Hach PAT700 User Manual

DOC026.53.80276
PAT700
02/2014, Edition 2
User Manual
Table of Contents
Specifications..............................................................................................................5
Safety information........................................................................................................6
Use of hazard information
............................................................................................ 6
Precautionary labels....................................................................................................6
Certification..................................................................................................................7
FCC conformance for RFID..................................................................................7
Product overview.........................................................................................................8
Product components....................................................................................................8
Installation.....................................................................................................................9
Installation guidelines..................................................................................................9
Mechanical installation............................................................................................... 10
Attach the instrument to a wall...........................................................................10
Attach the instrument to dual pipes....................................................................12
Electrical installation..................................................................................................13
Wiring safety information....................................................................................13
About quick-connect wiring connections............................................................13
Connect the power cord..............................................................................13
Connect the 4–20 mA output....................................................................... 13
Connect the discrete inputs.........................................................................14
Connect the discrete outputs....................................................................... 14
Connect a 40-column printer.......................................................................15
About conduit wiring connections.......................................................................17
Wiring for power..........................................................................................18
Connect to power........................................................................................18
Plumbing....................................................................................................................19
Plumb the instrument.......................................................................................... 19
User interface and navigation............................................................................20
Sliding toolbar............................................................................................................21
Setup menu...............................................................................................................22
Startup........................................................................................................................... 23
Connect to power....................................................................................................... 23
Set the sample flow rate............................................................................................23
Basic operation.........................................................................................................23
Select the run mode................................................................................................... 23
Calibration.................................................................................................................. 24
TOC calibration................................................................................................... 24
Conductivity calibration....................................................................................... 24
Do a system suitability test........................................................................................25
Measure a grab sample ............................................................................................26
Manage measurement data.......................................................................................26
Set the data filter................................................................................................26
Export or print data.............................................................................................27
Advanced operation................................................................................................27
TOC operating mode configuration............................................................................ 27
1
Configure the general settings............................................................................ 27
Configure the flow during idle mode...................................................................28
Configure digital control......................................................................................28
Enable the stabilization process.........................................................................29
Set the TOC offset..............................................................................................29
System configuration.................................................................................................29
Set the analyzer name, location and language..................................................29
Set the date, time and units................................................................................30
Configure the alarm sounds...............................................................................30
Configure the network settings...........................................................................30
Configure the backlight settings.........................................................................31
Configure the digital output alarms............................................................................31
Conductivity alarm limit
....................................................................................... 32
Configure the analog outputs..................................................................................... 33
Configure the printer settings..................................................................................... 33
Security configuration................................................................................................34
Enable security, data logging and encryption.....................................................34
Set password expiration and automatic log off...................................................34
Add, delete or change users............................................................................... 34
About backdoor passwords................................................................................35
Show the audit trail.............................................................................................35
Configure clean-in-place (CIP) options......................................................................36
Configure the auto sampler.......................................................................................36
Bottle mode configuration..........................................................................................37
Configure the TOC calibration settings............................................................... 37
Configure the conductivity calibration settings...................................................37
Set the analyzer to return to online mode........................................................... 38
Bottle mode operation................................................................................................ 38
Validate the TOC calibration............................................................................... 38
Do an auto sampler test.....................................................................................39
Get an excursion sample....................................................................................39
Flush the sample bottle lines..............................................................................39
Maintenance...............................................................................................................40
Maintenance schedule...............................................................................................40
Clean the instrument.................................................................................................. 40
Clean leaks and spills................................................................................................40
Replace a fuse...........................................................................................................41
Drain the instrument..................................................................................................41
Send the instrument for service.................................................................................42
Troubleshooting.......................................................................................................42
Alarms........................................................................................................................ 42
Alarm indication..................................................................................................42
Accept or remove an alarm ...............................................................................43
Error codes..................................................................................................43
Critical error codes......................................................................................44
Warning codes............................................................................................44
Measurement alarm codes..........................................................................46
Diagnostics................................................................................................................46
Table of Contents
2
Review the data log usage and firmware version...............................................46
Review the UV lamp information........................................................................46
Do a lamp test....................................................................................................47
Change to the other UV lamp.............................................................................47
Select a diagnostic test
....................................................................................... 47
Validate the RS232 connection...................................................................47
Validate the digital I/O communication........................................................47
Validate the 4-20 mA output........................................................................48
Do a printer test...........................................................................................48
Do a plumbing test......................................................................................48
Validate the pump operation.......................................................................49
Do an RFID test........................................................................................... 49
Do a fan test................................................................................................49
Calibrate the display....................................................................................50
See the calibration dates....................................................................................50
Technical support......................................................................................................50
Appendix......................................................................................................................50
Theory of operation.................................................................................................... 50
Analyzer startup sequence........................................................................................52
Run modes................................................................................................................52
The analysis cycle..............................................................................................52
Online TOC mode............................................................................................... 52
Conductivity mode..............................................................................................52
CIP Validation mode...........................................................................................52
Standby mode....................................................................................................53
Offline mode.......................................................................................................53
Manual TOC sample........................................................................................... 53
Self clean mode..................................................................................................53
Bottle mode................................................................................................................ 53
Calibration and validation...................................................................................53
Onboard Automated Standards Introduction System (OASIS)........................... 53
RFID technology.................................................................................................53
Data log.....................................................................................................................54
Default settings..........................................................................................................54
Replacement parts and accessories...............................................................57
Table of Contents
3
Table of Contents
4
Specifications
Specifications are subject to change without notice.
Specification Details
Dimensions (W x D x H) 59.7 x 22.9 x 25.4 cm (23.5 x 9 x 10 in.)
Enclosure Quick-connect wiring model: IP46
Conduit wiring model: IP56
Weight 13.6 kg (30 lb)
Pollution degree 2
Installation category II
Protection class I
Power requirements 100/230 VAC, ±10%, 65 W, 50/60 Hz
Altitude 4000 m (13,125 ft) maximum
Operating temperature 10 to 40 °C (50 to 104 °F)
Humidity 5–95%, non-condensing
Sample temperature 1 to 95 °C (34 to 203 °F)
Sample flow rate 60 to 300 mL/min
Sample pressure 0.70 to 7.0 kg/cm² (10 to 100 psi)
Temperature accuracy ±0.5 °C
UV lamps 2x, with UV Detect technology
User interface Color touch screen
TOC Range: 0.5 to 2000 ppb as carbon
Accuracy: ±1 ppb or ±5%, whichever is greater
Repeatability: ±0.3 ppb or ±1%, whichever is greater
Detection limit: 0.5 ppb
Conductivity Range for conductivity mode: 0.05 to 150 μS/cm at 25 °C
Accuracy: ±2% over the full range
Range for TOC mode: 0.05–5.0 µS/cm
Outputs Analog: Three isolated
Discrete: Four isolated, configurable
Inputs Discrete: Two isolated
Certifications CE Certified. Listed to UL and CSA safety standards by ETL.
Warranty 1 year
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website.
English 5
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
C A U T I O N
Indicates a potentially hazardous situation that may result in minor or moderate injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end­of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
This symbol indicates that a risk of electrical shock and/or electrocution exists.
This symbol indicates the need for protective eye wear.
This symbol indicates that the marked item requires a protective earth connection. If the instrument is not supplied with a ground plug on a cord, make the protective earth connection to the protective conductor terminal.
6 English
This symbol, when noted on the product, identifies the location of a fuse or current limiting device.
This symbol indicates the presence of mercury. Components that contain mercury must be disposed in accordance with all applicable local and regional law.
This symbol indicates a puncture or pinch hazard. Keep hands and fingers away.
This symbol indicates the presence of a UV light source that may cause eye and skin injury. Wear appropriate protective equipment and obey all safety protocols.
Certification
Canadian Radio Interference-Causing Equipment Regulation, IECS-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including interference that may cause
undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:
1. Disconnect the equipment from its power source to verify that it is or is not the source of the interference.
2. If the equipment is connected to the same outlet as the device experiencing interference, connect the equipment to a different outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the interference.
5. Try combinations of the above.
FCC conformance for RFID
This instrument may contain a registered radio frequency identification device (RFID). Refer to
Table 1.
English 7
Table 1 Registration information
Parameter Value
FCC identification number (FCC ID) VICPAT700TOC
IC 6149A-PT700TOC
For Japan
Frequency 13.56 MHz to ±7 kHz
RF output power <180 mW
Product overview
D A N G E R
Chemical or biological hazards. If this instrument is used to monitor a treatment process and/or chemical feed system for which there are regulatory limits and monitoring requirements related to public health, public safety, food or beverage manufacture or processing, it is the responsibility of the user of this instrument to know and abide by any applicable regulation and to have sufficient and appropriate mechanisms in place for compliance with applicable regulations in the event of malfunction of the instrument.
This analyzer uses UV light to oxidize water samples for TOC analysis in pure and ultra-pure water process applications. The analyzer is capable of compliance to 21 CFR Part 11 with password protection and an audit trail of all user actions.
For power and data collection, the analyzer connects to the facility systems through quick-connect wiring connections. Some analyzers have conduit openings instead of quick-connect wiring connections with internal connector blocks for conduit wiring connections.
The analyzer has an OASIS bottle bay to measure individual grab samples or do a calibration/validation. Bottle information is read automatically with the radio frequency identification device (RFID) wireless system and RFID tags on the bottles*.
Product components
Make sure that all components have been received. Refer to Figure 1. If any items are missing or damaged, contact the manufacturer or a sales representative immediately.
*
Not all models include the RFID wireless system
8 English
Figure 1 Instrument components
1 Wall mount bracket 5 Power cord 9 Connector, I/O, 5-pin (3x)
2 PAT700 6 Connector, I/O, 6-pin (1x) 10 Screwdriver, 2-mm wide blade
3 Tubing, 10 ft 7 USB flash drive 11 Stylus for touchscreen
4 Tubing, 5 ft 8 Graduated cylinder, 50-mL 12 Hex wrench, T-handle
Installation
Installation guidelines
D A N G E R
Fire hazard. This product is not designed for use with flammable liquids.
D A N G E R
Electrocution hazard. If this equipment is used outdoors or in potentially wet locations, a Ground Fault Circuit Interrupt (GFCI/GFI) device must be used for connecting the equipment to its main power source.
W A R N I N G
Electrical shock hazard. Externally connected equipment must have an applicable country safety standard assessment.
W A R N I N G
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
English 9
N O T I C E
To prevent measurement error, make sure that the instrument is attached in an upright, level position.
For environmental and sample requirements, refer to Specifications on page 5
.
• Put the instrument in a location that has access for operation, service and calibration.
• Install the instrument near the water system sample point.
• Make sure that there is a minimum of 30.48 cm (12 in.) of clearance for the enclosure doors to open. Refer to Figure 2 on page 11.
• Make sure that there is sufficient clearance to access the power switch, plumbing and electrical connections on the end panels. Refer to Figure 2 on page 11.
Mechanical installation
Attach the instrument to a wall
W A R N I N G
Personal injury hazard. Make sure that the wall mounting is able to hold 4 times the weight of the equipment.
Install the instrument on a wall or other stable, flat surface. Refer to Figure 2 for dimensions. Refer to
Figure 3 for installation.
Use user-supplied 8-mm (5/16-in.) diameter screws or bolts and nuts to attach the instrument to a wall or other flat, rigid surface. Use hardware that is approved for a process environment.
• Do not attach bolts to separate girders, beams or wall studs that can move independently.
• Do not attach the instrument to a surface that has vibration.
10 English
Figure 2 Installation dimensions
English 11
Figure 3 Wall mount
Attach the instrument to dual pipes
Refer to Figure 2 on page 11 for dimensions. Refer to Figure 4 for installation.
• Attach the analyzer mounting bracket to two instrument pipes that are attached to a common surface. Do not attach the instrument to two instrument pipes that could move independently.
