Section 2 General Information ..................................................................................................................................... 6
2.1 Safety Information ..................................................................................................................................................... 6
2.2 General Product Information ..................................................................................................................................... 6
2.3 Theory of Operation................................................................................................................................................... 7
3.1.2 Using the Optional Sun Shield ....................................................................................................................... 11
3.1.3 Mounting the Controller.................................................................................................................................. 12
3.2.1 Installation in Conduit..................................................................................................................................... 14
3.2.3 Wiring for Power at the Controller.................................................................................................................. 15
3.3 Alarms and Relays .................................................................................................................................................. 18
3.3.1 Connecting the Relays ................................................................................................................................... 18
3.3.2 Connecting the Analog Outputs ..................................................................................................................... 18
3.4 Connecting/Wiring the Sensor Cable ...................................................................................................................... 19
3.5 Connecting the Optional Digital Output ................................................................................................................... 21
3.6 Turbidimeter Installation Information ....................................................................................................................... 24
3.6.1 Mounting the Turbidimeter Body.................................................................................................................... 24
3.6.2 Installing the Head Assembly......................................................................................................................... 24
3.7 Installing a Sample Line .......................................................................................................................................... 24
4.1 Using the Keypad .................................................................................................................................................... 26
4.2 Controller Display Features ..................................................................................................................................... 27
4.2.1 Important Key Presses................................................................................................................................... 27
4.3.3 Specifying the Displayed Language............................................................................................................... 29
4.3.4 Setting the Time and Date ............................................................................................................................. 29
4.4 Changing the Sensor Name .................................................................................................................................... 30
4.4.1 Setting Up System Security ........................................................................................................................... 31
4.6.1 Relay Setup Menu (from System Setup) ....................................................................................................... 34
4.7 Data and Event Logging Options............................................................................................................................. 36
4.7.1 Data Logging Options .................................................................................................................................... 36
4.8 Digital Network Options ........................................................................................................................................... 36
4.9 Menu Structure ........................................................................................................................................................ 36
4.9.1 Sensor Diagnostics Menu .............................................................................................................................. 36
4.9.2 Sensor Setup Menu ....................................................................................................................................... 37
4.9.3 System Setup Menu....................................................................................................................................... 37
4.9.4 Test/Maint Menu ............................................................................................................................................ 38
2
Page 3
Table of Contents
Section 5 System Startup ........................................................................................................................................... 40
5.1 General Operation ................................................................................................................................................... 40
5.5 Calibration and Verification History ......................................................................................................................... 46
6.2 Removing a Sensor from the System...................................................................................................................... 47
6.3 Reinstalling a Sensor on the System....................................................................................................................... 47
Section 8 Replacement Parts and Accessories........................................................................................................ 55
Section 9 How to Order............................................................................................................................................... 57
9.1 Information Required ............................................................................................................................................... 57
9.2 International Customers .......................................................................................................................................... 57
9.3 Technical and Customer Service (U.S.A. only) ....................................................................................................... 57
Section 10 Compliance Information .......................................................................................................................... 60
Appendix A ModBUS Register Information .............................................................................................................. 62
Index ............................................................................................................................................................................. 70
3
Page 4
Section 1Specifications
Specifications are subject to change without notice.
Table 1 Turbidimeter Specifications
Nephelometric light scatter at 90 degrees relative to the incident light beam. The incident light
Method of Detection
Range0.001–100 nephelometric turbidity units (NTU)
beam is composed of a tungsten filament light source with a color temperature between 2200 and
3000K.
Accuracy
Linearity
Resolution (Displayed)0.0001 NTUup to 9.9999 NTU; 0.001 NTU from 10.000 to 99.999 NTU
Repeatability Better than ±1.0% of reading or ±0.002 NTU, whichever is greater for each range.
Response Time
Sample Flow Required 200 to 750 mL/minute (3.2 to 11.9 gal/hour)
Sensor Storage
Temperature
Operating Temperature0 to 50 °C (32–122 °F) for single sensor system, 0 to 40 °C (32–104 °F) for two sensor system
± 2% of reading or ± 0.015 NTU (whichever is greater) from 0 to 40 NTU;
± 5% of reading from 40 to 100 NTU
Better than 1% 0–40 NTU on formazin. Allows for accurate calibration at high turbidity values.
Temperature dependent ±2 °C.
For a full-scale step change, initial response in 1 minute, 15 seconds. Varies with flow rate,
see the table below. The response time is also dependent on the signal averaging time, which is
user selectable.
2.Formazin – user-prepared primary or wet calibration of the instrument
3.Multi-sensor calibration – Performed with a specialized calibration procedure for up to eight
sensors on a single set of fresh StablCal
1.StablCal
Verification (Wet) Method
range of measurement. For regulatory verification, standards of 0.1. 0.3. 0.5 and 1.0 NTU
are available.
2.Formazin – fresh user-prepared standard
1.ICE-PIC™ Verification Module with factory-set values of 20.0 or 1.0 ±25%. Unique value is
Verification (Dry) Method
assigned when dry verification is done immediately after calibration and is used as pass/fail
criteria for subsequent verifications.
1.Mandatory before calibration
Recommended Cleaning
Intervals
2.Optional before verification
3.Mandatory upon verification failure
LanguagesEnglish (default), German, Spanish, Nederlands
Installation EnvironmentIndoor
Primary Compliance Method USEPA 180.1; Hach Method 8195; ASTM D 6698; Standard Methods 2130B
®
(stabilized formazin) – primary or wet calibration of the instrument
®
standards.
®
(stabilized formazin) – recommended for verification in the appropriate application
Specifications
Limit of Detection 0.0032 NTU (according to criteria specified by ISO 15839)
Table 2 Controller Specifications
Component Description
Controller Operating Temperature
Controller Storage Temperature–20 to 70 °C (–4 to 158 °F); 95% relative humidity, non-condensing
EnclosureNEMA 4X/IP66 metal enclosure with a corrosion-resistant finish
Power Requirements
Pollution Degree/Installation CategoryII; II
Outputs
Relays
Microprocessor-controlled measuring unit with measured value display,
temperature display (for some parameters), and menu-driven system
–20 to 60 °C (–4 to 140 °F); 95% relative humidity, non-condensing with
sensor load <7 W; –20 to 40 °C (–4 to 104 °F) with sensor load <25 W
100–230 V ac ±10%, 50/60 Hz;
Power: 11 W with 7 W sensor load, 35 W with 25 W sensor load
Two (Analog outputs, each selectable for 0–20mA or 4–20 mA), maximum
impedance 500 ohm. Output span programmable over any portion of the 0–100
1
NTU range. Optional digital network connection
. Infrared Data Acquisition (IrDA).
Three SPDT, user-configurable contacts rated 100–230 V ac, 5 Amp
resistive maximum.
Controller Dimensions½ DIN—144 x 144 x 150 mm (5.7 x 5.7 x 5.9 inches)
Controller Weight1.6 kg (3.5 lb)
1. See Replacement Parts and Accessories on page 55.
5
Page 6
Section 2General Information
2.1Safety Information
Please read this entire manual before unpacking, setting up, or operating this
equipment. Pay attention to all danger and caution statements. Failure to do so
could result in serious injury to the operator or damage to the equipment.
To ensure that the protection provided by this equipment is not impaired, do not
use or install this equipment in any manner other than that specified in
this manual.
Use of Hazard Information
DANGER: Indicates a potentially or imminently hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION: Indicates a potentially hazardous situation that may result in minor or
moderate injury.
Note: Information that requires special emphasis.
Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or damage to
the instrument could occur if not observed.
This symbol, if noted on the instrument, references the instruction manual for operation
and/or safety information.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or
electrocution exists.
This symbol, if noted on the product, indicates the need for protective eye wear.
This symbol, when noted on the product, identifies the location of the connection for Protective Earth
(ground).
2.2General Product Information
The controller enclosure is NEMA4X/IP66-rated and has a corrosion-resistant
finish designed to withstand corrosive environmental constituents such as salt
spray and hydrogen sulfide. The controller display shows the current turbidity
reading if connected to a single sensor, or two readings when two sensors
are connected.
The 1720E Turbidimeter is a continuous-reading nephelometric turbidimeter
designed for low-range turbidity monitoring. This process turbidimeter is capable
of measuring turbidity from 0.001 to 100.0 NTU. Calibration is based on formazin,
the primary turbidity reference standard adopted by the APHA Standard Methods for the Examination of Water and Wastewater and the U.S. Environmental
Protection Agency (EPA) and on StablCal
standard.
®
which is also recognized as a primary
6
Page 7
2.3Theory of Operation
Figure 190 Degree Detector
General Information
The 1720E Turbidimeter measures turbidity by directing a strong beam of
collimated light from the sensor head assembly down into the sample in the
turbidimeter body. Light scattered at 90° relative to the center line of incident light
by suspended particles in the sample is detected by the submerged photocell
(see Figure 1).
The amount of light scattered is proportional to the turbidity of the sample. If the
turbidity of the sample is negligible, little light will be scattered and detected by the
photocell and the turbidity reading will be low. High turbidity, on the other hand, will
cause a high level of light scattering and result in a high reading.
Sample enters the turbidimeter body and flows through the baffle network of the
bubble trap. The flow allows bubbles to either cling to surfaces of the baffle system
or rise to the surface and vent to atmosphere. After traveling through the bubble
trap, sample enters the center column of the turbidimeter body, rises into the
measuring chamber and spills over the weir into the drain port. A reading is taken
once per second.
7
Page 8
Section 3Installation
DANGER
Only qualified personnel should conduct the installation tasks described in
this section of the manual. The 1720E/sc100 product configuration is not
intended for installation in hazardous locations.
3.Bracket for panel and pipe mounting,
Cat. No. 1000C4F3217-101
4.Gasket for panel mounting, rubber,
Cat. No. 1000A4F3249-101
5.Hex nut, M6 (4), Cat. No. 586730010. Pan head screws (4), M6 x 1.0 x 150 mm, Cat. No. 5867600
3
2
4
7.Flat washer, ¼-inch I.D. (4), Cat. No. 8H1346
8.Pan head screws (4), M6 x 1.0 x 20 mm, Cat. No. 58674-00
9.Pan head screws (4), M6 x 1.0 x 100 mm, Cat. No. 5867500
5
6
7
8
9
10
Table 3 Customer-supplied Items
Item
14-AWG wire for electrical power connections in conduit or if allowed by local electrical codes, 115 or 230 V ac power cord plus a
NEMA 4X-rated strain relief
High-quality, shielded instrumentation cable for connecting the analog outputs plus a NEMA 4X-rated strain relief
Mounting hardware for the sensor
Sun shield for mounting configurations where the sun strikes the front of the display (available from the manufacturer, order
separately). See Figure 8 on page 11.
Common hand tools
3.1Mechanical Installation
Install in an environment that is protected from corrosive fluids. The sensor is
adversely affected by ClO
corrosive liquids or gasses.
