Measuring accuracy (with standard solution)2% of the measured value + 0.015 mg/L
Repeatability (with standard solution)0.7% of the measured value + 0.005 mg/L
1
Response time (90 %)10 minutes
Adjustable measuring interval10 to 120 minutes
Power supplyWith power cable, connected to a SC1000 controller or
SC200 with power box.
Analyzer and drain tubing: 115 V or 230 V versions
Data transmissionSC standard
Electrical power consumption500 VA
Electrical fuse protectionSupplied by SC1000 controller or SC200 controller with
power box.
Maximum of two analyzers for each SC Controller.
OutputsRelay, analog outputs, network interface through
SC1000/SC200 controller2.
Operating temperaturePHOSPHAX sc LR: –20 to 40 °C (–4 to 104 °F); 95% relative
humidity, non-condensing, non-corrosive
PHOSPHAX indoor sc LR: 5 to 40 °C (41 to 104 °F); 95%
relative humidity, non-condensing, non-corrosive
Storage temperature–20 to 60 °C (–4 to 140 °F); 95% relative humidity, non-
condensing
Data and power cable lengths2 m (80 in.) from edge of enclosure
CertificationsCE marked. Listed to UL and CSA safety standards by TÜV.
Warranty1 year (EU: 2 years)
1
At room temperature. ± 4% for values > 1 mg/L PO4–P along the full temperature range.
2
Refer to the controller documentation for more information about the relay, analog and digital outputs.
3
Specifications
1.1Sample requirements
The water from the sample source(s) must agree with the specifications that follow.
SpecificationDescription
Sample flow rate1.0 to 20.0 L/h
Sample pressureWith continuous sample preparation: –30 mbar to +50 mbar at overflow vessel.
Sample temperature4 to 45 °C (39 to 113 °F)
Sample qualityUltra filtrated or comparable
Sample pH5 to 9
1.2Interferences
Table 1 shows that the ions were individually examined to the given concentrations and
do not cause interference. No cumulative effects or influences of other ions were found.
Verify the measurement results with sample dilutions or standard additions.
Table 1 Interfering substances
Interfering substanceInterference level
–
Cl
5000 mg/L
4
Section 2General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or
consequential damages resulting from any defect or omission in this manual. The
manufacturer reserves the right to make changes in this manual and the products it
describes at any time, without notice or obligation. Revised editions are found on the
manufacturer’s website.
2.1Safety information
The manufacturer is not responsible for any damages due to misapplication or misuse of this
product including, without limitation, direct, incidental and consequential damages, and disclaims
such damages to the full extent permitted under applicable law. The user is solely responsible to
identify critical application risks and install appropriate mechanisms to protect processes during a
possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment.
Pay attention to all danger and caution statements. Failure to do so could result in serious
injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or
install this equipment in any manner other than that specified in this manual.
2.1.1 Use of hazard information
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death
or serious injury.
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in
death or serious injury.
Indicates a potentially hazardous situation that may result in minor or moderate injury.
N O T I C E
D A N G E R
W A R N I N G
C A U T I O N
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that
requires special emphasis.
2.1.2 Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol on the instrument is referenced in the
manual with a precautionary statement.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal
systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
This symbol indicates that a risk of electrical shock and/or electrocution exists.
N O T I C E
5
General information
This symbol indicates the need for protective eye wear.
This symbol indicates that the marked item requires a protective earth connection. If the instrument is not supplied
with a ground plug on a cord, make the protective earth connection to the protective conductor terminal.
This symbol, when noted on the product, identifies the location of a fuse or current limiting device.
This symbol identifies a risk of chemical harm and indicates that only individuals qualified and trained to work with
chemicals should handle chemicals or perform maintenance on chemical delivery systems associated with the
equipment.
This symbol indicates that the marked item can be hot and should not be touched without care.
This symbol indicates the presence of devices sensitive to Electro-static Discharge (ESD) and indicates that care
must be taken to prevent damage with the equipment.
This symbol indicates that the object is heavy.
This symbol indicates that the marked item should not be touched.
2.1.3 Chemical and biological safety
Chemical or biological hazards. If this instrument is used to monitor a treatment process
and/or chemical feed system for which there are regulatory limits and monitoring
requirements related to public health, public safety, food or beverage manufacture or
processing, it is the responsibility of the user of this instrument to know and abide by any
applicable regulation and to have sufficient and appropriate mechanisms in place for
compliance with applicable regulations in the event of malfunction of the instrument.
2.1.4 Certification
This equipment is not intended for use in residential environments and may not provide adequate
protection to radio reception in such environments.
Canadian Radio Interference-Causing Equipment Regulation, IECS-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian Interference-
Causing Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation
canadienne sur les équipements provoquant des interférences.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device complies with Part 15 of
the FCC Rules. Operation is subject to the following conditions:
D A N G E R
C A U T I O N
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including interference that
may cause undesired operation.
