PAT700 Total Organic Carbon Analyzer – Before You Begin Page 1
Chapter 1Before You Begin
1.1About this manual
This instruction manual explains how to install and operate the Anatel® PAT700 Total Organic
Carbon Analyzer.
•This chapter provides an overview of the PAT700 analyzer; explains your responsibility
for reading and following all instructions; explains safety procedures, cautions, and
warnings, which must be adhered to at all times; defines terms that are used
throughout this manual; and tells you how to contact customer service.
•Chapter 2 explains how to install the analyzer.
•Chapter 3 explains analyzer behavior at startup, how to use the lock and unlock icons
to sign on or sign off the analyzer, and how to use the home screen to read Total
Organic Carbon (TOC), conductivity, and temperature.
•Chapter 4 explains how to use the software to monitor the TOC analysis cycle.
•Chapter 5 explains how to use the setup dialog box to navigate and change analyzer
settings.
•Chapter 6 explains how to use the bottles in the sample cartridge to run calibrations,
validations, or system suitability tests; analyze samples from a bottle; install excursion
bottles to capture samples when TOC measurements exceed alarm limits; or schedule
tests to be run.
•Chapter 7 explains how to read, export, and filter historical sample data.
•Chapter 8 explains how to perform tests on the system to verify it is operating correctly.
•Chapter 9 explains how to enter or change passwords that provide data security for the
analyzer.
•Chapter 10 explains troubleshooting and service procedures for the analyzer.
•Appendix A: provides specifications for the analyzer.
1.2Anatel® PAT700 analyzer overview
The Anatel® PAT700 analyzer provides TOC analysis for pure and ultra-pure water
processing.
The PAT700 oxidizes a water sample to determine the TOC in the sample. The analyzer traps
a sample in the analysis cell, exposes the sample to ultraviolet (UV) light, and monitors
changes in temperature and conductivity until the sample has completely oxidized. Once full
oxidation has occurred, the analyzer reports TOC, temperature, conductivity, and the
oxidation curve. The PAT700 incorporates the OASIS™ onboard, automated standards
introduction system that simplifies analyzer performance testing using standards bottles
tagged with RFID technology.
A touch-screen interface provides access to all analyzer functions. The user interface consists
of a home screen from which all user operation initiates. The home screen displays the
current status of the analyzer, TOC level, and environmental data. The software can display
data in a number of user-selected formats, including TOC, compensated and uncompensa ted
conductivity, and temperature in engineering units, with graphical representations of the data.
Compliance with 21 CFR Part 11 is ensured with password protection and an audit trail of all
WGM - 26 July 2007 - Edition 01-d4
user intervention.
AnatelOperator Manual
Page 2 Before You Begin – PAT700 Total Organic Carbon Analyzer
•You can interact with the analyzer remotely, including downloading data records, by
querying the analyzer through an RS-232 communications port.
®
•You can communicate with the analyzer using Modbus
protocol via the Ethernet port.
Alternatively, the user may connect a memory stick to the unit and download the
contents of the data buffer and bottle analysis reports.
•You can print data on a remote printer via the RS-232 serial port.
•Printing may be done automatically or on demand through an RS-232 port to a serial
printer.
1.3RFID technology
The Aantel PAT700 on-line TOC analyzer Onboard Automated Standards Introduction
System (OASIS™) employs Radio Frequency Identification (RFID) technology, a registered
radio frequency device. The RFID system in the PAT700 operates over a very short distance
to eliminate any interference with other wireless communications.
The term RFID describes a system that transmits data wirelessly, using radio waves, over a
very short distance. An RFID system is comprised of a “tag” and a “reader/writer”. In the
PAT700 analyzer, the tag is attached to the standards bottles used in calibrations
(conductivity and TOC), validations, and system suitability tests. The RFID tag consists of a
microchip attached to a radio antenna mounted on a substrate. The RFID tag is attached to
the bottom of the standards bottle. The microchip contains data about the standard contained
in the bottle. The PAT700 OASIS system contains four RFID reader/writers permanently
mounted inside the analyzer that align with the bottle RFID tags when the bottles are fully
loaded in the analyzer. The reader/writer retrieves the data stored on the RFID tags located
on the bottom of the standards bottles. The RFID reader/writers have antennas that emit radio
waves and receive signals back from the tags on the standards bottles. The information
provided from the tag includes the identity of the standard, the standard’s concentration, date
of expiration and other pertinent data.