• Use pipes that extend at least 254 mm (10 in.) from the same rigid base.
• Use pipes with a diameter of 50.8 mm (2 in.) or less.
• Use four user-supplied 5/16-inch U-bolts (two for each 2-inch pipe) and eight user-supplied matching nuts. Use U-bolts and nuts that can withstand the process environment.
• Use U-bolts and nuts that do not extend more than 15 mm (0.60 in.) from the inside of the bracket.
Figure 4 Pipe installation
12 English
Electrical installation
Wiring safety information
D A N G E R
Electrocution hazard. Always remove power to the instrument before making electrical connections.
About quick-connect wiring connections
D A N G E R
Electrocution and fire hazards. For instruments with a power cord, make sure that there is easy access to the local power disconnect.
Figure 5 shows the quick-connect connections.
Figure 5 Power supply and wire connections
1 Power switch 5 RS232 connections
2 Power cord connection 6 Network connection
3 Digital output 7 Digital input
4 USB port 8 4–20 mA output connection
Connect the power cord
1. Connect the power cord to the instrument. Refer to Figure 5 on page
13, item 2.
2. Tighten the compression fitting by hand.
Connect the 4–20 mA output
Table 2 shows the terminal designations for the 4–20 mA quick-connect connector. The default
parameter represented on each 4–20 mA output is shown in Table 2. To change the default parameter, refer to Configure the analog outputs on page 33.
English 13
Table 2 4–20 mA terminal designations
Connector Terminal Description Variable Designation
1 4–20 mA source output for analog output
1 (+)
TOC AO1+
2 4–20 mA sink output for analog output 1 (–) TOC AO1–
3 4–20 mA source output for analog output
2 (+)
Conductivity or resistivity AO2+
4 4–20 mA sink output for analog output 2 (–) Conductivity or resistivity AO2–
5 4–20 mA source output for analog output
3 (+)
Temperature AO3+
6 4–20 mA sink output for analog output 3 (–) Temperature AO3–
7
8
Connect the discrete inputs
Table 3 shows the discrete input designations for the quick-connect connector.
Table 3 Discrete input designations
Connector Terminal Description Variable Designation
1 Common connection for external
source (+)
Common DI1/2+
2 Digital input 1 (–) Start a single TOC analysis DI1–
3 Digital input 2 (–) TOC or conductivity mode
selection
DI2–
4
1
12 VDC power supply + Output power supply +12 V
5 12 VDC power supply – Output power supply GND
1
Connect terminals 4 and 5 only if the input device requires a power supply.
Connect the discrete outputs
The terminal designations for the discrete output quick connections are shown in Table 4 and
Table 5. The terminal designations for the discrete output connector block is shown in Table 6.
14 English
Table 4 Quick connect – Discrete output 1 and 2
Wiring connector 1-2 Terminal Description Default value Designation
1 Digital output 1 (+) TOC alarm (configurable) DO1+
2 Digital output 2 (+) Conductivity alarm (configurable) DO2+
3 Common return (–) Digital output 1 and 2 common
return
DO1/2–
4 12 VDC power supply + Input power supply +12 V
5 12 VDC power supply – Input power supply GND
Table 5 Quick connect – Discrete output 3 and 4
Wiring connector 3-4 Terminal Description Default value Designation
1 Digital output 3 (+) Error alarm (configurable) DO3+
2 Digital output 4 (+) Valve state (TOC analysis start,
configurable)
DO4+
3 Common return (–) Digital output 3 and 4 common
return
DO3/4–
4 12 VDC power supply + Input power supply +12 V
5 12 VDC power supply – Input power supply GND
Table 6 Connector block – Discrete outputs
Wiring connector block Terminal Description Default value Designation
1 Digital output 1 (+) TOC alarm (configurable) DO1+
2 Digital output 2 (+) Conductivity alarm (configurable) DO2+
3 Common return (–) Digital output 1 and 2 common
return
DO1/2 –
4 Digital output 3 (+) Error alarm (configurable) DO3+
5 Digital output 4 (+) Valve state (TOC analysis start,
configurable)
DO4+
6 Common return (–) Digital output 3 and 4 common
return
DO3/4 –
7 12 VDC power supply + Input power supply +12 V
8 12 VDC power supply – Input power supply GND
Connect a 40-column printer
Connect a printer to the 9-pin D-subminiature serial port as necessary. This RS232 interface communicates at 9600 baud, 8 data bits, 1 stop bit and no parity. A 40-column printer is necessary for the printout report. Refer to Configure the printer settings on page
33 to change the default
printer settings.
If the printer is purchased from the manufacturer of the analyzer, the DIP switches on the printer are set at the factory. If a Seiko Instruments DPU-414 Type II thermal printer is installed, set the DIP switches. Refer to Set the printer DIP switches on page 16.
English 15
Set the printer DIP switches
1. Set the printer power switch to OFF.
2. Push and hold ONLINE, then switch the power switch to ON. Release the button when a printout
of the current settings starts.
3. Push ONLINE again. "DIP SW1" shows on the printout. This is a prompt to change DIP switch
1 (switches 1–8).
4. Enter the settings for DIP switch 1 (switches 1–8) in succession. Push ONLINE for on. Push
FEED for off. Refer to Table 7, Table 8
and Table 9.
The ONLINE light comes on to identify an ON entry. The OFFLINE light goes off to identify an OFF entry. Make sure to enter a setting for all eight switches. The setup mode cannot be cancelled after it is started.
5. After all eight switches have been set for DIP switch 1, a printout prompt asks whether to
continue or repeat the procedure for DIP switches 2 and 3 or save the current settings.
After the settings have been saved, a printout is made and the printer goes back to normal operation. Do not turn the printer off while it is saving settings. Wait until “Dip SW setting complete!” is on the printout before the printer power is set to off.
Table 7 Printer DIP switch 1 settings
Switch Setting Function
1 Off Serial input
2 On High print speed
3 On Auto loading enabled
4 Off Auto linefeed disabled
5 On Enabled setting
6 Off Print density
7 On Print density
8 On Print density
Table 8 Printer DIP switch 2 settings
Switch Setting Function
1 On 40-column printing
2 On Backup font enabled
3 On Normal characters
4 Off Zero = Slash
5 On International character set = USA
6 On International character set = USA
7 On International character set = USA
8 Off International character set = USA
Table 9 Printer DIP switch 3 settings
Switch Setting Function
1 On 8 data bits
2 Off Parity setting = Yes
16 English
Table 9 Printer DIP switch 3 settings (continued)
Switch Setting Function
3 On Parity condition = Odd
4 Off Flow control (X on/X off)
5 Off 9600 baud rate
6 On 9600 baud rate
7 On 9600 baud rate
8 On 9600 baud rate
About conduit wiring connections
Some analyzers have three ¾-inch female NPT conduit openings instead of quick-connect wiring connections. These analyzer have internal connector blocks for power, three 4–20 mA outputs, two discrete inputs and three discrete outputs. Refer to Figure 6.
The connector blocks are located on the I/O circuit board behind the user interface. Use an Allen wrench to open the compartment.
Use one conduit opening for power supply wiring. Use the other two conduit openings for 4-20 mA and/or discrete I/O wiring. Install twisted-pair unshielded wiring, 18 to 14 AWG (1.0 to 2.5 mm²) for all wiring except the power supply wiring.
For the connector block terminal designations, refer to Table 2 on page
14, Table 3 on page 14 and
Table 6 on page 15.
Note: The connector blocks can be unplugged from the analyzer for easier installation of wiring.
Figure 6 Conduit wiring connections
1 Power supply wiring enclosure 4 Discrete output connector block (J22)
2 4-20 mA output connector block (J17) 5 I/O circuit board
3 Discrete input connector block (J24)
English 17
Wiring for power
D A N G E R
Electrocution hazard. Protective Earth Ground (PE) connection is required.
D A N G E R
Electrical shock and fire hazards. Make sure to identify the local disconnect clearly for the conduit installation.
W A R N I N G
Potential Electrocution Hazard. If this equipment is used outdoors or in potentially wet locations, a Ground Fault Interrupt
device must be used for connecting the equipment to its mains power source.
N O T I C E
Install the device in a location and position that gives easy access to the disconnect device and its operation.
Supply power to the instrument with conduit or a power cable. Make sure that a circuit breaker with sufficient current capacity is installed in the power line. The circuit breaker size is based on the wire gauge used for installation.
For installation with conduit:
Install a local disconnect for the instrument within 3 m (10 ft) of the instrument. Put a label on the disconnect that identifies it as the main disconnect device for the instrument.
• Make sure that the power and safety ground service drops for the instrument are 18–12 AWG and the wire insulation is rated for 300 VAC or higher.
• Connect equipment in accordance with local, state or national electrical codes.
• If metal conduit is used, make sure that the conduit hub is tightened so that metal conduit is securely bonded to the safety ground wire through the conduit backing plate.
For installation with a power cable, make sure that the power cable is:
• Less than 3 m (10 ft) in length
• Rated for at least 60 °C (140 °F) and applicable to the installation environment
• Rated for the power requirement specifications. Refer to Specifications on page 5.
• 14–10 AWG with applicable insulation colors for local code requirements
• A power cable with a three-prong plug (with ground connection) that is applicable to the supply connection
• Does not have a locking type device on the plug
Connect to power
Connect power with conduit or a power cable. Refer to Figure 7 and Table 10. An AC power cable is not supplied with the analyzer.
After power supply wiring has been installed, make sure that the field wiring box cover is installed and the screw is tightened with a screwdriver so that the cover is correctly bonded to protective earth.
After the wiring is completed, tighten the conduit fittings. Make sure that the door is tightly closed so that the safety protections and the enclosure environmental ratings are maintained.
18 English
Figure 7 Wiring for power
1 Line 2 Neutral 3 Ground
Table 10 AC wiring information
Connection Color—North America Color—EU
Hot (L) Black Brown
Neutral (N) White Blue
Protective earth ground Green Green and yellow
Plumbing
Plumb the instrument
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
Refer to Figure 8 and the steps in this procedure to plumb the instrument.
User-supplied items:
Wrench, 7/16-in.
• Isolation valve
• Tubing, ¼-in. OD polypropylene, PFA or 316 stainless steel for inlet and outlet
1. Install the isolation valve on the process pipe. Open and close the isolation valve several times to let water into the system to remove debris.
2. Push the tube from the isolation valve into the WATER IN port until it stops. Do not crimp or bend the tubing.
3. Tighten the compression nut on the WATER IN port by hand. Make a mark on the compression nut and tube for reference.
Note: For the dual stream model, do step 2 and 3 for the second WATER IN port.
4. Tighten the compression nut another 1¼ turns.
English 19
5. Push one end of a 10-ft tube into the WATER OUT port until it stops. Do not crimp or bend the tubing.
6. Tighten the compression nut on the water outlet side by hand. Make a mark on the compression nut and tube for reference.
7. Tighten the compression nut another 1¼ turns.
8. Do a leak test:
a. Slowly open the isolation valve to let water into the instrument. b. Open and close the valve several times to pulse water through the connections. c. Examine the compression fittings for leaks. d. If necessary, slowly tighten the compression fittings to stop leaks.
Figure 8 Plumb the instrument
1 Process pipe 4 Drain tubing, ¼-in.OD
2 Isolation valve, customer-supplied 5 Drain
3 Water inlet tubing, ¼-in. OD
User interface and navigation
After startup, the user interface shows the home screen. Refer to Figure 9.
20 English
Figure 9 Home screen
1 Run mode 5 Lamp status
2 Current date and time 6 Bottle mode access
3 Sliding toolbar 7 Data section
4 Animated process graphic 8 View tabs
Sliding toolbar
Use the sliding toolbar to open the main functions of the instrument. Table 11 shows the options on the sliding toolbar.
