mounting
3. Lock washer, ¼-inch I.D., Cat. No. 8H1336
12
Page 13
Figure 11 Panel Mounting the Controller
Installation
1
8
2
9
8
3
4
5
6
7
11
1. Controller 7. Lock washer, ¼-inch I.D., (4) Cat. No. 8H1336
2. Gasket, rubber, panel mount, Cat. No. 1000A4F3249-101 8. Hex nut (4), Cat. No. 5867300
3. Panel (maximum thickness is 9.5 mm (3/8 inch)) 9. Flat washer (4), Cat. No. 8H1346
4. Mounting Foot (2), Cat. No. 1000B4F3222 10. Pan head screw, M6 x 1.0 x 150 mm (4), Cat. No. 5867600
5. Mounting bracket, controller, Cat. No. 1000C4F3217-101 11. It may be necessary to remove the sensor connectors. See
6. Pan head screw (4), Cat. No. 5867400
procedure below.
10
To remove the sensor connectors before inserting the controller enclosure into the panel cut-out:
1. Disconnect power to the controller.
1. Disconnect the wires at terminal block J5, see Figure 21 on page 21.
2. Loosen and remove the nut securing the sensor connector inside the
enclosure. Remove the sensor connector and wires. Repeat step 1 and 2 for the other sensor connector.
3. After the controller is in place in the panel, reinstall the sensor connectors and reconnect the wiring to terminal J5 as shown in Figure 21 on page 21.
13
Page 14
Installation
+
DATA
+
OUT 2
+
V
+
OUT 1
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
1
1
+
DATA+DATA
+
OUT 2+OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+V+
V
+
OUT 1+OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANALOG OUTPUTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
J1
J2
J4
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
S1
3.2 Electrical Installation
DANGER The instrument must be installed by qualified technical personnel for adherence to all applicable electrical codes. The 1720E/sc100
High-voltage wiring for the controller is conducted behind the high voltage barrier in the controller enclosure. The barrier must remain in place unless a qualified installation technician is installing wiring for power, alarms, or relays. See Figure 12 for barrier removal information.
product configuration is not intended for installation in hazardous locations.
3.2.1 Installation in Conduit
In hard-wired electrical applications, the power and safety ground service drops for the instrument must be 18 to 12 AWG. See Figure 13 on page 15 for strain relief and conduit opening sealing plug information. See section 3.2.3 on page 15 for wiring information.
3.2.2 Installation Using a Power Cord
DANGER Use of power cords is not permitted in hazardous locations.
Where permitted by local electrical codes, a sealing-type strain relief to maintain the NEMA 4X/IP66 environmental rating and a power cord less than 3 meters (10 feet) in length with three 18-gauge conductors (including a safety ground wire) can be used, see Replacement Parts and Accessories on page 55. See Figure 13 on page 15 for strain relief and conduit opening sealing plug assembly. See section 3.2.3 on page 15 for wiring information.
Figure 12 Removing Voltage Barrier
J1
S1
J4
NETWORK
INTERFACE
1. High voltage barrier 2. Unsnap the barrier latch then pull out to remove the barrier.
J2
U5
U9
CARD
J3
1
+
DATA
– DATA
2
SERVICE REQUEST
3 4
+
PROBES
J5
PCB
CONNECTOR
J6
ANALOG OUTPUTS
V
5
GND
6
1
+
OUT 2
– OUT 2
2
SHIELD/CHASSIS GND
3 4
+
OUT 1
5
– OUT 1
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
1
2
14
Page 15
Figure 13 Using the Optional Strain Relief and Conduit Plug
Installation
2
1. Power cord strain relief 2. Conduit strain relief 3. Conduit opening sealing plug
3
Figure 14 Proper Wire Preparation and Insertion
1
2
1. Strip ¼-inch of insulation. 2. Seat insulation against connector with no bare wire exposed.
3.2.3 Wiring for Power at the Controller
Wire the instrument for line power by hard-wiring in conduit or by wiring to a power cord if local code allows. Regardless of the type of wire used, the connections are made at the same terminal. A local disconnect designed to meet local electrical code is required and must be identified for all types of installation. See Figure 16 and Figure 17 on page 17 for suggested local disconnect configurations.
1. Obtain appropriate fittings with NEMA 4X/IP66 environmental rating.
2. Loosen the screws using a phillips-head screwdriver and open the hinged
controller cover.
3. Remove the high-voltage barrier (see Figure 12 on page 14).
4. Insert the wires through the strain relief fitting or conduit hub located in the
right-rear access hole in the bottom of the enclosure. Tighten the strain relief if used, to secure the cord.
5. Properly prepare each wire (Figure 14) and insert each wire into the terminal according to Table 4. Tug gently after each insertion to ensure the connection is secure.
6. Seal any unused openings in the controller box with conduit opening sealing plugs, see Replacement Parts and Accessories on page 55.
7. Reinstall the high-voltage barrier and latch to secure.
15
Page 16
Installation
Table 4 Power Wiring Information
Terminal Number Terminal Description Wire Color Code for North America Wire Color Code for Europe
1 Hot (L1) Black Brown 2 Neutral (N) White Blue 3 Protective Earth (PE) Green Green w/yellow tracer
Figure 15 Wiring Connections
312
45
J1
J3
14
S1
J2
CARD
U5
U9
11 10
F1
J5
NONO
J6
F2
NCNCNC
NO
COMCOMCOM RELAYCRELAYBRELAYA
89
13
12
J4
NETWORK
INTERFACE
1. J1—Network connector 8. Sensor connector
2. J2—Header for optional network interface card 9. Sensor connector
3. J5—Relay A connector 10. J6—Analog output (4–20 mA) connector
4. J6—Relay B connector 11. J5—Sensor connector for hard-wiring
5. J7—Relay C connector 12. Position for network interface card
6. Fuses (F1, F2) 13. Service port
7. J8—ac Power connections 14. Sensor terminator selector/service port configuration
6
7
16
Page 17
Figure 16 Local Disconnect for Power Cord
RELAYC
J1
S1
J2
J3
Installation
1
CARD
U5
U9
J5
RELAYBRELAYA
J6
F1
NCNCNC
NONO
NO
COMCOMCOM
RELAYC
J4
NETWORK
INTERFACE
1. Power terminal 2. Power cord strain relief
Figure 17 Local Disconnect for Hard-wired Line Power
J1
S1
J2
J3
F2
2
1
J4
NETWORK
INTERFACE
U5
U9
CARD
1. Power terminal 2. Conduit strain relief
F1
J5
F2
2
NCNCNC
NONO
NO
COMCOMCOM
RELAYCRELAYBRELAYA
J6
17
Page 18
Installation
NCNCNC
COMCOMCOM
NO
F1
F2
NONO
RELAYCRELAYBRELAYA
J1
J2
J4
S1
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
J5J5
3.3 Alarms and Relays
The controller is equipped with three unpowered relays rated 100–230 V ac, 50/60 Hz, 5 amp resistive maximum.
3.3.1 Connecting the Relays
The relay connector accepts 18–12 AWG wire (as determined by load application). Wire gauge less than 18 AWG is not recommended.
Danger: Relay loads must be resistive. User must externally limit current to the relays to 5 Amps by use of a fuse or breaker.
Danger: ac power terminals are designed for single wires. Do not use more than one wire in each terminal.
The controller contains three relays designed for use with high voltage (greater than 30V-RMS and 42.2V-PEAK or 60 V dc). Refer to Figure 18 for connection information. The wiring is not designed for low voltage connections. Relay must not be powered from the same wiring used to power the controller. See section 4.6 on page 34 for relay setup details.
The Normally Open (NO) and Common (COM) relay contacts will be connected when an alarm or other condition is active. The Normally Closed (NC) and Common relay contacts will be connected when an alarm or other condition is inactive or when power is removed from the controller.
Figure 18 Alarm and Relay Connections
J1
S1
J4
NETWORK
INTERFACE
J2
U5
U9
CARD
J3
J5J5
J5
J5 J6 J7
3
2
1
RELAY A RELAY B RELAY C
F1
F2
NCNCNC
NONO
NO
COMCOMCOM
RELAYCRELAYBRELAYA
3.3.2 Connecting the Analog Outputs
18
J6
Disconnect Power
Two isolated analog outputs (1 and 2) are provided, see Figure 19. Each output can be set to 0–20 or 4–20 mA, and can be assigned to represent the measured parameter or secondary measurement such as temperature. Make connections with twisted-pair shielded wire and connect the shield at the controlled component end or at the control loop end. Do not connect the shield at both ends of the cable. Use of non-shielded cable may result in radio frequency emission or susceptibility levels higher than allowed. Maximum loop resistance is 500 ohm. Refer to
section 4.5 on page 32 for output software setup.
Page 19
Make wiring connections at the analyzer end as shown in Figure 19.
RELAYC
RELAYA
+
DATA
+
OUT 2
+
V
+
OUT 1
ANALOG OUTPUTS
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
NCNCNC
COMCOMCOM
NO
F1
F2
NONO
RELAYCRELAYCRELAYBRELAYARELAY A
J1
J2
J4
S1
NETWORK INTERFACE
CARD
J3
J5
J6
U5
U9
1
1
+
DATA+DATA
+
OUT 2+OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+V+
V
+
OUT 1+OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANALOG OUTPUTSANALOG OUTPUTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
J5J5
Table 5 Output Connections at Terminal Block J6
Recorder Wires Circuit Board Position
Output 2 + 1 Output 2 – 2
Shield 3 Output 1 + 4 Output 1 – 5
Figure 19 Analog Output Connections
J1
S1
J2
J3
Installation
CARD
U5
J5J5
U9
J5
J4
NETWORK INTERFACE
J6
1 2 3 4 5
J6
3.4 Connecting/Wiring the Sensor Cable
The sensor cable is supplied with a keyed quick-connect fitting for easy attachment to the controller, see Figure 20. Retain the connector cap to seal the connector opening in case the sensor must be removed.
The 1720E sensor cable may be extended by a maximum of 7.62 m (25 ft), see Replacement Parts and Accessories on page 55.
Modify the controller for sensor hard-wiring as follows:
1. Remove all power to the controller.
2. Open the controller cover.
PCB
CONNECTOR
PROBES
ANALOG OUTPUTS
RELAYBRELAYA
1 2 3 4 5 6
1 2 3 4 5
NCNCNC
NONO
COMCOMCOM
RELAYC
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
+
DATA
– DATA
SERVICE REQUEST
+
V
GND
F1
+
OUT 2
– OUT 2
SHIELD/CHASSIS GND
+
OUT 1
– OUT 1
NO
F2
19
Page 20
Installation
RELAYA
+
DATA
+
OUT 2
+
V
+
OUT 1
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
3. Disconnect and remove the existing wires between the quick connect and terminal block J5, see Figure 21 on page 21.
4. Remove the quick connect fitting and wires and install the threaded plug on the opening to maintain the environmental rating.
Table 6 Wiring the Sensor at Terminal Block J5
Terminal Number Terminal Designation Wire Color
1 Data (+) Blue 2 Data (–) White 3 Service Request No Connection 4 +12 V dc Brown 5 Circuit Common Black 6 Shield Shield (grey wire in existing quick disconnect fitting)
Figure 20 Attaching the Sensor using the Quick-connect Fitting
S1
J4
From Probe
NETWORK
INTERFACE
CARD
J1
J2
U5
U9
J3
1
+
DATA
– DATA
2
SERVICE REQUEST
3 4
+
5 6
1 2 3 4 5
NCNCNC
NONO
COMCOMCOM
RELAYCRELAYBRELAYA
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
V
GND
F1
+
OUT 2 – OUT 2 SHIELD/CHASSIS GND
+
OUT 1
– OUT 1
NO
F2
PROBES
J5J5
J5
PCB
CONNECTOR
J6
5. Cut the connector from the sensor cable.
6. Reinstall the plug on the sensor access opening to maintain the
environmental rating.
7. Strip the insulation on the cable back 1-inch. Strip ¼-inch of each individual
ANALOG OUTPUTS
wire end.
20
8. Wire as shown in Table 6 .
Page 21
9. Pass the cable through conduit and a conduit hub or a strain relief fitting and
+
DATA
+
OUT 2
+
V
+
OUT 1
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
NCNCNC
COMCOMCOM
NO
F1
F2
NONO
RELAY3RELAY 2RELAY1
1
1
+
DATA+DATA
+
OUT 2+OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+V+
V
+
OUT 1+OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANALOG OUTPUTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
J1
J2
J4
NETWORK
INTERFACE
CARD
J3
J6
U5
U9
S1
J5
10. Close and secure the cover.
Figure 21 Hard-wiring the Sensor
J1
S1
J2
U5
J4
U9
NETWORK INTERFACE
CARD
J3
J5J5
J5
J6
Installation
an available access hole in the controller enclosure. Tighten the fitting.