. Install the sensor in an area well ventilated from any
2
8
Page 9
3.1.1 Controller Dimension Illustrations
Figure 3Controller Dimensions
144.0 mm (5.67 inches)150.0 mm (5.91 inches)
sc100
Installation
144.0 mm (5.67 inches)
Figure 4Controller Mounting Dimensions
144.02 mm (5.67 inches)
72.01 mm
(2.84 inches)
72.01 mm
(2.84 inches)
144.02 mm
(5.67 inches)
40.14 mm
(1.58 inches)
80.27 mm
(3.16 inches)
40.14 mm (1.58 inches)
M6 x 1.0
80.27 mm
(3.16 inches)
9
Page 10
Installation
Figure 5Panel Mount Cut-out Dimensions
75 mm 2.955 inches
(ref only)
144 mm 5.67 inches
(ref only)
144 mm
5.67 inches
(ref only)
72 mm
2.835 inches
(ref only)
66.67 mm 2.625 inches
Figure 6Conduit Hole Dimensions
75.07 mm 2.955 inches
28.57 mm 1.125 inches
135 mm
5.31 inches
67.4 mm
2.65 inches
133 mm 5.25 inches
144 mm 5.67 inches
68.96 mm 2.715 inches
28.57 mm 1.125 inches
10
150 mm
5.91 inches
99.31 mm
3.91 inches
50.8 mm
2.00 inches
6.35 mm 0.25 inch
15.24 mm 0.60 inch
25.4 mm 1.00 inches25.4 mm 1.00 inches
5.00 inches
(ref only)
127 mm
Page 11
3.1.2 Using the Optional Sun Shield
The optional sun shield was designed to increase the readability of the display
by screening it from direct sunlight. See Replacement Parts and Accessories on
page 55 for ordering information.
Figure 7Sun Shield Kit Components
Installation
12
3
4
6
5
7
8
1.Sun shield5.Pipe mounting brackets (2), includes items 6 and 7,
Cat. No. 9H1079
3.Panel (maximum thickness is 9.5 mm (3/8 inch))9.Flat washer (4), Cat. No. 8H1346
4.Mounting Foot (2), Cat. No. 1000B4F322210. Pan head screw, M6 x 1.0 x 150 mm (4), Cat. No. 5867600
5.Mounting bracket, controller, Cat. No. 1000C4F3217-10111. It may be necessary to remove the sensor connectors. See
6.Pan head screw (4), Cat. No. 5867400
procedure below.
10
To remove the sensor connectors before inserting the controller enclosure into the
panel cut-out:
1. Disconnect power to the controller.
1. Disconnect the wires at terminal block J5, see Figure 21 on page 21.
2. Loosen and remove the nut securing the sensor connector inside the
enclosure. Remove the sensor connector and wires. Repeat step 1 and 2 for
the other sensor connector.
3. After the controller is in place in the panel, reinstall the sensor connectors and
reconnect the wiring to terminal J5 as shown in Figure 21 on page 21.
13
Page 14
Installation
+
DATA
+
OUT 2
+
V
+
OUT 1
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
1
1
+
DATA+DATA
+
OUT 2+OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+V+
V
+
OUT 1+OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANALOG OUTPUTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
J1
J2
J4
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
S1
3.2Electrical Installation
DANGER
The instrument must be installed
by qualified technical personnel
for adherence to all applicable
electrical codes. The 1720E/sc100
High-voltage wiring for the controller is conducted behind the high voltage
barrier in the controller enclosure. The barrier must remain in place unless a
qualified installation technician is installing wiring for power, alarms, or relays.
See Figure 12 for barrier removal information.
product configuration is not
intended for installation in
hazardous locations.
3.2.1 Installation in Conduit
In hard-wired electrical applications, the power and safety ground service drops
for the instrument must be 18 to 12 AWG. See Figure 13 on page 15 for strain
relief and conduit opening sealing plug information. See section 3.2.3 on page 15
for wiring information.
3.2.2 Installation Using a Power Cord
DANGER
Use of power cords is not
permitted in hazardous locations.
Where permitted by local electrical codes, a sealing-type strain relief to
maintain the NEMA 4X/IP66 environmental rating and a power cord less than
3 meters (10 feet) in length with three 18-gauge conductors (including a safety
ground wire) can be used, see Replacement Parts and Accessories on page 55.
See Figure 13 on page 15 for strain relief and conduit opening sealing plug
assembly. See section 3.2.3 on page 15 for wiring information.
Figure 12Removing Voltage Barrier
J1
S1
J4
NETWORK
INTERFACE
1.High voltage barrier2.Unsnap the barrier latch then pull out to remove the barrier.
J2
U5
U9
CARD
J3
1
+
DATA
– DATA
2
SERVICE REQUEST
3
4
+
PROBES
J5
PCB
CONNECTOR
J6
ANALOG OUTPUTS
V
5
GND
6
1
+
OUT 2
– OUT 2
2
SHIELD/CHASSIS GND
3
4
+
OUT 1
5
– OUT 1
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
1
2
14
Page 15
Figure 13Using the Optional Strain Relief and Conduit Plug
1.Strip ¼-inch of insulation.2.Seat insulation against connector with no bare wire exposed.
3.2.3 Wiring for Power at the Controller
Wire the instrument for line power by hard-wiring in conduit or by wiring to a power
cord if local code allows. Regardless of the type of wire used, the connections are
made at the same terminal. A local disconnect designed to meet local electrical
code is required and must be identified for all types of installation. See Figure 16
and Figure 17 on page 17 for suggested local disconnect configurations.
1. Obtain appropriate fittings with NEMA 4X/IP66 environmental rating.
2. Loosen the screws using a phillips-head screwdriver and open the hinged
controller cover.
3. Remove the high-voltage barrier (see Figure 12 on page 14).
4. Insert the wires through the strain relief fitting or conduit hub located in the
right-rear access hole in the bottom of the enclosure. Tighten the strain relief if
used, to secure the cord.
5. Properly prepare each wire (Figure 14) and insert each wire into the terminal
according to Table 4. Tug gently after each insertion to ensure the connection
is secure.
6. Seal any unused openings in the controller box with conduit opening sealing
plugs, see Replacement Parts and Accessories on page 55.
7. Reinstall the high-voltage barrier and latch to secure.
15
Page 16
Installation
Table 4 Power Wiring Information
Terminal NumberTerminal DescriptionWire Color Code for North AmericaWire Color Code for Europe
2.J2—Header for optional network interface card9.Sensor connector
3.J5—Relay A connector10. J6—Analog output (4–20 mA) connector
4.J6—Relay B connector11. J5—Sensor connector for hard-wiring
5.J7—Relay C connector12. Position for network interface card
6.Fuses (F1, F2)13. Service port
7.J8—ac Power connections14. Sensor terminator selector/service port configuration
6
7
16
Page 17
Figure 16Local Disconnect for Power Cord
RELAYC
J1
S1
J2
J3
Installation
1
CARD
U5
U9
J5
RELAYBRELAYA
J6
F1
NCNCNC
NONO
NO
COMCOMCOM
RELAYC
J4
NETWORK
INTERFACE
1.Power terminal2.Power cord strain relief
Figure 17Local Disconnect for Hard-wired Line Power
J1
S1
J2
J3
F2
2
1
J4
NETWORK
INTERFACE
U5
U9
CARD
1.Power terminal2.Conduit strain relief
F1
J5
F2
2
NCNCNC
NONO
NO
COMCOMCOM
RELAYCRELAYBRELAYA
J6
17
Page 18
Installation
NCNCNC
COMCOMCOM
NO
F1
F2
NONO
RELAYCRELAYBRELAYA
J1
J2
J4
S1
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
J5J5
3.3Alarms and Relays
The controller is equipped with three unpowered relays rated 100–230 V ac,
50/60 Hz, 5 amp resistive maximum.
3.3.1 Connecting the Relays
The relay connector accepts 18–12 AWG wire (as determined by load
application). Wire gauge less than 18 AWG is not recommended.
Danger: Relay loads must be
resistive. User must externally
limit current to the relays to
5 Amps by use of a fuse or
breaker.
Danger: ac power terminals are
designed for single wires. Do not
use more than one wire in each
terminal.
The controller contains three relays designed for use with high voltage
(greater than 30V-RMS and 42.2V-PEAK or 60 V dc). Refer to Figure 18 for
connection information. The wiring is not designed for low voltage connections.
Relay must not be powered from the same wiring used to power the controller.
See section 4.6 on page 34 for relay setup details.
The Normally Open (NO) and Common (COM) relay contacts will be connected
when an alarm or other condition is active. The Normally Closed (NC) and
Common relay contacts will be connected when an alarm or other condition
is inactive or when power is removed from the controller.
Figure 18Alarm and Relay Connections
J1
S1
J4
NETWORK
INTERFACE
J2
U5
U9
CARD
J3
J5J5
J5
J5J6J7
3
2
1
RELAY ARELAY BRELAY C
F1
F2
NCNCNC
NONO
NO
COMCOMCOM
RELAYCRELAYBRELAYA
3.3.2Connecting the Analog Outputs
18
J6
Disconnect
Power
Two isolated analog outputs (1 and 2) are provided, see Figure 19. Each output
can be set to 0–20 or 4–20 mA, and can be assigned to represent the measured
parameter or secondary measurement such as temperature. Make connections
with twisted-pair shielded wire and connect the shield at the controlled component
end or at the control loop end. Do not connect the shield at both ends of the cable.
Use of non-shielded cable may result in radio frequency emission or susceptibility
levels higher than allowed. Maximum loop resistance is 500 ohm. Refer to
section 4.5 on page 32 for output software setup.
Page 19
Make wiring connections at the analyzer end as shown in Figure 19.
RELAYC
RELAYA
+
DATA
+
OUT 2
+
V
+
OUT 1
ANALOG OUTPUTS
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
NCNCNC
COMCOMCOM
NO
F1
F2
NONO
RELAYCRELAYCRELAYBRELAYARELAYA
J1
J2
J4
S1
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
1
1
+
DATA+DATA
+
OUT 2+OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+V+
V
+
OUT 1+OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANALOG OUTPUTSANALOG OUTPUTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
J5J5
Table 5 Output Connections at Terminal Block J6
Recorder WiresCircuit Board Position
Output 2 +1
Output 2 –2
Shield3
Output 1 +4
Output 1 –5
Figure 19Analog Output Connections
J1
S1
J2
J3
Installation
CARD
U5
J5J5
U9
J5
J4
NETWORKINTERFACE
J6
1
2
3
4
5
J6
3.4Connecting/Wiring the Sensor Cable
The sensor cable is supplied with a keyed quick-connect fitting for easy
attachment to the controller, see Figure 20. Retain the connector cap to seal the
connector opening in case the sensor must be removed.
The 1720E sensor cable may be extended by a maximum of 7.62 m (25 ft),
see Replacement Parts and Accessories on page 55.
Modify the controller for sensor hard-wiring as follows:
1. Remove all power to the controller.
2. Open the controller cover.
PCB
CONNECTOR
PROBES
ANALOG OUTPUTS
RELAYBRELAYA
123456
12345
NCNCNC
NONO
COMCOMCOM
RELAYC
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
+
DATA
– DATA
SERVICE REQUEST
+
V
GND
F1
+
OUT 2
– OUT 2
SHIELD/CHASSIS GND
+
OUT 1
– OUT 1
NO
F2
19
Page 20
Installation
RELAYA
+
DATA
+
OUT 2
+
V
+
OUT 1
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
3. Disconnect and remove the existing wires between the quick connect and
terminal block J5, see Figure 21 on page 21.