To open the sliding toolbar, select the << icon in the top left of the measurement screen.
Table 11 Sliding toolbar options
Icon Function Description
Run mode Set or change the operation mode.
Setup Show the Setup menu.
Bottle mode Start a calibration, validation or grab sample.
Alarm Review and acknowledge alarms.
English 21
Table 11 Sliding toolbar options (continued)
Icon Function Description
Diagnostics General—Monitor lamp use and test, change or replace lamps.
Diagnostic tests—Test the pump, 4–20 mA output, digital I/O, plumbing, RS232, RFID, fan and touchscreen calibration.
Calibration—See the calibration dates.
Data review Review, print or export data.
Sign on or off If password security is enabled, users must use a password to make operation
changes. Log off to prevent access to system changes.
Note: If a user is signed on remotely, the user cannot be signed off locally. If a user is signed on locally, the user cannot be signed off by a remote user.
Setup menu
Table 12 shows the options on the setup menu.
Table 12 Setup menu options
Icon Description
TOC Change the settings for the TOC operating mode. Refer to TOC operating mode configuration
on page 27.
System Change the system settings for the analyzer (instrument name, language, time, measurement
units, sounds, network and backlight). Refer to System configuration on page 29.
Alarms Change the settings for the digital output alarms. Refer to Configure the digital output alarms
on page 31.
Analogs Change the settings for the analog outputs. Refer to Configure the analog outputs
on page 33.
Printer Change the printer settings. Refer to Configure the printer settings on page 33.
Security Enable or disable security, data logging and/or encryption of exported data. Set password
expiration and auto logout. Add, change or delete user accounts. Show the audit trail. Refer to
Security configuration on page 34.
Bottles Change the auto acceptance and test completion criteria for bottle mode operations. Refer to
Bottle mode configuration on page 37.
Password Change the password of the user that is logged in. This icon is only enabled when security is
enabled.
CIP Set the conductivity threshold and CIP timeout value. Refer to Configure clean-in-place (CIP)
options on page 36.
AutoSampler Enter the autosampler model. Refer to Configure the auto sampler on page 36.
Factory Shows the factory settings which can only be changed by factory authorized users. Push the X
icon to show the factory settings (e.g., model number, serial number and factory calibration date) or copy the oxidation buffers (3 maximum) to a USB flash drive.
22 English
Startup
Connect to power
After plumbing and wiring are complete, connect the instrument to power. For the location of the power switch and power cord, refer to Figure 5 on page
13.
1. Connect the power cord to a power outlet with a safety ground.
2. Toggle the power switch to on.
3. Make sure that the startup screen shows on the display.
Set the sample flow rate
Adjust the flow rate before initial operation. Refer to Specifications on page 5 for sample requirements. For hot water applications, a flow rate of 60 mL/minute is recommended.
1. Push << to show the sliding menu.
2. Push RUN MODE.
3. Select Standby.
Note: In Standby, water continues to flow through the instrument.
4. Use a graduated cylinder to measure the amount of water that drains from the instrument in 15 seconds.
5. Use the flow control valve to adjust the flow rate. Refer to Figure 10.
Figure 10 Flow control valve
1 Flow control valve 2 Water filter
Basic operation
Select the run mode
1. Push << to show the sliding menu.
2. Push RUN MODE.
3. Select an operation. Refer to Run modes on page 52.
Option Description
Online TOC Monitor TOC continuously in process
Conductivity Monitor conductivity continuously in process
CIP validation Start the "Clean in Place" validation
English 23
Option Description
Standby Put the instrument in standby mode (valve open)
Offline Put the instrument in offline mode (valve closed)
TOC manual sample Manual TOC measurement for online sample
Self clean Configure and start a clean cycle
4. Push ENTER.
Calibration
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
TOC calibration
Calibrate the instrument for TOC measurements with default settings or with custom settings. Use default settings with packaged standards kits that use RFID tags on the bottles. Use custom settings for less replicates or less bottles.
1. Make sure that the TOC standards are at room temperature.
2. Push << to show the sliding menu.
3. Push BOTTLE MODE, then RUN STANDARDS.
4. Push TOC CALIBRATION.
5. Select default or custom settings.
6. Obey the screen prompts to set the number of repetitions and confirm the sucrose standard
values.
Note: To save the new settings as the default, push SAVE AS DEFAULT.
7. Install the standards in the order shown on the screen. If the bottle bay door closes too soon, push UNLOCK to open the door.
8. Push START.
9. When the calibration is complete, the results are shown. If a printer is attached, the report can be
sent to a printer. Select the necessary action to complete the calibration.
Push CHECK to accept the calibration and to remove the bottles. This option will not show if the Autoaccept setting is enabled.
• Push CANCEL to repeat the calibration and discard the results.
• Push EXPORT to send the data to the supplied USB flash drive.
Conductivity calibration
Use the optional calibration resistor for this procedure. Figure 11 shows the location of the calibration resistor.
1. Make sure that the conductivity standards are at room temperature.
2. Push << to show the sliding menu.
3. Push BOTTLE MODE, then RUN STANDARDS.
4. Push CONDUCTIVITY CALIBRATION.
5. Obey the screen prompts.
24 English
6. Install the standards in the order shown on the screen. If the bottle bay door closes too soon, push UNLOCK
to open the door.
7. Push START. Install the resistor when the screen prompt shows.
Note: When the calibration is complete, the results show on the screen. If a printer is attached, the report prints automatically.
8. Select the necessary action to complete the calibration.
• Push CHECK to accept the calibration and to remove the bottles. This option will not show if
the Autoaccept setting is enabled.
• Push CANCEL to repeat the calibration and discard the results.
• Push EXPORT to send the data to the supplied USB flash drive.
Figure 11 Calibration resistor
1 Calibration resistor (optional)
Do a system suitability test
A system suitability test examines analyzer measurements according to the guidelines established in USP Method <643> using 500 ppb sucrose and 500 ppb 1,4-benzoquinone standards.
1. Make sure that the system suitability test standards are at room temperature.
2. Push << to show the sliding menu.
3. Push BOTTLE MODE, then RUN STANDARDS.
4. Push SYSTEM SUITABILITY.
5. Install the standards in the order shown on the screen.
Note: If the bottle bay door closes too soon, push UNLOCK to open the door.
6. Close the bottle bay door, then push the forward icon.
If bottles have been properly loaded and the RFID tags can be read, the review test setup dialog box shows. If not, enter the bottle information as prompted.
7. Push START.
Three samples are drawn from each bottle. After all the repetitions are complete, the analyzer repeats the process for the rest of the bottles.
Once all the bottles have been analyzed, the analyzer compares the average reading of the sucrose standard to the average reading of the 1,4 benzoquinone. If the two values are within ±15% of each other, the test passes. The analyzer then shows the test summary dialog box. The formulas that follow are used to calculate the response efficiency.
English 25
re = 100 × (SR ÷ LR)
Where:
SR = suitability response of the analyzer defined as (rss – rw)
LR = limit response of the analyzer defined as (rs – rw).
re = response efficiency of the PAT700
rw = average TOC response for the blank
rs = average TOC response for the sucrose standard
rss = average TOC response for the benzoquinone standard
8. When the suitability test is completed, the results are shown. If a printer is attached, the report prints automatically. Select the necessary action to complete the validation.
Push CHECK to accept the system suitability test and to remove the bottles. This option will not show if the Autoaccept setting is enabled.
• Push CANCEL to repeat the system suitability test and discard the results.
• Push EXPORT to send the data to the supplied USB flash drive.
Measure a grab sample
Use the grab sample test to measure samples taken from other locations. Grab samples do not need bottles with RFID tags.
1. Push << to show the sliding menu.
2. Select BOTTLE MODE, then GRAB SAMPLE.
3. Obey the screen prompts to enter the values.
Note: To save the new settings as the default, push SAVE AS DEFAULT.
4. Install the standards in the order shown. If the bottle bay door closes too soon, push UNLOCK to open the door.
5. Set the volume of each sample bottle.
6. Set an ID for each bottle.
7. Push START.
Note: When the measurement is complete, the results show on the screen. If a printer is attached, the report prints automatically.
8. Select the necessary action to complete the calibration.
• Push CHECK to accept the measurement and remove the bottles. This option does not show if
the autoaccept setting is enabled.
• Push CANCEL to repeat the measurement and discard the results.
• Push EXPORT to send the data to the supplied USB flash drive.
Manage measurement data
Measurement data can be reviewed on the instrument, printed, exported and filtered.
1. Push << to show the sliding menu.
2. Select DATA REVIEW. The data review window shows. Icon command options show on the
bottom of the screen.
Set the data filter
Data can be filtered by test type, date range or special ranges.
1. Push << to open the sliding menu.
2. Select DATA REVIEW, then FILTER.
26 English
3. Select all of the data types that apply. All types are selected by default.
4. Select time range or special range. Special range options sets the data to sort first by type, then
by time. The time range option sets the data to sort based on the specified date range.
Option Description
Last accepted Includes only the last accepted tests for the selected data types
Last done Includes only the last completed tests for the selected data types
Last five Includes only the last five completed tests for the selected data types
Export or print data
1. Push << to show the sliding menu.
2. Select DATA REVIEW.
3. Select an option.
Push EXPORT to send data as formatted text to the supplied USB flash drive.
• Push PRINT to send data to an attached printer.
Advanced operation
TOC operating mode configuration
Configure the general settings
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push TOC.
4. Select the General tab.
5. Select an option.
Option Description
Flush Time Set the amount of time the cell is flushed before a TOC analysis is started (hh:mm:ss).
Options: 00:00:00 to 23:59:59 (default = 00:01:00)
Cycle Time Set the minimum amount of time between automatic TOC analyses. Any value less than
the actual elapsed time causes the next analysis to start immediately (hh:mm:ss). Options: 00:00:00 to 99:59:59 (default = 00:00:00)
Use pump Enable the sample pump for online sampling (default = disabled). Do not enable the
sample pump on a pressurized system or pump damage may occur.
Bypass flow during analysis
Enable bypass flow. When disabled, the bypass flow valve stays closed. When enabled, sample water flows continuously through the analyzer (bypasses the analysis cell) during TOC analysis. The bypass valve is open when the sample is captured and closed when the analysis is complete.
Average last Average the last x number of TOC analysis. The averaged value is shown on the home
screen and in the log view, logged to the data log and sent to the printer based on the printer settings.
Switch Lamps when Marginal
Enable the analyzer to switch to the other UV lamp when the UV lamp diagnostics sense a marginal level of UV intensity on the operating UV lamp.
English 27
Configure the flow during idle mode
Select the state the analyzer will enter while it waits for the next analysis (cycle) to start. The analyzer will only enter idle mode if the cycle time is more than the combined flush time and oxidation time.
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push TOC.
4. Select the Idle tab.
5. Select an option.
Option Description
Without conductivity The bypass valve opens during idle mode. There is continuous flow through the
bypass loop. No conductivity is reported.
With conductivity The sample valve opens during idle mode. There is continuous flow through the
sample cell. Conductivity of the water is reported.
None The analyzer sample valve closes during idle mode. There is no flow through the
sample cell.
Configure digital control
Enable the digital inputs and set the states for digital input control as necessary.
The two digital inputs are optically isolated. The inputs are rated for 5–30 VDC, 2–15 mA. Terminals 4 and 5 of the discrete input terminal supply a 12 VDC output that can be used to drive each digital input. Both digital input 1 and digital input 2 share a common positive power connection. In the default setting:
Digital Input 1—A high to low (power supplied for at least 500 milliseconds) transition on digital
Input 1 stops the current operation and immediately starts a single TOC analysis. The analysis goes to completion regardless of whether subsequent trigger values are received. When the analysis is complete, the instrument goes back to the previous run mode.