1
+
DATA
– DATA
2
SERVICE REQUEST
3 4
+
5 6
1 2 3 4 5
NCNCNC
NONO
COMCOMCOM RELAY3RELAY 2RELAY1
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
V
GND
F1
+
OUT 2 – OUT 2 SHIELD/CHASSIS GND
+
OUT 1 – OUT 1
NO
F2
PCB
CONNECTOR
PROBES
ANALOG OUTPUTS
3.5 Connecting the Optional Digital Output
At this time, the manufacturer supports ModBUS RS485 and ModBUS RS232 communication protocols. The optional digital output card is installed in the location indicated in Figure 22 on page 22. Terminal block J1 provides user connection to the optional network card. The terminal connection is based on the selected network card. Refer to the instructions supplied with the network card for more details.
Table 7 Network Connections at Terminal Block J1
Terminal Number ModBUS RS485 ModBUS RS232
1In + — 2In — 3Out + — 4Out — 5 Common Common 6 No connection No connection 7 Shield Shield
21
Page 22
Installation
+
DATA
+
OUT 2
+
V
+
OUT 1
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
NCNCNC
COMCOMCOM
NO
F1
F2
NONO
RELAYCRELAYBRELAYA
J1
J2
J4
S1
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
1
1
+
DATA+DATA
+
OUT 2+OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+V+
V
+
OUT 1+OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANALOG OUTPUTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
Figure 22 Network Card Position in the Controller
4
J1
S1
J2
U5
J4
U9
NETWORK
INTERFACE
CARD
J3
J5
J6
PCB
CONNECTOR
PROBES
ANALOG OUTPUTS
1 2 3 4 5 6
1 2 3 4 5
NCNCNC
NONO
COMCOMCOM
RELAYCRELAYBRELAYA
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
+
DATA – DATA SERVICE REQUEST
+
V
GND
F1
+
OUT 2 – OUT 2 SHIELD/CHASSIS GND
+
OUT 1
– OUT 1
NO
F2
1
2
3
1. J2—Network card header 2. Mounting hole (3) 3. Network card placement 4. J1 Terminal
22
Page 23
Installation
23
Figure 23 1720E Dimensions
POWER
STATUS
127.8 mm
(5.0 inches)
255.3 mm (11.05 inches)
217.2 mm (8.55 inches)
143.2 mm (5.64 inches)
40.6 mm
(1.60 inches)
72.04 mm (2.84 inches)
107.95 mm (4.25 inches)
343.67 mm (13.53 inches)
19.72 mm (0.78 inches)
238.35 mm (9.38 inches)
312.48 mm (12.30 inches)
273.05 mm (10.75 inches)
157.00 mm
(6.18 inches)
172.4 mm (6.79 inches)
2 Slots for
0.25 fastener² Sample Inlet
¼-NPT to Compression fitting¼-
Cable 2 meters (6.6 feet)
Sample Drain
NPT to ½-ID Tubing hose Barb½-
Service Drain
1.00 I.D.
164.2 mm (6.46 inches)
193.7 mm (7.625 inches)
195.6 mm (7.70 inches)
Allow at least 254 mm (10.00 inches) above the instrument for removal of the head assembly.
Page 24
Installation
3.6 Turbidimeter Installation Information
The turbidimeter body is designed for wall-mounting (although it may be mounted on the optional floor stand). The turbidimeter sensor must be mounted within six feet of the controller unless an extension cable is used. Maximum cable length is 9.6 m (31 feet).
3.6.1 Mounting the Turbidimeter Body
Locate the turbidimeter as close to the sampling point as possible. A shorter distance for the sample to travel results in a faster response time.
Clean the turbidimeter body and bubble trap before installation using the instructions supplied in section 6.4.3, Cleaning the Turbidimeter Body and Bubble
Trap on page 48. Slotted mounting brackets are integral parts of the turbidimeter
body. Install customer-supplied hardware appropriate for the installation environment using the criteria detailed below:
Install in a location that is isolated from vibration.
Allow at least 22 cm (approximately 10 inches) clearance for removal of the
head assembly and bubble trap cover from the top of the turbidimeter body.
Leave enough room below the turbidimeter body to remove the bottom plug and to place a container under the drain when calibrating or cleaning.
Note: Make sure the top of the turbidimeter body is level.
Install two ¼-20 bolts 10-3/4 inches apart (on center). Leave at least ¼-inch of the bolt head exposed.
Make sure the bolts are installed level.
Slide the slotted mounting brackets of the turbidimeter body onto the bolts.
3.6.2 Installing the Head Assembly
After the turbidimeter body has been mounted, install the bubble trap cover, then place the head assembly on the turbidimeter body with the label facing the front. Move the head assembly back and forth slightly to ensure it is properly seated on the body of the instrument. Failure to properly seat the head will result in light leakage and erroneous readings.
The rear portion of the head assembly has a molded “lip” which may be used to hang the head assembly on the turbidimeter body edge for routine maintenance.
3.7 Installing a Sample Line
DANGER This turbidimeter is not designed for use in hazardous locations or with samples that are flammable or explosive in nature. If any sample solution other than water is used in this product, test the sample/product compatibility to ensure user safety and proper product performance.
One-fourth inch O.D. rigid or semi-rigid tubing is recommended for sample lines. Run them as directly as possible between the turbidimeter body and the sampling point to minimize sample flow lag time.
Install sample line taps into larger process pipes to minimize interference from air bubbles or pipeline bottom sediment. A tap projecting into the center of the pipe is ideal. Figure 24 shows examples of sample tap installations.
Note: When setting the flow rate, take care to avoid sweeping air “micro-bubbles” through
the internal bubble trap. Observe the sample flow inside the turbidimeter body. If small air bubbles can be seen flowing up through the center, reduce the flow rate.
24
Page 25
Figure 24 Sampling Techniques
Installation
Air (Typical)
Sediment (Typical)
Poor
3.8 Sample Connections
Sample inlet and drain connections are made on the turbidimeter body. The sample inlet fitting installed in the body is a ¼-inch NPT fitting. One additional fitting supplied with the instrument is a ½-inch NPT-to-hose fitting for use with ½-inch ID flexible plastic tubing on the drain.
Note: For samples with high solids content (high turbidity), operate at the highest flow rate possible. For samples with low solids content (low expected turbidity), operate at a low flow rate (200–300 mL/min).
Figure 25 Sample Connections
The required flow rate is 200 to 750 mL/minute (4.0 to 11.9 gal/hour). Flow rate into the turbidimeter may be controlled with a flow restriction device on the inlet line. Flow rates below 200 mL/min will reduce response time and cause inaccurate readings. Flow rates above 750 mL/min will cause the turbidimeter to overflow, indicating the flow rate is too high.
Poor
Good Best
x ¼-inch compression
STATUS
POWER
1. Sample Inlet, ¼-28 NPT x ¼-inch Compression fitting 2. Drain, ½-inch NPT fitting 3. Service Drain
1
2
3
25
Page 26
Section 4 Operation
4.1 Using the Keypad
The front of the controller is shown in Figure 26. The keypad consists of the eight keys described in Table 8 .
Figure 26 Front of the Controller
1
sc100
5
2
3
4
1. Instrument Display 4. Right, Left, Up, and Down keys 7. Enter key
2. Back key 5. IrDA (Infrared Data Acquisition) window
3. Menu key 6. Home key
Table 8 Controller Key Functions/Features
Number Key Function
2 Move back one level in the menu structure.
3
Move to the main menu from other menus. This key is not active in menus where a selection or other input must be made.
6
7
4 Navigate through the menus, change settings, and increment and decrement digits.
5
6 Accept an input value, updates, or accepts displayed menu options.
Move to the Main Measurement screen from any other screen. This key is not active in menus where a selection or other input must be made.
26
Page 27
4.2 Controller Display Features
When the controller is in measurement mode, measurements for each connected sensor are displayed.
The display will flash on startup, when the hold outputs function has been activated, and when the filter function (signal average) is changed to a different value.
An active system warning will cause the warning icon (a triangle with an exclamation point inside) to be displayed on the right side of the display.
Figure 27 Display
Operation
1
2
SENSOR NAME
0.023
NTU
3
1. Status bar. Indicates the sensor name and status of relays. The relay letter is displayed when the relay is energized.
2. Main measurement 5. Warning icon area
3. Secondary measurement/output information 6. Measurement units
4.2.1 Important Key Presses
OUTPUT1: 11.23 mA
4. Energized relay indicator
Press the HOME key then the RIGHT or LEFT key to display two readings
when two sensors are connected. Continue to press the RIGHT or LEFT key to toggle through the available display options as shown below.
4
5
6
RTC:MM/DD/YY
24:00:00
OUTPUT1: 11.23 mA
SENSOR NAME:
02
0.023
ppm
OUTPUT1: 11.23 mA
MAIN MENU
SENSOR DIAG SENSOR SETUP
SYSTEM SETUP
TEST/MAINT
NTU
SENSOR NAME:
0.031
OUTPUT1: 11.23 mA
MAIN MEASURE
SENSOR NAME:
NTU
SENSOR NAME:
0.023 NTU
0.031 NTU
SENSOR NAME
Turbidity:
0.023 NTU
TEMP:
22.9°C
SENSOR NAME
Turbidity:
0.031 NTU
TEMP:
22.9°C
Press the UP and DOWN keys to toggle the status bar at the bottom of the
measurement display to display the output information.
In Menu mode, an arrow may appear on the right side of the display to indicate that more items are available. Press the UP or DOWN key (corresponding to the arrow direction) to display additional menus.
SYSTEM SETUP
OUTPUT SETUP RELAY SETUP NETWORK SETUP
DISPLAY SETUP
SYSTEM SETUP
DISPLAY SETUP SECURITY SETUP LOG SETUP
CALCULATION
SYSTEM SETUP
SECURITY SETUP LOG SETUP CALCULATION
ERROR HOLD MODE
27
Page 28
Operation
4.3 Instrument Setup
4.3.1 Software Text Abbreviation Conventions
Abbreviation Meaning Abbreviation Meaning
Adj Adjust P/F Pass/Fail Cal Calibration Pass Password
Cont. Continue Preped Prepared
Cyl Cylinder SN Serial Number Dflt Default Std Standard
Diag Diagnostic Temp Temperature
Int Internal Ver Verification
Meas. Measurement Xfer Transfer
4.3.2 Adjusting Display Contrast
Step Select Menu Level/Instructions Confirm
1
2
3
4
ADJ CONTRAST
5
6
MAIN MENU or Main Measurement Screen
MAIN MENU
SYSTEM SETUP
DISPLAY SETUP
(+0–50)
28
Page 29
Operation
4.3.3 Specifying the Displayed Language
Step Select Menu Level/Instructions Confirm
1
2
3
4
5
6
Main Menu or Main Measurement Screen
MAIN MENU
SYSTEM SETUP
DISPLAY SETUP
LANGUAGE
select language
4.3.4 Setting the Time and Date
4.3.4.1 Setting the Time
Note: The time is available only in 24-hour (military) format.
Step Select Menu Level/Instructions Confirm
1
2
3
4
5
MAIN MENU
SYSTEM SETUP
DISPLAY SETUP
SET DATE/TIME
highlight TIME
select character to edit
6
choose appropriate number
7
Main Menu or Main Measurement Screen
29
Page 30
Operation
4.3.4.2 Setting the Date Format and Date
Step Select Menu Level/Instructions Confirm
1
2
3
4
5
6
highlight DATE FORMAT
choose appropriate date format
7
MAIN MENU
SYSTEM SETUP
DISPLAY SETUP
SET DATE/TIME
Highlight DATE
select character to edit
8
choose appropriate number
9
Main Menu or Main Measurement Screen
4.4 Changing the Sensor Name
Step Select Menu Level/Instructions Confirm
1
2
3
highlight sensor of interest if more than one sensor is attached
4
5
6
choose appropriate alpha/numeric digit
7
Main Menu or Main Measurement Screen
MAIN MENU
SENSOR SETUP
CONFIGURE
EDIT NAME
select character to edit
30
Page 31
Operation
4.4.1 Setting Up System Security
The sc100 has a passcode feature to restrict unauthorized access to Network Setup, Security Setup, Log Setup, and Test/Maint menus. In addition, the passcode also regulates function selection for relay options. The passcode is factory set to sc100 (the five digits must be followed by a space to remove the trailing asterisk). The passcode may be changed, see section 4.4.1.1.
The following two options are available:
Disabled: All configuration settings and calibrations can be changed. This is the default setting.
Enabled: All configuration settings can be displayed but not changed. Network Setup, Security Setup, Log Setup, and Test/Maint menus cannot be accessed without the passcode.
Step Select Menu Level/Instructions Confirm
1
2
3
4
SET PASSCODE
5
6
Main Menu or Main Measurement Screen
MAIN MENU
SYSTEM SETUP
SECURITY SETUP
highlight ENABLED
4.4.1.1 Editing the Passcode
If the passcode is enabled, it may be edited. The passcode can consist of up to six digits (alpha and/or numeric and available characters). If a passcode is forgotten, obtain the Master passcode from the Technical Consulting Services Department, see Technical and Customer Service (U.S.A. only) on page 57.