6
Changes or modifications to this equipment not expressly approved by the party
responsible for compliance could void the user's authority to operate the equipment. This
equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications. Operation of this equipment in
a residential area is likely to cause harmful interference, in which case the user will be
required to correct the interference at their expense. The following techniques can be
used to reduce interference problems:
1. Disconnect the equipment from its power source to verify that it is or is not the source
of the interference.
2. If the equipment is connected to the same outlet as the device experiencing
interference, connect the equipment to a different outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the interference.
5. Try combinations of the above.
2.2Product overview
General information
The PHOSPHAX sc LR is a one-channel analyzer that measures ortho-phosphate ions
3—
(PO
) in waste water and surface water. The analyzer does not measure diphosphates
4
or polyphosphates. The measuring principle is based on the Vanadat-Molybdat method,
also known as the Yellow method.
The analyzer is used with an SC Controller. The measured value shows on the controller
display as PO4–P (default) or PO
3—
.
4
There are two analyzer models available:
•PHOSPHAX sc LR: weather-resistant enclosure for outdoor installation
•PHOSPHAX indoor sc LR: for indoor installation
The reagents and standards necessary for the chemical analysis are internally installed in
the analyzer enclosure. The analyzer uses pumps, valves and syringes to move the
sample and reagents to the measuring cell on the analytics panel. When the
measurement cycle is complete, the analyzer discards the sample through the drain line.
The analyzer can automatically do cleaning cycles and calibrations for better
measurement performance. Refer to Figure 1.
The sample must be prepared and filtered before analysis. Refer to Sample requirements
on page 4. Based on the system configuration, one or two analyzers can connect to a
FILTRAX instrument for filtration (or an applicable sample feed) and measure one or two
parameters. Refer to System configuration options on page 16.
7
General information
Figure 1 Product overview
1 PHOSPHAX sc LR6 Piston pump11 Valve block
2 Status indicator light
3
7 Safety glass12 Overflow vessel
3 Door lock8 Reagent pumps13 Transport lock
4 PHOSPHAX indoor sc LR9 Measuring cell14 Lever
5 Pocket for manual (only in outdoor
10 Analytics panel
models)
3
Refer to Status indicator light on page 9.
4
Refer to Remove the transport lock on page 14.
8
4
General information
2.2.1 Status indicator light
The status indicator light shows the analyzer condition. Refer to Table 2.
Table 2 Status indicator description
ColorStatus
GreenThe analyzer is in operation with no warnings, errors or reminders.
OrangeThe analyzer is in operation with active warnings or reminders.
RedThe analyzer is not in operation because of an error condition. An important problem has occurred. It is necessary
to resolve the error before further operation is possible.
Flashing There is no communications between the analyzer and the controller.
2.3Product components
Make sure that all components have been received. Refer to Figure 2. If any items are
missing or damaged, contact the manufacturer or a sales representative immediately.
2 Door keys11 Console bracket, wall mount20 Sealing disc
3 Tubing plug (3x)12 Blind plug21 Blind plug
4 Sealing plug (3x)13 Sample tube22 Cleaning Solution C
5 Sealing plug type 2 (not used)14 Drain tube23 Standard Solution S
6 Screw, M5 × 40 (2x)15 Tube cutter24 Reagent A
7 Screw, M5 × 8 (2x)16 Collecting tray25 Reagent B
8 Lock washer, M5 (2x)17 Ferrule (4x)
9 Lock washer, M6 (2x)18 Flangeless nut (4x)
10
Section 3Installation
Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
3.1Installation guidelines
Install the instrument:
•As near the sample source as possible to decrease analysis delay
•In a clean, dry, well ventilated and temperature controlled location
•In a location with minimum vibrations that has no direct exposure to sunlight
•In an environmental enclosure that supplies protection from precipitation and direct
sunlight, good ventilation and temperature control if installed outdoors
•In a location where the power switch and power cord are visible and easily accessible
•In a location where there is sufficient clearance around it to make plumbing and
electrical connections
3.2Mechanical installation
3.2.1 Attach the instrument to a wall
Risk of injury or death. Make sure that the wall mounting is able to hold 4 times the weight
of the equipment.
D A N G E R
D A N G E R
C A U T I O N
Personal injury hazard.
Instruments or components are heavy. Use assistance to install or move.
The object is heavy. Make sure that the instrument is securely attached to a wall, table or
floor for a safe operation.
Attach the instrument upright and level on a flat, vertical surface. Keep a minimum
clearance of 813 mm (32 in.) in front of the instrument to open the door. Refer to Figure 3.
Mounting hardware is supplied by the user. Make sure that the fastening has sufficient
load bearing capacity (approximately 160 kg). The wall plugs must be selected and
approved to suit the properties of the wall. Refer to the illustrated steps that follow.
For rail mount and stand mount installations, refer to the documentation supplied with the
mounting hardware.
11
Installation
Figure 3 Mounting dimensions
12
Installation
3.2.2 Open the enclosure
Open the analyzer enclosure to get access to the wiring connections and plumbing.
C A U T I O N
Personal injury hazard. The object is heavy. Make sure that the instrument is securely
attached to a wall, table or floor for a safe operation.