Digital Input 2— When digital input 2 is high (no power supplied), the analyzer enters online TOC
analysis mode. If digital input 2 is held low (power supplied), the analyzer goes to conductivity mode.
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push TOC.
4. Select the Digital Control tab.
5. Select an option.
Option Description
Control with digital inputs
Enable the two digital inputs to control the TOC analysis and run mode (default = disabled). When enabled, the normal operating modes shown on the Run Mode window do not apply.
Note: When a TOC analysis is started with digital input 1, the analyzer ignores all other commands through the digital inputs.
Return to defined run mode
Completes the current TOC analysis, then goes to the previous run mode. When not selected, the normal operating modes in the run mode dialog box do not apply. All control is activated with the digital inputs, which lets the user toggle between TOC mode and conductivity mode.
28 English
Option Description
Invert Digital Inputs
Invert the default state for control of the digital inputs.
Digital Input 1—When power is removed across pins 1 and 2 for at least 500 milliseconds, the analyzer stops the current operation and does a single TOC analysis.
Digital Input 2—When power is supplied across pins 1 and 3 (low state), the analyzer enters online TOC analysis mode. When power is not supplied across pins 1 and 3 (high state), the analyzer goes to conductivity mode.
Make TOC analog output persistent
Hold the last reported value at the output when the analyzer is in standby mode or offline mode.
Invert Digital Outputs
Invert the default state of each digital output (1–4). For the default state of each digital output, refer to Table 13 on page 32.
Enable the stabilization process
When the analyzer switches to do an online TOC analysis after it operates in continuous flow modes (such as conductivity mode or standby) for extend periods of time, a stabilization process should be done. The stabilization process can decrease the number of TOC analysis cycles necessary to give stable results.
During the stabilization process, water is trapped in the sample cell and the UV lamp is set to on to thermally stabilize and clean the sample cell.
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push TOC.
4. Select the Stabilization tab.
5. Select an option.
Option Description
Stabilize cell prior to analyzing
Enable automatic stabilization of the sample cell before a reportable TOC analysis is done.
Time between TOC analyses
Set the minimum elapsed time between TOC analysis before the automatic stabilization process is activated (hh:mm:ss). Options: 00:00:01 to 23:59:59 (default = 02:00:00)
Stabilization time Set the amount of time to do the stabilization process (hh:mm:ss). Options:
00:00:01 to 23:59:59 (default = 00:10:00)
Set the TOC offset
Use the TOC offset feature to make the TOC output between two instruments the same and better than the ±5% specification. The TOC offset value is applied to the reported TOC analysis result.
1. Push <<
to show the sliding menu.
2. Push the SETUP icon.
3. Push TOC.
4. Select the Offset tab.
5. Enter the TOC offset. Options: +5.0 and –5.0.
System configuration
Set the analyzer name, location and language
1. Push << to show the sliding menu.
2. Push the SETUP icon.
English 29
3. Push SYSTEM.
4. Select the General tab.
5. Select an option.
Option Description
Instrument Name
Enter the name that will be used for the analyzer by the system. Options: 1 to 13 alphanumeric or symbol characters (default = PAT700 + the serial number)
Note: The system will reboot when the instrument name is changed.
Location Enter the location of the instrument.
Language Set the language for all interfaces (default = English). For Japanese and Chinese
translations, this option is not available. The applicable language will be installed at the factory.
Note: A change to the language setting changes the operating system setting in the registry. When the system is rebooted, the new setting is used.
Set the date, time and units
Set the date, time, temperature units and conductivity units that show on the display.
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push SYSTEM.
4. Select the Display tab.
5. Select an option.
Option Description
Date/Time Push Set Time. Use the arrow keys to enter the date and time.
Time Format Set the time format (default = 24-Hour).
Temperature display Set the temperature units (default = Deg. C).
Conductivity display Set the conductivity units (default = µS/cm U = uncompensated conductivity).
Configure the alarm sounds
Select the alarm sound for each alarm type.
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push SYSTEM.
4. Select the Sound tab.
5. Select a sound for each alarm type. Push the speaker icon to hear the selected sound.
6. Push the volume slider to set the speaker volume.
Configure the network settings
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push SYSTEM.
30 English
4. Select the Network tab.
5. Select an option.
Option Description
Static TCP/IP Address
Set the TCP/IP address for the analyzer. Options: 0 to 255 for each octet (default =
192.168.0.105)
Static Subnet Mask Set the subnet mask for the analyzer. Options: 0 to 255 for each octet (default =
255.255.255.0)
Static Gateway Set the default gateway for the analyzer. Options: 0 to 255 for each octet (default =
blank)
DNS Server Set the DNS server address for the analyzer. Options: 0 to 255 for each octet
(default = blank)
WINS Server Set the WINS server address for the analyzer. Options: 0 to 255 for each octet
(default = blank)
DHCP Enable DHCP. When selected, the TCP/IP address, subnet mask and default
gateway entries are disabled (default = enabled).
Configure the backlight settings
1. Push << to open the sliding menu.
2. Push the SETUP icon.
3. Push SYSTEM.
4. Select the Backlight tab.
5. Configure the backlight options.
Option Description
Automatically turn off backlight Have the backlight switch off automatically after a period of inactivity.
Turn off after Select the number of minutes of inactivity before the backlight switches
off. Options: 2, 5, 10, 15 or 30 minutes
6. To adjust the brightness of the backlight, push the brightness slider.
Configure the digital output alarms
Assign output functionally for each digital output as necessary. Refer to Table 13. To invert the digital outputs, refer to Configure digital control on page 28
.
Note: The TOC alarm can be assigned to multiple digital outputs (1–4) at the same time to set a multi-level alarm on the TOC result.
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push ALARMS.
4. Assign the output functionality for each digital output. Refer to Table 13.
Note: Conductivity alarm, Error state and Valve state can only be assigned to one digital output at a time.
5. Select the checkbox to enable the alarm, if present.
6. For TOC alarm, select the upper TOC limit for the alarm. Options: 1 to 2500 (default = 500). This
value applies only to online TOC analysis, not to bottle or grab sample analysis.
English 31
Table 13 Digital output – default states
Alarm type Description
TOC alarm High state—The TOC level is below the selected upper limit.
Low state—The TOC level is above the selected upper limit.
When enabled, any online TOC reading that is more than the alarm upper limit triggers a TOC alarm.
Conductivity alarm High state—The conductivity level is below the conductivity limit.
Low state—The conductivity level is above the conductivity limit.
The actual (uncompensated) conductivity limits are in accordance with the specifications set forth in the method “<645> Water Conductivity” of the USP 25–NF 20 (January
2002). Refer to Conductivity alarm limit on page 32.
For proper operation of the conductivity alarm, uncompensated conductivity must be selected under the Display tab in the System settings.
Error state High state—No error conditions are present.
Low state—An error condition is present.
Valve state (TOC analysis start)
High state—The inlet sample valve is open.
Low state—The inlet sample valve is closed.
Conductivity alarm limit
The actual sample water temperature is rounded down to the nearest 5 degrees to set the conductivity limit. This makes sure that any potential excursions are identified early. For example, if the measured water temperature is 24.5 °C, the 20 °C limit is applied and an alarm is triggered if the conductivity is more than 1.1
μS/cm U.
To make sure of an agreement between all data outputs, the TOC, conductivity and temperature values are all rounded to have the same accuracy as the analyzer for alarm limit checking, display, printouts and any digital outputs.
Table 14 Conductivity alarm limit
Temperature (°C) Uncompensated conductivity
(µS/cm)
Temperature (°C) Uncompensated conductivity
(µS/cm)
0 0.6 55 2.1
5 0.8 60 2.2
10 0.9 65 2.4
15 1.0 70 2.5
20 1.1 75 2.7
25 1.3 80 2.7
30 1.4 85 2.7
35 1.5 90 2.7
40 1.7 95 2.9
45 1.8 100 3.1
50 1.9
32 English
Configure the analog outputs
Set the parameter (TOC, conductivity or temperature), upper limit and lower limit for each analog output. Set the value at the analog outputs when an error occurs.
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push ANALOGS.
4. Select the General tab.
5. Set the zero scale and full scale values for each parameter (TOC, conductivity and temperature).
Zero scale—The lower limit parameter value that will be represented as 4 mA at the analog output. Options: TOC = 0 to 1999 ppb, Conductivity = 0 to 19 µS/cm, Temperature = 0 to 99 °C
Full scale—The upper limit parameter value that will be represented as 20 mA at the analog output. Options: TOC = 1 to 2000 ppb, Conductivity = 1 to 20 µS/cm, Temperature = 1 to 100 °C
6. Set the value at the analog outputs that will be present when an error occurs. Options: 2 mA (default), last output or 22 mA
7. To enable analog output during bottle mode, select "Analog Outputs during Bottle Mode".
Configure the printer settings
Configure the analysis result printing and the printing frequency for the online TOC mode and conductivity mode.
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push PRINTER.
4. Select the General tab.
5. Select the type of printer.
6. Select the TOC tab.
7. Select an option.
Option Description
Disabled Disable the printing of online TOC readings.
Continuous Enable continuous printing. A TOC printout occurs at the end of each analysis cycle.
Percent Change Enable a printout to occur only when the change in TOC between successive analyses is
more than the value in the Limit field. Enter a value in the Limit field. Options: 1 (default) to 100%
8. Select the Conductivity tab.
9. Select an option.
Option Description
Disabled Disable the printing of online conductivity readings.
Timed Enable continuous printing. A conductivity printout occurs at the end of each interval.
Enter an interval in the Interval field. Options: 1 minute (default) to 24 hours
Percent Change Enable a printout to occur only when the change in conductivity between successive
analyses is more than the value in the Limit field. Enter a value in the Limit field. Options: 1 (default) to 100%
English 33
Security configuration
Enable security, data logging and encryption
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push SECURITY.
4. Select the General tab.
5. Select an option.
Option Description
Enable Security Enable security (default = disabled). Login is necessary before changes are made. Only
the administrator may access the security settings.
Note: When not selected, no login is necessary and the audit trail is disabled.
Enable Data Logging
Enable data logging (default = enabled). All data (maximum of 5000 records) is automatically kept in the internal data buffer.
Encrypt Exported Data
Enable encryption of exported data files for 21 CFR Part 11 compliance (default = disabled). Encrypted files are read with the TOC Dataview PC software. Decrypted data can be saved as PDF for file storage.
6. To set all the settings back to the factory defaults, push Restore Factory Defaults.
Set password expiration and automatic log off
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push SECURITY.
4. Select the Settings tab.
5. Select an option.
Option Description
Passwords expire in Set the number of days before a password must be changed. Options: 30 to
365 days (default = 60)
Remember last Set the number of passwords that are kept in memory to prevent the use of an
old password. Options: 0 to 5 (default = 3)
Automatically logoff after Set the number of minutes of inactivity before automatic logoff occurs. Options:
0 (disable auto logoff) to 60 (default = 30)
Add, delete or change users
A maximum of 10 operator users can be added. Only operator users may be added or deleted. The administrator and factory users cannot be deleted.
There are four types of users:
Guest—Any user who is not logged on to the system. This user may access most areas and view information but cannot change any settings or start any operations. Guests have no access to the security settings dialog box.
Operator—Has full access to most dialog boxes. Operators have no access to the security settings dialog box and read-only access to the factory settings and the bottle mode settings.
Administrator—Has full access to most dialog boxes. The administrator has read-only access to the factory dialog box. The default administrator entry has a user name of “Administrator”, a user ID of “Admin” and a default password of “123456”.
Factory—Has full access to all dialog boxes.
Note: Push the Print icon to print the user settings for all the users except the administrator and factory users.
34 English
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push SECURITY.
4. Select the Users tab. A list of the operator users that have been added shows. The administrator
and factory users are not shown.