Step Select Menu Level/Instructions Confirm
1
2
3
4
MAIN MENU
SYSTEM SETUP
highlight SECURITY SETUP
ENTER (current or default) PASSCODE
EDIT PASSCODE
31
Page 32
Operation
Step Select Menu Level/Instructions Confirm
5
6
Change the existing passcode
Main Menu or Main Measurement Screen
4.5 Output Options
The controller provides two isolated analog outputs (Output 1 and Output 2). Customize the outputs using the table in section 4.5.1 on page 32.
4.5.1 Output Setup Menu (from System Setup)
1. Select OUTPUT 1 or 2
SELECT SOURCE
Press ENTER to access a list of all connected sensors. Choose the sensor to associate with the output.
SET PARAMETER
Highlight the appropriate displayed parameter and press ENTER
SET FUNCTION
Select LINEAR CONTROL for current output to track the measurement value. Select PID CONTROL for the sc100 to operate as a PID controller.
SET TRANSFER
Each analog output is normally active, responding to the measured value of its assigned parameter. However, during calibration, each output can be transferred to this preset transfer value. Default: 4mA; Range: 0–20
.
SET FILTER
Allows the user to average the analog outputs over time Default: 0; Range: 0–120 seconds
SCALE 0 mA/4 mA
Select 0 mA or 4 mA for minimum current (outputs will be set to 0–20 mA or 4–20 mA).
ACTIVATION
FUNCTION set to LINEAR CONTROL If LINEAR CONTROL was selected in SET FUNCTION, set the low and the high values for the current output here.
Defaults: Low = 0; High = 100; Low Value Range: 0–100, High Value Range: 0–100. FUNCTION set to PID CONTROL If PID CONTROL was selected in SET FUNCTION, configure the PID Control as follows:
1. Set MODE: AUTO or MANUAL. Manual output default: 100%
2. Set PHASE: DIRECT or REVERSE controller operation.
3. SET SETPOINT: enter the set point the PID control will control the process to. Default: 100; Range: 0–100
4. PROP BAND: control the proportional band for the PID control. Default: 5.00; Range: 0–1000
5. INTEGRAL: control the integral action time period in minutes. Range: 0–999
6. DERIVATIVE: control the settings for the rate control. Range: 0–999
32
Page 33
Operation
4.5.2 Hold/Transfer Outputs
When cleaning or servicing the instrument, the analog outputs can be held at the last measured values. To hold the output until released:
Step Select Menu Level/Instructions Confirm
1
2
3
4
5
SET OUTMODE
Select HOLD OUTPUTS or XFER OUTPUTS
6
7
8
MAIN MENU
TEST/MAINT
Enter Passcode if enabled
HOLD OUTPUTS
SET CHANNELS
Select ALL or 1720E
ACTIVATION
9
10
LAUNCH
Main Menu or Main Measurement Screen
reading will
flash
During calibration, the analog outputs can remain active, be held, or be transferred to a preset mA value. When output hold or transfer is enabled during a calibration, the hold or transfer is automatically released when calibration is completed. See section 5.3, Sensor Calibration and Verification on page 40.
4.5.3 Release Outputs
Step Select Menu Level Confirm
1
2
3
4
MAIN MENU
TEST/MAINT
HOLD OUTPUTS
ACTIVATION
33
Page 34
Operation
Step Select Menu Level Confirm
5
6
RELEASE
Main Menu or Main Measurement Screen
4.6 Relay Options
Step Select Menu Level Confirm
1
2
3
4
Customize the options using the information in section 4.6.1
4.6.1 Relay Setup Menu (from System Setup)
1. Select Relay A, B, or C
MAIN MENU
SYSTEM SETUP
RELAY SETUP
SELECT SOURCE
Choose from the available options (none, installed sensors, real time clock (RTC)).
SET PARAMETER
Choose from the available options.
SET FUNCTION
Source set to sensor Alarm: Operates relays in response to the measured parameter. Contains separate High and Low Alarm points, deadbands, and
ON/OFF delay. Defaults: Low = 0.000 NTU, high = 100.00 NTU, low deadband = 5.000 NTU, high deadband = 5.000 NTU, on/off delays default to zero seconds; Range: 0–999 sec.
Feeder Control: Operates in response to the measured parameter. Can be set for phasing, set point, deadband, overfeed timer, and ON/OFF delay.
Event Control: Controls a cleaning system (or equivalent) on a timed basis. Warning: Activated when the analyzer detects a sensor warning.
Source set to RTC
Timer: Sets the timer for a cleaning system (or equivalent). Controls the output hold, interval, duration and off delay.
SET TRANSFER
Normally, each control or alarm relay is active, responding to the measured value of its assigned parameter. During calibration, however, the relay can be transferred to a preset on/off state to suit the application requirements. Select Energize or De-energize and press ENTER.
34
Page 35
Operation
4.6.1 Relay Setup Menu (from System Setup) (continued)
ACTIVATION
Function set to ALARM
Sets the value where the relay will turn on in response to decreasing measured value. For example: if the
LOW ALARM
HIGH ALARM
LOW DEADBAND
HIGH DEADBAND
OFF DELAY Sets a time to delay the relay from normally turning off. Off Delay Range: 0–999 seconds ON DELAY Sets a time to delay the relay from normally turning on. On Delay Range: 0–999 seconds
Function set to FEEDER CONTROL
PHASE
SET SETPOINT Sets the value where the relay will turn on. Default: 100 NTU; Range: 0–100
DEADBAND
OVERFEED TIMER
OFF DELAY Sets a time to delay the relay from normally turning off. Default: 0 seconds; Range: 0–999 seconds ON DELAY Sets a time to delay the relay from normally turning on. Default: 0 seconds; Range: 0–999 seconds
Function set to EVENT CONTROL
PHASE
SET SETPOINT Sets the value where the relay will turn on. Default: 100; Range: 0–100
DEADBAND
OnMax TIMER Sets the time to limit the time the relay can remain “on.” Default: 0 seconds; Range: 0–999 seconds OffMax TIMER Sets a time to delay the relay from normally turning off. Default: 0 seconds; Range: 0–999 seconds OnMin TIMER Sets the time to limit the time the relay can remain “on.” Default: 0 seconds; Range: 0–999 seconds OffMin TIMER Sets the time to limit the time the relay can remain “off.” Default: 0 seconds; Range: 0–999 seconds
Function set to TIMER CONTROL (RTC selected in SELECT SOURCE)
HOLD OUTPUTS Set OUTMODE to select output hold operation and select the channels that cause the outputs to be held. INTERVAL Set the off time for the relay. Default: 5 minutes; Range: 0–999 minutes DURATION Set the on time for the relay. Default: 30 seconds; Range: 0–999 seconds
OFF DELAY
Function set to WARNING CONTROL
WARNING LEVEL
low alarm is set for 1.0 and the measured value drops to 0.9, the relay will be activated. Range: 0.00–100 NTU
Sets the value where the relay will turn on in response to increasing measured value. For example: if the high alarm is set for 4.0 and the measured value increases to 4.2, the relay will be activated. Range: 0.00–100 NTU
Sets the range where the relay remains on after the measured value increases above the low alarm value. Default is 20% of the range. For example: if the low alarm is set for 1.0 and the low deadband is set for 0.5, then the relay remains on between 1.5 and 1.0. Range: 0.00–100 NTU
Sets the range where the relay remains on after the measured value decreases below the high alarm value. For example: if the high alarm is set for 4.0 and the high deadband is set for 0.5, then the relay remains on between 3.5 and 4.0. Range: 0.00–100 NTU
“High” phase assigns the relay setpoint to respond to an increasing measured value; conversely, a “Low” phase assigns the relay setpoint to respond to a decreasing measured value.
Sets the range where the relay remains on after the measured value decreases below the setpoint value (high phase relay) or increases above the setpoint value (low phase relay). Default: 5 NTU; Range: 0–100 NTU
Sets the time to limit how long the relay can remain “on.” Default: 5 minutes; Range: 0–999 minutes
“High” phase assigns the relay setpoint to respond to increasing measured value; conversely, a “Low” phase assigns the relay setpoint to respond to decreasing measured value.
Sets the range where the relay remains on after the measured value decreases below the setpoint value (high phase relay) or increases above the setpoint value (low phase relay). Default: 5 NTU; Range: 0–100 NTU,
Set the time for additional hold/output time after the relay has been turned off. Default: 1 second; Range: 0–999 seconds
Set the warning level that will trigger a relay. Range: 0–32 (warnings assigned to 1-9 for 1720E). For example: Set the warning level to 0 to allow all warnings to trigger the relay; set the warning level to 5 to allow warnings 6 and above to trigger the relay. Set the warning level to 9 or greater to not trigger the relay on any warning. See Table 10 on page 52 for a full list of warnings.
35
Page 36
Operation
4.7 Data and Event Logging Options
The sc100 provides two data logs (one for each sensor) and two event logs (one for each sensor). The data logs store the measurement data at selected intervals. The event log stores a variety of events that occur on the devices such as configuration changes, alarms, and warning conditions. The data logs are stored in a packed binary format and the event logs are stored in a CSV format. The logs can be downloaded through either the digital network port or the IrDA port using the file transfer program available from the manufacturer.
The default datalogging frequency is 15 minutes. If the datalogging frequency is set to 15 minute intervals, the instrument can continue to store data for approximately six months.
4.7.1 Data Logging Options
Sensor Data Log:
Step Select Menu Level/Instructions Confirm
1
2
3
4
5
6
Highlight sensor of interest if more that one sensor is attached.
4.8 Digital Network Options
The sc100 provides two digital communication methods with the controller (the digital network port and the IrDA port). Either of the digital ports can be used to access setup data, measurement data, or data/event logs. For the features available for each individual digital network port, refer to the instruction sheet supplied with the selected network card.
MAIN MENU
SENSOR SETUP
CONFIGURE
DATALOG INTERVAL
(select from 30 sec., 1 min., 5 min., 10 min., 15 min.)
Main Menu or Main Measurement Screen
4.9 Menu Structure
4.9.1 Sensor Diagnostics Menu
SELECT SENSOR
ERROR LIST Displays a list of errors that are present. See section 7.1 on page 52. WARNING LIST Displays a list of warnings that are present. See section 7.2 on page 52.
36
Page 37
4.9.2 Sensor Setup Menu
CALIBRATE
SELECT SENSOR (if more than one sensor is attached)
USER PREPD CAL Calibration using 4000 NTU stock solution diluted to 20.00 NTU formazin. STABLCAL CAL Calibration using 20 NTU StablCal Stabilized Formazin Standard VERIFICATION Perform a verification, set the pass/fail criteria, and view the verification history. 0 ELECTRONICS Zero electronics SET DFLT GAIN Return instrument to default calibration.
CAL HISTORY
CONFIGURE
BUBBLE REJECT Choose Yes or No to enable/disable bubble reject. Default: Yes
SIGNAL AVG
MEAS UNITS
EDIT NAME
SET RESOLUTION Set the number of significant digits to display. Default is three significant digits.
DATALOG INTRVL
DIAG/TEST
View the last 12 entered calibrations. Press the ENTER key to move to the next history entry. See section 5.5 on page 46 for more information.
Choose no averaging or specify the amount of time for signal averaging. Available options are: no averaging, 6 sec., 30 sec., 60 sec., or 90 sec. Default is 30 seconds.
Select the appropriate measurement units to display. Choose from mg/L, NTU, TE/F, and FTU. Default: NTU
Enter up to a 12-digit name in any combination of symbols and alpha or numeric characters. Press ENTER when the entry is complete. The name will be displayed on the status line above the measurement value on the main display.
Choose the amount of time between saving data points to the data log. Default: 15 min.; Options: 30 sec., 1 minute, 5 minutes, 10 minutes, or 15 minutes.
Operation
SOFTWARE VERS. Displays the software version number. DRIVER VERS Displays the software driver version number. SERIAL NUMBER Displays the serial number of the sensor. INT TEMP Displays the internal temperature of the sensor electronics in °C. DEFAULT SETUP Restores the sensor’s factory default settings and invalidates the current calibration. POWER CHECK Displays the electrical statistics for the sensor. CAL VALUE Displays the gain and zero electronics values for the current calibration.
4.9.3 System Setup Menu
OUTPUT SETUP (see section 4.5 on page 32 for expanded menu information)
SELECT OUTPUT 1 or 2
SELECT SOURCE Press ENTER to access a list of all connected sensors and select the sensor that will drive the output. SET PARAMETER Press ENTER to select from the displayed parameters.
SET FUNCTION
SET TRANSFER
SET FILTER Average measurements over time (0–120 seconds). Default: 0 seconds. SCALE 0 mA/4 mA Select 0 mA or 4 mA for minimum current (outputs will be set to 0–20 mA or 4–20 mA). ACTIVATION Dependent on Function selected previously. See section 4.5 on page 32 for additional information.
Select LINEAR CONTROL for current output to track the measurement valve. Select PID CONTROL for the sc100 to operate as a PID controller.
Each analog output is normally active, responding to the measured value of its assigned parameter. However, during calibration, each output can be transferred to this preset transfer value.