Figure 1-1 PAT700 with bottle bay open
Operator ManualAnatel
WGM - 26 July 2007 - Edition 01-d4
PAT700 Total Organic Carbon Analyzer – Before You Begin Page 3
The reader/writer takes the data received from the tag and passes the information in digital
form to the PAT700 processor. The RFID system in the PAT700 can read and write to the
RFID tags attached to the standards bottle. The writing feature allows the PAT700 to write
data to the bottles showing that the bottle has been used and provides for writing of data for
the exclusive excursion sampling feature.
The RFID system in the PAT700 does not require any user intervention to operate. The
system operates automatically when bottle tests are performed. The system automatically
turns on and off to read and write data only when necessary. When enabled, the radio
frequency modulation emitted from the antennas is fixed at 13.56 Mhz. There are no user
serviceable parts associated with the RFID system. The RFID reader/writer board assembly
should only be serviced by a Hach Ultra certified service representative.
Since the RFID labels on the standards bottles contain the data necessary for each standard,
the labels must not be removed from the bottle. Without the label, the RFID reader/writer has
no data to read and will not operate as intended.
1.4Reading and following instructions
You must comply with all instructions while you are installing, operating, or maintaining the
analyzer. Failure to comply with the instructions violates standards of design, manufacture,
and intended use of the analyzer. Hach Ultra disclaims all liability for the customer's failure to
comply with the instructions.
•Read instructions — Read all instructions before installing or operating the product.
•Retain instructions — Retain the instructions for future reference.
•Follow instructions — Follow all installation, operating and maintenance instructions.
•Heed warnings and cautions — Adhere to all warnings and caution statements on the
product and in these instructions.
•Parts and accessories — Install only those replacement p arts and acce ssories that are
recommended by Hach Ultra Analytics, Inc. Substitution of parts is hazardous.
1.5Safety
•Read the Anatel PAT700 TOC Analyzer Operator Manual thoroughly before installing
or operating the analyzer. Although the analyzer is designed for rugged use, you
should care for and maintain it as described in this manual. Following proper safety
and handling instructions promotes accident free operation and prolong product life.
•For any questions regarding the analyzer, phone Hach Ultra at 800.866.7889 or
+1 541.472.6500.
•All service procedures should be conducted by properly trained service personnel.
•Follow all procedures in Return procedures, page 109 before shipping the analyzer to
a service center for repair or recalibration.
•Make sure the analyzer is properly installed and all connections are correctly installed
before operating the analyzer. Adhere to all instructions provided in caution and
warning statements.
•Any changes or modifications not expressly approved by Hach Ultra Analytics, Inc.
could void the user's authority to operate this equipment.
WGM - 26 July 2007 - Edition 01-d4
AnatelOperator Manual
Page 4 Before You Begin – PAT700 Total Organic Carbon Analyzer
1.6Caution and warning statements
This manual contains caution and warning statements with which you must comply to prevent
inaccurate measurement, property damage, or personal injury.
CAUTION
Caution statements alert you to hazards or unsafe practices that could result in minor
personal injury or property damage.
Each caution statement explains what you must do to prevent or avoid the potential result of
the specified hazard or unsafe practice.
WARNING
Warning statements alert you to hazards or unsafe practices that could result in
severe property damage or personal injury due to electrical shock, fire, or explosion.
Each warning statement explains what you must do to prevent or avoid the potential result of
the specified hazard or unsafe practice.
Caution and warning statements comply with American Institute of Standards Z535.1-2002
through Z535.5-2002, which set forth voluntary practices regarding the content and
appearance of safety signs, symbols, and labels.
Each caution or warning statement explains:
1)The specific hazard that you must prevent or unsafe practice that you must avoid,
2)The potential result of your failure to prevent the specified hazard or avoid the
unsafe practice, and
3)What you must do to prevent the specified hazardous result.
1.7FCC conformance
The PAT700 contains a registered Radio Frequency device (RFID)
FCC ID: VICPAT700TOC
IC: 6149A-PT700TOC
Japan
Frequency: 13.56 Mhz - +/- 7Khz
RF Output power <180mW
WGM - 26 July 2007 - Edition 01-d4
Operator ManualAnatel
P AT700 Total Organic Carbon Analyzer – Before You Begin Page 5
This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular installation. If
this equipment does cause harmful interference to radio or television reception, which can be
determined by turning the equipment off and on, you should try to correct the in terference by
one or more of the following measures:
•Reorient or relocate the receiving antenna.