5. To add a user:
a. Push the + icon. b. Enter the full name of the user (5 to 20 alphanumeric or symbol characters and must be
unique).
c. In the ID field, enter a unique user identifier (5 to 15 alphanumeric or symbol characters). d. Enter a password for the user (5 to 15 alphanumeric or symbol characters). Passwords are
case sensitive.
e. Enter the password for the user again to confirm. f. To make the user change their password at the next logon, select Change Password.
Note: Password changes are recorded in the audit trail and password history list. New and old passwords are not recorded in the audit trail.
6. To change a user:
a. Select a user. b. Push the triangle icon. c. Edit the fields as necessary.
Note: To exit and not save, push the "x" icon.
7. To delete a user:
a. Select a user. b. Push the "–" icon. c. Push Yes to confirm.
About backdoor passwords
If administrators or factory personnel forget their passwords, they may get temporary access with a backdoor password supplied by factory service. Upon successful sign on, the analyzer prompts the user to change the password.
Contact technical support to get a backdoor password for the analyzer. To get the password, the factory asks for the serial number and current date shown on the analyzer. Have this information ready.
Show the audit trail
Only administrator and factory users can see the audit trail. Any additions, changes or deletions to the analyzer configuration are recorded in the audit trail with the:
Keyword ("added", "modified" or "deleted")
• Information affected
• Old value if applicable
• New value if applicable
• Full name of the user that did the operation
• Date and time of the change
System events such as logging in and exporting data are also recorded in the audit log.
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push SECURITY.
4. Select the Audit trail tab. The audit trail shows. The audit trail shows the filtered audit trail data.
English 35
Daily header—Shown once per day before the first entry and at the change from one day to the next.
Entry header—Shown at the start of each audit trail entry. It includes the time of the entry in long format (hh:mm:ss) and the name of the person associated with the entry. If the analyzer time is configured for 24-hour format, the AM/PM entry is blank.
Audit trail actions—shown after the entry header. There are four types of entries possible for this section (data added, data modified, data deleted or user action). Data added—the keyword “Added” shows before the item name. The next line contains the keyword “New Value:” and then the value name.
Data modified—the keyword “Modified” shows before by one space and the item name. The next line contains the keyword “Old Value:” and then the value name. The next line contains the keyword “New Value:” and then the value name.
Data deleted—the keyword “Deleted” shows before two spaces and the item name. The next line contains the keyword “Old Value:” and then the value name.
5. To send the audit trail data to the printer, push PRINTER.
6. To send the audit trail data to the USB flash drive, push EXPORT. The audit trail sends the entire
audit trail in formatted clear text with a checksum to make sure that there is data integrity.
7. To delete the audit trail, push CLEAR DATA LOG.
8. To filter the audit trail, push the Filter icon.
Configure clean-in-place (CIP) options
Set a CIP validation to be done according to a selected time duration or threshold conductivity value. This feature supplies local control on a CIP system when automation is not supported. This feature is typically used during the development of cleaning validation procedures.
1. Push << to open the sliding menu.
2. Push SETUP.
3. Push CIP.
4. Select Threshold Conductivity or CIP Timeout.
Option Description
Threshold Conductivity
Push the CIP button when in CIP Validation mode to start a TOC analysis when the conductivity of the water is less than or equal to the selected conductivity threshold value. In the field, enter a conductivity threshold value for CIP validation (µS/cm).
CIP Timeout Set the amount of time before a TOC analysis is started.
In the field, enter the time duration for CIP validation in minutes and seconds (MM:SS).
Configure the auto sampler
1. Push << to open the sliding menu.
2. Push SETUP.
3. Push AUTOSAMPLER.
4. Select an option.
Option Description
Auto Sampler Model Select the auto sampler model. Only ASX-7200 is currently supported.
Equipped with Rinse When the rinse function is available on the auto sampler, select this checkbox to
enable a needle rinse between samples.
36 English
Bottle mode configuration
Configure the TOC calibration settings
1. Push << to show the sliding menu.
2. Push the SETUP icon.
3. Push BOTTLES.
4. Select the TOC Calibration tab.
5. Select an option.
Option Description
User must manually accept results after test is complete
The results of the TOC calibration are not used until they are manually accepted.
Return to online mode The analyzer goes back to the online mode after the TOC calibration. When not
selected, the analyzer stays at the bottle mode dialog box. All valves are closed. The pump is off. The lamp is off.
Note: This option is only available when "User must accept results after test is complete" is selected.
Auto accept results and return online
The Slope Limit and R2 Limit options that follow are enabled. If the calibration is accepted, the new calibration factors are used.
Slope Limit <= Set the limit of the slope for the calibration to pass. Options: 1 to 15% (default =
15 %)
Note: This option is only available when "Auto accept results and return online" is selected.
R2 Limit >= Set the limit for the determination coefficient. Options: 0.9000 to 0.9999 (default =
0.9900)
Note: This option is only available when "Auto accept results and return online" is selected.
Display 'as Received' reports
Enable 'as Received' reports to be shown in the calibration results.
Configure the conductivity calibration settings
1. Push <<
to show the sliding menu.
2. Push the SETUP icon.
3. Push BOTTLES.
4. Select the Conductivity Calibration tab.
5. Select an option.
Option Description
User must manually accept results after test is complete
The results of the conductivity calibration are not used until they are manually accepted.
Return to online mode The analyzer goes back to the online mode after the conductivity calibration.
When not selected, the analyzer stays at the bottle mode dialog box. All valves are closed. The pump is off. The lamp is off.
Note: This option is only available when "User must accept results after test is complete" is selected.
Auto accept results and return online
The Slope Limit options that come next are enabled. If the calibration is accepted, the new calibration factors are used.
English 37
Option Description
Slope Limit from previous cal <=
Set the percent difference between the resulting gain factor and the current value. Options: 1 to 5% (default = 5%)
Note: This option is only available when "Auto accept results and return online" is selected.
Slope Limit from factory cal <=
Set the percent difference between the resulting gain factor and the factory value. Options: 1 to 10% (default = 10 %)
Note: This option is only available when "Auto accept results and return online" is selected.
Use Calibration Resistor The calibration check resistor is used during the conductivity calibration
process.
Set the analyzer to return to online mode
Set the analyzer to return to the online mode after TOC validation, the System suitability test and/or the analysis of one/multiple grab sample(s), or stay at the bottle mode dialog box (default).
1. Push <<
to show the sliding menu.
2. Push the SETUP icon.
3. Push BOTTLES.
4. Select the TOC Validation, System Suitability or Grab Sample tab.
5. To have the analyzer return to online mode after the operation is completed, select the Return to
online mode checkbox.
When not selected, the analyzer stays at the bottle mode dialog box after the operation. All valves are closed. The pump is off. The lamp is off.
Bottle mode operation
Validate the TOC calibration
Validate the TOC calibration with the default settings or custom settings. A TOC validation can be a single-point analysis or multi-point analysis. Use the default settings with the packaged standards kits that use RFID tags on the bottles. Use the custom settings to add bottles or to use less replicates.
The instrument response for the validation standard must have a deviation of less than 15% of the calibrated response to be accepted.
1. Make sure that the TOC standards are at room temperature.
2. Push << to show the sliding menu.
3. Push BOTTLE MODE, then RUN STANDARDS.
4. Push TOC VALIDATION.
5. Select the default or custom settings. The default setting is a blank bottle and one standard of
500 ppb sucrose. To do a multi-point validation, select the custom settings and enter the necessary configuration. At least one bottle must be selected.
6. Obey the screen prompts to set the number of repetitions and confirm the sucrose standard values.
Note: To save the new settings as the default, push SAVE AS DEFAULT.
7. Install the standards in the order shown on the screen.
Note: If the bottle bay door closes too soon, push UNLOCK to open the door.
8. Close the bottle bay door, then push the forward icon.
If bottles have been correctly loaded and the RFID tags can be read, the review test setup dialog box shows. If not, enter the bottle information as prompted.
9. Push START.
38 English
The dialog box shows the current status of the test and a lists of the results of each repetition and bottle as it is being read.
10. When the validation is complete, the results are shown. If a printer is attached, the report prints automatically. Select the necessary action to complete the validation.
Push CHECK to accept the TOC validation and to remove the bottles. This option will not show if the Autoaccept setting is enabled.
• Push CANCEL to repeat the TOC validation test.
• Push EXPORT to send the data to the supplied USB flash drive.
Do an auto sampler test
Do an auto sampler test after the analyzer and auto sampler are prepared and connected to each other. Refer to the auto sampler quick start guide for the auto sampler installation procedure and the auto sampler test instructions.
Get an excursion sample
Use excursion mode to withdraw sample from the source stream to fill two 65 mL RFID tagged bottles when a TOC alarm, conductivity alarm or specific alarm conditions occurs. An excursion sample can also be implemented manually.
Note: Excursion mode does not operate without RFID tagged excursion bottles. A minimum flow rate of 160 mL/minute is necessary for excursion capture.
Items needed: two RFID tagged excursion bottles
1. Push << to show the sliding menu.
2. Push BOTTLE MODE, then EXCURSION MODE.
3. Install the RFID tagged bottles as shown on the screen.
Note: If the bottle bay door closes too soon, push UNLOCK to open the door.
4. Close the bottle bay door, then push the forward icon.
5. Enter the TOC limit in the Excursion mode setup screen. Options: 1 to 2000 ppb (default =
500 ppb)
6. To immediately fill the excursion bottles, push START.
7. Push DONE to go to the home screen.
When an excursion condition occurs, the analyzer collects a sample. When both bottles are filled, the instrument automatically does a confirmation analysis on the contents of the first bottle. This offline analysis is then used to validate the online result.
When the analysis of the contents of the first bottle is completed, the TOC results are reported and if the results (TOC or Cond > limit) are the same, the message "Excursion is valid" is reported. If the results are not the same (TOC or Cond < limit), the message "Excursion not valid" is reported.
The water sample in the second bottle is available for lab analysis to identify the cause of the water system excursion.
Flush the sample bottle lines
1. Push << to show the sliding menu.
2. Push BOTTLE MODE, then BOTTLE FLUSH.
3. Install bottles of pure water as shown on the screen.
Note: If the bottle bay door closes too soon, push UNLOCK to open the door.
4. Close the bottle bay door, then push the forward icon.
The analyzer starts the flushing cycle and will identify on the display which bottle line is in the cleaning process.
English 39
Maintenance
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
D A N G E R
Electrocution hazard. Remove power from the instrument before doing maintenance or service activities.
Maintenance schedule
Table 15
shows the recommended schedule of maintenance tasks. Facility requirements and
operating conditions may increase the frequency of some tasks.
Table 15 Maintenance schedule
Task 6 months As necessary
Clean the instrument on page 40 X
Clean leaks and spills on page 40 X
Replace the water filter
1
X
Replace the UV lamps
1
X
Replace a fuse on page 41 X
1
Refer to the instructions supplied with the replacement part.
Clean the instrument
Clean the exterior of the instrument with a moist cloth and a mild soap solution.
Clean leaks and spills
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations.
W A R N I N G
Personal injury hazard. The sample line contains water under high water pressure that can burn skin if hot. Qualified personnel must remove the water pressure and wear personal protective equipment during this procedure.
N O T I C E
Tighten a new ferrule 1¼ turns the first time it is tightened. Then, only tighten the ferrule 1/8 turn to make a connection. Do not tighten the fittings more as this may cause damage to the fittings or ferrules and cause leaks.
If a leak occurs in the tubing, the sample water will drain from small holes in the bottom of the enclosure.
For leaks caused by a valve, worn tubes or other parts, send the instrument to the manufacturer for service. Refer to Send the instrument for service on page 42
.