37
Page 38
Operation
4.9.3 System Setup Menu (continued)
RELAY SETUP (See section 4.6 on page 34 for expanded menu information.)
SELECT RELAY A, B, or C
SELECT SOURCE Select from none, any connected sensor, or the real time clock (RTC). SET PARAMETER Press ENTER to select from the displayed parameters.
SET FUNCTION
SET TRANSFER Sets the relay to Energize or De-energize (user-selectable). ACTIVATION Activate the relays from this menu (dependent on Function selected).
NETWORK SETUP (this menu appears only if a network card is installed in the controller)
MODBUS ADDRESS
BAUD RATE Select a baud rate of 9600, 19200, 38.4K, 57.6K, or 115.2K. Default: 19200 STOP BITS Select 1 or 2 stop bits. Default: 1 MODBUS MODE Select RTU or ASCII. Default: RTU DATA ORDER Select NORMAL or SWAPPED.
DISPLAY SETUP
ADJ CONTRAST
LANGUAGE
SET DATE/TIME
Select from the available options to customize the relay functions. See section 4.6.1 on page 34 for additional details.
Highlight sc100 Analyzer, or either connected sensor then press ENTER to select. Choose a number between 1 and 247 as the address (each source must have a different address) then press ENTER.
Use the UP and DOWN keys to increase or decrease the contrast, see section 4.3.2 on page 28. Range = 0–50
The default is English. Choose from the available options to allow all menus to appear in the selected language.
Use this menu to select the date format and to set the date and time (24-hour (military) format), see
section 4.3.4 on page 29.
SECURITY SETUP (Enter a 6-digit passcode)
SET PASSCODE
ENABLE Enables system security. See section 4.4.1 on page 31. DISABLE Disables system security. See section 4.4.1 on page 31.
LOG SETUP (Not used for 1720E system. Enable datalogging from the sensor setup menu for 1720E)
DATALOG SETUP Set up datalogging of data and events. See section 4.7.1 on page 36.
ERROR HOLD MODE
HOLD OUTPUTS Holds outputs when unable to communicate with the sensor. XFER OUTPUTS Goes to transfer state when unable to communicate with the sensor.
4.9.4 Test/Maint Menu
STATUS
Indicates the status of each relay and indicates which sensors are connected to the controller.
OUTPUT CAL
SELECT OUTPUT 1 or 2
Calibrate Analog Output by specifying values to correspond to 4 mA and 20 mA.
HOLD OUTPUTS
SET OUTMODE Choose Hold Outputs or Xfer Outputs. SET CHANNELS Choose any individual attached sensor or all attached sensors to be held or transferred. ACTIVATION Select Launch or Release.
38
Page 39
4.9.4 Test/Maint Menu (continued)
OVERFEED RESET
Reset the overfeed time out.
TEST OUTPUT
SELECT OUTPUT 1 or 2
User selectable mA value. 0–20 mA
TEST RELAY
SELECT RELAY A, B, or C
Energize or de-energize the selected relay.
RESET CONFIG
Reset to default configuration of the controller
SIMULATION
SELECT SOURCE, SET PARAMETER, SET SIM VALUE
Simulate sensor measurement values for testing the outputs and relays.
SCAN SENSORS
Manually scans for sensors to determine if sensors have been added or removed.
MODBUS STATS
Indicates the communication statistics for use with an external network.
Operation
CODE VERSION
Indicates the controller software version.
39
Page 40
Section 5 System Startup
5.1 General Operation
Plug the sensor into the unpowered controller by aligning the orientation tab on the cable connector with the channel in the controller connector. Push in and turn to secure the connection. Tug gently to check the connection.
After all plumbing and electrical connections have been completed and checked, place the head on the body and supply power to the system. Ensure the head is seated on the body when power is applied, since dark readings are measured at this time. If power is applied while the sensor head is off the turbidimeter body, cycle the power with the sensor head on the body.
The first time a controller is powered up, a language selection menu will appear. The user must select the correct language from the displayed options. Use the UP and DOWN keys to highlight the appropriate language and press ENTER to select.
Following language selection and upon power-up, the controller will search for connected sensors. The display will show the main measurement screen. Press the MENU key to access the menus.
5.2 Starting Sample Flow
Start sample flow through the instrument by opening the sample supply valve. Allow the turbidimeter to run long enough for the tubing and body to become completely wetted and the reading on the display to stabilize. One to two hours or longer may be required initially for complete stabilization. Allow measurements to become stable through adequate conditioning before completing instrument settings or performing calibrations.
5.3 Sensor Calibration and Verification
The manufacturer offers two EPA-approved calibration methods one using user­prepared formazin and the other using StablCal® stabilized formazin. Two verification methods (wet and dry) are also offered.
The 1720E Turbidimeter is factory-calibrated using StablCal before shipment. The instrument must be recalibrated before use to meet published accuracy specifications. In addition, recalibration is recommended after any significant maintenance or repair and at least once every three months during normal operation. The turbidimeter body and bubble trap must be thoroughly
cleaned and rinsed before initial use and prior to each calibration.
Tips to achieve the most accurate calibrations:
Optimum performance is achieved when calibration is performed in the
turbidimeter body. Accurately prepare the standard then add it to the turbidimeter body at the appropriate step in the procedure. Do not prepare the standard in the body.
®
Stabilized Formazin
40
Stop sample flow, drain, and clean the turbidimeter body before beginning the calibration procedure.
Always clean the photocell window per the instructions in section 6.4.2 on
page 48. Rinse the photocell with deionized water and dry with a soft, lint-free
cloth before calibrating.
Page 41
System Startup
Always clean the turbidimeter body or calibration cylinder per the instructions
in section 6.4.3 on page 48. Rinse with deionized water before calibrating.
Store the calibration cylinder upside-down to minimize contamination between calibrations.
Pour the calibration standard into the turbidimeter body at the inflow end (left side when facing the instrument).
Gently invert StablCal standards for 1 minute before opening. Do not shake. This ensures a consistent turbidity of the standard.
If the 20.0 NTU StablCal standard is allowed to sit in the calibration cylinder or turbidimeter body for more than 15 minutes, it must be remixed (gently swirled in the calibration cylinder) before use to ensure a consistent turbidity.
Discard all standards after use per the instructions on the container. Never transfer the standard back into its original container. Contamination will result.
Always recalibrate after restoring default settings.
5.3.1 User-prepared Calibration
Before starting the calibration, read and apply the tips in section 5.3.
Follow the procedure as written (using 1 L of deionized water and 5.0 mL of 4000 NTU formazin) if using a calibration cylinder for calibration.
If using the turbidimeter body for the user-prepared calibration follow the procedure below using 20 NTU formazin in step 6c. Prepare the standard as follows:
1. Stop the sample flow, then drain and clean the body.
2. Prepare a 20 NTU standard by adding 5.0 mL of 4000 NTU formazin to a
1-L flask. Dilute to the mark with deionized water and invert gently to mix.
3. Drain the deionized water and pour the prepared 20 NTU standard into the turbidimeter body at step 6c. Do not add additional 4000 NTU formazin.
Step Select Menu Level/Instructions Confirm
1
2
MAIN MENU
SENSOR SETUP
3
4
5
USER PREPD CAL
SELECT SENSOR (if more than one sensor is connected)
CALIBRATE
41
Page 42
System Startup
Step Select Menu Level/Instructions Confirm
6
a
Stop sample flow. Drain body and clean the body and bubble trap.
b
c
ADD 5 ML OF 4000 NTU FORMAZIN INTO CAL CYLINDER.
d
Select ACTIVE, HOLD, or TRANSFER
FILL CYL WITH 1 L DI WATER. REPLACE HEAD.
Measured reading (based on a gain of 1.0) displayed
Measured reading (based on a gain of 1.0) displayed
e
f
OUTPUT MODE
(Remove head)
GOOD CAL!
GAIN: X.XX
ENTER TO CONT
Verify CAL?
(see Note below)
(to store)
to verify
exit no verify
7
8
RETURN SENSOR TO MEASURE MODE
9
5.3.2 Calibration with StablCal
Select VERIFICATION type (begin at step 7 in section 5.4.1 on page 44 or
section 5.4.2 on page 45) or enter initials to complete calibration.
Main Menu or Main Measurement Screen
Note: If a dry verification is performed directly after a calibration, the measured value is assigned as the expected value for future verifications (when using the dry verification device with the same serial number). As long as the verification exists within the verification history, the expected value will be retained. Otherwise, the expected value will be the nominal labeled value associated with the dry verification device.
®
Before starting the calibration, read and apply the tips in section 5.3 on page 40.
Step Select Menu Level/Instructions Confirm
1
MAIN MENU
2
42
SENSOR SETUP
Page 43
System Startup
Step Select Menu Level/Instructions Confirm
3
—CALIBRATE
4
5
6a
b
c
d
STABLCAL CAL
OUTPUT MODE
Select ACTIVE, HOLD, or TRANSFER
Drain/clean/rinse the turbidimeter body or cal cylinder.
POUR 20 NTU STD INTO CYL/BODY. REPLACE HEAD
Measured reading and reading based on 1.0 gain displayed
GOOD CAL!
GAIN: X.XX
ENTER TO CONT
Verify CAL? (see Note below)
(to store)
to verify/
e
7
RETURN SENSOR TO MEASURE MODE
Select VERIFICATION type (begin at step 7 in section 5.4.1 on page 44 or section 5.4.2
8
Note: If a dry verification is performed directly after a calibration, the measured value is assigned as the expected value for future verifications (when using the dry verification device with the same serial number). As long as the verification exists within the verification history, the expected value will be retained. Otherwise, the expected value will be the nominal value associated with the dry verification device.
5.4 Instrument Verification
Instrument verification is intended as a simple check to ensure turbidimeter functionality between calibrations. A verification is initially performed directly after a calibration and subsequent independent verifications are referenced to the initial verification. The pass/fail criteria is set and subsequent verifications are deemed good or bad, when compared to the initial verification. All verifications are based on the current calibration and must be repeated when the instrument is recalibrated or when the pass/fail criteria is not met.
exit no verify
on page 45) or enter initials to complete calibration.}
Main Menu or Main Measurement Screen
43
Page 44
System Startup
Two types of verifications are offered. The dry verification is performed using a “dry” calibration device such as the ICEPIC™. A wet verification is performed using a standard with a predetermined value such as StablCal
®
Stabilized Formazin or a user-prepared standard with a value that has been verified on an independent device such as a laboratory turbidimeter.
5.4.1 Dry Verification
Step Select Menu Level/Instructions Confirm
1
2
3
4
5
6
7
8
PERFORM VER
Choose ACTIVE, HOLD, or TRANSFER
VER TYPE Select DRY
select 1 NTU STD or 20 NTU STD or verify SN of previously used calibration device
MAIN MENU
SENSOR SETUP
CALIBRATE
VERIFICATION
OUTPUT MODE
DRY VERIFY
Set Head on Standard
9
10
RETURN SENSOR TO MEASURE MODE
11
44
Reading Displayed
(to accept)
GOOD VER!
(to store)
ENTER INITIALS (user input)
MAIN MENU or Main Measurement Screen
Page 45
System Startup
5.4.2 Wet Verification
Before starting the verification, read and apply the appropriate tips in section 5.3
on page 40.
Step Select Menu Level/Instructions Confirm
1
2
3
—CALIBRATE
4
5
6
PERFORM VER
Choose ACTIVE, HOLD, TRANSFER
7
a.
DRAIN AND CLEAN SENSOR BODY. ENTER TO CONT
MAIN MENU
SENSOR SETUP
VERIFICATION
OUTPUT MODE
Select VER TYPE
Select WET
Enter Std Turbidity
8
9
10
11
b.
c.
d.
POUR STANDARD INTO CYL/BODY. PLACE HEAD ON. ENTER TO CONT
Reading Displayed
GOOD VER!
ENTER INITIALS
RETURN SENSOR TO MEASURE MODE
Main Menu or Main Measurement Screen
(to accept)
45
Page 46
System Startup
5.5 Calibration and Verification History
The calibration and verification history logs contain information on the last 12 calibrations and the last 12 verifications. The calibration history log shows the gain value, the time and date of the calibration, and the initials of the operator performing verification.
Note: Restoring default settings from the DIAG/TEST menu will return the turbidimeter to its non calibration state (gain = 1.0) but it will not remove the previous calibration history from memory.
The calibration history log is accessed from the Calibrate menu. The verification history log is accessed from the Verification menu (a submenu of the Calibrate menu).
Each verification history entry shows the serial number of the verification device (dry verification) or the value of the verification standard (wet verification), the time and date of the verification, and the initials of the operator performing the verification.
Scroll through the entries by pressing the ENTER key. After scrolling through all 12 histories, the display will return to the calibration menu level.
When the instrument is received from the factory, default values or blank spaces will be shown for the calibration and verification history information. Those values will be replaced with real data as the history log is filled.
The data is retained as first in, first out. When the log is full, the newest entry is stored and the oldest entry in the log is deleted.
46
Page 47
Section 6 Maintenance
DANGER Only qualified personnel should conduct the maintenance tasks described in this section of the manual.