4. Remove the quick connect fitting and wires and install the threaded plug on
the opening to maintain the environmental rating.
Table 6 Wiring the Sensor at Terminal Block J5
Terminal NumberTerminal DesignationWire Color
1Data (+)Blue
2Data (–)White
3Service RequestNo Connection
4+12 V dcBrown
5Circuit CommonBlack
6ShieldShield (grey wire in existing quick disconnect fitting)
Figure 20Attaching the Sensor using the Quick-connect Fitting
S1
J4
From Probe
NETWORK
INTERFACE
CARD
J1
J2
U5
U9
J3
1
+
DATA
– DATA
2
SERVICE REQUEST
3
4
+
5
6
1
2
3
4
5
NCNCNC
NONO
COMCOMCOM
RELAYCRELAYBRELAYA
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
V
GND
F1
+
OUT 2
– OUT 2
SHIELD/CHASSIS GND
+
OUT 1
– OUT 1
NO
F2
PROBES
J5J5
J5
PCB
CONNECTOR
J6
5. Cut the connector from the sensor cable.
6. Reinstall the plug on the sensor access opening to maintain the
environmental rating.
7. Strip the insulation on the cable back 1-inch. Strip ¼-inch of each individual
ANALOG OUTPUTS
wire end.
20
8. Wire as shown in Table 6 .
Page 21
9. Pass the cable through conduit and a conduit hub or a strain relief fitting and
+
DATA
+
OUT 2
+
V
+
OUT 1
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
NCNCNC
COMCOMCOM
NO
F1
F2
NONO
RELAY3RELAY2RELAY1
1
1
+
DATA+DATA
+
OUT 2+OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+V+
V
+
OUT 1+OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANALOG OUTPUTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BERATED TO
80° CMINIMUM
FIELD WIRING
INSULATION MUST
BERATED TO
80° CMINIMUM
J1
J2
J4
NETWORK
INTERFACE
CARD
J3
J6
U5
U9
S1
J5
10. Close and secure the cover.
Figure 21Hard-wiring the Sensor
J1
S1
J2
U5
J4
U9
NETWORKINTERFACE
CARD
J3
J5J5
J5
J6
Installation
an available access hole in the controller enclosure. Tighten the fitting.
1
+
DATA
– DATA
2
SERVICE REQUEST
34
+
56
12345
NCNCNC
NONO
COMCOMCOMRELAY3RELAY2RELAY1
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
V
GND
F1
+
OUT 2– OUT 2SHIELD/CHASSIS GND
+
OUT 1– OUT 1
NO
F2
PCB
CONNECTOR
PROBES
ANALOG OUTPUTS
3.5Connecting the Optional Digital Output
At this time, the manufacturer supports ModBUS RS485 and ModBUS RS232
communication protocols. The optional digital output card is installed in the
location indicated in Figure 22 on page 22. Terminal block J1 provides user
connection to the optional network card. The terminal connection is based on the
selected network card. Refer to the instructions supplied with the network card for
more details.
Allow at least 254 mm (10.00 inches) above the instrument
for removal of the head assembly.
Page 24
Installation
3.6Turbidimeter Installation Information
The turbidimeter body is designed for wall-mounting (although it may be mounted
on the optional floor stand). The turbidimeter sensor must be mounted within
six feet of the controller unless an extension cable is used. Maximum cable length
is 9.6 m (31 feet).
3.6.1 Mounting the Turbidimeter Body
Locate the turbidimeter as close to the sampling point as possible. A shorter
distance for the sample to travel results in a faster response time.
Clean the turbidimeter body and bubble trap before installation using the
instructions supplied in section 6.4.3, Cleaning the Turbidimeter Body and Bubble
Trap on page 48. Slotted mounting brackets are integral parts of the turbidimeter
body. Install customer-supplied hardware appropriate for the installation
environment using the criteria detailed below:
•Install in a location that is isolated from vibration.
•Allow at least 22 cm (approximately 10 inches) clearance for removal of the
head assembly and bubble trap cover from the top of the turbidimeter body.
•Leave enough room below the turbidimeter body to remove the bottom plug
and to place a container under the drain when calibrating or cleaning.
Note: Make sure the top of the
turbidimeter body is level.
•Install two ¼-20 bolts 10-3/4 inches apart (on center). Leave at least ¼-inch of
the bolt head exposed.
•Make sure the bolts are installed level.
Slide the slotted mounting brackets of the turbidimeter body onto the bolts.
3.6.2 Installing the Head Assembly
After the turbidimeter body has been mounted, install the bubble trap cover, then
place the head assembly on the turbidimeter body with the label facing the front.
Move the head assembly back and forth slightly to ensure it is properly seated on
the body of the instrument. Failure to properly seat the head will result in light
leakage and erroneous readings.
The rear portion of the head assembly has a molded “lip” which may be used to
hang the head assembly on the turbidimeter body edge for routine maintenance.
3.7Installing a Sample Line
DANGER
This turbidimeter is not designed
for use in hazardous locations or
with samples that are flammable
or explosive in nature. If any
sample solution other than water
is used in this product, test the
sample/product compatibility to
ensure user safety and proper
product performance.
One-fourth inch O.D. rigid or semi-rigid tubing is recommended for sample lines.
Run them as directly as possible between the turbidimeter body and the sampling
point to minimize sample flow lag time.
Install sample line taps into larger process pipes to minimize interference from
air bubbles or pipeline bottom sediment. A tap projecting into the center of the
pipe is ideal. Figure 24 shows examples of sample tap installations.
Note: When setting the flow rate, take care to avoid sweeping air “micro-bubbles” through
the internal bubble trap. Observe the sample flow inside the turbidimeter body. If small
air bubbles can be seen flowing up through the center, reduce the flow rate.
24
Page 25
Figure 24Sampling Techniques
Installation
Air (Typical)
Sediment (Typical)
Poor
3.8Sample Connections
Sample inlet and drain connections are made on the turbidimeter body. The
sample inlet fitting installed in the body is a ¼-inch NPT
fitting. One additional fitting supplied with the instrument is a ½-inch NPT-to-hose
fitting for use with ½-inch ID flexible plastic tubing on the drain.
Note: For samples with high solids
content (high turbidity), operate at
the highest flow rate possible. For
samples with low solids content (low
expected turbidity), operate at a low
flow rate (200–300 mL/min).
Figure 25Sample Connections
The required flow rate is 200 to 750 mL/minute (4.0 to 11.9 gal/hour). Flow rate
into the turbidimeter may be controlled with a flow restriction device on the inlet
line. Flow rates below 200 mL/min will reduce response time and cause
inaccurate readings. Flow rates above 750 mL/min will cause the turbidimeter to
overflow, indicating the flow rate is too high.
The front of the controller is shown in Figure 26. The keypad consists of the eight
keys described in Table 8 .
Figure 26Front of the Controller
1
sc100
5
2
3
4
1.Instrument Display4.Right, Left, Up, and Down keys 7.Enter key
2.Back key5.IrDA (Infrared Data Acquisition) window
3.Menu key6.Home key
Table 8 Controller Key Functions/Features
Number KeyFunction
2Move back one level in the menu structure.
3
Move to the main menu from other menus. This key is not active in menus where a selection or other
input must be made.
6
7
4Navigate through the menus, change settings, and increment and decrement digits.
5
6Accept an input value, updates, or accepts displayed menu options.
Move to the Main Measurement screen from any other screen. This key is not active in menus where a
selection or other input must be made.
26
Page 27
4.2Controller Display Features
When the controller is in measurement mode, measurements for each connected
sensor are displayed.
The display will flash on startup, when the hold outputs function has been
activated, and when the filter function (signal average) is changed to a different
value.
An active system warning will cause the warning icon (a triangle with an
exclamation point inside) to be displayed on the right side of the display.
Figure 27Display
Operation
1
2
SENSOR NAME
0.023
NTU
3
1.Status bar. Indicates the sensor name and status of relays. The relay
letter is displayed when the relay is energized.
2.Main measurement5.Warning icon area
3.Secondary measurement/output information6.Measurement units
4.2.1 Important Key Presses
OUTPUT1: 11.23 mA
4.Energized relay indicator
•Press the HOME key then the RIGHT or LEFT key to display two readings
when two sensors are connected. Continue to press the RIGHT or LEFT key
to toggle through the available display options as shown below.
4
5
6
RTC:MM/DD/YY
24:00:00
OUTPUT1: 11.23 mA
SENSOR NAME:
02
0.023
ppm
OUTPUT1: 11.23 mA
MAIN MENU
SENSOR DIAG
SENSOR SETUP
SYSTEM SETUP
TEST/MAINT
NTU
SENSOR NAME:
0.031
OUTPUT1: 11.23 mA
MAIN MEASURE
SENSOR NAME:
NTU
SENSOR NAME:
0.023 NTU
0.031 NTU
SENSOR NAME
Turbidity:
0.023 NTU
TEMP:
22.9°C
SENSOR NAME
Turbidity:
0.031 NTU
TEMP:
22.9°C
•Press the UP and DOWN keys to toggle the status bar at the bottom of the
measurement display to display the output information.
•In Menu mode, an arrow may appear on the right side of the display to
indicate that more items are available. Press the UP or DOWN key
(corresponding to the arrow direction) to display additional menus.
SYSTEM SETUP
OUTPUT SETUP
RELAY SETUP
NETWORK SETUP
DISPLAY SETUP
SYSTEM SETUP
DISPLAY SETUP
SECURITY SETUP
LOG SETUP
CALCULATION
SYSTEM SETUP
SECURITY SETUP
LOG SETUP
CALCULATION
ERROR HOLD MODE
27
Page 28
Operation
4.3Instrument Setup
4.3.1 Software Text Abbreviation Conventions
AbbreviationMeaningAbbreviationMeaning
AdjAdjustP/FPass/Fail
CalCalibrationPassPassword
Cont.ContinuePrepedPrepared
CylCylinderSNSerial Number
DfltDefaultStdStandard
DiagDiagnosticTempTemperature
IntInternalVerVerification
Meas.MeasurementXferTransfer
4.3.2 Adjusting Display Contrast
StepSelectMenu Level/InstructionsConfirm
1
2
3
4
—ADJ CONTRAST
5
6
MAIN MENU or Main Measurement Screen—
MAIN MENU—
SYSTEM SETUP
DISPLAY SETUP
(+0–50)
28
Page 29
Operation
4.3.3 Specifying the Displayed Language
StepSelectMenu Level/InstructionsConfirm
1
2
3
4
5
6
Main Menu or Main Measurement Screen—
MAIN MENU—
SYSTEM SETUP
DISPLAY SETUP
LANGUAGE
select language
4.3.4 Setting the Time and Date
4.3.4.1 Setting the Time
Note: The time is available only in 24-hour (military) format.