•Increase the separation between the equipment and receiver.
•Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
•Consult the dealer or authorized service person for help.
Any changes or modifications not expressly approved by Hach Ultra Analytics, Inc. could void
the user's authority to operate the equipment.
1.8Definitions of terms
The following terms are used throughout this manual:
Conductivity. A measure of the ability to conduct current through water. Conductivity, very
low with high-purity deionized water, is the reciprocal of resistivity and is measured in
microsiemens per centimeter (μS/cm).
Megohm (M). A measurement of 1,000,000 ohms.
Microsiemens (μS). A unit of conductance in the metric system equivalent to one millionth of
-6
an ampere per volt (1 µS = 1 x 10
A/V).
Normalize. To make conform to a standard. In this case, the user performs calculations on
the raw data to produce a result in a standard unit of measure.
Organic. Organic chemicals, including carbon atoms in complex molecules, but not including
carbonate (CO
, HCO3, CO3) or cyanide (CN) related compounds. Oxygen is very common
2
in organics, and they almost always include hydrogen.
Oxidation. The loss of electrons by a chemical species, typically with oxygen. Organic
carbon, for example, oxidizes to carbon dioxide and water.
Parts per billion (ppb). A term used to report trace chemical analyses. It refers to the
concentration of the element or compound present within one billion parts of water. One
microgram per liter (µg/L) is equal to 1 ppb).
Resistivity. Resistance, measured in megohmcentimeters (M-cm), to the flow of electrical
current through high-purity water. Resistivity is a means of continuously measuring the
purity of the water and is the reciprocal of conductivity.
Temperature compensation. Conductivity and resistivity measurements normalized to
25 °C (77 °F) for reporting purposes. Normalized conductivity and resistivity
measurements reflect the values that would be reported at a sample temperature of 25 °C
(77 °F) using a model based on sodium chloride in water. Conductivity and resistivity of
electric current in water depend on the temperature of the water. The lower the
temperature, the lower the conductivity and the higher the resistivity.
WGM - 26 July 2007 - Edition 01-d4
AnatelOperator Manual
Page 6 Before You Begin – PAT700 Total Organic Carbon Analyzer
Total organic carbon (TOC). A measurement of the organics present in water based on its
carbon content. High-purity water measures TOC in parts per billion (ppb). Total organic
carbon is used interchangeably with total oxidizable carbon.
VGA. Video Graphics Adapter.
1.9Customer service
For customer service:
•Voice:US:1.970.663.9760 or 1.800.373.0531
EU:41.22.594.6400
•FAX:US:1.970.663.9761
EU:41.22.594.6488
•Support hot line: US:1.877.4 ANATEL (1.877.426.2835)
•Email:www.hachultra.com
Operator ManualAnatel
WGM - 26 July 2007 - Edition 01-d4
P AT700 Total Organic Carbon Analyzer – Installation Page 7
Chapter 2Installation
2.1Installation requirements
Install the analyzer in a dry, dust-free environment. The analyzer has a splash-resistant
NEMA 12-compliant water resistant IP56 enclosure surrounding the electronics portion of the
analyzer. All components in the hydraulics portion of the analyzer are rated IP56.
•Install the analyzer in a properly ventilated location where ambient air temperature
does not exceed 35 °C (95 °F).
•The analyzer has 3/4-inch NPT conduit fittings or quick-connect fittings for wiring.
See Figure 2-1, page 8 for mounting dimensions. The analyzer weighs 27 lb (12.25 kg).
The analyzer ship kit includes the following parts:
•This instruction manual
•5-foot (1.5 meter) PFA inlet tubing
•10-foot (3 meter) polypropylene outlet tubing
•Allen wrench
•Stylus
•Instrument screwdriver
•RFID tagged excursion sample bottle
For units with quick connect I/O fittings the following items are also included:
•Quick connect power cord
•Male 8-pin quick connect cord fitting
•Three male 5-pin quick connect cord fitting
2.2Orientation and mounting
CAUTION
Failure to mount the analyzer in an upright position can cause measurement error.