6.1 Maintenance Schedule
Maintenance Task Frequency
Clean the sensor
Calibrate Sensor (as required by regulatory agency) Per agency-dictated schedule.
Scheduled periodic maintenance requirements of the 1720E are minimal and include calibration and cleaning of the photocell window, bubble trap, and body. Check and clean the bubble trap and turbidimeter body (as described in
section 6.4.3) if visual inspection shows that it is necessary. Perform other
maintenance on a regular basis; experience will dictate scheduling and may depend on the installation, sample type, and season.
Before each calibration and as needed. Depends on sample characteristics.
It is very important to maintain the cleanliness of the interior and exterior of the turbidimeter body, head assembly, the integral bubble trap, and the surrounding area. Doing so will ensure accurate, low-level turbidity measurements.
Clean the body before calibration and verification (especially when measurements are being made at 1.0 NTU or lower).
Check and/or perform a calibration periodically (as experience dictates) using one of the methods described in section 5.3 on page 40. A calibration history menu option is available under Sensor Setup/Calibrate.
6.2 Removing a Sensor from the System
Prior to physically removing a sensor from the system, record all user defined settings such as relays, signal averaging, etc. Disconnect the sensor connector at the controller.
6.3 Reinstalling a Sensor on the System
To return the system to normal operation following a software upgrade or sensor repair perform the following procedure:
1. Detach all sensors from the sc100 controller.
2. From the main menu, press the down arrow key to highlight TEST/MAINT. Press ENTER.
Note: Clean sensors before reinstallation on the system.
3. Use the down arrow key to scroll to SCAN SENSORS and press ENTER.
4. Remove attached sensors by selecting the corresponding serial number or
select “All”.
5. Power down the sc100 then attach the 1720E.
6. Supply power to the sc100. The system will initialize automatically.
47
Page 48
Maintenance
6.4 Cleaning
6.4.1 Cleaning the Controller
With the enclosure securely closed, wipe the exterior with a damp cloth.
6.4.2 Cleaning the Photocell Window
Occasional cleaning of the photocell window is required. The frequency will depend on the nature and concentration of dissolved and suspended solids in the sample. Biological activity is a primary factor in mineral scale deposit on the window and the amount differs with sample temperature. In general, more growth will occur in warm temperatures and less in cold.
Note: Take care to not scratch the photocell window.
Inspect the photocell window often to determine cleaning needs. Remove any organic growth or film on the photocell window before standardization or calibration. Use a cotton swab and isopropyl alcohol or a mild detergent (such as Liqui-nox cleaning with a mild acid applied with a cotton swab followed by a detergent wash.
Do not use abrasive cleaners.
®
) to remove most sediment and dirt. Mineral scale buildup may require
6.4.3 Cleaning the Turbidimeter Body and Bubble Trap
Sediment may collect in the turbidimeter body after extended use. Noise (fluctuation) in the reading could indicate the need to clean the body and/or bubble trap. The 1720E bubble trap and bottom plate may be removed to make cleaning easier. Drain and clean the turbidimeter body before each calibration. Establish a regular schedule or perform cleaning as determined by visual inspection.
Cleaning the Turbidimeter Body
Note: The turbidimeter body, bubble trap, and detector must be cleaned before each calibration.
1. Turn off sample flow to the turbidimeter body.
2. Remove the head assembly and bubble trap cover from the body. Remove the
bubble trap by lifting it vertically. Set it aside to be cleaned separately.
3. Drain the body by removing the plug from the bottom of the body.
4. Replace the drain plug and fill the body to the weir with cleaning solution. This
cleaning solution can consist of dilute chlorine solution (25 mL of household bleach in 3.78 liters of water) or a laboratory detergent such as Liqui-nox (1 mL detergent in 1 liter of water).
48
5. Use a soft brush to clean the inside surfaces of the body.
6. Remove the drain plug again and thoroughly flush the turbidimeter body with
ultra-filtered deionized water. Clean and replace the plug.
Cleaning the Bubble Trap
1. Prepare a cleaning solution (as in step 4 above) in a container large enough
to submerge the entire bubble trap.
2. Using a test tube brush such as Cat. No. 690-00, clean each surface.
3. Rinse the bubble trap thoroughly with ultra-filtered deionized water and
reinstall it in the turbidimeter body.
4. Replace the bubble trap cover and head assembly on the top of the body.
Page 49
Maintenance
5. Restore sample flow to the instrument.
6. Calibrate the instrument using one of the methods in section 5.3 on page 40.
If the above cleaning procedures have been performed and the turbidimeter readings are still noisy, the bottom plate and gasket may need to be removed and cleaned. Carefully perform the following procedure to ensure the turbidimeter body integrity is maintained.
1. Turn off sample flow to the turbidimeter body.
2. Remove the head assembly, bubble trap cover, and bubble trap (by lifting it
vertically) from the body.
3. Drain the body by removing the plug from the bottom of the body.
4. Lift the body off of its mounting screws.
5. With the body turned upside-down, remove the two Phillips-head screws
holding the bottom plate.
6. Lift the bottom plate off the body; set the gasket aside for use in reassembly.
7. Use a soft brush and a dilute cleaning solution (as prepared above) to clean
the bottom plate and inside surfaces of the turbidimeter body. Rinse the entire body and bottom plate with ultra-filtered deionized water.
8. Reassemble by inserting the gasket into the molded channel in the bottom plate.
9. Fit the bottom plate onto the turbidimeter body.
10. Reinstall both screws and carefully tighten to 15 inch-lb maximum.
11. Reinstall the turbidimeter onto the wall mounting screws.
12. Replace the bubble trap, bubble trap cover, and head assembly on the top of
the body.
13. Restore sample flow to the instrument.
6.4.4 Replacing the Lamp Assembly
The Lamp Assembly is located on the head assembly. Under normal use, Hach recommends replacing the lamp once a year to maintain peak performance. Replacement bulbs have been “burned-in” at the factory and are ready for installation and use.
To change the lamp, refer to Figure 28 on page 50 and perform the following steps:
1. Disconnect power to the controller to remove all power to the turbidimeter.
2. Disconnect the lamp leads by unplugging the connector.
3. After the bulb has cooled, remove as follows:
49
Page 50
Maintenance
Figure 28 Lamp Replacement
a. Wear cotton gloves to protect your hands and to avoid fingerprints on
the bulb.
b. Grasp the bulb.
c. Twist the bulb in a counterclockwise direction, pulling out slightly, until it is
released from the housing.
d. Pull the lamp leads and connector through the hole in the lamp housing.
Do not touch the new bulb with bare hands. Etched glass and reduced lamp life will result. Wear cotton gloves or grasp the lamp assembly with a tissue to avoid contamination. If contamination occurs, clean the glass bulb portion with isopropyl alcohol.
Replace the bulb by reversing the above instructions. The bulb base only fits one way; align the notch in the metal bulb flange with the hole in the lamp holder.
1
1. Lamp Housing 2. Lamp Assembly 3. Photo Detector
2
3
50
Page 51
6.5 Fuse Replacement
The controller contains two mains fuses. Failed fuses are an indication that an equipment problem could exist. Problem resolution and fuse replacement should only be performed by qualified service personnel. Refer to Figure 29 and perform the following steps to replace the fuses:
1. Disconnect power to the controller (including power to relays if powered).
2. Open the hinged controller cover by completely loosening all four captive
screws in the cover.
3. Remove the high voltage barrier; pull out on the lever of the captive fastener then pull straight out on the barrier. Set the barrier aside for reinstallation.
4. Remove the fuses and install new fuses of the same type and rating (T, 1.6A, 250 V, slow blow).
5. Reinstall the high voltage barrier.
6. Close the controller cover and hand-tighten the four screws.
Maintenance
Figure 29 Fuse Replacement
J1
S1
J4
NETWORK
INTERFACE
J2
U5
U9
CARD
7. Reconnect all power to the instrument.
J3
J5
NCNCNC
NONO
RELAYCRELAYBRELAYA
J6
1
F1
F2
NO
COMCOMCOM
1. Fuses F1 and F2, 1.6 A
Disconnect Power
51
Page 52
Section 7 Troubleshooting
7.1 Error Codes
When a sensor is experiencing an error condition, the sensor reading on the measurement screen will flash and all relays and analog outputs associated with the sensor will be held.
Highlight the Probe Diag menu and press ENTER. Highlight Errors and press
ENTER to determine the case of the error. Errors are defined in Table 9.
Table 9 Error Codes
Displayed Error Definition
A/D Fail Failed A/D converter. Call the Service Department.
Lamp Fail
Flash Fail Datalog and event log will not work.
7.2 Warnings
A sensor warning will leave all menus, relays, and outputs functioning normally, but will cause a warning icon to flash on the right side of the display. Highlight the Probe Diag menu and press ENTER to determine the cause of the error. Warnings are defined in Table 10.
The light source has failed. See section 6.4.4 on page 49 for instructions for replacing it.
A warning may be used to trigger a relay and users can set warning levels to define the severity of the warning. Errors are defined in Table 10.
Table 10 Warning Codes
Warning
Number
1 Dark Reading Warning Dark reading detects too much light.
2 Temperature Warning
3 Data Log Full Warning
4 Event Log Full Warning
5 5 Volt Warning
6 Voltage in Warning
7 Lamp V Warning
8 Lamp I Warning
9 Output Mode Not Normal
Displayed Warning Definition/Resolution
Sensor head internal temperature is higher than specified. Contact the Service Department.
Sensor data log is full. No additional data will be logged until sensor log is downloaded into controller memory.
Sensor data log is full. No additional data will be logged until sensor log is downloaded into controller memory.
Monitored voltage is outside the range of 4.5 – 5.5 V @ microprocessor input -> 698 to 854 counts from the microprocessor ADC.
Monitored voltage is outside the range of 9.08 – 14.3 V @ microprocessor pin -> 279 to 435 counts from the microprocessor ADC.
Monitored voltage is outside the range of 3.96 – 4.48 V @ microprocessor pin -> 614 to 385 counts from the microprocessor ADC.
Monitored voltage is outside the range of 1.67 – 2.75 V @ microprocessor pin -> 39 to 64 counts from the microprocessor ADC.
Activated when the sensor is not in normal measurement mode (such as when in calibration or verification mode).
52
Page 53
Tabl e 11 presents sensor warnings displayed in the Event Log, possible causes,
and corrective actions.
Table 11 Troubleshooting
Sensor Warning Possible Cause Corrective Action
Lamp burned out Replace the lamp. See section 6.4.4 on page 49. Lamp unplugged Restore connection
Bad Lamp
Low Signal
Bad System Voltage
A/D Converter Timeout
High Dark Counts
+12 V connection loose at controller Restore connection Dislodged lamp Reinstall lamp Bad circuit board in turbidimeter head Contact the Customer Service Department.
Photocell coated/dirty
Photocell wires disconnected Reconnect wires
Photocell broken/cracked
Lens coated/dirty Clean the lens using isopropyl alcohol and a cotton swab. Obstructed light path Remove obstruction Sample turbidity >100 NTU Switch to a high range turbidimeter See Bad Lamp causes above See Bad Lamp corrective actions above
Improper wiring at controller
Turbidimeter head cable shortened to improper length
Fluctuation in voltage Turn instrument power off and back on. Bad circuit board in turbidimeter head Contact the Service Department Fluctuation in voltage Turn instrument power off and back on. Bad circuit board in turbidimeter head Contact the Customer Service Department. Light Leak—Turbidimeter head not on
turbidimeter body or calibration cylinder during Power Up or Zero Electronics
Light Leak—Turbidimeter head not properly aligned on the turbidimeter body or calibration cylinder during Power Up or Zero Electronics
Photocell broken/cracked Contact the Customer Service Department.
See Cleaning the Photocell Window on page 48. Contact the Customer Service Department.
Replace photocell Contact the Customer Service Department.
See the controller wiring instructions in section 3.2,
Electrical Installation on page 14.
Contact the Customer Service Department.
Make sure the turbidimeter head is on the turbidimeter body and properly aligned and repower instrument or perform ZERO ELECTRONICS in the CALIBRATION MENU.
Make sure the turbidimeter head is properly aligned and repower instrument or perform ZERO ELECTRONICS in the CALIBRATION MENU.
Troubleshooting
Table 12 Additional Malfunctions Not Recorded in the Event Log
Symptom Possible Cause Corrective Action
Continuous Underrange (negative reading)
Continuous Overrange (100 NTU)
Calibration standards not in the correct order or incorrect dilution
Calibration standards not in the correct order or incorrect dilution
The following table presents additional malfunctions which may not be recorded in the Event Log.
Verify the accuracy of calibration standards and recalibrate the instrument. See Low Signal in Table 11.
Verify the accuracy of calibration standards and recalibrate the instrument.
53
Page 54
Troubleshooting
Table 12 Additional Malfunctions Not Recorded in the Event Log (continued)
Symptom Possible Cause Corrective Action
Erratic Readings Calibration standards have the
same value Inadequate bubble removal from sample
High Readings Deionized water turbidity is
greater than 0.5 NTU
Verify the accuracy of calibration standards and recalibrate the instrument. Increase the signal averaging time to a longer interval. Make sure the Bubble Reject feature is turned on. Slow the flow of sample into the instrument.