StepSelectMenu Level/InstructionsConfirm
1
2
3
4
5
MAIN MENU—
SYSTEM SETUP
DISPLAY SETUP
SET DATE/TIME
highlight TIME
select character to edit
6
choose appropriate number
7
Main Menu or Main Measurement Screen—
29
Page 30
Operation
4.3.4.2 Setting the Date Format and Date
StepSelectMenu Level/InstructionsConfirm
1
2
3
4
5
6
—highlight DATE FORMAT
choose appropriate date format
7
MAIN MENU—
SYSTEM SETUP
DISPLAY SETUP
SET DATE/TIME
Highlight DATE
select character to edit—
8
choose appropriate number
9
Main Menu or Main Measurement Screen—
4.4Changing the Sensor Name
StepSelectMenu Level/InstructionsConfirm
1
2
3
highlight sensor of interest if more than one sensor is attached
4
5
6
choose appropriate alpha/numeric digit
7
Main Menu or Main Measurement Screen—
MAIN MENU—
SENSOR SETUP
CONFIGURE
EDIT NAME
select character to edit—
30
Page 31
Operation
4.4.1 Setting Up System Security
The sc100 has a passcode feature to restrict unauthorized access to Network
Setup, Security Setup, Log Setup, and Test/Maint menus. In addition, the
passcode also regulates function selection for relay options. The passcode is
factory set to sc100 (the five digits must be followed by a space to remove the
trailing asterisk). The passcode may be changed, see section 4.4.1.1.
The following two options are available:
Disabled: All configuration settings and calibrations can be changed. This is the
default setting.
Enabled: All configuration settings can be displayed but not changed. Network
Setup, Security Setup, Log Setup, and Test/Maint menus cannot be accessed
without the passcode.
StepSelectMenu Level/InstructionsConfirm
1
2
3
4
—SET PASSCODE
5
6
Main Menu or Main Measurement Screen—
MAIN MENU—
SYSTEM SETUP
SECURITY SETUP
highlight ENABLED
4.4.1.1 Editing the Passcode
If the passcode is enabled, it may be edited. The passcode can consist of up to six
digits (alpha and/or numeric and available characters). If a passcode is forgotten,
obtain the Master passcode from the Technical Consulting Services Department,
see Technical and Customer Service (U.S.A. only) on page 57.
StepSelectMenu Level/InstructionsConfirm
1
2
3
4
MAIN MENU—
SYSTEM SETUP
highlight SECURITY SETUP
ENTER (current or default) PASSCODE—
EDIT PASSCODE
31
Page 32
Operation
StepSelectMenu Level/InstructionsConfirm
5
6
Change the existing passcode
Main Menu or Main Measurement Screen—
4.5Output Options
The controller provides two isolated analog outputs (Output 1 and Output 2).
Customize the outputs using the table in section 4.5.1 on page 32.
4.5.1 Output Setup Menu (from System Setup)
1.Select OUTPUT 1 or 2
SELECT SOURCE
Press ENTER to access a list of all connected sensors. Choose the sensor to associate with the output.
SET PARAMETER
Highlight the appropriate displayed parameter and press ENTER
SET FUNCTION
Select LINEAR CONTROL for current output to track the measurement value. Select PID CONTROL for the sc100 to operate as
a PID controller.
SET TRANSFER
Each analog output is normally active, responding to the measured value of its assigned parameter. However, during calibration,
each output can be transferred to this preset transfer value. Default: 4mA; Range: 0–20
.
SET FILTER
Allows the user to average the analog outputs over time Default: 0; Range: 0–120 seconds
SCALE 0 mA/4 mA
Select 0 mA or 4 mA for minimum current (outputs will be set to 0–20 mA or 4–20 mA).
ACTIVATION
FUNCTION set to LINEAR CONTROL
If LINEAR CONTROL was selected in SET FUNCTION, set the low and the high values for the current output here.
Defaults: Low = 0; High = 100; Low Value Range: 0–100, High Value Range: 0–100.
FUNCTION set to PID CONTROL
If PID CONTROL was selected in SET FUNCTION, configure the PID Control as follows:
1.Set MODE: AUTO or MANUAL. Manual output default: 100%
2.Set PHASE: DIRECT or REVERSE controller operation.
3.SET SETPOINT: enter the set point the PID control will control the process to. Default: 100; Range: 0–100
4.PROP BAND: control the proportional band for the PID control. Default: 5.00; Range: 0–1000
5.INTEGRAL: control the integral action time period in minutes. Range: 0–999
6.DERIVATIVE: control the settings for the rate control. Range: 0–999
32
Page 33
Operation
4.5.2 Hold/Transfer Outputs
When cleaning or servicing the instrument, the analog outputs can be held at the
last measured values. To hold the output until released:
StepSelectMenu Level/InstructionsConfirm
1
2
3
4
5
—SET OUTMODE
Select HOLD OUTPUTS or XFER OUTPUTS
6
7
8
MAIN MENU—
TEST/MAINT
Enter Passcode if enabled
HOLD OUTPUTS
SET CHANNELS
Select ALL or 1720E
ACTIVATION
9
10
—LAUNCH
Main Menu or Main Measurement Screen
reading will
flash
During calibration, the analog outputs can remain active, be held, or be
transferred to a preset mA value. When output hold or transfer is enabled during a
calibration, the hold or transfer is automatically released when calibration is
completed. See section 5.3, Sensor Calibration and Verification on page 40.
4.5.3 Release Outputs
StepSelectMenu LevelConfirm
1
2
3
4
MAIN MENU—
TEST/MAINT
HOLD OUTPUTS
ACTIVATION
33
Page 34
Operation
StepSelectMenu LevelConfirm
5
6
—RELEASE
Main Menu or Main Measurement Screen—
4.6Relay Options
StepSelectMenu LevelConfirm
1
2
3
4
—Customize the options using the information in section 4.6.1—
4.6.1 Relay Setup Menu (from System Setup)
1.Select Relay A, B, or C
MAIN MENU—
SYSTEM SETUP
RELAY SETUP
SELECT SOURCE
Choose from the available options (none, installed sensors, real time clock (RTC)).
SET PARAMETER
Choose from the available options.
SET FUNCTION
Source set to sensor
Alarm: Operates relays in response to the measured parameter. Contains separate High and Low Alarm points, deadbands, and
ON/OFF delay. Defaults: Low = 0.000 NTU, high = 100.00 NTU, low deadband = 5.000 NTU, high deadband = 5.000 NTU, on/off
delays default to zero seconds; Range: 0–999 sec.
Feeder Control: Operates in response to the measured parameter. Can be set for phasing, set point, deadband, overfeed timer,
and ON/OFF delay.
Event Control: Controls a cleaning system (or equivalent) on a timed basis.
Warning: Activated when the analyzer detects a sensor warning.
Source set to RTC
Timer: Sets the timer for a cleaning system (or equivalent). Controls the output hold, interval, duration and off delay.
SET TRANSFER
Normally, each control or alarm relay is active, responding to the measured value of its assigned parameter. During calibration,
however, the relay can be transferred to a preset on/off state to suit the application requirements. Select Energize or De-energize
and press ENTER.
34
Page 35
Operation
4.6.1 Relay Setup Menu (from System Setup) (continued)
ACTIVATION
Function set to ALARM
Sets the value where the relay will turn on in response to decreasing measured value. For example: if the
LOW ALARM
HIGH ALARM
LOW DEADBAND
HIGH DEADBAND
OFF DELAYSets a time to delay the relay from normally turning off. Off Delay Range: 0–999 seconds
ON DELAY Sets a time to delay the relay from normally turning on. On Delay Range: 0–999 seconds
Function set to FEEDER CONTROL
PHASE
SET SETPOINT Sets the value where the relay will turn on. Default: 100 NTU; Range: 0–100
DEADBAND
OVERFEED
TIMER
OFF DELAYSets a time to delay the relay from normally turning off. Default: 0 seconds; Range: 0–999 seconds
ON DELAYSets a time to delay the relay from normally turning on. Default: 0 seconds; Range: 0–999 seconds
Function set to EVENT CONTROL
PHASE
SET SETPOINT Sets the value where the relay will turn on. Default: 100; Range: 0–100
DEADBAND
OnMax TIMERSets the time to limit the time the relay can remain “on.” Default: 0 seconds; Range: 0–999 seconds
OffMax TIMERSets a time to delay the relay from normally turning off. Default: 0 seconds; Range: 0–999 seconds
OnMin TIMERSets the time to limit the time the relay can remain “on.” Default: 0 seconds; Range: 0–999 seconds
OffMin TIMERSets the time to limit the time the relay can remain “off.” Default: 0 seconds; Range: 0–999 seconds
Function set to TIMER CONTROL (RTC selected in SELECT SOURCE)
HOLD OUTPUTS Set OUTMODE to select output hold operation and select the channels that cause the outputs to be held.
INTERVALSet the off time for the relay. Default: 5 minutes; Range: 0–999 minutes
DURATIONSet the on time for the relay. Default: 30 seconds; Range: 0–999 seconds
OFF DELAY
Function set to WARNING CONTROL
WARNING LEVEL
low alarm is set for 1.0 and the measured value drops to 0.9, the relay will be activated.
Range: 0.00–100 NTU
Sets the value where the relay will turn on in response to increasing measured value. For example: if the
high alarm is set for 4.0 and the measured value increases to 4.2, the relay will be activated.
Range: 0.00–100 NTU
Sets the range where the relay remains on after the measured value increases above the low alarm value.
Default is 20% of the range. For example: if the low alarm is set for 1.0 and the low deadband is set for 0.5,
then the relay remains on between 1.5 and 1.0. Range: 0.00–100 NTU
Sets the range where the relay remains on after the measured value decreases below the high alarm value.
For example: if the high alarm is set for 4.0 and the high deadband is set for 0.5, then the relay remains on
between 3.5 and 4.0. Range: 0.00–100 NTU
“High” phase assigns the relay setpoint to respond to an increasing measured value; conversely, a “Low”
phase assigns the relay setpoint to respond to a decreasing measured value.
Sets the range where the relay remains on after the measured value decreases below the setpoint value
(high phase relay) or increases above the setpoint value (low phase relay). Default: 5 NTU;
Range: 0–100 NTU
Sets the time to limit how long the relay can remain “on.” Default: 5 minutes; Range: 0–999 minutes
“High” phase assigns the relay setpoint to respond to increasing measured value; conversely, a “Low” phase
assigns the relay setpoint to respond to decreasing measured value.
Sets the range where the relay remains on after the measured value decreases below the setpoint value
(high phase relay) or increases above the setpoint value (low phase relay). Default: 5 NTU;
Range: 0–100 NTU,
Set the time for additional hold/output time after the relay has been turned off. Default: 1 second;
Range: 0–999 seconds
Set the warning level that will trigger a relay. Range: 0–32 (warnings assigned to 1-9 for 1720E). For
example: Set the warning level to 0 to allow all warnings to trigger the relay; set the warning level to 5 to
allow warnings 6 and above to trigger the relay. Set the warning level to 9 or greater to not trigger the relay
on any warning. See Table 10 on page 52 for a full list of warnings.
35
Page 36
Operation
4.7Data and Event Logging Options
The sc100 provides two data logs (one for each sensor) and two event logs (one
for each sensor). The data logs store the measurement data at selected intervals.