To ensure proper operation of the measurement cell between analysis cycles, mount the
analyzer in an upright position. See Figure 2-2.
•Secure the analyzer using the factory-supplied mounting bracket (see Figure 2-3,
page 10). Use the mounting holes in the bracket to mount it to a flat, stable surface.
•Hook the analyzer onto the mounting bracket and latch it into place using the twist
latches.
•An optional handle, attached to the analyzer, enables it to be moved from place to
place.
WGM - 26 July 2007 - Edition 01-d4
AnatelOperator Manual
Page 8 Installation – PAT700 Total Organic Carbon Analyzer
Figure 2-1 PAT700 mounting dimensions
Operator ManualAnatel
WGM - 26 July 2007 - Edition 01-d4
P AT700 Total Organic Carbon Analyzer – Installation Page 9
Figure 2-2 PAT700 analyzer orientation
Mounting general considerations
Follow these guidelines when installing the analyzer:
•Locate the analyzer where it is accessible for operation, service and calibration.
•Minimize the distance between the water system sample point and the analyzer to
allow for representative sampling.
•Locate the analyzer where the ambient temperature remains between 15 and 35 °C
(59 and 95 °F).
•The transmitter mounting bracket is installed independently from the analyzer. Once
mounted, the analyzer hooks on the mounting bracket and locks into place with twist
latches.
•Mount the transmitter mounting bracket to a stable, flat surface or dual instrument
poles.
•The front of the analyzer requires 12 inches (30.48 cm) of clearance to allow the doors
to open completely.
•The analyzer is available with ¾-inch conduit openings or quick-connect fittings for
power and I/O wiring. If conduit is used, install fittings that ensure a complete seal with
the openings and are properly sealed to keep the analyzer electronics compartment
watertight.
Mounting to a wall
Follow these guidelines and refer to Figure 2-1 and Figure 2-3 to mount the transmitter to a
wall or other flat, rigid surface:
•Use 5/16-inch (8 mm) diameter screws or bolts and nuts to mount the analyzer to a
wall or other flat, rigid surface. Use hardware that can withstand the process
environment. Hach Ultra does not supply hardware.
•To minimize stress on the analyzer mounting bracket, secure all mounting bolts to the
same structure, which should be flat and should not vibrate or move excessively. Do
not secure bolts to separate girders, beams, or wall studs, which can move
independently.
WGM - 26 July 2007 - Edition 01-d4
AnatelOperator Manual
Page 10 Installation – PAT700 Total Organic Carbon Analyzer
Figure 2-3 PAT700 wall mount
Mounting to dual instrument poles
Follow these guidelines and refer to Figure 2-1 and Figure 2-4 to mount the transmitter to
instrument poles:
•Mount the analyzer mounting bracket to two instrument poles that are attached to a
common surface. Do not secure to two instrument poles that could move
independently.
•Use four 5/16-inch U-bolts, two for each 2-inch pipe, and eight matching nut s, to mount
the analyzer mounting bracket to two rigid instrument poles. Use U-bolts and nuts that
can withstand the process environment. Hach Ultra does not supply U-bolts or nuts.
•The instrument poles should extend at least 10 inches (254 mm) from a common rigid
base and should be no larger than 2 inches (50.8 mm) in diameter.
Operator ManualAnatel
WGM - 26 July 2007 - Edition 01-d4
P AT700 Total Organic Carbon Analyzer – Installation Page 11
Figure 2-4 PAT700 instrument-pole mount
2.3Plumbing connections
CAUTION
Improper plumbing connections to the analyzer can cause improper flow through the
analyzer, resulting in measurement error.
To avoid measurement error due to improper flow if temperature and pressure exceed the
limits listed in Table 2-1, page 12, install high grade 1/4” (OD) PTFE, FEP, PVDF, or 316
stainless steel tubing. Otherwise, install the factory-supplied 1/ 4” PFA (perfluoroalkoxy resin)
sample tubing and 1/4” OD polypropylene drain tubing. Hach Ultra supplies 5 feet (1.5 meters)
of PFA tubing and 10 feet (3 meters) of polypropylene tubing with the analyzer.
WGM - 26 July 2007 - Edition 01-d4
AnatelOperator Manual
Page 12 Installation – PAT700 Total Organic Carbon Analyzer
Isolation valve
To enable manual isolation of the analyzer from input flow, connect the analyzer to the sample
supply through a customer-supplied upstream isolation valve. See Figure 2-5.