Clean the instrument. Access Calibration History for turbidity value of ultra-filtered water. Verify the flow is between 200–750 mL/min. Recalibrate the instrument.
7.3 Event Codes
Event codes are not displayed on the screen of the sc100 and must be downloaded from the event log to be viewed. See section 4.7 on page 36 for additional information. Troubleshooting actions are given in section Table 11 on
page 53.
Table 13 Event Log List
Event Event # Data1 Data2 Data3
Bubble Reject Change 0
Filter Size Change 1
Data Log Interval Change 2
Power On 3 1 — Calibration 4 Cal Gain Initials — Verification 5 Expected Value Meas Value Initials Dark Reading Warning 6 Measured Value — Temperature Warning 7 Temperature Value
Voltage Warning 8
Lamp Warning 9
A/D Fail 10 1 — Lamp Fail 11 2 — Flash Fail 12 3
Output Mode Change 13
0 = OFF 1= ON
0 = No avg 1 = 6 sec 2 = 30 sec 3 = 60 sec 4 = 90 sec
0 = 30 sec 1 = 1min 2 = 5 min 3 = 10 min 4 = 15 min
16 = +5V high 32 = +5V low 64 = V in high 128 = V in low
1 = lamp V high 2 = lamp V low 4 = lamp I high 8 = lamp I low
1 = Hold 2 = Transfer
——
——
——
——
——
——
54
Page 55
Section 8 Replacement Parts and Accessories
Replacement Items
Item Cat. No.
Fuse, T, 1.6 A, 250 V .......................................................................................................................................... 52083-00
Instruction Manual, sc100 Controller, English .................................................................................................... 58600-18
Instruction Manual, 1720E Turbidimeter System, English .................................................................................. 60100-18
Installation kit, sc100 Controller.......................................................................................................................... 58672-00
Lamp Assembly .................................................................................................................................................. 18950-00
sc100 Controller ................................................................................................................................................. 58600-00
Optional Accessories
Cable, sensor extension, 7.7 m (25 ft)................................................................................................................ 57960-00
Cap, Connector Receptacle ...................................................................................................... each ................ 52100-00
Deionized (demineralized) water ..................................................................................................1 L .................... 272-56
Digital Output Card for ModBUS RS232 communication ................................................................................... 59200-00
Digital Output Card for ModBUS RS485 communication ................................................................................... 59200-01
Drain plug for the 1720E body................................................................................................... each................ 44116-00
Filter, 0.45 µm, to produce ultra-filtered water for cleaning and calibration ............................... each ................ 26705-00
Filter, 0.2 µm, to produce ultra-filtered water for calibration standard preparation .................... each................ 23238-10
Formazin Calibration Kit includes:
Calibration Cylinder, TenSette
Floor Stand ............................................................................................................................... each ................ 57432-00
Flow meter, 500–700 mL/min .................................................................................................... each................ 40282-00
ICE-PIC Module for Calibration and Calibration Verification
20 NTU ................................................................................................................................. each ................ 52250-00
1 NTU................................................................................................................................... each ................ 52215-00
0.5 NTU................................................................................................................................ each ................ 52225-00
Lid, Bubble Trap, 1720E ............................................................................................................ each............... 52012-00
Photocell Replacement Kit for the 1720E .................................................................................. each ................ 52180-00
Pipet tips for 19700-01 TenSette Pipet ................................................................................... 50/pkg ................ 21856-96
Pipet tips for 19700-10 TenSette Pipet ................................................................................... 50/pkg ................ 21997-96
Plug, conduit opening ......................................................................................................................................... 58687-00
Power Cord with strain relief, 115 V.................................................................................................................... 54488-00
Power Cord with strain relief, 230 V ................................................................................................................... 54489-00
Power Cord Kit, 10A-125V, 1.8 m (6 ft), UL/CSA listed ............................................................. each ................ 46306-00
Power Cord Kit, 10A-230V, 1.8 m (6 ft), European-style plug, VDE approved .......................... each ................ 46308-00
StablCal Calibration Set for the 1720 Series Turbidimeter
Includes: StablCal Standards, < 0.1 NTU, 20.0 NTU........................................................... 1 L/each ................ 26596-00
StablCal Standard, 0.1 NTU ......................................................................................................... 1 L ................ 27233-53
StablCal Standard, 0.3 NTU ......................................................................................................... 1 L ................ 26979-53
StablCal Standard, 0.5 NTU ......................................................................................................... 1 L ................ 26980-53
StablCal Standard, 1.0 NTU ......................................................................................................... 1 L ................ 26598-53
Strain relief, Heyco .................................................................................................................................................. 16664
Sun shield........................................................................................................................................................... 58690-00
Swabs, Cotton, presterilized for cleaning the photodetector ................................................ 100/pkg ................ 25543-00
TenSette Pipet, 0.1 to 1.0 mL .................................................................................................... each................ 19700-01
TenSette Pipet, 1.0 to 10.0 mL .................................................................................................. each................ 19700-10
Tubing, Inlet, ¼ inch O.D., Polyethylene ................................................................................per foot ................ 51322-00
Tubing, Outlet, ½ inch I.D., ¾ inch O.D., Tygon R3603..........................................................per foot................ 51263-00
®
Pipet, 4000 NTU Formazin Primary Standard (500 mL).... each................ 44156-00
55
Page 56
Replacement Parts and Accessories
Calibration and Verification Standards and Accessories
Item Qty Cat. No.
Calibration Cylinder....................................................................................................................each.................44153-00
Formazin Calibration Standards
Formazin, 4000 NTU Stock Solution................................................................................ 500 mL...................2461-49
Calibration/Verification Modules
ICE-PIC™ Module, 1 NTU .................................................................................................1 each.................52215-00
ICE-PIC™ Module, 20 NTU ...............................................................................................1 each.................52250-00
StablCal
®
StablCal StablCal StablCal StablCal StablCal StablCal
Calibration Standards
®
Stabilized Formazin Standard, 1 NTU.................................................................... 1 L.................26598-53
®
Stabilized Formazin Standard, 20 NTU ................................................................. 1 L.................26601-53
®
Stabilized Formazin Standard, <0.1 NTU............................................................... 1 L.................26597-53
®
Stabilized Formazin Set, four 1-L bottles 20-NTU and four 1-L bottles <0.1 NTU1 L each...........26596-00
®
Stabilized Formazin Standard, 40 NTU............................................. 1 gallon (3.78 L).................27463-56
®
Stabilized Formazin Standard, 0.1 NTU............................................ 1 gallon (3.78 L).................27233-56
56
Page 57
Section 9 How to Order
U.S.A. Customers
By Telephone:
6:30 a.m. to 5:00 p.m. MST Monday through Friday (800) 227-HACH (800-227-4224)
By Fax:
(970) 669-2932
By Mail:
Hach Company P.O. Box 389 Loveland, Colorado 80539-0389 U.S.A.
Ordering information by e-mail: orders@hach.com
9.1 Information Required
Hach account number (if available) billing address
Your name and phone number Shipping address
Purchase order number Catalog number
Brief description or model number Quantity
9.2 International Customers
Hach maintains a worldwide network of dealers and distributors. To locate the representative nearest you, send e-mail to intl@hach.com or contact:
Hach Company World Headquarters; Loveland, Colorado, U.S.A. Telephone: (970) 669-3050; Fax: (970) 669-2932
9.3 Technical and Customer Service (U.S.A. only)
Hach Technical and Customer Service Department personnel are eager to answer questions about our products and their use. Specialists in analytical methods, they are happy to put their talents to work for you.
Call 1-800-227-4224 or e-mail techhelp@hach.com
57
Page 58
Section 10 Repair Service
Authorization must be obtained from Hach Company before sending any items for repair. Please contact the Hach Service Center serving your location.
In the United States:
Hach Company Ames Service 100 Dayton Avenue Ames, Iowa 50010 (800) 227-4224 (U.S.A. only) FAX: (515) 232-3835
In Canada:
Hach Sales & Service Canada Ltd. 1313 Border Street, Unit 34 Winnipeg, Manitoba R3H 0X4 (800) 665-7635 (Canada only) Telephone: (204) 632-5598 FAX: (204) 694-5134 E-mail: canada@hach.com
In Latin America, the Caribbean, the Far East, the
Indian Subcontinent, Africa, Europe, or the Middle East: Hach Company World Headquarters, P.O. Box 389 Loveland, Colorado, 80539-0389 U.S.A. Telephone: (970) 669-3050 FAX: (970) 669-2932 E-mail: intl@hach.com
58
Page 59
Section 11 Limited Warranty
Hach Company warrants its products to the original purchaser against any defects that are due to faulty material or workmanship for a period of one year from date of shipment unless otherwise noted.
In the event that a defect is discovered during the warranty period, Hach Company agrees that, at its option, it will repair or replace the defective product or refund the purchase price, subject to the pro-rated schedule above, excluding original shipping and handling charges. Any product repaired or replaced under this warranty will be warranted only for the remainder of the original product warranty period.
This warranty does not apply to consumable products such as chemical reagents; or consumable components of a product, such as, but not limited to, lamps and tubing.
Contact Hach Company or your distributor to initiate warranty support. Products may not be returned without authorization from Hach Company.
Limitations
This warranty does not cover:
Damage caused by acts of God, natural disaster, labor unrest, acts of war (declared or
undeclared), terrorism, civil strife or acts of any governmental jurisdiction
Damage caused by misuse, neglect, accident or improper application or installation
Damage caused by any repair or attempted repair not authorized by Hach Company
Any product not used in accordance with the instructions furnished by Hach Company
Freight charges to return merchandise to Hach Company
Freight charges on expedited or express shipment of warranted parts or product
Travel fees associated with on-site warranty repair
This warranty contains the sole express warranty made by Hach Company in connection with its products. All implied warranties, including without limitation, the warranties of merchantability and fitness for a particular purpose, are expressly disclaimed.
Some states within the United States do not allow the disclaimer of implied warranties and if this is true in your state the above limitation may not apply to you. This warranty gives you specific rights, and you may also have other rights that vary from state to state.
This warranty constitutes the final, complete, and exclusive statement of warranty terms and no person is authorized to make any other warranties or representations on behalf of Hach Company.
Limitation of Remedies
The remedies of repair, replacement or refund of purchase price as stated above are the exclusive remedies for the breach of this warranty. On the basis of strict liability or under any other legal theory, in no event shall Hach Company be liable for any incidental or consequential damages of any kind for breach of warranty or negligence.
59
Page 60
Section 10 Compliance Information
Hach Co. certifies this instrument was tested thoroughly, inspected and found to meet its published specifications when it was shipped from the factory.
The Model sc100 with 1720E Sensor has been tested and is certified as indicated to the following instrumentation standards:
Product Safety
UL 61010A-1 (ETL Listing # 65454) CSA C22.2 No. 1010.1 (ETLc Certification # 65454) Certified by Hach Co. to EN 61010-1 Amds. 1 & 2 (IEC1010-1) per 73/23/EEC, supporting test records by Intertek Testing Services.
Immunity
This equipment was tested for Industrial level EMC per:
EN 61326 (EMC Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use) Hach Company, certified compliance by Hach Company.
per 89/336/EEC EMC: Supporting test records by
Emissions
Standards include:
IEC 1000-4-2:1995 (EN 61000-4-2:1995) Electro-Static Discharge Immunity (Criteria B) IEC 1000-4-3:1995 (EN 61000-4-3:1996) Radiated RF Electro-Magnetic Field Immunity (Criteria A) IEC 1000-4-4:1995 (EN 61000-4-4:1995) Electrical Fast Transients/Burst (Criteria B) IEC 1000-4-5:1995 (EN 61000-4-5:1995) Surge (Criteria B) IEC 1000-4-6:1996 (EN 61000-4-6:1996) Conducted Disturbances Induced by RF Fields (Criteria A) IEC 1000-4-11:1994 (EN 61000-4-11:1994) Voltage Dip/Short Interruptions (Criteria B)
Additional immunity Standard/s include:
ENV 50204:1996 Radiated Electro-Magnetic Field from Digital Telephones (Criteria A)
This equipment was tested for Radio Frequency Emissions as follows:
Per 89/336/EEC EMC: EN 61326:1998 (Electrical Equipment for measurement, control and laboratory use—EMC requirements) Class “A” emission limits. Supporting test records by Hewlett Packard, Fort Collins, Colorado Hardware Test Center (A2LA # 0905-01) and certified compliance by Hach Company.
60
Standards include:
EN 61000-3-2 Harmonic Disturbances Caused by Electrical Equipment EN 61000-3-3 Voltage Fluctuation (Flicker) Disturbances Caused by Electrical Equipment
Page 61
Compliance Information
Additional Emissions Standard/s include:
EN 55011 (CISPR 11) Class “A” emission limits
Canadian Interference-causing Equipment Regulation, IECS-003, Class A
Supporting test records by Hewlett Packard, Fort Collins, Colorado Hardware Test Center (A2LA # 0905-01) and certified compliance by Hach Company.