40 English
1. Put on protective eye wear.
2. Set the power switch to off.
3. Disconnect the sample supply. Let the sample cool if hot.
4. Open the tubing enclosure.
5. Use sample flow that is not hot to find the leak.
6. Tighten fittings as necessary. Look for additional leaks and repair if possible.
7. After the leaks are repaired, remove excess water that is on the inside and underside of the
enclosure with paper towels. Let the enclosure dry completely with the door open.
8. Use a towel or mop to remove spills around the analyzer.
9. Close the enclosure door. Set the power switch to on.
10. Do a plumbing test to make sure that the analyzer can operate without leaks. Refer to Do a
plumbing test on page 48.
Replace a fuse
D A N G E R
Fire hazard. Use the same type and current rating to replace fuses.
Fuse failure may be an indication that the instrument has a problem that requires service. Figure 12 shows the fuse location.
Item to collect:
1.25 A, 250 V EN60127 type T fuse
1. Toggle the power switch to off. Disconnect the instrument from the power outlet.
2. Use the supplied tool to open the display-side door.
3. Use a flat-blade screwdriver to remove the fuse holder.
4. Replace the fuse with a fuse of the same type and rating.
5. Use the flat-blade screwdriver to install the fuse holder.
6. Close the display-side door with the supplied tool. Make sure that the door is tightly closed to
keep the enclosure rating and the safety protections.
7. Connect the instrument to the power outlet. Toggle the power switch to on.
Figure 12 Fuse location
1 Fuse (2x)
Drain the instrument
To prepare the instrument for storage or shipment, remove all water from the instrument. Water left in the instrument may freeze, expand and cause instrument damage. Water in the measurement cell may contain contaminants that will cause damage to instrument components.
English 41
To drain all water from the instrument:
1. Remove the instrument from service.
2. Disconnect the instrument from the water system.
3. Set the power switch to on.
4. Push << to show the sliding menu.
5. Select Diagnostics>Tests>Plumbing Test.
6. From the drop-down menu, select Flow through cell with pump.
7. Let the test continue until all of the water has drained from the instrument.
8. When the water flow stops, push the green arrow to stop the plumbing test.
Send the instrument for service
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations.
1. Drain the instrument. Refer to Drain the instrument on page
41.
2. Clean the instrument fully.
3. Get a return authorization from the manufacturer. Make sure that the return authorization number
is on the shipping label.
Troubleshooting
The analyzer incorporates extended diagnostic and reporting to inform the user about the nature of a problem.
Alarm codes, with user-initiated diagnostic tests, serve to isolate the source of the problem. The analyzer constantly monitors its operation and measurement values for alarm conditions.
Alarm conditions include error alarms, critical error alarms, warning alarms and measurement alarms.
Alarms
Alarm indication
The analyzer has two basic operational modes: online and bottle. The online TOC and conductivity are the two primary online modes. In online TOC and conductivity modes, the measurement screen shows an animation that identifies the general state of water flow through the analyzer.
In online TOC, conductivity, or manual TOC modes, the animation area is also an alarm condition indicator. When an alarm occurs in an online mode:
• The animation area flashes red.
• The value of the variable that has caused the alarm becomes red.
• The alarm sound** is heard if enabled.
• The alarm condition is recorded to the data log and audit trail.
• The alarm icon is enabled.
• Error, critical error or warning: The digital output that represents the error state identifies an error condition. Refer to Configure the digital output alarms on page 31.
• Error or critical error: The three analog outputs identify an error condition (2 mA or 22 mA). Refer to Configure the analog outputs on page 33.
**
Each type of alarm (stop, warning, alarm) has a different sound. If multiple alarms occur, only the highest-level alarm is heard.
42 English
After an alarm is acknowledged:
The animation area shows steady red.
• The alarm sound is not heard.
• Once a full cycle has occurred without any errors, the error is removed and the normal animation shows. The digital output will no longer show an error state and the analog outputs represent the new valid result.
• If on the next cycle the same or a new error occurs, the animation area again flashes red and must be acknowledged.
• As long as alarms are present in the alarm log, the alarm log is available.
Accept or remove an alarm
When a warning or an error that is not critical occurs, the animation area of the measurement screen flashes red.
1. Push << to show the sliding menu.
2. Select ALARMS. The Alarm Acknowledge screen opens.
• Push ACCEPT ALARMS (bell with green checkmark) to acknowledge the alarm.
• Push DELETE ALARMS (bell with a red X) to delete all alarms in the list.
3. Push ENTER to return to the measurement screen. If the error condition still exists but has been
accepted, the animation area shows steady red.
Error codes
Table 16 shows the error codes that occur any time one of the non-critical subsystems fail or a failure
occurs from which the analyzer tries to recover.
When an error occurs, the animation area flashes red to identify when an error condition is present. In addition, the digital output that represents the error condition and the three analog outputs (2 or 22 mA) change their state to identify when an error condition is present. Refer to Alarm indication on page 42.
When the error condition is removed, the analyzer attempts to recover and continue to operate. Once a valid TOC measurement is completed, the analog outputs once again represent the measured values.
Table 16 Error codes
Code Message Description
SC027 FalseConductivity The measured conductivity is less than 0.
SC029 BadTocConvert The temperature/conductivity is out of bounds for TOC conversion.
SC035 LowOxidationRate The TOC is too low (out of range).
SC036 NegativeTOC Negative TOC (out of range). Organic acids may be present.
SC037 LowTemperature The temperature is out of range low.
SC038 HighTemperature The temperature is out of range high.
SC039 LowResistance The resistivity is out of range low.
SC040 HighResistance The resistivity is out of range high.
SC112 UnableTo-StartPrinterModule The analyzer can not start the printer module.
SC113 UnableTo-StartRs232Module The analyzer can not start the RS-232 module.
SC116 UnableTo-ReadDataLog The analyzer can not read the data log.
SC117 UnableTo-WriteToDataLog The analyzer can not write to the data log.
SC128 UvLampMonitorFailed The UV lamp monitor has failed.
English 43
Table 16 Error codes (continued)
Code Message Description
SC151 LampExtinct The current lamp has failed. Replace the lamp.
SC163 UnableTo-StopCurrentOperation The analyzer can not stop its current operation.
SC164 UnknownError The analyzer has experienced an unknown error.
SC167 WatchDogIoCard The I/O card has stopped responding.
SC168 WatchDogTocAlgoithm The TOC algorithm has stopped responding.
SC169 WatchDogPrinter The printer has stopped responding.
SC170 WatchDogRs232 The RS-232 module has stopped responding.
SC177 Exception An exception was thrown.
SC188 TOCOverLimit The TOC measurement is more than the instrument range.
Critical error codes
In a critical error condition, the instrument cannot function. The error shows as a full-screen flashing message. The digital output that represents the error condition and the three analog outputs (2 or 22 mA) change their state to identify when an error condition is present. Refer to Alarm indication on page
42.
Operation stops until the error condition is corrected. Touch the screen interface to accept the alarm and see more information. Refer to Table 17.
Table 17 Critical error codes
Code Message Description
SC110 UnableTo-StartIoModule The analyzer can not start the IO module.
SC111 UnableTo-StartTocModule The analyzer can not start the IO module.
SC114 UnableTo-StartEthernetModule The analyzer can not start the Ethernet module.
SC152 LampUnableToSwitch The analyzer can not switch from one lamp to the other.
SC173 RangeTable-FailedToLoad The RangeTable.txt file could not be loaded.
SC174 InstrumentTemperature The ambient temperature is more than the instrument specifications.
Decrease the instrument and environmental temperature.
SC175 BothLampsBad Both UV lamps have failed. Replace both lamps.
SC176 NoCond-TempReadings The analyzer can not measure conductivity and temperature. Restart
the analyzer.
SC179 HardwareInitializationFailure An error occurred during the hardware initialization.
Warning codes
When a warning condition is present, a warning indicator (flashing red rectangle on the display) shows on the display to identify that a warning message is pending in the alarm log. The digital output that represents an error condition changes its state to identify when a warning condition is present. The three analog outputs do not identify when a warning condition is present. Refer to Alarm
indication on page
42.
44 English
Operation may continue when a warning condition is present. Refer to Table 18.
Table 18 Warning codes
Code Message Description
SC100–SC103 IncorrectBottle-LoadedinPositionX Incorrect bottle detected in position X. The analyzer can
not do the bottle test operation. Examine and change the bottle as necessary.
SC104–SC107 NotEnough-FluidinBottleX There is not enough fluid in bottle to complete the
operation. The analyzer aborts operation and returns to its normal operating mode.
SC108 UnableTo-ReadSettings Some or all of the settings could not be read from the
settings file.
SC109 UnableTo-WritetoSettings The analyzer can not write to the settings file.
SC118 UnableTo-ChangeLanguageSettings The analyzer can not change the language settings.
SC119 UnableTo-ReadLanguageSettings The analyzer can not read the language settings
SC120 UnableTo-ChangeEthernetSettings The analyzer can not change the Ethernet settings.
SC121 UnableTo-ReadEthernetSettings The analyzer can not read the Ethernet settings.
SC122 UnableTo-PrintSystemSettings The analyzer can not print the system settings.
SC125 UnableTo-StartLampTest The analyzer can not start the lamp test.
SC126 UnableTo-StopLampTest The analyzer can not stop the lamp test.
SC127 UvLampLifeExceeded The maximum lamp hours of usage have been
exceeded or the UV Detect™ output is below the acceptable limit. The lamp has been set to a status of marginal. Replace the lamp.
SC129 UnableTo-
ReadInformationFromBottle
The analyzer can not read the information from the bottle.
SC130 UnableTo-WriteInformationToBottle The analyzer can not write the information to the bottle.
SC132 UnableTo-StartPumpTest The analyzer can not start the pump test.
SC133 UnableTo-StopPumpTest The analyzer can not stop the pump test.
SC134 UnableTo-CalibratePump The analyzer can not calibrate the pump.
SC135 UnableTo-ExportDataLog The analyzer can not export the data log.
SC136 UnableTo-ExportAuditTrail The analyzer can not export the audit trail.
SC137 UnableTo-AddUser The analyzer can not add a user.
SC138 UnableTo-EditUser The analyzer can not edit a user.
SC139 UnableTo-DeleteUser The analyzer can not delete a user.
SC141 UnableTo-AcknowledgeAlarms The analyzer can not acknowledge alarm.
SC142 UnableTo-SilenceAlarms The analyzer can not delete alarms.
SC143 UnableTo-StartSelfClean The analyzer can not start the self clean.
SC144 UnableTo-StopSelfClean The analyzer can not stop the self clean.
SC145 UnableTo-StartOnlineManualSample The analyzer can not start the online manual sample.
SC147 UnableTo-RestoreFactoryDefaults The analyzer can not restore the factory defaults.
SC150 UnableTo-PrintDataLog The analyzer can not print the data log.
English 45
Table 18 Warning codes (continued)
Code Message Description
SC153–SC156 BottleXMarkedAsUsed The bottle installed in location is marked as used.
Replace the bottle with the correct standard.
SC160 ExcursionBottle4NotEmpty The excursion bottle is not empty. Replace the
excursion bottle with an empty bottle.
SC165 LampSwitch One lamp is weak or has failed. Replace the lamp.
SC172 RougeDetected One or more of the 500 ppb standards took longer than
expected to oxidize. Clean the cell.
SC178 ExcursionCaptured An excursion sample was captured in an excursion
bottle. Remove the sample from the analyzer.
SC180 ConductivityMeterAlarm The analyzer has not passed the conductivity meter
test.
SC181 BottleTestCancelled The bottle test was cancelled by the user. Start the test
again.
SC182–SC185 BottleXExpired The bottle installed in the location has expired. Replace
the bottle with the correct standard.
SC186 BottleTestCanceled-ExpiredStandard The bottle test was canceled because there is an
expired standard bottle in the bottle bay.
SC189 IncorrectRFID The standards bottle has a incompatible RFID tag.