The event log stores a variety of events that occur on the devices such as
configuration changes, alarms, and warning conditions. The data logs are stored
in a packed binary format and the event logs are stored in a CSV format. The logs
can be downloaded through either the digital network port or the IrDA port using
the file transfer program available from the manufacturer.
The default datalogging frequency is 15 minutes. If the datalogging frequency is
set to 15 minute intervals, the instrument can continue to store data for
approximately six months.
4.7.1 Data Logging Options
Sensor Data Log:
StepSelectMenu Level/InstructionsConfirm
1
2
3
4
5
6
Highlight sensor of interest if more that one sensor is attached.
4.8Digital Network Options
The sc100 provides two digital communication methods with the controller
(the digital network port and the IrDA port). Either of the digital ports can be used
to access setup data, measurement data, or data/event logs. For the features
available for each individual digital network port, refer to the instruction sheet
supplied with the selected network card.
ERROR LISTDisplays a list of errors that are present. See section 7.1 on page 52.
WARNING LISTDisplays a list of warnings that are present. See section 7.2 on page 52.
36
Page 37
4.9.2 Sensor Setup Menu
CALIBRATE
SELECT SENSOR (if more than one sensor is attached)
USER PREPD CAL Calibration using 4000 NTU stock solution diluted to 20.00 NTU formazin.
STABLCAL CALCalibration using 20 NTU StablCal Stabilized Formazin Standard
VERIFICATIONPerform a verification, set the pass/fail criteria, and view the verification history.
0 ELECTRONICSZero electronics
SET DFLT GAINReturn instrument to default calibration.
CAL HISTORY
CONFIGURE
BUBBLE REJECTChoose Yes or No to enable/disable bubble reject. Default: Yes
SIGNAL AVG
MEAS UNITS
EDIT NAME
SET RESOLUTION Set the number of significant digits to display. Default is three significant digits.
DATALOG INTRVL
DIAG/TEST
View the last 12 entered calibrations. Press the ENTER key to move to the next history entry.
See section 5.5 on page 46 for more information.
Choose no averaging or specify the amount of time for signal averaging. Available options are: no
averaging, 6 sec., 30 sec., 60 sec., or 90 sec. Default is 30 seconds.
Select the appropriate measurement units to display. Choose from mg/L, NTU, TE/F, and FTU.
Default: NTU
Enter up to a 12-digit name in any combination of symbols and alpha or numeric characters. Press
ENTER when the entry is complete. The name will be displayed on the status line above the
measurement value on the main display.
Choose the amount of time between saving data points to the data log. Default: 15 min.; Options: 30
sec., 1 minute, 5 minutes, 10 minutes, or 15 minutes.
Operation
SOFTWARE VERS. Displays the software version number.
DRIVER VERSDisplays the software driver version number.
SERIAL NUMBERDisplays the serial number of the sensor.
INT TEMPDisplays the internal temperature of the sensor electronics in °C.
DEFAULT SETUPRestores the sensor’s factory default settings and invalidates the current calibration.
POWER CHECKDisplays the electrical statistics for the sensor.
CAL VALUEDisplays the gain and zero electronics values for the current calibration.
4.9.3 System Setup Menu
OUTPUT SETUP (see section 4.5 on page 32 for expanded menu information)
SELECT OUTPUT 1 or 2
SELECT SOURCEPress ENTER to access a list of all connected sensors and select the sensor that will drive the output.
SET PARAMETERPress ENTER to select from the displayed parameters.
SET FUNCTION
SET TRANSFER
SET FILTERAverage measurements over time (0–120 seconds). Default: 0 seconds.
SCALE 0 mA/4 mASelect 0 mA or 4 mA for minimum current (outputs will be set to 0–20 mA or 4–20 mA).
ACTIVATIONDependent on Function selected previously. See section 4.5 on page 32 for additional information.
Select LINEAR CONTROL for current output to track the measurement valve. Select PID CONTROL for
the sc100 to operate as a PID controller.
Each analog output is normally active, responding to the measured value of its assigned parameter.
However, during calibration, each output can be transferred to this preset transfer value.
37
Page 38
Operation
4.9.3 System Setup Menu (continued)
RELAY SETUP (See section 4.6 on page 34 for expanded menu information.)
SELECT RELAY A, B, or C
SELECT SOURCESelect from none, any connected sensor, or the real time clock (RTC).
SET PARAMETERPress ENTER to select from the displayed parameters.
SET FUNCTION
SET TRANSFERSets the relay to Energize or De-energize (user-selectable).
ACTIVATIONActivate the relays from this menu (dependent on Function selected).
NETWORK SETUP (this menu appears only if a network card is installed in the controller)
MODBUS
ADDRESS
BAUD RATESelect a baud rate of 9600, 19200, 38.4K, 57.6K, or 115.2K. Default: 19200
STOP BITSSelect 1 or 2 stop bits. Default: 1
MODBUS MODESelect RTU or ASCII. Default: RTU
DATA ORDERSelect NORMAL or SWAPPED.
DISPLAY SETUP
ADJ CONTRAST
LANGUAGE
SET DATE/TIME
Select from the available options to customize the relay functions. See section 4.6.1 on page 34 for
additional details.
Highlight sc100 Analyzer, or either connected sensor then press ENTER to select. Choose a number
between 1 and 247 as the address (each source must have a different address) then press ENTER.
Use the UP and DOWN keys to increase or decrease the contrast, see section 4.3.2 on page 28.
Range = 0–50
The default is English. Choose from the available options to allow all menus to appear in the selected
language.
Use this menu to select the date format and to set the date and time (24-hour (military) format), see
section 4.3.4 on page 29.
SECURITY SETUP (Enter a 6-digit passcode)
SET PASSCODE
ENABLEEnables system security. See section 4.4.1 on page 31.
DISABLEDisables system security. See section 4.4.1 on page 31.
LOG SETUP (Not used for 1720E system. Enable datalogging from the sensor setup menu for 1720E)
DATALOG SETUPSet up datalogging of data and events. See section 4.7.1 on page 36.
ERROR HOLD MODE
HOLD OUTPUTSHolds outputs when unable to communicate with the sensor.
XFER OUTPUTSGoes to transfer state when unable to communicate with the sensor.
4.9.4 Test/Maint Menu
STATUS
Indicates the status of each relay and indicates which sensors are connected to the controller.
OUTPUT CAL
SELECT OUTPUT 1 or 2
Calibrate Analog Output by specifying values to correspond to 4 mA and 20 mA.
HOLD OUTPUTS
SET OUTMODEChoose Hold Outputs or Xfer Outputs.
SET CHANNELSChoose any individual attached sensor or all attached sensors to be held or transferred.
ACTIVATIONSelect Launch or Release.
38
Page 39
4.9.4 Test/Maint Menu (continued)
OVERFEED RESET
Reset the overfeed time out.
TEST OUTPUT
SELECT OUTPUT 1 or 2
User selectable mA value. 0–20 mA
TEST RELAY
SELECT RELAY A, B, or C
Energize or de-energize the selected relay.
RESET CONFIG
Reset to default configuration of the controller
SIMULATION
SELECT SOURCE, SET PARAMETER, SET SIM VALUE
Simulate sensor measurement values for testing the outputs and relays.
SCAN SENSORS
Manually scans for sensors to determine if sensors have been added or removed.
MODBUS STATS
Indicates the communication statistics for use with an external network.
Operation
CODE VERSION
Indicates the controller software version.
39
Page 40
Section 5System Startup
5.1General Operation
Plug the sensor into the unpowered controller by aligning the orientation tab on
the cable connector with the channel in the controller connector. Push in and turn
to secure the connection. Tug gently to check the connection.
After all plumbing and electrical connections have been completed and checked,
place the head on the body and supply power to the system. Ensure the head is
seated on the body when power is applied, since dark readings are measured at
this time. If power is applied while the sensor head is off the turbidimeter body,
cycle the power with the sensor head on the body.
The first time a controller is powered up, a language selection menu will appear.
The user must select the correct language from the displayed options.
Use the UP and DOWN keys to highlight the appropriate language and press ENTER to select.
Following language selection and upon power-up, the controller will search for
connected sensors. The display will show the main measurement screen. Press
the MENU key to access the menus.
5.2Starting Sample Flow
Start sample flow through the instrument by opening the sample supply valve.
Allow the turbidimeter to run long enough for the tubing and body to become
completely wetted and the reading on the display to stabilize. One to two hours or
longer may be required initially for complete stabilization. Allow measurements to
become stable through adequate conditioning before completing instrument
settings or performing calibrations.
5.3Sensor Calibration and Verification
The manufacturer offers two EPA-approved calibration methods one using userprepared formazin and the other using StablCal® stabilized formazin. Two
verification methods (wet and dry) are also offered.
The 1720E Turbidimeter is factory-calibrated using StablCal
before shipment. The instrument must be recalibrated before use to meet
published accuracy specifications. In addition, recalibration is recommended after
any significant maintenance or repair and at least once every three months during
normal operation. The turbidimeter body and bubble trap must be thoroughly
cleaned and rinsed before initial use and prior to each calibration.
Tips to achieve the most accurate calibrations:
•Optimum performance is achieved when calibration is performed in the
turbidimeter body. Accurately prepare the standard then add it to the
turbidimeter body at the appropriate step in the procedure. Do not prepare the
standard in the body.
®
Stabilized Formazin
40
•Stop sample flow, drain, and clean the turbidimeter body before beginning the
calibration procedure.
•Always clean the photocell window per the instructions in section 6.4.2 on
page 48. Rinse the photocell with deionized water and dry with a soft, lint-free
cloth before calibrating.
Page 41
System Startup
•Always clean the turbidimeter body or calibration cylinder per the instructions
in section 6.4.3 on page 48. Rinse with deionized water before calibrating.
•Store the calibration cylinder upside-down to minimize contamination between
calibrations.
•Pour the calibration standard into the turbidimeter body at the inflow end (left
side when facing the instrument).
•Gently invert StablCal standards for 1 minute before opening. Do not shake.
This ensures a consistent turbidity of the standard.
•If the 20.0 NTU StablCal standard is allowed to sit in the calibration cylinder or
turbidimeter body for more than 15 minutes, it must be remixed (gently swirled
in the calibration cylinder) before use to ensure a consistent turbidity.
•Discard all standards after use per the instructions on the container.
Never transfer the standard back into its original container. Contamination
will result.
•Always recalibrate after restoring default settings.
5.3.1 User-prepared Calibration
Before starting the calibration, read and apply the tips in section 5.3.
Follow the procedure as written (using 1 L of deionized water and 5.0 mL of
4000 NTU formazin) if using a calibration cylinder for calibration.
If using the turbidimeter body for the user-prepared calibration follow the
procedure below using 20 NTU formazin in step 6c. Prepare the standard as follows:
1. Stop the sample flow, then drain and clean the body.
2. Prepare a 20 NTU standard by adding 5.0 mL of 4000 NTU formazin to a
1-L flask. Dilute to the mark with deionized water and invert gently to mix.
3. Drain the deionized water and pour the prepared 20 NTU standard into the
turbidimeter body at step 6c. Do not add additional 4000 NTU formazin.