This Class A digital apparatus meets all requirements of the Canadian Interference- Causing Equipment Regulations.
Cet appareil numÈrique de la classe A respecte toutes les exigences du RËglement sur le matÈriel brouilleur du Canada.
FCC PART 15, Class “A” Limits
Supporting test records by Hewlett Packard, Fort Collins, Colorado Hardware Test Center (A2LA # 0905-01) and certified compliance by Hach Company.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at his own expense. The following techniques of reducing the interference problems are applied easily.
1. Disconnect the Model sc100 Controller from its power source to verify that it is or is not the source of the interference.
2. If the Model sc100 Controller is connected into the same outlet as the device with which it is interfering, try another outlet.
3. Move the Model sc100 Controller and 1720E sensor away from the device receiving the interference.
4. Reposition the device receiving the interference.
5. Try combinations of the above.
61
Page 62
Appendix A ModBUS Register Information
Table 14 Controller ModBUS Registers
Group Name Tag Name Register # Data Type Length R/W Description
Measurements
Setup Language 40003
Setup Date Format 40004
Setup
Setup/Analog Output 1 Source 40006
Setup/Analog Output 1 Sensor Select 40007
Setup/Analog Output 1
Setup/Analog Output 1 Type 40009
Setup/Analog Output 1 Transfer Value 40010 Float 2 R/W Sets the transfer value
Setup/Analog Output 1 Filter 40012
Setup/Analog Output 1
Setup/Analog Output 1/Linear
Setup/Analog Output 1/Linear
Setup/Analog Output 1/PID
Setup/Analog Output 1/PID
Setup/Analog Output 1/PID
Setup/Analog Output 1/PID
Setup/Analog Output 1/PID
Setup/Analog Output 1/PID
Setup/Analog Output 1/PID
Setup/Analog Output 2 Source 40028
Setup/Analog Output 2 Sensor Select 40029
Setup/Analog Output 2
Calculated Value
Error Hold Mode
Measurement Select
0mA - 4mA Select
Min Setting 40014 Float 2 R/W Sets the min output value
Max Setting 40016 Float 2 R/W Sets the max output value
PID Mode 40018
PID Manual Set
PID Setpoint 40021 Float 2 R/W Sets the PID setpoint
PID Phase 40023
PID Proportional Band
PID Integral Time
PID Derivative Time
Measurement Select
40001 Float 2 R
Unsigned
Integer
Unsigned
Integer
40005
40008
40013
40019 Float 2 R/W
40024 Float 2 R/W Sets the PID proportional band
40026
40027
40030
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
1 R/W Current System Language
1R/W
1R/W
1R/W
1R/W
1R/W
1R/W
1R/W
1R/W
1 R/W Sets the PID mode (0 = auto; 1 = manual)
1R/W
1 R/W Sets the PID integral time (min)
1 R/W Sets the PID derivative time (min)
1R/W
1R/W
1R/W
Value calculated from two sensor measurements
Current Data Display Format (0 = DD/MM/YY; 1 = MM/DD/YY; 2 = DD-MM-YY; 3 = MM-DD-YY)
Error Mode Hold State (0 = Hold outputs; 1 = Transfer outputs to predefined value)
Selects data source for this output (0 = None; 2 = sensor; 4 = Calculation)
Selects sensor source when Source = Sensor (0 = sensor1; 1 = sensor2)
Selects measurement on the sensor (0 = Meas1.. 3 = Meas4)
Selects output type (0 = Linear output; 1 = PID control)
Sets the output filter value in seconds (0 to 120 sec.)
Selects 0mA/4mA for min output (0 = 0mA; 1 = 4mA)
Sets the PID manual output value (0.0 to 100.0%)
Sets the PID phase (0 = Direct; 1=Reverse)
Selects data source for this output (0 = None; 2 = Sensor; 4 = Calculation)
Selects sensor source when Source = Sensor (0 = Sensor1; 1 = Sensor2)
Selects measurement on the Sensor (0 = Meas1 .. 3 = Meas4)
62
Page 63
ModBUS Register Information
Table 14 Controller ModBUS Registers (continued)
Group Name Tag Name Register # Data Type Length R/W Description
Setup/Analog Output 2 Type 40031
Setup/Analog Output 2 Transfer Value 40032 Float 2 R/W Sets the transfer value
Setup/Analog Output 2 Filter 40034
Setup/Analog Output 2
Setup/Analog Output 2/Linear
Setup/Analog Output 2/Linear
Setup/Analog Output 2/PID
Setup/Analog Output 2/PID
Setup/Analog Output 2/PID
Setup/Analog Output 2/PID
Setup/Analog Output 2/PID
Setup/Analog Output 2/PID
Setup/Analog Output 2/PID
Setup/Relay 1 Source 40050
Setup/Relay 1 Sensor Select 40051
Setup/Relay 1
Setup/Relay 1 Type 40053
Setup/Relay 1
Setup/Relay 1/Alarm High Alarm 40055 Float 2 R/W Sets the high alarm setpoint Setup/Relay 1/Alarm Low Alarm 40057 Float 2 R/W Sets the low alarm setpoint
Setup/Relay 1/Alarm
Setup/Relay 1/Alarm
Setup/Relay 1/Alarm On Delay 40063
Setup/Relay 1/Alarm Off Delay 40064
Setup/Relay 1/Control Setpoint 40065 Float 2 R/W Sets the controller setpoint
0mA - 4mA Select
Min Setting 40036 Float 2 R/W Sets the min output value
Max Setting 40038 Float 2 R/W Sets the max output value
Mode 40040
Manual Set 40041 Float 2 R/W
Setpoint 40043 Float 2 R/W Sets the PID setpoint
Phase 40045
Proportional Band
Integral Time 40048
Derivative Time
Measurement Select
Transfer Setting
High Deadband
Low Deadband
40035
40046 Float 2 R/W Sets the PID proportional band
40049
40052
40054
40059 Float 2 R/W Sets the high alarm deadband
40061 Float 2 R/W Sets the low alarm deadband
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
1R/W
1R/W
1R/W
1 R/W Sets the PID mode (0 = auto; 1 = manual)
1R/W
1 R/W Sets the PID integral time (min)
1 R/W Sets the PID derivative time (min)
1R/W
1R/W
1R/W
1R/W
1R/W
1 R/W Sets the on delay time
1 R/W Sets the off delay time
Selects output type (0 = Linear output; 1 = PID control)
Sets the output filter value in seconds (0 to 120 sec)
Selects 0mA/4mA for min output (0 = 0mA; 1 = 4mA)
Sets the PID manual output value (0.0 to 100.0%)
Sets the PID phase (0 = Direct; 1 = Reverse)
Selects data source for this relay (0 = None; 1 = Real Time Clock; 2 = Sensor; 4 = Calculation)
Selects Sensor source when Source = Sensor (0 = Sensor1; 1 = Sensor2)
Selects measurement on the Sensor (0 = Meas1 .. 3 = Meas4)
Selects the relay type (0=Alarm;1=Control; 2 = Status; 3 = Timer; 4 = Event)
Selects the transfer value for the relays (0 = De-energized; 1 = Energized)
63
Page 64
ModBUS Register Information
Table 14 Controller ModBUS Registers (continued)
Group Name Tag Name Register # Data Type Length R/W Description
Setup/Relay 1/Control Phase 40067
Setup/Relay 1/Control Deadband 40068 Float 2 R/W Sets the controller deadband
Setup/Relay 1/Control
Setup/Relay 1/Control On Delay 40071
Setup/Relay 1/Control Off Delay 40072
Setup/Relay 1/Control
Setup/Relay 1/Event Setpoint 40074 Float 2 R/W Sets the event setpoint
Setup/Relay 1/Event Phase 40076
Setup/Relay 1/Event Deadband 40077 Float 2 R/W Sets the event deadband
Setup/Relay 1/Event On Max Time 40079
Setup/Relay 1/Event On Min Time 40080
Setup/Relay 1/Event Off Max Time 40081
Setup/Relay 1/Event Off Min Time 40082
Setup/Relay 1/Timer Hold Type 40083
Setup/Relay 1/Timer Sensor Select 40084
Setup/Relay 1/Timer Hold Mode 40085
Setup/Relay 1/Timer Duration Time 40086
Setup/Relay 1/Timer Period Time 40087
Setup/Relay 1/Timer Off Delay 40088
Setup/Relay 1/Status Level 40089
Setup/Relay 2 Source 40090
Setup/Relay 2 Sensor Select 40091
Setup/Relay 2
Overfeed Timer
Reset Overfeed Timer
Measurement Select
40070
40073
40092
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
1R/W
1 R/W Sets the overfeed timer value (mins)
1 R/W Sets the on delay time (sec)
1 R/W Sets the off delay time (sec)
1 R/W Resets the overfeed timer
1 R/W Sets the event phase (0 = Low; 1 = High)
1 R/W Sets the max on time (mins)
1 R/W Sets the min on time (mins)
1 R/W Sets the max off time (mins)
1 R/W Sets the min off time (mins)
1R/W
1R/W
1R/W
1 R/W Sets the timer on duration time (sec)
1R/W
1R/W
1R/W
1R/W
1R/W
1R/W
Sets the controller phase (0=Low;1=High)
Sets which Sensor outputs are affected during timer on time (0 = None; 2 = Selected Sensor; 13 = All Sensors)
Selects which Sensor outputs are being held/transferred during the timers on time (this is used when Hold type is set for single Sensor)
Selects hold outputs vs. set transfer value during timers on time
Sets the period between timer on events (mins)
Sets the time the affected Sensor outputs are held/transferred after the timer turns off (sec)
Sets the status level which will trigger the relay
Selects data source for this relay (0 = None; 1 = Real Time Clock; 2 = Sensor; 4 = Calculation)
Selects Sensor source when Source = Sensor (0 = Sensor1; 1 = Sensor2)
Selects measurement on the Sensor (0 = Meas1 .. 3 = Meas4)
64
Page 65
ModBUS Register Information
Table 14 Controller ModBUS Registers (continued)
Group Name Tag Name Register # Data Type Length R/W Description
Setup/Relay 2 Type 40093
Setup/Relay 2
Setup/Relay 2/Alarm High Alarm 40095 Float 2 R/W Sets the high alarm setpoint Setup/Relay 2/Alarm Low Alarm 40097 Float 2 R/W Sets the low alarm setpoint
Setup/Relay 2/Alarm
Setup/Relay 2/Alarm
Setup/Relay 2/Alarm On Delay 40103
Setup/Relay 2/Alarm Off Delay 40104
Setup/Relay 2/Control Setpoint 40105 Float 2 R/W Sets the controller setpoint
Setup/Relay 2/Control Phase 40107
Setup/Relay 2/Control Deadband 40108 Float 2 R/W Sets the controller deadband
Setup/Relay 2/Control
Setup/Relay 2/Control On Delay 40111
Setup/Relay 2/Control Off Delay 40112
Setup/Relay 2/Control
Setup/Relay 2/Event Setpoint 40114 Float 2 R/W Sets the event setpoint
Setup/Relay 2/Event Phase 40116
Setup/Relay 2/Event Deadband 40117 Float 2 R/W Sets the event deadband
Setup/Relay 2/Event On Max Time 40119
Setup/Relay 2/Event On Min Time 40120
Setup/Relay 2/Event Off Max Time 40121
Setup/Relay 2/Event Off Min Time 40122
Setup/Relay 2/Timer Hold Type 40123
Setup/Relay 2/Timer Sensor Select 40124
Setup/Relay 2/Timer Hold Mode 40125
Transfer Setting
High Deadband
Low Deadband
Overfeed Timer
Reset Overfeed Timer
40094
40099 Float 2 R/W Sets the high alarm deadband
40101 Float 2 R/W Sets the low alarm deadband
40110
40113
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
1R/W
1R/W
1 R/W Sets the on delay time
1 R/W Sets the off delay time
1R/W
1 R/W Sets the overfeed timer value (mins)
1 R/W Sets the on delay time (sec)
1 R/W Sets the off delay time (sec)
1 R/W Resets the overfeed timer
1 R/W Sets the event phase (0 = Low; 1 = High)
1 R/W Sets the max on time (mins)
1 R/W Sets the min on time (mins)
1 R/W Sets the max off time (mins)
1 R/W Sets the min off time (mins)
1R/W
1R/W
1R/W
Selects the relay type (0 = Alarm; 1 = Control; 2 = Status; 3 = Timer; 4 = Event)
Selects the transfer value for the relays (0 = De-energized; 1 = Energized)
Sets the controller phase (0=Low;1=High)
Sets which Sensor outputs are affected during timer on time (0 = None; 2 = Selected Sensor; 13 = All Sensors)
Selects which Sensor outputs are being held/transferred during the timers on time (this is used when Hold type is set for single Sensor)
Selects hold outputs vs. set transfer value during timers on time
65
Page 66
ModBUS Register Information
Table 14 Controller ModBUS Registers (continued)
Group Name Tag Name Register # Data Type Length R/W Description
Setup/Relay 2/Timer Duration Time 40126
Setup/Relay 2/Timer Period Time 40127
Setup/Relay 2/Timer Off Delay 40128
Setup/Relay 2/Status Level 40129
Setup/Relay 3 Source 40130
Setup/Relay 3 Sensor Select 40131
Setup/Relay 3
Setup/Relay 3 Type 40133
Setup/Relay 3
Setup/Relay 3/Alarm High Alarm 40135 Float 2 R/W Sets the high alarm setpoint Setup/Relay 3/Alarm Low Alarm 40137 Float 2 R/W Sets the low alarm setpoint
Setup/Relay 3/Alarm
Setup/Relay 3/Alarm
Setup/Relay 3/Alarm On Delay 40143
Setup/Relay 3/Alarm Off Delay 40144
Setup/Relay 3/Control Setpoint 40145 Float 2 R/W Sets the controller setpoint
Setup/Relay 3/Control Phase 40147
Setup/Relay 3/Control Deadband 40148 Float 2 R/W Sets the controller deadband
Setup/Relay 3/Control
Setup/Relay 3/Control On Delay 40151
Setup/Relay 3/Control Off Delay 40152
Setup/Relay 3/Control
Setup/Relay 3/Event Setpoint 40154 Float 2 R/W Sets the event setpoint
Setup/Relay 3/Event Phase 40156
Setup/Relay 3/Event Deadband 40157 Float 2 R/W Sets the event deadband
Setup/Relay 3/Event On Max Time 40159
Measurement Select
Transfer Setting
High Deadband
Low Deadband
Overfeed Timer
Reset Overfeed Timer
40132
40134
40139 Float 2 R/W Sets the high alarm deadband
40141 Float 2 R/W Sets the low alarm deadband
40150
40153
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
1 R/W Sets the timer on duration time (sec)
1R/W
1R/W
1R/W
1R/W
1R/W
1R/W
1R/W
1R/W
1 R/W Sets the on delay time
1 R/W Sets the off delay time
1R/W
1 R/W Sets the overfeed timer value (mins)
1 R/W Sets the on delay time (sec)
1 R/W Sets the off delay time (sec)
1 R/W Resets the overfeed timer
1 R/W Sets the event phase (0 = Low; 1 = High)
1 R/W Sets the max on time (mins)
Sets the period between timer on events (mins)