SC192 ExcursionNotConfirmed The excursion is not confirmed.
Measurement alarm codes
Measurement alarms occur when TOC or conductivity levels in the process water are more than the set limits. Refer to Table 19.
The digital outputs that identify a TOC alarm or conductivity alarm change their state when a measurement alarm is present.
Refer to Configure the digital output alarms on page 31. The three
analog outputs continue to represent the last measured value when a measurement alarm is present.
Table 19 Measurement alarm conditions
Code Message Description
SC161 TOCAlarm The measured TOC is more than the user-defined TOC level.
SC162 ConductivityAlarm The measured conductivity is more than the USP645 conductivity level.
Diagnostics
Review the data log usage and firmware version
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the General Tab.
The data log percent usage and the firmware version show.
Review the UV lamp information
See the lamp status and the number of operational hours until the next lamp replacement.
46 English
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the General Tab.
Option Description
Status The color indicator shows the lamp status. Green=normal operation, yellow=marginal operation,
red=lamp fault
Hours left Shows the hours of useful lamp life until the next replacement.
Do a lamp test
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the General tab.
3. Push LAMP TEST.
4. Obey the screen prompts to start the lamp test.
Note: Push the X icon at any time to stop the test.
Change to the other UV lamp
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS.
3. Push SWITCH LAMPS. The lamp status light changes to the new selection.
Select a diagnostic test
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the Tests tab.
3. Select the necessary test. Obey the screen prompts to start the test.
Note: Push X at any time to stop the test.
Validate the RS232 connection
Use the RS232 test icon to test serial communication from the serial port. The printer port is used as a return port. The analyzer sends serial data out one serial port and reads the data on the other. If the read data matches the sent data, the test passes.
1. Push <<
to show the sliding menu.
2. Push DIAGNOSTICS, then select the Tests tab.
3. Push RS-232 TEST.
4. Disconnect the printer and the RS232 serial connections.
5. Connect a serial cable between the serial ports of the analyzer (from one serial port to the other).
6. Push START.
Validate the digital I/O communication
Use the digital I/O test icon to test digital I/O communication. The test toggles the state on each digital output, then reads the state on each of the digital inputs. The results are shown on the screen.
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the Tests tab.
3. Push DIGITAL I/O TEST.
4. Connect the digital inputs of the analyzer to a system that can send a digital signal.
5. Configure one digital input with a high signal and the other with a low signal.
English 47
6. Connect the digital outputs to a system that can read digital signals.
7. Push START.
Validate the 4-20 mA output
Use the 4–20 mA output test icon to test the signal on the 4–20 mA outputs.
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the Tests tab.
3. Push 4-20 mA OUTPUT TEST.
4. Connect a digital multi-meter to one of the 4–20 mA outputs.
5. Push START for the output that is connected to the multi-meter.
A 4 mA signal and then a 20 mA signal are sent from the output for 3 seconds.
Do a printer test
Use the printer test to send a test printout to the printer. Before the test, make sure that the analyzer is correctly connected to the printer.
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the Tests tab.
3. Push PRINTER TEST.
Do a plumbing test
Use the plumbing test to test the various analyzer plumbing systems.
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the Tests tab.
3. Push PLUMBING TEST.
If a current analysis is being done, the analysis is stopped.
4. Select an option.
Option Description
None Set the pump to off. Close all the valves.
Bypass Open only the bypass valve
Flow through cell Open only the cell valve
Flow through cell with pump Open the cell valve and the pump valve. Set the pump to on.
Sample from bottle Open the selected bottle valve and the pump valve. Set the pump to on.
Back flush through bottle Open the selected bottle valve and the cell valve
5. Push LOAD BOTTLES.
6. Select the bottles to load, then push the forward icon.
7. Install the standards in the order shown on the screen.
Note: If the bottle bay door closes too soon, push UNLOCK to open the door.
8. Close the bottle bay door.
9. Push START.
10. Select the applicable test.
11. When the test is completed, push UNLOAD BOTTLES.
12. If a test is selected immediately, select YES to install a bottle as necessary. Enter the volume of
the bottle. Push the forward icon to start the test.
48 English
Validate the pump operation
Use the pump test icon to test and/or calibrate the pumps.
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the Tests tab.
3. Push PUMP TEST.
4. Put a 50-mL graduated cylinder at the water outlet of the analyzer. Do not use tubing.
5. Make sure that a bottle is installed in position 2 to enable the pump to be primed.
6. Select an option.
Option Description
Calibrate Pump Calibrate the pump.
Start/Stop Pump Start or stop the pump. Start Pump—Open the pump valve and the grab sample bottle
valve, then set the pump to on. The pump stays on until Stop Pump is pushed. Stop Pump—Set the pump to off and close both valves.
Prime Pump Open the pump valve and the grab sample valve, then set the pump to on. The pump
operates long enough to prime. Then, the pump switches off and both valves close.
Load Bottle Install bottles in the bottle bay.
Unload Bottle Remove bottles from the bottle bay.
7. Push START.
If Calibrate Pump was selected, the analyzer pumps water for 30 seconds.
8. For pump calibration, enter the volume pumped as measured in the graduated cylinder.
The test results give the:
Test Time— Always 30 seconds
Expected Results—The calculated amount of water that should be measured during the test. The formula is: V = Flow rate ÷ Test time, where Flow Rate = The target flow rate of the pump as entered in the Actual Results text box.
Actual Results—The actual amount of water measured during the test.
9. Push the previous icon to save the calibration results.
Do an RFID test
Use the RFID test to test the RFID reader system.
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the Tests tab.
3. Push RFID TEST.
4. Install four standards bottles equipped with RFID tags in the bottle bay. The bottle can be new or
used standards bottles.
5. Push START.
The system automatically senses those locations where RFID tags are present.
6. To cause a test failure, push the unlock icon during the test to open the bottle bay door. The system will return a failed test because the bottles are no longer positioned properly with the door open.
Do a fan test
Use the fan test to test the analyzer fans. The analyzer contains an internal online heat exchanger that operates to maintain proper water temperature to the analysis cell. Two fans cool the heat exchanger.
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the Tests tab.
English 49
3. Push FAN TEST.
4. Push the fan icon to set the fans to on and off.
Calibrate the display
If the touch screen display does not allow an accurate item selection, calibrate the display.
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the Tests tab.
3. Push SCREEN CALIBRATION.
4. Obey the screen prompts to calibrate.
See the calibration dates
1. Push << to show the sliding menu.
2. Push DIAGNOSTICS, then select the Calibration Dates tab. The screen shows the last
completed calibration dates for TOC, conductivity and the system suitability test.
Technical support
Technical support engineers can provide high quality advice and recommendations for applications, product operation, measurement specifications, hardware and software, factory and customer site training.
Any questionable problems should be referred to the manufacturer. Supply the information that follows in any correspondence:
The serial number of the instrument
• Sample resistivity and temperature
• Current and historical TOC data
• Current and historical alarm codes
• A setup printout
• Process conductivity or resistivity and temperature
• Current and historical calibration, validation and system suitability results
• Any other significant changes in operating conditions
If existing problems warrant equipment return, contact the manufacturer or a sales representative immediately.
Appendix
Theory of operation
The analyzer enables TOC analysis for pure and ultra-pure water processing. The analyzer oxidizes a water sample to find the TOC in the sample. The analyzer traps a water sample in the analysis cell, exposes the sample to ultraviolet (UV) light to oxidize a sample of water (photolysis) and monitors changes in temperature and conductivity until the sample has completely oxidized. Once full oxidation has occurred, the analyzer reports TOC.
Conductivity (in µS/cm) or resistivity (in MΩ-cm)—unadjusted for temperature or corrected to 25°— and sample temperature are measured and shown along with the oxidation curve type and TOC trend. The analyzer includes an integral inline filter, flow control valve and heat exchanger to make sure that the proper water flow rate and temperature are accurate for TOC analysis. An internal pump is available for sampling water from a supply with no pressure and to deliver standards to the analysis cell from the Onboard Automated Standards Introduction System (OASIS™). Refer to
Figure 13.
50 English
Figure 13 PAT700 theory of operation
1 Cell 5 Filter 9 Pump
2 Dual UV lamps 6 Online inlet 10 Vent
3 Heat exchanger 7 Bypass valve
4 Flow needle valve 8 To drain outlet
OASIS™
(onboard automated standards introduction system) simplifies analyzer performance tests with standards bottles tagged with Radio Frequency Identification (RFID) technology, which is a registered radio frequency device. The RFID system transmits data wirelessly with radio waves and operates over a very short distance to prevent interference with other wireless communications.
The RFID system communicates with tags and reader/writers. In the analyzer, the tag is attached to the bottom of the standards bottles used in calibrations (conductivity and TOC), validations, system suitability tests and excursion mode. The tag has a microchip attached to a radio antenna connected to a substrate. The microchip keeps data about the standard contained in the bottle.
The OASIS system has four RFID reader/writers permanently attached inside the analyzer that align with the bottle RFID tags when the bottles are fully loaded in the analyzer. The reader/writer retrieves data from the tags. The tag keeps data about the identity of the standard, the concentration of the standards, expiration date and other pertinent data. The reader/writer receives data from the tag and sends the information digitally to the processor.
The RFID system can read and write to the tags attached to the standards bottle. This enables the analyzer to write data to the bottles that shows the bottle has been used and also supplies data for the exclusive excursion sampling feature.
The system operates without any user intervention and operates automatically when bottle tests are done. The system powers on and off to read and write data only when necessary. The RFID labels on the standards bottles contain the data necessary for each standard, the labels must not be removed from the bottle. Without the label, the reader/writer has no data to read and will not operate as intended, although bottles without labels can be used if the user manually enters data.
A touch-screen interface gives access to all analyzer functions. The user interface has a home screen from which all user operations initiate. The home screen shows the current status of the analyzer and analysis results based on the current run mode selection. The software can show data in a number of user-selected formats, e.g., TOC, compensated and uncompensated conductivity, resistivity and temperature in engineering units with data logging and graphing of the data.
A user may connect a USB flash drive to the analyzer and download the contents of the data buffer, audit trail, alarm log and bottle analysis reports. Query the analyzer through an RS232 communications port to remotely interact with the analyzer and also download data records. Or communicate with the analyzer with Modbus® protocol via the Ethernet port. Print automatically or on demand with an RS232 port to a serial printer.
English 51
The analyzer permits compliance to 21 CFR Part 11 with password protection and an audit trail of all user actions.
Analyzer startup sequence
The analyzer goes through the sequence that follows when the power switch is set to on.
1.
Splash window: The splash window shows at startup while the system launches the application.
2. Verify lamp monitor: After the system has initialized, it verifies that the lamp monitor works. The
main lamp is set to on and the analyzer makes sure that a voltage is returned.
3. Sample bottle check: The system identifies if the bottles loaded in the bottle bay match the last known configuration. If the bottles contain RFID tags, the analyzer reads the tags and compares them against the information saved in the settings. If the RFID tag does not match what is saved in the settings, the settings are updated with the new information and a warning is triggered. If the settings list a non-RFID tagged bottle in one of the bottle positions and an RFID tag cannot be detected in that position, the analyzer assumes the bottle is still present.
4. Start sampling: If the analyzer is configured for online TOC, it starts to take a sample after power is supplied.
5. Print header: After the system has started up, it prints a daily header if a printer is attached.
Run modes
The analysis cycle
An analysis cycle includes a flush of the analysis cell, oxidation and idle time. Idle time occurs only after an online TOC analysis is completed and only if the elapsed time (combined flush and oxidation time) is not more than the cycle time. The oxidation time changes based on the amount of total organic carbon in the sample. The flush time is configurable. The idle time changes based on the cycle time.
During idle time, the analyzer can operate in one of three ways: flow without conductivity, flow with conductivity or no flow. Refer to Configure the general settings on page 27.