Table 2-1 Maximum temperature and pressure limits for PFA tubing
Maximum water temperatureMaximum pressure
75 °C (167 °F)90 psig (620 kPa)
85 °C (185 °F)80 psig (550 kPa)
90 °C (194 °F)70 psig (480 kPa)
Figure 2-5 PAT700 connection to isolation valve
Operator ManualAnatel
WGM - 26 July 2007 - Edition 01-d4
P AT700 Total Organic Carbon Analyzer – Installation Page 13
Water inlet and outlet
Tubing connects to the 1/4” inlet (WATER IN) and outlet (WATER OUT) 316 stainless steel
compression fittings on the analyzer. See Figure 2-6. Installation requires a 7/16” wrench.
A
B
AWater inlet
BWater outlet
Figure 2-6 PAT700 water inlet and outlet connections
To connect plumbing to the 1/4” inlet (WATER IN) port, follow these steps:
1)To avoid introducing debris through the inlet plumbing, flush the isolation valve by
opening and closing it fully several times.
2)T aking care not to crimp or da mage the tubing, push one end of the inlet tubing into
the water inlet port until it cannot be inserted any farther.
3)Taking care not to pull on the tubing, hand-tighten the compression nut.
4)Mark both the compression nut and the WATER IN connector as a reference for
tightening the nut.
5)Tighten the compression nut 1-1/4 turns to secure the connection.
WGM - 26 July 2007 - Edition 01-d4
AnatelOperator Manual
Page 14 Installation – PAT700 Total Organic Carbon Analyzer
CAUTION
Over tightening fittings can damage the ferules, causing leaks that can result in
measurement error or property damage.
After tightening a ferule 1-1/4 turns, do not tighten it more than another 1/4 turn to seal the
connection.
To connect plumbing to the 1/4” outlet (WATER OUT) port, follow these steps:
1)Noting the flow direction, attach the 10-foot long, 1/4” OD polypropylene drain
tubing to the WATER OUT port of the analyzer.
2)Taking care not to crimp or damage the tubing, push one end of the tubing into the
water outlet port until it cannot be inserted any farther.
3)Taking care not to pull on the tubing, hand-tighten the compression nut.
4)Mark both the compression nut and the WATER OUT connector as a reference for
tightening the nut.
5)Tighten the compression nut 1-1/4 turns to secure the connection.
6)Leak test the connections by slowly opening the upstream isolation valve to
introduce water into the analyzer. Pulse the valve several times by opening and
closing it, then recheck the fittings.
7)If necessary, slowly tighten the compression fittings to stop any leaks.
2.4Wiring connections for PAT700 with conduit openings
Refer to this section if the PAT700 has three 3/4-inch female NPT conduit openings.
•One opening accommodates power supply wiring.
•The other two openings accommodate 4-20 mA or discrete I/O wiring.
•Power supply and I/O wiring connects to terminals in the compartment behind the
analyzer’s display unit. Figure 2-8 illustrates power supply and I/O wiring terminals. To
access the terminals:
1)Use the factory-supplied Allen wrench to unlatch the door that contains the display.
2)Unlatch the door and swing it open on its hinges.
Operator ManualAnatel
WGM - 26 July 2007 - Edition 01-d4
P AT700 Total Organic Carbon Analyzer – Installation Page 15
Figure 2-7 Using Allen wrench to access wiring connections
Figure 2-8 PAT700 power supply and I/O wiring terminals
WGM - 26 July 2007 - Edition 01-d4
AnatelOperator Manual
Page 16 Installation – PAT700 Total Organic Carbon Analyzer
Power supply wiring
WARNING
Improper grounding can cause property damage or personal injury.
• Adhere to ground network requirements for the facility.
• Maintain all exposed conductors at earth ground.
Figure 2-9 illustrates power supply wiring terminals.
Table 2-2 lists specifications for power supply wiring.
Figure 2-9 PAT700 power supply wiring terminals
Table 2-2 Power supply wiring specifications
DescriptionSpecification
Voltage input100 to 240 VAC universal
Frequency50 to 60 Hz
Power900 Watts maximum, 1.25 A @ 250 VAC
I/O wiring
The analyzer has three 4-20 mA outputs, two discrete inputs, and four discrete outputs.