Sets the time the affected Sensor outputs are held/transferred after the timer turns off (sec)
Sets the status level which will trigger the relay
Selects data source for this relay (0 = None; 1 = Real Time Clock; 2 = Sensor; 4 = Calculation)
Selects Sensor source when Source = Sensor (0 = Sensor1; 1 = Sensor2)
Selects measurement on the Sensor (0 = Meas1 .. 3 = Meas4)
Selects the relay type (0 = Alarm; 1 = Control; 2 = Status; 3 = Timer; 4 = Event)
Selects the transfer value for the relays (0 = De-energized; 1 = Energized)
Sets the controller phase (0=Low;1=High)
66
Page 67
ModBUS Register Information
Table 14 Controller ModBUS Registers (continued)
Group Name Tag Name Register # Data Type Length R/W Description
Setup/Relay 3/Event On Min Time 40160
Setup/Relay 3/Event Off Max Time 40161
Setup/Relay 3/Event Off Min Time 40162
Setup/Relay 3/Timer Hold Type 40163
Setup/Relay 3/Timer Sensor Select 40164
Setup/Relay 3/Timer Hold Mode 40165
Setup/Relay 3/Timer Duration Time 40166
Setup/Relay 3/Timer Period Time 40167
Setup/Relay 3/Timer Off Delay 40168
Setup/Relay 3/Status Level 40169
Comm/Net Card Mode 40170
Comm/Net Card Baud 40171
Comm/Net Card Stop Bits 40172
Comm/Net Card Data Order 40173
Min
Comm/Net Card
Comm/Net Card
Comm/Net Card/Addresses
Comm/Net Card/Addresses
Comm/Net Card/Addresses
Comm/Net Card/Stats
Comm/Net Card/Stats
Comm/Net Card/Stats % Good Mesg 40183 Float 2 R/W % of good messages
Response Time
Max Response Time
sc100 40176
Sensor 1 40177
Sensor 2 40178
Good Messages
Bad Messages
40174
40175
40179
40181
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
1 R/W Sets the min on time (mins)
1 R/W Sets the max off time (mins)
1 R/W Sets the min off time (mins)
Sets which Sensor outputs are affected
1R/W
1R/W
1R/W
1 R/W Sets the timer on duration time (sec)
1R/W
1R/W
1R/W
1R/W
1R/W
1 R/W Sets the number of stop bits (1,2)
1R/W
1R/W
1R/W
1 R/W Sets the sc100 Modbus Address
1 R/W Sets the Sensor 1 Modbus Address
1 R/W Sets the Sensor 2 Modbus Address
2 R/W Number of good messages
2 R/W Number of failed messages
during timer on time (0 = None; 2 = Selected Sensor; 13 = All Sensors)
Selects which Sensor outputs are being held/transferred during the timers on time (this is used when Hold type is set for single Sensor)
Selects hold outputs vs. set transfer value during timers on time
Sets the period between timer on events (mins)
Sets the time the affected Sensor outputs are held/transferred after the timer turns off (sec)
Sets the status level which will trigger the relay
Sets the Modbus mode (0=RTU;1=ASCII)
Sets the Modbus baud rate (0 = 9600;
1 = 19200; 2 = 38400; 3 = 57600;
4 = 115200)
Sets the register data order for floats (0 = Normal; 1 = Reversed)
Sets the minimum response time (0 to 30 sec)
Sets the maximum response time (100 to 1000 sec)
67
Page 68
ModBUS Register Information
Table 14 Controller ModBUS Registers (continued)
Group Name Tag Name Register # Data Type Length R/W Description
Comm/Service Port Mode 40185
Comm/Service Port Baud 40186
Comm/Service Port Stop Bits 40187
Comm/Service Port Data Order 40188
Min
Comm/Service Port
Comm/Service Port
Comm/Service Port/Addresses
Comm/Service Port/Addresses
Comm/Service Port/Addresses
Comm/Service Port/Stats
Comm/Service Port/Stats
Comm/Service Port/Stats
Comm/Sensor/ Sensor1 Stats
Comm/Sensor/ Sensor1 Stats
Comm/Sensor/ Sensor1 Stats
Comm/Sensor/ Sensor2 Stats
Comm/Sensor/ Sensor2 Stats
Comm/Sensor/ Sensor2 Stats
Calibration
Calibration
Calibration
Calibration
Response Time
Max Response Time
sc100 40191
Sensor 1 40192
Sensor 2 40193
Good Messages
Bad Messages
% Good Mesg 40198 Float 2 R/W % of good messages
Good Messages
Bad Messages
% Good Mesg 40204 Float 2 R/W % of good messages
Good Messages
Bad Messages
% Good Mesg 40210 Float 2 R/W % of good messages
Output1 4mA count
Output1 20mA count
Output2 4mA count
Output2 20mA count
40189
40190
40194
40196
40200
40202
40206
40208
40212
40213
40214
40215
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned Integer
Unsigned Integer
Unsigned Integer
Unsigned Integer
Unsigned Integer
Unsigned Integer
Unsigned Integer
Unsigned Integer
Unsigned Integer
Unsigned Integer
Unsigned Integer
Unsigned Integer
Unsigned Integer
Unsigned Integer
1R/W
1R/W
1 R/W Sets the number of stop bits (1,2)
1R/W
1R/W
1R/W
1 R/W Sets the sc100 Modbus Address
1 R/W Sets the Sensor 1 Modbus Address
1 R/W Sets the Sensor 2 Modbus Address
2 R/W Number of good messages
2 R/W Number of failed messages
2 R/W Number of good messages
2 R/W Number of failed messages
2 R/W Number of good messages
2 R/W Number of failed messages
1 R/W Calibration counts for the 4mA output 1
1 R/W Calibration counts for the 20mA output 1
1 R/W Calibration counts for the 4mA output 2
1 R/W Calibration counts for the 20mA output 2
Sets the Modbus mode (0 = RTU; 1 = ASCII)
Sets the Modbus baud rate (0 = 9600; 1 = 19200; 2 = 38400; 3 = 57600; 4 = 115200)
Sets the register data order for floats (0 = Normal; 1 = Reversed)
Sets the minimum response time (0 to 30 sec)
Sets the maximum response time (100 to 1000 sec)
68
Page 69
Table 15 Sensor ModBUS Registers
ModBUS Register Information
Group Name Tag Name
Measurements Cal Gain 40013 float 2 R none 0.5 to 2.0 Verification PF Criteria 40062 Integer 1 R/W — Measurements Turbidity 40001 Float 2 R NTU 0/100
Diagnostics Temperature 40005 Float 2 R Deg C
Diagnostics Dark Reading 40009 Float 2 R NTU 0/100 Diagnostics Raw Turbidity 40011 Float 2 R NTU — Diagnostics Lamp Voltage 40018 Float 2 R Volts — Diagnostics Lamp Current 40020 Float 2 R Amps — Diagnostics Plus 5V 40022 Float 2 R Volts — Diagnostics Voltage In 40024 Float 2 R Volts — Setup Software Version 40015 Float 2 R — Setup Bubble Rej 40017 Integer 1 R/W On/Off
Setup DataLog Interval 40026 Integer 1 R/W Sec or Min
Setup Sensor Name 40027 String 6 R/W
Setup Filter Size 40033 Integer 1 R/W sec
Setup Sensor Ser Num 40036 String 6 R/W 12 digits Setup Output Mode 40042 Integer 1 R/W
Register #Data
Typ e
Length R/W Units (U) Range
30 sec, 1 min, 5 min, 10 min, 15 min
no averaging, 6, 30, 60, 90
Setup Set Resolution 40061 Integer 1 R/W decimal places 4, 3, or 2
69
Page 70
Index
Numerics
90 Degree Detector ....................................................... 7
A
Accuracy ........................................................................ 4
Alarms Analog Outputs
......................................................................... 18
............................................................ 18
B
Bubble trap .................................................................... 7
bubble trap
.................................................................... 7
C
Calibration ................................................................... 40
History StablCal
User-prepared Calibration Methods Cleaning Components
Customer Supplied
System Controller Dimensions
................................................................... 46
................................................................ 42
....................................................... 41
...................................................... 5
...................................................................... 48
................................................. 8
.................................................................... 8
................................................... 9
D
Data Logging ............................................................... 36
Date
Setting Digital Output Dimensions
Conduit Hole Display
Contrast
................................................................... 30
............................................................... 21
......................................................... 10
......................................................................... 27
................................................................ 28
E
Electrical Installation .................................................... 14
Conduit
Connecting/Wiring the Sensor Cable
Hard-wiring the Sensor
Power Cord Enclosure, Controller Event Codes Event Logging
.................................................................. 14
.................... 19
......................................... 19
........................................................... 14
..................................................... 6
................................................................ 54
............................................................. 36
F
Flow Rate .................................................................... 25
Fuse
Rating
.................................................................... 51
Replacement
......................................................... 51
H
Head Assembly ........................................................... 24
I
Installation ..................................................................... 8
Controller Panel Panel Mounting Pipe Mounting Wall Mounting
.............................................................. 12
..................................................................... 10
..................................................... 13
....................................................... 12
....................................................... 12
K
Key
Functions
.............................................................. 26
L
Language
Displayed
.............................................................. 29
M
Maintenance
Cleaning Fuse Replacement
Schedule ModBUS RS232 Mounting
................................................................ 48
................................................ 51
............................................................... 47
.......................................................... 21
..................................................................... 24
O
Output
Connections
Setup Outputs
Hold/Transfer
Release
.......................................................................... 5
.......................................................... 19
.............................................................. 32, 37
........................................................ 33
................................................................. 33
P
Photocell Window ........................................................ 48
Power Requirements
..................................................... 4
R
Range ............................................................................ 4
Relay
Setup
.............................................................. 34, 38
Setup Menu Relays
Response Time
......................................................................... 18
Connecting
Specifications
.......................................................... 34
............................................................ 18
.......................................................... 5
............................................................. 4
S
Safety Information ......................................................... 6
Sample
Connections
Flow
.................................................................. 4, 40
.......................................................... 25
70
Page 71
Index
Sample Line Installation Security
Passcode Setup System
Sensor
Setup Starting Sample Flow Sun Shield System
Setup
.............................................................. 31
............................................................... 31, 38
.................................................................. 31
..................................................................... 37
.................................................. 40
................................................................... 11
..................................................................... 37
.............................................. 24
System Power System Startup
............................................................. 40
............................................................ 40
T
Test/Maintenance Menu .............................................. 38
Theory of Operation Time
Setting
Turbidimeter Body
................................................................... 29
...................................................... 7
....................................................... 48
V
Verification .................................................................. 40
W
Warnings
Sensor
Wire preparation
.................................................................. 53
.......................................................... 15
71
Page 72
Visit http: //www.hach.com
Loading...