StepSelectMenu Level/InstructionsConfirm
1
2
MAIN MENU—
SENSOR SETUP
3
4
5
—USER PREPD CAL
SELECT SENSOR (if more than one sensor is connected)
CALIBRATE
41
Page 42
System Startup
StepSelectMenu Level/InstructionsConfirm
6
a
Stop sample flow. Drain body and clean the body and bubble trap.
b
c
ADD 5 ML OF 4000 NTU FORMAZIN INTO CAL CYLINDER.
d
Select ACTIVE, HOLD, or TRANSFER
FILL CYL WITH 1 L DI WATER. REPLACE HEAD.
Measured reading (based on a gain of 1.0) displayed
Measured reading (based on a gain of 1.0) displayed
e
f
OUTPUT MODE
(Remove head)
GOOD CAL!
GAIN: X.XX
ENTER TO CONT
Verify CAL?
(see Note below)
(to store)
to verify
exit no verify
7
8
—RETURN SENSOR TO MEASURE MODE
9
5.3.2 Calibration with StablCal
Select VERIFICATION type (begin at step 7 in section 5.4.1 on page 44 or
section 5.4.2 on page 45) or enter initials to complete calibration.
Main Menu or Main Measurement Screen—
Note: If a dry verification is performed directly after a calibration, the measured value is
assigned as the expected value for future verifications (when using the dry verification
device with the same serial number). As long as the verification exists within the verification
history, the expected value will be retained. Otherwise, the expected value will be the
nominal labeled value associated with the dry verification device.
®
Before starting the calibration, read and apply the tips in section 5.3 on page 40.
StepSelectMenu Level/InstructionsConfirm
1
MAIN MENU—
2
42
SENSOR SETUP
Page 43
System Startup
StepSelectMenu Level/InstructionsConfirm
3
—CALIBRATE
4
5
6a
b
c
d
STABLCAL CAL
OUTPUT MODE
Select ACTIVE, HOLD, or TRANSFER
Drain/clean/rinse the turbidimeter body or cal cylinder.
POUR 20 NTU STD INTO CYL/BODY. REPLACE HEAD
Measured reading and reading based on 1.0 gain displayed
GOOD CAL!
GAIN: X.XX
ENTER TO CONT
Verify CAL? (see Note below)
(to store)
to verify/
e
7
—RETURN SENSOR TO MEASURE MODE
Select VERIFICATION type (begin at step 7 in section 5.4.1 on page 44 or section 5.4.2
8
Note: If a dry verification is performed directly after a calibration, the measured value is
assigned as the expected value for future verifications (when using the dry verification
device with the same serial number). As long as the verification exists within the verification
history, the expected value will be retained. Otherwise, the expected value will be the
nominal value associated with the dry verification device.
5.4Instrument Verification
Instrument verification is intended as a simple check to ensure turbidimeter
functionality between calibrations. A verification is initially performed directly after
a calibration and subsequent independent verifications are referenced to the initial
verification. The pass/fail criteria is set and subsequent verifications are deemed
good or bad, when compared to the initial verification. All verifications are based
on the current calibration and must be repeated when the instrument is
recalibrated or when the pass/fail criteria is not met.
exit no verify
on page 45) or enter initials to complete calibration.}
Main Menu or Main Measurement Screen—
43
Page 44
System Startup
Two types of verifications are offered. The dry verification is performed using a
“dry” calibration device such as the ICEPIC™. A wet verification is performed
using a standard with a predetermined value such as StablCal
®
Stabilized
Formazin or a user-prepared standard with a value that has been verified on an
independent device such as a laboratory turbidimeter.
5.4.1 Dry Verification
StepSelectMenu Level/InstructionsConfirm
1
2
3
4
5
6
7
8
—PERFORM VER
Choose ACTIVE, HOLD, or TRANSFER
—VER TYPE Select DRY
select 1 NTU STD or 20 NTU STD or verify SN of previously used calibration device
MAIN MENU—
SENSOR SETUP
CALIBRATE
VERIFICATION
OUTPUT MODE
DRY VERIFY
Set Head on Standard
9
10
—RETURN SENSOR TO MEASURE MODE
11
44
Reading Displayed
(to accept)
GOOD VER!
(to store)
ENTER INITIALS (user input)
MAIN MENU or Main Measurement Screen—
Page 45
System Startup
5.4.2 Wet Verification
Before starting the verification, read and apply the appropriate tips in section 5.3
on page 40.
StepSelectMenu Level/InstructionsConfirm
1
2
3
—CALIBRATE
4
5
6
—PERFORM VER
Choose ACTIVE, HOLD, TRANSFER
7
a.
DRAIN AND CLEAN SENSOR BODY. ENTER TO CONT
MAIN MENU—
SENSOR SETUP
VERIFICATION
OUTPUT MODE
Select VER TYPE
Select WET
Enter Std Turbidity
8
9
10
11
b.
c.
d.
POUR STANDARD INTO CYL/BODY. PLACE HEAD ON. ENTER TO CONT
Reading Displayed
GOOD VER!
ENTER INITIALS
RETURN SENSOR TO MEASURE MODE
Main Menu or Main Measurement Screen—
(to accept)
45
Page 46
System Startup
5.5Calibration and Verification History
The calibration and verification history logs contain information on the last
12 calibrations and the last 12 verifications. The calibration history log shows the
gain value, the time and date of the calibration, and the initials of the operator
performing verification.
Note: Restoring default settings
from the DIAG/TEST menu will
return the turbidimeter to its non
calibration state (gain = 1.0) but it
will not remove the previous
calibration history from memory.
The calibration history log is accessed from the Calibrate menu. The verification
history log is accessed from the Verification menu (a submenu of the Calibrate
menu).
Each verification history entry shows the serial number of the verification device
(dry verification) or the value of the verification standard (wet verification), the time
and date of the verification, and the initials of the operator performing the
verification.
Scroll through the entries by pressing the ENTER key. After scrolling through all
12 histories, the display will return to the calibration menu level.
When the instrument is received from the factory, default values or blank spaces
will be shown for the calibration and verification history information. Those values
will be replaced with real data as the history log is filled.
The data is retained as first in, first out. When the log is full, the newest entry is
stored and the oldest entry in the log is deleted.
46
Page 47
Section 6Maintenance
DANGER
Only qualified personnel should conduct the maintenance tasks described
in this section of the manual.
6.1Maintenance Schedule
Maintenance TaskFrequency
Clean the sensor
Calibrate Sensor (as required by regulatory agency) Per agency-dictated schedule.
Scheduled periodic maintenance requirements of the 1720E are minimal and
include calibration and cleaning of the photocell window, bubble trap, and body.
Check and clean the bubble trap and turbidimeter body (as described in
section 6.4.3) if visual inspection shows that it is necessary. Perform other
maintenance on a regular basis; experience will dictate scheduling and may
depend on the installation, sample type, and season.
Before each calibration and as needed. Depends on sample
characteristics.
It is very important to maintain the cleanliness of the interior and exterior of the
turbidimeter body, head assembly, the integral bubble trap, and the surrounding
area. Doing so will ensure accurate, low-level turbidity measurements.
Clean the body before calibration and verification (especially when measurements
are being made at 1.0 NTU or lower).
Check and/or perform a calibration periodically (as experience dictates) using one
of the methods described in section 5.3 on page 40. A calibration history menu
option is available under Sensor Setup/Calibrate.
6.2Removing a Sensor from the System
Prior to physically removing a sensor from the system, record all user defined
settings such as relays, signal averaging, etc. Disconnect the sensor connector at
the controller.
6.3Reinstalling a Sensor on the System
To return the system to normal operation following a software upgrade or sensor
repair perform the following procedure:
1. Detach all sensors from the sc100 controller.
2. From the main menu, press the down arrow key to highlight TEST/MAINT. Press ENTER.
Note: Clean sensors before
reinstallation on the system.
3. Use the down arrow key to scroll to SCAN SENSORS and press ENTER.
4. Remove attached sensors by selecting the corresponding serial number or
select “All”.
5. Power down the sc100 then attach the 1720E.
6. Supply power to the sc100. The system will initialize automatically.
47
Page 48
Maintenance
6.4Cleaning
6.4.1 Cleaning the Controller
With the enclosure securely closed, wipe the exterior with a damp cloth.
6.4.2 Cleaning the Photocell Window
Occasional cleaning of the photocell window is required. The frequency will
depend on the nature and concentration of dissolved and suspended solids in the
sample. Biological activity is a primary factor in mineral scale deposit on the
window and the amount differs with sample temperature. In general, more growth
will occur in warm temperatures and less in cold.
Note: Take care to not scratch the
photocell window.
Inspect the photocell window often to determine cleaning needs. Remove any
organic growth or film on the photocell window before standardization or
calibration. Use a cotton swab and isopropyl alcohol or a mild detergent (such as
Liqui-nox
cleaning with a mild acid applied with a cotton swab followed by a detergent wash.
Do not use abrasive cleaners.
®
) to remove most sediment and dirt. Mineral scale buildup may require
6.4.3 Cleaning the Turbidimeter Body and Bubble Trap
Sediment may collect in the turbidimeter body after extended use. Noise
(fluctuation) in the reading could indicate the need to clean the body and/or bubble
trap. The 1720E bubble trap and bottom plate may be removed to make cleaning
easier. Drain and clean the turbidimeter body before each calibration. Establish a
regular schedule or perform cleaning as determined by visual inspection.
Cleaning the Turbidimeter Body
Note: The turbidimeter body, bubble
trap, and detector must be cleaned
before each calibration.
1. Turn off sample flow to the turbidimeter body.
2. Remove the head assembly and bubble trap cover from the body. Remove the
bubble trap by lifting it vertically. Set it aside to be cleaned separately.
3. Drain the body by removing the plug from the bottom of the body.
4. Replace the drain plug and fill the body to the weir with cleaning solution. This
cleaning solution can consist of dilute chlorine solution (25 mL of household
bleach in 3.78 liters of water) or a laboratory detergent such as Liqui-nox
(1 mL detergent in 1 liter of water).
48
5. Use a soft brush to clean the inside surfaces of the body.
6. Remove the drain plug again and thoroughly flush the turbidimeter body with
ultra-filtered deionized water. Clean and replace the plug.
Cleaning the Bubble Trap
1. Prepare a cleaning solution (as in step 4 above) in a container large enough
to submerge the entire bubble trap.
2. Using a test tube brush such as Cat. No. 690-00, clean each surface.
3. Rinse the bubble trap thoroughly with ultra-filtered deionized water and
reinstall it in the turbidimeter body.
4. Replace the bubble trap cover and head assembly on the top of the body.
Page 49
Maintenance
5. Restore sample flow to the instrument.
6. Calibrate the instrument using one of the methods in section 5.3 on page 40.
If the above cleaning procedures have been performed and the turbidimeter
readings are still noisy, the bottom plate and gasket may need to be removed and
cleaned. Carefully perform the following procedure to ensure the turbidimeter
body integrity is maintained.
1. Turn off sample flow to the turbidimeter body.
2. Remove the head assembly, bubble trap cover, and bubble trap (by lifting it
vertically) from the body.