C A U T I O N
Electrical shock hazard. Make sure that no water can enter the enclosure or come into
contact with the circuit boards.
13
Installation
Figure 4 Open the door
Use the door hook to open the door safely. Refer to the illustrated steps that follow.
As an alternative, remove the door for better access during installation and maintenance
procedures. Refer to Figure 5. Make sure to install and close the door before operation.
Figure 5 Remove the door
3.2.3 Remove the transport lock
Remove the transport lock from the analyzer. Refer to Figure 1 on page 8.
Note: Keep the transport lock for storage or shipping.
14
3.3Electrical connectors and plumbing access ports
Figure 6 shows the electrical connectors and fittings on the instrument. Use the tubing
plug to put tubing or cables through the analyzer access ports. To keep the environmental
rating of the enclosure, make sure that there is a sealing plug in the access ports that are
not used.
Refer to System configuration options on page 16 to find the correct plumbing
installation.
Figure 6 Electrical connectors and fittings
Installation
1 Analyzer access ports4 Tubing plug
2 Power and data connections (factory installed)5 Sealing plug5.
3 Tubing or cables
3.4Plumbing
D A N G E R
Fire hazard. This product is not designed for use with flammable liquids.
N O T I C E
Do not install reagents until all plumbing is complete.
Make sure to use the specified tubing size.
3.4.1 Sample line guidelines
Select a good, representative sampling point for the best instrument performance. The
sample must be representative of the entire system.
To prevent erratic readings:
•Collect samples from locations that are sufficiently distant from points of chemical
additions to the process stream.
5
Use the sealing plug to close access ports that are not used.
15
Installation
•Make sure that the samples are sufficiently mixed.
•Make sure that all chemical reactions are complete.
3.4.2 Drain line guidelines
Correct installation of the drain lines is important to make sure that all of the liquid is
removed from the instrument. Incorrect installation can cause liquid to go back into the
instrument and cause damage.
•Make the drain lines as short as possible.
•Make sure that the drain lines have a constant slope down.
•Make sure that the drain lines do not have sharp bends and are not pinched.
•Make sure that the drain lines are open to air and are at zero pressure.
3.4.3 Tubing considerations
Plan cable and tubing path to prevent sharp bends and tripping hazards. The analyzer
uses different tubing types for plumbing connections. The type of tubing is based on the
analyzer configuration:
•Ø 3.2 mm: sample line tubing
•Ø 6 mm: unheated drain tubing
•Ø 22 mm: heated drain tubing
Always put the drain tubing so that there is a continuous fall (minimum 3°) and the outlet
is open to air (not pressurized). Make sure the drain tubing is less than 2 meters (6.56 ft).
3.5System configuration options
Before plumbing or electrical installation, find the correct system configuration option
based on the number of analyzers, sample filtration, drain line and number of measured
parameters6. Refer to Table 3.
Table 3 System configuration options
Location FiltrationDrainNumber of
analyzers
OutdoorFILTRAXHeated1 1 Refer to Plumbing one
2 heated2 2 Refer to Plumbing two
Number of
parameters
Plumbing option
outdoor analyzer
on page 18.
outdoor analyzers
on page 19.
6
Refer to Two-parameter configuration on page 29.
16
Table 3 System configuration options (continued)
Installation
Location FiltrationDrainNumber of
analyzers
OutdoorContinuous
sample feed
IndoorFILTRAXUnheated 1 1 Refer to Plumbing one
Heated1 1 Refer to Plumbing one
2 heated2 2 Refer to Plumbing two
Number of
parameters
Plumbing option
outdoor analyzer with
continuous sample feed
on page 21.
outdoor analyzers with
continuous sample feed
on page 22.
indoor analyzer
on page 24.
Unheated 2 2 Refer to Plumbing two
indoor analyzers
on page 25.
17
Installation
Table 3 System configuration options (continued)
Location FiltrationDrainNumber of
analyzers
IndoorContinuous
sample feed
Unheated 1 1 Refer to Plumbing one
Unheated 2 2 Refer to Plumbing two
3.5.1 Plumbing one outdoor analyzer
This system configuration option uses one sc analyzer with the FILTRAX instrument for
the sample line. The waste from the analyzer is released into an open drain through the
optional heated drain hose.
Do the steps that follow to install one outdoor analyzer. Refer to Figure 7.
Number of
parameters
Plumbing option
indoor analyzer with
continuous sample feed
on page 27.
indoor analyzers with
continuous sample feed
on page 27.
1. Install the FILTRAX into the sample stream. Refer to the FILTRAX User Manual for
more information.
2. Use a tubing plug to put the heated sample hose from the FILTRAX through the
analyzer access port.
3. Use a tubing plug to put the heated drain hose through the analyzer access port.
Note: The two samples lines of the heated drain hose are not used.
4. Use the sealing plug to close access ports that are not used.
5. Connect the heated drain connections. Refer to Connect the optional heated drain
on page 34.
6. Connect the heated drain tube to the sample outlet T-fitting.
18