To select the run mode, refer to Select the run mode on page 23. Descriptions of the run modes follow.
Online TOC mode
In online TOC mode, the cell valve opens to flush the cell. When flushing is completed, the sample valve closes, the analysis starts, and initial conductivity and temperature readings are sent to the TOC algorithm. Conductivity is monitored until the oxidation is completed.
When oxidation is completed, the analyzer reports the results to the home screen, the data log and the printer, based on the printer settings and if the printer is connected. If the elapsed time is more than or equal to the cycle time in settings, the next online TOC analysis is started. If not, the analyzer goes to the idle mode.
Conductivity mode
In conductivity mode, the home screen shows the conductivity and temperature readings.
In conductivity mode, the analyzer first makes sure that the UV lamp is off and the cell valve is open to flush the cell. The analyzer reports the current readings for conductivity and temperature every time a new value is available. The reported value is an average of the data over the last two seconds. The average reading is also sent to the printer, based on the printout settings and to the data log. One data point is sent to the log every 60 seconds.
CIP Validation mode
In CIP Validation mode, the home screen highlights the TOC reading, with conductivity, temperature, sample time and curve type shown in smaller text. Additionally, the initiate CIP sampling Run button is shown on the right side of the display.
When the CIP Validation button is pushed, the analyzer measures conductivity until the readings goes below a specified threshold or until a specified time interval has expired. Then, the instrument starts one TOC analysis. When the TOC analysis is completed, the analyzer goes to the Idle state.
52 English
Standby mode
In standby mode, the analyzer makes sure that the lamp is off and the valves are open. To exit standby mode, select a different run mode.
Offline mode
In offline mode, the analyzer makes sure that the lamp is off, the valves are closed and flow through the analyzer has stopped. To exit offline mode, select a different run mode.
Manual TOC sample
A TOC analysis can be done manually. Push the TOC manual sample
icon, then select the source stream for analysis if a dual-stream analyzer. The analyzer immediately stops its current operation and then does one TOC analysis. When the analysis is complete, the analyzer goes back to the mode selected on the run mode screen.
Self clean mode
Select the source stream for the self-clean operation if a dual-stream analyzer. If Timed is not selected, the self cleaning continues until it is stopped by the user.
In manual self clean mode, first the elapsed time is set to 0, the cell valve is opened to flush the cell and the lamp is set to on. If Timed is selected, the time that remains shows. If Timed is not selected, the elapsed time shows.
To stop the cleaning, push the red X or select a different run mode.
Bottle mode
Calibration and validation
Bottle mode is used to do calibration, validation and system suitability test procedures. Proper calibration of the analyzer is critical to optimum instrument performance. Such functions can be done automatically with the Onboard Automated Standards Introduction System (OASIS™) with the touchscreen interface. This method minimizes user intervention with the use of a prepackaged standards kit with RFID technology. Calibration, validation and system suitability test procedures can be done immediately or scheduled to be done later.
• Calibration assures the accuracy of instrument readings at regular intervals. This instrument is
calibrated at the factory before shipment. It is not necessary to do a TOC calibration before the analyzer is put into service.
• Validation ensures the accuracy of the calibration. A standard solution of known value is analyzed
to challenge the calibration.
• Calibration and validation involve the analyses of one or more standard solutions and blanks
(protocols) to identify the analyzer response.
• In bottle mode, the analog output stays at the last online TOC, conductivity and temperature
values. The TOC and conductivity alarms are deactivated.
Five functions are available under bottle mode: run standards, grab sample, excursion mode, bottle flush and unload bottles.
Onboard Automated Standards Introduction System (OASIS)
The PAT700 Onboard Automated Standards Introduction System (OASIS) is used to do standards tests, grab sampling and excursion sampling. Through the bottle mode, various bottle test functions can be selected. The OASIS system is automated through the use of RFID technology. Refer to
RFID technology on page 53. When standards bottles with RFID tags are installed in the system,
the analyzer uses RFID technology to read and transmit data about the standards to the analyzer. This system removes the need for manual data entry.
Bottles without RFID tags may also be used. The standards data is entered manually when prompted by the analyzer.
RFID technology
The PAT700 analyzer Onboard Automated Standards Introduction System (OASIS™) uses Radio Frequency Identification (RFID) technology, a registered radio frequency device. The RFID system in
English 53
the PAT700 analyzer operates over a very short distance to prevent any interference with other wireless communications.
The term RFID describes a system that transmits data wirelessly using radio waves. An RFID system is includes a “tag” and a “reader/writer”. The tag is attached to the standards bottles used in calibrations (conductivity and TOC), validations, system suitability tests and excursion mode. The RFID tag includes a microchip attached to a radio antenna mounted on a substrate. The RFID tag is attached to the bottom of the standards bottle. The microchip contains data about the standard contained in the bottle. The PAT700 OASIS system contains four RFID reader/writers that are permanently mounted inside the analyzer. The RFID reader/writers align with the bottle RFID tags when the bottles are fully loaded in the analyzer.
The reader/writer retrieves the data stored on the RFID tags located on the bottom of the standards bottles. The RFID reader/writers have antennas that emit radio waves and receive signals back from the tags on the standards bottles. The information provided from the tag includes the identity of the standard, the concentration of the standard solution, date of expiration and other pertinent data.
The reader/writer takes the data received from the tag and passes the information in digital form to the PAT700 processor. The RFID system in the PAT700 can read and write to the RFID tags attached to the standards bottle. The writing feature lets the PAT700 write data to the bottles identifying that the bottle has been used and provides for writing of data for the exclusive excursion sampling feature.
The RFID system in the PAT700 does not require any user intervention to operate. The system operates automatically when bottle tests are done. The system automatically switches on and off to read and write data only when necessary. When enabled, the radio frequency modulation emitted from the antennas is fixed at 13.56 Mhz. There are no user-serviceable parts associated with the RFID system. The RFID reader/writer board assembly should only be serviced by a Hach certified service representative.
Since the RFID labels on the standards bottles contain the data necessary for each standard, the labels must not be removed from the bottle. Without the label, the RFID reader/writer has no data to read and will not operate as intended. However, bottles without labels can be run if the user manually enters data.
Data log
The data log holds up to 5,000 records. The data log contains analysis results, alarms, bottle test results and audit trail entries. Since the audit trail is kept in the data log, each audit trail entry uses one of the 5,000 data log entries.
The data log operates in a first in first out approach. Once the data log contains 5,000 entries the oldest entry is replaced with the newest entry.
The contents of the data log shows in the Data Review screen. Refer to Manage measurement data on page
26. The Data Review screen is populated only when data logging is selected in the security settings. The Data Review screen shows a list of all the data in the data log as will show on a printout.
The data shown on the Data Review screen is filtered based on the settings in the filter dialog box. Refer to Set the data filter on page 26. Data is shown in chronological order from earliest to most recent. The list scrolls to display the earliest data.
Default settings
Table 20 shows the parameters and default settings of the analyzer.
Table 20 Parameters and default settings
Parameter Default setting
TOC Setup
Flush time 1 minute
Cycle time 0
Use pump Disabled
54 English
Table 20 Parameters and default settings (continued)
Parameter Default setting
Bypass during analysis Disabled
Average last results Disabled
Flow during idle With conductivity
Control with digital inputs Disabled
Stabilization – time between analysis 2 hours
Stabilization – time 10 minutes
System settings
Instrument name PAT700SNXXXX (XXXX = serial number)
Language English
Time format 24-hour
Temperature display °C
Conductivity display Uncompensated conductivity
Static TCP/IP address 192.168.0.105
Static gateway 255.255.255.0
DNS server Blank
WINS server Blank
DHCP Enabled
Alarm setup
TOC upper limit 500 ppb
TOC alarm Disabled
Conductivity alarm Disabled
Analog outputs setup
TOC zero 0.0
TOC full scale 2000
Conductivity zero 0.0
Conductivity full scale 20.0
Temperature zero 0.0
Temperature full scale 100
On error, return 2.0
Printer setup
Printer model DPU-414
TOC print Continuous
Conductivity print Disabled
Conductivity print interval 1 minute
Conductivity % change 1 %
English 55
Table 20 Parameters and default settings (continued)
Parameter Default setting
Security settings
Enable security Disabled
Enable data logging Enabled
Password expires in # days 60
Remember last # passwords 3
Automatically log off after # minutes 30
Administrator ID Admin
Administrator password 123456
User's full name Blank
User ID Blank
User password Blank
Change password Enabled
Bottles change settings
TOC calibration:
Manually accept results Enabled
Return to online mode Disabled
Slope limit 15%
R2 limit 0.99
Conductivity calibration:
Manually accepts results Enabled
Return to online mode Disabled
Slope limit from previous 5%
Slope limit from factory calibration 10%
Use calibration resistor Enabled
TOC validation:
Return to online mode Disabled
System suitability:
Return to online mode Disabled
Bottle mode
TOC calibration setup Use default settings
TOC calibration custom settings All standards selected, 3 repetitions
Schedule bottle test Current time
56 English
Replacement parts and accessories
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the instrument or equipment malfunction. The replacement parts in this section are approved by the manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information.
Consumables
Description Item no.
PAT700 Calibration Standards Kit, includes: calibration blank and 0.250, 0.500 and 0.750 mg C/L sucrose calibration standards in RFID bottles
FG7019202
PAT700 Validation Standards Kit, includes: calibration blank and 0.500 mg C/L sucrose calibration standard in RFID bottles
FG7019222
PAT700 System Suitability Standards Kit, includes:
0.500 mg C/L sucrose, reagent water and USP 1,4-benzoquinone solution in RFID bottles
FG7018402
PAT700 Conductivity Standards Kit, includes: NIST calibration traceable electrolyte solution in RFID bottle
FG7002602
PAT700 Validation Protocol Reagent Kit, liquid reagent for the installation and operation qualification protocol, includes: calibration standards kits (3x), validation standards kits (2x), system suitability standards kits (4x), conductivity standards kit and NIST traceable calibration resistor
FG7019232
Replacement parts
Description Item no.
UV lamp, PAT700 FG7001014
Water filter, internal, PAT700 FG7001015
Accessories
Description Item no.
Conductivity meter calibration resistor, PAT700 FG7001009
PAT700 Septa Cap Kit, includes: 100 pre-cleaned caps for PAT700 grab sample bottles (60 mL)
FG7025011
PAT700 Excursion Bottle Assembly FG7025000
PAT700, Excursion Sample Bottle Kit, includes: ten bottles with RFID tags (60 mL)
FG7025100
Portable carrying handle for PAT700 FG7001005
Printer, 120 VAC with cable for PAT700 FG7001016
Printer, 230 VAC with cable for PAT700 FG7001018
PAT700 AutoSampler with rinse station (pump), model ASX-7200 FG7006501
PAT700 TOC Blank Kit, includes: four 60 mL bottles
FG7018501
English 57
Accessories (continued)
Description Item no.
HDPE Bottle Kit, includes: twenty-five clean, 60 mL grab sample bottles with septa caps
FG0025031
IQ/QD protocol on CD, PAT700 FG7001010
58 English
HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Tel. (970) 669-3050 (800) 227-4224 (U.S.A. only) Fax (970) 669-2932 orders@hach.com www.hach.com
HACH LANGE GMBH
Willstätterstraße 11 D-40549 Düsseldorf, Germany Tel. +49 (0) 2 11 52 88-320 Fax +49 (0) 2 11 52 88-210 info@hach-lange.de www.hach-lange.de
HACH LANGE Sàrl
6, route de Compois 1222 Vésenaz SWITZERLAND Tel. +41 22 594 6400 Fax +41 22 594 6499
©
Hach Company/Hach Lange GmbH, 2014.
All rights reserved. Printed in U.S.A.
Loading...