3. Drain the body by removing the plug from the bottom of the body.
4. Lift the body off of its mounting screws.
5. With the body turned upside-down, remove the two Phillips-head screws
holding the bottom plate.
6. Lift the bottom plate off the body; set the gasket aside for use in reassembly.
7. Use a soft brush and a dilute cleaning solution (as prepared above) to clean
the bottom plate and inside surfaces of the turbidimeter body. Rinse the entire
body and bottom plate with ultra-filtered deionized water.
8. Reassemble by inserting the gasket into the molded channel in the
bottom plate.
9. Fit the bottom plate onto the turbidimeter body.
10. Reinstall both screws and carefully tighten to 15 inch-lb maximum.
11. Reinstall the turbidimeter onto the wall mounting screws.
12. Replace the bubble trap, bubble trap cover, and head assembly on the top of
the body.
13. Restore sample flow to the instrument.
6.4.4 Replacing the Lamp Assembly
The Lamp Assembly is located on the head assembly. Under normal use, Hach
recommends replacing the lamp once a year to maintain peak performance.
Replacement bulbs have been “burned-in” at the factory and are ready for
installation and use.
To change the lamp, refer to Figure 28 on page 50 and perform the
following steps:
1. Disconnect power to the controller to remove all power to the turbidimeter.
2. Disconnect the lamp leads by unplugging the connector.
3. After the bulb has cooled, remove as follows:
49
Page 50
Maintenance
Figure 28Lamp Replacement
a. Wear cotton gloves to protect your hands and to avoid fingerprints on
the bulb.
b. Grasp the bulb.
c. Twist the bulb in a counterclockwise direction, pulling out slightly, until it is
released from the housing.
d. Pull the lamp leads and connector through the hole in the lamp housing.
Do not touch the new bulb with bare hands. Etched glass and reduced lamp life
will result. Wear cotton gloves or grasp the lamp assembly with a tissue to avoid
contamination. If contamination occurs, clean the glass bulb portion with isopropyl
alcohol.
Replace the bulb by reversing the above instructions. The bulb base only fits one
way; align the notch in the metal bulb flange with the hole in the lamp holder.
1
1.Lamp Housing2.Lamp Assembly3.Photo Detector
2
3
50
Page 51
6.5Fuse Replacement
The controller contains two mains fuses. Failed fuses are an indication that an
equipment problem could exist. Problem resolution and fuse replacement should
only be performed by qualified service personnel. Refer to Figure 29 and perform
the following steps to replace the fuses:
1. Disconnect power to the controller (including power to relays if powered).
2. Open the hinged controller cover by completely loosening all four captive
screws in the cover.
3. Remove the high voltage barrier; pull out on the lever of the captive fastener
then pull straight out on the barrier. Set the barrier aside for reinstallation.
4. Remove the fuses and install new fuses of the same type and rating
(T, 1.6A, 250 V, slow blow).
5. Reinstall the high voltage barrier.
6. Close the controller cover and hand-tighten the four screws.
Maintenance
Figure 29Fuse Replacement
J1
S1
J4
NETWORK
INTERFACE
J2
U5
U9
CARD
7. Reconnect all power to the instrument.
J3
J5
NCNCNC
NONO
RELAYCRELAYBRELAYA
J6
1
F1
F2
NO
COMCOMCOM
1.Fuses F1 and F2, 1.6 A
Disconnect
Power
51
Page 52
Section 7Troubleshooting
7.1Error Codes
When a sensor is experiencing an error condition, the sensor reading on the
measurement screen will flash and all relays and analog outputs associated with
the sensor will be held.
Highlight the Probe Diag menu and press ENTER. Highlight Errors and press
ENTER to determine the case of the error. Errors are defined in Table 9.
Table 9 Error Codes
Displayed ErrorDefinition
A/D FailFailed A/D converter. Call the Service Department.
Lamp Fail
Flash FailDatalog and event log will not work.
7.2Warnings
A sensor warning will leave all menus, relays, and outputs functioning normally,
but will cause a warning icon to flash on the right side of the display. Highlight the
Probe Diag menu and press ENTER to determine the cause of the error.
Warnings are defined in Table 10.
The light source has failed. See section 6.4.4 on page 49 for
instructions for replacing it.
A warning may be used to trigger a relay and users can set warning levels to
define the severity of the warning. Errors are defined in Table 10.
Table 10 Warning Codes
Warning
Number
1Dark Reading WarningDark reading detects too much light.
2Temperature Warning
3Data Log Full Warning
4Event Log Full Warning
55 Volt Warning
6Voltage in Warning
7Lamp V Warning
8Lamp I Warning
9Output Mode Not Normal
Displayed WarningDefinition/Resolution
Sensor head internal temperature is higher than specified. Contact the Service
Department.
Sensor data log is full. No additional data will be logged until sensor log is
downloaded into controller memory.
Sensor data log is full. No additional data will be logged until sensor log is
downloaded into controller memory.
Monitored voltage is outside the range of 4.5 – 5.5 V @ microprocessor input ->
698 to 854 counts from the microprocessor ADC.
Monitored voltage is outside the range of 9.08 – 14.3 V @ microprocessor pin ->
279 to 435 counts from the microprocessor ADC.
Monitored voltage is outside the range of 3.96 – 4.48 V @ microprocessor pin ->
614 to 385 counts from the microprocessor ADC.
Monitored voltage is outside the range of 1.67 – 2.75 V @ microprocessor pin ->
39 to 64 counts from the microprocessor ADC.
Activated when the sensor is not in normal measurement mode (such as when in
calibration or verification mode).
52
Page 53
Tabl e 11 presents sensor warnings displayed in the Event Log, possible causes,
and corrective actions.
Table 11 Troubleshooting
Sensor WarningPossible CauseCorrective Action
Lamp burned out Replace the lamp. See section 6.4.4 on page 49.
Lamp unpluggedRestore connection
Bad Lamp
Low Signal
Bad System Voltage
A/D Converter Timeout
High Dark Counts
+12 V connection loose at controllerRestore connection
Dislodged lampReinstall lamp
Bad circuit board in turbidimeter headContact the Customer Service Department.
Photocell coated/dirty
Photocell wires disconnectedReconnect wires
Photocell broken/cracked
Lens coated/dirty Clean the lens using isopropyl alcohol and a cotton swab.
Obstructed light pathRemove obstruction
Sample turbidity >100 NTUSwitch to a high range turbidimeter
See Bad Lamp causes aboveSee Bad Lamp corrective actions above
Improper wiring at controller
Turbidimeter head cable shortened to
improper length
Fluctuation in voltageTurn instrument power off and back on.
Bad circuit board in turbidimeter headContact the Service Department
Fluctuation in voltageTurn instrument power off and back on.
Bad circuit board in turbidimeter headContact the Customer Service Department.
Light Leak—Turbidimeter head not on
turbidimeter body or calibration cylinder
during Power Up or Zero Electronics
Light Leak—Turbidimeter head not
properly aligned on the turbidimeter
body or calibration cylinder during
Power Up or
Zero Electronics
Photocell broken/crackedContact the Customer Service Department.
See Cleaning the Photocell Window on page 48.
Contact the Customer Service Department.
Replace photocell
Contact the Customer Service Department.
See the controller wiring instructions in section 3.2,
Electrical Installation on page 14.
Contact the Customer Service Department.
Make sure the turbidimeter head is on the turbidimeter body
and properly aligned and repower instrument or perform
ZERO ELECTRONICS in the CALIBRATION MENU.
Make sure the turbidimeter head is properly aligned and
repower instrument or perform ZERO ELECTRONICS in
the CALIBRATION MENU.
Troubleshooting
Table 12 Additional Malfunctions Not Recorded in the Event Log
SymptomPossible CauseCorrective Action
Continuous Underrange
(negative reading)
Continuous Overrange
(100 NTU)
Calibration standards not in the
correct order or incorrect dilution
Calibration standards not in the
correct order or incorrect dilution
The following table presents additional malfunctions which may not be recorded in
the Event Log.
Verify the accuracy of calibration standards and recalibrate the
instrument. See Low Signal in Table 11.
Verify the accuracy of calibration standards and recalibrate the
instrument.
53
Page 54
Troubleshooting
Table 12 Additional Malfunctions Not Recorded in the Event Log (continued)
SymptomPossible CauseCorrective Action
Erratic ReadingsCalibration standards have the
same value
Inadequate bubble removal from
sample
High ReadingsDeionized water turbidity is
greater than 0.5 NTU
Verify the accuracy of calibration standards and recalibrate the
instrument.
Increase the signal averaging time to a longer interval.
Make sure the Bubble Reject feature is turned on.
Slow the flow of sample into the instrument.
Clean the instrument.
Access Calibration History for turbidity value of ultra-filtered water.
Verify the flow is between 200–750 mL/min.
Recalibrate the instrument.
7.3Event Codes
Event codes are not displayed on the screen of the sc100 and must be
downloaded from the event log to be viewed. See section 4.7 on page 36 for
additional information. Troubleshooting actions are given in section Table 11 on
page 53.
Table 13 Event Log List
EventEvent #Data1Data2Data3
Bubble Reject Change0
Filter Size Change1
Data Log Interval Change2
Power On31——
Calibration4Cal GainInitials—
Verification5Expected ValueMeas ValueInitials
Dark Reading Warning6Measured Value——
Temperature Warning7Temperature Value——
0 = 30 sec
1 = 1min
2 = 5 min
3 = 10 min
4 = 15 min
16 = +5V high
32 = +5V low
64 = V in high
128 = V in low
1 = lamp V high
2 = lamp V low
4 = lamp I high
8 = lamp I low
1 = Hold
2 = Transfer
——
——
——
——
——
——
54
Page 55
Section 8Replacement Parts and Accessories
Replacement Items
ItemCat. No.