Connect wiring to the Y-shaped connector blocks located on the I/O circuit board.
•The connector blocks are a 2-part assembly.
•The terminal block connector can be unplugged from the analyzer for easier
installation of wiring.
2
•Install twisted-pair unshielded wiring, 18 to 14 AWG (1.0 to 2.5 mm
).
Operator ManualAnatel
WGM - 26 July 2007 - Edition 01-d4
P AT700 Total Organic Carbon Analyzer – Installation Page 17
4-20 mA outputs
The analyzer has three 4-20 mA outputs. Wiring connects to terminals on connector block
J17, as listed in Table 2-3.
•Analog output 2 represents conductivity or resistivity (user selected).
•Analog output 3 represents temperature in °F or °C (user selected).
Table 2-3
J17 BlockTerminalDescriptionVariableDesignation
J17 connector block: 4-20 mA output wiring terminal designations
14-20 mA source output for analog output 1 (+)TOCAO1+
24-20 mA sink output for analog output 1 (–)TOCAO134-20 mA source output for analog output 2 (+)Conductivity or resistivityAO2+
44-20 mA sink output for analog output 2 (–)Conductivity or resistivityAO254-20 mA source output for analog output 3 (+)Temperature AO3+
64-20 mA sink output for analog output 3 (–)Temperature AO3-
Discrete inputs
The analyzer has two discrete inputs. Wiring connects to terminals on connector block J24, as
listed in Table 2-4.
•Discrete input 1 initiates TOC analysis. When the output state switches from high to
low, the ana lyzer stops the cu rrent operation and runs a single T O C analysis. Aft er the
analysis is complete, the analyzer returns to the mode defined by digital in put 2. During
the TOC analysis, the analyzer ignores all subsequent trigger values.
•Discrete input 2 switches the analyzer modes. If digital output 2 is in high state, the
analyzer is in TOC mode. If digit al output 2 is in lo w state, the analyzer runs in flow with
conductivity mode.
1Common connection for external sourcing (+) CommonDI1/2+
2Digital input 1 voltage source return input (–)Initiates TOC analysisDI1–
3Digital input 2 voltage source return input (–)Analyzer modeDI2–
4• 12 VDC power supply +
• Connect only if output device requires
power supply
5• 12 VDC power supply –
• Connect only if output device requires
power supply
WGM - 26 July 2007 - Edition 01-d4
Output power supply+12V
Output power supplyGND
AnatelOperator Manual
Page 18 Installation – PAT700 Total Organic Carbon Analyzer
Discrete outputs
The analyzer has four discrete outputs. Wiring connects to terminals on connector block J22,
as listed in Table 2-5.
•Discrete output 1 is a TOC alarm. The output reports the TOC level as above (low
state) or below (high state) the user-specified alarm limit.
•Discrete output 2 is an uncompensated conductivity alarm. A low state indicates the
conductivity level is above the alarm limit. A high state indicates the conductivity level
is below the alarm limit.
•Discrete output 3 reports a warning or failure condition. A low state indicates an error
condition exists. A high state indicates no error conditions exist.
•Discrete output 4 reports a TOC analysis start (by indicating the state of the sample
inlet valve). A low state indicates the sample valve is closed. A high state indicates the
sample valve is open.
14-20 mA source output for analog output 1 (+)TOCAO1+
24-20 mA sink output for analog output 1 (–)TOCAO1–
34-20 mA source output for analog output 2 (+)Conductivity or
resistivity
44-20 mA sink output for analog output 2 (–)Conductivity or
resistivity
54-20 mA source output for analog output 3 (+)Temperature AO3+
64-20 mA sink output for analog output 3 (–)Temperature AO3–
AO2+
AO2–
Discrete inputs
The analyzer has two discrete inputs, as listed in Table 2-8.
•Discrete input 1 initiates TOC analysis. When the output state switches from high to
low, the ana lyzer stops the cu rrent operation and runs a single T O C analysis. Aft er the
analysis is complete, the analyzer returns to the mode defined by digital in put 2. During
the TOC analysis, the analyzer ignores all subsequent trigger values.
•Discrete input 2 switches the analyzer modes. If digital output 2 is in high state, the
analyzer is in TOC mode. If digit al output 2 is in lo w state, the analyzer runs in flow with
conductivity mode.