Fuse, T, 1.6 A, 250 V .......................................................................................................................................... 52083-00
Instruction Manual, sc100 Controller, English .................................................................................................... 58600-18
Instruction Manual, 1720E Turbidimeter System, English .................................................................................. 60100-18
Cable, sensor extension, 7.7 m (25 ft)................................................................................................................ 57960-00
Cap, Connector Receptacle ...................................................................................................... each ................ 52100-00
Deionized (demineralized) water ..................................................................................................1 L .................... 272-56
Digital Output Card for ModBUS RS232 communication ................................................................................... 59200-00
Digital Output Card for ModBUS RS485 communication ................................................................................... 59200-01
Drain plug for the 1720E body................................................................................................... each................ 44116-00
Filter, 0.45 µm, to produce ultra-filtered water for cleaning and calibration ............................... each ................ 26705-00
Filter, 0.2 µm, to produce ultra-filtered water for calibration standard preparation .................... each................ 23238-10
Formazin Calibration Kit includes:
Calibration Cylinder, TenSette
Floor Stand ............................................................................................................................... each ................ 57432-00
ICE-PIC Module for Calibration and Calibration Verification
20 NTU ................................................................................................................................. each ................ 52250-00
1 NTU................................................................................................................................... each ................ 52215-00
0.5 NTU................................................................................................................................ each ................ 52225-00
Photocell Replacement Kit for the 1720E .................................................................................. each ................ 52180-00
Pipet tips for 19700-01 TenSette Pipet ................................................................................... 50/pkg ................ 21856-96
Pipet tips for 19700-10 TenSette Pipet ................................................................................... 50/pkg ................ 21997-96
Power Cord with strain relief, 115 V.................................................................................................................... 54488-00
Power Cord with strain relief, 230 V ................................................................................................................... 54489-00
Power Cord Kit, 10A-125V, 1.8 m (6 ft), UL/CSA listed ............................................................. each ................ 46306-00
Power Cord Kit, 10A-230V, 1.8 m (6 ft), European-style plug, VDE approved .......................... each ................ 46308-00
StablCal Calibration Set for the 1720 Series Turbidimeter
Sun shield........................................................................................................................................................... 58690-00
Swabs, Cotton, presterilized for cleaning the photodetector ................................................ 100/pkg ................ 25543-00
TenSette Pipet, 0.1 to 1.0 mL .................................................................................................... each................ 19700-01
TenSette Pipet, 1.0 to 10.0 mL .................................................................................................. each................ 19700-10
Tubing, Inlet, ¼ inch O.D., Polyethylene ................................................................................per foot ................ 51322-00
Tubing, Outlet, ½ inch I.D., ¾ inch O.D., Tygon R3603..........................................................per foot................ 51263-00
®
Pipet, 4000 NTU Formazin Primary Standard (500 mL).... each................ 44156-00
55
Page 56
Replacement Parts and Accessories
Calibration and Verification Standards and Accessories
6:30 a.m. to 5:00 p.m. MST
Monday through Friday
(800) 227-HACH (800-227-4224)
By Fax:
(970) 669-2932
By Mail:
Hach Company
P.O. Box 389
Loveland, Colorado 80539-0389 U.S.A.
Ordering information by e-mail: orders@hach.com
9.1Information Required
•Hach account number (if available)•billing address
•Your name and phone number•Shipping address
•Purchase order number•Catalog number
•Brief description or model number•Quantity
9.2International Customers
Hach maintains a worldwide network of dealers and distributors. To locate the
representative nearest you, send e-mail to intl@hach.com or contact:
Hach Company World Headquarters; Loveland, Colorado, U.S.A.
Telephone: (970) 669-3050; Fax: (970) 669-2932
9.3Technical and Customer Service (U.S.A. only)
Hach Technical and Customer Service Department personnel are eager to
answer questions about our products and their use. Specialists in analytical
methods, they are happy to put their talents to work for you.
Call 1-800-227-4224 or e-mail techhelp@hach.com
57
Page 58
Section 10Repair Service
Authorization must be obtained from Hach Company before sending any items for
repair. Please contact the Hach Service Center serving your location.
In the United States:
Hach Company
Ames Service
100 Dayton Avenue
Ames, Iowa 50010
(800) 227-4224 (U.S.A. only)
FAX: (515) 232-3835
In Latin America, the Caribbean, the Far East, the
Indian Subcontinent, Africa, Europe, or the Middle East:
Hach Company World Headquarters,
P.O. Box 389
Loveland, Colorado, 80539-0389 U.S.A.
Telephone: (970) 669-3050
FAX: (970) 669-2932
E-mail: intl@hach.com
58
Page 59
Section 11Limited Warranty
Hach Company warrants its products to the original purchaser against any defects that are due
to faulty material or workmanship for a period of one year from date of shipment unless
otherwise noted.
In the event that a defect is discovered during the warranty period, Hach Company agrees that, at
its option, it will repair or replace the defective product or refund the purchase price, subject to
the pro-rated schedule above, excluding original shipping and handling charges. Any product
repaired or replaced under this warranty will be warranted only for the remainder of the original
product warranty period.
This warranty does not apply to consumable products such as chemical reagents; or
consumable components of a product, such as, but not limited to, lamps and tubing.
Contact Hach Company or your distributor to initiate warranty support. Products may not be
returned without authorization from Hach Company.
Limitations
This warranty does not cover:
•Damage caused by acts of God, natural disaster, labor unrest, acts of war (declared or
undeclared), terrorism, civil strife or acts of any governmental jurisdiction
•Damage caused by misuse, neglect, accident or improper application or installation
•Damage caused by any repair or attempted repair not authorized by Hach Company
•Any product not used in accordance with the instructions furnished by Hach Company
•Freight charges to return merchandise to Hach Company
•Freight charges on expedited or express shipment of warranted parts or product
•Travel fees associated with on-site warranty repair
This warranty contains the sole express warranty made by Hach Company in connection with its
products. All implied warranties, including without limitation, the warranties of merchantability
and fitness for a particular purpose, are expressly disclaimed.
Some states within the United States do not allow the disclaimer of implied warranties and if this
is true in your state the above limitation may not apply to you. This warranty gives you specific
rights, and you may also have other rights that vary from state to state.
This warranty constitutes the final, complete, and exclusive statement of warranty terms and no
person is authorized to make any other warranties or representations on behalf of Hach
Company.
Limitation of Remedies
The remedies of repair, replacement or refund of purchase price as stated above are the
exclusive remedies for the breach of this warranty. On the basis of strict liability or under any
other legal theory, in no event shall Hach Company be liable for any incidental or consequential
damages of any kind for breach of warranty or negligence.
59
Page 60
Section 10Compliance Information
Hach Co. certifies this instrument was tested thoroughly, inspected and found to
meet its published specifications when it was shipped from the factory.
The Model sc100 with 1720E Sensor has been tested and is certified as
indicated to the following instrumentation standards:
Product Safety
UL 61010A-1 (ETL Listing # 65454)
CSA C22.2 No. 1010.1 (ETLc Certification # 65454)
Certified by Hach Co. to EN 61010-1 Amds. 1 & 2 (IEC1010-1) per
73/23/EEC, supporting test records by Intertek Testing Services.
Immunity
This equipment was tested for Industrial level EMC per:
EN 61326(EMC Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use)
Hach Company, certified compliance by Hach Company.
per 89/336/EEC EMC: Supporting test records by
Emissions
Standards include:
IEC 1000-4-2:1995 (EN 61000-4-2:1995) Electro-Static Discharge Immunity
(Criteria B)
IEC 1000-4-3:1995 (EN 61000-4-3:1996) Radiated RF Electro-Magnetic Field
Immunity (Criteria A)
IEC 1000-4-4:1995 (EN 61000-4-4:1995) Electrical Fast Transients/Burst
(Criteria B)
IEC 1000-4-5:1995 (EN 61000-4-5:1995) Surge (Criteria B)
IEC 1000-4-6:1996 (EN 61000-4-6:1996) Conducted Disturbances Induced
by RF Fields (Criteria A)
IEC 1000-4-11:1994 (EN 61000-4-11:1994) Voltage Dip/Short Interruptions
(Criteria B)
Additional immunity Standard/s include:
ENV 50204:1996 Radiated Electro-Magnetic Field from Digital Telephones
(Criteria A)
This equipment was tested for Radio Frequency Emissions as follows:
Per 89/336/EEC EMC: EN 61326:1998(Electrical Equipment for measurement,
control and laboratory use—EMC requirements) Class “A” emission limits.
Supporting test records by Hewlett Packard, Fort Collins, Colorado Hardware Test
Center (A2LA # 0905-01) and certified compliance by Hach Company.
60
Standards include:
EN 61000-3-2 Harmonic Disturbances Caused by Electrical Equipment
EN 61000-3-3 Voltage Fluctuation (Flicker) Disturbances Caused by Electrical
Equipment
Page 61
Compliance Information
Additional Emissions Standard/s include:
EN 55011 (CISPR 11) Class “A” emission limits
Canadian Interference-causing Equipment Regulation, IECS-003, Class A
Supporting test records by Hewlett Packard, Fort Collins, Colorado Hardware Test
Center (A2LA # 0905-01) and certified compliance by Hach Company.
This Class A digital apparatus meets all requirements of the Canadian
Interference- Causing Equipment Regulations.
Cet appareil numÈrique de la classe A respecte toutes les exigences du
RËglement sur le matÈriel brouilleur du Canada.
FCC PART 15, Class “A” Limits
Supporting test records byHewlett Packard, Fort Collins, Colorado Hardware Test
Center (A2LA # 0905-01) and certified compliance by Hach Company.
This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions:
(1) this device may not cause harmful interference, and (2) this device must
accept any interference received, including interference that may cause undesired
operation.
Changes or modifications to this unit not expressly approved by the party
responsible for compliance could void the user's authority to operate the
equipment.
This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with
the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference, in which case the user will be required to correct the interference at
his own expense. The following techniques of reducing the interference problems
are applied easily.
1. Disconnect the Model sc100 Controller from its power source to verify that it is
or is not the source of the interference.
2. If the Model sc100 Controller is connected into the same outlet as the device
with which it is interfering, try another outlet.
3. Move the Model sc100 Controller and 1720E sensor away from the device
receiving the interference.
4. Reposition the device receiving the interference.
5. Try combinations of the above.
61
Page 62
Appendix AModBUS Register Information
Table 14 Controller ModBUS Registers
Group NameTag NameRegister # Data TypeLengthR/WDescription
Measurements
SetupLanguage40003
SetupDate Format40004
Setup
Setup/Analog Output 1Source40006
Setup/Analog Output 1Sensor Select40007
Setup/Analog Output 1
Setup/Analog Output 1Type40009
Setup/Analog Output 1Transfer Value40010Float2R/WSets the transfer value
Setup/Analog Output 1Filter40012
Setup/Analog Output 1
Setup/Analog Output
1/Linear
Setup/Analog Output
1/Linear
Setup/Analog Output
1/PID
Setup/Analog Output
1/PID
Setup/Analog Output
1/PID
Setup/Analog Output
1/PID
Setup/Analog Output
1/PID
Setup/Analog Output
1/PID
Setup/Analog Output
1/PID
Setup/Analog Output 2Source40028
Setup/Analog Output 2Sensor Select40029
Setup/Analog Output 2
Calculated
Value
Error Hold
Mode
Measurement
Select
0mA - 4mA
Select
Min Setting40014Float2R/WSets the min output value
Max Setting40016Float2R/WSets the max output value
PID Mode40018
PID Manual
Set
PID Setpoint40021Float2R/WSets the PID setpoint
PID Phase40023
PID
Proportional
Band
PID Integral
Time
PID
Derivative
Time
Measurement
Select
40001Float2R
Unsigned
Integer
Unsigned
Integer
40005
40008
40013
40019Float2R/W
40024Float2R/WSets the PID proportional band
40026
40027
40030
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
Unsigned
Integer
1R/WCurrent System Language
1R/W
1R/W
1R/W
1R/W
1R/W
1R/W
1R/W
1R/W
1R/WSets the PID mode (0 = auto; 1 = manual)
1R/W
1R/WSets the PID integral time (min)
1R/WSets the PID derivative time (min)
1R/W
1R/W
1R/W
Value calculated from two sensor
measurements
Current Data Display Format
(0 = DD/MM/YY; 1 = MM/DD/YY;
2 = DD-MM-YY; 3 = MM-DD-YY)
Error Mode Hold State (0 = Hold outputs;
1 = Transfer outputs to predefined value)
Selects data source for this output
(0 = None; 2 = sensor; 4 = Calculation)